1502A 1504 Datasheet
1502A 1504 Datasheet
Battery option
If you want freedom from AC power in the
field or on the plant floor, order Model 2502
and we’ll install a DC power board in your
Tweener. Then you can connect your own
12-volt DC power or order Hart’s 9313 Bat-
tery Pack. Our battery gives you three to
eight hours between charges. It includes a
charger and a nylon pouch with a belt clip.
Calibration choices
Each Tweener and its accompanying
probe (sold separately) have their own
individual calibration reports. Overall sys-
tem error can be calculated from the indi-
vidual errors, rendering the added cost of
system data unnecessary. However, for
3 those requiring it, system data is avail-
able at two or more temperatures of your
choice. (See calibration options on page
● Two Tweeners to choose from—reading PRTs or thermistors 164.)
● Battery packs available
● Best price/performance package
One of Hart’s best-selling products is the subranges of the ITS-90, you simply enter
Tweener thermometer, and there’s a rea- the coefficients directly into your
son. No other company, not one, has a Tweener.
thermometer that comes close to the per- Each thermometer comes complete
formance and features of the Tweener for with an RS-232 interface for automation
anywhere near its price. of temperature data collection, calibra-
tions, or process control functions. An
1502A Tweener PRT Readout IEEE-488 interface is available as an
The 1502A Tweener features accuracy up option.
to ±0.006 °C (the 1504 is even more ac- The 1502A is calibrated digitally using
curate, up to ±0.002 °C). In addition, it the front-panel buttons. You never have
The thermistor version of the “Tweener” gives you
reads 100-ohm, 25-ohm, and 10-ohm to open the box to calibrate it. This cali- more variety in sensor configurations and even
probes, has a resolution of 0.001 °C bration protocol further reduces the cost higher accuracy over a limited temperature range.
across its entire range, and is the small- of the 1502A. It goes where you go and
est unit in its class. It also has an optional works the way you want it to.
battery pack for completely portable
operation. 1504 Tweener Thermistor Readout
Each Tweener is programmable to If you need more accuracy in a limited
match a probe’s constants for maximum temperature range, the Model 1504
linearity and accuracy. All probe con- Tweener gives it to you as a thermistor
stants and coefficients are programmed readout. Thermistors are less fragile than
through simple, front-panel keystrokes. PRTs and less likely to be impacted by
Temperature is displayed in °C, °F, K, or mechanical shock. Thermistors are more
resistance in ohms. sensitive to temperature, have faster re-
The 1502A accurately measures the sponse times, and come in many shapes
resistance of the probe and then converts for different applications.
the resistance to a temperature value us- Typical accuracy of a 1504 is
ing its built-in algorithms. ±0.002 °C with a resolution of 0.0001 °C.
For convenience, the 1502A reads the
common industrial grade IEC-751 or Software
“385” ALPHA RTD without any program- With our 9934 LogWare, both Tweener
ming. Enter the actual R0 and ALPHA of models may be used for real-time data
the individual probe for increased accu- acquisition. Collect data and analyze it
racy. For maximum accuracy, use the ITS- graphically or statistically. Additionally,
90 formulas. The Tweener accepts the Tweeners may be used as reference ther-
subranges 4 and 6 through 11. mometers with our MET/TEMP II soft-
ITS-90 formulas reside in the ware. (See our software section starting
Tweener’s firmware. If your probe has on page 82.)
been calibrated for any of the above
52 Thermometry
Tweener Thermometer Readouts
Ordering Information
1502A Tweener PRT Readout 1935 System Cal Report, Thermistors
1504 Tweener Thermistor Readout (see pages 164 to 167)
2502 DC Power Option 1930 System Cal Report, PRT
(see pages 164 to 167)
2505 Spare Connector
See pages 60 to 80 for a selection of probes
2506 IEEE Option to use with Tweeners and other Hart
2507 Mini-Printer readouts.
2508 Serial Cable Kit
9934-S LogWare, Single Channel, Single
User
9934-M LogWare, Single Channel, Multi
User
9313 Battery Pack
9301 Carrying Case, fits Tweener and
12 in probe
www.hartscientific.com 53