Ak95S Error Code List
Ak95S Error Code List
C CB......................................................................................................................................................... 1
C CC......................................................................................................................................................... 6
C CF ....................................................................................................................................................... 13
C CO....................................................................................................................................................... 36
C FRFF.................................................................................................................................................. 81
P PTRA .................................................................................................................................................. 83
C Cx 000................................................................................................................................................. 85
C CSYS .................................................................................................................................................. 86
C EBUF .................................................................................................................................................. 91
C SYS and P SYS................................................................................................................................... 92
P ENV .................................................................................................................................................... 95
P POW.................................................................................................................................................... 96
P PPR...................................................................................................................................................... 97
P PSYS ................................................................................................................................................... 98
P PREC................................................................................................................................................... 99
P SDT ................................................................................................................................................... 100
W WAD ............................................................................................................................................... 102
W WPR ................................................................................................................................................ 103
W WSYS .............................................................................................................................................. 104
? UDEF................................................................................................................................................. 105
Computer start-up sequence ………………………………………………………………………..…106
This error code list is intended for use together with AK 95 / AK 95 S, program version 12.20
Please report to Gambro ITS the occurrence of all “Internal Software Error”, use fax number below.
Important!
This error code list is valid for all versions of electronics. Whenever a component reference have
changed a note to N.E. (new electronics) is supplied.
0 1 2 3 4 98 99
Array index
In the AK 95 S, all variables of interest are placed in arrays. The array names refer
to the type of variables stored in the arrary. An example: the FRI-array stores
variables from the fluid control.
ex. FRI(FI_DEGASS$PRESS)
F is source TRACO name
R is RAM
I is the type of the variable, in this case an integer
(FI_DEGASS$PRESS) is the index in the FRI array
B: Blood
F: Fluid
O: Operator's
C: Common
Range of variables
Byte 0 +255
Integer -32768 +32767
Word 0 +65535
Logic FALSE or TRUE
Real ±8.43E-37 ±3.37E+38
The AK 95 S Arrays
Control system Protective system
Operator's comm vars:
ORB(OB_ ) CPU C CPU P
ORD(OD_) EEPROM vars:
ORI(OI_ ) EB(EEB_ )
ORL(O_ ) EI(EEI_ )
ORR(OR_ ) EL(EEL_ )
ORW(OW_ ) ER(EER_ ) I2C
OF(OF_) EW(EEW_ )
Mailed from P to C
Fluid vars: Common vars: Protective vars:
FF(FF_ ) CRL(CL_) PE (P to C event)
FRD(FD_) CRI(CI_) PRL(P_) CPU W
Watchdog
FRB(FB_ ) CRB(CB_) PRB(PB_)
FRI(FI_ ) CRR(CR_) PRI(PI_)
FRL(F_ ) CRW(CW_) PRW(PW_)
FRR(FR_ ) CRD(CD_) PRR(PR_)
FRW(FW_ ) CF(CF_)
C CB 010 002 ESTIMATED AND MEASURED HEPARIN VOL. DIFFER TOO MUCH
Conditional error, confirm by zero-set heparin flow
C CFF 006 004 TOO BIG DIFFERENCE BETWEEN HPG AND DEGASS PRESSURE
Unconditional error, blood pump available
C CO 072, Conductivity
C CO 072 000 CASE ERROR
Transient error
C CO 074, UF Profiling
C CO 074 000 CASE ERROR
Transient error
C CO 080, Attention
C CO 080 000 CASE ERROR
C CO 096, Profiling
C CO 096 000 CASE ERROR
Transient error
C COFF 088 004 BLOOD LEAK DETECTOR TEST ERROR, VALUE TOO HIGH
Unconditional error, blood pump available
C COFF 088 005 BLOOD LEAK DETECTOR TEST ERROR, VALUE OUT OF LIMIT
Unconditional error, blood pump available
C COFF 088 006 BLOOD LEAK DETECTOR TEST ERROR, UNSTABLE OFF
Unconditional error, blood pump available
C COFB 088 008 BLOOD PUMP TEST ERROR, PUMP NOT RUNNING (FIRST TEST)
Unconditional error, blood pump not available
C COFB 088 009 BLOOD PUMP TEST ERROR, PUMP NOT RUNNING (SECOND TEST)
Unconditional error, blood pump not available
C COFB 088 010 BLOOD PUMP TEST ERROR, PUMP NOT STOPPED (CPU P)
C COFB 088 011 BLOOD PUMP TEST ERROR, PUMP NOT RUNNING
ROTATION INDICATOR, (FIRST CHECK)
Unconditional error, blood pump not available
C COFB 088 012 BLOOD PUMP TEST ERROR, PUMP NOT RUNNING
ROTATION INDICATOR, (SECOND CHECK)
Unconditional error, blood pump not available
Unconditional error, blood pump not available
C COFB 088 013 BLOOD PUMP TEST ERROR, PUMP NOT RUNNING
TACHO SIGNALS MISSING, (FIRST CHECK)
Unconditional error, blood pump not available
C COFB 088 015 BLOOD PUMP TEST ERROR, PUMP NOT STOPPED (CPU W)
Unconditional error, blood pump not available
C COFB 088 019 BUZZER TEST ERROR, BUZZER NOT OFF (CPU P)
Unconditional error, blood pump not available
C COFF 088 030 FLOW SWITCH ERROR, FLOW DETECTED (INVA CLOSED)
Unconditional error, blood pump available
C COFF 088 031 FLOW SWITCH ERROR, NO FLOW (INLET PRESSURE OK)
Unconditional error, blood pump available
C COFF 088 035 HEATER TEST ERROR, HEATER ENERGIZED (INVA CLOSED)
Unconditional error, blood pump available
C COFF 088 037 HEATER TEST ERROR, HEATER NOT ENERGIZED (CPU C)
Unconditional error, blood pump available
C COFF 088 038 HEATER TEST ERROR, HEATER NOT ENERGIZED (INVA OPEN)
Unconditional error, blood pump available
C COFF 088 039 HEATER TEST ERROR, HEATER NOT ENERGIZED (CPU P)
Unconditional error, blood pump available
C COFB 088 040 HIGH VENOUS PRESSURE TEST ERROR, TOO HIGH VEN. PRESS.
Unconditional error, blood pump not available
C COFB 088 041 HIGH VENOUS PRESSURE TEST ERROR, TOO BIG DIFFERENCE
Unconditional error, blood pump not available
C COFB 088 043 PRIMING DETECTOR TEST ERROR, CURRENT OUT OF LIMIT
Unconditional error, blood pump not available
C COFB 088 046 PRIMING DETECTOR TEST ERROR, NO BLOOD DETECTED (CPU P)
Unconditional error, blood pump not available
C COFB 088 047 PRIMING DETECTOR TEST ERROR, NO BLOOD DETECTED (CPU W)
Unconditional error, blood pump not available
C COFB 088 050 SAFETY RELAY TEST ERROR, REFERENCE CURRENT TOO HIGH
Unconditional error, blood pump not available
C COFF 088 062 DIVA/TAVA TEST ERROR, DIALYSIS PRESSURE OUTSIDE LIMITS
IN THE LOW PRESSURE TEST (SECOND TEST)
Unconditional error, blood pump available
C COFF 088 063 DIVA/TAVA TEST ERROR, DIALYSIS PRESSURE OUTSIDE LIMITS
IN THE HIGH PRESSURE TEST (FIRST TEST)
Unconditional error, blood pump available
C COFF 088 065 DIVA/TAVA TEST ERROR, LEAKAGE IN LOW PRESSURE TEST
Unconditional error, blood pump available
C COFF 088 066 DIVA/TAVA TEST ERROR, DIALYSIS PRESSURE OUTSIDE LIMITS
IN LOW PRESSURE TEST (FIRST TEST)
Unconditional error, blood pump available
C COFF 088 067 DIVA/TAVA TEST ERROR, DIALYSIS PRESSURE OUTSIDE LIMITS
IN HIGH PRESSURE TEST (SECOND TEST)
Unconditional error, blood pump available
C COFB 088 069 VENOUS CLAMP TEST ERROR, CLAMP NOT OPEN AT START
Unconditional error, blood pump not available
C COFB 088 070 VENOUS CLAMP TEST ERROR, CLAMP NOT OPEN (CPU P)
Unconditional error, blood pump not available
C COFB 088 071 VENOUS CLAMP TEST ERROR, CLAMP NOT OPEN (CPU W)
Unconditional error, blood pump not available
C COFB 088 072 VENOUS CLAMP TEST ERROR, CLAMP NOT CLOSED (CPU P)
Unconditional error, blood pump not available
C COFB 088 073 VENOUS CLAMP TEST ERROR, CLAMP NOT CLOSED (CPU W)
Unconditional error, blood pump not available
C COFF 088 084 FAILURE WHEN REGULATING HPG PRESSURE (DIVA/TAVA test)
Unconditional error, blood pump available
C COFF 088 087 FAILURE WHEN REGULATING HPG PRESSURE (DIVA/TAVA test)
C COFF 088 088 FAILURE WHEN REGULATING HPG PRESSURE (DIVA/TAVA test)
Unconditional error, blood pump available
C COFF 088 089 FAILURE WHEN REGULATING HPG PRESSURE (DIVA/TAVA test)
Unconditional error, blood pump available
C COFF 088 090 FAILURE WHEN REGULATING HPG PRESSURE (DIVA/TAVA test)
Unconditional error, blood pump available
C COFF 088 092 FAILURE WHEN REGULATING TEMPERATURE (high temp. test)
C COFF 088 093 FAILURE WHEN REGULATING TEMPERATURE (high temp. test)
C COFF 088 094 FAILURE WHEN REGULATING TEMPERATURE (low temp. test)
C COFF 088 099 DIVA OPEN FAILURE DURING VALVES CURRENT TEST
Unconditional error, blood pump not available
C COFF 088 100 DIVA P CLOSED FAILURE DURING VALVES CURRENT TEST
Unconditional error, blood pump not available
C COFF 088 101 DIVA C CLOSED FAILURE DURING VALVES CURRENT TEST
Unconditional error, blood pump not available
C COFF 088 102 BYVA OPEN FAILURE DURING VALVES CURRENT TEST
Unconditional error, blood pump not available
C COFF 088 103 BYVA C OPEN FAILURE DURING VALVES CURRENT TEST
Unconditional error, blood pump not available
C COFF 088 104 BYVA P OPEN FAILURE DURING VALVES CURRENT TEST
Unconditional error, blood pump not available
C COFF 088 105 TAVA OPEN FAILURE DURING VALVES CURRENT TEST
Unconditional error, blood pump not available
C COFF 088 106 EVVA OPEN FAILURE DURING VALVES CURRENT TEST
Unconditional error, blood pump not available
C COFF 088 110 DRVA OPEN FAILURE DURING VALVES CURRENT TEST
Unconditional error, blood pump not available
C COFF 088 111 REVA OPEN FAILURE DURING VALVES CURRENT TEST
Unconditional error, blood pump not available
C COFF 088 112 ZEVA OPEN FAILURE DURING VALVES CURRENT TEST
Unconditional error, blood pump not available
C COFF 088 113 RIVA OPEN FAILURE DURING VALVES CURRENT TEST
Unconditional error, blood pump not available
C COFF 088 114 INVA OPEN FAILURE DURING VALVES CURRENT TEST
Unconditional error, blood pump not available
P PTRA 000 101 ERROR READING PARAMETERS FROM CONDUCTIVITY CELL P (COND)
Unconditional error, blood pump not available
P PTRA 000 102 ERROR READING PARAMETERS FROM CONDUCTIVITY CELL P (TEMP)
Unconditional error, blood pump not available
P PTRA 000 103 ERROR READING PARAMETERS FROM DIALYSIS PRESSURE TRANSDUCER
Unconditional error, blood pump not available
P PTRA 000 106 ERROR READING PARAMETERS FROM ARTERIAL PRESSURE TRANSDUCER
Unconditional error, blood pump not available
P SDT 004 001 CPU C DOES NOT SEND SHUT DOWN ACKNOWLEDGE
P SDT 004 002 CPU W DOES NOT SEND SHUT DOWN ACKNOWLEDGE
Initialization. of I/O ports and Initialization. of I/O ports and Internal memory and register
timers. timers. verification. (2)
Cold or warm start? (1) Cold or warm start? (1) Program verification. (3)
If warmstart jump to A If warmstart jump to A
CPU register verification. (2) CPU register verification. (2) Data memory verification. (4)
Data memory init. (5)
Program verification. (3) Program verification. (3)
SRI
SRL
SRP
SRR
SRB
P 12.20
Page: 1
P 12.20
Page: 2
P 12.20
Page: 3
P 12.20
modes
Page: 4
P 12.20
by the converter.
1 ( FDTRC$SELECT$PROC ), Scale_Factor_1
2 ( FDTRC$HEATER$ENABLE$CTRL$PROC ) Scale_Factor_1,
3 ( FDTRC$SAGS$TEST$PROC ), Scale_Factor_1,
4 ( FDTRC$SAGS$TEST$DRAIN$PROC ), Scale_Factor_1,
5 ( FDTRC$VALVE$TEST$PROC ), Scale_Factor_1,
6 ( FDTRC$DISINFECTION$PROC ), Scale_Factor_1,
7 ( FDTRC$CONC$FILLUP$PROC ), Scale_Factor_1,
8 ( FDTRC$CENTR$FILLUP$PROC ), Scale_Factor_1,
Page: 5
P 12.20
9 ( FDTRC$DIL$FILLUP$PROC ), Scale_Factor_1,
10 ( FDTRC$CLEAN$CART$PRIM$PROC ), Scale_Factor_1,
11 ( FDTRC$CLEAN$CART$FILL$PROC ), Scale_Factor_1,
12 ( FDTRC$CITRIC$DWELL$PROC ), Scale_Factor_1,
13 ( FDTRC$RINSE$PROC ), Scale_Factor_1,
14 ( FDTRC$HEAT$DIS$TEST$PROC ), Scale_Factor_1,
15 ( FDTRC$HEAT$DIS$FLUSH$PROC ), Scale_Factor_1,
16 ( FDTRC$DRAIN$PROC ), Scale_Factor_1,
17 ( FDTRC$DISINFECTION$TIME$PROC ) Scale_Factor_1,
18 ( FDTRC$STICK$A$RESOLVER$PROC ), Scale_Factor_1,
19 ( FDTRC$STICK$B$RESOLVER$PROC ), Scale_Factor_1,
20 ( FDTRC$BIC$LATCH$RESOLVER$PROC) Scale_Factor_1,
21 ( FDTRC$CHEM$FILLUP$RDY$LATCH$PROC ), Scale_Factor_1,
22 ( FDTRC$DISINF$CTRL$PROC ), Scale_Factor_1,
23 ( FDTRC$DRAIN$BEF$CHEM$DIS$PROC Scale_Factor_1,
24 ( FFTRC$FCH$MAIN$PROC ), Scale_Factor_1,
25 ( FFTRC$DISINFECT$VALVES$PROC ), Scale_Factor_1,
26 ( FFTRC$DEGASS$PR$TEST$PROC ), Scale_Factor_1,
27 ( FSTRC$UF$CALIBRATION$PROC ), Scale_Factor_1,
28 ( FSTRC$DRAIN$REF$VOL$PROC ), Scale_Factor_1,
29 ( FSTRC$UF$CALIB$PREP$PROC ), Scale_Factor_1,
30 ( FSTRC$UF$MAIN$FL$OFFSET$PROC ) Scale_Factor_1,
31 ( FSTRC$UF$CH1$FLOW$CAL$PROC ), Scale_Factor_1,
32 ( FSTRC$UF$DIFF$CAL$PROC ), Scale_Factor_1,
33 ( FSTRC$VERIFY$UF$PROC ), Scale_Factor_1,
34 ( FSTRC$UF$ISOL$VERIFY$PROC ), Scale_Factor_1,
35 ( @FFCNSTATE(FF_VARFLOWMOTOR ), Scale_Factor_1,
Page: 6