Circuit Breaker Analyzer TM1700 - UG - en PDF
Circuit Breaker Analyzer TM1700 - UG - en PDF
User’s Manual
WWW.MEGGER.COM
User’s Manual
Contents
1 Safety MANUAL.......................................................... 27
___________________________________ 6 4.3 Navigation & display buttons.............................. 28
1.1 General................................................................ 6 4.4 Main and submenus.......................................... 28
Symbols on the instrument................................. 6 Selected items bar............................................ 28
1.2 Safety instructions................................................ 6 4.5 Repair CABA Local............................................. 29
1.3 Protective earthing procedure in HV 4.6 Updating CABA Local........................................ 29
environments.................................................... 10 For instrument with built-in screen.................... 29
2 Introduction For instrument without screen.......................... 29
__________________________________ 12 4.7 Optional software.............................................. 30
2.1 General.............................................................. 12 Printer and Back-up media................................ 30
Benefits............................................................ 12 Antivirus........................................................... 30
1 Safety
tective ground wire with connection to the Neglecting these rules may result in fire, elec-
Protective Conductor Terminal on top of the tric shock, or other hazards.
instrument. Check the continuity of the pro- The Timing M/R section inputs are designed
tective ground wire before each use. Make to CAT I Environment – Secondary (signal
sure the connector is fastened properly to level) or battery operated circuits of electron-
the instrument Protective Conductor Termi- ic equipment. Not connected to mains.
nal. Make sure the connection point at the DO NOT use the Timing M/R inputs to meas-
ground system is fastened properly. Route ure on circuits in measurement category II, III
the wire so that it is unlikely to be walked on or IV.
or that it may loosen accidentally by some-
6] Heat. The instrument must be situated away
one or something moving near it.
from any heat sources such as radiators, heat
Note: The protective ground wire must not
registers, stoves, or other products that pro-
be loosened while any input connector is at-
duce heat.
tached to the contacts of a high voltage cir-
cuit breaker or another device being subject 7] Ventilation. Slots and openings in the in-
to inductive or capacitive coupled interfer- strument are provided for ventilation. They
ence from surrounding high voltage wires. ensure reliable operations of the instrument,
keeping it from overheating. These open-
5] Connections and disconnections to the ings must not be blocked nor covered during
inputs. All input or output connections and
operation. Do not install the instrument in a
disconnections must be made only while
closed space, such as a rack or cabinet, unless
the circuit breaker poles are connected to
proper ventilation is provided through fol-
ground on both sides. Always connect the in-
lowing the manufacturer’s installation proce-
strument to ground before making any other
dure. Blocking the ventilation slots prevents
connections to the instrument.
proper airflow and may result in fire, electric
The Timing M/R section inputs are designed
shock or equipment damage.
to handle coupled interference current from
surroundings by active interference suppres- 8] Water and moisture. Do not use the instru-
sion. The protection, though, is relaying on ment near water. To prevent fire or shock
a current path from the connected inputs to hazard, do not expose the instrument to rain
ground. or moisture.
The current passing through the wires may Do not touch the plug with wet hands. Doing
be harmful if there is no current path to so may result in electrical shock.
ground since, in that case, the high voltage 9] Accessories. Do not use any accessories not
may cause an arc. recommended by the instrument manufac-
Before connecting or disconnecting the turer as they may cause hazards.
instrument to a high voltage circuit breaker’s
contacts, make sure the circuit breaker is 10] EMC Warning. The instrument generates,
closed and connected to ground on both uses, and can generate radio frequency en-
sides. ergy. If not installed and used in accordance
The rules regarding connection to protec- with this manual it may cause interference
tive ground given above must be followed to radio communications. The instrument has
closely. been tested and found to comply with the
21] Servicing. Do not attempt to service the ing. Periodically clean the instrument with a
instrument yourself; opening or removing soft cloth. Stubborn stains may be removed
covers can expose you to dangerous voltage with a cloth lightly dampened with a mild
and other hazards. Please refer all servicing detergent solution.
to qualified service personnel as opening or Periodically clean the area around the mains
removing covers may result in fire, electric plug and at all other connectors.
shock or equipment damage.
If, for some reason, you need to return your
instrument, please use either the original
transport box or one of equivalent strength.
22] Damage requiring service. Unplug the instru-
ment from all connections and refer servicing
to qualified service personnel under the fol-
lowing conditions:
- When any connector is damaged, including
mains plug.
- If liquid has been spilled or objects have fallen
into the instrument.
- If the instrument has been exposed to rain or
moisture.
- If the instrument does not operate normally
(follow operating instructions).
- If the instrument has been dropped or dam-
aged in any way.
- When the instrument exhibits a distinct
change in performance. This indicates a need
for service.
If the instrument begins to emit smoke,
smells like something is burning or makes
strange noise, disconnect all power connec-
tions immediately and contact your dealer
for advice.
23] Replacement parts. When replacement parts
are required, be sure the service technician
has used replacement parts specified by the
manufacturer or that they have the same
characteristics as the original parts. Unau-
thorized substitutions may result in fire,
electric shock, or other hazards.
24] Safety check. Upon completion of any repairs
to the instrument, ask the service techni-
cian to perform safety checks to determine
that the unit is in safe operating condition.
Except for visual inspection and function
tests, dielectric strength test and Protective
Ground (Earth) & Bonding impedance tests
are required for safe operation after service.
Refer to manufacturer’s production test
instructions. Neglecting these measurements
after service may result in fire, electric shock,
or other hazards.
25] Cleaning. Unplug the instrument before
cleaning it. Do not use liquid cleaners or aero-
sol cleaners. Use only a damp cloth for clean-
Figure 1.3-1
2 Introduction
Benefits
▪▪Integrated software support with predefined breaker
templates and application oriented software for testing,
assessment and archiving of results. Results in faster
and easier breaker testing. Reduced need for operator
training, shorter connection and testing time.
▪▪Stand-alone functionality reduces the number of items
and boxes the tester needs to bring to the field.
2.4 Shipping
If you are going to send the instrument with parcel
service use the original transport box or one with
equivalent strength. Megger can not take responsibil-
ity for transport damages.
3 System overview
❶ POSITION
The TM1760 is equipped as a TM1750 plus three Indicates the position of the circuit breaker main
analog channels. contacts if the coil circuit is connected to the Control
section and “Auto detect” is enabled.
“O” indicates Open position
TM1760 with 3 extra analogue “I” indicates Closed position
channels
Note Position indicates per operating
mechanism(s). If the Circuit Breaker has
common operating mechanism only one
LED will indicate the CB position.
❷ SEQUENCE
Indicates the next operation of the circuit breaker. If
Auto-detect breaker state is enabled in CABA Local
or CABA Win, only possible sequences for the circuit
breaker are selectable.
❸ ON/OFF
Power on/off switch
❹ OPERATE/MEASURE
Initiates the selected operation sequence and makes
the measurement. Green “READY” LED must be lit
The TM1760 with three extra analog channels (totally before turning the rotary switch. The yellow “OP-
6 analog channels). ERATING” LED is blinking as long as the sequence is
performed.
❺ ❺ B Tree structure
C Main menu
❻ D Screen buttons
E ?
By pressing the “Question mark, “?”, button you
invoke the help function. There are to options to get
help:
A] Click “What’s this”, move the pointer to the
object in question and click. A PDF file will
❼ ❽ ❾ open and show the responding information.
B] Click “Manual” to open the User’s manual
❺ NAVIGATION BUTTONS and look for information in the PDF-file.
Most of the CABA Local functions are controlled by
the ten navigation buttons. The function of a naviga- ❼
tion button is stated on the display button on the
screen right next to it.
You can activate the function either by pressing the
navigation button or by clicking on the display button Touch screen ON/OFF
using an external mouse.
❽
Note A navigation button can be used when a
screen button is shown next to it. If no but-
tons are displayed the built in mouse is used
to navigate in the menus.
On screen keyboard ON/OFF
❻ SCREEN
❾
A B C
D E
❶ ❷ ❸ ❹ ❺ ❻
⓫
❼ 8 9 ❿
BL1448GE ZP-BL16E TM1700-series 19
3 System overview
Terminal configuration
The terminals are grouped in three similar parts A,B
and C suited for 3-phase operation (L1 L2 L3, RST).
As an option this section can be configured with three
extra analog channels.
Cable with female XLR contact needed, included in
the delivery of the instrument.
0 earth
IN input value from transducer Using the DCM1700 MODULE and the DualGround technol-
OUT output from unit to transducer ogy makes testing a lot safer.
❻ Digital
Digital channels are for measurement of motion, both
linear and rotary with incremental/digital transducer.
▪▪Number of channels 6
▪▪Measurement ranges transducer resolution up to +/-
32000 pulses.
DCM1700 Once the TUNE button is released the tuning will start
in opposite channel order, i.e. beginning with channel
C2 and ending with channel A1.
The reversed tuning order is an indication of that the
DCM1700 is in GIS mode.
▪▪The normal mode shall be used on open air circuit DCM capability
breakers.
Limitations
▪▪The GIS mode shall be used for GIS circuit breakers but it The technology utilised in DCM is applicable on all
can also be used for open air circuit breakers having one circuit breaker technologies. Still some limitations are
interrupter per phase. present for certain circuit breaker configurations.
Fig. 4.2-3
❿ TRIG IN
Fig. 4.2-2
Used for external trig of the unit. Starts an initiated
measuring sequence from a remote destination by
either a contact make or break or voltage signal. Set-
tings are made from CABA Local or CABA Win.
⓫ Mains input
Always use the power cord supplied with the unit.
⓬ Protective conductor terminal
Connection to earth (ground), see “1.2 Safety instruc-
tions”.
Network settings
Services
▪▪Client for Microsoft Networks
▪▪File and Print Sharing for Microsoft Networks
Protocols
▪▪TCP/IP Protocol
Settings for TCP/IP
▪▪Dynamic IP Address
▪▪Dynamic DNS Server Address
▪▪NetBIOS Enabled over TCP/IP
Printer
Software for printers and back-up media can be
installed without acceptance from Megger. Please see
the Microsoft Windows XP HCL (Hardware Compat-
ible List) for printers and back-up media that are
hardware compatible.)
See also section “Printer” in 4.14 System settings
menu”.
Backup media
As backup media you can use USB memory stick, PC
flash card, CD recorder or similar. USB memory stick
are used for software updates
4 CABA Local
Shut down
1] Press the ON/OFF button.
Note The OnLine Help is not available in the “Ana- 3] Click on desired topic.
lyzer Window”.
The keyboard button F1 does not invoke the Tip
OnLine Help function. You can also go to the manual index and
look for the desired topic.
WHAT´S THIS?
Using this option will automatically direct you to the
current topic in the OnLine Help.
1] Press the button “WHAT´S THIS?”.
2] Move the pointer to desired object and click
left mouse button.
The user manual will be opened.
3]
❶ ❶
❺
❸
❷
❷
Analyser View
Here you see the view of how the channels are used
in current test. You can also continuously view the
state of each channel, MONITOR mode.
Transducers
Here you define and calibrate your transducers.
System Settings
Here you do the system settings for the instrument
and CABA Local
4.9 Templates menu pulse and graphs settings as well as conversion tables
if applicable. Every setting that is needed for the test
but the selection of transducers.
There are three kinds of templates, the Max Config
template, Generic templates (delivered with the instru-
ment) and custom made templates. The Max Config
template and Generic templates are non editable.
The “MAX CONFIG TEMPLATE” is set automatically The “MAX CONFIG TEMPLATE” is automatically up-
to maximum measuring possibilities of the instrument dated every time you start the analyzer. This procedure
depending on the hardware configuration, without is done to make sure that the “MAX CONFIG TEM-
motion measurements. PLATE” always comply with the instrument configura-
tion.
What is a template Note If you work with a certain type of breaker dur-
The template contains all test plan information and ing a period in time you can set its template
settings. This helps you to test the breaker in exactly as default. By this you do not have to choose
the same way each time, which ensures accurate com- template. As soon as you turn the operation
knob the instrument uses your template.
parisons under same test conditions and settings.
If you do not select any breaker before you run a mea-
The template contains settings for the instrument; surement (i.e. turn the operation directly after starting
channels in use and configuration, connecting, timing
1] Navigate to the folder in which you want to 4] Press the “OK” button.
create a subfolder.
2] Press the “CREATE SUBFOLDER” button. Set a template as default
The default template is used when you want to run a
3] Write the name (ABB in this example) of the
quick test (see “5.4 Quick test” on page 30).
folder in the Folder ID text field.
4] Press the “OK” button. 1] Navigate to the template you want to set as
default.
Rename a subfolder
1] Navigate to the folder you want to rename.
2] Press the “CHANGE/DELETE” button.
3] Enter the new name of the folder in the
Folder ID text field.
4] Press the “OK” button.
If you want to rename it see the next section, “Edit Note To change set-up of a breaker please refer to
breaker ID”. chapter “6.5 Breaker view menu”.
2] Press the “EDIT BREAKER ID” button. 2] Press the “DELETE BREAKER” button.
Import a breaker
The default label for ID1 is Breaker ID1. In this example it
has been set to Substation. The four labels for the ID:s can 1] Navigate to the root level folder labeled
be user defined, see section “4.14 System Settings menu” Circuit Breaker(s).
on page 32.
2] Press the “IMPORT/EXPORT OF BREAKER”
3] Press the “Substation” (Breaker ID1) button. button.
Export a breaker
1] Navigate to the root level folder labeled
Circuit Breaker(s).
2] Press the “IMPORT/EXPORT OF BREAKER”
button.
3] Navigate to the breaker you want to export.
4] Press the “EXPORT BREAKER” button.
4] Press the “BACK”button in order to view the
test.
Edit Test ID
9] Press the “Date” button if you for some
1] Navigate to the test designation that you reason want to alter the date of the test oc-
want to edit. casion.
2] Press the “EDIT TEST ID” button. 10] Fill in the “Year” text field. Press the
3] Fill in the “Test ID” text field with the new “Month” and “Date” buttons repeatedly un-
designation. If this field is left empty the til the desired month and day appear in the
date and time designation will be shown in respective text fields.
the breaker list. 11] Press the “BACK” button.
4] Press the “Type of test” button. 12] Press the “Time” button.
5] Use either of the “Scroll up/Scroll down” 13] Press the “Hour”, “Minute” and “Second”
navigation buttons to choose from the user buttons repeatedly to alter the time.
defined drop down menu. Go to the next
step if you wish to add a new type. 14] Press the “BACK” button twice to return to
the Breaker List main menu.
6] Fill in the “Type of Test” text field.
7] Press the “Add to list” button. (Use the “De- Delete a test
lete from list” button to remove items from
1] Navigate to the test that you wish to delete.
the list.)
2] Press the “DELETE TEST” button.
8] Press the “BACK” button.
Scroll up
Scroll down
Scroll left
Scroll right
Zoom in
Zoom out
View a recording
Print Protocol
1] Navigate to the measurement recording des-
ignation that you want to view.
Label names TCIR[n] Common phase, trip coil current circuit [n]
In the graph window the recordings have a label for ATCIR[n] Phase A, trip coil current circuit [n]
identification. BTCIR[n] Phase B, trip coil current circuit [n]
The label names in the graph view and parameter CTCIR[n] Phase C, trip coil current circuit [n]
view are abbreviated. The first character shows the
measured quantity, the rest of the label name is the CMD[n] Common phase, close&trip command [n]
object ID. A CMD[n] Phase A, close&trip command [n]
B CMD[n] Phase B, close&trip command [n]
General text object ID
C CMD[n] Phase C, close&trip command [n]
CIR Operation coil circuit CCMD[n] Common phase, close command [n]
CMD Command, control pulse ACCMD[n] Phase A, close command [n]
T Trip BCCMD[n] Phase B, close command [n]
C Close (note C is also used for phase C) CCCMD[n] Phase C, close command [n]
M Main contact timing TCMD[n] Common phase, trip command [n]
R Resistor contact timing (not R is also used for right ATCMD[n] Phase A, trip command [n]
part of Head) BTCMD[n] Phase B, trip command [n]
R Right part of head CTCMD[n] Phase C, trip command [n]
L Left part of head
Head naming is a way of naming the breaking cham- AUXa[n] Common phase, a-contact [n]
bers on a general stand/structure. AAUXa[n] Phase A, a-contact [n]
BAUXa[n] Phase B, a-contact [n]
A Phase A CAUXa[n] Phase C, a-contact [n]
B Phase B
C Phase C AUXb[n] Common phase, b-contact [n]
AAUXb[n] Phase A, b-contact [n]
The first character in the label name BAUXb[n] Phase B, b-contact [n]
CAUXb[n] Phase C, b-contact [n]
Current I
Voltage V
- Common phase
Angle w
Motion M
A Phase A
Resistance R
B Phase B
Time t
C Phase C
Velocity v
Acceleration a
A [nn] Phase A, break No. [nn]
B [nn] Phase B, break No. [nn]
Object ID list
C [nn] Phase C, break No. [nn]
[n] - the letter n stand for a digit, 0 - 9
Parameters
The Parameters view shows a list of the recorded Display settings
parameters. Navigation and operation of this screen can be done
using the trackball and mouse buttons. Alternatively
the arrow keys can be used.
Edit notes
1] Press “EDIT NOTES”.
2] Make your notes, and press “BACK”.
Print preview
This image shows an example of a test report, see
“5.7 Reporting” on page 38.
Delete recording
1] Press “DELETE RECORDING”
Number of Phases
Here you enter whether the test object is a single or a Setting the pulse and delay times
three-phase breaker. 1] In Breaker View press the “PULSE AND DE-
LAY TIMES” button to make the settings. The
Number of interrupters per phase following submenu appears.
By repeatedly pressing the button you alter the value
to what is necessary for your test. The highest possible
value is two if number of phases is three and six if
number of phases is one.
Resistor contact
Here you select whether the test object is equipped
with resistor contacts (PIR, Pre-insertion resistor con-
tact) or not.
Operating mechanism
This button is used to select whether the test object
has a common or one separate operating mechanism
per phase.
Number of a-contacts per operating 2] For each operation sequence the settings
are to be done separately. Press the desired
mech.
sequence button. For each sequence the set-
By repeatedly pressing the button you alter the num- tings are done as follows.
ber of a-contacts per operating mech.
Number of b-contacts per operating mech.
By repeatedly pressing the button you alter the num-
ber of b-contacts per operating mech.
1] Use the keyboard to enter the desired value. 1] Use the keyboard to enter the desired value.
The arrow-up and arrow-down button can
2] Press the “BACK” button to leave this screen. be used for quick navigation between the
input fields.
Open
2] Press the “BACK” button to leave this screen.
Settings
1] In Breaker View/Measurement Preferences
press the “SETTINGS” button to make the
settings. The following screen appears.
Coil resistance
Pressing the button enables/disables coil resistance
measurements.
Transducer type
By Pressing the button you select the transducer type
to be used for testing this breaker.
2] Settings can either be done by pressing the
“Absolute” and “Relative”
respective buttons repeatedly or by selecting
the desired value from the drop down lists ▪▪The buttons are used to toggle between the transducer
using the mouse/trackball modes.
If Relative is selected you must enter a value for
the “Nominal stroke”.
Pre-trig time ▪▪If “Absolute” is selected you can optionally specify a
This is the time duration the recording starts before “Conversion” of transducer values into main contact
the sequence is initialized. values.
Compression
When you check this box the instrument will use data
compression. This will affect the measuring time and
resolution.
When you set the “Preferred measurement time” and
“Preferred sample interval” the instrument calculates
if compression is necessary. If it is necessary you will
get a message where you are asked if you want to use
compression. If you answers “NO” the test will run
although and data will be recorded until the recording
memory is full.
Hook-up instructions
1] Mark any channel and press the “i” button
in the middle of the right hand side of the
screen upon which the system provides you
with the information on how to connect.
Below is an example.
Select transducer
If you in Breaker View, Motion measurement prefer-
ences have selected Absolute (see section 6.12) “Se-
lect Transducer” is blinking on the active channel.
1] For selecting a transducer press the “SELECT”
button whereupon the transducer list ap-
pears.
Monitor
1] By pressing the “MONITOR” button you will
see information about the actual breaker
measurement and status.
Define new transducer ▪▪“Digital” transducer outputs are pulses depending on the
position of the transducer.
1] Press the “DEFINE NEW TRANSDUCER” but-
ton and the following screen appears. ▪▪“4 – 20 mA” transducer output is a current depending on
the position of the transducer.
6] Now the transducers (digital reference & 2] Enter the calibration values in the fields
resistive) should be mechanically connected shown in the upper right corner. For quick
in a way that the slider of the resistive one navigation between the fields use the arrow
is following exactly the reference. Move the buttons next to it.
transducer sliding contact to the desired posi- 3] Type in a significant name for the transducer
tion and keep it there. into the “Transducer ID” field.
The distance/position value in the fields
shown above the graphic is displayed au- 4] Press the “BACK” button to leave this screen.
tomatically using the data from the digital
transducer. The value in the corresponding Calibrating 4 - 20 mA motion transducers
voltage field will follow automatically. 1] Press the “Transducer type” button repeat-
edly to select the option “4 - 20mA”. The
7] Move the cursor to the entry field for the
following screen appears.
next position. For quick navigation between
the fields use the arrow buttons next to it.
The calibration data of each position is
stored automatically when moving the cursor
to the next field.
8] Proceed in the same way to calibrate the remain-
ing positions.
9] Type in a significant name for the transducer
into the “Transducer ID” field.
10] Press the “Finish Calibration” button.
11] Press the “BACK” button to leave this screen.
8] Proceed in the same way to calibrate the 3] Enter the transducers ratio (current and volt-
remaining positions. age) and offset values.
9] Type in a significant name for the transducer 4] Type in a significant name for the transducer
into the “Transducer ID” field. into the “Transducer ID” field.
10] Press the “Finish Calibration” button. 5] Press the “BACK” button to leave this screen.
11] Press the “BACK” button to leave this screen. Defining voltage transducers
Defining angle transducers 1] Press the “Voltage” button on the “DEFINE
NEW TRANSDUCER” screen.
1] Press the “Angular” button on the “DEFINE
NEW TRANSDUCER” screen.
6] Select all desired transducers or press the 5] Press the “BACK” button to finish.
“Toggle All” button to select all transducers
6] Confirm the “Do you want to save changes?”
from the list.
message by pressing the “Yes” button.
7] Press the “Import” button to import selected
transducers. To abort, press the “Exit” but-
ton.
Delete transducer
1] Navigate to the transducer you want to
delete.
Recalibrate transducer
1] Navigate to the transducer you want to
recalibrate.
2] Press the “RECALIBRATE” button.
3] Enter new calibration values (screens are the
same as when defining new transducer).
4] Press the “Calibrate” button.
BL1448GE ZP-BL16E TM1700-series 55
4 CABA Local
General
2] To set “Time zone”, “Month” and “Date”:
Press the respective button repeatedly to
alter the settings or use the keyboard (arrow
keys, up and down).
3] To set Year, Hour, Minute and Second:
Enter the desired value via the keyboard to
change the settings.
Company data
Language
1] By pressing the button repeatedly you alter
the CABA Local language.
Phase labels
1] By pressing the button repeatedly you alter
the Phase labels.
Date format
1] Press the button repeatedly to alter the Date
format. Label for ID1 – ID4
CABA Local uses 4 different labels to unify a test
Time format object (CB). The naming of the labels can be defined
1] Press the button repeatedly to alter the Time here by typing in your own significant terms.
format.
Company
This field can be used to enter your company name.
Report manager
Here you can assign a report template to a specific
breaker. When it is done the report template will
always be used with that breaker.
Time
This button alters the time unit.
Distance
This button to alters the distance unit.
Pressure
Report templates are grouped in three subfolders.
This button to alters the pressure unit.
Selected Breaker’s
Printer Contains all templates that come originally with the
breaker e.g. from an imported CABA Win breaker.
Templates from this folder are available for the in
Breaker List selected breaker only.
Standard
Contains all templates that come with the standard
delivery of the instrument or CABA Local respectively.
Templates from this folder are available for all breakers
in CABA Local.
Imported
Contains all imported templates e.g. from breakers
created in CABA Win. Templates from this folder are
available for all breakers in CABA Local. This folder is
empty at delivery.
REPORT MANAGER
Press this button for going to the Report Manager
Assign a report template to the selected
screen cicuit breaker
Information about the breaker presently selected is
SELECT PRINTER shown in the “Present breaker identity” field at bot-
Press this button for going to the Select Printer screen tom right of picture above. Selecting breakers takes
place in the Breaker List.
1] Navigate to the dedicated template using
the arrow keys on the left hand side of this
System Versions
Note In case of service and support question you Stop Hardware Device
might be asked for the above-mentioned
1] Press the button prior removing plug & play
information. In that case you can provide this
devices such as USB memory stick.
via Fax (printout) or via Email (text file) at your
choice.
Backup/Restore
Hardware Data You should choose an external storage for your
backup. We advise you to use the USB memory stick
to transport the backup of data from the instrument
to an external computer.
It is also possible to connect an external computer
directly to the instrument but we only recommend it if
you are familiar with how do it.
IMPORTANT
Do not remove the USB stick until you
have “stopped hardware device”, see
below
Analyze Tools 1] Insert the USB memory stick into the USB
port on the top panel.
1] Press this button to toggle between Normal
and Advanced mode 2] Select System Settings and press “Backup/
Restore”.
Note In Advanced mode the very same analyze
tool as in the optional CABA Win software is
3] Press the “BACKUP” button.
used when clicking on View a recording in the
breaker list. Please refer to the appendix for
detailed information.
Auto-save
1] Press this button to enable/disable automatic
saving. Default setting is enable.
Restore
1] Press the button”RESTORE”.
2] Follow the instructions on the screen.
5 Perform testing
IMPORTANT
The import function only supports circuit
breaker set-up/test plans created or ap-
proved by Megger Sweden AB. Megger
Sweden AB takes no responsibility for
circuit breaker set-up/test plans manually
edited or created by users.
5.7 Reporting
The instrument has an integrated Report Manager
that can view and print user defined or default re-
ports. See section “4.14 System Settings menu” on
page 64“, “Printer”, “Report Manager”.
Print preview
Here is an example of a print out, this is the parameter
list. If you also want to have the graph printed or se-
lect printer see section “4.14 System Settings menu”
on page 6456, “Printer”.
Print report
In System Settings, Printer you can select printer, de-
fault printer is the internal printer if it is installed. You
can also select if you want to print a graph, param-
eters or both from your measurements.
When viewing a recording you can choose to print a
report by clicking the button .
DRM example
Timing measurement
Simultaneous measurements within a single phase is
important in situations where a number of contacts
are connected in series. Here, the breaker becomes a
voltage divider when it opens a circuit. If the time dif-
ferences are too great, the voltage becomes too high
across one contact, and the tolerance for most types
of breakers is less than 2 ms.
The time tolerance for simultaneous measurements The distance throughout which the breaker’s electric
between phases is greater for a 3-phase power arc must be extinguished is usually called the arcing
transmission system running at 50 Hz since there is zone. From the motion curve, a velocity or accelera-
always 10 ms between zero-crossovers. Still, the time tion curve can be calculated in order to reveal even
tolerance is usually specified as less than 2 ms, even marginal changes that may have taken place in the
for such systems. It should also be noted that breakers breaker mechanics.
that perform synchronized breaking must meet more Damping is an important parameter for the high-
stringent requirements in both of the previously stated energy operating mechanisms used to open and close
situations. a circuit breaker. If the damping device does not func-
There are no generalized time limits for the time tion satisfactorily, the powerful mechanical strains that
relationships between main and auxiliary contacts, but develop can shorten breaker service life and/or cause
it is still important to understand and check their op- serious damage. The damping of opening operations
eration. The purpose of an auxiliary contact is to close is usually measured as a second speed, but it can also
and open a circuit. Such a circuit might enable a clos- be based on the time that elapses between two points
ing coil when a breaker is about to perform a closing just above the breaker’s open position.
operation and then open the circuit immediately after
the operation starts, thereby preventing coil burnout. Coil currents
The A contact must close well in advance of the clos- These can be measured on a routine basis to detect
ing of the main contact. The B contact must open potential mechanical and/or electrical problems in
when the operating mechanism has released its stored actuating coils well in advance of their emergence as
energy in order to close the breaker. The breaker actual faults. The coil’s maximum current (if current is
manufacturer will be able to provide detailed informa- permitted to reach its highest value) is a direct func-
tion about this cycle. tion of the coil’s resistance and actuating voltage.
This test indicates whether or not a winding has been
Motion measurement short-circuited.
A high-voltage breaker is designed to interrupt a When you apply a voltage across a coil, the current
specific short-circuit current, and this requires opera- curve first shows a straight transition whose rate of
tion at a given speed in order to build up an adequate rise depends on the coil’s electrical characteristic and
cooling stream of air, oil or gas (depending on the the supply voltage (points 1-2). When the coil arma-
type of breaker). This stream cools the electric arc suf- ture (which actuates the latch on the operating mech-
ficiently to interrupt the current at the next zero-cross- anism’s energy package) starts to move, the electrical
over. It is important to interrupt the current in such a relationship changes and the coil current drops (points
way that the arc will not re-strike before the breaker 3-5). When the armature hits its mechanical end posi-
contact has entered the so-called damping zone. tion, the coil current rises to the current proportional
to the coil voltage (points 5-8). The auxiliary contact
Speed is calculated between two points on the mo-
then opens the circuit and the coil current drops to
tion curve. The upper point is defined as a distance in
zero with a current decay caused by the inductance in
length, degrees or percentage of movement from a)
the circuit (points 8-9).
To do backup
Be sure you make regular backups of your record-
ings. See “Backup/Restore” in “4.14 System Settings
menu” on page 68.
6 Application examples
Transducer in use
▪▪Linear, resistive, 150 mm
Test quantities
▪▪Main contact timing measured by the Timing M/R section
▪▪Auxiliary contact timing (one a & one b contact) measured
by the Control section
▪▪Coil current & voltage measured by the Control section
Transducer in use
▪▪n/a
Others
▪▪No resistor contacts, no grading capacitors
Number of phases 3
Operating mechanism Separate
Number of interrupters per phase 2
Resistor contacts None
Number of a-contacts 1
Number of b-contacts 1
7 Specifications
8.1 Troubleshooting
Problem Possible cause Solution
When TM1700 starts in The internal battery is faulty Send the instrument to a service center
safe mode the date and
time are not correct
I’m not able to delete a The template is defined as Change the default setting
template default
I get no voltage from the The fuse is damaged Replace the damaged fuse with a F1H250V. A high
output breaking capacity type must be used
My auxiliary contacts do If the circuit breaker has AC If you have a Timing Aux section, set up your
not get detected auto- coils the control section can’t breaker in “Breaker view” to measure more then 1
matically detect the auxiliary contacts auxiliary contacts per mechanism. This means that
the Timing Aux section will measure the auxiliary
contact when you connect it to the a and b con-
tacts
The position indicators The instrument sense the position of the circuit
on the Control Panel only breaker via the control section i.e. the operating
show one phase mechanism position, therefore if common operat-
ing mechanism is selected only one LED indicates
the position for the whole breaker.
It is not possible to turn First press ctrl+alt+del and select “Task Manager”,
off the unit and/or the then under the “Processes” tab in the list locate
On/Off LED is blinking and highlight “HMI.exe”. Click the button “End
Process” in the lower right corner.
The desktop will then be displayed and you click
“Start” and “Shut Down”.
The internal screen will Megger Display mode is Change Megger Display mode to Enabled, see
flicker during start-up Disabled “How to activate the SVGA output” on page 24
8.2 FAQ - Frequently Asked The Breaker or Printer template I need does
not exist?
Questions Please refer to “create a new template” in the users
manual or contact support for technical assistance.
Can I use DC supply to TM1700?
The instrument is not rated for DC inlet supply
but there are several types of DC to AC converters
available on the market. Please contact us for more
information.
9 Data analysis
Print
Open the printer dialogue box for printing the com-
puter screen.
Edit menu 3] To reset the cursors to time zero, use the icon
buttons on the toolbar.
Notes
Used to edit notes for the selected operation.
Time 2 reset.
View menu
The table displays measurement values at time 1 and
Cursor Data 2 and also the difference. Please note that for the
Opens an on-chart-marking window to facilitate de- velocity graph the “difference” is instead the average
tailed analysis. velocity between the two cursor positions.
The two cursors can be moved using the mouse. Speed calculation points
1] Hold down left button, pick up at left edge Select if the speed calculation points should be dis-
of window or with left/right arrow keys. played.
Zooms the x-axis (time). Select zoom in, position the Lower Scroll bar
mouse at the left of zooming area, move the mouse Changes x-axis offset.
to right of zooming area while holding down left but-
ton. Release left button. This function is also available Right Scroll bar
on the tool bar. Changes y-axis offset.
Zoom Out
Layout menu
Restores the display to the original setting.
Graph
Toolbar Opens dialogue box used for graph settings. On/
Show/hide toolbar. off, colours, position, scaling etc. Selecting “store as
current” or “store as original layout” will save the set-
Status Bar tings while you work with the actual test or save the
Show/hide status bar. settings as original settings for the test.
Display
Parameter Window Opens dialogue box used for display settings. Grid on/
Opens a parameter window used for measured/calcu- off, x/y-scales and offsets, colour/monochrome etc.
lated parameters. To view all parameters use the scroll
It is also possible to change the graph and display
bar in the parameter window
directly on the screen by double-click on a label (graph
10 Quick Guide
Index
Time format.................................................. 56
Timing measurement.................................... 69
TM1700 as reference for calibrating a resistive
transducer.................................................. 49
Top panel...................................................... 19
Transducer................................................... 47
Transducers menu........................................ 48
Trig in settings.............................................. 43
Troubleshooting............................................ 78
Tune the system........................................... 21
U
Updating CABA Local................................... 29
USB memory................................................ 23
User’s Registration....................................... 13
Using circuit breaker created in CABA Win.. 64
V
View a recording........................................... 37
Voltage transducers...................................... 53
Subject to change without notice. Printed matter No. ZP-BL16E Doc. BL1448GE V06a 2017