Fabius GS Premium: Instructions For Use
Fabius GS Premium: Instructions For Use
Fabius GS premium
Any text shown on the screen and any labeling on The actual screen layout or the device may differ in
the device are printed in bold and italics, e.g., appearance or in configuration from the
PEEP or Man/Spon. illustrations.
Trademarks
WARNING CAUTION
A WARNING statement provides important A CAUTION statement provides important
information about a potentially hazardous information about a potentially hazardous
situation which, if not avoided, could result in situation which, if not avoided, may result in
death or serious injury. minor or moderate injury to the user or patient or
in damage to the medical device or other
property.
NOTE
A NOTE provides additional information intended
to avoid inconvenience during operation.
Contents
For your safety and that of your patients.... 7 Fastening the manual resuscitator .................. 81
General safety information .............................. 8 Fitting the COSY cover.................................... 82
Product-specific safety information.................. 12 Instructions for mounting the accessories ....... 83
WARNING WARNING
Risk of operating errors and incorrect use Risk of device malfunction
Strictly observe the instructions for use of all This medical device can be operated in
accessory parts, e.g.,: combination with other Dräger devices or
– Water traps with devices from other manufacturers. If a
– Flow sensors device combination is not approved by
– CLIC adapter Dräger, the safety and the functional state of
– CLIC absorber the individual devices can be compromised.
– Soda lime – The operating organization must ensure
– Breathing hoses that the device combination complies with
– Masks the applicable editions of the relevant
– Filter standards for medical devices.
– Endotracheal suction – Strictly observe the assembly instructions
– Vaporizer and instructions for use of each
– Manual resuscitator connected device.
– AGSS terminal unit
WARNING WARNING
Risk of electric shock and device malfunction Risk of explosion and fire
Any connected devices or device This medical device is neither approved nor
combinations not complying with the certified for use in areas where oxygen
requirements in these instructions for use concentrations greater than 25 Vol%,
may compromise correct functioning of the combustible or explosive gas mixtures are
medical device. likely to occur.
Before using the medical device, refer to and
strictly comply with the instructions for use
of all connected devices and device
combinations.
CAUTION
Risk of patient injury
Do not make therapeutic decisions based solely
on individual measured values and monitoring
parameters.
WARNING WARNING
Risk of misinterpretation Risk of patient injury
Misinterpretation of measured values or other Every user has the obligation to assess
parameters or misdiagnosis can endanger independently which components are
the patient. required corresponding to the specific
prerequisites for the anesthesia workstation.
Do not make therapeutic decisions based
In accordance with the general safety
solely on individual measured values and
standards for anesthesia systems, additional
monitoring parameters. Therapeutic
monitoring of the concentrations of CO2 and
decisions must be made solely by the user.
anesthetic agent is required when operating
the device. To guarantee patient safety,
WARNING however, the following components must
Risk of burns always be used:
– O2 monitor
Conductive breathing hoses or face masks – Pressure monitor
can cause burns during HF surgery. – Volume monitor
Do not use this type of hose and mask
combined with HF surgery. WARNING
Risk of malfunction
WARNING
Unallowed modifications to the medical
Risk of malfunction device lead to malfunctions.
Device failure or user error can compromise This medical device may not be changed
the correct functioning of the device. The without permission from Dräger.
medical device does not react automatically
to certain changes in the patient condition,
WARNING
operating errors, or failure of components.
Risk of accidental movement of the medical
Continuously monitor the medical device so
device
that corrective measures can be initiated
immediately. During operation the medical device can
move accidentally.
WARNING Activate the castor brakes.
Risk of device failure
The device can fail if the power supply is
interrupted.
Always connect the device on an
uninterruptible power supply.
WARNING WARNING
Risk of tipping over during transport Risk due to failure of flow measurement
The medical device may tip over if handled Deposits that were not removed during
incorrectly. Observe the following points reprocessing can damage the measuring
when transporting medical devices: wires in the flow sensor or cause a fire.
– The medical device may only be moved by – Before inserting the flow sensor check for
people who have the physical ability to do visible damage, soiling, and particles.
so. Repeat this check regularly.
– To improve the maneuverability, transport – Replace flow sensors when damaged,
the device with 2 persons. soiled, or not particle-free.
– When transporting over inclines, around
corners, or over thresholds (e.g., through WARNING
doors or in elevators), make sure that the
medical device does not bump against Risk of insufficient ventilation
anything. Device failure or operating errors can lead to
– Remove any devices mounted to the ventilation failure.
holding arms or the top of the device. – To ensure immediate remedial action in
– Clear the writing tray and fold it down case of device failure, only operate the
completely or slide it into the device. device under permanent supervision of
– Do not pull the medical device over hoses, users.
cables, or other obstacles lying on the – The general safety standards for
floor. anesthesia systems require that a manual
– Do not activate the brake while the resuscitator be kept at the ready for
medical device is being moved. emergency ventilation.
– Always use the handles on the device to
push or pull it. WARNING
WARNING WARNING
Risk of crushing Risk of electric shock
If the writing tray is not correctly locked in This device is only intended for use in rooms
place, objects can fall down or fingers and in which the power lines correspond to the
breathing hoses, for example, can be national applicable safety standards for
pinched. patient's rooms in hospitals. Observe the
following points to avoid electric shock:
Make sure that the writing tray is correctly
– The covers of the components must not
locked when folding down or sliding into the
be removed.
device.
– Maintenance work must only be
performed by DrägerService. Use only
WARNING grounded electrical connections and
Risk of crushing power cables that meet hospital
standards.
Movable device parts or attached – Before connecting the medical device,
components may cause crushing due to make sure that external devices are
clamping. Pay special attention to edges, grounded to meet the hospital standard (in
movable parts, and corners when working accordance with national applicable
with the following components: regulations).
– Breathing system cover – Before cleaning work or maintenance
– Drawers work is performed, disconnect all plugs
– Extensible writing tray for the power supply.
– Swivel arms for mounted devices – If the medical device has come in contact
– Accessories such as gas cylinders, with liquids, let it dry completely before it
vaporizers, CLIC absorber, and CLIC is reconnected to the power supply.
adapter – Check that the power cable is securely
clamped to the power inlet.
– Only connect additional devices if they
have been approved by Dräger.
WARNING
Risk of device failures
If the anesthesia workstation is used in a
tipped position, parts can be damaged or
their function can be comprised.
Do not use the anesthesia workstation at an
inclination angle over 5°.
NOTE
The device software of Fabius must be installed
by experts. Dräger recommends having the
software installation performed by
DrägerService.
Application
Indications/Contraindications ...................... 17
Indications ....................................................... 17
Contraindications............................................. 17
Further information on application.............. 18
Environment of use.......................................... 18
The MEDIBUS and Vitalink protocols .......... 19
Intended use
* optional
NOTE NOTE
CO2 values and anesthetic gas values can be O2 monitoring can be deactivated on site by an
monitored if the Fabius is combined with a gas authorized service partner. More information can
monitor (e.g., Vamos) or a gas analyzer (e.g., be found in chapter ''Deactivating the O2
Scio with Dräger patient monitor). monitoring'' on page 122. If O2 monitoring is
deactivated, use external O2 monitoring.
Indications/Contraindications
Indications
Contraindications
NOTE
Fabius applies medical gases such as O2, N2O,
or volatile anesthetic agents. For
contraindications to the applied medical gases,
strictly observe the instructions for use of the
medical gas.
Environment of use
WARNING
Risk of explosion
This medical device is neither approved nor
certified for use in areas where oxygen
concentrations greater than 25 Vol%,
combustible or explosive gas mixtures are
likely to occur.
WARNING
Risk of device malfunctions and/or patient
injury and user injury
Magnetic fields can negatively influence the
correct functioning of the medical device and
therefore endanger the patient or user.
Do not use the medical device near magnetic
resonance imagers.
WARNING
Do not use soda lime based on potassium
hydroxide. Otherwise, there is a risk of CO
formation.
WARNING
Risk of patient injury
All data transferred via the MEDIBUS
interface are for information only and must
not be used as the sole basis for diagnostic
or therapeutic decisions. The data accessible
via this interface are not intended for use with
a distributed alarm system in accordance
with IEC 60601-1-8:2012 (in the sense of
remote monitoring).
WARNING
For the protection of patients and users from
electrical risk, it is required that all systems
that consist of medical devices and other
electrical devices, such as computers,
printers, etc., are assembled exclusively by
trained personnel.
Overview
Rear view........................................................ 25
Pin-index connection ....................................... 25
Screw connections .......................................... 26
Interface panel ............................................... 27
Vaporizer ........................................................ 28
Interfaces ....................................................... 31
Recommended device configuration ............... 32
External fresh-gas outlet .............................. 34
Using the external fresh-gas outlet as a
common gas outlet .......................................... 35
Use of the external fresh-gas outlet with an
additional switch .............................................. 35
Abbreviations ................................................ 36
Symbols ......................................................... 38
Product labels................................................ 40
A B C D
E
Q
P
O
F
G
H
N
M
L I
* optional
A B C D E
O
M F
L
G
K J I
20983
E Flow-sensor guard or COSY guard (not O Holder for sample line (optional)
illustrated)
NOTE
F Expiratory port
The optional cover for the compact breathing
G Connection for breathing bag
system is not shown in the graphic.
H Inspiratory port
I Inspiratory valve
J Fresh gas decoupling valve
K Connection for APL bypass valve
* optional
21014
C
21015
A Power inlet
B LED indicator for power supply
C Cable clamp
Rear view
Pin-index connection
A B
C
2.8-6.0 kPa 2.8-6.0 kPa
x 100 (bar) x 100 (bar)
N2O
2.8-6.0 kPa
x 100 (bar)
N2O
2.8-6.0 kPa
x 100 (bar)
D F
AIR AIR
J 2.8-6.0 kPa
x 100 (bar)
2.8-6.0 kPa
x 100 (bar)
E
O2 O2
I
G
25843
A Power inlet J Connections for central supply hoses
B Interface panel
C Connection for ventilator hose
D On/Off switch
E Fuse
F Compact breathing system (COSY)
G Potential equalization pin
H Gas cylinder
I Pin-index connection*
* optional
Screw connections
N2O N2O
AIR AIR
O2 O2
25844
A Connections for central supply hoses
B Connections for compressed gas hoses for gas
cylinders with O2, O2 and N2O, or O2 and Air
(screw connections)*
C Gas cylinders
* optional
Interface panel
A
B
J
C
D
E
I G
20991
* optional
Vaporizer
26207
A
20987
APL valve
WARNING NOTE
Risk of patient injury The APL valve is automatically separated from
the breathing system as soon as an automatic
Wires and cables can get caught under the
ventilation mode is selected.
APL valve adjustment knob and block the
APL valve.
WARNING
Lay all cables and wires, e.g., sample line so
that they do not get caught. Risk of excessively high airway pressures
If the ventilator fails, the device switches into
The APL valve has 2 functions: the ManSpont ventilation mode.
– During manual ventilation, the maximum
The APL valve should also be set to a
airway pressure is limited.
pressure limitation value suitable for the
– During manual ventilation and spontaneous patient when using automatic ventilation
breathing, excess gas is discharged into the modes. If the ventilator fails, ventilate the
anesthetic gas scavenging system. patient manually.
The functional state is only guaranteed if the
ventilator is in ManSpont mode or is bypassed.
20956
Interfaces
CAUTION
Risk of electrical safety
To ensure electrical safety, only connect devices
to the serial ports (COM 1 and COM 2) with a
A maximum nominal voltage of 24 Vdc that meet
one of the following standards:
– IEC 60950-1: ungrounded SELV circuits
– IEC 60601-1 (as of 2nd edition): exposed
secondary circuits
CAUTION
Risk of device failure
Only use monitors, mounting parts, and
connection cables that have been approved by
Dräger.
NOTE
20948
A There are 2 ports on the rear of Fabius for The data transmission of the gas analysis data
communication with external devices. The must be activated by DrägerService or the
ports are marked with COM 1 and COM 2* . authorized local service partner.
These ports are used for data communication
using the Dräger MEDIBUS or Vitalink data
protocols.
WARNING
Risk of electric shock
Connecting devices to the MEDIBUS
interfaces can lead to an increased leakage
current. If the protective ground of one of
these devices fails, the leakage current may
rise above the permissible values.
– Only connect with the approval of the
respective device manufacturer.
– Have the leakage current checked by
service personnel.
– If the permissible value is exceeded,
disconnect the devices from the MEDIBUS
interface.
* optional
Configuration 1
C
C
D
A
A
21915
– Fabius
20958
Configuration 2
B
B
20958
– Fabius
– Multiparameter monitor with a COM connection
1 Connect monitor to COM 1 (B).
* optional
Configuration 4 Configuration 5
F G
F
E H H G
E
20962
20959
– Fabius
20963
– Fabius
– Multiparameter monitor with 2 COM interfaces
– Automatic anesthesia protocol system
1 Connect monitor (I) to the anesthesia protocol
system.
2 Connect the anesthesia protocol system to
COM 1 (J).
3 Connect monitor to COM 2 (K).
* optional
WARNING
Risk of misinterpretation of measured values
The values for O2, pressure, and volume
displayed on Fabius do not correspond to the
values for the patient connected to external
fresh-gas outlet as they are based on
measurements taken at the compact
breathing system.
When using the external fresh-gas outlet,
change into the Standby mode.
Using the external fresh-gas outlet as a Use of the external fresh-gas outlet with
common gas outlet an additional switch*
A B
F
C
E
D
C
21001
21002
C Short fresh-gas hose (on compact breathing
system) A External fresh-gas outlet with an additional
D Compact breathing system switch
E Sample line B Long fresh-gas hose (on Fabius)
F Non-rebreathing system (e.g., Bain) C Short fresh-gas hose (on compact breathing
system)
D Compact breathing system
E Sample line
F Non-rebreathing system (e.g., Bain)
* optional
Abbreviations
Abbreviation Explanation
S-ORC Sensitive Oxygen Ratio Control-
ler, maintains a minimum O2
concentration
STAPD Standard Temperature, Ambient
Pressure, Dry
20 °C (68 °F), ambient pres-
sure, dry gas
TI, TINSP Inspiratory time
TE Expiratory time
TI:TE Ratio of inspiratory time to expi-
ratory time
TIP:TI Ratio of inspiratory pause to
inspiratory time
Trigger Trigger
UMDNS Universal Medical Device
Nomenclature System, nomen-
clature for medical devices
USB Universal Serial Bus, computer
interface
VT Tidal volume
∆PPS Differential pressure of the pres-
sure support in Pressure Sup-
port mode
Symbols
Relative humidity
WEEE label, Directive
2002/96/EC
Atmospheric pressure
Consult instructions for use
Do not use if package damaged
Warning! Strictly follow these
instructions for use
Do not reuse
Caution! Observe the accompa-
nying documentation!(symbol)
Alarm inactive
Attention! (safety sign)
The alarm tone is suppressed for
Caution! Risk of electric shock. 2 minutes.
Do not remove cover.
Mains power
Applied part, protection class BF
(Body Floating)
Partial power supply switched on
Applied part of protection class B
Total power supply switched on
ESD warning label, observe the
warning statement, see ''Infor- Connection for potential equal-
mation on electromagnetic com- ization
patibility'' on page 11
Auto Exclusion Plug-in connec-
tion
Risk of crushing
Gas cylinder connection
Label on device surfaces where
the risk of tipping is increased by
e.g., leaning on or against the CO2 absorber bypass
surface or pushing
Read the flow at the center of the
Serial number float.
Battery charge
Ventilator connection
Fuse
Do not oil!
Close menu
Product labels
Operating concept
Control panel
C
B
C
E
C
21569
The control panel contains the following main Key Function
elements:
In the mode Standby:
A Screen
Opens a menu for configuration of the
B Key to select ventilation modes
system settings and default settings, see
– Volume Control chapter ''Configuration in standby mode''
– Pressure Control on page 128
– Pressure Support* In a ventilation mode:
– SIMV/PS *
Opens a menu for displaying and chang-
– ManSpont ing the monitoring settings, see chapter
C Key for more functions ''Configuration during operation'' on page
144
Key Function
Changes from displayed screen to the
Switches the workstation light and ventila- main screen.
tor illumination on and off.
The alarm tone of all active alarms is sup-
Opens the window with alarm limits. pressed for 2 minutes.
Switches to Standby mode. Monitoring is
switched off and the ventilator stops.
* optional
Softkeys
Screen display
D
H
G
21572
D Volume monitoring
Display of:
– Respiratory rate of the patient in breaths
per minute (Freq)
– Tidal volume (VT)
– Minute volume (MV)
– Upper and lower alarm limits of the minute
volume
E Airway pressure monitoring
Display of:
– Positive end-expiratory pressure (PEEP)
– Mean airway pressure (MEAN)
– Plateau pressure (PLAT)
– Peak pressure (PEAK)
F Pressure waveform for airway pressure
G Softkeys (labeling depending on ventilation
mode)
H Virtual flow tubes for O2, Air, N2O
A
B
21574
B
21577
Volume Control
C
D
21579
B
E
A H
C
F
I
D
G
J
L
26210
The total flow tube and the pressure gauge are I Pressure gauge for the central Air supply
located on the front of the device below the screen. J Pressure gauge for the central O2 supply
There are 2 flow control valves for Air and O2 on K Pressure gauges for gas cylinders (only for
the gas mixer for 2 gases. There is a third flow trolley-mounted devices)*
control valve for N2O on the gas mixer for 3 gases.
L LED warning indicator for low O2 pressure
All flow control valves are labeled and provided
with a color coding, see chapter ''Color coding for If the pressure drops below 20 psi (1.4 kPa
anesthetic agents and medical gases'' on page 51. x100), the LED warning indicator lights up.
The O2 flow control valve is additionally provided The displayed fresh-gas flow is in the range
with grooves on the gripping surface. between 0 L/min and 12 L/min.
A Total flow tube that displays the sum of the – If the value of the fresh-gas flow is higher than
individual flows of all used gases (O2, Air, 12 L/min, the electronic fresh-gas flow display
N2O). flashes (B, C, D).
B Electronic N2O fresh-gas flow display – The electronic fresh-gas flow display
C Electronic air fresh-gas flow display shows "+".
Total flow tube If a flow is higher than 9.99 L/min, the standard
resolution is activated.
NOTE Standby
The total flow tube is calibrated for a mixture of
N2O and O2 in a 50/50 ratio. The accuracy of the
total flow tube might be reduced with other gas
mixtures. More information can be found in A
chapter ''Technical data'' on page 186.
21582
The total flow tube remains functional if the
following faults occur: If the standard resolution is configured, the
– Fault in the electronic flow measurement electronic fresh-gas flow displays show the flow in
increments of 100 mL/min (format xx.x L/min). The
– Fault in the digital display virtual flow tubes (A) of the screen indicate a range
– Fault in the current switching circuit from 0 to 12 L/min.
In these cases, the measurement in the total flow
Standby
tube indicates the total flow before the fault
occurred.
To adapt the fresh-gas ratio during an exiting fault,
close all flow control valves and reset the B
individual flows one after another. The O2 valve
can remain open.
Example: After the flow control valves are closed,
the total flow tube indicates 2 L/min. This
corresponds to an O2 flow of 2 L/min.
If 1 L/min N2O is required, open the N2O flow
21587
LED indicators
There are several LED indicators on the front of LEDs for alarm status
the device.
Fabius has alarm LEDs that signal alarms and
indicate the alarm priority. For additional
LEDs for operating status information, see chapter ''Alarms''.
Alarm LEDs
A
C B
B A
21590
21590
supply, the mains power LED (A) is lit.
– In addition, the standby key (B) and all keys of – Red LED (A) flashing: signals a warning
the ventilation modes (C) have small LEDs that – Yellow LED (B) flashing: signals caution
light up when the respective mode is active. – Yellow LED (B) is lighted continuously: signals
a note
WARNING
Risk of device malfunction
If the battery is not sufficiently charged and
the mains power supply fails, operation
cannot be maintained long enough.
Before first operation or after storage, charge
the battery for at least 8 hours.
WARNING
Risk due to reduced power supply from the
internal battery
Batteries are wear parts. The capacity of the
battery diminishes with the period of use.
Check the functional state of the battery by
performing preventive maintenance on a
regular basis.
A
21017
Connecting the mains power supply 2 Plug the power cable into the mains power
socket on the wall.
The mains voltage must correspond to the voltage LED on the front of the device lights up
range indicated by the rating plate on the rear of green.
the device:
3 Set On/Off switch (A) to position .
100 V to 240 V
4 Check the status bar for the battery indicator.
A
20951
WARNING
Risk of electric shock and device malfunction
If the device is connected to a power socket
with incorrect mains voltage or without a
protective ground, the user can be injured
and the device damaged.
Only connect the power cable to power
sockets with a protective ground, see
''Technical data''.
NOTE
The mains plug must be freely accessible so that
the power supply to Fabius can be quickly
interrupted in the event of device failure.
N2O N2O
AIR AIR
O2 O2
26220
connected devices.
1 Connect the potential equalization cable to the
WARNING potential equalization pin (A) on the left side
(front view).
Risk of device malfunction
2 Connect the potential equalization cable to a
If high-frequency surgical devices are potential equalization connector of the hospital
connected to the auxiliary power sockets, the (e.g., wall, ceiling supply unit, operating table).
leakage current can damage the electronics
3 Establish potential equalization to the auxiliary
of the medical device and lead to a failure.
devices.
Do not connect any high-frequency surgical
devices to the auxiliary power sockets of the NOTE
medical device.
If the good condition of the protective ground or
its correct connection with the medical device
cannot be ensured, the device must be operated
via the internal power supply (battery).
WARNING
Central gas supply
N2O N2O
system.
26223
CAUTION
Gas cylinders with pin-index system
(optional)
Risk of insufficient gas pressure
If the supply pressure of the central gas supply is WARNING
too low, the functionality of the anesthesia
workstation is compromised. Risk due to incorrect mounting of the gas
cylinder
The pressure gauge of the individual gases must
display a constant pressure between 41 and When using several sealing washers between
87 psi (2.8 and 6 kPa x 100). the gas cylinder and the gas inlet of the
cylinder holder, the pin-index safety system
NOTE is compromised.
If a gas cylinder is connected, always check
A failure of the gas supply can lead to a failure of
whether the pin-index pins are present. Never
the connected devices.
attempt to bypass the pin-index safety
system.
WARNING
Risk of explosion
If the gas cylinder valves are opened too
quickly, a sudden increase in pressure may
occur.
– Open and close the gas cylinders valve
slowly by hand.
– Do not use tools.
CAUTION
Risk due to gas supply failure
Leave gas cylinders as a backup at the
anesthesia workstation, even if there is a
connection to the central gas supply.
NOTE
Leaky and stiff gas cylinder valves must be
repaired according to manufacturer's
specifications.
I J A
B
H
C 2.8-6.0 kPa
x 100 (bar)
N2O
2.8-6.0 kPa
x 100 (bar)
N2O
D
2.8-6.0 kPa 2.8-6.0 kPa
x 100 (bar) x 100 (bar)
AIR AIR
O2 O2
E A
G
F
20973
26220
2 Insert a new sealing washer (D) at the cylinder The pressure specifications are based on gas
holder (J). cylinders of size E at 21 °C/70 °F. If the pressure in
3 Make sure that both pin-index pins (A) are a gas cylinder does not reach the recommended
present below the gas inlet (B). minimum pressure (PSI - MIN), the gas cylinder
4 Align the gas cylinder (F) so that the pin-index must be replaced by a full gas cylinder.
holes on the cylinder head (E) are pointing
Gas PSI - FULL (kPa x PSI - MIN
towards the pin-index pins (A) on the cylinder
100 - FULL) (kPa x 100 - MIN)
holder (J).
5 Insert the cylinder head (E) of the gas cylinder (normal full load)
(F) from below into the cylinder holder (J). Air 1900 (131) 1000 (69)
6 Allow the pin-index pins (A) to engage in the N2O 745 (51) 600 (42)
pin-index holes. O2 1900 (131) 1000 (69)
7 Turn the handle (I) on the cylinder holder (J)
clockwise. The tip of the threaded retaining pin 1 Open cylinder valves (A).
will then be turned into the visible recess on the Make sure that the pressure gauges on the gas
cylinder head. Make sure that the gas cylinder cylinders indicate the appropriate pressure
is suspended vertically. recommended in the following table.
8 Tighten the handle (I) of the cylinder holder (J). No hissing must sound when opening the cylinder
If required, the gas cylinder valve (C) can be valves.
opened with the supplied wrench (G). In this case, the connection is leaking. The gas
If the gas cylinder is removed, insert the plug (H) in cylinder must be mounted again.
the mounted gas cylinder holder and tighten. 2 Close the cylinder valves again.
Maximum permissible dimensions of the gas Gas cylinders with screw connections
cylinders (optional)
Gas cylin- Left Right
der con- WARNING
nection
Risk of explosion
Length Length
Ø Ø When pressurized, O2 is self-igniting in
[mm] [mm]
combination with oil or grease.
Max. Max.
Pin-index 800 800 Do not oil or grease the gas cylinder valve or
105 105
the pressure reducer of the O2 cylinder. Do
Pressure
102/142 800 102/142 800 not touch with oily or greasy fingers.
reducers
Min. Ø of WARNING
gas cylin- 85/110 85/110
der Risk of explosion
If the gas cylinder valves are opened too
Observe the following points: quickly, a sudden increase in pressure may
– Do not exceed the maximum length of the gas occur.
cylinder including the pressure reducer / pin- – Open and close the gas cylinders valve
index connection. slowly by hand.
– The gas cylinders must not protrude over the – Do not use tools.
lower edge of the trolley (observe maximum
length of the gas cylinders). CAUTION
– The O2 cylinder must not be mounted on the Risk due to gas supply failure
right side (looking from the rear).
Leave gas cylinders as a backup at the
anesthesia workstation, even if there is a
connection to the central gas supply.
CAUTION
Risk of patient injury
Do not connect gas cylinders without the
pressure reducers listed in the current list of
accessories.
NOTE
Keep gas cylinders closed as long as they are
not being used. There is the risk of accidental
emptying of the gas cylinders.
2.8-6.0 kPa
x 100 (bar)
AIR
2.8-6.0 kPa
x 100 (bar)
AIR
adapter for the following connectors:
2.8-6.0 kPa 2.8-6.0 kPa
– Selectatec
x 100 (bar) x 100 (bar)
O2 O2
– Interlock S
– Dräger Auto Exclusion
A The anesthetic vaporizers must be mounted in
accordance with the respective instructions for
use.
26224
WARNING
Risk due to improperly mounted vaporizers
Incorrectly mounted vaporizers can cause
leakage. This can cause the fresh-gas flow to
be too low or contaminate the ambient air.
Patient and user can be endangered.
– Make sure that the vaporizers are
mounted levelly.
– When using D-Vapor vaporizers, make
sure that the power cable is not pinched.
– After mounting the vaporizers, perform a
leakage test.
A
B C
D
E
20981
20974
1 Fit the receiving system, with the slots resting The top edge of the float (D) in the flow tube must
on the corresponding pegs, on Fabius and move between the two marks (E).
slide down.
More information can be found in the instructions
2 Use the screw plug (A) to seal the connection for use (9038579) of the anesthetic gas receiving
not in use. system.
3 Slide the transfer hose (B) to the provided port.
4 Connect the other end of the transfer hose to
the exhaust port on the bottom of the COSY.
5 Connect the scavenging hose (C) to the
corresponding port of the receiving system.
6 Connect the probe of the scavenging hose (C)
to the terminal unit of the anesthetic gas
scavenging system. Observe the associated
instructions for use of the AGSS terminal unit.
C
B
A
20993
If the soda lime is used too long, carbon If conventional, non-pelletized soda lime is used,
dioxide can no longer be completely then a soda lime dust filter must be used.
absorbed.
Check the color of the soda lime regularly, Reusable CO2 absorber
especially if the inspiratory CO2 value
increases unexpectedly. Replace if CAUTION
necessary. Risk of chemical burns
WARNING
Do not use soda lime based on potassium
hydroxide. Otherwise, there is a risk of CO
21023
formation.
1 Turn the CO2 absorber (A) clockwise and
remove it from below.
2 Remove and dispose of the soda lime dust
filter* (B).
* optional
NOTE
Make sure that no soda lime residues are
between the seals and the seal surfaces. This
21022
D
F
21021
20977
* optional
CAUTION
Risk of patient recovering consciousness
If the CO2 absorber is not correctly locked into G
place, system leakage may occur.
Make sure that the absorber container is
connected tightly to the compact breathing
system during the ventilation. H
20992
disposable CO2 absorber may also be used.
CLIC adapter allows the following single-use CO2 1 Press the button (G): The CLIC adapter flips
absorbers to be used: open.
– CLIC Absorber 800+ 2 Loosen the soda lime in the disposable CO2
absorber, e.g., turn over the disposable CO2
– CLIC Absorber Free absorber a few times.
– Infinity ID CLIC Absorber 800+ 3 Remove the seal of the new disposable CO2
– Infinity ID CLIC Absorber Free absorber. Push the disposable CO2 absorber in
the Clic adapter.
More information on the connection of Drägersorb
CLIC-Adapters can be found in the associated 4 Push the disposable CO2 absorber (H)
instructions for use. upwards into the CLIC adapter, until the CLIC
adapter engages.
WARNING
Risk of insufficient ventilation
If the CO2 absorber is not correctly locked
into place, system leakage may occur. The
CO2 absorber must audibly engage before
Fabius is switched on. This ensures that the
CO2 absorber is included into the leakage test
and compliance test of the device.
After mounting and replacing, make sure the
CO2 absorber is firmly locked into place.
Mounting the COSY heating and the The heater must be switched off when it is not in
power supply unit (optional) use. This will prevent the CO2 absorbent from
drying out.
The compact breathing system can be operated
heated. This reduces the condensation of moisture WARNING
in the breathing system during low-flow operation Risk due to soda lime drying out
(fresh-gas flow less than 1 L/min).
The soda lime loses moisture. If the moisture
Before installation of the COSY heating, the falls below the minimum moisture, the
leakage test must be performed, see chapter following adverse reactions occur
''Leakage test'' on page 132. independent of the type of soda lime and
Follow the assembly instructions for breathing inhalational anesthetic used: Decreased CO2
system heating (9038262 and 8605899). absorption, increased generation of heat in
the CO2 absorber resulting in increased
breathing gas temperature, formation of CO,
Establishing the power supply of the breathing absorption and/or degradation of the
system heating inhalational anesthetics.
– Do not use unnecessarily high fresh-gas
flows.
– Only use the supplemental O2 delivery if
necessary.
– Do not leave the flow control valves open
unnecessarily long.
A B
26295
20942
E
20943
system in the provided holder (B).
5 Connect the ventilator hose to the
3 Release the locking pin again and turn the corresponding connector (D) on the anesthesia
compact breathing system until the locking pin workstation.
engages.
6 Connect the other end of the ventilator hose to
the ventilator connector (E) of the compact
breathing system.
If Fabius has a screw connection, the sealing rings
of this screw connection must be undamaged and
clean.
Only hand-tighten the screw connection. Do not
use tools.
C
26248
WARNING
Risk of fire
Residual vapors of easily flammable
disinfectants (e.g., alcohols) and deposits
that were not removed during reprocessing
can ignite when the flow sensor is in use.
– Ensure particle-free cleaning and
disinfection.
– After disinfection, allow the flow sensor to
air for at least 30 minutes.
– Before inserting the flow sensor, check for A
visible damage and soiling such as
residual mucus, medication aerosols, and
particles.
– Replace flow sensors when damaged,
soiled, or not particle-free.
20968
Screw on the exhaust port (A) in the compact
breathing system from below.
A B C
20967
* optional
20975
CAUTION
Risk of patient recovering consciousness 1 Position and align the connection adapter (A)
of the flexible arm on the connection of the
A blocked or incorrectly positioned breathing bag
compact breathing system.
can lead to lack of fresh gas for patients. Manual
ventilation is also not possible. 2 Tighten the knurled screws (B). Check that the
arm is fixed securely.
Make sure that the breathing bag is connected
tightly to the bag holder during the ventilation. 3 Fasten the elbow (D) at the end of the flexible
arm.
The breathing bag can be mounted to the compact 4 Fasten the breathing bag (C) on the other end
breathing system with the following variants: of the elbow.
– On a flexible arm
– On a rigid arm
– On the bag elbow directly on the compact
breathing system
A
A
B
C
21013
21012
The endotracheal suction system for Fabius Connect the Air connection hose of the suction
consists of: system via a 3-way adapter to the gas supply
block or the central gas supply.
– Suction regulator (A)
For vacuum suction:
– Suction bottles (D)
Connect the vacuum hose of the suction
1 The suction regulator (A) is attached to a system directly to the central gas supply.
holder (B). The holder is fastened on the side
GCX rail (C) on Fabius. Prepare the suction system in accordance with the
associated instructions for use.
2 The suction bottle (D) is attached to a separate
holder.
WARNING
Risk of patient injury
B
Only use the suction system if the ventilation
mode ManSpont is active or the patient is
A disconnected from the Y-piece.
D
20949
WARNING
Risk due to particles and dust
In order to protect the patient from particles
and dust, a filter must be used between the
inspiratory limb of the breathing system and
the patient.
Use a patient-side filter or a filter at the
inspiratory port.
20969
NOTE
CAUTION If it is not possible to use an expiratory filter (e.g.,
Risk to patients by damaged breathing hoses due to an intrinsic PEEP due to air trapping),
hygienically reprocess the device after use with
If the coil reinforcement of a breathing hose is this patient, see chapter Cleaning, disinfection
damaged, there is risk of kinking or occlusion. and sterilization.
When attaching or removing the breathing hoses,
always hold them at the connection sleeve and WARNING
not at the coil reinforcement. Otherwise the coil
Risk of infection
reinforcement can become separated from the
connection sleeve. Check the breathing hoses If no microbial filter is used, the breathing
for damage before each use. system may become contaminated with
disease-causing germs.
In this case, hygienically reprocess the
breathing system after each patient.
NOTE
Fabius contains no components made of natural
rubber latex.
For latex-free use, use a latex-free breathing bag
and breathing hoses.
NOTE
When applying tidal volumes in the range of the
maximum or minimum values indicated for each
patient category, use the smaller breathing bag
and the smaller breathing circuit.
B A
D
E
21456
LQVS
LQVS
H[S
H[S
Or
Lateral connections for the sam-
Filter at inspiratory port and expiratory port, connection for sample ple line support CO2 measure-
line at Y-piece: ment and help to flush the dead-
space between Y-piece and hose
adapter.
LQVS
H[S
21458
* The resistance of the breathing system and connected accessories must be observed.
WARNING CAUTION
Risk of underpressure in the lungs Risk due to misleading data
If filters are blocked, the sample gas flow can Replacing breathing hoses, filters, vaporizers, or
immediately cause underpressure in the soda lime can change the determined leakage
lungs. values and compliance values of the anesthesia
machine and thus affect therapy.
When ventilating pediatric patients and
– Perform a leakage test and compliance test
neonates, do not use HMEF or other filters at
after replacing breathing hoses, particularly
the Y-piece in connection with a hose adapter
flex hoses, vaporizers, or soda lime.
that have a patient-side connection for a
– Perform a leakage test and compliance test
sample line.
after changing the length of extendable
For measurement purposes, a permanent hoses.
sidestream flow flows through the sample line to
the patient-gas measurement module. With CAUTION
blocked HME filter or filter in this position on the Y- Risk due to changed hose lengths
piece, the measurement system will cause
underpressure situations in the patients lungs. Changed hose lengths can change resistance
and compliance. For neonates, this can cause
increased or reduced breathing volumes.
Observing the resistance and compliance
For neonates in particular, do not use flex hoses.
WARNING
Higher resistance values during spontaneous
Risk due to accessory components in breathing lead to an increased work of breathing in
breathing circuit the patient.
When using additional components or hose During volume-controlled ventilation, an increased
configurations which deviate from the resistance during inspiration has a slight effect on
standard breathing circuit, the inspiratory the applied volume. The peak pressure increases,
and expiratory resistance values may be however, with constant plateau pressure.
increased beyond standard requirements. Therefore, the time constant increases during the
When using such configurations, the user expiratory phase. When applying too short
must pay special attention to the measured expiratory times, this can lead to an incomplete
values. emptying of the lungs. This leads to a dynamic
overinflation of the lungs (airtrapping).
WARNING During pressure-controlled ventilation, an
increased resistance can decrease the inspiratory
Risk of increased rebreathing
and expiratory volumes.
Leakages between the inner and outer hose
Before performing the selftest, all accessories* to
of a coaxial breathing circuit cannot be
be used must be connected. Extendable hoses
detected during the leakage test.
must be pulled out to the length intended by the
To prevent insufficient gas exchange or CO2 user. Only in this way is the compliance correctly
rebreathing, pay strict attention to the determined and with volume-controlled ventilation
measured gas concentration. a correct tidal volume applied.
20971
The following formula is used to calculate the
resistance (R): 1 Remove the O2 sensor housing from the
inspiratory valve.
RInspiration = RBreathing system_insp + RInsp hose +
2 Unscrew the screw cap (A) from the O2 sensor
RBreathing bag hose + RInsp filter(port) + RInsp filter(Y-
housing.
piece)
3 Take the new O2 sensor capsule from the
RExpiration = RBreathing system_exsp + RExsp hose + package.
RExsp filter(port) + RExsp(Y-piece)
4 Place the O2 sensor capsule (B) in the housing
Make sure that for the calculation of the resistance so that the ring-shaped conductor touches the
only the accessories resistance values and the contacts in the housing.
peak flows are used for the respective accessory
category and patient category, e.g., resistance 5 Screw on the screw cap (A) tightly by hand.
values for adults at 60 L/min, for children at 6 Reinsert the O2 sensor housing in the
30 L/min, and for neonates at 5 L/min. inspiratory valve.
WARNING
Inserting a new O2 sensor capsule Danger of erroneous O2 measurement.
An incorrectly mounted O2 sensor will lead to
WARNING incorrect measurement results.
Risk of electric shock Make sure that the O2 sensor is inserted
If the O2 sensor is replaced during operation, correctly in the inspiratory valve, see page 22.
it can lead to transferring of leakage current.
Do not touch the patient.
B A C
A
20954
20955
1 Insert the O2 sensor (A) into the opening of the 1 Connect the pressure measurement hose to
inspiratory valve on the compact breathing the corresponding connector (A) on the bottom
system. of the compact breathing system.
2 Connect the plug (B) of the O2 sensor cable Make sure that the pressure measurement
with the connection marked with on the hose is not pinched.
rear of the device.
2 Connect the other end of the pressure
measurement hose to the bacterial filter (B).
3 Plug in the bacterial filter into the connection
(C) marked with on the rear of the device.
D C
B
A
B
20966
20972
2 Connect the T-piece (B) with the pressure
measurement hose and the pressure gauge.
1 Connect the flow sensor cable to the
3 Connect the other end of the pressure corresponding connection (A) on the bottom of
measurement hose to the bacterial filter (C). the compact breathing system.
4 Plug in the bacterial filter into the connection 2 Plug in the other end of the flow sensor cable
(D) marked with on the rear of the device. into the connection (B) marked with on the
rear of the device.
B C
A
20986
NOTE
The APL bypass hose is thicker than the
PEEP/PMAX hose.
A
26251
26254
* optional
CAUTION WARNING
Observe the assembly instructions of the Risk of tipping and risk of injury
accessory.
If the maximum permissible weight is
exceeded or if monitors and other auxiliary
devices are placed on the medical device, the
device can fall. Especially if the medical
device is rolled over door thresholds and
similar obstacles.
Before moving the device, remove the
monitors and other additional devices.
To increase the tipping stability: Position the optional flexible arm for breathing
Remove all monitors and other additional bag close to the device.
devices from the top shelf. Prepare additional components as described in the
Dismount any additional devices mounted to respective instructions for use.
the swivel arms or the top of the device (e.g.,
patient monitors, data management systems,
syringe pumps).
Clear the writing tray and slide it completely
into the device.
Getting started
Switching on .................................................. 86
WARNING
Risk of device malfunction
Some safety systems are only checked
during start-up.
– A selftest should be performed once daily.
– Switch the Fabius on and off or press the
softkey Run System Test.
Switching on
Prerequisite: The device is prepared (see chapter After switching on, the anesthesia workstation
''Cleaning, disinfection and sterilization'' on page starts as follows:
166) and assembled ready for operation (see – A system test is performed that checks various
chapter ''Assembly and preparation'' on page 52). components. The entire results of the tests, as
To prevent condensation and resulting failures of well as the test results for each component, are
electrical components, do not switch on the device displayed on the screen.
after abrupt temperature changes for 1 to 2 hours – Passed: Pass
(e.g., after storage in unheated rooms).
– Failed: Fail
WARNING – During the system test, 2 test tones are
sounded to test the speakers. To hear these
Risk of explosion and fire
tones, do not stand any further than 4 meters
Do not set the device into operation if oxygen (13 feet) from the device.
leakage is suspected in the medical device or – After completion of the system test, the default
its vicinity. settings for ventilation are loaded.
Stop oxygen supply and contact service
personnel. CAUTION
Malfunction of the device function
WARNING
The user must check whether the test tones were
Risk of accidental movement of the medical actually sounded. The device only checks
device whether the speakers are connected. A complete
During operation the medical device can loss of the ventilation function and monitoring
move accidentally. function would possibly not be noticed if the
speakers failed.
Activate the castor brakes.
If no tone or only one tone is sounded, the device
is only conditionally ready for operation. Contact
DrägerService.
A
21017
NON-FUNCTIONAL
26257
FUNCTIONAL
CONDITIONALLY FUNCTIONAL
Operation
Standby page after start-up.......................... 89 Preparing for storage or transport .............. 114
O2 flush .......................................................... 92
Ventilation ...................................................... 95
Ventilation mode ManSpont ............................ 95
Ventilation mode Volume Control.................... 98
Ventilation mode Pressure Control (optional).. 100
Ventilation mode Pressure Support (optional). 102
Ventilation mode SIMV/PS (optional) .............. 104
Adopting ventilation settings during mode
change............................................................. 106
Safety functions of the ventilator ................ 107
Behavior with too low fresh-gas supply ........... 107
Behavior of Fabius if the user does not take
any action ........................................................ 107
Patient change............................................... 108
21582
The fresh-gas settings can be changed before a S-ORC inhibits the N2O flow under the following
ventilation mode is selected. conditions:
Set the fresh-gas flow. – N2O flow control valve is open, even though
the O2 flow control valve is closed.
– O2 flow is below 0.2 L/min.
S-ORC (Sensitive Oxygen Ratio
Controller) With an N2O failure, O2 can continue to be
supplied. An alarm is not triggered. The float in the
N2O flow tube drops to zero.
Fabius is equipped with a mechanical minimum O2
delivery (S-ORC). This safety device prevents S-ORC has no oxygen-specific monitoring function
hypoxic gas mixtures if N2O is selected as the and offers no protection for the consequences of
carrier gas. an accidental interchange of gases.
Starting at a flow of approx. 200 mL/min, the N2O Therefore, the O2 concentration must always be
concentration in the fresh gas can be set to a value monitored.
between 0% and 75%.
When a lack of O2 is present, the S-ORC limits the
N2O concentration in the fresh gas so that the O2
concentration will not fall below 23 Vol%.
CAUTION
Risk of inaccurate measured values
When the O2 supply is again established after a
failure of the O2 supply, a supply pressure of at
least 2.7 kPa x 100 must be maintained for at
least 20 seconds. New failure of the O2 supply
will only be detected if the supply pressure is
stable for at least 20 seconds.
Do not use any functions during this time that
require O2, for example:
– O2 flush
– O2 fresh-gas flow
– Endotracheal suction
WARNING
Risk of material damage and risk to health
A vaporizer must never remain in operation
without a fresh-gas flow. High concentrations
of the anesthetic gas flow into the ventilation
circuit and the ambient air. This can lead to
material damage and contamination of the
ambient air with anesthetic gases.
Never cut off the fresh-gas flow before the
vaporizer is switched off.
WARNING
Risk of patient injury
When the vaporizer is in the control dial
position T, has heated up due to high ambient
temperature, and is then used, a high
anesthetic gas concentration might be
delivered.
After connecting the vaporizer to the
anesthesia workstation, always turn the
control dial of the vaporizer to position 0 and
wait at least 15 seconds to enable a pressure
equalization.
O2 flush
+ A
20960
NOTE
In ManSpont mode, the pressure can increase
quickly and thus trigger the APL valve.
Low-flow anesthesia
Do not perform low-flow anesthesia on Unsuitable soda lime can result in disintegration
patients with ketoacidosis or patients under products from the anesthetic gases.
the influence of alcohol. The risk of Use suitable soda lime such as Drägersorb Free.
accumulation of acetone in the patient
increases in such cases.
Ventilation
The user is responsible for setting the gas delivery In the following examples and illustrations, the
and ventilation according to the individual patient change from Volume Control mode to ManSpont
status. Patient status must be continually mode is described:
monitored for any potential changes.
Changing to spontaneous breathing
WARNING
Risk of strangulation
Negligent placement of hoses, cables, and
similar device components can endanger the
patient.
Use particular caution when establishing
connections to the patient.
20945
Ventilation mode ManSpont
1 Turn the APL valve head counterclockwise to
ManSpont (Manual/Spontaneous, manual its final position stop.
ventilation/spontaneous breathing) is a non- The label Spont and both points are vertical to
automatic ventilation mode. However, ventilation each other. The valve head lifts.
monitoring and alarm monitoring are still active.
The pressure limitation is disabled and the
The selection between manual ventilation and valve is open for free spontaneous breathing.
spontaneous breathing is made on the APL valve.
2 Set suitable fresh-gas flow.
If the APL valve is in the Spont position,
spontaneous breathing is enabled.
Volume Control
WARNING
Risk of excessively high airway pressures
If the ventilator fails, the device switches into
the ManSpont ventilation mode.
The APL valve should also be set to a
pressure limitation value suitable for the
patient when using automatic ventilation
modes since in case of a ventilator failure the A
patient must be ventilated manually.
21574
B B
21711
20946
– Apnoea Pressure (see chapter ''Pressure
alarms in automatic ventilation modes'' on
1 Set the APL valve head to the desired
page 125)
maximum airway pressure.
– Volume Alarms (see chapter ''Volume alarms''
Settings between the grid marks are also
on page 124)
possible.
Volume Control
A
21574
Pressure release
ManSpont
10 20 30
20947
B B 21711
In the ManSpont mode, lifting the valve head
relieves pressure from the breathing system.
The following alarms can be activated or
deactivated in ManSpont mode with the ON/OFF
key (B):
– Apnoea Pressure (see chapter ''Pressure
alarms in automatic ventilation modes'' on
page 125)
– Volume Alarms (see chapter ''Volume
alarms'' on page 124)
4 To refill the breathing bag, press the key.
5 Set suitable fresh-gas flow.
6 Start manual ventilation with the breathing bag.
The pressure is limited to the value that is set
on the APL valve.
21577
This guarantees an exact compliance
compensation. In the following table, all parameters (B) in Volume
Control mode, together with the corresponding
NOTE setting ranges and factory settings, are listed.
If the ventilator works at its performance limit due Parameter Setting range Factory set-
to the volume control settings, Fabius cannot ting
make the compliance compensation. If the
Pressure limita- 15 to 70, 40
performance limit of the ventilator is reached, it is
tion PMAX min. PEEP +10
not possible to increase the setting for tidal
[cmH2O]
volume VT.
([hPa])
Tidal volume 20 to 1400 600
Changing to Volume Control mode VT [mL]
In the following examples and illustrations, the Respiratory 4 to 60 12
change from Pressure Control mode to Volume rate Freq [bpm]
Control mode is described: ([1/min])
Pressure Control
Inspiratory 4:1 to 1:4 1:2
time:expiratory
A
timeTI:TE
Inspiratory 0 to 50 10
pause
time:inspira-
tory time TIP:TI
[%]
B
PEEP [cmH2O] 0 to 20 0
21723
([hPa])
1 Press the Volume Control (A) key.
2 Adjust ventilation settings (B).
3 Confirm the new mode.
PAW
PMAX
PEAK
PLAT
PAW ORZ
PEEP
7, 7(
7LPH
Freq
)ORZ
Insp Flow
VT
7,3 7LPH
7,
21741
Ventilation mode Pressure Control In the following table, all parameters (B) in
(optional) Pressure Control mode, together with the
corresponding setting ranges and factory settings,
are listed.
Changing to Pressure Control mode
Parameter Setting range Factory set-
In the following examples and illustrations, the
ting
change from Volume Control mode to Pressure
Control mode is described: Inspiratory 5 to 65, 15
pressure min. PEEP +5
Volume Control PINSP
[cmH2O]
([hPa])
A
Respiratory 4 to 60 12
rate Freq [bpm]
([1/min])
Inspiratory 4:1 to 1:4 1:2
B time:expiratory
timeTI:TE
21725
Inspiratory 10 to 75 30
1 Press the Pressure Control (A) key.
flowInsp Flow
2 Adjust ventilation settings (B). [L/min]
3 Confirm the new mode. PEEP [cmH2O] 0 to 20 0
Due to the influence of compliance and resistance, ([hPa])
the set value for Freq Min in Pressure Control
mode might not exactly be applied.
Pressure Control
21727
PAW
PINSP
PAW ORZ
7, 7(
7LPH
)ORZ Freq
7LPH
21742
21731
these conditions. In addition, the use of muscle
relaxants negatively influences the triggering by 1 Press the Pressure Support (A) key.
the patient.
2 Adjust ventilation settings (B).
In Pressure Support ventilation mode, the apnea
3 Confirm the new mode.
ventilation function is available, which can ensure
a minimum ventilation. To activate apnea
Pressure Support
ventilation, for the setting Freq Min another setting
must be selected than OFF. If the detected
spontaneous respiratory rate of the patient drops
under the set value for Freq Min, a mechanical
breath is applied. Apnea ventilation is not intended
to be a primary ventilation mode.
For apnea ventilation, Fabius uses the settings for
the following parameters:
– ∆PPS
– Freq Min
– Insp Flow
21729
– PEEP
If 2 successive mechanical breaths occur with
apnea ventilation, the alarm message APNOEA
VENTILATION !! is displayed in the alarm
message field. The alarm message is deleted as
soon as a spontaneous breath is detected.
In the following table, all parameters (B) in If the inspiratory flow (Insp Flow) during
Pressure Support mode, together with the inspiratory effort is greater than the set trigger flow
corresponding setting ranges and factory settings, (Trigger), the device supports the patient with the
are listed. setting ∆PPS.
Parameter Setting range Factory set- The set inspiratory flow (Insp Flow) defines how
ting fast the ∆PPS pressure is reached. When 25 % of
the maximum inspiratory flow (Insp Flow) is
Support pres- 3 to 20, 10 reached (or after maximum 4 seconds), the
sure ∆PPS OFF inspiration is automatically ended. The value Freq
[cmH2O] Min (e.g., 3 bpm (1/min)) defines a safety period
([hPa]) (safety period = 1/Freq Min, e.g., 20 seconds). If
Minimum respi- 3 to 20, 3 no inspiratory effort is detected and the safety
ratory rate for OFF period has elapsed, the device generates a
apnea ventila- pressure-controlled breath with PINSP=∆PPS.
tion Freq Min
The lower alarm limit of the airway pressure PAW
[bpm] ([1/min])
low is used for the airway pressure monitoring to
Trigger sensi- 2 to 15 2 detect apnea (disconnection) and continuous
tivity Trigger pressure.
[L/min]
If the pressure waveform does not cross the
Inspiratory flow 10 to 85 30 pressure threshold either from above or below, an
Insp Flow alarm is sounded.
[L/min]
PEEP [cmH2O] 0 to 20 0
([hPa])
PAW
Trigger Trigger $SQHD
PD[ 9HQWLODWLRQ
VHF
∆PPS PAW ORZ
Freq Min
)ORZ Freq Min
7LPH
21733
– TINSP
– TIP:TI 1 Press the SIMV/PS (A) key.
– PEEP 2 Adjust ventilation settings (B).
To support the inspiratory effort of the patient in To set the following additional parameters,
ventilation SIMV mode, pressure support ∆PPS press the MORE (C) key:
can be switched on. The setting of ∆PPS to – Trigger
another value than OFF activates the mode – Insp Flow
Pressure Support, see chapter ''Ventilation mode – TINSP
Pressure Support (optional)'' on page 102.
– TIP:TI
3 Confirm the new mode.
SIMV/PS
21735
21744
)ORZWULJJHU PD[VHF
port ∆PPS OFF
[cmH2O] ([hPa]) The respiratory rate Freq defines the time between
PEEP [cmH2O] 0 to 20 0 the individual volume-controlled breaths. The
([hPa]) synchronization of the mechanical breaths is done
Trigger sensitiv- 2 to 15 2 with a trigger sensitivity (Trigger) that is activated
ity Trigger a specific time before administering a new
[L/min] mechanical breath: 5 s for respiratory rates (Freq)
below 12 bpm (1/min). For higher respiratory rates,
Inspiratory flow- 10 to 85 30 the synchronization is done immediately after the
Insp Flow preceding expiration. Between these mandatory
[L/min] breaths, the patient can breathe spontaneously.
Inspiratory time 0.3 to 4.0 1.7 Mandatory breaths are synchronized with the
TINSP spontaneous breaths of the patient. These
[seconds] spontaneous breaths can be supported with
∆PPS.
Inspiratory 0 to 50 10
pause The lower alarm limit of the airway pressure PAW
time:inspiratory low is used for the airway pressure monitoring to
time TIP:TI [%] detect apnea (disconnection) and continuous
pressure.
If the pressure waveform does not cross the
pressure threshold either from above or below, an
alarm is sounded.
factory setting.
The ventilation settings for the new ventilation – ∆PPS is set to the last used value or the
mode are automatically derived from the settings factory setting.
of the previous ventilation mode. The – Trigger is set to the last used value or the
corresponding settings (A) in the new ventilation factory setting.
mode are highlighted.
When changing from Volume Control to SIMV/PS
The settings for Freq, TI:TE and PEEP are
adopted directly from the settings of the previous – PMAX and PEEP are automatically adopted for
ventilation mode if necessary. the new ventilation mode from the previous
mode.
When changing from Volume Control to
Pressure Control: When changing from Pressure Support to
SIMV/PS:
– PINSP is set to the plateau pressure (PLAT)
that occurred in Volume Control. – ∆PPS , Insp Flow , Trigger, and PEEP are
automatically adopted for the new ventilation
– Insp Flow is set to the last used value or the mode from the previous mode.
factory setting.
When changing from SIMV/PS with Pressure
When changing from Pressure Support to Support to Pressure Support:
Pressure Control:
– ∆PPS and Insp Flow are automatically
– Insp Flow is set to the last used value or the adopted for the new ventilation mode from the
factory setting. previous mode.
When changing from Pressure Control to When changing from SIMV/PS to Pressure
Volume Control: Support:
– VT is set to the value that results from the – Trigger and PEEP are automatically adopted
division of the last minute volume (MV) by the for the new ventilation mode from the previous
respiratory rate (Freq). mode.
– TIP:TI is set to the last used value or the
factory setting.
– PMAX is set to a value that lies 10 cmH2O
(hPa) above the plateau pressure (PLAT) that
occurred in Pressure Control.
CAUTION
20979
– Overpressure safety valve (A) If the gas supply fails completely, further
operation of the anesthesia machine takes place
– Underpressure safety valve (B)
with gas supply with ambient air. Anesthetic
– Pressure sensor in the ventilator chamber agents are no longer delivered and the
inspiratory anesthetic gas concentration in the
breathing gas decreases.
Behavior with too low fresh-gas supply Monitor the gas mixture carefully and use
intravenous anesthetic agents if need be.
For very low fresh-gas flow or an extremely large
leakage in the breathing system circuit, there can This allows for emergency ventilation with limited
be insufficient fresh gas. This is detected by the VT even with extremely low fresh-gas supply.
gradual emptying of the breathing bag. There will be no sudden switch-off of the ventilator.
NOTE
To remedy this, the user must take actions, e.g.,
increasing the fresh-gas flow.
Patient change
WARNING
Preparation
21000
values for the patient connected to external
fresh-gas outlet as they are based on 1 Remove fresh-gas hose from the external
measurements taken at the compact fresh-gas outlet (A).
breathing system.
When using the external fresh-gas outlet, C B
change into the Standby mode.
WARNING
Risk of faulty gas delivery A
O2 and CO2 and any anesthetic gases must
also be monitored for non-rebreathing
systems.
21001
Ending operation
A
21020
21000
4 If necessary, connect the anesthetic gas
scavenging hose of the non-rebreathing 1 Close all flow control valves on the device.
system to the second connection (D) of the
anesthetic gas receiving system. 2 Disconnect the non-rebreathing system from
the external fresh-gas outlet.
Observe the instructions for use of the non-
3 Connect the fresh-gas hose to the external
rebreathing system and the anesthetic gas
fresh-gas outlet (A).
receiving system.
4 Screw the sample line back to the Y-piece on
the breathing circuit.
Operation
WARNING
Preparation
21002
The values for O2, pressure, and volume
displayed on Fabius do not correspond to the
1 Connect the non-rebreathing system (B) to the
values for the patient connected to external
external fresh-gas outlet (A).
fresh-gas outlet as they are based on
measurements taken at the compact 2 Screw on the sample line (C) to the Luer Lock
breathing system. connector of the ventilation mask or of the
breathing system filter and to the water trap on
When using the external fresh-gas outlet,
the gas monitor.
change into the Standby mode.
WARNING
Risk of faulty gas delivery
O2 and CO2 and any anesthetic gases must D
also be monitored for non-rebreathing
systems.
The sample line must be connected with the
connector of the non-rebreathing system and
the connector of the gas analyzer (e.g., Scio,
Vamos).
21020
COSY
20998
21003
Diverting the fresh-gas flow to the non-rebreathing
system: Diverting the fresh-gas flow to the compact
1 Place the switch lever to . breathing system:
The lever points in the direction of the non- 1 Place the switch lever to COSY.
rebreathing system. The lever points in the direction of the fresh-
2 Set the fresh-gas flow. gas inlet.
Ending operation
WARNING
Risk of material damage and risk to health
A vaporizer must never remain in operation
A without a fresh-gas flow. High concentrations
of the anesthetic gas flow into the ventilation
circuit and the ambient air. This can lead to
material damage and contamination of the
21590
ambient air with anesthetic gases.
1 Press the button (A) and confirm. Never cut off the fresh-gas flow before the
Ventilation monitoring and alarm monitoring are vaporizer is switched off.
switched off. The ventilator stops.
3 Close the flow control valves.
Power-saving mode is activated after 2.5
minutes.
4 Close the cylinder valves.
B
NOTE
Leave Fabius connected to the mains power
supply so that the battery is not discharged
20965
21017
medical device does not bump against
anything. 6 Switch off Fabius with the On/Off switch (B) on
– Remove any devices mounted to the the rear and pull out the plug.
holding arms or the top of the device. 7 Remove hoses of the anesthetic gas receiving
– Clear the writing tray and fold it down system.
completely or slide it into the device.
– Do not pull the medical device over hoses,
cables, or other obstacles lying on the
floor.
– Do not activate the brake while the
medical device is being moved. C
– Always use the handles on the device to
push or pull it.
2.8-6.0 kPa
2.8-6.0 kPa
x 100 (bar)
x 100 (bar)
N2O N2O
2.8-6.0 kPa
x 100 (bar) 2.8-6.0 kPa
x 100 (bar)
AIR AIR
2.8-6.0 kPa
x 100 (bar) 2.8-6.0 kPa
x 100 (bar)
O2 O2
A
26223
Alarms
Alarm signaling
Alarm priorities
Priority of the existing alarm Priority of the new alarm Reaction from Fabius
(!!!) WARNING (!!!) WARNING – Alarm tone sequence starts
from the beginning.
(!!!) WARNING (!!) CAUTION – Alarm tone sequence for the
existing alarm is not inter-
rupted.
– No acoustic alarm signal for
the new alarm
(!!) CAUTION (!!) CAUTION – Alarm tone sequence starts
from the beginning.
(!!) CAUTION (!!!) WARNING – Alarm tone sequence for the
new alarm is started.
The alarm tone can be suppressed for a maximum Reactivating the suppressed alarm tone
of 2 minutes.
Press the (A) key again.
The LED in the key (A) goes out.
B
A
Adjusting the alarm limits
Monitoring
Main screen
O2 monitoring
NOTE
21780
The standard alarm limits configured for the The O2 sensor must be exposed to the ambient air
ventilation mode can be used unchanged, see during the entire calibration. The calibration of the
chapter ''Restoring the factory settings'' on page O2 sensor is done in the scope of the daily
139. checkout of the readiness for operation of Fabius.
or The O2 sensor can be calibrated in the following
ventilation modes:
The alarm limits can be set individually for the
current case: – In Standby mode, see chapter ''Calibrating the
O2 sensor'' on page 131.
– During the ventilation (in all available
A ventilation modes), see chapter ''Calibrating
the O2 sensor'' on page 146
Consequences of incorrect O2
calibration
21590
The diagram illustrates the correlation between the Deactivating the O2 monitoring
air mixture during calibration and the accuracy of
the oxygen measurement.
Volume Control
100
A
90
80
C
70
60
A 50
40
30 D
21784
20
The breathing volume is measured by the flow Parameter field for breathing volume
sensor based on thermal anemometry. The values
of the flow sensor are converted in the following
parameters and displayed: Volume Control
21780
NOTE The following information is displayed in the
parameter field for breathing volume:
Sudden, irregular expiratory flow can cause
A The respiratory rate (Freq) indicates the
erratic changes in the display of the tidal volume
breaths during the past minute in breaths per
and the respiratory rate. Before reading the
minute (bpm) (1/min).
display again, wait at least one minute.
The display is activated after 2 breaths.
The display range is between 2 bpm (1/min)
and 99 bpm (1/min).
B The tidal volume (VT) indicates the expiratory
volume for every breath in milliliters (mL).
The display range is between 0 mL and
1400 mL.
C The measured value for the minute volume
(MV) continually indicates the volume of the
gas breathed out in the past minute in liters per
minute (L/min).
The display range is between 0.0 L/min and
99.9 L/min.
D Upper alarm limit of the minute volume in L/min
E Lower alarm limit of the minute volume in L/min
Volume alarms
Volume Control
21782
period, after 30 seconds the alarm APNOEA
FLOW !! with the priority CAUTION is triggered. If The dialog window (B) with the alarm limits opens.
this alarm is not remedied, the priority increases 2 Adjust upper and lower alarm limit values of the
after a further 30 seconds to WARNING. minute volume (MV ), see setting ranges of the
The volume alarm messages are automatically alarm parameters in chapter ''Changing the
activated when changing from Standby mode in a alarm limits'' on page 138.
ventilation mode. 3 Confirm new values.
Setting the minute volume alarm limits Deactivating the volume alarms
The standard alarm limits configured for the
ventilation mode can be used unchanged, see
chapter ''Changing the alarm limits'' on page 138,
A
or
the alarm limits can be set individually for the
current case.
21590
A
The volume alarms can be switched on and off
during operation by pressing the key (A), see
chapter ''Switching the volume alarms on and off''
on page 145.
21590
Parameter field and waveform window C The peak pressure (PEAK) indicates the
for airway pressure highest pressure value of each breath in
cmH2O (hPa). The display range is between 0
and 80 cmH2O (0 and 80 hPa).
Volume Control
D Upper alarm limit
E Pressure threshold
F Pressure waveform
G Pressure threshold as a line
A B C D The pressure threshold is used for the
detection of apnea (disconnection) and
E continuous pressure. If the pressure waveform
H G F does not cross the pressure threshold either
from above or below, an alarm is sounded.
H Scale of the pressure waveform with display
range from 0 to 20, 0 to 50, or 0 to 100 cmH2O
21577
NOTE
Fabius can be configured by DrägerService or an
authorized local service partner so that the mean
airway pressure (MEAN) is displayed instead of
the plateau pressure (PLAT).
A
21590
Volume Control
C
21782
Configuration
The following configuration functions are available 2 Confirm the new mode.
in Standby mode: The ventilator changes to Standby mode. The
– Calibrations previous screen is replaced by the start screen
– System tests Standby and the standby LED is continually lit.
A C
21590
21582
ManSpont
The following softkeys (C) are displayed on the
start screen:
– RunSystemTest
– CalibrateFlowSensor
– CalibrateO2Sensor
– Leak /ComplTest
B
– AccessAlarmLog
– RestoreSiteDefaults
21971
NOTE
If the confirmation is not done within 15 seconds,
the ventilator remains in the previous mode and
the pressure waveform window is restored.
WARNING
21973
If the flow control valves are not closed before Risk of patient injury
accessing Standby mode, the following message During the system test, the system is
(D) is shown on the start screen: pressurized.
Gas still flowing - shut off all gas flow control To prevent patient injury, do not perform the
valves to activate Sleep Mode. system test on the medical device if a patient
As soon as the flow control valves are closed, the is connected.
message disappears.
3 Close the flow control valves. Standby
Power-saving mode
21582
1 Press the RunSystem Test softkey (A).
26258
The test results (B) are displayed on the screen. Troubleshooting with failed flow calibration
After completion of the system test, the total test
Repeat the calibration.
results (C) are displayed, see chapter ''Checking
the readiness for operation'' on page 87. If the Replace the flow sensor.
system test was successful, the Standby mode is If the calibration continues to fail, contact
activated. DrägerService or the authorized local service
partner.
Standby
A
21582
Standby
A
21582
<SLHFHRFFOXGHG
/XHU/RFNFRQQHFWRUFORVHG
B )UHVKJDVGHFRXSOLQJ
6HWWLQJ
30$;3((3
9HQWLODWRU
$3/E\SDVV
9DSRUL]HU
$3/YDOYH
&2DEVRUEHU
)ORZWXEHV
21984
2
Results of the system leakage test Troubleshooting with failed test of the
overpressure safety valve
The test of the system leakage can have the
following results:
WARNING
System leakage Displayed result Risk of unexpected occurring overpressure
[mL/min] [mL/min]
A soiled or non-functioning overpressure
250 Measured value and safety valve is not able to compensate for
PASSED suddenly occurring overpressure in the
251 to 350 Measured value and breathing system.
FAILED
Perform the leakage test before start-up of
>350 >350 and FAILED the device. Observe the test results of the
overpressure safety valve.
<SLHFHRFFOXGHG
/XHU/RFNFRQQHFWRUFORVHG
Repeat leakage test. If the test of the
overpressure safety valve continues to fail,
)UHVKJDVGHFRXSOLQJ contact DrägerService or the authorized local
service partner.
6HWWLQJ
30$;3((3
9HQWLODWRU
$3/E\SDVV
9DSRUL]HU
$3/YDOYH
&2DEVRUEHU
)ORZWXEHV
2
2IOXVK
%UHDWKLQJEDJ 6FDYHQJLQJ
21464
A
21582
Standby
D
C
B
21982
Standby
A
21582
WARNING
Risk due to unsuitable ventilation settings
After the default settings are restored, check
whether the settings for the ventilation and
monitoring are suitable for the patient.
Pressing the key in Standby mode provides Changing the default settings
access to various default settings and
configuration settings. 1 On the Standby Set-up screen, select Default
Access is password-protected. On request the Settings (A) and confirm.
password can be deactivated or a personal
password can be defined. Standby Set-up
A
The settings made are saved as default settings
and configurations.
A
22117
22103
1 Press the (A) key. The screen with the password query opens.
The Standby Set-up screen opens
Standby Set-up
Standby Set-up
B C D
B 22107
22103
The screen with the default settings opens. Default settings for Volume Control
C D
E
F
22105
22111
The following settings can be changed: 1 Select Volume Settings (E) and confirm.
– Volume Settings 2 Press softkey (F) of the parameter to be
– Pressure Settings * changed.
* optional
I J
H
G K
22113
22109
1 Select Alarm Limits (G) and confirm. 1 Select Minimum Alarm Volume (K) and
confirm.
2 Select the alarm limits (H) to be changed and
confirm. The current minimum alarm volume (L) is
displayed on the screen.
3 Select new value and confirm.
2 Set the new minimum alarm volume to a value
4 If necessary, repeat steps 2 and 3 for other
between 1 (minimum) and 10 (maximum) and
alarm limits.
confirm.
5 Select the input arrow (I) and confirm.
The window is closed, the cursor is on the input
The window is closed. arrow (J).
In the following table, the values for the setting
ranges and factory settings are listed for all alarm
limits of Fabius.
Standby Set-up
Pressure Support ∆PPS= 10
Freq Min= 3
Trigger= 2
Insp Flow= 30
PEEP= 0
SIMV/PS PMAX= 40
M
VT= 600
N Freq= 12
22115 ∆PPS= 10
PEEP= 0
1 Select Restore Factory Defaults (M) and Trigger= 2
confirm.
Insp Flow= 30
2 Select Yes or No (N) and confirm.
TINSP= 1.7
If Yes is selected, the factory settings are restored.
The factory settings replace the current default TIP:TI= 10
settings. Alarm limits for O2 Upper value = 100
The values for the factory settings for Fabius are Lower value = 20
listed in the following table. Alarm limits for MV Upper value = 12.0
Parameter Factory setting Lower value = 3.0
Volume Control PMAX= 40 Upper alarm limit and Upper value = 40
pressure threshold for
VT= 600 Lower value = 8
PEAK
Freq= 12
Minimum Alarm Vol- Volume = 5
TI:TE= 1:2.0 ume
TIP:TI= 10
PEEP= 0
Pressure Control PINSP= 15
Freq= 12
TI:TE= 1:2.0
Insp Flow= 30
PEEP= 0
Standby Set-up B
22127
1 Select Time Set (B) and confirm.
The cursor is in the hour field.
22124
C
E
22133
22129
1 Select Time Format (C) and confirm. 1 Select Date Format (E) and confirm.
2 Select new format and confirm. 2 Select new format and confirm.
The window is closed. The window is closed.
D
F
22135
22131
1 Select Date Set (D) and confirm. 1 Select Language (F) and confirm.
2 Select new value and confirm. 2 Select language and confirm.
The window is closed. The window is closed.
G
H
22137
22136
1 Select Pressure Unit (G) and confirm. When the function Acoustic Confirmation is
switched on, a tone is sounded when pressing the
The following units can be selected:
rotary knob.
– hPa
1 Select Acoustic Confirmation (H) and
– cmH2O
confirm.
– mbar
2 Select On or Off and confirm.
– kPa
The window is closed.
2 Select new unit and confirm.
The window is closed.
I K
22139
22143
1 Select Waveform Display (I) and confirm. 1 Select Display Background (K) and confirm.
2 Select Normal or Filled curve display and 2 Select Light or Dark screen brightness and
confirm. confirm.
The window is closed. The window is closed.
Standby Set-up
J
22141
If Fabius is in one of the ventilation modes, the The following softkeys (C) are displayed on the
following configuration functions can be screen, e.g., in Volume Control mode:
performed: – VolumeAlarms ON/OFF
– Calibration of the O2 sensor – Auto Set
– Displaying and changing the monitoring – CalibrateO2 Sensor
settings
– DesComp ON/OFF
– Changing configurations
– AccessAlarm Log
– AccessAlarm Volume
Volume Control
C
22155
Switching the volume alarms on and off Automatic setting of the pressure
threshold
Volume Control
Volume Control
A
A
22155
22155
1 Press the VolumeAlarms ON/OFF softkey (A).
1 Press the AutoSet softkey (A).
The key label changes from Volume Alarms
ON to Volume Alarms OFF. The pressure threshold for the peak pressure
(PEAK) is set to 4 cmH2O (hPa) below the
Instead of the upper and lower alarm limits, the
current plateau pressure (PLAT).
symbol appears, indicating the alarm is
deactivated.
NOTE
The volume alarms are deactivated.
The pressure threshold must not be below
NOTE 5 cmH2O (5 hPa) or above 30 cmH2O (30 hPa).
NOTE
In SIMV/PS mode, the pressure threshold
depends on the pressure of the mandatory
breaths.
A
22155
Volume Control
B
22157
CAUTION
B
Risk of inaccurate measured values
22159
22161
To switch the desflurane compensation back on: 2 To scroll through the alarm logbook (D), turn
Press the DesComp OFF softkey (B). the rotary knob.
The label of the softkey changes to DesComp ON.
Clearing the alarm logbook
Accessing the alarm logbook Select Clear Alarm Log (C) and confirm.
Volume Control
Closing the alarm logbook
A
22155
Volume Control
A
22155
Volume Control
22163
Troubleshooting
Leakages can lead to a failure of the system test or Possible causes of leakage
the leakage test and must be remedied.
– The CO2 absorber or the CLIC adapter is not
CAUTION securely screwed to the breathing system.
Risk due to contamination – The APL valve is not correctly fitted to the
breathing system or is not set to 30 hPa
Anesthetic gas can get into the ambient air as a
(cmH2O).
result of leakages.
– Perform the leakage test before using the – The breathing bag, the breathing hoses, the Y-
device. piece, or the microbial filter is incorrectly fitted
– Remedy all leakages. or damaged.
– The holder for the breathing bag is incorrectly
CAUTION mounted to the breathing system. The sealing
ring is soiled or damaged.
Risk due to leakage at the valves
– The water trap is not connected.
Leakages at valves can allow ambient air to enter
the breathing system and alter the composition of – The sample line is not connected, is kinked, or
the breathing gas. is leaking.
– Perform the leakage test before using the – The connections for the sample line are
device. damaged.
– Check all valves for leakage.
– The O-rings on the inspiratory port or
expiratory port are damaged, soiled, or
CAUTION missing.
Risk of insufficient ventilation – The flow sensor is incorrectly installed or
damaged. The rear O-ring is missing.
Breathing gas may escape because of leakages,
with the result that the applied volume is less – The valves or seals of the breathing system are
than the set volume. damaged.
– Perform the leakage test before using the – The cone for occluding the Y-piece is scratched
device. or damaged.
– Remedy all leakages.
– The filling or emptying connections on the
vaporizer are leaking or are open. The
vaporizer is incorrectly fitted. The O-ring is
missing or damaged. The control dial is not in
the 0 position.
Component Measure
Sample line Remove the sample line.
Occlude the Luer Lock con-
nector on the Y-piece.
Breathing hoses Disconnect the breathing
hoses. Connect the inspiratory
port and expiratory port with a
hose that is known to be with-
out leakages. Connect the
breathing bag directly to the
breathing system.
Vaporizers Remove the vaporizers.
Mains power supply failure The monitoring functions remain in operation until
the battery is completely discharged and all
If mains power fails, Fabius automatically switches electronic components are switched off.
to the internal battery. With a fully charged battery
the supply of the ventilator and the internal monitor CAUTION
functions are maintained for up to 2 hours. Risk of device malfunction
The remaining battery charge is displayed in the If mains power fails, devices connected to the
status bar. auxiliary power sockets are not supplied from the
The operating time of the battery depends on the internal battery.
ventilation settings and the condition of the battery Ensure an alternative power supply for
(age and battery charge). A completely charged connected devices.
battery can ensure supply for at least 45 minutes.
In battery operation and in case of decreasing WARNING
battery charge, the following information is
displayed: Insufficient ventilation of the patient
– The battery symbol is displayed in the If the alarm message BATTERY LOW !!
status bar and the LED indicator for mains (remaining battery charge 10%) is displayed
power supply goes out. for the first time, the ventilator still remains in
operation for up to 10 minutes.
– The note POWER FAIL ! is displayed in the
alarm window. Restore mains power supply. Afterwards, the
automatic ventilation is available again.
– When the remaining battery charge drops
below 20%, the note BATTERY LOW ! is
displayed in the alarm window. WARNING
WARNING
Incorrect patient settings
When the power supply is restored and
Fabius is restarted, all ventilation and alarm
settings are reset to default settings.
After the restart of Fabius, check all settings
and adjust to the patient if necessary.
Ventilator failure
Alarm VENTILATOR FAIL !!! Fabius restarts and performs a selftest. More
information on the selftest can be found in section
If the ventilator does not return to its initial state, ''Checking the readiness for operation'' on page
the alarm VENTILATOR FAIL !!! activates. 87.
Only manual ventilation or spontaneous breathing 1 Select ventilation mode ManSpont.
is possible. 2 Set the APL valve to position Man.
No other ventilation modes can be selected. 3 Set the APL valve to the desired pressure.
In this case, proceed as follows: 4 Fill the breathing bag, if necessary with the aid
of the O2 flush key.
1 Change to ManSpont ventilation mode.
5 Manually ventilate the patient.
2 Set the APL valve to position Man.
Before starting the ventilation with an automatic
3 Set the APL valve to the desired pressure.
ventilation mode, contact DrägerService or the
4 Fill the breathing bag, if necessary with the aid authorized local service partner.
of the O2 flush key.
5 Manually ventilate the patient.
Cause Remedy
During the calibration, the O2 sensor was exposed Make sure that the O2 sensor is exposed to ambi-
to an air mixture with extremely high or low oxygen ent air during the entire calibration.
concentration.
During the calibration, the O2 sensor was exposed Make sure that the O2 sensor is exposed to ambi-
to an air mixture with fluctuating oxygen concentra- ent air during the entire calibration.
tion.
The O2 sensor was not exposed to ambient air long Expose the O2 sensor to ambient air for 2 minutes.
enough before the calibration. When a new O2 sensor is connected, expose the
new sensor to ambient air for 15 minutes.
The maximum period of use of the O2 sensor has Replace the O2 sensor. Expose the new O2 sensor
elapsed. to ambient air 15 minutes before calibration.
The O2 sensor is not connected. Check the O2 sensor. Connect the O2 sensor cor-
rectly and recalibrate.
Disassembly
Attached devices
– Endotracheal suction
– Hinged arms
– Monitors
– Sensors and cables
– IT systems
– AGS
– workstation illumination
WARNING
Risk of infection
Used sample lines may be infectious due to
the breathing gases that passed through
them.
Replace the sample lines regularly, see table
''Semi-critical medical devices''.
Before removing the compact breathing system, Removing the expiratory valve
the following hoses and cables must be removed:
– Flow sensor cable 1 Screw off the cap nut (G).
– O2 sensor cable and O2 sensor capsule 2 Remove the sight glass (H).
– Pressure measurement hose 3 Take out the valve plate (I).
– APL bypass hose 4 Remove the sealing ring (J) from the socket
(K).
– PEEP/PMAX hose
A G
B
C H
D I
E J
F K A
20980
A
B
A B C C
D
20982
20979
1 Loosen the expiratory port (A) and remove. 1 Open the ventilator door.
2 Remove the flow-sensor guard (B). 2 Remove the pressure sensor line (B) of the
3 Remove the flow sensor (C). ventilator chamber from the corresponding
connection.
3 Release the 3 clamps (D).
Removing the APL valve 4 Remove the cover (A).
5 Remove the ventilator membrane (C).
Reprocessing procedures
3 Remove disinfectant residues after the contact Strictly observe the manufacturer's specifications
time has elapsed. on the disinfectants. Manufacturers may change
the composition of disinfectants over time.
Procedures:
1 Disinfect components by immersing.
2 After the contact time has elapsed, rinse the
components thoroughly under running water
until disinfectant residues are no longer
discernible.
3 Inspect components for visible soiling and
damage. Repeat manual disinfection if
necessary.
4 Shake off all excess water. Allow components
to dry thoroughly.
Procedures: CAUTION
1 Strictly observe the instructions for use of the Risk from faulty accessories
washer-disinfector.
Even reusable accessories have a limited
2 Position the parts in the basket in a stable
maximum period of use, e.g., residues from
position. Make sure that all interior spaces and
disinfectants can attack the material in the
surfaces are completely flushed and water can
autoclave. Signs of external wear can show up,
drain off freely.
e.g., cracks, deformation, discoloration, or
3 Use a suitable cleaning agent. delamination).
4 Select a suitable program, preferably If signs of external wear occur, replace the
anesthesia program. affected accessory.
– Cleaning must be performed at 40 °C to
60 °C (104 °F to 140 °F) for at least CAUTION
5 minutes.
Risk of failure of flow measurement
– Thermal disinfection must be performed at
80 °C to 95 °C (176 °F to 203 °F) and with Improper reprocessing and soiling, such as
corresponding contact time. deposits or particles, can damage the flow
5 Carry out final rinsing with demineralized water. sensor.
– No machine cleaning or disinfection
6 Immediately remove the components from – No plasma sterilization or radiation
washer-disinfector. sterilization
7 Inspect components for visible soiling and – No water jets, compressed air, brushes or the
damage. If necessary, repeat the program or like
perform manual cleaning or manual – No ultrasonic bath
disinfection. – No hot-steam sterilization with Spirolog and
Infinity ID flow sensors
8 Allow components to dry thoroughly.
– Clean and disinfect the flow sensor in
accordance with the corresponding
WARNING
instructions for use.
Risk of device failure – For disinfecting the flow sensor use only
clean disinfectant solutions.
If the control areas located in the valve plate
are not sufficiently dried, this may
compromise the device function or may lead
to failure of the medical device.
After cleaning, the breathing system must be
sterilized with steam until it is completely dry.
WARNING
Risk of fire
Residual vapors of easily flammable
disinfectants (e.g., alcohols) and deposits
that were not removed during reprocessing
can ignite when the flow sensor is in use.
– Ensure particle-free cleaning and
disinfection.
– After disinfection, allow the flow sensor to
air for at least 30 minutes.
– Before inserting the flow sensor, check for
visible damage and soiling such as
residual mucus, medication aerosols, and
particles.
– Replace flow sensors when damaged,
soiled, or not particle-free.
Sterilization
Reprocessing list
Maintenance
Overview......................................................... 180
Inspection....................................................... 181
Safety checks .................................................. 181
Preventive maintenance ............................... 182
Overview
Term Definition
Maintenance All measures (inspection, preventive maintenance, repair) intended to maintain
and restore the functional condition of a medical device
Inspection Measures intended to determine and assess the actual state of a medical device
Service Recurrent specified measures intended to maintain the functional condition of a
medical device
Repair Measures intended to restore the functional condition of a medical device after a
device malfunction
Inspection
Safety checks
The safety checks are no substitute for service – Check the function of the pressure reducer
measures indicated by the manufacturer, including (optional) of the compressed gas cylinder.
the preventive replacement of wearing parts. 3 Check if the product is in good condition:
WARNING – Labels are complete and legible
– No visible damage
Risk of malfunction of the medical device
– Fuses which are accessible from the
If the safety checks are not regularly outside are in compliance with the specified
performed, the medical device might not values
function properly. – Country-specific labeling of gas types
Perform the safety checks in the specified 4 Check components and accessories for
intervals. completeness according to the instructions for
use.
1 Check the accompanying documents:
5 Check for electrical safety in compliance with
– Latest instructions for use are available IEC 62353.
2 Check the following functions according to the 6 Check safety features:
instructions for use:
– Check the functional state of the optical and
– Check the proper function of the flow acoustic alarm generators.
measurement.
– Check the functional state of the O2 failure
– Check the function of the pressure alarm.
measurement based on the parameters
– Check the locking device of the anesthetic
PAW, PEEP, PMAX.
vaporizer.
– Check the proper function of the O2
– Check the function of the power failure
measurement.
alarm and the battery function.
– Check the function of the anesthetic
– Check S-ORC functionality.
vaporizer according to the associated
instructions for use.
– Check the function of the O2 flush.
Preventive maintenance
WARNING WARNING
Risk of faulty components Risk of electric shock
Device failure is possible due to wear or Before performing any service work,
material fatigue of the components. disconnect all electrical connections and gas
connections from power and gas supplies.
To maintain the function of all components,
this device must be inspected and serviced at
the intervals specified by the manufacturer.
Repair
Disposal
WARNING
For countries subject to the EU
Directive 2002/96/EC
Risk of infection
The device and its components must be This device is subject to EU Directive 2002/96/EC
disinfected and cleaned before disposal! (WEEE). In order to comply with its registration
according to this directive, this device may not be
At the end of its service life: disposed of at municipal collection points for waste
electrical and electronic equipment. Dräger has
Have the medical device appropriately
authorized a company to collect and dispose of
disposed of in accordance with applicable laws
this device. To initiate collection or for further
and regulations.
information, visit Dräger on the Internet at
www.draeger.com. Use the search function with
the keyword "WEEE" to find the relevant
information. If access to the Dräger website is not
possible, contact the local Dräger organization.
Disposal of accessories
WARNING
Risk of explosion and chemical burns
Improper handling of batteries can result in
explosions and chemical burns.
– Do not throw batteries in the fire.
– Do not force batteries open.
Technical data
Ventilator........................................................ 193
General information
Ambient conditions
During operation
Temperature 10 to 35 °C (50 to 95 °F)
Air pressure 700 to 1060 cmH2O (hPa)
Relative humidity 20 to 80 % (without condensation)
Height Up to 3000 m (9843 ft)
The conditions for use when using additional devices can limit the environment of use of a system as a
whole. Vaporizers and anesthetic agents can limit the use of an anesthesia workstation with regard to its
temperature range and maximum fresh-gas flow. Therefore when using additional devices, follow the
associated instructions for use.
Device data
Opening pressure of the internal safety valve 70 psi (4.8 kPa x 100)
Protection class
Device I, in compliance with IEC 60601-1
Application parts, connections for breathing hoses Type BF
Power supply
The power rating cannot be configured, with optional 100 to 240 VAC, 50/60 Hz, 2.3 A max
Dräger power socket strip (see instructions for use of
power socket strip for medical devices, 9038776)
Internal battery
Power rating 24 V; 3.5 Ah
Type Closed, lead/acid, gel
Charge time 16 hours on mains power for full operating time
Backup time with fully charged battery Minimum 45 minutes
Weight
Basic device with compact breathing system, no gas 134.2 kg (296 lb)
cylinders or vaporizers
Dimensions W x H x D
With trolley, compact breathing system, and mount 103.5 x 132 x 84 cm (40.7 x 52 x 33 in)
for 3 vaporizers1)
With trolley, compact breathing system, and mount 76 x 132 x 84 cm (30 x 52 x 33 in)
for 2 vaporizers1)
1) Width varies depending on the position of the breathing system, arm
Fuses
Electrical safety
Relevant standards
In addition to the standards listed here, this medical
device meets various other standards, e.g., stan-
dards concerning special national requirements.
IEC 60601-1 2nd ed. Part 1:
Medical electrical equipment General requirements for safety
The following also apply for devices manufactured from July 2014 on:
IEC 60601-1 3rd ed. Part 1:
Medical electrical equipment General requirements for basic safety and essential
performance
Ventilator
Control ranges
Pressure limitation (PMAX) 15 to 70 cmH2O (resolution: 1 cmH2O)
(15 to 70 hPa (resolution: 1 hPa))
(setting must be at least 10 cmH2O (10 hPa)
above PEEP, in SIMV/PS mode, the PMAX set-
ting must be greater than ∆PPS+PEEP)
Tidal volume (VT) 20 to 1400 mL (resolution: 10 mL)
Tidal volume (VT) 20 to 1100 mL (resolution: 10 mL), in SIMV/PS
mode
Respiratory rate (Freq) 4 to 60 bpm (resolution: 1 bpm)
(4 to 60 1/min (resolution: 1/min))
Ratio of inspiratory time to expiratory time (TI:TE) 4:1 to 1:4
Inspiratory pause (TIP:TI) 0 % to 50 % (resolution: 1%)
Positive end-expiratory pressure (PEEP) 0 to 20 cmH2O (resolution: 1 cmH2O)
(0 to 20 hPa (resolution: 1 hPa))
Inspiratory pressure (PINSP) 5 to 65 cmH2O (resolution: 1 cmH2O)
(5 to 65 hPa (resolution: 1 hPa)) (setting must be
at least 5 cmH2O (5 hPa) above PEEP)
Inspiratory flow (Insp Flow) 10 to 75 L/min (resolution: 1 L/min) in Pressure
Control mode
10 to 85 L/min (resolution: 1 L/min) in Pressure
Support and SIMV/PS modes
Support pressure (∆PPS) 3 to 20 cmH2O (resolution: 1 cmH2O)
(3 to 20 hPa (resolution: 1 hPa)), in Pressure
Support mode
Support pressure (∆PPS) 3 to 20 cmH2O, OFF (resolution: 1 cmH2O)
(3 to 20 hPa OFF (resolution: 1 hPa)), in
SIMV/PS mode
Minimum respiratory rate for apnea ventilation (Freq 3 to 20 bpm (resolution: 1 bpm) and OFF
Min) (3 to 20 1/min (resolution: 1/min) and OFF)
Trigger value (Trigger) 2 to 15 L/min (resolution: 1 L/min)
Inspiratory time (TINSP) 0.3 bis 4.0 sec.
Accuracy
Pressure limitation (PMAX) ±5 cmH2O (±5 hPa) of the setting
Tidal volume (VT) ±5 % of the setting or 20 mL, depending on which
value is higher (discharged to atmosphere, no
compliance correction)
Respiratory rate (Freq) ±1 bpm (±1 1/min) of the setting or ±5 %, depend-
ing on which value is higher
Ratio of inspiratory time to expiratory time (TI:TE) ±5 % of the setting
Inspiratory pause (TIP:TI) ±25 % of the setting
Positive end-expiratory pressure (PEEP) ±2 cmH2O (±2 hPa) or ±20 % of the setting,
depending on which value is higher
Inspiratory pressure (PINSP) ±2 cmH2O (±2 hPa) or ±20 % of the setting,
depending on which value is higher
Underpressure safety valve (inlet valve for ambi- -7.5 to -9 cmH2O (-7.5 to -9 hPa)
ent air)
Measuring the system compliance 0.2 to 6.0 mL/cmH2O (0.2 to 6.0 mL/hPa)
±0.2 mL/cmH2O (±0.2 mL/hPa) or ±10 % of the
actual compliance depending which value is
higher
Pressure limit of the common gas outlet Maximum 13 psi (0.9 kPa x 100) ±5 %
Vaporizer interface
O2 sensor
Response time (T90) Less than 16 seconds Measured values are not pressure
compensated.
Heat-up time after 5 minutes Error with ≤ 3% of the measured value
Drift sensitivity ±1 % of the measured value/ 8 h
Cross sensitivity 1 Vol% O2 at 70 Vol% N2O and 5 Vol%
CO2
With 4 Vol% halothane
or with 5 Vol% enflurane
or with 15 Vol% desflurane
or with 5 Vol% isoflurane
or with 10 Vol% sevoflurane
Measurement deviation due to Max. ±0.02% of the measured value per % relative humidity
humidity
Maximum period of use of O2 >12 months at 25 °C (77 °F), 50 % relative humidity, 50 % O2 in fresh
sensor cell gas (or >5000 hours at 100 Vol% O2)
Breathing system
Absorber volume
Reusable CO2 absorber, filled 1500 mL
Disposable CO2 absorber CLIC absorber Free 1200 mL
Disposable CO2 absorber CLIC Absorber 800 Free 1200 mL
Compliance
including ventilator hose (without breathing hoses)
0.8 mL/cmH2O (0.8 mL/hPa)
Resistance
In accordance with ISO 80601-2-13, dry, with adult Inspiratory: Expira- -4.7 cmH2O (-4.7 hPa)
breathing hose set M301461) tory: 4.4 cmH2O (4.4 hPa)
In accordance with ISO 80601-2-13, dry, without hos- Inspiratory: Expira- -3.7 cmH2O (-3.7 hPa)
es1) tory: 3.7 cmH2O (3.7 hPa)
Control ranges
APL valve
Manual ventilation mode 5 to 70 cmH2O (hPa)
Spontaneous breathing 1.5 cmH2O (hPa)
mode
Accuracy from 5 to 15 L/min ±15 % of the set value or ±3 cmH2O (hPa) (the
higher value applies)
Pressure drop at 30 L/min 3.4 cmH2O (hPa)
(wet and dry)
1) Depending on the current ventilation settings, the indicated values may deviate by ±0.3 cmH2O (0.3 hPa)
,623UHVVXUHIORZFKDUDFWHULVWLFVRIWKHFRPSDFWEUHDWKLQJV\VWHP&26<
DYHUDJHUHVLVWDQFHVZLWKRXWEUHDWKLQJKRVHV
3UHVVXUH>K3D@
)ORZ>/PLQ@
,QVSLUDWRU\SUHVVXUH ([SLUDWRU\SUHVVXUH
22229
Alarm in the event of failure of power supply and Continuous tone (approx. 30 s) at approx. 57 dB(A)
battery supply
Confirmation of selection using rotary knob Single tone when rotary knob is pressed
(approx. 51 dB(A) at max. alarm tone volume)
Time exceeded when changing ventilation mode 3 tones adjustable from 46 dB(A) to approx.
56 dB(A)
At a flow of approx. 200 mL Set the N2O concentration in the fresh gas
between 0 and 75 %.
N2O flow control valve is open and at the same S-ORC prevents N2O flow.
time the O2 flow control valve is closed or set at
less than 0.2 L/min.
Device outlets
Pin assignment
Pin 1 n/c
Pin 2 TXD
Pin 3 RXD
Pin 4 n/c
Pin 5 GND
Pin 6 n/c
Pin 7 n/c
Pin 8 n/c
Pin 9 n/c
NOTE
In accordance with general safety standards,
additional components are required for a
complete anesthesia workstation.
EMC declaration
Electromagnetic emissions
Electromagnetic immunity
Max. PEIRP 150 kHz to 2.5 GHz All other frequen- Examples
(watts) cies
0.03 0.32 m (1.1 ft) 0.96 m (3.2 ft) WLAN 5250 / 5775 (Europe)
0.10 0.58 m (1.9 ft) 1.8 m (5.9 ft) WLAN 2440 (Europe)
0.17 0.76 m (2.5 ft) 2.3 m (7.6 ft) Bluetooth, RFID 2.5 GHz
0.20 0.82 m (2.7 ft) 2.5 m (8.2 ft) WLAN 5250 (not in Europe)
0.25 0.92 m (3.0 ft) 2.8 m (9.2 ft) UMTS mobiles
0.41 1.2 m (3.9 ft) 3.5 m (12 ft) Cordless DECT devices
0.82 1.7 m (5.6 ft) 5.0 m (16 ft) RFID 13.56 MHz
1.00 1.8 m (5.9 ft) 5.5 m (18 ft) WLAN 5600 (not in Europe)
1.64 2.4 m (7.9 ft) 7.1 m (23 ft) GSM 1800 / GSM 1900
3.3 3.3 m (11 ft) 10 m (33 ft) GSM 900 mobile phones, RFID 868 MHz
Device combinations
Connections to IT networks
Prerequisites
This device must only be connected to the network
by service personnel. The IT representative of the
hospital must be consulted in advance.
The following documents must be followed:
– Accompanying documents of this device
– Descriptions of the network interface
– Description of the network-based alarm
systems
Dräger recommends complying with IEC 80001-1
(risk management for IT networks with medical
devices).
Serial interfaces
The following interfaces are supported:
– RS232 interfaces complying with EIA RS--232
(CCITT V.24/V.28) for the following
applications:
– MEDIBUS, MEDIBUS.X
– Connections to medical devices from other
manufacturers
Illustrations
9HQWLODWRU 6HWWLQJ
30$;3((3
)UHVKJDVGHFRXSOLQJ
)UHVKJDV
$3/E\SDVV $3/YDOYH
&2DEVRUEHU
%UHDWKLQJEDJ %UHDWKLQJJDVVFDYHQJLQJ
22234
Annex
CAUTION
If one of the checks does not pass, the device
must not be used.
Contact DrägerService or the responsible service
partner.
NOTE
The following applies to this section: cmH2O =
mbar = hPa.
Checklist
Anesthetic vaporizers with Selectatec connec- Check all plug connections and screw con-
tion nections for a tight seat.
P Vaporizer is securely connected, locked Replace missing or damaged seals. If nec-
and is hanging vertically. essary, contact DrägerService or the
P Control dial is in position 0. authorized local service partner
Reconnect the pressure measurement P The displayed measured value for the
hose to the socket for the airway pressure peak pressure (PEAK) is between 24 to
sensor on the rear of the device 36 cmH2O (hPa).
The check must be performed once without P Connect the breathing bag to the Y-piece.
vaporizer and once with vaporizer. Control dial is P Press the Pressure Control key and con-
in the zero position. firm.
D-Vapor must be switched on for the leakage test. P The measured values of the ventilation
After the leakage test has ended, switch off the D- parameters are displayed.
Vapor again. P The ventilator piston functions.
Change to Standby mode and press the P The valve plates in the inspiratory valve
Leak / Compl Test softkey. Follow the and expiratory valve move.
instructions on the screen.
P The breathing bag fills and empties.
If the system has leakage (the pressure drops):
P Press the standby key and confirm. P Change to the Standby screen.
Monitoring functions and alarms Set the APL valve for spontaneous breathing:
The alarm function can be checked by setting an Turn the APL valve head counterclockwise
alarm limit that causes an alarm message for cer- until the Spont mark is reached.
tain. The alarm limits can be adjusted at the start Press the O2 flush key and hold it pressed.
and during a check.
The airway pressure with an occluded Y-
Check the settings of the alarm limits. piece is less than 10 cmH2O (hPa).
Simulate alarm conditions and check Release the O2 flush key.
whether the correct alarm signals are trig-
gered. The airway pressure is higher than or
equals 0 cmH2O (hPa).
Check O2 display and alarm.
Check volume display and alarm.
Manual resuscitator
Check pressure display and alarm.
When squeezing the bag an audible and
Press the standby key and confirm. noticeable air flow must escape from the
When Fabius is restarted, the default settings for mask connection (cone). After releasing,
the alarm limits are automatically restored. the bag must quickly assume its original
Check the default settings and adjust if shape.
necessary. Close off the mask connection (cone) with
the ball of the hand: The bag must only be
able to be slightly squeezed.
Other monitors (optional)
Make sure that external monitors (if present) are
connected correctly and have been checked in COSY heating (optional)
accordance with the associated instructions for Check cable connections.
use. Check power cable connections.
Test the alarm functions on all monitors. LED indicator for power supply lights up.
CO2 monitor and alarm module are func- Toggle switch is set to ON.
tional.
LED indicator for COSY heating lights up.
Anesthetic agent monitor and alarm mod-
ule are functional. 30 minutes after the LED indicator for the
COSY heating lights up, the temperature
of the bottom of COSY is approx. 35-
Anesthetic gas scavenging system 40 °C.
P Hoses are correctly connected.
P Set the flow control valve on the anes-
thetic gas receiving system so that the
float is located between the "Min." and
"Max." marks.
P Occlude the Y-piece. Close all flow control
valves.
If any one of the test points is not passed, the Signature for pre-use checkout
device must not be used. Contact DrägerService
or the authorized local service partner.
Name
Signature for the daily checkout
Date
Name
Date
Signature for pre-use checkout
Name
Signature for pre-use checkout
Date
Name
Date
Signature for pre-use checkout
Name
Signature for pre-use checkout
Date
Name
Date
Signature for pre-use checkout
Name
Signature for pre-use checkout
Date
Name
Date
Signature for pre-use checkout
Name
Signature for pre-use checkout
Date
Name
Date
Signature for pre-use checkout
Name
Date
Name Name
Date Date
Name Name
Date Date
Name Name
Date Date
Name Name
Date Date
Name Name
Date Date
Name Name
Date Date
Password
8088
Index
A C
Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . 36 Central gas supply. . . . . . . . . . . . . . . . . . . . . . . 56
Adopting ventilation settings during mode change Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 212
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106 CO2 absorber
Airway pressure monitoring . . . . . . . . . . . . . . . 125 Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Alarm logbook . . . . . . . . . . . . . . . . . . . . . . . . . 134 Reusable . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Accessing . . . . . . . . . . . . . . . . . . . . . . . . . . 148 CO2- absorber
Clearing . . . . . . . . . . . . . . . . . . . . . . . . . . . 148 Disposable with CLIC adapter . . . . . . . . . . . 65
Closing . . . . . . . . . . . . . . . . . . . . . . . . . . . . 148 Compact breathing system
Alarm messages. . . . . . . . . . . . . . . . . . . . . . . . 157 Connecting . . . . . . . . . . . . . . . . . . . . . . . . . 67
Alarm priorities . . . . . . . . . . . . . . . . . . . . . . . . . 117 Removing . . . . . . . . . . . . . . . . . . . . . . . . . 169
Alarm signaling. . . . . . . . . . . . . . . . . . . . . . . . . 116 Compact breathing system COSY . . . . . . . . . . 22
Alarm tone Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Suppressing . . . . . . . . . . . . . . . . . . . . . . . . 118 Configuration
Alarm volume Automatically setting the pressure threshold
Changing . . . . . . . . . . . . . . . . . . . . . . . . . . 149 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145
Ambient conditions . . . . . . . . . . . . . . . . . . . . . . 187 During operation . . . . . . . . . . . . . . . . . . . . 144
Anesthetic gas receiving system In standby mode . . . . . . . . . . . . . . . . . . . . 128
Connecting . . . . . . . . . . . . . . . . . . . . . . . . . . 61 Leakage test . . . . . . . . . . . . . . . . . . . . . . . 132
Anesthetic gas supply module . . . . . . . . . . . . . 195 Restoring default settings . . . . . . . . . . . . . 135
APL bypass hose System test . . . . . . . . . . . . . . . . . . . . . . . . 129
Connecting . . . . . . . . . . . . . . . . . . . . . . . . . . 80 Volume alarms . . . . . . . . . . . . . . . . . . . . . . 145
APL valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30 Configurations
Removing . . . . . . . . . . . . . . . . . . . . . . . . . . 170 Acoustic confirmation . . . . . . . . . . . . . . . . 142
Auxiliary power sockets . . . . . . . . . . . . . . . . . . . 55 Changing . . . . . . . . . . . . . . . . . . . . . . . . . . 140
Date format . . . . . . . . . . . . . . . . . . . . . . . . 141
B Date setting . . . . . . . . . . . . . . . . . . . . . . . . 141
Pressure unit . . . . . . . . . . . . . . . . . . . . . . . 142
Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 153 Screen brightness . . . . . . . . . . . . . . . . . . . 143
Activating . . . . . . . . . . . . . . . . . . . . . . . . . . . 53 Selecting language . . . . . . . . . . . . . . . . . . 141
Breathing bag Setting the time . . . . . . . . . . . . . . . . . . . . . 140
Connecting . . . . . . . . . . . . . . . . . . . . . . . . . . 69 Time format . . . . . . . . . . . . . . . . . . . . . . . . 141
Breathing bag holder Waveform display . . . . . . . . . . . . . . . . . . . 143
Bag elbow . . . . . . . . . . . . . . . . . . . . . . . . . . . 70 Contraindications. . . . . . . . . . . . . . . . . . . . . . . . 17
Flexible arm . . . . . . . . . . . . . . . . . . . . . . . . . 69 Control panel . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Rigid arm . . . . . . . . . . . . . . . . . . . . . . . . . . . 70 COSY cover . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
Breathing hose configurations . . . . . . . . . . . . . . 75 COSY heating
Breathing hoses and filters. . . . . . . . . . . . . . . . . 72 Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Breathing system . . . . . . . . . . . . . . . . . . . . . . . 198
Assembling . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Breathing volume monitoring . . . . . . . . . . . . . . 123
D I
Default settings Indications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Changing . . . . . . . . . . . . . . . . . . . . . . . . . . 136 Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Changing the alarm limits . . . . . . . . . . . . . . 138 Interfaces. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Changing the minimum alarm volume . . . . 138 IT networks . . . . . . . . . . . . . . . . . . . . . . . . . . . 209
Pressure Control. . . . . . . . . . . . . . . . . . . . . 137
Pressure Support . . . . . . . . . . . . . . . . . . . . 137 L
Restoring the factory settings . . . . . . . . . . . 139
SIMV/PS . . . . . . . . . . . . . . . . . . . . . . . . . . . 137 Leakages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 151
Volume control . . . . . . . . . . . . . . . . . . . . . . 137 LED indicators . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Desflurane compensation Low-flow anesthesia . . . . . . . . . . . . . . . . . . . . . 93
Automatic . . . . . . . . . . . . . . . . . . . . . . . . . . 147
Switching on and off . . . . . . . . . . . . . . . . . . 147 M
Device combinations . . . . . . . . . . . . . . . . . . . . 208
Device configurations . . . . . . . . . . . . . . . . . . . . . 32 Mains power supply
Device data. . . . . . . . . . . . . . . . . . . . . . . . . . . . 188 Connecting. . . . . . . . . . . . . . . . . . . . . . . . . . 54
Device outlets . . . . . . . . . . . . . . . . . . . . . . . . . . 202 Maintenance intervals . . . . . . . . . . . . . . . . . . . 182
Manual resuscitator . . . . . . . . . . . . . . . . . . . . . . 81
Manual ventilation . . . . . . . . . . . . . . . . . . . . . . . 96
E MEDIBUS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Electrical safety . . . . . . . . . . . . . . . . . . . . . . . . 190 Medical devices
Electromagnetic compatibility . . . . . . . . . . . . . . . 11 Classification . . . . . . . . . . . . . . . . . . . . . . . 171
EMC declaration . . . . . . . . . . . . . . . . . . . . . . . . 204 Minute volume . . . . . . . . . . . . . . . . . . . . . . . . . 124
Exhaust port
Connecting . . . . . . . . . . . . . . . . . . . . . . . . . . 68 N
External fresh-gas outlet. . . . . . . . . . . . . . . 34, 190
Additional switch . . . . . . . . . . . . . . . . . . . . . . 35 Non-critical medical devices
Auxiliary switch . . . . . . . . . . . . . . . . . . . . . . 111 Reprocessing. . . . . . . . . . . . . . . . . . . . . . . 172
Common gas outlet. . . . . . . . . . . . . . . . 35, 109
O
F O2 flush . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
Flow sensor O2 insufflation . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Calibrating. . . . . . . . . . . . . . . . . . . . . . . . . . 130 O2 monitoring . . . . . . . . . . . . . . . . . . . . . 120, 121
Connecting . . . . . . . . . . . . . . . . . . . . . . . . . . 79 O2 sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . 156
Inserting . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68 Calibrating . . . . . . . . . . . . . . . . . 121, 131, 146
Removing . . . . . . . . . . . . . . . . . . . . . . . . . . 170 Connecting. . . . . . . . . . . . . . . . . . . . . . . . . . 78
Fresh-gas delivery . . . . . . . . . . . . . . . . . . . . . . . 48 Replacing the sensor capsule . . . . . . . . . . . 77
Fresh-gas flow Operation
Monitoring resolutions. . . . . . . . . . . . . . . . . . 49 Ending . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113
Setting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 190 P
Password. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 220
G Patient change . . . . . . . . . . . . . . . . . . . . . . . . 108
Gas cylinders (screw connections). . . . . . . . . . . 59 PEEP/PMAX hose
Gas flow plan . . . . . . . . . . . . . . . . . . . . . . . . . . 210 Connecting. . . . . . . . . . . . . . . . . . . . . . . . . . 80
Gas supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56 Pin-index system . . . . . . . . . . . . . . . . . . . . . . . . 57
S
Safety checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Screen colors . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Screen display . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Semi-critical medical devices
Reprocessing . . . . . . . . . . . . . . . . . . . . . . . 172
Sensors and measurement lines . . . . . . . . . . . . 78
Setting the anesthetic gas concentration . . . . . . 91
Soda lime
Replacing . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
Softkeys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Spontaneous breathing . . . . . . . . . . . . . . . . . . . 95
Sterilization . . . . . . . . . . . . . . . . . . . . . . . . . . . . 174
Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
Suction system
Connecting . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . 171
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
S-ORC . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89, 202
T
Total flow tube . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
V
Vaporizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Vaporizer interface . . . . . . . . . . . . . . . . . . . . . . 196
Ventilation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
Manufacturer
As of 2015-08:
Dräger Medical GmbH
changes to
Drägerwerk AG & Co. KGaA
9054626 – GA 5330.650 en
Á9054626ZÈ
© Dräger Medical GmbH
Edition: 1 – 2014-06
Dräger reserves the right to make modifications
to the device without prior notice.
ision]_0000027779_No.2904