®
HT1000
APPLICATION NOTE T2O-AN-P260
NITRATE (P260) METHOD
The following application note explains the procedure for the detection of Nitrate (P260) using the
HT1000 Photometer.
Equipment:
HT1000 Photometer
Nitrate test tube
Nitrate test powder
Nitrate Test tablet
10ml vial
Stirring rod
Nitrite LR tablet
Safety:
Consult the safety data sheet for all of the reagents before use. Even if you have used
HydroTest reagents before, the formulation may have changed.
Getting started:
Switch the unit on using the power key.
Select the Nitrate method by depressing the cursor keys until ‘Nitrate T ’ is displayed or use
the shortcut by pressing & holding shift followed by 260
Release shift and press enter.
Trace2o Ltd Rev1.0
Blank analysis:
Ensure that the 10ml vial is clean.
Fill the vial with 10ml of the water sample.
Fit the cap and tighten.
Place the vial in the sample chamber, making sure that the arrow marks on the instrument
and vial are aligned.
Press the zero key
Wait until the following is displayed
Zero accepted
Prepare Test
Press TEST
Sample preparation:
Fill the nitrate test tube with 20ml of the water sample.
Add 1 level spoon of Nitrate Test powder
Close the tube tightly with the cap and swirl vigorously for one minute
Add one Nitrate Test tablet straight from the foil to the water sample.
Close the tube tightly with the cap and swirl vigorously for one minute
Stand the tube upright, allow contents to settle, then gently invert three to four times.
Allow to stand for a further 2 minutes, then open and wipe carefully around the rim to
remove solid particles.
Decant 10ml of the treated solution into the 10ml vial.
Add one Nitrite LR tablet straight from the foil to the water sample.
Crush the tablet using a clean stirring rod, until no large pieces are visible.
Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
Analysis :
Place the vial in the sample chamber, making sure that the marks on the instrument and vial
are aligned.
Press the Test key
Wait for a reaction period of 10 minutes
After the reaction period, the measurement starts and the result is displayed in mg/l Nitrate
LOD/Tolerance
The Lower LOD is 0.08 mg/l (80ppb), upper LOD is 1 mg/l (1000ppb).
Tolerance: ± 0.005mg/l.
Trace2o Ltd Rev1.0
Notes
For best results, rinse vials thoroughly between with deionised water. Ensure that the
outside of the vials are clean, dry and free from fingerprints. Always handle vials by the lid
where possible.
The tablets should be added direct from the foil avoiding contact with hands or surfaces.
Nitrite present in the sample will also react, leading to a higher result. To correct, carry out a
nitrite test (P270) on the same sample and subtract the result.
Nitrate concentrations above 1 mg/L can be diluted up to 100x, with the result multiplied up
to compensate.
Interferences may occur from presence of the following ions: Antimony (III), Iron (III), Lead,
Mercury (I), Silver, Chloroplatinate, Metavanadate, Bismuth.
Copper (II) ions may artificially lower test results.
Trace2o Ltd Rev1.0
Trace2o Ltd Rev1.0