ePM 10M/12M
Patient monitor
Physical Specifications QTc formula Bazett, Fridericia, Framingham, or Hodges
Weight ePM 10M: 4.0 Kg QT/QTc range 200 to 800 ms
ePM 12M: 4.8 Kg QT accuracy ± 30 ms
(Standard configuration, excluding modules, QT resolution 4 ms
recorder, battery and accessories.) QTc resolution 1 ms
Size ePM 10M: 269 x252 x159mm QT-HR range Adult: 15 to 150 bpm
ePM 12M: 310 x289 x169mm Pediatric/Neonate: 15 to 180 bpm
Display screen Capacitive screen, support multi-touch Respiration
operation. Lead I or II, auto
ePM 10M: 10.1-inch, 1280 x 800 pixels RR range 0 to 200 rpm
ePM 12M: 12.1-inch, 1280 x 800 pixels RR accuracy ± 1 rpm (0 to 120 rpm)
Display channel ePM 10M: Up to 8 waveform channels ± 2 rpm (121 to 200 rpm)
ePM 12M: Up to 10 waveform channels RR resolution 1 rpm
ECG Sweep speed 3 mm/s, 6.25 mm/s, 12.5 mm/s, 25 mm/s,
Meet standards of IEC 60601-2-27 and IEC 60601-2-25. 50 mm/s
Lead set 3-lead: I, II, III Apnea time 10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s
5-lead: I, II, III, aVR, aVL, aVF, V SpO2
** 6-lead: I, II, III, aVR, aVL, aVF, Va, Vb Meet standards of ISO 80601-2-61.
12-lead: I, II, III, aVR, aVL, aVF, V1 to V6 SpO2 module Mindray SpO2, Nellcor SpO2
Automatic 3/5/6/12 - lead recognition. SpO2 range 0 to 100 %
Input signal range ± 10 mV (p-p) SpO2 accuracy Adult/Pediatric: ± 2 % (70 to 100%)
Electrode offset potential tolerance ± 800 mV Neonate: ± 3 % (70 to 100%)
Sweep speed 6.25 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s Perfusion indicator (PI) Yes, for Mindray SpO2
Gain x 0.125, x 0.25, x 0.5, x 1, x 2, x 4, auto Pitch tone Yes
Waveform format Standard, Cabrera Dual-SpO2 Yes, SpO2, SpO2b, ∆SpO2
Bandwidth Diagnostic mode: 0.05 to 150 Hz Refreshing rate ≤1s
Monitor mode: 0.5 to 40 Hz PR
Surgical mode: 1 to 20 Hz PR range 20 to 300 bpm (from SpO2)
ST mode: 0.05 to 40 Hz 20 to 350 bpm (from IBP)
CMRR Diagnostic mode: > 90 dB 30 to 300 bpm (from NIBP)
Monitor, Surgical, ST mode: > 105 dB PR accuracy ± 3 bpm (20 to 300 bpm, from Mindray SpO2)
Pace Detection Amplitude: ± 2 mV to ± 700 mV ± 3 bpm (20 to 300 bpm, from Nellcor SpO2)
Width: 0.1 to 2 ms ±1 bpm or ±1 %, whichever is greater (from IBP)
Rise time: 10 to 100 μs ± 3 bpm or ±3 %, whichever is greater
Defib. protection Withstand 5000V (360J) defibrillation (from NIBP)
Recovery time <5 s Refreshing rate ≤1s
Provides glasgow resting 12-lead ECG algorithm. Temperature
Heart Rate Meet standard of ISO 80601-2-56.
HR rang Adult: 15 to 300 bpm Technique Thermal resistance
Pediatric/Neonate: 15 to 350 bpm Channels 2 channels
HR accuracy ± 1 bpm or ± 1%, whichever is greater. Temp range 0 to 50 °C (32 to 122 °F)
HR resolution 1 bpm Temp accuracy ± 0.1 °C or ± 0.2 °F (without probe)
Arrhythmia Analysis Temp resolution 0.1 °C
Intended use for adult, pediatric and neonate. Refreshing rate ≤1s
Multi-lead, 25 classifications. Asystole, VFib/VTac, Vtac, Vent. Brady, NIBP
Extreme Tachy, Extreme Brady, Vrhythm, PVCs/min, Pauses/min, Couplet, Meet standards of ISO 80601-2-30.
Bigeminy, Trigeminy, R on T, Run PVCs, PVC, Tachy, Brady, Missed Beats, Technique Oscillometry
PNP, PNC, Multif. PVC, Nonsus. Vtac, Pause, Irr. Rhythm., Afib (for adult Operation mode Manual, Auto, STAT, Sequence
only). Parameters Systolic, diastolic, mean
ST Segment Analysis Max measurement time Adult/Pediatric: 180 s, Neonate: 90 s
Intended use for adult, pediatric and neonate. Systolic range Adult: 25 to 290 mmHg
ST range - 2.5 to + 2.5 mV Pediatric: 25 to 240 mmHg
ST accuracy ± 0.02 mV or ± 10%, whichever is greater Neonate: 25 to 140 mmHg
(- 0.8 to + 0.8 mV) Diastolic range Adult: 10 to 250 mmHg
ST resolution 0.01 mV Pediatric: 10 to 200 mmHg
QT Analysis Neonate: 10 to 115 mmHg
Intended use for adult, pediatric, and neonate. Mean range Adult: 15 to 260 mmHg
Parameters QT, QTc, ∆QTc Pediatric: 15 to 215 mmHg
Neonate: 15 to 125 mmHg CO2 accuracy ±2 mmHg (0 to 38 mmHg)
NIBP accuracy Max mean error: ± 5 mmHg ±5 % of the reading (0.08 % increased in error
Max standard deviation: 8 mmHg for every 1 mmHg if the reading is more than
NIBP resolution 1 mmHg 38mmHg) (39 to 99 mmHg)
Assisting venous puncture Yes awRR range 0 to 150 rpm
IBP awRR accuracy ±1 rpm (0 to 70 rpm)
Meet standard of IEC 60601-2-34. ±2 rpm (71 to 120 rpm)
Channels Up to 4 channels ±3 rpm (121 to 150 rpm)
Sensitivity 5 μV/V/mmHg Apnea time 10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s
Impedance range 300 to 3000 Ω Capnostat Mainstream CO2
IBP range -50 to 360 mmHg Meet standard of ISO 80601-2-55.
IBP accuracy ±1 mmHg or ±2 %, whichever is greater Rise time < 60 ms
IBP resolution 1 mmHg Sweep speed 3 mm/s, 6.25 mm/s, 12.5 mm/s, 25 mm/s,
PPV range 0 to 50 % 50 mm/s
PAWP Yes. CO2 range 0 to 150 mmHg
ICP measurement Support CO2 accuracy ±2 mmHg (0 to 40 mmHg)
Support waveforms overlapping. ±5 % of the reading (41 to 70 mmHg)
C.O. ±8 % of the reading (71 to 100 mmHg)
Technique Thermodilution ±10 % of the reading (101 to 150 mmHg)
C.O. range 0.1 to 20 L/min awRR range 0 to 150 rpm
C.O. accuracy ±0.1 L/min or ±5%, whichever is greater awRR accuracy ±1 rpm
C.O. resolution 0.1 L/min Multi-gas
TB range 23 to 43 °C Meet standard of ISO 80601-2-55.
TI range 0 to 27 °C Technique Infrared absorption, paramagnetic
TB, TI accuracy ± 0.1 °C (without sensor) properties for O2 monitoring
TB, TI resolution 0.1 °C Gas CO2, O2, N2O, Des, Iso, Enf, Hal, Sev
Artema Sidestream CO2 Warm-up time ISO accuracy mode: 45 s
Meet standard of ISO 80601-2-55. Full accuracy mode: 10 min
**Options: Paramagnetic O2 sensor. Sample flow rate (with DRYLINE II TM watertrap)
CO2 sample flow rate Adult/pediatric watertrap: 200 ml/min
120 ml/min (DRYLINE II TM watertrap for adult/pediatric) Neonate watertrap: 120 ml/min
90/70 ml/min (DRYLINE II TM watertrap for neonate) Sample flow rate accuracy ±10 ml/min or ±10%, whichever is greater.
CO2 sample flow rate accuracy Delay time <4s
±15 ml/min or ±15 %, whichever is greater. Response time DRYLINE II TM watertrap for adult/pediatric,
CO2 Response time ≤ 5.0 s @ 120ml/min (for adult/pediatric) 200 ml/min:
≤4.5 s @ 90 ml/min (for neonate) CO2: ≤ 4.2 s
≤ 5.0 s @ 70 ml/min (for neonate) N2O: ≤ 4.3 s
O2 Response time ≤ 5.0 s @ 120 ml/min Enf/Iso/Hal/Sev/Des: ≤ 4.5 s
≤ 4.5 s @ 90ml/min O2: ≤ 4 s
Sweep speed 3 mm/s, 6.25 mm/s, 12.5 mm/s, 25 mm/s, DRYLINE II TM watertrap for neonate,
50 mm/s 120 ml/min:
CO2 range 0 to 150mmHg CO2: ≤ 4 s
CO2 accuracy Full accuracy mode: N2O: ≤ 4.2 s
0 - 40 mmHg: ± 2 mmHg O2: ≤ 4 s
41 - 76 mmHg: ± 5% of reading Enf/Iso/Hal/Sev/Des: ≤ 4.4 s
77 - 150 mmHg: ± 10% of reading CO2 range 0 to 30 %
ISO accuracy mode: CO2 accuracy ±0.1%ABS (0 to 1%)
Add ± 2 mmHg to the full accuracy mode ±0.2%ABS (1 to 5%)
CO2 resolution 1 mmHg ±0.3%ABS (5 to 7%)
O2 range 0 to 100 % ±0.5%ABS (7 to 10%)
O2 accuracy ±1 % (0 to 25 %) O2 range 0 to 100 %
±2 % (25.1 to 80 %) O2 accuracy ±1%ABS (0 to 25%REL)
±3 % (80.1 to 100 %) ±2%ABS (25 to 80%REL)
O2 resolution 0.1 % ±3%ABS (80 to 100%REL)
awRR range 0 to 150 rpm N2O range 0 to 100 %
awRR accuracy ± 1 rpm ( 0 to 60 rpm) N2O accuracy ±2%ABS (0 to 20%REL)
± 2 rpm (61 to 150 rpm) ±3%ABS (20 to 100%REL)
Apnea time 10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s Enf/Iso/Hal/Sev/Des range 0 to 30 %
Oridion Microstream CO2 awRR range 2 to 100 rpm
Meet standard of ISO 80601-2-55. awRR accuracy ±1 rpm (2 to 60 rpm)
Sample flow rate 50 -7.5+15 ml/min Apnea time 10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s
Initialization time 30 s (typical) Provide MAC value (support calibrated by age).
Response time 2.9 s (typical) Support two mixed gas identify and monitoring.
Sweep speed 3 mm/s, 6.25 mm/s, 12.5 mm/s, 25 mm/s, BISx/BISx4
50 mm/s Meet standard of IEC 60601-2-26.
CO2 range 0 to 150 mmHg Technique Bispectral Index
Impedance range >5 MΩ IEEE 802.11n: 6.5 to 72.2 Mbps
EEG bandwidth 0.25 to 100 Hz Output power < 20dBm (CE requirement: detection
BIS range 0 to 100 (BIS, BIS L, BIS R) mode- RMS)
SQI range 0 to 100 % (SQI, SQI L, SQI R) < 30dBm (FCC requirement: detection
ASYM 0 to 100% mode- peak power)
DSA trend Yes Operating mode Infrastructure
Data Review Data security WPA-PSK, WPA2-PSK, WPA-Enterprise,
For 2G storage WPA2-Enterprise (EAP-FAST. EAP-TLS, EAP-
Trends data Up to 120 hours @ 1min TTLS, PEAP-GTC, PEAP-MSCHAPv2, PEAP-TLS,
Events Up to 1000 events, including parameter alarms, LEAP)
arrhythmia events technical alarms, and so Encryption: TKIP and AES
on. Interfacing
NIBP Up to 1000 sets Main unit AC power connector (1)
For 16G storage VGA port (1)
Trends data Up to 240 hours @ 1min, 2400 hours @ 10 min Network connector (1), RJ45
Events Up to 2000 events, including parameter alarms, USB 2.0 connector (2)
arrhythmia events technical alarms, and so Analog output/nurse call/defib. Sync. Port (1)
on. Integrated module rack (1), for 2 slots
NIBP Up to 3000 sets Barcode scanner Support 1D and 2D barcode
For 2G & 16G storage Remote control Support
Interpretation of resting 20 sets of 12-lead ECG results Thermal recorder 3 traces (paper 50 mm width, 20 m length)
Full disclosure Up to 48 hours for all parameter waveforms. Network printer Support
The specific storage time depends on the Power
waveforms stored and the number of stored Line voltage 100 to 240 VAC (±10 %)
waveforms. Maximum current 2.0A
OxyCRG 400 OxyCRG events Frequency 50/60 Hz (±3 Hz)
ST review Up to 120 hours @ 5 min Battery Rechargeable lithium-ion battery,
Minitrend Yes 2600mAh/4500mAh
Alarms Rechargeable smart lithium-ion battery
Audible indicator Yes, 3 different alarm tones, and prompt 5600mAh
tone ePM 10M/12M:>2 hours run time (2600mAh)
Visible indicator Red/yellow/cyan LED, and alarm message ePM 10M/12M:>4 hours run time (4500mAh)
display ePM 10M:>6 hours run time (5600mAh x1)
Provide AlarmSight infographic alarm indicator. ePM 12M:>4.5 hours run time (5600mAh x1)
Special Functions ePM 12M:>9 hours run time (5600mAh x2)
Clinical Assistive Application (CAA): ST Graphic TM, EWS, GCS, 24h ECG Recharge time (power off ) 2.5 hours to 90%(2600mAh)
summary, NIBP analysis. 5 hours to 90% (4500mAh)
Calculations (drug, hemodynamic, Oxygenation, Ventilation, Renal), and 5 hours to 90% (5600mAh x1)
Titration table. 10 hours to 90% (5600mAh x2)
Wi-Fi Communications Environmental requirements
Protocol IEEE 802.11a/b/g/n Temperature Operating: 0 to 40 °C (without AG),
Modulation mode DSSS and OFDM 10 to 40 °C (with AG)
Operating frequency IEEE 802.11b/g/n (2.4G): Storage: -20 to 60 °C
ETSI/FCC/KC: 2.4 to 2.483 GHz Humidity Operating: 15 to 95 % (non condensing)
MIC: 2.4 to 2.495 GHz Storage: 10 to 95 % (non condensing)
IEEE 802.11a/n (5G): Barometric Operating: 427.5 to 805.5 mmHg
ETSI: 5.15 to 5.35 GHz, 5.47 to 5.725 GHz (57.0 to 107.4 kPa)
FCC: 5.15 to 5.35 GHz, 5.725 to 5.82 GHz Storage: 120 to 805.5 mmHg
MIC: 5.15 to 5.35 GHz (16.0 to 107.4 kPa)
KC: 5.15 to 5.35 GHz, 5.47 to 5.725 GHz,
5.725 to 5.82 GHz
Channel spacing 5 MHz @ 2.4 GHz, 20 MHz @ 5 GHz --------------------------------------------------------------
Wireless baud rate IEEE 802.11a: 6 to 54 Mbps Some of functions marked with an asterisk may not be available. Please
IEEE 802.11b: 1 to 11 Mbps contact your local Mindray sales representative for the most current
IEEE 802.11g: 6 to 54 Mbps information.
| Healthcare with reach are registered trademarks or trademarks owned by Shenzhen Mindray Bio-medical Electronics Co., LTD
Mindray Building, Keji 12th Road South, © 2018 Shenzhen Mindray Bio-medical Electronics Co., Ltd. All rights reserved. Specifications subject to changes without prior notice.
High-tech Industrial Park, Nanshan, Shenzhen 518057, P.R. China P/N: ENG-ePM 10M/12M Datasheet-210285x4P-20181213
Tel: +86 755 8188 8998 Fax: +86 755 26582680
E-mail: [email protected] www.mindray.com