0% found this document useful (0 votes)
340 views2 pages

Portable Patient Monitor Guide

The document provides technical specifications for the M1000 patient monitor including: - Monitor dimensions of 220mm x 180 mm x 90mm and weight of 1.6kg. It has an 8" color TFT LCD touchscreen display. - It can monitor and display parameters such as ECG (3/5-lead), NIBP, SpO2, temperature, heart rate, and respiration. - Sensors and modules are listed along with their measurement ranges, resolutions, accuracies, and other technical details. - The monitor has data storage capabilities and rechargeable lithium-ion batteries providing up to 10.5 hours of use. - Connectivity options and environmental requirements

Uploaded by

Xinwen Chen
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
340 views2 pages

Portable Patient Monitor Guide

The document provides technical specifications for the M1000 patient monitor including: - Monitor dimensions of 220mm x 180 mm x 90mm and weight of 1.6kg. It has an 8" color TFT LCD touchscreen display. - It can monitor and display parameters such as ECG (3/5-lead), NIBP, SpO2, temperature, heart rate, and respiration. - Sensors and modules are listed along with their measurement ranges, resolutions, accuracies, and other technical details. - The monitor has data storage capabilities and rechargeable lithium-ion batteries providing up to 10.5 hours of use. - Connectivity options and environmental requirements

Uploaded by

Xinwen Chen
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

Technical Specifications

NIBP
M1000 Method : Automatic Oscillometric
Monitor size : 220mm x 180 mm x 90mm Operation mode : Manual, Auto, STAT
Weight : 1.6kg, Parameters : Systolic, Diastolic, Mean
Standard parameters configuration, Systolic range Adu : 30 to 270 mmHg
including lithium battery Ped : 30 to 235 mmHg
Neo : 30 to 135 mmHg
Display Diastolic range Adu : 10 to 220 mmHg
Type : 8” color TFT LCD, Ped : 10 to 220 mmHg
with touchscreen Neo : 10 to 110 mmHg
Resolution : 800 x 600 pixels Mean range : Adu: 20 to 235 mmHg
Waveforms : up to 10 Ped : 20 to 225 mmHg
Neo : 20 to 125 mmHg
ECG Accuracy : Static: ±3mmHg or ±2%, whichever is greater
Lead set : 3-lead:I, II, III Clinical : average error:±5 mmHg
5-lead : I, II, III, aVR, aVL, aVF, V Max standard deviation : 8mmHg
Automatic 3/5 – lead recognition Resolution : 1 mmHg
Intervals for auto measurement : 1,2,3,4,5,10,15,20,30,45,60,90 minutes
Gain : x0.25, x0.5, x1, x2, x4, Auto
2,4,8 hours
Sweep speed : 6.25 mm/s, 12.5 mm/s,
Overpressure Protection : Hardware and software double protections
25 mm/s, 50 mm/s
Adult : 297±3 mmHg
Bandwidth : Diagnostic Mode: 0.05-120Hz Child : 252±3 mmHg
Monitor Mode : 0.5-40Hz Neonatal : 147±3 mmHg
Surgical Mode : 1-25Hz
Defibrillator protection : Withstand 4000V Temperature
Recovery time : <10s Channel : 1-ch
CMRR : Diagnostic Mode: >89dB Parameters : T1
Monitor, Surgical : >105dB Range : 0 to 50°C (32 to 122 °F)
ST analysis : Range:-2.0 to 2.0 mV Resolution : 0.1°C
Accuracy : ±0.02 mV or ±10 %, Accuracy : ±0.2°C (without probe, 45.1°C to 50°C)
±0.1°C (without probe, 25.0°C to 45°C)
whichever is greater (-0.8 to +0.8 mV)
±0.2°C (without probe, 0°C to 24.9°C)
Over ±0.8mV : unspecified
Resolution : 0.01mV Data Storage
Arr analysis : Yes, multi-lead, 27 classifications Trend data : 240 hrs (interval 1 min), 8 hrs (interval 5 sec)
QT analysis : Yes Alarm events : 2300 events and associated waveforms
Arr. events : 500 Arr. events and associated waveforms
Heart Rate NIBP : 2000 measurements
Range :
Adu : 10 to 300 bpm Battery
Ped/Neo : 10 to 350 bpm Type : 1 Build-in chargeable Lithium-ion battery
Resolution : 1 bpm Voltage : 11.1 VDC
Accuracy : ±1 bpm or ±1%, whichever is greater Capacity : 2200mAh (3400mAh optional)
Run time : 6 hours (2500mAh), 10.5hours (3400mAh optional)
Respiration Recharge time : 2200mAh: 3 hours maximum (power off )
Range : 0 to 150 rpm 3400mAh : 4 hours maximum (power off )
Resolution : 1 rpm
Accuracy : ±2 rpm or ±2%, whichever is greater Interfacing
Lead : I(RA-LA) or II(RA-LL) Connectors : 1 AC power connector
1 RJ45 network connector / or nurse call connector
Sweep speed : 6.25 mm/s, 12.5 mm/s, 25 mm/s
2 USB 2.0 connector
WiFi support : Yes, 5G/2.4G dual band
SpO2
Barcode Scanner : Support
Range : 0 to 100%
USB printer : Support
Resolution : 1%
Accuracy : ±2% (70-100%, Adu/Ped)
Unspecified (0-69%)
Environmental Requirements
Temperature : Operating: 5 to 40°C
Modular Patient Monitor
Storage : -20 to 55°C
Pulse Rate Humidity : Operating: <85% (non condensing)
Range :
Accuracy :
25 to 250 bpm
±1bpm or ±1%,
whichever is greater
Storage :
Barometric :
< 93% (non condensing)
Operating:700~1060hPa
M1000
Storage : 500~1060hPa
Resolution : 1 bpm
Compact Design
Easy to carry. Less than 1.6Kg weight, thecompact M1000 is extremely portable.
With built-in concealed handles. M1000 fits perfectly in one hand
leaving your other hand free to operate the touch screen.

High Durability
Extended battery life supports up to 10.5 hours of continued monitoring
Intensively strong plastic housing resists aging and affords more disinfectants.
IPX1 water-proof,
Compact system design and high quality components, low failure rate & long use life.

Easy to Use
High resolution 8”TFT LCD and optional touchscreen
Biolight's new generation user interface.
Easy connectivity to Biolight M6000C central monitoring system via wired and wireless
Various mounting solutions helps you for easier and more flexible work in the clinical environment.

Advanced Performance
Comprehensive measurements: 3 / 5 lead ECG, NIBP, SpO2, TEMP
Enhanced high-performance algorithms & Stronger data storage,
Latest technology is also implemented for modern clinical applications,
such as 2.4G / 5G dual band WiFi.

External power module is more durable & stable, Low power Provide a variety of clinical installation solutions, support trolley,
consumption, no internal fan design, make the clinical wall mount, bed rail hooks, rail clamp, meet all kind requirements
environment cleaner & quieter. of various clinical installation

Powerful data storage240 hours trend data (interval 1 min) Easy-to-use interface and optionaltouchscreen makes the operation
8 hours trend data (interval 5 sec)500 Arr. events and moresmoothly. And considerable ergonomicdesign of 12° screen
associated waveforms2300 Alarm events and associated angle gets you thebest perspective.
waveforms2000 NIBP measurements

New BM1000 platform, Highly integrated design, combined


motherboard & multi-parameter module into one board,CPU
Can connect Biolight's “Patient Data Review “software, manager
Speed increase 600%,RAM increase 400%, Optimized circuit
& analysis patient's data on your computer, Can be very
design, reduce energy consumption, One battery run time is
convenient to produce the relevant clinical reports
more than 10.5Hours.Platform integrated advanced sound
card, support MP3 format alarm sound