REPORT OF CALIBRATION
Calibration Date Calibration Due Date Issue Date S.R.F No. & Date Certificate No.
03-03-2022 02-09-2022 05-03-2022 MQS/229001/12 & 02-03-2022 MQS/229001/12/01
A. Company Name And Address: B. Description And Identification ff UUC*
شركة النصاف Inst. Name Coriolis Mass Flow Meter
Make/Model Endress+Hauser / Promass 83
I.D. No. ----
Sr. No. DB15CB02000
Range 0 to 8700 l/hr
Resolution/L.C 0.1 l/hr
Company ID C-548 Accuracy ± 0.1%
C. Environmental Condition Connection 2”
Temperature 25 ± 4 °C Laboratory Laboratory
Relative Humidity 50 ± 20 % Medium of Calibration Liquid
D. Details Of Standard Used In Calibration
Sr. Instrument Serial Source of
Make/ Model Range Uncertainty Valid up to
No. Name No./ID No. Traceability
Coriolis Mass Endress+Hauser / S71D70020 0 to 240
1. 0.05% 06/08/2022 SAS
Flow Meter PromassF 00 m3/hr
E. Principle/Methodology of Calibration:
Comparison Method as per Calibration Procedure: CP/M-27, API MPMS Chapter 4.8 & 5.6
F. RESULTS
Master Meter
Run 1 Run 2 Run 3 Run 4 Run 5 Average
Data
Count 9997 10023 10005 10032 10027 2003.36
Temperature (°C) 20.1 20.1 20.1 20.1 20.1 20.14
Pressure (bar) 2.2 2.2 2.2 2.2 2.2 2.19
3
Density (kg/m ) 998.2 998.2 998.2 998.2 998.2 998.18
Flow (kg/hr) 45876 45875 45215 45715 45127 45561.6
Test Meter Data Run 1 Run 2 Run 3 Run 4 Run 5 Average
Count 10024 10046 10032 10058 10052 2008.48
Temperature (°C) 20.1 20.1 20.1 20.1 20.1 20.14
Pressure (bar) 2.2 2.2 2.2 2.2 2.2 2.19
3
Density (kg/m ) 998.2 998.2 998.2 998.2 998.2 998.18
Meter Factor 0.99730 0.99771 0.99731 0.99741 0.99751 0.997451
Flow (kg/hr) 45876 45875 45215 45715 45127 45561.6
Calculation Data:
Prover Master
A. Total Counts 2003.36
B. Linear K Factor 1
C. Prover/Master Meter Mass (A/B) (Kg) 2003.36
Meter
G. Total Counts 2008.48
H. K factor 1
I. Meter Mass (kg) 2008.48
M. Meter Factor (C/I) 0.99745
N. Repeatability%=(Max-Min)/Min*100 0.04
Sarbasti (32 Park), 8th Avenue, Building No. 266, Erbil (Iraq)
Page 1 of 2 E: [email protected] | www.miqias.com
Format: 7.8F-01 M: +964-7725579191, 7738218454 Security Number: 8852
REPORT OF CALIBRATION
Remarks: 1) Reported values are mean of 5 values.
2) At all Cal. Points UUC readings are found within limits and hence calibration passes.
❖ This Certificate refers only to the particular item submitted for calibration.
❖ This certificate Shall not be reproduced except in full, without the written permission of the Laboratory
❖ Results Reported are valid at the time of and under the stated condition of Measurement
Note:
❖ UUC* : Unit Under Calibration, N/A*: Not Applicable
❖ Traceability: The Measurement was performed using traceable reference standards. The measurement made
by MIQIAS as reported in this certificate are traceable to S.I. unit through National/International Institutes
End of Calibration Results
Accreditation No Accreditation No Accreditation No
IAS2002Q1756 IAS2002E1757 IAS310301901
Calibrated By Checked By Approved By
Calibration Engineer Quality Manager C.E.O / TM