HMT330 User's Guide in English M210566EN
HMT330 User's Guide in English M210566EN
M210566EN-K
PUBLISHED BY
© Vaisala 2013
No part of this manual may be reproduced, published or publicly displayed in any form
or by any means, electronic or mechanical (including photocopying), nor may its
contents be modified, translated, adapted, sold or disclosed to a third party without prior
written permission of the copyright holder. Translated manuals and translated portions
of multilingual documents are based on the original English versions. In ambiguous
cases, the English versions are applicable, not the translations.
The contents of this manual are subject to change without prior notice.
This manual does not create any legally binding obligations for Vaisala towards
customers or end users. All legally binding obligations and agreements are included
exclusively in the applicable supply contract or the General Conditions of Sale and
General Conditions of Service of Vaisala.
_________________________________________________________________________________
Table of Contents
CHAPTER 1
GENERAL INFORMATION .......................................................................... 11
About This Manual ................................................................. 11
Contents of This Manual ..................................................... 11
Version Information ............................................................. 12
Documentation Conventions ............................................... 12
Safety ....................................................................................... 13
ESD Protection .................................................................... 13
Recycling ................................................................................ 14
Regulatory Compliances ....................................................... 14
EU Declaration of Conformity.............................................. 14
DNV Type Approval ............................................................ 15
Transmitters with LAN or WLAN Interface .......................... 16
Transmitters with WLAN Interface ...................................... 16
Patent Notice .......................................................................... 17
Trademarks ............................................................................. 17
Software License .................................................................... 17
Warranty .................................................................................. 17
CHAPTER 2
PRODUCT OVERVIEW ................................................................................ 19
Introduction to HMT330 ......................................................... 19
Basic Features and Options ................................................ 20
Structure of the Transmitter ................................................ 21
Probe Options ..................................................................... 23
Warmed Probe HMT337 ..................................................... 25
Filter Options .......................................................................... 25
Catalytic VHP Filter ............................................................. 25
CHAPTER 3
INSTALLATION ............................................................................................ 27
Mounting the Housing ........................................................... 27
Standard Mounting without Mounting Plate ........................ 27
Wall Mounting with Wall Mounting Kit ................................. 28
Mounting with DIN Rail Installation Kit ................................ 30
Pole Installation with Installation Kit for Pole or Pipeline .... 30
Mounting Rain Shield with Installation Kit ........................... 32
Panel Mounting Frame ........................................................ 33
Wiring ...................................................................................... 34
Cable Bushings ................................................................... 34
Grounding the Cables ......................................................... 35
Grounding the Transmitter Housing .................................... 36
Signal and Power Supply Wiring ......................................... 37
Connections to a 24 VAC Power Supply ............................ 38
Probe Mounting ...................................................................... 40
VAISALA _________________________________________________________________________ 1
User's Guide _______________________________________________________________________
CHAPTER 4
OPERATION ................................................................................................. 67
Getting Started ........................................................................ 67
Display/Keypad (Optional) ..................................................... 67
Basic Display ....................................................................... 67
Graphic History .................................................................... 68
Menus and Navigation ......................................................... 70
Changing the Language ................................................. 71
Rounding Setting ............................................................ 71
Display Backlight Setting ................................................ 71
Display Contrast Setting ................................................. 72
Using Display/Keypad ............................................... 72
Using Serial Line ....................................................... 72
Keypad Lock (Keyguard) ................................................ 73
Menu PIN Lock ............................................................... 73
Factory Settings ............................................................. 74
Configuring Display Alarms ................................................. 74
Using Display/Keypad .................................................... 75
Using Serial Line ............................................................ 77
MI70 Link Program for Data Handling .................................. 78
Serial Line Communication ................................................... 79
User Port Connection .......................................................... 80
User Port Operating Modes............................................ 80
Service Port Connection ...................................................... 81
Connection Cables ......................................................... 81
Installing the Driver for the USB Cable .......................... 81
Using the Service Port .................................................... 82
LAN Communication .............................................................. 82
IP Configuration ................................................................... 83
Using Display/Keypad .................................................... 84
2 ____________________________________________________________________ M210566EN-K
_________________________________________________________________________________
VAISALA _________________________________________________________________________ 3
User's Guide _______________________________________________________________________
CHAPTER 5
MODBUS .................................................................................................... 141
Overview of Modbus Protocol Support .............................. 141
Taking Modbus into Use .................................................... 142
Enabling Serial Modbus ....................................................... 143
Using Display/Keypad (Optional) ...................................... 143
Using Serial Line ............................................................... 143
Enabling Ethernet Modbus .................................................. 144
4 ____________________________________________________________________ M210566EN-K
_________________________________________________________________________________
CHAPTER 6
MAINTENANCE ......................................................................................... 149
Periodic Maintenance .......................................................... 149
Cleaning ............................................................................ 149
Changing the Probe Filter ................................................. 149
Changing the Sensor ........................................................... 150
Error States ....................................................................... 151
Technical Support ................................................................ 153
Product Returns ................................................................... 153
CHAPTER 7
CALIBRATION AND ADJUSTMENT......................................................... 155
Opening and Closing the Adjustment Mode ..................... 155
Relative Humidity Adjustment ............................................ 157
Using Push Buttons ........................................................... 157
Using Display/Keypad ....................................................... 158
Using Serial Line ............................................................... 159
Multipoint Adjustment Using Serial Line ........................... 161
MPC Command Syntax ............................................... 161
Example Multipoint Adjustment Procedure .................. 162
Relative Humidity Adjustment after Sensor Change ........ 164
Using Display/Keypad ....................................................... 164
Using Serial Line ............................................................... 164
FCRH ........................................................................... 164
Temperature Adjustment ..................................................... 165
Using Display/Keypad ....................................................... 165
Using Serial Line ............................................................... 165
Analog Output Adjustment.................................................. 167
Using Display/Keypad ....................................................... 167
Using Serial Line ............................................................... 167
ACAL ............................................................................ 167
Feeding Adjustment Information ........................................ 168
Using Display/Keypad ....................................................... 168
Using Serial Line ............................................................... 168
CTEXT ......................................................................... 168
CDATE ......................................................................... 168
CHAPTER 8
TECHNICAL DATA .................................................................................... 169
Specifications ....................................................................... 169
Performance ...................................................................... 169
Relative Humidity ......................................................... 169
Temperature (+ Operating Pressure Ranges) ............. 170
Optional Temperature Probe ....................................... 170
Calculated Variables .................................................... 171
VAISALA _________________________________________________________________________ 5
User's Guide _______________________________________________________________________
APPENDIX A
PROBE INSTALLATION KITS AND INSTALLATION EXAMPLES ......... 185
Duct Installation Kits (for HMT333/337/335) ....................... 185
Duct Installation Kit for Temperature Probe (for HMT337)
................................................................................................ 186
Pressure Tight Swagelok Installation Kits (for HMT337) .. 187
RH Probe Installation......................................................... 187
Temperature Probe Installation ......................................... 187
Examples of Vapor Tight Installations with Cable Gland . 188
RH-Probe Installations (for HMT333/337) ......................... 188
T- Probe Installations (HMT337) ....................................... 189
Example of Climate Chamber Installation ......................... 190
Example of Installation Through Roof ............................... 191
Ball Valve Installation Kit for HMT338 ................................ 192
Meteorological Installation Kit (for HMT337) ..................... 194
APPENDIX B
CALCULATION FORMULAS ..................................................................... 195
APPENDIX C
MODBUS REFERENCE ............................................................................. 199
Function Codes..................................................................... 199
Register Map ......................................................................... 200
Data Encoding ................................................................... 200
32-Bit Floating Point Format......................................... 200
6 ____________________________________________________________________ M210566EN-K
_________________________________________________________________________________
VAISALA _________________________________________________________________________ 7
User's Guide _______________________________________________________________________
List of Figures
Figure 1 Transmitter Body ...................................................................... 21
Figure 2 Inside the Transmitter ............................................................... 22
Figure 3 HMT331 Fixed Probe ............................................................... 23
Figure 4 HMT331 Short Cable Probe ..................................................... 23
Figure 5 Probe Options ........................................................................... 24
Figure 6 Standard Mounting ................................................................... 27
Figure 7 Mounting with Wall Mounting Kit .............................................. 28
Figure 8 Dimensions of the Plastic Mounting Plate (mm/inch) ............... 28
Figure 9 Dimensions of the Probe Holder Plate (mm/inch) .................... 29
Figure 10 Mounting with the DIN Rail Installation Kit ............................... 30
Figure 11 Vertical Pole.............................................................................. 30
Figure 12 Horizontal Pole ......................................................................... 31
Figure 13 Mounting with Metal Wall Mounting Plate ................................ 31
Figure 14 Dimensions of the Metal Mounting Plate (mm/inch) ................. 32
Figure 15 Mounting the Rain Shield with the Installation Kit .................... 32
Figure 16 Panel Mounting Frame ............................................................. 33
Figure 17 Panel Mounting Dimensions (mm/inch) .................................... 34
Figure 18 Cable Bushings......................................................................... 34
Figure 19 Grounding the Screen of Electrical Cable ................................ 35
Figure 20 Screw Terminal Block on Motherboard .................................... 37
Figure 21 Connections to 24 VAC Power Supply ..................................... 39
Figure 22 Measurement Error at 100 %RH .............................................. 40
Figure 23 Horizontal Mounting of Probe ................................................... 41
Figure 24 Vertical Mounting of Probe ....................................................... 42
Figure 25 HMT344 Probe ......................................................................... 44
Figure 26 Tightening the Nut .................................................................... 44
Figure 27 Cleaning of Tightening Cone .................................................... 45
Figure 28 HMT338 Probe ......................................................................... 47
Figure 29 Sealing of Fitting Body into Process ......................................... 47
Figure 30 Tightening the Clasp Nut .......................................................... 48
Figure 31 Power Supply Module ............................................................... 49
Figure 32 Galvanic Isolation Module ........................................................ 54
Figure 33 Third Analog Output ................................................................. 54
Figure 34 Third Analog Output Selection.................................................. 55
Figure 35 Relay Module ............................................................................ 57
Figure 36 RS-422/485 Module .................................................................. 58
Figure 37 4-Wire RS-485 Bus ................................................................... 60
Figure 38 2-Wire RS-485 Bus ................................................................... 61
Figure 39 LAN Interface Module ............................................................... 62
Figure 40 WLAN Interface Module ........................................................... 63
Figure 41 Data Logger Module ................................................................. 65
Figure 42 Pinout of the Optional 8-Pin Connector .................................... 66
Figure 43 Basic Display ............................................................................ 67
Figure 44 Graphical Display...................................................................... 68
Figure 45 Graphical Display with Data Logger ......................................... 69
Figure 46 Main Menus .............................................................................. 70
Figure 47 Alarm Limits Shown on Graph Screen ..................................... 75
Figure 48 Display Alarm Active ................................................................. 75
Figure 49 Display Alarms .......................................................................... 76
Figure 50 Modifying an Alarm Limit .......................................................... 76
Figure 51 Service Port Connector and User Port Terminal on
Motherboard ............................................................................. 79
Figure 52 Connection Example between PC Serial Port and User Port .. 80
8 ____________________________________________________________________ M210566EN-K
_________________________________________________________________________________
VAISALA _________________________________________________________________________ 9
User's Guide _______________________________________________________________________
List of Tables
Table 1 Manual Revisions ..................................................................... 12
Table 2 Application, Location Classes................................................... 15
Table 3 Quantities Measured by HMT330 ............................................. 19
Table 4 Optional Quantities Measured by HMT330 .............................. 19
Table 5 HMT338 Probe Dimensions ..................................................... 47
Table 6 Connecting the Twisted Pair Wires to the Screw Terminals .... 59
Table 7 4-Wire (Switch 3: On) ............................................................... 60
Table 8 2-Wire (Switch 3: Off) ............................................................... 61
Table 9 Observation Periods and Resolution ........................................ 64
Table 10 Wiring of the Optional 8-Pin Connector .................................... 66
Table 11 Periods for Trend and Max/Min Calculations ........................... 68
Table 12 Graph Information Messages in Cursor Mode ......................... 69
Table 13 ALSEL Parameters ................................................................... 77
Table 14 Default Serial Communication Settings for the User Port ........ 80
Table 15 Communication Settings for the Service Port ........................... 82
Table 16 IP Settings for the LAN and WLAN Interfaces .......................... 83
Table 17 Wireless LAN Settings .............................................................. 86
Table 18 Measurement Commands ........................................................ 93
Table 19 Communication Commands ..................................................... 93
Table 20 Formatting Commands ............................................................. 93
Table 21 Data Recording Commands ..................................................... 94
Table 22 Chemical Purge Commands ..................................................... 94
Table 23 Calibration and Adjustment Commands ................................... 94
Table 24 Setting and Testing the Analog Outputs ................................... 94
Table 25 Setting and Testing the Relays ................................................. 94
Table 26 Other Commands ..................................................................... 95
Table 27 FORM Command Modifiers .................................................... 102
Table 28 Conversion Factors for Pressure Units .................................. 105
Table 29 Filtering Levels........................................................................ 107
Table 30 Selection of Output Modes ..................................................... 114
Table 31 Relay State Examples ............................................................ 129
Table 32 Supported Modbus Variants ................................................... 141
Table 33 Error Messages ...................................................................... 152
Table 34 Indicator Led Functions .......................................................... 156
Table 35 Example List for Multipoint Correction .................................... 162
Table 36 Calculated Variables (Typical Ranges) .................................. 171
Table 37 Standard Probe Cable Lengths and Approximate
Transmitter Weight (in kg/lb) .................................................. 175
Table 38 Spare Parts and Accessories ................................................. 178
Table 39 Supported Function Codes ..................................................... 199
Table 40 HMT330 Modbus Register Blocks .......................................... 200
Table 41 Measurement Data Registers ................................................. 202
Table 42 Status Registers ..................................................................... 202
Table 43 Configuration Parameter Registers ........................................ 203
Table 44 Configuration Flag Registers .................................................. 203
Table 45 HMT330 Exception Status Outputs ........................................ 204
Table 46 HMT330 Modbus Diagnostics ................................................ 204
Table 47 HMT330 Modbus Device Identification ................................... 205
Table 48 HMT330 Modbus Exception Responses ................................ 206
10 ___________________________________________________________________ M210566EN-K
Chapter 1 _________________________________________________________ General Information
CHAPTER 1
GENERAL INFORMATION
This chapter provides general notes for the manual and the product.
VAISALA ________________________________________________________________________ 11
User's Guide _______________________________________________________________________
Version Information
Table 1 Manual Revisions
Manual Code Description
M210566EN-K December 2013. This manual. Applicable from
transmitter software version 5.14 onward. Added
description of VHP filter option. Added new
command descriptions: ALSEL, AOVER, CON,
DSEND, MODS, RGLIMIT, RHLIMIT, and MPC
(multipoint RH calibration). Updated FILT command
description. Various small corrections.
M210566EN-J Previous version. Updated Chapter 2 Product
Overview. Updated Appendix C Modbus Reference.
M210566EN-I Applicable from transmitter software version 5.10
onward. Added Modbus protocol. Updated serial line
command descriptions. Updated storage
temperature range.
Documentation Conventions
Throughout the manual, important safety considerations are highlighted
as follows:
WARNING Warning alerts you to a serious hazard. If you do not read and follow
instructions very carefully at this point, there is a risk of injury or even
death.
CAUTION Caution warns you of a potential hazard. If you do not read and follow
instructions carefully at this point, the product could be damaged or
important data could be lost.
12 ___________________________________________________________________ M210566EN-K
Chapter 1 _________________________________________________________ General Information
Safety
The Vaisala HUMICAP® Humidity and Temperature Transmitter Series
HMT330 delivered to you has been tested for safety and approved as
shipped from the factory. Note the following precautions:
CAUTION Do not modify the unit. Improper modification can damage the product,
lead to malfunction, or make the product noncompliant with applicable
legislation.
ESD Protection
Electrostatic Discharge (ESD) can cause immediate or latent damage to
electronic circuits. Vaisala products are adequately protected against
ESD for their intended use. However, it is possible to damage the product
by delivering electrostatic discharges when touching, removing, or
inserting any objects inside the equipment housing.
To make sure you are not delivering high static voltages yourself:
VAISALA ________________________________________________________________________ 13
User's Guide _______________________________________________________________________
Recycling
Recycle all applicable material.
Regulatory Compliances
EU Declaration of Conformity
Vaisala HUMICAP® Humidity and Temperature Transmitter Series
HMT330 is in conformity with the provisions of the following EU
directives:
14 ___________________________________________________________________ M210566EN-K
Chapter 1 _________________________________________________________ General Information
VAISALA ________________________________________________________________________ 15
User's Guide _______________________________________________________________________
This equipment generates, uses and can radiate radio frequency energy
and, if not installed and used in accordance with the instructions, may
cause harmful interference to radio communications. However, there is
no guarantee that interference will not occur in a particular installation. If
this equipment does cause harmful interference to radio or television
reception, which can be determined by turning the equipment off and on,
the user is encouraged to try to correct the interference by one or more of
the following measures:
To reduce potential radio interference to other users, the antenna type and
its gain should be so chosen that the equivalent isotropically radiated
power (EIRP) is not more than that permitted for successful
communication.
16 ___________________________________________________________________ M210566EN-K
Chapter 1 _________________________________________________________ General Information
Patent Notice
The Vaisala HUMICAP® Humidity and Temperature Transmitter Series
HMT330 is protected by, for example, the following patents and their
corresponding national rights:
Finnish patents 98861 and 99164, French patents 6650303 and 9504397,
German patents 69418174 and 19513274, Japanese patents 3585973 and
2801156, UK patents 0665303 and 2288465, and U.S. patent 5607564.
Trademarks
HUMICAP® is a registered trademark of Vaisala Oyj.
Software License
This product contains software developed by Vaisala. Use of the software
is governed by license terms and conditions included in the applicable
supply contract or, in the absence of separate license terms and
conditions, by the General License Conditions of Vaisala Group.
Warranty
Visit our Internet pages for standard warranty terms and conditions:
www.vaisala.com/warranty.
Please observe that any such warranty may not be valid in case of
damage due to normal wear and tear, exceptional operating conditions,
negligent handling or installation, or unauthorized modifications. Please
see the applicable supply contract or Conditions of Sale for details of the
warranty for each product.
VAISALA ________________________________________________________________________ 17
User's Guide _______________________________________________________________________
18 ___________________________________________________________________ M210566EN-K
Chapter 2 ___________________________________________________________ Product Overview
CHAPTER 2
PRODUCT OVERVIEW
Introduction to HMT330
The HMT330 transmitter provides reliable humidity measurement in a
wide range of applications. Analog outputs can be chosen between
current and voltage signals. Alternatively, digital outputs RS-232
(standard) or RS-422/485 (optional) can be selected.
VAISALA ________________________________________________________________________ 19
User's Guide _______________________________________________________________________
20 ___________________________________________________________________ M210566EN-K
Chapter 2 ___________________________________________________________ Product Overview
1104-001
VAISALA ________________________________________________________________________ 21
User's Guide _______________________________________________________________________
0508-010
22 ___________________________________________________________________ M210566EN-K
Chapter 2 ___________________________________________________________ Product Overview
Probe Options
The HMT331 is intended for demanding wall-mounted applications. The
standard version has a fixed probe.
1102-025
The HMT331 short cable probe is a special version for use with the
WLAN module and when LAN module and power supply module are
simultaneously installed. It has the HMT333 probe on a short cable
(21 cm), and a mounting plate with a probe holder.
1102-024
VAISALA ________________________________________________________________________ 23
User's Guide _______________________________________________________________________
0911-066
24 ___________________________________________________________________ M210566EN-K
Chapter 2 ___________________________________________________________ Product Overview
Filter Options
There are several filter types for HMT330. All filters are 12 mm in
diameter with a female thread, and are compatible with all HMT330
probe models. The filters recommended for a probe type can be selected
on the corresponding HMT330 order form.
Fore more information, see section Spare Parts and Accessories on page
178
The catalytic filter is based on a generally known method where the H2O2
is split to H2O and O2. This reaction is produced by the catalyst layer that
has been applied to the filter. As the filter prevents the VHP from
reaching the humidity sensor, the sensor responds to humidity only. The
catalytic filter affects the reading approximately +1 %RH at 300 pmm
VHP, or +3 %RH at 900 ppm VHP (at 23 °C / 73.4 °F).
VAISALA ________________________________________________________________________ 25
User's Guide _______________________________________________________________________
26 ___________________________________________________________________ M210566EN-K
Chapter 3 ________________________________________________________________ Installation
CHAPTER 3
INSTALLATION
This chapter provides you with information that is intended to help you
install the product.
0804-066
VAISALA ________________________________________________________________________ 27
User's Guide _______________________________________________________________________
0503-004
183 (7.20)
133 (5.24)
4.5 (0.18)
8.5 (0.33)
6.5 (0.26)
Ø6.2 (0.24)
10.5 (0.41)
116 (4.57)
104 (4.09)
84 (3.30)
59 (2.32)
Thickness
9.5 (0.37)
59 (2.32)
0804-065
28 ___________________________________________________________________ M210566EN-K
Chapter 3 ________________________________________________________________ Installation
The HMT331 short cable probe is designed to be wall mounted with the
probe holder plate (Vaisala order code 226252). The probe holder plate is
similar to the standard mounting plate, except for the probe holder at the
bottom.
0911-061
VAISALA ________________________________________________________________________ 29
User's Guide _______________________________________________________________________
1. Attach two spring holders to the plastic mounting plate by using the
screws provided in the installation kit.
2. Attach the HMT330 to the plastic mounting plate with 4 screws
(provided).
3. Press the transmitter onto the DIN rail so that the clip-fasteners
snap into the rail.
0503-002
0503-006
30 ___________________________________________________________________ M210566EN-K
Chapter 3 ________________________________________________________________ Installation
0503-007
Metal mounting plate is included in rain shield with installation kit and
installation kit for pole or pipeline.
0503-041
VAISALA ________________________________________________________________________ 31
User's Guide _______________________________________________________________________
0509-151
0503-008
32 ___________________________________________________________________ M210566EN-K
Chapter 3 ________________________________________________________________ Installation
The frame is used to hide any rough edges of the installation hole, and
provide a more finished look. Note that the panel mounting frame is not
intended to bear the weight of the transmitter, and does not include any
mounting supports.
1. Use the frame as a template to mark the required size for the
installation hole in the panel.
2. Cut the hole in the panel.
3. Mount the transmitter through the panel with suitable supports.
4. Remove the paper protecting the adhesive tape on the frame, and
attach the frame around the transmitter. Refer to Figure 16 below.
0704-002
VAISALA ________________________________________________________________________ 33
User's Guide _______________________________________________________________________
0804-083
Wiring
Cable Bushings
A single electrical cable with screen and three to ten wires is
recommended for power and analog/serial connections. The cable
diameter should be 8 ... 11 mm. The number of cable bushings depends
on the transmitter options. See the following recommendations for the
cable bushings:
0503-010
34 ___________________________________________________________________ M210566EN-K
Chapter 3 ________________________________________________________________ Installation
NOTE When there is high electric noise level (for example, near a powerful
electric motor) in the operating environment it is recommended to use
shielded cable or take care that the signal cables are separated from other
cables.
0605-027
VAISALA ________________________________________________________________________ 35
User's Guide _______________________________________________________________________
36 ___________________________________________________________________ M210566EN-K
Chapter 3 ________________________________________________________________ Installation
0506-028
1. Unfasten the four cover screws and open the transmitter cover.
2. Insert the power supply wires and signal wires through the cable
bushing in the bottom of the transmitter; see the grounding
instructions in the previous sections.
3. Connect the analog output cables to terminals: Ch1+, Ch1-, Ch2+,
Ch2-. Connect the RS-232 user port cables to terminals RxD, GND
and TxD. For more information about the RS-232 connection refer
to section Serial Line Communication on page 79.
4. When wiring the optional modules, see the corresponding section
for instructions:
- RS-422/485 Interface on page 57
- Relays on page 56
- Third Analog Output on page 54
- LAN Interface on page 62
- WLAN Interface on page 63
VAISALA ________________________________________________________________________ 37
User's Guide _______________________________________________________________________
38 ___________________________________________________________________ M210566EN-K
Chapter 3 ________________________________________________________________ Installation
0703-041
VAISALA ________________________________________________________________________ 39
User's Guide _______________________________________________________________________
Probe Mounting
In humidity measurement and especially in calibration it is essential that
temperature of the probe and measuring environment is the same. Even a
small difference in temperature between the environment and the probe
causes an error. As the curve below shows, if the temperature is +20 °C
and the relative humidity 100 %RH, a difference of ±1 °C between the
environment and the probe causes an error of ±6 %RH.
The graph below illustrates the measurement error at 100 %RH when the
difference between the ambient and sensor temperature is 1 °C.
0507-023
40 ___________________________________________________________________ M210566EN-K
Chapter 3 ________________________________________________________________ Installation
0507-024
VAISALA ________________________________________________________________________ 41
User's Guide _______________________________________________________________________
0507-022
For Vaisala probe installation kits and some installation examples, see
Appendix A on page 185.
42 ___________________________________________________________________ M210566EN-K
Chapter 3 ________________________________________________________________ Installation
The HMT333 provides for two measuring range options. The first probe
version is equipped with a flexible cable and can be used when
measuring in environments up to 80 ºC. The second version is suitable
for measuring in environments up to 120 ºC.
See Appendix A on page 185 for the following probe installation kits for
HMT333 and installation examples.
1. Remove the fitting screw from the nut and the probe.
2. Attach the fitting screw to the chamber wall with a sealing washer.
Tighten the fitting screw into the threaded sleeve with a torque
wrench. The tightening torque is 150 ± 10 Nm (110 ± 7 ft-lbs).
3. Insert the body of the probe into the fitting screw and attach the nut
manually to the fitting screw so that the connection feels tight.
4. Mark both the fitting screw and the nut hex.
VAISALA ________________________________________________________________________ 43
User's Guide _______________________________________________________________________
0506-029
5. Tighten the nut a further 30º (1/12) turn or if you have a torque
wrench tighten it with a torque of 80 ± 10 Nm (60 ± 7 ft-lbs).
0503-034
NOTE When re-tightening the nut after detachment the nut must be tightened
without increased effort.
6. Clean and grease the tightening cone of the fitting screw after every
tenth detachment. Change the sealing washer every time the fitting
screw is detached. Use high-vacuum grease (for example Dow
Corning) or similar grease.
44 ___________________________________________________________________ M210566EN-K
Chapter 3 ________________________________________________________________ Installation
0503-033
VAISALA ________________________________________________________________________ 45
User's Guide _______________________________________________________________________
The installation kits are available for both humidity and temperature
probe.
46 ___________________________________________________________________ M210566EN-K
Chapter 3 ________________________________________________________________ Installation
0508-077
0507-025
VAISALA ________________________________________________________________________ 47
User's Guide _______________________________________________________________________
0505-276
CAUTION Take care not to damage the probe body. A damaged body makes the
probe less tight and may prevent it from going through the clasp nut.
NOTE Take care not to over tighten the clasp nut to avoid difficulties when
opening it.
48 ___________________________________________________________________ M210566EN-K
Chapter 3 ________________________________________________________________ Installation
Optional Modules
Power Supply Module
The AC (mains) power connection may be connected to the power supply
module only by an authorized electrician. A readily accessible disconnect
device shall be incorporated in the fixed wiring.
0506-027
VAISALA ________________________________________________________________________ 49
User's Guide _______________________________________________________________________
Installation
1. Disconnect the power and open the transmitter cover.
2. Remove the protective plug from the cable gland and thread the
wires. In case the power supply module is installed in the factory,
continue with the step 5.
3. Attach the power module to the bottom of the housing with four
screws. See the position in Figure 2 on page 22.
4. Connect the wires from the terminals of the power supply module
marked with + and - to the terminals POWER 10 ... 35 V 24V on
the motherboard of the transmitter.
5. Connect the AC mains voltage wires to the power supply module
terminals marked with N and L.
6. Attach the grounding wire to the grounding terminal on the right-
hand side of the transmitter.
7. Connect the power. The LED on the cover of the transmitter is lit
continuously during normal operation.
WARNING Do not detach the power supply module from the transmitter when the
power is on.
WARNING Do not connect the mains power to power supply module when it is not
installed in the transmitter.
50 ___________________________________________________________________ M210566EN-K
Chapter 3 ________________________________________________________________ Installation
Warnings
Dieses Produkt entspricht der Niederspannungsrichtlinie (2006/95/EWG).
- Das Netzmodul darf nur von einem dazu befugten Elektriker angeschlossen werden.
- Trennen Sie das Netzmodul nicht vom Messwertgeber, wenn der Strom eingeschaltet ist.
- Verbinden Sie das Netzmodul nur mit der Spannungsquelle, wenn es im Messwertgeber
HMT330 montiert ist.
- Das Erdungskabel muss zum Schutz immer angeschlossen sein.
VAISALA ________________________________________________________________________ 51
User's Guide _______________________________________________________________________
- De stroom kan aan de stroomtoevoer module aangesloten worden alleen door een bevoegde
monteur.
- Het is niet toegestaan de stroomtoevoer module van de transmitter los te koppelen wanneer de
stroom aan is.
- Het is niet toegestaan de stroom aan de stroomtoevoer module aan te sluiten als deze niet in
een HMT330-transmitter is gemonteerd.
- Altijd beschermend aardcontact aansluiten!
52 ___________________________________________________________________ M210566EN-K
Chapter 3 ________________________________________________________________ Installation
- Elektros tinklą su energijos tiekimo moduliu sujungti gali tik įgaliotas elektrikas.
- Niekada neišimkite energijos tiekimo modulio iš siųstuvo, kai maitinimas yra įjungtas.
- Jei energijos tiekimo modulis nėra įmontuotas HMT330 siųstuve, nejunkite jo į elektros
tinklą.
- Visada prijunkite prie apsauginės įžeminimo jungties!
- Strāvas pieslēgumu var pieslēgt pie barošanas avota moduļa tikai autorizēts elektriķis.
- Neatvienot barošanas avota moduli no raidītāja, kad pieslēgta strāva.
- Nepievienot strāvu barošanas avota modulim, ja tas nav uzstādēts HMT330 raidītājā
- Vienmēr pievienot aizsargājošu iezemētu terminālu !
- Připojení síťového napájení k napájecímu modulu smí provádět pouze oprávněný elektrikář.
- Neodpojujte napájecí modul od snímače při zapnutém napájení.
- Nepřipojujte síťové napájení k napájecímu modulu, pokud není instalován ve snímači
HMT330.
- Vždy zapojte ochrannou zemnící svorku!
VAISALA ________________________________________________________________________ 53
User's Guide _______________________________________________________________________
NOTE Galvanic isolation module is not needed when using the AC power
supply module. Note also that these two modules are physically
incompatible to be installed at the same time. If both modules are
installed, there is no room to close the transmitter cover completely.
0506-030
0503-030
54 ___________________________________________________________________ M210566EN-K
Chapter 3 ________________________________________________________________ Installation
0508-029
VAISALA ________________________________________________________________________ 55
User's Guide _______________________________________________________________________
Relays
HMT330 can be equipped with one or two configurable relay modules.
Each module contains two configurable relays. See the contact ratings in
section Technical Specifications of Optional Modules on page 176.
CAUTION For installations in the USA: If your transmitter has both the relay
module and a LAN or WLAN module, the maximum voltage you are
allowed to connect to the relay module is 50 V.
NO Normally open
C Common relay
NC Normally closed
NOTE For instructions on how to operate the relay (for example, selecting
quantity for the relay output and setting the relay setpoints), see section
Operation of Relays on page 125.
56 ___________________________________________________________________ M210566EN-K
Chapter 3 ________________________________________________________________ Installation
0503-037
WARNING The relay module may contain dangerous voltages even if the
transmitter power has been disconnected. Before opening the transmitter
you must switch off both the transmitter and the voltage connected to
the relay terminals.
WARNING Do not connect the mains power to relay unit without grounding the
transmitter.
RS-422/485 Interface
The RS-422/485 interface enables communication between an RS-485
network and the HMT330 transmitter. The RS-485 interface is isolated
and offers a maximum communications rate of 115 200 bits/s. (For
maximum bus length of 1 km, use bit rate 19200 b/s or less.)
NOTE RS-232 User Port on HMT330 main board cannot be used and
connected when RS-485 module is connected. Service port operates
normally.
VAISALA ________________________________________________________________________ 57
User's Guide _______________________________________________________________________
1
1102-023
NOTE The data lines on earlier versions of the RS-422/485 module are marked
A and B instead of D1+ and D0-. When the line is idle, D1+ has positive
voltage compared to D0-.
When connecting the module, be prepared to swap the D1+ and D0-
wires if you have a communication problem.
58 ___________________________________________________________________ M210566EN-K
Chapter 3 ________________________________________________________________ Installation
NOTE If you use the internal termination of the transmitter at the end of the
RS-485 bus (instead of using separate terminators) removing that
transmitter will block the bus operation.
VAISALA ________________________________________________________________________ 59
User's Guide _______________________________________________________________________
Termination Termination
120R
Rx D0-
Rx D1+ 120R
Common Junction
Tx D0- box
Tx D1+ Common
Stub line
RS485 bus master
Twisted pair
Address NN
Switch Tx D0-
Term off Tx D1+ Junction
Term off Common
2/4 wire on Rx D0- box
RS422 off Rx D1+ Common
Stub line
RS485-1
Twisted pair
Address MM
Switch Tx D0-
Term off Tx D1+ Junction
Term off Common
2/4 wire on Rx D0- box
RS422 off Rx D1+ 120R
Stub 120R
RS485-1
60 ___________________________________________________________________ M210566EN-K
Chapter 3 ________________________________________________________________ Installation
Termination
120R
D0-
D1+
Common Junction
box
Stub
RS485 bus master
Twisted pair Common line
Address NN
Switch
Term off
Term off Common Junction
2/4 wire off D0- box
RS422 off D1+
Stub
RS485-1
Twisted pair Common line
Address MM
Switch
Term off Junction
Term off Common
2/4 wire off D0- box
RS422 off D1+
Stub 120R
RS485-1
Termination
1102-027
VAISALA ________________________________________________________________________ 61
User's Guide _______________________________________________________________________
LAN Interface
The optional LAN interface enables an Ethernet connection to the
transmitter. The user can establish a virtual terminal session using a
Telnet client program such as PuTTY or by using Modbus TCP protocol.
When the LAN interface is in use, serial communication using the RS-
232 User Port is disabled.
The LAN interface can use both static and dynamic network settings. If
the interface is configured to use dynamic settings, the network where the
LAN interface is connected must have a DHCP server that provides the
settings.
The network configuration can be done using the optional display and
keypad, or by using the service port. For instructions, see section LAN
Communication on page 82. The LAN interface also provides a web
configuration interface, which you can access by entering the IP address
of the LAN interface in the address field of a web browser. See section
Web Configuration for LAN and WLAN on page 89.
0709-003
62 ___________________________________________________________________ M210566EN-K
Chapter 3 ________________________________________________________________ Installation
WLAN Interface
The optional WLAN interface enables a wireless Ethernet connection
(IEEE 802.11b) to the transmitter. The user can establish a virtual
terminal session using a Telnet client program such as PuTTY or by
using Modbus TCP protocol. When the WLAN interface is in use, serial
communication using the RS-232 User Port is disabled.
The interface supports Wired Equivalent Privacy (WEP) and Wi-Fi
Protected Access (WPA, WPA2). For WEP, 64 and 128 bit encryption is
supported, with open system or shared key authentication. WPA is used
in the Pre-Shared Key (PSK) mode, with either TKIP or CCMP (also
known as AES) encryption.
Similarly to the LAN interface, the WLAN interface can use both static
and dynamic network settings. If the interface is configured to use
dynamic settings, the network where the WLAN interface is connected
must have a DHCP server that provides the settings.
The WLAN interface also provides a web configuration interface, which
you can access by entering the IP address of the WLAN interface in the
address field of a web browser. See section Web Configuration for LAN
and WLAN on page 89.
0802-103
NOTE The default country setting for the WLAN interface is United States,
which limits the operation to WLAN channels 1 ... 11. To enable
additional channels (12 ... 14), you can change the country setting using
the web configuration interface.
VAISALA ________________________________________________________________________ 63
User's Guide _______________________________________________________________________
NOTE The WLAN interface is not recommended for use with the HMT331
fixed probe model (for wall-mounted applications). The HMT331 short
cable probe is designed to be used with the WLAN interface.
The quantities that are logged are the same that have been selected for
measurement using the display/keypad or the serial line. When taking the
transmitter into use, verify that the desired quantities are selected. If you
change the quantities later, the transmitter will start logging the new
quantities, and stop logging the quantities that are no longer selected.
Changing the quantities does not delete any measurement data that is
64 ___________________________________________________________________ M210566EN-K
Chapter 3 ________________________________________________________________ Installation
The data logger module has a real time clock with a battery back-up. The
clock has been set to the Coordinated Universal Time (UTC) at the
factory, and its time cannot be set by the user. The data that is stored in
the logger's memory is timestamped using the logger's clock.
When date and time are set on the transmitter, they are stored to the
transmitter's memory as an offset from the time on the logger's clock.
When browsing the stored data, the time offset is applied to the
timestamps shown in the graphical history, and data outputted from the
serial port. The timestamps in the data logger's memory remain as they
were originally stored.
You can compensate for the clock drift (less than ±2 min/year) by setting
the time on the transmitter. This updates the time offset used on the
display and the serial port. You can set the time by using the
keypad/display or the serial commands.
0706-068
After a reset or a power up, it will usually take at least 10 seconds before
the data logger module is initialized. The instrument will not start up
before the data logger module is ready.
The indicator LED on the module will blink green during normal
operation. If the LED is lit in red color, there is a problem with the
module. The transmitter will also indicate the problem by activating the
"Internal error on add-on module" error. If the module is not operating
correctly, the transmitter must be sent to Vaisala for maintenance.
The data logger module must be installed at the factory (when ordering
the transmitter), or by a Vaisala Service Center. Once installed, the
module is automatically used by the transmitter. When the module
requires a new battery, the transmitter must be sent to Vaisala for
maintenance.
VAISALA ________________________________________________________________________ 65
User's Guide _______________________________________________________________________
8-Pin Connector
1104-126
66 ___________________________________________________________________ M210566EN-K
Chapter 4 _________________________________________________________________ Operation
CHAPTER 4
OPERATION
Getting Started
Within a few seconds after power-up the LED on the cover of the
transmitter is lit continuously indicating normal operation. If the data
logger module is installed, the startup may take up to 18 seconds.
When using the optional display and turning the transmitter on the first
time, the language selection menu window opens. Select the language
with ▼▲ arrow buttons and press the SELECT button (the left-hand
button).
Display/Keypad (Optional)
Basic Display
Display shows you the measurement values of the selected quantities in
the selected units. You can select 1 ... 4 quantities for the numerical basic
display (see section Changing Quantities and Units on page 100.)
VAISALA ________________________________________________________________________ 67
User's Guide _______________________________________________________________________
NOTE From any view, even in the absence of an EXIT button, a four-second
press on the right-hand function button takes you directly to the basic
display.
Graphic History
The graphical display shows the data trend or min/max graph of the
selected quantities, one at a time. The graph is updated automatically
while measuring.
0705-210
- Press the NEXT button to change between the trend graph and
max/min graph for the quantities selected for display.
- Press the EXIT button to return to the basic display.
- Press the ▼▲ arrow buttons to zoom in and out in the graph window.
68 ___________________________________________________________________ M210566EN-K
Chapter 4 _________________________________________________________________ Operation
- Press the ◄► arrow buttons move the cursor (vertical bar) along the
time axis. The cursor mode allows you to observe individual
measuring points. The numerical value at the cursor position is shown
at the left upper corner. The right upper corner shows the time from
the present to the chosen moment (without the logger module), or the
date and time at the cursor position (when the logger module is
installed).
- If the optional data logger module is installed, you can scroll the
cursor off the screen to move to a new point on the time axis. The new
date will be displayed, and the cursor will be centered at the date
where the cursor scrolled off the screen.
0706-030
If the data logger module is installed, you can move quickly on the time
axis: zoom out, move the cursor, and zoom back in.
The time that is shown below the graph is adjusted with the current time
offset of the transmitter. If you change the transmitter's date and time
setting, the displayed timestamps in the history graph change
accordingly. For an explanation of the effect of changing the date and
time manually, see section Data Logger Module on page 64.
A question mark after time tells you that at least one power failure
(dashed vertical line) has occurred after the chosen moment. In this case,
the exact time difference between the present and the cursor position is
not exactly known. If the data logger module is installed, the time is
known and the question marks are not present.
VAISALA ________________________________________________________________________ 69
User's Guide _______________________________________________________________________
0705-208, 0802-040
Some menu items, such as Purge in the Measuring menu, are only
shown if supported by the transmitter and the installed options.
70 ___________________________________________________________________ M210566EN-K
Chapter 4 _________________________________________________________________ Operation
Rounding Setting
Round off one decimal by using the Rounding function. The default
setting is rounding on. Rounding has no effect on quantities without
decimals.
VAISALA ________________________________________________________________________ 71
User's Guide _______________________________________________________________________
Using Display/Keypad
1. Open the Main Menu by pressing any of the ▼▲◄► arrow
buttons.
2. Select Display, press the ►arrow button.
3. Select Contrast, press the ADJUST button.
4. Adjust the contrast by pressing the ◄► arrow buttons.
5. Press OK and EXIT to return to the basic display.
CON [xxx]<cr>
where
xxx = Contrast value for display. Recommended range -9 ... 9,
default 0. Negative values make screen lighter, positive
values make it darker.
72 ___________________________________________________________________ M210566EN-K
Chapter 4 _________________________________________________________________ Operation
When you want to turn off the PIN lock, go to the menu by giving the
PIN code and select System, Menu PIN, press OFF button.
In case you have forgotten the PIN code, open the transmitter cover and
press the ADJ button once. Wait for a few seconds and the adjustment
menu opens. Select Clear menu PIN, press CLEAR.
NOTE You can also disable the keypad completely with serial command
LOCK. See section Locking Menu/Keypad Using Serial Line on page
111.
VAISALA ________________________________________________________________________ 73
User's Guide _______________________________________________________________________
Factory Settings
Use the display/keypad to restore the factory settings. This operation
does not affect the adjustments. Only settings available in the menus are
restored.
You can have the alarm activate inside or outside the specified range by
setting the limits correctly:
- To have the alarm activate when the measured quantity is outside the
specified range, set the Act. above limit higher than the Act. below
limit.
- To have the alarm activate when the measured quantity is inside the
range, set the Act. above limit lower than the Act. below limit.
The alarm limits are shown on the graph display as thicker dotted lines.
When a display alarm is activated, the automatic scaling of the graph
display always keeps the limits in view.
74 ___________________________________________________________________ M210566EN-K
Chapter 4 _________________________________________________________________ Operation
1102-012
0802-041
Multiple alarms can be active at the same time; the alarm that was
triggered first will be shown on the display. The next active alarm is
revealed when the currently shown alarm is acknowledged by pressing
the OK button.
Activated alarms are only shown on the screen; there are no alarm
messages output to the serial line. After an alarm has been
acknowledged, you must refer to the data graphs to see when the
measured quantities have exceeded the limits. You may use the optional
relay outputs as alarm outputs by configuring them separately; see
section Setting Relay Outputs on page 129.
Using Display/Keypad
1. Enter the Main Menu by pressing an arrow key on the keypad.
2. Use the arrow keys to select Display, followed by Alarms, to open
the Display Alarms menu. The Display Alarms menu shows the
currently enabled and disabled alarms.
VAISALA ________________________________________________________________________ 75
User's Guide _______________________________________________________________________
0802-069
NOTE Changes you do on the alarm editing page will take effect immediately,
and may cause an alarm to appear on the screen.
4. To select a quantity for the alarm, press the Change button and
select the quantity from the list.
5. To modify or remove the alarm limit values, move the selection
over the Act. above or Act. below field and press the Set button.
You will be prompted to Modify or Remove the value.
0802-070
When modifying the value, use the arrow up and down buttons to
change the value under the cursor. Left and right arrow buttons
move the cursor. Select the OK button to accept the modified
value, or Cancel to undo the modification.
6. Set a suitable Hysteresis value to prevent the alarm from being
triggered unnecessarily by small measurement changes that pass
the alarm limit repeatedly.
7. Set or clear the Alarm enable checkbox to enable or disable the
alarm.
8. Press the Exit button to leave the alarm configuration screen and
return to the basic view.
76 ___________________________________________________________________ M210566EN-K
Chapter 4 _________________________________________________________________ Operation
where
quantity1 = Quantity for display alarm 1. You can set any quantity
available on your transmitter; see Table 3 on page 19 and
Table 4 on page 19. In addition to quantities, you can also
select the following:
VAISALA ________________________________________________________________________ 77
User's Guide _______________________________________________________________________
Example (set Tdf and T as display alarm quantities and configure the
alarms as prompted):
>alsel tdf t
Al1 Tdf above: 80.00 'C ? 0
Al1 Tdf below: - ?
Al1 Tdf hyst : 1.00 'C ?
Al1 Tdf enabl: ON ?
Al2 T above: 40.00 'C ? 30
Al2 T below: - ?
Al2 T hyst : 1.00 'C ?
Al2 T enabl: ON ?
>
To use the MI70 Link program, you must connect your Windows PC to
the service port of the transmitter with a serial or USB cable. The MI70
Link program, and the optional connection cables, are available from
Vaisala. See section Spare Parts and Accessories on page 178.
Follow the steps below to connect your transmitter and start using the
MI70 Link:
NOTE For instructions on using the MI70 Link, see the program’s online help.
78 ___________________________________________________________________ M210566EN-K
Chapter 4 _________________________________________________________________ Operation
0605-039
VAISALA ________________________________________________________________________ 79
User's Guide _______________________________________________________________________
+
ANALOG OUTPUTS USER PORT POWER
24 V~
–
RxD
RS-232
GND
TxD
Ch1+
Ch1–
1 2 3 4 5
Ch2+
6 7 8 9
Ch2–
IOIOI
0506-033
80 ___________________________________________________________________ M210566EN-K
Chapter 4 _________________________________________________________________ Operation
1. Check that the USB service cable is not connected. Disconnect the
cable if you have already connected it.
2. Insert the media that came with the cable, or download the latest
driver from www.vaisala.com.
3. Execute the USB driver installation program (setup.exe), and
accept the installation defaults.
4. After the driver has been installed, connect the USB service cable
to a USB port on your PC. Windows will detect the new device,
and use the driver automatically.
5. The installation has reserved a COM port for the cable. Verify the
port number, and the status of the cable, using the Vaisala USB
Instrument Finder program that has been installed in the
Windows Start menu.
Windows will recognize each individual cable as a different device, and
reserve a new COM port. Remember to use the correct port in the
settings of your terminal program.
VAISALA ________________________________________________________________________ 81
User's Guide _______________________________________________________________________
LAN Communication
To enable LAN communication, a LAN or WLAN interface must be
physically connected to the network, and the networking settings must be
suitable for your network. For a description of interfaces, see sections
LAN Interface on page 62 and WLAN Interface on page 63.
The LAN and WLAN interfaces both operate by accessing the serial
interface (User Port) of the transmitter. All commands and protocols that
are available using the serial interface are available through the LAN and
WLAN interfaces; refer to section List of Serial Commands on page 93.
For instructions on how to connect using a terminal program, see section
Terminal Program Settings on page 90.
82 ___________________________________________________________________ M210566EN-K
Chapter 4 _________________________________________________________________ Operation
IP Configuration
The IP settings of the LAN and WLAN interfaces are described in Table
16. The current settings can be viewed on the serial line or using the
device information display; see section Device Information on page 108.
VAISALA ________________________________________________________________________ 83
User's Guide _______________________________________________________________________
Using Display/Keypad
You can configure the IP settings of the LAN and WLAN interfaces
using the display/keypad as follows:
1102-015
0709-004
84 ___________________________________________________________________ M210566EN-K
Chapter 4 _________________________________________________________________ Operation
where
REFRESH = Updates the network information and displays it
DISCONNECT = Disconnects all current sessions
DHCP = ON or OFF. Enables or disables the automatic IP
configuration.
WEB = ON or OFF. Enables or disables the Web
Configuration page.
IP = The four part network ID of the transmitter. Must
be set manually if automatic configuration is not
used.
SUBNET = Used together with the IP address to determine
which network the transmitter is a part of. Must be
set manually if automatic configuration is not used.
GATEWAY = IP address of the server that enables the transmitter
to access other networks. Must be set manually if
automatic configuration is not used.
Examples:
>net refresh
OK
DHCP : OFF
IP address : 192.168.0.101
Subnet mask : 255.255.255.0
Default gateway: 192.168.0.1
Web config. : OFF
MAC address : 00:40:9d:2c:d2:05
Status : Not connected
>
>net on off
DHCP : ON
IP address : 192.168.0.104
Subnet mask : 255.255.255.0
Default gateway: 192.168.0.1
Web config. : OFF
MAC address : 00:40:9d:2c:d2:05
Status : Connected
OK
>
VAISALA ________________________________________________________________________ 85
User's Guide _______________________________________________________________________
OPEN
OPEN/WEP
WPA-PSK/TKIP
WPA-PSK/CCMP
86 ___________________________________________________________________ M210566EN-K
Chapter 4 _________________________________________________________________ Operation
Using Display/Keypad
You can configure the Wireless LAN settings using the display/keypad
as follows:
0802-111
5. The Name entry on the page shows the SSID of the currently
selected wireless network. To change the SSID, press the SET
button. Use the ▲▼ arrow buttons to change the character under
the cursor, and ◄► arrow buttons to more the cursor. Press the
OK button when done.
0802-110
0802-112
VAISALA ________________________________________________________________________ 87
User's Guide _______________________________________________________________________
where
SSID = Network name in 1 … 32 characters
TYPE = Security type of the wireless network. The options are:
OPEN
OPEN/WEP
WPA-PSK/TKIP
WPA-PSK/CCMP
Examples:
>wlan ?
Network SSID : WLAN-AP
Type : OPEN
>
88 ___________________________________________________________________ M210566EN-K
Chapter 4 _________________________________________________________________ Operation
Communication Protocol
When a connection is established via the LAN or WLAN interface, the
session has the same communication mode, run interval, poll address and
echo settings as the serial port (User Port) session would have.
These settings can be changed using the display/keypad, using the serial
line (User Port or Service Port), or on-the-fly during a Telnet session.
The commands for changing the settings are SMODE, INTV, ADDR,
and ECHO.
When accessing the web configuration page, you must first log in.
Username: user
Password: vaisala
VAISALA ________________________________________________________________________ 89
User's Guide _______________________________________________________________________
1102-017
NOTE PuTTY cannot be used to access the transmitter through the User Port if
the transmitter is configured to use the Modbus protocol. However, you
can always use PuTTY to access the transmitter through the Service Port.
If you are using a Vaisala USB cable, you can check the port that it
uses by clicking the USB Finder... button. This opens the Vaisala
USB Instrument Finder program that has been installed along with
the USB drivers.
3. Check that the other serial/USB line settings are correct for your
connection, and change if necessary. The default settings (shown in
Figure 59 on page 91) are used by the Service Port of the HMT330.
90 ___________________________________________________________________ M210566EN-K
Chapter 4 _________________________________________________________________ Operation
0810-070
4. Click the Open button to open the connection window and start
using the serial line.
If PuTTY is unable to open the serial port you selected, it will show
you an error message instead. If this happens, restart PuTTY and
check the settings.
NOTE The instructions below assume that the LAN/WLAN interface on the
transmitter is correctly configured and the network connection is already
done.
VAISALA ________________________________________________________________________ 91
User's Guide _______________________________________________________________________
0810-071
4. Click the Open button to open the connection window and start
using the Telnet session.
92 ___________________________________________________________________ M210566EN-K
Chapter 4 _________________________________________________________________ Operation
VAISALA ________________________________________________________________________ 93
User's Guide _______________________________________________________________________
94 ___________________________________________________________________ M210566EN-K
Chapter 4 _________________________________________________________________ Operation
R<cr>
Example:
>r
RH= 60.5 %RH T= 23.7 'C Tdf= 15.6 'C Td= 15.6 'C a= 13.0
g/m3 x= 11.1 g/kg Tw= 18.5 'C H2O= 17889 ppmV pw=
17.81 hPa pws= 29.43 hPa h= 52.3 kJ/kg dT= 8.1 'C
If a value is too long to fit to the allocated space in the output, or if there
is an error in outputting the quantity, the value is displayed with stars ‘*’.
Example:
>r
RH=***.* %RH T= 31.0 'C
VAISALA ________________________________________________________________________ 95
User's Guide _______________________________________________________________________
S<cr>
SEND<cr>
Examples:
>send
RH= 98.4 %RH T= 31.1 'C
>send
RH= 98.4 %RH T= 31.1 'C Td= 36.0 'C Tdf= 36.0 'C a= 42.4
g/m3 x= 38.8 g/kg Tw= 30.8 'C ppm= 62414 pw= 59.53 hPa
pws= 60.52 hPa h= 130.7 kJ/kg
SCOM<cr>
You can remove the SCOM definition by pressing Esc when entering the
command name:
>scom
Send command : measure ? <esc>
>
96 ___________________________________________________________________ M210566EN-K
Chapter 4 _________________________________________________________________ Operation
NOTE The total execution time of the DSEND command depends on the bit rate
of the serial port connection. At slow speeds, it will take a considerable
amount of time for devices with large address number to respond. The
delay will be the same even if there are no other devices at all.
DSEND<cr>
OPEN [aa]
where
aa = Address of the transmitter (0 ... 255)
CLOSE
The CLOSE command switches the transmitter back to the POLL mode.
Example:
>OPEN 2 (opens the line to transmitter 2, other
transmitters stay in POLL mode)
VAISALA ________________________________________________________________________ 97
User's Guide _______________________________________________________________________
FTIME [x]<cr>
FDATE [x]<cr>
where
x = ON or OFF
Examples:
>send
RH= 98.4 %RH T= 31.0 'C
>ftime on
Form. time : ON
>send
03:47:59 RH= 98.4 %RH T= 31.0 'C
>fdate on
Form. date : ON
>send
2004-07-05 03:48:03 RH= 98.4 %RH T= 31.0 'C
>
98 ___________________________________________________________________ M210566EN-K
Chapter 4 _________________________________________________________________ Operation
FST
To output the state of optional probe heating and chemical purge in
connection with SEND and R commands enter:
FST [x]<cr>
where
x = ON or OFF (default)
Examples:
>fst on
Form. status : ON
>send
N 0 RH= 40.1 %RH T= 24.0 'C Td= 9.7 'C Tdf= 9.7 'C a=
8.7 g/m3 x= 7.5
g/kg Tw= 15.6 'C ppm= 11980 pw= 12.00 hPa pws= 29.91 hPa
h= 43.2 kJ/kg
>purge
Purge started, press any key to abort.
>send
S 134 RH= 40.2 %RH T= 24.1 'C Td= 9.8 'C Tdf= 9.8 'C a=
8.8 g/m3 x= 7.5
g/kg Tw= 15.7 'C ppm= 12084 pw= 12.10 hPa pws= 30.11 hPa
h= 43.5 kJ/kg
>
where the state of the probe is indicated by the following letters and values:
N ... xxx = Normal operation where h xxx = Probe heat power
X ... xxx = Sensor heating where xxx = Sensor temperature (ºC)
H ... xxx = Chemical purge where xxx = Sensor temperature (ºC)
S ... xxx = Sensor cooling where xxx = Sensor temperature (ºC)
after purge
VAISALA ________________________________________________________________________ 99
User's Guide _______________________________________________________________________
General Settings
Changing Quantities and Units
To change quantities and units use serial commands or the optional
display/keypad. See Table 3 on page 19 for available quantities and
Table 4 on page 19 for optional quantities.
NOTE Only the quantities selected when ordering the device can be selected as
display output quantities.
Using Display/Keypad
Use display/keypad to select the display output quantities.
NOTE Changing the display quantities/units (by using the display/keypad) has
no effect on the serial output data.
FORM
Use the serial line command FORM to select the output quantities and
format for serial line measurement messages.
FORM [x]<cr>
where
x = Formatter string
Example:
>form
3.1 "RH=" RH " " U4 3.1 "T=" T " " U3 3.1 "Tdf=" Tdf " " U3
3.1 "Td=" Td " " U3 3.1 "a=" a " " U7 4.1 "x=" x " " U6 3.1
"Tw=" Tw " " U3 6.0 "H2O=" H2O " " U5 4.2 "pw=" pw " " U4
4.2 "pws=" pws " " U4 4.1 "h=" h " " U7 3.1 "dT=" dT " " U3
\r \n
>
When entering the command, use the abbreviations of the quantities. For
more information on quantities, see Table 3 and Table 4 on page 19.
Examples:
>form "RH=" 4.2 rh U5 #t "T=" t U3 #r #n
OK
>send
RH= 16.03%RH T= 74.66'F
Command ‘FORM /’ will return the default output format. The default
output format depends on the device configuration.
>form /
OK
>send
RH= 98.4 %RH T= 31.1 'C
>
UNIT
Use the UNIT command to select metric or non-metric output units, and
change the unit for the H2O quantity:
UNIT [x]<cr>
where
x = Unit system to be used. Options are:
M (metric units)
N (non-metric units)
where
y = Unit to be used for H20. Options are:
PPMV (parts per million by volume)
PPMW (parts per million by weight)
NOTE This command changes both the serial output and display units. When
you want to output both metric and non-metric units simultaneously on
the display, select the display units later by using the display/keypad.
Examples:
>unit m
Output units : metric
>
Note that conversions from mmHg and inHg are defined at 0°C and for
mmH2O and inH2O at 4°C.
Using Display/Keypad
Use display/keypad to set the pressure compensation. To select the
pressure unit using display/keypad, see section Changing Quantities and
Units on page 100.
XPRES [aaaa.a]<cr>
where
aaaa.a = Absolute process pressure (hPa)
Example:
>pres
Pressure : 1013.00 hPa ?
>pres 2000
Pressure : 2000.00 hPa
>
RHLIMIT [ON/OFF]<cr>
where
ON = Allowed range of RH output is 0 ... 100 %.
OFF = Allowed range of RH output is -5 ... 110 % (default).
Example:
>time
Time : 13:42:49 ?
>date
Date : 2007-05-31 ?
NOTE If the optional Data Logger Module is not installed, time and date
available on the serial ports (only) are cleared to 2000-01-01 00:00:00 at
reset or at power failure.
Data Filtering
The averaging data filter calculates an average over a certain period of
time. The lowest measurement noise is achieved with the extended
filtering. There are three filtering levels available.
Using Display/Keypad
1. Press any of the arrow buttons to open the Main Menu.
2. Select Measuring by pressing the ► arrow button.
3. Select Filtering and press CHANGE to confirm your selection.
4. Select Off/Standard/Extended and press SELECT to confirm
your selection.
5. Press EXIT to return to the basic display.
where
level = Filtering level. The options are:
Device Information
Use the display/keypad or the serial line to display the device
information.
Press the INFO button in the basic display to see the following
information:
1103-064
?
Use the serial line command ? to check the current transmitter
configuration. Command ?? is similar but can also be used if the
transmitter is in POLL mode.
Example:
>?
HMT330 / 5.10.0
Serial number : D1140055
Batch number : D0750008
Adjust. date : 2010-12-17
Adjust. info : Helsinki / FIN
Date : 2011-02-09
Time : 09:29:35
Serial mode : STOP
Baud P D S : 19200 N 8 1
Output interval: 10 s
Address : 0
Echo : ON
Pressure : 1013.25 hPa
Filter : OFF
Module 1 : LOGGER-1 (1024 MB; HW: B; SW: 3)
Module 2 : not installed
>
LIGHT
Use the LIGHT command to view or set the backlight mode of the
display (optional). Issuing the command without specifying a mode
shows the current backlight mode.
LIGHT [mode]<cr>
where
mode = Operating mode of the display backlight. The options are:
Example:
>light
Backlight : OFF
>light auto
Backlight : AUTO
>
HELP
Use the HELP command to list the currently available commands. The
available commands are determined by the device configuration and
installed options.
Example:
>help
? ACAL ADDR AERR ALSEL
ASEL CDATE CLOSE CON CRH
CT CTEXT DATE DELETE DIR
DSEL DSEND ECHO ERRS FCRH
FILT FORM HELP INTV ITEST
LIGHT LOCK MODBUS MODS PLAY
PRES R RESET SCOM SDELAY
SEND SERI SMODE SYSTEM TEST
TIME UNDELETE UNIT VERS XPRES
>
ERRS
Use the ERRS command to display transmitter error messages, see Table
33 on page 152.
MODS
Use the MODS command to display information about the optional
modules that are connected to the transmitter.
Example:
>mods
Module 1 : WLAN-1 (POST: release_82000941_J; FW:
Version 82000977_K1 10/16/2008)
Module 2 : LOGGER-1 (1024 MB; HW: B; SW: 5)
>
VERS
Use the VERS command to display software version information.
Example:
>vers
HMT330 / 5.10
>
If a PIN code has been set, the user will be prompted to enter the code
when trying to access the menu. Entering the code correctly will disable
the lock until the user returns back to the basic view.
where
x = Keypad locking level, range 0...2. The options are:
0 - No lock (enables full access)
1 - Menu locked, but graphs are accessible
2 - Keypad completely disabled
yyyy = 4-digit PIN code. The code can only be set when keypad
locking level is 1.
Examples:
>lock 1 4444
Keyboard lock : 1 [4444]
>
>lock 1
Keyboard lock : 1
>
Using Display/Keypad
1. Press any of the arrow buttons to open the Main Menu.
2. Select Interfaces and press the ►arrow button to confirm your
selection.
3. Select Serial interface and press the ►arrow button to confirm
your selection.
4. Select Bit rate and Serial format by pressing the CHANGE
button. Use the ▲▼ arrow buttons to select and press SELECT to
confirm your selection.
5. Select the Protocol to be used by the serial output by pressing the
CHANGE button. Select the option using the ▲▼ arrow buttons
and press SELECT. This is the same serial mode setting that is
changed by the SMODE serial command.
- RUN mode continuously outputs the measurement message at
the set interval. If you select the RUN mode, set the desired
RUN interval also.
- POLL mode enables multiple transmitters to share the same
RS-485 line. If you select the POLL mode, select the Device
address also, as each transmitter on the line must have an
unique address.
- In STOP mode, transmitter outputs the transmitter software
version at reset or power-up and remains silent, waiting for
commands.
- The SEND mode is almost identical to the STOP mode. The
only difference is that the transmitter outputs a measurement
message at start-up, instead of software version.
- In MODBUS mode, only Modbus protocol communication is
available. See Chapter 5, Modbus, on page 141.
6. Select the RUN interval and the unit. Press OK to confirm.
7. Select the Device address and press SET to confirm.
8. Select ECHO, and press ON to turn to it on, OFF to turn it off.
9. Press EXIT to return to the basic display.The new user port
settings set using the display/keypad are effective immediately.
SERI
Use the SERI command to set the communication settings for the user
port. Changed settings will be activated at next reset or power up.
SERI [b p d s]<cr>
where
b = Bit rate (110, 150, 300, 600, 1200, 2400, 4800, 9600,19200,
38400, 57600, 115200)
p = Parity (n = none, e = even, o = odd)
d = Data bits (7 or 8)
s = Stop bits (1 or 2)
SMODE
Use the SMODE command to set the user port start-up operating mode.
SMODE [xxxx]<cr>
where
xxx = STOP, SEND, RUN, POLL, or MODBUS.
ADDR
Use the ADDR command to set the device address of the transmitter.
Addresses are required for POLL mode and MODBUS mode (serial
Modbus).
ADDR [aa]<cr>
where
aa = Device address of the transmitter, range 0 ... 255 (default = 0)
INTV
Use the INTV command to set the RUN mode output interval. The time
interval is used only when the RUN mode is active. Setting the interval to
zero enables the fastest possible output rate.
where
xxx = Delay, range 0 ... 255.
yyy = Unit: S, MIN or H.
SDELAY
With the SDELAY command you can set delay (response time) for user
port, or view currently set delay value. Adjusting the delay may be
needed when half-duplex communication (typically two-wire RS-485) is
used.
Example:
>sdelay
Serial delay : 0 ? 10
>sdelay
Serial delay : 10 ?
ECHO
Use the ECHO command to set the user port echo. The command either
enables or disables echo of characters received.
ECHO [x]<cr>
where
x = ON (enabled, default) or OFF (disabled)
NOTE When using the RS-485 interface with a 2-wire connection, always
disable echo. When using a RS-232, RS-422/485 4-wire connection,
LAN, or WLAN, you can enable or disable it as you wish.
Data Recording
Data recording function is always on and collects data automatically into
the memory of the device. If the optional data logger module is installed,
the transmitter uses it automatically. Recorded data does not disappear
from the memory when the power is switched off. Collected data can be
observed in a form of a graph in the graphical view of the display or it
can be listed out by using the serial line or MI70 Link program.
DSEL [xxx]<cr>
where
xxx = Data recording quantity. See Table 3 on page 19 and Table 4
on page 19 for the quantities.
Example:
>dsel rh t tdf
RH T Tdf
>
Enter the command without parameters and press ENTER to display the
current recording parameters.
You may also dump the logged data to the serial line in numeric form
with the following commands.
DIR
Use the serial line and enter the DIR command to check the available
files.
Without the data logger module, the device records six files (six
observation periods) for each selected quantity. The data logger raises the
number of recorded files to seven for each quantity. Thus, the total
amount of the files varies between 6 and 28. See Table 11 on page 68.
Select, for example, three quantities (RH, T, and Tdf). The last column
illustrates the number of data points that has been stored in the file.
Example (data logger module installed):
>dir
File description Oldest data available No. of points
1 RH (10 s intervals) 2007-05-30 08:26:50 13996800
2 RH (90 s intervals) 2007-05-30 05:25:30 1555200
3 RH (12 min intervals) 2007-05-29 05:48:00 194400
4 RH (2 h intervals) 2007-05-19 02:00:00 19440
5 RH (12 h intervals) 2007-03-23 12:00:00 3240
6 RH (3 d intervals) 2006-04-20 00:00:00 540
7 RH (12 d intervals) 2002-12-16 00:00:00 135
8 T (10 s intervals) 2007-05-30 08:26:50 13996800
9 T (90 s intervals) 2007-05-30 05:25:30 1555200
10 T (12 min intervals) 2007-05-29 05:48:00 194400
11 T (2 h intervals) 2007-05-19 02:00:00 19440
12 T (12 h intervals) 2007-03-23 12:00:00 3240
13 T (3 d intervals) 2006-04-20 00:00:00 540
14 T (12 d intervals) 2002-12-16 00:00:00 135
15 Tdf (10 s intervals) 2007-05-30 08:26:50 13996800
16 Tdf (90 s intervals) 2007-05-30 05:25:30 1555200
17 Tdf (12 min intervals) 2007-05-29 05:48:00 194400
18 Tdf (2 h intervals) 2007-05-19 02:00:00 19440
19 Tdf (12 h intervals) 2007-03-23 12:00:00 3240
20 Tdf (3 d intervals) 2006-04-20 00:00:00 540
21 Tdf (12 d intervals) 2002-12-16 00:00:00 135
>
Example (without data logger module):
>dir
File description Oldest data available No. of points
1 RH (10 s intervals) 2008-04-11 23:41:10 135
2 RH (90 s intervals) 2008-04-11 20:41:11 135
3 RH (12 min intervals) 2008-04-10 21:03:41 135
4 RH (2 h intervals) 2008-03-31 18:03:41 135
5 RH (12 h intervals) 2008-02-04 12:03:41 135
6 RH (3 d intervals) 2007-03-04 00:03:41 135
7 T (10 s intervals) 2008-04-11 23:41:11 135
8 T (90 s intervals) 2008-04-11 20:41:11 135
9 T (12 min intervals) 2008-04-10 21:03:41 135
10 T (2 h intervals) 2008-03-31 18:03:41 135
11 T (12 h intervals) 2008-02-04 12:03:41 135
12 T (3 d intervals) 2007-03-04 00:03:41 135
13 Tdf (10 s intervals) 2008-04-11 23:41:11 135
14 Tdf (90 s intervals) 2008-04-11 20:41:11 135
15 Tdf (12 min intervals) 2008-04-10 21:03:41 135
16 Tdf (2 h intervals) 2008-03-31 18:03:41 135
17 Tdf (12 h intervals) 2008-02-04 12:03:41 135
18 Tdf (3 d intervals) 2007-03-04 00:03:41 135
>
PLAY
Use the PLAY command to output the selected file to the serial line. If
the data logger module is installed, you can specify an interval to be
outputted.
where
x = Number of the data file that will be outputted, range 0...21.
The numbers correspond to the output of the DIR command;
refer to the example on page 116.
Selecting number 0 will output all data files.
start_date = Starting date of the interval to be outputted. Must be given in
the following format: yyyy-mm-dd.
start_time = Starting time of the interval to be outputted. Must be given in
the format hh:mm:ss or h:mm.
end_date = Ending date of the interval to be outputted. Must be given in
the following format: yyyy-mm-dd.
end_time = Ending time of the interval to be outputted. Must be given in
the format hh:mm:ss or h:mm.
Example:
>play 3 2007-05-05 00:00:00 2007-05-06 00:00:00
RH (12 min intervals) 2007-05-05 00:00:00 121
Date Time trend min max
yyyy-mm-dd hh:mm:ss %RH %RH %RH
2007-05-05 00:00:00 19.16 18.99 19.33
2007-05-05 00:12:00 19.30 19.09 19.55
2007-05-05 00:24:00 20.01 19.28 21.17
2007-05-05 00:36:00 21.21 20.98 21.44
2007-05-05 00:48:00 19.57 17.72 21.11
2007-05-05 01:00:00 19.09 18.62 19.84
...
You can use the <ESC> key to interrupt the output listing.
NOTE Output of large amounts of recorded data can result in huge data files
and take a long time, up to several days for the entire memory of the
data logger at 10 second resolution. To make it easier to process the data
it is recommended to select the largest suitable data interval, and to
specify the start and end times carefully.
Note that the transmitter automatically overwrites the old data when the
memory is full, so manual deletion of the recorded files is not necessary
in normal use.
CAUTION This function clears the entire data history of the transmitter, including
all graphs and the content of the optional data logger module.
UNDELETE
Similarly to the DELETE command, the UNDELETE command is used
without any arguments. It will recover all deleted data that has not been
overwritten yet.
0503-045
NOTE Only set one mode and one range switch to ON for one output channel.
The rest of the switches must remain OFF.
OFF ON Selection
█
1 2 3 4 5 6 7 8
█ Voltage output selected
█
█
█
█ 0 ... 5 V selected
█
█
█
█
█ 4 ... 20 mA selected
█
█
█
█
NOTE If you have customized the error output setting (AERR), check that the
set error values are still valid after changing the output mode/range. See
section Analog Output Fault Indication Setting on page 124.
AMODE/ASEL
Use the serial line to select and scale the analog output quantities. Check
the analog output modes with the AMODE command.
AMODE<cr>
Example:
>amode
Ch1 output : 0...1V
Ch2 output : 0...1V
>
Select and scale the quantities for the analog outputs with the command
ASEL. Note that the optional quantities can be selected only if they have
been selected when ordering the device.
where
xxx = Quantity of channel 1
yyy = Quantity of channel 2
zzz = Quantity of the optional analog output channel 3
Always enter the quantities for all outputs. For quantities and their
abbreviations see Table 3 on page 19 and Table 4 on page 19.
Use the ASEL [xxx yyy] command as shown in the example below when
using a device with two analog outputs.
Example:
>asel rh t
Ch1 (RH ) low : 0.00 %RH ? 0
Ch1 (RH ) high : 100.00 %RH ? 100
Ch2 (T ) low : -40.00 'C ? -50
Ch2 (T ) high : 60.00 'C ? 80
>
ITEST
Use the ITEST command to test the operation of the analog outputs. The
ITEST command forces the analog outputs to entered values. The analog
outputs remain at these values until you enter the command ITEST
without parameters or reset the transmitter.
where
aa.aaa = Current or voltage value to be set for channel 1 (mA or V)
bb.bbb = Current or voltage value to be set for channel 2 (mA or V)
cc.ccc = Current or voltage value to be set for channel 3 (optional)
(mA or V)
Examples:
>itest 20 5
Ch1 (Td ) : * 20.000 mA H'672A
Ch2 (T ) : * 5.000 mA H'34F9
>itest
Ch1 (Td ) : -23.204 'C 16.238 mA H'FFFE
Ch2 (T ) : 22.889 'C 8.573 mA H'5950
>
AERR
Use the serial line AERR command to change the error output.
Example:
>aerr
Ch1 error out : 0.000V ? 5.0
Ch2 error out : 0.000V ? 5.0
>
NOTE The error output value must be within a valid range of the output mode.
NOTE The error output value is displayed only when there are minor electrical
faults such as humidity sensor damage. When there is a severe device
malfunction, the error output value is not necessarily outputted.
AOVER [ON/OFF]<cr>
Example:
>aover on
Extended output: ON
>
Operation of Relays
Quantity for Relay Output
A relay monitors the quantity chosen for the relay output. Any of the
quantities available can be chosen.
1102-007
NOTE If the measurement of the selected quantity fails or the transmitter loses
its power, the relay is released.
Hysteresis
Hysteresis function is to prevent the relay switching back and forth when
the measured value is near to the setpoint values.
Relay is activated when the measured value passes the exact value of the
setpoint. When returning and passing the setpoint again relay is not
released before the value reaches the setpoint increased/decreased by the
hysteresis value.
Example: When the 'active above' value is 60 %RH and the hysteresis
value is 5 %RH, relay activates when the relative humidity reaches 60
%RH. As the humidity decreases, relay releases at 55 %RH.
NOTE If both setpoints are specified and "above" setpoint is lower than
"below" setpoint, the hysteresis works in the opposite direction, that is,
relay is released when the measured value passes the exact value of the
setpoint.
FAULT STATUS
Not measuring state (error state or power off): relay released (C and NC
outputs are closed)
ONLINE STATUS
Live measurement (data available): relay active (C and NO outputs are
closed)
1102-040
NOTE If transmitter loses its power, all status-based relays are released
similarly to the case of an instrument failure.
Power off
Measured RH
above 50 %
RH below 50 %*
Measured RH
below 50 %
Measured RH
above 50 %
RH above 50 %*
Measured RH
below 50 %
Measurement OK
Fault status
Error active
Measurement live
Online status Purge or sensor
heating active
* Hysteresis has an effect when relay switching is based on live measurement.
See section Hysteresis on page 127.
Enabling/Disabling Relays
You can deactivate the relay outputs for example for service purposes of
your system.
0706-003
RSEL
Use the serial line to select the quantity, setpoints and hysteresis or
enable/disable the relay outputs. Enter the RSEL command.
where
q1 = Quantity for the relay 1 or Fault/Online
q2 = Quantity for the relay 2 or Fault/Online
q3 = Quantity for the relay 3 or Fault/Online
q4 = Quantity for the relay 4 or Fault/Online
Factory setting: All relays disabled.
Use the module push buttons to activate the relays. Press the REL 1 or
REL 2 button to activate the corresponding relay.
RTEST
Use the serial line command RTEST to test the operation of the relays.
where
x = ON/OFF
Sensor Functions
Chemical Purge (Optional)
In some specific applications the sensor gain may decrease gradually due
to an interference caused by a particular chemical present in the
measured gas, for example. The decrease of sensor gain due to an
interfering chemical and the effect of the chemical purge process are
illustrated below, see Figure 66 below. The sensor polymer absorbs the
interfering chemical, and this reduces the ability of the polymer to absorb
water molecules. Consequently the sensor gain decreases. In chemical
purge, heating the humidity sensor to a temperature level of
approximately +160 °C for several minutes evaporates the interfering
chemical.
The purge function starts with heating stage, continues with settling and
when the temperature of the sensor is decreased the transmitter returns to
normal mode. The whole cycle takes about 6 minutes.
NOTE Chemical purge function locks the output values for about 6 minutes.
0508-035
- The sensor is protected with a PPS grid with stainless steel netting, a
stainless steel sintered filter or with membrane SST filter.
- The sensor temperature must be below 100 °C. At higher temperatures
the chemicals evaporate spontaneously from the sensor and the
chemical purge is not necessary.
- Subzero temperatures slow down the recovery of the sensor after the
chemical purge. The sensor is completely dry after the purge, and in
very cold temperatures it can take more than two hours for it to reach
ambient humidity again.
0508-011
1102-016
- If you have the data logger module installed, you can also specify
the Date and Time when the purge is performed next.
4. Press EXIT to return to the basic display.
0706-004
The prompt '>' appears when the heating period is over. However, the
transmitter outputs are locked to the values measured before performing
chemical purge until the settling time is over.
PUR
With PUR command you can enable or disable automatic and power-up
chemical purge and set the interval for automatic purge. If the sensor is
exposed to chemicals it is recommended to have the chemical purge done
at least once in 720 min (=12 hours). In applications where the chemical
exposure is not likely, the interval can be longer.
If the transmitter has a data logger module installed, you can also set the
starting date and time for the interval purge. After the set time has been
reached, the interval purge will be performed at the set interval. This
functionality can be used to make sure the chemical purge starts at a
specific time of the day.
NOTE To activate the new interval settings immediately, reset the transmitter.
NOTE When chemical purge in power-up is enabled, wait about 6 min after
power up before taking measurements. The output channels are locked
for the first operation minutes to the initial measured values
RGLIMIT
Use the RGLIMIT command to prevent the chemical purge in subzero
temperatures.
RGLIMIT [ON/OFF]<cr>
where
ON = Chemical purge disabled in < 0 °C temperatures.
OFF = Chemical purge allowed in < 0 °C temperatures.
Sensor Heating
This function is optionally available only in transmitters with
HUMICAP®180C or HUMICAP®180RC sensor. It should be used only
with the warmed probe.
After the heating cycle the humidity conditions are checked and new
sensor heating is performed if the predefined conditions are reached
again.
NOTE During the sensor heating the outputs are locked to the values measured
before the heating cycle.
XHEAT
Enables/disables the sensor heating.
XHEAT [xx]<cr>
where:
xx = ON / OFF
Example:
>xheat on
Extra heat : ON
>xheat off
Extra heat : OFF
>
Example:
>xheat
Extra heat : OFF
Extra heat RH : 95 ? 90
Extra heat temp: 100 ? 85
Extra heat time: 30 ? 10
>xheat on
Extra heat : ON
>
CHAPTER 5
MODBUS
The configuration instructions assume you are familiar with the use of
the transmitter. Additional details on the display/keypad, service port use,
and serial commands are available in other sections of this manual:
1101-033
4. After changing the settings, press the EXIT button. The Modbus
configuration is now complete, as configuration changes made with
the display and keypad are effective immediately.
>smode modbus
Serial mode : MODBUS
>
7. If necessary, use the SERI command to check and/or change the
serial interface settings for the User Port. For example, to set the
User Port serial settings to 19200 N 8 1, issue the following
command:
>seri 19200 N 8 1
NOTE The number of data bits must always be 8 for Modbus RTU.
Serial Modbus interface of the HMT330 does not work with baud rates
115, 150, and 300 b/s.
>addr 52
You must set a non-zero address or Modbus RTU will not work.
9. The Modbus configuration is now complete. Reset or power cycle
the transmitter to enable the Modbus mode, and proceed with the
installation of the transmitter and the wiring of the serial interface.
1101-034
Figure 71 IP Configuration
4. Navigate back to the Network Settings menu. If you are using the
WLAN interface, select Wireless LAN Settings.
5. On the Wireless LAN Settings screen, set the network name
(SSID) and security options, and exit to save the changes.
1101-036
Note that the device address setting is not relevant for Modbus
TCP. In the Modbus mode, the transmitter will respond to all valid
Modbus messages with any “unit identifier” value.
1101-034
>smode modbus
serial mode : MODBUS
>
7. Configure the networking parameters of the interface you are
using:
NOTE After transmitter startup, it may take a few minutes for the network
interface to become available for configuration.
>net
DHCP : ON ? OFF
IP address : 0.0.0.0 ? 143.154.142.102
Subnet mask : 0.0.0.0 ? 255.255.0.0
Default gateway: 0.0.0.0 ?
Web config. : ON ? OFF
Save changes (Y/N) ? y
OK
>
>wlan
Network SSID : NAME ? NETWORKID
Type : OPEN ? WPA-PSK/TKIP
WPA-PSK phrase ? thequickbrownfox
Save changes (Y/N) ? y
OK
>
1101-037
MODBUS<cr>
Example:
>modbus
Bus messages : 0
Bus comm. error: 0
Bus exceptions : 0
Slave messages : 0
Slave no resp. : 0
Last message :
>
Modbus RTU only: If the last received message was rejected because of
bad CRC checksum, the MODBUS command displays the message with
corrected CRC (last two bytes updated).
To clear the counters, select again the Modbus mode with the SMODE
command:
>smode modbus
Disabling Modbus
If you stop using Modbus with the transmitter, set the transmitter to some
other operating mode using the display/keypad option or the SMODE
command.
For example, to return the transmitter to the RUN mode where the
measurements are output at regular intervals, issue the following
command through the service port:
>smode run
Alternatively, you can enter the Main Menu using the display/keypad
option, and change the mode from the Interfaces submenu.
CHAPTER 6
MAINTENANCE
Periodic Maintenance
Cleaning
Clean the transmitter enclosure with a soft, lint-free cloth moistened with
mild detergent.
Error States
In error state the affected quantities are not measured and the output is
shown as follows:
1103-068
The error indicator disappears when the error state is over and you have
checked the error message. Press the INFO button to display the error
message.
You can also check the error message via the serial interface by using the
command ERRS. In case of constant error, please contact Vaisala. See
section Technical Support on page 153.
Technical Support
For technical questions, contact the Vaisala technical support by e-mail at
[email protected]. Provide at least the following supporting
information:
Product Returns
If the product must be returned for service, see www.vaisala.com/returns.
CHAPTER 7
CALIBRATION AND ADJUSTMENT
0508-013
0706-007
LiCl reference
NaCl reference
Using Display/Keypad
Note that the difference between the two humidity references must be at
least 50 %RH.
0706-005
6. Remove the filter from the probe and insert the probe into a
measurement hole of the dry end reference chamber (for example,
LiCl: 11 %RH in the humidity calibrator HMK15.) Use the adapter
fitting for the probes of HMT334, HMT335, HMT337 and
HMT338.
7. Wait at least 30 minutes for the sensor to stabilize. Follow the
stabilization from the GRAPH display.
8. Press READY when stabilized. If you have chosen the Other
reference value, enter now the reference value by using the arrow
buttons.
When carrying out the 2-point adjustment proceed to the next
adjustment point and carry out the procedure as described in the
previous steps.
9. Press YES to confirm the adjustment. Press OK to return to the
adjustment menu.
10. Press EXIT to close the adjustment mode and return to the basic
display. Before closing the adjustment mode, feed the adjustment
information into the device, see section Feeding Adjustment
Information on page 168.
>crh
RH : 11.25 Ref1 ? c
RH : 11.25 Ref1 ? c
RH : 11.25 Ref1 ? c
RH : 11.24 Ref1 ? c
RH : 11.24 Ref1 ? 11.3
Press any key when ready ...
9. Now the device is waiting for the high end reference. Insert the
probe into the measurement hole of the high end reference chamber
(for example, NaCl: 75 %RH chamber in the humidity calibrator
HMK15). Use the adapter fitting for the HMT334, HMT335,
HMT337 and HMT338 probes. Press any key when ready.
10. Let the probe stabilize for about 30 minutes. You can follow the
stabilization by typing C and pressing ENTER.
11. When stabilized, type the high end reference value after the
question mark and press ENTER.
>crh
RH : 11.25 Ref1 ? c
RH : 11.24 Ref1 ? c
RH : 11.24 Ref1 ? 11.3
Press any key when ready ...
RH : 75.45 Ref2 ? c
RH : 75.57 Ref2 ? c
RH : 75.55 Ref2 ? c
RH : 75.59 Ref2 ? 75.5
OK
>
12. OK indicates that the adjustment has succeeded and the new
calibration coefficients are calculated and stored. Enter the
adjustment information (date and text) to the memory of the
transmitter; see the commands CTEXT and CDATE.
13. Press the ADJ button on the motherboard to close the adjustment
mode.
14. Take the probe out of the reference conditions and replace the
filter.
NOTE Multipoint adjustment is applied “on top” of the standard user-made two
point adjustment (done with CRH command, push-buttons, or display). It
is recommended that before you start using the MPC command, you clear
the existing humidity adjustment. This is done as part of the example
multipoint adjustment procedure provided here.
where
quantity = Quantity to be adjusted. The only available quantity is
RH.
function = Desired function. The options are:
>purge
Purge started, press any key to abort...
>l
RH offset : 0.00000000E+00
RH gain : 1.00000000E+00
T offset : 0.00000000E+00
T gain : 1.00000000E+00
>li
RH offset : 0.001300000E+00 ? 0
RH gain : 1.040000000E+00 ? 1
T offset : 0.00000000E+00 ?
T gain : 1.00000000E+00 ?
To start, remove the filter from the probe and insert the probe into a
measurement hole of the driest reference chamber you intend to
use. Proceed from low to high humidity. Remember to allow at
least 30 minutes for the sensor to stabilize at each point.
7. When done preparing the list, remove the probe from the last
measurement chamber and re-attach the filter.
8. On the serial line, issue the MPC RH INIT command and start
entering the readings and reference points from the list you
prepared in step 6. When you have no more points to enter, press
enter to save the points.
>mpc rh init
1 Reading : ? 12.2
1 Reference: ? 11
2 Reading : ? 34
2 Reference: ? 33
3 Reading : ? 75.5
3 Reference: ? 75
4 Reading : ? 97.3
4 Reference: ? 97
5 Reading : ?
9. Issue the MPC RH LIST command to verify the list you entered:
>mpc rh list
MPC : OFF
# Reading Reference Correction
1 12.20 11.00 -1.20
2 34.00 33.00 -1.00
3 75.50 75.00 -0.50
4 97.30 97.00 -0.30
10. If you are satisfied with the correction list, enable multipoint
correction by issuing the command MPC RH ON.
>mpc rh on
FCRH
Example:
>fcrh
RH : 1.82 1. ref ? 0
Press any key when ready...
RH : 74.22 2. ref ? 75
OK
>
Temperature Adjustment
Using Display/Keypad
1. Press the ADJ button on the motherboard to open the
ADJUSTMENT MENU. If using a warmed probe for measuring,
probe heating will be interrupted when ADJ key is pressed. Wait
some time for the probe to reach ambient temperature.
2. Select ►Adjust T measurement, press ►key.
3. Select 1-point/ 2-point adjustment, press START.
4. Remove the filter from the probe and insert the probe into the
reference temperature.
5. Wait at least 30 minutes for the sensor to stabilize. Follow the
stabilization from the GRAPH display.
6. Press READY when stabilized. Give the reference temperature
using the arrow buttons.
>ct
>cta
>ct
T : 16.06 Ref1 ? c
T : 16.06 Ref1 ? c
T : 16.06 Ref1 ? c
T : 16.06 Ref1 ? c
T : 16.06 Ref1 ? c
T : 16.06 Ref1 ? 16.0
Press any key when ready ...
T : 16.06 Ref2 ?
OK
>
Using Display/Keypad
1. Press the ADJ button to open the ADJUSTMENT MENU.
2. Select Adjust analog outputs, press ► button.
3. Select the output to be adjusted Adjust analog output 1/2, press
START.
4. Measure the first analog output value with a multimeter. Give the
measured value by using the arrow buttons. Press OK.
5. Measure the second analog output value with a multimeter. Give
the measured value by using the arrow buttons. Press OK.
6. Press OK to return to the adjustment menu.
7. Press EXIT to close the adjustment mode and to return to the basic
display.
ACAL
Example (current outputs):
>acal
Ch1 I1 (mA) ? 2.046
Ch1 I2 (mA) ? 18.087
Ch2 I1 (mA) ? 2.036
Ch2 I2 (mA) ? 18.071
>
Using Display/Keypad
1. If you are not in the adjustment menu, press the ADJ button on the
motherboard (opens the ADJUSTMENT MENU).
2. Select Adjustment info, press the ► button.
3. Select Date, press SET. Give the date by using the arrow buttons.
Press OK.
4. Select i, press SET. Enter information text including 17 characters
at maximum by using the arrow buttons. Press OK.
5. Press EXIT to return to the basic display.
Example:
>ctext
Adjust. info : (not set) ? HMK15
>
CDATE
Use the CDATE command to enter date to adjustment information field.
Set the adjustment date in format YYYY-MM-DD.
Example:
>cdate
Adjust. date : (not set) ? 2004-05-21
>
CHAPTER 8
TECHNICAL DATA
Specifications
Performance
Relative Humidity
Measurement range 0 ... 100 %RH
Accuracy (including non-linearity, hysteresis and repeatability)
D °C
0.7
0.6
0.5
0.4
0.3
0.2
0.1
°C
0
-0.1
-0.2
-0.3
-0.4
-0.5
-0.6
-0.7
-80 -60 -40 -20 0 20 40 60 80 100 120 140 160 180
0507-021
Calculated Variables
0508-017
Operating Environment
Operating temperature range
for humidity measurement -70 ... +180 °C (-94 ... +356 °F)
see probe specifications
for transmitter body electronics -40 ... +60 °C (40 ... +140 °F)
with display 0 ... +60 °C (+32 ... +140 °F)
Mechanics
Cable bushing M20x1.5 for cable diameter
8 ... 11mm/0.31..0.43"
Conduit fitting 1/2"NPT
User cable connector (optional) M12 series 8- pin (male)
option 1 with plug (female) with 5 m /
16.4 ft black cable
option 2 with plug (female) with screw
terminals
Probe cable diameter
HMT333 80ºC 6.0 mm
Other probes 5.5 mm
Probe tube material
HMT331 Chromed ABS plastic
Other probes AISI 316L
Housing material G-AlSi 10 Mg (DIN 1725)
Housing classification
Without display IP 66 (NEMA 4X)
With display and keypad IP 65 (NEMA 4X)
Transmitter weight (with probe, cable, and modules) 1.0 ... 3.0 kg (2.2 ... 6.6 lb)
Table 37 Standard Probe Cable Lengths and Approximate
Transmitter Weight (in kg/lb)
Probe Type Probe Cable Length
2m 5m 10 m 15 m 20 m
HMT333 1.1/2.4 1.2/2.6 1.5/3.3 - 2.1/4.5
HMT334 1.4/3.1 1.6/3.5 1.9/4.2 - -
HMT335 1.3/2.9 1.4/3.1 1.7/3.7 2.0/4.3 -
HMT337 1.2/2.6 1.3/2.9 1.5/3.3 - 2.1/4.5
HMT338 232 mm 1.3/2.9 1.5/3.3 1.7/3.7 - -
HMT338 454 mm 1.4/3.1 1.6/3.5 1.9/4.2 - 2.5/5.6
External loads
current outputs RL< 500 ohms
Max load + cable loop resistance 540 ohms
0 ... .1 V RL> 2000 ohms
0 ... 5 V and 0 ... 10 V RL> 10 000 ohms
Storage temperature range -55 ... +80 °C (-67 ... +176 ºF)
3-pole screw terminal
max wire size 1.5 mm2 (AWG16)
Relay Module
Operating temperature range -40 ... +60 ºC (-40 ... +140 ºF)
Operating pressure range 500 ... 1300 mmHg
Power consumption at 24 V max 30 mA
Contacts SPDT (change over), for example,
Contact arrangement Form C
Imax 0.5 A 250 VAC
Imax 0.5 A 30 VDC
Safety standard for the relay component IEC60950 UL1950
Storage temperature range -55 ... +80 ºC (-67 ... +176 ºF)
3-pole screw terminal / relay
max wire size 2.5 mm2 (AWG14)
RS-485 Module
Operating temperature range -40 ... +60 ºC (-40 ... +140 ºF)
Operating modes 2-wire (1-pair) half duplex
4-wire (2-pair) full duplex
Operating speed max 115.2 kbaud
Bus isolation 300VDC
Power consumption at 24V max 50 mA
External loads
standard loads 32 RL> 10kohm
Storage temperature range -55 ... +80 ºC (-67 ... +176 ºF)
Max wire size 1.5 mm2 (AWG16)
Dimensions (mm/inch)
77 (3.03)
71 (2.80)
(1.08)
27.5
53.5 (2.10) 38 (1.50) 38 (1.50) 53.5 (2.10)
183 (7.20)
169 (6.65)
116 (4.57)
96 (3.78)
ø 7 (0.28)
0506-035
183 (7.20)
116 (4.57)
92 (3.62)
0804-035
HMT331
120 (4.72)
12 (0.47)
0508-030
42 (1.65)
0911-060
HMT333
0804-060
HMT334
0804-059
HMT335
0508-020
HMT337
0804-061
HMT338
0508-078
Temperature Probe
0804-062
APPENDIX A
PROBE INSTALLATION KITS AND
INSTALLATION EXAMPLES
0508-021
NOTE When the temperature difference between the duct and the air outside the duct is
remarkable, the supporting bar must be installed as deep in the duct as possible.
This prevents errors caused by the heat conduction in the bar and cable.
0507-018
0508-032
33 mm
18 mm min
P = max 10 bar
T = max 180 ºC max 30 mm
1 2 3 4
0508-016
0508-026
0508-018
0508-015
0508-022
0507-016
NOTE Let the cables hang loosely to prevent condensed water running to the
probe.
1
3 12 Outdoor
environment
11
10
9
8
Process:
4 7
For example
5 6 +85 ºC, 85 %RH
0507-015
0507-043
NOTE The probe can be installed in the process through the ball valve
assembly provided that the process pressure is less than 10 bars. This
way, the process does not have to be shut down when installing or
removing the probe. However, if the process is shut down before
removing the probe, the process pressure can be max. 20 bars.
NOTE When measuring temperature dependent quantities make sure that the
temperature at the measurement point is equal to that of the process,
otherwise the moisture reading may be incorrect.
Follow the steps below to install the HMT338 probe through a ball valve
assembly. After the installation, the probe should be sitting in the process
chamber or pipeline as shown in Figure 102 on page 192.
1. Shut down the process if the process pressure is more than 10 bars.
If the pressure is lower there is no need to shut down the process.
2. Close the ball valve.
3. Seal the threads on the fitting body; refer to Figure 29 on page 47.
4. Attach the fitting body to the ball valve and tighten it.
5. Slide the clasp nut of the probe toward the filter, as far as it will go.
6. Insert the probe to the fitting body, and manually tighten the clasp
nut to the fitting body.
7. Open the ball valve.
8. Push the probe through the ball valve assembly into the process. If
the pressure is high, use the pressing handle that is provided with
the probe. If you push the probe hard without using the handle, you
may damage the cable.
Note that the probe must be pushed so deep that the filter is
completely inside the process flow.
9. Mark the fitting screw and the clasp nut.
10. Tighten the clasp nut with a fork wrench a further 50 ... 60º
(ca. 1/6 turn). If you have a suitable torque wrench, tighten the nut
to max 45 ± 5 Nm (33 ± 4 ft-lbs). Refer to Figure 30 on
page 48.
NOTE Take care not to tighten the clasp nut more than 60° to avoid difficulties
when opening it.
If you wish to remove the probe from the process, note that you have to
pull the probe out far enough. You cannot close the valve if the groove
on the probe body is not visible.
0804-063
APPENDIX B
CALCULATION FORMULAS
This Appendix contains the formulas used for the calculated output
quantities.
Dewpoint:
Tn
Td =
m
−1 (1)
Pw
log
A
Pw is the water vapor pressure. The parameters A, m, and Tn depend on
temperature according to the following table:
t A m Tn
1)
<0 °C 6.1134 9.7911 273.47
0 ... 50 °C 6.1078 7.5000 237.3
50 ... 100 °C 5.9987 7.3313 229.1
100 ... 150 °C 5.8493 7.2756 225.0
150 ... 180 °C 6.2301 7.3033 230.0
Mixing ratio:
P
x = 621.99 × w
(2)
p-P w
Absolute humidity:
Pw
a = 216.68 ⋅ (3)
T
Enthalpy:
3
Θ = T − ∑ CiT i (5)
i =o
where:
T = temperature in K
Ci = coefficients
C0 = 0.4931358
C1 = -0.46094296 * 10-2
C2 = 0.13746454 * 10-4
C3 = -0.12743214 * 10-7
3
ln Pws = ∑ bi Θ i + b4 ln Θ (6)
i = −1
where:
bi = coefficients
b-1 = -0.58002206 * 104
b0 = 0.13914993 * 101
b1 = -0.48640239 * 10-1
b2 = 0.41764768 * 10-4
b3 = -0.14452093 * 10-7
b4 = 6.5459673
Pws
Pw = RH ⋅ (7)
100
Pw
ppmv = 106 ⋅
( p − Pw )
(8)
Symbols:
Td = dewpoint temperature (°C)
Pw = water vapor pressure (hPa)
Pws = water vapor saturation pressure (Pa)
RH = relative humidity (%)
x = mixing ratio (g/kg)
p = atmospheric pressure (hPa)
A = absolute humidity (g/m3)
T = temperature (K)
h = enthalpy (kJ/kg)
APPENDIX C
MODBUS REFERENCE
Function Codes
HMT330 supports all Modbus conformance class 0 and class 1 function
codes defined in Open Modbus/TCP Specification, Release 1.0.
Class 0 function codes are enough to access all measurement data and
configuration settings of the HMT330 transmitter. In addition, all class 1
and some class 2 commands are also supported giving better
compatibility and allowing more efficient communication when needed.
Modbus diagnostic and device identification data can be read out only
with the function codes dedicated for those purposes (08 and 43 / 14).
Register Map
All data available via the Modbus interface is grouped in six contiguous
blocks of registers as described in Table 40 below.
The addresses are 1-based decimal Modbus data model addresses without
the first digit (e.g. 0xxxx, 1xxxx, 3xxxx, or 4xxxx). Subtract 1 to get
address field values used in Modbus Protocol Data Unit (PDU).
The register map is the same for all Modbus function codes. For
example, function codes 03 and 04 return exactly same result, as do
01 and 02.
Data Encoding
All numeric values are available both in 32-bit IEEE floating point and
16-bit signed integer formats.
A “quiet NaN” value is returned for unavailable values. Writing any NaN
or infinite value is silently ignored. A quiet NaN is, for example,
0x7FC00000; however, the master should understand any NaN value.
NOTE A complete 32-bit floating point value should be read and written in a
single Modbus transaction (e.g. function codes 05, 06, and 22 do not
have an effect on floating point values).
If the scaled value does not fit in the range 0…65535 (16-bit range), the
value is “wrapped” to the range by adding or subtracting 65536 as many
times as needed.
For example, mixing ratio (x) value 658.92 g/kg is returned in 16-bit
integer format as 356 (0x0164). You will need to add suitable offset to
the 16-bit data in the Modbus master to get valid value out:
356 (16-bit register value) + 65536 (offset) = 65892 --> 658.92 g/kg
(scaled by ×0.01).
However, most of the measurement data values do not need any offset.
NOTE If your Modbus master supports 32-bit floating point values, always use
them instead of 16-bit integer registers.
The use of 16-bit integer values is not recommended in critical
applications because you cannot distinguish true zero values from zero
values generated by measurement failures.
In addition to this, 2’s complement makes negative values look like large
positive values, which must be taken into account when utilizing the
16-bit register values.
use the floating point registers to write values outside this 16-bit signed
integer range. Maximum writable value is 32767 also for configuration
registers that do not accept negative values.
NOTE Only real-time data can be read from the instrument. Data that is
recorded in the transmitter’s memory (internal memory or data logger
module) cannot be read through Modbus.
Error information can retrieved also with function code 08, sub-function
02.
NOTE Basic status information of the HMT330 transmitter is also available with
the Read Exception Status function code 07. For details, see section
Exception Status Outputs on page 204.
Configuration Registers
Configuration parameter registers are used to configure the measurement.
Writing out-of-range values is silently ignored.
Other configuration settings are not usually needed and are not available
using the Modbus interface. The service cable must be used to change
other configuration settings if needed.
Diagnostic Sub-Functions
HMT330 supports some of the Modbus diagnostic functions described in
the Modbus Application Protocol Specification V1.1b. These diagnostic
functions are accessed with function code 08. See Table 46 below for
details.
Exception Responses
Exception responses from the transmitter are according to the Modbus
Application Protocol Specification V1.1b.
*M210566EN*