0% found this document useful (0 votes)
75 views6 pages

Full Schematic

1. The document describes a room power supply and indicator circuit. 2. The circuit uses ICs U2, U3, and transistors Q1 and Q2 to power room lamps and indicators from a 12V supply, and generate call and nurse attention signals. 3. The room power supply regulates the 12V supply to 5V using IC U2 to power the indicator lamps and ICs, while transistors Q1 and Q2 generate blinking outputs to indicate call and nurse attention status.

Uploaded by

yb3hgf2222
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
75 views6 pages

Full Schematic

1. The document describes a room power supply and indicator circuit. 2. The circuit uses ICs U2, U3, and transistors Q1 and Q2 to power room lamps and indicators from a 12V supply, and generate call and nurse attention signals. 3. The room power supply regulates the 12V supply to 5V using IC U2 to power the indicator lamps and ICs, while transistors Q1 and Q2 generate blinking outputs to indicate call and nurse attention status.

Uploaded by

yb3hgf2222
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 6

A B C D E

Room Power Supply Room Indicator


VCC Room J1
VCC Power
1 U2 VCC Room
PxCall Lamp NurseAtt Lamp
1 1
7805 RL1 2
5V Gnd
VCC Power
1 3
VI VO
L1 L2
GND

C1 12V 12V
R7
100nF D3
10k
2

1N4001
B
PxCall Indicator NurseAtt Indicator
2 2
A Q1 Emergency Signal J2
BC547 1
Blinking Signal 2
VCC Room VCC Room VCC Room
3
D1
2

1
1N4001 VCC Room
R3 Call Signal
RL2
Q

Q
100k
PxCall R8 5V
4 9 13
R D Q
R2 10k
CLK

3 6
S
11
CLK D4 3
D

100k 1N4001
12
Q

R
S
3

R1 C2 C3
10k 100nF U3:A 1uF U3:B PxCall Lamp J3

10
4013 4013 1
Q2 VCC Room
2
D2 BC547 RL3 NurseAtt Lamp
1N4001
VCC Room 5V

4 4
2

D5
C5 1N4001
Q

NurseAtt. R9
4 100nF 9 13
R D Q
R5 10k
CLK

6 11
S CLK
D

100k Q3
12 BC547
R6 Q
R
S
3

R4 C4 U4:B
100k
10k 100nF U4:A 4013 C6
8

10

5 100nF
BUZZER 5
4013
R10
100k
TITLE
APPROVALS DATE
DRAWN
F. Bram Dewanto 01/01/2011
Nurse Call System
CHECKED
ROOM CONTROLLER
6 ENGINEER SIZE FSCM NO. DRAWING NO. REV. 6
PROJ ENGINEER A4 Room.DSN -
SCALE SHEET
APPROVED
NONE MODIFIED: 03/08/11 23:02:18
1 OF 5
A B C D E
Room Power Supply Call Switch on Bed
Central PSU
1 1

GND
2

Bed Call
2 2

Room Indicator

3 3

PxCall Indicator NurseAtt Indicator


Att
Bathroom Call
4 4
Pwr

5 5

TITLE
APPROVALS DATE
F. Bram Dewanto

6 SIZE FSCM NO. DRAWING NO. REV. 6

SCALE SHEET
MODIFIED: 01/16/19 6:22:43 PM OF
A B C D E

1 1
VCC Central

U2
9 22 Q1
RESET PC0/SCL BC557 VCC Power
23 VCC Power
PC1/SDA
13 24 R1 J2
XTAL1 PC2/TCK
12 25
XTAL2 PC3/TMS 1
26
PC4/TDO 10k 2
40 27
2 39
PA0/ADC0 PC5/TDI
28
2
38
PA1/ADC1 PC6/TOSC1
29 RL1
PA2/ADC2 PC7/TOSC2
37
PA3/ADC3 D1 5V
36 14 1N4001
PA4/ADC4 PD0/RXD
35 15
PA5/ADC5 PD1/TXD
34 16
PA6/ADC6 PD2/INT0
33 17
PA7/ADC7 PD3/INT1 Blinking Signal
18
PD4/OC1B
1 19
PB0/T0/XCK PD5/OC1A
2 20
PB1/T1 PD6/ICP1
3 21
3 4
PB2/AIN0/INT2 PD7/OC2 U3 J3 3
PB3/AIN1/OC0 VCC Central R2 Emergency Signal
5 6 1 1
PB4/SS
6 2
PB5/MOSI 560
7 32 5 3
PB6/MISO AREF
8 30 2
PB7/SCK AVCC
ATMEGA32 4

4N25

4 U4 BUZZER 4
R3 Call Signal
6 1
560
5
2

4
Central Power Supply
4N25

VCC Power VCC Central


5 5
J1 U1 7805
1 1 3
VI VO
2
GND

TITLE
C1 APPROVALS DATE
100nF DRAWN
F. Bram Dewanto 01/01/2011
Nurse Call System
2

CHECKED
CENTRAL CONTROLLER
6 ENGINEER SIZE FSCM NO. DRAWING NO. REV. 6
PROJ ENGINEER A4 Central.DSN -
SCALE SHEET
APPROVED
NONE MODIFIED: 03/08/11 22:54:44
2 OF 5
A B C D E
A B C D E

Room #1 Indicator J1
1 1

3
2
1
NurseAtt Lamp

PxCall Lamp
J2 J3 J4 J5
1 16 16 1 1 16 16 1
2 15 15 2 2 15 15 2
3 14 14 3 3 14 14 3
2 4 13 13 4 4 13 13 4
2
5 12 12 5 5 12 12 5
6 11 11 6 6 11 11 6
7 10 10 7 7 10 10 7
8 9 9 8 8 9 9 8

3 3

Room #N Indicator J1N


3
2
1

NurseAtt Lamp

4 4
PxCall Lamp
J2N J8 J9 J10
1 16 16 1 1 16 16 1
2 15 15 2 2 15 15 2
3 14 14 3 3 14 14 3
4 13 13 4 4 13 13 4
5 12 12 5 5 12 12 5
6 11 11 6 6 11 11 6
7 10 10 7 7 10 10 7
8 9 9 8 8 9 9 8
5 5

TITLE
APPROVALS DATE
DRAWN
F. Bram Dewanto 01/01/2011
Nurse Call System
CHECKED
CENTRAL INDICATOR
6 ENGINEER SIZE FSCM NO. DRAWING NO. REV. 6
PROJ ENGINEER A4 Indicator.DSN -
SCALE SHEET
APPROVED
NONE MODIFIED: 03/08/11 22:56:22
3 OF 5
A B C D E
4 3 2 1

12 VDC

Call Signal

Emergency Signal

Blinking Signal

D D
GND

Call Light (room #1)

Nurse Attendant Light (room #1)

Call Light (room #2)

Nurse Attendant Light (room #2)

Call Light (room #N)

C Nurse Attendant Light (room #N) C


1 2 3 4 5 1 2 3 4 5 6 7 1 2 3 4 5 6 7 1 2 3 4 5 6 7

Room #1
Lamp
Room #2 Call
Att

Room #N

Px Bed 1

B B
ROOM INDICATOR Nurse
Att

Px Bed N

ROOM CONTROL UNIT ROOM CONTROL UNIT ROOM CONTROL UNIT


(ROOM #1) (ROOM #2) (ROOM #N)

220 12 12 VDC

AC Nurse Call System


A 0 0 A
INSTALLATION & CONNECTION
Buzzer POWER SUPPLY SIZE FSCM NO DWG NO REV
A4 A
CENTRAL CONTROL UNIT
SCALE 1:1 SHEET 4 OF 5
4 3 2 1
A B C D E

J3
Push Button J4
1 BathRoom 1
Switch

1
Emergency Blinking
D2 Signal Signal

1
2

Q
1N4002

1
2
4 9 13
J1 C1 R D Q
R1 R3 RL1

CLK
2 6 11
S CLK

D
12VDC 1 12V
100k
8k2 1nF 12
Q

R
S
3

5
APPROVED

PROJ ENGINEER

ENGINEER

CHECKED

DRAWN

D1 R2 U1:A U1:B

10
2 5v1 10k 4013 4013 2
APPROVALS
F. Bram Dewanto

R4 C2

1
2
D3 100k
J2 100nF
Px Call Nurse Att
Lamp Lamp
A 1
2 1N4002 J5
B
01/01/2011

DATE

3 R5 3
10k
A4
SIZE
SCALE

TITLE

Mode 1
FSCM NO.
NONE

BATHROOM EMERGENCY SWITCH

J3 J4
4 BathRoom 4
Nurse Call System

Emergency Blinking
Switch
MODIFIED:

D1 Signal Signal
1N4002

1
2

1
2
R1
1
12VDC 2
04/06/11

RL1
DRAWING NO.

8k2
12V
J1
21:13:29

D2
5v1
BathRoom.DSN

5 5
SHEET

D3

1
2
J2
Px Call Nurse Att
A 1
2 1N4002
Lamp Lamp
B J5
5 OF 5

R2
10k
6 6

Mode 2
REV.
-

A B C D E

You might also like