Report of Calibration: 03-03-2022 02-09-2022 05-03-2022 MQS/229001/12 & 02-03-2022 MQS/229001/12/01

The document reports the results of a calibration test performed on a Coriolis mass flow meter. The test involved running the meter through 5 runs and comparing its measurements to a master reference meter. Key details include: - The meter was calibrated on March 3rd 2022 and is due for recalibration by September 2nd 2022. - Testing occurred at a temperature of 20.1°C and involved flowing liquid through the meter at rates between 0-8700 l/hr. - Results showed the meter's measurements averaged 0.997451 times those of the master meter, within the required 0.1% accuracy. - The calibration was performed according to established procedures
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
84 views2 pages

Report of Calibration: 03-03-2022 02-09-2022 05-03-2022 MQS/229001/12 & 02-03-2022 MQS/229001/12/01

The document reports the results of a calibration test performed on a Coriolis mass flow meter. The test involved running the meter through 5 runs and comparing its measurements to a master reference meter. Key details include: - The meter was calibrated on March 3rd 2022 and is due for recalibration by September 2nd 2022. - Testing occurred at a temperature of 20.1°C and involved flowing liquid through the meter at rates between 0-8700 l/hr. - Results showed the meter's measurements averaged 0.997451 times those of the master meter, within the required 0.1% accuracy. - The calibration was performed according to established procedures
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

REPORT OF CALIBRATION

Calibration Date Calibration Due Date Issue Date S.R.F No. & Date Certificate No.
03-03-2022 02-09-2022 05-03-2022 MQS/229001/12 & 02-03-2022 MQS/229001/12/01
A. Company Name And Address: B. Description And Identification ff UUC*
‫شركة النصاف‬ Inst. Name Coriolis Mass Flow Meter
Make/Model Endress+Hauser / Promass 83
I.D. No. ----
Sr. No. DB15CB02000
Range 0 to 8700 l/hr
Resolution/L.C 0.1 l/hr
Company ID C-548 Accuracy ± 0.1%
C. Environmental Condition Connection 2”
Temperature 25 ± 4 °C Laboratory Laboratory
Relative Humidity 50 ± 20 % Medium of Calibration Liquid
D. Details Of Standard Used In Calibration
Sr. Instrument Serial Source of
Make/ Model Range Uncertainty Valid up to
No. Name No./ID No. Traceability
Coriolis Mass Endress+Hauser / S71D70020 0 to 240
1. 0.05% 06/08/2022 SAS
Flow Meter PromassF 00 m3/hr
E. Principle/Methodology of Calibration:
Comparison Method as per Calibration Procedure: CP/M-27, API MPMS Chapter 4.8 & 5.6
F. RESULTS
Master Meter
Run 1 Run 2 Run 3 Run 4 Run 5 Average
Data
Count 9997 10023 10005 10032 10027 2003.36
Temperature (°C) 20.1 20.1 20.1 20.1 20.1 20.14
Pressure (bar) 2.2 2.2 2.2 2.2 2.2 2.19
3
Density (kg/m ) 998.2 998.2 998.2 998.2 998.2 998.18
Flow (kg/hr) 45876 45875 45215 45715 45127 45561.6
Test Meter Data Run 1 Run 2 Run 3 Run 4 Run 5 Average
Count 10024 10046 10032 10058 10052 2008.48
Temperature (°C) 20.1 20.1 20.1 20.1 20.1 20.14
Pressure (bar) 2.2 2.2 2.2 2.2 2.2 2.19
3
Density (kg/m ) 998.2 998.2 998.2 998.2 998.2 998.18
Meter Factor 0.99730 0.99771 0.99731 0.99741 0.99751 0.997451
Flow (kg/hr) 45876 45875 45215 45715 45127 45561.6
Calculation Data:
Prover Master
A. Total Counts 2003.36
B. Linear K Factor 1
C. Prover/Master Meter Mass (A/B) (Kg) 2003.36
Meter
G. Total Counts 2008.48
H. K factor 1
I. Meter Mass (kg) 2008.48
M. Meter Factor (C/I) 0.99745
N. Repeatability%=(Max-Min)/Min*100 0.04
Sarbasti (32 Park), 8th Avenue, Building No. 266, Erbil (Iraq)
Page 1 of 2 E: [email protected] | www.miqias.com
Format: 7.8F-01 M: +964-7725579191, 7738218454 Security Number: 8852
REPORT OF CALIBRATION

Remarks: 1) Reported values are mean of 5 values.


2) At all Cal. Points UUC readings are found within limits and hence calibration passes.
❖ This Certificate refers only to the particular item submitted for calibration.
❖ This certificate Shall not be reproduced except in full, without the written permission of the Laboratory
❖ Results Reported are valid at the time of and under the stated condition of Measurement
Note:
❖ UUC* : Unit Under Calibration, N/A*: Not Applicable
❖ Traceability: The Measurement was performed using traceable reference standards. The measurement made
by MIQIAS as reported in this certificate are traceable to S.I. unit through National/International Institutes
End of Calibration Results

Accreditation No Accreditation No Accreditation No


IAS2002Q1756 IAS2002E1757 IAS310301901

Calibrated By Checked By Approved By


Calibration Engineer Quality Manager C.E.O / TM

Sarbasti (32 Park), 8th Avenue, Building No. 266, Erbil (Iraq)
Page 1 of 2 E: [email protected] | www.miqias.com
Format: 7.8F-01 M: +964-7725579191, 7738218454 Security Number: 8852