LR71 e Investigating Process Parameter Mechanism
LR71 e Investigating Process Parameter Mechanism
Katharina Paulsen and Dirk Leister, Thermo Fisher Scientific, Material Characterization, Karlsruhe, Germany
Andreas Gryczke, BASF SE, Lampertheim, Germany
1
2 The feed rate for the different extruder sizes is calculated Results
in dependence on the equation of Schuler (Equation 1) [3]. To have a successful scale up, it is required to have the same
experience for the material on the lab scale extruder as on
the bigger, production scale extruder. Therefore it is assumed
that the residence time of the material within the extruder
must be the same, to allow melting and mixing on one hand
and to avoid degradation on the other.
Equation 1: Empirical equation of Schuler
Especially when working with very low feed rates, it needs
For all the experiments the screw setup was kept constant to make sure that there is no influence of the tracer itself
with two mixing sections, like shown in Figure 1. on the process and therefore on the measurement of the
residence time distribution. If we are thinking about the
low feed rates of the lab scale 11 mm extruder, where we
have only 0,17 kg/h, that means, that every second less then
50 mg is fed into the extruder. When then the amount of
tracer is too high then the value for the feed rate will be
Fig. 1: Set up of screws and barrel used for the scale up expe- higher at that moment the tracer is added and therefore
riments on the Pharma 11, Pharma 16 and Process 24. At the
feeding section the Soluplus® is added as well as the pigment also all the other parameters depending on the feed rate
is added at a given time T0. The degassing section at the left will change as well. So it is easy to imagine that the amount
hand side is an atmospheric degassing to allow water vapor of the tracer can have an influence. To determine the in-
to evaporate out of the polymer. fluence of the tracer concentration, with the same process
settings were measured the residence time distribution with
different amounts of tracer (Figure 3).
Measurement of the residence time
Obviously with increasing amount of tracer the distribution
The pigment is added as a tracer to the hopper of the feeding
gets broader and the mean residence time shifts to higher
section at a given time T0. The color concentration is
values. Therefore there should be always used a very small
measured at the die over the time.
amount and to get comparable results also always the
Picture method: a picture of the strand is taken every 0.2 sec.
same concentration of tracer.
On every picture a defined size of strand is detected regard-
For the scale up experiments at first, the feed rate was only
ing the amount of red pixel (Figure 2). calculated by the equation of Schuler. As shown in Figure 4
IR method: ExtruVis 2 is a colorimeter, developed by A. the throughput was increasing from 1 kg/h to 3 kg/h
Gryczke. It is measuring in line the concentration of the according to this equation when changing from a 11 mm
pigment in the melt at the die exit. screw diameter to an extruder with 16 mm.
Reference
[1] Breitenbach, J.: Melt extrusion, from process to
drug delivery technology, European Journal of
Pharmaceutics and Biopharmaceutics, 2002
[2] Douroumis, D.: Hot melt Extrusion - Pharma-
ceutical applications, Willey 2012
[3] Bogun, M.: Untersuchungen zur kontinuierlichen
Herstellung von Kautschukmischungen basierend
auf Rubber/Filler-Composites am Doppelschnecken-
extruder, Thesis Hallee Germany, 2005
[4] Kohlgrüber, K.: Der gleichläufige Doppelschnecken-
extruder, Carl Hanser Verlag, 2007
Acknowledgements
Fig. 8: Pharma 16 - Twin screw extruder with 16 mm screw This work was performed in collaboration with the BASF.
diameter In this collaboration the BASF and Thermo Fisher
Scientific are working close together to investigate the
dependency and influences of process parameters in Hot
Melt Extrusion Processes. Also the link between Rheology
and HME is investigated.
Especially to understand the process and the way to scale
up HME processes are focus of this work.
thermoscientific.com/mc
© 2013 Thermo Fisher Scientific Inc.⋅Copyrights in and to all photographs of instruments are owned by Thermo Fisher Scientific. This document
is for informational purposes only. Specifications, terms and pricing are subject to change. Not all products are available in every country. Please
consult your local sales representative for details.