BeneFusion SP5 Ex
Syringe Pump
Data Sheet
Physical Specifications History log up to 2000 records
Weight < 2.5kg Volume collection available in 4 methods: 24H total, current
Size 295 x 174 x 87mm total, period, timing
Dose Error Reduction available; definition of dose limits;
Screen 3.5-inch TFT color LCD; 480 x 272 pixels
System (DERS) automatic alarms when reaching dose
Brightness 1-8 levels, adjustable
limits
Display infusion rate, VTBI, total volume,
pressure limit & status, drug name,
Syringes
remaining time, alarm status, battery
Compatibility 2/3/5/10/20/30/50/60ml;
status, size of syringe, brand of syringe
automatic recognition of syringe size
Parameters Specifications
Alarms
Accuracy ≤ ±2%
Type audible and visual alarm
Mode rate mode, time mode, body weight mode
3 Levels High: occlusion / no power / syringe
ramp up/down mode, sequential mode,
empty /no syringe/ KVO finished/system
micro-infusion mode, loading dose mode,
error
intermittent mode
Flow rate 0.1-2000ml/h
Medium: abnormal system/ standby times
Increment 0.01ml(0.1-99.99ml/h);
expired
0.1ml(100-999.9ml/h);
1ml(1000-2000ml/h)
Low: reminder/ low battery/syringe near
Preset volume(VTBI) 0.10 - 9999.99ml (increment: 0.01ml)
empty/ VTBI near done/ No AC power
Preset time 00:00:01-99:59:59, adjustable
Sound volume 1-8 levels selectable, default is level 4
Accumulated volume 0.1-9999.99 ml
Reminder 1-5 minutes selectable;
KVO 0.1-5.0ml/h adjustable, step 0.1ml/h,
ON/OFF switchable
default is 1.0ml/h
Purge rate 0.10-2000ml/h
Battery
Bolus rate 0.1-2000ml/h (automatic or manual)
Operating time ≥ 10 hours at 5ml/h
Occlusion detection 12 levels selectable: 75-975mmHg,
Charging time ≤ 6 hours to full capacity
default is 450mmHg;
4 units of pressure selectable:
Interface RS232; nurse call connector; DC adapter
mmHg/kPa/bar/psi, default is mmHg
Anti-bolus unexpected bolus reduced when the
Power Supply voltage 100-240 V~ ; frequency 50/60Hz;
occlusion occurs
current 0.4-0.14A
Dose rate units ng/kg/min, μg/kg/h, ug/kg/min, ug/kg/h,
ug/kg/24h, mg/kg/min, mg/kg/h,
Work Environment
mg/kg/24h, g/kg/min, g/kg/h, mU/kg/min,
Temperature 5-40ºC for operating; -20-60 ºC for
mU/kg/h, U/kg/min, U/kg/h, U/kg/24h,
kU/kg/h, EU/kg/h, mmol/kg/h, mol/kg/h, storage
kcal/kg/h, kcal/kg/24h, mEq/kg/min, Relative humidity 15-95% for operating; 10-95% for storage
mEq/kg/h
Atmosphere pressure 57-106 kPa for operating; 50-106 kPa for
Auto-lock time 1 - 5 minutes selectable; ON/OFF
switchable storage
Level of resistance to water and dust: IP23
Drug library up to 2000 drugs; ON/OFF switchable
| Healthcare with reach are registered trademarks or trademarks owned by Shenzhen Mindray Bio-medical Electronics Co., LTD
Mindray Building, Keji 12th Road South, © 2018 Shenzhen Mindray Bio-medical Electronics Co., Ltd. All rights reserved. Specifications subject to changes without prior notice.
High-tech Industrial Park, Nanshan, Shenzhen 518057, P.R. P/N: ENG-BeneFusion SP5 Ex Datasheet-210285x2P-20190110
China
Tel: +86 755 8188 8998 Fax: +86 755 26582680
E-mail: [email protected] www.mindray.com