Agilent 1260 Infinity Binary Pump: User Manual
Agilent 1260 Infinity Binary Pump: User Manual
Binary Pump
User Manual
Agilent Technologies
Notices
© Agilent Technologies, Inc. 2005-2008, Warranty receive no greater than Restricted Rights as
2010-2013 defined in FAR 52.227-19(c)(1-2) (June
The material contained in this docu- 1987). U.S. Government users will receive
No part of this manual may be reproduced
ment is provided “as is,” and is sub- no greater than Limited Rights as defined in
in any form or by any means (including elec-
tronic storage and retrieval or translation
ject to being changed, without notice, FAR 52.227-14 (June 1987) or DFAR
into a foreign language) without prior agree- in future editions. Further, to the max- 252.227-7015 (b)(2) (November 1995), as
ment and written consent from Agilent imum extent permitted by applicable applicable in any technical data.
Technologies, Inc. as governed by United law, Agilent disclaims all warranties,
States and international copyright laws. either express or implied, with regard Safety Notices
to this manual and any information
Manual Part Number contained herein, including but not CAUTION
limited to the implied warranties of
G1312-90015 Rev. B merchantability and fitness for a par- A CAUTION notice denotes a
ticular purpose. Agilent shall not be hazard. It calls attention to an
Edition liable for errors or for incidental or
consequential damages in connection
operating procedure, practice, or
11/2013
with the furnishing, use, or perfor- the like that, if not correctly per-
Printed in Germany
mance of this document or of any formed or adhered to, could
Agilent Technologies information contained herein. Should result in damage to the product
Hewlett-Packard-Strasse 8 Agilent and the user have a separate
76337 Waldbronn written agreement with warranty
or loss of important data. Do not
This product may be used as a com- terms covering the material in this proceed beyond a CAUTION
ponent of an in vitro diagnostic sys- document that conflict with these notice until the indicated condi-
tem if the system is registered with terms, the warranty terms in the sep- tions are fully understood and
the appropriate authorities and com- arate agreement shall control.
met.
plies with the relevant regulations.
Otherwise, it is intended only for gen- Technology Licenses
eral laboratory use. The hardware and/or software described in WA R N I N G
this document are furnished under a license
and may be used or copied only in accor- A WARNING notice denotes a
dance with the terms of such license. hazard. It calls attention to an
Restricted Rights Legend operating procedure, practice,
If software is for use in the performance of a
or the like that, if not correctly
U.S. Government prime contract or subcon- performed or adhered to, could
tract, Software is delivered and licensed as result in personal injury or
“Commercial computer software” as
defined in DFAR 252.227-7014 (June 1995),
death. Do not proceed beyond a
or as a “commercial item” as defined in FAR WARNING notice until the indi-
2.101(a) or as “Restricted computer soft- cated conditions are fully under-
ware” as defined in FAR 52.227-19 (June
1987) or any equivalent agency regulation stood and met.
or contract clause. Use, duplication or dis-
closure of Software is subject to Agilent
Technologies’ standard commercial license
terms, and non-DOD Departments and
Agencies of the U.S. Government will
In This Guide...
This manual covers the Agilent 1260 Infinity Binary Pump (G1312B).
5 Optimizing Performance
This chapter gives information on how to optimize the performance of the
Binary Pump under special operational conditions.
7 Error Information
This chapter describes the meaning of error messages, and provides
information on probable causes and suggested actions how to recover from
error conditions.
9 Diagnostic Signals
This chapter explains all diagnostic signals of the binary pump.
10 Maintenance
This chapter describes the maintenance of the module.
12 Identifying Cables
This chapter provides information on cables used with the Agilent 1200
Infinity Series modules.
13 Hardware Information
This chapter provides detailed technical information about your binary
pump.
14 Appendix
This appendix provides general safety and environmental information.
Contents
5 Optimizing Performance 79
When to Use a Vacuum Degasser 80
When to Use the Active Seal Wash Option 81
When to Use Alternative Seals 82
When to Use the Low Volume Mixer 83
When to Remove Damper and Mixer 84
How to Optimize the Compressibility Compensation Setting 87
7 Error Information 99
What Are Error Messages 101
General Error Messages 102
Module Error Messages 109
10 Maintenance 147
Introduction to Maintenance and Repair 148
Warnings and Cautions 149
Cleaning the Module 151
Overview of Maintenance and Simple Repairs 152
Maintenance Procedures 153
Exchanging the Purge Valve Frit or the Purge Valve 154
Removing the Pump Head Assembly 157
Maintenance of a Pump Head without Seal Wash 159
Maintenance of a Pump Head with Seal Wash 163
Reinstalling the Pump Head Assembly 167
Seal Wear-in Procedure 169
Exchanging the Active Inlet Valve (AIV) or its Cartridge 170
Exchanging the Outlet Valve 173
Installation of the Solvent Selection Valve Upgrade Kit 175
Exchanging the Solvent Selection Valve 177
Exchanging the Optional Interface Board 180
Replacing Module Firmware 181
14 Appendix 239
General Safety Information 240
The Waste Electrical and Electronic Equipment Directive 243
Lithium Batteries Information 244
Radio Interference 245
Sound Emission 246
Agilent Technologies on the Internet 247
1
Introduction to the Binary Pump
Instrument and Operation 10
Introduction to the Pump 10
Principle of Operation 11
Overview of the Hydraulic Path 13
Leak and Waste Handling 18
Agilent Technologies 9
1 Introduction to the Binary Pump
Instrument and Operation
Outlet valve
Purge valve
Mixer
Principle of Operation
The binary pump is based on a two- channel, dual- piston in- series design
which comprises all essential functions that a solvent delivery system has
to fulfill. Metering of solvent and delivery to the high- pressure side are
performed by two pump assemblies which can generate pressure up to
600 bar.
Each channel comprises a pump assembly including pump drive, pump
head, active inlet valve with replaceable cartridge, and outlet valve. The
two channels are fed into a low- volume mixing chamber which is
connected via a restriction capillary coil to a damping unit and a mixer. A
pressure sensor monitors the pump pressure. A purge valve with
integrated PTFE frit is fitted to the pump outlet for convenient priming of
the pumping system.
Ejg\ZkVakZ 9VbeZg
B^mZg
EjbedjiaZi
EgZhhjgZ
hZchdg
idlVhiZ
DjiaZi DjiaZi
kVakZ kVakZ
>caZikVakZ >caZikVakZ
B^m^c\
8]VbWZg
[gdb [gdb
hdakZci HZVah HZVah hdakZci
WdiiaZ6 WdiiaZ7
E^hidch E^hidch
Ejbe]ZVY6 Ejbe]ZVY7
Figure 2 The Hydraulic Path of the Binary Pump with Damper and Mixer
Damper and mixer can be bypassed for lowest delay volume of the binary
pump. This configuration is recommended for low flow rate applications
with steep gradients, see the Agilent 1260 Infinity Binary LC System
Manual.
Figure 3 on page 12 illustrates the flow path in low delay volume mode.
For instructions on how to change between the two configurations, see
“Convert the Binary Pump to Low Delay Volume Mode” on page 85.
Bypassing the mixer while the damper remains in line is not a supported configuration and
NOTE may lead to undesired behavior of the binary pump.
B^mZg 9VbeZg
Ejg\ZkVakZ
EjbedjiaZi EgZhhjgZ
hZchdg
idlVhiZ
DjiaZikVakZ
DjiaZi
kVakZ
>caZi
kVakZ >caZikVakZ
B^m^c\X]VbWZg
[gdbhdakZci HZVah HZVah [gdbhdakZci
WdiiaZ6 WdiiaZ7
E^hidch E^hidch
Ejbe]ZVY6 Ejbe]ZVY7
Figure 3 The Hydraulic Path of the Binary Pump with Bypassed Damper and Mixer
The solvent from the bottle in the solvent cabinet enters the pump through
an active inlet valve. Each side of the binary pump comprises two
substantially identical pump units. Both pump units comprise a ball- screw
drive and a pump head with two sapphire pistons for reciprocating
movement.
Idb^m^c\X]VbWZg
>caZi DjiaZi
kVakZ kVakZ
;gdbhdakZciWdiiaZ
HZVa
E^hidc
7VaahXgZlYg^kZ
<ZVg
Bdidgl^i]ZcXdYZg
A servo- controlled variable reluctance motor drives the two ball- screw
drives in opposite directions. The gears for the ball- screw drives have
different circumferences (ratio 2:1) allowing the first piston to move at
double the speed of the second piston. The solvent enters the pump heads
close to the bottom limit and leaves it at its top. The outer diameter of
the piston is smaller than the inner diameter of the pump- head chamber
allowing the solvent to fill the gap in between. The first piston has a
stroke volume in the range of 20 µL to 100 µL depending on the flow
rate. The microprocessor controls all flow rates in a range of 1 µL/min to
5 mL/min. The inlet of the first pumping unit is connected to the active
inlet valve which is processor- controlled opened or closed allowing solvent
to be drawn into the first pump unit.
The outlet of the first pump chamber is connected by a 500 µL absorber
capillary to the second pump chamber. The outlets of the second chambers
of both pump channels joined via a small mixing chamber. A coiled
restriction capillary connects the mixing chamber via a pressure pulse
damper, a mixer and a pressure sensor to the purge valve assembly. The
outlet of the purge valve assembly is then connected to the attached
chromatographic system.
When turned on, the pump runs through an initialization procedure to
determine the upper dead center of the first piston of both pump
channels. The first piston moves slowly upwards to the mechanical stop of
the pump head and from there it moves back a predetermined path
length. The controller stores this piston position in memory. After this
initialization the pump starts operation with the set parameters for the
two pump channels.
The active inlet valve is opened and the down moving piston draws
solvent into the first pump head. At the same time the second piston is
moving upwards delivering into the system. After a controller defined
stroke length (depending on the flow rate) the drive motors are stopped
and the active inlet valve is closed. The motor direction is reversed and
moves the first piston up until it reaches the stored upper limit and at the
same time moving the second piston downwards.
Then the sequence starts again moving the pistons up and down between
the two limits. During the delivery stroke of the first piston the solvent in
the pump head is pressed through the outlet valve into the second
pumping unit. The second piston draws in half of the volume displaced by
the first piston and the remaining half volume is directly delivered into
the system. During the drawing stroke of the first piston, the second
piston delivers the drawn volume into the system.
For pump specifications, see “Performance Specifications” on page 26.
The 1200 Infinity Series has been designed for safe leak and waste
handling. It is important that all security concepts are understood and
instructions are carefully followed.
&
' 6
7
8
(
)
*
,
+
,
Figure 5 Leak and waste handling concept (overview - typical stack configuration as an
example)
2
Site Requirements and Specifications
Site Requirements 22
Physical Specifications 25
Performance Specifications 26
Agilent Technologies 21
2 Site Requirements and Specifications
Site Requirements
Site Requirements
Power Considerations
The module power supply has wide ranging capability. It accepts any line
voltage in the range described in Table 1 on page 25. Consequently there
is no voltage selector in the rear of the module. There are also no
externally accessible fuses, because automatic electronic fuses are
implemented in the power supply.
The module is partially energized when switched off, as long as the power cord is
WA R N I N G
plugged in.
Repair work at the module can lead to personal injuries, e.g. electrical shock, when
the cover is opened and the module is connected to power.
➔ Always unplug the power cable before opening the cover.
➔ Do not connect the power cable to the instrument while the covers are removed.
Power Cords
Different power cords are offered as options with the module. The female
end of all power cords is identical. It plugs into the power- input socket at
the rear. The male end of each power cord is different and designed to
match the wall socket of a particular country or region.
Bench Space
The module dimensions and weight (see Table 1 on page 25) allow you to
place the module on almost any desk or laboratory bench. It needs an
additional 2.5 cm (1.0 inches) of space on either side and approximately
8 cm (3.1 inches) in the rear for air circulation and electric connections.
If the bench shall carry a complete HPLC system, make sure that the
bench is designed to bear the weight of all modules.
The module should be operated in a horizontal position.
Condensation
Condensation within the module
CAUTION
Condensation will damage the system electronics.
➔ Do not store, ship or use your module under conditions where temperature
fluctuations could cause condensation within the module.
➔ If your module was shipped in cold weather, leave it in its box and allow it to warm
slowly to room temperature to avoid condensation.
Physical Specifications
Safety standards: Installation category II, Pollution degree 2 For indoor use only.
IEC, CSA, UL
Performance Specifications
Table 2 Performance Specifications of the Agilent 1260 Infinity Binary Pump (G1312B)
Type Specification Comments
Hydraulic system Two dual piston in series pumps with
servo-controlled variable stroke drive,
power transmission by gears and ball
screws, floating pistons
Setable flow range Set points 0.001 – 5 mL/min, in
0.001 mL/min increments
Flow range 0.05 – 5.0 mL/min
Flow precision ≤0.07 % RSD or ≤0.02 min SD, whatever is based on retention time at
greater constant room temperature
Flow accuracy ± 1 % or 10 µL/min, what ever is greater pumping degassed H2O at
10 MPa (100 bar)
Pressure operating Operating range 0 – 60 MPa (0 – 600 bar,
range 0 – 8700 psi) up to 5 mL/min
Pressure pulsation < 2 % amplitude (typically < 1.3 %), or
< 0.3 MPa (3 bar), whatever is greater, at
1 mL/min isopropanol, at all pressures
> 1 MPa (10 bar, 147 psi)
Low delay volume configuration:
< 5 % amplitude (typically < 2 %)
Compressibility Pre-defined, based on mobile phase
compensation compressibility
Recommended pH 1.0 – 12.5 , solvents with pH < 2.3 should
range not contain acids which attack stainless
steel
Gradient formation High-pressure binary mixing
Delay volume Standard delay volume configuration: measured with water at
600 – 800 µL, (includes 400 µL mixer), 1 mL/min (water/caffeine
dependent on back pressure tracer)
Low delay volume configuration:
120 µL
Table 2 Performance Specifications of the Agilent 1260 Infinity Binary Pump (G1312B)
Type Specification Comments
Composition range settable range: 0 – 100 %
recommended range: 1 – 99 % or
5 µL/min per channel, whatever is greater
Composition < 0.15 % RSD or < 0.04 min SD whatever at 0.2 and 1 mL/min; based
precision is greater on retention time at constant
room temperature
Composition ± 0.35 % absolute, at 2 mL/min, at (water/caffeine tracer)
accuracy 10 MPa (100 bar)
Control Agilent control software (e.g.
ChemStation, EZChrom, OL, MassHunter)
Local control Agilent Instant Pilot Revision B.02.00 or above
Analog output For pressure monitoring, 1.33 mV/bar,
one output
Communications Controller-area network (CAN), RS-232C,
APG Remote: ready, start, stop and
shut-down signals, LAN optional
Safety and Extensive support for troubleshooting and
maintenance maintenance is provided by the Instant
Pilot, Agilent Lab Advisor, and the
Chromatography Data System.
Safety-related features are leak detection,
safe leak handling, leak output signal for
shutdown of pumping system, and low
voltages in major maintenance areas.
GLP features Early maintenance feedback (EMF) for
continuous tracking of instrument usage
in terms of seal wear and volume of
pumped mobile phase with pre-defined
and user settable limits and feedback
messages. Electronic records of
maintenance and errors
Housing All materials are recyclable
For use with flow rates below 500 µl/min or for use without damper and mixer a vacuum
NOTE degasser is required.
All specification measurements are done with degassed solvents.
3
Installing the Pump
Unpacking the Binary Pump 30
Optimizing the Stack Configuration 33
Installation Information on Leak and Waste Handling 41
Installing the Binary Pump 45
Flow Connections with Solvent Selection Valve 48
Flow Connections without Solvent Selection Valve 51
Priming the System 54
Initial Priming 54
Regular Priming 56
Changing Solvents 57
This chapter gives information about the preferred stack setup for your system
and the installation of your binary pump.
Agilent Technologies 29
3 Installing the Pump
Unpacking the Binary Pump
Damaged Packaging
If the delivery packaging shows signs of external damage, please call your
Agilent Technologies sales and service office immediately. Inform your
service representative that the instrument may have been damaged during
shipment.
Delivery Checklist
Compare the delivery checklist with the contents of the shipping boxes to
ensure completeness of the shipment. The contents lists is shown below.
For parts identification check the illustrated parts breakdown in “Parts
and Materials for Maintenance” on page 183. Please report missing or
damaged parts to your local Agilent Technologies sales and service office.
p/n Description
G1312B Agilent 1260 Infinity Binary Pump
optionally with active seal wash and/or solvent selection valve
G1311-60003 (2x) Bottle-head assembly (OPTIONAL)
G1369C Interface board (LAN) (OPTIONAL)
G1312-67500 Calibration capillary assembly
G1312-90303 Agilent 1260 Infinity Binary LC System User Guide
not orderable
G4201-68707 HPLC Starter Kit incl. 0.17 mm i.d. cap (OPTIONAL)
G4202-68707 HPLC Starter Kit incl. 0.12 mm i.d. cap (OPTIONAL)
G4203-68708 HPLC System Tool Kit (OPTIONAL)
827975-902 Column: SB-C18, 4.6x50 mm, 1.8 µm, 600 bar (OPTIONAL)
959961-902 Column Eclipse Plus C18, 4.6 x 100 mm, 3.5 µm (OPTIONAL)
699975-302 Column Poroshell 120 EC-C18, 3.0 x 50 mm, 2.7 µm (OPTIONAL)
699975-902 Column Poroshell 120 EC-C18, 4.6 x 50 mm, 2.7 µm (OPTIONAL)
5067-4770 Solvent Cabinet Kit (OPTIONAL)
G4800-64500 Agilent 1200 Infinity Series User Documentation DVD (OPTIONAL)
not orderable (OPTIONAL)
M8500A Lab Advisor incl. license (OPTIONAL)
Power cord
Items identified as "optional" are additional accessories. They are not included in the
NOTE standard scope of delivery.
Items identified as "not orderable" can be downloaded from the Agilent website
NOTE http://www.agilent.com.
Accessory Kit
The Accessory Kit (G1311- 68755) contains the following items:
p/n Description
5063-6527 Tubing assembly, i.d. 6 mm, o.d. 9 mm, 1.2 m (to waste)
9Z\VhhZg
Ejbe
)%%bbm%#&,bb
=^E6AH
%#&,bbhZVi
&%*bbm%#&,bb
8ddaZg[dg6AH
I88 &%*bbm%#&,bb
OZgd9ZVYKdajbZjc^dc
&(¥aXZaa
969
9Z\VhhZg
Ejbe )%%bbm%#&,bb
=^E6AH
%#&'bbhZVi XV#('%bbm%#&'bb
8ddaZg[dg6AH CZlAdl9ZaVnKdajbZ
=ZVi":mX]Vc\Zg
CZlA9KedhiXdajbcXddaZg
I88 %#&'bb>9XVe#
'¥aXZaa
969
9Z\VhhZg
Ejbe )%%bbm%#&,bb
=^E6AH
%#&'bbhZVi XV#('%bbm%#&'bb
8ddaZg[dg6AH CZlAdl9ZaVnKdajbZ
=ZVi":mX]Vc\Zg
CZlA9KedhiXdajbcXddaZg
I88 %#&'bb>9XVe#
'¥aXZaa
969
9Z\VhhZg
Ejbe )%%bbm%#&,bb
=^E6AH
%#&'bbhZVi XV#('%bbm%#&'bb
8ddaZg[dg6AH CZlAdl9ZaVnKdajbZ
=ZVi":mX]Vc\Zg
CZlA9KedhiXdajbcXddaZg
I88 %#&'bb>9XVe#
'¥aXZaa
969
Reg. Pump
WPS
0.12 mm seat
Pump
Cooler f. WPS
The Agilent 1200 Infinity Series has been designed for safe leak and waste
handling. It is important that all security concepts are understood and
instructions are carefully followed.
&
' 6
7
8
(
)
*
,
+
,
Figure 12 Leak and waste handling (overview - typical stack configuration as an
example)
1 Solvent cabinet
2 Leak pan
3 Leak pan's outlet port (A), leak funnel (B) and corrugated waste tube (C)
7 Waste tube
For other cables see text below and “Overview” on page 202.
Module is partially energized when switched off, as long as the power cord is
WA R N I N G
plugged in.
Repair work at the module can lead to personal injuries, e.g. shock hazard, when the
cover is opened and the module is connected to power.
➔ Make sure that it is always possible to access the power plug.
➔ Remove the power cable from the instrument before opening the cover.
➔ Do not connect the power cable to the Instrument while the covers are removed.
HiVijh>cY^XVidg
EdlZghl^iX]
HZg^VacjbWZg
3 Connect the power cable to the power connector at the rear of the
module.
Configuration switch
RS232
Remote
Analog output
CAN-Bus
HP-IB
Power plug
5 Connect the capillary, solvent tubes and waste tubings (see “Flow
Connections with Solvent Selection Valve” on page 48 or “Flow
Connections without Solvent Selection Valve” on page 51).
6 Press the power switch to turn on the module.
The power switch stays pressed in and a green indicator lamp in the power switch is on
NOTE when the module is turned on. When the line power switch stands out and the green light
is off, the module is turned off.
1 Remove the front cover by pressing the snap fasteners on both sides.
10 Purge your system prior to the first use (see “Initial Priming” on
page 54).
7diiaZ"]ZVYVhhZbWan
HdakZciXVW^cZi
9Z\VhhZg
8]VccZa6& 8]VccZa6'
EgZhhjgZhZchdg HiVi^Xb^mZg
Ejg\ZkVakZ
Ejbe]ZVY6 Ejbe]ZVY7
6Xi^kZ^caZikVakZ6 6Xi^kZ^caZikVakZ7
AZV`YgV^c
LVhiZijW^c\
DjiaZiXVe^aaVgnidVjidhVbeaZg 8]VccZa7'
8]VccZa7& HdakZcihZaZXi^dckVakZ
1 Remove the front cover by pressing the snap fasteners on both sides.
8 Purge your system before first use (see “Initial Priming” on page 54).
7diiaZ"]ZVYVhhZbWan
HdakZciXVW^cZi
9Z\VhhZg
EgZhhjgZhZchdg
HiVi^Xb^mZg
Ejg\ZkVakZ
Ejbe]ZVY6 Ejbe]ZVY7
6YVeiZg
6Xi^kZ^caZikVakZ6 6Xi^kZ^caZikVakZ7
LVhiZijW^c\
LVhiZdjiaZi
DjiaZiXVe^aaVgnidVjidhVbeaZg
Initial Priming
When Before a degasser or solvent tubing can be used, it is necessary to prime the system. Isopropanol is
recommended as priming solvent due to its miscibility with nearly all HPLC solvents and its
excellent wetting properties.
Preparations Connect all modules hydraulically as described in the respective module manuals.
Fill each solvent bottle with 100 mL isopropanol
Switch the system on
The purge tool of the Lab Advisor can be used to purge the pump automatically.
NOTE
If the pump is not able to draw in the solvent from the bottles, use a syringe to move the
NOTE solvent manually through tubing and degasser.
When priming the vacuum degasser with a syringe, the solvent is drawn through the
NOTE degasser tubes very quickly. The solvent at the degasser outlet will therefore not be fully
degassed. Pump for approximately 10 minutes at your desired flow rate before starting an
analysis. This will allow the vacuum degasser to properly degas the solvent in the degasser
tubes.
Regular Priming
When When the pumping system has been turned off for a certain time (for example, overnight) air will
rediffuse into the solvent channel between the vacuum degasser and the pump. If solvents
containing volatile components are left in the degasser without flow for a prolonged period, there
will be a slight loss of the volatile components.
The purge tool of the Lab Advisor can be used for automatically purging the pump.
NOTE
Changing Solvents
When When the solvent of a channel is to be replaced by another solvent that is not compatible (solvents
are immiscible or one solvent contains a buffer), it is necessary to follow the procedure below to
prevent clogging of the pump by salt precipitation or residual liquid droplets in parts of the system.
To clean the system when HPLC grade water Best solvent to re-dissolve
using buffers buffer crystals
After changing aqueous HPLC grade water Best solvent to re-dissolve
solvents buffer crystals
4
Using the Pump
Leak and Waste Handling 60
Hints for Successful Use of the Binary Pump 61
Setting up the Pump with the G4208A Instant Pilot 62
Setting up the Agilent 1260 Infinity Binary Pump G1312B with the
Instrument Control Interface 63
Overview 63
Setup of Basic Pump Parameters 64
Pump Control 65
Auxiliary Pump Parameters 66
Data Curves 67
Bottle Filling 67
Solvent Information 69
Algae Growth in HPLC Systems 74
How to Prevent and/or Reduce the Algae Problem 75
Prevent Blocking of Solvent Filters 76
Checking the Solvent Filters 77
Cleaning the Solvent Filters 77
Agilent Technologies 59
4 Using the Pump
Leak and Waste Handling
• Place solvent cabinet with the solvent bottles always on top (or at a
higher level) of the pump.
• When using the binary pump without vacuum degasser, shortly degas
your solvents by putting the solvent to a suitable container and
applying a vacuum pressure for some time. If possible apply solvent
conditions that will decrease the gas solubility over time (for example,
warming up the solvents).
• The use of a vacuum degasser is mandatory for flow rates below
0.5 mL/min and for configurations without damper and mixer.
• When using the binary pump with vacuum degasser, flush the degasser
with at least 5 mL per channel before operating the pump, especially
when the pumping system had been turned off for a certain length of
time (for example, overnight) and volatile solvent mixtures are used in
the channels (see “Regular Priming” on page 56).
• Prevent blocking of solvent inlet filters (never use the pump without
solvent inlet filters). Growth of algae should be avoided (see “Prevent
Blocking of Solvent Filters” on page 76).
• Check purge valve frit and column frit in regular time intervals. A
blocked purge valve frit can be identified by black, yellow or greenish
layers on its surface or by a pressure greater than 10 bar in low delay
volume configuration and 20 bar in standard configuration when
pumping distilled water at a rate of 5 mL/min with an open purge
valve.
• Whenever possible use a minimum flow rate of 5 µL/min per solvent
channel to avoid crossflow of solvent into the unused pump channel.
• Whenever exchanging the pump seals, the purge valve frit should be
exchanged, too.
• When using buffer solutions, flush the system with water before
switching it off. The seal wash option should be used when buffer
solutions with concentrations of 0.1 M or higher are being pumped for
long periods of time.
• Check the pump pistons for scratches, grooves and dents when
changing the piston seals. Damaged pistons cause micro leaks and will
decrease the lifetime of the seals.
• After changing the piston seals, apply the seal wear- in procedure (see
“Seal Wear- in Procedure” on page 169).
• Place the aqueous solvent on channel A and the organic solvent on
channel B. The default compressibility settings are set accordingly.
Setting up the Agilent 1260 Infinity Binary Pump G1312B with the
Instrument Control Interface
Overview
Parameters described in following sections is offered by the instrument
control interface and can usually be accessed through Agilent instrument
control software. For details, please refer to manuals and online help of
respective user interfaces.
• Flow 0.001 – 5 mL/min Total flow rate of the pump. See “When to Remove Damper and Mixer” on
page 84 for pump hardware modifications to achieve lowest delay volume.
• Stop Time 0.01 min - no limit The stop time of the pump usually controls the run time of the whole LC system.
Use no limit to stop the run manually (useful for method development).
• Post Time off - 99999 min Time between the end of a run and the start of the next. Used for column
equilibration after a gradient.
• Pressure Max: 0 – 600 bar Max must be bigger than Min! Set max pressure to the maximum operating
Limits Min: 0 – 600 bar pressure of your column. A min pressure setting of e.g. 10 bar will turn off your
pump automatically when running out of solvent. A smarter way, however, is to
use the bottle fillings function (see “Bottle Filling” on page 67).
• Solvent A 0 – 100 % Although channel A can be set to 0 %, it cannot be turned off. This channel should
be used for the aqueous phase (water).
• Solvent B off - 100 % The percentage of channel B is automatically complemented by channel A to give
100 %.
• Solvent H2O, ACN, MeOH, IPA Select the solvent you are using in the respective solvent channel from the
type drop-down list. In case your solvent is not listed, perform a solvent compressibility
calibration (see “Running the Solvent Compressibility Calibration” on page 134.
For details on solvent compressibility see “Binary Pump Solvent Compressibility
Calibration” on page 133.
• Solvent Free text field for a description of the solvent. This description will show up in
Comment method printouts, etc.
• Timetable max. number of lines Use the timetable to build solvent gradients, flow gradients, or combinations of
depends on free both. Gradients are always linear. Use multiple timetable entries to mimic
space in pump exponential or parabolic gradients.
memory
• Display There are three ways to display the timetable:
• in tabular form
• as flow/pressure graph
• as solvent percentage plot
Values can only be changed in tabular view.
Pump Control
The pump can be switched between following states: On, Off or to Standby.
In Standby, the pump motor is still controlled. When the pump is switched
on from standby, it does not re- initialize.
Upon initialization, the pump ignores the Maximum Flow Gradient value.
CAUTION
This can result in a rapid and uncontrolled pressure increase.
➔ To prevent harm to the column, open the purge valve until the initialization is
finished.
The optional seal wash pump can be controlled by either switching it off,
using it for a single time or specifying frequency and duration of periodic
wash intervals.
Upon initialization, the pump ignores the Maximum Flow Gradient value.
CAUTION
This can result in a rapid and uncontrolled pressure increase.
➔ To prevent harm to the column, open the purge valve until the initialization is
finished.
• Maximum Flow 0.1 – 100 mL/min² With this parameter flow rate changes can be ramped up and down slowly to
Gradient default: avoid pressure shocks to the column. The default value is 100 mL/min² which
100 mL/min² in fact turns the function off.
• Minimum 20 – 100 µL The volume one pump piston delivers per stroke. In general, a smaller stroke
Stroke default: Auto volume results in lower pump ripple. The Auto setting adjusts the strokes
dynamically to the lowest possible value.
The strokes can be set individually for pump heads A and B.
• Compressibility 0 - 150·10-6 /bar For best performance, check option Use enhanced compressibility calibration.
bar or enhanced With this option, the pump will use solvent data libraries provided by Agilent or
compressibility data generated by using solvent compressibility calibrations (see “Running the
calibration Solvent Compressibility Calibration” on page 134 ).
default: use For details on solvent compressibility see “Binary Pump Solvent
enhanced comp. Compressibility Calibration” on page 133.
calibration For backward compatibility to 400 bar pumps, the solvent compressibility can
still be set manually for each channel when the box is unticked.
Data Curves
The binary pump provides the possibility to store the following operational
data in the data file of the Agilent data system:
• Solvent percentage for each channel,
• pump flow,
• pressure
The pressure data curve is generated from the pressure sensor readings, while %A, %B and
NOTE flow are calculated from the method settings of the pump.
For details, please refer to the online help or manual of your instrument
control software.
Bottle Filling
The pump offers a powerful feature to monitor the liquid level in the
solvent bottles. With total bottle volume and initial filling volume set
correctly, the pump subtracts the consumed volume continuously from the
initial value and stops the pump and method/sequence execution before
the system runs dry or an analysis is corrupted.
The bottle filling feature fails if multiple solvent inlets are put into one solvent bottle!
CAUTION
➔ In that case implement a minimum pressure limit (see Table 4 on page 64) to avoid
that the pump runs dry when solvents are empty.
• Total 0 – 1000 L This is the capacity (maximum possible volume) in liter of the solvent bottle. In
Volume default: 0 L combination with the actual volume, this parameter is used for calculating and
displaying the relative liquid level.
• Actual 0 – 1000 L After filling the solvent bottles, enter the actual volumes into these boxes. The
Volume default: 0 L Actual Volume must not be larger than the Total Volume of the bottle.
• Prevent default: unchecked If this option is checked, the pump won’t start a new run if the solvent level in
analysis....... one or more bottles is below the minimum volume. Enter a minimum volume in
liter, which considers the position of the solvent inlet and size/shape of the
solvent bottle such that no air is drawn if the actual volume gets close to this
limit.
• Turn pump default: unchecked If this option is checked, the pump will turn off before air is aspirated. However,
off... the residual solvent volume has been calculated for 1 L solvent bottles and may
be too small for large bottles or other vessels.
Solvent Information
Part Materials
Material Information
Materials in the flow path are carefully selected based on Agilent’s
experiences in developing highest quality instruments for HPLC analysis
over several decades. These materials exhibit excellent robustness under
typical HPLC conditions. For any special conditions, please consult the
material information section or contact Agilent.
Disclaimer
Subsequent data were collected from external resources and are meant as
a reference. Agilent cannot guarantee the correctness and completeness of
such information. Data is based on compatibility libraries, which are not
specific for estimating the long- term life time under specific but highly
variable conditions of UHPLC systems, solvents, solvent mixtures and
samples. Information can also not be generalized due to catalytic effects of
impurities like metal ions, complexing agents, oxygen etc. Apart from pure
chemical corrosion, other effects like electro corrosion, electrostatic
charging (especially for non- conductive organic solvents), swelling of
polymer parts etc. need to be considered. Most data available refers to
room temperature (typically 20 – 25 °C, 68 – 77 °F). If corrosion is
possible, it usually accelerates at higher temperatures. If in doubt, please
consult technical literature on chemical compatibility of materials.
PEEK
PEEK (Polyether- Ether Ketones) combines excellent properties regarding
biocompatibility, chemical resistance, mechanical and thermal stability.
PEEK is therefore the material of choice for UHPLC and biochemical
instrumentation.
It is stable in a wide pH range, and inert to many common solvents.
There is still a number of known incompatibilities with chemicals such as
chloroform, methylene chloride, THF, DMSO, strong acids (nitric acid >
10 %, sulphuric acid > 10 %, sulfonic acids, trichloroacetic acid), halogenes
or aequous halogene solutions, phenol and derivatives (cresols, salicylic
acid etc.).
Polyimide
Agilent uses semi- crystalline polyimide for rotor seals in valves and needle
seats in autosamplers. One supplier of polyimide is DuPont, which brands
polyimide as Vespel, which is also used by Agilent.
Polyimide is stable in a pH range between 1 and 10 and in most organic
solvents. It is incompatible with concentrated mineral acids (e.g. sulphuric
acid), glacial acetic acid, DMSO and THF. It is also degraded by
nucleophilic substances like ammonia (e.g. ammonium salts in basic
conditions) or acetates.
Polyethylene (PE)
Agilent uses UHMW (ultra- high molecular weight)- PE/PTFE blends for
yellow piston and wash seals, which are used in 1290 Infinity pumps and
for normal phase applications in 1260 Infinity pumps.
Polyethylene has a good stability for most common inorganic solvents
including acids and bases in a pH range of 1 to 12.5 . It is compatible to
many organic solvents used in chromatographic systems like methanol,
acetonitrile and isopropanol. It has limited stability with aliphatic,
aromatic and halogenated hydrocarbons, THF, phenol and derivatives,
concentrated acids and bases. For normal phase applications, the
maximum pressure should be limited to 200 bar.
Tantalum (Ta)
Tantalum is inert to most common HPLC solvents and almost all acids
except fluoric acid and acids with free sulfur trioxide. It can be corroded
by strong bases (e.g. hydroxide solutions > 10 %, diethylamine). It is not
recommended for the use with fluoric acid and fluorides.
• High concentrations of inorganic acids like nitric acid, sulfuric acid and
organic solvents especially at higher temperatures (replace, if your
chromatography method allows, by phosphoric acid or phosphate buffer
which are less corrosive against stainless steel).
• Halogenated solvents or mixtures which form radicals and/or acids, for
example:
2 CHCl3 + O2→ 2 COCl2 + 2 HCl
This reaction, in which stainless steel probably acts as a catalyst,
occurs quickly with dried chloroform if the drying process removes the
stabilizing alcohol.
• Chromatographic grade ethers, which can contain peroxides (for
example, THF, dioxane, di- isopropylether). Such ethers should be
filtered through dry aluminium oxide which adsorbs the peroxides.
• Solutions of organic acids (acetic acid, formic acid, and so on) in
organic solvents. For example, a 1 % solution of acetic acid in methanol
will attack steel.
• Solutions containing strong complexing agents (for example, EDTA,
ethylene diamine tetra- acetic acid).
• Mixtures of carbon tetrachloride with 2- propanol or THF.
Gold
Gold is inert to all common HPLC solvents, acids and bases within the
specified pH range. It can be corroded by complexing cyanides and
concentrated acids like aqua regia.
Platinum/Iridium
Platinum/Iridium is inert to almost all common acids, bases and solvents.
There are no documented incompatibilities for HPLC applications.
5
Optimizing Performance
When to Use a Vacuum Degasser 80
Operational Hints for the Vacuum Degasser 80
When to Use the Active Seal Wash Option 81
When to Use Alternative Seals 82
When to Use the Low Volume Mixer 83
When to Remove Damper and Mixer 84
Convert the Binary Pump to Low Delay Volume Mode 85
How to Optimize the Compressibility Compensation Setting 87
Solvent Compressibility Calibration 87
Optimization of Legacy Compressibility Settings 88
Agilent Technologies 79
5 Optimizing Performance
When to Use a Vacuum Degasser
Concentrated buffer solutions will reduce the lifetime of the seals and
pistons in your binary pump. The active seal wash option allows to
maintain the seal lifetime by flushing the low pressure side of the seals
with a wash solvent.
The seal wash option is strongly recommended if buffer concentrations of
0.1 M or higher are used regularly with the pump.
The active seal wash option kit can be ordered by quoting Active Seal
Wash Upgrade Product including Service (G1399A).
The seal wash option comprises a peristaltic pump, secondary seals,
gaskets, seal holders and tubing for both pump heads. A bottle of
premixed water/isopropanol (90 /10 vol%) is placed in the solvent cabinet
and connected to the peristaltic pump as described in the technical note
that comes with the active seal wash kit.
Always use a mixture of HPLC- grade water (90 %) and isopropanol (10 %)
as wash solvent. This mixture prevents bacteria growth in the wash bottle
and reduces the surface tension of the water.
In order to avoid accumulation of buffer salts or impurities, regularly replace the washing
NOTE solution using fresh solvents.
The operation of the peristaltic pump can be controlled from the data
system or the Instant Pilot.
For adding a seal- wash option, please contact your local Agilent
Technologies service representative.
The standard seals for the binary pump can be used for most applications.
However, normal phase applications (for example, hexane) are not
compatible with the standard seals. They cause extremely high abrasion
and significantly shorten seal life time.
For the use with normal phase applications special polyethylene pistons
seals (yellow color, PE seals (pack of 2) (0905- 1420)) are available. These
seals have less abrasion compared to the standard seals.
The seal wear-in procedure causes problems to the normal phase seals (yellow).
WA R N I N G
They will be destroyed by the procedure.
➔ DO NOT apply the seal wear-in procedure performed to normal phase seals.
Polyethylene seals have a limited pressure range of 0–200 bar. When used above 200 bar,
NOTE their lifetime will be significantly reduced.
The Low volume mixer ( 200 µL) (5067- 1565) is designed for use with the
1260 Infinity Binary LC System in low delay volume mode. This
configuration is typically used for 2.1 mm i.d., 1.8 µm particle size
columns, where emphasis is put on S/N ratio. The low volume mixer helps
mixing gradients starting with a low concentration of organic solvents,
which can cause noise on the baseline. The maximum benefit of the mixer
is achieved using the mixer together with FW revisions A.06.06 or higher.
The low volume mixer is installed with the corresponding holder (Low
Volume Mixer Holder (G1312- 00003)).
HiVcYVgYXdc[^\jgVi^dc AdlYZaVnkdajbZXdc[^\jgVi^dc
Preparations Flush the system (water if buffers were used, otherwise isopropanol).
Turn the flow off.
1 Remove the front cover by pressing the clip fastener on 2 Use the 1/4 inch hex driver to remove fitting B from port
both sides of the cover. 2 of the pressure sensor.
'
3 Fold capillary end B away. It remains 4 Connect fitting A to port 2 of the pressure sensor. Seal
unconnected.Disconnect fitting A from outlet 1 of the port 1 of the mixer with a plastic blank nut.
mixer.
6 6
7 7
& &
' '
Acetonitrile 115
Benzene 95
Chloroform 100
Cyclohexane 118
Ethanol 114
Heptane 120
Hexane 150
Isobutanol 100
Isopropanol 100
Methanol 120
1-Propanol 100
Toluene 87
Water 46
6
Troubleshooting and Diagnostics
Overview of the Module’s Indicators and Test Functions 92
Status Indicators 94
Power Supply Indicator 94
Module Status Indicator 95
User Interfaces 96
Agilent Lab Advisor Software 97
Agilent Technologies 91
6 Troubleshooting and Diagnostics
Overview of the Module’s Indicators and Test Functions
Status Indicators
The module is provided with two status indicators which indicate the
operational state (prerun, run, and error states) of the module. The status
indicators provide a quick visual check of the operation of the module.
Error Messages
In the event of an electronic, mechanical or hydraulic failure, the module
generates an error message in the user interface. For each message, a
short description of the failure, a list of probable causes of the problem,
and a list of suggested actions to fix the problem are provided (see
chapter Error Information).
Test Functions
A series of test functions are available for troubleshooting and operational
verification after exchanging internal components (see Tests and
Calibrations).
Pressure Test
The Pressure Test is a quick test designed to determine the pressure
tightness of the system (i.e. the high pressure flow path between pump
and column). After exchanging flow path components (e.g. pump seals or
injection seal), use this test to verify the system is pressure tight, see
“(System) Pressure Test” on page 126.
Diagnostic Signals
The pump has several signals (pressure, voltages and piston movement)
that can be used for diagnosing pressure stability, composition and flow
problems (see chapter Diagnostic Signals).
Status Indicators
Two status indicators are located on the front of the module. The lower
left indicates the power supply status, the upper right indicates the
module status.
HiVijh>cY^XVidg
EdlZghl^iX]
HZg^VacjbWZg
User Interfaces
7
Error Information
What Are Error Messages 101
General Error Messages 102
Timeout 102
Shutdown 103
Remote Timeout 104
Lost CAN Partner 104
Leak 105
Leak Sensor Open 105
Leak Sensor Short 106
Compensation Sensor Open 106
Compensation Sensor Short 107
Fan Failed 107
Open Cover 108
Module Error Messages 109
Solvent Zero Counter 109
Pressure Above Upper Limit 109
Pressure Below Lower Limit 110
Pressure Signal Missing 110
Valve Failed 111
Missing Pressure Reading 112
Wrong Pump Configuration 112
Electronic Fuse of SSV Open 113
AIV Fuse 113
Temperature Out of Range 114
Temperature Limit Exceeded 115
Motor-Drive Power 116
Encoder Missing 117
Inlet-Valve Missing 117
Agilent Technologies 99
7 Error Information
Agilent Lab Advisor Software
Timeout
Error ID: 0062
The timeout threshold was exceeded.
1 The analysis was completed successfully, Check the logbook for the occurrence and
and the timeout function switched off the source of a not-ready condition. Restart the
module as requested. analysis where required.
2 A not-ready condition was present during a Check the logbook for the occurrence and
sequence or multiple-injection run for a source of a not-ready condition. Restart the
period longer than the timeout threshold. analysis where required.
Shutdown
Error ID: 0063
An external instrument has generated a shutdown signal on the remote
line.
The module continually monitors the remote input connectors for status
signals. A LOW signal input on pin 4 of the remote connector generates
the error message.
1 Leak detected in another module with a Fix the leak in the external instrument before
CAN connection to the system. restarting the module.
2 Leak detected in an external instrument Fix the leak in the external instrument before
with a remote connection to the system. restarting the module.
4 The degasser failed to generate sufficient Check the vacuum degasser for an error
vacuum for solvent degassing. condition. Refer to the Service Manual for the
degasser or the 1260 pump that has the
degasser built-in.
Remote Timeout
Error ID: 0070
A not- ready condition is still present on the remote input. When an
analysis is started, the system expects all not- ready conditions (for
example, a not- ready condition during detector balance) to switch to run
conditions within one minute of starting the analysis. If a not- ready
condition is still present on the remote line after one minute the error
message is generated.
3 Defective components in the instrument Check the instrument for defects (refer to the
showing the not-ready condition. instrument’s documentation).
3 Defective main board in another module. Switch off the system. Restart the system, and
determine which module or modules are not
recognized by the system.
Leak
Error ID: 0064
A leak was detected in the module.
The signals from the two temperature sensors (leak sensor and
board- mounted temperature- compensation sensor) are used by the leak
algorithm to determine whether a leak is present. When a leak occurs, the
leak sensor is cooled by the solvent. This changes the resistance of the
leak sensor which is sensed by the leak- sensor circuit on the main board.
3 Loose or leaking purge valve, active inlet Ensure pump components are seated correctly.
valve, or outlet valve. If there are still signs of a leak, exchange the
appropriate seal (purge valve, active inlet valve,
outlet valve).
4 Defective pump seals. Exchange the pump seals.
2 Leak sensor incorrectly routed, being Please contact your Agilent service
pinched by a metal component. representative.
Fan Failed
Error ID: 0068
The cooling fan in the module has failed.
The hall sensor on the fan shaft is used by the main board to monitor the
fan speed. If the fan speed falls below a certain limit for a certain length
of time, the error message is generated.
This limit is given by 2 revolutions/second for longer than 5 seconds.
Depending on the module, assemblies (e.g. the lamp in the detector) are
turned off to assure that the module does not overheat inside.
Open Cover
Error ID: 0205
The top foam has been removed.
The sensor on the main board detects when the top foam is in place. If
the foam is removed, the fan is switched off, and the error message is
generated.
1 The top foam was removed during Please contact your Agilent service
operation. representative.
2 Foam not activating the sensor. Please contact your Agilent service
representative.
4 Rear of the module is exposed to strong Ensure that the rear of module is not directly
direct sunlight. exposed to strong sunlight.
1 Volume in bottle below specified volume. Refill bottles and reset solvent counters.
2 Incorrect setting. Make sure the set solvent volume matches the
actual bottle filling and set the shutoff limit to a
reasonable value (e.g. 100 mL for 1 L bottles)
1 Upper pressure limit set too low. Ensure the upper pressure limit is set to a value
suitable for the analysis.
2 Blockage in the flowpath (after the damper). Check for blockage in the flowpath. The
following components are particularly subject
to blockage: inline filter frit, needle
(autosampler), seat capillary (autosampler),
sample loop (autosampler), column frits and
capillaries with small internal diameters (e.g.
50 µm ID).
Valve Failed
Error ID: 2040
Valve 0 Failed: valve A1
Valve 1 Failed: valve A2
Valve 2 Failed: valve B2
Valve 3 Failed: valve B1
One of the solvent selection valves in the module failed to switch
correctly.
The processor monitors the valve voltage before and after each switching
cycle. If the voltages are outside expected limits, the error message is
generated.
2 Connection cable (inside instrument) not Please contact your Agilent service
connected. representative.
1 Active-inlet valve and pump encoder of Please contact your Agilent service
channel B disconnected. representative.
1 Defective solvent selection valve. Restart the pump. If the error message appears
again, exchange the solvent selection valve.
2 Defective connection cable (front panel to Please contact your Agilent service
main board). representative.
AIV Fuse
Error ID: 2044
Inlet- Valve Fuse 0: Pump channel A
Inlet- Valve Fuse 1: Pump channel B
One of the active- inlet valves in the module has drawn excessive current
causing the inlet- valve electronic fuse to open.
1 Defective active inlet valve. Restart the module. If the error message
appears again, exchange the active inlet valve.
2 Defective connection cable (front panel to Please contact your Agilent service
main board). representative.
1 High friction (partial mechanical blockage) Remove the pump-head assembly. Ensure there
in the pump drive assembly. is no mechanical blockage of the pump-head
assembly or pump drive assembly.
2 Partial blockage of the flowpath in front of Ensure the outlet valve is not blocked.
the damper.
Motor-Drive Power
Error ID: 2041, 2042
Motor- Drive Power: Pump channel A
B: Motor- Drive Power: Pump channel B
The current drawn by the pump motor exceeded the maximum limit.
Blockages in the flow path are usually detected by the pressure sensor in
the damper, which result in the pump switching off when the upper
pressure limit is exceeded. If a blockage occurs before the damper, the
pressure increase cannot be detected by the pressure sensor and the
module will continue to pump. As pressure increases, the pump drive
draws more current. When the current reaches the maximum limit, the
module is switched off, and the error message is generated.
1 Flow path blockage in front of the damper. Ensure the capillaries and frits between the
pump head and damper inlet are free from
blockage.
4 High friction (partial mechanical blockage) Remove the pump-head assembly. Ensure there
in the pump drive assembly. is no mechanical blockage of the pump-head
assembly or pump drive assembly.
Encoder Missing
Error ID: 2046, 2050, 2510
Encoder Missing: Pump channel A
B: Encoder Missing: Pump channel B
The optical encoder on the pump motor in the module is missing or
defective.
The processor checks the presence of the pump encoder connector every
2 s. If the connector is not detected by the processor, the error message is
generated.
Inlet-Valve Missing
Error ID: 2048, 2052
Inlet- Valve Missing: Pump channel A
B: Inlet- Valve Missing: Pump channel B
The active- inlet valve in the module is missing or defective.
The processor checks the presence of the active- inlet valve connector
every 2 s. If the connector is not detected by the processor, the error
message is generated.
1 Disconnected or defective cable. Ensure the pins of the active inlet valve
connector are not damaged. Ensure the
connector is seated securely.
3 Mechanical blockage of the module. Remove the pump-head assembly. Ensure there
is no mechanical blockage of the pump-head
assembly or pump drive assembly.
1 Pump head not installed correctly (screws Install the pump head correctly. Ensure nothing
not secured, or pump head not seated (e.g. capillary) is trapped between the pump
correctly). head and body.
Index Limit
Error ID: 2203, 2213
Index Limit: Pump channel A
B: Index Limit: Pump channel B
The time required by the piston to reach the encoder index position was
too short (pump).
During initialization, the first piston is moved to the mechanical stop.
After reaching the mechanical stop, the piston reverses direction until the
encoder index position is reached. If the index position is reached too
fast, the error message is generated.
1 Irregular or sticking drive movement. Remove the pump head, and examine the seals,
pistons, and internal components for signs of
wear, contamination or damage. Exchange
components as required.
Index Adjustment
Error ID: 2204, 2214
Index Adjustment: Pump channel A
B: Index Adjustment: Pump channel B
The encoder index position in the module is out of adjustment.
During initialization, the first piston is moved to the mechanical stop.
After reaching the mechanical stop, the piston reverses direction until the
encoder index position is reached. If the time to reach the index position
is too long, the error message is generated.
1 Irregular or sticking drive movement. Remove the pump head, and examine the seals,
pistons, and internal components for signs of
wear, contamination or damage. Exchange
components as required.
Index Missing
Error ID: 2205, 2215, 2505
Index Missing: Pump channel A
B: Index Missing: Pump channel B
The encoder index position in the module was not found during
initialization.
During initialization, the first piston is moved to the mechanical stop.
After reaching the mechanical stop, the piston reverses direction until the
encoder index position is reached. If the index position is not recognized
within a defined time, the error message is generated.
Stroke Length
Error ID: 2206, 2216
Stroke Length: Pump channel A
B: Stroke Length: Pump channel B
The distance between the lower piston position and the upper mechanical
stop is out of limits (pump).
During initialization, the module monitors the drive current. If the piston
reaches the upper mechanical stop position before expected, the motor
current increases as the module attempts to drive the piston beyond the
mechanical stop. This current increase causes the error message to be
generated.
Initialization Failed
Error ID: 2207, 2217
Initialization Failed: Pump channel A
B: Initialization Failed: Pump channel B
The module failed to initialize successfully within the maximum time
window.
A maximum time is assigned for the complete pump- initialization cycle. If
the time is exceeded before initialization is complete, the error message is
generated.
8
Test Functions and Calibration
(System) Pressure Test 126
Positioning the Blank Nut 127
Running the (System) Pressure Test 128
Evaluating the Results 129
Valve Test 131
Running the Valve Test 132
Evaluating the Results 132
Binary Pump Solvent Compressibility Calibration 133
Running the Solvent Compressibility Calibration 134
Pump Elasticity Calibration 135
Running the Pump Elasticity Calibration 136
Leak Rate Test 137
Running the Test 139
Evaluating the Results 140
Potential Causes of Leak Rate Test Failure 141
This chapter explains all test functions that are available for the binary pump.
This Lab Advisor test uses different names depending on the firmware revision used:
NOTE
• FW revision ≥ A.06.50: System Pressure Test
• FW revision < A.05.60: Pressure Test
Description
The (system) pressure test is a quick built- in test designed to demonstrate
the leak tightness of the system. The test involves monitoring the flow
profile while the pump delivers against a blank nut. The result is
presented as the leak rate of the module and provides information about
the leak tightness of the system between the outlet valves of the pump
and the blank nut.
The blank nut can be positioned anywhere between the purge valve of the pump and the
NOTE detector inlet to pressure test the desired part of the system.
Step 1
The test begins with the initialization of both pump heads. After
initialization, the pump is starting the compression phase and the required
flow rate is constantly monitored and adjusted. The pump continues to
pump until a system pressure of around 600 bar is reached.
Step 2
When the system pressure reaches 600 bar, the pump continues to pump
at a flow rate that keeps the pressure constant. The flow that is needed to
keep the pressure constant is directly translated into a leak rate.
When The test should be used when problems with small leaks are suspected, or after maintenance of
flow path components (e.g., pump seals, injection seal) to prove pressure tightness up to 600 bar
Preparations Place two bottles of HPLC-grade water in channels A and B (A1 and B1 if the pump is equipped with
a solvent selection valve)
Make absolutely sure that all parts of the flow path that are part of the test are very
NOTE thoroughly flushed with water before starting to pressurize the system! Any trace of other
solvents or the smallest air bubble inside the flow path definitely will cause the test to fail!
1 Select the (system) pressure test from the test selection menu.
2 Start the test and follow the instructions.
Make sure to release the pressure by opening the purge valve when the test has
NOTE completed. Otherwise the pump may generate an overpressure error.
Please notice the difference between an error in the test and a failure of the test! An error
NOTE is caused by the abnormal termination during the operation of the test whereas a failure of
a test indicates that the test results were not within the specified limits.
Often it is only a damaged blank nut itself (poorly shaped from overtightening) that causes
NOTE the test to fail. Before investigating on any other possible sources of failure make sure that
the blank nut you are using is in good condition and properly tightened!
• If the test fails again, insert the blank nut at the outlet of the previous
module in the stack (e.g. autosampler, port 6 of the injection valve), and
repeat the pressure test. Exclude each module one by one to determine
which module is leaking.
• If the pump is determined to be the source of the leak, run the valve
test to identify the defective pump component.
Damaged pump seals or pistons. Run the valve test to identify the defective
component.
Loose purge valve. Tighten the purge valve nut (14 mm wrench).
Damaged metering seal or piston. Exchange the metering seal. Check the piston
for scratches. Exchange the piston if required.
Valve Test
Description
The Valve Test provides a fast and accurate way to verify proper hydraulic
operation of the binary pump. Problems that are related to defective
valves, seals or pistons can be diagnosed and usually the defective part is
identified.
If a minimum firmware of A.06.50 is installed on this module and a minimum Lab Advisor
NOTE revision of B.02.03. is available, please run the improved Leak Rate Test instead (see “Leak
Rate Test” on page 137).
Step 1
The system is setup with water on both channels and a restriction
capillary is attached to the outlet of the pump. Pump head A is delivering
at 1 mL/min. The pressure signal is monitored and overlaid with the
piston movement plot. The pressure pattern and the slope of the pressure
signal are evaluated for the delivery strokes of both pistons.
Step 2
The procedure from step 1 is repeated on pump head B.
Step 3
The data from step 1 and 2 are evaluated. In case test failed, a conclusion
about the defective part is made.
Preparations Place two bottles of HPLC-grade water in channels A and B (A1 and B1 if the pump is equipped with
a solvent selection valve)
The Valve Test should only be performed when firmware revisions < A.06.50 are installed.
NOTE
For firmware revision A.06.50 or above, perform the Leak Rate Test instead.
Make absolutely sure that the pump is very thoroughly flushed with water before starting
NOTE the test! Any trace of other solvents or the smallest air bubble inside the flow path
definitely will cause the test to generate misleading results!
Make sure to release the pressure by opening the purge valve when the test has
NOTE completed. Otherwise the pump may generate an overpressure error.
Description
Each solvent or solvent mixture has unique compressibility at different
pressures. In order to deliver accurate flow with minimal pressure- and
composition ripple over the full operational pressure range, it is necessary
that the pump compensates precisely for the compressibility of the
solvents in use.
The binary pump comes with compressibility parameters for the most
common HPLC solvents and solvent mixtures. If a solvent is not available
in the list of pre- calibrated solvents, the solvent compressibility calibration
allows the appropriate compressibility data to be generated.
Technical background
The solvent compressibility calibration relies on an accurate elasticity
calibration of the pump. With a proper elasticity calibration in place, the
pump is switched into pressure control mode. A restriction capillary is
connected to the purge valve outlet. By varying the flow rate, the pump
maintains a certain pressure. The pump optimizes the compressibility
value of the solvent until the lowest possible pump ripple is reached. The
pump increases the flow rate and adjusts the pressure to the next
calibration step where the pump ripple is minimized again. This process is
repeated until solvent compressibility data for the whole operating
pressure range of the pump are available.
The compressibility data set for this solvent is stored in an XML- file in
C:\Documents and Settings\<username>\Application Data\Agilent Technologies\Agilent
Lab Advisor\2.02.0.0\data\. It can be shared with other G1312B pumps via
the controlling data system.
When If a solvent is not available in the list of pre-calibrated solvents, the solvent compressibility
calibration allows to generate appropriate compressibility data.
Preparations Place a bottles with solvent to be calibrated in channel A (resp. A1 if a solvent selection valve is
installed).
Make absolutely sure that the pump is very thoroughly flushed with the solvent to be
NOTE calibrated before starting the procedure! Any trace of other solvents or the smallest air
bubble inside the flow path definitely will cause the calibration to fail!
Make sure to release the pressure by opening the purge valve when the test has
NOTE completed. Otherwise the pump may generate an overpressure error.
Description
The flow path components of the binary pump have an inherent and
pressure dependent elasticity which differs from pump to pump. When
this elasticity/pressure function is known, a correction algorithm can be
applied. This results in significantly improved pump performance in low
delay volume mode (damper and mixer bypassed).
The pump elasticity calibration uses a solvent with well known properties
(HPLC- grade water) to determine the pump elasticity over the entire
operating pressure range and stores the calibration values in the
non- volatile RAM of the pump mainboard.
The initial calibration is done at the factory. It only needs to be repeated
after replacement of major pump parts (mainboard, pump drive). The test
allows to define which pump head will be calibrated.
Results of the pump elasticity calibration rely on known compressibility parameters for pure
NOTE water. If the water is not HPLC-grade, not well degassed or degasser and pump are not
flushed properly, the pump elasticity calibration will fail. The pump elasticity calibration has
to be performed for each pump head individually.
When The initial calibration is done at the factory. It only needs to be repeated after replacement of major
pump parts (mainboard, pump drive).
Make absolutely sure that the pump is very thoroughly flushed with the solvent to be
NOTE calibrated before starting the procedure! Any trace of other solvents or the smallest air
bubble inside the flow path definitely will cause the calibration to fail!
If a solvent selection valve is installed flush all four solvent channels to avoid that air from a
NOTE dry solvent intake tube is drawn into the flow path upon initialization.
1 Select the pump elasticity calibration from the test selection menu.
2 Start the test and follow the instructions.
Make sure to release the pressure by opening the purge valve when the test has
NOTE completed. Otherwise the pump may generate an overpressure error.
Introduction
The Leak Rate Test is used for verifying the internal tightness of the pump
and helps identifying parts which may have caused a leak.
System requirements
Minimum software revisions:
• Lab Advisor B.02.03
Minimum firmware revisions:
• A.06.50
This test does not work in emulation mode, that is when the G1312B Binary Pump emulates
NOTE the G1312A Binary Pump. In case of an emulated module, convert to the original type first
(see “Replacing Module Firmware” on page 181).
Test Principle
A solvent can be chosen from available solvent channels and a maximum
target pressure can be defined at which the test will be run. Typically,
this is the maximum pressure specified for the pump. The test can be run
with any solvent compatible to the pump.
Before the test, the pump is flushed with solvent in order to remove air
bubbles, as air bubbles are compressed during the test and therefore
would appear as leaks. Using a degasser is highly recommended.
e^hidc'
VigZVgedh^i^dc
e^hidc&
VigZVgedh^i^dc
e^hidc'
YZa^kZgh
e^hidc&
YZa^kZgh
VWdji&%%WVg
WZadliVg\Zi
egZhhjgZ
Initially, the pressure is increased to about 100 bar below the target
pressure, which has been set for the test.
Then piston 1 is brought to its rear position. An increasing flow is
delivered by piston 1. In case of a leak, the pressure will drop initially as
long as the flow rate delivered by the piston is lower than the leak rate.
As soon as the flow rate of the piston exceeds the leak rate, the measured
pressure will increase again. Therefore the minimum pressure of that
curve segment corresponds to the flow and leak rate at that time and the
leak rate is measured.
Subsequently, piston 2 is moved to its rear position, then piston 2 delivers
and the measurement is done as described for piston 1.
For a binary pump (G1312B/C, K1312B), the test is run for both pump
heads for channels A and B.
Make sure to release the pressure by slowly opening the purge valve when the test has
NOTE finished.
9VbeZg
EjbeX]VbWZg& EjbeX]VbWZg'
Ejg\ZkVakZ
DjiaZikVakZ IdXdajbc
>caZikVakZ
IdlVhiZ
HZVa
E^hidc&
E^hidc'
Secondary Leak
If a leak is found for movement of piston 2 (secondary leak), the following
reasons are possible:
6 Leak at piston 2 or seal in chamber 2 Inspect piston, replace piston and/or seal
Primary Leak
If a leak is found for movement of piston 1 (primary leak), any leak
described for piston movement 2 will cause a failure for piston 1 as well,
as the liquid can move through the outlet valve to chamber 2. Such cases
need to be identified as described before. Additionally, following causes
are possible:
2 Leak at inlet valve Replace inlet valve or inlet valve cartridge (AIV only)
9
Diagnostic Signals
Analog Pressure Output 144
Diagnostic Signals in Chromatographic Data Systems 145
Instrument Signals 145
A BNC connector at the rear of the binary pump provides the reading of
the pressure sensor as analog value with a resolution of 1.33 mV/bar. The
maximum reading of 660 bar equals 800 mV. The signal is available in
real time and can be fed into an appropriate recording device (e.g.
integrator or strip chart recorder) for troubleshooting purposes.
Analog pressure
output
Instrument Signals
In Chromatographic Data Systems (for example, Agilent OpenLAB CDS, or
any 3rd party software that supports the ICF (Instrument Control
Framework) standard), the following instrument parameters are accessible
during data acquisition and can be stored in the data file:
• current pump pressure
• solvent composition (gradient)
10
Maintenance
Introduction to Maintenance and Repair 148
Warnings and Cautions 149
Cleaning the Module 151
Overview of Maintenance and Simple Repairs 152
Maintenance Procedures 153
Exchanging the Purge Valve Frit or the Purge Valve 154
Removing the Pump Head Assembly 157
Maintenance of a Pump Head without Seal Wash 159
Maintenance of a Pump Head with Seal Wash 163
Reinstalling the Pump Head Assembly 167
Seal Wear-in Procedure 169
Exchanging the Active Inlet Valve (AIV) or its Cartridge 170
Exchanging the Outlet Valve 173
Installation of the Solvent Selection Valve Upgrade Kit 175
Exchanging the Solvent Selection Valve 177
Exchanging the Optional Interface Board 180
Replacing Module Firmware 181
The pump is designed for easy repair. The most frequent repairs such as
piston seal replacement and purge valve frit exchange can be done from
the front side without removing the pump from the system stack. These
repairs are described in “Overview of Maintenance and Simple Repairs” on
page 152 .
Electrical shock
WA R N I N G
Repair work at the module can lead to personal injuries, e.g. shock hazard, when the
cover is opened.
➔ Do not remove the cover of the module.
➔ Only certified persons are authorized to carry out repairs inside the module.
Agilent 1260 Infinity pump heads have been discontinued for Agilent 1260 Infinity
CAUTION
Binary Pumps
A number of 1260 Infinity Binary Pumps have been shipped with pump heads labeled
as 1260 Infinity pump heads and passive inlet valves. While these parts are as good
from a performance perspective, they are no longer used for 1260 Infinity Binary
Pumps. Parts listed in this manual are not compatible to 1260 Infinity pump heads and
may get damaged.
➔ Please contact your Agilent service representative.
To keep the module case clean, use a soft cloth slightly dampened with
water, or a solution of water and mild detergent.
Liquid dripping into the electronic compartment of your module can cause shock
WA R N I N G
hazard and damage the module
➔ Do not use an excessively damp cloth during cleaning.
➔ Drain all solvent lines before opening any connections in the flow path.
Figure 23 on page 152 shows the main user accessible assemblies of the
binary pump. The pump heads and its parts require normal maintenance
(for example, seal exchange) and can be accessed from the front (simple
repairs). Replacement of valve cartridges or filters don’t require to remove
the pump from the system stack.
Purge valve
Outlet valve
1 Purge valve, see “Exchanging the Purge Valve Frit or the Purge Valve” on page 154
3 Active inlet valve, see “Exchanging the Active Inlet Valve (AIV) or its Cartridge” on
page 170
4 Pump head, see “Removing the Pump Head Assembly” on page 157
5 Solvent selection valve, see “Installation of the Solvent Selection Valve Upgrade Kit” on
page 175
Maintenance Procedures
The procedures described in this section can be done with the binary
pump in place in the system stack.
“Exchanging the Purge Valve Frit or the Yearly, or if the frit shows A pressure drop of > 10 bar in low delay volume
Purge Valve” on page 154 indication of configuration and > 20 bar in standard
contamination or configuration across the frit (5 mL/min H 2O
blockage with purge valve open) indicates blockage
If internally leaking Solvent dripping out of waste outlet when valve
is closed
“Removing the Pump Head Assembly” on During yearly Necessary to get access to pump seals and
page 157 maintenance pistons
“Maintenance of a Pump Head without Seal Yearly, or if pump Leaks at lower pump head side, unstable
Wash” on page 159 performance indicates retention times, pressure ripple unstable — run
seal wear Valve Test for verification
Seal life time shorter than normally expected —
check pistons while changing the seals
“Maintenance of a Pump Head with Seal Yearly, or if pump Only necessary when Seal Wash Option is
Wash” on page 163 performance indicates installed. Leaks at lower pump head side, loss
seal wear of wash solvent
“Exchanging the Active Inlet Valve (AIV) or If leaking externally Error messages “Inlet Valve Fuse” or “Inlet
its Cartridge” on page 170 If solenoid is defective Valve Missing”
“Exchanging the Outlet Valve” on page 173 If internally leaking Pressure ripple unstable, run Valve Test for
verification
When • Frit – when piston seals are exchanged or when contaminated or blocked (pressure drop of
> 10 bar in low delay volume configuration and > 20 bar in standard configuration across the frit
at a flow rate of 5 mL/min of water with purge valve opened)
• Purge valve – if internally leaking
1 Using a 1/4 inch wrench disconnect the pump outlet capillary from the
purge valve. Disconnect the waste tube. Beware of leaking solvents due
to hydrostatic pressure.
EjbedjiaZiXVe^aaVgn
Ejg\ZkVakZ
LVhiZijWZ
2 Using the 14 mm wrench, unscrew the purge valve and remove it from
the purge valve holder.
3 Remove the seal cap from the purge valve.
4 Using a pair of tweezers or a toothpick remove the frit.
KVakZWdYn
EI;:[g^il^i]\gddkZ
HZVaXVe
5 Place a new frit into the purge valve with the orientation of the frit as
shown below (slit in frit points to the front). Reinstall the seal cap
including the gold seal.
Before reinstallation always check the gold seal in the seal cap. A deformed seal cap
NOTE should be exchanged.
6 Insert the purge valve into the purge valve holder and orient the waste
outlet nozzle downward.
7 Tighten the purge valve and reconnect outlet capillary and waste
tubing.
Both pump head assemblies use the same internal components. In addition, pump head A
NOTE is fitted with the purge valve. The following procedure describes the removal and
disassembly of pump head A (left). For pump head B (right) proceed in the same way and
skip steps that deal with the purge valve.
3 Using a 4- mm hexagonal key stepwise loosen and remove the two pump
head screws.
Ejbe]ZVYhXgZlh
Both pump head assemblies use the same internal components. In addition, pump head A
NOTE is fitted with the purge valve. The following procedure describes the removal and
disassembly of pump head A (left). For pump head B (right) proceed in the same way and
skip steps that deal with the purge valve.
1 Place the pump head on a flat surface. Loosen the lock 2 Remove the support rings from the piston housing and lift
screw (two turns) and while holding the lower half of the the housing away from the pistons.
assembly (piston housing) carefully lift up the pump
housing and remove it from the piston housing.
Lock screw
E^hidc]djh^c\
Piston housing
E^hidc
3 Check the piston surface and remove any deposits or 4 Using the steel side of the insert tool carefully remove
layers. Most suitable is polishing of the piston rod with the seal from the pump housing. Remove wear retainers,
toothpaste. Replace the piston if scratched or if dents are if still present.
visible.
Insert tool
Piston surface
Piston seal
Pump housing
NOTE
The best way to inspect a piston is to hold it up and
watch e.g. a light bulb through the piston rod. The
transparent sapphire acts as a very strong magnifier and
even smallest surface abnormalities become visible.
5 Using the plastic side of the insert tool insert the new 6 Place a seal wash gasket in the recess of the support
seals into the pump head and press them firmly in ring. Put the seal holder on top of the gasket.
position.
Gasket
7 Reassemble the pump head assembly. Note the correct 8 Insert the pistons and carefully press them into the seals.
position of the pin on the support ring.
E^hidc
Ejbe]djh^c\
Hjeedgig^c\h
E^hidc]djh^c\
NOTE
Reset the seal wear counter and liquimeter in the
Agilent Lab Advisor.
9 Tighten the lock screw. Next Steps:
10 Reinstall the pump head assembly, see “Reinstalling the
AdX`hXgZl Pump Head Assembly” on page 167.
11 If a standard seal has been installed, run the seal wear-in
procedure, see “Seal Wear-in Procedure” on page 169.
12 For the normal phase seal, the purge valve frit should be
replaced, see “Exchanging the Purge Valve Frit or the
Purge Valve” on page 154.
Both pump head assemblies use the same internal components. In addition, pump head A
NOTE is fitted with the purge valve. The following procedure describes the removal and
disassembly of pump head A (left). For pump head B (right) proceed in the same way and
skip steps that deal with the purge valve.
1 Place the pump head on a flat surface. Loosen the lock 2 Remove the seal holder and the seal wash support rings
screw (two turns) and while holding the lower half of the from the piston housing. Remove the seal holder from the
assembly (piston housing) carefully lift up the pump support ring assembly.
housing and remove it from the piston housing.
AdX`hXgZl
Seal holder
Ejbe]djh^c\
Piston housing
E^hidc]djh^c\
Piston
3 Check the piston surface and remove any deposits or layers. 4 Using the steel side of the insert tool carefully remove
Most suitable is polishing of the piston rod with toothpaste. the seal from the pump housing. Remove wear retainers,
Replace the piston if scratched or if dents are visible. if still present.
Insert tool
Piston surface
Piston seal
Pump housing
NOTE
The best way to inspect a piston is to hold it up and
watch e.g. a light bulb through the piston rod. The
transparent sapphire acts as a very strong magnifier and
even smallest surface abnormalities become visible.
164 1260 Infinity Binary Pump User Manual
Maintenance 10
Maintenance of a Pump Head with Seal Wash
5 Using the plastic side of the insert tool insert the new 6 Using the steel side of the insert tool remove the seal
seals into the pump head and press them firmly in wash gasket and the wash seal from the support ring.
position. The removed seal will be damaged and cannot be
re-used!
>chZgiidda
E^hidchZVah
Wash seal
7 Using the plastic side of the insert tool press the new 8 Place a seal wash gasket in the recess of the support
wash seal (spring pointing upwards) into the recess of ring. Put the seal holder on top of the gasket.
the support ring.
Seal holder
>chZgiidda
Support ring
LVh]hZVa
Hjeedgig^c\
9 Place the support rings on the piston housing (pistons 10 Insert the pistons and carefully press them into the seals.
not installed) and snap the pump housing and piston
housing together. Note the correct position of the pin on
the support ring.
E^hidc
Pump housing
Support ring
Piston housing
11 Tighten the lock screw. 12 Reinstall the pump head assembly, see “Reinstalling the
Pump Head Assembly” on page 167.
AdX`hXgZl
Ejg\ZkVakZ]daYZg
Ejbe]ZVYhXgZlh
Seal damage
CAUTION
This procedure is required for black PTFE seals (standard applications, p/n 5063-6589),
but it will damage the yellow PE seals (normal phase applications, p/n 0905-1420).
➔ Do not run the seal wear-in procedure if PE seals are installed in the pumphead.
1 Put a bottle with 100 ml of isopropanol in the solvent cabinet and place
the solvent intake filter of the pump head you want to wear in into this
bottle.
2 Screw the PEEK adapter 1/4- 28 to 10- 32 (0100- 1847) onto the active
inlet valve and connect the inlet tube from the bottle head directly to it.
3 Connect the Restriction capillary (5022- 2159) to the purge valve.
Connect its other end to a waste container.
4 Open the purge valve and purge the system for 5 min with isopropanol
at a flow rate of 2 mL/min.
5 Close the purge valve, set the flow to a value that gives a pressure of
350 bar. Pump 15 min at this pressure to wear the seals in. The
pressure can be monitored on the analog output connector of the pump,
with the Instant Pilot, chromatographic data system or any other
controlling device connected to your pump.
6 Turn OFF the pump, slowly open the purge valve to release the
pressure from the system, disconnect the restriction capillary and
reconnect the outlet capillary to the purge valve. Reconnect the intake
tubing to the solvent selection valve and the connecting tube from the
solvent selection valve (if installed) to the AIV.
7 Purge your system with the solvent used for your next application.
Binary pumps without solvent selection valve (SSV) have an adapter installed between the
NOTE solvent line and the active inlet valve (AIV). Disconnect the solvent tubes at the adapter
and remove the adapter from the AIV.
4 Using a 14 mm wrench, loosen the active inlet valve and remove the
valve from the pump head.
8dccZXidg
HdakZci^ciV`ZijW^c\
6Xi^kZ^caZikVakZ
5 Using a pair of tweezers, remove the valve cartridge from the defective
active inlet valve.
6Xi^kZ^caZikVakZXVgig^Y\Z
6Xi^kZ^caZikVakZ
7 Screw the new valve into the pump head. With the 14 mm wrench turn
the nut until it is hand tight.
8 Position the valve so that the solvent inlet tube connection points
towards the front.
9 Using the 14 mm wrench tighten the nut by turning the valve in its
final position (not more than a quarter turn). Do not overtighten the
valve.
10 Reconnect the Active Inlet Valve cable to the connector in the Z- panel
and the inlet tube to the valve.
11 Reinstall the front cover.
After an exchange of the valve it may be required to pump several mL of the solvent used in
NOTE the current application before the flow stabilizes at a pressure ripple as low as it used to be
when the system was still working properly.
4 Reinstall the outlet valve and tighten it using a torque wrench (12 Nm).
Absorber capillary
Outlet valve
When Applicable modules: This kit is compatible to the 1260 Infinity Binary Pumps G1312B and G1312C
and to the 1260 Infinity Binary Pump Clinical ed. K1312B.
The figures below show a Binary Pump G1312B. The kit can be used similarly for the Binary
NOTE Pump G1312C and for the Binary Pump Clinical ed. K1312B.
1 Remove the front cover sheet by removing both screws. 2 Plug in the connector of the solvent selection valve
Connector
SSV cable
Cover sheet
Screws
Solvent
selection valve
3 Install the solvent selection assembly by fixing both 4 Connect the outlet for solvent A (upper row) to the inlet
indicated screws. valve of the left pump head. Connect the outlet for
solvent B (lower row) to the inlet valve of the right pump
head. Put the solvent bottles into the solvent cabinet.
Connect the bottle heads of solvents A1 and A2 to the
inlets in the upper row, see labels on valve assembly.
Connect the bottle heads of solvents B1 and B2 to the
inlets in the lower row, see labels on valve assembly.
To solvent
bottles
Inlets
Outlets
When If leaking internally (crossflow between the ports), or if one of the channels is blocked
1 Lift solvent bottles out of the solvent cabinet and place 2 Disconnect all tubings from the solvent selection valve.
them on the table. Disconnect the solvent tubes from the
solvent selection valve and empty the tubes into the
bottles. Place the bottles back into the solvent cabinet.
3 Using a Pozidriv screwdriver #1 loosen the holding 4 Carefully pull the valve holder out and disconnect the
screws of the valve holder. valve cable at the connector.
5 Exchange the defective solvent selection valve. 6 Tighten the screws of the valve holder.
After an exchange of the valve it may be required to pump several mL of solvent before the
NOTE flow stabilizes at a pressure ripple as low as it used to be when the system was still
working properly.
Preparations Read update documentation provided with the Firmware Update Tool.
NOTE
The module G1312B exists both as part of the 1200 Series and 1260 Infinity product lines.
These modules use different solvent selection valves and main boards.
Latest firmware revisions in each set automatically recognize the main board type and
can be used without special handling.
• A.06.06 [020] These revisions recognize the board type and shall be used.
• A.06.10 [020] A.06.06 is the minimum version, which is compatible to 1260
• A.06.34 [001] Infinity Binary Pumps G1312B.
• A.06.53 [002] Older G1312B firmware is only compatible to 1200 Series Binary
Pumps SL G1312B.
All listed firmware revisions allow emulation of G1312A pumps.
Please consult the Firmware Update Guide for 1200 Infinity
Systems for details.
11
Parts and Materials for Maintenance
Hydraulic Path with Solvent Selection Valve 184
Hydraulic Path without Solvent Selection Valve 186
Pump Head Assembly Without Seal Wash 188
Pump Head Assembly with Seal Wash Option 190
Outlet Valve 192
Purge Valve Assembly 193
Active Inlet Valve Assembly 194
HPLC Starter Kit G4201-68707 195
HPLC Starter Kit G4202-68707 195
HPLC System Tool Kit 196
Bottle Head Assembly 197
Solvent Cabinet 198
This chapter lists all parts and tools that are required for maintenance.
13 G1312-87330 Mixer
&%
&* +
,
)
&,
&+ ' (
12 G1312-87330 Mixer
&,
.
&) *
) ,
'
+
(
Figure 26 Hydraulic Path without Solvent Selection Valve, with Active Seal Wash
7 5063-6589 Piston seal PTFE, carbon filled, black (pack of 2), default
9 0515-0175 Mounting screw for manual purge valve holder, M4, 20 mm long
15 G1312-23201 Adapter
&%
&& &
'
&' (
)
&( *
+
,
&*
&+ -
&)V!W
9 5063-6589 Piston seal PTFE, carbon filled, black (pack of 2), default
10 0515-0175 Mounting screw for manual purge valve holder, M4, 20 mm long
17 G1312-23201 Adapter
10
11
2
12
3
4
13 5
6
14 7
8
9
17
18 15
16 a, b
Outlet Valve
p/n Description
3
Figure 30 Purge Valve Assembly
p/n Description
5182-0716 Screw Cap Vial, 2 mL, amber glass, write-on spot, 100/pk
p/n Description
5182-0716 Screw Cap Vial, 2 mL, amber glass, write-on spot, 100/pk
The HPLC System Tool Kit (G4203- 68708) contains some accessories and
tools needed for installation and repair of the module.
p/n Description
)
( *
&$'
,
-
Solvent Cabinet
'
&
( )
Figure 33 Solvent Cabinet Parts (1)
*$+
12
Identifying Cables
Overview 202
Analog cables 204
Remote Cables 206
BCD Cables 209
CAN/LAN Cables 211
External Contact Cables 212
RS-232 Cable Kit 213
Agilent 1200 Module to Printer 214
This chapter provides information on cables used with the Agilent 1200 Infinity
Series modules.
Overview
Never use cables other than the ones supplied by Agilent Technologies to ensure proper
NOTE functionality and compliance with safety or EMC regulations.
Analog cables
p/n Description
Remote cables
p/n Description
BCD cables
p/n Description
CAN cables
p/n Description
LAN cables
p/n Description
5023-0202 Twisted pair network cable, shielded, 7 m (for point to point connection)
p/n Description
RS-232 cables
p/n Description
Analog cables
1 Not connected
2 Shield Analog -
3 Center Analog +
1 Not connected
2 Black Analog -
3 Red Analog +
Remote Cables
NC 5 - Pink Not
connected
13, 15 Not
connected
p/n 5061-3378 Pin 35900 A/D Pin Agilent Signal Name Active
module (TTL)
Pink 5 Not
connected
BCD Cables
One end of these cables provides a 15- pin BCD connector to be connected
to the Agilent modules. The other end depends on the instrument to be
connected to
p/n G1351-81600 Wire Color Pin Agilent Signal Name BCD Digit
module
Green 1 BCD 5 20
Violet 2 BCD 7 80
Blue 3 BCD 6 40
Yellow 4 BCD 4 10
Black 5 BCD 0 1
Orange 6 BCD 3 8
Red 7 BCD 2 4
Brown 8 BCD 1 2
not connected 14
p/n 03396-60560 Pin 3396 Pin Agilent Signal Name BCD Digit
module
1 1 BCD 5 20
2 2 BCD 7 80
3 3 BCD 6 40
4 4 BCD 4 10
5 5 BCD0 1
6 6 BCD 3 8
7 7 BCD 2 4
8 8 BCD 1 2
9 9 Digital ground
NC 15 +5V Low
CAN/LAN Cables
p/n Description
LAN Cables
p/n Description
5023-0202 Twisted pair network cable, shielded, 7 m (for point to point connection)
5 1
10 6
15 11
White 1 EXT 1
Brown 2 EXT 1
Green 3 EXT 2
Yellow 4 EXT 2
Grey 5 EXT 3
Pink 6 EXT 3
Blue 7 EXT 4
Red 8 EXT 4
p/n Description
p/n Description
5181-1529 Cable Printer Serial & Parallel, is a SUB-D 9 pin female vs. Centronics
connector on the other end (NOT FOR FW UPDATE). For use with
G1323 Control Module.
13
Hardware Information
Firmware Description 216
Electrical Connections 219
Rear view of the module 220
Interfaces 221
Overview Interfaces 224
Setting the 8-bit Configuration Switch (without On-board) LAN 228
Communication Settings for RS-232C 229
Special Settings 231
Optional Interface Boards 232
The Solvent Selection Valve (SSV) 236
Early Maintenance Feedback 237
Instrument Layout 238
This chapter provides detailed technical information about your binary pump.
Firmware Description
Resident System
This resident section of the firmware is identical for all Agilent
1100/1200/1220/1260/1290 series modules. Its properties are:
• the complete communication capabilities (CAN, LAN and RS- 232C)
• memory management
• ability to update the firmware of the 'main system'
Main System
Its properties are:
• the complete communication capabilities (CAN, LAN and RS- 232C)
• memory management
• ability to update the firmware of the 'resident system'
In addition the main system comprises the instrument functions that are
divided into common functions like
• run synchronization through APG remote,
• error handling,
• diagnostic functions,
• or module specific functions like
• internal events such as lamp control, filter movements,
• raw data collection and conversion to absorbance.
Firmware Updates
Firmware updates can be done using your user interface:
• PC and Firmware Update Tool with local files on the hard disk
• Instant Pilot (G4208A) with files from a USB Flash Disk
• Agilent Lab Advisor software B.01.03 and above
The file naming conventions are:
PPPP_RVVV_XXX.dlb, where
PPPP is the product number, for example, 1315AB for the G1315A/B DAD,
R the firmware revision, for example, A for G1315B or B for the G1315C
DAD,
VVV is the revision number, for example 102 is revision 1.02,
XXX is the build number of the firmware.
For instructions on firmware updates refer to section Replacing Firmware
in chapter "Maintenance" or use the documentation provided with the
Firmware Update Tools.
Update of main system can be done in the resident system only. Update of the resident
NOTE system can be done in the main system only.
Main and resident firmware must be from the same set.
BV^c;LjeYViZ
GZh^YZciHnhiZb BV^cHnhiZb
GZh^YZci;LJeYViZ
Some modules are limited in downgrading due to their main board version or their initial
NOTE firmware revision. For example, a G1315C DAD SL cannot be downgraded below firmware
revision B.01.02 or to a A.xx.xx.
Some modules can be re-branded (e.g. G1314C to G1314B) to allow operation in specific
control software environments. In this case the feature set of the target type are use and
the feature set of the original are lost. After re-branding (e.g. from G1314B to G1314C), the
original feature set is available again.
All these specific informations are described in the documentation provided with the
firmware update tools.
Electrical Connections
• The CAN bus is a serial bus with high speed data transfer. The two
connectors for the CAN bus are used for internal module data transfer
and synchronization.
• One analog output provides signals for integrators or data handling
systems.
• The interface board slot is used for external contacts and BCD bottle
number output or LAN connections.
• The REMOTE connector may be used in combination with other
analytical instruments from Agilent Technologies if you want to use
features such as start, stop, common shut down, prepare, and so on.
• With the appropriate software, the RS- 232C connector may be used to
control the module from a computer through a RS- 232C connection.
This connector is activated and can be configured with the
configuration switch.
• The power input socket accepts a line voltage of 100 – 240 VAC ± 10 %
with a line frequency of 50 or 60 Hz. Maximum power consumption
varies by module. There is no voltage selector on your module because
the power supply has wide- ranging capability. There are no externally
accessible fuses, because automatic electronic fuses are implemented in
the power supply.
Never use cables other than the ones supplied by Agilent Technologies to ensure proper
NOTE functionality and compliance with safety or EMC regulations.
Configuration switch
RS232
Remote
Analog output
CAN-Bus
HP-IB
Power plug
Interfaces
The Agilent 1200 Infinity Series modules provide the following interfaces:
Pumps
G4220A/B Bin Pump 2 No Yes Yes No Yes CAN-DC- OUT for CAN
G4204A Quat Pump slaves
G1361A Prep Pump 2 Yes No Yes No Yes CAN-DC- OUT for CAN
slaves
Samplers
Others
G4240A CHIP CUBE 2 Yes No Yes No Yes CAN-DC- OUT for CAN
slaves
THERMOSTAT for
G1330A/B (NOT
USED), K1330B
1
Requires a HOST module with on-board LAN (e.g. G4212A or G4220A with minimum firmware B.06.40 or C.06.40) or with ad-
ditional G1369C LAN Card
The detector (DAD/MWD/FLD/VWD/RID) is the preferred access point for control via
NOTE LAN. The inter-module communication is done via CAN.
Overview Interfaces
CAN
The CAN is inter- module communication interface. It is a 2- wire serial
bus system supporting high speed data communication and real- time
requirement.
LAN
The modules have either an interface slot for an LAN card (e.g. Agilent
G1369B/C LAN Interface) or they have an on- board LAN interface (e.g.
detectors G1315C/D DAD and G1365C/D MWD). This interface allows the
control of the module/system via a PC with the appropriate control
software. Some modules have neither on- board LAN nor an interface slot
for a LAN card (e.g. G1170A Valve Drive or G4227A Flex Cube). These are
hosted modules and require a Host module with firmware B.06.40 or later
or with additional G1369C LAN Card.
RS-232C (Serial)
The RS- 232C connector is used to control the module from a computer
through RS- 232C connection, using the appropriate software. This
connector can be configured with the configuration switch module at the
rear of the module. Refer to Communication Settings for RS- 232C.
There is no configuration possible on main boards with on-board LAN. These are
NOTE pre-configured for
• 19200 baud,
• 8 data bit with no parity and
• one start bit and one stop bit are always used (not selectable).
1 In DCD
2 In RxD
3 Out TxD
4 Out DTR
5 Ground
6 In DSR
7 Out RTS
8 In CTS
9 In RI
>chigjbZci E8
APG Remote
The APG Remote connector may be used in combination with other
analytical instruments from Agilent Technologies if you want to use
features as common shut down, prepare, and so on.
Remote control allows easy connection between single instruments or
systems to ensure coordinated analysis with simple coupling requirements.
The subminiature D connector is used. The module provides one remote
connector which is inputs/outputs (wired- or technique).
To provide maximum safety within a distributed analysis system, one line
is dedicated to SHUT DOWN the system’s critical parts in case any module
detects a serious problem. To detect whether all participating modules are
switched on or properly powered, one line is defined to summarize the
POWER ON state of all connected modules. Control of analysis is
maintained by signal readiness READY for next analysis, followed by START
of run and optional STOP of run triggered on the respective lines. In
addition PREPARE and START REQUEST may be issued. The signal levels are
defined as:
• standard TTL levels (0 V is logic true, + 5.0 V is false),
• fan- out is 10 ,
• input load is 2.2 kOhm against + 5.0 V, and
• output are open collector type, inputs/outputs (wired- or technique).
All common TTL circuits operate with a 5 V power supply. A TTL signal is defined as "low"
NOTE or L when between 0 V and 0.8 V and "high" or H when between 2.0 V and 5.0 V (with
respect to the ground terminal).
2 PREPARE (L) Request to prepare for analysis (for example, calibration, detector
lamp on). Receiver is any module performing pre-analysis activities.
4 SHUT DOWN (L) System has serious problem (for example, leak: stops pump).
Receiver is any module capable to reduce safety risk.
5 Not used
6 POWER ON (H) All modules connected to system are switched on. Receiver is
any module relying on operation of others.
7 READY (H) System is ready for next analysis. Receiver is any sequence
controller.
8 STOP (L) Request to reach system ready state as soon as possible (for
example, stop run, abort or finish and stop injection). Receiver is any
module performing run-time controlled activities.
9 START REQUEST (L) Request to start injection cycle (for example, by start key on any
module). Receiver is the autosampler.
Special Interfaces
There is no special interface for this module.
With the introduction of the Agilent 1260 Infinity, all GPIB interfaces have been removed.
NOTE The preferred communication is LAN.
The following tables represent the configuration switch settings for the modules without
NOTE on-board LAN only.
Mode Select 1 2 3 4 5 6 7 8
Reserved 1 0 Reserved
The LAN settings are done on the LAN Interface Card G1369B/C. Refer to the
NOTE documentation provided with the card.
Mode
1 2 3 4 5 6 7 8
Select
Use the following tables for selecting the setting which you want to use for
RS- 232C communication. The number 0 means that the switch is down
and 1 means that the switch is up.
3 4 5 3 4 5
0 0 0 9600 1 0 0 9600
0 0 1 1200 1 0 1 14400
0 1 0 2400 1 1 0 19200
0 1 1 4800 1 1 1 38400
0 7 Bit Communication
1 8 Bit Communication
Switches Parity
7 8
0 0 No Parity
0 1 Odd Parity
1 1 Even Parity
One start bit and one stop bit are always used (not selectable).
Per default, the module will turn into 19200 baud, 8 data bit with no
parity.
Special Settings
The special settings are required for specific actions (normally in a service case).
Boot-Resident
Firmware update procedures may require this mode in case of firmware
loading errors (main firmware part).
If you use the following switch settings and power the instrument up
again, the instrument firmware stays in the resident mode. It is not
operable as a module. It only uses basic functions of the operating system
for example, for communication. In this mode the main firmware can be
loaded (using update utilities).
Mode Select SW1 SW2 SW3 SW4 SW5 SW6 SW7 SW8
TEST/BOOT 1 1 0 0 1 0 0 0
Loss of data
CAUTION
Forced cold start erases all methods and data stored in the non-volatile memory.
Exceptions are calibration settings, diagnosis and repair log books which will not be
erased.
➔ Save your methods and data before executing a forced cold start.
If you use the following switch settings and power the instrument up
again, a forced cold start has been completed.
Mode Select SW1 SW2 SW3 SW4 SW5 SW6 SW7 SW8
TEST/BOOT 1 1 0 0 0 0 0 1
p/n Description
G1351-68701 Interface board (BCD) with external contacts and BCD outputs
The BCD board provides a BCD output for the bottle number of the
Agilent 1200 Series autosampler and four external contacts. The external
contact closure contacts are relay contacts. The maximum settings are:
30 V (AC/DC); 250 mA (fused).
7dVgY
^YZci^[^XVi^dc
G;>[^aiZg
&'
EgdXZhhdg 789 789
^ciZg[VXZ gZ\^hiZg XdccZXidg
G;>[^aiZg
A^cZYg^kZg
'*%b6
:miZgcVa )m :miZgcVa
XdciVXih XdciVXi
XdccZXidg
There are general purpose cables available to connect the BCD output, see
“BCD Cables” on page 209 and the external outputs, see “External Contact
Cables” on page 212 to external devices.
1 BCD 5 20
2 BCD 7 80
3 BCD 6 40
4 BCD 4 10
5 BCD 0 1
6 BCD 3 8
7 BCD 2 4
8 BCD 1 2
9 Digital ground
10 BCD 11 800
11 BCD 10 400
12 BCD 9 200
13 BCD 8 100
15 +5V Low
p/n Description
One board is required per Agilent 1260 Infinity instrument. It is recommended to add the
NOTE LAN board to the detector with highest data rate.
For the configuration of the G1369 LAN Communication Interface card refer to its
NOTE documentation.
The following cards can be used with the Agilent 1260 Infinity modules.
1
These cards may be no longer orderable. Minimum firmware of these Hewlett Packard JetDirect
cards is A.05.05.
p/n Description
5023-0202 Twisted pair network cable, shielded, 7 m (for point to point connection)
EMF Counters
EMF counters increment with use and can be assigned a maximum limit
which provides visual feedback in the user interface when the limit is
exceeded. Some counters can be reset to zero after the required
maintenance procedure.
Instrument Layout
14
Appendix
General Safety Information 240
The Waste Electrical and Electronic Equipment Directive 243
Lithium Batteries Information 244
Radio Interference 245
Sound Emission 246
Agilent Technologies on the Internet 247
Safety Symbols
Table 26 Safety Symbols
Symbol Description
The apparatus is marked with this symbol when the user should refer to
the instruction manual in order to protect risk of harm to the operator and
to protect the apparatus against damage.
Indicates eye damage may result from directly viewing the light produced
by the deuterium lamp used in this product.
The apparatus is marked with this symbol when hot surfaces are available
and the user should not touch it when heated up.
A WARNING
WA R N I N G
alerts you to situations that could cause physical injury or death.
➔ Do not proceed beyond a warning until you have fully understood and met the
indicated conditions.
A CAUTION
CAUTION
alerts you to situations that could cause loss of data, or damage of equipment.
➔ Do not proceed beyond a caution until you have fully understood and met the
indicated conditions.
Safety Standards
This is a Safety Class I instrument (provided with terminal for protective
earthing) and has been manufactured and tested according to
international safety standards.
Operation
Before applying power, comply with the installation section. Additionally
the following must be observed.
Do not remove instrument covers when operating. Before the instrument is
switched on, all protective earth terminals, extension cords,
auto- transformers, and devices connected to it must be connected to a
protective earth via a ground socket. Any interruption of the protective
earth grounding will cause a potential shock hazard that could result in
serious personal injury. Whenever it is likely that the protection has been
impaired, the instrument must be made inoperative and be secured against
any intended operation.
Make sure that only fuses with the required rated current and of the
specified type (normal blow, time delay, and so on) are used for
replacement. The use of repaired fuses and the short- circuiting of fuse
holders must be avoided.
Some adjustments described in the manual, are made with power supplied
to the instrument, and protective covers removed. Energy available at
many points may, if contacted, result in personal injury.
Any adjustment, maintenance, and repair of the opened instrument under
voltage should be avoided whenever possible. When inevitable, this has to
be carried out by a skilled person who is aware of the hazard involved. Do
not attempt internal service or adjustment unless another person, capable
of rendering first aid and resuscitation, is present. Do not replace
components with power cable connected.
Do not operate the instrument in the presence of flammable gases or
fumes. Operation of any electrical instrument in such an environment
constitutes a definite safety hazard.
Do not install substitute parts or make any unauthorized modification to
the instrument.
Capacitors inside the instrument may still be charged, even though the
instrument has been disconnected from its source of supply. Dangerous
voltages, capable of causing serious personal injury, are present in this
instrument. Use extreme caution when handling, testing and adjusting.
When working with solvents, observe appropriate safety procedures (for
example, goggles, safety gloves and protective clothing) as described in the
material handling and safety data sheet by the solvent vendor, especially
when toxic or hazardous solvents are used.
Abstract
The Waste Electrical and Electronic Equipment (WEEE) Directive
(2002/96/EC), adopted by EU Commission on 13 February 2003, is
introducing producer responsibility on all electric and electronic
appliances starting with 13 August 2005.
This product complies with the WEEE Directive (2002/96/EC) marking requirements. The
NOTE affixed label indicates that you must not discard this electrical/electronic product in
domestic household waste.
Product Category:
With reference to the equipment types in the WEEE Directive Annex I, this product is
classed as a Monitoring and Control Instrumentation product.
Lithium batteries may not be disposed-off into the domestic waste. Transportation of
WA R N I N G
discharged Lithium batteries through carriers regulated by IATA/ICAO, ADR, RID,
IMDG is not allowed.
Danger of explosion if battery is incorrectly replaced.
➔ Discharged Lithium batteries shall be disposed off locally according to national
waste disposal regulations for batteries.
➔ Replace only with the same or equivalent type recommended by the equipment
manufacturer.
Lithiumbatteri - Eksplosionsfare.
WA R N I N G
Ved udskiftning benyttes kun batteri som anbefalt av apparatfabrikanten.
➔ Brukt batteri returneres appararleverandoren.
Bij dit apparaat zijn batterijen geleverd. Wanneer deze leeg zijn, moet u ze niet weggooien
NOTE maar inleveren als KCA.
Radio Interference
Sound Emission
Manufacturer’s Declaration
This statement is provided to comply with the requirements of the
German Sound Emission Directive of 18 January 1991.
This product has a sound pressure emission (at the operator position) <
70 dB.
• Sound Pressure Lp < 70 dB (A)
• At Operator Position
• Normal Operation
• According to ISO 7779:1988/EN 27779/1991 (Type Test)
For the latest information on products and services visit our worldwide
web site on the Internet at:
http://www.agilent.com
Index
communication interface board 234 outlet valve 153, 173, 192 pulsation 26
leak rate test overpressure error 128, 132, 134, 136 priming
evaluating results 140 with a pump 56
potential causes of failure 141 P PTFE frit 153
running the test 139 pulsation
packaging
leak sensor open 105 damaged 30 pressure 17
leak sensor short 106 parts identification pump elasticity calibration 93, 136
leak 105 cables 201 pump head assembly 153
line frequency 25 parts pump head missing 119
line voltage 25 active inlet valve 194 pump head 153, 153
liquimeter 162 bottle head assembly 197 pump seals
lithium batteries 244 damaged 31 alternative material 82
lost CAN partner 104 hydraulic path with solvent selection for normal phase solvents 82
valve 184 purge valve 61, 153
M hydraulic path without solvent
selection valve 186 R
main assemblies, overview 152 missing 31
maintenance outlet valve 192 radio interference 245
feedback 237 system tool kit 196 reassembling the pump head 167
pump head with seal wash 153 performance recommended pH range 26
pump head without seal wash 153 specifications 26 remote
replacing firmware 181, 181 cable 206
pH range 26
message remove damper and mixer 84
physical specifications 25
ignition without cover 108, 108 removing the pump head 157
piston 13
remote timeout 104 removing
power consideration 22
minimum flow rate 61 pump head assembly 153
power consumption 25
missing parts 31 repair procedures 153
power cords 23
missing pressure reading 112 repairs
power supply indicator 94
mixer 11 definition of 148
power switch 46
when to remove 84 introduction 148
pressure above upper limit 109
motor-drive power 116 replacing firmware 181, 181
pressure below lower limit 110
warnings and cautions 148
N pressure pulsation 17, 88
results
pressure range 82 pressure test 132, 129
non-operating altitude 25
pressure ripple 133 ripple
non-operating temperature 25
pressure signal missing 110 composition 133
pressure test 92 pressure 133
O
results 132 RS-232C
operating Altitude 25 pressure cable 213
operating temperature 25 operating range 26 communication settings 229
In This Book
*G1312-90015*
*G1312-90015*
G1312- 90015
Rev. B
Agilent Technologies









