4008B Operacion Ingles 16 10 03 v52
4008B Operacion Ingles 16 10 03 v52
Hemodialysis Machine
Operating Instructions
0123
Caution!
! These Operating Instructions in pdf-
format are for information only.
They are not a replacement for the
Operating Instructions supplied with
the machine/device and options.
Part No.: M34 543 1 Fresenius Medical Care 4008 B 16/10.03 (OP) 0-1
0-2 Fresenius Medical Care 4008 B 16/10.03 (OP)
OPERATING INSTRUCTIONS
Hemodialysis Machine 4008 B
The present 4008 B Operating Instructions include the description of the profile function. 4008 B machines
are also available without profile function. Those parts of the Operating Instructions which are not
applicable for 4008 B machines without profile function have been marked as such.
These Operating Instructions are part of the machine documentation and thus an integral part of the
hemodialysis machine. They contain any information necessary for the use of the 4008 B hemodialysis
machine. All operators must have been thoroughly trained by the manufacturer and must have read the
entire Operating Instructions manual before using the machine.
The machine may only be operated by persons who have been instructed on the proper operation and
handling of the unit.
The Operating Instructions must be carefully studied before attempting to operate the machine.
The Brief Operating Instructions are intended as a reference only and do not replace the Operating
Instructions manual with which the operator must be fully familiar.
☞ Note
Informs the operator that in case of a
failure to follow the steps as de-
scribed, a specific function will be
executed incorrectly or will not be
executed at all, or will not produce
the desired effect.
Caution
Advises the operator against certain
procedures or actions that could
cause damage to the equipment or
may have adverse effects on opera-
tors and patients.
The parameters entered must be verified by the operator, i.e. the operator must check the values entered
for correctness. Should the desired value deviate from the parameters displayed, the setting must be
corrected before activating the function.
The actual values displayed must be compared with the desired values specified!
The 4008 B hemodialysis machine reflects the latest state of technology and complies with the require-
ments of EN 60601 (IEC 601-1).
Assembly, installation, adjustment, modification or repair are to be performed only by the manufacturer or
persons authorized by him.
Local Service:
Chapter Page
2 Preparation .................................................................................................................... 2-
2.1 Hemodialysis Machine .................................................................................................... 2-3
2.2 Functional Test ............................................................................................................... 2-12
2.3 Blood Pump .................................................................................................................... 2-25
2.4 Heparin Pump ................................................................................................................. 2-27
2.5 Air Detector ..................................................................................................................... 2-31
2.6 Extracorporeal Blood Circuit ........................................................................................... 2-33
2.7 Priming the Extracorporeal Blood Circuit With a Saline Bag ......................................... 2-35
2.8 Dialysate Side ................................................................................................................. 2-40
3 Treatment ....................................................................................................................... 3-
3.1 Ultrafiltration .................................................................................................................... 3-3
3.2 Double Needle Dialysis ................................................................................................... 3-11
3.3 Single Needle Dialysis .................................................................................................... 3-16
3.4 Single Needle Click-Clack .............................................................................................. 3-21
3.5 Isolated Ultrafiltration (ISO-UF) ...................................................................................... 3-24
3.6 Sodium and UF Profiles .................................................................................................. 3-31
3.7 DIASAFE ®plus (option) ..................................................................................................... 3-41
5 Disconnection ............................................................................................................... 5-
5.1 End of Treatment ............................................................................................................ 5-3
5.2 Reinfusion ....................................................................................................................... 5-6
5.3 Stopping the Treatment .................................................................................................. 5-10
5.4 Emptying the bibag® (option) .......................................................................................... 5-17
5.5 Emptying the Dialyzer ..................................................................................................... 5-19
6 Cleaning ......................................................................................................................... 6-
6.1 General Overview ........................................................................................................... 6-3
6.2 External Cleaning ........................................................................................................... 6-5
6.3 Basic Conditions for Cleaning Programs ........................................................................ 6-5
6.4 Disinfection and Decalcification ...................................................................................... 6-6
6.5 Cleaning / Degreasing Program ..................................................................................... 6-13
6.6 Hot Rinse ........................................................................................................................ 6-18
6.7 Rinsing ............................................................................................................................ 6-21
6.8 Priming During Cleaning Programs ................................................................................ 6-23
6.9 Precirculation During the Cleaning Programs ................................................................. 6-27
Chapter Page
The 4008 B dialysis machine is designed for performing chronic and acute hemodialysis. It can be
used in home dialysis, limited care centers and clinical hemodialysis.
Dialysis therapy occasionally causes hypotension, nausea, vomiting and cramping in some
patients. Please take note of the package inserts enclosed with the hemodialysis concentrates,
dialyzers, etc.
1.1.3 Contraindications
The hemodialysis machine allows dialysis treatments without any additional equipment. The
machine operates and monitors the dialysate circuit and the extracorporeal blood circuit.
In the dialysate circuit, purified water is mixed with hemodialysis concentrate, heated, degassed
and delivered to the dialyzer. Inflowing and outflowing quantities are balanced volumetrically. The
pressure at the dialyzer is adjusted depending on the ultrafiltration rate selected and the type of
dialyzer used.
The DIASAFE®plus option permits hemodialysis treatment with high-purity dialysate. The
DIASAFE®plus option basically consists of one filter stage. For the entire time of its use the filter is
an integral part of the hemodialysis machine and is rinsed and disinfected in conjunction with the
machine. The integrity of the filter is tested by means of a pressure holding test which is
integrated in the functional test of the hemodialysis machine. The filter can be used for a
maximum period of 12 weeks. Should the filter fail to pass the filter test, it must be replaced at an
earlier date. For the use of this option, the quality of the water and the dialysate must comply with
the applicable standards. (See Chapter 1.6 Initial Start-Up.)
In the extracorporeal blood circuit, the blood is continuously heparinized and passes through the
dialyzer. An air detector prevents infusion of air . Any dangerous loss of blood is prevented by a
blood leak detector and a device monitoring the venous return pressure. The arterial pressure
monitoring unit detects any intake of the needle in the vessel.
The hemodialysis machine is designed for both acetate dialysis and bicarbonate dialysis.
The mixing ratio of concentrate to purified water can be programmed (default setting 1+34 –
selection of other mixing ratios is possible).
Depending on the concentrate used, the Na+ concentration can be readjusted within a range of 125
to 150 mmol.
With the bibag® option the bicarbonate solution is produced from a bag (sodium hydrogen
carbonate powder, NaHCO3). The powder is diluted in the machine to obtain ready-to-use
bicarbonate.
Na+ and UF profiles can be programmed by using the built-in variation function.
The basic version of the hemodialysis machine is designed for double needle dialysis.
Freely selectable programs are available for cleaning and disinfecting the machine.
The hemodialysis machine is equipped with all safety systems required for proper function and
patient safety EN 60601-1 (IEC 601-1).
The automatic test of the safety systems (functional test, also called T1 test) must be performed
prior to each dialysis treatment. This test also checks the integrity of the closed system (pressure
holding test PHT).
In the following cases, the hemodialysis machine will request the user to perform the functional test:
– after a start with external power supply (not power failure),
– after one of the cleaning programs.
For additional safety, cyclic pressure holding tests (CPHT) are also performed during the dialysis
treatment to detect possible leakages in the closed system.
The CPHT is performed automatically every 12.5 minutes. For the CPHT the dialysis machine is
switched for a few seconds into bypass and ultrafiltration is stopped. Balancing is not affected by
these CPHTs, but the effective dialysis time will be reduced by approx. 2 minutes per hour of
treatment.
Caution
For any additional equipment connected to the analog and digital interfaces of
the machine, evicence must be provided that this equipment complies with the
applicable EN specifications (e.g. EN 60950 for data processing equipment, and
EN 60601 (IEC 601) for electro-medical equipment). In addition, all configura-
tions must comply with the system standard EN 60601-1-1 (IEC 601-1-1).
Anyone connecting additional equipment to the parts of signal input or signal
output component configurates the system and is, thus, responsible for compli-
ance with the system standard EN 60601-1-1 (IEC 601-1-1).
These accessories have been approved within the scope of an EC prototype test.
1.2.1 Transportation
● Inside buildings
● Outside buildings
– Never push the hemodialysis machine across an uneven paving (e.g. cobblestone pavement).
Always lift it.
– When transporting the hemodialysis machine in vehicles, accommodate it vertically or hori-
zontally and protect it with an appropriate padding.
– When transporting the hemodialysis machine outside of buildings for a prolonged period of
time, the range of storage temperature must be observed (if necessary, fill the machine with
antifreeze).
The hemodialysis machine must be stored vertically in a well-aired room with low variations in
temperature.
Storage temperature:
– without antifreeze +5 °C to +70 °C
– with antifreeze –20 °C to +70 °C
Relative humidity:
30 % to 75 %, temporarily 95 %
Should the hemodialysis machine be stored with antifreeze, the antifreeze used should have the
following composition:
– 48.75 % water
– 48.75 % glycerine
– 2.5 % Teta-plus
I/O
Luftdetektor Arterieller Venöser TMP Ultrafiltration Leitfähigkeit
Druck Druck
UF Menge ml
Spülen Reset
Konz. Bic.
kPa mmHg kPa mmHg kPa mmHg
9
Blutleck 30 60 60 UF Rate ml/h 15.5
200
Desinfektion
20 50 400 50 400 15
10
100 40 300 40 300 14.5
0 0 30 30
Single 200 200 14
Needle –10
–100 20 20
UF Ziel ml
–20 10
100 10
100 13.5
Überbrücken
–200
Test
–30 0 0 0 0 13
–40 –300 mS/cm
(25°C)
Vorbereiten
UF Restzeit h:min
Variation
I/O Prog.
Dialyse Start
Tempera-
UF Fluß tur
Pven. 8
Start Start
Stop Stop
2
Rate: ml/h
(Bolus: ml)
( :h.min)
Bolus
Rate
Start
Stop
2a
7
3 6
Legend
1 Monitor 3 Hydraulics
2 Modules – Level 1 4 Brake
(from left to right) 5 Bicarbonate suction tube (blue)
Blood pump (arterial) 6 Concentrate suction tube (red)
Blood pump (SN) 7 Rinse chamber
Air detector 8 Interlock shunt for dialyzer connecting tubes
2a Modules – Level 2 9 IV pole
(from left to right)
Heparin pump
Rinse chamber, bibag® connector (option)
not used
3 14
4
5
15
13
12
6 11
10
9
8
Legend
20b
Fresenius Medical Care 4008 B 20c 20d
20a
Confirm Select
19 20
19a 19b 19c
1 1a I/O 10 12 13 14 15 16
Air Detector Arterial Venous Transmembrane Ultrafiltration Conductivity
1b Pressure Pressure Pressure
UF Volume ml
15b 16a
10a 12a 13a 14a Reset
2 2a Rinse 15a Volume Conc Bic
2b kPa mmHg kPa mmHg kPa mmHg
16c 16d
3 3a Hotrinse 11 280 500 500 15d
Blood Leak 30 UF Rate ml/h 15.5
3b 200 60 60
20b
Fresenius Medical Care 4008 B 20c 20d
20a
Confirm Select
19 20
19a 19b 19c
1 1a I/O 10 12 13 14 15 16
Air Detector Arterial Venous Transmembrane Ultrafiltration Conductivity
1b Pressure Pressure Pressure
UF Volume ml
15b 16a
10a 12a 13a 14a Reset
2 2a Rinse 15a Volume
2b kPa mmHg kPa mmHg kPa mmHg
mS/cm
(25°C)
1 2 3 4 5 6 7 8 9 10 11
15
16
17
14 13 12
Caution
Apart from the options mentioned on page 1-8, additional equipment should only
be connected to the ports on the rear of the monitor, if the resulting overall
system complies with the requirements of EN 60601-1-1 (IEC 601-1), or if their
applicability with regard to technical safety has been proven by a specific
certificate issued by a testing agency authorized to test the ready-for-use
machine.
1 Fan
2 Service switch
3 Alarm tone volume control
4 Connection socket for traffic light / status indicator (optional)
5 RS232 interface (galvanically decoupled by optocoupler) (optional)
6 Input / output port (for external equipment)
7 Connection socket for screen
8 Operating hours indicator
9 ALARM IN port (can be used to connect an external pushbutton. This pushbutton can be
used to initiate a machine stop.)
10 ALARM OUT port (can be used to connect an external alarm indicator, nurse call)
11 Symbols. Caution: consult accompanying documents.
12 Power supply unit (voltage supply for the hemodialysis machine)
13 Power switch (main switch)
14 Power cord (to supply, the hemodialysis machine with line voltage, 230 V AC; 50 – 60 Hz)
15 Loudspeaker for audible alarm
16 Port for connection of the 4008 Remote Control (optional)
17 Key interface (patient identification in conjunction with FINESSE) (optional)
1 Rate: ml/min
(Ø: mm)
2
11
3
10
4 Start
Stop 9
1 Display (indicates the delivery rate, the line diameter or an error code)
2 Alarm LED (red)
3 Operation LED (green)
4 Pressure connector (Luer-lock connector of the arterial pressure measuring line)
5 Clamping piece (holds the installed line segment securely in place)
6 Blood pump cover
7 Rotor (delivers the patient’s blood by means of the rollers and the lines installed; the rotor
can be removed for cleaning)
8 Sensor (blood pump cover open or closed)
9 Start/Stop key (to turn the blood pump on and off)
10 ▼ key (to reduce the delivery rate or the line diameter)
11 ▲ key (to increase the delivery rate or the line diameter)
12
Rate: ml/h
(Bolus: ml)
1
11
( :h.min)
10
2 Bolus 9
3 Rate 8
Start
4 Stop 7
Pven. 1
1 Keys (to raise (▲) and lower (▼) the fluid level in the venous bubble catcher)
2 Venous pressure connector (Luer-lock connector of the venous pressure measuring line)
3 Holder for the venous bubble catcher with ultrasonic sensors
4 Venous line clamp
5 Optical detector
The float switch (5) located in the heater block (66) controls the inflow of fresh water via the water
inlet valve (41).
The upstream pressure reducing valve (61) ensures a constant water inlet pressure.
In the chamber (88c) of the multifunction block (88), the concentrate pump (23) adds acetate-
containing concentrate. During bicarbonate dialysis, it adds acid concentrate.
When peforming bicarbonate dialysis the bicarbonate is pumped by the membrane pump (25)
from the bicarbonate suction tube (95) via the air separation chamber (131a) into the chamber
(88c) of the multifunction block (88) adding exact quantities of bicarbonate. The conductivity
measuring cell (132) and the temperature sensor (133) measure and monitor the bicarbonate
conductivity.
The heat exchanger (77) transmits part of the heat of the outflowing dialysate to the cold inflowing
water.
The water is heated in the heater rod chamber (66b). The degassing pump (29) and the
degassing orifice (89) are used to generate a negative pressure to deaerate the water. The air is
collected in the primary air separator. From there, it is directed to the heater block (66) via the
loading pressure valve (65) and then discharged into the atmosphere via the float chamber (66c).
The dialysate flows from the chamber (88c) via the mixing chamber (131b) to the balancing
chamber (68) from where it passes through the conductivity measuring cell (7) and the dialyzer
valve (24) to the dialyzer.
In the conductivity cell (7), both the conductivity and the temperature (temperature sensor 3) are
measured and indicated on the monitor.
If temperature and conductivity are inside the alarm limits, the two dialyzer valves (24 and 24b)
are open and the bypass valve (26) is closed.
Should one of the two values be outside the alarm limits (temperature or CD alarm), the bypass
valve (26) opens and the dialyzer valve (24) closes.
The used dialysate coming from the dialyzer is monitored by a blood leak detector (8), from where
it flows into the secondary air separator (88b) of the multifunction block (88). The fluid coming
from this chamber is forced by the flow pump (21) into the balancing chamber (68). This ensures
that the dialysate that is expelled is always replaced by the same amount of fresh dialysate.
The pressure transducer (9) is located below the secondary air separator (88b) of the multifunc-
tion block (88). The pressure measured here is taken as a quantity in the computation of the TMP.
A membrane pump (UF pump 22) with a volume of 1 ml per stroke withdraws fluid from the
system at a specified rate. Since the system is closed, the same volume is removed as ultrafiltrate
from the blood via the dialyzer.
After the balancing chamber (68), the used dialysate is directed to the drain via the valve (30), the
heat exchanger (77) and the drain valve (87).
With the DIASAFE®plus option (114), the dialysate passes through a capillary filter with a
polysulfone membrane and then flows to the dialyzer. The filter holds back pyrogenic substances
and microorganisms.
The pressure measuring points A, B, C, and D allow direct measurement of various pressures,
without having to separate lines and tubings. Measuring point A measures the water inlet
pressure; B the loading pressure of the balancing chamber, C the pressure of the flow pump, and
D the negative degassing pump pressure.
During the hot rinse and disinfection programs, the machine is operated in recirculation mode. In
this case, the drain valve (87) is closed and the recirculation valve (86) is open. This method of
recirculation serves to save energy (hot rinse) and disinfectant. The disinfectant is introduced into
the circuit by the UF pump (22) via the disinfectant valve (84) and the sensor (115). The sensor
(115) detects presence or absence of disinfectant.
The float switch (5) located in the heater block (66) controls the inflow of fresh water via the water
inlet valve (41).
The upstream pressure reducing valve (61) ensures a constant water inlet pressure.
The water is heated in the heater rod chamber (66b). The degassing pump (29) and the
degassing orifice (89) are used to generate a negative pressure to deaerate the water. The air is
collected in the primary air separator (88c). From there, it is directed into the heater rod chamber
(66b) via the loading pressure valve (65) and then discharged into the atmosphere via the float
chamber (66c).
The preheated and deaerated water is delivered from the primary air separator (88c) to the
concentrate / bicarbonate mixing point (205).
Concentrate containing acetate (or in bicarbonate-based dialysis acid concentrate) flows from
the concentrate suction tube (94) through the concentrate air separator (201) and the concen-
trate pump (23) to the concentrate/bicarbonate mixing point (205). In bicarbonate-based dialysis,
bicarbonate is transported via the bicarbonate suction tube (95), the air separator (203) and the
bicarbonate pump (25) to the concentrate / bicarbonate mixing point (205).
The evacuation valve (188) and the bibag® drain valve (130) are provided for deaeration.
The conductivity measuring cell (132) und the temperature sensor (133) measure and monitor the
bicarbonate conductivity.
At the concentrate/bicarbonate mixing point (205) the appropriate concentrates are mixed with
water. The dialysate then flows across the balancing chamber (68), the dialysate filter (114) and
the conductivity cell (7) to the dialyzer.
The dialysate filter (114) holds back pyrogenic substances and microorganisms.
In the conductivity cell (7), both the conductivity and the temperature (temperature sensor 3) are
measured and indicated on the monitor. The temperature sensor (109) controls the temperature.
If temperature and conductivity are inside the alarm limits, the two dialyzer valves (24 and 24b)
are open and the bypass valve (26) is closed.
Should one of the two values be outside the alarm limits (temperature or CD alarm), the bypass
valve (26) opens and the dialyzer valve (24) closes.
The integrity of the DIASAFE®plus (114) membrane is checked by ventilating the internal capillary
space via the vent valve.
The retentate valve (189) is used to rinse retentate from the internal capillary space of the
DIASAFE®plus (114).
The pressure transducer (9) is located below the secondary air separator (88b) of the multifunc-
tion block (88). The pressure measured here is taken as a quantity in the computation of the TMP.
A membrane pump (UF pump 22) with a volume of 1 ml per stroke withdraws fluid from the
system at a specified rate. Since the system is closed, the same volume is removed as
ultrafilatrate from the blood via the dialyzer.
After the balancing chamber (68), the used dialysate is directed to the drain via the valve (30), the
heat exchanger (77) and the drain valve (87).
The heat exchanger (77) transmits part of the heat of the outflowing dialysate to the cold inflowing
water.
Larger amounts of air entering the system, as a result of major deaeration or during priming, are
detected by the level sensor (6). This air is discharged into the atmosphere via the air separation
pump (97).
The pressure measurement points A, B, C, and D allow direct measurement of various pressures,
without having to separate lines and tubings. Measurement point A measures the water inlet
pressure; B the loading pressure of the balancing chamber, C the pressure of the flow pump, and
D the negative degassing pump pressure.
During the hot rinse and disinfection programs, the machine is operated in recirculation mode. In
this case, the drain valve (87) is closed and the recirculation valve (86) is open. This method of
recirculation serves to save energy (hot rinse) and disinfectant. The disinfectant is introduced into
the circuit by the UF pump (22) via the disinfectant valve (84) and the sensor (115). The sensor
(115) detects presence or absence of disinfectant.
When using the CDS option (central concentrate delivery), concentrate containing acetate, or in
bicarbonate-based dialysis acid concentrate, flows from the concentrate connector (121) through
the valve (102) to the rinse chamber (90a). Bicarbonate is delivered in the same way from the
bicarbonate connector (122) through the valve (104) to the rinse chamber (90b).
The desired type of concentrate supply must be selected in the SETUP menu (see Technical
Manual). The following 5 possibilities are available:
1. No central delivery system (factory setting)
2. Central bicarbonate delivery
3. Central acid delivery
4. Central bicarbonate delivery and central acid delivery
5. Central acetate delivery
When selecting a type of central delivery, the user is not bound to the centrally supplied
concentrate. Other concentrates can also be connected by pulling or inserting the concentrate
suction tubes. The valves V102 and V104 are controlled automatically. The hemodialysis machine
also recognizes the type of dialysis performed (acetate or bicarbonate dialysis)and activates the
concentrate or the bicarbonate pump, as appropriate.
☞ Note
Aspiration of concentrates from containers has priority (see graphical represen-
tations on the following pages).
If, with central delivery, the concentrate suction tube is repeatedly pulled off, the
rinse chamber will flow over. The rinse chamber will not be emptied before the
next cleaning program.
The user has the option to switch off the central delivery system. (See Chapter 2.1.4 Concentrate
Supply.)
☞ Note
The operator is responsible for correct installation and function of the CDS.
The following illustrations show the possible combinations of connection with regard to the CDS.
(The bibag ® option can be used as an alternative to bicarbonate from containers.)
● Mode of operation
The hemodialysis machine permits the optional performance of acetate or bicarbonate dialysis.
The type of treatment is determined by:
– the concentrate(s),
– the positioning of the concentrate suction tubes.
The concentrates are delivered by two pumps and are mixed with RO (reverse osmosis) water in
the hemodialysis machine.
The concentrate suction tubes are connected to the appropriate containers.
With CDS, the appropriate concentrates are delivered by the central delivery system.
The mixing process is volumetric and proportional, i.e. the concentrate pumps add an exact
quantity of concentrate to a fixed volume of water.
The mixing ratio can be varied. The desired ratio can be selected in the dialysate menu. To change
the ratio, the operator must press the Dialysate Menu key.
During bicarbonate dialysis, precipitation of calcium and magnesium salts can occur. The machine
should therefore be decalcified and disinfected after each bicarbonate dialysis, using a suitable
agent (e.g. Puristeril® 340 or Citrosteril®).
During a cleaning program, the concentrate suction tubes must be inserted in the rinse ports
located on the hemodialysis machine.
The correct ion concentration in the dialysate can only be achieved with concentrates that
match the mixing system.
Acid concentrate
SK-F 203 (PGS 21)/SK-F 003 (PGS 01), 35-fold, 6 l in 10-l container.
Bicarbonate concentrate
8.4% sodium hydrogen carbonate solution, 8 l in 10-l container.
☞ Note
The contents of opened bicarbonate concentrate containers cannot be stored.
After mixing, the acetic acid from the acid concentrate reacts with the equivalent quantity of
bicarbonate, producing carbonic acid and sodium acetate.
This produces dialysate which has the following ion composition during the basic setting
of the pumps for:
Procedure
The dialysate is continuously prepared from bicarbonate concentrate, acid concentrate, and
reverse-osmosis water.
Two membrane pumps, which are independent of each other, deliver the acid concentrate solution
and the bicarbonate solution in a volumetrically proportional mixing system.
Caution
Use only specially keyed containers and mating suction tubes.
This system does, however, not exclude the possibility of incorrect use of
uncoded containers.
Only the dedicated bibag® may be connected to the bibag® connector (option).
to dialyzer
Degassing
via balancing chamber
pump
Concentrate Bicarbonate
pump pump
○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○
○ ○ ○ ○ ○ ○ ○
RO Concentrate A Concentrate B
water (acetate or acid) (Bicarbonate) or (bibag ® Option)
to dialyzer
via balancing chamber
Concentrate Bicarbonate
pump pump
Degassing
pump
○ ○ ○ ○ ○ ○ ○ ○ ○ ○
○ ○ ○ ○ ○
Measurement of the conductivity and the bypass function prevent that dialysate of the wrong con-
centration will enter the dialyzer.
Monitoring of the desired conductivity to within ±5% by setting alarm limits provides a safeguard
against errors in the mixing system.
In the event of dialysate of the wrong composition, the conductivity will be outside the alarm limits.
The CD alarm, which is then generated, will cause the machine to go into bypass, accompanied
by an audible and a visual alarm.
An initial error in the mixing system will be detected before the deviation of the dialysate values from
the theoretical value becomes dangerous.
B* Belgian bicarbonate
1
for detailed information refer to the Technical Manual
☞ Note
For treatment with the bibag ® (option) the mix ratio must be set to 1+34 or 1+44.
The total conductivity of the ready-to-use dialysate (in mS/cm) is determined by the acidic and
basic part of the bicarbonate hemodialysis concentrates.
The total conductivity of the ready-to-use dialysate with the appropriate mixing ratio is indicated
on each package of the acidic bicarbonate hemodialysis concentrate from Fresenius Medical
Care.
In the event of a ready-to-use dialysate with 32 mmol/l bicarbonate and 3 mmol/l acid (e.g.
acetate content), the contribution of the basic component of the bicarbonate concentrate to the
total conductivity is 2.2 mS/cm. The contribution of the acidic component of the bicarbonate
concentrate to the conductivity depends on the different electrolytic contents.
Based on a basic sodium value of e.g. 140 mmol, the sodium value changes by 5 %, if the
prescribed sodium value is increased to 147 mmol. The total conductivity of the ready-to-use
dialysate then also increases by 5 %.
If the bicarbonate proportion is changed, the acid proportion is changed in the opposite direction,
in order to maintain the prescribed sodium value constant. This may result in a minor CD alteration.
Example:
A prescribed sodium value of 140 mmol/l corresponds to 105 mmol/l from the acid concentrate and
35 mmol/l from the bicarbonate concentrate. If the bicarbonate is readjusted by +3 mmol/l
(= 38 mmol/l on the bicarbonate pump), the acid concentrate pump is readjusted by –3 mmol/l to
102 mmol/l. The prescribed sodium value of 102 + 38 = 140 mmol/l is preserved.
In order to improve the individual tolerance towards dialysis treatments, ultrafiltration profiles and
sodium profiles can be combined, either depending upon each other (e.g. high ultrafiltration with
high sodium concentration) or independently of each other.
Separate documentation is available from the manufacturer on request.
Sodium
1 2 3 4 5 6
1
Ultrafiltration
● System Components
Hydrophobic filter
Balancing Cond./Temp.
chamber monitoring Vent valve
DIASAFE® plus
Dialyzer Venous
valve 1 bubble catcher
Air detector
V24
UF Flow
pump pump Dialyzer
V24b
V26 Bypass Venous
valve line
clamp
Dialyzer
valve 2
Arterial
blood
pump
The filter test (DIASAFE®plus test) is incorporated in the functional test provided it has been
preselected by the DIP switch setting on LP 632.
When the filter test is started, the internal capillary spaces of the DIASAFE®plus must be ventilated
by the vent valve. A special balancing chamber switching cycle is employed to remove fluid from
the interal capillary space across the membrane. The hydrophylic membrane rejects air so that a
negative pressure builds up in the system. Balancing chamber switching stops at a negative
pressure of about – 300 mmHg (peak value). The UF pump increases the negative pressure to
about – 450 mmHg. A differential pressure measurement is started. The result computed by the
microprocessor must not exceed a pressure change of ± 16.7. The maximum duration of the test
is 5 minutes. In the event of a more significant change in pressure, the UF pump will adjust the
negative pressure to the initial pressure. Differential measurement is then restarted.
Another test criterion is the fluid volume removed. A deviation of ± 30 ml from the expected value
of 145 ml will cause the test to stop.
An error message will be displayed if the filter fails to pass the test.
On successful completion of the functional test the hemodialysis machine switches into bypass
mode for a defined period of time. During this period the internal capillary space of the
DIASAFE ®plus is automatically refilled with fluid.
Hydrophobic filter
Balancing Cond./Temp.
chamber monitoring Vent valve
DIASAFE® plus
Dialyzer
valve 1
V24
UF Flow
pump pump Interlock shunt
V24b
V26 Bypass
valve
Dialyzer
valve 2
The fresh dialysate flows from the balancing chamber via the conductivity and temperature
monitoring system to the DIASAFE®plus.
This filter is integrated in the hydraulic system to ensure that in a no alarm state (no dialysate
alarm: V26 closed, V24 and V24b open) the "fresh" dialysate flows across the filter membrane.
In bypass mode (dialysate alarm: V26 open, V24 and V24b closed) fluid flows through the inner
capillary space of the DIASAFE ®plus . Possible retentate is rinsed from the capillaries and
transported to the drain thus overcoming the disadvantages of a dead end filtration.
Hydrophobic filter
Balancing Cond./Temp.
chamber monitoring Vent valve
DIASAFE® plus
Dialyzer Venous
valve 1 bubble catcher
Air detector
V24
UF Flow
pump pump Dialyzer
V24b
V26 Bypass Venous
valve line
clamp
Dialyzer
valve 2
Arterial
blood
pump
During the cleaning programs the dialysate lines (red/blue) are connected to the interlock shunt.
The cleaning agent flows through the balancing chamber to the DIASAFE®plus. Valves V24, V24b
and V26 are clocked so that the fluid alternately flows through the internal capillary space
(bypass) and across the filter membrane.
This ensures that in the bypass mode possible retentate is rinsed from the capillaries and is
transported to the drain. The filter is thus thoroughly cleaned, disinfected or rinsed.
Hydrophobic filter
Balancing Cond./Temp.
chamber monitoring Vent valve
DIASAFE® plus
Dialyzer
valve 1
V24
UF Flow
pump pump Interlock shunt
V24b
V26 Bypass
valve
Dialyzer
valve 2
● System components
DIASAFE ®plus
The DIASAFE®plus is a hollow-fibre filter. Its excellent filtering characteristics can be attributed to
the Fresenius Polysulfon® membrane.
Pressure transducer 2
This transducer is a sensor mounted as a standard component in the hydraulic unit. It is used to
control and evaluate the membrane integrity test.
Compressor
The compressor is required for testing the integrity of the membrane of the DIASAFE®plus . Within
a minimum of time it ventilates one side of the filter membrane until the test pressure is achieved.
Test valve
The test valve is only open as long as the compressor ventilates the filter membranes and the test
pressure has built up.
The integrity of the membrane of the DIASAFE ®plus is tested by means of a pressure holding test.
This test is integrated in the functional test. For this test, the outer capillaries of the DIASAFE ®plus
are ventilated. This is done by the compressor which pumps air via the test valve into the
hydraulic system. The inflowing air displaces the fluid in the system via the hydrophilic filter
membrane and the retentate valve into the drain. This process continues until the entire outer
capillary space is filled with air. As air cannot pass across an intact membrane, a positive
pressure will build up on the ventilation side. The pressure transducer 2 controls the compressor
and evaluates the integrity test. When the test pressure has been achieved, the compressor is
switched off and the test valve closes. The pressure drop per unit of time is used as a measure for
the evaluation of the integrity of the membrane.
Drain Pressure
valve DIASAFE® P transducer 2
plus Cond./Temp. (182)
(87) monitoring
(114)
(3/7/109)
Retentate Bypass
valve valve
(189) (26)
Pressure
transducer Dialyzer
(9) valve 2
P (24b)
The fresh dialysate flows from the balancing chamber to the DIASAFE®plus.
This filter is integrated in the hydraulic system to ensure that in a no alarm state (no dialysate
alarm: V26 closed, V24 and V24b open) the "fresh" dialysate flows across the filter membrane.
In bypass mode (dialysate alarm: V26 open, V24 and V24b closed) fluid flows through the inner
capillary space of the DIASAFE ®plus . Possible retentate is rinsed from the capillaries and
transported to the drain thus overcoming the disadvantages of a dead end filtration.
In all cleaning programs the dialyzer lines (red / blue) must be connected to the interlock shunt.
The cleaning agent, water or disinfectant flow via the balancing chamber to the DIASAFE®plus. As
the retentate valve, V26 and V24/24b open alternatingly, fluid flows alternatingly through the
inner capillary space and the membrane of the filter.
Drain Pressure
valve DIASAFE® P transducer 2
plus Cond./Temp. (182)
(87) monitoring
(114)
(3/7/109)
Retentate Bypass
valve valve
(189) (26)
Pressure
transducer Dialyzer
(9) valve 2
P (24b)
Legend
The arterial blood pump delivers blood from the patient’s vascular access to the dialyzer. The
pressure on the suction side of the pump is measured and monitored by means of the arterial
pressure monitor, which is separated by a hydrophobic filter. If the pressure is outside the set limits,
a blood alarm will be released: the arterial blood pump stops, the venous line clamp is closed, a
visual and audible alarm is initiated. Ultrafiltration is stopped.
Heparin doses can be added to the blood at an adjustable rate via the heparin pump.
After the dialyzer, the blood enters the venous bubble catcher. Here, the venous back pressure is
measured and monitored within tight limits. The monitor for the venous back pressure is separated
by two hydrophobic filters. The venous bubble catcher is located in the level detector, which serves
as a protection against infusion of air. Should the level drop or should there be frothy blood in the
venous bubble catcher, a blood alarm is emitted: the arterial blood pump stops, the venous line
clamp is closed, a visual and audible alarm is emitted. Ultrafiltration is stopped.
After the bubble catcher, the blood passes the optical detector (OD), which differentiates as
follows:
– OD senses non-opaque fluid (saline or air in the tubing system) or
– OD senses opaque fluid (blood in the line tubing system).
From the optical detector, the blood is then returned to the patient (venous needle).
☞ Note
This procedure should be used in exceptional cases only, since the stroke
volumes and, thus, the corresponding recirculation volume may be very unfavor-
able.
Should blood flow problems arise in one of the vascular accesses during double needle dialysis,
it is still possible to continue the treatment using the installed tubing system, by selecting SN click-
clack. After the switching pressure values have been entered, the venous pressure transducer
controls the blood pump and the venous line clamp. Both the arterial and the venous blood line are
connected to the remaining vascular access by means of a Y-piece.
The venous line clamp is closed, as long as the arterial blood pump is delivering blood. As soon
as the upper limit at the venous pressure transducer has been reached in the blood lines, the blood
pump stops and the venous line clamp opens. The pressure, which has developed in the tubing
system, returns the stored blood to the patient. As soon as the set lower limit has been reached,
the venous line clamp closes, the arterial blood pump starts running, and the same procedure is
repeated periodically.
Should the cycle time be exceeded by 15 s, an alarm will be emitted. The venous line clamp is
closed. The blood pump and the UF pump stop running.
Legend
During SN dialysis, blood is alternately collected from and returned to the patient through one
needle. This is achieved by means of an additional blood pump module which includes the single
needle control electronics (2-pump procedure with pressure-pressure control).
With SN (Q SN) in the twin-pump SN system, the mean blood flow is computed as follows:
QSN = (BPRart x BPRSN) / (BPRart + BPRSN) [theoretical value]
The following approximation formula is sufficient for practical use:
QSN = (BPRart + BPRSN) / 4
The arterial blood pump delivers blood into the external compliance chamber. The blood volume
is dependent on the set stroke volume (10 to 50 ml). The SN blood pump is not running. The venous
line clamp is closed. The external compliance chamber is connected to a pressure transducer in
the SN blood pump. The compliance chamber, which is filled with air, is located inside the SN blood
pump. As soon as the pressure exceeds the set upper limit, the arterial blood pump stops. The SN
blood pump starts running and delivers the blood through the dialyzer, the venous bubble catcher
and the venous line clamp into the patient. As soon as the pressure in the internal compliance
chamber drops below the set lower limit, the described procedure starts again.
The lower changeover point is fixed to 75 mmHg. The upper changeover point depends on the
stroke volume entered.
The position of the SN blood pump before the dialyzer prevents level fluctuations in the venous
bubble catcher. The TMP stability is also improved.
The therapeutical result (clearance) depends on the effective blood flow and the stroke volume.
The higher the stroke volume, the lower the proportionate recirculation.
For this reason, the largest possible blood flow and the largest possible stroke volume should be
set. These settings differ from patient to patient and, owing to the individual vascular accesses,
have different limits.
All other components and their functions correspond to those already described for double needle
dialysis.
The stop recognition of both blood pumps is activated. Should one of the pumps stop longer than
for the set time (factory setting: 30 seconds), a blood pump stop alarm is emitted.
☞ Note
The design (closed system) of 4008 hemodialysis machines ensures that pres-
sure fluctuations during SN operation will not affect the UF rate or the UF
volume.
The delivery rate display on the blood pump indicates the theoretical blood flow. This value is
computed from the rotating speed of the rotor and the inside diameter of the blood line used.
The actual blood flow (effective blood flow) may deviate slightly from this value, because it
depends on the different input pressures (arterial pressure) on the pump.
The computer of the monitor contains an algorithm which corrects the theoretical blood flow,
dependent on the arterial pressure. By means of the menu option 4. Bloodflow.calc., the value
can be indicated on the text display of the monitor. (See Chapter 7.4.1 Displaying the Effective
Blood Flow.)
Using the actual blood flow and the elapsed treatment time, the monitor computes the accumu-
lated blood volume. This information can also be displayed under menu option 4.
1 Rate: ml/min
(Ø: mm)
2
11
3
10
4 Start
Stop 9
1 Display (indicates the delivery rate, the line diameter, the SN stroke volume, or an error
code)
2 Alarm LED (red)
3 Operation LED (green)
4 Pressure connector (Luer-lock connector of the SN pressure measuring line)
5 Clamping piece (holds the installed line segment securely in place)
6 Blood pump cover
7 Rotor (delivers the patient’s blood by means of the rollers and the lines installed. The rotor
can be removed for cleaning.)
8 Sensor (blood pump cover open or closed)
9 Start/Stop key (to toggle the blood pump on and off)
10 ▼ key (to reduce the delivery rate, the value of the line diameter, or the SN stroke volume)
11 ▲ key (to increase the delivery rate, the value of the line diameter, or the SN stroke volume)
The following points must be strictly observed to assure safe operation of the hemodia-
lysis machine:
● Electric installation
– The national standards and regulations (e.g. in Germany DIN VDE 0100-710) are to be
observed when connecting the machine to the local power supply system.
– When using safety class I machines, the quality of the protective ground of the installation is of
particular importance. It must be taken into consideration that in many countries regulations
have been enacted by national authorities.
– If the power cord needs to be replaced, use only the original power cord listed in the spare
parts catalog.
– Connect the equipotential bonding to the rear of the machine, if this is required by the legal
regulations of the place of installation (e.g. according to DIN VDE 0100-710 in rooms of
application group 2).
Dialysis Additional
machine electromedical
device
all types!
(B / BF / CF)
3. When determining the room design, a group 1 room is sufficient, i.e., the
mains may be turned off in case of a single fault condition, and the treatment
may be aborted or repeated.
Additionally, the room must comprise a potential equalization.
For further information refer to the national regulations.
● Batteries
– Only water, which is suitable for hemodialysis, should be used for operating the hemodialysis
machine. The quality of the water should comply with the local regulations (e.g. European
Pharmacopoeia - 4th Edition, ANSI/AAMI - RD5 - 1992 - Part 3.2).
– The ranges of the following parameters are specified in Chapter 1.10 Specifications: water
inlet pressure, supply pressure of the dialysate concentrate, temperature, and flow rates,
which are required for operating the hemodialysis machine.
– The national rules and regulations with regard to the prevention of reverse flow into the water
supply network and the air gap between drain connection and sewage connection must be
observed when installing and operating the machine (e.g. in Germany VDE 0753, part 4). For
further information refer to Chapter 1.10 Specifications.
When using the DIASAFE®plus attention should be given to the points listed below with
regard to the quality of product water, concentrate and dialysate:
Caution
Use only reverse osmosis water.
Always ensure that both product water and dialysate meet the applicable
standard (see below).
For microbiologic quality standards / recommendations refer to the following
table.
General notes
The microbiologic purity of dialysate prepared in the dialysis center is of critical importance.
Concentrate quality
While acid concentrate may be considered bacteriostatic, bicarbonate concentrate is an ideal
nutrient medium for microorganisms. The use of sterile or microorganism-poor concentrates is
recommended. On-line dry concentrates (bibag®) aid the preparation of dialysate with high
microbiological quality. The use of residues from open canisters and the preparation of bicarbo-
nate concentrate in the dialysis centre are not recommended. Central delivery systems for
concentrate or dialysate must meet high hygienic standards and require regular inspection.
– Prior to each treatment, the prescribed functional checks must be performed. (See Chapter
2.8.1 Functional Checks Prior to Each Dialysis Treatment.)
● Important documents
– On request, circuit diagrams, descriptions, spare parts lists, and other documents are made
available by the manufacturer. These are intended to support trained personnel in servicing
and repairing the machine.
☞ Note
The machine may only be used as specified in the machine documentation.
Only under these conditions will the manufacturer consider himself responsible
for any effects on the safety, the reliability, and the performance of the machine.
Initial start-up of the machine should be performed by the technical service of the Fresenius
Medical Care or a person authorized by them.
– If the machine is transported from a cooler into a warmer room, allow the hemodialysis
machine to reach room temperature for approx. 2 hours, before turning power on.
– Prior to putting the machine into operation for the first time, the DIP switch positions must be
checked and, if necessary, the switches set correctly. The setup settings must also be
checked.
(See the Technical Manual)
The selected DIP switch settings must then be entered in the Medical Device Register.
Caution
When connecting the hemodialysis
machine for the first time, make sure
to observe the technical specifica-
tions in Chapter 1.10.
☞ Note
To avoid problems during the man-
datory rinse (to remove the anti-
freeze) the service switch must be
set to ON (up).
On completion of the mandatory
rinse set the service switch back to
OFF (down) again.
Hotrinse 280
Blood Leak 30 60
200
20 50
Disinfection
10
100 40
0 0 30
Single
Needle –10
–100 20
–20
☞ Note
Installation of the DIASAFE®plus must
be entered in the Medical Device Re-
gister (date, batch number).
Verify, and if necessary adjust, the
DIP switch settings prior to initial
start-up. Record the settings in the
Medical Device Register.
● Basic requirements
☛ I/O
Luftdetektor
Override
Arterieller
Druck
Ve
D
–30
–200
0
Test
–40 –300
The Test LED is flashing.
If the Test LED is not flashing, one of
Prime the conditions for the test to start is not
fulfilled.
Dialysis Start
Reset Alarm
Alarm Tone
Mute
☛
–30
–200
0
Test
–40 –300
Press the Test key.
Prime The Test LED is illuminated.
Dialysis Start
Reset Alarm
Alarm Tone
Mute
☞ Note
If the machines fails to pass the
functional test, repeat the test by
pressing the Test key.
If a DIASAFE®plus failure (FXX) is
displayed after the functional test,
this test step has to be skipped.
Then run a heat disinfection program
or a disinfection program followed by
a hot rinse.
Caution
Disinfect prior to the first treatment.
☞ Note
Installation of the DIASAFE®plus must
be entered in the Medical Device Re-
gister (date, batch number).
Verify, and if necessary adjust, the
DIP switch settings prior to initial
start-up. Record the settings in the
Medical Device Register.
● Basic requirements
Caution
The machine must be disinfected
prior to initating the first treatment.
The length of the individual cleaning programs can be freely selected (see Technical Manual,
adjustment instructions in the SETUP menu).
☞ Note
Flow alarms during cleaning programs will increase the length of the cleaning
program by the time of the alarm. Should the hemodialysis machine remain in
the no-program state for more than 10 minutes after completion of a cleaning
program, the machine turns off automatically.
Exception: The cleaning program was started with the Auto-On function.
The cleaning program was stopped (e.g. by turning the suction tube).
The following components and functions are monitored during the cleaning programs:
– Dialyzer valve
– Bypass valve
– Balancing chamber switching pulse (flow)
– Float switch
– UF pump
– Temperature
– 24 V and 12 V supply voltages
– Software watchdog
– Shunt door
– Dialyzer lines connected
– Concentrate and bicarbonate suction tubes connected
– CDS
– V84
– Retentate valve
Caution
It is imperative to disinfect the machine after each dialysis treatment.
☞ Note
If the machine has been idle for more than 72 hours, it is recommended to run a
disinfection program before initiating the next treatment.
The agents for disinfection (Puristeril® 340) and for heat disinfection (Citrosteril® or Diasteril®) are
connected on the rear of the machine (disinfection programs 1 through 4).
Irrespecitve of the container connected to the rear of the machine, the special program 5 can be
used , if necessary, to draw in a hypochlorite-containing disinfectant (e.g. Sporotal® 100) from the
front via the concentrate suction tube.
☞ Note
The heat disinfection program may only be performed with Citrosteril ® or Dia-
steril®. If a different type of disinfectant is connected to the rear of the machine
when changing the program (e.g. Puristeril® 340 instead of Citrosteril® or Dia-
steril®), the previously used disinfectant must be thorougly rinsed from the
suction tube.
When the disinfectant is to be drawn in from the front (PGM 5), ensure that a
container with a sufficient amount of disinfectant is connected. Avoid aspiration
of air.
–HR– –HR–
Temp.: 84 °C Temp.: 84 °C
Flow: 450 ml/min Flow: 450 ml/min
Time: adjustable from Time: adjustable from
15 to 30 min in the 15 to 30 min in the
SETUP menu SETUP menu
Timing: Timing:
from > 80 °C from > 80 °C
Perform the maintenance procedures specified in the Technical Manual. The technical safety
checks (TSC) are included in the maintenance intervals.
The items listed in this chapter are an extract from our extensive product range. Further
products on request.
1.9.1 Disposables
Caution
The dialyzers, blood lines and syringes listed in this chapter are disposables,
identified by the symbol: 2 = do not re-use.
The 4008 B hemodialysis system has been approved for use with the disposables listed below
within the scope of the EC design certification.
● Dialyzers
Caution
These blood lines match the pressure limiting system of the blood pumps.
In order to avoid cross infections at the pressure connectors, the use of
hydrophobic filters is required.
Optimum single needle operation is only achieved by using the original tubing
systems.
During continuous operation, the blood lines must be exchanged after 24 hours
at the latest.
Double-Needle dialysis:
Single-Needle dialysis:
● Disposable syringes
● Hemodialysis concentrates
● Saline solution
Solutions for disinfecting and removing calcium deposits from the dialysate circuit
Caution
Strictly observe the following information:
1. The "use before" date printed on the container.
2. The storage conditions for the disinfectant concentrates.
3. The manufacturer’s instructions for use.
Caution
Prior to the use of disinfectants other than those listed hereafter, their efficacy
and compatibility with the materials affected in the hemodialysis machine and
the DIASAFE®plus (option) must be ensured.
Improper use of disinfectants (concentration, temperature range, dwell time)
may result in damage to the dialysis machine and the DIASAFE®plus (option).
Caution
The following must be observed when using Puristeril®plus:
In case of treatments with High-Flux membranes, we additionally recommend to
perform a regular alkaline cleaning with Sporotal® 100.
Puristeril®plus must not be used on machines with the DIASAFE (old) and ON-
LINE-HDF options.
2. Fresenius Puristeril®plus:
(Water temperature in the cleaning program: 37 °C)
Active substance base: peracetic acid
Active substance concentration: < 0.1 % (diluted)
☞ Note
Applicable for DIASAFE®plus (option) only:
After installation, the DIASAFE ®plus can be cleaned with Sporotal® 100 for a
maximum of 11 times.
In machines with basic hydraulics, the cleaning procedures with Sporotal® 100
must be documented.
In machines with advanced hydraulics the cleaning procedures with Sporotal ®
100 will be monitored by the hemodialysis machine.
4. Fresenius Citrosteril®:
(Water temperature in the cleaning program: 84 °C)
Active substance base: citric acid hydrate
Active substance concentration: approx. 0.8% (diluted)
5. Fresenius Diasteril®:
(Water temperature in the cleaning program: 84 °C)
Active substance base: hydroxyacetic acid
Active substance concentration: approx. 0.8% (diluted)
2. Fresenius Puristeril®plus :
(Water temperature in the cleaning program: 37 °C)
Active substance base: peracetic acid
Active substance concentration: < 0.1 % (diluted)
3. Fresenius Citrosteril®:
(Water temperature in the cleaning program: 37 °C)
Active substance base: citric acid hydrate
Active substance concentration: approx. 0.8% (diluted)
4. Fresenius Diasteril®:
(Water temperature in the cleaning program: 37 °C)
Active substance base: hydroxyacetic acid
Active substance concentration: approx. 0.8% (diluted)
Caution
It is recommended to perform the check for residual disinfectants at the drain of
the hemodialysis machine.
In case other test methods are applied, the requesting person bears the respon-
sibility for this decision.
Potassium-iodide-starch paper:
Caution
Always use fresh (white) indicator paper.
Caution
Test for approx. 10 to 15 sec. In the event of a positive pH reaction (yellow color
≤ 4.1), run a rinse program for a minimum of 5 min., then re-determine the pH.
Solutions for external disinfection and cleaning of the machine and the filter housing of
the DIASAFE®plus
Fresenius ClearSurf™:
Type of disinfection: scouring/wiping disinfection
Active substance base: cationic tensides
Use concentration: see product description
0.5 % – 1 hour
1.0 % – 15 minutes
Caution
Use only the original DIASAFE®plus specified herein.
Dialyzers
Blood lines
Double-Needle dialysis
AV-Set FMC (FA 204B / FV 204B) 500 056 1 22 mm ETO
AV-Set FMC (FA 204B / FV 234B) 500 057 1 30 mm ETO
AV-Set FMC (FA 204B / FV 204B) S 601 421 1 22 mm Steam
AV-Set FMC (FA 204B / FV 234B) S 601 431 1 30 mm Steam
AV-Set FMC (FA 204B / FV 204B) G 501 721 1 22 mm Gamma
AV-Set FMC (FA 204C / FV 234B) G 501 706 1 30 mm Gamma
Single-Needle dialysis
AV-SN-Set FMC (FA 514 SN B / FV 204B) 501 741 1 22 mm ETO
AV-SN-Set FMC (FA 514 SN B / FV 234B) 501 742 1 30 mm ETO
Arterial – FA 514 SN B S 605 014 1 – Steam
Venous – FV 204 B S 604 804 1 22 mm Steam
Venous – FV 234 B S 604 834 1 30 mm Steam
AV-SN-Set FMC (FA 514 SN B / FV 204B) G 501 703 1 22 mm Gamma
Disposable syringes
Type Fresenius 30-ml heparin syringe 503 032 1
Type Fresenius 50-ml heparin syringe 900 071 1
bibag® (option)
Sodium bicarbonate 700 g 508 507 1 GB, D, F, NL, E, P, I
Sodium bicarbonate 650 g 508 992 1 GB, D, F, NL, E, P, I
Sodium bicarbonate 700 g 508 510 1 S, SF, DK, N
Sodium bicarbonate 950 g 508 991 1 GB, D, F, NL, E, P, I
Sodium bicarbonate 950 g 508 732 1 S, SF, DK, N
Line voltage 230 V AC, +6%, –10%, 50/60 Hz Symbol: (alternating current)
240 V AC, ±10%, 50/60 Hz Symbol: (alternating current)
110 V AC, ±10%, 50/60 Hz Symbol: (alternating current)
Operating current:
– Dialysis: Approx. 6 A, at a water inlet temperature of 17 °C
dialysate temperature: 37 °C
dialysate flow: 500 ml/min
4008 B
1
230 V~ 50/60 Hz
2 11 A
3 7
4 6
5
Fresenius Medical Care AG
D-61346 Bad Homburg
Tel.: ++49 (0) 61 72 / 609-0
Legend
Water drain: 0 up to 100 cm above the ground, no less than 5 cm free fall. The
water drain must be located at a lower level than that of the dialyzer
position.
Concentrate supply: 0 to –100 mbar; max. suction height: 1 m, max. height of fall: 0 m
(with central delivery system – optional: 0 up to 500 mbar)
Heat dissipation: Dialysis: approx. 400 watts (at an ambient temperature of 20 °C)
Hot rinsing: approx. 520 watts (at an ambient temperature of 20 °C)
Range of operating 15 °C up to 35 °C
temperature:
Temporary downtime: It is recommended to run a hot rinse program without cooling rinse
(PGM 2) before putting a machine temporarily out of use and to run
a disinfection program prior to reuse.
Stability: 10°
Caution
Any additional equipment connected to the analog and digital interfaces of the
machine must comply with the applicable EN specifications (e.g. EN 60950 for
data processing equipment, and EN 60601 (IEC 601) for electro-medical equip-
ment). Apart from this requirement, all configurations must comply with the
system standard EN 60601-1-1 (IEC 601-1-1). The connection of additional
equipment to the signal input or output component affects the system configura-
tion and anyone connecting additional equipment is therefore responsible for
compliance with the system standard EN 60601-1-1 (IEC 601-1-1).
Condition indicator: Can be used for the connection of a condition indicator (optional).
Preparation: Defined by the optical detector below the venous bubble catcher.
Preparation is terminated as soon as the optical detector senses
opaque fluid in the blood line.
Priming the blood lines: Automatic priming of the blood lines when pressing a key (Prime).
Priming is automatically terminated as soon as the venous bubble
catcher is filled, however, not later than after 1 to 5 minutes (time
adjustable in the SETUP menu).
Single needle With use being made of the arterial blood pump and the venous line
click-clack dialysis: clamp; pressure-pressure-controlled with adjustable pressure re-
verse values.
Exceptional procedure in case of problems with the vascular access
during double needle dialysis.
Single needle dialysis Two-pump procedure with adjustable upper pressure reverse value
(optional): (adjustable via stroke volume).
Flush, rinsing the water With a water inflow rate of 1.3 l/min and an inlet pressure of 1.5 bar,
supply hose (option) the water supply hose will be rinsed with a volume of 260 ml.
Blood leak detector: Threshold of response lower than and equal to 0.5 ml blood loss per
minute into the dialysate at a hematocrit of 0.25.
(Flow rate 300 ml/min to 800 ml/min)
With the dialysate flow turned off, a blood leak alarm response is delayed. The
delayed response is dependent on the fluid volume in the hydraulic compartment
between dialyzer and blood leak detector (approx. 110 ml total fluid volume) and
on the size of the membrane rupture in the dialyzer. In addition, the response of a
blood leak alarm is, dependent on the set ultrafiltration rate.
Degassing: Residual gas volume in the dialysate at the inlet to the dialyzer
approx. 11 ml air per 1 liter of water (at 1000 hPa and 0 °C)
When operating with reverse osmosis water of 22 °C and an air
content of approx. 21 ml air per 1 liter of water (at 1000 hPa and
0 °C)
Procedure: negative pressure
– Acetate: Standard setting 1+34 (further mixing ratios can be adjusted via
the SETUP menu). Range of adjustment: 125 to 150 mmol/l, de-
pending on the concentrate used.
Volumetric proportional mixing system. The rated concentration is achieved after
10 minutes at the latest.
Arterial range:
– Arterial pressure Display range: –300 to +280 mmHg
measurement: Resolution: 20 mmHg
Accuracy: 10 mmHg
The OD in the LD senses clear fluid:
Alarm window width: –300 to +280 mmHg
The OD in the LD senses dark fluid:
Alarm window width: 120 mmHg, adjustable over the entire
display range
With single needle dialysis and single needle click-clack dialysis:
the OD in the LD senses either dark or clear fluid; upper limit fixed
to 280 mmHg; lower limit adjustable from 0 to –300 mmHg.
Monitoring of the arterial pressure.
– Air detector: Threshold of response: drop in the fluid level, air bubbles, froth or
microfroth (blood/air mixture) over the entire blood flow range from
0 up to 600 ml/min.
An alarm will be initiated when 5 ml of air over a rate of
600 ml/min are detected.
The indication of the air volume here refers to microbubbles with a
diameter <10 µm when using the tubing systems approved for the
machine.
Procedure: ultrasonic transmission
Audible alarm:
– Setting range of the Adjustable with the volume control on the rear of the monitor,
loudness of the minimum setting: ≥ 65 dBA
audible alarm:
Only environmentally compatible and recyclable materials are used for the manufacture of hemo-
dialysis machines.
Prior to the disposal of hemodialysis machines, any possible risk of infection must be eliminated
by appropriate disinfection.
Electronic boards can be disposed of as set forth in the regulation on the disposal of electronic
scrap.
For the disposal of lead-acid batteries, the respective regulations in force at the time of disposal
apply.
Further information on disposal can be obtained on request.
Materials used:
(Materials accompanied by an asterisk come into contact with water, dialysate and dialysate con-
centrate.)
Plastics
Metals
Aluminum 1.4401*
Sheet metal, Cu55, 1.1203 1.4404*
Sheet metal, DIN 1541, MK75, unhardened Spring steel 1.4568 – DIN 17224
Sheet metal, DIN 1783, AlMg Titanium – TiAl 4 V6*
Sheet metal, DIN 1798, AlMgSi o.5 Spring steel 1.4310 – DIN 59381
DIN 174 – St37K Steel, 1.4300 (for systems)
1.4301 Steel, 9S MnPb, 28K
1.4305 – DIN 178/668 Galvanized steel
Motors Cables
Copper Copper
Cast steel PVC
Cast zinc Teflon
Miscellaneous
Adhesives
Loctite
Scotch-Weld DP 499
Scotch-Weld V 23
Cyanolit
Elastosil
Blood alarm Group of alarms which cause the blood systems to stop:
– Arterial pressure
– Venous back pressure
– Transmembrane pressure
– Blood leak
– Air
– BP stop alarm
Blood systems Systems which maintain and control the function of the extracor-
poreal circuit.
Bubble catcher Device integrated in the blood line for separating gases which
are not dissolved in the blood.
Bypass Process during which the dialysate is diverted away from the
dialyzer.
Compliance chamber Device for levelling variations in pressure and volume, e.g. in the
single needle mode.
Dialysate pressure The pressure present in a defined section of the system carrying
the dialysate, e.g. at the dialyzer outlet.
Dialyzer blood connector Arterial or venous blood connector provided at the dialyzer.
Extracorporeal circuit A section of the blood circulation which takes place outside of the
body.
Occlusion clamp Device for automatically clamping the blood lines, e.g. in the
event of alarms or during single needle operation.
Pressure measurement Connector for the pressure measuring line, e.g. Luer-lock.
connector
Pressure measuring line Line connecting the tubing system to the pressure monitoring
unit.
Product water Water suitable for hemodialysis (e.g. water purified by reverse
osmosis).
Pump line holder Device for fixing the blood pump line segment.
Pump stator Stationary support for the pressure pulleys of the pump rotor.
Single needle method Technique in which the vascular access for the treatment proce-
dure is established by means of one puncture only.
Transmembrane pressure Difference between the pressures acting upon the dialyzer mem-
brane (blood pressure, pressure in the dialysate).
Venous back pressure Pressure in the extracorporeal circuit before the venous cannula
(e.g. in the venous bubble catcher).
Water alarms Group of alarms which do not cause the blood systems to be
turned off: conductivity (bypass operation), temperature (bypass
operation), flow.
1.12.1 Abbreviations
AC = alternating current
BC = balancing chamber
BIC = bicarbonate
BP = blood pump
CD = conductivity
DC = direct current
HD = hemodialysis
HDF = hemodiafiltration
LD = level detector
OD = optical detector
SI = shunt interlock
SN = single needle
SV = solenoid valve
UF = ultrafiltration
Alternating current
Earth terminal
Equipotentiality
I/O ON / OFF
Performance of the technical safety checks must be recorded in the Medical Device Register.
Numbers, which are not listed here, are not included in the TSC, but are part of the maintenance
procedures.
For further details on technical safety checks and maintenance procedures, refer to the Technical
Manual.
TSC checklist
4008 for the Technical Safety Checks
to be performed annually (every 12 months) Fresenius Medical Care
Page 1/2
Customer/Customer no.:
System type:
With option(s):
Page 2/2
No. Description Measured ✓
value
8 Checking the electrical safety
In Germany according to DIN VDE 0751 standard –1st Edition 10/2001.
In other countries, observe the local regulations!
8.1 Visual inspection performed according to item 1. ❏
8.2 Protective earth resistance maximum 0.3 ohms (with power cord) ………… Ω ❏
8.3 Measurement of the leakage current (device leakage current) ❏
❏ Differential current measurement according to fig. C.6
or
❏ Direct measurement according to fig. C.5
Remarks:
1.14.1 EC Certificate
Publications, investigating report on polysulfone membranes, with the enclosed literature refer-
ences.
– “No evidence for endotoxin transfer across high flux polysulfone membranes”, among others
by Professor Dr. med. Jürgen Bommer from the Ludolf-Krehl-Klinik in Heidelberg.
– “Filtration of dialysate using an ON-LINE dialysate filter”, dated June 5, 1990, among others by
Francise Dumler MD, Division of Nephrology & Hypertension, Henry Ford Hospital, 2799 West
Grand Blvd., Detroit, MI, 48202, USA as well as Fresenius brochures on DIASAFE and
polysulfone high-flux with literature.
– “Hemodiafiltration with ON-LINE Production of Bicarbonate Infusate: a new standard for high
efficiency, low-cost dialysis in elderly and uncompliant patients” among other by Dr. B.
Canaud, Division of Nephrology, Lapeyronie University Hospital, 555, Route de Ganges,
34059 Montpellier, France
Chapter Page
2.7 Priming the Extracorporeal Blood Circuit With a Saline Bag .................................. 2-35
2.7.1 Priming ............................................................................................................................ 2-35
2.7.2 Setting the Delivery Rate of the Blood Pump ................................................................. 2-36
2.7.3 Stopping Delivery ............................................................................................................ 2-37
2.7.4 Interrupting the Priming Program .................................................................................... 2-39
2.7.5 Premature Termination ................................................................................................... 2-39
☛ I/O
Luftdetektor Arterieller
Druck
Ve
D
al Care 4008 B
Indication of the machine type and the
** 4008 B / VX.X ** current software version.
Rate: ml/min
The LED segments, the alarm (red) LED
(Ø: mm)
and the operation (green) LED light for
1 second.
Start
Stop
Rate: ml/min
Indication of the line diameter for approx.
(Ø: mm)
2 seconds.
Start
Stop
Rate: ml/min
Indication of the delivery rate.
(Ø: mm)
Start
Stop
Rate: ml/h
(Bolus: ml)
The LED segments, the alarm (red)
LED, the operation (green) LED and the
( :h.min)
stop-watch time LED light for 1 second.
Bolus
Rate
Start
Stop
Rate: ml/h
(Bolus: ml)
Indication of the syringe size.
( :h.min)
Bolus
Rate
Start
Stop
Rate: ml/h
(Bolus: ml)
Indication of the delivery rate.
( :h.min)
Bolus
Rate
Start
Stop
● Acetate Dialysis
Rate: ml/h
(Bolus: ml)
Bolus
Rate
( :h.min)
● Bicarbonate Dialysis
Rate: ml/h
(Bolus: ml)
Bolus
Rate
( :h.min)
Bicarbonate Acid
concentrate concentrate
Caution
Assure that the containers contain
sufficient concentrate and bicarbo-
nate to complete the treatment.
General Notes
☛
Caution
Wait until the cooling rinse program
following a hot rinse (or heat desin-
fection) has been completed before
connecting the bibag ®.
☞ Note
During treatment with the bibag®, the
bicarbonate suction tube must be in-
serted in the rinse port.
Removing the suction tube may
cause a rinse chamber overflow.
☞ Note
The bibag® will be filled when the
bibag ® and the bicarbonate suction
tube have been connected.
The bibag® has priority over bicarbo-
nate supplied by the CDS.
Rate: ml/h
( :h.min)
Bolus
Rate
Start
Stop
tube for acid and acetate in the acid con-
tainer.
Acid
concentrate
( :h.min)
Bolus
Rate
Start
Stop
automatic program will be started
(lasting approx. 35 seconds).
This program will first remove all air from
the bibag® and will then fill the bag with
water.
Acid
concentrate
General notes
☞ Note
When using CDS, connect the con-
centrate suction tubes to match the
selected type of concentrate delivery
(SETUP) (See Chapter 1.3.5, Central
Delivery System).
Default setting in SETUP: CDS off
☞ Note
With bicarbonate and acid being de-
livered via the CDS, the CDS must
be turned off in the user menu when
performing acetate dialysis with con-
tainers.
– The functional test can be performed with or without previously installing the tubing system.
If the tubing system was filled during the cleaning program, the dialyzer must be deaerated
before starting the functional test. This is done by briefly opening and then closing the upper
screw cap of the dialyzer.
– Depending on the setting selected in the SETUP, the functional test will either start automat-
ically or must be started manually with the Test key.
– Starting conditions:
– The dialysate connectors (red/blue) are in the interlock shunt.
– The interlock shunt is closed.
– The optical detector in the air detector module senses clear fluid.
– The concentrate suction tubes are inserted in the appropriate container, or they are in-
serted according to the type of supply with CDS.
Caution
If the XX not calibrated message is
displayed on starting the functional
test, the system must not be used to
perform a treatment.
Override
–30
–200
0
Test
–40 –300
The Test LED is flashing.
Internal tests.
Prime
Dialysis Start
Reset Alarm
Alarm Tone
Mute
Override
–30
–200
0
Test
–40 –300
The functional test is started automatic-
ally.
Prime
The Test LED is illuminated.
Dialysis Start If one of the starting conditions is not
Reset Alarm
fulfilled, an audible alarm will be
sounded.
Alarm Tone
Mute Use the Alarm Tone Mute key to ac-
knowledge the audible alarm, make
sure all starting conditions are fulfilled,
then press the Test key to start the
test.
☞ Note
To prevent an automatic start of the
test proceed as follows:
– Open the shunt door.
– Turn the hemodialysis machine
on.
– Start the functional test as usual.
– Close the shunt door.
Override
–30
–200
0
Test
–40 –300
The Test LED is flashing.
Should the Test LED not be flashing,
Prime one of the conditions required for the
test to start is not fulfilled.
Dialysis Start
Reset Alarm
Alarm Tone
Mute
Override
☛
–30
–200
0
Test
–40 –300
Press the Test key.
Dialysis Start
Reset Alarm
Alarm Tone
Mute
● Basic condition
Override
–30
–200
0
Test
–40 –300
The functional test has been selected.
Dialysis Start
Reset Alarm
Alarm Tone
Mute
● Start
Override
–30
–200
Test
0
Press the Prime key.
–40 –300
or
Confirm Select
Prime / Circulation
Rate: ml/min
The operation (green) LED is illumin-
(Ø: mm)
ated.
The blood pump is delivering.
☛ The arterial blood pump fills the tubing
Start
system up to the venous bubble catcher.
Stop A delivery rate > 180 ml/min will auto-
matically be returned to 180 ml/min.
The operator may select any delivery
rate desired.
● Delivery stops
al Care 4008 B
Indication of the test steps
Test . . .
The red (alarm) status indicator is illu-
Arterial Venous Transmembrane
minated.
Air Detector Ultrafiltrati
Pressure Pressure Pressure
UF Volume ml
Override
–30
–200
Test
0
The Test LED is illuminated.
–40 –300
Prime The Alarm Tone Mute LED is flashing.
Alarm Tone
Mute
Override
–30
–200
0
Test
–40 –300
Press the Dialysis Start key.
The blood pump starts running, and
Prime the venous line clamp is opened. Fluid
is recirculating.
☛
Dialysis Start
Reset Alarm
The Test LED is illuminated.
Alarm Tone
Mute The Prime LED is illuminated.
The Alarm Tone Mute is dark.
al Care 4008 B
Temporary display message
Circulation
The yellow (warning/info) status
Arterial Venous Transmembrane
indicator is illuminated.
Air Detector Ultrafiltrati
Pressure Pressure Pressure
UF Volume ml
Rate: ml/min
Press the Start/Stop key.
(Ø: mm)
Start
Stop
☛
● To stop priming
Override
–30
–200
0
Test
–40 –300
Press either the Dialysis Start or the
Prime key.
☛
Prime
Confirm Select
Test Version X.X
Confirm Select
Test in Progress
Confirm Select
Test Display
Override
–30
–200
0
Test
–40 –300
The Test LED is dark.
Override
–30
–200 The Dialysis Start LED is illuminated.
0
Test
–40 –300
Prime
Dialysis Start
Reset Alarm
Alarm Tone
Mute
Override
–30
–200
0
Test
–40 –300
The Test LED is flashing.
Alarm Tone
Mute
☛
–30
–200
0
Test
–40 –300
Press the Test key.
Prime
The Test LED is flashing.
Dialysis Start
Reset Alarm The Alarm Tone Mute LED is flashing.
Audible signal
Alarm Tone
Mute
Confirm Select
F.. .... Indication of the error code.
Override
☛
–30
–200
0
Test
–40 –300 Press the Test key.
Prime
The Test LED is illuminated.
Dialysis Start
Reset Alarm The Alarm Tone Mute LED is dark.
Alarm Tone
Mute
Confirm Select
Test ??? The test step which the machine failed to
pass is displayed and repeated.
If the test step is successfully passed,
the test is completed.
Confirm Select
F.. .... Should the error recur, it will be indicated
on the display together with the respect-
ive error code.
☞ Note
The following error messages can be
acknowledged by pressing the Test
key:
Skip Battery?
Skip Diasafe?
☞ Note
If the message Skip Battery? is
acknowledged by pressing the Test
key, a power failure eventually might
no longer be signalled by an audible
alarm.
Alarm Tone
Mute
Override
–30
–200
0
Test
–40 –300
Press the Test key.
☛
Override
–30
–200
0
Test
–40 –300
Press the Test key.
– One of the starting conditions listed in Chapter 2.2.2 Start of the Test is no longer fulfilled.
– After the problem has been corrected the test continues.
☛ I/O
Luftdetektor Arterieller
Druck
Ve
D
The test can be prematurely terminated
only by turning the machine off.
Rate: ml/min
(Ø: mm)
➀ Open the blood pump door.
➁ Press and hold the Start/Stop key
until the rotor stops.
Start
Stop
☛
➁
Rate: ml/min
(Ø: mm)
➀ Install the line segment on the left
side.
➁ Press and hold the Start/Stop key
Start until the rotor stops.
Stop
➄ ☛ ➂ Install the line segment on the right
➁ side.
➃ Close the blood pump door.
➄ Connect the pressure measuring line
to the pressure connector.
➀ ➂
➃
Rate: ml/min
If the rotor of the blood pump is deliver-
(Ø: mm)
ing:
Start
Stop
The operation (green) LED is dark.
☛
Rate: ml/min
Simultaneously press the ▲ and ▼ keys.
(Ø: mm)
The current line size is indicated in mm
and flashing digits.
☛
☛
Start
Stop
Rate: ml/min
Use the ▲ key or the ▼ key to select the
(Ø: mm)
desired value.
(Pressing the keys longer than 3 sec-
onds causes the changing rate to ac-
Start
☛ celerate.)
Range of adjustment:
Stop 2 to 10 mm in 0.2 mm increments
Rate: ml/min
Acknowledge the value by pressing the
(Ø: mm)
Start/Stop key.
(The value is indicated for approx.
2 seconds.)
The delivery rate of the blood pump
is set to 0.
Start
Stop
☛
Rate: ml/min
Use the ▲ key to adjust the delivery rate.
(Ø: mm)
☛
Start
Stop
Rate: ml/h
(Bolus: ml)
Press the ▲ key.
The slide carriage moves to the fully
open position.
( :h.min)
Bolus
Rate
☛
☞ Note
When the slide carriage moves to the
fully open position, the bolus data
are reset to 0.
Rate: ml/h
(Bolus: ml)
➀ Connect a syringe filled with heparin
to the heparin line of the arterial line
( :h.min)
set.
Bolus ➁ Insert the syringe.
:h.min)
Note
The plate of the syringe plunger must
engage in the slide carriage.
Bolus
Rate
Start
Stop
☞ Note
Make sure that the syringe wings are
in the syringe wing slot.
Rate: ml/h
(Bolus: ml)
Press the Rate key.
The display is flashing.
( :h.min)
Bolus
Rate
☛ Start
Stop
Rate: ml/h
(Bolus: ml)
Set the desired rate by pressing the
▲ and ▼ keys.
(Pressing the keys longer than 3 sec-
( :h.min)
onds causes the changing rate to ac-
Bolus celerate.)
Range of adjustment: 0.1 to 10 ml/h in
Rate ☛ increments of 0.1 ml/h
Start
Stop
Rate: ml/h
(Bolus: ml)
Press the Rate key to acknowledge the
selected rate.
The display is flashing until the se-
( :h.min)
lected rate is acknowledged by press-
Bolus ing the Rate key.
During this time, the heparin pump de-
Rate livers at the previous rate.
☛
Start
Stop
☞ Note
Administer the heparin dose accord-
ing to the physician's instructions.
☞ Note
For the injection of an initial heparin
dose, use the Bolus key.
Rate: ml/h
(Bolus: ml)
Press the key.
The display is flashing.
( :h.min)
Bolus
Rate
☛ Start
Stop
Rate: ml/h
(Bolus: ml)
Select the desired time by pressing the
▲ and ▼ keys.
(Pressing the keys longer than 3 sec-
( :h.min)
onds causes the changing rate to jump
Bolus from 1-minute to 10-minute incre-
ments.)
Rate ☛ Range of adjustment: 1 minute to
9 hours 59 minutes in increments of
Start 1 minute
Stop
Rate: ml/h
(Bolus: ml)
Press the key to acknowledge the se-
lected time.
The display is flashing until the se-
( :h.min)
lected time is acknowledged by press-
Bolus
ing the key.
During this time, the previous delivery
Rate time remains active.
☞ Note
Place the clot catcher at a level
approximately 1 cm below the lower
edge of the bubble catcher holder.
Caution
Do not use ultrasound-conducting
media on the air detector.
➁
➀
➂
Caution
Do not remove the blood line from
the optical detector during treatment.
Regarding the regulations for the application of hemodialysis machines the following has
to be observed:
Caution
Use aseptic technique for all blood
side connections.
Use hydrophobic filters in the pres-
sure return lines to prevent contami-
nation. A wetted hydrophobic filter
must be replaced!
Caution
When inserting the lines make sure
that the blood lines are not kinked.
Use the line holders provided.
Caution
During the treatment verify at appro-
priate intervals:
– the condition of the patient.
– the function of the hemodialysis
machine and the extracorporeal
blood circuit. Pay particular atten-
tion to the venous insertion site,
as a possible dislocation of the
venous cannula may not always
be detected by the pressure
monitoring system.
Caution
It is important to set the lower ve-
nous alarm limit to be as close as
possible to the actual value of the ve-
nous pressure.
I/O I/O
Luftdetektor Arterieller Venöser TMP Ultrafiltration Leitfähigkeit Luftdetektor Arterieller Venöser TMP Ultrafiltration Leitfähigkeit
Druck Druck Druck Druck
UF Menge ml UF Menge ml
Vorbereiten Vorbereiten
UF Restzeit h:min UF Restzeit h:min
Variation Variation
Alarm I/O Prog. I/O Set Alarm I/O Prog. I/O Set
Ton Aus Ton Aus
Pven. Pven.
( :h.min) ( :h.min)
Bolus Bolus
Rate Rate
Start Start
Stop Stop
2.7.1 Priming
Override
–30
–200
0
Test
–40 –300
Press the Prime key.
☛ Prime
Dialysis Start
Reset Alarm
The Prime LED is illuminated.
Alarm Tone
Mute
al Care 4008 B
The yellow (warning/info) status
Prime indicator is illuminated.
Rate: ml/min
The operation (green) LED is illumin-
(Ø: mm)
ated.
The blood pump is delivering.
☛ The arterial blood pump fills the lines up
Start
to the venous bubble catcher.
Stop A delivery rate > 180 ml/min will auto-
matically be returned to 180 ml/min.
The operator may select any delivery
rate desired.
Display message
Confirm Select Pre-circulation will be started automat-
Preparation ically on completion of the priming pro-
cedure.
Override
–30
–200
0
Test
–40 –300
The Dialysis Start LED is flashing.
Override
–30
–200
0
Test
–40 –300
Press the Dialysis Start key.
The blood pump starts running and the
Prime venous line clamp opens.
☛
Allow fluid to recirculate until all air has
Dialysis Start been removed from the tubing system
Reset Alarm
and the dialyzer.
Alarm Tone
Mute The Dialysis Start LED is illuminated.
The Alarm Tone Mute LED is dark.
Override
–30
–200
0
Test
–40 –300
Should the venous bubble catcher not be
☛
filled after 2 minutes (depending on the
Prime
setting selected in the SETUP), proceed
Dialysis Start as follows:
Reset Alarm
Press the Prime key again.
Alarm Tone
Mute The Prime LED is illuminated.
The Dialysis Start LED is dark.
Rate: ml/min
Press the Start/Stop key.
(Ø: mm)
Start
Stop
☛
2.7.5 Premature Termination
Override
–30
–200
Test
0
Press the Dialysis Start key or the
–40 –300
Prime key.
☛
Prime
Continue priming by pressing the Prime
Dialysis Start
Reset Alarm key again.
Alarm Tone
Mute
Caution
The functional checks must be performed by the operator prior to each dialysis
treatment.
1. Concentrate container:
Use only the dedicated keyed containers for bicarbonate dialysis.
2. Verification of the dialysate composition:
Check the conductivity displayed on the monitor with the specifications on the
acid / concentrate container. (Observe the settings for sodium and bicarbon-
ate in the Dialysate menu!)
Caution
If the mixing ratio was set in the SETUP to be individually adjustable, always
make sure:
1. that the mixing ratio set by the technician complies with the mixing speci-
fications for the concentrates used.
2. to check the dialysate for correct physiological composition before initiating
dialysis.
3. that the individually adjustable mixing ratio and the bibag® option are not
used simultaneously.
Reset
Conc Bic
kPa mmHg
☛
500 15.5
☞ Note
If acetate dialysis (mixing ratio 1+34)
is desired and there is an other mix-
ing ratio set, check the basic condi-
tions:
Acetate supply established. Bicarbo-
nate suction tube in rinse chamber;
bicarbonate CDS switched off;
bibag ® not connected.
Confirm Select
Last set value.
Base Na+ xxx mmol
☛ Adjust the base Na+ by pressing the ▲
and ▼ keys.
(Range of adjustment: 125 to 150 mmol)
The base Na+ is specified on the con-
tainers and/or the package circular.
☞ Note
The respective sodium and bicarbo-
nate ion concentrations are specified
in the corresponding product infor-
mation.
Reset
Conc Bic
kPa mmHg
500 15.5
☛
Confirm Select Basic setting: ±0 (or as defined in the
Bicarbonate x mmol SETUP setting)
☛ Adjust the bicarbonate pump by pressing
the ▲ and ▼ keys.
(Range of adjustment: –8 to +8 mmol)
Confirm Select
Press the Select key.
Preparation
☛
Confirm Select Use the ▲ and ▼ keys to select
7. Treatment param.
☛ 7. Treatment param..
Press the Confirm key.
(25 C)
Tempera-
UF Flow ture
UF Rate ml/h
Use the ▲ key to set the UF Rate to the
desired value.
1000 ☛ Use the ▼ key to set the Time Left to 0.
UF Goal ml
0000
I/O Prog.
☛
☞ Note
When the optical detector senses
clear fluid and the dialysate pressure
is positive during preparation, the UF
rate is automatically set to 500 ml/h.
When entering UF parameters and
starting UF, only the values entered
are relevant.
(25 C)
☛
UF Flow ture
Chapter Page
☞ Note
If the dialyzer has been prepared
using a UF rate, the UF volume must
be erased.
UF Goal ml
Press the Flow I/O key.
0000
0:00
UF
☛ I/O Prog.
● Starting position
0750
UF Goal ml
3000
4:00
UF
I/O Prog.
or
Ultrafiltration
UF Volume ml
0000 Reset
Volume
UF Rate ml/h
0000
UF Goal ml
0000
0:00
UF
I/O Prog.
UF Goal ml
3000
☛
Time Left h:min
4:00
UF
I/O Prog.
UF Goal ml
3000
☛
Time Left h:min
4:00
UF
I/O Prog.
☛
UF Rate ml/h
Use the ▲ and ▼ keys to enter the de-
sired time.
0750
☛
UF Goal ml
0000
I/O Prog.
☛
● Rate function
0000
I/O Prog.
☛
UF Goal ml
☛
0000
I/O Prog.
☛
3.1.4 Starting Ultrafiltration
(25 C)
☛
UF Flow ture
UF Goal ml
4000
1:00
☛
UF
I/O Prog.
(25 C)
☛
UF Flow ture
UF Goal ml
3000
4:00
☛
UF
I/O Prog.
(25 C)
☛
UF Flow ture
Rate: ml/min
Press the Start/Stop key on the blood
(Ø: mm)
pump module.
The operation (green) LED is dark.
The arterial blood pump has stopped.
Start
Stop
☛
Connect the arterial line to the patient.
If infusion of the saline contained in the
line is desired, connect now the venous
line.
Rate: ml/min
Set the delivery rate of the arterial blood
(Ø: mm)
pump to the desired value.
Start
☛
Stop
Caution
When using high-flux dialyzers make
sure that the alarm limits for monito-
ring the TMP are as narrow as possi-
ble.
Override
–30
–200 The blood pump stops.
0
Test
–40 –300 The venous line clamp closes.
Prime
Audible signal
The Dialysis Start LED is flashing.
Dialysis Start
Reset Alarm
The Alarm Tone Mute LED is flashing.
Connect the venous connector of the tu-
Alarm Tone
Mute bing system to the patient (provided this
has not been done yet).
Override
–30
–200
0
Test
–40 –300
Press the Dialysis Start key.
☛ Dialysis Start
Reset Alarm
Alarm Tone
Mute
al Care 4008 B
Display message
Dialysis
The green (operation) status indicator is
Arterial Venous Transmembrane
illuminated.
Air Detector Ultrafiltrati
Pressure Pressure Pressure
UF Volume ml
Caution
When using high-flux dialyzers and
low UF rates are selected there is a
possibility of local backfiltration.
Backfiltration is dependent upon:
– the type of high-flux dialyzer
used,
– the difference in the flow resi-
stance on the dialysate side and
the blood side,
– the viscosity of the blood.
Rate: ml/h
(Bolus: ml)
Press the Start/Stop key on the heparin
pump.
( :h.min)
The operation (green) LED is illumin-
Bolus ated.
The heparin pump is delivering.
Rate ☛
Start
Stop
(25 C)
☛
UF Flow ture
Rate: ml/h
(Bolus: ml)
To check the bolus volume:
Press the Bolus key on the heparin
( :h.min)
pump module.
☛ Bolus
Rate
The bolus volume already adminis-
tered will be indicated.
Start
Stop
Rate: ml/h
(Bolus: ml)
To administer the bolus:
Press the Bolus key and hold it for more
( :h.min)
than 2 seconds.
☛ Bolus
Rate
Indication: 0.0
The heparin pump delivers the bolus in
0.1 ml increments.
Start After the key has been released, the
Stop display still indicates the bolus volume
delivered up to that point for another
5 s.
(The bolus administration is added to
the stored bolus volume.) Then the de-
livery rate is indicated.
– Maximum bolus administration: 5 ml
– Maximum total bolus:
1x syringe capacity
3.3.1 Preparation
Start
Stop
☞ Note
Do not install the single needle pump
line segment!
Start
☛
Rate: ml/h Rate: ml/min
Use the ▲ or ▼ key to adjust the desired
(Bolus: ml) (Ø: mm)
stroke volume.
Range of adjustment: 10 to 50 ml in
( :h.min)
5 ml increments.
Bolus
Start
☛
Rate Stop
Start
Start
Rate Stop
Start
☛
Rate: ml/min
Press the Start/Stop key on the blood
(Ø: mm)
pump module.
The operation (green) LED is dark.
The arterial blood pump has stopped
running.
Start
Stop
☛
Connect the arterial line to the patient.
If infusion of the saline contained in the
line is desired, connect now the venous
line.
Rate: ml/min
Use the ▲ or ▼ key to adjust the de-
(Ø: mm)
livery rate of the arterial blood pump.
Start
☛
Stop
Start
Rate Stop
Start
Stop
Start
☛ pump: 10 to 15% higher than that of
the arterial blood pump.
Rate Stop
Start
Rinse
Press the Single Needle key.
kPa mmHg kPa
The Single Needle LED is illuminated.
Hotrinse 280
Blood Leak 30 60
200
20 50
Disinfection
10
100 40
☛
0 0 30
Single
Needle –10
–100 20
–20
Override
–30
–200
0
Test
–40 –300
Press the Dialysis Start key.
☞ Note
If single needle dialysis is performed,
it is important to dialyze with the
maximum blood flow.
The effective blood flow can be dis-
played in the user menu under
option 4. Bloodflow. calc.
This procedure should be used in exceptional cases only, since the stroke volumes and, thus, the
respective recirculation may vary widely.
Both the arterial and the venous blood line segments are to be connected to the vascular access
by means of a Y-piece.
Confirm Select
Use the ▲ and ▼ keys to set the lower
set lower ven. Limit value in the Venous Pressure window.
☛ Press the Confirm key.
–20 10
100 10
100
Override
–200
–30 0 0 0 0
–40 –300
–20 10
100 10
100
Override
–200
–30 0 0 0 0
–40 –300
This treatment mode can be started at any time and can be repeated as often as necessary.
The parameters entered at the beginning of the treatment (UF Goal and Time Left) must be taken
into consideration.
In general, the total removal (UF Goal) as well as the total treatment time (Time Left) or UF Goal
and UF Rate must be programmed. The UF monitor is off.
Confirm Select
ISO-UF value
UF Rate ml/h
UF Goal ml
1000
☛
Time Left h:min
0:30
UF
I/O Prog.
☛
Confirm Select
Display message
Total: V=5000 t=4.00 e.g.:
☛ 5000 total withdrawal (UF Goal)
4:00 h total time (Time Left)
Confirm Select
Display message
ISO-UF start ?
Ultrafiltration
Press the UF I/O key.
UF Volume ml
The UF LED is illuminated.
0000 Reset
Volume
UF Rate ml/h
UF Goal ml
5000
4:00
☛
UF
I/O Prog.
UF Rate ml/h
UF Goal ml
5000
3:59
UF
I/O Prog.
Reset
Conc Bic
kPa mmHg
50 400 15
40 300 14.5
30
200 14
20
UF Goal ml
10
100 13.5
0 0 13
mS/cm
(25°C)
I/O Prog.
Tempera-
UF Flow ture
☞ Note
After completion of sequential dia-
lysis, hemodialysis is automatically
continued with dialysate flow and the
computed UF parameters.
The conductivity alarm and the tem-
perature alarm are suppressed for 3
minutes.
● On the UF monitor
UF Goal ml
5000
3:29
☛
UF
I/O Prog.
● In the menu
UF Rate ml/h
UF Goal ml
1000
☛
Time Left h:min
0:30
UF
I/O Prog.
☛
(25 C)
The Flow LED is dark.
Time Left h:min
Variation
UF Flow
Tempera-
ture
Display message
I/O Prog. I/O Set
☛
UF Rate ml/h
2000
UF Goal ml
1000
0:30
☛
UF
I/O Prog.
10
100 13.5
0 0 13
mS/cm
(25°C)
I/O Prog.
Tempera-
UF Flow ture
Before selecting profiles, the following values should have been entered:
– Base Na+, prescr. Na+ and bic.
– UF Goal (minimum UF goal: 200 ml)
– Time Left (minimum UF profile time: 2 hours)
– UF Rate:
– Minimum continuous UF rate:
Profiles 1 to 6 100 ml/h
– Maximum continuous UF rate:
Profiles Rate (ml/h)
1 3000
2 3000
3 2660
4 2050
5 2050
6 1250
Sodium
1 2 3 4 5 6
1
Ultrafiltration
The machine will prompt the operator in the event of missing or incorrectly entered data.
Confirm Select
Max. UF-Rate reached
When selecting a profile, the machine will prompt the operator to enter missing or to correct
incorrect data.
(25 C)
Tempera-
UF Flow ture
☛
Confirm Select Use the ▲ and ▼ keys to select the
UF Profile Nr. X desired profile number (1 to 6).
☛ Press the Confirm key.
Confirm
☛ Select Use the ▲ key to set the desired value.
Start Na+ 138 mmol
The start Na+ must be higher than the
prescribed Na+ (by at least 1 mmol).
Start
Start Na+
Soll Na+
Prescribed
Base Na+
Basis
(container value)
(Kanisterwert)
(25 C)
UF
I/O Prog.
Flow
I/O
Tempera-
ture
Set
☛
Start
Start Na+
Na+
Soll Na+
Prescribed
Base Na+
Basis
(container value)
(Kanisterwert)
Confirm Select
Press the Confirm key.
Start Na+ xxx mmol
☛
Confirm Select If no UF profile is desired:
UF Profile Nr. X
☛ Use the ▲ and ▼ keys to select
UF Profile Off.
☞ Note
Missing initial parameters are in-
quired when entering a profile.
(25 C)
☛
Tempera-
UF Flow ture The UF and Variation LEDs are illumin-
I/O Prog. I/O Set ated (depending on the preselected pro-
☛ file).
The TMP and CD limits will automat-
or ically track up.
Manual changes are still possible.
(25 C)
UF Flow ture
or
al Care 4008 B
Display message
UF-Goal reached
The yellow (warning/info) status indi-
Arterial Venous Transmembrane
cator is illuminated.
Air Detector Ultrafiltrati
Pressure Pressure Pressure
UF Volume ml
The UF and Variation LEDs are dark
R
(depending on the preselected profile).
The UF pump is not running.
Audible signal
Override
–30
–200
0
Test
–40 –300
The Dialysis Start LED is flashing.
Override
–30
–200
0
Test
–40 –300
Press the Dialysis Start key.
Prime
The Dialysis Start LED is illuminated.
☛ Dialysis Start
Reset Alarm
Alarm Tone
Mute
The Alarm Tone Mute LED is dark.
(25 C)
☛
Tempera-
UF Flow ture
al Care 4008 B
Profiles paused
Display message
The yellow (warning/info) status indi-
Air Detector Arterial
Pressure
Venous
Pressure
Transmembrane
Pressure
Ultrafiltrati cator is illuminated.
UF Volume ml
☛ Dialysis Start
Reset Alarm
Alarm Tone
Mute
Profiles paused
Confirm Select
Display message
The program has been stopped.
(25 C)
Tempera-
UF Flow ture
or
Select
Stop UF-Profile ?
Confirm
Safety prompt
in case of UF and Na+ profiles or UF pro-
or file only
Confirm Select or in case of Na+ profile.
Stop Na-Profile ? (brief indication)
The stopped program is automatically
continued.
(25 C)
☛
Tempera-
UF Flow ture The UF and/or Variation LEDs are
I/O Prog. I/O Set flashing.
(depending on the preselected profile)
☛
or UF stop, bypass, CD value is preserved.
Profiles paused
Confirm Select Display message
(25 C)
Tempera-
UF Flow ture
or
Stop UF-Profile ?
Confirm Select Safety prompt
in case of UF and Na+ profile or UF pro-
or file only
Confirm Select
or in case of Na+ profile.
Stop Na-Profile ? (brief indication)
(25 C)
Tempera-
UF Flow ture
Ultrafiltration is continued at continuous
I/O Prog. I/O Set
rate.
☛
Chapter Page
4.9 Error Messages When Using the Download Function ............................................. 4-50
–20 10
100 10
100
Override
–200
–30 0 0 0 0
–40 –300
Override
–30
–200
0
Test
–40 –300
Press the Dialysis Start key.
Prime
☛ Dialysis Start
Reset Alarm
Alarm Tone
Mute
☛
limits will be retained.
Dialysis Start
Reset Alarm
Alarm Tone
Mute
–20 10
100 10
100
Override
–200
–30 0 0 0 0
–40 –300
–20 10
100 10
100
Override
–200
–30 0 0 0 0
–40 –300
(25 C)
2x
–20 10
100 10
100
Override
–200
–30 0 0 0 0
–40 –300
280
The position of the TMP alarm window
500 500
Blood Leak 30
200 60 60 may automatically be shifted up during
20 50 400 50 400 treatment.
10
100 40 300 40 300
0 0 30 30
200 200
–10
–100 20 20
–20 10
100 10
100
Override
–200
–30 0 0 0 0
–40 –300
–20 10
100 10
100
Override
–200
–30 0 0 0 0
–40 –300
–20 10
100 10
100
Override
–200
–30 0 0 0 0
–40 –300
Confirm Select
Press the Confirm key.
set art. Limits ☛
4-6 Fresenius Medical Care 4008 B 16/10.03 (OP)
4.1.4 Changing the Venous Alarm Window Size
–20 10
100 10
100
Override
–200
–30 0 0 0 0
–40 –300
–20 10
100 10
100
Override
–200
–30 0 0 0 0
–40 –300
Confirm Select
Press the Confirm key.
Set ven. limits
☛
Fresenius Medical Care 4008 B 16/10.03 (OP) 4-7
4.1.5 Changing the TMP Alarm Limits
● Window position
–20 10
100 10
100
Override
–200
–30 0 0 0 0
–40 –300
Confirm Select
Use the ▲ and ▼ keys to select
TMP window position ☛ TMP window position.
Press the Confirm key.
Confirm Select
Use the ▲ and ▼ keys to shift the po-
Set TMP Limits ☛ sition of the TMP window.
–20 10
100 10
100
Override
–200
–30 0 0 0 0
–40 –300
● Window size
Caution
Should the user change the window
size, it is important to select as small
a window size as possible to ensure
adequate safety.
–20 10
100 10
100
Override
–200
–30 0 0 0 0
–40 –300
Confirm Select
Use the ▲ and ▼ keys to select
TMP window size TMP window size.
☛ Press the Confirm key.
Confirm Select
Use the ▲ and ▼ keys to change the
Set TMP Limits size of the TMP window.
☛ Double needle: ±30 to ±80 mmHg
Single needle: ±40 to ±90 mmHg
SN click-clack: ±40 to ±90 mmHg
–20 10
100 10
100
Override
–200
–30 0 0 0 0
–40 –300
Confirm Select
Press the Confirm key.
Set TMP Limits
☛
☞ Note
Automatic adjustment of the alarm
limits is now deactivated.
Confirm Select
Use the ▲ and ▼ keys to select
TMP window size TMP window size.
☛ 2x Press the Confirm key twice.
This procedure allows adjustment of the needle without false alarms. Both the arterial and the
venous alarm limits are inactive for 2 minutes. The full scale values as well as the lower venous
alarm limit of 20 mmHg are still active.
4.2.1 Start
–20 10
100 10
100
Override
–200
–30 0 0 0 0
–40 –300
Caution
When overriding a safety system the
responsibility for the patient’s safety
rests with the operator of the ma-
chine.
Fresenius Medical Care 4008 B The red (alarm) status indicator is illu-
minated.
The Dialysis Start LED is flashing.
I/O
Air Detector Arterial V The Alarm Tone Mute LED is flashing.
Pressure Pr
The blood pump(s) stop(s).
Rinse
kPa mmHg kPa The venous line clamp closes.
Hotrinse 280
Blood Leak 30
200 60 The UF pump(s) stop(s).
20 50
Disinfection
10
100 The nurse call is active.
40
0 0 30 Audible signal
Single
Needle –10
–100 20
–20 10
Override
–30
–200
0
Test
–40 –300
Prime
Dialysis Start
Reset Alarm
Alarm Tone
Mute
Cause:
kPa mmHg kPa mmHg kPa mmHg
The arterial pressure is outside the alarm
280 500 500
Blood Leak 30 60 60
limits.
200
20 50 400 50 400
10
100 40 300 40 300
0 0 30 30
200 200
–10
–100 20 20
–20 10
100 10
100
Override
–200
–30 0 0 0 0
–40 –300
Override
–30
–200
0
Test
–40 –300
Press the Dialysis Start key.
Prime
☛
Dialysis Start
Reset Alarm
Alarm Tone
Mute
Display message
New art limits? The alarm limits are flashing for 8 sec-
onds.
Limits are inactive.
Arterial Venous Transmembrane Ultrafiltra
Pressure Pressure Pressure
UF Volume ml
Override
–30
–200
0
Test
–40 –300
Press the Dialysis Start key again.
If the key is not pressed within approx.
Prime
8 seconds, the original limits will be re-
–20 10
100 10
100
Override
–200
–30 0 0 0 0
–40 –300
–20 10
100 10
100
Override
–200
–30 0 0 0 0
–40 –300
Caution
If the filter in the pressure measuring
line is wetted, disconnect the line
and use a syringe to purge the filter
or replace the filter.
Cause:
kPa mmHg kPa mmHg kPa mmHg
The venous pressure is outside the
280 500 500
Blood Leak 30 60 60
alarm limits.
200
20 50 400 50 400
10
100 40 300 40 300
0 0 30 30
200 200
–10
–100 20 20
–20 10
100 10
100
Override
–200
–30 0 0 0 0
–40 –300
Override
–30
–200
0
Test
–40 –300
Press the Dialysis Start key.
Prime
☛ Dialysis Start
Reset Alarm
Alarm Tone
Mute
Display message
new
New ven
ven limits
limits?? The alarm limits are flashing for 8 sec-
onds.
Limits are inactive.
Arterial Venous Transmembrane Ultrafiltra
Pressure Pressure Pressure
UF Volume ml
Override
–30
–200
0
Test
–40 –300
Press the Dialysis Start key again.
If the key is not pressed within approx.
Prime
8 seconds, the original limits will be re-
–20 10
100 10
100
Override
–200
–30 0 0 0 0
–40 –300
–20 10
100 10
100
Override
–200
–30 0 0 0 0
–40 –300
–20 10
100 10
100
Override
–200
–30 0 0 0 0
–40 –300
Override
–30
–200
Test
0
Press the Dialysis Start key.
–40 –300
Prime
☛ Dialysis Start
Reset Alarm
Alarm Tone
Mute
Display message
New TMP
ven limits?
limits? The limits are flashing for a period deter-
mined by the calculated UFC.
Limits are inactive.
Arterial Venous Transmembrane Ultrafiltra
Pressure Pressure Pressure
UF Volume ml
Override
–30
–200
0
Test
–40 –300
Press the Dialysis Start key again.
If the key is not pressed within approx.
Prime
8 seconds, the original limits will be re-
–20 10
100 10
100
Override
–200
–30 0 0 0 0
–40 –300
–20 10
100 10
100
Override
–200
–30 0 0 0 0
–40 –300
Caution
Sudden TMP changes are indicative
of defect.
Override
–30
–200
0
Test
–40 –300
Press the Dialysis Start key.
See also Chapter 4.5 Warnings,
Prime Message Fillprogram
☛ Dialysis Start
Reset Alarm
Alarm Tone
Mute
I/O
Air Detector Arterial V
Pressure Pr The Blood Leak alarm indicator is flash-
ing.
Rinse
kPa mmHg kPa Cause:
Hotrinse 280 Dialyzer membrane rupture.
Blood Leak 30 60
200
Disinfection
20 50 Press the Override key.
100
10 40 Override time: always 2 minutes
0 0 30
As soon as the leak closes, the alarm in-
Single
Needle –10
–100 20
dicator turns dark.
–20 10
Override If the problem persists, replace the dia-
–30
–200
0
Test
–40 –300 lyzer.
☛
Rate: ml/min
The Alarm LED (red) on the blood pump
(Ø: mm)
module is illuminated.
E.XX
Error message E.XX on the blood pump.
Start
Stop
Rate: ml/min
(Ø: mm)
Press the Start/Stop key.
E.XX
Start
Stop
☛
Override
–30
–200
0
Test
–40 –300
Press the Dialysis Start key.
Prime
☛ Dialysis Start
Reset Alarm
Alarm Tone
Mute
☞ Note
If the problem persists, call the
service.
Rate: ml/h
(Bolus: ml)
The Alarm LED (red) on the heparin
pump is illuminated.
E.XX
( :h.min)
Error message E.XX on the heparin
Bolus pump.
Rate
Start
Stop
Override
–30
–200
0
Test
–40 –300
Press the Dialysis Start key.
Prime
☛ Dialysis Start
Reset Alarm
Alarm Tone
Mute
☞ Note
If the problem persists, call the
service.
I/O
Air Detector Arterial V
Pressure Pr The Air Detector alarm indicator is
flashing.
Rinse
kPa mmHg kPa Cause:
Hotrinse 280 Air or frothy blood in the venous bubble
Blood Leak 30 60
200 catcher.
20 50
Disinfection
10
100 40
0 0 30
Single
Needle –10
–100 20
–20 10
Override
–30
–200
0
Test
–40 –300
Override
–30
–200
0
Test
–40 –300
Press the Dialysis Start key.
Prime
☛ Dialysis Start
Reset Alarm
Alarm Tone
Mute
Confirm Select
BP-stop-alarm Display message
Rate: ml/min
The Alarm LED (red) on the blood pump
(Ø: mm)
module is illuminated.
Cause:
The rotor has stopped for more than 30
(15) seconds.
Start
Stop
Override
–30
–200
Test
0
Press the Dialysis Start key.
–40 –300
Prime
☛ Dialysis Start
Reset Alarm
Alarm Tone
Mute
Rate: ml/min
The Alarm LED (red) on the arterial
(Ø: mm)
blood pump or the SN blood pump mod-
ule is illuminated.
Cause:
Start
The rotor of one of the blood pumps is
Stop not running.
The response time is determined by
the speed.
☛
0 0 30
Single
Needle –10
–100 20
–20
Start
Stop
Start
Stop
☛
Override
–30
–200
0
Test
–40 –300
Press the Dialysis Start key.
☛ Dialysis Start
Reset Alarm
Alarm Tone
Mute
Start
Rate Stop
Start
Stop
Rate: ml/min
Stop the arterial blood pump by pressing
(Ø: mm)
the Start/Stop key.
The operation LED is dark.
Start
Stop
☛
Start
Rate
➁ Stop
Start
Stop
➀
Start
Rate ➀ Stop
Start
Stop
☛
0 0 30
Single
Needle –10
–100 20
–20
Rate: ml/min
Use the ▲ and ▼ keys on the arterial
(Ø: min)
blood pump to readjust the delivery rate.
Start the arterial blood pump by pressing
the Start/Stop key.
Start
Stop
The operation LED is illuminated.
☛
Override
–30
–200
0
Test
–40 –300
Press the Dialysis Start key.
Alarm Tone
Mute
–20 10
100 10
100
Override
–200
–30 0 0 0 0
–40 –300
Override
–30
–200
0
Test
–40 –300
Press the Dialysis Start key.
Alarm Tone
Mute
Fresenius Medical Care 4008 B The red (alarm) status indicator is illu-
minated.
The Alarm Tone Mute LED is flashing.
Dialysis Start
Reset Alarm
Alarm Tone
Mute
Transmembrane
The Conductivity alarm indicator is illu-
Ultrafiltration Conductivity
Pressure
UF Volume ml
minated.
Reset
Bypass mode
Conc Bic
kPa mmHg
Cause:
500 UF Rate ml/h 15.5 The actual conductivity value is outside
60
I/O Prog.
Tempera-
UF Flow ture
or
Confirm Select Display message
Water alarm
Cause:
Water supply problem.
Action required:
Check the water supply.
Call the service.
0 0 30
Single
Needle –10
–100 20
–20 10
Override
–30
–200
0
Test
–40 –300
Prime
Dialysis Start
Reset Alarm
Alarm Tone
Mute
or
0 0 30
The blood system and the dialysate sys-
Single
Needle –10 tem are inactive.
–100 20
–20
Override
–200
10
Audible signal
–30 0
Test
–40 –300
Prime
or
Dialysis Start
Reset Alarm
Alarm Tone
Mute
Flow - off warning The dialysate flow Press the Dialysis 30 min 30 min
has been switched Start key (unless flow 45 min 45 min
off. is desired at the mo- 60 min 60 min
ment), or switch on (SETUP) (SETUP)
the dialysate flow.
Heparin pump not on The heparin pump Press the Dialysis Approx. –
module is not Start key (no 11 min
switched on. heparinization), or
switch on the
heparin pump.
Profiles paused Profiles have been Restart the profiles, 10 min 10 min
(not applicable for stopped. or continue to stop
4008 B machines with- them by pressing
out profiles) the Dialysis Start
key.
Shunt Cover open The interlock shunt Close the interlock 1 min –
has been opened. shunt.
al Care 4008 B
The red (alarm) status indicator is illu-
Emergency Operation minated.
Hydraulics off.
UF pump(s) stopped.
The extracorporal blood circuit is main-
tained.
Override
–30
–200
0
Test
–40 –300
The Dialysis Start LED is illuminated.
Dialysis Start
Reset Alarm
Alarm Tone
Mute
☛ Alarm Tone
Mute
The Alarm Tone Mute LED turns dark.
No audible signal
Emergency operation is possible for
approx. 15 minutes.
Should power supply not return, the
treatment must be stopped.
As soon as power has returned, the
program can be continued as usual.
Caution
In the event of a prolonged power
failure, use the hand crank to manu-
ally return the blood to the patient.
For manual reinfusion, remove the
venous patient line from the venous
line clamp. Visually check the line for
air (emergency operation)!
In Single Needle mode, also remove
the BP segment for the BP (SN) from
the blood pump!
Override
–20 10 The Alarm Tone Mute LED is flashing.
–30
–200
0
Test
–40 –300 The dialysate system is inactive.
Prime Audible signal
Dialysis Start
Reset Alarm
Alarm Tone
Mute
Arterial pressure Arterial pressure outside the Check the tubing system.
(only while priming/ alarm limits. Connect the tubing system.
precirculation is in Purge the hydrophobic filter.
progress during the Increase the blood pump
cleaning programs) speed (>100 ml/min).
Call the service.
Bic line not conn The bicarbonate connector is Insert the bicarbonate connec-
not inserted in the rinse cham- tor in the rinse chamber.
ber.
Blood Sensed The air detector detects blood Check the optical detector.
by OD at the start of a cleaning pro- Call the service.
gram.
Bypass Valve failed The air detector detects blood Can be acknowledged by
at the start of a cleaning pro- pressing the respective clean-
gram. ing program key.
Call the service.
Conc line not conn The concentrate connector is Insert the concentrate connec-
not inserted in the rinse cham- tor in the rinse chamber.
ber.
Disinfectant empty ? Disinfection valve sensor 115 Check the disinfectant con-
does not sense any conductiv- tainer.
ity. Can be acknowledged by
Concentrate level sensor does pressing the Disinfection key.
not detect fluid. Call the service.
Disinf-Temp. to high After the rinsing clear cycle, The rinse-clear cycle con-
the temperature is higher than tinues until the temperature
40 °C. has dropped below 40 °C.
Automatically clearing.
Call the service.
Internal Error XXX Run time error in the program. Press the I/O switch to toggle
the machine off and on again.
Call the service.
Level Detect. failed Air detector test failed. Problem cannot be corrected
(only while priming/ in the cleaning program
precirculation is in (precirculation impossible).
progress during the After cleaning: perform a func-
cleaning programs) tional test.
Call the service.
opt. Detector failed Optical detector test failed. Problem cannot be corrected
(only while priming/ in the cleaning program
precirculation is in (precirculation impossible).
progress during the After cleaning: perform a func-
cleaning programs) tional test.
Call the service.
Power Failure Power failure during operation. After power has returned, the
resp. program can be con-
tinued as usual.
Rinse Failure F02 V84 error. Press the I/O switch to toggle
the machine off and on again.
Call the service.
In disinfection program In disinfection program
(PGM 5): Concentrate level (PGM 5): Connect disinfectant.
sensor does not detect fluid.
Rinse Failure F04 The float switch does not Press the I/O switch to toggle
recognize any fluid. the machine off and on again.
Call the service.
Rinse Failure F05 PSW 124 does not open. Call the service.
Observe the pressure exerted
on the ring line (no more than
500 mbar).
Rinse Failure F06 PSW 123 does not open. Call the service. Observe the
pressure exerted on the ring
line (no more than 500 mbar).
Rinse Failure F07* Pressure loss on PSW 124. Call the service.
Rinse Failure F08* Pressure loss on PSW 123. Call the service.
* This error message may be indicated alternately with the message DO NOT SWITCH OFF !!
Turn the machine off after 1 min only!
Rinse Failure F09 PSW 123/124 does not open Call the service.
after pressure reduction.
Rinse Failure F11 Mandatory rinse 3 min before Call the service.
end. PSW 123 has not opened
after pressure reduction.
Rinse Failure F14 Flow problems in the rinsing- Call the service.
area test.
Rinse Failure F21 Maximum number of strokes Press the I/O switch to toggle
for drawing in disinfectant the machine off and on again.
exceeded. Replace the disinfectant con-
tainer.
Call the service.
Rinse required ! The mandatory rinse was Select the rinse program.
stopped.
Shunt cover defect Dialyzer lines connected to Change the blood lines.
alternating with dialyzer although the hemo- Check the dialyzer couplings.
Change blood lines ! dialysis machine is in the Check the interlock shunt.
(only with priming/ cleaning mode. (If cleaning can be started al-
precirculation is in though the two dialyzer coup-
progress during the lings are not in the interlock
cleaning programs) shunt: Call the service.)
Upper Flow Alarm The cleaning flow rises above Can be acknowledged by press-
1000 ml/min. ing the respective cleaning pro-
gram key.
Call the service.
V91/V100 Failure The valve does not open. Call the service.
V99 Failure The valve does not open. Call the service.
Venous pressure? Venous pressure outside the Check the tubing system.
(only while priming/ alarm limits. Connect the tubing system.
precirculation is in Purge the hydrophobic filter.
progress during the Increase the blood pump
cleaning programs) speed (>100 ml/min).
Call the service.
Voltage Failure One of the voltages is outside Acknowledge with the respect-
of limits (+12 V, +24 V). ive cleaning program key.
Press the I/O switch to toggle
the machine off and on again.
Call the service.
Water alarm Float switch for t>30 s high. Check the water supply.
Float switch for t>10 s low. Cleared automatically.
Call the service.
HPU Error Malfunction in hydraulic Use the I/O key to turn the
processing unit. machine off and on.
Call the service.
Keyboard Error The key on the monitor is Use the I/O key to turn the
pressed upon power up. machine off and on.
Short-circuit on the keyboard. Call the service.
Watchdog Error This error message will only Use the I/O key to turn the
be displayed immediately machine off and on.
after turning power on. Call the service.
Download Error F001 Optical detector does not sense Ensure basic requirements are
clear fluid. fulfilled. Restart data transfer.
Download Error F002 Data transfer disabled or im- Call the service.
possible.
(DIP switch 1 LP763 set to OFF)
Download Error F003 Patient key not connected. Connect the patient key.
Restart data transfer, if
necessary.
Download Error F004 CRC check for communication Call the service.
inactive.
Download Error F007 Not in Preparation mode. Ensure basic requirements are
fulfilled. Restart data transfer.
Download Error F008 Illegal patient name Enter the patient name.
(19 blanks) Restart data transfer.
Download Error F011 Illegal UF or Na+ profile number. Check external data.
Restart data transfer.
Download Error F013 Value range for Base Na + or Check external data.
prescribed Na+ exceeded or Restart data transfer.
plausibility ±13 mmol.
Download Error F014 Value range for Start Na+ Check external data.
exceeded. Restart data transfer.
Download Error F015 Value range for bicarbonate Check external data.
exceeded. Restart data transfer.
Download Error F019 Start Na+ plausibility check. Check external data.
Restart data transfer.
Download Error F020 Block check UF rate, UF goal, Check external data.
UF time, UF profile. Restart data transfer.
Na+ profile, Base Na+,
prescribed Na+ , start Na+.
Download Error F022 Na+ profile data plausibility. Check external data.
Restart data transfer.
Download Error F023 Invalid CRC of the patient key. Check patient key.
Restart data transfer.
Chapter Page
al Care 4008 B
The yellow (warning/info) status indi-
UF-Goal reached cator is illuminated.
–20 10
100 10
100
Override
–200
–30 0 0 0 0
–40 –300
Override
–30
–200
0
Test
–40 –300
The Dialysis Start LED is flashing.
Alarm Tone
Mute
Override
–30
–200
0
Test
–40 –300
Press the Dialysis Start key.
Bolus
Start
Rate Stop
Start
☛
Rate: ml/h
(Bolus: ml)
Rate: ml/min
(Ø: mm)
➀ Open the cover.
( :h.min) ➁ Press the Start/Stop key, until the ro-
Bolus tor stops.
Start
Rate Stop
Start
Stop
☛
➁
Rate: ml/h
(Bolus: ml)
Rate: ml/min
(Ø: mm)
➀ Remove the line segment on the left
side.
( :h.min)
➀ ➂
☛
0 0 30
Single
Needle –10
–100 20
–20
Confirm Select
Press the Confirm key.
Reinfusion ?
☛
al Care 4008 B The yellow (warning/info) status indica-
tor is illuminated.
The blood pump stops.
–20 10
100 10
100
Override
–200
–30 0 0 0 0
–40 –300
Rate: ml/min
Use the ▼ key to reduce the delivery
(Ø: mm)
rate of the blood pump.
Start
Stop
☛
☛ Dialysis Start
Reset Alarm
Alarm Tone
Mute
The Dialysis Start LED is illuminated.
Override
–30
–200
0
Test
–40 –300 Press the Dialysis Start key.
Prime The Dialysis Start LED is illuminated.
☛ Dialysis Start
Reset Alarm
Alarm Tone
Mute
Dialysis Start
Reset Alarm
Alarm Tone
Mute
Confirm Select
Reinfusion
Override
–30
–200
0
Test
–40 –300
Restart reinfusion by pressing the
Dialysis Start key.
Prime
The Dialysis Start LED is illuminated.
☛ Dialysis Start
Reset Alarm
Alarm Tone
Mute
Override
–30
–200
Test
0
Press the Dialysis Start key.
–40 –300
Prime The Dialysis Start LED is illuminated.
☛ Dialysis Start
Reset Alarm
Alarm Tone
Mute
(25 C)
Rate: ml/min
Switch off the blood pump.
(Ø: mm)
Rate: ml/h
(Bolus: ml)
Switch off the heparin pump.
The operation LED is dark.
( :h.min)
Bolus
Rate
Start
Stop
☛
Rate: ml/min
Switch on the blood pump with reduced
(Ø: mm)
blood flow.
The operation LED is illuminated.
Start
Stop
☛
R
Display message
The blood pump is not running.
The venous line clamp closes.
Audible signal
Empty the bibag® (option).
See Chapter 5.4 Emptying the bibag®
(option).
–20 10
100 10
100
Override
–200
–30 0 0 0 0
–40 –300
Override
–200
Test
–30 0 The Dialysis Start and the Alarm Tone
–40 –300 Mute LEDs are flashing.
Prime
Press the Dialysis Start key.
☛ Dialysis Start
Reset Alarm
Alarm Tone
Mute
Confirm Select
Display message
Preparation
Audible signals are suppressed.
Reinfuse the remaining patient blood,
until an air detector alarm occurs.
Dialysis Start
Reset Alarm
Alarm Tone
Mute
(25 C)
☛
● Via reset volume key
UF Rate ml/h
☛
0800
UF Goal ml
4000
1:00
UF
I/O Prog.
Confirm Select
Press the Confirm key.
Reinfusion ?
☛
–20 10
100 10
100
Override
–200
–30 0 0 0 0
–40 –300
Override
–30
–200
0
Test
–40 –300
The Dialysis Start LED is illuminated.
Rate: ml/h
(Bolus: ml)
Switch off the heparin pump.
The operation LED is dark.
( :h.min)
Bolus
Rate
Start
Stop
☛
Override
–30
–200
0
Test
–40 –300
Press the Dialysis Start key.
☛ Dialysis Start
Reset Alarm
Alarm Tone
Mute
Override
–30
–200
0
Test
–40 –300
Press the Dialysis Start key.
☛
Dialysis Start
Reset Alarm
Alarm Tone
Mute
Fresenius Medical Care 4008 B The red (alarm) status indicator is illu-
minated.
The Prime LED is flashing.
I/O
Air Detector Arterial V The Dialysis Start LED is flashing.
Pressure Pr
The Alarm Tone Mute LED is flashing.
Rinse
kPa mmHg kPa The Air Detector LED is flashing.
Hotrinse 280
Blood Leak 30
200 60 The blood pump is not running.
20 50
Disinfection
10
100 The venous line clamp closes.
40
0 0 30
Disconnect the patient on the venous
Single
Needle –10
–100 20
side.
–20 10
Override
–30
–200
0
Test
–40 –300
Prime
Dialysis Start
Reset Alarm
Alarm Tone
Mute
Override
–30
–200
Test
0
Press the Dialysis Start key.
–40 –300
Prime The Dialysis Start LED lights.
☛
Dialysis Start
Reset Alarm
Alarm Tone
Mute
Override
–30
–200
Test
0
The drain program can be skipped by
–40 –300
pressing the Dialysis Start key a sec-
Prime ond time.
☛
Dialysis Start
Reset Alarm
Alarm Tone
Mute
☞ Note
Always clean the sealing area of the
connector after dialysis to remove
bicarbonate precipitate.
Caution
It is imperative to disinfect the ma-
chine after each dialysis treatment.
☞ Note
If the machine has been idle for more
than 72 hours, it is recommended to
run a disinfection program before ini-
tiating the next treatment.
Chapter Page
Ende
End der Dialyse
of dialysis
Spülen
Rinse Heißreinigen
Hot rinsing Desinfektion
Disinfection (Reinigung)
(cleaning)
–R–
PGM 1: –S– –F–HR–C–
PGM 1: –F–HR–A– PGM 1: –F–D–Z–
–F–D–M–
–R–endl.
PGM 2: –S– endless –F–HR–
PGM 2: –F–HR– PGM 2: –F–HDES–Z–
–F–HDIS–M–
PGM 3: –IHR– PGM 3: –F–D–Z–HR–
–F–D–M–HR–
PGM 4: –IHR–C– PGM 4: –F–HDES–Z–HR–
–F–HDIS–M–HR–
PGM 5: –F–R(V)–Z–
–F–D(F)–M–
Dialyse
Dialysis
PGM Program
R Rinse
R endless Endless rinse
F Rinsing clean
HR Hot rinsing
C Cooling rinse
D Disinfection
D(F) Cleaning
Disinfectant is drawn in from the front (concentrate suction tube).
HDIS Hot disinfection
M Mandatory rinse
IHR Integrated hot rinsing
☞ Note
The remaining time indicated during
the cleaning programs are calculated
values.
Different environmental conditions
may result in correction of the calcu-
lated time.
☞ Note
Applicable for machines with
DIASAFE®plus (option) and advanced
hydraulics only:
When selecting a cleaning program,
a display message will inform the op-
erator when the filter needs to be
changed.
(See Chapter 7.6.1 Indication of a
Filter Change.)
After the treatment, the exterior of the hemodialysis machine must be cleaned with a cleaning and
disinfecting agent. The active substance concentration of the cleaning and disinfecting agents
must not exceed <5 % for amphoteric surfactants and <48 % for alcohol.
(See also Chapter 1.9 Disposables and Consumables.)
Any failure to comply with the basic conditions will be indicated by a message on the display.
Caution
Applicable for DIASAFE®plus (option)
only:
Contact of the filter with organic sol-
vents can affect the properties of the
material of housing, seals and capil-
laries. Safe function can then no lon-
ger be ensured and the manufacturer
shall no longer have any liability.
(See DIASAFE®plus package insert.)
Factory setting: Citrosteril® and Diasteril ® respectively. This setting must be changed by the
technician if a different agent is used.
Caution
During the hot disinfection cycle
scalding water or steam may be
emitted via the vent tubing.
Caution
Applicable for DIASAFE®plus (option)
only:
High surface temperatures on the
DIASAFE ®plus during heat disinfec-
tion.
☛
60
200
20 50
Disinfection
10
100 40
0 0 30
Single
Needle –10
–100 20
–20
al Care 4008 B
The yellow (warning/info) status indi-
PGM 2: –F–HDIS–M– cator is illuminated.
The text display indicates the program
Air Detector Arterial
Pressure
Venous
Pressure
Transmembrane
Pressure
Ultrafiltrati preset in the SETUP menu,
UF Volume ml
R
e.g. PGM 2: —F—HDIS—M—
The program starts automatically after
approx. 5 seconds.
Caution
It is recommended to perform the
check for residual disinfectants at the
drain of the hemodialysis machine.
In case other test methods are app-
lied, the requesting person bears the
responsibility for this decision.
Caution
After disinfection with Puristeril® 340
or Puristeril®plus, the machine must
be checked for residual disinfectant
after the mandatory rinse has been
completed (e.g. potassium iodide pa-
per).
If the test shows a residual concen-
tration of peracetic acid, the rinse
program must be restarted. (In this
case it is not necessary to run the
entire rinse program.)
Caution
After a disinfection with Diasteril®,
the following must be observed:
After termination of the rinse pro-
gram (mandatory rinse), the hemo-
dialysis machine must be checked
for residual disinfectant using the
pH test strips.
If the pH is too high (yellow color
pH ≤ 4.1), run an rinse program for a
minimum of 5 min., then re-deter-
mine the pH.
☛
60
200
20 50
Disinfection
10
100 40
0 0 30
Single
Needle –10
–100 20
–20
Confirm Select
PGM 3: –F–D–M–HR
Confirm Select
PGM 4: –F–HDIS–M–HR
Confirm Select
PGM 5: –F–D(F)–M
Caution
Do not stop a disinfection program
before its completion, as premature
termination will compromise the effi-
cacy of the disinfection.
☛ Rinse
Hotrinse
Blood Leak
kPa
30
mmHg
280
200
kPa
60
20 50
Disinfection
10
100 40
0 0 30
Single
Needle –10
–100 20
–20
>F<HDIS—M— : 0 h 29
Confirm Select The Alarm Tone Mute LED is flashing.
Audible signal
☞ Note
If the mandatory rinse is stopped, the
mandatory rinse time will restart from
the beginning.
☞ Note
Do not remove the concentrate and
bicarbonate suction tubes from their
ports to prevent leakage of fluid
which might then enter the interior of
the machine.
Cleaning program with the cleaning agent (Sporotal® 100) being drawn in from the front through
the concentration suction tube.
The use of cleaning agents with another composition may cause the message Disinfectant
empty ?, although the container may be full.
☞ Note
Applicable for DIASAFE®plus (option) only:
After installation, the DIASAFE®plus may be cleaned with Sporotal® 100 for a
maximum of 11 times.
In machines with basic hydraulics the cleaning procedures with Sporotal® 100
must be documented.
In machines with advanced hydraulics the cleaning procedures with Sporotal®
100 are monitored by the hemodialysis machine.
(See Chapter 7.6.1 Indication of a Filter Change.)
6.5.1 Start
☛
30 60
200
20 50
Disinfection
10
100 40
0 0 30
Single
Needle –10
–100 20
–20
Rate: ml/h
( :h.min)
Bolus
Rate
Start
Stop
the cleaning agent container
(Sporotal® 100).
Cleaning
agent
Caution
Assure that the container used con-
tains sufficient cleaning agent to
complete the rinse program.
Avoid aspiration of air.
Confirm Select
Display message
Please Wait
The yellow (warning/info) status indi-
cator is illuminated.
Cleaning agent is being drawn in.
Confirm Select
Display message
Conc line not conn
The red (alarm) status indicator is illu-
minated.
The Alarm Tone Mute LED is flashing.
Audible signal
Insert the red concentrate suction tube in
the rinse chamber.
Caution
It is recommended to perform the
check for residual disinfectants at the
drain of the hemodialysis machine.
In case other test methods are app-
lied, the requesting person bears the
responsibility for this decision.
Caution
Check the machine for residual disin-
fectant (e.g. with potassium iodide
paper)
If the test shows a residual concen-
tration, the respective rinse program
must be restarted.
(In this case it is not necessary to run
the entire rinse program.)
Caution
During the hot disinfection cycle
scalding water or steam may be
emitted via the vent tubing.
Caution
Applicable for DIASAFE®plus (option)
only:
High surface temperatures on the
DIASAFE®plus during hot rinse pro-
grams!
6.6.1 Start
☛
kPa mmHg kPa
0 0 30
Single
Needle –10
–100 20
–20
al Care 4008 B
The yellow (warning/info) status indi-
PGM 1: –F–HR–C– cator is illuminated.
Confirm Select
PGM 1: —F—HR—C—
☛ Rinse
Hotrinse
Blood Leak
kPa
30
mmHg
280
200
kPa
60
20 50
Disinfection
10
100 40
0 0 30
Single
Needle –10
–100 20
–20
>F<HR—C— : 0 h 29
Confirm Select The Alarm Tone Mute LED is flashing.
Audible signal
☞ Note
Do not remove the concentrate and
bicarbonate suction tubes from their
ports to prevent leakage of fluid
which might then enter the interior
portions of the machine.
6.7.1 Start
☛ Rinse
Hotrinse
Blood Leak
kPa
30
mmHg
280
200
kPa
60
Press the Rinse key.
The Rinse LED is illuminated.
20 50
Disinfection
10
100 40
0 0 30
Single
Needle –10
–100 20
–20
al Care 4008 B
The yellow (warning/info) status indi-
PGM 1: –R– cator is illuminated.
R
preset in the SETUP menu,
e.g. PGM 1: —R—
6.7.2 Sequence
Confirm Select
>R< : x h xx
☛ Rinse
Hotrinse
Blood Leak
kPa
30
mmHg
280
200
kPa
60
20 50
Disinfection
10
100 40
0 0 30
Single
Needle –10
–100 20
–20
Confirm Select
>R< : 0 h 13
>R< : 0 h 13
Confirm Select The Alarm Tone Mute LED is flashing.
Audible signal
☞ Note
Do not remove the concentrate and
bicarbonate suction tubes from their
ports to prevent leakage of fluid
which might then enter the interior
portions of the machine.
☛ Rinse
Hotrinse
Blood Leak
kPa
30
mmHg
280
200
kPa
60
Select a cleaning program, e.g. Rinse.
Press the Rinse key.
20 50
The Rinse LED is illuminated.
Disinfection
10
100 40
0 0 30
Single
Needle –10
–100 20
–20
6.8.2 Start
0 0 30
Single
Needle –10
–100 20
–20 10
Override
–30
–200
0
Test
–40 –300
☛ Prime
Dialysis Start
Reset Alarm
Alarm Tone
Mute
al Care 4008 B
The yellow (warning/info) status indi-
Prime cator is illuminated.
Rate: ml/min
The operation (green) LED is illu-
(Ø: mm)
minated.
The blood pump is delivering.
☛ The arterial blood pump fills the tubing
Start
system up to the venous bubble catcher.
Stop A delivery rate > 180 ml/min will auto-
matically be reset to 180 ml/min. The
delivery rate can be changed by the
user as desired.
al Care 4008 B
Display message
>R< : 0 h 14
The red (alarm) status indicator lights for
Arterial Venous Transmembrane
1 - 5 seconds.
Air Detector Ultrafiltrati
Pressure Pressure Pressure
UF Volume ml
0 0 30
Single
Needle –10
–100 20
–20
Rate: ml/min
Press the Start/Stop key.
(Ø: mm)
Start
Stop
☛
6.8.6 Premature Termination
0 0 30
Single
Needle –10
–100 20
–20 10
Override
–30
–200
0
Test
–40 –300
☛ Prime
Dialysis Start
Reset Alarm
Alarm Tone
Mute
☛ Prime
Dialysis Start
Reset Alarm
Alarm Tone
Mute
☞ Note
The dialysate couplings must be in
the interlock shunt.
6.9.2 Preselection
0 0 30
Single
Needle –10
–100 20
–20 10
Override
–30
–200
0
Test
–40 –300
Prime
☛ Dialysis Start
Reset Alarm
Alarm Tone
Mute
al Care 4008 B
The yellow status indicator (warning/
Prime/Circulation
info) is illuminated.
0 0 30
Single
Needle –10
–100 20
–20 10
Override
–30
–200
0
Test
–40 –300
Prime
☛ Dialysis Start
Reset Alarm
Alarm Tone
Mute
al Care 4008 B
The yellow status indicator (warning/
Circulation
info) is illuminated.
Rate: ml/min
The operation (green) LED is illumin-
(Ø: mm)
ated.
The blood pump is delivering.
☛ Use the ▲ or ▼ key to select the desired
Start
delivery rate.
Stop (Pressing the key for more than 3 sec-
onds accelerates the changing rate.)
al Care 4008 B
Display message
>R< : 0 h 14
The red (alarm) status indicator lights for
Air Detector Arterial
Pressure
Venous
Pressure
Transmembrane
Pressure
Ultrafiltrati
2 seconds.
UF Volume ml
0 0 30
Single
Needle –10
–100 20
–20
Rate: ml/min
Press the Start/Stop key.
(Ø: mm)
Start
Stop
☛
6.9.7 Premature Termination
0 0 30
Single
Needle –10
–100 20
–20 10
Override
–30
–200
0
Test
–40 –300
Prime
☛ Dialysis Start
Reset Alarm
Alarm Tone
Mute
Chapter Page
7.5 Changing the DIASAFE ®plus (option) on Machines with Basic Hydraulics ............ 7-13
7.5.1 Filter Change Criteria ...................................................................................................... 7-13
7.5.2 Changing the Filter ......................................................................................................... 7-13
7.6 Changing the DIASAFE ®plus (option) on Machines with Advanced Hydraulics ...... 7-18
7.5.1 Filter Change Criteria ...................................................................................................... 7-18
7.6.2 Indication of a Filter Change ........................................................................................... 7-18
7.6.3 Overriding a Filter Change .............................................................................................. 7-23
7.6.4 Changing the Filter ......................................................................................................... 7-24
*
CumBd. 3.1 Arterial 4.1 Eff. 5.1 Brightness 6.1 System 7.1 Reset Emptying Dialysis
XX0ml Start? limit bloodflow clock UF data BIBAG
2.2 Sequ. 3.2 Venous 4.2 Total 5.2 Auto–On 6.2 Auto- 7.2 Show
dialysis limits bld.volume program matic on ISO data
2.3 SN– 3.3 TMP 4.3 Clear 5.3 Key- 6.3 Ret. to 7.3 Reset
click–clack limits total vol. click mainprg ISO data
Menu Structures Overview
*
2.5 Needle- 5.5 Filter 7.5 Clear
Adj-Pgm change eff. time
*
2.6 4008 HDF 5.6 Ret. to 7.6 Dialysate
mainprg flow
7-3
7-4 Fresenius Medical Care 4008 B 16/10.03 (OP)
7.2 Settings on the Monitor
The user can adjust or change the following settings on the hemodialysis machine via the
operator menu.
The key click menu can be used to accompany each key actuation by an audible tone.
The system clock should be checked before using the menu option Automatic on. Observe
summer time and winter time.
Confirm
☛
Select Press the Select key.
Dialysis
☛
Confirm Select Use the ▲ and ▼ keys to select
6. Clock settings 6. Clock settings.
☛ Press the Confirm key.
Confirm Select
5.2.3 —F—D—M—
Confirm Select
5.2.4 —F—HDIS—M—
Confirm Select
5.2.5 –F–D–M–HR–
Confirm Select
5.2.6 –F–HDIS–M–HR–
Confirm Select
5.2.7 —F—HR—C—
Confirm Select
5.2.8 —F—HR—
Confirm Select
5.2.9 —IHR—
Confirm Select
5.2.10 —M—
Entry of the time when the cleaning program selected in the 5.2 Auto-On program menu is to
start.
Confirm ☛
Select Press the Select key.
Dialysis ☛
Confirm Select Use the ▲ and ▼ keys to select
6. Clock settings
☛ 6. Clock settings.
Press the Confirm key.
The current delivery rate is computed in relation with the arterial pressure.
Confirm Select
Press the Select key.
Dialysis
☛
Confirm Select
Use the ▲ and ▼ keys to select
4. Bloodflow.calc. 4. Bloodflow.calc..
☛ Press the Confirm key.
Confirm Select
Use the ▲ and ▼ keys to select
4.2 Total bld.volume 4.2 Total bld.volume.
☛ Press the Confirm key.
Depending on the presetting defined in the SETUP menu, the user can display the effective
dialysis time or the effective UF time by selecting the 7.4 Show eff. time menu.
The effective dialysis time does not include Preparation and Alarm times.
– The maximum service life of the filter (12 weeks) has been reached.
– The filter is defective (T1 Test failed).
☞ Note
Replacement of the DIASAFE®plus
must be entered in the Medical De-
vice Register (date, batch number).
☛ I/O
Luftdetektor Arterieller
Druck
Ve
D
☛ I/O
Luftdetektor Arterieller
Druck
Ve
D
Override
–30
–200
0
Test
–40 –300
The Test LED is flashing.
If the Test LED is not flashing, one of
Prime the conditions for the test to start is not
fulfilled.
Dialysis Start
Reset Alarm
Alarm Tone
Mute
☛
–30 0
Test
–40 –300
Press the Test key.
Prime The Test LED is illuminated.
Dialysis Start
Reset Alarm
Alarm Tone
Mute
☞ Note
If the machine fails to pass the
functional test, repeat the test by
pressing the Test key.
If a DIASAFE®plus failure (FXX) is dis-
played after the functional test, this
test step has to be skipped. Then run
a heat disinfection program or a dis-
infection program followed by a hot
rinse.
Caution
Run a disinfection program after
each filter change.
– The maximum service life of the filter (12 weeks) has been reached.
– The filter is defective (T1 test failed).
– The maximum admissible number of treatments (100) has been reached.
(Applicable for the ONLINE plus™ option only.)
☛
60
200
20 50
Disinfection
10
100 40
0 0 30
Single
Needle –10
–100 20
–20
Confirm Select
Press the Confirm key to select the
PGM 4: –F–HDIS–M–HR cleaning program.
Confirm Select
FILTER max. Cle.:
Confirm Select
The DIASAFE®plus can then still be
PGM 2: –F–HDIS–M rinsed with the available cleaning pro-
grams until the end of its service life or
until the maximum number of treatments
Confirm Select
has been reached.
PGM 3: –F–D–M–HR
Press the Confirm key to change the fil-
ter (see Chapter 7.6.3 Changing the Fil-
ter)
PGM 4: –F–HDIS–M–HR
Confirm Select
or
press the ▲ or ▼ key to select a dif-
ferent cleaning program.
Confirm Select
Filter change !
☛
kPa mmHg kPa
Hotrinse 280
Blood Leak 30 60 Press the Hot rinse key.
200
20 50
Disinfection
10
100 40
0 0 30
Single
Needle –10
–100 20
–20
Filter change !
Confirm Select
Press the Confirm key to change the fil-
ter (see Chapter 7.6.3 Changing the Fil-
ter)
or
select the desired cleaning program to
override the filter change. (See Chapter
7.6.2 Overriding a Filter Change.)
☛
kPa mmHg kPa e.g. Hot rinse
Hotrinse 280
Blood Leak 30
200 60 Press the Hot rinse key.
20 50
Disinfection
10
100 40
0 0 30
Single
Needle –10
–100 20
–20
Filter change !
Confirm Select
Press the Confirm key to change the fil-
ter (see Chapter 7.6.3 Changing the Fil-
ter)
or
select the desired cleaning program to
override the filter change. (See Chapter
7.6.2 Overriding a Filter Change.)
☛
kPa mmHg kPa e.g. Hot rinse
Hotrinse 280
Blood Leak 30
200 60 Press the Hot rinse key.
20 50
Disinfection
10
100 40
0 0 30
Single
Needle –10
–100 20
–20
☞ Note
Replacement of the DIASAFE®plus
must be entered in the Medical De-
vice Register (date, batch number).
Basic conditions
– The optical detector senses clear flu-
id.
– Both dialyzer lines are in the interlock
shunt.
– The shunt door is closed.
– The concentrate suction tubes are
locked in the rinse chamber.
Caution
Disinfect after each filter change.
☞ Note
Do not remove the bicarbonate suc-
tion tube from its port to prevent lea-
kage of fluid which might then enter
the interior portions of the machine.
Caution
Wait until the cooling rinse program
following a hot rinse or a heat desin-
fection has been completed before
connecting the bibag ®.
7.8.1 Description
The function to transfer (download) treatment parameters allows transfer of patient-related treat-
ment parameters to the dialysis machine.
The following treatment parameters are eligible for transfer:
An external system (data source) can be used to transfer the treatment parameters.
The treatment parameters downloaded to the dialysis machine must be accepted in total. It is not
possible to accept individual parameters only.
☞ Note
After the treatment parameters have been transferred, the downloaded data can
only be accepted on the dialysis machine, provided:
the UF unit has been switched off.
the UF volume has been set to 0000.
Caution
The data received must be verified by the user, i.e. the user must check the data
received to ensure they are correct. Only then may the data be accepted with the
Conf. key.
☞ Note
If parameters are changed before the data are accepted, other parameters may
also change.
(e.g.: A change of the UF time will cause a change of the UF rate).
UF Rate ml/h
0750
UF Goal ml
3000
4:00
UF
I/O Prog.
Confirm Select
Bicarbonate: x mmol
Confirm Select
Dialysate flow xxx ml/min
Confirm Select
set Temp. xx,x oC