The document provides a checklist of tests to perform on a medical imaging device every 6 months and 1 year to ensure electrical safety and proper functioning. The 6 month checklist includes verifying functions of the C-arm, compression paddle, beam alignment at various angles and positions, manual mode compliance testing, and quality checks using phantoms. The 1 year checklist involves inspecting wires and grounding and the table assembly.

Uploaded by

Alexandra Janic
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOCX, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
330 views2 pages

Multicare Platinum

The document provides a checklist of tests to perform on a medical imaging device every 6 months and 1 year to ensure electrical safety and proper functioning. The 6 month checklist includes verifying functions of the C-arm, compression paddle, beam alignment at various angles and positions, manual mode compliance testing, and quality checks using phantoms. The 1 year checklist involves inspecting wires and grounding and the table assembly.

Uploaded by

Alexandra Janic
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOCX, PDF, TXT or read online on Scribd
You are on page 1/ 2

Has electrical safety

6 months

Check if all lead screws and gears are clean and lubricated.

Verify the Table emergency stop and limit switches function correctly.

Verify the C-arm and assemblies function correctly: Up/down, rotation, X-Y movement and brakes, and
all lamps.

Verify the Compression functions correctly: Movement, brakes, lock, and breast tray position lock.

Inspect the Compression paddle clamp and lock for correct function and wear.

Verify the compression force is adjusted to correct specifications.

Beam alignment – Verify the Pivot Point (at 0, 90 and 180 degrees).

Beam alignment – Verify the Tube L/R alignment.

Beam alignment – Verify the Biopsy and Ancillary paddle alignment.

Beam alignment – Verify the Aperture Adjustment.

Beam alignment – Inspect the Breast platform integrity.

Beam alignment – QAS (10/20/30) test (perform at 0, 90 and 180 degrees).

Beam alignment – Verify filament standby voltage.

Beam alignment – Perform the kV accuracy.

Beam alignment – Verify the X-ray tube output rate.

Beam alignment – Verify the mA output at each kV station.

Beam alignment – Verify the Backup timer setting.

Beam alignment – Perform the Dark and White fields adjustment.

Manual mode compliance testing – HVL.

Manual mode compliance testing – Reproducibility.

Manual mode compliance testing – Linearity.

Manual mode compliance testing – Limiting spatial resolution. (Line pairs)

Manual mode compliance testing – Signal to Noise ration.

Quality checks – Phantom test – Fiber test. (512 mode)


Quality checks – Phantom test – Fiber test. (1024 mode)

Quality checks – Phantom test – Fiber test – Check that the DSM image clearly shows the 5 th fiber
(mammographic phantom) or the 3rd fiber (mini-phantom).

Quality checks – Phantom test – Speck group test. (512 mode)

Quality checks – Phantom test – Speck group test. (1024 mode)

Quality checks – Phantom test – Speck group test – Check that the DSM image clearly shows the 4 th
speck group (mammographic phantom) or the 3rd speck group (mini-phantom).

Quality checks – Phantom test – Mass test. (512 mode)

Quality checks – Phantom test – Mass test. (1024 mode)

Quality checks – Phantom test – Mass test – Check that the DSM image clearly shows 3.5 masses
(mammographic phantom) or 2.5 masses (mini-phantom).

1 year

Inspect the input wires for safety and integrity.

Inspect all ground wires for integrity.

Inspect the table up/down assemblies (motors, pulley, and belts).

SMPTE test (Laser printer only).

Record the Exposure count (do not reset).