Manual C28 Plus en

This document provides instructions for using a compressor nebulizer device. It outlines important safety information, instructions for use, cleaning and maintenance procedures. Key details include intended users are medical experts or patients under guidance; cleaning and disinfecting the nebulizer parts before and after each use is important; do not use water only and always dispose of leftover medication; check that all parts are correctly attached before use.

Uploaded by

Sveto Sl
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
456 views

Manual C28 Plus en

This document provides instructions for using a compressor nebulizer device. It outlines important safety information, instructions for use, cleaning and maintenance procedures. Key details include intended users are medical experts or patients under guidance; cleaning and disinfecting the nebulizer parts before and after each use is important; do not use water only and always dispose of leftover medication; check that all parts are correctly attached before use.

Uploaded by

Sveto Sl
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 28

NE-C28P-E_A_M04_110921.

pdf
NE-C28P-E_main.book Page 1 Friday, October 7, 2011 2:48 PM

TM

Compressor Nebulizer
Model NE-C28P

• Instruction Manual EN
• Mode d’emploi FR
• Gebrauchsanweisung DE
• Manuale di istruzioni IT
• Manual de instrucciones ES
• Gebruiksaanwijzing NL
• РУКОВОДСТВО ПО ЭКСПЛУАТАЦИИ RU
AR

IM-NE-C28P-E-01-09/2011
5334069-0A
NE-C28P-E_main.book Page 2 Friday, October 7, 2011 2:48 PM

Before using the unit

Contents
Before using the unit
Introduction .................................................................................................. 3
Intended use ................................................................................................. 3
Important safety instructions ...................................................................... 4
1. Know your unit ....................................................................................... 7
2. Guide to using the COMP AIR .............................................................. 9
Operating instructions
3. Preparing the nebulizer for use .......................................................... 10
4. Taking medication ............................................................................... 14
Care and maintenance
5. Cleaning and daily disinfecting .......................................................... 17
5.1 Cleaning ............................................................................................ 17
5.2 Disinfecting the nebulizing parts ....................................................... 18
6. Maintenance and storage .................................................................... 21
7. Changing the air filter .......................................................................... 22
8. Troubleshooting ................................................................................... 23
9. Technical data ...................................................................................... 24
10. Accessories .......................................................................................... 27

2
NE-C28P-E_main.book Page 3 Friday, October 7, 2011 2:48 PM

Introduction
Thank you for purchasing the OMRON COMP AIR.
This product was developed in conjunction with respiratory therapists for the
successful treatment of asthma, chronic bronchitis, allergies and other respiratory
disorders. The compressor forces air to the nebulizer kit. When the air enters the
nebulizer kit, it converts the prescribed medication into an aerosol of microscopic
droplets that can easily be inhaled.
This is a medical device. Operate the device only as instructed by your doctor and/
or respiratory therapist.

Intended use

Medical Purpose This product is intended to be used for inhaling medication


for respiratory disorders.
Intended User • Legally certified medical experts, such as doctor, nurse and
therapist, or healthcare personnel or patient under the
guidance of qualified medical experts.
• The user should also be capable of understanding general
operation of NE-C28P and the content of instruction
manual.
Intended Patients This product should not be used by patients, who are
unconscious or are not breathing spontaneously.
Environment This product is intended for use in a medical facility, such as
Hospital, clinic and doctor’s office, and in a room of general
household.
Durable period Durable periods are as follows, provided the product is used
to nebulize saline 3 times a day for 10 minutes each time at
room temperature (23°C). EN
Durable period may vary depending on usage environment.
Compressor 5 years
Nebulizer Kit 1 year
Mouthpiece 1 year
Nosepiece 1 year
Air Tube (PVC, 200cm) 1 year
Air Filter 60 days
Adult Mask (PVC) 1 year
Child Mask (PVC) 1 year
Precautions for use Warnings and cautions described in the instruction manual
should be observed.

3
NE-C28P-E_main.book Page 4 Friday, October 7, 2011 2:48 PM

Important safety instructions


Read all the information in the instruction manual and any other literature included
in the box before using the device.
When using an electrical product, especially when children are present, basic
safety precautions should always be followed including the following:

Warning:
Indicates a potentially hazardous situation which, if not avoided, could result in
serious injury.
(Usage)
• For type, dose, and regime of medication follow the instructions of your doctor or
respiratory therapist.
• If you feel anything unusual during use, stop using the device immediately and
consult your doctor.
• Do not use only water in the nebulizer, for inhaling purposes.
• Clean and disinfect the nebulizer kit, mouthpiece and nosepiece or masks before
using them for the first time after purchase, if the device has not been used for a
long period of time, or if more than one person uses the same device.
• Be sure to wash the parts after use, and ensure that they are thoroughly
disinfected and dried, and stored in a clean location.
• Keep the device out of the reach of unsupervised infants and children. The
device may contain small pieces that can be swallowed.
• Store the device and accessories in a clean location.
• Do not store the air tube while there is moisture or medication remaining inside it.
• Always rinse the nebulizing parts with clean hot tap water after disinfecting them.
Never leave the cleaning solution in the medication tank or mouthpiece.
• Do not use or store the device where it may be exposed to noxious fumes or
volatile substances.
• Do not use the device where it may be exposed to flammable gas.
• Do not block the ventilation slots. Never place the device where the ventilation
slots may be obstructed during operation. Do not cover the compressor with a
blanket or towel etc., during use.
• Make sure that the nebulizer kit is clean before use.
• Always dispose of any remaining medication after use, and use fresh medication
each time.
• Do not use in anaesthetic or ventilator breathing circuits.
(Risk of electrical shock)
• Do not use the compressor or power cord while they are wet, or connect them to
a power outlet or other devices with wet hands.
• The compressor and power cord are not waterproof. Do not spill water, or other
liquids, on these parts. If liquid does spill on these parts, immediately unplug the
power cord and wipe off the liquid with gauze or other absorbent material.

4
NE-C28P-E_main.book Page 5 Friday, October 7, 2011 2:48 PM

Important safety instructions


• Do not immerse the compressor in water or other liquid.
• Do not use or store the device in humid locations, such as a bathroom.
• Do not operate the device with a damaged power cord or plug.
• Keep the power cord away from heated surfaces.
Read and follow the “Important information regarding Electro Magnetic
Compatibility (EMC)” (page 26).

Caution:
Indicates a potentially hazardous situation which if not avoided, may result in minor
or moderate injury, or physical damage.
(Usage)
• Provide close supervision when this device is used by, on, or near children or
invalids.
• Make sure that the parts are attached correctly.
• Make sure that the vaporiser head is correctly installed before use.
• Make sure that the air filter is correctly attached.
• Make sure that the air filter is clean. If the air filter has changed colour, or has
been used on average for more than 60 days, replace it with a new one.
• Do not spill liquid or medication on the compressor.
• Do not tilt the nebulizer kit at an angle of greater than 45° in all directions, or
shake it while in use.
• Do not use or store the device while the air tube is creased.
• Use only original nebulizing parts, air tube, air filter and filter cover.
• Do not add more than 7 ml of medication to the medication tank.
• Do not carry or leave the nebulizer kit while the medication tank contains
medication.
• Do not leave the device unattended with infants or persons who cannot express
their consent.
• Do not subject the device or the parts to any strong shocks such as dropping the
device on the floor. EN
• Do not distort the vaporiser head, or jab the nozzle of the medication tank with a
pin etc.
• Do not insert fingers or objects inside the compressor.
• Do not disassemble or attempt to repair the compressor or power cord.
• Do not leave the device or its parts in extreme hot or cold temperature, or under
direct sunlight.
• Do not block the air filter cover.
• When disinfecting parts by boiling, make sure that the container does not boil dry.
• Never place, or attempt to dry the compressor, the power cord or nebulizer kit in a
microwave oven.
• Do not use the device while sleeping or if drowsy.

5
NE-C28P-E_main.book Page 6 Friday, October 7, 2011 2:48 PM

Important safety instructions


• To avoid injury to the nose mucosa, do not squeeze the nosepiece into the back
of the nose.
• Do not block the slit between the cap and the inhalation air inlet.
• Do not use a microwave oven or hair dryer to dry the nebulizing parts.
• Approved for human use only.
• When using the device, the main unit may become hot.
• Do not touch the main unit for other than necessary operation such as turning off
the power while nebulizing.
• To avoid the medication residue on the face, be sure to wipe the face after
removing the Mask.
• Do not use an autoclave, EOG gas disinfection or low temperature plasma
sterilizer to disinfect the device.
(Risk of electrical shock)
• Always unplug the device from the power outlet after use and before cleaning.
• Plug the device into the appropriate voltage outlet. Do not overload power outlets
or use extension cords.
• Do not wind the power cord around the compressor.
• Changes or modifications not approved by OMRON HEALTHCARE will void the
user warranty.
• Do not pull the power cord strongly.
General Safety Precautions:
• Inspect the device and parts before using them each time, and check that there
are no problems. In particular, be sure to check the following:
- That the vaporiser head, nozzle or air tube are not damaged.
- That the nozzle is not blocked.
- That the compressor operates normally.
• When using this device, there will be some noise and vibration caused by the
pump in the compressor. There will also be some noise caused by the emission
of compressed air from the nebulizer kit. This is normal and does not indicate a
malfunction.
• Operate the device only as intended. Do not use the device for any other
purpose.
• Do not use the device at temperatures greater than +40°C.
• Make sure that the air tube is securely attached to the compressor (main unit)
and nebulizer kit, and does not come loose. Twist the air tube plug slightly when
inserting it into the connectors to avoid the air tube disconnecting during use.
• To completely isolate the device from the power source unplug the plug from the
power source.
Read and follow the “Correct Disposal of This Product” (page 26) when disposing
of the device and any used accessories or optional parts.

Save these instructions for future reference.

6
NE-C28P-E_main.book Page 7 Friday, October 7, 2011 2:48 PM

1. Know your unit


A C D

A1

D1 D2 D3 D4

A5 E
A2 E1
A3
B1 A4
E2
E3
B2
B

Rear View E4

A5 E6
E5

A6
A7 E7

A. Compressor (Main Unit) D. Inhalation Attachments


A1.Power Switch D1.Mouthpiece
A2.Air Tube Connector D2.Adult Mask (PVC)
A3.Air Filter Cover (* Air filter inside.) D3.Child Mask (PVC)
EN
A4.Nebulizer Kit Holder D4.Nosepiece
A5.Ventilation Slots E. Nebulizer Kit
A6.Power Cord E1.Cap
A7.Power Plug E2.Inhalation Air Inlet
B. Air Tube (PVC, 200cm) E3.Inhalation Top
B1.Air Tube Plug E4.Vaporiser Head
B2.Air Tube Plug E5.Medication Tank
C. Nebulizing Parts E6.Nozzle
E7.Air Tube Connector

7
NE-C28P-E_main.book Page 8 Friday, October 7, 2011 2:48 PM

1. Know your unit

Accessories

TM

Compressor Nebulizer
Model NE-C28P

• Instruction Manual EN
• Mode d’emploi FR
• Gebrauchsanweisung DE
• Manuale di istruzioni IT
• Manual de instrucciones ES
• Gebruiksaanwijzing NL
• RU
AR

IM-NE-C28P-E-01-09/2011
5334069-0A

F G H I

F. Spare Air Filters x 5 H. Instruction Manual


G. Carrying Bag I. Warranty Card

8
NE-C28P-E_main.book Page 9 Friday, October 7, 2011 2:48 PM

2. Guide to using the COMP AIR

Preparing the Nebulizer for Use - Section 3 (page 10-13)


Warning:
Clean and disinfect the nebulizer kit, mouthpiece and nosepiece or
masks before using them for the first time after purchase, if the
device has not been used for a long period of time, or if more than
one person uses the same device.

Taking Medication - Section 4 (page 14-16)


Caution:
• Do not tilt the nebulizer kit at an angle of greater
than 45° in all directions.
• Do not crease or fold the air tube.

Cleaning Parts After Use - Section 5


(page 17-20)
Note: Failure to clean the nebulizing parts
after use may lead to infection.

How the nebulizer kit works


Baffle EN
The medication that is pumped up
through the medication channel is Aerosol Aerosol
mixed with compressed air which is
generated by a compressor pump. Nozzle Medication
The compressed air mixed with Channel
medication is tuned into fine Medication Medication
particles and sprayed by being
impacted to the baffle.

Compressed Air

9
NE-C28P-E_main.book Page 10 Friday, October 7, 2011 2:48 PM

Operating instructions
3. Preparing the nebulizer for use

Warning:
Clean and disinfect the nebulizer kit, mouthpiece and nosepiece or masks
before using them for the first time after purchase, if the device has not been
used for a long period of time, or if more than one person uses the same
device.
For directions on cleaning and disinfecting, refer to Section 5 (page 17-20).

Note: Make sure that the air filter is changed regularly. Refer to Section 7
“Changing the air filter” (page 22).

1. Make sure that the power switch is turned off.


Check that the power switch is in the position.

2. Plug the power plug into a power outlet.


Power Plug

3. Remove the mouthpiece and cap from the


nebulizer kit.

4. Remove the inhalation top from the medication 2


tank.
1) Rotate the inhalation top anti-clockwise.
1
2) Lift the inhalation top out of the medication
tank.

10
NE-C28P-E_main.book Page 11 Friday, October 7, 2011 2:48 PM

3. Preparing the nebulizer for use

5. Add the correct amount of prescribed


medication to the medication tank.

Caution:
The capacity of the medication tank is 2 to 7 ml.

Note: Scales on the nebulizer kit are for


reference only. The scale on the outside
of the tank is with no vaporiser head in
place. Please use the scale on your syringe or vial for accurate
measurement of medication.

6. Make sure that the vaporiser head is securely


attached to the inhalation top.

7. Put the inhalation top back inside the


medication tank.
1) Align the protrusion on the inhalation top
with the indent on the medication tank, as
shown.
2) Rotate the inhalation top clockwise and align 2
with the medication tank until they click in
place.

EN

11
NE-C28P-E_main.book Page 12 Friday, October 7, 2011 2:48 PM

3. Preparing the nebulizer for use

8. Attach the desired inhalation attachment, as described below.

Attaching the mouthpiece 2


1) Attach the cap to the inhalation air inlet.
2) Attach the mouthpiece to the inhalation top.
1

Attaching the adult mask (PVC)


Attach the mask to the inhalation top.

Note: If the amount of aerosol is too great,


attach the cap.

Attaching the child mask (PVC)


1) Attach the cap to the inhalation air inlet.
2) Attach the child mask to the inhalation top. 2
Note: If the amount of aerosol is too small, it is
not necessary to place the cap. 1

Attaching the nosepiece


1) Attach the cap to the inhalation air inlet. 2
2) Attach the nosepiece to the inhalation top.
1

12
NE-C28P-E_main.book Page 13 Friday, October 7, 2011 2:48 PM

3. Preparing the nebulizer for use

9. Attach the air tube.


Twist the air tube plugs slightly and push it firmly into the air tube
connectors.

Notes:
• Make sure that the air tube is securely attached to the compressor (main
unit) and nebulizer kit, and does not come loose. Twist the air tube plug
slightly when inserting it into the air tube connectors to avoid the tube
disconnecting during use.
• Take care not to spill the medication when attaching the air tube. Keep
the nebulizer kit upright.
• Use the nebulizer kit holder as a
temporary holder for the nebulizer kit.

EN

13
NE-C28P-E_main.book Page 14 Friday, October 7, 2011 2:48 PM

4. Taking medication

1. Hold the nebulizer kit as indicated on the right.

Caution:
Do not tilt the nebulizer kit at
an angle of greater than 45° Right angle
in all directions. Medication 45° 45°
may flow into the mouth.

2. Press the power switch to the


on ( ) position.
The compressor starts and
nebulization starts.

Note: Observe whether aerosol is


generated.

To stop nebulization
Press the power switch to the off ( ) position.

Warning:
Do not block the ventilation slots on the compressor. This could
result in the compressor overheating or malfunctioning.

Caution:
• Do not block the air filter cover.
• When using the device, the main unit may become hot.
• Do not touch the main unit for other than necessary operation such as
turning off the power while nebulizing.
Note: Do not leave the compressor on for a long period of time. This could
result in the compressor overheating or malfunctioning.

14
NE-C28P-E_main.book Page 15 Friday, October 7, 2011 2:48 PM

4. Taking medication

3. Inhale medication according to your doctor’s and/or health care adviser’s


instructions.

Using the mouthpiece


Insert the mouthpiece into your mouth and
inhale the medication breathing normally.
Exhale as normal through the mouthpiece.

Using the nosepiece


Inhale the medication through the nosepiece, as
shown and exhale through the mouth.

max
5
3

Using the adult mask (PVC)


Attach the mask so that it covers your nose and
mouth and inhale the medication.
Exhale through the mask.

Note: If the amount of aerosol is too great,


attach the cap.

Using the child mask (PVC)


Attach the mask so that it covers your nose and EN
mouth and inhale the medication.
Exhale through the mask.

15
NE-C28P-E_main.book Page 16 Friday, October 7, 2011 2:48 PM

4. Taking medication

4. When treatment is complete, turn the power off. Check that there is no
condensation or moisture in the air tube and unplug the device from the
power outlet.

Power Plug

Warning:
Condensation may occur in the air tube. Do not store the air tube
while it contains condensation or moisture. This could result in
infection via bacteria.
If there is moisture or liquid remaining in the air tube, be sure to follow the
procedure below to remove any moisture from within the air tube.
1) Make sure that the air tube is still connected to the air connector on the
compressor.
2) Unplug the air tube from the nebulizer kit.
3) Turn on the compressor and pump air through the air tube to expel the
moisture.

16
NE-C28P-E_main.book Page 17 Friday, October 7, 2011 2:48 PM

Care and maintenance

5. Cleaning and daily disinfecting

5.1 Cleaning
Following the cleaning instructions after each use will prevent any remaining
medication in the tank from drying, resulting in the device not nebulizing effectively
and also for preventing infections.

Warning:
Thoroughly rinse the nebulizer kit and the mask, nosepiece or mouthpiece
with clean hot tap water after each use, and disinfect the parts of the
nebulizer kit and applicable inhalation attachments after the last treatment of
the day.
Note: Do not start operation before the cleaned parts are completely dried.

1. Make sure that the power switch is in the (off) position.

2. Unplug the power cord from the electrical outlet.

3. Remove the nebulizer kit from the air tube and disassemble the nebulizer
kit. Refer to Section 3 (page 10-13).

4. Discard any remaining medication in the


medication tank.

EN
5. Wash all the nebulizing parts in warm water and mild detergent and rinse in
clean hot tap water.
For instructions on how to disinfect the parts, see the following section.

6. After cleaning and disinfecting the parts, rinse them thoroughly with clean
water and allow the parts to air dry in a clean environment.

17
NE-C28P-E_main.book Page 18 Friday, October 7, 2011 2:48 PM

5. Cleaning and daily disinfecting

7. If the compressor is dirty, wipe it clean with a

Mild Detergent
soft cloth moistened with water or mild
detergent.

Caution:
The compressor is not waterproof.

8. If the outside of the air tube is dirty, wipe it clean with a soft cloth moistened
with water or mild detergent.

9. Reassemble the nebulizer kit. Put the parts in a dry, sealed bag and place
in the storage pocket of the carrying bag.

Note: Nebulizer kit should be replaced after using for one year.

5.2 Disinfecting the nebulizing parts


Always clean before disinfection.
Always disinfect the medication tank, mouthpiece and other nebulizing parts after
the last treatment of the day. For appropriate method of disinfecting each part,
refer to page 20.
A. Use a commercially available disinfectant. Follow the instructions provided by
the disinfectant manufacturer.

1. Submerge the parts in the cleansing solution for the specified period.

2. Remove the parts and discard the solution.

3. Rinse the parts with clean hot tap water, shake off excess water and allow
to air dry in a clean environment.
B. Parts may be boiled between 15 to 20 minutes.
After boiling, carefully remove the parts, shake off excess water and allow to air dry
in a clean environment.

Note: Do not boil Air tube, Adult mask (PVC), Child mask (PVC), Air filter and Air
filter cover.

Caution:
Do not use an autoclave, EOG gas disinfection or low temperature plasma
sterilizer to disinfect the device.

18
NE-C28P-E_main.book Page 19 Friday, October 7, 2011 2:48 PM

5. Cleaning and daily disinfecting

Handling the Vaporiser Head


The vaporiser head is an important part that is used to vaporise medication.
Be sure to observe the following precautions when handling it.

Caution:
• Always wash the vaporiser head after each use. Baffle
• Do not use a brush or pin etc., to clean the parts.
• When disinfecting the parts by boiling, be sure to
boil them in plenty of water.
• Do not boil vaporiser head together with other
objects, other than applicable nebulizer
accessories.
• Do not use a microwave oven, dish dryer or hair
dryer to dry parts.

Vaporiser head

EN

19
Use the table below as a guideline to select a method for disinfecting. O: applicable ×: not applicable
Sodium Quaternary Amphoteric
Alcohol hypochlorite ammonium Chlorhexidine surfactant
Parts Model Materials Boiling
Disinfecting Milton* Osvan* Hibitane* Tego*
ethanol (0.1%) (0.1%) (0.5%) (0.2%)
Mouthpiece C28-3P-E PP O O O O O O
Cap: PP
C801S- Inhalation Top: PP
Nebulizer Kit Set O O O O O O
NSET3-E Vaporiser Head: PC
Medication Tank: PP
Nosepiece C28-7-E PP O O O O O O
Air Filter C28-2-E Polyester × × × × × ×
Air Filter Cover C28-5-E ABS × O O O O O
NE-C28P-E_main.book Page 20 Friday, October 7, 2011 2:48 PM

Mask: PVC
Adult Mask (PVC) C28-8-E (Phthalate free)
Child Mask (PVC) C28-9-E
× O O O O O
Band: Rubber
(non-latex)
5. Cleaning and daily disinfecting

Adult Mask (SEBS) Mask: SEBS


(optional) U10-1P-E Band: Rubber
U10-2P-E
O O O O O O
Child Mask (SEBS) Set (non-latex)
(optional) Adapter: PP
NEB-TP- PVC
Air Tube (PVC, 200cm) L20E × × × × × ×
(Phthalate free)
Air Tube (Silicone, 100cm) Tube: Silicone
C28-6-E O O O O O O
(optional) Connector: PP
* an example of commercially available disinfectant.

20
NE-C28P-E_main.book Page 21 Friday, October 7, 2011 2:48 PM

6. Maintenance and storage


To keep your device in the best condition and protect the device from damage
follow these directions:

Caution:
• Do not leave the unit unattended with infants or persons who cannot express
their consent.
• Do not subject the device or the parts to any strong shocks such as dropping the
device on the floor.
• Do not carry or leave the nebulizer while the medication tank contains
medication.

Notes:
• Do not store the device in extreme hot or cold temperatures, high humidity or
under direct sunlight.
• Do not crease or fold the air tube.
• Dispose of the device and any used accessories or optional parts according to
applicable local regulations.
• Never clean with benzene, thinner or a flammable chemical.

Benzene

Thinner
Always transport the device in its carrying bag, and store the
device in the bag if it will not be used for a long period.

EN

21
NE-C28P-E_main.book Page 22 Friday, October 7, 2011 2:48 PM

7. Changing the air filter


If the air filter has changed colour, or has been used on average for more than 60
days, replace it with a new one.

1. Pull the air filter cover off the compressor.

2. Change the air filter.


Remove the old air filter with a sharp object,
such as a toothpick, and insert the new air
filter.

3. Put the air filter cover back in place.


Notes:
• Use only OMRON air filters designed for this
device. Do not operate without a filter.
• There is no front or back orientation for the
air filters.
• Check that air filters are clean and free of dust before inserting them.
• Do not attempt to wash or clean the filter. If the air filter becomes wet,
replace it. Damp air filters can cause blockages.
• To prevent the air filter cover from becoming blocked, wash the cover
regularly. Do not boil the air filter cover. Remove the air filter cover, wash
it, and make sure it is dry and reattached correctly.
• Please contact your OMRON dealer for additional air filters, or if the parts
supplied with the device are damaged or missing.

22
NE-C28P-E_main.book Page 23 Friday, October 7, 2011 2:48 PM

8. Troubleshooting
Check the following if your unit should fail during operation. You can also refer to
the pages of this manual for complete instructions.
Symptom Cause Remedy
Check that the plug is
Nothing happens when the Is the power plug plugged inserted in an electrical
power switch is pressed. into an electrical outlet? outlet. Unplug then reinsert
the plug if necessary.
Is there medication in the
medication tank? Add the correct amount of
Is there too much/little medication to the medication
medication in the medication tank.
tank?
Is the vaporiser head
Attach the vaporiser head
missing and not assembled
correctly.
correctly?
Is the nebulizer kit Assemble the nebulizer kit
assembled correctly? correctly.
Make sure that the nozzle is
Is the nozzle blocked?
No nebulization or low free of blockages.
nebulization rate, when the Make sure that the nebulizer
power is on. Is the nebulizer kit tilted at a
kit is not tilted at an angle of
sharp angle?
more than 45 degrees.
Make sure that the air tube is
Is the air tube connected correctly connected to the
correctly? compressor and nebulizer
kit.
Is the air tube folded or Make sure that the air tube
damaged? does not contain kinks.
Make sure that the air tube is EN
Is the air tube blocked?
free of blockages.
Replace the air filter with a
Is the air filter dirty?
new one.
The device is abnormally Is the air filter cover attached Attach the air filter cover
loud. correctly? correctly.
Are the ventilation slots Do not block the ventilation
The device is very hot.
blocked? slots.

Note: If the suggested remedy does not solve the problem, do not try to repair the
device - no parts of the unit are user serviceable.
Return the unit to an authorized OMRON retail outlet or distributor.

23
NE-C28P-E_main.book Page 24 Friday, October 7, 2011 2:48 PM

9. Technical data
Product Description: Compressor Nebulizer
Model: NE-C28P (NE-C28P-E)
Electrical: 220 - 240V, 50/60 Hz
Power Consumption: 192VA
Operating Temperature/ +10°C to +40°C 30% to 85% RH
Humidity:
Storage Temperature/ -20°C to +60°C 10% to 95% RH 700 - 1060 hPa
Humidity/Air Pressure:
Weight: Approx. 1.9kg (compressor only)
Dimensions: Approx. 170 (W) × 103 (H) × 182 (D) mm
Contents: Compressor, Nebulizer Kit, Air Tube (PVC, 200cm),
5 pcs Spare Air Filters, Mouthpiece, Adult Mask (PVC), Child
Mask (PVC), Nosepiece, Carrying Bag, Instruction Manual,
Warranty Card.
Classification: Class ll equipment, Type B applied part

= Type B Read the instruction


= Class ll equipment
applied part manual carefully

= Power off = Power on

Notes:
• Subject to technical modification without prior notice.
• This OMRON product is produced under the strict quality system of OMRON
HEALTHCARE Co., Ltd., Japan.
• The device may not work if the temperature and voltage conditions are different
to those defined in the specifications.
• Do not use the device where it may be exposed to flammable gas.
• This device fulfils the provisions of the EC directive 93/42/EEC (Medical Device
Directive) and the European Standard EN13544-1:2007+A1:2009, Respiratory
therapy equipment - Part1: Nebulizing systems and their components.

24
NE-C28P-E_main.book Page 25 Friday, October 7, 2011 2:48 PM

9. Technical data
Technical data for the OMRON NE-C28P Compressor with the OMRON V.V.T.
(Virtual Valve Technology) Nebulizer kit N-01:

Particle Size: *MMAD **approximately 3µm


Medication Tank 7 ml maximum
Capacity:
Appropriate Medication 2 ml minimum - 7 ml maximum
Quantities:
Sound: Noise level (at 1 m distance) 60 dB
Nebulization Rate: 0.5 ml/min (by weight loss)
Aerosol Output: **0.4 ml (2ml, 1%NaF)
Aerosol Output Rate: **0.09 ml/min (2ml, 1%NaF)

Result of cascade impactor measurements for particle size**


Cumulative % particle mass of sodium fluoride under size
100
90
80

Cumulative Undersize %
70
60
50
40
Individual tests 30
Mean
20
MMAD: 3µm 10
0
0.1 1 10 100
Particle Size Dp (µm)

*MMAD = Mass Median Aerodynamic Diameter


**Independently measured by Prof. Dr. Hiroshi Takano, Department of Chemical
Engineering and Materials Science, Faculty of Science and Engineering,
Doshisha University, Kyoto, Japan, according to EN13544-1:2007+A1:2009
Notes: EN
• Performance may vary with drugs such as suspensions or high viscosity. See
drug supplier’s data sheet for further details.
• See website of OMRON HEALTHCARE EUROPE to update technical
information.
URL: www.omron-healthcare.com

25
NE-C28P-E_main.book Page 26 Friday, October 7, 2011 2:48 PM

9. Technical data

Important information regarding Electro Magnetic Compatibility (EMC)


With the increased number of electronic devices such as PC’s and mobile
(cellular) telephones, medical devices in use may be susceptible to
electromagnetic interference from other devices. Electromagnetic interference
may result in incorrect operation of the medical device and create a potentially
unsafe situation.
Medical devices should also not interfere with other devices.
In order to regulate the requirements for EMC (Electro Magnetic Compatibility)
with the aim to prevent unsafe product situations, the EN60601-1-2:2007
standard has been implemented. This standard defines the levels of immunity
to electromagnetic interferences as well as maximum levels of electromagnetic
emissions for medical devices.
This medical device manufactured by OMRON HEALTHCARE conforms to
this EN60601-1-2:2007 standard for both immunity and emissions.
Nevertheless, special precautions need to be observed:
• Do not use mobile (cellular) telephones and other devices, which generate
strong electrical or electromagnetic fields, near the medical device. This may
result in incorrect operation of the unit and create a potentially unsafe situa-
tion. Recommendation is to keep a minimum distance of 7 m. Verify correct
operation of the device in case the distance is shorter.
Further documentation in accordance with EN60601-1-2:2007 is available at
OMRON HEALTHCARE EUROPE at the address mentioned in this instruction
manual.
Documentation is also available at www.omron-healthcare.com.

Correct Disposal of This Product


(Waste Electrical & Electronic Equipment)

This marking shown on the product or its literature, indicates that it should not
be disposed of, with other household wastes at the end of its working life. To
prevent possible harm to the environment or human health from uncontrolled
waste disposal, please separate this product from other types of wastes and
recycle it responsibly to promote the sustainable reuse of material resources.
Household users should contact either the retailer where they purchased this
product, or their local government office, for details of where and how they can
return this item for environmentally safe recycling.
Business users should contact their supplier and check the terms and
conditions of the purchase contract. This product should not be mixed with
other commercial wastes for disposal.

26
NE-C28P-E_main.book Page 27 Friday, October 7, 2011 2:48 PM

10.Accessories

Omron V.V.T. Nebulizer Kit Air Tube (PVC, 200cm) Mouthpiece


N-01 Set Model: NEB-TP-L20E Model: C28-3P-E
Model: C801S-NSET3-E Order No.: 9520606-3 Order No.: 9956273-5
Order No.: 9515573-6

Mouthpiece

Air Tube (Silicone, 100cm) Nosepiece


Model: C28-6-E Model: C28-7-E
Order No.: 9956269-7 Order No.: 9956274-3

Nebulizer Kit No. : N-01

Adult Mask (PVC) Child Mask (PVC) Adult Mask (SEBS)


Model: C28-8-E Model: C28-9-E Model: U10-1P-E
Order No.: 9956275-1 Order No.: 9956276-0 Order No.: 9956312-0

with elastic band

Child Mask (SEBS) Set Air Filters (x 5)


Model: U10-2P-E Model: C28-2-E
Order No.: 9956281-6 Order No.:9956271-9

EN

Child Mask (SEBS)


with elastic band

Adapter for Child Mask (SEBS)

27
NE-C28P-E_A_M.fm Page 28 Friday, October 7, 2011 4:09 PM

10. Accessories

Manufacturer OMRON HEALTHCARE Co., Ltd.


53, Kunotsubo, Terado-cho, Muko, Kyoto,
617-0002 JAPAN

EU-representative OMRON HEALTHCARE EUROPE B.V.


Scorpius 33, 2132 LR Hoofddorp, THE NETHERLANDS
www.omron-healthcare.com

Production facility OMRON (DALIAN) Co., Ltd.


Dalian, CHINA
OMRON HEALTHCARE UK LTD.
Opal Drive
Fox Milne, Milton Keynes, MK15 0DG, U.K.
OMRON MEDIZINTECHNIK HANDELSGESELLSCHAFT mbH
Subsidiary John-Deere-Str. 81a
68163 Mannheim, GERMANY
www.omron-medizintechnik.de
OMRON SANTÉ FRANCE SAS
14, rue de Lisbonne, 93561 Rosny-sous-Bois Cedex, FRANCE
Made in China