Control Unit 9.3 Control Unit
Control Unit 9.3 Control Unit
The most important functions of the Dornier Compact Delta II, Delta II are executed
via the control unit. The control unit is integrated in a housing and connected to the
Dornier Compact Delta II, Delta II via a helix cable.
Figure 9-1 Control Unit of the Dornier Compact Delta II, Delta II with the most important
Elements
CS3.SM_GB_A 9.3
Control Unit
9.4 CS3.SM_GB_A
Control Unit
CS3.SM_GB_A 9.5
Control Unit
Releases shock-waves
S17 (shock-wave release continues as long as button is
Shock-wave release pressed)
button
S10 blocked
S11 blocked
9.6 CS3.SM_GB_A
Control Unit
H6 Status display
Status LED Off: System O.K.
Green flashing: System O.K. and
isocentric mode activated
Green: System O.K. and
image oriented movement
activated
Orange: Message
Orange flashing: Message,, acknowledge with
Reset
Red: Error Message
Red flashing: Error message, acknowledge
with Reset
CS3.SM_GB_A 9.7
Control Unit
9.8 CS3.SM_GB_A
Control Unit
CS3.SM_GB_A 9.9
Control Unit
9.10 CS3.SM_GB_A
Control Unit
CS3.SM_GB_A 9.11
Control Unit
The data interface is a serial interface according to RS 422 standard (CCITT/X27 resp.
DIN 66259, part 3). The receiving lines are terminated in a 68 Ohm impedance to
ground.
The interface parameters are the following:
Transmission: asynchronous, full duplex
Transfer rate: 9600 Baud
Character length: 8 bit
Frame: 1 start bit, 1 stop bit, parity even
Handshake: software handshake XON/XOFF
The XOFF signal is sent by the controller as soon as the receiving buffer is almost full.
The receiving buffer has a total depth of 20 bytes.
9.12 CS3.SM_GB_A