Automated Peritoneal: - The System You Can Trust
Automated Peritoneal: - The System You Can Trust
Power Up the Machine, when await for the display to show "Press Go to start" Back Panel Display Screen Heater
operation manual
change your options
Press Display shows:"Total Vol : XXXX ml"
press
Temperature Sensor Button
Press Display shows: "Therapy time : : " 2. p r e s s t h e or
Press Display shows: "Fill Vol : XXXX ml" for change
options and numbers.
Serial Number
–The system you can trust
Press Display shows: "The last fill vol : XXXX ml"
3. After changes, press Up/Down Buttons
Press Display shows: "Dextrose : same / different" the to confirm. Door
Press Automatically display shows "Cycles : XX" Handle
"Dwell time : : "
Alarm and Alarm issues Enter Button (Blue)
When the machine runs through something abnormal, it will alarm and stop filling. Then the error Stop Button (Red)
message appears on the display. When there is a continuous beeping sound:
Go Button (Green)
1. Press the red key to silence the beeping sound
2. If an error message appears on the display, please refer to in-home guidance manual
and go to the warnings chapter to deal with the problem. Disposable Set Configuration
3. After the problem is solved, press the Green Key and the machine will continue
treatment. Drain Line Cassette
Heater Line (Red Clamp)
Handling Power Failures
Supply Lines (White Clamps)
When power failure occurs (or if the power switch is turned off) during set-up, the HomeChoice
System will close the occluder and the display will turn off. Patient Line
When the power is restored, the HomeChoice System will return to "PRESS GO TO START".
● If the power is restored within 30 minutes: Final Line (Blue Clamp)
1. The system will automatically resume therapy without any alarms.
2. During the first 30 minutes of a power, you can press any key to retrieve Organizer
therapy information. If you do so, you will have to Press and to start
therapy after power is restored.
● If the power is not restored within 30 minutes:
1. An alarm will occur after 30 minutes without power. Press to mute the alarm.
Lines
2. Retrieve therapy information by Press
● If power is restored within approximately two hours:
1. The therapy can be restarted from where it left off. An alarm will occur again when
power is restored.
2. Press and to restart therapy. 百特醫療產品股份有限公司
● If power is not restored within two hours, you must End Therapy.
台北市106敦化南路二段216號15樓
You can then restart your therapy from the beginning, if needed.
腹膜透析諮詢專線:0800-009989
2007-08-HC-1-英
Automated peritoneal dialysis machine
Characteristics
● Proven reliability
Press Display show "self testing…" Please clean your
3 Therapy Completed
● Demonstrated acceptance-over 500,000 hands again.When "self testing…" is completed, the display will
patients treated The next morning after the completion of
change to "Connect bags"
● Simplified training and ease of use therapy, the diaplay will show that the "End of
1. Remove (red clamp) line from the organizer and connect it
● Customized PD therapy for specific to solution bag (in the heating pan), and then snap green therapy" (See figure 6)
patient needs frangible and open red clamp.(See figure 3) Press Display will show "close all clamps"
2. Remove (white clamp) line from the organizer and connect it 1. Close the Transfer Set.
Automated peritoneal dialysis machine operation steps
to supply bag and then snap green frangible and open white
3 clamp. 2. Close the patient line clamp.
6
If you are using different dextrose solution,please connect
(blue clamp) line to solution bag. Snap green frangible and
0 Before on machine then open the blue clamp. Press Display will show "Disconnect yourself".
1. Preparation Materials 3. Open patient line clamp 1. Wear a face mask and make sure hands are
clean and dry.
● Fresh dialysis solution
2. Open a new Minicap, disconnect patient line
● Drain waste bucket (or extension tube) tube and mount Minicap on the Transfer Set
● Disposal set ● Face mask ● Minicaps*2 Press Display show "PRIMING",please clean your immediately. (See figure 7)
● Alcohol free hand cleaning gel hands again.When "PRIMING" is completed,the display will
change to "Connect yourself" and "check patient line" Press When the display shows "Turn me off", please
2. Clean table surface record the following information on“APD daily
Prior to connect yourself, 7
*Sanitize (using only 75% alcohol) the machine log”.
before power on. 1. Make sure the patient line is fully primed (up to the ring cap)
ad then close clamp.
4 Records
1 When Using the Machine. 2. Remove the patient line and connect it to Transfer Set and luer
lock it until is completely closed. (See figure 4)
Turn on the power. Await for diaplay to show "Press GO to start". 4 3. Open patient line clamp and the Transfer set. Press I-Drain Vol : XXXX ML (See figure 8)
1. Place the dialysis solution on the heating pan. Make sure the (Please record the value on the diaplay shows)
dialysis fluid must be covered the temperature sensor, and
leave the other ones on adjustment of the machine.(See figure 1) Press Total UF : XXXX ML
2. Make sure a face mask is worn and hands are washed.
(please record the value on the display shows)
1
3. Check up on the dialysis fluid capacity, concentration, expired
date, export ring, green frangible and appearance of dialysis
2 Starting therapy
fluid. Press Ave dwell time : XX : XX (hours : minutes)
4. Inspect the disposal sets to make sure the protective caps are 8 (press record the value on the display shows)
intact and that all the clamps are closed. Press Display will show "Initial drain" (See figure 5)