0% found this document useful (0 votes)
966 views88 pages

IFC 100 IFC 100 IFC 100 IFC 100: Signal Converter For Electromagnetic Flowmeters

Uploaded by

Tinto Ten
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
966 views88 pages

IFC 100 IFC 100 IFC 100 IFC 100: Signal Converter For Electromagnetic Flowmeters

Uploaded by

Tinto Ten
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 88

IFC 100 Handbook

Signal converter for electromagnetic flowmeters


: IMPRINT ::::::::::::::::::::::::::::::::::::::::::
IFC 100

All rights reserved. It is prohibited to reproduce this documentation, or any part thereof, without
the prior written authorisation of KROHNE Messtechnik GmbH & Co. KG.

Subject to change without notice.

Copyright 2007 by KROHNE Messtechnik GmbH & Co.KG - Ludwig-Krohne-Straße 5 - 47058


Duisburg

2 www.krohne.com 01/2008 • 4000041002 - MA IFC 100 R02 en


IFC 100 CONTENTS

1 Safety instructions......................................................................... 6
1.1 Safety instructions from the manufacturer......................................................... 6
1.1.1 Copyright and data protection ................................................................................... 6
1.1.2 Disclaimer.................................................................................................................. 6
1.1.3 Product liability and warranty ................................................................................... 7
1.1.4 Information concerning the documentation ............................................................. 7
1.1.5 Display conventions ................................................................................................... 8
1.2 Safety instructions for the operator .................................................................... 8
2 Instrument description.................................................................. 9
2.1 Scope of delivery .................................................................................................. 9
2.2 Instrument description ...................................................................................... 10
2.2.1 Wall version ............................................................................................................. 11
2.3 Nameplates ........................................................................................................ 12
2.3.1 Nameplate (example) .............................................................................................. 12
3 Installation................................................................................... 13
3.1 Notes on installation.......................................................................................... 13
3.2 Storage ............................................................................................................... 13
3.3 Installation specifications .................................................................................. 13
3.4 Transport............................................................................................................ 14
3.5 Mounting of the compact version ...................................................................... 14
3.6 Mounting the wall-mounted housing, remote version...................................... 15
3.6.1 Wall mounting.......................................................................................................... 15
4 Electrical connections ................................................................. 17
4.1 Safety instructions ............................................................................................. 17
4.2 Important notes on electrical connection ......................................................... 17
4.3 Electrical cables for remote device versions, notes ......................................... 18
4.3.1 Notes on signal cable A ........................................................................................... 18
4.3.2 Notes on field current cable C ................................................................................ 18
4.3.3 Using other signal cables ........................................................................................ 19
4.4 Preparing the signal and field current cables................................................... 20
4.4.1 Signal cable A (type DS 300), construction ............................................................. 20
4.4.2 Preparing signal cable A, connection to signal converter...................................... 21
4.4.3 Length of signal cable A .......................................................................................... 23
4.4.4 Preparing field current cable B, connection to signal converter ........................... 24
4.4.5 Prepare signal cable A, connect to measuring sensor........................................... 25
4.4.6 Preparing field current cable C, connection to measuring sensor ........................ 27
4.5 Connecting the signal and field current cables ................................................ 28
4.5.1 Connecting the signal and field current cables to the signal converter, remote versi-
on ................................................................................................................................. 29
4.5.2 Connection diagram for signal and field current cable .......................................... 31

01/2008 • 4000041002 - MA IFC 100 R02 en www.krohne.com 3


CONTENTS IFC 100

4.6 Grounding the measuring sensor...................................................................... 32


4.6.1 Classical method ..................................................................................................... 32
4.7 Connecting the power ........................................................................................ 33
4.8 Description of the outputs ................................................................................. 35
4.8.1 Current output ......................................................................................................... 35
4.8.2 Pulse and frequency output .................................................................................... 36
4.8.3 Status output and limit switch................................................................................. 37
4.9 Electrical connection of the outputs.................................................................. 37
4.9.1 Electrical connection of the outputs ....................................................................... 38
4.9.2 Laying electrical cables correctly ........................................................................... 39
4.10 Connection diagrams of outputs........................................................................ 40
4.10.1 Description of the electrical symbols ..................................................................... 40
4.10.2 Basic outputs ........................................................................................................... 41
4.10.3 HART® connection................................................................................................... 44
5 Start-up ....................................................................................... 46
5.1 Switching on the power...................................................................................... 46
5.2 Starting the signal converter ............................................................................. 46
6 Operation ..................................................................................... 47
6.1 Display and operator input elements ................................................................ 47
6.1.1 Time-out functions .................................................................................................. 48
6.1.2 Display in measuring mode..................................................................................... 48
6.1.3 Display for selection of menu and functions........................................................... 48
6.1.4 Display for setting of data and functions ................................................................ 49
6.1.5 Display after change of data and functions............................................................. 49
6.2 Program structure ............................................................................................. 50
6.3 Function tables .................................................................................................. 51
6.3.1 Menu A, quick setup ................................................................................................ 51
6.3.2 Menu B, test............................................................................................................. 52
6.3.3 Menu C, setup .......................................................................................................... 54
6.3.4 Set free units ........................................................................................................... 67
6.3.5 Reset counter in Quick Setup menu........................................................................ 68
6.3.6 Deleting error messages in the Quick Setup menu................................................ 68
6.4 Status messages and diagnostic information ................................................... 69
7 Service ......................................................................................... 72
7.1 Spare parts availability ...................................................................................... 72
7.2 Availability of services........................................................................................ 72
7.3 Returning the device to the manufacturer ........................................................ 73
7.3.1 General information ................................................................................................ 73
7.3.2 Form (for copying) to accompany a returned instrument ...................................... 74
7.4 Disposal.............................................................................................................. 74

4 www.krohne.com 01/2008 • 4000041002 - MA IFC 100 R02 en


IFC 100 CONTENTS

8 Technical data ............................................................................. 75


8.1 Technical data .................................................................................................... 75
8.2 Flow tables ......................................................................................................... 79
8.3 Accuracy ............................................................................................................. 81
8.4 Dimensions and weights.................................................................................... 82
8.4.1 Housing .................................................................................................................... 82
8.4.2 Mounting plate, wall-mounted version ................................................................... 85
9 KROHNE Product Overview ......................................................... 88

01/2008 • 4000041002 - MA IFC 100 R02 en www.krohne.com 5


1 SAFETY INSTRUCTIONS IFC 100

1.1 Safety instructions from the manufacturer


1.1.1 Copyright and data protection

The contents of this document have been created with great care. Nevertheless, we provide no
guarantee that the contents are correct, complete or up-to-date.

The contents and works in this document are subject to German copyright. Contributions from
third parties are identified as such. Reproduction, processing, dissemination and any type of use
beyond what is permitted under copyright requires written authorisation from the respective
author and/or the manufacturer.

The manufacturer tries always to observe the copyrights of others, and to draw on works created
in-house or works in the public domain.

The collection of personal data (such as names, street addresses or e-mail addresses) in the
manufacturer's documents is always on a voluntary basis whenever possible. Whenever
feasible, it is always possible to make use of the offerings and services without providing any
personal data.

We draw your attention to the fact that data transmission over the Internet (e.g. when
communicating by e-mail) may involve gaps in security. It is not possible to protect such data
completely against access by third parties.

We hereby expressly prohibit the use of the contact data published as part of our duty to publish
an imprint for the purpose of sending us any advertising or informational materials that we have
not expressly requested.

1.1.2 Disclaimer

The manufacturer will not be liable for any damage of any kind by using its product, including,
but not limited to direct, indirect, incidental, punitive and consequential damages.

This disclaimer does not apply in case the manufacturer has acted on purpose or with gross
negligence. In the event any applicable law does not allow such limitations on implied warranties
or the exclusion of limitation of certain damages, you may, if such law applies to you, not be
subject to some or all of the above disclaimer, exclusions or limitations.

Any product purchased from the manufacturer is warranted in accordance with the relevant
product documentation and our Terms and Conditions of Sale.

The manufacturer reserves the right to alter the content of its documents, including this
disclaimer in any way, at any time, for any reason, without prior notification, and will not be liable
in any way for possible consequences of such changes.

6 www.krohne.com 01/2008 • 4000041002 - MA IFC 100 R02 en


IFC 100 SAFETY INSTRUCTIONS 1

1.1.3 Product liability and warranty

The operator shall bear responsibility for the suitability of the flowmeters for the specific
purpose. The manufacturer accepts no liability for the consequences of misuse by the operator.
Improper installation and operation of the flowmeters (systems) will cause the warranty to be
void. The respective "Standard Terms and Conditions" which form the basis for the sales
contract shall also apply.

1.1.4 Information concerning the documentation

To prevent any injury to the user or damage to the device it is essential that you read the
information in this document and observe applicable national standards, safety requirements
and accident prevention regulations.

If this document is not in your native language and if you have any problems understanding the
text, we advise you to contact your local the manufacturer office for assistance. The
manufacturer can not accept responsibility for any damage or injury caused by
misunderstanding of the information in this document.

This document is provided to help you establish operating conditions, which will permit safe and
efficient use of this device. Special considerations and precautions are also described in the
document, which appear in the form of underneath icons.

01/2008 • 4000041002 - MA IFC 100 R02 en www.krohne.com 7


1 SAFETY INSTRUCTIONS IFC 100

1.1.5 Display conventions

The following symbols are used to help you navigate this documentation more easily:

WARNING!
These warning signs must be observed without fail. Even only partial disregarding such
warnings can result in serious health damage, damage to the device itself or to parts of the
operator ’s plant.

DANGER!
This symbol designates safety advice on handling electricity.

CAUTION!
These warnings must be observed without fail. Even only partial disregarding such warnings can
lead to improper functioning of the device.

LEGAL NOTICE!
This symbol designates information on statutory directives and standards.

NOTE!
This symbol designates important information for the handling of the device.

• HANDLING
This symbol designates all instructions for actions to be carried out by the operator in the
specified sequence.
i CONSEQUENCE
This symbol designates all important consequences of the previous actions.

1.2 Safety instructions for the operator


WARNING!
In general, devices from the manufacturer may only be installed, commissioned, operated and
maintained by properly trained and authorized personnel.
This document is provided to help you establish operating conditions, which will permit safe and
efficient use of this device.

8 www.krohne.com 01/2008 • 4000041002 - MA IFC 100 R02 en


IFC 100 INSTRUMENT DESCRIPTION 2

2.1 Scope of delivery


NOTE!
Inspect the cartons carefully for damage or signs of rough handling. Report damage to the
carrier and to your local office.

NOTE!
Check the packing list to make sure that you have received your complete order.

NOTE!
Please check on the device nameplates, that the device is supplied according to your order.
Check for the correct mains voltage printed on the nameplate. If not, contact your local
representative for advice.

Figure 2-1: Scope of delivery


1 Device in the version as ordered
2 Documentation (calibration report, Quick Start directions, CD-ROM with product documentation for measuring sensor
and signal converter)
3 Signal cable (only for remote version)

01/2008 • 4000041002 - MA IFC 100 R02 en www.krohne.com 9


2 INSTRUMENT DESCRIPTION IFC 100

2.2 Instrument description


Electromagnetic flowmeters are exclusively suitable for measurement of flow rates and the
conductivity of electrically conductive liquid media.

Your measuring device is supplied ready for operation. The factory settings for the operating
data have been made in accordance with your order specifications.

The following versions are available:


• Compact version (the signal converter is mounted directly on the measuring sensor)
• Remote version (electrical connection to the measuring sensor via field current and signal
cable)

Figure 2-2: Device versions


1 Compact 45° version
2 Compact 0° version
3 Measuring sensor with outlet box
4 Wall-mounted version

10 www.krohne.com 01/2008 • 4000041002 - MA IFC 100 R02 en


IFC 100 INSTRUMENT DESCRIPTION 2

2.2.1 Wall version

Figure 2-3: Structure of wall-mounted version


1 Open housing cover
2 Terminal compartment for power with safety cover (shock-hazard protection)
3 Connection terminals of the outputs
4 GDC bus interface (for Service only)
5 Connection terminals of the field current cable
6 Connection terminals of the signal cable
7 Connection screw for grounding the housing
8 Cable entries; example for remote version (for compact version the two outer cable entries are omitted)

01/2008 • 4000041002 - MA IFC 100 R02 en www.krohne.com 11


2 INSTRUMENT DESCRIPTION IFC 100

2.3 Nameplates
NOTE!
Please check on the device nameplates, that the device is supplied according to your order.
Check for the correct mains voltage printed on the nameplate. If not, contact your local
representative for advice.

2.3.1 Nameplate (example)

Figure 2-4: Example of a nameplate


1 Manufacturer
2 Power supply information
3 Software version
4 Tag name
5 Approvals-related pressure and temperature thresholds
6 GK/GKL values (measuring sensor constants); size (mm /inches); field frequency; protection category; materials of
parts in contact with media
7 Product designation, serial number and date of manufacture

12 www.krohne.com 01/2008 • 4000041002 - MA IFC 100 R02 en


IFC 100 INSTALLATION 3

3.1 Notes on installation


NOTE!
Inspect the cartons carefully for damage or signs of rough handling. Report damage to the
carrier and to your local office.

NOTE!
Check the packing list to make sure that you have received your complete order.

NOTE!
Please check on the device nameplates, that the device is supplied according to your order.
Check for the correct mains voltage printed on the nameplate. If not, contact your local
representative for advice.

3.2 Storage
• Store the device in a dry, dust-free location.
• Avoid continuous direct sunlight.
• Store the device in its original packaging.

3.3 Installation specifications


NOTE!
The following precautions must be taken to ensure reliable installation.

• Make sure that there is adequate space to the sides.


• Protect the signal converter from direct sunlight and install a sun shade if necessary.
• Signal converters installed in control cabinets require adequate cooling, for example fans or
heat exchangers.
• Do not subject the signal converter to heavy vibrations. The flowmeters are tested for a
vibration level in accordance with IEC 68-2-3.

01/2008 • 4000041002 - MA IFC 100 R02 en www.krohne.com 13


3 INSTALLATION IFC 100

3.4 Transport
Signal converter
• No special requirements.

Compact versions
• Do not lift the flowmeter by the signal converter housing.
• Do not use lifting chains.
• To transport flange devices, use lifting straps. Wrap these around both process connections.

3.5 Mounting of the compact version


NOTE!
The signal converter is mounted directly on the measuring sensor. For installation of the
flowmeter, please observe the instructions in the supplied product documentation for the
measuring sensor.

14 www.krohne.com 01/2008 • 4000041002 - MA IFC 100 R02 en


IFC 100 INSTALLATION 3

3.6 Mounting the wall-mounted housing, remote version


NOTE!
Mounting materials and tools are not part of the scope of supply. Use the mounting material and
tools in compliance with the applicable occupational health and safety directives.

3.6.1 Wall mounting

Figure 3-1: Mounting the wall-mounted housing

1 Prepare the holes with the aid of the mounting plate. Additional information refer to Mounting
plate, wall-mounted version on page 85.
2 Fasten the device securely to the wall with the mounting plate.

01/2008 • 4000041002 - MA IFC 100 R02 en www.krohne.com 15


3 INSTALLATION IFC 100

NOTE!
Mounting multiple devices next to each other:

[mm] [inches]
a Ø6.5 Ø0.26
b 87.2 3.4
c 241 9.5
d 310 12.2
e 257 10.1

16 www.krohne.com 01/2008 • 4000041002 - MA IFC 100 R02 en


IFC 100 ELECTRICAL CONNECTIONS 4

4.1 Safety instructions


DANGER!
All work on the electrical connections may only be carried out with the power disconnected. Take
note of the voltage data on the nameplate!

DANGER!
Observe national installation regulations!

WARNING!
Observe the regional occupational health and safety regulations without fail. Only work on the
device electrics if you are appropriately trained.

NOTE!
Please check on the device nameplates, that the device is supplied according to your order.
Check for the correct mains voltage printed on the nameplate. If not, contact your local
representative for advice.

4.2 Important notes on electrical connection


DANGER!
Electrical connection is carried out in conformity with the VDE 0100 directive "Regulations for
electrical power installations with line voltages up to 1000 V" or equivalent national regulations.

NOTE!
• Use suitable cable entries for the various electrical cables.
• The measuring sensor and signal converter have been calibrated together at the factory! The
devices should therefore always be installed in pairs. Make sure that the measuring sensor
constants GK/GKL have identical settings (see nameplates).
• In the event of separate delivery or the installation of devices that have not been calibrated
together, the signal converter must be set to the DN size and GK/GKL of the measuring
sensor, refer to Function tables on page 51.

01/2008 • 4000041002 - MA IFC 100 R02 en www.krohne.com 17


4 ELECTRICAL CONNECTIONS IFC 100

4.3 Electrical cables for remote device versions, notes


4.3.1 Notes on signal cable A

NOTE!
Signal cable A (type DS 300) with double shielding ensures proper transmission of measured
values.

Observe the following notes:


• Lay the signal cable with fastening elements.
• It is permissible to lay the signal cable in water or in the ground.
• The insulating material is flame-retardant to EN 50625-2-1, IEC 60322-1.
• The signal cable does not contain any halogens and is unplasticized, and remains flexible at
low temperatures.
• The connection of the inner shield is carried out via the stranded drain wire (1).
• The connection of the outer shield (60) is carried out via the stranded drain wire (6).

4.3.2 Notes on field current cable C

DANGER!
A shielded two-wire copper cable is used as the field current cable. The shielding MUST be
connected in the housing of the measuring sensor and signal converter.

NOTE!
The field current cable is not part of the scope of supply.

18 www.krohne.com 01/2008 • 4000041002 - MA IFC 100 R02 en


IFC 100 ELECTRICAL CONNECTIONS 4

4.3.3 Using other signal cables

NOTE!
When other signal cables are used, the following electrical values must be observed.

Electrical safety
• To EN 60811 (Low Voltage Directive) or equivalent national regulations.

Capacitance of the insulated conductors


• Insulated conductor / insulated conductor < 50 pF/m
• Insulated conductor / shield < 150 pF/m

Insulation resistance
• Riso > 100 GΩ x km
• Umax < 24 V
• Imax < 100 mA

Test voltages
• Insulated conductor / inner shield 500 V
• Insulated conductor / insulated conductor 1000 V
• Insulated conductor / outer shield 1000 V

Twisting of the insulated conductors


• At least 10 twists per meter, important for screening magnetic fields.

01/2008 • 4000041002 - MA IFC 100 R02 en www.krohne.com 19


4 ELECTRICAL CONNECTIONS IFC 100

4.4 Preparing the signal and field current cables


NOTE!
Mounting materials and tools are not part of the scope of supply. Use the mounting material and
tools in compliance with the applicable occupational health and safety directives.

4.4.1 Signal cable A (type DS 300), construction

• Signal cable A is a double-shielded cable for signal transmission between the measuring
sensor and signal converter.
• Bending radius: ≥ 50 mm / 2"

Figure 4-1: Construction of signal cable A

1 Stranded drain wire (1) for the inner shield (10), 1.0 mm2 Cu / AWG 17 (not insulated, bare)
2 Insulated wire (2), 0.5 mm2 Cu / AWG 20
3 Insulated wire (3), 0.5 mm2 Cu / AWG 20
4 Outer sheath
5 Insulation layers
6 Stranded drain wire (6) for the outer shield (60)

20 www.krohne.com 01/2008 • 4000041002 - MA IFC 100 R02 en


IFC 100 ELECTRICAL CONNECTIONS 4

4.4.2 Preparing signal cable A, connection to signal converter

NOTE!
Mounting materials and tools are not part of the scope of supply. Use the mounting material and
tools in compliance with the applicable occupational health and safety directives.

• Connection of the two shields in the signal converter is carried out via the stranded drain
wires.
• Bending radius: ≥ 50 mm / 2"

Required materials
• PVC insulation tubing, Ø2.5 mm / 0.1"
• Heat-shrinkable tubing
• 2x wire end ferrules to DIN 46 228: E 1.5-8 for the stranded drain wires (1, 6)
• 2x wire end ferrules DIN 46 228: E 0.5-8 for the insulated conductors (2, 3)

01/2008 • 4000041002 - MA IFC 100 R02 en www.krohne.com 21


4 ELECTRICAL CONNECTIONS IFC 100

Figure 4-2: Preparation of signal cable A

• a = 80 mm / 3.15"

1 Strip the conductor to dimension a.


2 Cut off the inner shield (10) and the outer shield (60). Make sure not to damage the stranded
drain wires (1, 6).
3 Slide the insulation tubing over the stranded drain wires (1, 6)
4 Crimp the wire end ferrules onto the stranded drain wire.
5 Crimp the wire end ferrules onto the conductors (2, 3).
6 Pull the heat-shrinkable tubing over the prepared signal cable

22 www.krohne.com 01/2008 • 4000041002 - MA IFC 100 R02 en


IFC 100 ELECTRICAL CONNECTIONS 4

4.4.3 Length of signal cable A

NOTE!
For temperatures of the medium above 150°C / 300°F, a special signal cable and a ZD
intermediate socket are necessary. These are available including the changed electrical
connection diagrams.

Measuring sensor Size Min. electrical Curve for signal


conductivity cable A
DN [mm] [inch] [µS/cm]
OPTIFLUX 1000 F 10...150 3/8...6 5 A1
OPTIFLUX 2000 F 25...150 1...6 20 A1
200...1200 8...48 20 A2
OPTIFLUX 4000 F 10...150 3/8...6 1 A1
200...1200 8...48 1 A2
OPTIFLUX 5000 F 25...100 1...4 1 A1
150...250 6...10 1 A2
OPTIFLUX 6000 F 25...150 1...6 1 A1

Figure 4-3: Maximum length of signal cable A


1 Maximum length of signal cable A between the measuring sensor and signal converter [m]
2 Maximum length of signal cable A between the measuring sensor and signal converter [ft]
3 Electrical conductivity of the medium being measured [μS/cm]

01/2008 • 4000041002 - MA IFC 100 R02 en www.krohne.com 23


4 ELECTRICAL CONNECTIONS IFC 100

4.4.4 Preparing field current cable B, connection to signal converter

DANGER!
A shielded two-wire copper cable is used as the field current cable. The shielding MUST be
connected in the housing of the measuring sensor and signal converter.

NOTE!
Mounting materials and tools are not part of the scope of supply. Use the mounting material and
tools in compliance with the applicable occupational health and safety directives.

• Field current cable C is not part of the scope of supply.


• Bending radius: ≥ 50 mm / 2"

Required materials:
• Shielded, at least 2-wire copper cable with suitable heat-shrinkable tubing
• Insulating tubing, size according to the cable being used
• DIN 46 228 wire end ferrules: size according to the cable being used

Length and cross-section of field current cable C


Length Cross-section AF (Cu)

[m] [ft] [mm2] [AWG]


0...150 0...500 2 x 0.75 Cu 1 2 x 18
150...300 500...1000 2 x 1.50 Cu 1 2 x 14
300...600 1000...2000 2 x 2.50 Cu 1 2 x 12
1 Cu = copper cross-section

24 www.krohne.com 01/2008 • 4000041002 - MA IFC 100 R02 en


IFC 100 ELECTRICAL CONNECTIONS 4

Figure 4-4: Field current cable C, preparation for the signal converter

• a = 80 mm / 3.15"

1 Strip the conductor to dimension a.


2 If a stranded drain wire is present, remove the shield that is present. Make sure not to damage
the stranded drain wire.
3 Slide an insulating tube over the stranded drain wire.
4 Crimp a wire end ferrule onto the stranded drain wire.
5 Crimp wire end ferrules onto the conductors.
6 Pull a shrinkable tube over the prepared cable.

4.4.5 Prepare signal cable A, connect to measuring sensor

NOTE!
Mounting materials and tools are not part of the scope of supply. Use the mounting material and
tools in compliance with the applicable occupational health and safety directives.

• The outer shield (60) is connected in the terminal compartment of the measuring sensor
directly via the shield and a clip.
• Bending radius: ≥ 50 mm / 2"

Required materials
• PVC insulation tubing, Ø2.0...2.5 mm / 0.08...0.1"
• Heat-shrinkable tubing
• Wire end ferrule to DIN 46 228: E 1.5-8 for the stranded drain wire (1)
• 2x wire end ferrules DIN 46 228: E 0.5-8 for the insulated conductors (2, 3)

01/2008 • 4000041002 - MA IFC 100 R02 en www.krohne.com 25


4 ELECTRICAL CONNECTIONS IFC 100

Figure 4-5: Prepare signal cable A, connect to measuring sensor

• a = 50 mm / 2"
• b = 10 mm / 0.39"

1 Strip the conductor to dimension a.


2 Trim the outer shield (60) to dimension b and pull it over the outer sheath.
3 Remove the stranded drain wire (6) of the outer shield and the inner shield (10). Make sure not
to damage the stranded drain wire (1) of the inner shield.
4 Slide an insulating tube over the stranded drain wire (1).
5 Crimp the wire end ferrules onto conductors 2 and 3 and the stranded drain wire (1).
6 Pull the heat-shrinkable tubing over the prepared signal cable

26 www.krohne.com 01/2008 • 4000041002 - MA IFC 100 R02 en


IFC 100 ELECTRICAL CONNECTIONS 4

4.4.6 Preparing field current cable C, connection to measuring sensor

NOTE!
Mounting materials and tools are not part of the scope of supply. Use the mounting material and
tools in compliance with the applicable occupational health and safety directives.

• The field current cable is not part of the scope of supply.


• The shield is connected in the terminal compartment of the measuring sensor directly via the
shield and a clip.
• Bending radius: ≥ 50 mm / 2"

Required materials
• Shielded 2-wire insulated copper cable
• Insulating tubing, size according to the cable being used
• Heat-shrinkable tubing
• DIN 46 228 wire end ferrules: size according to the cable being used

Figure 4-6: Preparation of field current cable C

• a = 50 mm / 2"
• b = 10 mm / 0.4"

1 Strip the conductor to dimension a.


2 Trim the outer shield to dimension b and pull it over the outer sheath.
3 Crimp wire end ferrules onto both conductors.
4 Pull a shrinkable tube over the prepared cable.

01/2008 • 4000041002 - MA IFC 100 R02 en www.krohne.com 27


4 ELECTRICAL CONNECTIONS IFC 100

4.5 Connecting the signal and field current cables


DANGER!
The signal and field current cables may only be connected when the auxiliary power supply is
switched off

DANGER!
The device must be grounded in accordance with regulations in order to protect personnel
against electric shocks.

DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the special Ex
instructions.

WARNING!
Observe the regional occupational health and safety regulations without fail. Only work on the
device electrics if you are appropriately trained.

28 www.krohne.com 01/2008 • 4000041002 - MA IFC 100 R02 en


IFC 100 ELECTRICAL CONNECTIONS 4

4.5.1 Connecting the signal and field current cables to the signal converter, remote
version

INFORMATION!
The compact version is supplied preassembled from the factory.

Open housing

Figure 4-7: Open housing

1 Loosen the 4 screws with a suitable tool


2 Lift the housing at the top and bottom at the same time.
3 Slide the housing cover upward.
4 The housing cover is guided and held by the inside hinge.

01/2008 • 4000041002 - MA IFC 100 R02 en www.krohne.com 29


4 ELECTRICAL CONNECTIONS IFC 100

Connecting the signal and field current cables

Figure 4-8: Function of the electrical connection terminal

Connect the electrical conductors as follows:


1 Push the lever downwards with a screwdriver in good condition (blade: 3.5 mm wide and
0.5 mm thick).
2 Insert the electrical conductor into the plug.
3 The conductor will be clamped as soon as the lever is released.

Figure 4-9: Connecting the signal and field current cables


1 Cable entry for field current cable
2 Cable entry for signal cable
3 Connection of the shield for the field current cable
4 Electrical conductor (7)
5 Electrical conductor (8)
6 Stranded drain wire (1) of the inner shield (10) of the signal cable
7 Electrical conductor (2)
8 Electrical conductor (3)
9 Stranded drain wire (S) of the outer shield (60)

30 www.krohne.com 01/2008 • 4000041002 - MA IFC 100 R02 en


IFC 100 ELECTRICAL CONNECTIONS 4

4.5.2 Connection diagram for signal and field current cable

DANGER!
The device must be grounded in accordance with regulations in order to protect personnel
against electric shocks.

• A shielded two-wire copper cable is used as the field current cable. The shielding MUST be
connected in the housing of the measuring sensor and signal converter.
• The outer shield (60) is connected in the terminal compartment of the measuring sensor
directly via the shield and a clip.
• Bending radius of signal and field current cable: ≥ 50 mm / 2"
• The following illustration is schematic. The positions of the electrical connection terminals
may vary depending on the housing version.

Figure 4-10: Connection diagram for signal and field current cable
1 Electrical terminal compartment for signal and field current cable in signal converter.
2 Signal cable A
3 Field current cable C
4 Electrical terminal compartment in measuring sensor
5 Functional ground FE

01/2008 • 4000041002 - MA IFC 100 R02 en www.krohne.com 31


4 ELECTRICAL CONNECTIONS IFC 100

4.6 Grounding the measuring sensor


4.6.1 Classical method

DANGER!
There should be no difference in potential between the measuring sensor and the housing or
protective earth of the signal converter!

• The measuring sensor must be properly grounded.


• The grounding cable should not transmit any interference voltages.
• Do not use the grounding cable to connect more than one device to ground at the same time.
• The measuring sensors are connected to ground by means of a functional grounding
conductor FE.
• Special grounding instructions for the various measuring sensors are provided in the
separate installation instructions for the measuring sensors.
• The installation instructions for the measuring sensors also contain descriptions on how to
use grounding rings and how to install the measuring sensors in metal or plastic pipes or in
pipes which are coated on the inside.

32 www.krohne.com 01/2008 • 4000041002 - MA IFC 100 R02 en


IFC 100 ELECTRICAL CONNECTIONS 4

4.7 Connecting the power


DANGER!
The device must be grounded in accordance with regulations in order to protect personnel
against electric shocks.

• The housings of the flowmeters, which are designed to protect the electronic equipment from
dust and moisture, should be kept well closed at all times. Creepage distances and
clearances are dimensioned to VDE 0110 and IEC 664 for pollution severity 2. Supply circuits
are designed for overvoltage category III and the output circuits for overvoltage category II.
• Fuse protection (IN ≤ 16 A) for the infeed power circuit, and also a disconnecting device
(switch, circuit breaker) to isolate the signal converter must be provided.

Figure 4-11: Terminal compartment for power supply


1 Retaining band of the cover
2 Cable entry for power supply, remote version
3 Cable entry for power supply, compact version

01/2008 • 4000041002 - MA IFC 100 R02 en www.krohne.com 33


4 ELECTRICAL CONNECTIONS IFC 100

• Open the cover of the electrical terminal compartment by pressing down and pulling forwards
at the same time.

L N

Figure 4-12: Connection to power


1 100...230 VAC (-15% / +10%), 8 VA
2 24 VDC (-55% / +30%), 4 W
3 24 VAC/DC (AC: -15% / +10%; DC: -25% / +30%), 8 VA and 4 W

• Close the cover after the power has been connected.

100...230 VAC (tolerance range: -15% / +10%)


• Note the power supply voltage and frequency (50...60 Hz) on the nameplate.

NOTE!
240 VAC+5% is included in the tolerance range.

24 VDC (tolerance range: -55% / +30%)


24 VAC/DC (tolerance ranges: AC: -15% / +10%; DC: -25% / +30%)
• Note the data on the nameplate!
• When connecting to functional extra-low voltages, provide a facility for protective separation
(PELV) (acc. to VDE 0100 / VDE 0106 and/or IEC 364 / IEC 536 or relevant national
regulations).

NOTE!
For 24 VDC, 12 VDC-10% is included in the tolerance range.

34 www.krohne.com 01/2008 • 4000041002 - MA IFC 100 R02 en


IFC 100 ELECTRICAL CONNECTIONS 4

4.8 Description of the outputs


4.8.1 Current output

• All outputs are electrically isolated from each other and from all other circuits.
• All operating data and functions can be set.
• Passive mode: external power Uext ≤ 32 VDC I ≤ 22 mA
• Active mode: load impedance RL ≤ 750 Ω at I ≤ 22 mA
• Self-monitoring: interruption or load impedance too high in the current output loop
• Error message possible via status output, error indication on LCD display.
• Current value error detection can be set.
• Automatic range function via threshold. The setting range for the threshold is between 5%
and 80%of Q100%, ± 0...5% hysteresis (corresponding ratio from smaller to larger range of
1:20 to 1:1.25).
Signaling of the active range possible via a status output (settable).
• Forward / reverse flow measurement (F/R mode) is possible

INFORMATION!
Additional information refer to Connection diagrams of outputs on page 40 and refer to Technical
data on page 75.

DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the special Ex
instructions.

01/2008 • 4000041002 - MA IFC 100 R02 en www.krohne.com 35


4 ELECTRICAL CONNECTIONS IFC 100

4.8.2 Pulse and frequency output

• All outputs are electrically isolated from each other and from all other circuits.
• All operating data and functions can be set.
• Passive mode:
External power supply required: Uext ≤ 32 VDC
I ≤ 20 mA at f ≤ 10 kHz (overflow up to fmax ≤ 12 kHz)
I ≤ 100 mA at f ≤ 100 Hz
• Scaling:
Frequency output: in pulses per unit time (e.g. 1000 pulses/s at Q100%);
Pulse output: in pulses per unit volume (e.g. 100 pulses/m3).
• Pulse width:
symmetrical (pulse duty factor 1:1, independent of output frequency)
automatic (with fixed pulse width, duty factor approx. 1:1 at Q100%), or
fixed (pulse width adjustable as required from 0.05 ms...2 s)
• Forward / reverse flow measurement (F/R mode) is possible
• The pulse and frequency outputs can also be used as a status output / limit switch.

INFORMATION!
Additional information refer to Connection diagrams of outputs on page 40 and refer to Technical
data on page 75.

DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the special Ex
instructions.

36 www.krohne.com 01/2008 • 4000041002 - MA IFC 100 R02 en


IFC 100 ELECTRICAL CONNECTIONS 4

4.8.3 Status output and limit switch

• The status outputs / limit switches are electrically isolated from each other and from all
other circuits.
• The output stages of the status outputs / limit switches behave like relay contacts.
• All operating data and functions can be set.
• Passive mode: external power required:
Uext ≤ 32 VDC; I ≤ 100 mA
• For information on the operating states refer to Function tables on page 51 that can be set.

INFORMATION!
Additional information refer to Connection diagrams of outputs on page 40 and refer to Technical
data on page 75.

DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the special Ex
instructions.

4.9 Electrical connection of the outputs


NOTE!
Mounting materials and tools are not part of the scope of supply. Use the mounting material and
tools in compliance with the applicable occupational health and safety directives.

01/2008 • 4000041002 - MA IFC 100 R02 en www.krohne.com 37


4 ELECTRICAL CONNECTIONS IFC 100

4.9.1 Electrical connection of the outputs

DANGER!
All work on the electrical connections may only be carried out with the power disconnected. Take
note of the voltage data on the nameplate!

Figure 4-13: Connection of outputs


1 Cable entry, remote version
2 Cable entry, compact version
3 Terminal S for shield

• Open the housing cover


• Push the prepared cables through the cable entries and connect the necessary conductors.
• Connect the shield.
• Close the housing cover.

NOTE!
Ensure that the housing seal is properly fitted, clean and undamaged.

38 www.krohne.com 01/2008 • 4000041002 - MA IFC 100 R02 en


IFC 100 ELECTRICAL CONNECTIONS 4

4.9.2 Laying electrical cables correctly

Figure 4-14: Protect housing from dust and water

1 For compact versions with nearly horizontally-oriented cable entries, lay the necessary elec-
tric cables with a drip loop as shown in the illustration.
2 Tighten the screw connection of the cable entry securely.
3 Seal cable entries that are not needed with a plug.

01/2008 • 4000041002 - MA IFC 100 R02 en www.krohne.com 39


4 ELECTRICAL CONNECTIONS IFC 100

4.10 Connection diagrams of outputs


4.10.1 Description of the electrical symbols

Milliammeter
0...20 mA or 4...20 mA and others
RL also includes the line resistances

DC voltage source (Uext), external power supply, any connection polarity

DC voltage source (Uext), connection polarity as shown in the diagrams

Internal DC voltage source

Controlled internal power source in the device

Electronic or electromagnetic counter


At frequencies above 100 Hz, shielded cables must be used to connect the
counters.
Ri Internal resistance of the counter

Button, NO contact or similar

Table 4-1: Description of symbols

40 www.krohne.com 01/2008 • 4000041002 - MA IFC 100 R02 en


IFC 100 ELECTRICAL CONNECTIONS 4

4.10.2 Basic outputs

INFORMATION!
Additional information refer to Description of the outputs on page 35 and refer to HART®
connection on page 44.

Current output active (HART®)


• Uint,nom = 20 VDC
• I ≤ 22 mA
• RL ≤ 750 Ω

Figure 4-15: Current output active Ia

Current output passive (HART®)


• Uext ≤ 32 VDC
• I ≤ 22 mA
• U0 ≤ 2 V at I = 22 mA

Figure 4-16: Current output passive Ip

01/2008 • 4000041002 - MA IFC 100 R02 en www.krohne.com 41


4 ELECTRICAL CONNECTIONS IFC 100

INFORMATION!
At frequencies above 100 Hz, shielded cables must be used. Shielding takes place at the
electrical connection (S) of the output terminal block.

Pulse/frequency output passive


• Uext ≤ 32 VDC
• 100 Hz < fmax ≤ 10 kHz:
I ≤ 20 mA
open:
I ≤ 0.1 mA at Uext = 5 V
I ≤ 0.5 mA at Uext = 24 V
I ≤ 0.7 mA at Uext = 32 V
closed:
U0 ≤ 0.8 V at I = 1 mA
U0 ≤ 1.5 V at I = 10 mA
U0 ≤ 3.5 V at I = 100 mA
• R is required if the internal resistance of the counter does not reach the maximum load
resistance RL.
f ≤ 1 kHz: RL ≤ 10 kΩ
f ≤ 10 kHz: RL ≤ 2 kΩ
• Can also be set as a status output; for the electrical connection see status output connection
diagram.

Figure 4-17: Pulse frequency output passive Pp

42 www.krohne.com 01/2008 • 4000041002 - MA IFC 100 R02 en


IFC 100 ELECTRICAL CONNECTIONS 4

Status output / limit switch passive


• Uext ≤ 32 VDC
• I ≤ 100 mA
• RL ≤ 10 kΩ
• open:
I ≤ 0.1 mA at Uext = 5 V
I ≤ 0.5 mA at Uext = 24 V
I ≤ 0.7 mA at Uext = 32 V
closed:
U0 ≤ 0.8 V at I = 1 mA
U0 ≤ 1.5 V at I = 10 mA
U0 ≤ 3.5 V at I = 100 mA
• The output is closed when the device is de-energized.

Figure 4-18: Status output / limit switch passive Sp

01/2008 • 4000041002 - MA IFC 100 R02 en www.krohne.com 43


4 ELECTRICAL CONNECTIONS IFC 100

4.10.3 HART® connection

INFORMATION!
• The current output at connection terminals A+/A-/A always has HART® capability.

HART® connection active (point-to-point)

Figure 4-19: HART® connection active (Ia)

1 HART® communicator

The shunt resistor for the HART® communicator must have R ≥ 230 Ω.

44 www.krohne.com 01/2008 • 4000041002 - MA IFC 100 R02 en


IFC 100 ELECTRICAL CONNECTIONS 4

HART® connection passive (multidrop mode)


• I: I0% = 4 mA
• Multidrop I: Ifix = 4 mA
• Uext ≤ 32 VDC
• RL ≤ 230 Ω

Figure 4-20: HART® connection passive (Ip)

1 HART® communicator
2 Additional devices with HART® capability

01/2008 • 4000041002 - MA IFC 100 R02 en www.krohne.com 45


5 START-UP IFC 100

5.1 Switching on the power


Before connecting to power, please check that the system has been correctly installed. This
includes:
• The flowmeter must be mechanically safe and mounted in compliance with the regulations.
• The power connections must have been made in compliance with the regulations.
• The electrical terminal compartments must be secured and the covers have been screwed
on.
• Check that the electrical operating data of the power supply are correct.

• Switch the power on.

5.2 Starting the signal converter


The measuring device, consisting of the measuring sensor and the signal converter, is supplied
ready for operation. All operating data have been set at the factory in accordance with your order
specifications.

When the power is switched on, a self test is carried out. After that the flowmeter immediately
begins measuring, and the current values are displayed.

Figure 5-1: Display in measuring mode, examples

It is possible to change between the 1st and 2nd measured values windows, the trend display and
(if present) the list with the status messages by pressing the keys ↑ and ↓. Possible status
messages, their meaning and cause refer to Status messages and diagnostic information on
page 69.

46 www.krohne.com 01/2008 • 4000041002 - MA IFC 100 R02 en


IFC 100 OPERATION 6

6.1 Display and operator input elements

Figure 6-1: Display and operational elements, example with flow indication
1 Signals a status message in the status bar
2 The measuring point number (tag) is only indicated if this number was entered previously by the operator.
3 Indicates when a key has been pressed
4 1st and 2nd display line for indication of different measured variables (here large depiction of only one measured vari-
able)
5 Symbols of the operator input keys
6 Interface to the GDC bus (not present in all signal converter versions)
7 Infrared sensor (not present in all signal converter versions)
8 Operator input keys (see table below for description)
9 3rd display line (here bargraph)

Symbol Meas. mode Menu mode Function mode Data mode


> Switch from Entry into selected Entry into displayed For numerical
measuring mode to menu displayed, selected function or values, move cursor
menu mode; press then 1st function of subfunction (blue) one place to
key for 2.5 s, then the menu the right
"Quick Start" menu
displayed
^ - Return to measuring Press 1 - 3 times, Return to function or
mode, preceded by return to menu subfunction with
query whether mode with data acceptance of data
changed data to be acceptance
accepted
↓↑ Alternate between Select menu Select function or Blue cursor...
display measured subfunction - change number
value pages 1 + 2, - change unit
trend and any status - change property
list(s), if provided - change decimal
point
Esc (> + ↑) - - Return to menu Return to function or
mode without data subfunction without
acceptance acceptance of data
Table 6-1: Function of keys

01/2008 • 4000041002 - MA IFC 100 R02 en www.krohne.com 47


6 OPERATION IFC 100

6.1.1 Time-out functions

In Operator Control mode


• After 5 minutes without key op., return to meas. mode.
Prev. changed data are not accepted.

In Test Menu mode


• After 60 minutes without key op., return to meas. mode.
Prev. changed data are not accepted.

6.1.2 Display in measuring mode

Figure 6-2: Example for display in measuring mode


1 Signals a status message in the status list
2 Measuring point number (tag); is only indicated if this number was assigned previously by the operator.
3 1st measured value (depiction of the 1st measured value via the 1st and 2nd display lines)
4 3rd display line as a bargraph
5 3 measured values (depiction of 3 different measured values in display lines 1...3)

6.1.3 Display for selection of menu and functions

Figure 6-3: Display for selection of menu or function


1 Menu description
2 Function number (only in setup menu C)
3 Position within the menu or function list (here the beginning)
4 Current function (open: →; forward/back: ↑↓)
5 Position within the menu or function list (here the end)

48 www.krohne.com 01/2008 • 4000041002 - MA IFC 100 R02 en


IFC 100 OPERATION 6

6.1.4 Display for setting of data and functions

Figure 6-4: Display for setting of data and functions


1 Current menu
2 Function number (only in setup menu C)
3 Identifies factory settings
4 Identifies permitted value range
5 Next function or permissible value range for numeric values
6 Value, unit or function set at present (in white lettering on a blue background when selected →)
The value is changed here.
7 Current (sub)function (open: →)
8 Factory setting of the current (sub)function (for information only, cannot be changed)

6.1.5 Display after change of data and functions

Figure 6-5: Display after change of data and functions


1 Current menu
2 Function number (only in setup menu C)
3 Indicates a change to a (sub)function; simple check of the changed data by paging through the (sub)function lists
4 Next function or permissible value range for numeric values
5 Value, unit or function set at present (in white lettering on a blue background when selected →)
6 Current (sub)function (open: →)
7 Factory setting of the current (sub)function (for information only, cannot be changed)

01/2008 • 4000041002 - MA IFC 100 R02 en www.krohne.com 49


6 OPERATION IFC 100

6.2 Program structure

Measuring mode Select Select function and/or subfunction Set data


menu ↓↑ ↓↑
↓↑
^ Press >
2.5 s
A Quick setup > A1 Language >
^ ^
A2 Tag
A3 Reset
A4 Analog outputs
A5 Digital outputs
^ B Test > B1 Simulation >
^ ^
B2 Actual values
B3 Information
^ C Setup > C1 Process input > 1.1 Calibration >
^ ^ ^
1.2 Filter
1.3 Self test
1.4 Information
1.5 Simulation
^ > C2 Inputs/outputs (I/Os) > 2.1 Hardware >
^ ^ ^
2.… Current output X
2.… Frequency output X
2.… Pulse output X
2.… Status output X
2.… Limit switch X
^ > C3 I/O counter > 3.1 Counter 1 >
^ ^ ^
3.2 Counter 2
^ > C4 I/O HART > 4.1 PV is... >
^ ^ ^
4.2 SV is...
4.3 TV is...
4.4 4V is...
^ > C5 Device > 5.1 Device info >
^ ^ ^
5.2 Display
5.3 Meas. page 1
5.4 Meas. page 2
5.5 Graphic page
5.6 Special functions
5.7 Units
5.8 HART
5.9 Quick setup
↓↑ ↓↑ ↓↑ ↓↑>

50 www.krohne.com 01/2008 • 4000041002 - MA IFC 100 R02 en


IFC 100 OPERATION 6

6.3 Function tables


INFORMATION!
Depending on the device version, not all functions are available.

6.3.1 Menu A, quick setup

No. Function Setting / description

A1 Language
A1 Language Language selection depends on the device version.

A2 Tag
A2 Tag Measuring point identifier (Tag No.) (also for HART® operation), appears in
the LCD display header (up to 8 places).

A3 Reset
A3 Reset
A3.1 Error reset Reset error? Select: No/Yes
A3.2 Reset counter 1 Reset counter? Select: No / Yes (available if activated in C5.9.1)
A3.3 Reset counter 2 Reset counter? Select: No / Yes (available if activated in C5.9.2)

A4 Analog outputs (only for HART®)


A4 Analog outputs Applicable to current output (term. A), frequency output (term. D), limit
switches (term. C and / or D) and display page 1 / line 1
A4.1 Measurement 1) Measurement selection: volume flow / mass flow / diagnosis value / flow
speed / coil temperature / conductivity
2) Use for all outputs? (also use this output for the fcts. A4.2...A4.5!)
Setting: No (applies only to the main current output ) / Yes (applies to all
analog outputs)
A4.2 Unit Selection of the unit from a list, depending on the measured variable
A4.3 Range 1) Setting for main current output (measuring range: 0...100%)
Setting: 0…x.xx (format and unit, depending on the measured variable, see
A4.1 and A4.2 above)
2) Use for all outputs? Make setting, see Fct. A4.1 above!
A4.4 Low flow cutoff 1) Setting for main current output (sets output value to "0")
Setting: x.xxx ± x.xxx% (range: 0.0…20%)
(1st value = operating point / 2nd value = hysteresis); condition:
2nd value ≤ 1st value
2) Use for all outputs? Make setting, see Fct. A4.1 above!
A4.5 Time constant 1) Setting for main current output (applicable to all flow rate measurements)
Setting: xxx.x s (setting range: 000.1…100 s)
2) Use for all outputs? Make setting, see Fct. A4.1 above!

01/2008 • 4000041002 - MA IFC 100 R02 en www.krohne.com 51


6 OPERATION IFC 100

No. Function Setting / description

A5 Digital outputs
A5 Digital outputs Applicable to pulse output (term. D) and counter 1
A5.1 Measured value 1) Select measurement: volume flow / mass flow
2) Use for all outputs? (also use this output for the fcts. A5.2 to A5.4!)
Setting: No (only for pulse output D) / Yes (for all digital outputs)
A5.2 Pulse value unit Selection of the unit from a list, depending on the measured variable
A5.3 Value p. pulse 1) Setting for pulse output D (volume or mass value per pulse)
Setting: xxx.xxx in l/s or kg/s
2) Use for all outputs? Make setting, see Fct. A5.1 above!
A5.4 Low flow cutoff 1) Setting for pulse output D (sets output value to "0")
Setting: x.xxx ± x.xxx% (setting range: 0.0…20%)
(1st value = operating point / 2nd value = hysteresis); condition:
2nd value ≤ 1st value
2) Use for all outputs? Make setting, see Fct. A5.1 above!

6.3.2 Menu B, test

No. Function Setting / description

B1 Simulation
B1 Simulation Displayed values are simulated
B1.1 Flow speed Flow speed simulation
Break (exit function without simulation)
Set value (range: -12…+12 m/s; unit selection in Fct. C5.7.7)
Query: Start simulation?
Settings: No (exit function without simulation) / Yes (start simulation)
B1.2 Volume flow Volume flow simulation, sequence and settings similar to B1.1, see below!
[X stands for one of the connection terminals A, C or D]
… stands for Fct. No. B1.3...1.6
B1.… Current out X Simulation X
Sequence and settings similar to B1.1, see below!
B1.… Pulse output X
[X stands for one of the connection terminals A, C or D]
B1.… Frequency out X For pulse output a set number of pulses is output once in 1s!
B1.… Limit switch X
B1.… Status output X

52 www.krohne.com 01/2008 • 4000041002 - MA IFC 100 R02 en


IFC 100 OPERATION 6

No. Function Setting / description

B2 Actual values
B2 Actual values Display the current values; exit the displayed function with the ^ key.
B2.1 Operating hours
B2.2 Act. flow speed
B2.3 Act. coil temp. See also Fcts. C1.1.7...C1.1.8
B2.4 Electronic temperature
B2.5 Act. conductivity See also Fcts. C1.3.1...C1.3.2
B2.6 Act. electr. noise See also Fcts.C1.3.13...C1.3.15
B2.8 Act. coil resistance Current resistance of the field coils according to the current coil
temperature

B3 Information
B3 Information LCD display
1st line: ID No. of the circuit board
2nd line: software version
3rd line: date of calibration/manufacture
B3.1 C number Electronics type
B3.2 Process input Process input section
B3.3 Device Electronics and HART® software
B3.4 Display User interface
B3.5 "Interface" "Bus interface" (in preparation)

01/2008 • 4000041002 - MA IFC 100 R02 en www.krohne.com 53


6 OPERATION IFC 100

6.3.3 Menu C, setup

No. Function / subfunction Settings / descriptions

C1 Process input

C1.1 Calibration
C1.1 Calibration
C1.1.1 Zero calibration Display of current zero value (zero)
Query: calibrate zero?
Setting: break (return with ^) /standard (factory setting) /
manual (display last value, set new value, range: -1.00…+1 m/s) /
automatic (shows the current value as the new zero value)
C1.1.2 Size Select from size table; range: DN2.5...1200 mm / 1/10...48"
C1.1.5 GKL Set value acc. to nameplate; range: 0.5…12 (20)
C1.1.7 Coil resistance Rsp Field coil resistance at 20°C; range: 10.00...220 Ω
C1.1.8 Calib. coil temp. The coil temperature is derived from the coil resistance at the reference
temperature.
Set coil temperature:
Break return with ^ key
Standard (= 20°C)
Automatic (set current temperature); range: -40.0…+200°C
Set coil temperature:
Break (return with ^ key)
Standard (= setting from Fct. C1.1.7)
Automatic (= calibration with the current resistance)
C1.1.9 Density Calculation of mass flow with constant density of product; range: 0.1…5 kg/l
C1.1.10 Target conduct. Reference value for on-site calibration; range: 1.000...50000 µS/cm
C1.1.11 EF electr. factor For calculation of the conductivity based on the electrode impedance
Query: calibrate EF?
Break (return with ^ key)
Set value in the following:
Standard (with factory setting) /
Manual (set desired value) /
Automatic (determines EF according to the setting in Fct. C1.1.10)
C1.1.13 Field frequency Setting as on measuring sensor nameplate =
Line frequency x value (from the following list):
2; 4/3; 2/3; 1/2; 1/4; 1/6; 1/8; 1/12; 1/18; 1/36; 1/50
C1.1.14 Select settling Select settling (special function)
Select: Standard (fixed allocation) /
Manual (manual time setting for the settling time for the field current)
C1.1.15 Settling time Only when "Manual" selected under Fct. C1.1.14; range: 1.0…250 ms
C1.1.16 Line frequency Set line frequency
Automatic (measuring & setting; for DC systems fixed setting 50 Hz)
Select: 50 Hz or 60 Hz (fixed setting)
C1.1.17 Act. coil resistance Display of the current value for calculation of the temperature

54 www.krohne.com 01/2008 • 4000041002 - MA IFC 100 R02 en


IFC 100 OPERATION 6

No. Function / subfunction Settings / descriptions

C1.2 Filter
C1.2 filter
C1.2.1 limitation Limitation of all flow values, before smoothing by time constant, affects all
outputs
Settings: -xxx.x / +xxx.x m/s; condition: 1st value < 2nd value
Range 1st value: -100.0 m/s ≤ value ≤ -0.001 m/s
Range 2nd value: +0.001 m/s ≤ value ≤ +100 m/s
C1.2.2 flow direction Define polarity of flow values
Forwards (according to the arrow on the measuring sensor) or backwards
(in the opposite direction to the arrow)
C1.2.3 Time constant For all flow measurements and outputs
xxx.x s; range: 0.0…100 s
C1.2.4 Pulse filter Suppresses noise due to solids, air/gas bubbles and sudden changes in pH
Select: Off (without pulse filter) / On (with pulse filter)
C1.2.5 Pulse width Only when pulse filter switched on, Fct. C1.2.4
Length of interference and delays to be suppressed on sudden changes in
flow
xx.x s; range: 0.01...10 s
C1.2.6 Pulse limitation Dynamic limitation from one measured value to the next, only when pulse
filter switched on, Fct. C1.2.4 effective.
xx.x s; range: 0.01...100 m/s
C1.2.7 Noise filter Suppresses noise at low conductivity, high solids content, air and gas
bubbles, and chemically inhomogeneous media
Select: Off (without noise filter) / On (with noise filter)
C1.2.8 Noise level Range within which changes are evaluated as noise, and outside of which
changes are evaluated as flow (only with noise filter switched on, Fct. C1.2.7)
xx.xx m/s; range 0.01…10 m/s
C1.2.9 Noise suppression Set noise suppression (only when noise filter switched on, Fct. C1.2.7)
Range: 1…10, noise suppression factor [min = 1...max = 10]
C1.2.10 Low flow cutoff Sets output value of all outputs to "0":
x.xxx ± x.xxx%; range: 0.0…20%
(1st value = operating point / 2nd value = hysteresis); condition:
2nd value ≤ 1st value

01/2008 • 4000041002 - MA IFC 100 R02 en www.krohne.com 55


6 OPERATION IFC 100

No. Function / subfunction Settings / descriptions

C1.3 Self test


C1.3 Self test
C1.3.1 Empty pipe Switch conductivity measurement off and on (measurement of the electrode
resistance)
Select: off / on
Additional setting necessary when "on":
Cond. + empty pipe [F] (conductivity measurement and empty pipe
indication, error category [F] application);
Flow indication "= 0" when pipe empty
Cond. + empty pipe [S] (conductivity measurement and empty pipe
indication, error category [S] measurement outside of specification);
Flow indication "≠ 0" when pipe empty
C1.3.2 Act. conductivity Only available when empty pipe activated [..] in Fct. C1.3.1
Current conductivity is indicated. Activation takes place only after setting
mode is exited!
C1.3.3 Limit empty pipe Only available when empty pipe activated [..] in Fct. C1.3.1
Range: 0.0…9999 µS (set max 50% of the lowest occurring conductivity in
operation. Conductivity below this value = signal as empty pipe)
C1.3.9 Coil current Automatic test switched off / on, select: off / on
C1.3.13 Electrode noise Automatic test switched off/on, select: off / on
C1.3.14 Limit electr. noise Only with electrode noise activated, see Fct. C1.3.13
Range: 0.000…12 m/s (noise above this threshold generates an error of
category [S])
C1.3.15 Act. electr. noise Only available when electrode noise "on" activated in Fct. C1.3.13. Activation
takes place only after the setting mode is exited!
C1.3.17 Diagnosis value Select diagnosis value for testing the various analog outputs.
Select: off (no diagnosis) / electrode noise (activate Fct. C1.3.13)
Terminal 2 (electrode DC voltage) / Terminal 3 (electrode DC voltage)

C1.4 Information
C1.4 Information
C1.4.1 Liner Shows material of the liner
C1.4.2 Electr. material Shows material of the electrodes
C1.4.3 Calibration date Not available at this time
C1.4.4 Serial no. sensor Shows serial no. of the measuring sensor
C1.4.5 V no. sensor Shows the order number of the measuring sensor
C1.4.6 Sensor electr. info Shows the serial no. and calibration date of the electronics and the software
version

56 www.krohne.com 01/2008 • 4000041002 - MA IFC 100 R02 en


IFC 100 OPERATION 6

No. Function / subfunction Settings / descriptions

C1.5 Simulation
C1.5 Simulation
C1.5.1 Flow speed Sequence see Fct. B1.1
C1.5.2 Volume flow Sequence see Fct. B1.2

No. Function / subfunction Settings / descriptions

C2 Inputs/outputs (I/Os)

C2.1 Hardware
C2.1 Hardware Assignment of connection terminals dependent on signal converter version:
active / passive / NAMUR
C2.1.1 Terminal A Select: off (switched off) / current output / frequency output / pulse output /
status output / limit switch
C2.1.3 Terminal C Select: off (switched off) / current output / status output / limit switch
C2.1.4 Terminal D Select: off (switched off) / frequency output / pulse output / status output /
limit switch

C2.… Current output X


C2.… Current output X X stands for connection terminal A
… stands for Fct. No. C2.2 (A)
C2.….1 Range 0%…100% Current range for the selected measured variable, e.g. 4...20 mA,
corresponds to 0…100%
xx.x … xx.x mA;
Range: 0.00…20 mA (condition: 0 mA ≤ 1st value ≤ 2nd value ≤ 20 mA)
C2.….2 Extended range Exceeding the min. and max. limits
xx.x…xx.x mA; range: 03.5…21.5 mA (condition: 0 mA ≤ 1st value ≤ 2nd value
≤ 21.5 mA)
C2.….3 Error current Specify error current
xx.x mA; range: 0.00…22 mA (condition: 0 mA ≤ value ≤ 25 mA, outside of
extended range)
C2.….4 Error condition The following error conditions can be selected
Select: error in device (error category [F]) / application error (error category
[F]) / out of specification (error category [S])
C2.….5 Measurement Measured variables for activating the output
Select: volume flow / mass flow / diagnosis value / flow speed / coil
temperature / conductivity
C2.….6 Range 0…100% of the measured variable set in Fct. C2.….5
0…xx.xx _ _ _ (format and unit depend on the measured variable, see above)
C2.….7 Polarity Set measured value polarity, please note flow direction in C1.2.2!
Select: both polarities (plus and minus values are displayed) / positive
polarity (display for negative values = 0) / negative polarity (display for
positive values = 0) / absolute value (use for the output)

01/2008 • 4000041002 - MA IFC 100 R02 en www.krohne.com 57


6 OPERATION IFC 100

No. Function / subfunction Settings / descriptions


C2.….8 Limitation Limitation before applying the time constant
±xxx … ±xxx%; range: -150…+150%
C2.….9 Low flow cutoff Sets output value to "0"
x.xxx ± x.xxx%; range: 0.0…20%
(1st value = operating point / 2nd value = hysteresis), condition:
2nd value ≤ 1st value
C2.….10 Time constant Range: 000.1…100 s
C2.….11 Special function Automatic range; select:
off (switched off)
automatic range (range is changed automatically, extended lower range,
only makes sense together with a status output)
C2.….12 Threshold Appears only when Fct. C2.….11 is activated Threshold between extended
and normal range. The automatic range function always changes from the
extended to the normal range when 100% current is reached
The upper 100% value of the hysteresis is then = 0. The threshold is then the
hysteresis value, instead of "threshold ± hysteresis" as shown in the display.
Range: 5.0%…80%
(1st value = operating point / 2nd value = hysteresis), condition:
2nd value ≤ 1st value
C2.….13 Information Serial number of the I/O circuit board, software version number and
calibration date of the circuit board
C2.….14 Simulation Sequence see B1.… Current output X
C2.….15 4mA trimming Trimming of the current at 4 mA
Resetting to 4 mA restores the factory calibration.
Used for HART® setting.
C2.….16 20mA trimming Trimming of the current at 20 mA
Resetting to 20 mA restores the factory calibration.
Used for HART® setting.

C2.… Frequency output X


C2.… Frequency out X X stands for connection terminal D
… stands for Fct. No. C2.5 (D)
C2.….1 Pulse shape Specify the pulse shape
Select: symmetric (about 50% on and 50% off) / automatic (constant pulse
with about 50% on and 50% off at 100% pulse rate) / fixed (fixed pulse rate,
see below Fct. C2.….3 100% pulse rate)
C2.….2 Pulse width Only available if set to "fixed" in Fct. C2.….1
Range: 0.05…2000 ms
Note: max. setting value Tp [ms] ≤ 500 / max. pulse rate [1/s], Gives the
pulse width = time where the output is activated
C2.….3 100% pulse rate Pulse rate for 100% of the measuring range
Range: 0.0…10000 1/s
Limitation 100% pulse rate ≤ 100/s: Imax ≤ 100 mA
Limitation 100% pulse rate > 100/s: Imax ≤ 20 mA

58 www.krohne.com 01/2008 • 4000041002 - MA IFC 100 R02 en


IFC 100 OPERATION 6

No. Function / subfunction Settings / descriptions


C2.….4 Measurement Measured variables for activating the output
Select: volume flow / mass flow / diagnosis value / flow speed / coil
temperature / conductivity
C2.….5 Range 0…100% of the measured variable set in Fct. C2.….4
0…xx.xx _ _ _ (format and unit depend on the measured variable, see above)
C2.….6 Polarity Set measured value polarity, please note flow direction in C1.2.2!
Select: both polarities (plus and minus values are displayed) / positive
polarity (display for negative values = 0) / negative polarity (display for
positive values = 0) / absolute value (use for the output)
C2.….7 Limitation Limitation before application of the time constant
± xxx … ± xxx%; range: -150…+150%
C2.….8 Low flow cutoff Sets output value to "0":
x.xxx ± x.xxx%; range: 0.0…20%
(1st value = operating point / 2nd value = hysteresis); condition:
2nd value ≤ 1st value
C2.….9 Time constant Range: 000.1…100 s
C2.….10 Invert signal Select:
off (activated output generates a high current at the output, switch closed)
on (activated output generates a low current at the output, switch open)
C2.….12 Information Serial number of the I/O circuit board, software version number and
calibration date of the circuit board
C2.….13 Simulation Sequence see B 1.… Frequency out X

C2.… Pulse output X


C2.… Pulse output X X stands for connection terminal D
… stands for Fct. No. C2.5 (D)
C2.….1 Pulse shape Specify the pulse shape
Select: symmetric (about 50% on and 50% off) / automatic (constant pulse
with about 50% on and 50% off at 100% pulse rate) / fixed (fixed pulse rate,
setting see below Fct. C2.….3 100% pulse rate)
C2.….2 Pulse width Only available if set to "fixed" in Fct. C2.….1
Range: 0.05…2000 ms
Note: max. setting value Tp [ms] ≤ 500 / max. pulse rate [1/s], Gives the
pulse width = time where the output is activated
C2.….3 100 % pulse rate Pulse rate for 100% of the measuring range
Range: 0.0…10000 1/s
Limitation 100% pulse rate ≤ 100/s: Imax ≤ 100 mA
Limitation 100% pulse rate > 100/s: Imax ≤ 20 mA
C2.….4 Measurement Measured variables for activating the output
Select: volume flow / mass flow
C2.….5 Pulse value unit Selection of the unit from a list, depending on the measured variable
C2.….6 Value p. pulse Set value for volume or mass per pulse
xxx.xxx, meas. range in [l] or [kg] (volume or mass for current output
C2.….6)
At max. pulse rate see above 2.….3 Pulse output

01/2008 • 4000041002 - MA IFC 100 R02 en www.krohne.com 59


6 OPERATION IFC 100

No. Function / subfunction Settings / descriptions


C2.….7 Polarity Set measured value polarity, please note flow direction in C1.2.2!
Select: both polarities (plus and minus values are displayed) / positive
polarity (display for negative values = 0) / negative polarity (display for
positive values = 0) / absolute value (use for the output)
C2.….8 Low flow cutoff Sets output value to "0"
x.xxx ± x.xxx%; range: 0.0…20%
(1st value = operating point / 2nd value = hysteresis); condition:
2nd value ≤ 1st value
C2.….9 Time constant Range: 000.1…100 s
C2.….10 Invert signal Select:
off (activated output generates a high current at the output, switch closed)
on (activated output generates a low current at the output, switch open)
C2.….12 Information Serial number of the I/O circuit board, software version number and
calibration date of the circuit board
C2.….13 Simulation Sequence see B 1.… Pulse output X

C2.… Status output X


C2.… Status output X X (Y) stands for one of the connection terminals C or D
… stands for Fct. No. C2.4 (C) / C2.5 (D)
C2.….1 Mode The output shows the following measuring conditions:
Out of specification (output activated, signals application error or error in
device refer to Status messages and diagnostic information on page 69 /
Application error (output activated, signals application error or error in
device refer to Status messages and diagnostic information on page 69 /
Polarity flow (polarity of the current flow) /
Over range flow (over range of the flow) /
Counter 1 preset (activates counter X when preset value is reached) /
Counter 2 preset (activates counter X when preset value is reached) /
Output A (activated by the status of output Y, additional output data see
below) /
Output C (activated by the status of output Y, additional output data see
below) /
Output D (activated by the status of output Y, additional output data see
below) /
Off (switched off) /
Empty pipe (when empty pipe, output active) /
Error in device (when error, output activated)
C2.….2 Current output Y Only appears if output A is set under "mode (see above)", and this output is a
"current output".
Polarity (is signaled)
Over range (is signaled)
Automatic range signals lower range
C2.….2 Frequency out Y and pulse Only appears if output D is set under "mode (see above)" and this output is a
output Y "frequency / pulse output".
Polarity (is signaled)
Over range (is signaled)
C2.….2 Status output Y Only appears if output C or D is set under "mode (see above)" and this output
is a "status output".
Same signal (like other connected status output, signal can be inverted, see
below)

60 www.krohne.com 01/2008 • 4000041002 - MA IFC 100 R02 en


IFC 100 OPERATION 6

No. Function / subfunction Settings / descriptions


C2.….2 Limit switch Y Only appears if output C or D is set under "mode (see above)", and this
output is a "limit switch".
Status off (is always selected here if status output X is connected with a limit
switch.
C2.….2 off Only appears if output A, C or D is set under "mode (see above)" and this
output is switched off.
C2.….3 Invert signal off (activated output supplies a high current, switch closed)
on (activated output supplies a low current, switch open)
C2.….4 Information Serial number of the I/O circuit board, software version number and
calibration date of the circuit board
C2.….5 Simulation Sequence see B 1.… Status output X

C2.… Limit switch X


C2.… Limit switch X X stands for one of the connection terminals C or D
… stands for Fct. No. C2.4 (C) / C2.5 (D)
C2.….1 Measurement Select: volume flow / mass flow / diagnosis value / flow speed / coil
temperature / conductivity
C2.….2 Threshold Switching level, set threshold with hysteresis
xxx.x ±x.xxx (format and unit depend on the measured variable, see above)
(1st value = threshold / 2nd value = hysteresis); condition:
2nd value ≤ 1st value
C2.….3 Polarity Set measured value polarity, please note flow direction in C1.2.2!
Select: both polarities (plus and minus values are displayed) / positive
polarity (display for negative values = 0) / negative polarity (display for
positive values = 0) / absolute value (use for the output)
C2.….4 Time constant Range: 000.1…100 s
C2.….5 Invert signal Select:
off (activated output generates a high current, switch closed)
on (activated output generates a low current, switch open)
C2.….6 Information Serial number of the I/O circuit board, software version number and
calibration date of the circuit board
C2.….7 Simulation Sequence see B 1.… Limit switch X

01/2008 • 4000041002 - MA IFC 100 R02 en www.krohne.com 61


6 OPERATION IFC 100

No. Function / subfunction Settings / descriptions

C3 I/O counter
C3.1 Counter 1 Set function of counter …
… stands for 1, 2 (= counter 1, 2)
C3.2 Counter 2
The basic version (standard) has only 2 counters!
C3.….1 Function of counter Select: sum counter (counts positive and negative values) / +counter (counts
only the positive values) / -counter (counts only the negative values) / off
(counter is switched off) /
C3.….2 Measurement Selection of the measured variable for counter …
Select: volume flow / mass flow
C3.….3 Low flow cutoff Sets output value to "0".
Range: 0.0%…20%
(1st value = operating point / 2nd value = hysteresis); condition:
2nd value ≤ 1st value
C3.….4 Time constant Range: 0.1…100 s
C3.….5 Preset value If this value is reached (positive or negative), a signal is generated that can
be used for a status output at which "preset counter X" has to be set
Preset value (max. 8 places) x.xxxxx in selected unit, see C5.7.10 + 13
C3.….6 reset counter Sequence see Fcts. A 3.2 and A 3.3
C3.….7 Set counter Set counter … to the desired value
Select: break (exit function) / set value (opens the editor to make the entry)
Query: set counter?
Select: no (exit function without setting the value) / yes (sets the counter and
exits the function)
C3.….8 Stop counter Counter … stops and holds the current value
Select: no (exits the function without stopping the counter) / yes (stops the
counter and exits the function)
C3.….9 Start counter Start counter … after that counter is stopped
Select: no (exits the function without starting the counter) / yes (starts the
counter and exits the function)
C3.….10 Information Serial number of the I/O circuit board, software version number and
calibration date of the circuit board

62 www.krohne.com 01/2008 • 4000041002 - MA IFC 100 R02 en


IFC 100 OPERATION 6

No. Function / subfunction Settings / descriptions

C4 I/O HART
C4 I/O HART Selection / display of the 4 dynamic variables (DV) for HART®
The HART® current output (Term. A basic I/Os or Term. C EExi I/Os) always
has a fixed link to the primary variables (PV). Fixed links of the other DVs (1-
3) are only possible if 1 additional analog output (frequency) is available; if
not, the measured variable may be freely selected from the following list:
see Fct. A4.1 "Measurement"
… stands for 1, 3 or 4
X stands for connection terminals A, C or D
C4.1 PV is _ _ _ Current output (primary variable)
C4.2 SV is _ _ _ (secondary variable)
C4.3 TV is _ _ _ (tertiary variable)
C4.4 4V is _ _ _ (4th variable)
C4.….1 Current output X Shows the current analog measured value of the linked frequency output.
The measured variable cannot be changed!
C4.….1 Frequency output X Shows the current analog measured value of the linked frequency output, if
present. The measured variable cannot be changed!
C4.….1 HART dynam. var. Measurements of the dynamic variables for HART®
Linear measured variables: volume flow / mass flow / diagnosis value / flow
speed / coil temperature / conductivity
Digital measured variables: counter 1 / counter 2 / operating hours

No. Function / subfunction Settings / descriptions

C5 Device

C5.1 Device info


C5.1 Device info
C5.1.1 Tag Settable characters (max. 8 places):
A…Z; a…z; 0…9; / - , .
C5.1.2 C number Electronic type, cannot be changed
C5.1.3 Device serial no. Serial no. of the system
C5.1.4 Electronic serial no. Serial no. of the electronic assembly, cannot be changed
C5.1.5 Information Serial no. of the circuit board, version no. of the main software, date of
manufacture of the circuit board

01/2008 • 4000041002 - MA IFC 100 R02 en www.krohne.com 63


6 OPERATION IFC 100

No. Function / subfunction Settings / descriptions

C5.2 Display
C5.2 Display
C5.2.1 Language Language selection depends on the device version.
C5.2.2 Contrast Adjust display contrast for extreme temperatures.
Setting: -9…0…+9
This change takes place immediately, not just when setting mode is exited!
C5.2.3 Default display Specification of the default display page that is returned to after a short
delay period.
Select: None (the current page is always active) / 1. meas. page (show this
page) / 2. meas. page (show this page) / Status page (show only status
messages) / Graphics page (trend display of the 1st measurement)
C5.2.4 Self test Not available at this time
C5.2.5 Information Serial no. of the circuit board, user software version no., date of
manufacture of the circuit board

C5.3 and C5.4 meas. page 1 and 2


C5.3 1. meas. page … stands for 3 = meas. page 1 and 4 = meas. page 2
C5.4 2. meas. page
C5.….1 Function Specify number of measured value lines (font size)
Select: one line / two lines / three lines
C5.….2 Measurement 1.line Specify measured variable for 1st line
Select: volume flow / mass flow / diagnosis value / flow speed / coil
temperature / conductivity
C5.….3 Range 0…100% of the measured variable set in Fct. C5.….2
0…xx.xx _ _ _ (format and unit depend on the measured variable)
C5.….4 Limitation Limitation before application of the time constant
xxx%; range: -120…+120%
C5.….5 Low flow cutoff Sets output value to "0": x.xxx ± x.xxx % Range: 0.0…20 %
(1st value = operating point / 2nd value = hysteresis)
Condition: 2nd value ≤ 1st value
C5.….6 Time constant Range: 0.1…100 s
C5.….7 Format 1.line Specify decimal places
Select: Automatic (adaptation carried out automatically) /
X (= none) …X.XXXXXXXX (max. 8 places)
C5.….8 Measurement 2.line Specify measured variable for 2nd line (only available if this 2nd line is
activated)
Select: bar graph (for the measured variable selected in the 1st line) /
volume flow / mass flow / diagnosis value / flow speed / counter 1 /
counter 2 / conductivity / coil temperature
C5.….9 Format 2.line Specify decimal places
Select: Automatic (adaptation carried out automatically) /
X (= none) …X.XXXXXXXX (max. 8 places)

64 www.krohne.com 01/2008 • 4000041002 - MA IFC 100 R02 en


IFC 100 OPERATION 6

No. Function / subfunction Settings / descriptions


C5.….10 Measurement 3.line Specify measured variable for 3rd line (only available if this 3rd line is
activated)
Select: volume flow / mass flow / diagnosis value / flow speed / coil
temperature / conductivity / counter 1 / counter 2
C5.….11 Format 3.line Specify decimal places
Select: Automatic (adaptation carried out automatically) /
X (= none) …X.XXXXXXXX (max. 8 places)

C5.5 Graphic page


C5.5 Graphic page
C5.5.1 Select range Graphic page always shows trend curve of the measurement of the
1st page / 1st line, see Fct.. C5.3.2
Select: Manual (set range in Fct. C5.5.2) /
Automatic (automatic depiction based on the measured values)
Reset only after parameter change or after switching off and on.
C5.5.2 Range Set the scaling for the Y axis. Only available if "Manual" is set in C5.5.1.
+xxx ±xxx%; range: -100…+100%
(1st value = lower limit / 2nd value = upper limit), condition: 1st value ≤
2nd value
C5.5.3 Time scale Set the time scaling for the X axis, trend curve
xxx min; range: 0…100 min

C5.6 Special functions


C5.6 Special functions
C5.6.1 Reset errors Reset error?
Select: No / Yes
C5.6.2 Save settings Save current settings
Select: break (exit function without saving) / backup 1 (save in storage
location 1) / backup 2 (save in storage location 2)
Query: go on with copy? (cannot be undone)
Select: No (exit function without saving) / Yes (copy current settings to
storage backup 1 or backup 2)
C5.6.3 Load settings Load saved settings
Select: break (exit function without loading) / factory settings (load in state
as delivered) / backup 1 (load data from storage location 1) / backup 2 (load
data from storage location 2)
Query: go on with copy? (cannot be undone)
Select: No (exit the function without saving) Yes (load data from the selected
storage location)
C5.6.4 Password quick setup Password required to change data in the quick setup menu
0000 (= to quick setup menu without password)
xxxx (password required); range 4place: 0001…9999
C5.6.5 Password setup Password required to change data in the setup menu
0000 (= to quick setup menu without password)
xxxx (password required); range 4place: 0001…9999

01/2008 • 4000041002 - MA IFC 100 R02 en www.krohne.com 65


6 OPERATION IFC 100

No. Function / subfunction Settings / descriptions

C5.7 Units
C5.7 Units
C5.7.1 Volume flow m³/h; m³/min; m³/s; l/h; l/min; l/s (l = liters);
ft³/h; ft³/min; ft³/s; gal/h; gal/min; gal/s;
free unit (set factor and text in the next two functions, sequence see below)
C5.7.2 Text free unit Text refer to Set free units on page 67 to be specified:
C5.7.3 [m³/s] × factor Specification of the conversion factor, based on m³/s:
xxx.xxx refer to Set free units on page 67
C5.7.4 Mass flow kg/s; kg/min; kg/h; t/min; t/h; g/s; g/min; g/h;
lb/s; lb/min; lb/h; ST/min; ST/h (ST = Short Ton); LT/h (LT = Long Ton);
free unit (set factor and text in the next two functions, sequence see below)

C5.7.5 Text free unit Text refer to Set free units on page 67 to be specified:
C5.7.6 [kg/s] × factor Specification of the conversion factor, based on kg/s:
xxx.xxx refer to Set free units on page 67
C5.7.7 Flow speed m/s; ft/s
C5.7.8 Conductivity µS/cm; S/cm
C5.7.9 Temperature °C; °F; K
C5.7.10 Volume m³; l (liters); hl; ml; gal; IG; in³; ft³; yd³;
free unit (set factor and text in the next two functions, sequence see below)
C5.7.11 Text free unit Text refer to Set free units on page 67 to be specified:
C5.7.12 [m³] × factor Specification of the conversion factor, based on m³:
xxx.xxx refer to Set free units on page 67
C5.7.13 Mass kg; t; mg; g; lb; ST; LT; oz;
free unit (set factor and text in the next two functions, sequence see below)
C5.7.14 Text free unit Text refer to Set free units on page 67 to be specified:
C5.7.15 [kg] × factor Specification of the conversion factor, based on kg:
xxx.xxx refer to Set free units on page 67
C5.7.16 Density Kg/cm³; kg/l; kg/m³; lb/ft³; lb/gal;
free unit (set factor and text in the next two functions, sequence see below)
C5.7.17 Text free unit Text refer to Set free units on page 67 to be specified:
C5.7.18 [kg/m³] × factor Specification of the conversion factor, based on kg/m³:
xxx.xxx refer to Set free units on page 67

66 www.krohne.com 01/2008 • 4000041002 - MA IFC 100 R02 en


IFC 100 OPERATION 6

No. Function / subfunction Settings / descriptions

C5.8 HART®
C5.8 HART
C5.8.1 HART Switch HART® communication on/off:
Select: on (HART® activated) current = 4…20 mA / off (HART® not activated)
current = 0…20 mA
C5.8.2 Address Set address for HART® operation:
Select: 00 (point to point operation, current output has normal function,
current = 4…20 mA) / 01...15 (multidrop operation, current output has a
constant setting of 4 mA)
C5.8.3 Message Set required text:
A…Z ; a…z ; 0…9 ; / - + , . *
C5.8.4 Description Set required text:
A…Z ; a…z ; 0…9 ; / - + , . *

C5.9 Quick setup


C5.9 Quick setup Activate quick access in quick setup menu:
Select: yes (switched on) / no (switched off)
C5.9.1 Reset counter 1 Reset counter 1 in quick setup menu?
Select: yes (activated) / no (switched off)
C5.9.2 Reset counter 2 Reset counter 2 in quick setup menu?
Select: yes (activated) / no (switched off)

6.3.4 Set free units

Free units Sequences to set tests and factors


Texts
Volume flow, mass flow and density: 3 places before and after the slash
xxx/xxx (max. 3 places before / after the slash)
Volume, mass xxx (max. 3 places)
Permissible characters: A…Z; a…z; 0…9; / - + , . *; @ $ % ~ () [] _
Conversion factors
Desired unit = [unit see above] × conversion factor
Conversion factor Max. 9 places
Shift decimal point: ↑ to the left and ↓ to the right

01/2008 • 4000041002 - MA IFC 100 R02 en www.krohne.com 67


6 OPERATION IFC 100

6.3.5 Reset counter in Quick Setup menu

INFORMATION!
It may be necessary to activate resetting of the counter in the Quick Setup menu in function C5.9.

Key Display Description and setting


→ Quick setup Time counts backwards from 2.5...0.0 s,
after that release the key.
→ Language
2x↓ Reset?
→ Reset error?
↓ Reset counter 1 Select desired counter
↓ Reset counter 2
→ Reset counter
→ No
↓ or ↑ Yes
^ Reset counter Counter has been reset
2x^ Measuring operation

6.3.6 Deleting error messages in the Quick Setup menu

NOTE!
The detailed list of the possible error messages refer to Status messages and diagnostic
information on page 69.

Key Display Description and setting


→ Quick setup Time counts backwards
from 2.5...0.0 s, after that
release the key.
→ Language
2x↓ Reset?
→ Reset error?
→ Reset? no
↓ or ↑ Reset? Yes
^ Reset errors Error has been reset
2x^ Measuring operation

68 www.krohne.com 01/2008 • 4000041002 - MA IFC 100 R02 en


IFC 100 OPERATION 6

6.4 Status messages and diagnostic information


Operational faults in device
Messages on the display Description Actions

Status: F _ _ _ _ _ Operational fault in device, mA output ≤ 3.6 Repair necessary.


mA or set fault current (depending on the
seriousness of the fault), status output open,
pulse / frequency output: no pulses
F error in device Fault or failure of device. Parameter or Group message, when one of the following
hardware error. No measurement possible. or some other severe error occurs.
F IO 1 Error, operational fault in IO 1. Parameter Load settings (Fct. C4.6.3) (Backup 1,
or hardware error. No measurement Backup 2 or factory settings). If status
possible. message still does not disappear, replace
electronic unit.
F parameter Error, operational fault of data manager,
parameter or hardware error. Parameters
no longer usable.
F IO 2 Error, operational fault in IO 2. Parameter
or hardware error. No measurement
possible.
F configuration Invalid configuration: display software, bus If device configuration unchanged:
parameter or main software do not match defective, replace electronic unit.
existing configuration.
F display Error, operational fault in display. Defective, replace electronic unit.
Parameter or hardware error. No
measurement possible.
F sensor electronic Error, operational fault in sensor Defective, replace electronic unit.
electronics. Parameter or hardware error.
No measurement possible.
F sensor global Data error in the global data of the sensor Load settings (Fct. C5.6.3, (Backup 1, 2 or
electronic equipment. factory settings). If status message still
does not disappear, replace electronic unit.
F sensor local Data error in the local data of the sensor Defective, replace electronic unit.
electronic equipment.
F field current local Data error in the local data of the field Defective, replace electronic unit.
current supply
F current output A Error, operational fault in current output. Defective, replace electronic unit
Parameter or hardware error. No
measurement possible.
F software user interface Fault revealed by CRC check of operator Replace electronic unit.
software.
F hardware settings The set hardware parameters do not match Answer queries in dialogue mode, follow
the identified hardware. A dialogue appears directions.
in the display. Defective, replace electronic unit.
F hardware detection Existing hardware cannot be identified. Replace electronic unit.
F RAM/ROM error IO1 A RAM or ROM error is detected during the Defective, replace electronic unit.
CRC check.
F RAM/ROM error IO2
F Fieldbus Malfunction of the fieldbus interface

01/2008 • 4000041002 - MA IFC 100 R02 en www.krohne.com 69


6 OPERATION IFC 100

Measurements out of specification


Messages on the display Description Actions

Status: S _ _ _ _ _ Out of specification, measurement Maintenance required.


continues, accuracy possibly less.
S uncertain measurement Device maintenance necessary; measured Group message, when errors as described
values only conditionally usable. below or other influences occur.
S empty pipe 1 or 2 measuring electrodes are not in Filling level of EMF less than 50% or
contact with the medium: measured value is electrodes completely insulated. If "0" to be
set to zero. Measurement continues. indicated when pipe is empty, activate under
Fct. C 1.3.1 "cond.+empty pipe [F]".
The two empty pipe messages cannot appear at the same time. The difference lies in
whether the measured value is also set to zero upon the detection of an empty pipe. The
sensor electronics will use one or the other function (setting to zero or further
measurement) depending on a selection made by the user.
Electrode noise Noise on the electrodes too high. Measured a) Electrodes extremely soiled;
values are still supplied. No message when b) Conductivity too low: activate noise or
empty pipe. pulse filter Fct. C1.2.4, C1.2.7;
c) Gas bubbles, solids or chem. reactions in
medium: activate noise or pulse filter Fct.
C1.2.4, C1.2.7;
d) Electrode corrosion (if message also
appears when flow is zero): use sensor with
suitable electrode material
S gain error Preampfl. not equal to the calibrated value; Defective, replace electronic unit.
check calibration. Measured values are still
supplied.
S electrode symmetry Impedance of the two measuring electrodes Deposits in measuring tube or electrode
not equal. Measured values are still short-circuit to ground. Clean and check
supplied. measuring tube!
S field coil broken Field coil resistance too high. Check field coil connections to the
electronic module (for remote versions:
S field coil bridged Field coil resistance too low.
field current cable) for open circuit / short
circuit
S field current deviation Measured field current not equal to the Check field current connections. If OK:
calibrated value. Check calibration. defect, replace electronic unit.
Measured values are still supplied. No
message if coil broken or bridged.
S electronic temperature Upper limit for the permissible electronic Ambient temperature too high, direct solar
temperature has been exceeded. radiation or, for C version, process temp.
too high.
S coil temperature Upper limit for the permissible coil Process and ambient temperature too high.
temperature has been exceeded. No
message if coil broken/bridged.
S overflow counter 1 This is counter 1 or FB2 (with Profibus).
Counter has overrun and started again at
zero.
S overflow counter 2 This is counter 2 or FB3 (with Profibus).
Counter has overrun and started again at
zero.
S backplane invalid The data record on the backplane is invalid. No data can be loaded from the backplane
The CRC check has revealed a fault. when the electronics are replaced. Replace
housing.

70 www.krohne.com 01/2008 • 4000041002 - MA IFC 100 R02 en


IFC 100 OPERATION 6

Simulation of the measured values


Messages on the display Description Actions

Status: C _ _ _ _ _ Output values partially simulated or fixed Maintenance required.


C checks in progress Test mode of the device. Measured values Message depending on the situation via
are possibly simulated values or values with HART® or FDT.
fixed settings.
Test sensor Test function of the measuring sensor
electronics active.
Field bus simulation Values on the field bus interface foundation
are simulated.

Information
Messages on the display Description Actions

Status: I _ _ _ _ _ Information (current measurement OK)


I counter 1 stopped This is counter 1 or FB2 (with Profibus). The If counter to continue counting, activate
counter has stopped. "yes" in Fct. C2.y.9 (Start counter).
I counter 2 stopped This is counter 2 or FB3 (with Profibus). The
counter has stopped.
I power fail The device was not in operation for an Temporary line failure, during which
unknown period of time, because the counters stop counting.
emergency current was switched off. This
message is for information only.
I over range display 1 1st line on page 1 (2) of display limited by Menu display Fct. C4.3 and/or C4.4, select
filter setting. meas. page 1 or 2 and increase values in
functions C4.z.3 Meas. range and/or C4.z.4
Limitation
I backplane sensor The sensor data on the backplane are not
usable because they have been generated
with an incompatible version.
I backplane settings The global settings on the backplane are not
usable because they have been generated
with an incompatible version.
I backplane difference The data on the backplane differ from the
data in the display. If the data are usable, a
dialogue is indicated in the display.
I write cycles overfl. The maximum number of write cycles of the
EEPROM or FRAMS on the Profibus DP PCB
has been exceeded.
I baudrate search The baudrate of the Profibus DP interface is
searched for.
I no data exchange There is no data exchange between the
signal converter and the Profibus.
I conductivity off Conductivity measurement switched off. Changing of settings in Fct. C1.3.1.
I diagnosis channel off Diagnosis value switched off. Changing of settings in Fct. C1.3.17.

01/2008 • 4000041002 - MA IFC 100 R02 en www.krohne.com 71


7 SERVICE IFC 100

7.1 Spare parts availability


The manufacturer adheres to the basic principle that operational spare parts for each flowmeter
or each important accessory part will be kept available for period of 10 (ten) years after delivery
of the last production run for that device.

Operational spare parts are defined as parts that are subject to faults in normal operation.

7.2 Availability of services


The manufacturer offers a range of services to support the customer after expiration of the
warranty. These include repair, technical support and training.

NOTE!
For more precise information, please contact your local representative.

72 www.krohne.com 01/2008 • 4000041002 - MA IFC 100 R02 en


IFC 100 SERVICE 7

7.3 Returning the device to the manufacturer


7.3.1 General information

This device has been carefully manufactured and tested. If installed and operated in accordance
with these operating instructions, it will rarely present any problems.

CAUTION!
Should you nevertheless need to return a device for inspection or repair, please pay strict
attention to the following points:
• Due to statutory regulations on environmental protection and safeguarding the health and
safety of our personnel, manufacturer may only handle, test and repair returned devices that
have been in contact with products without risk to personnel and environment.
• This means that the manufacturer can only service this device if it is accompanied by the
following certificate (see next section) confirming that the device is safe to handle.

CAUTION!
If the device has been operated with toxic, caustic, flammable or water-endangering products,
you are kindly requested:
• to check and ensure, if necessary by rinsing or neutralizing, that all cavities are free from
such dangerous substances,
• to enclose a certificate with the device confirming that is safe to handle and stating the
product used.

01/2008 • 4000041002 - MA IFC 100 R02 en www.krohne.com 73


7 SERVICE IFC 100

7.3.2 Form (for copying) to accompany a returned instrument

Company: Address:

Department: Name:

Tel. no.: Fax no.:

Manufacturer's order no. or serial no.:

The device has been operated with the following medium:

This medium is: water-hazardous


toxic
caustic
flammable
We checked that all cavities in the device are free from such
substances.
We have flushed out and neutralized all cavities in the
device.

We hereby confirm that there is no risk to persons or the environment through any residual media
contained in the device when it is returned.

Date: Signature:

Stamp:

7.4 Disposal
CAUTION!
Disposal must be carried out in accordance with legislation applicable in your country.

74 www.krohne.com 01/2008 • 4000041002 - MA IFC 100 R02 en


IFC 100 TECHNICAL DATA 8

8.1 Technical data

Measuring system
Measurement principle Faraday's law of induction
Function Continuous measurement of current volume flow, flow velocity,
conductivity, mass flow (at constant density), coil temperature of the
measuring sensor
Modular construction The measurement system consists of a measuring sensor and a
signal converter
Signal converter
Compact version (C) IFC 100 C (0° & 45° version)
Remote version (W) IFC 100 W
Measuring sensor
OPTIFLUX 1000 IFC 100 C & W: DN10...150 / 3/8"…6"
OPTIFLUX 2000 IFC 100 C & W: DN25...1200 / 1"…48"
OPTIFLUX 4000 IFC 100 C: DN2.5...1200 / 1/10"…48";
IFC 100 W: DN10...1200 / 3/8"…48";
OPTIFLUX 5000 IFC 100 C: DN2.5...250 / 1/10"…12";
IFC 100 W: DN10...250 / 3/8"…12"
OPTIFLUX 6000 IFC 100 C: DN2.5...150 / 1/10"…6";
IFC 100 W: DN10...150 / 3/8"…6"
Communication
Outputs Current (incl. HART®), pulse, frequency, status output and/or limit
switch
Counter 2 internal counters with a max. of 8 counter places (e.g. for counting
volume and/or mass units)
Verification Integrated verification, diagnosis functions: flowmeter, empty pipe
detection, stabilization
Display and user interface
Graphic display LC display, backlit white; size: 128x64 pixels, corresponds to 59x31
mm = 2.32"x1.22"
Display functions 2 measured value pages, 1 status page, 1 graphic page (measured
values and depictions adjustable as required)
Units Metric, British and US units selectable as required from lists for
volume / mass flow and counting, flow speed, electrical conductivity,
temperature
Language of display texts English, French, German (others on request)
Operating elements 4 keys for operator control of the signal converter without opening
the housing

01/2008 • 4000041002 - MA IFC 100 R02 en www.krohne.com 75


8 TECHNICAL DATA IFC 100

Measuring accuracy
Maximum measuring error ±0.3% of the measured value ±1 mm/s, depending on the measuring
sensor (see accuracy curves)
Repeatability ±0.1%

Operating conditions
Temperature
Process temperature See also data sheet for the measuring sensor
Ambient temperature -40…+65°C / -40…+149°F (ambient temperature 55°C / 131°F and
higher: protect electronics against self-heating, because an increase
in the electronics temperature in 10°C / 50°F steps leads to a
corresponding reduction of the electronics' service life by a factor of
two.)
Storage temperature -50…+70°C / -58…+158°F
Electrical conductivity
All media except for water Min. 5 µS/cm (see also data sheet for the measuring transformer)
Water Min. 20 µS/cm

Materials
Die-cast aluminium Standard
(polyurethane-coated)

Electrical connection
Voltage Standard: 100…230 VAC (-15% / +10%), 50/60 Hz
Option 1: 24 VDC (-55% / +30%)
Option 2: 24 VAC/DC (AC: -15% / +10%; DC: -25% / +30%)
Power consumption Standard: 8 VA
Option 1: 4 W
Option 2: AC 8 VA; DC: 4 W
Signal cable Only for remote versions
A: type DS 300 Max. length: 600 m / 1950 ft (depending on electrical conductivity and
measuring sensor version)
Cable entries Standard: M20 x 1.5
Option: ½" NPT, PF ½

76 www.krohne.com 01/2008 • 4000041002 - MA IFC 100 R02 en


IFC 100 TECHNICAL DATA 8

Outputs
Current output
Function Measurement of volume and mass (at constant density), HART®
communication
Settings Without HART® With HART®
Q = 0%: 0…15 mA Q = 0%: 4…15 mA
Q = 100%: 10…21.5 mA Q = 100%: 10…21.5 mA
Error identification: 0…22 mA Error identification:
3.5…22 mA
Operating data
Active Uint,nom = 24 VDC
I ≤ 22 mA
RL ≤ 750 Ω
Passive Uext ≤ 32 VDC
I ≤ 22 mA
U0 ≤ 2 V at I = 22 mA
Pulse or frequency output
Function Can be set as a pulse output (e.g.- for volume or mass counting) or
frequency output
Settings For Q = 100%: 0.01...10000 pulses per second or pulses per unit
volume
Pulse width: setting automatic, symmetric or fixed (0.05...2000
ms)manual
Operating data
Passive Uext ≤ 32 VDC
100 Hz < fmax ≤ 10 kHz:
I ≤ 20 mA
open:
I ≤ 0.1 mA at Uext = 5 V
I ≤ 0.5 mA at Uext = 24 V
I ≤ 0.7 mA at Uext = 32 V
closed:
U0 ≤ 0.8 V at I = 1 mA
U0 ≤ 1.5 V at I = 10 mA
U0 ≤ 3.5 V at I = 100 mA
f ≤ 1 kHz: RL ≤ 10 kΩ
f ≤ 10 kHz: RL ≤ 2 kΩ

01/2008 • 4000041002 - MA IFC 100 R02 en www.krohne.com 77


8 TECHNICAL DATA IFC 100

Status output / limit switch


Function and settings Settable as automatic measuring range change, indicator for
direction of flow, overflow, error, operating point or empty pipe
detection
Valve control with activated dosing function
Status and/or control: ON or OFF
Operating data
Passive Uext ≤ 32 VDC
I ≤ 100 mA
open:
I ≤ 0.05 mA at Uext = 32 VDC
closed:
U0 ≤ 0.2 V at I = 10 mA
U0 ≤ 2 V at I = 100 mA
Low-flow cutoff
On 0...±9.999 m/s; 0...20.0%, settable in 0.1% steps, separately for each
current and pulse output
Off 0...±9.999 m/s; 0...19.0%, settable in 0.1% steps, separately for each
current and pulse output
Time constant
Function Can be set together for all flow indicators and outputs, or separately
for: current, pulse and frequency output, and for limit switches and
the 2 internal counters
Time setting 0…100 seconds, settable in 0.1 second steps

Approvals
Hazardous areas
Non-Ex Standard
EEx - Zone 1/2 In preparation
SAA version Ex Zone 1/2 In preparation
TIIS - Zone 1/2 In preparation
Protection category to IEC 529 / EN 60529
All versions IP 66 / 67 (corresponds to NEMA 4X/6)

78 www.krohne.com 01/2008 • 4000041002 - MA IFC 100 R02 en


IFC 100 TECHNICAL DATA 8

8.2 Flow tables


Flow rate in m/s and m3/h
Q100 % in m3/h

v [m/s] 0.3 1 3 12

DN [mm] Min. flow Nominal flow Max. flow


2.5 0.01 0.02 0.05 0.21
4 0.01 0.05 0.14 0.54
6 0.03 0.10 0.31 1.22
10 0.08 0.28 0.85 3.39
15 0.19 0.64 1.91 7.63
20 0.34 1.13 3.39 13.57
25 0.53 1.77 5.30 21.21
32 0.87 2.90 8.69 34.74
40 1.36 4.52 13.57 54.29
50 2.12 7.07 21.21 84.82
65 3.58 11.95 35.84 143.35
80 5.43 18.10 54.29 217.15
100 8.48 28.27 84.82 339.29
125 13.25 44.18 132.54 530.15
150 19.09 63.62 190.85 763.40
200 33.93 113.10 339.30 1357.20
250 53.01 176.71 530.13 2120.52
300 76.34 254.47 763.41 3053.64
350 103.91 346.36 1039.08 4156.32
400 135.72 452.39 1357.17 5428.68
450 171.77 572.51 1717.65 6870.60
500 212.06 706.86 2120.58 8482.32
600 305.37 1017.90 3053.70 12214.80
700 415.62 1385.40 4156.20 16624.80
800 542.88 1809.60 5428.80 21715.20
900 687.06 2290.20 6870.60 27482.40
1000 848.22 2827.40 8482.20 33928.80
1200 1221.45 3421.20 12214.50 48858.00

01/2008 • 4000041002 - MA IFC 100 R02 en www.krohne.com 79


8 TECHNICAL DATA IFC 100

Flow rate in ft/s and gallons/min


Q100 % in .US gallons/min

v [ft/s] 1 3.3 10 40

DN [inch] Min. flow Nominal flow Max. flow


1/10 0.02 0.09 0.23 0.93
1/8 0.06 0.22 0.60 2.39
1/4 0.13 0.44 1.34 5.38
3/8 0.37 1.23 3.73 14.94
1/2 0.84 2.82 8.40 33.61
3/4 1.49 4.98 14.94 59.76
1 2.33 7.79 23.34 93.36
1.25 3.82 12.77 38.24 152.97
1.5 5.98 19.90 59.75 239.02
2 9.34 31.13 93.37 373.47
2.5 15.78 52.61 159.79 631.16
3 23.90 79.69 239.02 956.09
4 37.35 124.47 373.46 1493.84
5 58.35 194.48 583.24 2334.17
6 84.03 279.97 840.29 3361.17
8 149.39 497.92 1493.29 5975.57
10 233.41 777.96 2334.09 9336.37
12 336.12 1120.29 3361.19 13444.77
14 457.59 1525.15 4574.93 18299.73
16 597.54 1991.60 5975.44 23901.76
18 756.26 2520.61 7562.58 30250.34
20 933.86 3112.56 9336.63 37346.53
24 1344.50 4481.22 13445.04 53780.15
28 1829.92 6099.12 18299.20 73196.79
32 2390.23 7966.64 23902.29 95609.15
36 3025.03 10082.42 30250.34 121001.37
40 3734.50 12447.09 37346.00 149384.01
48 5377.88 17924.47 53778.83 215115.30

80 www.krohne.com 01/2008 • 4000041002 - MA IFC 100 R02 en


IFC 100 TECHNICAL DATA 8

8.3 Accuracy
Reference conditions
• Medium: water
• Temperature: 20°C / 68°F
• Pressure: 1 bar / 14.5 psi
• Inlet run: ≥ 5 DN

• X [m/s]: flow velocity


• Y [%]: deviation from the actual measured value (mv)

DN [mm] DN [inch] Accuracy Curve


OPTIFLUX 2100 / 4100 / 5100 / 10…1200 3/8…48 0.3% of mv +1 mm/s 1
6100
OPTIFLUX 1100 10…150 3/8…6 0.4% of mv +1 mm/s as 1 + 0.1%
OPTIFLUX 4100 / 5100 / 6100 2.5…6 1/10…1/4

01/2008 • 4000041002 - MA IFC 100 R02 en www.krohne.com 81


8 TECHNICAL DATA IFC 100

8.4 Dimensions and weights


8.4.1 Housing

Wall-mounted version

Dimensions and weight in mm and kg


Dimensions [mm] Weight
[kg]
a b c d e f g h i k
Wall-mounted 161 40 87.2 120 155 241 95.2 257 19.3 39.7 Std: 1.9
version Ex: 2.4

Dimensions and weight in inches and lbs


Dimensions [inches] Weight
[lbs]
a b c d e f g h i k
Wall-mounted 6.34 1.57 3.43 4.72 6.10 9.50 3.75 10.12 0.76 1.56 Std: 4.2
version Ex: 5.3

82 www.krohne.com 01/2008 • 4000041002 - MA IFC 100 R02 en


IFC 100 TECHNICAL DATA 8

Compact 0° version

1 4xM6

Dimensions and weight in mm and kg


Dimensions [mm] Weight
[kg]
a b c d e f g h
0° version 161 40 155 81.5 257 - - Ø72 Std: 1.9
Ex: 2.4

Dimensions and weight in inches and lbs


Dimensions [inches] Weight
[lbs]
a b c d e f g h
0° version 6.34 1.57 6.1 3.21 10.12 - - Ø2.83 Std: 4.2
Ex: 5.3

01/2008 • 4000041002 - MA IFC 100 R02 en www.krohne.com 83


8 TECHNICAL DATA IFC 100

Compact 45° version

1 4xM6

Dimensions and weight in mm and kg


Dimensions [mm] Weight
[kg]
a b c d e f g h
45° version 161 40 155 184 27.4 45° 186 Ø72 Std: 2.1
Ex: 2.6

Dimensions and weight in inches and lbs


Dimensions [inches] Weight
[lbs]
a b c d e f g h
45° version 6.34 1.57 6.10 7.24 1.08 45° 7.32 Ø2.83 Std: 4.6
Ex: 5.7

84 www.krohne.com 01/2008 • 4000041002 - MA IFC 100 R02 en


IFC 100 TECHNICAL DATA 8

8.4.2 Mounting plate, wall-mounted version

Dimensions in mm and inches

[mm] [inches]
a Ø6.5 Ø0.26
b 87.2 3.4
c 241 9.5

01/2008 • 4000041002 - MA IFC 100 R02 en www.krohne.com 85


8 TECHNICAL DATA IFC 100

86 www.krohne.com 01/2008 • 4000041002 - MA IFC 100 R02 en


IFC 100 TECHNICAL DATA 8

01/2008 • 4000041002 - MA IFC 100 R02 en www.krohne.com 87


IFC 100 nnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnn

KROHNE Product Overview


• Electromagnetic flowmeters • Level measuring instruments
• Variable area flowmeters • Pressure gauges
• Mass flowmeters • Temperature measuring instruments
• Ultrasonic flowmeters • Water solutions & analysis
• Vortex flowmeters • Oil and gas turnkey solutions
• Flow controllers

Addresses:

Germany KROHNE sales Great Britain


KROHNE Ltd.
Norway
KROHNE Norway A.S.
Representatives
companies Rutherford Drive
Park Farm Industrial Estate
Ekholtveien 114
NO-1521 Moss
Algeria
Northern sales office Wellingborough Phone:+47 (0)69 264 860
Argentina
KROHNE Messtechnik GmbH & Co. KG Northants NN8 6AE Fax:+47 (0)69 267 333
Cameroon
Bremer Str. 133
D-21073 Hamburg
International Phone:+44 (0)19 33 408 500
Fax:+44 (0)19 33 408 501
[email protected]
Poland
Canada
Chile
Phone:+49 (0)40 767 3340 [email protected] KROHNE Polska Sp.z.o.o. Columbia
Fax:+49 (0)40 767 33412 CIS ul. Stary Rynek Oliwski 8a Croatia
Australia
[email protected] Kanex KROHNE Engineering AG 80-324 Gdansk Denmark
KROHNE Australia Pty Ltd
ZIP code: 10000 - 29999, 49000 - 49999 Business-Centre Planeta Phone: +48 (0)58 520 9211 Ecuador
Quantum Business Park 10/287
Office 404 ul. Fax.:+48 (0)58 520 9212 Egypt
Western and middle sales office Victoria Rd Rydalmere NSW 2116
Marxistskaja 3 [email protected] Finland
Phone: +61 2 8846 1700
KROHNE Messtechnik GmbH & Co. KG 109147 Moscow/Russia Switzerland Gabon
Fax: +61 2 8846 1755
Ludwig-Krohne-Straße Phone:+7 (0)095 911 7165 KROHNE AG Ghana
[email protected]
D-47058 Duisburg Fax:+7 (0)095 742 8873 Uferstr. 90 Greece
Phone:+49 (0)203 301 416 Austria
[email protected] CH-4019 Basel Hong Kong
Fax:+49 (0)203 301 10416 KROHNE Gesellschaft m.b.H.
India Phone:+41 (0)61 638 30 30 Hungary
[email protected] Modecenterstraße 14
Krohne Marshall Ltd. Fax:+41 (0)61 638 30 40 Indonesia
ZIP code: 30000 - 34999, 37000 - A-1030 Vienna
A-34/35, M.I.D.C. Industrial Area, [email protected] Iran
48000, 50000 - 53999, 57000 - 59999, Phone:+43 (0)1/203 45 32
H-Block Ireland
Fax:+43 (0)1/203 45 32 99 Singapore
98000 - 99999 Pimpri Poona 411018 Israel
[email protected] Tokyo Keiso - KROHNE (Singapore)
Phone:+91 (0)202 744 2020 Ivory Coast
Southern sales office Belgium Pte. Ltd.
Fax:+91 (0)202 744 2020 Japan
14, International Business Park,
KROHNE Messtechnik GmbH & Co. KG KROHNE Belgium N.V. Jordan
[email protected] Jurong East
Landsberger Str. 392 Brusselstraat 320 Kuwait
B-1702 Groot Bijgaarden Iran Chiyoda Building, #01-01/02
D-81241 Munich Libya
KROHNE Liaison Office Singapore 609922
Phone:+49 (0)89 121 5620 Phone:+32 (0)2 4 66 00 10 Lithuania
North Sohrevardi Ave. 26, Phone: (65) 6567 4548
Fax:+49 (0)89 129 6190 Fax:+32 (0)2 4 66 08 00 Malaysia
[email protected] Sarmad St., Apt. #9 Fax : (65) 6567 9874
[email protected] Mauritius
Tehran 15539 [email protected]
ZIP code: 0 - 9999, 80000 - 89999, Brazil Mexico
Phone: +9821 8874 5973 Republic of South Africa
90000 - 97999 KROHNE Conaut Controles Morocco
Fax: +9821 8850 1268 KROHNE Pty. Ltd.
Automaticos Ltda. New Zealand
[email protected] Bushbock Close
Southwestern sales office Estrada Das Águas Espraiadas, 230 Peru
Italy Corporate Park South
Subject to change without notice

KROHNE Messtechnik GmbH & Co. KG C.P. 56 06835 - 080 EMBU - SP Portugal
KROHNE Italia Srl. Midrand, Gauteng Romania
Rüdesheimer Str. 40 Phone:+55 (0)11-4785-2700
Fax:+55 (0)11 4785-2768 Via V. Monti 75 P.O. Box 2069 Saudi Arabia
D-65239 Hochheim/Main
I-20145 Milan Midrand, 1685 Senegal
Phone: +49(0)6146) 827 30 [email protected]
Phone:+39 02 4300 661 Tel.: +27 (0)11 314 1391 Slovakia
Fax:+49 (0)6146 827 312 China
Fax:+39 02 4300 6666 Fax: +27 (0)11 314 1681 Slovenia
[email protected] KROHNE Measurement Instruments [email protected] [email protected] Sweden
ZIP code: 35000 - 36999, 54000 - (Shanghai) Co. Ltd., (KMIC)
56999, 60000 - 79999 Korea Spain Taiwan
Room 1501
KROHNE Korea I.I. KROHNE IBERIA, S.r.l. Thailand
1033 Zhaojiabang Road
Instrumentation and control Room 508 Miwon Bldg 43 Poligono Industrial Nilo Tunisia
Shanghai 200030
equipment catalog Yoido-Dong Youngdeungpo-Ku Calle Brasil, nº. 5 Turkey
Phone: +86 21 6487 9611
TABLAR Messtechnik GmbH Seoul, Korea 28806 Alcalá de Henares Madrid Venezuela
Fax:+86 21 6438 7110
Ludwig-Krohne-Straße 5 Phone: 00-82-2-782-1900 Phone: +34 (0)91 883 2152 Yugoslavia
[email protected]
D-47058 Duisburg Fax: 00-82-2-780-1749 Fax: +34 (0)91 883 4854
Czech Republic [email protected] [email protected]
Phone:+49 (0)2 03 305 880
Krohne CZ, spol. s r.o.
Other countries
4000041002 - MA IFC 100 R02 en

Fax:+49 (0)2 03 305 8888 Netherlands USA


Sobìsická 156
[email protected] www.tablar.de KROHNE Nederland B.V. KROHNE, Inc.
63800 Brno
Kerkeplaat 14 7 Dearborn Road
Phone: +420 (0)545.242 627
NL-3313 LC Dordrecht Peabody, MA 01960
Fax: +420 (0)545 220 093 KROHNE Messtechnik GmbH & Co. KG
Phone:+31 (0)78 630 6200 Phone: +1 (800) FLOWING
[email protected] Ludwig-Krohne-Str. 5
Fax:+31 (0)78 630 6405 Phone: +1 (978) 535 6060 (in MA)
France D-47058 Duisburg
Service Direct: +31 (0)78 630 6222 [email protected]
KROHNE S.A.S. Phone:+49 (0)203 301 0
[email protected]
Les Ors BP 98 Fax:+49 (0)203 301 389
F-26103 ROMANS Cedex [email protected]
Phone:+33 (0)4 75 05 44 00
Fax:+33 (0)4 75 05 00 48
[email protected]
© KROHNE 01/2008

www.krohne.com

You might also like