Chapter 4: Characterization Techniques For Nanomaterials
Chapter 4: Characterization Techniques For Nanomaterials
FOR NANOMATERIALS
90
CHAPTER 4
NANOMATERIALS
4.1 Introduction
Nanomaterials behave differently as the size changes with respect to the bulk.
a) Chemical characterization
b) Structural characterization
done.
i) Optical Spectroscopy:
b) Photoluminescence (PL)
91
c) Fourier Transform Infrared Spectroscopy (FTIR)
d) Raman Spectroscopy
about the shape, lattice constants and crystallinity are important. To know the
92
c) Small Angle X-ray Scattering (SAXS)
phonons) are usually in the IR and are studied either using IR absorption or
93
(i) Optical Absorption Spectroscopy
UV-visible spectroscopy
into one of the empty anti-bonding orbitals. In each possible case, an electron
is excited from a full orbital into an empty anti-bonding orbital. Each jump takes
energy from the light, and a big jump needs more energy than a small one.
Figure 4.1 showing the possible electron jumps that light may cause.
theory.htmlj
Each wavelength of light has a particular energy associated with it. If that
particular amount of energy is just right for making one of these energy jumps,
94
then that wavelength will be absorbed its energy will have been used in
promoting an electron.
The relationship between the frequency of light absorbed and its energy
E=hv (4.1)
Where, E is the energy of each quanta of light, h is the Planck’s constant and v
is the frequency of light For higher energy jump, light of a higher frequency is
to be absorbed.
wavelength of light passing through the spectrometer, the intensity of the light
.The intensity of the light, I passing through the sample cell is also measured
for that wavelength. If l is less than l0l then obviously the sample has absorbed
The Beer-Lambert Law gives, the relationship between A (the absorbance) and
light at that wavelength has been absorbed that means that the intensity is 10%
95
of what it would otherwise be. A spectrophotometer can be either single beam
or double beam.
kaist ac kr/vt/chem-ed./spec/uv-vis/uv-vls.htm],
The basic parts of a spectrophotometer are a light source, a holder for the
tungsten filament (300-2500 nm), a deuterium arc lamp, continuous over the
ultraviolet region (190-400 nm), xenon arc lamps, continuous from 160-2,000
nm; or more recently, light emitting diodes (LED) for the visible wavelengths [4],
photomultiplier tubes are used with scanning monochromators, which filter the
light so that only light of a single wavelength reaches the detector at one time.
96
arrays. As both of these devices consist of many detectors grouped into one or
two dimensional arrays, they are able to collect light of different wavelengths on
In a single beam spectrophotometer the light passes through the sample cell. I0
light is split into two beams before it reaches the sample. One beam is used as
the reference; the other beam passes through the sample. The reference beam
measurement displayed is the ratio of the two beam intensities. Some double
beam instruments have two detectors (photodiodes), and the sample and
reference beam are measured at the same time shown in figure 4.3. In other
instruments, the two beams pass through a beam chopper, which blocks one
beam at a time. The detector alternates between measuring the sample beam
and the reference beam in synchronism with the chopper. There may also be
one or more dark intervals in the chopper cycle. In this case the measured
http://www.fiie8.chem.vt.edu/chem-ed/spec/uv-vis/dualbeam.html]
97
The Tauc’s plot is used to determine optical bandgap. Tauc plot shows the
quantity hv (the photon energy) on the abscissa and the quantity (ahv)r on the
ordinate, where a is the absorption coefficient of the material [5]. The value of
the exponent r denotes the nature of the transition; for example, r = 14 for
indirect transitions [5], The resulting plot has a distinct linear regime which
denotes the onset of absorption. Thus, extrapolating this linear region to the
abscissa yields the energy of the optical bandgap of the material. However, if
the material does not have a single phase, it will likely not have a single distinct
Tauc’s plot.
photon.
If a light particle (photon) has an energy greater than the band gap energy,
then it can be absorbed and thereby raise an electron from the valence band
up to the conduction band across the forbidden energy gap. In this process of
photoexcitation, the electron generally has excess energy which it loses before
coming to rest at the lowest energy in the conduction band. At this point the
electron energy eventually falls back down to the valence band. As this
98
happens, the energy it loses is converted back into a luminescent photon which
is emitted from the material. Thus the energy of the emitted photon is a direct
measure of the band gap energy, Eg. The process of photon excitation followed
about 10ns, and no significant transition of the internal energy of the chemical
also fast, but some of the original energy is dissip'ated so that the emitted light
a triplet state. When the energy is trapped in the triplet state, transition back to
the lower singlet energy states is forbidden quantum mechanically. This leads
to a slow process of radiative transition back to singlet state that last from
rarer than fluorescence, since a molecule in the triplet state has a good chance
can occur.
99
A spectrometer is an instrument used for measuring the intensity of light as a
prism) to disperse the light, thereby spreading out the light of differing
wavelengths into different positions. The spectrometer unit has an internal CCD
to measure the light intensity at various positions along its length. From the
Band gap is the energy difference between the lowest state in the conduction
and the highest state in the valence bands, in semiconductors. The spectral
determine the electronic band gap and this provides a means to quantify the
often reveals spectral peaks associated with impurities contained within the
100
c) Surface structure and excited states
d) Recombination mechanisms
radiative rates are typically associated with impurities and the amount of
the valence band to conduction band occurs and, the electrons losses excess
the conduction band. The electrons may radioactively recombine with holes of
the valence band and if the sample is completely free of impurities i.e. pure an
exciton gets formed between these two carriers with a small binding energy.
The characteristic of the energy levels is the energy of the emitted photons due
transitions [9].
If the sample is impure or doped, radiative recombination also may occur via
transition may occur, conduction band to acceptor level, donor level to valence
Energy Dispersive Spectrometry (EDS) was first introduced in the late 1960s,
when solid state detectors were first interfaced to microanalyzers [10]. The
spectroscopy (EDS). EDS systems are typically integrated into either an SEM
nitrogen is use for cooling, and software to collect and analyze energy spectra.
The detector is mounted in the sample chamber of the main instrument at the
end of a long arm, which is itself cooled by liquid nitrogen.ln all EDS all photons
state X-ray detector. The common EDS detector is a lithium- drifted silicon,
Si(Li) [1]. The detectors made of Si(Li) crystals that operate at low voltages to
so-called "silicon drift detectors" that operate at higher count rates without liquid
An EDS detector contains a crystal that absorbs the energy of incoming X-rays
by ionization, yielding free electrons in the crystal that become conductive and
produce an electrical charge bias. The X-ray absorption thus converts the
102
Figure 4.4 Schematic of Energy Dispersive Spectroscopy [http://serc. carleton. edu/
research_education/.../eds.html]
in sample occurs and these are the characteristics of the elements present and
fi^lrl Thic* mot il+o ir» r% Iapo r\f anorrn/ r\f +he» ortH ^p/arpw or>r\<aoro no
MV_IW. i i m o iooui to III u iu jo ui V^I 1^1 yy ui u i v_, JUUll ui i ui iu u icu v^i iui yy U|j[jUUIO uo
^Antoine nhnlnnc of oil Dnornioc till tho onomu of tho orioinol olootron oe on
Wl IIWII l*J |JI IWWI lw» VI 1411 W I ui y • W \J till II IU WIIWI^jT Ml UIV VI l^ll IMI VlUVll WI I , UJ Mil
electron can lose any energy, from zero to the energy of primary electrons. The
smoothly varying photon intensity. From the X-ray lines the atom can be
EDS has certain limitations, the energy peak overlaps among different
elements; in this case, the user can apply deconvolution methods to try peak
103
separation, or simply consider which elements make "most sense" given the
known context of the sample [11]. EDS cannot detect the lightest elements,
X-ray Diffraction (XRD) is one of the classical methods for identification and
figure 4.5.The X-ray source, sample and the detector are placed in a particular
The source is stationary and the sample and the detector are mobile.When the
sample moves by an angle 0, the detector moves by angle 20. The rotation rate
is kept at 1°/min and the sample is scanned for 10°-80° scan. The sample is
104
Figure 4.5 X-ray Diffractometer[http://epswww.unm.edu/xrd/xrdclass/01-XRD-lntro.pdf]
XRD-lntro.pdf]
atomic basis. The sharpness and shape of the spots are related to the
(i) The presence or absence of a certain set of planes gives us the crystal
structures.
(ii) The shift of the peaks from its original positions in case of bulk crystals
20(Degrees)
including ceramics, metals, insulators, organic, polymers, thin film powders etc.
X-ray diffractometers can be used either for single crystals or for powders.
While single crystal diffractometers are used for the study of molecular
structure, powder diffractometers are used for analysis of phases, though the
Two approaches are generally used for the analysis of X-ray diffraction data.
crystal may be treated in the same way as the diffraction by an optical grating.
106
Upon projection, the grating is like an array of points similar to a crystal. The
periodic in all three directions and three independent laue equations can be
occur [1],
several lattice points. When X-rays are incident on a crystal, different lattice
reflected beams are in phase when the path length between the beams is an
will be in phase only when this condition is satisfied. This means that distance
ABC=nA or 2dSin0=nA. But for all other angles other than © destructive
2dsin0 = nA (4.3)
Where d is the separation between the planes of the crystal,0 is the angle of
For all angle other than 0, destructive interference will occur leading to
107
is usually complete. However, in cases where number of diffraction planes is
the diffraction peak positions of the XRD [2]. The shift in the peak positions is
due to homogeneous strain that leads to change of lattice constants and the
that increases with increase in Sin0. Peak broadening is also caused by the
the peak width can be determined. The crystallite size D, can be estimated
K?>
D= (4.4)
PCos8
108
Where p is the Full Width at Half maximum in radians, 9 is the peak position, D
the particles is obtained by using the formula [5] between strain and particle
size,given by,
pcos© 1 t|Sm8
ADA, = “ + —— (4.5)
' #
Where p is the Full Width at Half maximum in radians, 0 is the peak position D
is the average grain size, A is the wavelength of incident X-ray, n is the effective
strain.The strain is determine from the slope of the plot between pcos0 / A and
sin0/A .whereas average particle size is obtained from the inverse of the y-axis
determining the lattice constant and structure and labeling each peak with its
The structure factor calculations give the relationship for allowed reflections for
109
4.3.2(b) Scanning Electron Microscope (SEM)
The first Scanning Electron Microscope was debuted in 1942 and commercially
the range of 0.5 nm [1], SEM images have a characteristic three dimensional
In SEM, a monochromatic electron beam with a very fine spot size of ~ 5nm
and having energy from a few hundred eV to 50 KeV is passed over the
surface of the specimen which induces various changes in the sample. The
resulting particles from the sample are used to create an image of the
The schematic of SEM is shown in the figure 4.9. SEM consists of an electron
gun at the top, the "Virtual source” that produces a stream of monochromatic
controlled by the “coarse probe current knob”). This lens is used to form beam
and also to limit the amount of current in the beam. The condenser aperture
along with the first condenser lens works to eliminate the high-angle electrons
from the beam. The second condenser lens forms electrons into a thin, tight,
coherent beam and is usually controlled by the “fine probe current knob”. The
objective aperture further eliminates high-angle electrons from the beam. A set
no
of coils “scan “ or sweep the beam in a grid like fashion, dwelling on points for a
period of time determined by the scan speed. The last objective lens, focuses
the scanning beam onto the part of the specimen desired. The beam of
electrons strikes the sample and interactions occur inside the sample and are
number of interactions. Then beam moves to the next dwell point. This process
is repeated until grid scan is finished and then repeated, the entire pattern cab
semoptic.html]
excitation beam, of the order of a few nanometer [1], The electron beam
causes various excitations in the sample that are the characteristic of the
in
as a result of core hole decay is used for elemental identification .The intensity
time consuming and cumbersome than EDS but improved energy resolution is
The first TEM was built by Max Knoll and Ernst Ruska with resolving power
greater than that of light, in 1933 and the first commercial TEM in 1939.TEM is
300k is possible in all TEM. Latest TEM has magnification of 50 million times.
basic components of TEM microscope are (i) an electron gun, that emits a
112
Figure 4.10 Schematics of TEM [www.indepthinfo.com/.../electron-microscope]
potential gradient placed across the filament (field emission, fine-tipped single -
optical axis of the microscope are focused to a small , thin, coherent beam by
knocking out high angle electrons.The beam then strikes the specimen and
out high-angle diffracted electrons, whereas the selected area aperture enables
atoms in the sample. The image is then passed down the column through the
intermediate and projector lenses. The image strikes the phosphor image
113
screen and light is generated, allowing the user to see the image.The darker
areas of the image represent those areas of the sample that fewer electrons
were transmitted through .The lighter areas of the image represent those areas
There are two image modes of TEM, Bright field mode and Dark field mode.ln
Bright field (BF) image mode an aperature is placed in the back focal plane of
the objective lens that allows only the direct beam to pass through [15].The
image results from weakening of the direct beam by its interaction with the
image formation. Thick areas, that are areas in which heavy atoms are
enriched, and crystalline area appear dark contrast. The image is interpreted by
field image mode, the direct beam is blocked by aperature while one or more
diffracted beams are allowed to pass the objective aperature. As the diffracted
beams strongly interacted with the specimen, useful informations about planar
In the TEM the electrons pass through the sample as it is thin.The electron
that pass through the sample is classified into three categories: the unscattered
inciding with the sample without any interaction with the sample atoms and
sample.So, thicker parts of the sample appear darker while the thinner part of
the sample appear lighter [17].The incident electrons that are deflected due to
114
their interaction with the sample but do not loss any energy , gives elastically
interaction and are scattered according to Bragg’s law .The scattered electrons
When the incident electrons lose energy due to their interaction with the sample
sample and this confirms the composition and also provide the bonding
HRTEM is useful for direct atomic level study like interface, dislocation, defects,
TEM.
image plane of the electron wave with itself [18].The contrast formation in high
(thickness < 20 nm), in most cases a crystal. During transmission the incident
115
electron wave is scattered (or diffracted in the case of a crystal) at the
potentials of the atoms, and thereby the phase of the-electron wave is changed.
At the exit surface of the specimen the object wave is formed, which carries
direct and highly resolved information on the object. The object wave is
magnified in the electron microscope and during this process the wave suffers
additional phase shifts due to imperfect lenses (aberrations). Finally, the image
image wave, which itself and it contains essentially phase contrast with all the
microscopic aberrations included. As the phase of the electron wave carries the
information about the sample and generates contrast in the image, and so
of electron intensities only the phase of the wave and hence an important
the microscope which is the optical resolution of the objective lens [19].
Each imaging electron interacts independently with the sample. Above the
positive atomic potentials of the atom cores, and channels along the atom
columns of the crystallographic lattice (s-state model). At the same time, the
116
interaction between the electron wave in different atom columns leads to Bragg
diffraction.As a result of the interaction with the sample, the electron exit wave
distances from the optical axis). The phase change of cpe (x,u) compared to the
incident wave peaks at the location of the atom columns. The exit wave now
further phase change and interferes as the image wave in the imaging plane
(photo plate or CCD).It is important to realize that the recorded image is NOT a
high intensity might or might not indicate the presence of an atom column in
that precise location. The relationship between the exit wave and the image
on the phase of the exit wave <pe(x,u) and propagates it to the image wave.
approximation (WPOA) holds (thin sample) the contrast transfer function (CTF)
becomes
CTF(u)=A(u)E(u)sm(z(u)) (4.7)
117
Where A(u) is the aperture function, E(u) describes the attenuation of the wave
for higher spatial frequency u, also called envelope function. ^ (u) is a function
The last, sinusoidal term of the CTF will determine the sign with which
components of frequency u will enter contrast in the final image. If one takes
rotationally symmetric about the optical axis of the microscope and thus only
Z(u)=jCsA3U4-My%u2 (4.8)
and Af is the defocus. In TEM, defocus can easily be controlled and measured
to high precision. Thus one can easily alter the shape of the CTF by defocusing
The aperture function cuts off beams scattered above a certain critical angle
(given by the objective pole piece for ex), thus effectively limiting the attainable
resolution. However it is the envelope function E(u) which usually dampens the
attainable with a microscope and is known as the information limit. E(u) can be
118
E(ii)=EXu)EXu)Ed{u)Ev(u)ED{u) (4.9)
coherency and defines Es(u) and the chromatical aberration, together with
current and voltage instabilities that define the temporal coherency in Ec(u).
One of the demerit with HRTEM is that image formation relies on phase-
which often can be estimated from the Fourier transform of the HRTEM image.
Some of the techniques described above are adopted for characterising the as-
synthesised samples and those results obtained are presented in the next
chapter.
119
REFERENCE
No. 2(2009)
4. http://en.wikipedia.org/wiki/ultraviolet-visible-spectroscopy
7. http://en.wikipedia.org/wiki/photoluminescece
Available at:http://cnx.org/content/col10699
Nanotechnology(PH\ 2009)
14. http://www.semitracks.com/index.php/reference-material/eds.html
16. http://www.microscopy.ethz.ch/TEM_BF.htm
17. Chin Wee Shong.Sow Chorng Haur, Andrew TS Wee, Science At the
18. http://en.wikipedia.org/wiki/Flighresolution_transmission_electron_microscopy
19. http://www.elmi.uni-bonn.de/en/sub/forschung/hrtem_e.html
121