0% found this document useful (0 votes)
89 views12 pages

Anton Paar Surpass 3: Release Notes

The SurPASS 3 software provides measurement and control capabilities for the SurPASS 3 electrokinetic analyzer. It communicates over CAN bus with the SurPASS 3 device, which is connected via USB to a Windows PC running the software. This release introduces a new measurement data viewer. Hardware requirements include at least an Intel i3 CPU, 4GB RAM, and 20GB disk space. The software requires Windows 7/8/8.1/10 and will install dependencies like Microsoft .NET 4.5.1 and Windows Installer 4.5.

Uploaded by

Karen
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
89 views12 pages

Anton Paar Surpass 3: Release Notes

The SurPASS 3 software provides measurement and control capabilities for the SurPASS 3 electrokinetic analyzer. It communicates over CAN bus with the SurPASS 3 device, which is connected via USB to a Windows PC running the software. This release introduces a new measurement data viewer. Hardware requirements include at least an Intel i3 CPU, 4GB RAM, and 20GB disk space. The software requires Windows 7/8/8.1/10 and will install dependencies like Microsoft .NET 4.5.1 and Windows Installer 4.5.

Uploaded by

Karen
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 12

Anton Paar SurPASS 3

Release Notes

The SurPASS 3 software is the measurement and control software for the SurPASS 3
electrokinetic analyzer for solid surface analysis. It controls all SurPASS 3 hardware
components and enables automated data acquisition and measurement series.
The software is running on a Windows PC and communicates over CAN bus with the
SurPASS 3 device which is connected to a USB port of the PC. For the first time, this version
of SurPASS 3 will feature a measurement data viewer.
Anton Paar SurPASS 3 – Release Notes

 Hardware Requirements ...................................................................................................... 3


 Software Requirements ....................................................................................................... 4
Operating Systems.................................................................................................................. 4
 Installation ........................................................................................................................... 5
Administrative Privileges and User Rights ............................................................................ 5
Power Options ........................................................................................................................ 5
Additional Software Components .......................................................................................... 5
Microsoft Windows Installer 4.5 ............................................................................................ 5
Microsoft .NET Framework 4.5.1 .......................................................................................... 6
PEAK PCAN-USB adapter .................................................................................................... 6
 Known Issues ...................................................................................................................... 7
 Version History ................................................................................................................... 8
Version 2.10.270 2018-11-09 .............................................................................................. 8
Version 2.10.260 2018-03-19 .............................................................................................. 8
Version 2.00.220 2017-11-15 .............................................................................................. 9
Version 2.00.170 2017-06-01 .............................................................................................. 9
Version 1.30.160 2017-03-17 ............................................................................................ 10
Version 1.20.150 2016-12-22 ............................................................................................ 10
Version 1.10.120 2016-06-27 ............................................................................................ 11
Version 1.00.110 2016-03-15 ............................................................................................ 11
Version 1.00.100 2015-12-16 ............................................................................................ 12

2
Anton Paar SurPASS 3 – Release Notes

 Hardware Requirements
This section describes the hardware requirements that your target machine must meet to successfully install and
run Anton Paar SurPASS 3. Those hardware requirements are intended to provide guidance in choosing the right
hardware to run SurPASS 3 smoothly; hence, they are more a kind of recommendation than real requirement.

Recommended Hardware
 Intel(R) Core(TM) i3 CPU
 4 GB RAM (depends on the operating system)
 Screen resolution of at least 1024 x 768 at True Color (32 bit), best experience with a monitor with an
aspect ratio of 16:9 or 16:10
 Use of a dedicated graphics card that supports Direct X 9 is recommended
 20 GB free disk space on system drive
 Mouse and keyboard

3
Anton Paar SurPASS 3 – Release Notes

 Software Requirements
This section describes the software requirements that your target machine must fulfill to successfully install
Anton Paar SurPASS 3. SurPASS 3 Setup automatically checks the software requirements and prevents
installation on unsupported systems.

Operating Systems
SurPASS 3 is built on Windows technologies and is only available for Windows operating systems. Thus, the list
of supported operating systems depends on which operating systems are officially supported by Microsoft.
Please keep in mind that the list of supported operating systems might change for future releases of SurPASS 3.

A detailed list of supported and unsupported operating systems is provided for your convenience.

Supported Operating Systems


 Microsoft Windows 7
o 32 bit and 64 bit are supported
o Service Pack 1 or later is required
o Windows 7 Starter Edition is not supported
 Microsoft Windows 8
o 32 bit and 64 bit are supported
 Microsoft Windows 8.1
o 32 bit and 64 bit are supported
 Microsoft Windows 10
o 32 bit and 64 bit are supported
 Microsoft Windows Server 2008 / 2008 R2
o 32 bit and 64 bit are supported
o Service Pack 2 or higher is required for Windows Server 2008

Unsupported Operating Systems


 Microsoft Windows 2000
 Microsoft Windows XP
 Microsoft Windows Vista
 Microsoft Windows Server 2003

4
Anton Paar SurPASS 3 – Release Notes

 Installation
This section provides information about the installation of Anton Paar SurPASS 3. Installing SurPASS 3 should
be an easy and straightforward task. Just start the provided setup package and follow the instructions of the setup
program.

Before starting with installation you should check that your target system meets the hardware and software
requirements.

Administrative Privileges and User Rights


You will need administrative privileges on the target machine to be able to install SurPASS 3. Thus, it might be
necessary to contact your IT department.

On Windows 7, Windows Server 2008 and Windows Server 2008 R2 you still need administrator privileges, but
you may start the setup package with the “Run as administrator” option from within a standard user account.

Installing SurPASS 3 from an administrator account and then switching to a standard user account for daily use
is a valid scenario that is fully supported. SurPASS 3 does not require administrator rights for normal operation;
just for installation.

Power Options
You should configure your operating system’s power options so that your target machine does not go into
sleep/standby mode or into hibernation after a few minutes of inactivity; otherwise SurPASS 3 will not be able to
continuously record measuring data.

Additional Software Components


SurPASS 3 depends on a few external software components that will be automatically installed by SurPASS 3
Setup. The installation of those components might take a few minutes up to an hour, depending on the number of
software components that need to be installed. For the first-time installation SurPASS 3 Setup might also require
to reboot the machine once or twice.

Following you find a list of software components that will be automatically installed.

Software Components
 Microsoft Windows Installer 4.5
 Microsoft .NET Framework 4.5.1

The subsequent sections will describe how to install those software components for the case that you need to
install them manually.

Microsoft Windows Installer 4.5


The Microsoft Windows Installer (MSI) is the application installation and configuration service for Windows.
The download package will update the version of Windows Installer on your system to version 4.5.
 Download Windows Installer 4.5 Redistributable from
http://www.microsoft.com/downloads/en/details.aspx?FamilyID=5a58b56f-60b6-4412-95b9-
54d056d6f9f4&displaylang=en
 Ensure that you download the appropriate setup package for your operating system.
5
Anton Paar SurPASS 3 – Release Notes

 Run the setup package to install Microsoft Windows Installer.


 Restart your system after installation.

Microsoft .NET Framework 4.5.1


The Microsoft .NET Framework 4 redistributable package installs the .NET Framework runtime and associated
files that are required to run and develop applications to target the .NET Framework 4.5.1

 Download Microsoft .NET Framework 4.5 (Standalone Installer) from


http://www.microsoft.com/en-us/download/details.aspx?id=40779
 Run the setup package to install Microsoft .NET Framework Version 4.5.1
 Restart your system after installation if necessary.

PEAK PCAN-USB adapter


The SurPASS 3 software uses a PEAK PCAN-USB adapter to communicate with the SurPASS 3 device.
The PCAN-USB adapter driver is provided by the SurPASS 3 software setup and will be installed automatically.
In order to repair or to update the driver take following steps

 Download PCAN-USB-Package from


http://www.peak-system.com/PCAN-USB.199.0.html?L=1
 Or navigate to folder C:\Program Files\Anton Paar\Surpass3\Setup\PCAN_USB
 Run the setup package to install the driver.

6
Anton Paar SurPASS 3 – Release Notes

 Known Issues
At the moment no known issues were reported.

7
Anton Paar SurPASS 3 – Release Notes

 Version History

Version 2.10.270 2018-11-09


New Features

SurPASS 3
 Cell constant for the conductivity electrode extended (45 - 50 1/m)

Bug Fixes

SurPASS 3
 SurPASS 3 (Error Code -2) Rinse failed. Fixed.
 SurPASS 3 s/w frozen. Fixed.

Version 2.10.260 2018-03-19


New Features

SurPASS 3
 Support of SurPASS 3 ECO.
 New measuring cell Clamping 10 x 10 was implemented.
 The Advanced measurement options were added to the measurement preparation pane.
When selected the Upper pressure limit value can be entered or set via a vertical annotation line in
the Measurement chart.
The Upper pressure limit value will be effective throughout the entire experiment. All
measurement values above that pressure limit will be marked as invalid and will not be used in
calculations.

Changes

SurPASS 3
 Measurement options which are not available when the titration unit is missing are disabled.
 Measurement option Adsorption will be disabled when no external pressure supply is selected.
 Improved static resistance measurement to eliminate influences of measuring cell geometries.
 Calculation for next pH step in Isoelectric Point was adapted.

Firmware
 New firmware version 2.03 for Digital Module.

Bug Fixes

Setup
 Windows File Explorer link to SurPASS 3 Viewer was broken after installation. Fixed.

SurPASS 3

8
Anton Paar SurPASS 3 – Release Notes

 Failures of the Pressure Control Unit were not reported. Fixed.

Version 2.00.220 2017-11-15


New Features

SurPASS 3
 Recognition of the verification device.

Changes

SurPASS 3
 Improved constant pressure limits for Adsorption Kinetics measurement.
 Improved pressure setting for nitrogen purge of the electrolyte solution.
 New tolerance for conductivity calibration.
 New default settings for better display of Adsorption Kinetics results.

Bug Fixes

Firmware
 New firmware version 2.02 for Digital Module to support the Constant Pressure Module.

Version 2.00.170 2017-06-01


Changes

SurPASS 3
 Upgrade chart to version 4.
 New reasonable start limits for Measurement Chart and improved limits for Result Chart axes.

Firmware
 New Constant Pressure Module available
 New Firmware version for Digital Module 2.00 to support the Constant Pressure Module.

Automated Pressure Control and Nitrogen Purge


 New Hardware: Automated Pressure Control
 Nitrogen Purge
 Control of the target pressure while performing Adsorption Kinetics Measurements

SurPASS Viewer
 SurPASS Viewer is a new component which allows the user to view data measured with SurPASS 3.
 Automatically installed with SurPASS 3 version 2.0
 View and compare multiple experiments and measurements at the same time
 Save and load packages of measurement data
 Export the experiment chart as .XPS or .EMF

Bugfixes

9
Anton Paar SurPASS 3 – Release Notes

 Fixed an error where the parameter FileCount was always 0.


 Fixed an error where the ExternalPressureSettings were set to true when starting SurPASS 3.

Version 1.30.160 2017-03-17


Changes

Firmware
 “Firmware update available” notification. If there is a new Firmware version available the user
receives a notification.
 New Firmware version for Measurement Module: 1.06 which fixes an internal timing issue.

Adsorption
 Adsorption is now available in two versions: Slow and fast Adsorption which enables the user to
easily perform an adsorption measurement.

Supporting a new Measuring Cell


 Ceramic Membrane

Version 1.20.150 2016-12-22


Changes

Excel file export


 Measurement values will now be stored in the user’s preferred unit

pH Scan and IEP


 The pH step-width used for pH-Scans and Iso Electric Point Measurements can now be set in the
user interface of SurPASS 3

Experiment preparation
 Improved the experiment preparation including a simplified way of handling the use of external
pressure supply
 The values for Rinse and Zeta Potential are also used for the respective Rinse cycles and Zeta
cycles during pH-Scans and Iso Electric Point Measurements

Zeta-Potential repetitions introduced to pH-Scan and IEP


 pH-Scan and Iso Electric Point Measurement support a new parameter: Zeta cycles (1-4)
 The last measured Zeta-Potential value will be used for Iso Electric Point calculation

Adsorption
 Adsorption Kinetics requires a Zeta Potential or Rinse beforehand

Optimized settings
 Optimized settings for cleaning and pH titration

10
Anton Paar SurPASS 3 – Release Notes

Version 1.10.120 2016-06-27


Changes

Supporting New Measuring Cells


 Contact Lens, Flexible Tubing, Hollow Fiber and Adsorption

Measurement Data Files


 Due to the integration of the new measuring cells the Permeability Index value was introduced.

User Interface
 The Clean the system procedure automatically performs two cleaning cycles, and does not have to
be started twice anymore.
 The Firmware Update dialog was improved and provides the actual firmware update files.

Bug Fixes
 The measurement time of a single Adsorption cycle was too long so that air was sucked in. Fixed.
 Clicking Auto scale in the result chart’s options caused obsolete data to reappear. Fixed.
 When SurPASS 3 performs any operation the computer will not enter the sleep mode anymore.

Version 1.00.110 2016-03-15


Changes

User Interface
 Obsolete measurement result values will be cleared as soon as another measurement is started.
 The actual resistance value will be displayed during the ‘Rinse’ operation.
New parameters for cleaning procedure ‘Empty the system’
 Start pressure = 800mbar
 Stop pressure = 50mbar
New parameters for cleaning procedure ‘Clean the system’
 Start pressure = 1200mbar
 Stop pressure = 10mbar
Settings for ‘Dosage’ operations to achieve a target pH
 up to pH 4.5 : dosage volume = 100µl
 up to pH 9.5 : dosage volume = 10µl
 up to pH 14 : dosage volume = 200µl
 Settle time increased to 5 seconds when dosing to lower pH values.
Software Update / Setup
 Support Windows 8.1 & Windows 10

Bug Fixes
 Firmware update failed to flash certain firmware packages.
 Firmware update operation failed to flash the Digital module.

11
Anton Paar SurPASS 3 – Release Notes

Version 1.00.100 2015-12-16


General
 First public release.

12