Agilent 1290 Infinity II LC

Specification Compendium

Agilent Technologies
Notices
© Agilent Technologies, Inc. 2014 Warranty receive no greater than Restricted Rights as
defined in FAR 52.227-19(c)(1-2) (June
No part of this manual may be reproduced
The material contained in this docu- 1987). U.S. Government users will receive
in any form or by any means (including elec-
ment is provided “as is,” and is sub- no greater than Limited Rights as defined in
tronic storage and retrieval or translation
into a foreign language) without prior agree-
ject to being changed, without notice, FAR 52.227-14 (June 1987) or DFAR
ment and written consent from Agilent in future editions. Further, to the max- 252.227-7015 (b)(2) (November 1995), as
Technologies, Inc. as governed by United imum extent permitted by applicable applicable in any technical data.
States and international copyright laws. law, Agilent disclaims all warranties,
either express or implied, with regard Safety Notices
Manual Part Number to this manual and any information
contained herein, including but not CAUTION
01200-90060 Rev. B limited to the implied warranties of
merchantability and fitness for a par- A CAUTION notice denotes a
Edition ticular purpose. Agilent shall not be hazard. It calls attention to an
11/2014 liable for errors or for incidental or
consequential damages in connection
operating procedure, practice, or
Printed in Germany the like that, if not correctly per-
with the furnishing, use, or perfor-
Agilent Technologies mance of this document or of any formed or adhered to, could
Hewlett-Packard-Strasse 8 information contained herein. Should result in damage to the product
76337 Waldbronn Agilent and the user have a separate
written agreement with warranty
or loss of important data. Do not
terms covering the material in this proceed beyond a CAUTION
document that conflict with these notice until the indicated condi-
terms, the warranty terms in the sep- tions are fully understood and
arate agreement shall control.
met.
Technology Licenses
The hardware and/or software described in WA R N I N G
this document are furnished under a license
and may be used or copied only in accor- A WARNING notice denotes a
dance with the terms of such license. hazard. It calls attention to an
Restricted Rights Legend operating procedure, practice,
If software is for use in the performance of a
or the like that, if not correctly
U.S. Government prime contract or subcon- performed or adhered to, could
tract, Software is delivered and licensed as result in personal injury or
“Commercial computer software” as
defined in DFAR 252.227-7014 (June 1995),
death. Do not proceed beyond a
or as a “commercial item” as defined in FAR WARNING notice until the indi-
2.101(a) or as “Restricted computer soft- cated conditions are fully under-
ware” as defined in FAR 52.227-19 (June
1987) or any equivalent agency regulation stood and met.
or contract clause. Use, duplication or dis-
closure of Software is subject to Agilent
Technologies’ standard commercial license
terms, and non-DOD Departments and
Agencies of the U.S. Government will

Specification Compendium
Contents

Contents

1 Pumps 5
Agilent 1290 Infinity II Flexible Pump (G7104A) 6
Agilent 1290 Infinity II High Speed Pump (G7120A) 9

2 Injectors 13
Agilent 1290 Infinity II Multisampler (G7167B) 14

3 UV-Detectors 19
Agilent 1290 Infinity II Variable Wavelength Detector (G7114B) 20
Agilent 1290 Infinity II DAD FS (G7117A) 23
Agilent 1290 Infinity II DAD (G7117B) 26

4 Special Detectors 29
Agilent 1290 Infinity II Evaporative Light Scattering Detector (G7102A) 30

5 Multicolumn Thermostat 33
Agilent 1290 Infinity II Multicolumn Thermostat (G7116B) 34

Specification Compendium 3
Contents

4 Specification Compendium
Specification Compendium

1
Pumps
Agilent 1290 Infinity II Flexible Pump (G7104A) 6
Physical Specifications 6
Performance Specifications 7
Agilent 1290 Infinity II High Speed Pump (G7120A) 9
Physical Specifications 9
Performance Specifications 10

Agilent Technologies 5
1 Pumps
Agilent 1290 Infinity II Flexible Pump (G7104A)

Agilent 1290 Infinity II Flexible Pump (G7104A)

Physical Specifications
Table 1 Physical Specifications

Type Specification Comments

Weight 16.1 kg (35.5 lbs)

Dimensions (height × width 180 x 396 x 436 mm (7.1 x 15.6 x 17.2


× depth) inches)

Line voltage 100 – 240 V~, ± 10 % Wide-ranging


capability

Line frequency 50 or 60 Hz, ± 5 %

Power consumption 120 VA / 110 W

Ambient operating 4 – 55 °C (39 – 131 °F)


temperature

Ambient non-operating -40 – 70 °C (-40 – 158 °F)


temperature

Humidity < 95 % r.h. at 40 °C (104 °F) Non-condensing

Operating altitude Up to 3000 m (9842 ft)

Non-operating altitude Up to 4600 m (15092 ft) For storing the module

Safety standards: IEC, EN, Installation category II, Pollution degree 2 For indoor use only.
CSA, UL

6 Specification Compendium
Pumps 1
Agilent 1290 Infinity II Flexible Pump (G7104A)

Performance Specifications
Table 2 Agilent 1290 Infinity II Flexible Pump (G7104A) Performance Specifications

Feature Specification

Hydraulic system Dual pistons in series pump with proprietary servo-controlled variable
stroke design and smooth motion control for active damping.

Pump resolution step 300 pL step size


size

Settable flow range 0.001 – 5 mL/min, in 0.001 mL/min increments (executed in 300 pL/step
increments).

Flow precision ≤0.07 % RSD or 0.01 min SD, whatever is greater

Flow accuracy ±1 % or ±10 µL/min, whatever is greater

Pressure range up to 130 MPa (1300 bar) at 0 – 2 mL/min


ramping down to 80 MPa (800 bar) at 5 mL/min

Pressure pulsation <1 % amplitude or <0.5 MPa (5 bar), whatever is greater

Compressibility Automatic
compensation

Recommended 1.0 – 12.5, solvents with pH <2.3 should not contain acid which attack
pH-range stainless steel.

Gradient formation Low pressure quaternary mixing

Delay volume As low as 350 µL

Composition range Settable range: 0 – 100 %


Recommended range: 1 – 99 % or 5 µL/min

Composition <0.15 % RSD or 0.02 min SD, whatever is greater


precision

Composition ±0.4 % absolute (1 – 99 % B)


accuracy

Number of solvent 4 out of maximum 26 solvents


Solvent selection Internal 4-solvent gradient formation valve included.
valve External 2x 12 solvent valve as option, fully integrated in the pump control
interface.

Specification Compendium 7
1 Pumps
Agilent 1290 Infinity II Flexible Pump (G7104A)

Table 2 Agilent 1290 Infinity II Flexible Pump (G7104A) Performance Specifications

Feature Specification

Degassing unit Integrated.


Number of channels: 4, Internal volume per channel: 1.5 mL

Materials in contact TFE/PDD copolymer, FEP, PEEK, PPS, stainless steel, polyimide
with solvent

Automatic Purge Included, allows automatic inline-filter back-flush and automatic mixer
Valve change, e.g. for optional TFA-mixer

Active Seal wash Included

Intelligent System Included


Emulation
Technology (ISET)

Communications Controller-area network (CAN), RS232C, APG remote: ready, start, stop and
shutdown signals, LAN

Safety and Extensive diagnostics, error detection and display through included
maintenance Agilent LabAdvisor, leak detection, safe leak handling, leak output signal
for shutdown of the pumping system. Low voltage in major maintenance
areas.

GLP features Early maintenance feedback (EMF) for continuous tracking of instrument
usage in terms of seal wear and volume of pumped mobile phase with
pre-defined and user settable limits and feedback messages. Electronic
records of maintenance and errors.

Housing All materials are recyclable.

8 Specification Compendium
Pumps 1
Agilent 1290 Infinity II High Speed Pump (G7120A)

Agilent 1290 Infinity II High Speed Pump (G7120A)

Physical Specifications
Table 3 Physical Specifications

Type Specification Comments

Weight 21.0 kg (46.3 lbs)

Dimensions (height × width 200 x 396 x 436 mm (7.9 x 15.6 x 17.2


× depth) inches)

Line voltage 100 – 240 V~, ± 10 % Wide-ranging


capability

Line frequency 50 or 60 Hz, ± 5 %

Power consumption 210 VA / 180 W

Ambient operating 4 – 55 °C (39 – 131 °F)


temperature

Ambient non-operating -40 – 70 °C (-40 – 158 °F)


temperature

Humidity < 95 % r.h. at 40 °C (104 °F) Non-condensing

Operating altitude Up to 3000 m (9842 ft)

Non-operating altitude Up to 4600 m (15092 ft) For storing the module

Safety standards: IEC, EN, Installation category II, Pollution degree 2 For indoor use only.
CSA, UL

Specification Compendium 9
1 Pumps
Agilent 1290 Infinity II High Speed Pump (G7120A)

Performance Specifications
Table 4 Agilent 1290 Infinity II High Speed Pump (G7120A) Performance Specifications

Feature Specification

Hydraulic system Two dual pistons in series, pumps with proprietary servo-controlled
variable stroke design and smooth motion control.

Pump resolution step 300 pL step size


size

Settable flow range 0.001 – 5 mL/min, in 0.001 mL/min increments (executed in 300 pL/step
increments).

Flow precision ≤0.07 % RSD or 0.005 min SD, whatever is greater

Flow accuracy ±1 % or 10 µL/min, whatever is greater

Pressure range up to 130 MPa (1300 bar) at 0 – 2 mL/min


ramping down to 80 MPa (800 bar) at 5 mL/min

Pressure pulsation <1 % amplitude or <0.5 MPa (5 bar), whatever is greater

Compressibility Automatic
compensation

Recommended 1.0 – 12.5, solvents with pH <2.3 should not contain acid which attack
pH-range stainless steel.

Gradient formation High pressure binary mixing

Delay volume As low as 45 µL (10 µL without mixer)

Composition <0.15 % RSD or 0.01 min SD, whatever is greater


precision

Composition ±0.35 % absolute


accuracy

Number of solvent 2 out of maximum 26 solvents

Solvent selection Internal 4-solvent selection valve included.


valve External 2x 12 solvent valve as option, fully integrated in the pump control
interface.

Integrated degassing Included


unit Number of channels: 2 Internal volume per channel: 1.5 mL
Materials in contact with solvent: TFE/PDD Copolymer, FEP, PEEK, PPS.

10 Specification Compendium
Pumps 1
Agilent 1290 Infinity II High Speed Pump (G7120A)

Table 4 Agilent 1290 Infinity II High Speed Pump (G7120A) Performance Specifications

Feature Specification

Automatic Purge Included


Valve

Active Seal wash Included

Intelligent System Included


Emulation
Technology (ISET)

Communications Controller-area network (CAN), RS232C, APG remote: ready, start, stop and
shutdown signals, LAN

Safety and Extensive diagnostics, error detection and display through included
maintenance Agilent LabAdvisor, leak detection, safe leak handling, leak output signal
for shutdown of the pumping system. Low voltage in major maintenance
areas.

GLP feature Early maintenance feedback (EMF) for continuous tracking of instrument
usage in terms of seal wear and volume of pumped mobile phase with
pre-defined and user settable limits and feedback messages. Electronic
records of maintenance and errors.

Housing All materials are recyclable.

Specification Compendium 11
1 Pumps
Agilent 1290 Infinity II High Speed Pump (G7120A)

12 Specification Compendium
Specification Compendium

2
Injectors
Agilent 1290 Infinity II Multisampler (G7167B) 14
Physical Specifications 14
Performance Specifications 15
Physical Specifications of the Sample Cooler 17

Agilent Technologies 13
2 Injectors
Agilent 1290 Infinity II Multisampler (G7167B)

Agilent 1290 Infinity II Multisampler (G7167B)

Physical Specifications
Table 5 Physical Specifications

Type Specification Comments

Weight 22 kg (48.5 lbs) w/o sample cooler

Dimensions (height × width 320 x 396 x 468 mm (12.6 x 15.6 x 18.4


× depth) inches)

Line voltage 100 – 240 V~, ± 10 % Wide-ranging


capability

Line frequency 50 or 60 Hz, ± 5 %

Power consumption 180 VA, 180 W

Ambient operating 4 - 40 °C (39 - 104 °F)


temperature

Ambient non-operating -40 – 70 °C (-40 – 158 °F)


temperature

Humidity < 95 % r.h. at 40 °C (104 °F) Non-condensing

Operating altitude Up to 3000 m (9842 ft)

Non-operating altitude Up to 4600 m (15092 ft) For storing the module

Safety standards: IEC, EN, Installation category II, Pollution degree 2 For indoor use only.
CSA, UL

14 Specification Compendium
Injectors 2
Agilent 1290 Infinity II Multisampler (G7167B)

Performance Specifications
Table 6 Agilent 1290 Infinity II Multisampler (G7167B) Performance Specifications

Feature Specification

Hydraulic System Flow through needle design for minimized sample carry-over and sample
discrimination. Unique dual needle feature1 with two independent
injection needles and loops optionally available. Different loop sizes for
optimized injection range and lowest delay-volume available,
user-exchangeable. Optional multi-wash feature for ultra-low carry-over.

Pressure range up to 130 MPa (1300 bar)

Injection range 0 – 20 µL or 0 – 100 µL single stroke injection at max. pressure


(Single Needle) Up to 0.1 – 1500 µL with optional kits

Injection range (Dual Mix of 20 µL, 40 µL, 100 µL, or 500 µL loops and analytical heads and
Needle1) multi-load
Either two identical set-ups for maximum injection speed or two different
set-ups for maximum application flexibility are possible.

Injection linearity (for R2 = 0.999 from 5 to max. µL


max. 100 µL loop and
metering device)

Precision <0.15 % RSD or SD<10 nL

Accuracy 0.7 % (10 µL, n=10)

Sample viscosity 0.2 – 5 cp

Sample capacity Up to 6144 with 16 x 384 shallow well micro titer plates
Up to 432 x 2 mL vials
Up to 8 single height drawers for 2 micro-titer plate footprint sample
containers, each. Free configuration of single height (e.g. for shallow well
micro titer plates), double height (e.g. for 2 mL vials or deep-well
micro-titer plates) or triple height drawers (e.g. for 6 mL vials) possible.

Injection cycle time As low as 10 s with Single-needle setup, virtually zero with Dual-needle
set-up
Carry Over <0.0009 % (9 ppm) with Multi-wash
<0.003 % (30 ppm) without Multi-wash

Multiwash Outer needle wash and seat backflush for carryover reduction with up to 3
different solvents

Specification Compendium 15
2 Injectors
Agilent 1290 Infinity II Multisampler (G7167B)

Table 6 Agilent 1290 Infinity II Multisampler (G7167B) Performance Specifications

Feature Specification

Communications Controller-area network (CAN),Local Area Network (LAN) ERI: ready, start,
stop and shut-down signals

GLP features Early maintenance feedback (EMF) for continuous tracking of instrument
usage with user-settable limits and feedback messages. Electronic
records of maintenance and errors

Housing All materials recyclable.

Injector Pretreatment functionality such as multi-draw, mixing, stacked injection


programming and dilution

Performance Specifications Agilent 1290 Infinity II Sample Cooler

Operating principle High performance, low-energy consumption micro-compressor based


cooler with ozone-friendly R134A coolant (42 g), user-upgradable.

Temperature range Settable from 4 – 40 °C in 1 °C increments (with a max. temperature of


5 °C below ambient)

Temperature ±2 °C at set-point of 4 °C
accuracy
1
Availability planned early 2015

16 Specification Compendium
Injectors 2
Agilent 1290 Infinity II Multisampler (G7167B)

Physical Specifications of the Sample Cooler


Cooling unit is designed as vapor- compression refrigeration system.
Contains fluorinated greenhouse gas (refrigerant) according to the Kyoto
protocol. For specifications of refrigerant, charge capacity, carbon dioxide
equivalent (CDE), and global warming potential (GWP) see instrument
label.

Table 7 Physical Specification of the Sample Cooler

Type Specification Comments

Weight < 6 kg

Dimensions (height × width × 205 mm X 340 mm X 370 mm


depth)

Refrigerant gas HFKW-134a (0.042 kg) Ozone depletion potential


(ODP) = 0

Line voltage 24 VDC (nominal)

Current 10 A max.

Ambient operating 4 – 40 ° C (39.2 – 131 ° F)


temperature

Ambient non-operating -40 – 70 ° C (-20 – 158 ° F)


temperature

Humidity < 95 % r.h. at 40 °C (104 ° F) Non-condensing

Operating altitude Up to 3000 m (9842 ft)

Non-operating altitude Up to 4600 m (15091 ft)

Safety standards: Installation category II, For indoor use only.


IEC, CSA, UL Pollution degree 2

Specification Compendium 17
2 Injectors
Agilent 1290 Infinity II Multisampler (G7167B)

Table 8 Performance Specifications Agilent 1290 Sample Cooler

Type Specifications

Temperature range from 4 °C to ambient

Temperature settable from 4 – 40 °C in 1 ° increments

Temperature accuracy (<25°C, <50% r.H.) 2 °C to 6 °C at a setpoint of 4 °C

Temperature accuracy at ambient temperatures 2 °C to 8 °C at a setpoint of 4 °C


and humidity <60 %

Temperature Stability +/- 1°C

18 Specification Compendium
Specification Compendium

3
UV-Detectors
Agilent 1290 Infinity II Variable Wavelength Detector (G7114B) 20
Physical Specifications 20
Performance Specifications 21
Agilent 1290 Infinity II DAD FS (G7117A) 23
Physical Specifications 23
Performance Specifications 24
Agilent 1290 Infinity II DAD (G7117B) 26
Physical Specifications 26
Performance Specifications 27

Agilent Technologies 19
3 UV-Detectors
Agilent 1290 Infinity II Variable Wavelength Detector (G7114B)

Agilent 1290 Infinity II Variable Wavelength Detector (G7114B)

Physical Specifications
Table 9 Physical Specifications

Type Specification Comments

Weight 11 kg (24.3 lbs)

Dimensions (height × width 140 x 396 x 436 mm (5.5 x 15.6 x 17.2


× depth) inches)

Line voltage 100 – 240 V~, ± 10 % Wide-ranging


capability

Line frequency 50 or 60 Hz, ± 5 %

Power consumption 80 VA, 70 W

Ambient operating 4 - 55 °C (39 - 131 °F)


temperature

Ambient non-operating -40 – 70 °C (-40 – 158 °F)


temperature

Humidity < 95 % r.h. at 40 °C (104 °F) Non-condensing

Operating altitude Up to 2000 m (6562 ft)

Non-operating altitude Up to 4600 m (15092 ft) For storing the module

Safety standards: IEC, EN, Installation category II, Pollution degree 2 For indoor use only.
CSA, UL

20 Specification Compendium
UV-Detectors 3
Agilent 1290 Infinity II Variable Wavelength Detector (G7114B)

Performance Specifications
Table 10 Agilent 1290 Infinity II Variable Wavelength Detector (G7114B) Performance
Specifications

Feature Specification

Detection type Double-beam photometer

Light source Deuterium lamp

Number of signals Single and dual wavelength detection

Maximum data rate 240 Hz (single wavelength detection)


2.5 Hz (dual wavelength detection)

Noise <±0.15·10-5 AU, at 230 nm (single wavelength detection)


<±0.80·10-5 AU, at 230 nm and 254 nm (dual wavelength detection)
Drift <1·10-4 AU/h, at 230 nm

Linearity >2.5 AU upper limit

Wavelength range 190 – 600 nm

Wavelength ±1 nm, self-calibration with deuterium lines, verification with holmium


accuracy oxide filter

Wavelength <±0.1 nm
precision

Slit width 6.5 nm typical over whole wavelength range


Time programmable Wavelength, polarity, peak width, lamp on/off

Flow cells Standard: 14 µL volume, 10 mm cell path length and 40 bar (588 psi)
pressure maximum
Micro: 2 µL volume, 3 mm cell path length and 120 bar (1760 psi) pressure
maximum
Semi-micro: 5 µL volume, 6 mm cell path length and 40 bar (588 psi)
pressure maximum
Preparative: 4 µL volume, 3 mm cell path length and 120 bar (1760 psi)
pressure maximum
Preparative: 0.3 mm cell path length and 50 bar (725 psi) pressure
maximum
Preparative: 0.06 mm cell path length and 50 bar (725 psi) pressure
maximum

Spectral tools Stop-flow wavelength scan

Specification Compendium 21
3 UV-Detectors
Agilent 1290 Infinity II Variable Wavelength Detector (G7114B)

Table 10 Agilent 1290 Infinity II Variable Wavelength Detector (G7114B) Performance


Specifications

Feature Specification

Analog output Recorder/Integrator 100 mV or 1 V, 1 output

Communication LAN, Controller-area network (CAN),


ERI: ready, start, stop and shut-down signals

GLP Early maintenance feedback (EMF) for continuous tracking of instrument


usage in terms of lamp burn time with user settable limits and feedback
messages. Electronic records of maintenance and errors. RFID for
electronics records of flow cell and UV lamp conditions (path length,
volume, product number, serial number, test passed, and usage).
Verification of wavelength accuracy with built-in holmium oxide filter.

Safety and Extensive diagnostics, error detection and display through Agilent Instant
maintenance Pilot and Agilent Lab Advisor software. Leak detection, safe leak handling,
leak output signal for shutdown of pumping system. Low voltages in major
maintenance areas. Tracking of flow cells and lamps with RFID (radio
frequency identification) tags

22 Specification Compendium
UV-Detectors 3
Agilent 1290 Infinity II DAD FS (G7117A)

Agilent 1290 Infinity II DAD FS (G7117A)

Physical Specifications
Table 11 Physical Specifications

Type Specification Comments

Weight 11.5 kg (25.4 lbs)

Dimensions (height × width 140 x 396 x 436 mm (5.5 x 15.6 x 17.2


× depth) inches)

Line voltage 100 – 240 V~, ± 10 % Wide-ranging


capability

Line frequency 50 or 60 Hz, ± 5 %

Power consumption 110 VA, 100 W

Ambient operating 4 – 40 °C (39 – 104 °F)


temperature

Ambient non-operating -40 – 70 °C (-40 – 158 °F)


temperature

Humidity < 95 % r.h. at 40 °C (104 °F) Non-condensing

Operating altitude Up to 2000 m (6562 ft)

Non-operating altitude Up to 4600 m (15092 ft) For storing the module

Safety standards: IEC, EN, Installation category II, Pollution degree 2 For indoor use only.
CSA, UL

Specification Compendium 23
3 UV-Detectors
Agilent 1290 Infinity II DAD FS (G7117A)

Performance Specifications
Table 12 Agilent 1290 Infinity II Diode Array Detector FS (G7117A) Performance Specifi-
cations

Feature Specification

Detector type 1024-element diode array

Light source Deuterium

Number of signals 8

Maximum 120 Hz (both spectra and signals)


sampling rate

Short-term noise with 10 mm Max-Light cartridge cell: <±3·10-6 AU at 230/4 nm, slit width
4 nm, TC 2 s, ASTM
with 60 mm Max-Light cartridge cell: <±0.6·10-6 AU/cm at 230/4 nm, slit
width 4 nm, TC 2 s, ASTM

Drift <0.5·10-3 AU/h at 230 nm


Linearity >2.0 AU (5 %) at 265 nm
Typically 2.5 AU (5 %)

Wavelength range 190 – 640 nm

Wavelength ±1 nm, self-calibration with deuterium lines


accuracy

Wavelength <±0.1 nm
precision

Diode width ≈0.5 nm

Wavelength Programmable, 2 – 400 nm, in steps of 1 nm


bunching

Flow cells User-exchangeable, self-aligning cartridge cells with RFID tags.


Max-Light Cartridge Cell (Standard): 10 mm, σV = 1.0 µL
Max-Light Cartridge Cell (High Sensitivity): 60 mm, σV = 4 µL
Max-Light Cartridge Ultra Low Dispersion (ULD) Cell: 10 mm, σV = 0.6 µL
Max-Light Cartridge High Dynamic Range (HDR) Cell: 3.7 mm, σV = 0.8 µL
Maximum Operating Pressure (MOP)1: 70 bar
Maximum Incidental Pressure (MIP)2: 150 bar

24 Specification Compendium
UV-Detectors 3
Agilent 1290 Infinity II DAD FS (G7117A)

Table 12 Agilent 1290 Infinity II Diode Array Detector FS (G7117A) Performance Specifi-
cations

Feature Specification

Spectral tools Data analysis software for spectra evaluation, including spectral libraries and
peak purity functions

Analog output Recorder/integrator: 100 mV or 1 V, output range 0.001 – 2 AU, one output

Communications LAN, controller-area network (CAN),


ERI: ready, start, stop and shut-down signals

GLP features Data recovery card to prevent data losses. RFID for electronics records of
flow cell and UV lamp conditions (path length, volume, product number, serial
number, test passed, usage)
Early maintenance feedback (EMF) for continuous tracking of instrument
usage in terms of lamp burn time with user settable limits and feedback
messages. Electronic records of maintenance and errors. Verification of
wavelength accuracy with deuterium lines.

Safety and Extensive diagnostics, error detection and display through Agilent Instant
maintenance Pilot and Agilent Lab Advisor software. Leak detection, safe leak handling,
leak output signal for shutdown of pumping system. Low voltages in major
maintenance areas.

Others Second generation of Electronic temperature control (ETC) for the complete
optical unit
1
Maximum operating pressure (MOP): Maximum pressure at which a system can operate continu-
ously under normal conditions.
2
Maximum incidental pressure (MIP): The maximum pressure which the system can experience
during a short time.

Specification Compendium 25
3 UV-Detectors
Agilent 1290 Infinity II DAD (G7117B)

Agilent 1290 Infinity II DAD (G7117B)

Physical Specifications
Table 13 Physical Specifications

Type Specification Comments

Weight 11.5 kg (25.4 lbs)

Dimensions (height × width 140 x 396 x 436 mm (5.5 x 15.6 x 17.2


× depth) inches)

Line voltage 100 – 240 V~, ± 10 % Wide-ranging


capability

Line frequency 50 or 60 Hz, ± 5 %

Power consumption 110 VA, 100 W

Ambient operating 4 – 40 °C (39 – 104 °F)


temperature

Ambient non-operating -40 – 70 °C (-40 – 158 °F)


temperature

Humidity < 95 % r.h. at 40 °C (104 °F) Non-condensing

Operating altitude Up to 2000 m (6562 ft)

Non-operating altitude Up to 4600 m (15092 ft) For storing the module

Safety standards: IEC, EN, Installation category II, Pollution degree 2 For indoor use only.
CSA, UL

26 Specification Compendium
UV-Detectors 3
Agilent 1290 Infinity II DAD (G7117B)

Performance Specifications
Table 14 Agilent 1290 Infinity II Diode Array Detector (G7117B) Performance Specifica-
tions

Feature Specification

Detector type 1024-element diode array

Light source Deuterium

Number of signals 8

Maximum sampling 240 Hz (both spectra and signals)


rate

Short-term noise with 10 mm Max-Light cartridge cell: <±3·10-6 AU at 230/4 nm, slit width
4 nm, TC 2 s, ASTM
with 60 mm Max-Light cartridge cell: <±0.6·10-6 AU/cm at 230/4 nm, slit
width 4 nm, TC 2 s, ASTM

Drift <0.5·10-3 AU/h at 230 nm


Linearity >2.0 AU (5 %) at 265 nm
Typically 2.5 AU (5 %)

Wavelength range 190 – 640 nm

Wavelength ±1 nm, self-calibration with deuterium lines


accuracy

Wavelength <±0.1 nm
precision

Slit width Programmable: 1, 2, 4, 8 nm

Diode width ~0.5 nm

Wavelength Programmable, 2 – 400 nm, in steps of 1 nm


bunching

Spectral tools Data analysis software for spectra evaluation, including spectral libraries
and peak purity functions

Specification Compendium 27
3 UV-Detectors
Agilent 1290 Infinity II DAD (G7117B)

Table 14 Agilent 1290 Infinity II Diode Array Detector (G7117B) Performance Specifica-
tions

Feature Specification

Flow cells User-exchangeable, self-aligning cartridge cells with RFID tags.


Max-Light Cartridge Cell (Standard): 10 mm, σV = 1.0 µL
Max-Light Cartridge Cell (High Sensitivity): 60 mm, σV = 4 µL
Max-Light Cartridge Ultra Low Dispersion (ULD) Cell: 10 mm, σV = 0.6 µL
Max-Light Cartridge High Dynamic Range (HDR) Cell: 3.7 mm, σV = 0.8 µL
Maximum Operating Pressure (MOP)1: 70 bar
Maximum Incidental Pressure (MIP)2: 150 bar
Analog output Recorder/integrator: 100 mV or 1 V, output range 0.001 – 2 AU, one output

Communications LAN, controller-area network (CAN), ERI: ready, start, stop and shut-down
signals

GLP features Data recovery card to prevent data losses. RFID for electronics records of
flow cell and UV lamp conditions (path length, volume, product number,
serial number, test passed, usage)
Early maintenance feedback (EMF) for continuous tracking of instrument
usage in terms of lamp burn time with user settable limits and feedback
messages. Electronic records of maintenance and errors. Verification of
wavelength accuracy with deuterium lines.

Safety and Extensive diagnostics, error detection and display through Agilent Instant
maintenance Pilot and Agilent Lab Advisor software. Leak detection, safe leak handling,
leak output signal for shutdown of pumping system. Low voltages in major
maintenance areas.

Others Second generation of Electronic temperature control (ETC) for the


complete optical unit
1 Maximum operating pressure (MOP): Maximum pressure at which a system can operate continu-
ously under normal conditions.
2 Maximum incidental pressure (MIP): The maximum pressure which the system can experience
during a short time.

28 Specification Compendium
Specification Compendium

4
Special Detectors
Agilent 1290 Infinity II Evaporative Light Scattering Detector
(G7102A) 30
Physical Specifications 30
Performance Specifications 31

Agilent Technologies 29
4 Special Detectors
Agilent 1290 Infinity II Evaporative Light Scattering Detector (G7102A)

Agilent 1290 Infinity II Evaporative Light Scattering Detector


(G7102A)

Physical Specifications
Table 15 Physical Specifications

Type Specification Comments

Weight 11 kg (non-cooled), 13 kg (cooled)

Dimensions (height × width 415 x 200 x 450 mm (16.3 x 7.9 x 17.7


× depth) inches)

Line voltage 100 – 240 V~, ± 10 % Wide-ranging


capability

Line frequency 50 or 60 Hz, ± 5 %

Power consumption 150 W (max) Maximum

Ambient operating 10–35 °C (50–95 °F)


temperature

Ambient non-operating -40 – 70 °C (-40 – 158 °F)


temperature

Humidity < 80 % r.h. at 40 °C (104 °F) Non-condensing

Operating altitude Up to 2000 m (6562 ft)

Non-operating altitude Up to 4600 m (15092 ft) For storing the module

Safety standards: IEC, CSA, Installation category II, Pollution degree 2 For indoor use only.
UL

30 Specification Compendium
Special Detectors 4
Agilent 1290 Infinity II Evaporative Light Scattering Detector (G7102A)

Performance Specifications
The instrument is suitable for indoor use only and is classified suitable
under the following categories (EN 61010- 1):2010
• Installation category II
• Pollution degree 2
• Safety class 1

Table 16 Agilent 1290 Infinity II Evaporative Light Scattering Detector (G7102A) Perfor-
mance Specifications

Type Specification

Light Source LASER 405 nm, 10 mW (Class 3B)

Detector Dual PMT with digital signal processing

Nebuliser OFF, 25 – 90 °C

Evaporator

Non-cooled OFF, 25 – 120 °C

Cooled OFF, 10 – 80 °C

Gas Flow Range 0.9 – 3.25 SLM (controlled gas shut-off)

Dynamic Raange 4 orders of magnitude

Short Term Noise <0.1 LSU/h (1 mL/min water).

Drift <1 LSU/h (1 mL/min water).

Operating Pressure 60 – 100 psi ( 4.1 – 6.9 bar)

Eluent Flow range 0.2 – 5.0 mL/min

Digital Output 10, 40 or 80 Hz ( 24 bit)

Remote Operation Remote Start Input

Communication Ethernet
Serial (RS232)
Remote Start Input
Pump Stop: 1 Contact closure

Specification Compendium 31
4 Special Detectors
Agilent 1290 Infinity II Evaporative Light Scattering Detector (G7102A)

Table 16 Agilent 1290 Infinity II Evaporative Light Scattering Detector (G7102A) Perfor-
mance Specifications

Type Specification

PC Control ELSD driver for OpenLAB ChemStation edition


ELSD driver for OpenLAB EZChrom edition

Safety and maintenance Gas shut-off Valve, Leak Detection, Laser Interlock

32 Specification Compendium
Specification Compendium

5
Multicolumn Thermostat
Agilent 1290 Infinity II Multicolumn Thermostat (G7116B) 34
Physical Specifications 34
Performance Specifications 35
Extended Specifications 36

Agilent Technologies 33
5 Multicolumn Thermostat
Agilent 1290 Infinity II Multicolumn Thermostat (G7116B)

Agilent 1290 Infinity II Multicolumn Thermostat (G7116B)

Physical Specifications
Table 17 Physical Specifications

Type Specification Comments

Weight 12.5 kg (27.6 lbs)

Dimensions (height × width 160 x 435 (472) x 436 mm (6.3 x 17.1 (18.6)
× depth) x 17.2 inches)

Line voltage 100 – 240 V~, ± 10 % Wide-ranging


capability

Line frequency 50 or 60 Hz, ± 5 %

Power consumption 150 VA, 150 W

Ambient operating 4–55 °C (39–131 °F)


temperature

Ambient non-operating -40 – 70 °C (-40 – 158 °F)


temperature

Humidity < 95 % r.h. at 40 °C (104 °F) Non-condensing

Operating altitude Up to 2000 m (6562 ft)

Non-operating altitude Up to 4600 m (15092 ft) For storing the module

Safety standards: IEC, EN, Installation category II, Pollution degree 2 For indoor use only.
CSA, UL

34 Specification Compendium
Multicolumn Thermostat 5
Agilent 1290 Infinity II Multicolumn Thermostat (G7116B)

Performance Specifications
Table 18 Agilent 1290 Infinity II Multicolumn Thermostat (G7116B) Performance Specifi-
cations

Feature Specification

Operating principle Dual, independent Peltier-element thermostatted column compartment.


Solvent pre-heating and still-air operation for reduction of
chromatographic band-broadening under UHPLC-conditions. Up to three
devices can be clustered and controlled by a single user interface for
additional flexibility1.

Temperature range 4 °C to 110 °C, (minimum 20 °C below ambient)

Temperature stability ±0.05 °C

Temperature ±0.5 °C (with calibration)


accuracy

Independent 2 (in single device)


Temperature zones up to 6 in clustered configuration1
Column capacity 8 columns of 100 mm length plus Quick-Connect fittings or pre-columns
4 columns of 300 mm length plus Quick-Connect fittings or pre-columns
Selection of columns by single optional integrated 8-column selection
valve (1300 bar)
Maximum of
24 columns of 100 mm length plus Quick-Connect fittings or pre-columns
12 columns of 300 mm length plus Quick-Connect fittings or pre-columns
with clustering1 of three devices.
Heat-up/cool-down 5 min from ambient to 40 °C
time 10 min from 40 °C to 20 °C

Solvent heat Individually quick-installable for every column.


exchangers Available at 1.0 µL (ultra-low dispersion), 1.6 µL (standard) and 3 µL
(high-flow) volume.

Valve options 1x integrated valve drive as option


2x external valve drives as option
to host user-exchangeable Quick-Change valve heads of different formats,
materials and pressure ratings (up to 1300 bar):
2-position/6-port, 2-position/10-port, 6-column selection (6-pos/14-port),
8-column selection (8-pos/18-port).
Equipped with tags, valve heads are automatically identified by SW

Specification Compendium 35
5 Multicolumn Thermostat
Agilent 1290 Infinity II Multicolumn Thermostat (G7116B)

Table 18 Agilent 1290 Infinity II Multicolumn Thermostat (G7116B) Performance Specifi-


cations

Feature Specification

Communications Controller-area network (CAN).

Safety and Extensive diagnostics, error detection and display (through Instant Pilot
maintenance control module and Agilent LabAdvisor), leak detection, safe leak handling,
leak output signal for shutdown of pumping system. Low voltages in main
maintenance areas. Door-open sensor.

GLP Valve heads carrying tags with serial number, pressure rating, number of
switches and valve type.
1
Availability planned early 2015

Extended Specifications
The G7116B MCT comes along with one 1.6 µL Low Dispersion Heat
Exchanger that is suitable for most applications.
Additional Heater devices are available for optimization regarding better
heating performance at higher flow rates (>2.5 mL) or for reducing the
dispersion volume for low flow applications.

36 Specification Compendium
Multicolumn Thermostat 5
Agilent 1290 Infinity II Multicolumn Thermostat (G7116B)

Specification Compendium 37
www.agilent.com

In This Book

The compendium contains specifications of


Agilent 1290 Infinity II modules:
• pumps
• injector
• UV- detectors
• special detectors
• multicolumn thermostat

© Agilent Technologies 2014


Printed in Germany
11/2014

*01200-90060*
*01200-90060*
01200-90060
Rev. B

Agilent Technologies