Infusion Pump HP-60 Service Manual: Medcaptain Medical Technology Co., LTD
Infusion Pump HP-60 Service Manual: Medcaptain Medical Technology Co., LTD
Service Manual
Before maintaining the HP-60 infusion pump, please read this Manual carefully and
follow the safety precautions and maintenance instructions contained herein.
Statement
Version Information
V1.0
Contents
STATEMENT ....................................................................................................................................... 2
4 HARDWARE ........................................................................................................................ 18
6 MAINTENANCE .................................................................................................................. 28
8 CALIBRATION .................................................................................................................... 34
1.4 DISMANTLING THE PERISTALSIS PUMP MOTOR SPEED MEASURING PCBA AND PUMP FINGER POSITION
DETECTION PCBA .................................................................................................................................. 55
1.5 DISMANTLING THE SLIDING BLOCK AND HP-60 DOOR POTENTIOMETER ASSEMBLY ................................... 57
1.6 DISMANTLING THE HP-60 TUBE DETECTION PCBA, ANTI-FREE-FLOW CLAMP PCBA, PATIENT SIDE OCCLUSION
PRESSURE DETECTION PCBA AND ITS CONNECTION FFC ................................................................................ 59
1.7 DISMANTLING THE PERISTALSIS PUMP ............................................................................................... 60
1.8 DISMANTLING THE PERISTALSIS PUMP MOTOR ................................................................................... 61
1.9 DISMANTLING THE DOOR DRIVE ASSEMBLY ........................................................................................ 63
1.10 DISMANTLING THE HP-60 FLUID SIDE OCCLUSION PRESSURE DETECTION PCBA .................................. 65
1.11 DISMANTLING THE SEPARATE AIR BUBBLE SENSOR MODULE A ......................................................... 65
1.12 DISMANTLING THE HP-60 PUMP DOOR PCBA .............................................................................. 66
1.13 DISMANTLING THE SEPARATE AIR BUBBLE SENSOR MODULE B.......................................................... 67
1.14 DISMANTLING THE HP-60 VENEER .............................................................................................. 68
1.15 DISMANTLING THE HP-60 PUMP DOOR BOARD’S CONNECTION FFC ................................................. 69
1.16 DISMANTLING THE HP-60 MAIN PCBA ....................................................................................... 70
1.17 DISMANTLING THE HP-60 DROP SENSOR INTERFACE PCBA ............................................................. 72
1.18 DISMANTLING THE FASTENER ASSEMBLY ....................................................................................... 73
1.19 ASSEMBLY............................................................................................................................... 74
1 Important Information
- The TSC requires that the instructions in the manuals are observed
- Depending on the unit type, the Service Program must be called which may lead to a dangerous
unit condition in case of inappropriate operation. Furthermore, a special service connector may be
necessary.
Maintenance in violation of the requirements in this service manual may seriously damage
the device.
The service personnel must be trained and permitted in writing by MEDCAPTAIN to do so.
The maintenance must be performed under electrostatic discharge (ESD) protective
conditions. Do not touch PCBA or semiconductors by hand without any protective measure.
Do not touch the display by using sharp objects. Otherwise, the display may be damaged.
Do not autoclave the infusion pump.
Before using the internal battery, check the battery to ensure that sufficient power is available.
Recharge the battery if required.
Liquid intrusion into the AC power socket or nurse call socket may cause short-circuit.
Before connecting the power cable, check if the connecting parts are dry. If liquid has spilled
onto the infusion pump, clean the pump with a dry cloth and then perform an inspection and
maintenance.
Use the maintenance parts and accessories provided by MEDCAPTAIN for replacement and
maintenance.
Do not maintain or use the infusion pump in a flammable environment.
High-frequency surgical equipment, mobile phones, wireless devices, and defibrillators may
cause interference on the infusion pump. Therefore, keep the infusion pump away from these
devices when using the pump.
After maintenance, perform a safety test and clean the pump according to this manual.
1.5 Quality Control
MEDCAPTAIN meets the requirements for quality management system specified in
ISO9001 and ISO13485 standards. Complying with the requirements of the medical device
directive (MDD), the product mentioned in this manual has also passed CE certification.
1.6 Safety Warnings
WARNING:
The power supply of this device may reach up to 100-240V. Maintenance in violation of the
requirements in this service manual may cause electric shock, serious injury, and even death.
Maintenance in violation of the requirements in this service manual may seriously damage
the device.
1.7 Acronyms and Abbreviations
ESD Electro-static discharge
PCBA Printed Circuit Board Assembly
ISO International Standardization Organization
CE Council of Europe
MDD Medical Device Directive
Important Information
Non-ionizing
Nurse caller
electromagnetic radiation
DISPOSAL: Do not
dispose of this product as
Refer to instruction unsorted municipal waste.
manual/booklet Separate collection of such
waste for special treatment
is necessary.
Atmospheric pressure
Temperature limit
limitation
CE Mark: conforms to
essential requirements of the
Class I Class-I equipment
Medical Device Directive
93/42/EEC
HOME button. Press this
button to access the setting ON/OFF button. Press this
interface or return to the button to power on/off the
infusion preparation pump.
interface.
If you have any question when using the infusion pump, please contact local
distributor or directly contact us at any time.
Address: 12th Floor, Baiwang Research Building, No.5158 Shahe West Road,
Xili, Nanshan, 518055 Shenzhen, Guangdong, PEOPLE'S REPUBLIC OF
CHINA
Fax: +86-755-26001651
Website: http://www.medcaptain.com/en/
E-mail: [email protected]
MEDCAPTAIN and its distributors all over the world have established after-sales service agencies
to effectively solve your problems in time.
System Overview
2 System Overview
This product is intended to be used in conjunction with the IV set to control the dose
of liquid infused into the patient’s body in clinical departments.
2.2 Composition
Main unit + handle + pole clamp + power cable + nurse caller (optional) + drop sensor
(optional)
2.3 Appearance
a) Front view
7 – Alarm indicator
System Overview
clamp block
pressure sensor
pressure sensor
c) Rear view
Accessories
Table 4-2 List of accessories
Accessory Description Part Number
Pole clamp
1202-00003-01
Handle
1404-00105-01
AC power cable European standard, 240V
1462-00004-01
AC 10A
3 Product Specifications
Model HP-60
3. IP23
Battery operating time is not smaller than 10 hours if the infusion rate
is 25 ml/h.
Key backlight
Interfaces Micro USB 2.0: enables connection to the external drop sensor.
USB 3.0: enables connection to a USB flash drive for software and
When infusion rate is not smaller than 1 ml/h: infusion accuracy ≤ ±3%
Dedicated IV set:
System Overview
Occlusion Level Patient side occlusion: 150 to 975 mmHg, 12 occlusion levels available
(default: level 6)
(default: level 5)
7 detection sensitivity levels: 25, 50, 100, 200, 300, 500, and 800 (ul)
Applicable IV Set Various brands of 20d/ml and 60d/ml IV sets conforming to relevant
standards
Infusion Modes Rate mode, time mode, weight mode, sequence mode, trapezia mode,
loading dose mode, drip mode, micro mode, and relay mode (used
Drug Library A maximum of 5000 drug types can be stored (default: 60 preset drug
types).
High-level alarms:
Infusion End, BAT Empty, Patient Side OCCL, Infusion End KVO
Start, KVO End, Relay Failed, Fluid Side OCCL, Air Bubble In Line,
No Drop, Too Many Drops, Too Few Drops, Drip Chamber Overflow,
Middle-level alarms:
Low-level alarms:
Door Failure
Special Functions Repeated alarming: After the sound of an alarm is muted, this alarm is
System Overview
playback.
barcode scanning.
power failure.
Text size adjustment: Three text size levels are available for selection.
Wireless Networking The wireless networking function enables the infusion pump to be
4 Hardware
Interface
Drop J3 J25 J11
board for
sensor
drop sensor
J13
Motor speed
Battery Speaker
detection board
Figure 4
Hardware Description
4.2 Power Board
a. Circuit description:
The AC/DC power board adopts a medical-grade switching power supply that features low leakage current
and high isolation voltage and meets the EMC requirements. The input AC voltage range of the AC/DC power
board is 100-240V, 50/60Hz. The output single DC power supply is 15.0V, and the maximum output current is
2A.
b. Main test point:
No. Test Name Range Unit Remarks
Point
1 TB2 DC output voltage 14.0-16.0 V /
c. Circuit board socket numbering and definition
The maximum voltage of the AC/DC power board may reach up to about 400 V. The workbench must be
insulated to avoid touch of the high-voltage circuit part. High voltage may cause electric shock and
personal injury.
When performing a test with a multimeter or oscilloscope, note that the test probe must not short-circuit the
live part on the circuit board to avoid damage of the circuit board.
19
Hardware Description
hospital number. The USB3.0 interface is used to connect to the nurse call system for implementing
the nurse call function.
(4) Audible alarm: The beeper and loudspeaker alarm circuit is used to implement the audible alarm
function.
(5) Motor drive control: The motor drive chip manufactured by TRINAMIC is used, which supports
24V/2A drive output. In addition, the stepper motor supports a subdivision of up to 256, ensuring
infusion accuracy and stability at low infusion rate. The SPI communication interface is used for
connection to M0. Motor start/stop, motor rotation speed, and motor acceleration/deceleration are
all controlled by SPI signals.
(6) Air bubble detection circuit: Air bubble drive and monitoring circuit is provided to monitor whether
air bubbles exist in the infusion tube.
(7) Fluid side/patient side occlusion pressure detection circuit: The fluid side/patient side occlusion
pressure detection board is connected to test the fluid side/patient side occlusion pressure.
(8) WiFi function: WiFi circuit and protocol are used for communication with the nurse station,
implementing output of infusion data.
(9) Wired network function: Wired network is used for communication with the nurse station,
implementing output of infusion data.
b. Main test point:
No. Test Name Range Unit Remarks
Point
1 TP15 24V DC output 22-26 V /
voltage
2 TP25 5V DC output voltage 4.9-5.1 V /
3 TP31 3.3V DC voltage 3.2-3.4 V /
a. Circuit description: Two optical coupling circuits are used to detect the motor rotation speed. The rotation
speed and rotation direction can be determined based on signal existence/absence and sequence of optical
couplers A and B.
b. Main test point: None
c. Circuit board socket numbering and definition
No. Socket PIN Name Description
Quantity
1 J13 6 Socket for connection to J12 /
of HP-60 main board
2 J2 3 Socket for connection to J2 /
of HP-60 pump finger
position detection board
Note: The sliding resistor is a very sensitive and precise device. When disassembling or assembling the resistor,
be careful not to touch peripheral components and not to pull the connection cable.
24
Sensor Information
5 Sensor Information
This chapter only describes the software UIs related to servicing and test. For the description of other
software functions, see the corresponding operation manual.
c) Pump finger signal: When the pump finger board optocoupler is shielded, the status displayed is 1.
Otherwise, the status displayed is 0. If the display status is incorrect, check the pump finger
optocoupler and related connection cable.
d) Air bubble detection: When no tube is installed or the tube is empty, a value in the range of 0-100 is
displayed. When a full tube is installed, a value larger than 1000 is displayed. Otherwise, check
whether the tube is installed in position and whether the air bubble sensor and detection board are
faulty.
e) Dedicated consumable: Check whether the consumable used is a dedicated consumable. If yes, the
status displayed is 1. Otherwise, the status displayed is 0.
f) AC test: Check whether a value in the range of 14V-16V is displayed when the pump is connected to
AC power supply and a value smaller than 2V is displayed when the pump is not connected to AC
power supply. If the displayed value is incorrect, the device generates the ERR007 alarm, which may
be caused by a power module failure.
25
Sensor Information
g) DC test: This value varies with the input DC input voltages. Connect the infusion pump to the DC
power supply and check whether the test value falls within the range of 14-16V. Then, disconnect the
pump from the DC power supply and check whether the test value is about 6V. If the voltage is
incorrect, the power module may be faulty.
h) Battery test: Install the battery in the infusion pump and check whether the test value is stable and falls
within the range of 9-13V. If the test value keeps changing, the battery may be damaged or the main
board is faulty.
i) VDD test: The normal value is 3.3±0.2V. If the test value is abnormal, the main board may be faulty.
j) VAA test: The normal value is 24±2V. If the test value is abnormal, the main board may be faulty.
k) VCC test: The normal value is 3.3±0.2V. If the test value is abnormal, the main board may be faulty.
26
Sensor Information
l) Fluid/patient side pressure test: Press the pressure sensor to check whether the pressure value changes.
If the pressure value does not change, the pressure sensor may be faulty.
m) Pump door Hall signal: Check whether 1 is displayed when the door is open and 0 is displayed when
the door is closed. If the displayed value is incorrect, check whether the electromagnet is properly
installed and whether the Hall switch functions properly.
n) Number of drops: Check whether the actual number of drops is consistent with that detected by the
drop sensor with naked eyes. If there is a large difference between them, check whether the drop
sensor is properly installed and whether the drop sensor is faulty.
27
Maintenance
6 Maintenance
6.1 Overview
To ensure long-term steady operation of the infusion pump, service personnel must check, maintain, and
test the pump at regular intervals. This chapter describes the basic test method of the infusion pump and
recommends appropriate test frequency and test tools. Service personnel must select appropriate test tools to
maintain and test the infusion pump based on actual requirements.
The test and inspection methods provided in this chapter are used to verify that the performance of the
infusion pump meets the specification requirements. During the test, if the test results do not meet requirements,
a certain functional module of the infusion pump is faulty. In this case, immediate maintenance or replacement
must be performed. In case of any other doubts, please contact the after-sales service department of Medcaptain
in time.
Caution
3. Switch on unit.
5. Remove battery.
28
Maintenance
When the battery always gives a low battery indication or the battery is not installed correctly, a
battery check is needed.
Use a screwdriver to loosen the battery cover fixing screws at the bottom of the pump.
CAUTION:
Remove the battery if the infusion pump is not likely to be used for some time
WARNING:
l The battery’s replacement must be done by specialist who has been trained to finish such operation.
Otherwise there will be a risk of danger.
l Please strictly follow the instruction to replace the battery, and the battery should be provided by the
manufacturer. Otherwise there will be a risk of danger.
Do not disassemble, short circuit or throw the battery into fire in case of the danger caused by linkage or
explosion.
29
Maintenance
30
Cleaning and Disinfection
7 Cleaning and Disinfection
It is highly recommended that the materials and methods listed in this chapter be used for
cleaning and disinfection of the device. If other materials or methods are used, the device may be
damaged or its service life may be shortened.
CAUTION:
In case of any doubts about the use of the detergent or disinfectant, please consult the local
distributor.
Please dispose of the wastes generated after the cleaning and disinfection according to the
relevant regulations of the local hospital.
7.1 Preparations
1. Before the cleaning and disinfection, disconnect the device from the patient.
2. Power off the device and disconnect the device from the AC or DC power supply.
3. Remove the infusion consumable and accessories (for example, drop sensor and barcode
scanner) connected to the pump.
4. Wear a pair of rubber gloves and a gauze mask to prevent contaminants from splashing
onto your skin during the cleaning and disinfection.
5. You are not allowed to disassemble this device for cleaning and disinfection. To
disassemble this device for further cleaning and disinfection, please contact the local distributor.
6. Prepare several pieces of soft medical gauze, a detergent container, and a disinfectant
container.
7.2 Cleaning
WARNING:
Do not immerse the device in the detergent solution.
Prevent the solution from seeping into the device.
Do not use organic solvent, halogenated solvent, petroleum-based solvent, glass detergent,
acetone, or other irritant detergents.
Only manual cleaning is allowed to be adopted for this device. Do not adopt the automatic
cleaning mode for this device.
Cleaning procedure:
31
Cleaning and Disinfection
1. Completely immerse a piece of soft medical gauze in neutral or slightly alkaline detergent
solution, wring out the gauze, and then use the gauze to wipe the device surface.
2. Wipe all the surfaces of the device in sequence until all the contaminants are removed
from the device surface.
3. Ensure that all the edges and corners of the device are completely cleaned.
4. After the cleaning, use a piece of dry medical gauze to remove the residual detergent
solution.
The following table lists the detergents recommended for the device.
Table 6-1 recommended detergents
7.3 Disinfection
WARNING:
Do not immerse the device in the disinfectant solution.
Prevent the solution from seeping into the device.
Use the disinfectant according to its operation manual.
Do not autoclave the device.
Only manual disinfection is allowed to be adopted for this device. Do not adopt the automatic
disinfection mode for this device.
Disinfection procedure:
1. Before the disinfection, clean the device according to the method provided in section 9.2.
2. Completely immerse a piece of soft medical gauze in the intermediate-efficiency or
high-efficiency disinfectant solution, wring out the gauze, and then use the gauze to wipe
the device surface.
3. Wipe all the surfaces of the device in sequence. For the contact time of the disinfectant, see
the operation manual of the disinfectant.
4. Ensure that all the edges and corners of the device are completely disinfected.
32
Cleaning and Disinfection
5. After the disinfection, immerse another piece of soft medical gauze in clean water, wring
out the gauze, and then use the gauze to wipe the device surface for removing the residual
disinfectant solution.
The following table lists the disinfectants recommended for the device and the required
contact time for the disinfection.
WARNING:
Do not dry the device by using a drying machine or similar products.
Connect the device to the power supply again after the device is completely dry.
1. After cleaning and disinfection, place the device in a shady, cool, and ventilated
environment for air drying.
2. If you are not going to use the device soon after air drying, place the device in its original
package for storage and transportation.
33
Calibration
8 Calibration
34
Calibration
(Schematic diagram)
35
Calibration
(Picture of real product)
36
Calibration
37
Calibration
150 225 300 375 450 525 600 675 750 825 900 975
3 Manually input the rates of the 18 points of a known IV set, fluid side occlusion pressure, and level 11
patient side occlusion pressure.
39
Calibration
2 Select a brand of consumable to be calibrated (if a desired consumable does not exist, execute method
1 to add this consumable).
40
Calibration
3 Tap Purge to remove the air bubbles and fill the entire infusion tube with liquid.
4 Tap to access the next page, and tap IV Set Cali. to access the IV set calibration page.
41
Calibration
42
Alarms and Troubleshooting
9 Alarms and Troubleshooting
Warnings:
During troubleshooting by service personnel, there is a 100-240V AC high voltage in the HP-60 and
the switching power supply circuit may generate a 400V DC high voltage. An accidental touch may
cause serious personal injury. Please follow the requirements in related regulations and ensure that
the maintenance is performed by professional qualified personnel.
When a device is used to measure a circuit board or other parts, short circuit or incorrect
measurement method may cause serious personal injury or serious device damage.
The device safety and functions must be verified after being disassembled for maintenance and then
assembled each time. For details, see chapter 7.
Basic tools for maintenance and inspection:
PC (used for downloading software and filling table)
Software downloading connector
Digital multimeter
20-50 ml measuring cylinder (accuracy: 0.2 ml) or electronic balance
Time keeping device like a stopwatch
Notes:
Among the aforementioned tools, the measuring cylinder or electronic balance is used to check the
infusion accuracy.
For the tools required for disassembling and assembling the device, see chapter 8.
Dedicated safety test devices must be used to test safety items after disassembling and then
assembling the device.
43
Alarms and Troubleshooting
46
Alarms and Troubleshooting
9.4 System Faults and Troubleshooting
System Faults and Troubleshooting
Alarm Alarm
Alarm Content Alarm Cause Troubleshooting
Message Level
1. Check whether the infusion tube is
twisted or extruded and whether the
retaining needle leaks.
2. Check whether the occlusion level
setting is reasonable (an older patient
indicates a higher drug liquid viscosity
and higher infusion rate, and the
occlusion level also needs to be
higher).
3. Calibrate the pressure value of the
Real-time pressure in the consumable.
Patient Side
Patient Side OCCL High tube ≥ occlusion alarm 4. Calibrate the pressure value of
OCCL
value configured the pressure sensor when it does
not touch the infusion tube, and
check that the KB value is correct.
(Generally, the K value is
404mV/N±50 and B value is about
500mV±200.)
5. Check the pressure sensor.
6. Check the drive head board, nut
displacement detection board, main
board, and related connection cable.
1. Check the optocoupler and related
1. The motor is locked. connection cable.
System error: motor
2. The motor rotation 2. Check the motor and related
drive locked
ERR 001 High speed is too low. connection cable.
(motor rotation
3. The motor rotation 3. Check the speed measurement board
speed is abnormal)
speed is too high. and related connection cable.
4. Check the main board.
System error:
reversed rotation of 1. Check the optocoupler and related
The motor rotation
motor drive connection cable.
ERR 002 High direction is opposite to the
(motor rotation 2. Check the speed measurement board
expected direction.
direction is and related connection cable.
abnormal)
System error: 1. Check the pressure sensor and
pressure sensor related connection cable.
The pressure sensor status
ERR 003 failure High 2. Check the drive head board and
is incorrect.
(pressure sensor is related connection cable.
faulty) 3. Check the main board.
1. Check the main board and related
Main chip voltage
The monitored voltage is connection cable (generally, check
3.3V
ERR 005 High smaller than 1.67V or whether the C71 voltage on the main
(CPU voltage is
greater than 2.25V. board falls within the range of
abnormal)
1.67-2.25V).
DC input voltage The monitored voltage is 1. Check the power board and related
ERR 006 (DC voltage is High smaller than 2.15V or connection cable.
abnormal) greater than 3.3V. 2. Check the main board.
AC input voltage The monitored voltage is 1. Check the power board and related
ERR 007 High
(AC power supply smaller than 2.12V or connection cable.
47
Alarms and Troubleshooting
is abnormal) greater than 2.87V.
48
Alarms and Troubleshooting
1. Check the fluid/patient side
Drive
occlusion pressure board and related
communication A communication error
connection cable.
ERR 013 error: M4 fails to High occurs between M0 and
2. Check the connection cable between
detect the data of M4.
the main board and pump door board.
M0.
3. Check the main board.
System error: fluid
1. Calibrate the pressure sensor.
side pressure sensor The AD value of the fluid
2. Check the pressure sensor and
ERR 009 failure High side pressure sensor
related connection cable.
(fluid side pressure exceeds the normal range.
3. Check the main control board.
sensor is faulty)
1. Check whether the drop sensor is
properly installed.
1. No drop sensor is 2. If no drop sensor is installed, check
Fluid side occlusion connected or the drop whether the drop detection function is
Fluid Side (mutual exclusion detection is disabled. disabled.
High
OCCL with the drop 2. Real-time occlusion 3. Calibrate the pressure sensor.
sensor) pressure ≥ preset occlusion 4. Check the pressure sensor and
alarm value related connection cable.
5. Check the main control board and
related connection cable.
1. Check whether air bubbles exist in
the tube.
2. Check the air bubble sensor and
Air bubbles/air The detected air bubble
Air Bubble In related connection cable.
bubble High value is greater than the
Line 3. Check the pump door board and
accumulation preset air bubble threshold.
related connection cable.
4. Check the main control board and
related connection cable.
1. Check whether dust or dirt exists on
When the second pump the pump finger optocoupler.
Pump finger failure finger reaches the highest 2. Check the pump finger detection
(peristalsis pump position, the pump finger board and related connection cable.
ERR 011 High
finger drive encoding disk does not 3. Check the motor speed measurement
failure) reach the specified board and related connection cable.
position. 4. Check the main control board and
related connection cable.
1. Check whether the drop detection
function is enabled while no drop
sensor is connected.
The drop detection 2. Check whether the drop sensor
No Drop
No Drop Sensor Middle function is enabled but no works properly.
Sensor
drop sensor is connected. 3. Check the drop sensor interface
board and related connection cable.
4. Check the main control board and
related connection cable.
1. Check whether the drop detection
1. The infusion tube is
function is enabled while no drop
inserted into the drop
sensor is connected.
sensor.
2. Check whether the drip chamber is
2. Drop detection is
No Drop No Drop High dripping liquid.
enabled.
3. Check whether the drop sensor is
3. The drop signal does not
properly installed.
generate any signal all the
4. Check whether the drop sensor is
time.
faulty.
49
Alarms and Troubleshooting
5. Check the drop sensor interface
board and related connection cable.
6. Check the main control board and
related connection cable.
Notes:
For the alarm message ERR 00X, X indicates a character. The specific alarm message may be ERR 001,
ERR 002, and etc.
For all problems related to the potentiometer, judge whether the hardware is faulty. Access the AD value on
the Maintenance screen, move the potentiometer, and observe whether the AD value changes to judge
whether the sensor and related wires are damaged.
Access the AD value displayed on the screen, press the pressure sensor, and observe whether the AD value
displayed changes to judge whether the pressure sensor and related wires are damaged.
50
Disassembly and Assembly
4 Battery 1
1. Use tweezers to remove the five lower shell hole plugs (2).
2. Use a Philips screwdriver to remove the four M3 screws (3).
3. Take down the face shell (1).
Confidential Information of MEDCAPTAIN MEDICAL TECHNOLOGY CO., LTD. 51 / 93
Disassembly and Assembly
4. Press the fastener on the battery connector to separate the plug and socket, and take out the
battery (4).
1. Remove the AC power supply connection cable, ground wire, and loudspeaker connection
cable.
2. Use a Phillips screwdriver to remove the two PT3*8 cross recessed pan head drilling PT thread
tapping screws (1).
3. Take down the rear panel socket assembly.
1.2 Dismantling the Loudspeaker and AC Socket
2 Loudspeaker 1
3 AC socket 1
8 Interface socket 1
1. Remove the five cross recessed pan head drilling PT thread tapping screws (4), and take
down the AC socket and loudspeaker.
2. Screw out the two cross recessed countersunk drilling tapping screws (7), and take down
the AC socket.
2 HP-60 AC shield 1
4 AC-DC module 1
4. Use a Phillips screwdriver to remove the two PT3*8 cross recessed pan head drilling PT
thread tapping screws (3).
5. Take out the AC-DC module (4).
1.4 Dismantling the Peristalsis Pump Motor Speed Measuring PCBA and Pump Finger Position
Detection PCBA
1. Remove the connection cable (7008000021) of the pump finger position detection board.
2. Screw out the cross recessed pan head, spring washer, and flat washer combination screw M2x6
GB 9074.4 (1), and take down the HP-60 pump finger position detection PCBA (2).
3. Remove the connection cable (7000000025) of the HP-60 motor speed measuring board.
4. Screw out the cross recessed pan head, spring washer, and flat washer combination screw M2x6
GB 9074.4 (4), and take down the HP-60 motor speed measuring PCBA (3).
1.5 Dismantling the Sliding Block and HP-60 Door Potentiometer Assembly
1. Turn the HP-60 door drive gear G3 (1) to make the HP-60 sliding block (2) slide to the
rightmost shown in the figure.
2. Screw out the HP-60 door drive push rod (3) and the two HP-60 sliding block set pin (4).
3. Take down the HP-60 sliding block.
4. Remove the connection cable plug on the HP-60 door potentiometer assembly.
5. Screw out the two PM2*3 cross recessed small pan head machine thread screws (6), and take
Confidential Information of MEDCAPTAIN MEDICAL TECHNOLOGY CO., LTD. 58 / 93
Disassembly and Assembly
1.6 Dismantling the HP-60 Tube Detection PCBA, Anti-Free-Flow Clamp PCBA, Patient Side
Occlusion Pressure Detection PCBA and its Connection FFC
1. Take down the HP-60 rear panel socket assembly and HP-60 AC shield.
2. Remove the HP-60 patient side occlusion signal connection cable (7008000019).
3. Screw out the two PT2.3*6 cross recessed pan head drilling PT thread tapping screws (2), and
take down the HP-60 tube detection PCBA (1).
4. Screw out the two PT2.3*6 cross recessed pan head drilling PT thread tapping screws (4), and
take down the HP-60 anti-free-flow clamp PCBA (3).
5. Remove the HP-60 patient side occlusion pressure detection board’s connection FFC (7).
6. Screw out the two PT2.3*6 cross recessed pan head drilling PT thread tapping screws (6), and
take down the HP-60 patient side occlusion pressure detection PCBA (5).
1.7 Dismantling the Peristalsis Pump
2 PM3*6 screw 4
1. Take down the HP-60 sliding block according to the method in section 8.6.
2. Take down the HP-60 common consumable middle plate (1).
3. Screw down the four PM3*6 screws (2).
4. Remove the HP-60 door potentiometer assembly (9121000008), HP-60 motor speed measuring
board’s connection cable (7000000025), and plug of the peristalsis pump motor’s connection
cable.
5. Take down the peristalsis pump assembly (3).
1.8 Dismantling the Peristalsis Pump Motor
9 Stepper motor 1
1. Take down the peristalsis pump assembly according to the method in section 8.8.
2. Remove the three cross recessed pan head, spring washer, and flat washer combination screws
M3x6 GB 9074.4 (1 and 5).
3. Remove the two HP-60 peristalsis pump springs (6), HP-60 pump adjustment nut (7), and HP-60
pump spring support (8).
4. Screw out the only PM2.5*6 cross recessed small pan head machine thread screw+spring
washer+flat washer assembly (4).
5. Take down the HP-60 speed measuring encoding disk (3).
6. Screw out the four PM2.5*10 cross recessed small pan head machine thread screw+spring
washer+flat washer assembly (2).
7. Take down the stepper motor (9).
1.9 Dismantling the Door Drive Assembly
1. Remove the two cross recessed pan head, spring washer, and flat washer combination screws
M2x6 GB 9074.4 (2) and the PT2.3*6 cross recessed pan head drilling PT thread tapping screw
(3).
2. Remove the motor connection cable, and take down the door drive assembly.
3. Remove the two PM2.5*6 cross recessed small pan head machine thread screw+spring
1. Remove the three PT2.3*6 cross recessed pan head drilling PT thread tapping screws (1).
2. Take down the HP-60 fluid side occlusion pressure detection PCBA (2).
1.11 Dismantling the Separate Air Bubble Sensor Module A
1. Disconnect the separate air bubble sensor module A and the main board connection plug.
2. Remove the two PT2.3*6 cross recessed pan head drilling PT thread tapping screws (2).
3. Take down the separate air bubble sensor module A (1).
1.12 Dismantling the HP-60 Pump Door PCBA
1. Use tweezers to remove the seven pump door hole plugs (1).
2. Remove the seven PT2.3*6 cross recessed pan head drilling PT thread tapping screws (2).
3. Take down the HP-60 cover plate (3).
4. Remove the seven PT2.3*6 cross recessed pan head drilling PT thread tapping screws (5).
5. Take out the HP-60 pump door PCBA (4).
1.13 Dismantling the Separate Air Bubble Sensor Module B
1. Screw out the four cross recessed pan head screws M2x4 GB 818 (1).
2. Take down the HP-60 patient side occlusion pressure detection board (2).
3. Take down the separate air bubble sensor module B (3).
1.14 Dismantling the HP-60 Veneer
1 HP-60 veneer 1
1. Use tweezers to strip off the veneer (1) from the seam between the veneer and shell.
1. Use tweezers to remove the HP-60 pump door hole plug (1).
2. Remove the PT2.3*6 cross recessed pan head drilling PT thread tapping screws (2).
3. Take down the HP-60 bottom shell FPC cover (4).
4. Disconnect one end of the HP-60 pump door board’s connection FFC (3).
No. Part Name Quantity
5. Take down the HP-60 cover plate (60Z0000026) according to the method in section 8.13.
6. Take down the HP-60 pump door board connection FFC (5).
1.16 Dismantling the HP-60 Main PCBA
2 PM3*6 screw 8
1. Remove the eight HP-60 lower shell hole plugs (1) and eight PM3*6 screws (2).
2. Take down the HP-60 face shell (3).
5 HP-60 AC shield 1
3. Take out the HP-60 middle plate assembly (10) and HP-60 pump door assembly (11).
4. Remove the two PT3*8 cross recessed pan head drilling PT thread tapping screws (6), and take
down the HP-60 AC shield (5).
5. Remove the two PT3*8 cross recessed pan head drilling PT thread tapping screws (8), and take
down the HP-60 rear panel socket assembly (7).
6. Remove the five cross recessed pan head, spring washer, and flat washer combination screws
M3x6 GB 9074.4 (9), and take down the HP-60 main PCBA (4).
1.17 Dismantling the HP-60 Drop Sensor Interface PCBA
1. Take down the HP-60 middle plate assembly (3) and HP-60 pump door assembly (4) according
to the method in section 8.17.
2. Remove the two PT2.3*6 cross recessed pan head drilling PT thread tapping screws (1).
3. Take down the HP-60 drop sensor interface PCBA (2).
1.18 Dismantling the Fastener Assembly
4 HP-60 fastener 1
1. Take down the shifting block plate (1). The shifting block plate is connected through the fastener.
Be careful not to break off the fastener when taking down the shifting block plate.
2. Remove the two PT2.3*6 cross recessed pan head drilling PT thread tapping screws.
3. Take down the fastener seat (3) together with the fastener (4) and shifting block plate pressure
spring (5).
1.19 Assembly
Perform assembly in the reversed order of the disassembly procedure. When performing assembly, note
the following points:
3. For the parts where lubricating grease has been applied but already erased, the operator must
4. When installing the main PCBA, do not forget to install the main board shielding case and main
5. When installing the main motor, pay attention to the direction of the outgoing line.
6. After installing the main motor and reduction gear assembly, manually make the gear rotate for
over one circle to ensure that the gear assembly rotates steadily and smoothly. Ensure that the
speed measuring encoding disk and speed measuring optocoupler do not rub each other.
7. When installing the pump door board, ensure that the surface of the display screen and transparent
8. All cables and wires must be connected correctly and firmly. The wire arrangement must be tidy,
and the wires must be bundled and fastened as requested
Note: The list of spare parts is updated as requested based on product changes and maintenance at regular
intervals.
11 Software Upgrade
83
Appendix
1) Connect the white end of the USB-to-RS232 connection cable to the USB port of the process
equipment PC.
Note: A drive needs to be installed at the first use. After drive installation, determine a
fixed USB port for connection of the USB-to-RS232 connection cable in the future. If you
change the port for connection of this line, you need to reinstall the drive.
2) Connect the black end of the USB-to-RS232 connection cable to the external USB port of the
HP-30/HP-60 rear panel assembly.
4) Restart the integrated machine or power on the board. The system displays a message
indicating that the connection is successful.
84
Appendix
7) Click start download. A download progress bar is displayed. It takes about 20 minutes for
programming of the “main program resource”.
85
Appendix
86
Appendix
2. Standard screwdriver with Philips head (Diameter Φ 3mm, All length is ok)
4. Tweezers
27
Appendix
88
Appendix
months
Hospital: Department: Contact person:
89
Appendix
□ Check that no crack or □Shell Leakage Current Test □Check air bubble sensor
damage exists 1Connect the 601 safety analyzer to the Check method: make a 3mm
AC power supply. air bubble in the tube, then set
Connect the applied part of the tested a speed to infusion; it would
device to the SUM end of the process be alarm when air bubble
equipment and the RA end of the safety passing the pump’s sensor.
analyzer.
Connect the tested device to the auxiliary
power output socket of the 601 safety
analyzer through a power cable.
Connect one end of the red test lead to the
red input terminal of the safety analyzer
and clamp the other end of the red test
lead onto the metal foil attached to the
shell of the tested device.
Plug in the 601 safety analyzer, and press
5-Enclosure leakage on the panel of the
safety analyzer to access the shell leakage
current test page.
Normally, the shell leakage current is not
greater than 100 μA. In case of a single
fault, the shell leakage current is not
greater than 500 μA.
27
Appendix
27
Appendix
92
MEDCAPTAIN MEDICAL TECHNOLOGY CO., LTD.
After-sales service address: 12th Floor, Baiwang Research Building, No.5158 Shahe West
Road, Xili, Nanshan, 518055 Shenzhen, Guangdong, PEOPLE'S REPUBLIC OF CHINA
Fax: +86-755-26001651
Postal code: 518055
Website: http://www.medcaptain.com
Email: [email protected]
©2017 MEDCAPTAIN MEDICAL TECHNOLOGY CO., LTD. All Rights Reserved.
27