Microlab 300 Applications
Microlab 300 Applications
MICROLAB 300
AA-MICROLAB300-CAT-10
A Versions M
ALBUMIN .............................................. ALBU-4 ................ PRO NEW MAGNESIUM XYLIDYL ......... MAGX-1................ PRO
Update ALP (DEA) SL ....................................... PASL-4 ................ PRO MICROPROTEIN PLUS ........ PRTU-3 ................ PRO
Update ALT / GPT 4+1 SL ................................ ALSL4+1-4........... PRO
AMYLASE SL ....................................... AMSL-6................ PRO
Update AST / GOT 4+1 SL................................ ASSL4+1-5 .......... PRO
P
B PHOSPHORUS ..................... PHOS-4 ................ PRO
REF: ALBU-XXXX
INSTRUMENT SETTINGS
GENERAL
NAME ALBU
PREPARATION OF WORKING REAGENT:
UNITS g/dL
The reagent is ready for use. R-BLANK YES
S-BLANK NO
REPLICATION 1
SETTINGS
PROCEDURE: MODE END-POINT
WAVELENGTH 620
Pipette 1000 µL of reagent and 10 µL of sample VOLUME 500
into a test tube. T-FACTOR 1.000
Mix and after 5 minutes incubation at 37°C, DELAY 3
aspirate the mixing by the analyser and start BI-CHROM NO
measurement. WAVELENGTH 2 -
LIMITS
Note: Before use, the reagent should be brought L-ABS **
to the working temperature (37°C). H-ABS **
REF-LOW **
REF-HIGH **
DECIMALS 2
L-ABS-RB 0.000*
H-ABS-RB 0.300*
CALIBRATION
METHOD 1-POINT
NAME **
FACTOR 1
REPLICATION 3
DEVIATION **
L-ABS **
H-ABS **
STANDARDS
CON 1 #
(03/2019)
AA-MICROLAB300-ALBU-4
Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
ALP (DEA) SL
ELITech Clinical Systems SAS
REF: PASL-XXXX
INSTRUMENT SETTINGS
GENERAL
PREPARATION OF WORKING REAGENT (WR): NAME PASL
UNITS U/L
Mix 4 volumes of reagent 1 with 1 volume of R-BLANK NO
reagent 2.
S-BLANK NO
REPLICATION 1
SETTINGS
PROCEDURE: MODE KINETIC
WAVELENGTH 405
Pipette 1000 µL of WR and 20 µL of sample into a VOLUME 500
test tube. T-FACTOR 1.000
Mix and after 30 secondes incubation at 37°C,
aspirate the mixing by the analyser and start
DELAY 30
measurement. MEAS-TIME 30
NON-LIN 10
LIMITS
Note: Before use, the working reagent WR should L-ABS **
be brought to the working temperature (37°C). H-ABS **
REF-LOW **
REF-HIGH **
DECIMALS 0
CALIBRATION
METHOD FACTOR
FACTOR 2750
Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
ALT/GPT 4+1 SL
ELITech Clinical Systems SAS
REF: ALSL-XXXX
EXTENDED LINEARITY
APPLICATION MICROLAB 300
PROPOSAL
(03/20129)
AA-MICROLAB300-ALSL4+1-4
Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
AMYLASE SL
ELITech Clinical Systems SAS
REF: AMSL-XXXX
(03/2019)
AA-MICROLAB300-AMSL-6
Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
AST/GOT 4+1 SL
ELITech Clinical Systems SAS
REF: ASSL-XXXX
EXTENDED LINEARITY
APPLICATION MICROLAB 300
PROPOSAL
Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
BILIRUBIN
ELITech Clinical Systems SAS DIRECT4+1
REF: BITD-XXXX
IMPROVED PRODUCT
Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
BILIRUBIN
ELITech Clinical Systems SAS TOTAL 4+1
REF: BITD-XXXX
IMPROVED PRODUCT
(03/2019)
AA-MICROLAB300-BITO-Improved Product-5
Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
CALCIUM
ELITech Clinical Systems SAS
ARSENAZO
Be carrefull: This application requires a REF: CALA-XXXX
filter at 660 nm.
(03/2019)
AA-MICROLAB300-CALA-5
Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
CHLORIDE
ELITech Clinical Systems SAS
REF: CHLO-XXXX
Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
CHOLESTEROL SL
ELITech Clinical Systems SAS
REF: CHSL-XXXX
(03/2019)
AA-MICROLAB300-CHSL-5
Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
CHOLESTEROL HDL
ELITech Clinical Systems SAS SL 2G
REF: HDLL-XXXX
Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
CHOLESTEROL LDL
ELITech Clinical Systems SAS SL 2G
REF: LDLL-XXXX
(03/2019)
AA-MICROLAB300-LDLL-5
Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
CHOLINESTERASE
ELITech Clinical Systems SAS
REF: CHES-XXXX
Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
CK-MB SL
ELITech Clinical Systems SAS
REF: CMSL-XXXX
(03/2019)
AA-MICROLAB300-CMSL-4
Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
CK NAC SL
ELITech Clinical Systems SAS
REF: CKSL-XXXX
(03/2019)
AA-MICROLAB300-CKSL-4
Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
CREATININE
ELITech Clinical Systems SAS
JAFFE
REF: CRCO-XXXX
(03/2019)
AA-MICROLAB300-CRCO-5
Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
CREATININE PAP SL
ELITech Clinical Systems SAS
REF/ CRSL-XXXX
Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
GAMMA-GT PLUS SL
ELITech Clinical Systems SAS
REF: GISL-XXXX
(03/2019)
AA-MICROLAB300-GISL-3
Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
GLUCOSE HK SL
ELITech Clinical Systems SAS
REF: GHSL-XXXX
(03/2019)
AA-MICROLAB300-GHSL-5
Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
GLUCOSE PAP SL
ELITech Clinical Systems SAS
REF: GPSL-XXXX
(03/2019)
AA-MICROLAB300-GPSL-5
Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
LACTATE
ELITech Clinical Systems SAS
REF: LACT-XXXX
(03/2019)
AA-MICROLAB300-LACT-4
Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
LDH-L SL
ELITech Clinical Systems SAS
REF: LLSL-XXXX
.
** Enter data by the operator
Modification from the previous version
(03/2019)
AA-MICROLAB300-LLSL-3
Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
LIPASE SL
ELITech Clinical Systems SAS
REF: LPSL-XXXX
03/2019)
AA-MICROLAB300-LPSL-2
Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
MAGNESIUM
ELITech Clinical Systems SAS
XYLIDYL
REF: MAGX-XXXX
INSTRUMENT SETTINGS
GENERAL
PREPARATION OF WORKING REAGENT (WR): NAME MAGX
UNITS mg/dL
Reagent is ready to use.
R-BLANK YES
S-BLANK NO
REPLICATION 1
SETTINGS
PROCEDURE: MODE END-POINT
WAVELENGTH 620
Pipette 1000 µL of reagent and 10 µL of sample VOLUME 500
into a test tube. T-FACTOR 1.000
Mix and after 5 minutes incubation at 37°C, DELAY 3
aspirate the mixing by the analyser and start BI-CHROM YES
measurement. WAVELENGTH 2 505
LIMITS
Note: Before use, the working reagent WR should L-ABS -0.999
be brought to the working temperature (37°C). H-ABS 3.000
REF-LOW **
REF-HIGH **
DECIMALS 2
L-ABS-RB 0.300
H-ABS-RB 1.000
CALIBRATION
METHOD 1-POINT
NAME **
FACTOR 1
REPLICATION 3
DEVIATION **
L-ABS **
H-ABS **
STANDARDS
CON 1 #
.
** Enter data by the operator.
# Enter standard/calibrator value.
Modification from the previous version
(03/2019)
AA-MICROLAB300-MAGX-1
Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
MICROPROTEIN
ELITech Clinical Systems SAS
PLUS
REF: PRTU-XXXX
(03/2019)
AA-MICROLAB300-PRTU-3
Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
PHOSPHORUS
ELITech Clinical Systems SAS
REF: PHOS-XXXX
Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
TOTAL PROTEIN
ELITech Clinical Systems SAS
PLUS
REF: PROB-XXXX
(03/2019)
AA-MICROLAB300-PROB-3
Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
TRIGLYCERIDES
ELITech Clinical Systems SAS MONO SL NEW
REF: TGML-XXXX
Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
UREA UV SL
ELITech Clinical Systems SAS
REF: URSL-XXXX
LIMITS
L-ABS **
H-ABS **
REF-LOW **
REF-HIGH **
DECIMALS 1
CALIBRATION
METHOD 1-POINT
NAME **
FACTOR 1
REPLICATION 3
DEVIATION **
L-ABS **
H-ABS **
STANDARDS
CON 1 #
Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
URIC ACID
ELITech Clinical Systems SAS
MONO SL
REF: AUML-XXXX
Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
URIC ACID SL
ELITech Clinical Systems SAS
REF: AUSL-XXXx
INSTRUMENT SETTINGS
GENERAL
PREPARATION OF WORKING REAGENT (WR): NAME AUSL
UNITS mg/dL
Mix 4 volumes of reagent 1 with 1 volume of
R-BLANK YES
reagent 2.
S-BLANK NO
REPLICATION 1
PROCEDURE: SETTINGS
MODE END-POINT
Pipette 1000 µL of WR and 25 µL of sample into a WAVELENGTH 546
test tube. VOLUME 500
Mix and after 5 minutes incubation at 37°C, T-FACTOR 1.000
aspirate the mixing by the analyser and start DELAY 3
measurement. BI-CHROM NO
WAVELENGTH 2 -
LIMITS
Note: Before use, the working reagent WR should L-ABS **
be brought to the working temperature (37°C). H-ABS **
REF-LOW **
REF-HIGH **
DECIMALS 2
L-ABS-RB 0.000*
H-ABS-RB 0.250*
CALIBRATION
METHOD 1-POINT
NAME **
FACTOR 1
REPLICATION 3
DEVIATION **
L-ABS **
H-ABS **
STANDARDS
CON 1 #
Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
CRP IP
ELITech Clinical Systems SAS
REF: ICRP-XXXX
INSTRUMENT SETTINGS
GENERAL
NAME ICRP
PREPARATION OF WORKING REAGENT :
UNITS mg/dL
Reagent 1: Ref: ICRP-6125 Ready to use R-BLANK NO
Reagent 2: Ref: ICRP-5025 Ready to use S-BLANK YES
REPLICATION 1
SETTINGS
CALIBRATION: MODE END-POINT
WAVELENGTH 340
CRP IP CALIBRATOR SET VOLUME 500
Ref: ICRP-0043 Ready to use. T-FACTOR 1.000
Add a zero point (NaCl 9 g/L solution). DELAY 2
BI-CHROM NO
WAVELENGTH 2 -
PROCEDURE:
LIMITS
L-ABS **
Sample blank:
H-ABS **
Pipette 320 µL of Reagent 1 and 20 µL of sample
into a test tube. REF-LOW **
Mix and after 1 minute incubation at 37°C, REF-HIGH **
aspirate the mixing by the analyser and start DECIMALS 2
measurement. CALIBRATION
METHOD MULTI-POINT
Sample: NAME **
Pipette 320 µL of Reagent 1, 32 µL of Reagent 2 NBR/ACCURACY 6/1%
and 20 µL of sample into a test tube. REPLICATION **
Mix and after 5 minutes of incubation at 37°C, DEVIATION **
aspirate the mixing by the analyser and start L-ABS **
measurement. H-ABS **
STANDARDS
CON 1 #
CON 2 #
CON 3 #
CON 4 #
Note: Before use, the reagents should be brought CON 5 #
to the working temperature (37°C). CON 6 #
Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
HAPTOGLOBIN
ELITech Clinical Systems SAS
IP
REF: IHAP-XXXX
INSTRUMENT SETTINGS
PREPARATION OF WORKING REAGENT :
GENERAL
Reagent 1: Ref: IHAP-6125 Ready to use NAME IHAP
Reagent 2: Ref: IHAP-5025 Ready to use UNITS mg/dL
R-BLANK NO
CALIBRATION: S-BLANK YES
REPLICATION 1
PROTEIN IP CALIBRATOR SET SETTINGS
Ref: IPRO-0043 Ready to use. MODE END-POINT
Add a zero point (NaCl 9 g/L solution). WAVELENGTH 340
VOLUME 500
T-FACTOR 1.000
PROCEDURE: DELAY 2
BI-CHROM NO
Sample blank: WAVELENGTH 2 -
Pipette 840 µL of Reagent 1 and 6 µL of sample
LIMITS
into a test tube.
L-ABS **
Mix and after 1 minute incubation at 37°C,
H-ABS **
aspirate the mixing by the analyser and start
REF-LOW **
measurement.
REF-HIGH **
Sample: DECIMALS 0
Pipette 840 µL of Reagent 1, 120 µL of Reagent 2 CALIBRATION
and 6 µL of sample into a test tube. METHOD MULTI-POINT
Mix and after 5 minutes of incubation at 37°C, NAME **
aspirate the mixing by the analyser and start NBR/ACCURACY 6/1%
measurement. REPLICATION **
DEVIATION **
L-ABS **
H-ABS **
Note: Before use, the reagents should be brought STANDARDS
to the working temperature (37°C). CON 1 #
CON 2 #
CON 3 #
CON 4 #
CON 5 #
CON 6 #
** Enter data by the operator.
# Enter standard/calibrator value.
Modification from the previous version
(03/2019)
AA-MICROLAB300-IHAP-3
Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
IgA IP
ELITech Clinical Systems SAS
REF: IIGA-XXXX
INSTRUMENT SETTINGS
GENERAL
PREPARATION OF WORKING REAGENT : NAME IIGA
UNITS mg/dL
Reagent 1: Ref: IIGB-6125 Ready to use
R-BLANK NO
Reagent 2: Ref: IIGA-5025 Ready to use
S-BLANK YES
REPLICATION 1
SETTINGS
CALIBRATION: MODE END-POINT
WAVELENGTH 340
PROTEIN IP CALIBRATOR SET VOLUME 500
Ref: IPRO-0043 Ready to use. T-FACTOR 1.000
Add a zero point (NaCl 9 g/L solution). DELAY 2
BI-CHROM NO
WAVELENGTH 2 -
PROCEDURE:
LIMITS
Sample blank: L-ABS **
Pipette 900 µL of Reagent 1 and 5 µL of sample H-ABS **
into a test tube. REF-LOW **
Mix and after 1 minute incubation at 37°C, REF-HIGH **
aspirate the mixing by the analyser and start DECIMALS 0
measurement. CALIBRATION
METHOD MULTI-POINT
Sample: NAME **
Pipette 900 µL of Reagent 1, 150 µL of Reagent 2 NBR/ACCURACY 6/1%
and 5 µL of sample into a test tube. REPLICATION **
Mix and after 5 minutes of incubation at 37°C, DEVIATION **
aspirate the mixing by the analyser and start L-ABS **
measurement. H-ABS **
STANDARDS
CON 1 #
Note: Before use, the reagents should be brought CON 2 #
to the working temperature (37°C). CON 3 #
CON 4 #
CON 5 #
CON 6 #
Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
IgG IP
ELITech Clinical Systems SAS
REF: IIGG-XXXX
INSTRUMENT SETTINGS
GENERAL
NAME IIGG
PREPARATION OF WORKING REAGENT :
UNITS mg/dL
Reagent 1: Ref: IIGB-6125 Ready to use R-BLANK NO
Reagent 2: Ref: IIGG-5025 Ready to use S-BLANK YES
REPLICATION 1
CALIBRATION: SETTINGS
MODE END-POINT
PROTEIN IP CALIBRATOR SET WAVELENGTH 340
Ref: IPRO-0043 Ready to use. VOLUME 250
Add a zero point (NaCl 9 g/L solution). T-FACTOR 1.000
DELAY 2
BI-CHROM NO
PREPARATION OF SAMPLES AND CONTROLS WAVELENGTH 2 -
LIMITS
Dilute samples and controls at 1/10 in NaCl L-ABS **
9g/L solution. H-ABS **
REF-LOW **
PROCEDURE: REF-HIGH **
DECIMALS 0
Sample blank: CALIBRATION
Pipette 500 µL of Reagent 1 and 10 µL of diluted METHOD MULTI-POINT
sample into a test tube.
NAME **
Mix and after 1 minute incubation at 37°C,
NBR/ACCURACY 6/1%
aspirate the mixing by the analyser and start
REPLICATION **
measurement.
DEVIATION **
Sample: L-ABS **
Pipette 500 µL of Reagent 1, 50 µL of Reagent 2 H-ABS **
and 10 µL of diluted sample into a test tube. STANDARDS
Mix and after 5 minutes of incubation at 37°C, CON 1 #
aspirate the mixing by the analyser and start CON 2 #
measurement. CON 3 #
CON 4 #
CON 5 #
Note: Before use, the reagents should be brought CON 6 #
to the working temperature (37°C).
** Enter data by the operator.
Modification from the previous version # Enter standard/calibrator value.
(03/2019)
AA-MICROLAB300-IIGG-3
Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
IgM IP
ELITech Clinical Systems SAS
REF: IIGM-XXXX
Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
µALBUMIN IP
ELITech Clinical Systems SAS
REF: IMAL-XXXX
INSTRUMENT SETTINGS
GENERAL
NAME IMAL
PREPARATION OF WORKING REAGENT :
UNITS mg/dL
Reagent 1: Ref: IMAL-6125 Ready to use R-BLANK NO
Reagent 2: Ref: IMAL-5025 Ready to use S-BLANK YES
REPLICATION 1
SETTINGS
CALIBRATION: MODE END-POINT
WAVELENGTH 340
µALBUMIN IP CALIBRATOR SET VOLUME 250
Ref: IMAL-0043 Ready to use. T-FACTOR 1.000
Add a zero point (NaCl 9 g/L solution). DELAY 2
BI-CHROM NO
WAVELENGTH 2 -
PROCEDURE:
LIMITS
Sample blank: L-ABS **
Pipette 300 µL of Reagent 1 and 20 µL of sample H-ABS **
into a test tube. REF-LOW **
Mix and after 1 minute incubation at 37°C, REF-HIGH **
aspirate the mixing by the analyser and start DECIMALS 2
measurement. CALIBRATION
METHOD MULTI-POINT
Sample: NAME **
Pipette 300 µL of Reagent 1, 50 µL of Reagent 2 NBR/ACCURACY 6/1%
and 20 µL of sample into a test tube. REPLICATION **
Mix and after 5 minutes of incubation at 37°C, DEVIATION **
aspirate the mixing by the analyser and start L-ABS **
measurement. H-ABS **
STANDARDS
CON 1 #
CON 2 #
Note: Before use, the reagents should be brought CON 3 #
to the working temperature (37°C).
CON 4 #
CON 5 #
CON 6 #
Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
OROSOMUCOID IP
ELITech Clinical Systems SAS
REF: IORO-XXXX
INSTRUMENT SETTINGS
GENERAL
PREPARATION OF WORKING REAGENT : NAME IORO
UNITS mg/dL
Reagent 1: Ref: IORO-6125 Ready to use
R-BLANK NO
Reagent 2: Ref: IORO-5025 Ready to use
S-BLANK YES
REPLICATION 1
CALIBRATION: SETTINGS
MODE END-POINT
PROTEIN IP CALIBRATOR SET WAVELENGTH 340
Ref: IPRO-0043 Ready to use. VOLUME 500
Add a zero point (NaCl 9 g/L solution). T-FACTOR 1.000
DELAY 2
BI-CHROM NO
PROCEDURE:
WAVELENGTH 2 -
LIMITS
Sample blank:
Pipette 750 µL of Reagent 1 and 5 µL of sample L-ABS **
into a test tube. H-ABS **
Mix and after 1 minute incubation at 37°C, REF-LOW **
aspirate the mixing by the analyser and start REF-HIGH **
measurement. DECIMALS 0
CALIBRATION
Sample: METHOD MULTI-POINT
Pipette 750 µL of Reagent 1, 100 µL of Reagent 2 NAME **
and 5 µL of sample into a test tube. NBR/ACCURACY 6/1%
Mix and after 5 minutes of incubation at 37°C, REPLICATION **
aspirate the mixing by the analyser and start DEVIATION **
measurement. L-ABS **
H-ABS **
STANDARDS
CON 1 #
Note: Before use, the reagents should be brought CON 2 #
to the working temperature (37°C). CON 3 #
CON 4 #
CON 5 #
CON 6 #
Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
PREALBUMIN IP
ELITech Clinical Systems SAS
REF: IPAL-XXXX
INSTRUMENT SETTINGS
GENERAL
PREPARATION OF WORKING REAGENT : NAME IPAL
UNITS mg/dL
Reagent 1: Ref: IPAL-6125 Ready to use
R-BLANK NO
Reagent 2: Ref: IPAL-5025 Ready to use
S-BLANK YES
REPLICATION 1
CALIBRATION: SETTINGS
MODE END-POINT
PROTEIN IP CALIBRATOR SET WAVELENGTH 340
Ref: IPRO-0043 Ready to use. VOLUME 250
Add a zero point (NaCl 9 g/L solution). T-FACTOR 1.000
DELAY 2
BI-CHROM NO
PROCEDURE:
WAVELENGTH 2 -
LIMITS
Sample blank:
Pipette 550 µL of Reagent 1 and 5 µL of sample L-ABS **
into a test tube. H-ABS **
Mix and after 1 minute incubation at 37°C, REF-LOW **
aspirate the mixing by the analyser and start REF-HIGH **
measurement. DECIMALS 0
CALIBRATION
Sample: METHOD MULTI-POINT
Pipette 550 µL of Reagent 1, 50 µL of Reagent 2 NAME **
and 5 µL of sample into a test tube. NBR/ACCURACY 6/1%
Mix and after 5 minutes of incubation at 37°C, REPLICATION **
aspirate the mixing by the analyser and start DEVIATION **
measurement. L-ABS **
H-ABS **
STANDARDS
CON 1 #
Note: Before use, the reagents should be brought CON 2 #
to the working temperature (37°C). CON 3 #
CON 4 #
CON 5 #
CON 6 #
Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
TRANSFERRIN IP
ELITech Clinical Systems SAS
REF: ITRF-XXXX
INSTRUMENT SETTINGS
GENERAL
PREPARATION OF WORKING REAGENT : NAME ITRF
UNITS mg/dL
Reagent 1: Ref: ITRF-6125 Ready to use
R-BLANK NO
Reagent 2: Ref: ITRF-5025 Ready to use
S-BLANK YES
REPLICATION 1
CALIBRATION: SETTINGS
MODE END-POINT
PROTEIN IP CALIBRATOR SET
WAVELENGTH 340
Ref: IPRO-0043 Ready to use.
VOLUME 500
Add a zero point (NaCl 9 g/L solution).
T-FACTOR 1.000
DELAY 2
PREPARATION OF SAMPLES AND CONTROLS BI-CHROM NO
WAVELENGTH 2 -
Dilute samples and controls at 1/20 in NaCl LIMITS
9g/L solution. L-ABS **
H-ABS **
PROCEDURE: REF-LOW **
REF-HIGH **
Sample blank: DECIMALS 0
Pipette 880 µL of Reagent 1 and 12 µL of diluted CALIBRATION
sample into a test tube. METHOD MULTI-POINT
Mix and after 1 minute incubation at 37°C,
NAME **
aspirate the mixing by the analyser and start
NBR/ACCURACY 6/1%
measurement.
REPLICATION **
DEVIATION **
Sample:
Pipette 880 µL of Reagent 1, 40 µL of Reagent 2 L-ABS **
and 12 µL of diluted sample into a test tube. H-ABS **
Mix and after 5 minutes of incubation at 37°C, STANDARDS
aspirate the mixing by the analyser and start CON 1 #
measurement. CON 2 #
CON 3 #
CON 4 #
CON 5 #
Note: Before use, the reagents should be brought CON 6 #
to the working temperature (37°C).
** Enter data by the operator.
# Enter standard/calibrator value.
Modification from the previous version
(03/2019)
AA-MICROLAB300-ITRF-3
Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.