SERVICE BULLETIN
TECHNICAL INFORMATION
X-RAY
NUMBER: XR-200502 DATE: 01/26/05
1) At the B-40 console, press “setup” key and hold for 3 sec.
2) Press the button 1, 2, 3 as shown below:
3 2
Program System
AEC Console
1 Exit
Service
Setup Adjust
Exit
4) Press Setup key. The following page appears. (Initial settings screen)
.
Setup
RATIO COMMUNICATION
TECH OPTION
TUBE SIGNAL
FLUO
Exit
5) Press SIGNAL. Use the up and down arrow to go to TABLE I/F out-1. With the cursor at
TABLE I/F out-1, select “Edit”. At item #9, change the number from 0 to 542. This allows
X1B pin 18 and 19 on Table I/F board becomes relay close contact when performing prep and
exposure.
1 2 3 4
5 6 7 8
714 715 0 0
TABLE I/F 2002 TABLE I/F 2002 TABLE I/F 2002 TABLE I/F 2002
Input 542
9