Microlab 300 Applications

This document provides instructions and settings for running an albumin assay on the ELITech Clinical Systems SAS MICROLAB 300 analyzer. It details the reagent preparation, assay procedure, instrument settings, calibration, and quality control parameters for the ALBU-4 albumin reagent version for use with the analyzer. The recommended settings are based on experience with the analyzer and reagent, but each lab should validate the procedure prior to clinical use.

Uploaded by

Shahid Iqbal
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
100% found this document useful (6 votes)
8K views41 pages

Microlab 300 Applications

This document provides instructions and settings for running an albumin assay on the ELITech Clinical Systems SAS MICROLAB 300 analyzer. It details the reagent preparation, assay procedure, instrument settings, calibration, and quality control parameters for the ALBU-4 albumin reagent version for use with the analyzer. The recommended settings are based on experience with the analyzer and reagent, but each lab should validate the procedure prior to clinical use.

Uploaded by

Shahid Iqbal
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 41

ELITech Clinical Systems SAS

MICROLAB 300
AA-MICROLAB300-CAT-10

A Versions M

ALBUMIN .............................................. ALBU-4 ................ PRO NEW MAGNESIUM XYLIDYL ......... MAGX-1................ PRO
Update ALP (DEA) SL ....................................... PASL-4 ................ PRO MICROPROTEIN PLUS ........ PRTU-3 ................ PRO
Update ALT / GPT 4+1 SL ................................ ALSL4+1-4........... PRO
AMYLASE SL ....................................... AMSL-6................ PRO
Update AST / GOT 4+1 SL................................ ASSL4+1-5 .......... PRO

P
B PHOSPHORUS ..................... PHOS-4 ................ PRO

BILIRUBIN TOTAL 4+1 ......................... BITO-5 ................. PRO


BILIRUBIN DIRECT 4+1 ....................... BIDI-5 .................. PRO

C TOTAL PROTEIN PLUS ......... PROB-3 ................ PRO


TRIGLYCERIDES MONO SL .. TGML-4 ................ PRO
CALCIUM ARSENAZO ......................... CALA-5 ................ PRO
CHLORIDE ........................................... CHLO-4 ............... PRO
CHOLESTEROL SL .............................. CHSL-5 ................ PRO
CHOLESTEROL HDL SL 2G ................ HDLL-5 ................ PRO
CHOLESTEROL LDL SL 2G ................. LDLL-5 ................. PRO
CHOLINESTERASE ............................. CHES-4 ............... PRO
CK - MB SL ........................................... CMSL-4 ............... PRO U
CK NAC SL ........................................... CKSL-4 ................ PRO
Update CREATININE JAFFE ............................ CRCO-5 ............... PRO Update UREA UV SL ........................... URSL-5 ................ PRO
CREATININE PAP SL........................... CRSL-5 ................ PRO URIC ACID MONO SL ............. AUML-5 ................ PRO
URIC ACID SL ......................... AUSL-5................. PRO

GAMMA-GT PLUS SL........................... GISL-3 ................. PRO


GLUCOSE HK SL ................................. GHSL-5................ PRO SPECIFIC PROTEINS
GLUCOSE PAP SL ............................... GPSL-5 ................ PRO

CRP IP ...................................... ICRP-3 .............. PRO


HAPTOGLOBIN IP .................... IHAP-3 .............. PRO
IgA IP ........................................ IIGA-3................ PRO
IgG IP ........................................ IIGG-3 ............... PRO
L Versions IgM IP ........................................ IIGM-3 ............... PRO
µALBUMIN IP ............................ IMAL-5 .............. PRO
LACTATE................................................ LACT-4 .................PRO OROSOMUCOID IP ................. IORO-3.............. PRO
LDH-L SL ................................................. LLSL-3 ..................PRO PREALBUMIN IP¨...................... IPAL-3 ............... PRO
LIPASE SL ............................................... LPSL-2 .................PRO TRANSFERRIN IP ..................... ITRF-3 ............... PRO
ALBUMIN
ELITech Clinical Systems SAS

REF: ALBU-XXXX

APPLICATION MICROLAB 300


PROPOSAL

For more details, see the Instructions for use :


and consult the analyzer manual.

Working temperature : 37°C

INSTRUMENT SETTINGS

GENERAL
NAME ALBU
PREPARATION OF WORKING REAGENT:
UNITS g/dL
The reagent is ready for use. R-BLANK YES
S-BLANK NO
REPLICATION 1
SETTINGS
PROCEDURE: MODE END-POINT
WAVELENGTH 620
Pipette 1000 µL of reagent and 10 µL of sample VOLUME 500
into a test tube. T-FACTOR 1.000
Mix and after 5 minutes incubation at 37°C, DELAY 3
aspirate the mixing by the analyser and start BI-CHROM NO
measurement. WAVELENGTH 2 -
LIMITS
Note: Before use, the reagent should be brought L-ABS **
to the working temperature (37°C). H-ABS **
REF-LOW **
REF-HIGH **
DECIMALS 2
L-ABS-RB 0.000*
H-ABS-RB 0.300*
CALIBRATION
METHOD 1-POINT
NAME **
FACTOR 1
REPLICATION 3
DEVIATION **
L-ABS **
H-ABS **
STANDARDS
CON 1 #

* This value could be optimized.


Modification from the previous version ** Enter data by the operator.
# Enter standard/calibrator value.

(03/2019)
AA-MICROLAB300-ALBU-4

Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
ALP (DEA) SL
ELITech Clinical Systems SAS

REF: PASL-XXXX

APPLICATION MICROLAB 300


PROPOSAL

For more details, see the Instructions for use :


and consult the analyzer manual.

Working temperature : 37°C

INSTRUMENT SETTINGS

GENERAL
PREPARATION OF WORKING REAGENT (WR): NAME PASL
UNITS U/L
Mix 4 volumes of reagent 1 with 1 volume of R-BLANK NO
reagent 2.
S-BLANK NO
REPLICATION 1
SETTINGS
PROCEDURE: MODE KINETIC
WAVELENGTH 405
Pipette 1000 µL of WR and 20 µL of sample into a VOLUME 500
test tube. T-FACTOR 1.000
Mix and after 30 secondes incubation at 37°C,
aspirate the mixing by the analyser and start
DELAY 30
measurement. MEAS-TIME 30
NON-LIN 10
LIMITS
Note: Before use, the working reagent WR should L-ABS **
be brought to the working temperature (37°C). H-ABS **
REF-LOW **
REF-HIGH **
DECIMALS 0
CALIBRATION
METHOD FACTOR
FACTOR 2750

** Enter data by the operator

Modification from the previous version


(03/2019)
AA-MICROLAB300-PASL-4

Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
ALT/GPT 4+1 SL
ELITech Clinical Systems SAS

REF: ALSL-XXXX

EXTENDED LINEARITY
APPLICATION MICROLAB 300
PROPOSAL

For more details, see the Instructions for use :


and consult the analyzer manual.

Working temperature : 37°C

PREPARATION OF WORKING REAGENT (WR): INSTRUMENT SETTINGS

Mix 4 volumes of reagent 1 with 1 volume of


reagent 2. GENERAL
NAME ALSL
UNITS U/L
R-BLANK NO
PROCEDURE: S-BLANK NO
REPLICATION 1
Pipette 1000 µL of WR and 50 µL of sample into a SETTINGS
test tube. MODE KINETIC
Mix and after 30 secondes incubation, aspirate WAVELENGTH 340
the mixing by the analyser and start VOLUME 500
measurement. T-FACTOR 1.000
DELAY 30
Note: Before use, the working reagent WR should
MEAS-TIME 30
be brought to the working temperature (37°C). NON-LIN 10
LIMITS
L-ABS **
H-ABS **
REF-LOW **
REF-HIGH **
DECIMALS 0
CALIBRATION
METHOD FACTOR
FACTOR -3333
** Enter data by the operator.

Modification from the previous version

(03/20129)
AA-MICROLAB300-ALSL4+1-4

Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
AMYLASE SL
ELITech Clinical Systems SAS

REF: AMSL-XXXX

APPLICATION MICROLAB 300


PROPOSAL

For more details, see the Instructions for use :


and consult the analyzer manual.

Working temperature : 37°C

PREPARATION OF WORKING REAGENT: INSTRUMENT SETTINGS

The reagent is ready for use.


GENERAL
NAME AMSL
UNITS U/L
R-BLANK NO
S-BLANK NO
PROCEDURE: REPLICATION 1
SETTINGS
Pipette 1000 µL of reagent and 10 µL of sample MODE KINETIC
into a test tube. WAVELENGTH 405
Mix, aspirate the mixing by the analyser and
VOLUME 500
start measurement.
T-FACTOR 1.000
DELAY 60
MEAS-TIME 30
Note: Before use, the reagent should be brought
to the working temperature (37°C). NON-LIN 10
LIMITS
L-ABS **
H-ABS **
REF-LOW **
REF-HIGH **
DECIMALS 0
CALIBRATION
METHOD FACTOR
FACTOR 9227

** Enter data by the operator

Modification from the previous version

(03/2019)
AA-MICROLAB300-AMSL-6

Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
AST/GOT 4+1 SL
ELITech Clinical Systems SAS
REF: ASSL-XXXX

EXTENDED LINEARITY
APPLICATION MICROLAB 300
PROPOSAL

For more details, see the Instructions for use :


and consult the analyzer manual.

Working temperature : 37°C

PREPARATION OF WORKING REAGENT (WR): INSTRUMENT SETTINGS

Mix 4 volumes of reagent 1 with 1 volume of


reagent 2. GENERAL
NAME ASSL
UNITS U/L
PROCEDURE: R-BLANK NO
S-BLANK NO
Pipette 1000 µL of WR and 50 µL of sample into a REPLICATION 1
test tube. SETTINGS
Mix and after 30 secondes incubation, aspirate MODE KINETIC
the mixing by the analyser and start WAVELENGTH 340
measurement. VOLUME 500
T-FACTOR 1.000

Note: Before use, the working reagent WR should


DELAY 30
be brought to the working temperature (37°C).
MEAS-TIME 30
NON-LIN 10
LIMITS
L-ABS **
H-ABS **
REF-LOW **
REF-HIGH **
DECIMALS 0
CALIBRATION
METHOD FACTOR
FACTOR -3333
** Enter data by the operator.

Modification from the previous version


(03/2019)
AA-MICROLAB300-ASSL4+1-5

Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
BILIRUBIN
ELITech Clinical Systems SAS DIRECT4+1
REF: BITD-XXXX
IMPROVED PRODUCT

APPLICATION MICROLAB 300


PROPOSAL

For more details, see the Instructions for use :


and consult the analyzer manual.

Working temperature : 37°C

PREPARATION OF WORKING REAGENT: INSTRUMENT SETTINGS

The reagents are ready for use.


GENERAL
NAME BIDI
UNITS mg/dL
R-BLANK NO
PROCEDURE: S-BLANK YES
REPLICATION 1
Sample blank: SETTINGS
Pipette 1000 µL of reagent 1 (Direct) and 100 µL MODE END-POINT
of sample into a test tube. WAVELENGTH 546
Mix and after 5 minutes incubation at 37°C,
VOLUME 500
aspirate the mixing by the analyser and start
T-FACTOR 1.000
measurement.
DELAY 3
BI-CHROM NO
Sample:
Pipette 800 µL of reagent 1 (Direct) and 100 µL of WAVELENGTH 2 -
sample into a test tube. Mix and wait 5 minutes LIMITS
at 37°C, and add 200 µL of reagent 2. L-ABS **
Mix and wait 50 secondes at 37°C then aspirate H-ABS **
the mixing by the analyser and start REF-LOW **
measurement. REF-HIGH **
DECIMALS 2
Note: Before use, the reagent should be brought CALIBRATION
to the working temperature (37°C). METHOD FACTOR
FACTOR 21.0

** Enter data by the operator.


EXTRA CLEANING (after each samples)
- 1x Cleaning cycle (Bleach solution 5%)
- 1x rinse cycle (distilled water)
The analyzer cannot clean automatically after a
sample. Please do the cleaning manually by
using these 2 solutions in place of a sample
blank and sample measurement.

Modification from the previous version


(03/2019)
AA-MICROLAB300-BITD Improved-5

Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
BILIRUBIN
ELITech Clinical Systems SAS TOTAL 4+1
REF: BITD-XXXX
IMPROVED PRODUCT

APPLICATION MICROLAB 300


PROPOSAL

For more details, see the Instructions for use :


and consult the analyzer manual.

Working temperature : 37°C

PREPARATION OF WORKING REAGENT: INSTRUMENT SETTINGS

The reagents are ready for use.


GENERAL
NAME BITO
UNITS mg/dL
R-BLANK NO
PROCEDURE: S-BLANK YES
REPLICATION 1
Sample blank: SETTINGS
Pipette 1000 µL of reagent 1 (Total ) and 50 µL of MODE END-POINT
sample into a test tube. WAVELENGTH 546
Mix and after 5 minutes incubation at 37°C,
VOLUME 500
aspirate the mixing by the analyser and start
T-FACTOR 1.000
measurement.
DELAY 3
BI-CHROM NO
Sample:
Pipette 800 µL of reagent 1 (Total) and 50 µL of WAVELENGTH 2 -
sample into a test tube. Mix and wait 5 minutes LIMITS
at 37°C, and add 200 µL of reagent 2. L-ABS **
Mix and wait 5 minutes at 37°C then aspirate the H-ABS **
mixing by the analyser and start measurement. REF-LOW **
REF-HIGH **
Note: Before use, the reagents should be brought DECIMALS 2
to the working temperature (37°C). CALIBRATION
METHOD FACTOR
FACTOR 31

** Enter data by the operator.

Modification from the previous version

(03/2019)
AA-MICROLAB300-BITO-Improved Product-5

Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
CALCIUM
ELITech Clinical Systems SAS
ARSENAZO
Be carrefull: This application requires a REF: CALA-XXXX
filter at 660 nm.

APPLICATION MICROLAB 300


PROPOSAL

For more details, see the Instructions for use :


and consult the analyzer manual.

Working temperature : 37°C

PREPARATION OF WORKING REAGENT: INSTRUMENT SETTINGS

The reagent is ready for use. GENERAL


NAME CALA
UNITS mg/dL
R-BLANK YES
S-BLANK NO
PROCEDURE:
REPLICATION 1
SETTINGS
Pipette 1000 µL of reagent and 20 µL of sample
into a test tube. MODE END-POINT
Mix and aspirate the mixing by the analyser and WAVELENGTH 660
start measurement. VOLUME 500
T-FACTOR 1.000
Note: Before use, the reagent should be brought DELAY 60
to the working temperature (37°C). BI-CHROM NO
WAVELENGTH 2 -
 LIMITS
 L-ABS **
 H-ABS **
REF-LOW **
 REF-HIGH **
 DECIMALS 2
 L-ABS-RB 0.000*
H-ABS-RB 2.300*
 CALIBRATION
 METHOD 1-POINT
 NAME **
FACTOR 1
 REPLICATION 3
 DEVIATION **
 L-ABS **
H-ABS **
 STANDARDS
 CON 1 #

* This value could be optimized.
Modification from the previous version ** Enter data by the operator.
# Enter standard/calibrator value.

(03/2019)
AA-MICROLAB300-CALA-5

Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
CHLORIDE
ELITech Clinical Systems SAS

REF: CHLO-XXXX

APPLICATION MICROLAB 300


PROPOSAL

For more details, see the Instructions for use :


and consult the analyzer manual.

Working temperature : 37°C

PREPARATION OF WORKING REAGENT (WR) INSTRUMENT SETTINGS

The reagent is ready for use. GENERAL


NAME CHLO
UNITS mEq/L
R-BLANK YES
S-BLANK NO
PROCEDURE:
REPLICATION 1
SETTINGS
Pipette 1000 µL of reagent and 10 µL of sample
into a test tube. MODE END-POINT
Mix and after 5 minutes incubation at 37°C, WAVELENGTH 505
aspirate the mixing by the analyser and start VOLUME 500
measurement. T-FACTOR 1.000
DELAY 3
Note: Before use, the reagent should be brought BI-CHROM NO
to the working temperature (37°C). WAVELENGTH 2 -
LIMITS
L-ABS **
H-ABS **
REF-LOW **
REF-HIGH **
DECIMALS 1
L-ABS-RB 0.000*
H-ABS-RB 0.200*
CALIBRATION
METHOD 1-POINT
NAME **
FACTOR 1
REPLICATION 3
DEVIATION **
L-ABS **
H-ABS **
STANDARDS
CON 1 #

* This value could be optimized.


** Enter data by the operator.
# Enter standard/calibrator value.
Modification from the previous version
(03/2019)
AA-MICROLAB300-CHLO-4

Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
CHOLESTEROL SL

ELITech Clinical Systems SAS 

REF: CHSL-XXXX

APPLICATION MICROLAB 300


PROPOSAL

For more details, see the Instructions for use :


and consult the analyzer manual.

Working temperature : 37°C

PREPARATION OF WORKING REAGENT INSTRUMENT SETTINGS

The reagent is ready for use. GENERAL


NAME CHSL
UNITS mg/dL
R-BLANK YES
S-BLANK NO
PROCEDURE:
REPLICATION 1
SETTINGS
Pipette 1000 µL of reagent and 10 µL of sample
into a test tube. MODE END-POINT
Mix and after 5 minutes incubation at 37°C, WAVELENGTH 505
aspirate the mixing by the analyser and start VOLUME 500
measurement. T-FACTOR 1.000
DELAY 3
Note: Before use, the reagent should be brought BI-CHROM NO
to the working temperature (37°C). WAVELENGTH 2 -
LIMITS
L-ABS **
H-ABS **
REF-LOW **
REF-HIGH **
DECIMALS 0
L-ABS-RB 0.000*
H-ABS-RB 0.100*
CALIBRATION
METHOD 1-POINT
NAME **
FACTOR 1
REPLICATION 3
DEVIATION **
L-ABS **
H-ABS **
STANDARDS
Modification from the previous version CON 1 #

* This value could be optimized.


** Enter data by the operator.
# Enter standard/calibrator value.

(03/2019)
AA-MICROLAB300-CHSL-5

Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
CHOLESTEROL HDL
ELITech Clinical Systems SAS SL 2G
REF: HDLL-XXXX

APPLICATION MICROLAB 300


PROPOSAL

For more details, see the Instructions for use :


and consult the analyzer manual.

Working temperature : 37°C

PREPARATION OF WORKING REAGENT : INSTRUMENT SETTINGS

Reagent 1: Ready to use GENERAL


Reagent 2: Ready to use. NAME HDLL
UNITS mg/dL
R-BLANK NO
PROCEDURE:
S-BLANK YES
REPLICATION 1
Sample blank: SETTINGS
Pipette 1200 µL of Reagent 1 and 10 µL of sample MODE END-POINT
into a test tube. WAVELENGTH 578
Mix and after 4 minutes incubation at 37°C, VOLUME 500
aspirate the mixing by the analyser and start T-FACTOR 1.000
measurement. DELAY 3
BI-CHROM NO
WAVELENGTH 2 -
Sample: LIMITS
Pipette 900 µL of Reagent 1, 10 µL of sample, mix L-ABS **
and after 5 minutes of incubation at 37°C, add H-ABS **
300 µL of reagent 2 into a test tube. REF-LOW **
Mix and after 4 minutes incubation at 37°C, REF-HIGH **
aspirate the mixing by the analyser and start DECIMALS 0
measurement. CALIBRATION
METHOD 1-POINT
NAME **
FACTOR 1
REPLICATION 3
DEVIATION **
L-ABS **
H-ABS **
Note: Before use, the reagents should be brought STANDARDS
to the working temperature (37°C). CON 1 #

** Enter data by the operator.


# Enter standard/calibrator value.

Modification from the previous version


(03/2019)
AA-MICROLAB300-HDLL-5

Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
CHOLESTEROL LDL
ELITech Clinical Systems SAS SL 2G
REF: LDLL-XXXX

APPLICATION MICROLAB 300


PROPOSAL

For more details, see the Instructions for use :


and consult the analyzer manual.

Working temperature : 37°C

PREPARATION OF WORKING REAGENT : INSTRUMENT SETTINGS

Reagent 1: Ready to use GENERAL


Reagent 2: Ready to use. NAME LDLL
UNITS mg/dL
R-BLANK NO
PROCEDURE:
S-BLANK YES
REPLICATION 1
Sample blank: SETTINGS
Pipette 1200 µL of Reagent 1 and 10 µL of sample MODE END-POINT
into a test tube. WAVELENGTH 578
Mix and after 4 minutes incubation at 37°C, VOLUME 500
aspirate the mixing by the analyser and start T-FACTOR 1.000
measurement. DELAY 3
BI-CHROM NO
WAVELENGTH 2 -
Sample: LIMITS
Pipette 900 µL of Reagent 1, 10 µL of sample, mix L-ABS **
and after 5 minutes of incubation at 37°C, add H-ABS **
300 µL of reagent 2 into a test tube. REF-LOW **
Mix and after 4 minutes incubation at 37°C, REF-HIGH **
aspirate the mixing by the analyser and start DECIMALS 0
measurement. CALIBRATION
METHOD 1-POINT
NAME **
FACTOR 1
REPLICATION 3
DEVIATION **
L-ABS **
H-ABS **
Note: Before use, the reagents should be brought STANDARDS
to the working temperature (37°C). CON 1 #

** Enter data by the operator.


# Enter standard/calibrator value.

Modification from the previous version

(03/2019)
AA-MICROLAB300-LDLL-5

Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
CHOLINESTERASE

ELITech Clinical Systems SAS 
REF: CHES-XXXX

APPLICATION MICROLAB 300


PROPOSAL

For more details, see the Instructions for use :


and consult the analyzer manual.

Working temperature : 37°C

PREPARATION OF WORKING REAGENT: INSTRUMENT SETTINGS

Working reagent 1 (WR1)


Dissolve the reagent 1 with the suitable volume of GENERAL
distilled water as stated on the label. NAME CHES
Wait 15 minutes before use. UNITS U/L
R-BLANK NO
Working reagent 2 (WR2) S-BLANK NO
Dissolve the reagent 2 with the suitable volume of REPLICATION 1
distilled water as stated on the label. SETTINGS
Wait 15 minutes before use.
MODE KINETIC
WAVELENGTH 405
VOLUME 500
T-FACTOR 1.000
DELAY 2
PROCEDURE: MEAS-TIME 90
NON-LIN 10
Pipette 1500 µL of WR1, 10 µL of sample and 50 LIMITS
µL of WR2 into a test tube. L-ABS **
Mix and aspirate the mixing by the analyser and H-ABS **
start measurement. REF-LOW **
REF-HIGH **
Note: Before use, the working reagent WR1 DECIMALS 1
should be brought to the working temperature CALIBRATION
(37°C). METHOD FACTOR
FACTOR 23460

** Enter data by the operator

Modification from the previous version


(03/2019)
AA-MICROLAB300-CHES-4

Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
CK-MB SL
ELITech Clinical Systems SAS

REF: CMSL-XXXX

APPLICATION MICROLAB 300


PROPOSAL

For more details, see the Instructions for use :


and consult the analyzer manual.

Working temperature : 37°C

PREPARATION OF WORKING REAGENT (WR): INSTRUMENT SETTINGS

Mix 4 volumes of reagent 1 with 1 volume of


reagent 2. GENERAL
NAME CMSL
UNITS U/L
R-BLANK NO
PROCEDURE: S-BLANK NO
REPLICATION 1
Pipette 1000 µL of WR and 40 µL of sample into a SETTINGS
test tube. MODE KINETIC
Mix, aspirate the mixing by the analyser and WAVELENGTH 340
start measurement.
VOLUME 500
T-FACTOR 1.000
DELAY 180
Note: Before use, the working reagent WR should
MEAS-TIME 180
be brought to the working temperature (37°C).
NON-LIN 10
LIMITS
L-ABS **
H-ABS **
REF-LOW **
REF-HIGH **
DECIMALS 0
CALIBRATION
METHOD FACTOR
FACTOR 8254

** Enter data by the operator.

Modification from the previous version

(03/2019)
AA-MICROLAB300-CMSL-4

Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
CK NAC SL
ELITech Clinical Systems SAS

REF: CKSL-XXXX

APPLICATION MICROLAB 300


PROPOSAL

For more details, see the Instructions for use :


and consult the analyzer manual.

Working temperature : 37°C

PREPARATION OF WORKING REAGENT (WR): INSTRUMENT SETTINGS

Mix 4 volumes of reagent 1 with 1 volume of


reagent 2. GENERAL
NAME CKSL
UNITS U/L
R-BLANK NO
PROCEDURE: S-BLANK NO
REPLICATION 1
Pipette 1000 µL of WR and 40 µL of sample into a SETTINGS
test tube. MODE KINETIC
Mix, aspirate the mixing by the analyser and WAVELENGTH 340
start measurement.
VOLUME 500
T-FACTOR 1.000
Note: Before use, the working reagent WR should
DELAY 120
be brought to the working temperature (37°C).
MEAS-TIME 180
NON-LIN 10
LIMITS
L-ABS **
H-ABS **
REF-LOW **
REF-HIGH **
DECIMALS 0
CALIBRATION
METHOD FACTOR
FACTOR 4127

** Enter data by the operator.

Modification from the previous version

(03/2019)
AA-MICROLAB300-CKSL-4

Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
CREATININE
ELITech Clinical Systems SAS
JAFFE
REF: CRCO-XXXX

APPLICATION MICROLAB 300


PROPOSAL

For more details, see the Instructions for use :


and consult the analyzer manual.

Working temperature : 37°C

PREPARATION OF WORKING REAGENT (WR) INSTRUMENT SETTINGS

Mix 1 volume of the reagent 1 with 1 volume of GENERAL


the reagent 2. NAME CRCO
UNITS mg/dL
R-BLANK NO
S-BLANK NO
PROCEDURE:
REPLICATION 1
SETTINGS
Pipette 1000 µL of WR and 100 µL of sample into
a test tube. MODE KINETIC
Mix and aspirate the mixing by the analyser and WAVELENGTH 505
start measurement. VOLUME 500
T-FACTOR 1.000
Note: Before use, the working reagent WR should DELAY 10
be brought to the working temperature (37°C). MEAS-TIME 30
NON-LIN 10
LIMITS
L-ABS **
H-ABS **
REF-LOW **
REF-HIGH **
DECIMALS 2
CALIBRATION
METHOD 1-POINT
NAME **
FACTOR 1
REPLICATION 3
DEVIATION **
L-ABS **
H-ABS **
STANDARDS
CON 1 #
Modification from the previous version
** Enter data by the operator.
# Enter standard/calibrator value.

(03/2019)
AA-MICROLAB300-CRCO-5

Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
CREATININE PAP SL
ELITech Clinical Systems SAS

REF/ CRSL-XXXX

APPLICATION MICROLAB 300


PROPOSAL

For more details, see the Instructions for use :


and consult the analyzer manual.

Working temperature : 37°C

PREPARATION OF WORKING REAGENT: INSTRUMENT SETTINGS

The reagents are ready for use.


GENERAL
NAME CRSL
UNITS mg/dL
R-BLANK NO
S-BLANK NO
PROCEDURE: REPLICATION 1
SETTINGS
Pipette 600 µL of reagent 1 and 14 µL of sample MODE KINETIC
into a test tube. Mix, wait 5 minutes at 37°C and WAVELENGTH 546
add 200 µL of reagent 2.
VOLUME 500
Mix, aspirate the mixing by the analyser and
T-FACTOR 1.000
start measurement.
DELAY 60
MEAS-TIME 120
Note: Before use, the reagent 1 should be
brought to the working temperature (37°C). NON-LIN 20
LIMITS
L-ABS **
H-ABS **
REF-LOW **
REF-HIGH **
DECIMALS 2
CALIBRATION
METHOD 1-POINT
NAME **
FACTOR 1
REPLICATION 3
DEVIATION **
L-ABS **
H-ABS **
STANDARDS
CON 1 #

** Enter data by the operator.


# Enter standard/calibrator value.

Modification from the previous version


(03/2019)
AA-MICROLAB300-CRSL-5

Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
GAMMA-GT PLUS SL
ELITech Clinical Systems SAS

REF: GISL-XXXX

APPLICATION MICROLAB 300


PROPOSAL

For more details, see the Instructions for use :


and consult the analyzer manual.

Working temperature : 37°C

PREPARATION OF WORKING REAGENT (WR): INSTRUMENT SETTINGS

Mix 4 volumes of reagent 1 with 1 volume of


reagent 2. GENERAL
NAME GISL
UNITS U/L
R-BLANK NO
PROCEDURE: S-BLANK NO
REPLICATION 1
Pipette 1000 µL of WR and 75 µL of sample into a SETTINGS
test tube. MODE KINETIC
Mix, aspirate the mixing by the analyser and WAVELENGTH 405
start measurement.
VOLUME 500
T-FACTOR 1.000
Note: Before use, the working reagent WR should
DELAY 60
be brought to the working temperature (37°C).
MEAS-TIME 30
NON-LIN 10
LIMITS
L-ABS **
H-ABS **
REF-LOW **
REF-HIGH **
DECIMALS 1
CALIBRATION
METHOD FACTOR
FACTOR 1510*

* This factor should be validated by using Elical 2


calibrator.
** Enter data by the operator

Modification from the previous version

(03/2019)
AA-MICROLAB300-GISL-3

Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
GLUCOSE HK SL
ELITech Clinical Systems SAS

REF: GHSL-XXXX

APPLICATION MICROLAB 300


PROPOSAL

For more details, see the Instructions for use :


and consult the analyzer manual.

Working temperature : 37°C

PREPARATION OF WORKING REAGENT (WR): INSTRUMENT SETTINGS

Mix 4 volumes of reagent 1 with 1 volume of GENERAL


reagent 2. NAME GHSL
UNITS mg/dL
R-BLANK YES
S-BLANK NO
PROCEDURE:
REPLICATION 1
SETTINGS
Pipette 1000 µL of WR and 10 µL of sample into a
test tube. MODE END-POINT
Mix and after 5 minutes incubation at 37°C, WAVELENGTH 340
aspirate the mixing by the analyser and start VOLUME 500
measurement. T-FACTOR 1.000
DELAY 3
BI-CHROM NO
Note: Before use, the working reagent WR should WAVELENGTH 2 -
be brought to the working temperature (37°C). LIMITS
 L-ABS **
H-ABS **
 REF-LOW **
 REF-HIGH **
 DECIMALS 1
L-ABS-RB 0.000*
 H-ABS-RB 0.850*
 CALIBRATION
METHOD 1-POINT
 NAME **
 FACTOR 1
 REPLICATION 3
DEVIATION **
 L-ABS **
 H-ABS **
 STANDARDS
CON 1 #

 * This value could be optimized.
Modification from the previous version ** Enter data by the operator.
# Enter standard/calibrator value.

(03/2019)
AA-MICROLAB300-GHSL-5

Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
GLUCOSE PAP SL
ELITech Clinical Systems SAS
REF: GPSL-XXXX

APPLICATION MICROLAB 300


PROPOSAL

For more details, see the Instructions for use :


and consult the analyzer manual.

Working temperature : 37°C

PREPARATION OF WORKING REAGENT: INSTRUMENT SETTINGS

The reagent is ready for use. GENERAL


NAME GPSL
UNITS mg/dL
R-BLANK YES
PROCEDURE:
S-BLANK NO
REPLICATION 1
Pipette 1000 µL of reagent and 10 µL of sample
SETTINGS
into a test tube.
Mix and after 10 minutes incubation at 37°C, MODE END-POINT
aspirate the mixing by the analyser and start WAVELENGTH 505
measurement. VOLUME 500
T-FACTOR 1.000
Note: Before use, the reagent should be brought DELAY 3
to the working temperature (37°C). BI-CHROM NO
WAVELENGTH 2 -
LIMITS
L-ABS **
H-ABS **
REF-LOW **
REF-HIGH **
DECIMALS 1
L-ABS-RB 0.000*
H-ABS-RB 0.250*
CALIBRATION
METHOD 1-POINT
NAME **
FACTOR 1
REPLICATION 3
DEVIATION **
L-ABS **
H-ABS **
STANDARDS
Modification from the previous version CON 1 #

* This value could be optimized.


** Enter data by the operator.
# Enter standard/calibrator value.


(03/2019)
AA-MICROLAB300-GPSL-5

Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
LACTATE
ELITech Clinical Systems SAS

REF: LACT-XXXX

APPLICATION MICROLAB 300


PROPOSAL

For more details, see the Instructions for use :


and consult the analyzer manual.

Working temperature : 37°C

PREPARATION OF WORKING REAGENT (WR): INSTRUMENT SETTINGS

Dissolve the reagent 2 in the suitable volume of GENERAL


reagent 1. NAME LACT
Wait about 15 minutes before use. UNITS mg/dL
R-BLANK YES
S-BLANK NO
REPLICATION 1
PROCEDURE:
SETTINGS
Pipette 1000 µL of WR and 10 µL of sample into a MODE END-POINT
test tube. WAVELENGTH 546
Mix and after 5 minutes incubation at 37°C, VOLUME 500
aspirate the mixing by the analyser and start T-FACTOR 1.000
measurement. DELAY 3
BI-CHROM NO
Note: Before use, the working reagent WR should WAVELENGTH 2 -
be brought to the working temperature (37°C). LIMITS
 L-ABS **
H-ABS **
 REF-LOW **
 REF-HIGH **
 DECIMALS 1
L-ABS-RB 0.000*
 H-ABS-RB 0.300*
 CALIBRATION
METHOD 1-POINT
 NAME **
 FACTOR 1
 REPLICATION 3
DEVIATION **
 L-ABS **
 H-ABS **
Modification from the previous version STANDARDS
CON 1 #

* This value could be optimized.


** Enter data by the operator.
# Enter standard/calibrator value.

(03/2019)
AA-MICROLAB300-LACT-4

Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
LDH-L SL
ELITech Clinical Systems SAS

REF: LLSL-XXXX

APPLICATION MICROLAB 300


PROPOSAL

For more details, see the Instructions for use :


and consult the analyzer manual.

Working temperature : 37°C

PREPARATION OF WORKING REAGENT (WR): INSTRUMENT SETTINGS

Reagents are ready to use.


GENERAL
NAME LLSL
UNITS U/L
PROCEDURE: R-BLANK NO
S-BLANK NO
Pipette 1000 µL of reagent 1 (R1) and 35 µL of REPLICATION 1
sample into a test tube.Mix, wait 3minutes at SETTINGS
37°C and add 250 µL of reagent 2 (R2). MODE KINETIC
Mix, aspirate the mixing by the analyser and WAVELENGTH 340
start measurement.
VOLUME 500
T-FACTOR 1.000
DELAY 100
Note: Before use, the reagents should be brought
MEAS-TIME 180
to the working temperature (37°C).
NON-LIN 10
LIMITS
L-ABS **
H-ABS **
REF-LOW **
REF-HIGH **
DECIMALS 0
CALIBRATION
METHOD FACTOR
FACTOR 5828

.
** Enter data by the operator
Modification from the previous version

(03/2019)
AA-MICROLAB300-LLSL-3

Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
LIPASE SL
ELITech Clinical Systems SAS

REF: LPSL-XXXX

APPLICATION MICROLAB 300


PROPOSAL

For more details, see the Instructions for use :


and consult the analyzer manual.

Working temperature : 37°C

PREPARATION OF WORKING REAGENT (WR): INSTRUMENT SETTINGS

Reagents are ready to use.


GENERAL
NAME LPSL
UNITS U/L
PROCEDURE: R-BLANK NO
S-BLANK NO
Pipette 1000 µL of reagent 1 (R1) and 20 µL of REPLICATION 1
sample into a test tube.Mix, wait 3 minutes at SETTINGS
37°C and add 600 µL of reagent 2 (R2). MODE KINETIC
Mix, aspirate the mixing by the analyser and WAVELENGTH 578
start measurement.
VOLUME 500
T-FACTOR 1.000
DELAY 60
Note: Before use, the reagents should be brought
MEAS-TIME 150
to the working temperature (37°C).
NON-LIN 10
LIMITS
L-ABS **
H-ABS **
REF-LOW **
REF-HIGH **
DECIMALS 0
CALIBRATION
METHOD FACTOR
FACTOR 1685

** Enter data by the operator


Modification from the previous version

03/2019)
AA-MICROLAB300-LPSL-2

Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
MAGNESIUM
ELITech Clinical Systems SAS
XYLIDYL
REF: MAGX-XXXX

APPLICATION MICROLAB 300


PROPOSAL

For more details, see the Instructions for use :


and consult the analyzer manual.

Working temperature : 37°C

INSTRUMENT SETTINGS

GENERAL
PREPARATION OF WORKING REAGENT (WR): NAME MAGX
UNITS mg/dL
Reagent is ready to use.
R-BLANK YES
S-BLANK NO
REPLICATION 1
SETTINGS
PROCEDURE: MODE END-POINT
WAVELENGTH 620
Pipette 1000 µL of reagent and 10 µL of sample VOLUME 500
into a test tube. T-FACTOR 1.000
Mix and after 5 minutes incubation at 37°C, DELAY 3
aspirate the mixing by the analyser and start BI-CHROM YES
measurement. WAVELENGTH 2 505
LIMITS
Note: Before use, the working reagent WR should L-ABS -0.999
be brought to the working temperature (37°C). H-ABS 3.000
REF-LOW **
REF-HIGH **
DECIMALS 2
L-ABS-RB 0.300
H-ABS-RB 1.000
CALIBRATION
METHOD 1-POINT
NAME **
FACTOR 1
REPLICATION 3
DEVIATION **
L-ABS **
H-ABS **
STANDARDS
CON 1 #

.
** Enter data by the operator.
# Enter standard/calibrator value.
Modification from the previous version
(03/2019)
AA-MICROLAB300-MAGX-1

Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
MICROPROTEIN
ELITech Clinical Systems SAS
PLUS
REF: PRTU-XXXX

APPLICATION MICROLAB 300


PROPOSAL

For more details, see the Instructions for use :


and consult the analyzer manual.

Working temperature : 37°C

PREPARATION OF WORKING REAGENT: INSTRUMENT SETTINGS

The reagent is ready for use. GENERAL


NAME PRTU
UNITS mg/dL
R-BLANK YES
PROCEDURE:
S-BLANK NO
REPLICATION 1
Pipette 1000 µL of reagent and 20 µL of sample
SETTINGS
into a test tube.
Mix and after 5 minutes incubation at 37°C, mix MODE END-POINT
again before aspirate ,aspirate the mixing by the WAVELENGTH 620
analyser and start measurement. VOLUME 500
T-FACTOR 1.000
Note: Before use, the reagent should be brought DELAY 3
to the working temperature (37°C). BI-CHROM NO
WAVELENGTH 2 -
LIMITS
L-ABS **
H-ABS **
REF-LOW **
REF-HIGH **
DECIMALS 1
L-ABS-RB 0.040
H-ABS-RB 0.200
CALIBRATION
METHOD 1-POINT
NAME **
FACTOR 1
REPLICATION 3
DEVIATION **
L-ABS **
H-ABS **
STANDARDS
CON 1 #

Modification from the previous version ** Enter data by the operator.


# Enter standard value.

(03/2019)
AA-MICROLAB300-PRTU-3

Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
PHOSPHORUS
ELITech Clinical Systems SAS

REF: PHOS-XXXX

APPLICATION MICROLAB 300


PROPOSAL

For more details, see the Instructions for use :


and consult the analyzer manual.

Working temperature : 37°C

PREPARATION OF WORKING REAGENT): INSTRUMENT SETTINGS

The reagent is ready for use. GENERAL


NAME PHOS
UNITS mg/dL
R-BLANK YES
PROCEDURE:
S-BLANK NO
REPLICATION 1
Pipette 1000 µL of reagent and 10 µL of sample
SETTINGS
into a test tube.
Mix and after 5 minutes incubation at 37°C, MODE END-POINT
aspirate the mixing by the analyser and start WAVELENGTH 340
measurement. VOLUME 500
T-FACTOR 1.000
Note: Before use, the reagent should be brought DELAY 3
to the working temperature (37°C). BI-CHROM NO
WAVELENGTH 2 -
LIMITS
L-ABS **
H-ABS **
REF-LOW **
REF-HIGH **
DECIMALS 2
L-ABS-RB 0.050*
H-ABS-RB 0.350*
CALIBRATION
METHOD 1-POINT
NAME **
FACTOR 1
REPLICATION 3
DEVIATION **
L-ABS **
H-ABS **
STANDARDS
CON 1 #

* This value could be optimized.


** Enter data by the operator.
# Enter standard/calibrator value.
Modification from the previous version
(03/2019)
AA-MICROLAB300-PHOS-4

Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
TOTAL PROTEIN
ELITech Clinical Systems SAS
PLUS
REF: PROB-XXXX

APPLICATION MICROLAB 300


PROPOSAL

For more details, see the Instructions for use :


and consult the analyzer manual.

Working temperature : 37°C

PREPARATION OF WORKING REAGENT: INSTRUMENT SETTINGS

The reagent is ready for use. GENERAL


NAME PROB
UNITS g/dL
R-BLANK YES
PROCEDURE:
S-BLANK NO
REPLICATION 1
Pipette 1000 µL of reagent and 10 µL of sample
SETTINGS
into a test tube.
Mix and after 10 minutes incubation at 37°C, MODE END-POINT
aspirate the mixing by the analyser and start WAVELENGTH 546
measurement. VOLUME 500
T-FACTOR 1.000
Note: Before use, the reagent should be brought DELAY 3
to the working temperature (37°C). BI-CHROM NO
WAVELENGTH 2 -
LIMITS
L-ABS **
H-ABS **
REF-LOW **
REF-HIGH **
DECIMALS 2
L-ABS-RB 0.000*
H-ABS-RB 0.100*
CALIBRATION
METHOD 1-POINT
NAME **
FACTOR 1
Modification from the previous version REPLICATION 3
DEVIATION **
L-ABS **
H-ABS **
STANDARDS
CON 1 #

* This value could be optimized.


** Enter data by the operator.
# Enter standard/calibrator value.

(03/2019)
AA-MICROLAB300-PROB-3

Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
TRIGLYCERIDES
ELITech Clinical Systems SAS MONO SL NEW


REF: TGML-XXXX

APPLICATION MICROLAB 300


PROPOSAL

For more details, see the Instructions for use :


and consult the analyzer manual.

Working temperature : 37°C

PREPARATION OF WORKING REAGENT: INSTRUMENT SETTINGS

The reagent is ready for use. GENERAL


NAME TGML
UNITS mg/dL
R-BLANK YES
PROCEDURE:
S-BLANK NO
REPLICATION 1
Pipette 1000 µL of reagent and 10 µL of sample
SETTINGS
into a test tube.
Mix and after 10 minutes incubation at 37°C, MODE END-POINT
aspirate the mixing by the analyser and start WAVELENGTH 505
measurement. VOLUME 500
T-FACTOR 1.000
Note: Before use, the reagent should be brought DELAY 3
to the working temperature (37°C). BI-CHROM NO
WAVELENGTH 2 -
LIMITS
L-ABS **
H-ABS **
REF-LOW **
REF-HIGH **
DECIMALS 0
L-ABS-RB 0.000*
H-ABS-RB 0.200*
CALIBRATION
METHOD 1-POINT
NAME **
FACTOR 1
REPLICATION 3
DEVIATION **
L-ABS **
H-ABS **
STANDARDS
CON 1 #

* This value could be optimized.


** Enter data by the operator.
# Enter standard/calibrator value.
Modification from the previous version
(03/2019)
AA-MICROLAB300-TGML-4

Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
UREA UV SL
ELITech Clinical Systems SAS

REF: URSL-XXXX

APPLICATION MICROLAB 300


PROPOSAL

For more details, see the Instructions for use :


and consult the analyzer manual.

Working temperature : 37°C

PREPARATION OF WORKING REAGENT (WR): INSTRUMENT SETTINGS

Mix 4 volumes of reagent 1 with 1 volume of GENERAL


reagent 2. NAME URSL
UNITS mg/dL
R-BLANK NO
PROCEDURE:
S-BLANK NO
REPLICATION 1
Pipette 1000 µL of WR and 10 µL of sample into a
SETTINGS
test tube.
Mix, aspirate the mixing by the analyser and MODE KINETIC
start measurement. WAVELENGTH 340
VOLUME 500
T-FACTOR 1.000
Note: Before use, the working reagent WR should DELAY 30
be brought to the working temperature (37°C). MEAS-TIME 30
NON-LIN 20

LIMITS
L-ABS **
H-ABS **
REF-LOW **
REF-HIGH **
DECIMALS 1
CALIBRATION
METHOD 1-POINT
NAME **
FACTOR 1
REPLICATION 3
DEVIATION **
L-ABS **
H-ABS **
STANDARDS
CON 1 #

** Enter data by the operator.


# Enter standard/calibrator value.

Modification from the previous version


(03/2019)
AA-MICROLAB300-URSL-5

Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
URIC ACID
ELITech Clinical Systems SAS
MONO SL
REF: AUML-XXXX

APPLICATION MICROLAB 300


PROPOSAL

For more details, see the Instructions for use :


and consult the analyzer manual.

Working temperature : 37°C

PREPARATION OF WORKING REAGENT: INSTRUMENT SETTINGS

The reagent is ready for use. GENERAL


NAME AUML
UNITS mg/dL
R-BLANK YES
PROCEDURE:
S-BLANK NO
REPLICATION 1
Pipette 1000 µL of reagent and 25 µL of sample
SETTINGS
into a test tube.
Mix and after 5 minutes incubation at 37°C, MODE END-POINT
aspirate the mixing by the analyser and start WAVELENGTH 546
measurement. VOLUME 500
T-FACTOR 1.000
Note: Before use, the reagent should be brought DELAY 3
to the working temperature (37°C). BI-CHROM NO
WAVELENGTH 2 -
LIMITS
L-ABS **
H-ABS **
REF-LOW **
REF-HIGH **
DECIMALS 2
L-ABS-RB 0.000*
H-ABS-RB 0.100*
CALIBRATION
METHOD 1-POINT
NAME **
FACTOR 1
REPLICATION 3
DEVIATION **
L-ABS **
H-ABS **
STANDARDS
CON 1 #

* This value could be optimized.


** Enter data by the operator.
# Enter standard/calibrator value.
Modification from the previous version
(03/2019)
AA-MICROLAB300-AUML-5

Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
URIC ACID SL
ELITech Clinical Systems SAS
REF: AUSL-XXXx

APPLICATION MICROLAB 300


PROPOSAL

For more details, see the Instructions for use :


and consult the analyzer manual.

Working temperature : 37°C

INSTRUMENT SETTINGS

GENERAL
PREPARATION OF WORKING REAGENT (WR): NAME AUSL
UNITS mg/dL
Mix 4 volumes of reagent 1 with 1 volume of
R-BLANK YES
reagent 2.
S-BLANK NO
REPLICATION 1
PROCEDURE: SETTINGS
MODE END-POINT
Pipette 1000 µL of WR and 25 µL of sample into a WAVELENGTH 546
test tube. VOLUME 500
Mix and after 5 minutes incubation at 37°C, T-FACTOR 1.000
aspirate the mixing by the analyser and start DELAY 3
measurement. BI-CHROM NO
WAVELENGTH 2 -
LIMITS
Note: Before use, the working reagent WR should L-ABS **
be brought to the working temperature (37°C). H-ABS **
REF-LOW **
REF-HIGH **
DECIMALS 2
L-ABS-RB 0.000*
H-ABS-RB 0.250*
CALIBRATION
METHOD 1-POINT
NAME **
FACTOR 1
REPLICATION 3
DEVIATION **
L-ABS **
H-ABS **
STANDARDS
CON 1 #

* This value could be optimized.


** Enter data by the operator.
# Enter standard/calibrator value.
Modification from the previous version
(03/2019)
AA-MICROLAB300-AUSL-5

Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
CRP IP
ELITech Clinical Systems SAS

REF: ICRP-XXXX

APPLICATION MICROLAB 300


PROPOSAL

For more details, see the Instructions for use :


and consult the analyzer manual.

Working temperature : 37°C

INSTRUMENT SETTINGS

GENERAL
NAME ICRP
PREPARATION OF WORKING REAGENT :
UNITS mg/dL
Reagent 1: Ref: ICRP-6125 Ready to use R-BLANK NO
Reagent 2: Ref: ICRP-5025 Ready to use S-BLANK YES
REPLICATION 1
SETTINGS
CALIBRATION: MODE END-POINT
WAVELENGTH 340
CRP IP CALIBRATOR SET VOLUME 500
Ref: ICRP-0043 Ready to use. T-FACTOR 1.000
Add a zero point (NaCl 9 g/L solution). DELAY 2
BI-CHROM NO
WAVELENGTH 2 -
PROCEDURE:
LIMITS
L-ABS **
Sample blank:
H-ABS **
Pipette 320 µL of Reagent 1 and 20 µL of sample
into a test tube. REF-LOW **
Mix and after 1 minute incubation at 37°C, REF-HIGH **
aspirate the mixing by the analyser and start DECIMALS 2
measurement. CALIBRATION
METHOD MULTI-POINT
Sample: NAME **
Pipette 320 µL of Reagent 1, 32 µL of Reagent 2 NBR/ACCURACY 6/1%
and 20 µL of sample into a test tube. REPLICATION **
Mix and after 5 minutes of incubation at 37°C, DEVIATION **
aspirate the mixing by the analyser and start L-ABS **
measurement. H-ABS **
STANDARDS
CON 1 #
CON 2 #
CON 3 #
CON 4 #
Note: Before use, the reagents should be brought CON 5 #
to the working temperature (37°C). CON 6 #

** Enter data by the operator.


# Enter standard/calibrator value.
Modification from the previous version
(03/2019)
AA-MICROLAB300-ICRP-3

Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
HAPTOGLOBIN
ELITech Clinical Systems SAS
IP
REF: IHAP-XXXX

APPLICATION MICROLAB 300


PROPOSAL

For more details, see the Instructions for use :


and consult the analyzer manual.

Working temperature : 37°C

INSTRUMENT SETTINGS
PREPARATION OF WORKING REAGENT :
GENERAL
Reagent 1: Ref: IHAP-6125 Ready to use NAME IHAP
Reagent 2: Ref: IHAP-5025 Ready to use UNITS mg/dL
R-BLANK NO
CALIBRATION: S-BLANK YES
REPLICATION 1
PROTEIN IP CALIBRATOR SET SETTINGS
Ref: IPRO-0043 Ready to use. MODE END-POINT
Add a zero point (NaCl 9 g/L solution). WAVELENGTH 340
VOLUME 500
T-FACTOR 1.000
PROCEDURE: DELAY 2
BI-CHROM NO
Sample blank: WAVELENGTH 2 -
Pipette 840 µL of Reagent 1 and 6 µL of sample
LIMITS
into a test tube.
L-ABS **
Mix and after 1 minute incubation at 37°C,
H-ABS **
aspirate the mixing by the analyser and start
REF-LOW **
measurement.
REF-HIGH **
Sample: DECIMALS 0
Pipette 840 µL of Reagent 1, 120 µL of Reagent 2 CALIBRATION
and 6 µL of sample into a test tube. METHOD MULTI-POINT
Mix and after 5 minutes of incubation at 37°C, NAME **
aspirate the mixing by the analyser and start NBR/ACCURACY 6/1%
measurement. REPLICATION **
DEVIATION **
L-ABS **
H-ABS **
Note: Before use, the reagents should be brought STANDARDS
to the working temperature (37°C). CON 1 #
CON 2 #
CON 3 #
CON 4 #
CON 5 #
CON 6 #
** Enter data by the operator.
# Enter standard/calibrator value.
Modification from the previous version
(03/2019)
AA-MICROLAB300-IHAP-3

Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
IgA IP
ELITech Clinical Systems SAS

REF: IIGA-XXXX

APPLICATION MICROLAB 300


PROPOSAL

For more details, see the Instructions for use :


and consult the analyzer manual.

Working temperature : 37°C

INSTRUMENT SETTINGS

GENERAL
PREPARATION OF WORKING REAGENT : NAME IIGA
UNITS mg/dL
Reagent 1: Ref: IIGB-6125 Ready to use
R-BLANK NO
Reagent 2: Ref: IIGA-5025 Ready to use
S-BLANK YES
REPLICATION 1
SETTINGS
CALIBRATION: MODE END-POINT
WAVELENGTH 340
PROTEIN IP CALIBRATOR SET VOLUME 500
Ref: IPRO-0043 Ready to use. T-FACTOR 1.000
Add a zero point (NaCl 9 g/L solution). DELAY 2
BI-CHROM NO
WAVELENGTH 2 -
PROCEDURE:
LIMITS
Sample blank: L-ABS **
Pipette 900 µL of Reagent 1 and 5 µL of sample H-ABS **
into a test tube. REF-LOW **
Mix and after 1 minute incubation at 37°C, REF-HIGH **
aspirate the mixing by the analyser and start DECIMALS 0
measurement. CALIBRATION
METHOD MULTI-POINT
Sample: NAME **
Pipette 900 µL of Reagent 1, 150 µL of Reagent 2 NBR/ACCURACY 6/1%
and 5 µL of sample into a test tube. REPLICATION **
Mix and after 5 minutes of incubation at 37°C, DEVIATION **
aspirate the mixing by the analyser and start L-ABS **
measurement. H-ABS **
STANDARDS
CON 1 #
Note: Before use, the reagents should be brought CON 2 #
to the working temperature (37°C). CON 3 #
CON 4 #
CON 5 #
CON 6 #

** Enter data by the operator.


# Enter standard/calibrator value.
Modification from the previous version
(03/2019)
AA-MICROLAB300-IIGA-3

Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
IgG IP
ELITech Clinical Systems SAS

REF: IIGG-XXXX

APPLICATION MICROLAB 300


PROPOSAL

For more details, see the Instructions for use :


and consult the analyzer manual.

Working temperature : 37°C

INSTRUMENT SETTINGS

GENERAL
NAME IIGG
PREPARATION OF WORKING REAGENT :
UNITS mg/dL
Reagent 1: Ref: IIGB-6125 Ready to use R-BLANK NO
Reagent 2: Ref: IIGG-5025 Ready to use S-BLANK YES
REPLICATION 1
CALIBRATION: SETTINGS
MODE END-POINT
PROTEIN IP CALIBRATOR SET WAVELENGTH 340
Ref: IPRO-0043 Ready to use. VOLUME 250
Add a zero point (NaCl 9 g/L solution). T-FACTOR 1.000
DELAY 2
BI-CHROM NO
PREPARATION OF SAMPLES AND CONTROLS WAVELENGTH 2 -
LIMITS
Dilute samples and controls at 1/10 in NaCl L-ABS **
9g/L solution. H-ABS **
REF-LOW **
PROCEDURE: REF-HIGH **
DECIMALS 0
Sample blank: CALIBRATION
Pipette 500 µL of Reagent 1 and 10 µL of diluted METHOD MULTI-POINT
sample into a test tube.
NAME **
Mix and after 1 minute incubation at 37°C,
NBR/ACCURACY 6/1%
aspirate the mixing by the analyser and start
REPLICATION **
measurement.
DEVIATION **
Sample: L-ABS **
Pipette 500 µL of Reagent 1, 50 µL of Reagent 2 H-ABS **
and 10 µL of diluted sample into a test tube. STANDARDS
Mix and after 5 minutes of incubation at 37°C, CON 1 #
aspirate the mixing by the analyser and start CON 2 #
measurement. CON 3 #
CON 4 #
CON 5 #
Note: Before use, the reagents should be brought CON 6 #
to the working temperature (37°C).
** Enter data by the operator.
Modification from the previous version # Enter standard/calibrator value.

(03/2019)
AA-MICROLAB300-IIGG-3

Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
IgM IP
ELITech Clinical Systems SAS

REF: IIGM-XXXX

APPLICATION MICROLAB 300


PROPOSAL

For more details, see the Instructions for use :


and consult the analyzer manual.

Working temperature : 37°C

PREPARATION OF WORKING REAGENT : INSTRUMENT SETTINGS

Reagent 1: Ref: IIGB-6125 Ready to use GENERAL


Reagent 2: Ref: IIGM-5025 Ready to use NAME IIGM
UNITS mg/dL
R-BLANK NO
CALIBRATION:
S-BLANK YES
PROTEIN IP CALIBRATOR SET REPLICATION 1
Ref: IPRO-0043 Ready to use. SETTINGS
Add a zero point (NaCl 9 g/L solution). MODE END-POINT
WAVELENGTH 340
VOLUME 500
PROCEDURE: T-FACTOR 1.000
DELAY 2
Sample blank: BI-CHROM NO
Pipette 900 µL of Reagent 1 and 5 µL of sample WAVELENGTH 2 -
into a test tube. LIMITS
Mix and after 1 minute incubation at 37°C,
L-ABS **
aspirate the mixing by the analyser and start
H-ABS **
measurement.
REF-LOW **
REF-HIGH **
Sample:
DECIMALS 0
Pipette 900 µL of Reagent 1, 130 µL of Reagent 2
and 5 µL of sample into a test tube. CALIBRATION
Mix and after 5 minutes of incubation at 37°C, METHOD MULTI-POINT
aspirate the mixing by the analyser and start NAME **
measurement. NBR/ACCURACY 6/1%
REPLICATION **
DEVIATION **
L-ABS **
Note: Before use, the reagents should be brought H-ABS **
to the working temperature (37°C). STANDARDS
CON 1 #
CON 2 #
CON 3 #
CON 4 #
CON 5 #
CON 6 #

** Enter data by the operator.


# Enter standard/calibrator value.
Modification from the previous version
(03/2019)
AA-MICROLAB300-IIGM-3

Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
µALBUMIN IP
ELITech Clinical Systems SAS

REF: IMAL-XXXX

APPLICATION MICROLAB 300


PROPOSAL

For more details, see the Instructions for use :


and consult the analyzer manual.

Working temperature : 37°C

INSTRUMENT SETTINGS

GENERAL
NAME IMAL
PREPARATION OF WORKING REAGENT :
UNITS mg/dL
Reagent 1: Ref: IMAL-6125 Ready to use R-BLANK NO
Reagent 2: Ref: IMAL-5025 Ready to use S-BLANK YES
REPLICATION 1
SETTINGS
CALIBRATION: MODE END-POINT
WAVELENGTH 340
µALBUMIN IP CALIBRATOR SET VOLUME 250
Ref: IMAL-0043 Ready to use. T-FACTOR 1.000
Add a zero point (NaCl 9 g/L solution). DELAY 2
BI-CHROM NO
WAVELENGTH 2 -
PROCEDURE:
LIMITS
Sample blank: L-ABS **
Pipette 300 µL of Reagent 1 and 20 µL of sample H-ABS **
into a test tube. REF-LOW **
Mix and after 1 minute incubation at 37°C, REF-HIGH **
aspirate the mixing by the analyser and start DECIMALS 2
measurement. CALIBRATION
METHOD MULTI-POINT
Sample: NAME **
Pipette 300 µL of Reagent 1, 50 µL of Reagent 2 NBR/ACCURACY 6/1%
and 20 µL of sample into a test tube. REPLICATION **
Mix and after 5 minutes of incubation at 37°C, DEVIATION **
aspirate the mixing by the analyser and start L-ABS **
measurement. H-ABS **
STANDARDS
CON 1 #
CON 2 #
Note: Before use, the reagents should be brought CON 3 #
to the working temperature (37°C).
CON 4 #
CON 5 #
CON 6 #

** Enter data by the operator.


# Enter standard/calibrator value.
Modification from the previous version
(03/2019)
AA-MICROLAB300-IMAL-5

Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
OROSOMUCOID IP
ELITech Clinical Systems SAS

REF: IORO-XXXX

APPLICATION MICROLAB 300


PROPOSAL

For more details, see the Instructions for use :


and consult the analyzer manual.

Working temperature : 37°C

INSTRUMENT SETTINGS

GENERAL
PREPARATION OF WORKING REAGENT : NAME IORO
UNITS mg/dL
Reagent 1: Ref: IORO-6125 Ready to use
R-BLANK NO
Reagent 2: Ref: IORO-5025 Ready to use
S-BLANK YES
REPLICATION 1
CALIBRATION: SETTINGS
MODE END-POINT
PROTEIN IP CALIBRATOR SET WAVELENGTH 340
Ref: IPRO-0043 Ready to use. VOLUME 500
Add a zero point (NaCl 9 g/L solution). T-FACTOR 1.000
DELAY 2
BI-CHROM NO
PROCEDURE:
WAVELENGTH 2 -
LIMITS
Sample blank:
Pipette 750 µL of Reagent 1 and 5 µL of sample L-ABS **
into a test tube. H-ABS **
Mix and after 1 minute incubation at 37°C, REF-LOW **
aspirate the mixing by the analyser and start REF-HIGH **
measurement. DECIMALS 0
CALIBRATION
Sample: METHOD MULTI-POINT
Pipette 750 µL of Reagent 1, 100 µL of Reagent 2 NAME **
and 5 µL of sample into a test tube. NBR/ACCURACY 6/1%
Mix and after 5 minutes of incubation at 37°C, REPLICATION **
aspirate the mixing by the analyser and start DEVIATION **
measurement. L-ABS **
H-ABS **
STANDARDS
CON 1 #
Note: Before use, the reagents should be brought CON 2 #
to the working temperature (37°C). CON 3 #
CON 4 #
CON 5 #
CON 6 #

** Enter data by the operator.


# Enter standard/calibrator value.
Modifcation from the previous version
(03/2019)
AA-MICROLAB300-IORO-3

Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
PREALBUMIN IP
ELITech Clinical Systems SAS

REF: IPAL-XXXX

APPLICATION MICROLAB 300


PROPOSAL

For more details, see the Instructions for use :


and consult the analyzer manual.

Working temperature : 37°C

INSTRUMENT SETTINGS

GENERAL
PREPARATION OF WORKING REAGENT : NAME IPAL
UNITS mg/dL
Reagent 1: Ref: IPAL-6125 Ready to use
R-BLANK NO
Reagent 2: Ref: IPAL-5025 Ready to use
S-BLANK YES
REPLICATION 1
CALIBRATION: SETTINGS
MODE END-POINT
PROTEIN IP CALIBRATOR SET WAVELENGTH 340
Ref: IPRO-0043 Ready to use. VOLUME 250
Add a zero point (NaCl 9 g/L solution). T-FACTOR 1.000
DELAY 2
BI-CHROM NO
PROCEDURE:
WAVELENGTH 2 -
LIMITS
Sample blank:
Pipette 550 µL of Reagent 1 and 5 µL of sample L-ABS **
into a test tube. H-ABS **
Mix and after 1 minute incubation at 37°C, REF-LOW **
aspirate the mixing by the analyser and start REF-HIGH **
measurement. DECIMALS 0
CALIBRATION
Sample: METHOD MULTI-POINT
Pipette 550 µL of Reagent 1, 50 µL of Reagent 2 NAME **
and 5 µL of sample into a test tube. NBR/ACCURACY 6/1%
Mix and after 5 minutes of incubation at 37°C, REPLICATION **
aspirate the mixing by the analyser and start DEVIATION **
measurement. L-ABS **
H-ABS **
STANDARDS
CON 1 #
Note: Before use, the reagents should be brought CON 2 #
to the working temperature (37°C). CON 3 #
CON 4 #
CON 5 #
CON 6 #

** Enter data by the operator.


# Enter standard/calibrator value.
Modification from the previous version
(03/2019)
AA-MICROLAB300-IPAL-3

Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.
TRANSFERRIN IP
ELITech Clinical Systems SAS

REF: ITRF-XXXX

APPLICATION MICROLAB 300


PROPOSAL

For more details, see the Instructions for use :


and consult the analyzer manual.

Working temperature : 37°C

INSTRUMENT SETTINGS

GENERAL
PREPARATION OF WORKING REAGENT : NAME ITRF
UNITS mg/dL
Reagent 1: Ref: ITRF-6125 Ready to use
R-BLANK NO
Reagent 2: Ref: ITRF-5025 Ready to use
S-BLANK YES
REPLICATION 1
CALIBRATION: SETTINGS
MODE END-POINT
PROTEIN IP CALIBRATOR SET
WAVELENGTH 340
Ref: IPRO-0043 Ready to use.
VOLUME 500
Add a zero point (NaCl 9 g/L solution).
T-FACTOR 1.000
DELAY 2
PREPARATION OF SAMPLES AND CONTROLS BI-CHROM NO
WAVELENGTH 2 -
Dilute samples and controls at 1/20 in NaCl LIMITS
9g/L solution. L-ABS **
H-ABS **
PROCEDURE: REF-LOW **
REF-HIGH **
Sample blank: DECIMALS 0
Pipette 880 µL of Reagent 1 and 12 µL of diluted CALIBRATION
sample into a test tube. METHOD MULTI-POINT
Mix and after 1 minute incubation at 37°C,
NAME **
aspirate the mixing by the analyser and start
NBR/ACCURACY 6/1%
measurement.
REPLICATION **
DEVIATION **
Sample:
Pipette 880 µL of Reagent 1, 40 µL of Reagent 2 L-ABS **
and 12 µL of diluted sample into a test tube. H-ABS **
Mix and after 5 minutes of incubation at 37°C, STANDARDS
aspirate the mixing by the analyser and start CON 1 #
measurement. CON 2 #
CON 3 #
CON 4 #
CON 5 #
Note: Before use, the reagents should be brought CON 6 #
to the working temperature (37°C).
** Enter data by the operator.
# Enter standard/calibrator value.
Modification from the previous version
(03/2019)
AA-MICROLAB300-ITRF-3

Parameters are based on experience and knowledge of the Microlab300 and using ELITech reagents.
It is recommended each laboratory validate the procedure, prior to routine use, to determine suitability for the patient population served.
For further information consult the reagent IFU.
For technical support or assistance, please contact your local ELITech distributor.