SERVICE MANUAL UPDATE RECORD
MODEL: XS-series
Rev ECR/TB Applicable Released Summary Page
NO. NO. S/N Date
1 200610 Not Applicable 2006/04/06 Updating XS Series Section 7 entire pages, A-2-1 thru A-2-30
Service Manual
2 306E031 XS-1000i S/N 11653 2006/06/27 Adding Fixing Metal A-1-46, A-1-47
and thereafter No.365 to prevent from
XS-800i S/N 11487 dropping O-ring
and thereafter
3 306C082 XS-1000i S/N 11315 2006/06/29 Changing blank value of 4-30, 4-31
and thereafter Hgb and aspiration sensor
XS-800i S/N 11143
and thereafter
4 306C104-1 S/N not applicable 2006/08/11 Replacing Pharmed XS-series Service Manual, pages
Parts will be Tubing due to part being A-1-23, A-1-28(2/2), A-1-31, A-1-65, A-1-87
changed when our discontinued
stock is depleted.
Note:
Due to document
size and number of
instruments
involved, this ECR
has been divided
into 3 sections.
These sections will
be identified as ECR
306C104-1, -2, and
-3. Refer to Table
above for models
covered in this
section of ECR
306C104.
5 306G079 XS-1000 S/N 11733 2006/09/07 Changing aspiratin sensor XS-series Service Manual, pages
and thereafter setting value (includes A-1-23, A-1-28(2/2), A-1-31, A-1-65, A-1-87
XS-800i S/N 11615 updates to service manual)
and thereafter
6 306H048 XS-1000i S/N 11946 2006/10/13 Changing XS DP for A-1-28 (1/2)
and thereafter charging to prevent
XS-800i S/N 11763 stucking
and thereafter
Parts will be
changed when our
stock is depleted.
7 306F053 S/N not fixed yet 2006/10/13 Upgrading the XS-series Entire appendix C
program to version 00-08
8 306J058 XS-1000i S/N: 2006/10/31 Changing aspiration 5-30, 5-31
11735, 11739, sensor threshold level
11744, 11788,
11816, 11820,
11825, 11849,
11850, 11869,
11895, 11908,
11925 and
thereafter
XS-Series - 1/2 -
9 306E 066-1 S/N: Not Applicable 2006/11/07 Changing spacers used in XS-series Service Manual, pages
Spare parts will be Sysmex instruments A-1-8(2/3), A-1-80, A-2-11, A-2-17, A-2-29
changed when our
stock is depleted.
10 200635 S/N not applicable 2006/12/04 Updating the XS-series 1-16, 4-32, 4-33, B TOC, B-33 through B-50
service manual
11 306K086 XS-1000i S/N 2006/12/21 Countermeasure against A-1-7, A-1-8, 7-2
11560, 11561, frequent IG flag
11613, 11615
through 11617,
11646, 11647,
11660, 11661,
11704,11705,
11710 , 11712,
11714 through
11717, 11756,
11758 through
11760, 11767,
11818 through
11835, 11837
through 11999,
12000 and
thereafter
XS-800i S/N 11887
and thereafter
12 306I072 XS-1000i S/N 11979 2006/12/25 Changing Releaf Valve to A-1-51
and thereafter prevent from stuck
XS-800i S/N 11825
and thereafter
13 306I093 S/N not fixed yet 2007/01/22 Changing RBC isolation A-1-31
chamber to prevent PLT
sampling error
14 306J014 XS-1000i S/N 62139 2007/03/15 Changing XS-series cover A-1-6(1/2), A-2-5
and thereafter color
XS-800i S/N 62038
and thereafter
OPSU-11 S/N
61769 and
thereafter
15 306L001 S/N not applicable. 2007/03/20 Changing Power Cord A-1-107, A-1-109
Parts will be included in Sysmex supply
changed when our parts to conform to the
stock is depleted. RoHS directive (European
markets only)
16 200714 Serial No: Not 2007/03/29 Adding information to Appen C TOC, C-19, C-27
Applicable XS-series Service Manual
Appendix:C SNCS
17 306K011 S/N not fixed yet 2007/05/17 Upgrading the XS-series 5-26, 5-28
program to version 00-10
18 307D079 S/N not fixed yet 2007/05/21 Changing slide rail used A-1-43, A-2-13, A-2-15
on Sysmex products
19 307D033 S/N not fixed yet 2007/05/28 Changing the color of A-2-3
sample rack
20 307B082 S/N: Not fixed yet. 2007/05/30 Addition of Wiring Cord A-1-14
No.3159 for XS-Series
XS-Series - 2/2 -