UN6000 Operating Instr. ECT
UN6000 Operating Instr. ECT
ABB reserves all rights to this document and to the information and topics
contained in it. This also applies to any possible claims to copyright or
patents. Forwarding and/or the duplicating of this document without the
express permission of ABB is forbidden.
This document has been prepared and checked with great care. If
however it still contains errors, then the user is asked to report these to
ABB.
Table of Contents
Table of Contents ................................................................................................................ 3
Installation .................................................................................................... 15
3.1. Panel Configuration.......................................................................................................15
3.2. Channel IP Addresses for local TCP/IP Network ...........................................................15
Settings ......................................................................................................... 17
4.1. Screen after Start-up .....................................................................................................17
4.2. Selection of the Operation Level ...................................................................................18
4.3. Options, Brief Description..............................................................................................20
4.4. ECT Network Configuration ...........................................................................................23
List of Figures
Figure 2-1 Example of a typical Excitation System ......................................................................... 13
Figure 3-1 Overview of a possible IP Address Configuration .......................................................... 15
Figure 4-1 Typical Screen after Start-up ......................................................................................... 17
Figure 4-2 Application Settings ....................................................................................................... 20
List of Tables
Table 2-1 Windows User .................................................................................................................11
Table 3-1 Panel Configuration .........................................................................................................15
Table 3-2 Default IP Address ..........................................................................................................16
Table 4-1 Operation Levels, Default Factory Settings......................................................................18
Table 4-2 Application Settings .........................................................................................................21
Table 5-1 Static Display Elements ...................................................................................................24
Table 5-2 Display of the Status Bar .................................................................................................25
Table 5-3 Available Screens of the ECT, Selection Buttons ............................................................26
Table 5-4 Folder and Parameter Name of the Limiter Release ........................................................32
Table 5-5 Settings for Fast Trending ...............................................................................................45
Table 5-6 Event Type ......................................................................................................................59
Table 5-7 Description of the Available Tags ....................................................................................63
Table 5-8 Setting and Saving Procedure of the Boot Parameters ....................................................75
Table 5-9 Name of Product Group to the Screen Name ..................................................................79
Table 5-10 List of Settings Relating to the Operation Level ...........................................................103
Table 5-11 Description of the Device Configuration Settings .........................................................105
General Information
Manufacturer’s Address
If any questions arise, consult the local ABB representative or the factory:
IMPORTANT!
If calling ABB, please leave your name, department and phone number.
This will allow the responsible ABB representative to call back without
delay.
Safety Instructions
1.1. General
The safety instructions must be followed during commissioning, operation
and maintenance of the excitation system. Read these instructions and
the excitation system User Manual carefully before operating the device
and keep this document for future reference.
Operating Personnel The operator must be familiar with the functions of the ECT in order to
operate the excitation system.
Maintenance Only specially instructed personnel can carry out maintenance and repair
Personnel work. The maintenance personnel must be informed about the
emergency shutdown measures and must be capable of turning off the
system in case of emergency.
The maintenance personnel must be familiar with the accident prevention
measures at their workplace and must be instructed in first aid and
firefighting.
Responsibilities It is the owner’s responsibility to ensure that each person involved in the
installation and commissioning has received the appropriate training or
ABB can provide neither warranty nor liability for any injury which can be
caused by the installation and operation of unauthorized software.
Device Description
2.1. Introduction
The Excitation Control Terminal (ECT) is used for controlling and
monitoring the excitation system with a different regulation configuration.
The ECT can be located in the door panel of the excitation system for
local control and/or in the control room for remote control.
The ECT is a powerful industrial PC with a user friendly Human Machine
Interface that runs independently of the control CPU’s. The ECT can be
added to an excitation system to support the operating and maintenance
staff. The TFT (Thin Film Transistor Liquid Crystal Display, LCD
technology) touch screen provides the operator with a range of selectable
screens showing information about the current status of the system in
graphical and numerical form. Additionally, authorized engineers can read
and adjust all parameter values. Further, analysis tools for
troubleshooting are available as well.
The ECT includes long-time recordings and auto save functions. These
are indispensable features for quick and precise analyses.
Important data are saved and are ready to be shown on the ECT or to be
exported for further investigation.
In order to protect the system against unauthorized manipulation, different
password-protected Operation Levels can be defined.
2.4. Operation
The operator has a range of screens at his disposal that inform about the
status of the excitation system. A detailed description of all screens can
be found in Chapter 5 - Functional Description.
If the operator has logged in "Active Control Mode" (Password protected)
he is entitled to perform active control such as opening and closing
breakers, adjusting setpoints, selecting an operation mode etc.
These operating principles follow the rules below to ensure easy and fail-
safe operation:
ECT requires a hardlock (Dongle) otherwise the panel is not working.
To control the system with the ECT:
A password is necessary according to the Operation Level
"Active Control Mode" must be selected (Panel control
button on "green")
The command buttons must be touched in the release or
execute mode
Users without password can supervise the system but have no
control over it.
All indications on the screens are real feedback information from the
system. This means that all data shown on the screen are retrieved
directly from the selected channel.
NOTICE!
A failure of the ECT or an interruption of the hardware link has no
influence on the current operation. The default value of the operator
location is remote. In this case the access to the ECT is immediately
disabled.
2.7. Hardware
Structure
The ECT is mechanically robust. The unit cover is made of galvanized
sheet-steel.
The connections, slots and interfaces are on the left sidewall of the case.
Installation
The seamless sealed front panel protects reliably against ingress and dust,
dirt and spray water. The blue LED indicates power supply OK
Mounting
The ECT can be installed in racks, cubicle doors and desks.
It is designed as a fanless system but due to the temperature inside the
cubicle the pure convection cooling could be insufficient.
See Chapter 9.3.1 General Data TPC-1551T
Dimensions
The mechanical and fitting dimensions are described in
Chapter 9.3.3 Mechanical Dimensions
Structure
The ECT is mechanically robust. The unit cover is made of anodized
aluminum.
The connections, slots and interfaces are on the bottom of the case.
Installation
The seamless sealed front panel protects reliably against ingress and dust,
dirt and spray water.
Mounting
The ECT can be installed in racks, cubicle doors and desks.
It is designed as a fanless system but due to the temperature inside the
cubicle the pure convection cooling could be insufficient.
See Chapter 9.3.2 General Data IRTPPC15
Ports
- USB 2.0 port: The software of the ECT does not work without a hardlock
(Dongle).
- Ethernet port: RJ-45
- 24 V Connector (green): Phoenix Contact FKC 2,5/3-STF-5,08
Dimensions
The mechanical and fitting dimensions are described in
Chapter 9.3.3 Mechanical Dimensions
Installation
3.1. Panel Configuration
First, perform the following adjustments:
- Set the parameter for the monitoring release of the ECT to true.
- Touch Parameters, path [Control IT \ In \ ...]
ECT
(remote 2)
Ethernet
IP Address: 172.16.0.64 ECT-PC
switch
with ECT-SW
(remote 1)
IP Address: 172.16.0.63
ECT
Ethernet
switch
(local)
172.16.0.61 IP Address:
ECT 6x x =1 Local
x =2 …
x =3 Remote 1
x =4 Remote 2
…
…
Settings
This Chapter describes the following settings:
Description and setting of the password-dependent Operation Levels
Settings of the ECT configuration (Options, Brief Description)
Settings of the IP Address from local TCP/IP Network.
Note: The ECT has a screen saver installed which turns off the screen.
- Touch the screen to turn it on again.
The measure extends the lifetime of the screen and the backlight.
Microsoft Windows 10: Start\Settings\Personalization\Screen Saver\
password or blank Sample = Personal User with a Personal Operation Level can
(e.g.) be defined by the Administrator
[Login]
- Open the User ID
[Users]
- Accept or reject the User
ID ,
[Text Editor]
- Accept or reject the
password ,
[Login]
- Touch Options
Operation Level ≥ 5
Import settings from an external file.
(Use a memory stick to transfer the ECT configuration
data to another ECT)
Export settings to an external file.
(Use a memory stick to transfer the ECT configuration
data to another ECT)
Discard current changes and reset to active settings.
+ Application
- Data Folder Possibility to change the location
- Disk Space Manager Set free disk space to prevent the data overload
+ Configuration
- Configurable Modules Not used in this application
+ Operations
- Command Editor Configuration of the command buttons with fast ECT-
communication monitoring included.
- Copy Settings Copy settings from source to target device
- Keep-Alive Signaling Slow ECT-Communication monitoring for normal data traffic in
the Keep-Alive entry
- Top Level Control - Configuration of the setpoint up/down buttons and the value
Configuration indication
+ Tools
- Pec Software Information Display installed components of connected target (PECs)
- CPU Load Show CPU load information of connected devices
- Time Overwrite the channel time
- Data Collection Collection of all available system information for
troubleshooting
- Process Selector Selector switch to activate another application
- Configuration information Not used in this application
+ Parameters
- Dockable panel A parameter set of My Groups is visible, docked and operable
on all the selection screens
Not used in standard applications
- Save to flash Configuration of the button features
- Parameter Configuration Change or show/hide columns order in the parameter screen
+ Diagrams
- Diagrams Configuration Select the visible diagram
+ Trending
- Fast Trending Setting of the Chart Update Interval (ms),
Default Number of Samples,
Minimum and Maximum Number of Samples
- Slow Trending Setting of the Chart Update Interval (ms),
Maximum Log Size, Default Range (s),
Minimum and Maximum Range (s)
+ Transient Recorder
- Settings Default number of samples (ms)
+ Security
- User Management Changing the personal password,
Administrator access: Editing user accounts
- Properties
- Properties
Functional Description
5.1. Static Display Elements
In order to ensure that the operator has a clear process overview, the
most important system display elements are permanently visible on the
screen.
Independent of the actual screen selection, the following important parts
are always visible:
System status bar
Panel control button
Function, screen selection
Setpoint indication and setpoint command [up] and [down] buttons.
Screen selection,
Function selection buttons
Process selector
(not standard application)
The status of the excitation system is displayed at the top of the screen.
This important information is presented on all user screens, as it indicates
the actual condition of the system.
The order of the screen selection can be changed or unused buttons can
be hidden. See Chapter 5.12.1 - General / CT Modules.
- Touch Operations
or
Before any of the command buttons can be used, the following conditions
must be fulfilled:
Note: These settings have to be made in each Control Channel (CH 1, CH 2 and BU)
- Touch Operations
[Data Point...]
- Close the window of
parameter/signal information
- Confirm the message OK
[0601 Measu..]
- Finally confirm or reject the
parameter/signal
Threshold
See the red bars 0.7 to 0.9
and 1.1 to 1.3
CosPHi Type
Different scale adjustable
[Cancel]
- Close the pull-down menu
- Touch Powerchart
P
Q
P(q) Q(p)
P Q [cro
Q P
- Touch Diagrams
or
The excitation system is shown as a
UN6000 Medium UN6000 X-power
simplified single line diagram. This window
contains information about:
Diagrams:
- Not ordered: The button has to be canceled. See Chapter 5.12.1 - General, section CT Modules.
- Ordered: The Diagrams have to be activated. See Chapter 5.12.9 - Diagrams.
- Scroll to CH 1 or CH 2
(gray highlighted line)
[Meas_Sys…]
- Touch the table with the
signal groups and scroll
to the desired group (gray
highlighted line)
[Control IT \ Out \
Meas_SystemMeasurement]
Cancel a signal:
- Scroll to the desired signal
line (gray highlighted line) and
touch this button
Custom Groups
G = Group number 1 to 6
I = Signal number of the 1 to 12
graph
Data Point = Signal name
Configure Signal
CIOConfigWord:
Override database
descriptor
Not used in this application
The bit-setup is only useful for bitwise rendering (Chapter 5.5.3 - Scaling of the Signals in the
Graph, page 38) and displayed in a separate group (see next section).
Group Configuration:
Data format*
User-defined min / max: ,
Fill out both values
Data format*
Logarithmic scaling: ,
Fill out both values
Data format*
Stacked boolean signals:
For boolean signals, select Signal 2
another Group Signal 1
Signal 0
Toolbar display
"Legend" buttons
Single Chart:
All recorded signals are displayed in one graph.
Split Mode:
Each signal of the recording is displayed
separately in a small graphic.
User Groups:
Combines individual signals into groups, which
are displayed in one graphic.
Upload display:
The design of the display can be saved and
reused as a template for an identical display.
Note: Only useful for the same parameter set.
The Slow Trending has already started after the selection of the signals
Stop, Start
Timer
,
The times and the duration are shown in the
box next to the timer buttons. The measured
range is colored dark-gray in the graphic.
Note: First a recording has to be started or loaded. Otherwise the following message will appear:
"No data available for export. Please load log files or configure data points for recording."
,
- Write a mark Text Editor: e.g. Slow Trending....
The marks are stored in
the file called remarks.txt
in the above-mentioned
defined folder
[Export Data]
- Finally confirm the export
data
,
Take notice of the file format in the folder "Data":
[Chart Export..]
- Finally confirm the export [Select File] Take notice of the file format in the folder "Export":
data
,
- Export only the chart Compared to the above export recording, only the
chart file is exported.
See Chapter 5.12.10 - Trending, Section Slow
Trending
- Save settings Settings are applied after a restart when the box
"Restart trending after application restart auto-
matically" is checked.
See Chapter 5.12.10 - Trending, Section Slow
Trending
Load the saved log file from hard disk ECT or from another device (graph visible on the
screen)
[Delete]
- Confirm the message Yes
or [Log Browser]
- Open the file from
another device (USB stick)
Note:
All columns of a table can be sorted in ascending or descending order.
Touch the headline of the corresponding column and the sort order will immediately change.
See Chapter 5.5.2 - Selection of the Signals and Start of the Recording. The trending does not start
automatically. See later.
or
- Cancel the fast trending
Start, Stop
Timer
[Chart Export..]
- Finally confirm the [Select File] Export in COMTRADE format creates a folder
export data
,
Or save the data file from the Fast Trending Recorders screen
Load the log file from hard disk ECT or from another device (graph visible on the screen)
- Fine tuning
Channel Selection
- Or open "Select
Recorder…"
Note: Only the "Configurable Transient Recorder" is configurable. The "Transient Recorder" has
fixed allocated signals and cannot be controlled. It is not part of this description.
Note: The Configurable Transient Recorder can only sample Simulink signals.
This signal name starts with SFS_…or STS_... (Example: SFS_Meas_UgRel)
Trigger Conditions
Number of Samples:
Min/Max value: 0 to 1500
Pretrigger:
Min/Max value: 0 to 1500
[402 State
Machine_STS]
- Mark and confirm the line
,
[Configure Recorder:
Configurable...]
- Confirm the configuration
[Configure Recorder:
Configurable...]
- Confirm the configuration
Auto Trigger
After the trigger conditions are set, the trigger trips after the conditions are fulfilled.
See 5.7.1 - Parameter Settings and Trigger Conditions
Manual Trigger
or
- Touch the "Select
Recorder…" to open
the dialog [Transient
Recorder…] and select
the line "Configurable
Transient Recorder".
[Transient..]
- Confirm the triggering
Auto Read saves the recorded data of the selected Channel (CH 1, CH 2 or BU) to the hard disk of
the ECT. Otherwise the data is only saved to the Channel. The graph is not visible on the screen.
The "Auto Read" function is allowed to be active only in one ECT of an excitation system. This
concerns both Transient Recorders and all devices such as CH 1, CH 2 and BU.
- Select "Configurable
Transient Recorder".
to
- Mark Auto Read
check box
[Transient
- Confirm the trigger Rec...]
condition
Or select "Select
Recorder…" to open
the dialog [Transient
Recorder…] and select
"Configurable Transient
Recorder".
Timer
[Select Rec..]
Recorded data from the Channel:
- Select the desired recorded file
[Select Rec…]
- Confirm the selection
Note: Each recorded file from the Channel can be saved to the hard disk.
See the section "Save and export" at the beginning of this Chapter.
- Fine tuning
Bit analysis
Each bit is shown in a separate
graph
[Events list]
- Close the Events list
- See Chapter 5.8 - Events
- Touch Events
The Event Logger contains a table with the last 2000 entries. As soon as this window is selected, an
automatic upload will start. If a new entry occurs, the panel reads this out of the UNITROL System
and the table will automatically be updated.
Actual alarm
Reversed alarm
Event Toolbar
- Export the logged data to Default folder location: See Chapter 9 - Appendix
the ECT hard disk or to an C:\...\Control Terminal\ECT\Export\Alarms and Events\..
external storage device
Print a PDF
Save in Directory:
Windows 10:C:\Users\[User]\Documents
Operation Level 2
Operation Level 4
- Confirm the message Yes
Parameter changes
Warning!
! Only specially trained ABB personal are permitted to make parameter
changes.
- Touch Parameters
Parameter toolbar top line (after touching Parameters the first time):
Compare snapshots
Tag Filters:
Undo button
Note:
- The range of tags depend on the database, which is different for the various software releases of
UNITROL 6000 Medium or X-power.
- The tags are only visible if the database is activated.
Change the mode from AUTO to MANUAL with the parameter ForceManual (Example)
Control IT [down]
In [right]
…
Brk_Breaker [right]
SM_StateMachine [down]
The change from "False" to "True" and back to "False" is the principle of a toggle switch.
Check the change: See the "ON State Message" of the command button "Mode" in the menu
Operations.
Note: The Editor dialog depends on the data format of the parameter.
The user can define separate directories (group) and subdirectories for frequently used parameters.
Note: These settings must be made in each Channel (CH 1, CH 2 and BU)
[My Groups]
-Touch the + button to open the
Text Editor
Note: A new directory is located one level below the gray highlighted directory.
It is not possible to create a directory on the same level as My Groups.
Control IT [down]
In [right]
…
Brk_Breaker [right]
SM_StateMachine [down]
- Scroll to TestConfiguration
- Shift to ForceChannel
The parameters from the Channel are uploaded to the hard disk or an external storage device.
The snapshot file is saved to the hard disk or an external storage device.
Channel CH 1
Unless two snapshots are opened, the icon of the compare function remains gray.
See last Chapter.
Note: The written parameters are only active if the Panel Control Button is green.
Note: In this case the result of the snapshot comparison must be analyzed before the button "Write
data to the target" has been touched. But this is not part of this description.
The snapshot file is only written to the target, but not saved in the flash memory.
Channel CH 1
Toolbar top line
(snapshot view)
- Touch this button to open the
dialog
Processing it takes a few min.
- Export the logged data to the Default folder location: See Chapter 9 -
ECT hard disk or to an external Appendix
storage device C:\...\Control Terminal\ECT\Export\..
[Different Data..]
The warnings have to be checked. The result of this analysis should provide the basis for the
decision. But this is not part of this description.
[Update Tar..]
- Confirm the update
Yes or No
- Confirm the message
- Scroll to the CH 2
After updating the target (Download) the parameters have to be saved in the flash memory to make
changes permanent.
Description
The live data view of the toolbar
top line has to be present
The save button is not available in releases older than 2.1.xx. The description how to do it manually
is described on the next page.
Select Channel CH 1
In [right]
…
Brk_Breaker [right]
Sys_ExcitationSystem [down]
If the field is already cleared (false state) then mark it (true state)
Note: The Panel Control Button must be green to start the boot sequence.
The update of the boot parameters is only possible in the configuration sequence.
The table below shows how to make the settings and save them permanently to the flash memory.
Table 5-8 Setting and Saving Procedure of the Boot Parameters
Note: Finish the complete sequence otherwise the system remains in the configuration sequence.
The installer allows to backup the archive respectively the image of the flash memory of the PEC
control or to restore the data.
- Touch Installer
Download archive
Stop applications and reboot controller after download otherwise the new application is not
active.
Backup controller
Presettings
Digital In
- Mark Force
Force value: - Select On for simulation
Digital Out
HW Output:
Possibility to invert only the SW signal
- Mark Force
Force value: - Select On for simulation
Analog Out
SW Value: value from the control unit
HW Output: - Select Preset and confirm or
define other min/max values
- Mark Force
Force: - Set a value
5.12. Options
Operation Level = Administrator
- Touch Options
The menu bar "Application Settings" and the toolbar bottom line are described in
Chapter 4.3 - Options, Brief Description, page 20.
Note: The settings of decimals values are limited between the minimum and maximum value
specified in the editor menu.
Note: The name of the product group does not match the screen name.
Table 5-9 Name of Product Group to the Screen Name
Name of Product Group Screen Name
UN6000 X-power UN6800
UN6000 Light or Medium UN6080
5.12.1. General
Device Configuration
Device Information:
- General
- Custom Properties
- Disconnect the device Note: Touch this button to reset the present device
from the system (Channel 1, after a connection loss (e.g. supply interrupt).
2 or BU)
Configuration of a new
device
- Open the dialog
"Supported Devices"
Device Configuration:
General:
- Open the Text Editor and
enter the Device Name
[Text Editor]
- Confirm the Text
Editor
[Device Confi]
- Finally confirm the
settings
Note 1) For detailed information see Table 5-11 Description of the Device Configuration Settings,
page 105 at the end of this Chapter.
CT Modules
CT Channel Mapping
Unit Name:
- Open the Text Editor
and enter the Name
- Scroll to the
Application Type
- Repeat it for CH 2,
BU
In a single channel system blinks CH 2 in the status bar "Connect" if CH 2 is mistakenly mapped
Language Settings
Note: After changing the language, the ECT must be restarted. Otherwise the setting will not
become active.
5.12.2. Application
Data Folder
Possibility to change
the location
Default: C:\...\Control Terminal\ECT\Data or Export or Settings
If no value is entered, the default location is used.
User-defined: Replace only the first part of the path name. C:\...
The second part appears automatically. (Control Terminal\...etc.)
For the first part see Chapter 9 - Appendix
The minimum value with an ECT is calculated automatically by ECT tool based on the disk space
available on the PC hard disk.
5.12.3. Configuration
5.12.4. Operations
Command Editor
- Define a new
command button
Example:
Command button
called ForceBackup
Name:
- Open the Text Editor -
Enter the name in two
lines (word length too
long for the button
field)
ON State Message:
- Open the Text Editor
- Enter "FB ON"
- Confirm Command
Editor
- Define a new
feedback indication
- Select tab FB-Signal Add Ind Display format on the command lines:
1, 2, 3 or 4 FB-Signal 1
FB-Signal 2
FB-Signal 3
FB-Signal 4
Example:
Tab FB-Signal 1,
Feedback Data Point:
- Open the parameters
and go to [Control IT \
Out \ Reg_PSS]
After changing the attribute, the screen "Operations" has to be reloaded (toggle the Function
selection buttons).See Chapter 5.2.1 Command Buttons.
- Select settings
category
Keep-Alive Signaling
(Setting of the SlowCommMonRamp)
- Configure the
commands:
Setpoint Up,
Setpoint Down,
Value indication
Overrides:
- Open the Overrides Editor
(User Events)
Category:
- Scroll to the desired gray
highlighted line
[Event Prop...]
- Confirm the settings
- Cancel an entry
(gray highlighted line)
[Overrides...]
- Confirm the settings
New Event Mappings: Only required if plant specific event type is created in this application.
Mappings:
- Open the Text Editor
[Editor...]
[Event Pro..]
- Confirm the settings
Event Filters
Masking of the Event
Types in the Event Logger
Table
(not marked means
hidden)
5.12.6. Tools
[Select File]
- Confirm the settings
CPU Load
Chart View:
Display of the processor
performance
Update display
Read sampled values (2) (1)
- (1) Start of refresh all 5 s
- (2) Refresh once
(2) (1)
- (1) Stop auto refresh
- (2) not active (button gray)
Core 1: percentage of
processor load
Time
Data Collection
This feature collects all data in a ZIP-file.
Collection:
Start…Running…Finished
It takes some minutes to
prepare the ZIP-file.
Note: The output directory must be existing, otherwise a message will be displayed.
(See tab "Misc" on the next section)
Misc:
Number of last files
TR = Transient Recorder
ST = Slow Trending
FT = Fast Trending
Output directory:
- Choose your directory
Device groups:
- Define a new group/
device
- Edit the group/device
- Delete a group/device
Process Selector
5.12.7. Parameters
Dockable panel
(Not used in standard
applications)
Save to flash
- Open Control IT
The command and
feedback are found in the
parameter list
- Touch Parameters,
Target: [Control IT\ In\
Sys_ExcitationSystem..]
Parameter Configuration
Default setting
5.12.8. Powerchart
Powerchart Configuration
There is no saturation
calculation and no
values are taken over.
5.12.9. Diagrams
Diagrams Configuration
Fast Trending
Slow Trending
Sampling interval
- Default value: ≤1 s
- Slow: 10 s
(reduces the data
volume of the file)
Automatic restart
Settings
Plot Trigger
Timestamp:
It marks additionally
the timestamp at a
defined position
Save Format
5.12.12. Security
User Management
Dialog User
Management
- Create a User ID
(e.g. Sample)
- Create the password
or leave the line blank
if there is no password
desired
Or cancel a role
- Scroll to the desired
line (gray highlighted
line) and move the
role to the unassigned
user roles
[User Settings]
- Confirm the settings
Upgrading:
- Scroll in the table of
unassigned roles to
the desired role
[User Settings]
- Confirm the settings
[Application...]
- Confirm the settings
User Account
The following table lists the default settings of the different operation levels
Table 5-10 List of Settings Relating to the Operation Level
Settings of the user profile Level 2 Level 3 Level 4 Level 5 Administrator
Archive Installer: Edit / Operate Backup - - - X X
Archive Installer: Edit / Operate Download - - - X X
Archive Installer: Visible - - - X X
BitField: Read - - X X X
BitField: Write - - X X X
Configuration: Configurable Modules - - - X X
Device Testing: Edit - - - X X
Device Testing: Run - - X X X
Device Testing: Visible - - - - X
Diagrams: Configure - - - X X
Events: Clear Event Logger - - X X X
Events: Reset / Acknowledge Events X X X X X
Events: Save / Print / Freeze X X X X X
General: Minimize / Close CT - - - X X
IO Config: Setup - - - X X
IO Config: View - - - X X
IO Config: Visible - - - X X
Operations: Access Command Buttons X X X X X
Operations: Configure Meters X X X X X
Operations: Operate sliders 2) - - - - X
Operations: View sliders 2) - - - - X
Options: Access Options Dialog X X X X X
Options: Alarms and Events - - - X X
Options: Application - - - X X
Options: Configuration Information - - - - X
Options: Data Collection - Execute - - - - X
Options: Data Collection - Setup - - - - X
Options: Data Collection - Visible - - - - X
Options: General - - X X X
Options: Import Settings - - - X X
Options: Number Display 2) - - - - X
Options: Operations - - - X X
Options: Tools X X X X X
Options: Trending - - - X X
Parameter: Edit Parameter Configuration - - X X X
Parameter: View Parameter Configuration - - X X X
Parameters: Configure save to flash - - X X X
Parameters: Create Snapshot - - X X X
Parameters: Edit Parameter Values - - X X X
Parameters: Edit Parameter Values Level 1 1) - - - X X
Parameters: Edit Parameter Values Level 2 2) - - - - X
Parameters: Edit Parameter Values Level 3 2) - - - - X
Parameters: Edit Parameter Values Level 4 2) - - - - X
Parameters: Edit Parameter Values Level 5 2) - - - - X
Parameters: Execute save to flash - - X X X
Parameters: Manage "My Groups" - - X X X
Parameters: Open Snapshot - - X X X
Parameters: Save Snapshot - - X X X
Parameters: View Parameter Values Level 1 1) - - X X X
Parameters: View Parameter Values Level 2 2) - - - - X
Note 1): Each parameter can be made invisible for a specified user group, allocated in the database.
Note 2): Not used for this application.
Note: After changing an attribute, the ECT must be restarted. Otherwise the setting will not become
active.
Examples of adjustment
[Editor]
- Confirm the CT Location
[Editor]
- Confirm the setting
[Device Configuration]
- Confirm the settings
6.2. Grounding
PIN 3 of the Phoenix Contact connector
Conductor cross-section: up to 2.5 mm2 (AWG 13)
6.3. Disposal
Improper disposal of electrical devices can lead to environmental
damage. It is therefore important that qualified personnel carry out
the disposal of electronic equipment.
The metallic casing does not present any risk to the environment
and can be recycled.
The printed circuit boards are simple to remove. The boards must
be removed and should be disposed of by a licensed disposal
company.
Environmentally hazardous elements such as capacitors must be
separated from the boards.
Commissioning
Safety Regulations
The instructions according to Chapter 1 - Safety Instructions, page 9 shall
be followed.
Commissioning
Before switching on, check whether the connection terminals are wired up
according to the plant schematic.
See Chapter 3 - Installation, page 15
Maintenance
Manual Handling: Overflow of the hard disk.
Check the directory "My Documents \ …." number and size of the files to
prevent an overflow of the hard disk.
Windows 10:
C:\ProgramData\ABB\Control Terminal\[Release]\ECT\Logs\..
Recommended Restart
It is recommended to restart the ECT device each month.
A reboot of the device reinitializes the system.
Fault Message
There is a communication mistake between the ECT and channel
(Control) system.
Probably there is a mismatch between the configuration file version and
the application file version.
Application Exception
If this message appears, follow the instructions in the screenshot below.
To recover the ECT from the "unexpected error", disconnect the 24 V
supply to restart the ECT.
Restart
ECT operates independently of the control system and it always starts up
in passive mode. The hand of the panel control button is gray.
See Chapter 5.1.2 - Panel Control Button, page 26.
Repair
It is not permitted to open the case of an ECT.
In this case all warranty claims are void.
A defective unit should be sent in for repair. See manufacturer’s details
on page 8.
Appendix
9.1. Appendix A
Directory Structure
Windows 10 has the following directory structure and name. The first part
of the path to the folder Control Terminal\ECT\Data or Export:
9.2. Appendix B
ECT SW Rel: Always use the latest SW from the Excitation Support DB.
For Win10 it is recommended to use ECT SW rel ≥ 2.6.xx
9.3. Appendix C
Product Information
Product designation Human-Machine-Interface
Type 15" TPC-1551T
Mechanical Data
Touch screen, diagonal (XGA: 1024 x 768) 15"
Weight 4.4 kg (9.7 lbs)
Protection class, front side IP 66
Dimensions (WxHxD) see Chapter 9.3.3
Climatic Stability
Temperature range for operation -20 to +60 °C
Temperature range for storage -30 to +70 °C
Electrical Data
Supply voltage 24 V dc ±20%
Power consumption, average / max 50.8 W (typical)
Operating System
Microsoft Windows 10 Enterprise LTSC
The ECT Software was designed and tested in this operating system.
Relevant Standards
Test certificates CE, FCC Class A, UL, CCC, BSMI
Product Information
Product designation Human-Machine-Interface
Type 15" IRTPPC15
Mechanical Data
Touch screen, diagonal (XGA: 1024 x 768) 15"
Weight 4.5 kg (9.9 lbs)
Protection class, front side IP 65
Dimensions (WxHxD) see Chapter 9.3.3
Electrical Data
Supply voltage 24 V dc ±20%
Operating System
Microsoft Windows 10 IoT Enterprise Entry 2019
The ECT Software was designed and tested in this operating system.
Software Release ≥CT 2.6.x
Relevant Standards
Test certificates CE, FCC Class A, UL, CCC, BSMI