Binder 1
Binder 1
General Index
0 Index
Page
1 Technical characteristics
1.1 Applications and Destined Use 2
1.2 Compatibility 2
1.3 Technical characteristics 3
1.3.1 Electrical data 3
1.3.2 Operating Characteristics 3
1.3.3 Environmental Data 4
1.3.4 Radiological Characteristics 4
1.3.4.1 Table of current value 5
1.3.5 Mechanical Characteristics 7
1.3.6 Collimator 7
1.3.7 Inverter Monobloc 8
1.3.8 kV and mA relationship 9
1.3.9 Identification Plates 9
1.4 Dimension and weights 10
2 Acceptance and Installation
2.1 Introduction 2
2.2 Electrical Connections 3
2.3 Acceptance Test 3
2.3.1 Cable and Connector Check 3
2.3.2 Mechanical Checks 3
2.3.3 Electrical Checks 4
2.3.4 X-Ray field Dimension Adjustment 6
3 Maintenance
3.1 Ordinary Maintenance 2
3.1.1 General Recommendations 2
3.1.2 Frequent Checks and Inspections 2
3.1.3 General Checks and Inspections 3
3.1.4 Cleaning and Disinfection 3
3.2 Extraordinary Maintenance 4
3.2.1 Information for Assistance 4
3.2.2 Troubleshooting 5
3.2.3 Replacement of Components 6
3.2.4 Service Information Checklist 7
3.2.5 Programmed Maintenance Card 8
Sect.0 - Page 1 OF 2
Doc.000534
SIMPLY HP 32
4 Adjustments
4.1 Introduction 2
4.2 Set-Up Procedure 3
4.3 kV Adjustment 3
4.3.1 “Set kV” Voltage Adjustment 4
4.4 Maximum Filament Current Adjustment 4
4.4.1 Filament Current Adjustment 4
4.5 mAs Adjustment 7
4.6 Reading Battery Voltage 7
4.7 Checking Radiographic Parameters 8
4.7.1 Checking kV and mA in Radiography 8
4.8 Collimator Centring 9
5 Appendix
5.1 Spare Parts 2
5.2 Configuration 3
5.4 List of Fuses, LEDs and Potentiometers 5
Sect.0 - Page 2 OF 2
Doc.000534
SIMPLY HP
Sect.1 - Page 1 OF 10
Doc.000535
SIMPLY HP
1. TECHNICAL CHARACTERISTICS
1.1 APPLICATIONS AND DESTINED USE
The apparatus is a MOBILE RADIOGRAPHIC UNIT aimed at satisfying a vast range of clinical
applications. It must ONLY be used by qualified, competent, trained personnel informed of
the risks linked to the use of ionising radiation
The system does not come under the devices where service continuity is required.
The system is not used in contact with the patient - only accidental contact with some parts of
the device with the patient is possible and its contact with the operator is foreseen.
Contact with the patient is not of the invasive type.
Contact with the operator is foreseen for reasons strictly linked to the conditions of use
(Operating Manoeuvres).
The unit is destined to be used to carry out radiological examinations and diagnostic
investigations in the sphere of :
• Operating Theatres
• Sports Medicine
• Plaster Room
• Emergency Department
• Paediatrics
• Orthopaedics
1.2 COMPATIBILITY
The System is supplied in a standard Configuration which cannot be implemented in the field.
External accessories cannot be interfaced .
Sect.1 - Page 2 OF 10
Doc. 000535
SIMPLY HP
1.3 TECHNICAL CHARACTERISTICS
1.3.1 ELECTRICAL DATA
The information regarding the Technical data of the System is given in Tables 1.3.1A and 1.3.1B .
TABLE 1.3.1A
Description Data
Voltage 230 Vac
Frequency 50 Hz/60Hz
Maximum Absorbed Current Cont. Oper. < 1A Impulsive <12A
Line Compensation Automatic
Line Resistance < 2.5 Ω
Power Supply Socket (Standard) 16A Power supply at 230 Vac
Insulation Class EN60601-1 Class I with type B parts applied
Conditions of use Continuous Operation with intermittent load
TABLE 1.3.1B
TABLE 1.3.2
Description Data
Keyboard with Touch Screen for all the operating parameters and
Operator Interface messages for any possible anomalous conditions.
Managed by Microcontroller.
Sect.1 - Page 3 OF 10
Doc.000535
SIMPLY HP
TABLE 1.3.4
Description Data
Maximum Power 32 kW @ 100 kV
Inverter Frequency 100 khz
H.V. Piloting Inverter Piloted by IGBT
Max Ripple ≤ 3 % at Maximum Power
Rising Time ≤ 1 ms
Exposure Times 3ms to 1.5s
KV Variation Range 40 ÷ 125 (Steps of 1kV)
0,2/0,5/0,8/1/1,3/1,6/2/2,5/3,2/4/5/6,3/8/10/13/16/20/25/32/40/50/
mAs Range ( 27 steps )
63/80/100/130/160/200
Sect.1 - Page 4 OF 10
Doc. 000535
SIMPLY HP
1.3.4 TAB 1 TABLE OF CURRENT VALUE 32 KW
LARGE FOCUS
mAs mA
200 320 / / / / / / / / / / / /
160 320 / / / / / / / / / / / /
130 320 320 320 / / / / / / / / / /
100 320 320 320 320 320 / / / / / / / /
80 320 320 400 400 320 320 320 250 / / / / /
63 320 320 400 400 320 320 320 320 320 250 200 / /
50 320 320 400 400 320 320 320 320 320 250 200 200 200
40 320 320 400 400 400 400 320 320 320 250 200 200 200
32 320 320 400 400 400 400 320 320 320 320 250 250 200
25 320 320 400 400 400 400 320 320 320 320 250 250 200
20 320 320 400 400 400 400 400 320 320 320 250 250 200
16 320 320 400 400 400 400 400 320 320 320 250 250 250
13 320 320 400 400 400 400 400 320 320 320 250 250 250
10 320 320 400 400 400 400 400 320 320 320 250 250 250
8 320 320 400 400 400 400 400 320 320 320 250 250 250
6.3 320 320 400 400 400 400 400 320 320 320 250 250 250
5 320 320 400 400 400 400 400 320 320 320 250 250 250
4 320 320 400 400 400 400 400 320 320 320 250 250 250
3.2 320 320 400 400 400 400 400 320 320 320 250 250 250
2.5 320 320 320 320 320 320 320 320 320 320 250 250 250
2 320 320 320 320 320 320 320 320 320 320 250 250 250
1.6 320 320 320 320 320 320 320 320 320 320 250 250 250
1.3 250 250 250 250 250 250 250 250 250 250 250 250 250
1 250 250 250 250 250 250 250 250 250 250 250 250 250
0.8 200 200 200 200 200 200 200 200 200 200 200 200 200
0.5 160 160 160 160 160 160 160 160 160 160 160 160 160
0.2 63 63 63 63 63 63 63 63 63 63 63 63 63
40-45 46-49 50 51-60 61-65 66-70 71-80 81 82-90 91-100 101-103 104-110 111-120 KV
The current value in bold type, can be reached selecting, on the touch screen, the button:
Max charge
Sect.1 - Page 5 OF 10
Doc.000535
SIMPLY HP
TABLE OF CURRENT VALUE 32 KW
1.3.4 TAB 2
SMALL FOCUS
mAs mA
200 125 / / / / / / / / / / / /
160 125 / / / / / / / / / / / /
130 160 160 160 / / / / / / / / / /
100 160 160 160 125 100 / / / / / / / /
80 160 160 160 125 100 100 80 80 / / / / /
63 160 160 160 125 100 100 100 80 80 80 63 / /
50 160 160 160 125 125 125 100 80 80 80 63 63 50
40 160 160 160 125 125 125 100 80 80 80 63 63 63
32 160 160 160 160 125 125 100 100 100 80 80 80 63
25 160 160 200 160 125 125 100 100 100 80 80 80 63
20 160 160 200 160 125 125 100 100 100 80 80 80 63
16 160 160 200 160 125 125 100 100 100 80 80 80 80
13 160 160 200 160 125 125 125 100 100 80 80 80 80
10 160 160 200 160 125 125 125 100 100 100 80 80 80
8 160 160 200 160 160 160 125 100 100 100 80 80 80
6.3 160 160 200 160 160 160 125 100 100 100 80 80 80
5 160 160 200 160 160 160 125 100 100 100 100 100 80
4 160 160 200 160 160 160 125 100 100 100 100 100 80
3.2 160 160 200 160 160 160 125 100 100 100 100 100 80
2.5 160 160 200 160 160 160 125 100 100 100 100 100 80
2 160 160 200 160 160 160 125 100 100 100 100 100 80
1.6 160 160 200 160 160 160 125 100 100 100 100 100 80
1.3 160 160 200 160 160 160 125 125 125 100 100 100 80
1 160 160 160 160 160 160 125 125 125 100 100 100 80
0.8 160 160 160 160 160 160 125 125 125 100 100 100 80
0.5 160 160 160 160 160 160 125 125 125 100 100 100 80
0.2 63 63 63 63 63 63 63 63 63 63 63 63 63
40 - 45 46 - 49 50 51 - 60 61 - 65 66 - 70 71 - 80 81 82 - 90 91 - 100 101 - 103 104 - 110 111 - 120 KV
The current value in bold type, can be reached selecting, on the touch screen, the button:
Max charge
Sect.1 - Page 6 OF 10
Doc. 000535
SIMPLY HP
TABLE 1.3.5
Description Data
Width 1245 mm
Length 650 mm
Height 1500 mm
Focus’s height max 2000 mm
Weight 200 mm
Handling 200 kg
Cassette holder 35x43 Dead man brake
1.3.6 COLLIMATOR
The information regarding the Collimator characteristics is given in TABLE 1.3.6.
1.3.6
TABLE 1.3.6
Description Data
Make and Model R105
Multiple level blades parallel and perpendicular
Blade handling Manual
Dimensions of radiated field with focus-film distance of 1 m: 43 x 43 cm
Focus-film distance indication with retractable tape measure (2m)
Light brightness (lux) 160 lux at 1 m
Inherent Collimator filtration 2 mm Al
Sect.1 - Page 7 OF 10
Doc.000535
SIMPLY HP
TABLE 1.3.7A
Description Data
Model E 100 R
Insert Model X 22
Type of Anode Rotating (3000 RPM)
Size of focuses 0.6 x 1.3
Rated Anode Power FP 11kW/ FG 32kW
Anode Thermal Capacity 80 K J
Anode Target Angle 15°
Anode Max Continuous dissipation 300W
Monobloc Thermal Capacity 600 kJ
Monobloc Max Cont. dissipation 40 W
Min. Inherent Filtration 1.8 mm Al
Total Filtration > 2.7 mm Al
TABLE 1.3.7B
Description Data
Model HF1 100
Maximum Power 32 KW
Power Supply Voltage 230 Vac ± 10%
Power Supply Frequency 50/60 Hz
Maximum Current 12A
Digital Outputs High Active
Digital Inputs High Active
Logic Circuit Power Supply ± 15Vdc
Weight Internal Electronics
Cooling /
Sect.1 - Page 8 OF 10
Doc. 000535
SIMPLY HP
280,00
240,00
101
200,00
160,00
120,00 51 FP
80,00 81
40,00
111
35 45 55 65 75 85 95 105 115 120 125
kV
IBIS s.r.l.
SL - Via Montegrappa 7 I-24121 SL - Via Montegrappa 7 I-24121
Bergamo BG ITALY Bergamo (BG) - ITALY
SO - Via Cascina bruciata 3 24068 SO - Via Cascina bruciata I-24068
Seriate BG ITALY Seriate (BG) - ITALY
TEL
TEL +39 035 42.36.298 TEL +39 035 42.36.298
FAX +39 035 42.36.343 FAX +39 035 42.36.343
Plate located on the MOBILE UNIT FAC-SIMILE plate placed near the
X-RAY MONOBLOC
Sect.1 - Page 9 OF 10
Doc.000535
SIMPLY HP
1.4 Dimension and weights
Sect.1 - Page 10 OF 10
Doc. 000535
SIMPLY HP
Sect.2 - Page 1 OF 6
Doc.000536
SIMPLY HP
2. INSTALLATION
2.1 INTRODUCTION
The installation procedure consists of different operations to be carried out in the order indicated in
the flow chart given below.
U N PACKING
Electrical Connections
Acc
Ac ceptance Test
The different operations indicated are described in detail in the paragraphs below.
ATTENTION
The radiological system is normally pre-installed and configured in the factory according to the
specific requirements of the customer. However, certain Adjustments may be necessary in the case of
replacement of components .
The whole adjustment procedure of the machine parameters is described in detail
in this TECHNICAL MANUAL (SECTION 5 and SECTION 6).
Sect.2 - Page 2 OF 6
Doc.000536
SIMPLY HP
! Make sure that power supply socket is approved for 230 Vac – 16 A .
Sect.2 - Page 3 OF 6
Doc.000536
SIMPLY HP
The single checking stages are listed below.
1 Rolling the radiographic unit wheels
2 Parking brake of the radiographic unit
3 Correct operation of the anti-tilting groups
4 Integrity of the control console
5 Integrity and fixing of the casings
6 Balancing of the pantograph arm
ATTENTION!
! Remove the casings to facilitate the Electrical Checks.
4 CHECKING PRESENCE OF THE SET-UP DATA IN THE MEMORY: Enter SET-UP and check that the SET-UP
data agrees with what was set
(See 5.3 Table 5.2
5.2);
5 X-RAY TUBE FORMATION : When the unit is not used for a long period,
before using it normally, carry out
“formation” of the X-Ray tube with a
series of exposures.
Use the following data, leaving a pause of
minute between one exposure and the next
every kV and mAs value .
Sect.2 - Page 4 OF 6
Doc.000536
SIMPLY HP
Settings 1^ 2^ 3^ 4^ 5^ P 1^ 2^ 3^ 4^ 5^ 6^ 7^ 8^
A
kv 60 70 80 90 100 U 40 50 60 70 80 90 100 110
S
mAs 10 10 10 10 10 E 100 80 64 64 50 50 40 40
During this formation, by means of the oscilloscope, check the values of the kV card S 19 Tp3 (+
+) and
Tp11 (earth) ; mA card S 19 Tp6 (+
+) and Tp11 (earth) on the inverter.
Set the oscilloscope in the following way:
• Channel A:
A 1V/DIV 1V read corresponds to 50 mA
• Channel B: 1V/DIV 1V read corresponds to 20 kV
• Time Base: 20 ms/DIV .
CAUTION!
Should the wave form of the kV show high voltage discharges or discharges are
heard inside the X-
X-Ray tube, interrupt the sequence and let the tube rest for
!
at least 30 minutes, then repeat the formation from the beginning .
6 CHECKING EXPOSURE TIME: During X-Ray exposure, check that the relative LED for
passage of X-Rays lights up and that at the end of the
exposure the display shows the effective exposure time in
seconds.
Sect.2 - Page 5 OF 6
Doc.000536
SIMPLY HP
2.3.4 X-
X-RAY FIELD DIMENSION ADJUSTMENT
Adjust the Focus / Film distance - which can be measured using the RETRACTABLE METER (4
4) -
moving the X-RAY UNIT (Par. 3.2.2).
3.2.2
Turn the COLLIMATOR light on by using the TURN COLLIMATOR LIGHT ON pushbutton on the
CONTROL PANEL or using the TURN COLLIMATOR LIGHT ON (1 1) pushbutton located on the
COLLIMATOR itself .
At this point, adjust the luminous field (which coincides with the X-RAY FIELD) to centre the
area to be radiographed, by means of the LONGITUDINAL DIAPHRAGM ADJUSTMENT (5 5) and
2) handles.
TRANSVERSE DIAPHRAGM ADJUSTMENT (2
2 5
3 3
4
Sect.2 - Page 6 OF 6
Doc.000536
SIMPLY HP
Sect.3 - Page 1 OF 8
Doc.000537
SIMPLY HP
3. MAINTENANCE
WARNING !
Should it be necessary to replace parts which may negatively affect the
!
safety of the machine, only use original spare parts .
Sect.3 - Page 2 OF 8
Doc.000537
SIMPLY HP
TABLE 4.1
PERIOD CHECK
DAILY Check operation of the signals, the touch-screen and the luminous LEDs
CHECKS Check operation of the parking brake.
Check integrity of the warning and danger labels.
WEEKLY CHECKS Check that there are no oil leaks from the Monobloc.
Check there are no unusual noises in the Monobloc during X-ray emission.
SIX-MONTHLY Check correct operation and the relative value of the whole earthing circuit.
CHECKS Check the power supply voltage value.
Check the value of the direct voltages generated inside the system.
Check fixing and general state (dust and corrosion) of the cards.
Check centring of the monobloc-collimator group.
ANNUAL CHECKS Call the Service to carry out the safety and reproducibility tests, as indicated in the IEC
1223-2 and IEC 1223-11 standards, and the other functional checks of the apparatus,
as described in the programmed maintenance plan
(Technical manual)
Sect.3 - Page 3 OF 8
Doc.000537
SIMPLY HP
To disinfect the system, only use methods in accordance with the regulations in force regarding
disinfection procedures and protection against explosions.
To carry out the cleaning and disinfection operations, follow the precautions indicated below:
• Turn the system off and disconnect the mains power supply cable;
• Make sure that no liquid enters the apparatus, in order to prevent short-circuits and
corrosion of the electrical and electro-mechanical components.
WARNING!
Should products which form explosive gaseous mixtures be used for disinfection, make sure that
! the gases have dispersed before using the apparatus again for x-
x-ray emission..
The apparatus is not suitable for use in the presence of flammable anaesthetics, disinfection
! products of cleaning products.
• DIGITAL OSCILLOSCPE (2 Channels) “This must not have a GND earth connection”
• mAs TRUE RMS METER
• mA TRUE RMS METER
• TRU RMS MULTIMETER
• kVp TIME METER
Sect.3 - Page 4 OF 8
Doc.000537
SIMPLY HP
3.2.2 TROUBLESHOOTING
Troubleshooting can be facilitated by the indications in TABLE 3.2 which illustrates the most usual
malfunctions and highlights their possible causes.
TABLE 3.2
The UNIT does not turn on No power supply on input. -Check Presence of Mains Power supply (230 Vac).
(ON LED lit up) -Check Fuse F230V (TR1 2AT).
1
The UNIT does not turn on No power supply on input. -Check Position of Thermomagnetic SW1 ON
(ON LED off) -Check/Replace Fuse IB02/F1-4AT.
2 The UNIT works but does not produce x-rays. Fault in the RADIOGRAPHY PUSHBUTTON. -Check the RADIOGRAPHY PUSHBUTTON cable and
(No alarm on the UNIT display) connector
3 The Touch Screen of the UNIT indicates Fault at +15V or –15V of the feeder of the IB02 -Check operation of the feeder (voltages
“POWER SUPPLY SYSTEM FAULT” Micro card. +15VA/-15VA/+24VA of the IB02 Micro card)
4 The Touch Screen of the UNIT indicates Filament IB05 Card Power supply Absent. -Check Power supplies and Connections of
“FILAMENT ALARM OFF” Filament S15 Card.
5 The UNIT indicates If the indication appears while X-rays are being -Turn off and on again. If the indication persists,
carried out: fault in the X-ray power circuit contact the technical service.
“INVERTER ERROR ALARM”
FBatt fuse (100A) blown -Replace FBatt
Advisable to contact the technical service
6 The UNIT indicates “KV ERROR” During a radiograph the rms kV are lower than -Turn off and on again. If the indication persists,
85% of the ones set: fault in the X-ray power circuit contact the technical service.
7 The UNIT indicates “MA ERROR” During a radiograph the mA values is lower than -Repeat the radiograph. If the indication persists,
the permitted limit contact the technical service.
8 The UNIT indicates “MONOBLOC THERMAL ALARM” Monobloc overheated -Do not carry out X-rays and let the Monobloc
cool down.
Breakage of the thermal sensor -Replace the thermal sensor positioned on the
MONOBLOC and/or check relative cabling with
the IB02 card.
9 The UNIT indicates “MAN X-RAY STOP” During a radiograph with cassette, the X-ray -Repeat the radiograph.
control pushbutton was released early.
10 The UNIT indicates “MAX TIME” Radiography time was higher than the maximum -Check correctness of mA.
time allowed for exposure -Check mA reading on S19/B Card
Tp6(+) and Tp11(-).
11 The UNIT indicates Fault in the X-ray control pushbutton -Check/Replace the X-ray control pushbutton.
“X-RAY CONTROL ALREADY PRESSED”
12 The UNIT indicates The Exposure Time is not Controlled by the -Turn off and on again. If the indication persists,
“WATCH DOG INTERVENTION” Microprocessor contact the technical service.
13 The UNIT indicates Check if Inverter Fuses F1 and F2 are faulty, and -Turn off and on again. If the indication persists,
“NO X-RAYS” Inverter power supplies.
contact the technical service.
14 The UNIT indicates During exposure the anode does not rotate. Check fuses card IB06 F1 and F2.
“ANODE STARTER FAULT” Probable fault in Starter IB06 Card If the indication persists, contact the technical
service.
15 The UNIT indicates Fault in the Power circuit Contact the technical service.
“CAPACITOR FAULT”
Sect.3 - Page 5 OF 8
Doc.000537
SIMPLY HP
TABLE 3.2.3
COMPONENT REPLACED REFERENCE RIFERIMENTO
FILAMENT S15 CARD SECT.5 SEZ.5
MONOBLOC S 17 CARD SECT.5 SEZ.5
INVERTER S 19 CARD SECT.5 SEZ.5
MICRO IB02 CARD SECT.5 SEZ.5
SECT.5 SEZ.5
SECT.5 SEZ.5
TOUCH SCREEN IB07 CARD / /
STARTER IB06 CARD SECT.5 SEZ.5
BATTERY IB04 CARD SECT.5 SEZ.5
ATTENTION
Before carrying out monobloc replacement, carry out the following operations:
• Add some counter-weight on the front part of the unit to prevent the unit tilting.
! • Secure the pantograph arm to the base of the unit (using ropes) to prevent the
arm making undesired movements caused by the lack of weight which will occur
when the monobloc is removed.
Sect.3 - Page 6 OF 8
Doc.000537
SIMPLY HP
EQUPIMENT:
EQUPIMENT:
INSTALLATION DATE:
TYPE:
TYPE:
SERIAL N°:
Sect.3 - Page 7 OF 8
Doc.000537
SIMPLY HP
3.2.5 PROGRAMMED
ROGRAMMED MAINTENANCE
NTENANCE CARD
Checking the good state of all the cables and sheaths. □ POSITIVE □ NEGATIVE
Checking oil leakage from the monobloc. □ POSITIVE □ NEGATIVE
Checking the good state of the casings. □ POSITIVE □ NEGATIVE
Checking general presence of screws and bolts. □ POSITIVE □ NEGATIVE
Checking the painting. □ POSITIVE □ NEGATIVE
Test of Mechanical Operation Result
Checking Touch Screen operation. □ POSITIVE □ NEGATIVE
Checking Arm Balance and sliding. □ POSITIVE □ NEGATIVE
Checking rotation and balance of the Monobloc-Collimator Group. □ POSITIVE □ NEGATIVE
Checking Mobility of the Wheels. □ POSITIVE □ NEGATIVE
Checking Operation of the Brakes. □ POSITIVE □ NEGATIVE
Checking fixing of the Suspended parts (Monobloc, Collimator). □ POSITIVE □ NEGATIVE
Checking General Tightening of screws and bolts. □ POSITIVE □ NEGATIVE
Radiological Checks Result Result
Checking resistance of the earth protection □ POSITIVE □ NEGATIVE
Checking the current dispersed to earth and into the housing. □ POSITIVE □ NEGATIVE
Sect.3 - Page 8 OF 8
Doc.000537
SIMPLY HP
Sect.4 - Page 1 OF 9
Doc.000539
SIMPLY HP
4. ADJUSTMENTS
4.1 INTRODUCTION
The radiological system is entirely regulated in the factory: it is therefore only necessary to
install it to be able to use it.
All the information contained in this SECTION 5 and in SECTION 6 should be used when an Extraordinary
Maintenance operation becomes necessary.
Table 5.1 indicates the Potentiometers which do not require any further
adjustments since they have already been adjusted and sealed in the factory.
TABLE 4.1
Sect. 4 - Page 2 OF 9
Doc.000539
SIMPLY HP
4.2 SET-UP PROCEDURE
In the following parts of this chapter, reference is made to the SET-UP PROCEDURE of the system.
To access this mode, only work from the UNIT console.
Selection of Set-up is made when the system is turned on, as illustrated below.
• Turn the Unit on by pressing the ON and mAs decrease keys at the same time .
• The Set-up screen will appear. Press the Set-up Symbol
• Now enter password 1441 on the alphanumerical keyboard which will appear and you will
enter the General Set-up screen .
4.3 KV ADJUSTMENTS
Ser kV ADJ to40 kV
The SET kV voltage is supplied by the microprocessor and must be adjusted
according to the ratio kv/20 ( 1V=20 kV). 2,00 V
• Press the “Set 40 kV” field. + -
• Connect the probes of a digital multimeter in Vdc on the S 19/B card CP1-10 CP1-2
Sect.4 - Page 3 OF 9
Doc.000539
SIMPLY HP
ATTENTION !
By FILAMENT CURRENT, the current passing through the primary
winding of the power supply transformer of the filament itself is meant
(COOLIDGE TRANSFORMER).
A) MEASUREMENTS
To measure the FILAMENT CURRENT, connect the probes of a digital multimeter in AmA
in series between S15-CP1/2
S15 CP1/2 (+) and the green wire of the electric cable of filament.
To measure the anode mA in RADIOGRAPHY, connect the probe of an oscilloscope
between test points TP6(+) and TP11(-
TP11(-) of the S 19/B card. Correspondence in
reading is 1V=50mA, whereas between test points TP3(+) and TP11(-TP11(-) of the S 19/B
card to display the kV wave form. Correspondence in reading is 1V=20kV ;
TP6 TP11
Adjust the set value of the “DDAC”, again by means of the mAs selection keys, until + -
Sect. 4 - Page 4 OF 9
Doc.000539
SIMPLY HP
C) ADJUSTMENT OF THE NARROW FOCUS CURRENT
Activate the SET UP procedure ;
Select “40 KV 125mA ”
Adjust the set value of the “D DAC” by means of the mAs selection keys until you
obtain a filament current,
current displayed on the digital voltmeter, which is the same as
the one indicated in table 5.4.
Command X-Ray and check that the anode mA are 125; 125 if necessary re-adjust the SET
UP until you obtain the desired value ;
Select “80 KV 50mA”.
Adjust the set value of the “D
DAC”, again by means of the mAs selection keys until you
obtain a filament current,
current displayed on the digital voltmeter, which is the same as
the one indicated in table 5.4.
1V=50mA
+ -
(E.g. for 80 kV - 50 mA - Typical DAC 1170 ) TP6 TP11
Command X-Ray and check that the anode mA are 50 mA; if necessary re-adjust the
SET UP until you obtain the desired value; 1V=20kV
Select “80 KV 125mA” -
+
Adjust the set value of the “D DAC” by means of the mAs selection keys until you TP3 TP11
obtain a filament current,
current displayed on the digital voltmeter, which is the same as
the one indicated in table 5.4.
Sect.4 - Page 5 OF 9
Doc.000539
SIMPLY HP
Can 3 1V=100 mA
Can 1 1V=20 kV
OSCILLOSC0PES
FIGURE
IGURE 4.4.1 - ADJUSTMENT OF NF CURRENT 40KV 400MA
Sect. 4 - Page 6 OF 9
Doc.000539
SIMPLY HP
A) MEASUREMENT
Remove the bridge between terminals CP3-
CP3-1 and CP3-
CP3-3 of the S19/Bcard.
Connect a mAs meter between the above terminals.
B) PROCEDURE
50mAs
Command X-Ray;
On the instrument, check that the mAs value is really the one set (50
50 mAs – Typ. DAC 0480);
0480
if necessary re-adjust the SET UP until you obtain the desired value;
exit SET UP. mAs
A) MEASUREMENTS
To measure the true voltage in the Battery,
Battery connect a digital Multimeter with RMS reading
between the XT2-C(+) XT2-C(-) of the IB04 card.
ATTENTION !
350 Vdc voltage present – take the necessary precautions!
Check that the maximum charge value does not exceed 355 Vdc.
Should it exceed this value, contact the technical service.
Sect.4 - Page 7 OF 9
Doc.000539
SIMPLY HP
4.7 CHECKING RADIOGRAPHIC PARAMETERS
After having completed the operations indicated in the previous paragraphs, it is
necessary to check the values of the proportional signals at the kV and mA which arrive
at the INVERTER. The kV signal tolerance is ± 5%, whereas for the mA it is ± 15%.
15%
ATTENTION!
If the kV value is incorrect, it is necessary to repeat
the SET kV adjustment.
ATTENTION!
For the following checks, an oscilloscope with two channels is required.
• CHANNEL A:
A TP3(+) - TP11(-
TP11(-) of the S 19/B card (kV
kV AT THE TUBE);
• TIME BASE: 20 ms
Ch. A
If necessary, connect the probe of the oscilloscope as follows : + -
TP3 TP11
• CHANNEL B:
B TP6(+) - TP11(-
TP11(-) of the S 19/B card (mA) ;
• TIME BASE: 20 ms Ch. B
+ -
Command X-Ray and check that the values read on the oscilloscope correspond with
those given in TABLE 5.6.1;
5.6.1
Repeat the previous operations setting the kV values at 80 kV and 100 kV and 120kV.
120kV
Sect. 4 - Page 8 OF 9
Doc.000539
SIMPLY HP
Sect.4 - Page 9 OF 9
Doc.000539
SIMPLY HP
Sect.5 - Page 1 OF 6
Doc.000540
SIMPLY HP
5. APPENDIX
5.1 SPARE PARTS
Should it be necessary to replace parts which may affect the safety of the machine in a negative way, only
use original spare parts.
ATTENTION!
When requesting Spare Parts, it is necessary to provide not only
! the code of the piece, but also the serial number of the unit .
Tables 5.1 and 5.2 indicate the electronic and mechanical components which can become faulty or
worn and which must therefore be replaced:
TABLE 5.1 – LIST OF SPARE PARTS
ATTENTION!
For the parts not indicated in tables 6.1 and 6.2
6.2, please contact the
technical assistance service.
Sect.5 - Page 2 OF 6
Doc.000540
SIMPLY HP
5.2 CONFIGURATION
The signals can be displayed in Italian, English, Turkish, German or Russian
on the Touch Screen of the UNIT console.
Selection of the language used is made when the system is turned on, in the way
shown below .
• Turn the UNIT on using the ON key and the decrease mAs key at the
same time.
mAs -
• The Set-up screen will appear. Press the “ Set-up” symbol.
• Now enter the “2442 ENTER” password on the alphanumerical
keyboard which will appear and you will enter the “System
Configuration” screen
• By pressing the language present in the “LANGUAGE” box by means
of the decrease mAs key, select the desired language.
• Finally, to confirm the selection, press Kv increase and the kV
decrease keys at the same time.
Following this, the initial Message will appear in the language selected. kV +
(E.g. READY - 40 kV/0.2mAs)
kV -
Sect.5 - Page 3 OF 6
Doc.000540
SIMPLY HP
Sect.5 - Page 4 OF 6
Doc.000540
SIMPLY HP
Sect.5 - Page 5 OF 6
Doc.000540
SIMPLY HP
Sect.5 - Page 6 OF 6
Doc.000540
SIMPLY HP 32
General Index
0 - Index
Page
Page
1 General Description
1.1 Applications and Destined use 2
1.2 Key 2
1.2.1 Overall View 2
1.2.2 Control Console 4
1.2.3 Collimator 6
2 Safety
2.1 Introduction 2
2.2 Declaration of Conformity 3
2.3 Safety Devices of the System 3
2.3.1 Mechanical Safety Devices 3
2.3.2 Electrical Safety Devices 3
2.4 Protection against Ionising Radiation 4
2.5 Residual Risks 4
2.6 Signals 6
2.6.1 Symbols Used 6
2.6.2 Acoustic Signal Messages 7
2.6.3 Signal and Alarm Messages 7
2.7 Compatibility 8
2.7.1 Electromagnetic Compatibility 8
4 Maintenance
4.1 Ordinary Maintenance 2
4.1.1 General Recommendations 2
4.1.2 Checks and Inspections by the User 2
4.2 Cleaning and Disinfection 3
4.3 Elimination of the Device 4
5 Configuration
5.1 Changing the Unit Console Language 2
6 Technical Characteristics
6.1 Electrical Data 2
6.2 Operating Characteristics 2
6.3 Environmental Data 3
6.4 Radiological Characteristics 3
6.5 Mechanical Characteristics 6
6.6 Collimator 6
6.7 Monobloc and Inverter 7
6.8 kV mA relationship 8
6.9 Identification Plates 8
6.10 Declaration of responsibility 9
Sect.0 - Page 2 OF 2
Doc.000545
SIMPLY HP
Sect.1 - Page 1 OF 6
Doc.000546
SIMPLY HP
1. GENERAL DESCRIPTION
1.1 APPLICATIONS AND DESTINED USE
The apparatus is a MOBILE RADIOGRAPHIC UNIT aimed at satisfying a vast range of clinical applications.
It must ONLY be used by qualified, competent, trained personnel informed of the risks linked to the use
of ionising radiation .
The system does not come under the devices where service continuity is required.
The system is not used in contact with the patient - only accidental contact with some parts of the device
with the patient is possible and is contact with the operator is foreseen.
Contact with the patient is not of the invasive type.
Contact with the operator is foreseen for reasons strictly linked to the conditions of use
(Operating Manoeuvres).
The unit is destined to be used to carry out radiological examinations and diagnostic investigations in the
sphere of :
• Operating Theatres
• Sports Superivision Medicine
• Plaster Room
• Emergency Department
• Paediatrics
• Orthopaedics
1.2 KEY
1.2.1 OVERALL VIEW
16
2
3 4
1
5
7
15
9
10
14
13 11
Sect.1 - Page 2 OF 6
Doc.000546
SIMPLY HP
12
2
1 4
17
Sect.1 - Page 3 OF 6
Doc.000546
SIMPLY HP
1.2.2 CONTROLL
ONTROLL CONSOLE
ONSOLE
1 9
10
2
4 5 8 6 7
12 18
READY mA=63
mA=63 ms= 3
kV 49 mAs 0.2
THORAX LT. SIDE 14
m m m
11
13 15 16 17
Sect.1 - Page 4 OF 6
Doc.000546
SIMPLY HP
1 KV+
KV+ Increase kV 6 PREP X-ray Preparation Control
2 KV-
KV- Decrease kV 7 X-Ray X-ray Control
14 A ANATOMICAL
ANATOMICAL Selection of Anatomical Techniques
16 m MEDIUM
MEDIUM Selection of Medium Patient Body Size
Sect.1 - Page 5 OF 6
Doc.000546
SIMPLY HP
1.2.3 COLLIMATOR
2 5
3 3
4
COLLIMATOR
1 Turning Collimator Light On
4 Retractable Measure
Sect.1 - Page 6 OF 6
Doc.000546
SIMPLY HP
Sect.2 - Page 1 OF 8
Doc.000547
SIMPLY HP
2. SAFETY
2.1 INTRODUCTION
The purpose of this INSTRUCTION MANUAL is to provide operating instructions for qualified radiology
technicians, medical and paramedical personnel so that use of the RADIOGRAPHIC UNIT is simple and safe.
IBIS reserves the right to integrate and improve this information when it is available and to update the
apparatus in accordance with continual technological development.
This apparatus emits X RADIATION and must only be used in conformity with the safety instructions
indicated in this MANUAL and must not be used for purposes other than those foreseen.
The radiological apparatus must only be used by personnel with the necessary knowledge in the field of
protection against radiation and with the required training in use of X-ray apparatus.
The user is always responsible for conformity with the applicable standards
regarding installation and the use of the apparatus!
Furthermore, it requires periodic checks and maintenance according to what is specified in the
MAINTENANCE SECTION of this INSTRUCTION MANUAL.
IBIS declines any responsibility for damage to persons and/or things caused by
improper use of the same!
Sect.2 - Page 2 OF 8
Doc.000547
SIMPLY HP
Sect.2 - Page 3 OF 8
Doc.000547
SIMPLY HP
2.4 PROTECTION AGAINST IONISING RADIATION
! Before carrying out any x-ray exposure, make sure that all the necessary
precautions against radiation have been taken!
During emission of radiation, the personnel present in the x-ray room must follow the instructions in
force on the subject of protection against radiation. For this purpose, bear in mind the following rules:
• where necessary, use protective accessories against radiation in addition to the devices
already provided on the apparatus.
• always use the special radio-protective gowns: a radio-protective material equivalent to 0.35
mm of lead (0.35
0.35 mm Pb)
Pb attenuates 99.95% of the radiation produced at 50 kV and 94.5% of
that produced at 100 kV;
kV
• the best protection against radiation is distance: therefore always keep as far away as
possible both from the radiation source and from the object of the exposure, using the
special remote radiography control with extension cable;
• avoid moving within or staying in the direct beam of radiation.
• always use the smallest exposure field possible: the radiation dispersed considerably
depends on the volume of the object radiated;
• Use always the smallest field of exposure possible, appropriately closing the collimator
diaphragms. The radiation dispersed does, in fact, depend greatly on the volume of the object
radiated.
The MOBILE UNIT has been designed and constructed to avoid turning over up to an inclination of
10° from horizontal in its transport position.
Sect.2 - Page 4 OF 8
Doc.000547
SIMPLY HP
It is therefore advisable:
• not to remain, move or place the APPARATUS on surfaces with an inclination of more than 10°;
• not to try to move the APPARATUS without working on the special handle for “releasing the
brakes”;
• pay attention, during movement of the UNIT, to any obstacles on the floor (cables, steps and
differences in level of any kind).
For the Residual Risks due to any defect in the protection measures used, remember that:
• Protection against electric shocks is made by connecting the metal parts of the apparatus to
earth: it is therefore necessary to periodically check correct operation of the whole earthing
circuit - according to the ORDINARY MAINTENANCE PROGRAMME described in SECTION 4 of this
manual.
➠ During movements of the apparatus, take care that the parts do not hit the patient or the
operator.
!
➠ Avoid excessively fast movements: the kinetic energy accumulated
➠ could be a hazard for personnel close to the unit.
Sect.2 - Page 5 OF 8
Doc.000547
SIMPLY HP
2.6 SIGNALS
PRESENCE
RESENCE OF HIGH VOLTAGE
CIRCUIT-BREAKER CLOSED
CIRCUIT-BREAKER OPEN
PHYSIOLOGICAL EFFECTS
WIDE FOCUS
NARROW FOCUS
POSITION OF X-
X-RAY FOCUS
RADIOGRAPHY
EARTHING PROTECTION
MAX CHARGE
Sect.2 - Page 6 OF 8
Doc.000547
SIMPLY HP
THERMAL ALARM Monobloc temperature too high Wait for the monobloc to cool down.
The radiography control pushbutton has been Assess whether the image obtained is good and, if
MAN X-
X-RAY STOP
released before the end of the exposure necessary, repeat the exposure.
The unit has interrupted the radiography exposure Assess whether the image obtained is good and, if
MAX TIME since the maximum admissible exposure time has necessary, repeat the exposure increasing the value
been reached of the kV.
The Capacitor Bank is not charged or discharged Turn the device off and on again.
CAPACITOR FAULT
correctly by the device. If the alarm is repeated, contact the technical service
Error in Power supply circuit of the Monobloc
STARTER ANODE FAULT Call the technical service
stator
ALARM
MESSAGES
Sect.2 - Page 7 OF 8
Doc.000547
SIMPLY HP
2.7 COMPATIBILITY
2.7.1 ELECTROMAGNETIC COMPATIBILITY (EMC)
According to the foreseen use, this electronic apparatus conforms with the EMC 89/336 Directive,
which defines the maximum permitted emission levels for electronic apparatus and immunity from
electromagnetic fields.
In any case, it is not possible to absolutely exclude the possibility that radio signals coming from
transmitters, such as mobile phones or similar mobile radio systems, themselves conforming to the EMC
standard, may influence correct operation of the electro medical apparatus if they are used near the
systems and have a relatively high transmission power. Consequently the use of this type of radio
apparatus in the immediate vicinity of medical systems controlled electronically must be avoided,
avoided to
eliminate any risk of interference.
EXPLANATION
Electronic apparatus which satisfies the EMC standard is designed so that, under normal conditions,
there is no risk of electromagnetic interference. However, in the case where radio signals coming from
high frequency transmitters with a relatively high transmission power are used near the electronic
apparatus, the risk of electromagnetic incompatibility cannot be totally kept under control.
Sect.2 - Page 8 OF 8
Doc.000547
SIMPLY HP
Sect.3 - Page 1 OF 17
Doc.000548
SIMPLY HP
3.1 TRANSPORT
• Movement of the UNIT must only be made after releasing the parking brakes using the special
handle (1
1).
TRANSPORT POSITION
PRESS
Sect.3 - Page 2 OF 17
Doc.000548
SIMPLY HP
Sect.3 - Page 3 OF 17
Doc.000548
SIMPLY HP
3.2 POSITIONING
• Do not try to move the system when the brakes are on.
• For movements, use the special handles.
• To connect the apparatus to the power supply network, only use 16A !
sockets with earth terminal.
A B
Sect.3 - Page 4 OF 17
Doc.000548
SIMPLY HP
The monobloc / Collimator group can rotate in all directions, as shown in the figures below.
To position it, use the MONOBLOC POSITIONING HANDLE (1
1).
The inclination of the X-RAY UNIT is indicated by the two GONIOMETERS placed in the front (3
3)
and lateral (2
2) position respectively.
Sect.3 - Page 5 OF 17
Doc.000548
SIMPLY HP
Sect.3 - Page 6 OF 17
Doc.000548
SIMPLY HP
3.3 TURNING ON
To make the system operational:
• Connect the UNIT to the power supply mains;
!
To connect the apparatus to the power supply mains, only use
16A sockets with earth terminal.
• The “Connection to the Network” LED on the stand will light up next to
the ON pushbutton.
• Turn the UNIT on.
• The initial automatic TEST of the unit will start immediately:
for about 5/6 seconds the touch-screen will display the control
software version installed. (E.g. IBIS_X-ray)
IBIS_RX
Sect.3 - Page 7 OF 17
Doc.000548
SIMPLY HP
• The word “CHARGE” will appear on the Touch-Screen together with the charge value in a
percentage for the time required to charge the capacitor bank.
•
➠ Wait for the capacitors to be charged before carrying out any other
operation!
• The Touch-Screen will display the values set for radiography and the working technique
(technique set manually);
• In this situation the machine is ready for the x-ray command.
• Therefore select the type of exposure required according to what is indicated on the
following pages.
READY mA=200
mA=200 ms= 4
kV 40 mAs 0.2
Sect.3 - Page 8 OF 17
Doc.000548
SIMPLY HP
• Adjust the Focus / Film distance - which can be measured using the RETRACTABLE
METER (4
4) - moving the X-RAY UNIT
• Turn the COLLIMATOR light on by using the TURN COLLIMATOR LIGHT ON pushbutton
on the TOUCH SCREEN or using the
TURN COLLIMATOR LIGHT ON (1
1) pushbutton located on the COLLIMATOR
itself.
• At this point, adjust the luminous field (which coincides with the X-RAY FIELD) to
centre the area to be x-rayed, by means of the LONGITUDINAL DIAPHRAGM
ADJUSTMENT (5
5) and TRANSVERSE DIAPHRAGM ADJUSTMENT (2
2) knobs.
2 5
3 3
4
Sect.3 - Page 9 OF 17
Doc.000548
SIMPLY HP
3.5 X-
X-RAY EXPOSURE
Before making any exposure, make sure that all the necessary
precautions against radiation have been taken.
Following a long period of inactivity (3 or more months), it is very
important to proceed with FORMATION OF THE X-RAY TUBE.
This is to prevent high voltage discharges which might damage the !
X-ray tube.
The formation procedure is described in the Technical Manual
3.5.1 MANUAL X-
X-RAY EXPOSURE TWO POINT TECHNIQUE ( KV & MAS )
Connect the unit to the mains network. The presence of power supply
voltage is indicated by the ON pushbutton LED lighting up
Activate the maximum battery charge only in case you need the max
power of the unit. (high mAs, high KV)
On selection of the KV and mAs values desired, the theoretical exposure time
associated with the data Set will automatically be displayed
Theoretical time
READY mA=200
mA=200 ms= 4
Symbol
Activate max charge
Sect.3 - Page 10 OF 17
Doc.000548
SIMPLY HP
Remove the X-RAY CONTROL pushbutton from the X-RAY CONTROL SUPPORT.
SUPPORT This
CLICK A and CLICK B - see figures) and must
device is a pushbutton with double click (C
be used as shown below.
Use the X-RAY CONTROL pushbutton and carry out the first click (A
A). The message
“PREPARATION
PREPARATION”
PREPARATION appears on the Display. Keep the pushbutton in this position until
the wording READY FOR X-
X-RAYS appears on the DISPLAY, and consequently make the
second click (B
B) with the X-RAY CONTROL pushbutton and keep it in that position.
At this moment X-RAY RADIATION emission starts and the EMISSION IN PROGRESS LED
lights up.
At the end of X-RAY RADIATION emission, the EMISSION IN PROGRESS LED turns off and
the acoustic signal ends. At this point it is possible to release both the clicks of the
X-RAY CONTROL pushbutton. At the same time the TOUCH-SCREEN will display the
positive outcome of the exposure and the actual emission time.
Releasing the first click (A) early does not have any consequences
Early release of the second click (B) interrupts X-RAY RADIATION emission. The DISPLAY !
will show the wording “MAN X-RAY STOP” for about 10 seconds and the UNIT will emit an
Sect.3 - Page 11 OF 17
Doc.000548
SIMPLY HP
3.5.2 MANUAL X-
X-RAY EXPOSURE THREE POINT TECHNIQUE ( KV & MAS )(OPTION)
Connect the unit to the mains network. The presence of power supply
voltage is indicated by the ON pushbutton LED lighting up
Set the three points x-ray technique, touching the botton on the screen,
as shown.
Set the mA value, touching the write on the screen, by means of the
mAs push-buttons selection. mS
Set the mS value, touching the write on the screen, by means of the mAs
push-buttons selection.
On selection of the KV and mAs values desired, the theoretical exposure time
associated with the data Set will automatically be displayed
Theoretical time
READY mA=200
mA=200 ms= 4
kV 40 mA 100
mAs
KV mS 100
A
Symbol
Three Point
Technique
Sect.3 - Page 12 OF 17
Doc.000548
SIMPLY HP
3.5.3 ANATOMICAL X-
X-RAY EXPOSURE TECHNIQUES
2003
It is normal practice to call the boxes of this menu with general names
of anatomical parts (e.g. upper limbs).
To programme the boxes, select the one desired by touching it with your finger.
The selection will be displayed in the top rectangle of the display. Touching this
with your finger, a keyboard for entering the desired name for the anatomical part
will be displayed. Once writing is completed, press enter and the wording will
appear on the menu.
Repeat the procedure for all the boxes you want to name.
At this point there are lots of boxes on the menu, with the names of as many anatomical
parts as were written in. Now you must proceed to programme the specific anatomical
parts and set their values.
Select the desired box by touching it with your finger. The wording it contains will
appears in the top rectangle on the display. Press enter and a submenu of ten (10)
free boxes ready to be programmed will be displayed.
To programme the boxes, select the one desired by touching it with your finger.
The selection will be displayed in the top rectangle of the display. Touching this
with your finger, a keyboard for entering the desired name for the anatomical part
will be displayed. Once writing is completed, press enter and the wording will
appear on the submenu.
Repeat the procedure for all the boxes you want to name.
Sect.3 - Page 13 OF 17
Doc.000548
SIMPLY HP
After having selected the names of the submenu, it is necessary to assign the kV and
mAs or kV, mA and second, values to each of them.
Select the desired submenu from those shown on the display by touching it with
your finger. The name will be displayed in the top rectangle of the display.
By pressing enter, a screen for setting the values will be displayed. Three (3) sets m
of values can be associated with each anatomical part
m m
(commonly called small, medium or large). P M G
To set these values, touch the associated box with your finger and use the keys on the
right of the keyboard until the value required is reached.
To select of the type of focus to be used, press the key until the one wanted is
displayed.
A
Once all the desired values have been set, exit by touching the “ ESC ” key at
the top left of the screen . A
ESC
To return to normal operation of the unit, press the “ ESC ” key at the top on the left
on the screen several times.
Sect.3 - Page 14 OF 17
Doc.000548
SIMPLY HP
At this point, it is possible to recall the various anatomical techniques (which must
A
have been programmed previously).
Press the Anatomical selection key at the bottom in the centre on the display.
At this point, all the values previously assigned to that anatomical technique will
appear on the display.
m m Am
To select the desired set (commonly called small, medium or large), touch the
associated key with your finger. P M G
The unit is ready for radiography.
READY mA=63
mA=63 ms= 3
kV 49 mAs 0.2
LEFT LATERAL THORAX
A m mA A
m A
Sect.3 - Page 15 OF 17
Doc.000548
SIMPLY HP
3.6 OPTIONAL COMPONENTS
OMPONENTS
3.6.1 DOSEMETER
OSEMETER
When the Dosemeter is installed, the messages shown on the unit console display
change as explained below:
When the unit is turned on, the Dosemeter prepares to read the dose by displaying mGy*cm2
an initial calibration check value. ( e.g. 385) 385
This initial calibration check value is read and checked by the Microprocessor and
displayed in the Dose value display window. ( e.g. 385)
When the calibration dose check is completed, press the RESET button and the mGy*cm2
dosemeter prepares to read the dose and the value is ZERO. (mGy*cm² = 0) 0
mGy*cm2
When an x-ray exposure is carried out, the dosemeter measures the x-ray dose and
displays it in the Dose value display window . xxxx
To reset the dose value, press the RESET DOSE button until the dose value displayed mGy*cm2
If the RESET DOSE button is not pressed, the dose value displayed is summed with
that of the following exposure.
When the unit is turned off, the dose value returns to zero.
DOSE READING
WINDOW
RESET DOSE
READY mA=63
mA=63 = 3
kV 49 mAs 0.2
mGy cm² LATERAL THORAX A
i m m m A
Sect.3 - Page 16 OF 17
Doc.000548
SIMPLY HP
Do not remove the connector from the power supply socket unless the unit
has been turned off.
!
A B
Sect.3 - Page 17 OF 17
Doc.000548
SIMPLY HP
Sez.4 - Pagina 1 DI 4
Doc.000549
SIMPLY HP
4. MAINTENANCE
In this INSTRUCTION MANUAL reference is only made to ordinary maintenance. For
extraordinary maintenance operations, interventions in the case of faults and/or
replacement of components, it is necessary to refer to the TECHNICAL MANUAL.
WARNING
Should it be necessary to replace parts which could jeopardise safety
of the machine, only use original spare parts
Sez.4 - Pagina 2 DI 4
Doc.000549
SIMPLY HP
TABLE 4.1
PERIOD CHECK
DAILY Check operation of the signals, the touch-screen and the luminous LEDs
CHECKS Check operation of the parking brake.
Check integrity of the warning and danger labels.
WEEKLY CHECKS Check that there are no oil leaks from the Monobloc.
Check there are no unusual noises in the Monobloc during X-ray emission.
SIX-MONTHLY Check correct operation and the relative value of the whole earthing circuit.
CHECKS Check the power supply voltage value.
Check the value of the direct voltages generated inside the system.
Check fixing and general state (dust and corrosion) of the cards.
Check centring of the monobloc-collimator group.
ANNUAL CHECKS Call the Service to carry out the safety and reproducibility tests, as indicated in the
IEC 1223-2 and IEC 1223-11 standards, and the other functional checks of the
apparatus, as described in the programmed maintenance plan
(Technical manual)
WARNING
The apparatus is not suitable for use in the presence of anaesthetics and/or flammable
disinfection and cleaning products.
! Should products which form explosive gaseous mixtures be used,
make sure that the gases are dispersed before turning the unit on again.
For any replacements of the unit components (Monobloc, collimator), please
refer to the Technical Manual – Section 3 – Maintenance.
Sez.4 - Pagina 3 DI 4
Doc.000549
SIMPLY HP
The symbol
Indicates that the product has the requirements requested by new directives introduced for the
environmental protection (2002/95/EC, 2002/96/EC, 2003/108/EC) and it must be disposed
properly once its life cycle is ended.
When the device has reached the end of its life it must be disposed at the proper centres for the
separate collection of electrical and electrical wastes, or it must be returned to the reseller or to
the manufacturer in case you want to replace the product with another equivalent new one.
The proper separate collection helps to avoid possible negative effect on the environment and
on health and it facilitates the second use and the re-cycle of materials of which the device is
composed of. Ask for further information to the local authorities about the areas dedicated to the
wastes disposal.
Who does not dispose the product following what here above mentioned will be responsible for
the enforced rules.
Sez.4 - Pagina 4 DI 4
Doc.000549
SIMPLY HP
Sect.5 - Page 1o OF 2
Doc.000550
SIMPLY HP
5 CONFIGURATION
5.1 CHANGING THE LANGUAGE FOR THE UNIT CONSOLE
The signals can be displayed on the Unit console Touch Screen in Italian,
English, Turkish, Polish, German or Russian.
Selecting the language to be used is done on turning the system on as shown
below: mAs -
• Turn the UNIT on using the ON key and the mAs decrease key at the same
time.
• The Set-up screen will appear. Press the “ Setup” symbol.
• Now enter the “2442 ENTER” password on the alphanumerical keyboard
which will appear and you enter the “System Configuration” screen. By
pressing the language present in the “LANGUAGE” box by means of the
decrease mAs key, select the desired language .
• Finally, to confirm the selection, press Kv increase and the kV decrease
keys at the same time. kV +
Following this, the initial Message will appear in the language selected. kV -
(E.g. READY - 40 kV/0.2mAs)
Sect.5 - Page 2 OF 2
Doc.000550
SIMPLY HP
Sect.6 - Page 1 OF 9
Doc.000551
SIMPLY HP
6. TECHNICAL CHARACTERISTICS
6.1 ELECTRICAL DATA
The information regarding the Technical data of the System is given in TABLE 6.1A and
6.1B.
6.1B
TABLE 6.1A
Description Data
Voltage 230 Vac
Frequency 50 Hz/60Hz
Maximum Absorbed Current Contin. Oper. <1A / Inter. Oper. 12A
Line Compensation Automatic
Line Resistance < 2.5 Ω
Power Supply Socket (Standard) 16A Power supply at 230 Vac
Insulation Class EN60601-1 Class I with type B parts applied
Conditions of use Continuous Operation with intermittent load
TABLE 6.1B
TABLE 6.2
Description Data
Rear-lit Touch Screen for all the operating
parameters and messages for any possible
Operator Interface
anomalous conditions.
Managed by Microcontroller.
TABLE 6.4
Description Data
Maximum Power 32 kW @ 100 kV
Inverter Frequency 100 khz
H.V. Piloting Inverter Piloted by IGBT
Max Ripple ≤ 3 % at Maximum Power
Rising Time ≤ 1 ms
Exposure Times 3ms to 1.5s
KV Variation Range 40 ÷ 125 (Steps of 1kV)
0,2/0,5/0,8/1/1,3/1,6/2/2,5/3,2/4/5/6,3/8/10/13/16/20/25/32/40
mAs Range ( 27 steps )
/50/63/80/100/130/160/200
Sect.6 - Page 3 OF 9
Doc.000551
SIMPLY HP
6.4 TAB 1 TABLE OF CURRENT VALUE 32 KW
LARGE FOCUS
mAs mA
200 320 / / / / / / / / / / / /
160 320 / / / / / / / / / / / /
130 320 320 320 / / / / / / / / / /
100 320 320 320 320 320 / / / / / / / /
80 320 320 400 400 320 320 320 250 / / / / /
63 320 320 400 400 320 320 320 320 320 250 200 / /
50 320 320 400 400 320 320 320 320 320 250 200 200 200
40 320 320 400 400 400 400 320 320 320 250 200 200 200
32 320 320 400 400 400 400 320 320 320 320 250 250 200
25 320 320 400 400 400 400 320 320 320 320 250 250 200
20 320 320 400 400 400 400 400 320 320 320 250 250 200
16 320 320 400 400 400 400 400 320 320 320 250 250 250
13 320 320 400 400 400 400 400 320 320 320 250 250 250
10 320 320 400 400 400 400 400 320 320 320 250 250 250
8 320 320 400 400 400 400 400 320 320 320 250 250 250
6.3 320 320 400 400 400 400 400 320 320 320 250 250 250
5 320 320 400 400 400 400 400 320 320 320 250 250 250
4 320 320 400 400 400 400 400 320 320 320 250 250 250
3.2 320 320 400 400 400 400 400 320 320 320 250 250 250
2.5 320 320 320 320 320 320 320 320 320 320 250 250 250
2 320 320 320 320 320 320 320 320 320 320 250 250 250
1.6 320 320 320 320 320 320 320 320 320 320 250 250 250
1.3 250 250 250 250 250 250 250 250 250 250 250 250 250
1 250 250 250 250 250 250 250 250 250 250 250 250 250
0.8 200 200 200 200 200 200 200 200 200 200 200 200 200
0.5 160 160 160 160 160 160 160 160 160 160 160 160 160
0.2 63 63 63 63 63 63 63 63 63 63 63 63 63
40-45 46-49 50 51-60 61-65 66-70 71-80 81 82-90 91-100 101-103 104-110 111-120 KV
The current value in bold type, can be reached selecting, on the touch screen, the button:
Max charge
Sect.6 - Page 4 OF 9
Doc.000551
SIMPLY HP
mAs mAs
200 125 / / / / / / / / / / / /
160 125 / / / / / / / / / / / /
130 160 160 160 / / / / / / / / / /
100 160 160 160 125 100 / / / / / / / /
80 160 160 160 125 100 100 80 80 / / / / /
63 160 160 160 125 100 100 100 80 80 80 63 / /
50 160 160 160 125 125 125 100 80 80 80 63 63 50
40 160 160 160 125 125 125 100 80 80 80 63 63 63
32 160 160 160 160 125 125 100 100 100 80 80 80 63
25 160 160 200 160 125 125 100 100 100 80 80 80 63
20 160 160 200 160 125 125 100 100 100 80 80 80 63
16 160 160 200 160 125 125 100 100 100 80 80 80 80
13 160 160 200 160 125 125 125 100 100 80 80 80 80
10 160 160 200 160 125 125 125 100 100 100 80 80 80
8 160 160 200 160 160 160 125 100 100 100 80 80 80
6.3 160 160 200 160 160 160 125 100 100 100 80 80 80
5 160 160 200 160 160 160 125 100 100 100 100 100 80
4 160 160 200 160 160 160 125 100 100 100 100 100 80
3.2 160 160 200 160 160 160 125 100 100 100 100 100 80
2.5 160 160 200 160 160 160 125 100 100 100 100 100 80
2 160 160 200 160 160 160 125 100 100 100 100 100 80
1.6 160 160 200 160 160 160 125 100 100 100 100 100 80
1.3 160 160 200 160 160 160 125 125 125 100 100 100 80
1 160 160 160 160 160 160 125 125 125 100 100 100 80
0.8 160 160 160 160 160 160 125 125 125 100 100 100 80
0.5 160 160 160 160 160 160 125 125 125 100 100 100 80
0.2 63 63 63 63 63 63 63 63 63 63 63 63 63
40 - 45 46 - 49 50 51 - 60 61 - 65 66 - 70 71 - 80 81 82 - 90 91 - 100 101 - 103 104 - 110 111 - 120 KV
The current value in bold type, can be reached selecting, on the touch screen, the button:
Max charge
Sect.6 - Page 5 OF 9
Doc.000551
SIMPLY HP
TABLE 6.5
Description Data
Width 1245 mm
Length 650 mm
Height 1500 mm
Focus’s Height max 2000 mm
Diameter of rear wheels 200 mm
Weight 200 kg
Handling Dead man brake
Cassette holder 35x43 35X45
6.6 COLLIMATOR
The information regarding the information relative to the collimator characteristics is given in
Table 6.6.
TABLE 6.6
Description Data
Make and Model R105
Multiple level blades parallel and perpendicular
Blade handling Manual
Dimensions of radiated field with focus-film distance of 1 m: 43 x 43 cm
Focus-film distance indication with retractable tape measure (2m)
Light brightness (lux) 160 lux at 1 m
Inherent collimator filtration 2 mm Al
Sect.6 - Page 6 OF 9
Doc.000551
SIMPLY HP
AND 6.8.
Sect.6 - Page 7 OF 9
Doc.000551
SIMPLY HP
280,00
240,00
101
200,00
160,00
120,00 51 FP
80,00 81
40,00
111
35 45 55 65 75 85 95 105 115 120 125
kV
IBIS s.r.l.
SL - Via Montegrappa 7 I-24121 SL - Via Montegrappa 7 I-24121
Bergamo BG ITALY Bergamo (BG) - ITALY
SO - Via Cascina bruciata 3 24068 SO - Via Cascina bruciata I-24068
Seriate BG ITALY Seriate (BG) - ITALY
TEL
TEL +39 035 42.36.298 TEL +39 035 42.36.298
FAX +39 035 42.36.343 FAX +39 035 42.36.343
FACSIMILE of plate placed on the MOBILE UNIT FACSIMILE of plate placed on the
X-RAY MONOBLOC
Sect.6 - Page 8 OF 9
Doc.000551
SIMPLY HP
6.10 DECLARATION OF RESPONSIBILITY
IBIS shall only be held responsible for the safety of its products in the case where their
maintenance, repair and/or modification has been carried out by IBIS or by personnel
expressly authorised by IBIS.
IBIS
IBIS shall not be held responsible for any malfunctions, damage and/or hazards due to
incorrect use of the system or due to non-conformity with the Maintenance Rules.
The PERSON IN CHARGE of the plant the system is inserted in shall be responsible for making
sure that the plant is only used by suitably trained and qualified operators.
The Simply HP System conforms with the national and international standards in force.
Sect.6 - Page 9 OF 9
Doc.000551