Hemotherm MR 56234-S
Hemotherm MR 56234-S
MODEL 400MR
Cincinnati Sub-Zero Products, LLC • 12011 Mosteller Road • Cincinnati, Ohio 45241, U.S.A.
www.cszmedical.com
I
SYMBOLS
Caution: Read operation
instructions and manual Water Temperature
before operating
Pump ~ AC Voltage
Cool Outlet
Heat Return
Drain
Compressor
Equipotentiality
Test Indicators
5V Remote Port
Silence Alarm
Low Limit
Danger; Risk of Electric Shock
High Heater
Disconnect Power before Servicing
Low Heater
Operating instructions or
“Important /Caution Information Risk of Explosion: Do not use in the
Presence of Flammable Anesthetics
Decreasing Temperature
Fill to Screen
Type B Equipment
Cincinnati Sub-Zero Products, Inc., reserves the right to make equipment changes and
improvements which may not be reflected in this manual.
This document may not be reproduced in whole or in part without written permission from
Cincinnati Sub-Zero Products, Inc.
CAUTION
Federal law restricts this device to use only on the order of a physician.
CAUTION
• For safe handling and use of chemicals follow manufacturer guidelines.
• Use only sterile water or water that has been passed through a filter of less
than or equal to 0.22 microns. Do not use deionized water. Do not use
alcohol. Alcohol may cause heat exchanger, blanket, and/or unit deterioration.
Do not use tap water to rinse, fill, refill or top-off water tanks.
• Do not overfill. Overfilling may result in overflow when the water in the blanket
drains back into the system when the system is turned off.
III
DANGER
• Power interruption will cause the Hemotherm unit to revert to FILL MODE Follow
instructions for desired mode to resume operation.
Failure to resume therapy could result in serious injury or death.
DANGER
The repair, calibration, and servicing of the Hemotherm unit should be performed
by qualified Medical Equipment Service Technicians, Certified Biomedical
Electronics Technicians, or Certified Clinical Engineers familiar with good repair
practices for servicing medical devices, and in accordance with instructions
contained in this manual. Improper repair can result in patient injury and
damage to the Hemotherm unit.
DANGER
Before returning the Hemotherm unit to patient use after repairs, the INITIAL SET-UP
in Section 2.1 must always be performed.
IV
WARNING
When warming blankets are used in conjunction with the Hemotherm unit, the
following applies:
• A physician's order is required for setting blanket water temperature and use of
equipment. At least every 20 minutes, or as directed by a physician, check
patient's temperature and skin condition of areas in contact with blanket; also,
check blanket water temperature. Pediatric, temperature-sensitive, patients with
vascular disease and operating room patients should be checked more
frequently. Notify the physician promptly of any change in order to avoid
serious injury or death.
• Do not place additional heat sources between the patient and blanket.
Skin damage may result.
• The area between the patient and the blanket should be kept dry to avoid injury
to patient.
• Upon receipt of the HEMOTHERM, the unit should be disinfected per the
WATER SYSTEM DISINFECTION PROCEDURE. Additionally, any unit
removed from storage should be disinfected. Failure to do so could result in
patient and/or caregiver infection.
• Immediately remove from service and clean & disinfect devices that show
discoloration or cloudiness in the fluid lines/circuits. Failure to do so could
result in patient and/or caregiver infection.
V
TABLE OF CONTENTS
Page No.
TECHNICAL HELP ............................................................................................................ 1
AUTHORIZED EUROPEAN REPRESENTATIVE: ........................................................................... 1
BEFORE YOU CALL FOR SERVICE... .............................................................................. 1
IN-WARRANTY REPAIR AND PARTS ............................................................................... 1
RECEIVING INSPECTION................................................................................................. 1
IMPORTANT SAFETY INFORMATION............................................................................. 1
SECTION 1. GENERAL INFORMATION.......................................................................... 2
1.0 GENERAL SAFETY PRECAUTIONS ..................................................................... 2
1.1 FUNCTION OF HEMOTHERM MODEL 400MR COOLER/HEATER ..................... 2
1.2 COOLING SYSTEM ............................................................................................... 2
1.3 HEATING SYSTEM ................................................................................................ 3
1.4 CIRCULATION SYSTEM ....................................................................................... 4
1.5 GENERAL OPERATION ........................................................................................ 5
1.6 GENERAL SAFETY PRECAUTIONS ..................................................................... 9
1.7 EQUIPMENT DESCRIPTION AND SPECIFICATIONS .......................................... 9
1.8 REMOTE CONTROL OPTION ............................................................................... 9
1.9 CLINICAL APPLICATIONS .................................................................................. 10
1.10 HEMOTHERM SYSTEM EQUIPMENT & ACCESORIES ..................................... 10
1.11. HEMOTHERM SYSTEM EQUIPMENT SPECIFICATIONS.................................. 12
1.12. HEMOTHERM CERTIFICATIONS ....................................................................... 13
SECTION 2. OPERATING INSTRUCTIONS .................................................................. 14
2.1 INITIAL SET-UP ................................................................................................... 14
2.2 CONNECTION PROCEDURE ............................................................................. 15
2.3 SUGGESTIONS FOR PROTECTING THE RESERVOIR FROM BACTERIA....... 16
2.4 BUILT-IN SAFETY PRECAUTIONS ..................................................................... 16
SECTION 3. GENERAL MAINTENANCE AND REPAIR ................................................ 17
3.1 GENERAL ............................................................................................................ 17
3.2 EXTERNAL CLEANING INSTRUCTIONS ........................................................... 17
3.3 MAINTENANCE OF THE WATER RESERVOIRS ............................................... 18
3.4 MAINTENANCE OF THE WATER CIRCULATING SYSTEM ............................... 19
3.5 MAINTENANCE OF THE WATER FILTER........................................................... 23
3.6 MAINTENANCE OF THE CONDENSER GRILLE ................................................ 23
3.7 MAINTENANCE OF HYPER-HYPOTHERMIA BLANKETS ................................. 25
3.8 INSTRUCTIONS FOR REPLACING HEATERS ................................................... 26
3.9 INSTRUCTIONS FOR REPLACING PUMP HOUSING OR PUMP ....................... 26
3.10 REPLACING THE SENSOR................................................................................. 27
3.11 REPLACING THE MICROPROCESSOR BOARD ............................................... 28
3.12 REPLACING THE MEMBRANE CONTROL PANEL ............................................ 29
3.13 REPLACING THE MANIFOLD ............................................................................. 29
3.14 TROUBLESHOOTING THE COMPRESSOR (115, 230 AND 240 VAC UNITS) ....... 30
3.15 TROUBLESHOOTING THE COMPRESSOR (100 VAC UNITS ONLY).................... 31
3.16 INSTRUCTIONS FOR LEAKAGE CURRENT ...................................................... 32
3.17 HI LIMIT TEST PROCEDURE .............................................................................. 32
VI
3.18 LOW LIMIT TEST PROCEDURE ......................................................................... 33
3.19 REQUIRED QUARTERLY PREVENTIVE MAINTENANCE CHECKLIST ............ 34
3.20 REQUIRED MONTHLY PREVENTIVE MAINTENANCE CHECKLIST ................. 34
3.21 TROUBLESHOOTING GUIDE ............................................................................. 35
3.22 MICROPROCESSOR BOARD ERROR CODE TROUBLESHOOTING ............... 42
SECTION 4 PARTS INFORMATION ............................................................................... 43
4.1 ORDERING INFORMATION FOR REPLACEMENT PARTS ............................... 44
4.2 RECOMMENDED REPLACEMENT PARTS INVENTORY .................................. 44
4.3 RETURNING PARTS UNDER WARRANTY ........................................................ 44
4-4. SHIPPING PARTS ............................................................................................... 45
4-5. EQUIPMENT AND ACCESSORIES ORDER INFORMATION ............................. 45
WORLDWIDE ORDER PLACEMENT ............................................................................. 60
VII
LIST OF ILLUSTRATIONS
Page No.
FIGURE 4.1 HEMOTHERM FRONT & SIDE VIEW ................................................................... 5
FIGURE 4.2 PARTS LIST FOR THE HEMOTHERM - FRONT & SIDE VIEW .................................. 6
FIGURE 4.3 HEMOTHERM – MEMBRANE CONTROL .............................................................. 7
FIGURE 4.4 PARTS LIST FOR THE HEMOTHERM MEMBRANE CONTROL PANEL 100/115V ....... 8
FIGURE 4.5 HEMOTHERM - ELECTRICAL SUB-PANEL ........................................................... 9
FIGURE 4.6 PARTS LIST FOR THE HEMOTHERM ELECTRICAL SUB-PANEL ............................ 50
FIGURE 4.7 HEMOTHERM - REAR VIEW ............................................................................ 51
FIGURE 4.8 PARTS LIST FOR HEMOTHERM - REAR VIEW .................................................... 52
FIGURE 4.9 HEMOTHERM - ELECTRICAL SCHEMATIC 115 VOLT .......................................... 54
FIGURE 4.10 HEMOTHERM ELECTRICAL SCHEMATIC 230 VOLT .......................................... 55
FIGURE 4.11 HEMOTHERM - ELECTRICAL SCHEMATIC 240 VOLT ........................................ 56
FIGURE 4.12 HEMOTHERM - ELECTRICAL SCHEMATIC 100 VOLT ........................................ 57
FIGURE 4.13 HEMOTHERM - W ATER CIRCULATION DIAGRAM ............................................. 58
FIGURE 4.14 HEMOTHERM - REFRIGERATION FLOW DIAGRAM............................................ 59
VIII
OPERATING INSTRUCTIONS HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
TECHNICAL HELP
CEpartner4U
Esdoornlaan 13, 3951 DB Maarn
The Netherlands
RECEIVING INSPECTION
After unpacking the Hemotherm System, be sure to inspect the system for concealed
damage. Retain all packing material and carefully describe or photograph any damage.
Notify the carrier at once and ask for an inspection (in writing). Failure to do this within 15
days may result in loss of claim. Do not return the equipment to Cincinnati Sub-Zero.
Call our Medical Technical Service department for further instructions.
Refer to this manual for instructions and caregiver information. Read and understand all
precautionary information before using, prescribing, or servicing the Hemotherm unit.
In order to minimize risk of contamination and/or infection, the unit should be disinfected
per the WATER SYSTEM CLEANING & DISINFECTION PROCEDURE prior to its first
use and at a minimum on a quarterly basis.
1
OPERATING INSTRUCTIONS HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
To provide the patient maximum safety during the use of the HEMOTHERM
Cooler/Heater system, a thorough knowledge and understanding, and the correct
application and operating use of the system is required. Each person who is responsible
for the use or the direction of use of the system, such as physicians, nurses, technicians
and operators, must read and understand this operating manual, and all precautions and
warnings prior to use. It is recommended that this manual be reviewed at least semi-
annually as a refresher to safe operation and application.
The CSZ HEMOTHERM Model 400MR Cooler/Heater system is used to lower/raise the
blood temperature and/or maintain blood temperature, as required, through conductive
heat transfer. The conductive heat transfer is accomplished through any type of Blood
Oxygenator/Heat Exchanger unit, along with the use of a blanket under the patient for the
re-warming phase, if desired. The complete system usually comprises of the
HEMOTHERM Cooler/Heater unit and reusable or disposable blanket on the O. R. Table-
Adult, Pediatric, or Infant size. Available blankets are : the reusable PLASTIPAD®
blankets (polyurethane) and the disposable MAXI-THERM®, MAXI-THERM® LITE, and
TEMP PAD® blankets.
The operator of the unit regulates the desired fluid temperature. Frequent observation of
the actual blood temperature is required.
With a capacity to lower the water temperature at a 2°C/minute rate, a reduction of blood
temperature is produced. It is important to remember that there is no direct relationship
2
OPERATING INSTRUCTIONS HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
between this and the rate of change in the blood temperature. The re-circulation pump
and water flow circuitry have been designed for high flow rate to maximize the blood-to-
water heat transfer efficiency.
The cooling system's temperature controller allows the operator to set the temperature
range from +3°C to +32°C. An independent Low Limit Safety in the microprocessor board
provides the low limit safety function and is factory preset at 2°C.
When in the cool mode, the 300 Watt heater will pre-warm the heat bath to the desired
temperature set for that reservoir. While in the heat mode, the bath is heated with both
heaters or 1500 watts of heat. The water temperature will increase at a rate of
approximately 2.5°C/minute. It is important to remember that there is no direct
relationship between this and the rate of change in the blood temperature.
The high temperature safety is designed to shut off the heaters at 44°C. If the controller
should happen to fail at the 42°C set point, a high temperature LED warning light on the
control panel will illuminate and alarm at 44°C. If this should occur, the system should be
inspected by the hospital's maintenance or engineering department. The heating system
will come back on after approximately a 1°C temperature drop.
* 230/240 VAC units have one 800 watt heater and one 1200 watt heater.
3
OPERATING INSTRUCTIONS HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
Selection of the operation mode (heat or cool) determines which water flow circuit is
used.
HEATING MODE: Pump draws water from the warm water reservoir and
circulates it through the water filter to the external outlet couplings connected to
the device in use (Blood Oxygenator/Heat Exchanger and/or blanket). Returning
water flows back into the same (warm) water reservoir.
COOLING MODE: Pump draws water from the cooled water reservoir and
circulates it through the water filter to the connected external device(s). Returning
water flows back into the cooled water reservoir.
CAUTION
Use only sterile water or water that has been passed through a filter of less than or equal
to 0.22 microns. Do not use deionized water. Do not use alcohol. Alcohol may cause
heat exchanger, blanket, and/or unit deterioration. Do not use tap water to rinse, fill, refill
or top-off water tanks.
* Flow rate will vary depending upon brand and model number of heat exchanger used
and also upon length of tubing leading to and from heat exchanger.
4
OPERATING INSTRUCTIONS HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
4. Check the level of the water in both fluid reservoirs to make certain they are
filled to where water can be seen in bottom of the strainer. (Heat reservoir
holds 6 qts. (5.7 liters) and the Cool reservoir holds 8 qts. (7.6 liters). It is
necessary to add water after each use to replace water left in disposable heat
exchanger and Hyper-Hypothermia blanket. Use only sterile water or water
that has been passed through a filter of less than or equal to 0.22 microns.
6. Inspect the hospital grade plug for bent or missing prongs. Insert the plug
into a properly grounded, securely mounted receptacle. Grounding reliability
can only be achieved when connected to an equivalent hospital grade
receptacle. Do not bypass the third or grounding prong. An electrical hazard
may result if it is bypassed, removed, or otherwise rendered useless.
7. Lay the Hyper-Hypothermia blanket flat on the O.R. table with hose
attachment coming off the table as close as possible to where the
HEMOTHERM unit will be placed. Cover the blanket with a sheet, or bath
blanket.
5
OPERATING INSTRUCTIONS HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
8. Test-Indicators: Energize and De-energize all display segments and all red
LED's. Press the power switch to the "ON" position.
NOTE: The HEMOTHERM unit will go through a two second self calibration
procedure with both displays blank and then through a fill mode for
approximately forty seconds. The displays will show FI on the Heat side and
LL on the Cool side spelling out FILL. This allows the unit to calibrate itself
and prime the pump automatically.
B. Select Cool - cool display will flash when below 10°C until pump switch is
activated. Then it will display actual water temperature and the
compressor will start up.
C. If you wish to change the setting, depress the "Temp Set" switch. The
display will flash the previous set point temperature or 3°C if the unit has
just been turned on.
NOTE: You have five seconds to depress one of the arrows or the
temperature readout will return to the actual reservoir temperature.
Depress either the "up" or "down" arrow to raise or lower the set point
temperature to the desired setting.
D. Depress the "Cool" switch. When ready to circulate the water, depress the
pump switch.
WARNING
The temperature cannot be set below 10°C without activating the "Pump" switch. If
the operator attempts to do so, the set point temperature will flash until the "Pump"
switch is activated. Then it will display the actual temperature and the compressor
will activate.
NOTE: You have five seconds to depress one of the arrows or the
temperature readout will return to the actual reservoir temperature.
Depress either the "up" or "down" arrow to raise or lower the set point
temperature to the desired setting.
6
OPERATING INSTRUCTIONS HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
NOTE: When the compressor cycles off, both heaters come on until the
temperature of the heat reservoir is within 1°C of the set point. If the
compressor cycles back on, only the 300 watt heater remains on for 100
and 115 VAC units. (See * for 230/240 VAC units.)
F. The unit will not cool below 10°C unless the pump is activated.
G. Select the "HEAT" switch. Depress the "Temp Set" switch. The display
will flash the previous set point temperature or 42°C if the unit has just
been turned on.
NOTE: You have five seconds to depress one of the arrows or the
temperature readout will return to the actual reservoir temperature.
Depress either the "up" or "down" arrow to raise or lower the desired
setting. Depress the heat switch.
9. After the cool bath is preconditioned, proceed to connect your heat exchanger
to the 1/2" self-sealing Hansen fittings.
NOTE: You may want to turn off the pump or the flow indicator shut-off valve
first if you choose not to circulate cool water through the heat exchanger at
this time.
11. At this point, you may choose to maintain normal body temperature (Step 12)
or you may choose to cool the blood (Step 13).
NOTE: Both the 300 watt low heater and the 1200 watt high heater will cycle
on and off to maintain the set point temperature for 115 VAC units. (See * for
230/240 VAC units.)
* 230/240 VAC units have an 800 watt and a 1200 watt heater.
7
OPERATING INSTRUCTIONS HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
13. Cool the blood by setting the desired fluid temperature. Using the cool mode
"temp set", you may choose 3°C for better efficiency. The "COOL"
temperature displayed is the fluid temperature circulating through the heat
exchanger.
NOTE: The compressor will cycle on and off to maintain the set point
temperature.
14. Maintain final hypothermia temperature by resetting the Cool set point
temperature to 10°C and turning off the pump when you reach the
hypothermia temperature desired. If the temperature should tend to rise,
reactivate the pump allowing the cool water to circulate through the heat
exchanger and as a result bring the blood temperature down again.
NOTE: Preset your hot bath temperature to the desired initial rewarming
temperature.
15. For 115 VAC units, rewarm by activating the "HEAT" switch and resetting the
set point temperature to the desired water temperature. Both the 300 watt
and the 1200 watt heaters will activate giving you a maximum of 1500 watts
of heat to raise the hot bath fluid temperature to the set point temperature. At
this point, if desired, connect your Hyper-Hypothermia blanket to the 1/8"
Hansen fittings of the HEMOTHERM unit to maximize rewarming efficiency.
(See * for 100 VAC and 230/240 VAC units.)
16. If you happen to overshoot your desired "COOL" bath temperature, you may
push the "HEAT" switch for five to ten seconds allowing the water in the water
lines and heat exchanger to empty into the "HEAT" bath and fill up with warm
water. Switching back to the "COOL" mode will allow warm water in the water
lines and heat exchanger to be dumped into the "COOL" bath, in turn
warming the temperature of that bath a few degrees (1°- 6°C) depending on
the difference in temperature between the two reservoirs. The same can be
done if you overshoot your desired hot bath temperature.
17. To discontinue operation, simply turn off the power switch and disconnect all
lines leading to the heat exchanger and blanket from the HEMOTHERM unit.
* 100/115 VAC units have a 300 watt and a 1200 watt heater. 230/240 VAC
units have one 800 watt and one 1200 watt heater.
8
OPERATING INSTRUCTIONS HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
If water is found leaking into or around the HEMOTHERM unit prior to or during the
operating procedure discussed here, immediately disconnect power to the unit and
correct the malfunction before proceeding.
Before performing any disassembly procedure, unplug power cord from receptacle.
High and low temperature limiting thermostats are preset at the factory and should
perform satisfactorily for the life of the HEMOTHERM unit. Except for refrigeration
component repairs, the repair and servicing of the HEMOTHERM unit requires no special
tools. However, no attempt should be made to perform any of the repairs or service
procedures outlined in this manual unless the proper skills and knowledge are
possessed.
All HEMOTHERM units manufactured after serial # 891-9327M are equipped with a
remote control port. The 25 pin sub-d port connection is located on the right side of the
unit, above the circuit breaker. If the remote control option is purchased, the "assembly"
will include a mating 25 foot cable, the remote control unit, and a mounting clamp.
To install the remote control option, first turn the HEMOTHERM unit power switch off.
Mount the remote control in a convenient place utilizing the universal mounting clamp
provided. Connect the cable between the two devices and lock in place by tightening the
thumb screws attached to the cable. The remote control will operate the same as the
main membrane control panel located on the HEMOTHERM.
NOTE: The display on the remote control will be "blank" during the initial forty second
fill mode.
9
OPERATING INSTRUCTIONS HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
The HEMOTHERM Dual Reservoir Cooler/Heater was designed specifically for use in the
extracorporeal circuit. The unit is a self contained temperature control device designed to
deliver chilled or heated water to the blood heat exchanger in the cardiopulmonary
bypass circuit.
There are many variables that affect the heating or cooling of the patient in the
extracorporeal circuit. Some of these include the patient's weight, blood flow, gas flow
and oxygenator/heat exchangers. The HEMOTHERM Cooler/Heater has been designed
to provide high water flow rates at low pressure for safe and highly efficient operation with
all currently available heat exchangers.
10
OPERATING INSTRUCTIONS HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
11
OPERATING INSTRUCTIONS HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
PHYSICAL ELECTRICAL
Maximum Pressure:
12.5 PSI - Heat Exchanger SAFETY
Fluid Cooling Range: 32°C to 3°C Fluid Temp Control Accuracy: ± .5°C
Fluid Heating Range: 25°C to 42°C 1 year parts and labor if returned to CSZ
13
OPERATING INSTRUCTIONS HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
Steps 1-10 should be followed the first time the HEMOTHERM unit is set-up. For
subsequent operation, refer to General Operation Instructions, Section 1.5.
WARNING
Upon receipt of the HEMOTHERM, the unit should be disinfected per the WATER SYSTEM DISINFECTION
PROCEDURE. Additionally, any unit removed from storage should be disinfected. Failure to do so could
result in patient and/or caregiver infection.
A. HEMOTHERM unit.
B. Extracorporeal circuit with blood heat exchanger.
C. Hyper-Hypothermia blanket (optional).
D. 17 qts. (16 liters) of sterile water or water that has been passed through a
filter of less than or equal to 0.22 microns. [The cool reservoir holds 8 qts.
(7.6 liters), the heat reservoir holds 6 qts. (5.7 liters), the blanket (adult)
holds 2 qts. (1.9 liters) and approximately 1.5 qts. (1.4 liters) will be
required to fill the connecting lines and the heat exchanger]. No alcohol
should be used.
3. Lift the reservoir lid and remove individual reservoir covers. Fill the cool
reservoir with 8 qts. (7.6 liters) of sterile water or water that has been passed
through a filter of less than or equal to 0.22 microns and the heat reservoir
with 6 qts. (5.7 liters) of sterile water or water that has been passed through a
filter of less than or equal to 0.22 microns. The level should be
approximately 1" from the top of each reservoir or just covering the
bottom of the strainer.
7. Attach the heat exchanger and the blanket connecting lines (if desired) to the
HEMOTHERM unit.
8. Read operating instructions on the key pad before starting the unit.
Familiarize yourself with the name and location of all features and controls
shown.
9. Fill the heat exchanger and blanket by pressing the power switch to the ON/”I”
position. The HEMOTHERM will go through a self-calibration procedure with
both displays blank and then through a fill mode. The displays will show FI on
the Heat side and LL on the Cool side spelling out FILL. Then activate the
heat/cool mode switch and then the pump switch. Add more sterile water or
water that has been passed through a filter of less than or equal to 0.22
microns to the reservoir of the mode selected when filling.
1. Take the set of 1/2" Hansen self-sealing fittings and attach them to (2), 1/2"
I.D. plastic tubes long enough to reach from the HEMOTHERM unit to the
Heat Exchanger. The shorter the length of tubing, the greater the flow rate.
The female fitting should be attached to the tube leading to the Heat
Exchanger's water inlet and the male fitting should be attached to the Heat
Exchanger's water outlet.
2. After attaching the tubes to the heat exchanger, attach the 1/2" Hansen self-
sealing fittings to the matched fittings on the lower right side of the
HEMOTHERM unit (see Figure 4.1).
3. Attach the 1/8" Hansen self-sealing fittings leading from the Hyper-
Hypothermia blanket to the matching set of 1/8" Hansen self-sealing fittings
on the lower right hand side of the HEMOTHERM unit.
4. The air in the lines will be bled out after the pump switch is activated. Check
the water level of the reservoirs after filling connecting water lines.
15
OPERATING INSTRUCTIONS HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
1. A Low Limit Thermostat will automatically shut off the compressor and the
compressor indicator light at 2°C ±.5°C. The low temp light will come on and
the alarm will sound. However, the circulating pump continues to operate. In
the unlikely event that the thermostat should malfunction, the water will
freeze, provided no antifreeze solution has been added to prevent it. This will
also stop the circulation of water.
2. A High Limit Thermostat will automatically shut off the heaters should the
water temperature reach 44°C. Again, the circulating pump will continue to
operate. If this happens, a high temperature warning light on the front panel
will also illuminate and the alarm will sound. Then, the unit should be
returned to the maintenance or engineering department to inspect the high
temperature safety system.
3. Circuit Breaker - The HEMOTHERM unit (right side) is equipped with an over-
current protection circuit breaker built into the power switch to protect the unit
and the hospital wiring system against possible overload.
16
GENERAL MAINTENANCE HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
3.1 GENERAL
The HEMOTHERM Cooler/Heater unit is designed and built to be field serviceable. The
repair, calibration and servicing of this unit requires no special tools, except for
refrigeration repairs. However, no attempt should be made to perform any of these
procedures unless the proper skills and knowledge are possessed. Repair or service of
the HEMOTHERM by qualified hospital personnel will not void the warranty of the unit.
WARNING
Before performing any disassembly procedure, be sure the power cord is unplugged
from the receptacle.
All internal operating components are readily exposed by removing the rear panel. This
is accomplished by removing the five retaining screws around the perimeter of the panel.
Pull rear panel away from the unit.
Access to the condenser grille is attained by removing the air intake vent at the front of
the HEMOTHERM unit. To do this, loosen the four screws securing the vent to the unit.
The HEMOTHERM unit and Remote Control are constructed of powder coated steel with
a plastic membrane keypad. For cleaning and disinfecting, use tuberculocidal wipes.
Avoid 100% alcohol and other strong, undiluted disinfectants. These may cause staining
of the device’s outer skin. Wipe down the entire exterior of the device paying extra
attention to crevices on the device.
NOTE: Surfaces are to remain wet for the duration specified by the wipe’s instructions.
Use wipes that have bleach (sodium hypochlorite) as an active ingredient to
avoid discoloration of the unit.
17
GENERAL MAINTENANCE HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
CAUTION
For safe handling and use of chemicals follow manufacturer guidelines.
The reservoirs should be maintained at least once a month, more often if possible.
To empty the reservoirs, first make sure the power switch is in the OFF/"O” position.
Then, attach the 1/4" female coupling of the drain tube over either of the 1/4" male fittings
marked "DRAIN" on the side of the unit. Allow gravity to drain the water from the one
reservoir and then repeat for the other one. If draining unit into a container, be sure that
the container has a capacity of at least five gallons (13.3 liters).
Remove the reservoir covers by lifting the cover at the strainer handle. Prior to refilling,
clean the underside of both reservoir lids using a tuberculocidal wipe. Clean any residue
from the bottom and sides of the empty reservoir and wipe down with tuberculocidal
wipes. Pay extra attention to the areas not in the water pathways, such as the lid of each
reservoir, and the top section of the water reservoir.
NOTE: Surfaces are to remain wet for the duration specified by the wipe’s instructions.
Use wipes that have bleach (sodium hypochlorite) as an active ingredient to
avoid discoloration of the unit.
Place the reservoir covers back on the proper reservoir. Refill the reservoir with sterile
water or water that has been passed through a filter of less than or equal to 0.22 microns.
The capacity of the reservoirs is - 6 qts (5.7 liters) for the heat reservoir and 8 qts. (7.6
liters) for the cool reservoir. Under all circumstances, the reservoirs should be filled
before starting the HEMOTHERM unit. It is not necessary to drain the reservoirs after
each use.
18
GENERAL MAINTENANCE HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
CAUTION
For safe handling and use of chemicals follow manufacturer guidelines.
Always drain the HEMOTHERM to a sanitary drain because bio-contaminants
may be present in the unit’s water supply.
Use only sterile water or water that has been passed through a filter of less than
or equal to 0.22 microns. Do not use deionized water. Do not use alcohol.
Alcohol may cause blanket and/or unit deterioration. Do not use tap water to
rinse, fill, refill or top-off water tanks.
The following procedure disinfects the fluid circuit in these products. The required
tools/supplies are:
• Drain hose (P/N 93807) and Bypass hose, if applicable, (P/N 93817)
• Agency registered tuberculocidal wipes1
• Cleaning Agent2: Prolystica® 2X Concentrate Enzymatic Presoak and Cleaner
• Disinfectant3: 6% bleach or 8.25% bleach
• Sterile water or water that has been passed through a filter of less than or equal to
0.22 microns
• pH strips
• Appropriate AC electrical power.
Procedure:
1. Turn unit off (“O” position).
2. Remove any blood heat exchangers, Hyper-Hypothermia blankets, and hoses that
are connected to the unit.
3. Drain the water from each reservoir as described in MAINTENANCE OF THE
WATER RESERVOIRS.
4. Using a tuberculocidal wipe, wipe down the reservoirs and the underside of both
reservoir lids. Pay extra attention to the areas not in the water pathways, such as
the lid of each reservoir, and the top section of the water reservoir.
5. Drain the water from each reservoir as described in MAINTENANCE OF THE
WATER RESERVOIRS.
1
ARTG registered, EPA Registered, Health Canada licensed
2
Australian Register of Therapeutic Goods (ARTG) #151419
3
Health Canada Drug Identification Number (DIN) 6% bleach (DIN 02459116) or 8.25%
bleach (DINs 02459108 or 02438100)
19
GENERAL MAINTENANCE HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
6. Using a tuberculocidal wipe, wipe down the reservoirs and the underside of both
reservoir lids. Pay extra attention to the areas not in the water pathways, such as
the lid of each reservoir, and the top section of the water reservoir.
NOTE: Surfaces are to remain wet for the duration specified by the wipe’s instructions.
Use wipes that have bleach (sodium hypochlorite) as an active ingredient to
avoid discoloration of the unit.
20
GENERAL MAINTENANCE HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
21. RETURNING TO SERVICE: Ensure that the water reservoirs are filled with sterile
water or water that has been passed through a filter of less than or equal to 0.22
microns. Turn off the unit and proceed to Step 24.
22. FOR DRY STORAGE: If unit is being placed in dry storage, do the following steps.
23. Drain the water from each reservoir as described in the Section 3.3.
24. When all fluid has been removed from the unit, disconnect the drain hose(s) and
wipe unit clean.
26. Remove any accessories attached to the unit. Document unit maintenance per
Cincinnati Sub-Zero and hospital protocol. (Maintenance records can be found in
Sections 3.19 & 3.20).
When you are ready to return unit from storage, repeat the above disinfection procedure.
Perform all the checks as described in this Manual.
22
GENERAL MAINTENANCE HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
WARNING
Always unplug the unit before accessing internal components during service. Failure to
unplug the unit could result in electric shock.
The HEMOTHERM circulating system includes a water filter designed to clear the line of
particulate matter as the water is pumped through the system. The water filter assembly
is a plastic T-shaped fitting that intersects the hose from the pump to the outlet manifold.
A stainless steel wire-mesh screen is located inside the clear plastic cap of the water filter
assembly.
Once every three months, or more often if deemed necessary, the water filter should be
disassembled and cleaned. To do so, the rear enclosure panel must be removed. The
cap of the water filter assembly should then be unscrewed. The wire mesh and the
plastic cap should then be disinfected. The parts are then to be reassembled.
3. Drain both of the unit’s water reservoirs as described in the Section 3.3.
4. Remove the five screws holding on the rear enclosure panel. Guide the panel
outward towards the floor.
23
GENERAL MAINTENANCE HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
6. Firmly grasp the notched rim of the cap of the assembly and remove. The
cap and hose will contain water if the unit was used prior to cleaning. The
wire mesh may be lodged in the top of the fitting or it may be resting in the
plastic cap.
NOTE: Water in hose lines and cap present a slip and fall hazard
8. Clean the wire mesh and the plastic cap with a tuberculocidal wipe. Be careful
not to lose the black O-ring in the rim of the plastic cap.
NOTE: Surfaces are to remain wet for the duration specified by the wipe’s instructions. Use
wipes that have bleach (sodium hypochlorite) as an active ingredient to avoid
discoloration of the unit.
9. Replace the wire mesh in the plastic cap and position the wire mesh, O-ring
and plastic cap under the fitting.
NOTE: If the black O-ring is not present in the plastic cap when the cap is returned to the unit,
water can leak from the device.
NOTE: If the plastic cap is not screwed back on at the correct angle, or is not screwed on tightly,
water can leak from the device.
Cool air is taken in through the grille at the front of the HEMOTHERM unit. Warm air is
expelled through the bottom of the unit. Both the air intake and outlet must be kept clear
when the machine is in operation.
If the condenser grille becomes covered with dust and lint, the cooling capacity of the unit
will be reduced. AT LEAST EVERY MONTH THE CONDENSER SHOULD BE
CLEANED. To do this, first loosen the four screws securing the condenser grille over the
condenser intake. Then, remove accumulated dust or lint with a brush or vacuum. Wipe
down both sides of the condenser grille with a tuberculocidal wipe. Finally, replace
condenser grille and tighten all four screws.
24
GENERAL MAINTENANCE HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
NOTE: Surfaces are to remain wet for the duration specified by the wipe’s instructions. Use
wipes that have bleach (sodium hypochlorite) as an active ingredient to avoid
discoloration of the unit.
This section describes the general maintenance and for the reusable and disposable
blankets. General maintenance tasks include cleaning, draining, and storing the blankets.
For further maintenance information on Cincinnati Sub-Zero blankets refer to the blanket’s
corresponding Instructions for Use.
NOTE: Reusable blankets have the potential to re-contaminate the device that is connected to
them.
NOTE: Surfaces are to remain wet for the duration specified by the wipe’s instructions. Use
wipes that have bleach (sodium hypochlorite) as an active ingredient to avoid
discoloration of the unit.
Thoroughly rinse product with clear water to remove any residue from cleaning solutions.
DO NOT use gas sterilization or autoclaving for cleaning and disinfecting the blankets.
To drain the water from the reusable blankets shut off the power to the unit. The blanket
will drain by gravity back into the unit. Allow it to drain as much as possible before
disconnecting the blanket from the unit or removing from bed, stretcher, or table.
Refer to Section 3.4 for the internal disinfection of reusable blankets and hoses.
To store PlastiPad blankets with a permanently attached hose, loosely coil the hose
lengthwise into the center of the blanket. Fold the blanket lengthwise into the center, 1/3
from the left side and 1/3 from the right side. Do not fold the blankets width-wise, as doing
so will crimp the tubing.
To store PlastiPad blankets without a permanently attached hose, gently fold the corner
containing the tubing towards the center of the blanket. Roll or fold blanket loosely
avoiding sharp creases.
Store the Gelli-Roll Pads in a dry area, in flat or rolled position (never folded). Keep pads
out of direct sunlight. Keep cabinet doors and other sharp objects from coming in contact
with pad surfaces at all times.
4. Remove rear enclosure panel by removing the five screws securing the panel
to the unit.
5. Locate heaters at lower portion of the right reservoir as seen when viewing
the unit from the back. The 300 watt heater is on the left and the 1200 watt
heater on the right.*
10. Refill reservoir with sterile water or water that has been passed through a
filter of less than or equal to 0.22 microns and check for leaks.
* 100 VAC units have one 300 watt heater and one 1200 watt heater.
230/240 VAC units have one 800 watt heater and one 1200 watt heater.
26
GENERAL MAINTENANCE HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
3. Locate the pump housing towards the lower back side of the inside of the
HEMOTHERM unit.
8. Insert the fitting from the defective pump housing in step 7 above into the
identical orifice of the new pump housing. Seal the fitting with Teflon tape or
equivalent.
9. Secure new pump housing by replacing the seven screws removed in step 6
above. Do not tighten screws sequentially in either clockwise or
counterclockwise direction. Tighten one screw, then a screw opposite it, etc...
10. Remount pump with the three bolts attached to the base.
3. Drain proper reservoir, see Section 3.3, depending on which sensor is being
replaced. There is one dual thermistor temperature sensing probe located in
each water reservoir.
4. Remove rear enclosure panel by removing the five screws securing the panel
to the unit.
27
GENERAL MAINTENANCE HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
5. Disconnect wire plug located approximately 6" from the sensor fitting.
Remove the sensor from the proper reservoir. For location, see Figure 4.7.
9. Refill water reservoir (see Section 2.1) and check for leaks.
3. Unseat the membrane control panel by removing the four screws securing it
to the HEMOTHERM unit.
4. Disconnect the 12 pin, 9 pin, 10 pin, 25 pin sub D and ground connectors
leading from the control panel and the electrical compartment.
5. Unseat the microprocessor board by removing the six nuts and washers
securing it to the membrane control panel. Remove the ground wire.
6. Take the new microprocessor board and reconnect the 12 pin, 9 pin, 10 pin,
25 pin sub D and ground wire.
7. Install the new microprocessor board in its original position by connecting the
ground wire and replacing and tightening the six nuts and washers.
8. Secure the face panel in its original position by replacing and tightening the
four screws.
28
GENERAL MAINTENANCE HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
3. Unseat the membrane control panel by removing the four screws securing it
to the unit.
4. Unseat the microprocessor board by removing the six nuts and washers
securing it to the membrane control panel. Remove the ground wire.
6. Secure the membrane control panel in its original position by replacing and
tightening the four screws.
4. Remove rear enclosure panel by removing the five screws securing it to the
unit.
5. Locate manifolds in the lower left side when viewing the unit from the back.
7. Unscrew both outlet/male fittings and both inlet/female fittings (see Figure
4.1) from outside of unit.
8. Remove the whole manifold. Replace with new manifold. Replace all fittings
with the male fittings on the top and the female fittings on the bottom and seal
using Teflon tape or equivalent.
29
GENERAL MAINTENANCE HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
10. Refill reservoirs with sterile water or water that has been passed through a
filter of less than or equal to 0.22 microns.
3. Remove rear enclosure panel by removing the five screws securing it.
4. Locate compressor (see Figure 4.7) directly under the reservoir. Remove
screws from around the cover protecting the overload protector and the
compressor relay directly under the starting capacitor.
A. Using a screw driver, unclip the starting capacitor from its mounting
bracket.
B. Remove the two (2) screws from the compressor electric box cover and
set the cover aside.
C. Remove one (1) capacitor lead wire from terminal S of the starting relay
and the other lead wire from terminal S of the electrical junction board.
D. Install the new capacitor.
E. Attach new capacitor wires to both S terminals.
F. Snap the new capacitor into place.
G. Reinstall the electric box cover.
A. Remove the two (2) screws from the compressor's electric box cover and
set the cover aside.
B. Remove the three (3) wires from the "slide-on" terminals of the starting
relay.
C. Remove the two (2) screws that mount the starting relay to the
compressor's electric box.
D. Install the new starting relay using the screws you removed in step c.
E. Reconnect the three (3) wires to the "slide-on" terminals of the new
starting relay.
F. Reinstall the electric box cover.
A. Remove the two (2) screws from the compressor's electric box cover and
set the cover aside.
30
GENERAL MAINTENANCE HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
B. Remove one overload protector lead wire from terminal C of the electrical
junction board and the other lead wire from "slide-on" terminal C of the
compressor motor leads.
C. Using a screwdriver, disengage the overload protector's mounting clip and
remove the overload protector.
D. Install the new overload protector and fasten it with the mounting clip.
E. Reconnect the two lead wires to the C terminals from which you removed
them in step B.
F. Reinstall the electric box cover.
3. Remove rear enclosure panel by removing the five screws securing it.
4. Locate compressor (see Figure 4.7) directly under the reservoir . Remove
screws from around cover protecting the overload protector and compressor
relay, directly under the starting capacitor.
A. Using a screwdriver, remove the clip from each end of the compressor's
electric box cover and set the cover aside.
B. Remove the two capacitor lead wires from the terminals on the left side of
the starting relay.
C. Remove the starting relay by pulling toward you with a side to side motion.
D. Install the new starting relay.
E. Reconnect both the capacitor lead wires and the compressor wires to their
proper terminals.
F. Reinstall the electric box cover and spring clip.
31
GENERAL MAINTENANCE HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
Excessive leakage current is most commonly caused by a defective heater, but other
components can also fail in such a way as to increase leakage current. To find the
source of excessive leakage current, suspect components should be sequentially
disconnected until the source of the high leakage is isolated.
3. Remove the four self-tapping screws from the membrane control panel.
4. Arrange the control panel in such a manner that the electric compartment is
easily accessible.
5. Locate the two (2) solid state relays for the 1200 watt heater, number 11 and
12 in Figure 4.5.
6. Electrically jump (short) the primary side of relay number 11 and relay number
12 using test leads with alligator clips or a comparable item. On relay number
11, jump between terminals #1 and #9. On relay #12, jump between terminals
#10 and #2.
32
GENERAL MAINTENANCE HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
9. After the unit goes through the FILL mode, press the heat and pump keys. The
unit will now heat past its set point until it reaches 44°C ± .5°C. At this point,
the unit will indicate HI TEMP (both visual and audible).
3. Remove the four self-tapping screws from the membrane control panel.
4. Arrange the control paneling in such a manner that the electric compartment
is easily accessible.
5. Locate the two (2) solid state relays for the compressor, number 4 and 5 in
Figure 4.6.
6. Electrically jump (short) the primary side of relay number 4 and relay number
5 using test leads with alligator or a comparable item. On relay number 4,
jump between terminals #4 and #2. On relay number 5, jump between
terminals #1 and #3
9. After the unit goes through the FILL mode, press the COOL and PUMP keys.
The unit will now cool down past its set point until it reaches 2°C ± .5°C. At
this point, the unit will indicate LOW TEMP (both visual and audible).
33
GENERAL MAINTENANCE HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
1. External cabinet and controls in good condition (i.e. no dents or missing parts).
2. All warning labels properly affixed.
3. Quick disconnect couplings are tight, straight, and not leaking.
4. Power cord (i.e. no cuts or exposed wire) and plug (i.e. no bent or missing pins) are in
good condition.
5. All indicator lights are operative (i.e. heat & cool modes, heaters, pump, power).
6. Clean water filter (See Section 3.5).
7. Clean & disinfect water system (See Section 3.4).
8. Leakage current check (all readings should be less than 300 micro amps for 115V units
and 500 micro amps for 230V units).
• OFF normal polarity
• OFF reverse polarity
• ON normal polarity (heat)
• ON reverse polarity (heat)
• ON normal polarity (cool)
• ON reverse polarity (cool)
9. Perform low limit safeties check. (See Section 3.18)
10. Perform high limit safeties check. (See Section 3.17)
11. Check Condition of blankets, hoses, couplings (check for leaks). (See Section 3.7)
34
GENERAL MAINTENANCE HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
Check when
Hospital I.D. No. _________________________ completed
_________________________________ __________________
Signature of Inspector Date of Inspection
35
GENERAL MAINTENANCE HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
36
GENERAL MAINTENANCE HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
37
GENERAL MAINTENANCE HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
Troubleshoot or replace
the microprocessor board.
Refrigeration system If electrical power is not
inoperative - water in reaching the compressor,
COOL water reservoir troubleshoot circuitry (See
not being chilled. wiring diagrams Figs. 4.9.
and 4.12.)
If electrical power is
reaching compressor,
troubleshoot compressor
per Sec. 3.13.
NOTE: DO NOT
OVERCHARGE.
38
GENERAL MAINTENANCE HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
Defective compressor
control relays (2SSR).
6. Blood Water flow partially Check for partial
oxygenator/ obstructed obstruction in external
heat exchanger liquid circuit or partially
cool-capacity engaged quick disconnect
low fittings.
Cool water reservoir Switch to HEAT mode for
partially or fully iced up ten seconds.
Water flow control Rap sharply. Clean or
solenoid valve is stuck replace as soon as
possible.
Partially blocked cooling Move external equipment
airflow from front of cabinet air
grille.
39
GENERAL MAINTENANCE HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
40
GENERAL MAINTENANCE HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
41
GENERAL MAINTENANCE HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
42
GENERAL MAINTENANCE HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
43
PARTS INFORMATION HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
44
PARTS INFORMATION HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
All parts are covered by a one year warranty. To replace parts during the warranty
period*, send the part prepaid to:
To qualify for credit, warranty parts should be tagged with the following information:
* You must first obtain a RMA number by calling the factory prior to returning an item.
.
NOTE: Shipping damage will be the responsibility of the shipper. Insure if necessary.
Equipment and accessories can be ordered directly from our authorized dealer for you
area. For information on your dealer, you can call the Customer Service Department at
the factory, at (513) 772-8810 or 800-989-7373.
45
PARTS INFORMATION HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
46
PARTS INFORMATION HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
Figure 4.2 Parts List for the HEMOTHERM - Front & Side View
47
PARTS INFORMATION HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
100/115V
230/240V
48
PARTS INFORMATION HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
Figure 4.4 Parts List for the HEMOTHERM Membrane Control Panel 100/115V
49
PARTS INFORMATION HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
50
PARTS INFORMATION HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
51
PARTS INFORMATION HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
53
PARTS INFORMATION HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
54
PARTS INFORMATION HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
55
PARTS INFORMATION HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
56
PARTS INFORMATION HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
57
PARTS INFORMATION HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
58
PARTS INFORMATION HEMOTHERM MODEL 400MR,
OPERATION/TECHNICAL MANUAL
100 V ONLY
59
Worldwide Order Placement