GlucoLab Auto-Coding - MeterManual

The document is an instruction manual for the GLUCOLAB Auto-coding Blood Glucose Monitoring System. It contains 13 sections that provide information about intended use, kit contents, operating instructions, maintenance, troubleshooting, specifications and warranty for the blood glucose meter and test strips. The manual emphasizes that the system is for self-testing blood glucose levels outside the body and should only be used with fresh capillary whole blood samples under a healthcare provider's guidance.

Uploaded by

Mario Barra
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
597 views22 pages

GlucoLab Auto-Coding - MeterManual

The document is an instruction manual for the GLUCOLAB Auto-coding Blood Glucose Monitoring System. It contains 13 sections that provide information about intended use, kit contents, operating instructions, maintenance, troubleshooting, specifications and warranty for the blood glucose meter and test strips. The manual emphasizes that the system is for self-testing blood glucose levels outside the body and should only be used with fresh capillary whole blood samples under a healthcare provider's guidance.

Uploaded by

Mario Barra
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 22

Lancet : Sae Han Med Corp.

973-16, Sungsuk-Dong, Ilsandong-Gu,


Goyang-City,
Gyeonggi-Do, Korea

Lancing Device : GMMC


Room No. 1112, Ace Tower 9th Bldg.,
345-30, Kasan-Dong, Kumchon-Ku,
Seoul, Korea

CUSTOMER CARE LINE: 01376 500 720


http://neondiagnostics.co.uk/

INFM22AGB

IMM20-520 Rev. 2013-05-06


Dear GLUCOLABTM Auto-coding Blood Glucose
Monitoring System Owner Test Principle

Thank you for choosing the GLUCOLABTM Auto-coding Blood Glucose The enzyme glucose oxidase on the blood glucose test strip reads specifically
Monitoring System. This booklet has important information you need to know with the blood glucose. The current generated is converted and displayed as
about the GLUCOLABTM Auto-coding Blood Glucose Monitoring System. blood glucose value.
Please read it carefully. The GLUCOLABTM Auto-coding Blood Glucose Monitoring System is plasma-
calibrated to allow easy comparison of results with laboratory methods.
We understand that self-testing of blood glucose level provides a way to control
your diabetes and may give you peace of mind by testing regularly. As a result A blood glucose meter which is calibrated against a whole blood method may
have different results in comparison to GLUCOLABTM Auto-coding.
GLUCOLABTM Auto-coding was developed to provide you with a fast and
The laboratory system used for calibration of the GLUCOLABTM Auto-coding
accurate reading with a convenient and simple process. Our goal at GLUCO-
Blood Glucose Monitoring System is YSI 2300 STAT plus which is equipped
LABTM Auto-coding is to provide the best quality healthcare products coupled with a glucose oxidase system.
with superior customer service. Always consult with your healthcare profes-
sional before making any changes to your diabetes management. The GLU-
Note
COLABTM Auto-coding Blood Glucose Monitoring System is for in vitro diag-
< Maintaining the GLUCOLABTM Auto-coding blood glucose test meter >
nostic use only.
1. Do not assemble or disassemble the blood glucose test meter.
2. Please keep in a dry place.
Important information 3. Please keep clean and dry.
4. Please do not store in the same place as toxic chemicals or gases.
The GLUCOLABTM Auto-coding Blood Glucose Monitoring System is intended
5. Keep out of direct sunlight.
for use outside the body (in vitro diagnostic use only). It should be used only for
testing blood glucose and only with fresh capillary whole blood samples. It
should not be used for the diagnosis of diabetes.
CAUTION
Consult your physician or diabetic healthcare professional about daily man-
agement of your diabetes and proper use of the glucometer. 1. If the code of the blood glucose test meter and of the blood glucose
test strip vial does not match, you must set the code number.
Please pay close attention when handling blood. Improper procedures may 2. If the code on the LCD window and on the blood glucose test vial
cause serious hazards to your health. do not match, follow the coding procedure for accurate results.
The GLUCOLABTM Auto-coding Blood Glucose Monitoring System contains
small parts. Please keep your monitoring system out of reach of children.

4 www.infopia21.com 5
Intended use of GLUCOLABTM Auto-coding Blood
Contents Glucose Test Meter

1. Learning the System The GLUCOLAB Auto-codingTM Blood Glucose Monitoring System is intended
Intended use of Blood Glucose Test Meter 7 for in vitro diagnostic use only. GLUCOLAB Auto-codingTM Blood Glucose
GLUCOLABTM Auto-coding Blood Glucose Monitoring System Kit Contents 8 Monitoring System is for self-testing.
GLUCOLABTM Auto-coding Blood Glucose Test Meter 9 It should not be used for the diagnosis of diabetes or for the testing of newborns.
GLUCOLABTM Auto-coding Blood Glucose Test Meter Display 10
The GLUCOLABTM Auto-coding was designed to test a blood glucose (sugar)
GLUCOLABTM Auto-coding Blood Glucose Test Strip Information 11-12
level and can be used with a capillary whole blood sample.
2. Before Testing Do not use the GLUCOLABTM Auto-coding Blood Glucose Monitoring System for
Setting the blood glucose test meter 13-14 any purpose other than blood glucose testing.
Auto-Coding Function 15
Performing a Glucose control solution test 16-17
3. Testing Your Blood Glucose
Testing Your Blood-Preparation 18
Collecting a Drop of Blood 19-21
CAUTION : Before using any product, please read all instructions.
Testing Procedure 22-24
Consult your physician or diabetic healthcare professional about daily
Blood Glucose Test Strip Ejector Function 26
management of your diabetes and proper use of the glucometer.
4. Reviewing Your Results 27 If you have any questions about the GLUCOLABTM Auto-coding
5. Alternate Site Testing 28 product, please contact Neon Diagnostics Ltd. Freephone patient
6. Performing a Blood Test using and Alternate Site 29-30 helpline 0800 009 3378.
7. GLUCOLABTM Auto-coding Blood Glucose Test Meter Software 31
8. Taking Care of Your Blood Glucose Test Meter
Cleaning Your Blood Glucose Test Meter and Maintenance 32
Dispose of meter and batteries 33
9. Troubleshooting 34-37
10. Specifications 38
11. Conversion table mmol/L mg/dL 39
12. Symbol reference 40
13. Warranty 41

6 www.infopia21.com 7
GLUCOLABTM Auto-coding Blood Glucose Monitoring
System KIT Contents GLUCOLABTM Auto-coding Blood Glucose Test Meter
Blood glucose test strip Port:
Insert the GLUCOLABTM Auto-
coding blood glucose test strip
here.
Blood glucose test strip ejector

Display Screen:
Displays your test results, sym-
bols and messages.

Function Button:
Powers the meter on/off, prompts
memory data and acts as an
enter button during setup and
after tests.

Up/Down Arrow Buttons:


Adjusts setting, scrolls through
BLOOD GLUCOSE TEST METER
For use with Blood Glucose Test Strip only
memory
TM
1. GLUCOLAB Auto-coding Blood 6. Warranty Registration Card INFM22A

Serial Number
GL-5C13-1C0-03 Rev.2009-04

Glucose Test Meter 7. Patient Logbook


2. GLUCOLABTM Auto-coding Blood 8. Carrying Case
BLOOD GLUCOSE TEST METER

Glucose Test Strips 9. Two 3V Li-CR2032 Batteries Battery Cover


3. Lancing Device 10. Control Solution Normal Bd. General Wahis 53
1030 Brussels, Belgium GL-5C13-1C1-03 Rev.2009-04

4. Lancets
5. Operation Manual
CAUTION
1. Do not use GLUCOLABTM Auto-coding blood glucose test meter in a dry
Your GLUCOLABTM Auto-coding Blood Glucose Monitoring System has environment, especially if synthetic materials are present. Synthetic
clothes, carpets, etc., may cause damaging static discharges in a dry envi-
been sealed and should contain all of the parts shown above. Please ronment.
return the system to the place of purchase or contact your nearest GLU-
2. Do not use GLUCOLABTM Auto-coding blood glucose test meter near cellu-
COLABTM Auto-coding representative if the container seals are broken or lar or cordless telephones, walkie talkies, garage door openers, radio
parts are missing. transmitters, or other electrical or electrical equipment that are sources of
electromagnetic radiation, as these may interfere with the proper operation
of the blood glucose test meter.
8 www.infopia21.com 9
GLUCOLABTM Auto-coding Blood Glucose GLUCOLABTM Auto-coding Blood Glucose
Test Meter Display Test Strip Information

Storage and Handling Caution


Low Battery Warning Symbol 1. Keep out of direct sunlight.
Strip Code Symbol 2. Store the GLUCOLABTM Auto-coding Blood Glucose Test Strip vials in a cool,
dry place between 2-30°C (36-86? ). Do not freeze.
Blood drop symbol for test 3. Make a note of the discard date on the vial label when you first open it.
Discard remaining GLUCOLABTM Auto-coding blood glucose test strips 3
Alarm Symbol months after first opening the vial.
4. Close the vial cap tightly immediately after removing a GLUCOLABTM Auto-
Before Having a Meal
coding blood glucose test strip.
After Having a Meal 5. Do not use test strips after the expiration date printed on the package or vial
as this may cause inaccurate results.
After Sports Activity 6. Avoid getting dirt, food or water on the blood glucose test strip. Do not bend,
cut, or alter the GLUCOLABTM Auto-codingblood glucose test strip.
After Taking Medication
7. GLUCOLABTM Auto-coding blood glucose test strips are for single use
Temperature / Date only.
8. Keep the blood glucose test strip vial away from children.
Time / Test Result Amount 9. Please refer to the user’s manual for GLUCOLABTM Auto-coding blood glu-
cose test strip for additional information.
Test Result Unit Symbol
10. Discard the used blood glucose test strip carefully, to prevent any infection.
Control Solution Symbol
AVR Average Test Result System Use only GLUCOLABTM Auto-coding brand blood glucose test strips with your
GLUCOLABTM Auto-coding blood glucose test meter.
CHK Troubleshooting Symbol
? Apply Blood Target
? Contact Bar
? Confirmation Window

10 www.infopia21.com 11
GLUCOLABTM Auto-coding Blood Glucose
Test Strip Information Setting Your Blood Glucose Test Meter

Limitations of Blood glucose measuring system : The GLUCOLABTM Auto-coding Blood Glucose Test Meter has a vari-
The blood glucose test strips provide accurate results when the following conditions ety of excellent functional modes; beep sound to alert when function is
are observed:
set, date, time, five daily alarms, ability to designate activities related to
- Use the GLUCOLABTM Auto-coding blood glucose test strips with the GLUCOLABTM the test result, and the ability to designate three unique data averages.
Auto-coding blood glucose test meter or GLUCOLABTM blood glucose test meter.
- Use fresh capillary whole blood only.
- Do not use neonate samples.
- The blood glucose test strips are for single use only. Do not reuse.
- Dehydration may lower test results.
- Inaccurate results may occur when in shock, hypotensive, hyperglycemic, or hyper-
osmolar state, with or without ketosis.
- Blood glucose test strips used above altitudes of 10,000 feet will have an effect on
test results.

Doctors - Please note the following interferences that may affect test results : Power On User Activity Option Year
- GLUCOLABTM Auto-coding test strip results can be used with hematocrit levels in the Press for 3 seconds After pressing and releas-
range of 20% to 60%. After pressing and
at least. ing either ▲ or ▼ buttons
If the hematocrit range is out of the range(20~60%), then the test results may be to turn the user Activity releasing either ▲ or ▼
lower or higher than actual value. option on/off, confirm with buttons to set the year
- Interferences : Acetaminophen, Uric and, Ascorbic acid (Vitamin C), and other reduc- . (from 2007 to 2099),
ing substances when occurring in normal blood or normal therapeutic concentrations confirm with .
do not significantly affect results. However, abnormally high concentrations may
cause inaccurately high results.
- Lipemic samples; Cholesterol up to 12.93mmol/L or triglyceride up to 33.9mmol/L do
not significantly affect the results. Values beyond these levels should be interpreted
with caution.
- Blood samples that contain a high concentration of dissolved oxygen may lower the
test result.

Date / Time Temperature unit Number of days for


Performance Characteristics : Press and release either average
The performance of the blood glucose test strips has been evaluated in laboratory and After pressing and
▲ or ▼ buttons to set ℉ After pressing and releas-
in clinical tests. releasing either ▲ or ▼ ing either ▲ or ▼ buttons
or ℃. Enter button.
buttons to set the date to set the number of days
Measurement Range : The measurement range of the GLUCOLABTM Auto-coding and time, confirm each for average calculation (3
Blood glucose measuring system is 0.5 to 33.3 mmol/L. different possible), confirm
with . with .

12 www.infopia21.com 13
Setting Your Blood Glucose Test Meter Auto-Coding Function

1. GLUCOLABTM Auto-coding has automatic


code recognition function.

Alarm on/off Alarm time Alarm


After pressing and releas- After programming the
After pressing and releas-
ing either ? or ? buttons number of desired alarms
ing either ? or ? buttons
to set the alarm hour and (5 alarms possible), con-
to turn the alarm on/off,
minute, confirm with . firm with .
confirm with .

2. It recognizes automatically the code num-


ber. This is a very convenient function pre-
venting the inconvenience of setting the
code number on the blood glucose test
meter every time.

CAUTION
1. **VERY IMPORTANT: Correct test unit.
Your GLUCOLABTM Auto-coding blood glucose test meter is pre-set as mmol/L as test unit. An incorrect CAUTION
test unit will lead to an incorrect test result display. Please consult your healthcare professional if you’re 1. Do not bend the blood glucose test strip to prevent the Automatic code recognition
not sure which test unit to use. Stored test results are displayed correctly after re-setting the test unit. See
also conversion table on page 39.
failure.
2. Without setting the date properly, the average glucose level and the results in memory will not show 2. If the code recognition label is damaged, the code recognition may fail. Please
proper values. It is recommended to set the blood glucose test meter before use and when new batteries check the code number on the LCD window with the code number on the blood
are installed. glucose test strip vial.
3. It is not possible to test your blood glucose in the setting mode.
4. To turn the meter off during setup, press for at least 5 seconds.

14 www.infopia21.com 15
Performing a Glucose Control Solution Test Performing a Glucose Control Solution Test

The glucose control solution should be used when there is suspicion that the blood glu- Note: - GLUCOLABTM Auto-coding glucose control solution is sold separately.
cose test meter or the blood glucose test strips are not working properly. For example: if
strips have been exposed to extreme conditions or if you drop the blood glucose test - It is recommended that glucose control solution is stored at room temper-
ature 20~25℃(68~77℉) before testing.
meter.
Check the expiration date before performing a glucose control solution test. For example
Do not used if expired. Once opened, solution expires after 3 months. If you use a normal glucose control solution, you must use the normal range printed on
the side of the blood glucose test strip vial.
The Glucose control solution should be used
•Whenever you suspect the blood glucose test meter or blood glucose
test strip is not functioning properly.
•If your blood glucose test results are not consistent with your symptoms or if
you think they are not accurate.
•If you have dropped the blood glucose test meter.
•For quality control in the point of care usage. 1.9-3.6
•For teaching or learning the system. 4.4-6.7
14.2-19.2
1. Shake the glucose control solution before use.
2. Discard the 1st drop of glucose control solution, this will eliminate any residue.
3. Place a drop of glucose control solution on a clean dry surface (ex. The lid of the blood
glucose test strip vial).
• Results that fall outside the expected range may indicate:
4. Insert a blood glucose test strip •Test not performed at room temperature.
into the port and press the up •An error in the way you are doing the test.
arrow. A glucose control solution •Expired or contaminated glucose control solution.
vial will appear on the bottom left
•A problem with the blood glucose test meter.
hand side of the screen, above
•Glucose control solution that is too warm or cold.
the date and time. This will allow
•Failure to discard the first drop of glucose control solution and to wipe the bottle
you to differentiate between a glu-
cose control solution test and an actual blood test for future reference. tip clean.
•Failure to shake the glucose control solution adequately.
5. Dip the blood glucose test strip into the glucose control solution.
6. Results appear in 5 seconds.
CAUTION
7. Compare the result to the range printed on the blood glucose test strip vial, the results • If your glucose control solution falls out of range, do not perform a glucose test, please
should fall within that range. contact your local representative.
8. If your glucose control solution falls out of range, do not perform a glucose test, please • The Glucose Control Solution range is not the recommended range for your blood
contact your local representative. glucose level.

16 www.infopia21.com 17
Testing Your Blood - Preparation Collecting a Drop of Blood

Before testing, be sure to read this section and blood glucose test strip package insert The lancing device is a convenient medical tool for collecting capillary blood samples
found in the blood glucose test strip box carefully, Make sure you have all the necessary for glucose monitoring or other tests which require one or two drops of blood. As a
more advanced lancing device, it provides ultimate safety and comfort for obtaining
items to begin testing: blood samples. An adjustable tip offers 5 levels of skin penetration for individual
user’s comfort and a lancet ejector enables the safe disposal of the used lancet. As
an added bonus it also has a transparent tip for AST (Alternative Site Testing).

• GLUCOLABTM Auto-codingblood PRECAUTIONS


•Never use a lancet that has been used by someone else. This could lead to con-
glucose test meter
tamination.
• GLUCOLABTM Auto-codingblood
•If the lancing device is to be used by another person, the unit must be properly dis-
glucose test strip
infected and a new tip and lancet must be used.
• Lancing Device
•Do not leave the lancet in the device after use.
• Sterile Lancets •A new lancet must only be placed into the lancing device directly before testing.
•This device has many small parts and could be a choking hazard for children if
swallowed.
•In the case of hospital use, hospitals need to consult their own infection control pro-
tocols in order to avoid any contamination.
•Always dispose of the used lancet in a biohazard garbage container.

COMPONENTS

Clear Tip

Ejection System Sliding Barrel


CAUTION
To reduce risk of infection :
1. The lancing device and sterile lancets should NOT be shared with others.
2. Always use a new, sterile lancet and a new blood glucose test strip.
3. Practice using the Lancing Device and become accustomed with its use.
4. Lancets and blood glucose test strips are for single use only. Release Button
5. Wash your hands with warm, clean water and soap before testing. Depth Indicator/Adjustable Tip
6. Avoid getting hand lotion, oils, dirt or debris on the lancets or on the lancing Lancet Holder
device.
7. Dry your hands completely before testing.

18 www.infopia21.com 19
Collecting a Drop of Blood Collecting a Drop of Blood

RECOMMENDED AVAILABLE LANCETS


Some general sterilized lancets are BD-Ultra Fine, Cleanlet Fine, E-Z Ject, G-P Lite, Microlet,
Monolet, One Touch, Soft Touch, Therasense, Ultra TLC and Unilet GP.
Please note: Some lesser known lancets may not work properly due to their dimensions. Please
contact your representative to make sure that you are using proper lancets.

1. Unscrew the tip of the lanc- 2. With two fingers, pull out 3. While holding the lancet USING THE CLEAR TIP (AST)
ing device by turning it count- the lancet carrier and hold it carrier, insert a new sterile
er-clockwise while holding in place. lancet into the bottom of 1. Place the clear tip in the skin at the chosen alternative site, press and hold the continuously for a
the base firmly. lancet carrier. few seconds, then push the release button to take a sample.
2. Watch through the clear tip until a sufficient blood sample is taken. If there is not enough blood,
gently massage the area until a sufficient sample has been collected.

STORAGE
Products must be stored at room temperature protected from sunlight or moisture.

CLEANING AND DISINFECTION


4. Twist the cap of the lancet 5. Put the tip of the lancing 6. Hold the tip firmly in one Clean and wipe the outside of the lancing device a minimum of once a week with a soft cloth,
off. device back on and turn it hand then pull out the slid-
clockwise. ing barrel with the other dampened with mild soap and water. Wash the adjustable tip and the clear tip once a week with
hand. This will load the mild soap. Do not immerse in liquid.
lancing device.

Click WARRANTY
The lancing device has a 2-year warranty from the date of purchase. A malfunctioning device
should be returned to the following address:

GMMC
Room No. 1112, Ace Tower 9th Bldg., 345-30, Kasan-Dong, Kumchon-Ku, Seoul, South Korea
7. Place the lancing device in 8. Unscrew the tip of the 9. Push the lancet ejector for- Tel : +82 2 2104 0470
place. Hold the firmly against lancing device by turning it ward with the thumb and
the side of the finger, with the counter-clockwise. simultaneously pull out the Fax : +82 2 2104 0472
cap resting on the finger. sliding barrel to dispose of GMMC, S.L.
(The harder it is pressed the the used lancet in a proper
deeper the puncture.) Press biohazard container. C/Jordi de S. Jordi 13, 6-23 46022 Valencia SPAIN
the release button to take a Tel : +34 963 691 484
sample.
Fax: +34 961 125 949
Discard your Blood glucose test Lancet
U.K. Distributor: Neon Diagnostics Ltd.
Untis 7&8, Swanbridge Industrial Park, Black Croft Road, Witham, Essex, CM8 3YN
CUSTOMER CARE LINE: 01376 500 720
www.neondiagnostics.co.uk

Push the needle into its protective cover. pull the lancet out and discard appropriately.

20 www.infopia21.com 21
Testing Procedure Testing Procedure

Correct Applying Blood

Completely Filled Correct

Step 1 Incorrect
•Firmly insert the blood glucose test strip into the blood glucose test
meter port with the “infopia” facing towards you.
•The blood glucose test meter automatically turns on displaying the code, Incorrect
ambient temperature and the date and time. Poorly filled
•The GLUCOLABTM Auto-coding will automatically recognize the blood
•GLUCOLABTM Auto-coding blood glucose test meter only requires
glucose test strip code number and adjust the blood glucose test meter 1.0㎕㎕ sample volume.
accordingly. •Apply your blood to the glucose test strip and do not take off your finger
•A blinking blood glucose test strip will appear at the top of the screen until you get the beep sound.The test will automatically start.
indicating the blood glucose test meter is ready for testing.
Note : Inserting the blood glucose test strip in the wrong direction will turn the Step 2
blood glucose test meter on, but you will be unable to test. •Place the top edge of the blood glucose test strip to the area you prepared for testing.
Caution •The blood will automatically be “drawn” into the blood glucose test strip channel.
1. If the code is not displayed after for a while, pull the blood glucose test strip out •When the chamber has enough blood, your blood glucose test meter will automatically
of the port, and re-start the procedure from the beginning.
count down.
2. If the code on the LCD display window and on the test vial does not match, try
another new blood glucose test strip. If the mismatch persists, please contact
your local representative for help. Caution
3. If the blood glucose test meter does not power on, pull the blood glucose test 1. If the countdown does not start, do not add more blood to the blood glucose test
strip out of the port, and re-insert the blood glucose test strip. strip! Discard the blood glucose test strip an restart testing.
4. Avoid testing under direct sunlight, for a more accurate test result. 2. If you do not conduct the test within 3 minutes, the blood glucose test meter will auto-
5. If you apply your blood sample too early, ERROR5 message will appear on matically power off to save battery life. In this case the test procedure should start
again from the beginning.
LCD ( refer to p35). 3. You may get an inaccurate result if the blood sample is not completely filled in a sin-
• For a more accurate test result avoid testing under direct sunlight. gle action.

22 www.infopia21.com 23
Testing Your Blood - Case of Setting User
Testing Procedure Option

Step 4
•If you selected to use the User Activity option dur-
ing set up, press ▲ and ▼ button to select the
activity that correlates with your result (refer to
p10), then press .

mmol/L

Step 3
•After the beeping sound, the test will begin automatically and your results will
appear in 5 seconds. It should begin counting down from 5 to 1 second on
the LCD display window. The LCD window will display the result of your
blood glucose level, temperature and time.
•Record the result value in your logbook. When blood glucose test strip is
removed, the blood glucose test meter turns off automatically.

CAUTION
1. If the test result is out of the test range, the Hi/Lo message will be shown on
the LCD. (refer to p35)
2. Safely discard used blood glucose test strip and lancet, to a proper place,
according to the local regulation.

24 www.infopia21.com 25
Test Strip Ejector Function Reviewing Your Result

The GLUCOLABTM Auto-coding blood glucose test meter stores up to 500 test
results in its built-in memory, along with the average sugar level for the number
of days which you had preset the blood glucose test meter to calculate. These
values can be recalled in the LCD display at any time.
•Check your previous result by pressing the arrow ▼ button.
•Check your average data by pressing the arrow ▲ button.

Press and Push forward Eject your blood glucose test strip

The third latest The second latest The latest


1. Once the test is complete, slide the ejector button forward to remove the blood glu-
cose test strip from the blood glucose test meter.
2. Discard the used blood glucose test strip and lancet accordingly.

7day average 14 day average 21day average


from the test date from the test date from the test date

Average test result


CAUTION Test Unit
1. If you push the ejector button too far forward, it may malfunction. Test result amount
2. Strong impacts may result in a malfunction of the blood glucose
meter. Average test result days

26 www.infopia21.com 27
Alternate Site Testing Performing a Blood Test Using and Alternate Site

Performing a Blood Test Using an Alternate Site To


Upper Arm ensure accurate results when lancing your forearm,
Forearm upper arm, hand, thigh, or calf, wash the test site with
Dorsal Hand Ventral Palm soap and water.

Fingertips Make sure there is no cream or lotion on the test


Thigh
site.Thoroughly dry your hands and test site.
Calf

Important Information About Using Alternative Sites Testing :


• Alternate sites where you can test are dorsal hand, ventral palm, upper
arm, forearm, calf, and thigh.
• Under certain conditions, blood glucose test results obtained using sam-
ples taken from your alternate sites may differ significantly from fingertip
samples
• The conditions in which these differences are more likely to occur are
when your blood glucose is changing rapidly such as following a meal, as
insulin dose, or associated with physical exercise.
• When blood glucose is changing rapidly, fingertip samples show these
changes more quickly than alternate sites samples.
• When your blood glucose is falling, testing with a fingertip sample may
identify a hypoglycemic (low blood glucose) level sooner than a test with a
alternate sites sample.
• Use alternate sites samples only for testing prior to or more than two hours
after meal, insulin doses, or physical exercise.
• Testing performed within two hours after meals, insulin doses, or physical
exercise or whenever you feel that your glucose levels may be changing
rapidly should be done from the fingertip.
• You should also use fingertip testing whenever you have a concern about
hypoglycemia (insulin reactions) such as when driving a car, particularly if
you suffer from hypoglycemic unawareness (lack of symptom to indicate
as insulin reaction), as forearm testing may fail to detect hypoglycemia. CAUTION
Important :
•To receive accurate test results, at least 1.0㎕ minimum sample volume
We recommend that you test on your fingertips if you are testing for hypoglycemia is required.
(low blood glucose) or if you are suffering from hypoglycemia unawareness.

28 www.infopia21.com 29
Performing a Blood Test Using and Alternate Site GLUCOLABTM Auto-coding Blood Glucose
Test Meter Software

To bring fresh blood to the surface of the test site, rub the test site vigor- You can transfer test results from the GLUCOLAB Auto-coding
TM

ously for a few seconds until you feel it getting warm. blood glucose test meter to a computer.
You can download the software from our web site.
(www.neondiagnostics.co.uk)
Lance the test site with the lancing device:
You can purchase the computer communication cable separately from
1. Hold the clear cap down against the top of your a local representative.
test site. Press the release button. Do not lift up.
[User] [PC]
2. Continue to hold the lancing device and gradually
increase pressure for several seconds.

3. While holding the lancing device on your test site,


look through the clear cap; until a drop of blood
appears approximate • size

4. Lift the lancing device straight up; be careful not


to smear the blood on your testing site.

5. Place the edge of the blood glucose test strip to


the drop of blood. cable
6. The blood will automatically draw into the chan-
nel of the blood glucose test strip.

You may get an inaccurate result if the blood


sample is not completely filled in a single
action.

30 www.infopia21.com 31
Cleaning Your Meter and Maintenance Replacing the Batteries

Blood glucose test meter : The low battery icon will appear in the upper left corner of the LCD screen to alert you
Your blood glucose test meter does not require special maintenance or clean- when the battery power is running low, indicating a new battery is needed.
ing. avoid getting dirt, dust, blood, glucose control solution, or liquids on the
blood glucose test meter, the test port, or data port. Your blood glucose test
meter operation temperature is 10~40℃(50~104℉).
It is recommended that you store the blood glucose test meter in its carrying
case after each use.
A cloth dampened with water and mild detergent can be used to wipe down the
outside of the blood glucose test meter. Your GLUCOLABTM Auto-coding blood Your GLUCOLABTM Auto-coding blood glucose test meter uses only two 3V Lithium
glucose test meter is a precision instrument. Please handle it with care. batteries (CR2032), which are included. When replacing the batteries, only CR2032 or
equivalent lithium battery should be used.
Lancing Device:
After changing the batteries, please be sure to recheck your blood glucose test meter
Clean the lancet device and caps with soap and tepid water. To disinfect the settings.
lancet device, prepare a disinfectant solution of one part household bleach to 10 Dispose of meter and batteries
parts water. Dampen a cloth with this solution and wipe the lancet device thor- Dispose of meter:
oughly. Soak only the cap for at least 30 minutes in the disinfectant solution. Do The meter must be disposed of according to the local regulations concerning the dis-
posal of electrical and electronic equipment.
not soak the lancet device in liquid. Rinse the lance device and cap with water The Waste Electrical and Electronic Equipment(WEEE) regulation implement provi-
and dry thoroughly. sions of the European Parliament and Council Directive 2012/19/EU aimed to reducing
the amount of EEE waste going for final disposal.
•For detailed instructions concerning GLUCOLABTM Auto-coding blood glucose The manufacturer, has specific instructions for the recovery of the meter.
test strips, refer to the blood glucose test strip package insert found in the Please contact your distributor.
Dispose of batteries :
blood glucose test strip box. Please recycle or dispose of used batteries using your local battery collection systems
•Please store the GLUCOLABTM Auto-coding blood glucose measuring system and in compliance with your local environmental laws and regulations.
in a cool, dry area, out of reach of children. Do not freeze. For a more accu- Batteries contain chemicals that, if released, may affect the environment and human
health. The crossed-out wheeled-bin symbol indicates the need for the separate col-
rate test result avoid testing under direct sunlight. lection for batteries.
•Do not soak the blood glucose test meter orblood glucose test strips in water
<Follow-up Action>
or liquid. 1. Check your meter with the glucose control solution (p16-17)
•Do not subject the meter or blood glucose test strips to excessive heat. 2. Perform your test again (p22 - 24)
•Use your meter according to the instructions in the manual. 3. If you are experiencing symptoms that are not consistent with your blood glu-
•If you need to purchase the lancing device, lancets or blood glucose test cose test results, call your healthcare professional.
strips, contact your local representative.
CAUTION
•Discard the used Lancets carefully, to prevent any infection. Remove both batteries from the compartment and dispose of them
according your institution's guidelines.
32 www.infopia21.com 33
Troubleshooting Troubleshooting

The following chart may help you identify certain problems, but may not solve all
Message Problem Action Required
problems that can occur. Please contact your local representative or customer
User applied the blood Please wait for the LCD to
support if the problem persists.
sample before the display icon blinking before
LCD display applying your blood sample
was blinking

Message Problem Action Required Please insert a new test strip


Problem with the and perform your test again.
Place the battery again, and set Auto-coding label If Er6 message persists,
Problem with the the blood glucose test meter. If please contact your local rep-
blood glucose test the problem persists, please resentative. If blinks with
"SUN", avoid to the direct
meter contact your authorized repre- sunlight and retest.
sentative or customer support
The ambient tempera- Place the blood glucose test
Please insert a new blood glu- ture is too low meter within the operating
Caused by either temperature range for 10 min-
used or wet blood cose test strip and perform
your test again utes or more and retest.
glucose test strip

The ambient tem- Place the meter within the


Please insert a new test strip perature is too high operating temperature
Problem with the and apply the blood sample range for 10 minutes or
test strip until confirmation window is more and retest.
completely filled.
Check the accuracy of your
The test result is lower blood glucose test strip by
Blood glucose test Please insert a new blood glu- than 0.5 mmol/L performing a glucose control
strip is damaged cose test strip and perform solution test. If the test results
your test again in a normal reading, re-test
your blood sample two or
three times. If "Lo" persists,
consult your doctor immedi-
ately.
34 www.infopia21.com 35
Troubleshooting Troubleshooting

Message Problem Action Required Message Problem Action Required


The test result is higher than Check the accuracy of your Low battery Battery needs replacing.
33.3 mmol/L blood glucose test strip by
performing a glucose control For your free replacement
solution test. If the test results batteries, please contact
in a normal reading, re-test Neon Diagnostics Ltd.
your blood sample two or three Low battery symbol Freephone Patient Helpline:
times. If "Hi" persists, consult appears in the 0800 009 3378
your doctor immediately. upper left corner of
The blood glucose Battery is dead or there Change the battery and if the the LCD screen
test meter does not is a problem with the problem persist, contact your local
Low battery Low battery Battery needs replacing.
power on blood glucose test representative
warning symbol For your free replacement
meter (refer to the P.33)
and 3 digit code batteries, please contact
Test does not start Poor blood sample Please insert a new blood glu- Neon Diagnostics Ltd.
after applying blood cose test strip and perform your Freephone Patient Helpline:
sample test again 0800 009 3378

The result is incon- There may be a problem Please insert a new blood glu-
sistent with the blood glucose cose test strip and perform your
test strip test again

No readings in memo-
ry to display

Not enough readings in


memory to display des-
ignated averages

36 www.infopia21.com 37
Specifications Conversion table mmol/L mg/d/L

Sample type Capillary whole blood


mmol/L 0.55 1.0 1.5 2.0 2.2 2.5 2.8 3.0
Sample volume 1.0ul
mg/dL 10 18 27 36 40 45 50 54
Test Range 0.55 ~ 33.3 mmol/L
Reading time 5 seconds mmol/L 3.3 3.9 4.0 4.4 4.7 5.0 5.5 6.0 6.1

Calibration Plasma - Equivalent mg/dL 60 70 72 80 85 90 100 108 110

Altitude ≤ 3048m (10,000 feet ) mmol/L 6.7 7.0 7.2 7.5 7.8 8.0 8.3 8.9
Operating Temperature 10-40? (50-104? )
mg/dL 120 126 130 135 140 145 150 160
Operating Humidity 10-90%
mmol/L 9.0 9.4 10.0 10.5 11.0 11.1 12.0 12.5 13.9
Strip storage temperature 2-30? (35.6-86? )

Display Type mg/dL 162 170 180 190 196 200 216 225 250
LCD

Dimension 85 X 50.5 X 19.34 (mm) mmol/L 14.4 15.0 16.0 16.6 17.0 18.0 19.0 20.0

Weight 49g ( Including Battery) mg/dL 260 270 288 300 306 325 342 360
Power rating 3 V Li Battery (CR2032) X 2
mmol/L 20.8 22.2 23.0 24.0 25.0 26.4 27.7 30.0 33.3
Battery Life One year after purchasing
mg/dL 375 400 414 432 450 475 500 540 600
PC communications port Communication is available to connect
computer with cable

38 www.infopia21.com 39
Symbol Reference Warranty

Symbol Description Three - Year Warranty


Consult Instructions for use
If, at any time during the first three years after purchase the blood
Used By glucose test meter does not work for any reason with the exception
This product fulfills the requirements of Directive 98/79/EC of obvious abuse, misuse, or disastrous damage, NEON DIAGNOS-
on in vitro diagnostic medical devices
TICS LTD will replace your GLUCOLABTM Auto-coding
Attention, See Instructions for use cose test meter or equivalent free of charge.
For In Vitro Diagnostic Use
Batch code
The Warranty Policy of NEON DIAGNOSTICS LTD applies to the
original purchaser of the blood glucose test meter and include the
Catalogue Number
batteries supplied with the blood glucose meter set.
Temperature limitation
Please fill in the warranty card attached to the set.
Do not reuse
Mail or fax it back to us.
SN Serial number
Date of Manufacture The GLUCOLABTM Auto-coding blood glucose test meter has a full
Manufacturer
three-year warranty from the original date of your purchase. For your
convenience, please include your date and place of purchase.
Keep away from sunlight

Self-testing Used

Waste electrical and electronic equipment

Authorized representative in the European Community

Waste batteries and accumulators

Biohazard

40 www.infopia21.com 41
MEMO MEMO

42 www.infopia21.com 43