S IDE 1 S IDE 2
M anual Operation E CG M onitoring A E D Operation
l Preparation
Insert Data Card (optional). 1. Apply monitoring electrodes.
Turn Energy Select knob to Manual On . If Patient is:
2. Select lead (optional).
Using Paddles wUnresponsive
Apply conductive matter. Press LEAD
SELECT
to choose desired lead. wNot Breathing
Apply paddles to patient's chest, using
the anterior-apex placement. 3. Set HR Alarm (optional).
wPulseless
Using Pads Connect Pads Patient Cable
Apply pads as directed by packaging Press HR ALARM to select alarm limits. Apply Pads
instructions.
Connect the pads to the pads patient To disable alarm, press HR ALARM . Insert Data Card (optional)
cable. Turn Energy Select to
Electrode Placement: 3 - Lead AED On
l
Defibrillation
RA/R LA/L
l
Select Energy. lll
1 Move the Energy Select knob to ll If Instructed, Press ANALYZE
the desired energy level.
Charge. LL/F
Press or yellow charge
CHARGE
button on apex paddle.
Electrode Placement: 5 - Lead
If Shock Advised If No Shock Advised
Check Patient
Shock. 1 2 3
Press SHOCK . If using external RA/R LA/L
4 5 6
paddles or switched internal V1
paddles, simultaneously press Press SHOCK
If needed, press PAUSE
shock buttons on paddles. and begin CPR
To disarm, press DISARM .
RL/N LL/F
Limb Leads V-LeadOptions
l Pacing if shock series is set to greater than one
if rhythm monitoring active
Select Manual On .
Apply electrodes and pads. WARNING
Press Pacer . Electric shock hazards exist internally. Do not remove assembly
screws. Refer servicing to qualified personnel.
- adjust Rate. Charge batteries between use or if the message LOW BATTERY
- press Start.
- adjust Output if needed. *M4735-97800* is displayed. Perform a Battery Capacity Test every 3 months.
The defibrillator will take longer to charge when used with only
AC power or a depleted battery. Use only with the M3516A
for AE D Operation
*C* M4735-97800 Rev C
Battery Pack.
M4735A
for Manual Operation, P acing & Monitoring