Has electrical safety
6 months
Check if all lead screws and gears are clean and lubricated.
Verify the Table emergency stop and limit switches function correctly.
Verify the C-arm and assemblies function correctly: Up/down, rotation, X-Y movement and brakes, and
all lamps.
Verify the Compression functions correctly: Movement, brakes, lock, and breast tray position lock.
Inspect the Compression paddle clamp and lock for correct function and wear.
Verify the compression force is adjusted to correct specifications.
Beam alignment – Verify the Pivot Point (at 0, 90 and 180 degrees).
Beam alignment – Verify the Tube L/R alignment.
Beam alignment – Verify the Biopsy and Ancillary paddle alignment.
Beam alignment – Verify the Aperture Adjustment.
Beam alignment – Inspect the Breast platform integrity.
Beam alignment – QAS (10/20/30) test (perform at 0, 90 and 180 degrees).
Beam alignment – Verify filament standby voltage.
Beam alignment – Perform the kV accuracy.
Beam alignment – Verify the X-ray tube output rate.
Beam alignment – Verify the mA output at each kV station.
Beam alignment – Verify the Backup timer setting.
Beam alignment – Perform the Dark and White fields adjustment.
Manual mode compliance testing – HVL.
Manual mode compliance testing – Reproducibility.
Manual mode compliance testing – Linearity.
Manual mode compliance testing – Limiting spatial resolution. (Line pairs)
Manual mode compliance testing – Signal to Noise ration.
Quality checks – Phantom test – Fiber test. (512 mode)
Quality checks – Phantom test – Fiber test. (1024 mode)
Quality checks – Phantom test – Fiber test – Check that the DSM image clearly shows the 5 th fiber
(mammographic phantom) or the 3rd fiber (mini-phantom).
Quality checks – Phantom test – Speck group test. (512 mode)
Quality checks – Phantom test – Speck group test. (1024 mode)
Quality checks – Phantom test – Speck group test – Check that the DSM image clearly shows the 4 th
speck group (mammographic phantom) or the 3rd speck group (mini-phantom).
Quality checks – Phantom test – Mass test. (512 mode)
Quality checks – Phantom test – Mass test. (1024 mode)
Quality checks – Phantom test – Mass test – Check that the DSM image clearly shows 3.5 masses
(mammographic phantom) or 2.5 masses (mini-phantom).
1 year
Inspect the input wires for safety and integrity.
Inspect all ground wires for integrity.
Inspect the table up/down assemblies (motors, pulley, and belts).
SMPTE test (Laser printer only).
Record the Exposure count (do not reset).