CRB (Couch Relay Board) LED's
Standby Norm
LED# Color Signal Name In / Out TP Ref Schem Ref Comments
State State
1 Grn VERTICAL UP RELAY In 2:A6 Across K1 Coil. From Couch Power Board.
2 Grn VERTICAL DOWN RELAY In 2:A5 Across K2 Coil. From Couch Power Board.
3 Grn VERTICAL ENABLE RELAY In 2:A4 Across K3 Coil. From Couch Power Board.
4 Grn VERTICAL SPEED CTRL ENABLE RELAY In 2:A3 Across K4 Coil. From Couch Power Board.
5 Grn ISO +12VDC 2:C4 Made from 5VDC from Power Board.
TWS:473621547.xls Philips Medical Confidential Page 1
CRB TEST POINTS
Standby Norm
TP # Name In / Out LED Ref Schem Ref Comments
State State
1 INVERTER COM 2:B1 Floating - DO NOT GROUND!!
2 ISO +12VDC 5 2:D5 Floating Inverter voltage, use only TP1 as Ref.
3 VERTICAL_SPEED_PCSTR 2:C2 Floating Inverter voltage, use only TP1 as Ref.
4 VERTICAL_SPEED_K&B 2:D2 Floating Inverter voltage, use only TP1 as Ref.
5 /SERVICE_MODE_ACTIVE In 2:B3 Low True. Use TP21 as Ref.
6 2 Not Populated with TP Hook.
7 2 Not Populated with TP Hook.
8 2 Not Populated with TP Hook.
9 Not Populated with TP Hook.
10 Not Populated with TP Hook.
11 Not Populated with TP Hook.
12 Not Populated with TP Hook.
13 Not Populated with TP Hook.
14 Not Populated with TP Hook.
15 PACESETTER SPEED 3:D6 Floating Inverter voltage, use only TP1 as Ref.
16 PACESETTER 10V REF 3:D6 Floating Inverter voltage, use only TP1 as Ref.
17 K&B 5V REF 3:D4 Floating Inverter voltage, use only TP1 as Ref.
18 K&B SPEED 3:D3 Floating Inverter voltage, use only TP1 as Ref.
19 PACESETTER 15V 3:B7 Floating Inverter voltage, use only TP1 as Ref.
20 +5VDC In 2:C8 From Power Board. Use TP21 as Ref.
21 Gnd (5VDC Return) In 2:B7 From Power Board
22 P736 pin 16 3:D4 Floating Inverter voltage, use only TP1 as Ref.
23 P738 pin 10 3:A4 Floating Inverter voltage, use only TP1 as Ref.
TWS: 473621547.xls Philips Medical Confidential Page 2
CRB CONNECTORS
Conn # # Pins Type Name To / From Schem Ref Comments
P735 12 Header Couch Power Board 2:B8
P736 16 MNL 3:D5
P738 10 Header 3:B3
tws: 473621547.xls Philips Medical Confidential Page 3
CRB FUSES and SWITCHES
Fuses Type Function
Schem
Switches # Poles Settings Ref Comments