This service bulletin provides instructions on how to connect a room light for the Shimadzu UD150B-40 x-ray system. The steps include entering the service mode screen, going to the initial settings screen, selecting the SIGNAL option, and changing the number at item #9 from 0 to 542. This allows the room light connection pins on the Table I/F board to act as a relay close contact during x-ray preparation and exposure.

Uploaded by

Richard Gomez
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
279 views3 pages

Shimadzu How To Connect Room Light For Radspeed PDF

This service bulletin provides instructions on how to connect a room light for the Shimadzu UD150B-40 x-ray system. The steps include entering the service mode screen, going to the initial settings screen, selecting the SIGNAL option, and changing the number at item #9 from 0 to 542. This allows the room light connection pins on the Table I/F board to act as a relay close contact during x-ray preparation and exposure.

Uploaded by

Richard Gomez
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 3

S H I M ADZ U M E DI CAL S Y S T E M S

Shimadzu Training And Remote Technologies


20101 S. Vermont Ave. Torrance, CA 90502
Tel# (800) 228-1429 Fax# (310) 217-0729
Web: www.shimadzuservice.com E-mail: [email protected]

SERVICE BULLETIN
TECHNICAL INFORMATION
X-RAY
NUMBER: XR-200502 DATE: 01/26/05

TITLE: How to connect room light for Radspeed


EQUIPMENT MODEL: UD150B-40/CH-200/BK-200/BR-120
SERIAL NUMBER AFFECTED:
PURPOSE:
ESTIMATED TIME:
TOOLS REQUIRED:
2 5 ' (5 )2 5 0 ,1 &/8 ' (' &2 1 ),5 0 $7,2 1 2 )&2 0 3/(7,2 1 )2 5 0 ,1 &/8 ' (' 6(5 9,&(5 (32 5 7,1 &/8 ' (' 

Issued by: Sid Nget ..………………….

Checked by: Randal Walker ……………………

Approved by: Randy Walker ..…………………..


How to connect room light for the UD150B-40

1) At the B-40 console, press “setup” key and hold for 3 sec.
2) Press the button 1, 2, 3 as shown below:

3 2

Program System

AEC Console

1 Exit

3) The following page appears on the display. (Service mode screen)

Service

Setup Adjust

Exit

4) Press Setup key. The following page appears. (Initial settings screen)
.
Setup
RATIO COMMUNICATION

TECH OPTION

TUBE SIGNAL

FLUO

Exit

5) Press SIGNAL. Use the up and down arrow to go to TABLE I/F out-1. With the cursor at
TABLE I/F out-1, select “Edit”. At item #9, change the number from 0 to 542. This allows
X1B pin 18 and 19 on Table I/F board becomes relay close contact when performing prep and
exposure.

TABLE I/F out-1

1 2 3 4

603 614 602 613

TABLE I/F TABLE I/F TABLE I/F TABLE I/F


2002 2002 2002 2002
X1A 1-2 X1A 3-4 X1A 5-6 X1A 7-8

KC at terminal X-ray ON for KC at terminal X-ray ON for


code 5 terminal code 5 code 4 terminal code 4

5 6 7 8

714 715 0 0

TABLE I/F 2002 TABLE I/F 2002 TABLE I/F 2002 TABLE I/F 2002

X1A 9-10 X1B 1-2 X1B 14-15 X1B 16-17

OUT 4A 4B OUT 5A 5B OUT 6A 6B OUT 7A 7B

Tube 2 Exposure Free Free


Tube 1 Exposure confirmation
confirmation signal
signal

Input 542
9

TABLE I/F 2002


X1B 18-19
OUT 8A 8B

Free