PaxScan® 4336Wv4
Flat Panel Detector
Version 2
Product Description
The PaxScan® 4336Wv4 is a light weight, wireless flat panel
detector designed for digital radiographic systems. The
PaxScan 4336Wv4 fits standard 14”x17” bucky trays and its
wireless communication enables easy migration between table,
above the table, chest stand, and mobile cart applications. The
PaxScan 4336Wv4 works with commercially available Access
Points, or as a stand-alone Access Point. The receptor SDK
allows for direct integration into existing systems.
Technical Specifications Patient Contact
Receptor Type . . . . Amorphous Silicon with TFT/PIN diode Technology Surface Temperature . . . . . . . . . . . rated to not exceed 42 degrees C
Conversion Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CsI, DRZ+ Weight Limit
Uniform load across entire carbon fiber surface . . . . . . . . . . . 150 kg
Pixel Area Concentrated 40mm diameter load at the center of the imager . . 100 kg
Total . . . . . . . . . . . . . . . . . . . . . . 42.7 (v) x 34.4 (h) cm (16.8 x 13.5”) AED will not false trigger due to mechanical impacts.
Active (DRZ+) . . . . . . . . . . . . . . . 42.4 (v) x 34.1 (h) cm (16.7 x 13.4”) Software
Active (CsI) . . . . . . . . . . . . . . . . . 42.4 (v) x 33.9 (h) cm (16.6 x 13.3”) The PaxScan 4336Wv4 embeds the M-series Varex Imaging Smart Panel
(VSP) software within the receptor. Developers interface with the recep-
Pixel Matrix tor through VSP COMM which resides on the workstation. The integrator
Total . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3,072 (v) x 2,476 (h) experience is simplified through the new M-series software interface. An
Active (DRZ+) . . . . . . . . . . . . . . . . . . . . . . . . . . 3,052 (v) x 2,456 (h) onboard Control Panel is used to manage receptor settings and configu-
Active (CsI) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3,032 (v) x 2,436 (h) ration. The ViVA™ sample imaging application is included. VSP COMM is
Pixel Pitch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 139 µm Windows® 7 (64 bit), Windows 8.1 (64-bit) and Window® 10 compatible.
Limiting Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.6 lp/mm
Automatic Exposure Detection (AED) . . . . . . . . . . . . . . . via vTrigger Computer Requirements
RAM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.00 GB
Main Functionalities CPU . . . . . . . . . . . . . . . . . 1 GHz or faster processor (32-bit or 64-bit)
Cycle Time @ 550ms . . . . . . . . . . . . . . . . . . . . . . 5.6 sec (MSR2, RCT) Power
(X-ray Window) Power Consumption . . . . . . . . . . . . . . . . . . . . . . . . . . Idle - 4.3 watts
X-ray window. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 350-3500 ms Acquisition - 8.1 watts
Image Transfer - 10.0 watts
Image Quality GADOX (typical) CSI (typical)
DQE @ 0 lp/mm 39% 78% Battery
DQE @ 1 lp/mm 28% 58% Lithium polymer smart battery prevents over charging
DQE @ 2 lp/mm 18% 42% Charge capability . . . . . . . . . . . . . . . . . . . . . . 1600 images over 8 hrs
DQE @ 3 lp/mm 8% 24% Expected Life . . . . . . . . . . . . . . . . . . . 500 cycles of charge/discharge
DQE @ Nyquist 4% 14% Battery Charge . . . . . . . . . . . . . . . . . . . . . 12 hours in standby mode
MTF @ 1 lp/mm 56% 57% Wireless
MTF @ 2 lp/mm 24% 28% Wireless Modes STA or AP . . . . . . . . . . . . . 802.11 a/g/n/ac, 2x2 MIMO
MTF @ 3 lp/mm 12% 16% Minimum Signal Strength Required . . . . . . . . . . . . . . . . . . >-80 dBm
MTF @ Nyquist 7% 11% or no image will be acquired
Sensitivity 0.6 LSB/nGy 0.86 LSB/nGy Mechanical
Pixel Noise (1000ms) 9.2 LSB 8.7 LSB Weight (values are typical) (includes battery)
Memory Effect 0.001 (@ 60sec) 0.004 (@ 60sec) DRZ+ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.9 kg ± 0.25 kg
CsI . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.0 kg ± 0.25 kg
Dose Range DRZ+ CsI Housing Material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Aluminum
Maximim Linear Dose 84 µGy 58 µGy Sensor Protection Material . . . . . . . . . . . . . . . . . . . Carbon fiber plate
NED 0.56 µGy 0.36 µGy
Environmental
Energy Range Standard . . . . . . . . . . . . . . . . . . . . . . . 40 - 150 kVp Shock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . High-shock tolerance
Fill Factor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60% Water Resistant . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IP54
Scan Method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Progressive Temperature Range - Operating (at back cover) . . . . 10°C to 35°C (max.)
Data Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Wireless (Ambient) - Storage . . . . . . . . -20°C to +70°C
A/D Conversion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16-bit Humidity - Operating & Storage (non-condensing) . . . . . . 10% to 90%
Exposure Control Inputs: . . . . . . . . . . . . Prepare, Expose-Request Atmospheric Pressure - Operating & Storage . . . . . 70 kPa to 106 kPa
Outputs: . . . . . . . . . . . . . . . . . . . . . . Expose-OK Regulatory
U.S. . . . . . . . . . . . . . . . . . . . . . . . . . . . . ANSI/AAMI ES 60601-1:2012
Canada . . . . . . . . . . . . . . . . . . . . . . . CAN/CSA C22.2 No. 60601-1:14
®PaxScan is a Registered Trademark of Varex Imaging Corporation
EU . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IEC/EN 60601-1:2012
Copyright © 2019, Varex Imaging Corporation. All Rights Reserved.
139064-000 Rev A 01/19
1
®
PaxScan® 4336Wv4
Flat Panel Detector
Dimensions are for reference only Version 2
Dimensions are in mm
Copyright © 2019, Varex Imaging Corporation. All Rights Reserved.
2
®
PaxScan® 4336Wv4
Flat Panel Detector
Dimensions are for reference only Version 2
Weight - 0.77 lbs (.35 kg) (nominal)
Copyright © 2019, Varex Imaging Corporation. All Rights Reserved.
3
®
PaxScan® 4336Wv4
Flat Panel Detector
Battery Charger (Optional) Version 2
Weight - 1.33 kg (nominal)
Single Bay Charger (Optional)
Weight - 0.3 kg (nominal)
®
Manufactured by Varex Imaging Corporation
Specifications subject to change without notice.
Salt Lake City, UT 1-801-972-5000
Note: All Varex Imaging Amorphous Silicon Receptors are
For a complete listing of our global offices, designed to be integrated into a complete X-ray system by a
visit www.vareximaging.com qualified system integrator. The system Integrator is responsi-
ble for obtaining FDA clearance for medical use.
Copyright © 2019, Varex Imaging Corporation. All Rights Reserved.