FORCED CONVECTION DRY OVEN .
INSTRUCTION MANUAL
Equipment Name: Forced Convection Dry Oven
Model : VT-DO 100
Manufacturer : VISION TEC
Serial No. : VT-DO100 20003VM001
Tel : +82 53 382 4131~3
fax : +82 53 382 4130
Web site : WWW. VISIONTEC21. CO.KR
E-MAIL :
[email protected]비 전 텍 열풍 건조기 매뉴얼 1 / 14
Before Use.
Caution! 본 Before using this equipment,reat the Manual carefully,then
using this equipment.
본 Improper handling and use not in conformity with the
Manual reduce performance of the equipment and may
cause accident and failure. So avoid using it for any other
purpose than its peculiar object.
Regarding the guidance concerning safety of this Manual,
please fully understand the instruction prior to use without
fail. Otherwise, it may cause the accident. In the event the
user handles the equipment at random resulting in any
problem, we are not responsible for it regardless of
warranty.
Description of the Manual is subject to change for the
purpose of improving performance or function of the
About the Manual equipment.
M This Manual is the property belonging to us, and therefore
any reproduction is prohibited.
In the event this Manual becomes unreadable resulted from
pollution, please request us new manual with the serial
number of the equipment.
If there is any inconvenience in use, please contact us and
file the dissatisfied thing and convenience. We will correct
them immediately.
.
비 전 텍 열풍 건조기 매뉴얼 2 / 14
Product Warranty
Procedure According to our product registration system, the
warranty period is applied from the delivery date. So
please check the quality of the product upon receipt.
Warranty area Service provided free of charge during the
warranty period is valid only within Korea in principal.
Warranty period Warranty period is one (1) year. (1 year from the
delivered date)
Free repair service In the event you duly abide by this Manual and
warning label, etc. and any failure occurred owning to
our mistake during the warranty period, you are entitled
to free repair service.
Exception from -Damage occurred owing to your carelessness in
warranty storage, installation, operation, handling, etc.
-Failure owing to user’s error or improper repair
and modification.
-Damage or failure caused by lack of voltage
capacity or out of the range of required voltage stated
in the Manual.
- Failure caused by installation of the equipment
at the particular place without any necessary measure.
- Damage to the equipment under the situation
of Force Majeure such as natural disaster, etc.
-Failure arising from the cause judged to be out
of our responsibility or handling that is judged to be
beyond a normal method.
-Failure arising from damage of the part that is
written as consumables.
비 전 텍 열풍 건조기 매뉴얼 3 / 14
Safety Instruction
Inflammables This equipment is not the explosion proof type. Avoid
using this equipment near the object that has a
combustible or inflammable risk.
Transport
Please contact us when the equipment already installed
should be transported to the other place.
Our service provided for the transportation of this
equipment shall be charged.
Installation
Installation Check List
This equipment is required to be installed in a horizontal
manner.
This equipment installed in a declining state may cause
malfunction, and result in unexpected accident.
Avoid installing this equipment near the place exposed to
direct rays or heat generating source. Well-ventilated or
stable environment without sudden change in temperature
or humidity is recommended.
Room temperature 15℃ ~ 25℃
Relative humidity 20% ~ 80%
Power connected to this equipment is AC 3800V 3 Phase.
Power connection Don’t connect with any other power or voltage, etc.
To prevent any risk, ground wire should be connected
without fail.
비 전 텍 열풍 건조기 매뉴얼 4 / 14
Safety Instruction
Maintenance/Check Maintenance/checking works of this equipment should
ing be carried out by one who has professional knowledge
only.
This equipment is being operated under hot and humid
environment.
During maintenance/checking works, be careful not to
touch the area marked as high temperature as there is a
danger of burns.
When the equipment will not be used for a long time,
Storage please keep it with power cable connection released and
cover closed.
Avoid keeping the equipment in hot and humid area. It
may cause short circuit when installing and using it later.
Modification In the event of any modification of the equipment by the
user, we are not responsible for any failure arising from
such modification as it is out of our warranty range
비 전 텍 열풍 건조기 매뉴얼 5 / 14
Contents
1. Installation of Equipment
1-1. Precautions for Installation
2. The name of each part
3. Equipment specifications
4. Name and function of control unit
5. Confirmation and operation method before use
6. Abnormal alarms actions
7. Part List (Electricity)
비 전 텍 열풍 건조기 매뉴얼 6 / 14
1. Installation of Equipment
1-1. Precautions for Installation
[ Top of equipment ]
► Install exhaust Duct.
[Ambient Temp. ]
► 15~25℃
[Horizontal confirmation]
► If not horizontal,vibration
[Ambient humi. ]
occurs
► 80% below.
[ Power connection ]
► 380V 3phase (10A)
► Install ground connection
[ Information to prevent user safety and property damage ]
( Read the following carefully and use correctly )
• Check the voltage used.( 380V 3Phase 50Hz / 10A).
→ Make sure the power connection is loose.
→ Do not use or check the equipment with wet hands.
• Do not place high-temperature thermals,flammable sprays,flammable materials,etc.
Or use or sore them in close proximity to them
(Be at risk of explosion and fire.
• Be careful not to get foreign or metal substance in the ventilation duct or control box..
• Turn off the power and clean the equipment
(Do not spray water directly or wipe with gasoline or alcohol.)
비 전 텍 열풍 건조기 매뉴얼 7 / 14
2.The name of each part
Exhaust Duct (60A)
Circulation fan(1/2HP)
OP Panel
Main circuit breaker
Additional details regarding facility
1) One way-directional circulation (Left → Right)
비 전 텍 열풍 건조기 매뉴얼 8 / 14
3. The list of specification
Classification Item Specification
Voltage 380V 3 phase
Utility Electric current 10A
Etc. Exhaust duct 60A
Out size 1160 x 920 x 1700 mm
Dimension Chamber internal size 800 x 600 x 800 mm
Materials SUS304 & Glass wool 100T
Range of temperature MAX 250℃
Temperature
Usage temperature 200℃ ± 3℃
Heater 6KW
Circulation fan 1/2hp * 1ea
Exhaust fan -
Sensor CA – K type * 2EA
Loading Method Shelf(2ea) Insertion method
Function Door 1 door (There is no view window attached)
Tower lamp -
Temperature controller TZ4M-24S
Timer LC7
Recorder -
Primary S.Device Main ‘temperature controller’control function
(+50℃) EV2
Safety Device
Secondary S.Device SUB Temperature controller (250℃)
Tertiary S. Device Heater heating section E.G.O sense (250℃)
비 전 텍 열풍 건조기 매뉴얼 9 / 14
4. Name and function ‘Control Unit’.
Ampere Meter : Shows the amount of current in use.
( When main breaker off, only at turn off)
Temp Controller : Touch type controller (TD500)
SWITCH
1. POWER : POWER ON/0FF
2. EMERGENCY : Emergency Stop
3. FAN ON : Circulation Fan operation
4. FAN OFF: Circulation Fan stop
5. HEATER ON : Heater operation
6. HEATER OFF : Heater stop
Buzzer and Buzzer S/W
비 전 텍 열풍 건조기 매뉴얼 10 / 14
5. Confirmation and Operation method before use.
Confirmation before use.
1. Check the correct level of equipment.
2.When the ‘Main circuit Breaker’ is turned on, power is supplied to the OP inner panel
And enters the standby of operation mode.
(Temp.controller),(Timer),(OFF SW)’→Supplied power→Turn ON.
3.Check the exhaust system of the outlet for blockages.
Operation Method
1. [POWER SW] →[ON]→Power is applied to the entire OP panel to able equipment use.
※ The controller and timer is state of complete set up.
2. [FAN ON SW]→Turn ON [Circulation Fan]→Rotation.
3. [HEATER ON SW]→Turn ON [HEATER]→Turn ON→ Temperature is rise.
※ Heater does not turn on when the Fan is not turned on.
4. [TIMER SW] It works as follows.
① ON : When the temperature reaches the set temperature the Timer operates and
The passes.
- When the time of the Timer is complete,the Heater is turned off first, and
The circulating fan is operated for a period of time before stopping.
(Purpose of cooling Heater)
② OFF :Regardness of the temperature value and timer setting ,the timer does not
Proceed.( Used for continuous operation).
5. [Emergency S/W] All functions of the equipment are interrupted and the Buzzer sounds.
비 전 텍 열풍 건조기 매뉴얼 11 / 14
6. Abnormal alarms actions
1. When the equipment is not working.
- Circuit Breaker TRIP, CP cut-off,Controller,S/W→Check operation ,etc.
2. Check the as following abnormalities.
Abnormal occurrence checklist.
Error signal Causes and check items
Unable to 1. (Inside operation panel)=circuit breaker →Short
Temperature rise - Primary : Circuit breaker turn on →Reconfirmation
- Secondary : Ask our company for a check
2. The ‘out lamp’of the temperature controller does not turn on.
- Primary : Conditions where the set temperature and the current
- Value are close.
- Secondary : Ask our company for a check.
3. Complete work (Check the timer time)
- Timer reset
4. Fan OFF state
- Primary : Rework after Fan ON.
- Secondary : Ask our company for a check.
Slow 1. Check for forced discharge through exhaust unit.
temperature rise 2. Heater failure : Ask our company for a check
Fan inoperative 1. Status of completion of work
- Primary : Restart after power reset
- Secondary : Ask our company for a check
Vibration of 1. Equipment is not equivalent of horizontal
equipment - Primary : Re-check the level of horizontal using a horizontal meter
- Secondary : Ask our company for a check
Circuit breaker 1. Heater failure or short circuit in equipment
short - Primary : Check the ground condition of the equipment
- Secondary : Ask our company for a check
Buzzer alarms 1. Status of completion of work
- Timer reset
2. Temperature deviation during operation
- Temperature controller→Auto tuning act.
3. Temperature excess rise detect
- Primary : Temperature controller→Auto tuning act.
- Secondary : Ask our company for a check
other,disorder 1. Check that the emergency S/W is pressed.
비 전 텍 열풍 건조기 매뉴얼 12 / 14
Part List
NO PART NAME MODELS AND SPEC. Q REMARKS
1 Power distributor breaker ABN-53c (30A) 1 LS
2 Short circuit breaker EBN-53c (30A) 1 LS
3 Magnet MC-22b 1 LS
4 Magnet MC-9b 1 LS
5 Over load MT-32 (1-1.6) 1 LS
6 Reverse phase preventer PMR-440N7 1 Schneider
7 CP LCP32-3A 1 LS
8 Temp. controller TZ4M-24S 1 Autonics
9 Temp. controller T3S-B4RK4C 1 Autonics
10 Relay HR-705-4PL 8
11 Relay socket KMY40 11
12 Timer LC7-P61NA 1
13 Mini Timer TTM-40 3
14 SSR 1
15 Select S/W K22-47W 1
16 Push S/W K22-27R 2
17 Push S/W K22-27R 2
18 Emergency S/W K22-81R 1
19 Toggle S/W WJT-2210C 1
20 Ground terminal 5P 1 -
21 Trans former WYAU-200DW Woon young
22 Temperature sensor CA sensor 2 -
23 Heater 1KW 6 -
24 Fan Motor 1/2HP 4pole 380V 3phase 1 Simens
25 Terminal 30A , 4P 1 Carcon
26 E.G.O 320℃ 3PIN 1 E.G.O
27 Buzzer BZ-PL 1
28 Radiator plate HSM-150 1
29 Radiator fan 220V 1 phase 80 1
30
31
32
33
34
비 전 텍 열풍 건조기 매뉴얼 13 / 14