VP300 Calibration Instructions

1. The document provides calibration instructions for product VP300. 2. It involves preparing the vaporizer, connecting it to test tools and a gas monitor, adjusting pressures and flows, and marking concentration scales at different flow settings. 3. The dial is removed by taking off a cover with a screwdriver, then removing 3 screws underneath to separate the dial.

Uploaded by

Ignacio Nicolas
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
251 views6 pages

VP300 Calibration Instructions

1. The document provides calibration instructions for product VP300. 2. It involves preparing the vaporizer, connecting it to test tools and a gas monitor, adjusting pressures and flows, and marking concentration scales at different flow settings. 3. The dial is removed by taking off a cover with a screwdriver, then removing 3 screws underneath to separate the dial.

Uploaded by

Ignacio Nicolas
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd

Calibration instruction

Product VP300 No. VP300-1-200


1.preparation works
1.1 Pour anesthetic into one vaporizer and make sure the liquid over the lowest level. Place the vaporizer over 72 hours and test then.
1.2 Place the vaporizer in the lab not less then 4 hours. (Lab temperature: 20 °C±3°C).
1.3 Adjust the reducer of test tools to 0.4MPa.(have done in BA)

1.Vaporizer
1
2.Test tools

3.Gas monitor
3
Other tools: Pencil and rubber
2

S/5 gas
monitor Test tools

AEON/QP13-D1-ZD 第1页 共6页


Calibration instruction
Product VP300 No. VP300-1-200
2.Procedure

Gas inlet
Exhaust tube

Sample line of
gas monitor Gas outlet

outlet switch

inlet switch

Gas outlet
2.Connect gas monitor
with test tools
Gas Inlet

1.Connect gas monitor


with test tools

3. Mount the vaporizer


on backbar. Make
sure tight connection.

AEON/QP13-D1-ZD 第2页 共6页


Calibration instruction
Product VP300 No. VP300-1-200

Flowmeter
Pressure gauge

5.Open the inlet and


outlet switch,adjust
4.Open the the pressure between
flowmeter 0.04-0.1 Mpa.

Close the inlet and


outlet switch.Observe
the pressure, make
sure the pressure
drop less than 0.04
MPa/min.

AEON/QP13-D1-ZD 第3页 共6页


Calibration instruction
Product VP300 No. VP300-1-200

Flowmeter 7. Mark the zero


1L/min position contrast Δ
before open the
vaporizer.

Turn the dial time


and again, make
6. Adjust flowmeter the dial turn
to 1L/min smoothly.

8.Turn the dial 1min, When the output


concentration is steady, then make a
Anesthetic mark.(The lower concentration output,
concentration the longer steady time last.
(Sev:0~8, others :0~5)

AEON/QP13-D1-ZD 第4页 共6页


Calibration instruction
Product VP300 No. VP300-1-200

Flowmeter
to 5L/min 10.Turn the dial and
wait 1min
1min. When the
concentration steady
several seconds,
confirm the
corresponding marks
and modify the
9.After mark with incorrect value .
1L/min flow, wait 5
mins, switch on Refer to the allowable
gas inlet and outlet, difference. (Sev:0~8,
adjust flowmeter to
others 0~5).
5L/min,check the
mark.
mark

11. When set flow to 1L/min ,the gas flow should be 1L/min ±0.1L/min.
Turn the dial and wait for 1 min, when the concentration is steady, mark the scale from low to high.
If there is no correction, record the scale or test it according to step12 after correction.

12. When set flow to 5L/min ,the gas flow should be 5 L/min ±0.5L/min.
Turn the dial and wait for 1 min, when the concentration is steady mark the scale from low to high.
If there is no correction, record the scale or test it according to step10 after correction

13. Check the scale and record it .

AEON/QP13-D1-ZD 第5页 共6页


Calibration instruction
Product VP300 No. VP300-1-200

3、How to take off the Dial

14.Take off the cover on top of the dial


with "-" Slotted screwdriver or similar tool.

16. You can


separate the
15. After taking dial after 3pcs
off the cover, of screws were
you can see 3pcs taken out.
of screws, take
out of the screws Caution!!
with "+" cross Don't move
screwdriver. other screws,
or it will give
damage to the
vaporizer.

AEON/QP13-D1-ZD 第6页 共6页