Control Unit 9.3 Control Unit

The control unit is integrated in a housing and connected to the Dornier Compact Delta II, Delta II via a helix cable. It contains displays, buttons to control shockwave intensity, pulse frequency, coupling pressure, and positioning of the therapy head and X-ray C-bow. The shockwave release button releases shockwaves as long as the button is pressed.

Uploaded by

Mahfoud M Hassan
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
328 views

Control Unit 9.3 Control Unit

The control unit is integrated in a housing and connected to the Dornier Compact Delta II, Delta II via a helix cable. It contains displays, buttons to control shockwave intensity, pulse frequency, coupling pressure, and positioning of the therapy head and X-ray C-bow. The shockwave release button releases shockwaves as long as the button is pressed.

Uploaded by

Mahfoud M Hassan
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 10

Control Unit

9.3 Control Unit

The most important functions of the Dornier Compact Delta II, Delta II are executed
via the control unit. The control unit is integrated in a housing and connected to the
Dornier Compact Delta II, Delta II via a helix cable.

Figure 9-1 Control Unit of the Dornier Compact Delta II, Delta II with the most important
Elements

1 Display (H1) 8 Coupling pressure (S6, S7)


2 Reset (S1) 9 Movement buttons (S8 to S13)
3 Display (H2) 10 Zero position (S16)
4 Pulse frequency (S2, S3) 11 Shock-wave release button (S17)
5 Display (H3) 12 C-arm CC+ (S14)
6 Shock-wave intensity (S4, S5) 13 C-arm CC- (S15)
7 Display (H4)

CS3.SM_GB_A 9.3
Control Unit

Table 9-1 Elements of Control Unit

Symbol Designation Description


- Display of shock-wave number or error message
H1
Display 1 - Display of the Service Menu

H2 - Display of the type of triggering / fixed trigger


Display 2 frequency
- Display of the Default values of the single
Service Menus

H3 Display of shock-wave intensity


Display 3

H4 Display of coupling pressure


Display 4

S1 - Pressing simultaneously with S4,


Reset Reset of shock-wave counter to zero

- Pressing simultaneously with S16,


switching between isocentric mode, the mode for
image oriented movement and the mode for
normal movement
- Acknowledge of error message
- Switch-over Front/Side Mode

S3 - Pulse frequency (+) switches step-by-step to the


Pulse frequency (+) next higher value,

S2 - Pulse frequency (-) switches step-by-step to the


Pulse frequency (-) next lower value.

- Selection between ECG triggering and fixed


frequency triggering.
Sequence: ECG 60, 70, 80, 90, 100, 110, 120, 150
and 180 SW/min
Press repeatedly to switch from one value to the
next. To switch to next value after 180, press pulse
frequency (-). In ECG mode, press pulse frequency
(+) to select next value.

S5 Shock-wave intensity (+)


Shock-wave intensity (+) increases shock-wave intensity

S4 Shock-wave intensity (-)


Shock-wave intensity (-) reduces shock-wave intensity

S7 Coupling pressure (+) increases the coupling


Coupling pressure (+) pressure.

S6 Coupling pressure (-) reduces coupling pressure.


Coupling pressure (-)
Coupling pressure is displayed with 0 as lowest
value to 5 as highest value on display H4.

9.4 CS3.SM_GB_A
Control Unit

Symbol Designation Description

S15 X-ray C-bow moves in preset position for CC-


Position for CC- locating locating as long as key is pressed. Intermediate
position is PA position. Press key once more to
move to CC- end position.
Release key during movement to enable
intermediate positions.

S14 X-ray C-bow moves in preset position for CC+


+
Position for CC locating locating as long as key is pressed. Intermediate
position is PA position. Press key once more to
move to CC+ end position.
Release key during movement to enable
intermediate positions.

S10 Move therapy head and X-ray C-bow in Z axis.


Z axis (+) Z axis (+) raises,
Z axis (-) lowers therapy head and X-ray C-bow.

With image oriented movement:


movement in direction T+ and T-
S11
Z axis (-)

S12 Move therapy head and X-ray C-bow in Y axis.


Y axis (+) Y axis (+) moves therapy head and X-ray C-bow
towards the unit,
Y axis (-) moves therapy head and X-ray C-bow
towards patient table.

When isocentric mode is activated:


S13 Movement of TKCF axis
Y axis (-)
With image oriented movement:
movement in direction V+ and V-

Move therapy head and X-ray C-bow in X axis.


X axis (-) moves therapy head and X-ray C-bow to
the left (view from the front onto the unit),
S9 S8 X axis (+) moves therapy head and X-ray C-bow
X axis (-) X axis (+) to the right (view from the front onto the unit).

When isocentric mode is activated:


Movement of TKCD axis

With image oriented movement:


movement in direction H+ and H-

CS3.SM_GB_A 9.5
Control Unit

Symbol Designation Description

Move therapy head and X-ray C-bow in any axis


up to set zero position (setup position). Hold key
pressed until zero position (setup position) is
S16 reached.
Zero position,
setup position, (only if a In combination with reset key:
table Dornier Relax + is Switching to the isocentric mode, the mode for
available) image oriented movement and the mode for
normal movement

Releases shock-waves
S17 (shock-wave release continues as long as button is
Shock-wave release pressed)
button

Key reset in combination with key


S1 + S16 zero position :
Switching between isocentric mode, the mode for
image oriented movement and the mode for
normal movement

With isocentric mode:

S9 Turning therapy head in CD axis to 190°.


TKCD (-)

S8 Turning therapy head in CD axis to +10°.


TKCD (+)

S12 With isocentric mode:


TKCF (+)
Move therapy head in CF axis up to
max. +45°.

Move therapy head in CF axis down to


S13 max. -8°.
TKCF (-)

With isocentric mode:

S10 blocked

S11 blocked

9.6 CS3.SM_GB_A
Control Unit

Symbol Designation Description

With image oriented movement:

S9 H- (moving the concrement to the left on the


ultrasound image)

S8 H+ (moving the concrement to the right on the


ultrasound image)

With image oriented movement:

S12 V+ (moving the concrement up on the


ultrasound image)

S13 V- (moving the concrement down on the


ultrasound image)

With image oriented movement:

S10 T+ (moving the concrement along the depth inside


on the ultrasound image)

S11 T- (moving the concrement along the depth


outside on the ultrasound image)

All movement and rotating axes are moved to a


defined position in order to bring the positioning
S16 + S17 into the transport position.
Transport position

H6 Status display
Status LED Off: System O.K.
Green flashing: System O.K. and
isocentric mode activated
Green: System O.K. and
image oriented movement
activated
Orange: Message
Orange flashing: Message,, acknowledge with
Reset
Red: Error Message
Red flashing: Error message, acknowledge
with Reset

CS3.SM_GB_A 9.7
Control Unit

Figure 9-2 Component Placement on PCB M370A

9.8 CS3.SM_GB_A
Control Unit

Figure 9-3 Circuit Diagram M370A

CS3.SM_GB_A 9.9
Control Unit

Figure 9-4 Circuit Diagram M370A

9.10 CS3.SM_GB_A
Control Unit

Figure 9-5 Circuit Diagram M370A

CS3.SM_GB_A 9.11
Control Unit

9.4 Data Interface

The data interface is a serial interface according to RS 422 standard (CCITT/X27 resp.
DIN 66259, part 3). The receiving lines are terminated in a 68 Ohm impedance to
ground.
The interface parameters are the following:
Transmission: asynchronous, full duplex
Transfer rate: 9600 Baud
Character length: 8 bit
Frame: 1 start bit, 1 stop bit, parity even
Handshake: software handshake XON/XOFF
The XOFF signal is sent by the controller as soon as the receiving buffer is almost full.
The receiving buffer has a total depth of 20 bytes.

Table 9-2 Pin Assignment on 1X2 and at the Control Unit

PIN Assignment at Function PIN Assignment at the Control


1X2 Unit
A +24 VDC 1
B GND 2
C T(A) 8
D T(B) 9
E R(A) 10
F R(B) 11
G MovEn 12
H screen screen + green/black
J HandTrig 2 3
K HandTrig 2 4
L HandTrig 1 5
M HandTrig 1 6

9.12 CS3.SM_GB_A