Experiment 8 Refractometer
Experiment 8 Refractometer
Experiment No. 8
Refractometer
I. Objectives:
1. To be able to calibrate and operate the LEICA ABBE MARK II Refractometer.
2. To be able to determine the refractive index (%solids) of sugar solution.
3. To be able to verify the relationship of the refractive index to temperature.
4. To compare experimental results with standard values of % solids of sugar solutions.
II. Theory:
The method that measures how light is refracted when it passes through a given substance
is what we called refractometry. The amount by which the light is refracted determines the
refractive index. The refractive index is a physical property that is characterize by a pure
compound. It can be used to confirm the identity of a compound and to assess its purity by
comparison with the given value.
In refractometry, the instrument that is used to obtain the refractive index of a compound
is called refractometer. Refractometers are instruments that work using the principle of light
refraction through liquids. The light passes through air in a fast manner but slows down when it
passes through liquids and this phenomenon is what gives the “bent-look” of the objects when
they are partially submerged in water or any liquid for that matter. In other words, the more
dissolved solids a liquid contains or shall we say water, the slower light travels through it and
therefore the more distinct the bending effect on light.
This image shows that as the pencil is submerged in the water, we see a bending look,
which in fact gives us the wrong position of the pencil in water and this is because of the
refraction of light in the water.
The general principle of refractometry states that any time a ray of light moves one
material to another at an angle, it changes direction. The ration of the angle of refraction in a
vacuum versus that in a material of interest is the refractive index of the material.
The Abbe refractometer passes light through a thin film in a sample solution, and this
illuminates the a reference mark, and movement of an adjustment knob to adjust the visual
alignment of the mark with a reference is associated with a calibration scale in order to make the
measurement on the sample solution. This refractometer measures with a precision of four
decimal places and the temperature at which the measurements is made, and remember that the
refractive index of a solution typically decreases with increasing temperature.
III. Materials/Apparatus:
A. Apparatus:
Leica Abbe II Refractometer, Analytical Balance, Beakers
B. Materials:
Distilled water, sugar, Methanol
IV. Procedure:
A. Calibration of the refractometer:
During the experiment, our group did not perform this procedure since the group that
have conducted this experiment before us had already calibrated the refractometer, therefore, we
proceeded on the next procedure.
B. Operating Instructions:
We prepared ten sugar solution samples of different concentrations expressed in % solids.
We first weighed ten different weights of sugar in the prepared beakers with the use of the
analytical balance. Then, we poured 20 ml of water in the ten beakers containing the sugars in
order to have solutions with different concentrations.
After preparing the materials, we moved on to the refractometer. We opened the prism
assembly cover and remove the protective lens tissue. Then, we cleaned the measuring prism
surface with methanol and then with distilled water and wiped it with a clean, soft tissue because
any residue left on the prism could affect the accuracy of the readings. We set the mode selector
to the desired position and applied the first sample solution on the measuring prism surface using
a dropper and closed the prism cover and positioned the illumination arm so that the exposed
face of the upper prism was fully illuminated.
Then we turned the dispersion correction wheel so that the crosshair adjustment access
hole was at the six o’clock position and rotated the adjustment control counterclockwise to
position the shadow line at the bottom of the field of view.
We rotated the eyepiece to bring the crosshair into focus, and moved the shadow line to
the crosshair reticule with the coarse adjustment control. After that we rotated the dispersion
correction wheel in order to eliminate any red or green color at the edge of the shadow line and
turned the adjustment control to center the shadow line to the crosshair.
And lastly, we pushed the READ button to start the reading the values of the test sample,
such as the temperature of the sample placed on the measuring prism and other values that will
be shown on the results.
V. Diagram:
Note: see attached page on the back portion.
A. The data that we had gathered from the experiment are the following:
1.39
1.38
1.37
1.36
1.35
1.34
1.33
1.32
1.31
5 10 15 20 25 30 35 40 45 50
The x-axis represents the values for the concentration (wt %) of the ten sample sucrose
solutions and the refractive index for the vertical axis.
1.39
1.38
1.37
1.36
1.35
1.34
1.33
1.32
1.31
1 2 3 4 5 6 7 8 9 10
VII. Observation and Discussion of Results:
During the experiment, I observed that the refractometer is a delicate instrument and we
must have known knowledge about it before using it, since its measuring prism, the part where
drops of the sample solution is placed, is sensitive to foreign objects. Dirt and dust are the
primary enemies of the prism and that’s why we cleansed the prism right after the readings of the
measurements are through for the sample.
Foreign objects must be wiped out on the prism because it may affect the readings of the
instrument and will give an inaccurate result for the sample. I noticed that the reading of the
refractometer for the % solids of the samples increases for more concentrated samples. The same
goes for the % solids-temperature compensated in which it also increases with increasing
concentration.
The results shows that the reading of the refractometer for the % solids-TC is higher
compare to the results in the % solids where temperature is not considered. This means that the
temperature is a factor and has big effect on the % solids. The data also shows that the refractive
index of the samples increases as the % solids and concentration increases. Therefore, we can
say that the relationship among the values of the measurement is directly proportional except
only for the temperature. This is shown on the graph before which shows the relationship of the
refractive index on both concentration and sample solution.
In the experiment, the temperature does not vary greatly since we used the room
temperature of the sample only, we did not alter their temperature therefore, and they have
mostly likely the same temperature which is also from the reading of the refractometer.
Therefore, the data that we have obtained did not verify the theory that states: “the refractive
index of a solution typically decreases with increasing temperature”, since the temperature
among the sample solutions is likely the same for they are taken at room temperatures only.
I also conclude that we were not able to verify the relationship of the refractive index to
temperature since the temperature that we used for all the sugar solutions was at room
temperature only. We have not proven whether the refractive index does really decreases as the
temperature increases. But based on the results, we can say that the refractive index of the
solutions increases with an increasing concentration, and therefore I conclude that they are
directly proportional to each other.
I recommend that the refractometer should be always handled with care and clean the
measuring prism from time to time. I also recommend for those who will be conducting this
experiment later on, that they should include in their procedure to have different temperatures for
each sample solutions, or to have different temperatures for the same solution and then test it
using the refractometer and obtain the refractive index, in order to verify if the refractive index
do decrease with increasing temperature.
IX. Applications:
Refractive index has a large number of applications. It is often applied for the
determination of identity of a particular substance, know its purity, and also measure its
concentration. It can be also used in determination of drug concentration in pharmaceutical
industry. It is applied for estimation of thermophysical properties of hydrocarbons and petroleum
mixtures.
X. References:
http://chemistry.oregonstate.edu/courses/ch361-464/ch362/refract02.htm
https://laboratoryresource.com.au/?navaction=getitem&id=174
http://academics.wellesley.edu/Chemistry/chem211lab/Orgo_Lab_Manual/A
ppendix/Instruments/Refractometry/refractometry.html
https://www.ucmo.edu/chemphys/about/documents/Refractometer.pdf
http://www.omicsonline.org/refractive-index-and-its-applications-2157-
7544.1000e117.php?aid=14768