Temperature Mapping Study
Temperature Mapping Study
To verify an area maintains temperature within permitted limits at all times (2-8 C)
Collect temperature data along the length and breadth of the room at different operating conditions for
couple of days (3-4 days per WHO protocol standards) and analyze the data
Identify hot and cold points of a room – those that are below and above the limits; isolate those
areas/not use to store medicines/resolve issues
If humidity is to be maintained within a maximum range (60-65%) or limits of 40-65% Rh. Install
dehumidifiers
For products which are temperature sensitive and which will get damaged if limits not maintained
Place data loggers in the room – location and number depends on the number and position of the doors,
racks etc; 1 data logger for every 5 cm3 and all racks; a data logger is a temperature monitoring system
Data collection
Power failure – to determine how long temp is maintained within limits after power failure
In addition to temperature mapping, all installed major and minor equipment are tested for
temperature limits
CAPA – Corrective Actions Preventive Actions