Troubleshooting: Lunar
Troubleshooting: Lunar
E (3/98)
Chapter 8
Troubleshooting
This chapter contains troubleshooting techniques for diagnostic failures, failing quality
assurance tests, mechanical failures and imaging problems.
Troubleshooting Chapter 8
LUNAR® DPX-IQ Service Manual Rev. E (3/98)
Chapter 8 Troubleshooting
LUNAR® DPX-IQ Service Manual Rev. E (3/98)
If the transverse motion stops during the scan, see section 8.7. In Total Body scanning, a
defect in the transverse mechanics may cause one of the transverse limit switches to be closed.
Closure of a limit switch prevents further operation of the motor. The shutter will close, the
shutter open lamp will go out and the end of exposure alarm will sound. Seconds later the error
message will appear on the screen. For further information, see section 8.7.
The transverse and longitudinal motion detection system is only operational during patient
scans. If the problem occurs during a quality assurance scan, for instance, then it is probable
that the high voltage system is arcing.
If the error occurs consistently on the first line of a patient scan, and the scanner is moving in
the transverse direction, then check the following:
Remove the table top and the front panel. Verify that the slotted disk at the front of the lower
arm rail is in the middle of the slot between the photo diode and photo transistor. The slotted
disk must be completely flat and remain in the center of the sensor slot during its entire rotation.
If the slotted disk has been in contact with the optical sensor, disassemble this mechanism, and
clean the sensor and the slots of the disk.
The pulses that normally enter the OINK board at J14 (the black wire at the center of the
connector) may have stopped. These pulses are necessary to keep the OINK board from
sending an interrupt to the SBC. These pulses can be seen with an oscilloscope (see figure 1).
Use a voltmeter to measure the voltage on the positive side of C16 (33 µf capacitor). This
voltage should never approach 2 Volts DC during a scan if the pulses are present to reset
U11A. If this voltage stays well below 2VDC during the scan, but the error 15 still occurs, the
interrupt was invalid. Check for arcing in the high voltage system or replace the OINK and SBC.
If the longitudinal motion stops in the middle of the scan, see "Longitudinal Motion Failure" in
section 8.7.
This supply is turned on by the X-ray Relay, so verify that the Relay is closing. If not, then either
the Relay is bad or it is not receiving the signal from the SBC via the OINK (see section 4.5).
The Tube Head Thermostat is wired in series with the Relay, so if it has opened, the Relay will
not be able to close. (see Failure #30 below)
The 28VDC should also be measured at the Terminal Block. If not, check the continuity of the
wiring and refasten all connections. Also, check the wire tie-downs for excess tension they may
be putting on the wires.
It may be necessary to check the wiring from the Terminal Block to the MAX board and to the
High Voltage Power Supplies.
If any of the QA test results fail, none of the results are considered valid. The results will be
stored in the Quality Assurance History file (typically C:\LUNAR\ANC\DPXS.DBF), but these
values will not be averaged with the other results for calibration purposes. In addition, after two
days the operator will find messages appearing on the screen warning that no valid Quality
Assurance Test has been run within the last 48 hours. After two weeks without a valid Quality
Assurance Test, the software will prevent patient scans. Recent valid QA's are necessary for
accurate results.
Lamp Operation
Just before the QA begins, the operator is asked if the Shutter Open, X-Ray On, and Power-On
lights are illuminated. If the Green Power-On light has burned out and yet the other two lights
are lit, it is recommended to ignore the Green light and answer "yes", allowing the QA to pass.
Answering "no" will fail the QA regardless of all the other tests passing, and failed QA's are not
stored in the database.
However, it is not recommended that scanner operation be continued if the X-ray On light fails,
and the scanner will not operate if the Shutter Open light fails. These LED's are the only
indicator of exposure to x-rays and must be replaced as soon as possible.
Peak Setting
Customers will note slight fluctuations in the Peak Setting, but only a shift of >50 units in
consecutive QA's will give a "fail" evaluation. Drift in the Peak setting should be monitored by
the customer.
Background
The background test will fail if other sources of radiation are located in the room or nearby on
the other side of walls.
Verify that the lead cup is installed immediately underneath the PMT and that other sources of
radiation are not causing the failure. Any other sources of radiation in the vicinity of the scanner
which causes failure of this background measurement must be removed for proper operation of
the scanner.
Spillover
An increase in the Spillover percentage over time is an indication that the PMT is losing
resolution, even if the Spillover test does not fail. Trends in excess of those specified on in
section 8.15 indicate service is required. A large change in the Spillover can be explained in
some cases by standard positioning (or a small change in the home position of the scanner). If
the beam is not fully eclipsed by the brass piece during the Spillover test, the Spillover value
increases dramatically (see section 8.3)
CHI Square
This is a statistical test to measure the spread of the Air Count data. Under normal conditions,
the number of counts obtained in a large number of samples should form a Gaussian
distribution. The Chi Square test is a measure of how the data fits to a Gaussian distribution,
based on a certain confidence factor. There is a small statistical probability that this test will fail
even though there is no problem. Rare failure of this test parameter should not cause concern.
The Air Ratio value should remain very constant over time. Check to see that the Air Ratio has
not varied from the value at the time of installation by +0.02 or greater. So if the Air Ratio was
0.59 at installation, then 0.58 and 0.60 are also acceptable, but 0.61 or 0.57 would be a cause
for concern.
Version 4.2a and higher software revisions have an additional test for air counts called “True Air
Ratio”. This test moves the beam off the standard block and into unobstructed air, collects data,
and calculates the count ratio. The normal air count ratio is sensitive to rollunder but not
rollover. The True Air Ratio is sensitive to rollover but not rollunder. The combination of the two
should detect any problems with the AGS system.
Unfortunately, Air Ratio changes are usually not noticed until the test results fail, and sometimes
the customer will not even notice the problem until they are warned that no valid Quality
Assurance Test has been run within the last 48 hours. Therefore, it is important that the Quality
Assurance trends be inspected during each service visit (see section 8.15).
It is not possible to view the Air Ratio trends in the Quality Assurance History file, but it is
possible to view the Low Channel Air Counts. A change in the Low Channel Air Counts will
usually be accompanied by a change in Air Ratio (see "Decreasing Air Counts" in section 8.5).
If the graph of the Low Channel Air Counts shows a trend, view the Quality Assurance History,
and calculate the initial and final Air Ratio by dividing HIGH by LOW. Compare these two
values to determine if the Air Ratio has changed 0.02 or more.
These are very difficult to diagnose by a method other than substitution of new components until
the Air Ratio returns to normal. The Scintillation Detector deterioration or unstable High Voltage
Power Supply, will usually cause the Air Matrix Scan (see section 5.1) to fail. The specification
for XORB Board transorbs is that they must allow less than 0.5 micro amperes reverse bias
current. At LUNAR, each transorb is measured by applying a 5 Volt reverse bias to the transorb
and a 100 kΩ 1% resistor connected in series. The voltage measured across the 100 kΩ
resistor must then be less than 50 mV.
Homing Problem
The scanner may not be finding the correct "Home" position. The correct "Home" position
aligns the center of the x-ray beam with the center of the Brass Piece when the Standard is
correctly positioned. The Air Counts must be obtained through the plastic part of the Standard
next to the Brass Piece. If the Air Counts are obtained with the x-ray beam passing through the
Brass Piece or missing the standard entirely, the Quality Assurance Scan will fail. For additional
information on these symptoms, see "Standard Positioning Problems", section 8.3.
The values recorded for these tests should remain fairly constant over time. Variations between
QA's of under 25 steps should not be a cause for concern as 1 transverse step = 0.05 mm and
1 longitudinal step = 0.1 mm, so the actual variation is only a few millimeters.
If the number of steps continually increases from QA to QA, this could indicate an impediment to
the scanner's motion and should be rectified.
Tissue Value
The tissue value should not be the sole failing parameter. The failure of this test is usually
accompanied by problems with Bone Mineral values (see below) or Air Ratio trends (see
above). If this is failing alone it is probably a problem with the counting system (Detector, AGS
or DCA's).
Another cause of failing %CV can be arcing during the QA. This can be detected by examining
the QA History graphically and looking for variation of the Large BM values (see Figures 1 and 2
below). Note the few data points that vary. These are the early signs of arcing.
This will also be apparent (but not obvious) on the QA Results printout. The arc occurred in the
third standard scan line and elevated the BM values. If a customer reports a failing QA because
of a %CV being too high, be aware that this could be an early warning of arcing. Obtain from
the site the DPXS.DBF file for analysis. Look for variation of the Large BM values. Arcs of
greater magnitude can also affect more than one line of the standard scan.
It is very important to notice these early warning signs of arcing so that the system can be re-
greased before any damage is done to the high voltage cable connectors or the tube head.
Limits
(High keV) 0 50
Chi Square 0 40
In the first example of low kV (see Figure 4), defective power supplies are most likely
responsible.
DETECTOR
1 2 3 4 5 MEAN SD %CV
LARGE BM 260.9 264.0 262.4 259.5 260.1 261.4 1.64 0.63
WIDTH 503 507 507 498 503 504 3.32 0.66
MEDIUM BM 192.7 194.6 193.7 194.6 194.3 194.0 0.73 0.37
WIDTH 437 437 436 433 434 435 1.62 0.37
SMALL BM 137.7 138.5 139.6 137.5 138.6 138.4 0.73 0.53
WIDTH 370 370 374 369 369 370 1.85 0.50
In the case of high voltage (Figure 5), an arcing tube head is the most likely cause although
defective power supplies could cause this as well.
DETECTOR
1 2 3 4 5 MEAN SD %CV
LARGE BM 261.0 261.6 261.9 263.6 262.5 262.1 0.87 0.33
WIDTH 493 495 495 495 494 494 0.80 0.16
MEDIUM BM 196.1 195.4 196.9 195.0 195.2 195.7 0.71 0.36
WIDTH 425 423 427 425 424 425 1.33 0.31
SMALL BM 140.7 139.4 139.5 140.0 139.3 139.8 0.53 0.38
WIDTH 362 358 359 360 358 359 1.50 0.42
NOTE: It would be prudent in both of these cases to take a positive and negative power
supply, AND a tube head. Also, the XORB test points may look normal
(indicating 76 KV) in each case as well. This usually occurs where the power
supplies are defective and so the monitor voltages returned to the XORB may not
show the true voltage. When the tube head arcs causing the voltage to go too
high, then the XORB test points may indeed show the actual voltages.
The clearest indicators of these problems are the shape of the Peak, the Air Counts, and Air
Ratio. Note how these differ from cases where the standard block is positioned incorrectly (see
section 8.3).
Example 1
Figure 1is a normal passing QA with the standard block properly placed.
DETECTOR
1 2 3 4 5 MEAN SD %CV
LARGE BM 264.6 265.2 265.4 264.4 264.7 264.9 0.35 0.13
WIDTH 499 502 505 500 502 502 2.06 0.41
MEDIUM BM 199.3 197.5 196.8 195.8 197.0 197.3 1.13 0.57
WIDTH 432 430 428 428 428 429 1.60 0.37
SMALL BM 142.7 140.4 140.0 141.3 142.0 141.3 0.98 0.70
WIDTH 364 362 361 364 364 363 1.26 0.35
Example 2
Figure 2 is a QA in which the standard was not on the table. Note the errant Spillover value and
Air Counts (the Low KeV counts indicates the counters rolled over). In this case Spillover will
usually be approximately 110%. However, this is unpredictable as individual AGS boards try to
handle the superimposed signals from the unattenuated x-ray beam.
DETECTOR
Lights - Pass
Peak Setting 495 units Pass
Background (Low keV) 2 cps Pass
(High keV) 0 cps Pass
Beam Stop Action - Pass
Percent Spillover 129.64% Fail
Chi Square 6 Pass
Air Counts (Low keV) 14087 cps Fail
(High keV) 3593432 cps Fail
Air Ratio 255.10 Fail
Transverse Mechanics 12398/12401 steps Pass
Longitudinal Mechanics 19755 /19755 steps Pass
Tissue Value 3.350 Fail
Collimation Ratio 2.119 Fail
1 2 3 4 5 MEAN SD %CV
LARGE BM 61.7 44.6 90.0 35.6 70.4 60.51 19.21 31.772
WIDTH 53 61 85 62 70 66 10.83 16.36 2
MEDIUM BM 6.5 49.2 35.1 38.7 27.5 31.41 14.26 45.44
WIDTH 42 71 47 42 65 53 12.21 22.86 2
SMALL BM 47.7 1.3 38.2 78.5 33.3 39.81 24.84 62.38
WIDTH 42 42 58 112 43 59 26.99 45.44
Example 3
Figure 3 is a QA run with the standard upside down and backwards. Spillover was measured
through the white delrin plastic plug and Tissue Value was measured through the brass
Spillover piece.
DETECTOR
Lights - Pass
Peak Setting 515 units Pass
Background (Low keV) 0 cps Pass
(High keV) 0 cps Pass
Beam Stop Action - Pass
Percent Spillover 95.40% Fail
Chi Square 5 Pass
Air Counts (Low keV) 143640 cps Fail
(High keV) 455003 cps Pass
Air Ratio 3.17 Fail
Transverse Mechanics 12398/12401 steps Pass
Longitudinal Mechanics 19755 /19754 steps Pass
Tissue Value 1.599 Fail
Collimation Ratio 2.389 Fail
1 2 3 4 5 MEAN SD %CV
1
LARGE BM 1974.4 1986.7 1968.4 1966.0 1983.2 1975.8 8.10 0.41
WIDTH 385 390 379 383 382 384 3.66 0.95
MEDIUM BM 2357.7 2362.0 2359.0 2343.3 2340.1 2352.41 8.90 0.38
WIDTH 450 447 451 449 454 450 2.32 0.51
SMALL BM 2746.2 2755.5 2729.8 2741.3 2728.9 2742.31 8.44 0.31
WIDTH 526 523 518 520 519 521 2.93 0.56
Example 4
Misplacing the standard by approximately 1" toward the foot end will give results as in figure 4.
Spillover is measured through the air and the Air Counts and Peak are measured through the
brass.
DETECTOR
Lights - Pass
Peak Setting 510 units Pass
Background (Low keV) 1 cps Pass
(High keV) 1 cps Pass
Beam Stop Action - Pass
Percent Spillover 129.84% Fail
Chi Square 3 Pass
Air Counts (Low keV) 318 cps Fail
(High keV) 14625 cps Fail
Air Ratio 45.93 Fail
Transverse Mechanics 12399/12401 steps Pass
Longitudinal Mechanics 19754 /19754 steps Pass
Tissue Value -0.623 Fail
Collimation Ratio 2.123 Fail
1 2 3 4 5 MEAN SD %CV
LARGE BM 0.2 0.1 -0.4 1.6 4.1 1.11 1.61 142.442
WIDTH 1 1 1 1 1 1 0.00 0.00 2
MEDIUM BM 0.2 0.1 -0.4 1.6 4.1 1.11 1.61 142.44
WIDTH 1 1 1 1 1 1 0.00 0.00
SMALL BM 0.0 0.0 0.0 0.0 0.0 0.01 0.00 0.00
WIDTH -48 -40 -22 -13 -13 -27 14.33 0.00
Example 5
The QA in figure 5 is what would happen in the rare instance that someone would inadvertently
remove the standard during the QA (after the Peak) and then replace it having realized the QA
was in progress (after the Tissue Value Test). Although the five lines are scanned across the
block as normal, the errant Air Counts will cause the BM values and widths to deviate from
expected.
DETECTOR
Lights - Pass
Peak Setting 515 units Pass
Background (Low keV) 1 cps Pass
(High keV) 1 cps Pass
Beam Stop Action - Pass
Percent Spillover 129.66% Fail
Chi Square 1 Pass
Air Counts (Low keV) 318 cps Fail
(High keV) 3605174 cps Fail
Air Ratio 11322.78 Fail
Transverse Mechanics 12398/12401 steps Pass
Longitudinal Mechanics 19756 /19754 steps Pass
Tissue Value 1.264 Fail
Collimation Ratio 2.127 Fail
1 2 3 4 5 MEAN SD %CV
LARGE BM 0.2 0.3 0.2 0.2 0.2 0.21 0.2 10.912
WIDTH 1 1 1 1 1 1 0.00 0.00
MEDIUM BM 0.2 0.3 0.2 0.2 0.2 0.21 0.2 10.912
WIDTH 1 1 1 1 1 1 0.00 0.00
SMALL BM 0.2 0.3 0.2 0.2 0.2 0.21 0.2 10.912
WIDTH 1 1 1 1 1 1 0.00 0.00
Example 6
Placing the standard backwards with the brass piece toward the rear and the delrin plug
forward, yet both toward the foot end, will result in Spillover and Tissue Value being taken
through the black plastic of the block and Air Counts and Peak are taken through the delrin plug
(see figure 6).
DETECTOR
Lights - Pass
Peak Setting 515 units Pass
Background (Low keV) 1 cps Pass
(High keV) 0 cps Pass
Beam Stop Action - Pass
Percent Spillover 115.22% Fail
Chi Square 2 Pass
Air Counts (Low keV) 318 cps Fail
(High keV) 316090 cps Pass
Air Ratio 992.74 Fail
Transverse Mechanics 12398/12400 steps Pass
Longitudinal Mechanics 19753 /19754 steps Pass
Tissue Value 0.203 Fail
Collimation Ratio 5.601 Fail
1 2 3 4 5 MEAN SD %CV
LARGE BM 5.0 592.9 621.3 -5.8 645.3 371.71 304.35 81.872
WIDTH 42 400 493 42 454 286 201.5670.43 2
MEDIUM BM 12.0 717.8 657.3 687.4 728.3 560.51 275.41 49.13
WIDTH 42 501 485 502 533 413 185.9545.07 2
SMALL BM -0.2 898.3 109.3 95.0 55.2 231.51 335.55 144.94
WIDTH 42 608 94 76 46 173 218.25 126.01
If a customer reports a QA failure on a scanner that seemed to be operating fine and showed no
signs of mechanical problems while running the QA, the QA results should be examined for
characteristics like the above examples. Ask the customer to check the position of the standard
if it is still on the table, or have them rerun the QA making sure the block is placed correctly.
Check the signal at TP4 of the AGS board with an oscilloscope. Look for noise on the wave
form. This test-point is the input for the AGS, so noise here means a problem somewhere
towards the PMT. Also, check TP19 on the AGS with an oscilloscope. This is the output of the
AGS and would be a good indication of the AGS board's condition. If a storage oscilloscope is
used to look at these wave forms, they will look similar to the channel 1 signal (see Figure 1). If
a normal oscilloscope is used, the signal will appear as a superimposed multitude of bipolar
signals with different amplitudes.
If the TP19 wave form looks abnormal, replace either the AGS or the AGSDCA printed circuit
board assemblies. If the spare part is not immediately available, then disconnect J8 on the AGS
and run a calibration. This will run the signal through the AGS at unity gain. If normal results
reappear after this modification, the cause of the problem is confined to a defective AGS or
AGSDCA board.
If the "End of Exposure Alarm" rings during the time the Air Matrix scan is running, see "Alarm
Pings During Scan" in section 8.11. If the Shutter Open or the X-ray On lamps on the front
panel blink, during the test, see section 8.11 also.
While the Air Matrix scan is running, measure the voltage at test points 1, 2, 5 and 6 of the
XORB board. They should have approximately the following values respectively: -
0.150VDC, 3.8VDC, 0.150VDC, and 3.8VDC. A pure DC voltage on an oscilloscope at test
point 1 and 5 of the XORB will verify that there is constant current through the filament of the X-
ray Insert. If there is AC ripple in excess of 0.2 Vpp, the power supplies should be replaced.
Start and stop the x-rays several times while observing the voltage at TP3 and TP7 of the
XORB board. This is the programming voltage from the SBC, and although this voltage is
dependent on the feedback information returned to the SBC, the voltages at the XORB board
test points should be approximately the same each time the x-rays are produced. The AC ripple
on this signal must be less than 0.2 Vpp. Replace the SBC board if the proper control signal is
not present.
When the x-rays are off between scans, verify that the shutter and collimator slide apertures line
up with the aperture to the Tube Head beneath. Take some alignment pictures and verify that
the image is a bright, round image. If the image is slightly elliptical, the collimator, shutter and
Tube Head apertures are not aligned. Open and close the shutter and collimator to make sure
that they return to the same position each time. Turn the x-rays back on, and make sure that
you obtain nearly the same count rate each time the shutter is opened or the collimator is
cycled. Either the shutter or the collimator slide may be adjusted by loosening the set screw on
the solenoid mounting bracket and rotating the solenoid. Tighten the set screw after
adjustment.
8.5.2 No Counts
When the table to the right of the peak graph on the Quality Assurance Results printout is
entirely filled with zeros, use the Static Counter program in the Diagnostics (see section 5.1) to
create x-rays at 76 kV and 150 uA and open the Shutter for sampling. Then check the following
items :
B. If not, the voltage or current ramping has probably failed. Are both the red and
green LED's on the MAX Board illuminated?
2. If the red and green MAX Board LED's are illuminated, verify the following
test point voltages:
c. X-ray On LED
Test the X-Ray On LED by inserting it into the Power On
receptacle. Replace if defective.
If either of the test points is at 1.0 VDC, the High Voltage Power
Supply is delivering as much current as it possibly can, and has
automatically limited the voltage.
Check TP2 and TP6. If the voltage is approximately 3.8 VDC, the
voltage has been set properly.
Feel the heat sinks on the back of the High Voltage Power
Supplies. If one of the power supplies is cold, this is usually the
defective one. If one is warm and the other is hot, replace the hot
one. The best troubleshooting technique may be to substitute
power supplies.
If either TP2 or TP6 are incorrect, the possibility exists that one of
the High Voltage Cables or the Tube Head is shorted. The short
may be possible to find with an ohm meter, but often it takes
several kV to break down the defective component. It will be
difficult to troubleshoot this problem by any method other than part
substitution.
CAUTION: Never disconnect any connectors from the AMP board when the
power is on, or AMP-U1 will be destroyed.
If there is a signal, check the continuity of the coaxial cable with the
yellow strain reliefs. Remember that this cable passes through a
connector at the Arm Bulkhead, and a poor connection there could be the
cause of the problem.
Mark the position of the large and small collimator apertures on the slide channel
edge. Then lift the Collimator, taking care not to damage the limit switch, and
mark the open and close position of the shutter aperture center. Lift the shutter
slide, taking care not to damage the limit switch, and visually determine if the
apertures above match up with the bottom aperture.
Take another picture with the alignment film holder. The x-ray image should be
round. If the image is even slightly elliptical, the apertures are not aligned.
Remove the Collimator assembly from the Tube Head, and hold it up to a light.
Look through the aperture with the Shutter open and closed and the Collimator in
each position.
Verify that the SBC is properly controlling the current. TP11 of the MAX board should be
approximately 0.75 volts. This voltage is dependent on feedback information from the
power supplies, which makes it difficult to troubleshoot by any means other than SBC
substitution.
Check the current through the X-ray Insert. The absolute value of the voltage at test points 1
and 5 on the XORB board is proportional to the current through the X-ray Insert. 1 millivolt is
equal to 1 micro-ampere of current. A current setting of 750 µA should give a reading of 0.750
VDC at test points 1 and 5. If either of these voltages vary from the expected by more than 10
millivolts, the MAX board could be at fault. If TP1 and TP5 are more than 15 millivolts apart,
substitute new high voltage power supplies.
8.6 Arcing
The X-ray Tube Head Insert is an evacuated glass enclosure. An AC current is applied to the
filament inside the insert. It glows like the filament in a light bulb, and electrons are boiled off
into the evacuated space. A high voltage is applied between the anode and the cathode
causing electrons to rush toward the anode, striking it and creating x-rays. As long as the insert
is properly evacuated, there can be no internal arc. However, no insert can be totally evacuated
and impurities can be ionized creating a lightning like effect; arcing. During the arc the
resistance of the insert is dramatically decreased and a large amount of current flows.
The significant amounts of electromagnetic energy released inside the scanner can cause
problems with the electronics of the scanner system and may result in abnormal operations of
the scanner. An arcing scanner will have one or more of the following symptoms :
• complete lock up of the system requiring rebooting the computer,
• a diagnostic error code message which is undefined or inappropriate,
• receiving a screen that says an error has occurred refer to appendix I in the manual,
• a stripe or artifact in the image of the patient's scan, or
• 15-4 Transverse Motion Failure message.
To specifically locate the source of an arc, it is necessary to find out which high voltage power
supply provides the excessive current. This may be done by recording the power supply current
monitors at TP1 and TP5 of the XORB board with a storage oscilloscope. An arc typically
draws 1.5 mA from the power supply for about 1 ms. The trigger voltage should therefore be
set at 1.5V for TP1 and for TP5.
Arcing can also occur inside the high voltage connectors. This will usually result in a plainly
visible black or brown carbon track through the grease on the connector. In any case, once the
high voltage connectors have been removed from the Tube Head and the power supplies, they
should not be re-connected without first being cleaned and regreased (see procedure
DXSV0002 in chapter 6 appendix).
After cleaning the old grease off of the connectors, they should be carefully inspected for carbon
tracks. Look for these tracks on both the rubber cable connectors and on the phenolic sockets
of the Tube Head. If such tracks are found after cleaning, the following are the options for
returning the scanner to service:
If carbon tracks are found on the rubber cable connectors, they can be removed by excising the
damaged section with a sharp blade. Severe tracks can burn quite deep into the rubber, so
care must taken to remove all the damaged rubber. After the carbon tracks have been
removed, additional grease must be used when repacking the connection to fill in the volume of
the removed rubber.
If the above procedure requires too much rubber to be removed, or the carbon track looks very
severe, another option is to replace the high voltage cable.
Arcing 8-29
LUNAR® DPX-IQ Service Manual Rev. E (3/98)
Arc tracks will be impossible to see on the sockets in the Tube Head. Therefore, if arc tracks
are seen on the rubber cable connectors, the sockets should be sanded with emory cloth as a
precaution. After sanding the socket, remember to flush the socket with cleaning solution to
remove any particles.
If the tracking inside the socket is too severe or the carbon track cannot be removed, another
option is to replace the Tube Head.
8-30 Arcing
LUNAR® DPX-IQ Service Manual Rev. E (3/98)
The High Voltage Cables can also impede transverse motion toward the front Limit Switch. This
is caused when the cables have been tied down without enough play for the Tube Head and
Detector to move all the way forward. These problems should be investigated by manually
tripping both Limit Switches while inspecting for cable conflicts, binding, or tension problems.
On Total Body scans, a limit switch out of position could allow the Source/Detector Carriages to
hit the frame or panels before the Limit Switch is actuated.
Wiring
In rare instances, the wires from the Shutter Solenoid and/or Fans can snag on the bolts that
protrude through the frame on the foot end of the scanner. This is solved by properly tying
down these wires.
Transverse Motor
Check the Transverse Motor for a broken wire in one of the internal coils, or a bad electrical
connection to its Centent Motor Controller.
Transverse Centent
The Centent Motor Controller, if defective, will cause transverse motion problems. Sometimes
the controller works well enough to acquire some scans, but it will not provide enough torque to
complete every fast scan. The CURRENT SET voltage at terminal 11 on a properly operating
controller should be 14 to 16 VDC (transverse motor wired in parallel).
Transverse Belt
The Transverse Belt should not be excessively tightened or this will cause excessive binding in
the transverse mechanism. It should be possible to deflect the belt by 4 cm when it is properly
tightened. Sometimes the spare belt material near the clamp on the Tube Head Carriage
comes into contact with the forward gear and prevents the scanner from going all the way to
Home position.
Drive Wheels
The wheels that support the Tube Head and Detector Carriages must be adjusted so that they
come into perfect contact with the Transverse Rails. These wheel are best inspected with the
table top, front panel and arm covers off so one can sight down the extrusions. However, it is
possible to test the wheels by preventing any wheel from turning and seeing if the carriage will
still move. By preventing any wheel from turning, you should be able to tell that it slides along
the extrusion while the others roll. This indicates that the wheel has not been excessively
tightened down. This is least likely to be the source of transverse motion problems, as it is
unlikely that the adjustment of the wheels would have become any tighter over time. They
would be more likely to loosen over time and fail alignment tests (see Air Matrix Test, section
5.1). Adjustments can be made by loosening and rotating the eccentric bearings of any of the
lower wheels.
Cable Track
The major impediment to longitudinal motion is the plastic Cable Track that runs through the
trough at the rear of the scanner.
This track is attached at two points: at the Rear Longitudinal Carriage and to the scanner frame
(low, rear and center) each spot by 4 bolts. Also, adequate slack must be left in the cables
inside the Cable Track or they will stop the arm from moving fully to the foot end.
Should the Cable Track detach from the scanner frame, it will slide freely in the trough and will
eventually cause trouble. This can allow the Cable Track to get in between the Rear
Longitudinal Carriage and the scanner frame on the foot end preventing the tripping of the limit
switch.
Slip Clutch
A slip clutch is part of the longitudinal motion system to limit torque. This is a feature to protect
the patient should he/she pinch an arm or leg between the back side of the scanner and the
Arm Column. If the Slip Clutch is set too loose, it will fail to move the belt and will just "slip" as
the motor turns. This may produce the following symptoms:
• failure of Quality Assurance scan Mechanics test
• compressed image in limited areas of the image
• a Diagnostic Failure #23-3
On Total Body scans, a limit switch out of position could allow the Source/Detector Carriages to
hit the frame or panels of the scanner before the Limit Switch.
If the mechanical stop is reached before the Limit Switch is actuated, check carefully to see
which part of the Arm Assembly is in contact with the Table Assembly. The front part of the
Lower Transverse Extrusion is clamped to the Longitudinal Drive Cable at the front of the
scanner. If the Lower Transverse Extrusion is not clamped in such a way that it forms a 90
degree angle with the length of the table, the rollers at the front end of the Lower Transverse
Extrusion may strike the end of the scan table before the Limit Switch is actuated.
Longitudinal Motor
Check the motor for a broken wire in one of the internal coils, or a bad electrical connection to
the Centent Motor Controller.
Longitudinal Centent
The longitudinal Centent Motor Controller may be the cause of a failure. The current set voltage
should be roughly 9-11 volts at terminal 11 on a properly operating controller. If this voltage is
not correct, replace the Centent Controller.
Longitudinal Belt
The Longitudinal Belt should not be tightened too much or this will cause the brackets holding
the gears to deform at either end of the scanner. When the belt is properly tightened, it should
be possible to deflect the upper and lower sides of the belt so that they touch within 8 cm of the
gears at either end.
Drive Wheels
The rollers in front and the wheels in back that support the Arm must be adjusted so that they
come into perfect contact with the Longitudinal Rails. Test them by preventing any wheel from
turning and see if the carriage will still move. By preventing any wheel from turning, it should be
possible to slide the carriage along the rail with one wheel dragging while the others roll. This
indicates that the wheel has not been excessively tightened down. Adjustments can be made
by loosening and rotating the eccentric bearings of any of the lower wheels.
If the transverse mechanics of the scanner become imprecise, it may cause a Limit Switch to be
tripped during a scan. This will usually only happen during a total body scan or during an Air
Matrix test since both of these scans involve full-width scanning. Once the switch is closed, the
motors stop, the End-of-Exposure Alarm sounds, the amber SHUTTER OPEN lamp turns off
and after a few seconds a Diagnostic Failure #15-4 appears on the screen.
During full-width scanning, the Detector Carriage comes very close to the Limit Switches, so any
imprecision will cause a Limit Switch to be tripped and the scan will be aborted with a Diagnostic
Failure message for Transverse Motion Failure.
A cause of this problem is a loosening of the first drive Reduction Belt which connects the
Transverse Motor to the first Reduction Pulley. This loosening causes the belt to "walk" on the
pulleys causing enough imprecision in the motion to trip a switch. To tighten the belt, first
remove the Pulley Shroud and loosen all four nuts that hold the motor in place. Then, while
holding the motor such that the belt is pulled taught, tighten the nuts to secure the motor in
place. Replace the shroud and test the scanner (an Air Matrix test works well).
If the Tube Head cables come into contact with the cable bundle entering the cable track, it may
be impossible for the scanner to complete all of the necessary transverse steps away from the
operator. Consequently, on each scan line the detector will move closer to the front transverse
limit switch, and the switch may be eventually closed. Form the cable bundle exiting from the
cable track into an arch such that the Tube Head cables will move under the arch rather than
running into the bundle. This arch must not be too high. If it is, there will not be sufficient slack
in the Tube Head cable bundle to allow the Tube Head to move to the front transverse limit
switch. Also, if the arch is too small, it may cause too much slack when the Tube Head is at the
forward side of the table. This causes the Tube Head cable bundle to rub against the inside of
the front panel causing a scraping noise to be heard. If the above does not solve the problem
then check the other mechanical components. Electrical components that could cause such a
failure are the Centent Motor Controller, the Transverse Motor, the OINK board or the SBC
board.
These lines are caused by Automatic Gain Stabilizer (AGS) trying to adjust the input signal
which is at an excessively high count rate. However, the AGS is not at fault, and the correct
action is for the customer to attenuate the x-ray beam. For an x-ray beam generated at 3 mA
the tissue areas need to be the equivalent of 12 cm. If the patient does not have 12 cm of
tissue, the operator must make up the difference with rice. Lunar normally recommends placing
a rice bag (on its side to create 12 cm of rice in the x-ray beam) along the side of the patient's
thigh for thin or osteoporotic patients.
In the case the last test fails, perhaps the voltage at AGSDCA-T4 needs to be increased 10 or
20 mV. Check the rollover at AGS-T12 with an oscilloscope and an x-ray beam at 150 µA to
verify that there is no roll-over or roll-under when this voltage is increased. If stability can not be
maintained in air at 150 µA and in 5-18 cm of water at 3 mA then it will be necessary to replace
some parts to determine which item is causing the instability. The following items can cause
this stability problem : detector, AGS board, both DCA boards, and the AMP-03 board.
If the signal changes suddenly at AGS-TP4, the cause must be the circuitry that controls the
high voltage to the Scintillation Detector, the Scintillation Detector itself, the circuitry on the
AMP-03, or the coaxial cables carrying the signals up the Scanner Arm. Measure the signal at
AGS-TP4, while taking voltage measurements at AMP-TP1 (the high voltage to the Scintillation
Detector) and SBC-TP20 (the control voltage from the SBC to the Bertan PMT-10A).
If the AMP-TP1 voltage changes at the same time the AGS-TP4 amplitude changes, but the
SBC-TP20 signal does not change, then the Bertan PMT-10A power supply or the coaxial cable
from the power supply to the AMP-03 must be faulty.
If the SBC-TP20 voltage changes at the same time the AGS-TP4 amplitude changes (AMP-TP1
will also change), replace the SBC.
Shutter Failures
Certain problems can cause the shutter to close intermittently completely attenuating the x-rays.
The shutter-open lamp should turn off and the end of exposure alarm will sound. A "Shutter
Failure" error message should appear on the monitor screen. Scanning should cease at this
moment, but if it does continue, blue lines should be expected in the image. Always ask the
operator if a "ping" sound (the end of exposure alarm) is heard, and whether the Shutter Open
or X-Ray On lamp turns off. In the case of a shutter problem, the Shutter Open light will turn off.
Unstable AC Line
If the customer's AC line conditions are unstable, voltages below 95 VAC will begin to drop the
high voltage on the x-ray tube. The counts will drop significantly with only a change of a few kV.
However, this normally causes black or dark grey lines in the scan image.
Arcing
If the tube head is arcing, the arcs will be more likely to occur at the higher current settings.
Arcing generally also causes other strange symptoms such as:
• Scanner stops in the middle of a patient or QA scan with no error message or with
Diagnostic failure 15-4 or 31-6 message.
• Scanner stops during a QA scan with the error message "Print Buffer Error"
• The QA scan fails intermittently because the "Mean value deviates from expected" or
because there is "Excessive variation in standard values"
• The scanner periodically leaves the message on the screen "Starting X-rays Please
Wait" for a longer than normal period of time (more than 10 seconds)
1. View the Quality Assurance History (see section 8.15). In particular note the trend in
BM Values as discussed in the Failing QA's section of this chapter.
3. Replacement of the tube head if no arc tracks are discovered, or if the repacking of the
connectors does not eliminate all arcing symptoms.
Summary
Except for the x-ray relay, the cause of the problem will be difficult to find. Try to obtain some
additional information. If the customer is hearing the end-of-exposure alarm, or seeing the
yellow x-ray on lamp flickering, you can deduce the tube head control cable is bad. Likewise, it
is possible to decide on the shutter/collimator/fan cable. In the case of no alarm or lamp
indications, or in the case where large deviations occur in the peak, the upper and lower cable
bundles should be replaced, as problems with the coaxial cable will be too expensive to
troubleshoot.
Check the size of the trochanter, neck and Ward's areas and verify that the sizes are
reasonable. The software may not have correctly selected the baseline causing the bone edges
to be improperly fixed. This false baseline could be due to one or more of the following
problems:
• There are not enough lines below the pelvis and above the trochanter (>25 lines) for an
adequate baseline to be established. The Operator's manual recommends starting the
scan 5.1 cm below the trochanter, and verifying that there are 15-20 lines before the
ischium appears in the image.
• If the x-ray beam passes into air during the femur scan, the software will incorrectly set
the baseline. White lines on the edge of the image, or unreasonable bone areas are
indications that the beam has passed into air. If this happens, the patient must be
rescanned with a soft tissue equivalent material (such as bages of uncooked rice) placed
against their hip.
• New systems will be delivered with LUNAR Tissue Bags (p/n 0784). These bags are to
be used as described above. Extra bags may be ordered through the LUNAR Customer
Service Department.
• Be sure the customer places the ROI box near the pelvis end of the neck with nearly
equal amounts of soft tissue in each end of the box and none of the ischium or
trochanter. The software moves the box down the neck towards the trochanter during
the search for the Ward's Triangle area.
To insure proper baselines, Lunar recommends that tissue equivalent material (as mentioned
above) be placed along side of and on top of the hips of very thin or osteoporotic patients.
In the case of a patient having unusually low bone mass, the program may have difficulty in
locating bone edges as it scans. In this case the only option is to slow down to Slow or even
Detail mode, and the automatic width option may have to be changed to manual if the bone
edges are still not tracked properly.
If the white lines that occur in the image when the beam passes into air continue to be white
when the beam passes back into the patient's body, the Automatic Gain Stabilizer (AGS) may
not be properly adjusted. There is a test for proper AGS operation (see ASG Calibration Test).
If manual analysis is used, the ROI boxes must be inside bone image only to obtain a correct
BMD. Include no soft tissue points in manual ROI's.
Category Limit
Chi Squared 45
Skew ±.07
Kurtosis ±.07
NOTE: All six categories MUST PASS during testing for new installations!
Chi Squared values ≤45 that fail should be considered in the light that the Air Matrix test is a
statistical evaluation, and therefore, there is a statistical chance that a properly functioning
scanner could fail the Air Matrix test. Skew and Kurtosis values will pass between -0.07 and
0.07. Anything outside of these limits is failing. These limits are valid only for scans made with
1/4" (6 mm) of aluminum in the beam. Limits for scans made with more or less aluminum have
not been determined.
If the failing Air Matrix has values that are close to passing (as described above), check the
Deviation Image (see below). If that looks good, run another Air Matrix (perhaps a shortened
test depending on time limits). The second test should pass. If not, a problem exists in the
scanner.
A misadjusted AGS DCA could be the cause of a failing Air Matrix. To determine if this is the
cause before parts substitution is used to solve the problem, refer to the AGS Adjustment
Procedure in Section 6.2. NOTE: This test and adjustment procedure will not be of use if the
PMT Detector is defective. Verify that the Detector is good by examining QA History trends in
%Spillover, Air Ratio and Air Counts (see Section 8.14).
Otherwise, the following parts should be substituted until passing Air Matrix results can be
obtained:
• Scintillation Detector
• Automatic Gain Stabilizer Board (AGS)
• AGS DCA
• DPX DCA
Deviation Image
The Deviation Image is a graphic illustration of the counts recorded during the Air Matrix scan.
A good image has a black field with evenly distributed grey dots and perhaps a few randomly
distributed dots of various colors (green or blue usually).
A poor Deviation Image will have colored lines, streaks or spots (indicating counts higher or
lower than expected). Colored lines or streaks going longitudinally in the image indicate an
alignment problem. Recheck the alignment of the scanner including the levelness of the
scanner. Also, check the scanner frame for any bending that may have occurred at the site or
during shipment.
Colored lines or streaks going transversely across the image indicate a problem with the AGS,
the AGS DCA or the DPX DCA (also see section 6.2). Again, replace parts until a passing Air
Matrix is obtained.
The amber LED is controlled by a comparator on the OINK Board. This LED illuminates when
approximately 40µA (or greater) of current is passing through the X-ray Insert. On Spellman
system scanners the amber light may appear to fade out slowly when x-rays are ramped down.
This due to the Spellman power supplies bleeding off current and is not a problem.
If the voltage ramping fails, the High Voltage Power Supplies must be shut completely off before
a second ramping attempt is made. If the amber X-RAY ON LED turns on momentarily, then
turns off for a few seconds before coming on steadily, the system is having difficulty ramping the
high voltage. One of the High Voltage Power Supplies may be defective.
The SBC does not monitor the Insert current once the proper level has been reached.
Therefore, if the current to the Insert is interrupted, the SBC will not recognize the problem and
will not alert the computer, or terminate the exposure.
If the X-RAY ON LED blinks once the x-ray high voltage has been set, there must be a problem
in one of the following areas:
• Tube Head Control Cable—The most likely cause of an intermittent problem in this
circuit is a broken wire in the Tube Head Control Cable. The Red wire is most likely the
broken one, as a broken blue or black wire should blow the MAX Board Fuse. Turn the
scanner off and test for continuity between MAX Board test points TP4, TP5, and TP13.
• MAX Board—The MAX Board may be operating intermittently. Verify that the wave
forms on TP4, and TP5 match the wave forms on page 9.17 and 9.18. Substitute a
replacement MAX Board.
• OINK Board—Noise may be present on the line extending from pin 3 of U5 to the SBC
board. This noise would cause the bell to ring after every scan line during Total Body
scans at 150A current modes. Adding a 0.1F capacitor between pins 1 and 3 of U5
on the OINK Board should attenuate this noise (see section 7, ECO #1393).
If the Shutter is actually opening and closing intermittently and uncontrollably, the problem is
either on the OINK, or a broken wire between the OINK and the Shutter/Collimator Assembly.
Remove connector J11 from the OINK Board and measure the resistance between pin 5 and
pin 6. This will be a measurement of the resistance of the Shutter Solenoid and the cable
through the Cable Track. If the problem occurs on every scan, run a scan while making this
continuity check. If the cable and solenoid seem to be working properly, substitute a new OINK
Board.
If no Diagnostic Failure Code is reported, the scanner may still be in motion continuing with the
scan. Abort the scan and remove the patient from the table. In any case, note the status of the
SHUTTER OPEN light and the X-RAY ON light.
• Both the SHUTTER OPEN light and the X-RAY ON light are on and steady. This would
indicate a faulty OINK board. Replace it.
• The SHUTTER OPEN light is off. This could indicate the following:
• The LED becomes defective during the scan and since the shutter solenoid and this
light are wired in series, the Shutter closed and the Alarm sounded.
• The cable to the amber Shutter Open light broke during the scan with the same
result as above.
• The shutter solenoid failed and the Shutter closed followed by the Alarm sounding .
• The OINK is faulty.
• The SBC is faulty.
• The X-RAY ON light is off. This could indicate the following:
• X-ray production has halted. This turns off the Xray On light and sounds the Alarm.
Check the I/O cable connections at the serial port on the back of the computer and at the SBC.
Be sure both connections are tight and that the thumb screws are used to hold the connectors
firmly together. Also, inspect the 25 pin connector on the SBC. This connector is very fragile
and may have been damaged. Be very careful when connecting the I/O cable to the SBC.
If the ports are properly configured and communication still fails, then the probable causes are:
• A faulty I/O cable.
• A faulty SBC.
• A faulty system board async (serial) port.
Replacement of these parts is the best troubleshooting method. The system board async port is
a part of the computer's system board.
Another possibility is that the SBC has lost its firmware or the firmware has become corrupted.
This can be remedied by downloading the program again. See Monitor Directory, section 5.1,
for instructions.
DOS Installation
CONFIG.SYS
rem LUNAR Corporation, 313 West Beltline Hwy, Madison, WI 53713. (608) 274-2663
rem File: config.sys DOS version: 6.20
files = 50
buffers = 45
stacks = 0,0
dos = high,umb
shell = C:\command.com C:\ /P /e:900
device = C:\dos\himem.sys
device = C:\dos\emm386.exe 64 ram
AUTOEXEC.BAT
PATH=C:\;C:\DOS
@rem LUNAR Corporation, 313 West Beltline Hwy, Madison, WI 53713. (608) 274-2663
@rem File: autoexec.bat DOS version: 6.20
@echo off
break off
prompt $p$g
C:\dos\smartdrv.exe 512
call C:\LUNAR\setdpx
:--- (check disk integrity and)
cls
echo RUNNING CHKDSK ON DRIVES
for %%f in (C:) do if EXIST %%f\nul chkdsk %%f /F
:--- (Create a temporary timestamp file for todays date)
echo. | date > C:\defrag.tmp
:--- (Check to see if the official timestamp file exists)
if not exist C:\defrag.dat goto CheckDate
:--- (A timestamp file exists, check if this is a new day by comparing files)
fc C:\defrag.tmp C:\defrag.dat | find "no differences encountered" > \nul
CONFIG.SYS
rem LUNAR Corporation, 313 West Beltline Hwy, Madison, WI 53713. (608) 274-2663
rem File: config.sys DOS version: 7.0
switches = /f
[menu]
menuitem=dos,Start MS-DOS
menuitem=win,Start Windows '95
menuitem=dpx,Start Lunar DPX
menudefault=dpx,8
[common]
files = 50
buffers = 45
device=C:\windows\himem.sys /testmem:off
dos=high,umb
shell=C:\windows\command.com /p
[dos]
device=C:\windows\emm386.exe ram highscan noems
device=C:\cdrom\nec_ide.sys /d:mscd001
[dpx]
[win]
AUTOEXEC.BAT
PATH=C:\;C:\WINDOWS;C:\WINDOWS\COMMAND;C:\LUNAR
@rem LUNAR Corporation, 313 West Beltline Hwy, Madison, WI 53713. (608) 274-2663
@rem File: autoexec.bat DOS version: 7.0
@echo off
break off
prompt $p$g
goto %config%
:dpx
:dos
lh C:\windows\smartdrv 512
call C:\LUNAR\setdpx
lh C:\mouse\mouse.exe /Q
lh C:\windows\command\doskey
if %config%==dos goto end
C:
cd \LUNAR
call dpx
goto end
:win
call C:\LUNAR\setdpx
win
:end
echo.
C:
cd \
MSDOS.SYS
[Paths]
WinDir=C:\WINDOWS
WinBootDir=C:\WINDOWS
HostWinBootDrv=C
[Options]
Logo=0
BootGUI=0
BootMulti=0
;
;The following lines are required for compatibility with other programs.
;Do not remove them (MSDOS.SYS needs to be >1024 bytes).
;xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxa
;xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxb
;xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxc
;xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxd
;xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxe
;xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxf
;xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxg
;xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxh
;xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxi
;xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxj
;xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxk
;xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxl
;xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxm
;xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxn
;xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxo
;xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxp
;xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxq
;xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxr
;xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxs
To use these features press F4 at the Main Menu to select Database Utilities and then press F2
to select QA History.
The Quality Assurance History screen displays three columns of data results that are stored in
the database. The 11 highlighted categories are selected by default. Using the arrow keys and
F4, F5 and F6 keys, the various categories can be selected or deselected depending on what
needs to be examined. Since only one category can be graphed at a time, simply move the
highlight to the item to be graphed and press F3.
Pressing F1, F2 or F3 selects View, Print or Graph and the next option is to enter the starting
and ending dates for the time period to be examined. Once that is done to view the data, press
[Esc] to start the printout or generate the graph.
When viewing the 11 default categories, or more, all the information will not fit on the screen.
Use the arrow keys to view the data off the screen to the right. All 11 default categories will fit
on one printed page. Printing more data will simply place the overflow on a separate page. In
graphing, it is possible to use the arrow, PgUp and PgDn keys to highlight a particular date's
QA. Note that the point on the graph for the highlighted date is also highlighted. The exact
value of the category is also displayed by the date.
LUNAR tries to analyze every system's QA History database file (DPXS.DBF) in order to
discover systems in need of service. This database file is converted to a Lotus 1-2-3
spreadsheet so that the data can be easily analyzed. LUNAR has determined, through these
analyses, acceptable limits for the variability observed for three important factors: Low Channel
Air Counts, Air Ratio and % Spillover.
The acceptable limit of change for Low Channel Air Counts can be expressed three ways: 1) no
more than 1% change/month, 2) no more than 10% change from the value at install or 3) no
change greater than 300 counts/QA.
The limits for Air Ratio are: 1) no change greater than 0.002/month, 2) no more than 0.02
change from the value at install or 3) no more than 0.0001 change/QA.
The limits for % Spillover are: 1) no change greater than 0.05/month or 2) no more than 0.002
change/QA.
The Peak setting normally drifts up or down. A long term upward trend is expected as the
Detector ages, but a downward trend is not unusual. The ideal operating peak is 500 units ±25.
However, Peaks between 400 and 600 units are acceptable. If the Peak is found to be further
from 500 than ±100 units, it should be adjusted (see section 6.1) until the Peak is 500 units
again (±25).
NOTE: The changes per month and per QA must be consistent over a three month
period or 63 QA's to be considered valid. Changes from time of install are invalid
if major parts were replaced since install, such as the Tube Head or PMT.
Changes should be compared to values after these replacements.
If any sudden jumps are noticed in the values of any of the categories, determine first if these
are due to service work such as a Tube Head replacement. If service work causes a shift in the
QA values, then compare present values to those obtained only after the service.
For complete analysis of the QA History database, Lunar suggests periodically sending a copy
of the DPXS.DBF file to the Lunar Customer Service Department.