Ficha Tecnica Turbidimetro
Ficha Tecnica Turbidimetro
Turbidity meter/Turbidimeter
(Turbidity meter, portable Turbidity meter, Lab Turbidity meter, bench-top
Turbidity meter, portable Turbidity Analyser, bench-top Turbidity Analyser)
SRTU, SRW Series scattered light turbidity meter is used for measuring the scattering degree of the lights generated by the insoluble grain matters
suspended in water or transparent liquids, with capability to characterize the content of these suspended grain matters.
It can be broadly applied in the measurement of turbidity in power plants, waterworks, treatment stations of domestic sewage, beverage factories,
environmental protection departments, industrial water, wine and pharmacy industries, epidemic prevention divisions, hospitals, and etc.
SRTU-1, SRTU-1A Turbidity Meter
Specification:
Model SRTU-1 SRTU-1A
Minimum Principle 90° Scattered light
SRTU-1B Portable Turbidity Meter
Specification:
Model SRTU-1B
Minimum Principle 90° Scattered light
www.china-total.com/Product/meter/Turbidity-meter/Turbidity_meter.htm 1/9
10/4/2019 Turbidity_meter
0~10.00;
0~10.00; 0~2.000;
Measuring range 0~100.0;
0~100.0; 0~20.00;
(NTU) 0~400.0
0~400.0 0~200.0
0~2000
0~10.00;
0~10.00; 0~2.000;
Measuring range 0~100.0;
0~100.0; 0~20.00;
(NTU) 0~400.0
0~400.0 0~200.0
0~2000
www.china-total.com/Product/meter/Turbidity-meter/Turbidity_meter.htm 2/9
10/4/2019 Turbidity_meter
Repeatability ≤1.0%
Computer interface RS232 data communication interface (optional)
High-precision; Good stability; Auto-switch measuring
Characteristics
range; Can be used for high and low turbidity analysis.
External dimension 272*205*118mm
SRTU-3, SRTU-3A Turbidity meter
Specification:
Model SRTU-3 SRTU-3A
Repeatability ≤1%
www.china-total.com/Product/meter/Turbidity-meter/Turbidity_meter.htm 3/9
10/4/2019 Turbidity_meter
SRTU-3BS Beer Turbidity meter
Specification:
Model SRTU-3BS
Repeatability ≤1%
SRTU-1M Bacteria Turbidity Meter
Specification:
Model SRTU-1M
Minimum readout (MCF) 0.01 MCF
Measuring range (MCF) 0~1.00MCF; 0~6.00MCF
Basic error F.S ≤3.0%F.S
Repeatability ≤1.0%
Computer interface RS232 data communication interface (optional)
Bench microcomputer type; Good stability; LCD
Characteristics
backlight display; Auto-switch measuring range.
External dimension 272*205*118mm
Common size: 15*100mm, 16*100mm, 18*50mm
Sample tube size Please indicate the specific size when place order.
(The tube size also can be customized)
Portable Turbidity meter
Product Characteristics
a) AC & DC power supply, low power consumption, cabinet and handsome contour design, facilitating field and
lab use;
b) Configured with auto power-off function and low power dissipation circuit, able to effectively prolong service
life of battery;
c) Adopting big-screen highly transparent LCD display, with comfortable reading, and free of ill influence from
sunlight;
d) Adopting low-drift and high-accuracy integrated circuit, efficient and timeproof light source, effectively
ensuring long reliable performance
e) Unique locative structure and high-accuracy light path system, effectively ensuring exactness and repeatability;
f) Low voltage indication, cuing users to change battery, to ensure exact and effective measurement;
g) Configured with output port of measured values, possible to connect with recording meter
Specifications
Model SRW-1B SRW-20B SRW-200B
Measure Principle 90° scattered light
Minimum readout (NTU) 0.1 0.01 0.01
Measuring range (NTU) 0~200 NTU 0~20 NTU 0~20, 0~200 NTU
Basic error F.S ±8% (+/-2.5% F.S)
Repeatability ≤1%
Zero draft (NTU) ±1%F.S
DC1.5V x 5 AA alkaline dry cells
Power supply
AC 220V/50Hz DC7.5V 0.2A adapter
Portable Turbidity meter
Product Characteristics
a, AC & DC power supply, low power consumption, cabinet and handsome contour design, facilitating field and lab
use;
b, Configured with auto power-off function and low power dissipation circuit, able to effectively prolong service life of
battery;
c, Microcomputer, touch keyboard, backlight LCD, standard parallel RS232 data communication interface
d, Custom-made high strength long service life lamp-house, provided with data storage and inquiry function, can
meet GLP requirements
www.china-total.com/Product/meter/Turbidity-meter/Turbidity_meter.htm 4/9
10/4/2019 Turbidity_meter
e, Data nonlinear processing and data smoothing function, quick and automatic multi-points calibration, self-
diagnostic information prompt, selectable span – automatic or manual switch
f, Adopting big-screen highly transparent LCD display, with comfortable reading, and free of ill influence from
sunlight;
g, Adopting low-drift and high-accuracy integrated circuit, efficient and time proof light source, effectively ensuring
long reliable performance
h. Unique locative structure and high-accuracy light path system, effectively ensuring exactness and repeatability;
i, Low voltage indication, cuing users to change battery, to ensure exact and effective measurement;
j, Configured with output port of measured values, possible to connect with recording meter
Specifications
Model SRW-500B SRW-2B SRW-3B SRW-4000B
Measure Principle 90°scattered light
Minimum readout (NTU) 0.01 0.001 0.01 0.001
0~50 0~10 0~10 0~10
0~500 0~100 0~100 0~100
Measuring range (NTU) 0~500 0~1000 0~500
0~1000
0~4000
Basic error (F.S.) ±2%
Repeatability ≤1%
Zero draft (F.S.) ±0.5% (0.1 NTU) ±0.5% (0.02 NTU)
Computer interface RS232 data communication interface
DC1.5V x 5 AA alkaline dry cells
Power supply
AC 220V/50Hz DC7.5V 0.2A adapter
Bench Turbidity meter
Product Characteristics
▶Streamline outward,Beautiful international fashilnable snow lacquer.
▶wide&bright LCD display.
▶Operation easily,wide measuring range,high rate of function price.
▶Special fixed position construction,high precision light parh.
▶Adopting low-drift and high-accuracy integrated circuit, efficient and timeproof light source, effectively
ensuring long reliable performance
▶Added chromaticity compensation system,avoided the disturber of sample color.
Specifications
Model SRW-20S SRW-1A SRW-200S
Measure Principle 90° scattered light
Minimum readout (NTU) 0.01 0.1 0.01
Measuring range (NTU) 0~20 NTU 0~200 NTU 0~20, 0~200 NTU
Basic error F.S ±8%(±2.5% F.S.)
Repeatability ≤ 0.8%
Zero draft ±0.8%F.S
Power supply AC 220V/50Hz, 3W
Benchtop one, Economical, Easy Operated,
Others
Good Stability.
Bench Turbidity meter
Product Characteristics
▶Streamline outward,Beautiful international fashilnable snow lacquer.
▶wide&bright LCD display.
▶Operation easily,wide measuring range,high rate of function price.
▶Special fixed position construction,high precision light parh.
▶Adopting low-drift and high-accuracy integrated circuit, efficient and timeproof light source, effectively ensuring long
reliable performance
▶Added chromaticity compensation system,avoided the disturber of sample color.
Specifications
Model SRW-1 SRW-200 SRW-800
Measure Principle 90º scattered light
Minimum readout NTU 0.1 0.01 0.1
Measuring range NTU 0-100 2.5%F.S 0-20 2%F.S 0-200 2.5%F.S
0-200
Basic error F.S 0-200 2.5%F.S 0-800 3%F.S
2.5%F.S
Repeatability 1%
Zero draft NTU 0.2 0.02 0.2
for low
for General- for high measuring
Using measuring
purchase rane
range
Bench Turbidity meter
Product Characteristics
▶Microcomputer, touch keyboard, backlight LCD, standard parallel RS232 data communication interface
▶Custom-made high strength long service life lamp-house, provided with data storage and inquiry function, can meet
GLP requirements
▶Data nonlinear processing and data smoothing function, quick and automatic multi-points calibration, self-diagnostic
information prompt, selectable span ¨C automatic or manual switch
www.china-total.com/Product/meter/Turbidity-meter/Turbidity_meter.htm 5/9
10/4/2019 Turbidity_meter
▶Set average measuring mode with shortcut to obtain correct data within quickest time, applicable particularly to
measure extreme-low turbidity and can be used to measure unstable water sample
▶Precise light route system, reliable positioning structure, effective tone compensation, direct turbidity reading
Specifications
Model SRW-200A SRW-2 SRW-2A SRW-100
Model (with printer) SRW-200AP SRW-2P SRW-2AP SRW-100P
Measure Principle 90º scattered light
Minimum readout NTU 0.01 0.001 0.01 0.001
0-10
0-10 2%F.S 0-100 2%F.S
2%F.S
0-100 0-100
Measuring range NTU 0-100 2%F.S 0-200 3%F.S
2%F.S 2%F.S
0-200
0-200 2%F.S 0-500 3%F.S
2%F.S
0-500
Basic error F.S
2%F.S
Repeatability 1%
Zero draft NTU 0.1 0.02 0.1 0.02
Computer interface RS232 data communication interface
Average measuring method, Year/Month/Day/Time
Characteristics display, Data record and auto range change, auto zero,
1~5 point calibration, RS232 port
Bench Turbidity meter
Product Characteristics
▶Microcomputer, touch keyboard, backlight LCD, standard parallel RS232 data communication interface
▶Custom-made high strength long service life lamp-house, provided with data storage and inquiry function, can meet
GLP requirements
▶Data nonlinear processing and data smoothing function, quick and automatic multi-points calibration, self-diagnostic
information prompt, selectable span ¨C automatic or manual switch
▶Set average measuring mode with shortcut to obtain correct data within quickest time, applicable particularly to
measure extreme-low turbidity and can be used to measure unstable water sample
▶Precise light route system, reliable positioning structure, effective tone compensation, direct turbidity reading
Specifications
Model SRW-3 SRW-3A
Model (with printer) SRW-3P SRW-3AP
Measure Principle 90 scattered light
Minimum readout NTU 0.01 0.001
Measuring range NTU 0-10/0-100/0-200/0-500/0-1000
Basic error F.S 2%F.S
Repeatability 1%
Zero draft NTU 0.02
Computer interface RS232 data communication interface
Laboratory Bench Turbidity meter
Product Characteristics
►Microcomputer, touch keyboard, backlight LCD, standard parallel RS232 data communication interface
►Custom-made high strength long service life lamp-house, provided with data storage and inquiry function,
can meet GLP requirements
►Data nonlinear processing and data smoothing function, quick and automatic multi-points calibration, self-
diagnostic information prompt, selectable span ¨C automatic or manual switch
►Set average measuring mode with shortcut to obtain correct data within quickest time, applicable particularly
to measure extreme-low turbidity and can be used to measure unstable water sample
►Precise light route system, reliable positioning structure, effective tone compensation, direct turbidity reading
, built-in or external printer is optional
Specifications
SRW-
Model SRW-2000 SRW-2000A SRW-4000
4000A
Measure Principle 90º scattered light
Minimum readout NTU 0.001 0.01 0.001 0.01
Bacterium turbidity
Bacterium turbidity analysis
analysis of direct
of direct reading ( MCF )
reading ( MCF )
Maxwell turbidity unit
Maxwell turbidity unit
Characteristics
Note: model with " P" with Note: model with " P"
a built-in printer with a built-in printer
Beer Turbidity Meter SRW-2PJ SRW-2PJP
Summary of functions
Used for measuring suspended in water or in transparent liquids insoluble particulate matter generated by the
degree of light scattering, and quantitative characterization of the suspended particulate matter content. The
instrument is mainly used specially for determination of turbidity in beer, wine etc. At the same time it can also
be applied to the power plant, the pure water plant, water plant, sewage treatment plants, beverage plants,
environmental protection departments, industrial water, wine making industry and the pharmaceutical industry,
hospitals, epidemic prevention departments and other departments of the turbidity value.
Product Characteristics
Microcomputer, the touch keyboard, LCD backlight LCD display, serial RS232 data communication interface
standard
Instrumental readings automatically in response to balance or at a desired time interval of automatic data
collection, can effectively stabilize readings
Instrumental fast setting average measurement mode, with the most short time to get the correct data,
measurable unstable water
Place the built-in clock memory storage system, can be long-term storage and call the recent correction and
measurement data
Multiple measurement model, measurement units were available, a special high intensity light source, the
running time up to 100000 hours
Precise optical system, the reliable positioning structure, effective chroma compensation, direct read EBC
turbidity value
Specifications
Model SRW-2PJ SRW-2PJP
Minimum Principle Scattering of light Scattering of light
SRTU-100 Portable Turbidity Meter
Features:
• Easy-to-use portable turbidimeter is equipped with a standard carrying case.
• 2 to 5 points calibration using the Formazin Standards.
• Selectable four turbidity units, including the NTU (Nephelometric Turbidity Unit), FNU (Formazine Turbidity
Unit), EBC (European Brewery Convention) and ASBC (American Society of Brewing Chemists).
• Single measurement mode automatically recognizes the stable reading.
• Continuous measurement mode that can used for indexing and matching the sample vials.
• Auto-Power Off helps user effectively conserve battery life.
• System menu allows setting the 7 parameters, including the date and time, resolution, etc.
• Reset feature automatically resumes all settings back to factory default settings.
• Expanded memory stores and recalls up to 100 readings.
• Stored data can be transferred into computer by USB communicationinterface.
• Multi-mode power scheme (battery, power adapter, USB port) ensures that use the meter smoothly.
Specifications:
Measurement Method ISO 7027 nephelometric method (90°)
Measurement Range 0~1100 NTU/FNU, 0~275 EBC, 0~9999 ASBC
0.01 (0~100 NTU), 0.1 (100~999 NTU), 1 (999~1100
Resolution
NTU)
Accuracy ±2% (0~500 NTU), ±3% (501~1100 NTU)
Calibration Standards 0.02, 10, 200, 500, 1000 NTU or User-Defined
Light Source Infrared-emitting diode (850 nm wavelength)
Detector Silicon Photodiode
Stray Light < 0.02 NTU
Sample Vials 60(H)×25(Dia.)mm
Accessories: Power Off Manual or Automatic (15 minutes)
Glass Sample Vials
Calibration Standards 0, 10, 200 and 1000 NTU Memory Stores up to 100 data sets
Lint-Free Cloth Output USB Communication Interface
9V Battery
Carrying Case Operating Temperature 0~60°C
Power Requirements 1×9V Battery
Dimensions 180(L)×85(W)×70(H)mm
Weight 300g
SRTU-200 Benchtop Turbidity Meter
Features:
• Standard benchtop turbidity meter is equipped with a 4.5" LCD display.
• 2 to 7 points calibration using the Formazin Standards.
• Selectable four turbidity units, including the NTU, FNU, EBC and ASBC.
• TSS conversion factor ensures accurate measurement for Total Suspended Solids.
• Continuous measurement mode that can used for indexing and matching the sample vials.
• Hold function freezes current reading for easy viewing and recording.
• Password protection prevents the unauthorized calibration and settings.
• Help message as a operational guide that helps user quickly begin using the meter.
• System menu allows setting the 9 parameters, including the date and time, resolution, etc.
• Reset feature automatically resumes all settings back to factory default settings.
• Expanded memory stores and recalls up to 200 readings.
• Stored data can be transferred into computer by USB communication interface.
Specifications:
Measurement Method ISO 7027 nephelometric method (90°)
Turbidity Range 0~2000 NTU/FNU, 0~500 EBC, 0~9999 ASBC
TSS Range Depends on conversion factor
0.01 (0~100 NTU), 0.1 (100~999 NTU), 1 (999~2000
Resolution
NTU)
Accuracy ±2% (0~500 NTU), ±3% (501~2000 NTU)
0.02, 10, 200, 500, 1000, 1500, 2000 NTU or User-
Calibration Standards
Defined
Light Source Infrared-emitting diode (850 nm wavelength)
Detector Silicon Photodiode
Stray Light < 0.02 NTU
Sample Vials 60(H) ×25(Dia)mm
Accessories:
Glass Sample Vials Power Off Manual or Automatic (2 hours after last key pressed)
www.china-total.com/Product/meter/Turbidity-meter/Turbidity_meter.htm 8/9
10/4/2019 Turbidity_meter
Calibration Standards 0, 10, 200, 500, 1000 and 1500 NTU Reset Function Yes
Lint-Free Cloth Memory Stores up to 200 data sets
DC12V Power Adapter
Output USB Communication Interface
Operating
0~60°C
Temperature
Power Requirements DC12V Power Adapter
Dimensions 250(L)×177(W)×96(H)mm
Weight 1.2kg
SRTU-2000SS Digital Turbidity/ TSS Online Analyzer
Features:
* Performance characteristics;
* High range TUR/ TSS two in one Turbidity analzer;
* RS485 Digital communication, Modbus RTU standard protocol;
* Combine arbitrarily with the HMI and the wireless module to the system integration;
* Build-in optical maintenance brush/ wiper, cycling/ presetting cleaning optical
window.
Specifications:
Turbidity: (0.01~4000)NTU;
Measurement range TSS: (0.001~50)g/L
Light source 850nm (infrared light)
<10NTU, 0.001NTU
Resolution <100NTU, 0.01NTU
<1000NTU, 0.1NTU
<4000NTU, 1NTU
Turbidity: (0-100)NTU ±3.0%FS
Accuracy (100-1000)NTU ±1.0%FS
(1000-4000)NTU ±1.0%FS
TSS: ±5% of reading
Repeatability 1%
Sample flow <3m/s
Cable length Standard length 10m (or customize)
Maintenance Build-in brush, settable period
Power supply DC 24V±15%
Installation Submersible installation
Application: Communication port RS485(Modbus RTU protocol)
The online Turbidity monitoring for the source water quality of the Pressure range ≤6bar
environment environmental monitor station, for the industrial water Working Environment (0~40)°C
treatment Recycling process, for the every water inlet and outlet in
Weight 1.5Kg
the sewage water treatment, and the TSS monitoring for the sludge
Consumption <10W
treatment process. Widely used in the water supply factory, paper
mill, coal washery and the like Turbidity online monitoring.
( If you need prices, pictures without our water mark, user manual or E-catalogue for above meters, please ask for at: [email protected] )