Vacuum Pump V-100
Technical data sheet
The Vacuum Pump V-100 is an economical and chemically resistant PTFE diaphragm vacuum
pump with a space saving design. It covers the essential needs in vacuum generation. On/off-
regulated by the Interface I-100, it is optimally suited to work with the Rotavapor ® R-100. The
ready to use vacuum pump will be delivered with a silencer and 2 m of vacuum tubing.
Scope of delivery
All models are delivered ready to use and are complete of:
Components V-100
Vacuum Pump V-100 1
Vacuum hose 1
Power cable 1
Silencer 1
Set hose connection 1
Interface I-100 *
Woulff bottle *
Condensation assembly *
Communication cable (Mini-DIN) *
*Optional, according to order code
Order code
A recommended solution with the Vacuum Pump V-100 in combination with Rotavapor® R-100, Interface I-100
and Recirculating Chiller F-100 / F105 can be found in the technical data sheet of the Rotavapor® R-100, “Ex-
emplary solution”.
Choose the configuration according to your needs:
1 1 V 1 0 0
Interface
0 Without
1 Interface I-100 (incl. power pack)**
Woulff bottle
0 Without**
1 Woulff bottle
Condensation assembly
0 Without
1 Secondary condenser with insulation
2 Secondary cold trap
**When using the vacuum pump with the Interface, the valve unit is not needed, but a Woulff bottle.
A Woulff bottle is recommended for any configuration. All required holders are included.
2
Recommended combinations of Rotavapor ® / vacuum systems
Available instrument. Recommended vacuum system
R-100 without I-100 V-100 with I-100 and Woulff bottle
R-100 with I-100 V-100 (including Woulff bottle) with additional Mini-DIN cable
(11060882)
Dimensions and weights
Dimensions (WxHxD) Weight
Shipping box V-100 incl. I-100 340 x 630 x 340 mm –
V-100, silencer 180 x 275 x 210 mm 5.6 kg
V-100, secondary condenser, silencer 220 x 365 x 335 mm 6.6 kg
V-100, cold trap, silencer 220 x 415 x 300 mm 6.6 kg
V-100, I-100, Woulff bottle, silencer 180 x 400 x 350 mm 6.6 kg
V-100, I-100, Woulff bottle, secondary condenser, silencer 220 x 400 x 350 mm 7.5 kg
V-100, I-100, Woulff bottle, cold trap, silencer 220 x 415 x 350 mm 7.5 kg
Technical data
Vacuum Pump V-100
Suction capacity (DIN 28432) 1.5 m3/h*
Number of steps (heads) 2 (2)
Final vacuum (absolute) 10 mbar (± 2 mbar)
Connections GL14
Power consumption 150 W
Power consumption in ECO2-mode* (70 %) 70 W
Electrical requirements 100 – 240 V, 50/60 Hz
Pump motor DC, brushless
Nominal speed max. 1280 rpm
Nominal speed in ECO2-mode* (70 %) 70 % of nominal speed
Sound level (EN 61010-1) 32 – 57 dB(A)
57 dB(A) with 100 % load
Material in contact PTFE, FEP, FFKM, PPS
Approvals CE
* The ECO2-mode is active after 1 h with suction capacity of 1 m3/h
3
Compatibility
Compatibility of vacuum pumps with vacuum controllers of different generations
Compatibility Working principle Communication cable
V-100 I-100 yes on/off control 038010 / 11060882*
V-850 / V-855 yes on/off control 038010 / 11060882*
V-800 / V-805 no - -
V-700 / I-100 yes on/off control 038010
V-710 V-850 / V-855 yes speed controlled 044989
V-800 / V-805 yes on/off control 038010
V-500 I-100 yes on/off control 038010
V-850 / V-855 yes on/off control 038010
V-800 / V-805 yes on/off control 038010
Regulated: On/Off function - the vacuum pump is switched ON and OFF automatically via the Vacuum Controller / Interface
* Length of communication cable: 038010 (1.5 m), 11060882 (0.6m)
Explanation of terms
On/Off control: the pump is automatically turned On and Off via the interface -> On/Off
communication
Speed controlled: the pump is turned On and Off as well as speed controlled via the interface
4
Functional principle
The diaphragm pump is a positive displacement pump that uses a combination of the reciprocating action of
a PTFE diaphragm and non-return check valves. The eccentric module lets the diaphragm move strictly axial
which increase the duability.
ECO2-mode
The ECO²-mode is automatically activated after 1 hour of operation, hence the pump does not work on its full
capacity. In the ECO2-mode the pump consumes less electrical power and works under reduced stress. As a
consequence, the overall maintenance costs will be reduced over the lifetime of the pump.
Performance characteristics
The graph illustrate the evacuation time of a laboratory as well as an industrial Rotavapor.
V-100: Evacuation time
1000
800
600
Pressure (mbar)
(Voulme = 3.7 L)
400
200
0
0 1 2 3 4 5 6
Time (min)
Upgrade
The Vacuum Pump V-100 can easily be upgraded to a regulated vacuum source by adding the Interface I-100.
In addition to the details listed below please check the technical data sheet of the Interface I-100 for further
information.
5
Vacuum pump accessories
V-100
Secondary condenser, insulation of condenser, holder, receiving flask, 500mL 047180
Condenses vapor that goes through the pump, is cooled by cooling liquid
Secondary cold trap, holder, receiving flask, 500mL 047190
Condenses vapor that goes through the pump, is cooled by dry ice
Valve unit, Woulff bottle, 125mL, P+G, holder for lab-stand and vacuum pump 047160
Non-return valve and condensate trap
Woulff bottle, 125mL, P+G, holder for vacuum pump 047170
For trapping particles and droplets and for pressure equalisation, recommended for stand-alone
systems and in combination with a vacuum pump
Woulff bottle, 3-neck, 800mL, P+G 025519
For trapping particles and droplets
Support set, connection tube, mounting device 11061448
To mount Interface on vacuum pump
Communication cable, Mini-DIN, 0.6m 11060882
Connection between Interface I-100 and Vacuum Pump V-100
Tubing, for vacuum, silicone rubber, Ø6/16mm, red, per m 017622
Tubing, for vacuum, PTFE (rigid), Ø8/10mm, white/transparent, per m 027277
Tubing, for vacuum, PEP, Ø6/8mm, transparent, per m 027900
Tubing, for cooling media, Nylflex, PVC-P, Ø8/14mm, transparent, per m 004113
Tubing, for cooling media, silicone rubber, Ø6/9mm, transaprent, per m 004133
Additional glassware
Additional glassware (e.g. receiving flask for secondary condenser and cold trap) can be found in the
“Laboratory Evaporation Glassware” brochure.
Vacuum Pump wear parts Order number
Set of check valves, O-ring (4pcs) 047156
O-ring, FKM/FEP, Ø28.25/2.62mm (for secondary condenser) 11057661
Set of O-rings, FFKM, Ø15.4/1.55mm, white (4pcs) 11057136
Membrane, membrane clamp, support ring 047153
Set of membranes, membrane clamp, support ring (10pcs) 11055214
Hose barb, bent, GL14, silicone seal 018916
Set of hose barbs, bent, GL14, silicone seal, incl. cap nut (4pcs each) 037287
Set of hose barbs, bent, GL14, FPM seal, incl. cap nut (4pcs each) 040295
Set of hose barbs, bent, GL14, EPDM seal, incl. cap nut (4pcs each) 043129
6
Vacuum Pump wear parts Order number
Set of hose barbs, straight, GL14, silicone seal, incl. cap nut (4pcs each) 037642
Set of hose barbs, straight, GL14, FPM seal, incl. cap nut (4pcs each) 040296
Set of hose barbs, straight, GL14, EPDM seal, incl. cap nut (4pcs each) 043128
Set of hose barbs, bent (1pc), straight (1pc), GL14, silicone seal, incl. cap nut (2pcs) 041939
Set of hose barbs, bent (4pcs), straight (2pcs), GL14, silicone seal, incl. cap nut 038000
(6pcs)
Set of hose barbs, bent, GL14, silicone seal (3pcs each) 041987
Set of cap nuts, screw cap with hole, GL14, with FEP seal (10pcs) 041999
Set of cap nuts, screw cap with hole, GL14 (10pcs) 041956
Set of cap nuts, screw cap without hole, GL14 (5pcs) 040624
Set of seals, for hose barb, GL14, silicone, red (20pcs) 040023
Set of seals, for hose barb, GL14, FPM, green (10pcs)* 040040
Set of seals, for hose barb, GL14, EPDM, black (10pcs)* 040029
Seal GL14, for cap nut, GL14, FEP 038225
11595196D V-100 Data Sheet en 1602 / Technical data are subject to change without
notice / Quality Systems ISO 9001
www.buchi.com/laboratory-evaporation Quality in your hands