Calibration instruction
Product VP300 No. VP300-1-200
1.preparation works
1.1 Pour anesthetic into one vaporizer and make sure the liquid over the lowest level. Place the vaporizer over 72 hours and test then.
1.2 Place the vaporizer in the lab not less then 4 hours. (Lab temperature: 20 °C±3°C).
1.3 Adjust the reducer of test tools to 0.4MPa.(have done in BA)
1.Vaporizer
1
2.Test tools
3.Gas monitor
3
Other tools: Pencil and rubber
2
S/5 gas
monitor Test tools
AEON/QP13-D1-ZD 第1页 共6页
Calibration instruction
Product VP300 No. VP300-1-200
2.Procedure
Gas inlet
Exhaust tube
Sample line of
gas monitor Gas outlet
outlet switch
inlet switch
Gas outlet
2.Connect gas monitor
with test tools
Gas Inlet
1.Connect gas monitor
with test tools
3. Mount the vaporizer
on backbar. Make
sure tight connection.
AEON/QP13-D1-ZD 第2页 共6页
Calibration instruction
Product VP300 No. VP300-1-200
Flowmeter
Pressure gauge
5.Open the inlet and
outlet switch,adjust
4.Open the the pressure between
flowmeter 0.04-0.1 Mpa.
Close the inlet and
outlet switch.Observe
the pressure, make
sure the pressure
drop less than 0.04
MPa/min.
AEON/QP13-D1-ZD 第3页 共6页
Calibration instruction
Product VP300 No. VP300-1-200
Flowmeter 7. Mark the zero
1L/min position contrast Δ
before open the
vaporizer.
Turn the dial time
and again, make
6. Adjust flowmeter the dial turn
to 1L/min smoothly.
8.Turn the dial 1min, When the output
concentration is steady, then make a
Anesthetic mark.(The lower concentration output,
concentration the longer steady time last.
(Sev:0~8, others :0~5)
AEON/QP13-D1-ZD 第4页 共6页
Calibration instruction
Product VP300 No. VP300-1-200
Flowmeter
to 5L/min 10.Turn the dial and
wait 1min
1min. When the
concentration steady
several seconds,
confirm the
corresponding marks
and modify the
9.After mark with incorrect value .
1L/min flow, wait 5
mins, switch on Refer to the allowable
gas inlet and outlet, difference. (Sev:0~8,
adjust flowmeter to
others 0~5).
5L/min,check the
mark.
mark
11. When set flow to 1L/min ,the gas flow should be 1L/min ±0.1L/min.
Turn the dial and wait for 1 min, when the concentration is steady, mark the scale from low to high.
If there is no correction, record the scale or test it according to step12 after correction.
12. When set flow to 5L/min ,the gas flow should be 5 L/min ±0.5L/min.
Turn the dial and wait for 1 min, when the concentration is steady mark the scale from low to high.
If there is no correction, record the scale or test it according to step10 after correction
13. Check the scale and record it .
AEON/QP13-D1-ZD 第5页 共6页
Calibration instruction
Product VP300 No. VP300-1-200
3、How to take off the Dial
14.Take off the cover on top of the dial
with "-" Slotted screwdriver or similar tool.
16. You can
separate the
15. After taking dial after 3pcs
off the cover, of screws were
you can see 3pcs taken out.
of screws, take
out of the screws Caution!!
with "+" cross Don't move
screwdriver. other screws,
or it will give
damage to the
vaporizer.
AEON/QP13-D1-ZD 第6页 共6页