0 ratings0% found this document useful (0 votes) 446 views41 pagesSMZ 800 Manual
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content,
claim it here.
Available Formats
Download as PDF or read online on Scribd
M248 E wsve2
Nikon
Stereoscopic Zoom Microscope
$MZ1000 / SMZ800
InstructionsThank you for purchasing the Nikon product
This instruction manual is written for the users of Nikon
Stereoscopic Zoom Microscopes.
To ensure correct usage, read this manual carefully before
ing the instrument.
© Itis prohibited to reproduce or transmit this manual in part or
whole without Nikon’s expressed permission.
© The contents of this manual are subject to change without notice.
© Although every effort has been made to ensure the accuracy of
this manual, if you note any points that are unclear or incorrect,
contact your nearest Nikon representative.
© Some of the products described in this manual may not be
included in the set you have purchased.
-—— Warning/Caution Symbols Used in This Manual
Though Nikon products are designed to provide you with the
utmost safety during use, incorrect usage or disregard of the
instructions may cause personal injury or property damage. For
your own safety, read the instruction manual carefully and
thoroughly before using the product. Do not discard this manual.
Always keep it near the product for easy reference.
Inside this instruction manual, safety instructions are indicated
with the symbols shown below. Be sure to follow the instructions
marked with these symbols for your safety.
Symbol Meaning
A WARNING —_ Distegarding instructions marked with this
symbol may lead to death or serious injury.
Disregarding instructions marked with this
Lorton symbol may lead to injury or property
damage.1. Intended use of this product
This product is intended only for microscopy. Do not
use it for any other purpose.
Do not disassemble
Disassembly may result in damage to the instrument.
Never disassemble any part except as described in this
operation manual. Contact your Nikon representative
you notice any malfunction of this instrument.
Check the input voltage
‘When using an illuminator, check that the input voltage
displayed on the power supply of the illuminator
matches the operating voltage. Contact your Nikon
representative if the displayed voltage does not match the
operating voltage. Use of an improperly matched
illuminator may result in damage to equipment.
Power cord
Always turn the power switch off and unplug the power
cord when replacing the lamp of an illuminator as failure
to do so may result in electric shock or equipment
damage.
Heat from the light source
When using an illuminator, do not place cloth or paper or
highly flammable materials, such as gasoline, benzene,
thinner or alcohol, near the lamp as there is a danger of
fire.
Always use lamps that are of the specified rating only.
For the lamp rating, please refer to the operation
manual.
Always turn the power switch off and unplug the
power cord when replacing the lamp of a illuminator
as failure to do so may result in electric shock or
equipment damage.L\cavtion
1. Caution when replacing lamps
The lamp and the surrounding area become very hot
during and immediately after using the illuminator. Be
careful not to burn yourself. In addition, when replacing
a lamp be sure the lamp has cooled sufficiently.
. Check the light source
Always use lamps that are of the specified rating only.
Use of a lamp having a different specification may result
in damage to equipment. For the lamp rating, please
refer to the operation manual1. Installation location
a
ee
a
Note the following points when installing the stereoscopic
microscope.
@ Install the microscope in a location with the temperature between
0° and 40°C, and humidity of less than 80%. If installed in a hot
and humid location, mold may form on the len or condensation
may occur inside, resulting in reduced performance or damage to
the microscope.
Do not install the microscope in a location subject to direct
sunlight.
Install the microscope in a location that is not subject to vibration.
Install the microscope in a location that is free from dust and dirt.
Handle the microscope carefully "9
The stereoscopic microscope is a precision optical instrument.
Handle it carefully and do not subject it to impact. Shock from
impact during transportation or operation as well as forcible
operations may cause damage to the instrument.
Cleaning the lenses (O00
Do not let dust, fingerprints, etc. to get on the lenses. Dirt on the
lenses will adversely affect the view of the image. If any lenses
become dirty, clean them as described below.
Use an air blower to blow dust away. If this does not suffice, brush
away the dust with a soft brush or gently wipe it away with a piece
of gauze.
© Only if the lenses become dirty with fingerprints or grease stains,
slightly dampen a piece of soft, clean cotton cloth with absolute
alcohol (ethyl alcohol or methyl alcohol) and gently wipe away the
dirt. Do not use the same part of the cloth more than once.
Since absolute alcohol is highly flammable, be careful in handling
it so that it does not ignite.
Observe the manufacturer’s handling instructions when handling
absolute alcohol.Be sure to always follow these guidelines. y
a
a
_|
Use of silicon cloth is recommended when cleaning painted parts,
plastic parts and printed parts. If such a part becomes excessively
dirty, wipe it gently with gauze dampened in a mild detergent
solution. Do not use organic solvents (such as alcohol, ether, or paint
thinner) as this may result in deformation of the part or remove of
lettering.
Store the microscope in a location with low humidity where mold is
unlikely to form. Do not store the microscope in a location subject
to direct sunlight or high temperature and humidity. During storage,
place a plastic cover over the equipment to prevent dust
accumulation.
|. LY
Regular
Regular inspections are recommended in order to maintain peak
performance. Please consult your Nikon representative for details
about regular inspections.:
Hl contents
Warning/Caution Symbols Used in This Manual....................
Be sure to always follow these guidelines. .......
Sy WARNING .
Qs CAUTION .
El Nomenclature ..
Hy Assembly ...
I usage
1 Prep:
1. Adjust the torque of the focus knob.
2. Adjust the interpupillary distance.
3. Adjust the diopter. .....
2 Focusing ...
1. Check the working dis
2. Focus on the sample.
3 Zoom.
1. Change the zooming magnification.
4 If You Cannot Focus On The Sample Though The Zooming
Body Is At The Highest Position.
Using Accessories ... car oa com oa 19
1. Reticle:
2. P-BERG tilted eyepiece tube
3. P-IER Eyelevel Riser
4 Plan 1X ERGO Objective
5 SM-S4L 4 X 4 Stage L....
Mluminator ...
1 Halogen Illuminators....
1. Lamp..
2. Attaching G-l LS. to the stand Ic. “PS &C PSC.
3. Arm for halogen illuminators
4. Power supplies ........5. Lighting area adjustment [G-LS]
6. Attaching filters [C-DSLS
2 P-ICI2 Coaxial Episcopic Illuminator.
Photomicrographic and TV-related Accessories ....
1. Photomicrographic magnification .....
2. Attaching the beam splitter...
3. Photo reticle:
4. Photomicrography procedure .
Table 1: Total Magnification and Real Field...
Table 2: Observable Sample Heights [mm] ........ ee seoeeeee BB,The figures on p.8 & 9 show a SMZ1000 zooming body, with P-BT
binocular eyepiece tube, C-W 10X eyepieces, Plan Apo 1X objective,
C-FMC focusing mount, and C-PS160 plain stand.
Diopter ring Eyepiece sleeve
Refer to “Adjust the \
diopter” on p.16. \ P-BT Binocular
eyepiece tube
“ P-BTL low eye-level
eyepiace tube and
P-BERG tited
Eyepieces \
———
410X, 15X, 20X and
30X eyepieces are \
available, ‘eyepiece tube are
oe also available.
clanglecra _ Zooming knob
: Changes the
Zooming body ‘magnification of the
clamp screw
sample image. Refer
to “Zoom” on p.17.
Coarse focus knob
Zooming body Refer to "Focusi
Outside diameter ofthe on p.16.
part to be rrounted the ——
focusing mount: 676
Fixes the zooming body
to the focusing mount.
Reter to “Focusing”
Objective ne
Plan 1X, Adhro 0.5X,
Plan 1X ERGO, Plan 4 Stand
‘Apo 0.5X, Plan Apo 1X,
ED Plan 1.8 and ED
Plan 2X objactives are
available. HR Plan Apo
series objectives for
SMZ1500 can be used
also, but the magnif-
cation in this case
becomes 1.25 times the
magnificaticn indicated
‘on the objective itselt.
Refer to Table 1 on p.36.
C-PS160 Stand base
Various stands are
available.
Screw for mounting
accessories
Holds the 1/4 lambda
plate and the like.
‘The screw on the
Groove for mounting
accessories
Holds the F uorescent /
ring illuminetor, Fiber /
optic ring illuminator / inside diameter:
andthe like, Outside / M55 x 0.75
diameter: 660 Stage plate
‘Stage plate clamp °
‘screw The black or the
white side can be
Fixes the stage plate chosen according to
to the stand base. the sample.Zooming knob clicks ON/OFF
screw (inside)
Refer to “ON/OFF of the Zooming
Knob Clicks” on p.17.
Clamp tool
/' Used for the stage plate clamp
/ | screw, eyepiece tube clamp
/ ‘screw, eto. Same size as the
/ ‘small hexagonal wrench
|| supplied with the zooming body.
Stored in the top of the pillar.
Pillar
Torque adjustment ring
(innermost ring)
‘Adjusts the torque of the
focus knob. (Refer to
“Adjust the focus knob
torque” on p.15.)
C-FMC Focusing mount
Screw hole for mounting
an Epi-arm for halogen
illuminator
Located on both sides
of the focusing mount.
Holds the G-EIA
flexible arm or Epi-arm.
Focusing mount
\ clamp screw
Fixes the focusing mount
to the pillar
Screw hole for
‘mounting accessories
Located on both sides
of the stand base. M5
Pillar clamp screw
\ (inside)The illustrations on p.10 & 11 show a SMZ800 zooming body, with P-BT
binocular eyepiece tube, C-W 10X eyepieces, Plan 1X objective, and
C-PS plain focusing stand.
Diopter ring Eyepiece sleeve
Refer to “Adust the
diopter on p.16. P-BT Binocular
eyepiece tube
~~ P-BTL low eye-level
eyepiece tube and
Eyepiece
410X, 15X, 20X and \
30X eyepieces are
available. P-BERG tilted
' eyepiece tube are
Conmecene also available
Zooming knob
Zooming body Changes the
clamp so magnification of the
Fixes the zooming sample image.
body to the arm, Reter to "Zoom" on
paz
Zooming body
Outside diameter of the Focus knob
part to be mounted on Reter to “Focusing”
the arm: 078 canta
Obj )
Plan 1X, Achro 0.5X, Plan
1X ERGO, Plan Apo 0.5X,
Plan Apo 1X, ED Plan
4.5X and ED Plan 2X
objectives a’e available,
HR Plan Apo series
‘objectives fer SMZ1500
‘can be used also, but the
magnification in this case
becomes 1.26 times the
magnification indicated on
the objective itselt. Refer
to Table 1 01 p.36.
Groove for mounting
accessories
Holds the Fluorescent
ring illuminator, Fiber
optic ring illuminator and,
the like.
Outside diameter: 660
Arm
\
S. Arm fixing screw
Holds the arm. The
position of the arm
can be chosen from
\ either high or low.
\ Vertical si
Hole for clip
Holds the clip for
retaining the sample.
C-PS Focusing
stand
Various stands
available.
Stage plate
‘The black or the
white side can be
chosen according to
the sample.
Screw for mounting
accessories
Holds the 1/4 lambda
plate and tha like.
‘The screw cn the inside
diameter: MBS x 0.75Nomenclature fj
Zooming knob clicks ON/OFF screw
(inside)
Refer to “ON/OFF of the Zooming
Knob Clicks" on p.17.
Hexagonal wrench
hanger
/, Holds the supplied
hexagonal wrenches.
Hexagonal wrench
(Large)
Supplied with the
focusing stand.
Used for changing
the height of the arm.
‘Screw hole for mounting an
Epi-arm for halogen illuminator
Screw holes are located on
both sides of the arm.
Holds the G-EIA flexible
arm or Epi-arm,
Hexagonal wrench
ob )\ Grail)
Supplied with the
zooming body. Used
for the eyepiece tube
clamp screw and
zooming knob clicks
ON/OFF screw.
‘Screw hole for mounting
6V-10W halogen illuminator
Holds the G-LS 6V-10W
halogen illuminator.
\ Pillarpu
Assembly _
Se 4
E& + An equipment model name in the bold-faced brackets [ ] indicates
that the following description is exclusively for that model.
[C-PS160]
__ Attach the pillar to the stand.
"Insert the pillar straight into the stand with the
__ groove of the pillar located down as far as the
__ pillar reaches to the bottom. Clamp the pillar oF
“using the clamp tool.
Place the stand on the level surface.
a Mount the stage plate.
_ [C-PS160]
Fit the stage plate into the stand base. Tighten
| the stage plate camp screw using the clamp
| tool to fix the stage plate.
| [C-PS & C-PSC]
___ Fitthe stage plate into the stand base while
| pushing it against the rim in the direction
|. shown by the arrow in the illustration.
[C-PS160]
© Fix the C-FMC focusing mount to the
__ pillar.
|. Face the focusing mount to the front and
| tighten the focusing mount clamp screw.
| Note) When fixing the focusing mount to the
ao upper part of the pillar, securely
| tighten the clamp screw so as not to
co fall the focusing mount."Use the hexagonal wrench (large - supplied
- Lightly tighten the zooming body clamp
[C-PS & C-PSC]
The stand arm can be lowered. (If you Groove
do not need to lower the arm, jump to
step
with the focusing stand) to loosen the arm
fixing screw. Reattach the arm using the
screw hole at the lower side of the vertical
slider. The arm will become 55 mm lower
than its original position. Before tightening
the screw, make sure that the two pins on the
arm fit in the grooves on the vertical slider.
Mount the zooming body.
screw to hold the zooming body on the
focusing mount or arm.
Note) Do not over-tighten the zooming body
clamp screw since it may cause
damage to the instrument,
Screw the objective into the bottom of
the zooming body.
Mount the eyepiece tube.
Align the groove on the eyepiece tube with
the positioning pin on the zooming body, push
the eyepiece tube toward the back to attach.
Tighten it to the zooming body using the
supplied hexagonal wrench (small) or the
clamp tool supplied with the C-PS160.Assembly 11]
a Insert the eyepieces into the eyepiece sleeves.
___ Be sure that it is inserted all the way until it touches the end of the sleeve.
=
| Note) When inserting the eyepiece, hold the rubber cover of the eyepiece.
| Holding the diopter ring causes a failure. Especially, be careful
a with the 10X eyepieces because their rubber covers obstruct the
ae view of the sleeve end.
S)
Efi] The hexagonal wrenches and clamp tool can be stored in the
at pillar of the stand.
| Refer to the figures on p.9 and p.11.a
Usage
= oo
1 Preparations For Observation
Adjust the torque of the focus knob.
___ Adjust the torque of the focus knob so as not to fall down the zooming
|. body on its own weight.
ee
—
| [c-FMc] Increasing the torque.
1
Turn the torque adjustment ring using the
_ clamp tool. Turning the ring in the direction
_. . ,
- of the arrow indicated on the side surface of
Eo the focusing mount will increase the torque.
/
_
© Note) * Do not twist the left and right focus
| knobs.
eo * Do not turn the knob beyond its
oo turning limit. :
- * Do not turn the fine focus knob while
@ holding the coarse focus knob or vice
versa. Otherwise, the damage to the
| instrument will result. Increasing the torque.
|
'
| [C-PS and C-PSC]
eA Rotating the left and right focus knobs in the
| opposite direction as shown in the illustration
| will increase the torque.
(To reduce the torque,
turn the knob in the direction
‘opposite to the arrow.)
a Adjust the interpupillary distance. >»
__ This adjustment should be performed every
"time the observer is changed since the O O
| interpupillary distance differs between (SF A
©. individuals. Adjust the interpupillary distance P\e
| __ so that the view field for each eye is merged =
| into one. Move while holding each sleeve cc
“— with both hands.=]
|
Adjust the diopter.
This adjustment should be performed every time the observer is changed
|___ since the eyesight differs between individuals.
1
2
Turn the diopter rings on both eyepieces to set them at the 0 position
(match the 0 line with the index line).
Turn the zooming knob to the highest magnification. Focus on the
sample using the focus knob. (Refer to “Focusing”.)
Turn the zooming knob to the lowest magnification. Peering
through the left eyepiece with the left eye, focus on the sample using
the diopter ring on the left eyepiece. Then, peer through the right
eyepiece with your right eye and focus on the sample using the
diopter ring on the right eyepiece.
Repeat steps 2 and 3 until the image is kept focused even though the
zooming magnification is changed. This adjustment ensures sharp
image throughout the zooming range.
2 Focus ng
Fa Check the working distance.
The distance between the focus plane and the bottom of the objective is
_
called the working distance. Focusing is easier if beforehand the distance
between the specimen and the bottom of the objective is s
t to the working
distance. The working distance differs depending on the objective being
used, Please refer to Table I on p.36.
Focus on the sample.
Turning the focus knob moves the zooming body up and down and brings
focus on the sample.
[c-FMC]
Using the fine focus knob will permit the delicate focusing. Especially, it
is effective for the high magnification observation.Usage [il]
3 Zoom
Change the zooming magnification.
“Turning the zooming knobs on the left and right side of the zooming body
"will change the magnification of the sample image.
| Mi Total Magnification
| __ The zooming knob has the indication of the zooming magnification.
The total magnification can be found by multiplying the magnification
___ of the objective times that of the eyepieces.
| Note) When the HR Plan Apo series objective for the SMZ1500 is used,
te the total magnification will be obtained by multiplying by 1.25.
7
| :
____ MON/OFF of the Zooming Knob Clicks
| The clicks on the zooming knob can be
eliminated.
| J 1 Insert the supplied hexagonal wrench
2 (small) or the clamp tool supplied \e.
with the C-PS160 into the screw hole “>
D4 on the rear of the zooming hody.
|] 2 Turing the inside serew in the LS
@ direction opposite to the arrow
- eliminates the clicks (OFF). Turning
: ;
| the same screw in the direction of the
'
ables arrow puts on back the clicks (ON).4
Usage jm]
If You Cannot Focus On The Sample Though
The Zooming Body Is At The Highest Position.
When you use the 0.5X objective or observe a tall sample, you may not be
able to focus on the sample though turning the focus knob to raise the
zooming body to its highest position.
[C-PS160]
Exchange the pillar for the C-EP extension pillar available on order.
Attach the C-EP extension pillar to the stand in the same manner as for the
standard pillar, (Hold the side having the groove downward. Refer to
item 1. on p.12.) Put the stopper ring on the pillar and then the focusing
mount. Tighten the focusing mount clamp screw.
Note) Contact the bottom surface of the focusing mount with the stopper
ring and securely tighten each clamp screw.
Extension pillar
Correct Wrong
IC-PS & C-PSC]
Use the C-ER auxiliary adapter available as
an option. (This adapter can’t be used for the auxiiary
SMZ1000.) adapter
- (AC
ES + Refer to the Table 2 on p.38 for the sample heights which may be
viewed.1 Reticles
Your reticles may be attached to the eyepiece.
Once remove the field ring (or lens room of
the 20X and 30X) from the eyepiece. Attach
the reticle with its pattern surface facing down
to the eyepiece and reattach the field ring (or
lens room).
BS + Refer to the Table | on p.36 for the size
of the reticle which may be attached.
Note) The 20X and 30X eyepieces adopt the
internal focusing system, and the
primary image has the magnification.
Use the reticle with 1.3X pattern for
20X eyepiece, and with 1.4X pattern
for 30X eyepiece.
A reticle with cross hairs, scaled cross
hairs, or photographic mask is
available on order for 10X eyepiece.
2 Eyepiece Tube
P-BTL low eye-level eyepiece tube
| When using a beam splitter or coaxial
|__ episcopic illuminator, it is possible to lower
| the eye level (eye level during observation) by
___ using this eyepiece tube in place of the
| binocular eyepiece tube.
| The installation procedure for this eyepiece
___ tube is the same as for the P-BT binocular
“eyepiece tube.
So
a,
Rete |
SS
Pattern |
surface,
Field ring
=—
Separately
sold reticle3
=
y
ZX carton Be careful not to put your
P-BERG tilted eyepiece tube
This eyepiece tube allows the angle of the
eyepiece sleeve to be continuously adjusted as
desired by the observer.
It is also possible to change the eye level by
rotating the interpupillary distance adjuster
(shown in gray in the illustration) 18
The installation procedure for this eyepiece
tube is the same as for the P-BT binocular
eyepiece tube,
fingers between the
interpupillary distance
adjuster and the lower part
of the eyepiece tube when
adjusting the angle of the
sleeve.
P-IER Eyelevel Riser
When the eye level is too low, it is possible to
raise the eye level by 25 mm by attaching a
P-IER Eyelevel riser between the zooming
body and the eyepiece tube. Itis possibleto LOC TT
use up to three Eyelevel risers at once. N
Avoid using the four or more Eyelevel risers 7
at once or using the other intermediate tube
together such as a beam splitter or coaxial =o
episcopic illuminator. It may cause the
vignetting around the image periphery.Using Accessories
4 Plan 1X ERGO Objective
Using the Plan 1X ERGO objective allows
you to continuously change the eye level
through a 40 mm range while maintaining a
constant working distance (50 mm).
1 Rotate the upper knurled ring on the
Plan 1X ERGO objective to attach it to
the zooming body.
2. Rotate the focus knobs on the focusing
stand to raise or lower the microscope to Plan 1X ERGO\,
set to the height best suited for viewing.
3 Rotate the lower knurled ring to focus
on the specimen. Use the microscope as
usual after setting the eye level.
Note) When attaching to the SMZ1000, vignetting will occur in the lateral
direction of the viewfield at the high eye level with 1X or lower
zoom magnification.Using Accessories
5 SM-S4L 4 X 4 Stage L
SM-S4L 4 X 4 stage L may be attached by 4X 4 stage Stage
using the C-4SA stage adapter. In thiscase, ——S=«S:C clamp screw
use the C-ER auxiliary adapter or C-EP
extension pillar because the position of the
sample will become higher. (Refer to “4. If
you cannot focus on the sample though the
zooming body is at the highest position.”.)
4 Remove the stage plate from the stand
base, and attach the C-4SA stage
adapter in place using the fixing screws Stage adapter
supplied with the adapter. a fixing screw \
2 Sufficiently loosen the stage clamp /stage adapter
screw of the SM-S4L 4 X 4 stage L. Fit
and mount the stage onto the stage
adapter with the side equipped with the
stage clamp screw facing front. Tighten
the stage clamp screw.1 Halogen Illuminators
There are two illuminators available: the G-LS of 6V 10W and the C-
DSLS of 6V 20W.
Note) These illuminators are not adequate to observe the sample having a
high reflective surface. Use the P-ICI2 Coaxial Episcopic
Mluminator on p.29 for this purpose.
G-LS
(6V10W)
Lamp house Lamp socket Plug
(may be moved
back and forth)
a)
-_
Mounting screw
C-DSLS
(6V20W)
Lamp house Plug
Lamp socket
Holder and Epi arm
mounting screw hole
Filter cap
983 mm fitter
attachableCAUTION The lamp and the surrounding area become very hot
during and immediately after using the illuminator.
Be careful not to burn yourself. In addition, be sure
the lamp has cooled sufficiently before replacing a
lamp.
Pull the lamp socket from the lamp house and G-LS.
insert the lamp straight into the socket as far
as it will go. Only use halogen lamps of the
specified rating.
Next, attach the lamp socket to the lamp
house.
| Socket
Lamp
(GE 787 6v-10W)
C-DSLS
Confirmation
window Lamp
/, Philips JCRIM
Via)
|
/
{
ight
shielding tube
Holder mounting |
Socket screw hole | | Short protrusion
Note) Do not touch the glass part of a lamp with your bare hands as your
skin may burn and stick to the lamp, possibly reducing its level of
light output or causing damage to the lamp. When replacing aMluminator
ae
7S
:
a
ae
Cee
lamp, handle it through a piece of cloth or through its packaging
until replacement is complete. If you get fingerprints on the lamp,
wipe gently with a piece of clean cloth dampened with alcohol.
Attaching G-LS to the stand [C-PS & C-PSC]
Set so that the protrusion on the lamp house
aligns with the inside of the screw hole on the
arm of the stand (screw hole for mounting
6V10W halogen illuminator) and secure it in
place from the outside of the arm using the
mounting screw.
Mounting
screw
Screw hole for
mounting 6V10W
halogen illuminator
Arm for halogen illuminators
It is possible to freely adjust the direction of illumination by using a
special arm.
1. G-EIA flexible arm [G-LS] GLs
Set so that the protrusion on the lamp 5 cD
house aligns with the inside of the hole cB Protru-
in the holder of the flexible arm and SES. sion
secure it in place from the outside of the m~ Arm,
holder using the mounting screw.
Mounting screw
Flexible arm
Secure by screwing the mounting screw
supplied with the arm into the arm
mounting screw hole in the stand (screw
hole for mounting an Epi-arm).
Mounting screw \,i
Aebbiol:
Hae
ee
a
2 Epiarm
Slightly loosen the clamp knob on the Clamp knob_
arm, Screw in the mounting screw at Simutancousy /
the end of the Epi arm into the screw ‘armsijoints. //
hole of the stand (screw hole for
mounting an Epi-arm) while turning the
joint. Next, securely tighten the nut
using the supplied wrench.
Epi arm may be attached also to the
screw hole for mounting accessories of
C-PS160.
Spanner
Screw the mounting screw at the other
end of the Epi arm into the screw hole
on the lamp house in the same manner
as described above. Set the lamp house
in a position that makes for easy
viewing and tighten the clamp knob.Mluminator
Power supplies
|____ Use the Power Supply XN for regions using 100/120 V power, and the
___ Power Supply TN-PSE30 for regions using 230 V power. Insert the plug
___ of the Halogen illuminator into the output connector located on the rear of
the power supply.
Input voltage display
‘Check that the voltage displayed matches the
voltage used in your area. If the voltages do
‘not match, do not use the power supply and
‘contact your Nikon representative
Power switch
Brightness adjuster
Brightness adjuster
Changes the illumination
brightness when rotated
oe
TO 4 |
Power switch
Power cord
Power cord
Fuse
‘Always use the fuse
specified.
Output connector
Connect the
lamphouse cable here.
(Rear view)
‘Changes the illumination brightness when rotated,
Power Supply XN (only for regions with 100/120 V power)
Input voltage display
Check that the voltage displayed
matches the voltage used in your area.
Ifthe voltages do nat match, do not use
the power supply and contact your
Nikon representative,
?
Output
connector
Connect the
lamphouse cable here.
AC input connector
Connect the specified power
‘cord here,
Power Supply TN-PSE30 (only for regions with 230 V power)EEETLELETEERETEEREIEEEEIEEEEIEIEE
LL
———
2
lll
Power Supply XN (only for regions with 100/120 V power)
Input rating
Power supply for 100V regions only:
100 V AC, 50/60 Hz, 30 W
Power supply for 120V regions onl
120 V AC, 50/60 Hz, 30 W
Output rating
3-6 V AC, 3.3A
Fuse rating
250V, 1A
Power Supply TN-PSE30 (only for regions with 230 V power)
Input rating
230V AC, 50/60Hz, 0.3A
Voltage fluctuation | +10%
Output rating AC 6V SA
Internal fuse Fast acting type, FIA/250V, 5.2 x 20 mm
Operating Altitude: 2000 m max.
environment
Temperature: 0° - 40°C
(indoor use only)
Relative humidity: 85% max.
(no condensation)
Pollution level: Degree 2
Installation: Category 2
Storage Temperature: _-20° - 60°C
environment Relative humidity: 90% max.
(no condensation)
Protection class
Class |
Power cord.
Use only the following power supply cord.
Using the wrong power cord could result in
danger or fire. The protection Class 1 equipment
should be connected to PE (protective earth)
terminal
© For 220-240V AC area
Approved according to EU/EN standards, 3
conductor grounding Type HOSVV-F, 3 m
long maximum, rated at 250V AC minimum.Mluminator
ighting area adjustment [G-LS]
It is possible to adjust the lighting area by moving the lamp socket back
and forth. Adjust the lighting area to allow easy viewing.
ue
—
a
Attaching filters [C-DSLS]
‘
___ Rotate and remove the filter cap at the end of
oo the lamp house and insert a filter. Use filters
___. that have a diameter of 33 mm and a width of \
ep Filter
— 15~2mm. Nessim
Filter cap
Light guide mount
Light guide clamp screw
Main body Light guide Fiber illuminator
Fiber iluminator (C-FI115 for
Eyepiece tube 100/120V region and C-FI230
clamp screw for 230V region) available on
frder is necessary for the
light source.
lead also the Instruction
‘Manual of tne Finer
luminator C-FI1 18/C-Fi230.
Luminous section
Incident section _/
Magnification of the P-ICI2 Coaxial opera a |incom
Episcopic Iuminator is 1.5X. Total aso se 3X or higher
magnification is obtained by multiplying 1.5. [Pian 1x ISX or higher
by the total magnification of the basic set of | Plan XERGO | 1.SX or higher
i Plan ApoO.SX | 4X or higher
the microscope. Plan Apo 1X 2X orhigher
D Plan 1.5% 2X orhi
Note) Usable magnification for the P-ICI2, [eB Tian 13% we
Coaxial Episcopic Mluminator is HR Plan Apo OSX | 25K or higher
restricted. Use the magnification HR Plan Apo IX _| 2.5X orhigher
within the range shown in the right HR Plan Apo L.6X_|_Notusableeen
-
she
»
table. If the magnification out of the
specified range is used, uneven
illumination will occur or the seeing of
the sample image will be degraded.
Mount the P-ICI2 onto the zooming body
aligning the groove on the bottom of the
P-ICI2 main body with the positioning pin on
the zooming body while tilting the P-ICI2
toward the rear. Tighten the eyepiece tube
clamp screw.
Note) When the other intermediate tube such
as a beam splitter is used together,
attach it on the P-ICI2.
/
/ Positioning pin
Screw in the 1/4 lambda plate to the end of
the objective. When the C-FIR ring fiber light
guide is used together, use the other 1/4
lambda plate separately available for ring
fiber illuminator.
Luminous
section
Ef fer the incident seetion of the light guide IO
=]
|
|
; |
|
, |
co
into the light guide mount of the fiber
illuminator and tighten the clamp screw. t a
Insert the luminous section of the light guide =
into the light guide mount of the P-ICI2 and
tighten the clamp screw.Mluminator
_—
The optical axis of the 1/4 lambda plate is at
the 45° position from the red dot. Brightness
__ is the maximum when the red dot is
_ positioned at the front and switches from
bright to dark by each 45° position. Adjust
the plate to the position where the brightness
and contrast are best suited for the
observation.
Insert the filter P-ICI33 NCB 14 into the fiber
illuminator to improve the color
reproducibility when taking a
photomicrogaph. Refer to the instruction
manual of Fiber Illuminator.
Bright
Dark Dark
Bright Bright
Dark Dark
BrightLLL
It is possible to photograph images or display them on a TV monitor by
attaching a beam splitter between the zooming body and the eyepiece
tube.
There are two beam splitters available: the S P-IBSS with one port and the
D P-IBSD with two ports.
Read also the instruction manual of the instrument such as a
photomicrographic equipment or TV camera to be used.
Photomicrographic
equipment FX-III series
Photomicrographic
equipment adapter __ Photo reticle
C-mount TV adapter 7
Eyepiece
Binocular
eyepiece tube
Port
Optical path Optical path
switching knob switching knob
Beam splitter D Beam splitter SPhotomicrographic and TV-related accessories
1
eo
nee
coger ER a
uo a
te
Photomicrographic magnification
Photomicrographic magnification = (PLI projection lens magnification)
x (Zoom magnification) x (Objective magnification)
When the P-ICI2 Coaxial Episcopic Illuminator is used:
Photomicrographic magnification = (PLI projection lens magnification)
x (Zoom magnification) x (Objective magnification) x 1.5
* When the HR Plan Apo series objective is used:
Photomicrographie magnification = (PLI projection lens magnification)
x (Zoom magnification) x (Objective magnification) x 1.25
Attaching the beam splitter
Mount the beam splitter onto the zooming
body aligning the groove on the bottom of the
beam splitter with the positioning pin on the
zooming body while tilting the beam splitter +
toward the rear. Tighten the eyepiece tube
clamp screw.
[P-IBSS] Mount it so as to position the port
at either the left or right.
P-BSS Port location
[P-IBSD] 1) Mount it so as to position the Side port
rear port at the rear, :
2) Attach the side port to the
beam splitter main body with
the positioning pin on the side
port fitted into the groove on
the beam splitter main body.
Tighten the side port clamp
screw.
P-IBSD Attaching side portPhoto reticles
Attaching the C-FP photo reticle to the C-W10X eyepiece permits you to
make focusing for phtotomicrography through the eyepiece. Select the
correct frame indicating the area being photographed according to the
magnification of the PLI projection lens. For more accurate focusing and
photographing area, use the finder of the photomicrographic equipment.
2X projection lens,
35-mm full size
2.5X projection lens,
35-mm half size
/— 2.5X projection lens,
36-mm full size
oat 5X projection lens,
e 36-mm full size
4X projection lens,
35-mm full size
Photomicrography procedure
1 Switching the optical path
[P-IBSS] Set the optical path switching knob to PHOTO/BINO.
100% 50% — 50%
PHOTO/BINO
[P-IBSD] Set the optical path switching knob to the left or right.
Setting it to the left allows you to use only the rear port
and setting it to the right, both the rear and side ports.
100%
100% 50%
100%Photomicrographic and TV-related accessories
2 Focusing
The following is the focusing procedure using the photo reticle. For
the focusing procedure using the finder of the photomicrographic
equipment, refer to the instruction manual of the phtotomicrographic
equipment.
1 Once remove the field ring of the &
eyepiece and attach the photo mask oo
reticle. (Refer to p.19 "1. Reticles".) cS
2 = Turn the diopter ring to focus on the
reticle. _ eA GY
3. Turn the focus knob to focus on both
the reticle and the image. “wy Y
Note) [P-IBSD] ‘Adjust so that the double
You can not observe the image crosshairs is clearly
through the left eyepiece if the divided in two
optical path switching knob is set
to the left. In this case, use the
finder of the photomicrographic
equipment for focusing.
Focusing with the right eyepiece
or focusing with the optical path
switching knob set to the position
other than left may result in
defocusing.
3 Adjusting the frame position
Turn the eyepiece to align each frame of the photo mask and the
photomicrographic equipment finder. Note that the image seen
through the finder is reversed from that seen through the eyepiece.
Finder and eyepiece images
Eyepiece Finderfeey pue uoneoubey [eIOL =, a1GeL
ro-rr | xosr~sr | xo~se | xoce~ce | or~oor | xore~te | ri~ser | Xoo~ ve X91 ody Wid YH
Lo~k X 008 ~ 0 X 002 ~ 07 O1~O9I | XOSI~SI TI~OTZ | XOO1~ OF +s XI ody weld YH
ri-pl | Xost~st x 001 ~ ve-oze | xse~se | rr~orb | xos~s oe X50 ody uid YH
ro-rr | xosr~sr | xo-se | xoze~ze | o1~ xor~tz | PI~ ser | XO9T~9F st Xt wid da
90~8S | XOK~9e | OT~FOL | XOFe~tz X08I~ si~est | Xoz~z str Xs1 wed da
60-88 | XOPT~ Fz 91-9ST | XOT~ O7%~OOT | XOT~TI BT~ SLT | XOB~8 OL Xt ody weld
ST~SLt | XOzI~ZI re-ete | xos~s or-oor | x09-9 ss-0ss | XOr~F ezl xso ody unig
60-88 | xOr~tz J oT~oSt | Xom~9 | oz~ooe | Xoz~zI | 8t~sez | XO8~8 os Joowa x1 wid
60-88 | xor~# | 91~9ST | XODI~ o-oo | xocl~zi | ee~suz | XOR~8 8 x! wid
ST-Sut | xoz~ re~ere | x08~8 or-oor | x09~9 ss~oss | Xor~r si xso ony
[ww | uoneoyubew | fw) | uoneoyuBew | wu] | uoneoyuBew | fw]
reow | elo. rou | elo, roy | eIOL
Xp’ uoneowuBeW xe"| uoneoyuBeW
zigssiewerp eoney | eidserowerp ajney | e1dserewerp epney | gzdserewerp ejoney | [wu] senyoelgo
Z soquinu piel $21 sequinu plol4 91 sequinu piel zz sequinu piel —_jeoueisip|
Xoe x oz XSi Xb Sunion
seoeideh3
000LZWSd Real Field ij
tion ani
:Total Magnificat
Table 1
YO~Se | XOBE~OO9}O1~€9 | XOTSZ~OOF| E~O8 | XOSSI~OOE XO9zI~O0Z| +7 X 9°] ody weld YH
6O-9S | XEVT~SLE]OT~ OO! | X SUSI ~O'se | OT~STI | XIBIT~s'1 | Ke~ HLT | X8'BL~STT rs XT ody weld YH
SITU | XVSII~s'st | TE~OO | XB8L~STI | Tr-9S7 | XToS~ro | oS~TSE | XPEE-~EO 9E1 X £0 ody wold MH
vo-se | xose~oo9}o1~e9 | xozsz~oor}e1~o8 | xoesl~ooe xo91~o0c | Sze Xt wud a
Lo~Lr | xsesc~osr | er~es | xoest~ooE | LI~Lor | xsttI~s7z] ec~eFT | XsPE~OST | Str Xs wid ag
1-OL | XO68I~OE | OT~STI | XOSZI~OOT| SzT~O9I | XSFH~O'ST | SE~OT | XOEI~OOT OL XI ody wed
TT~OFI | XSPO~OSI | OF~OST | XOEI~OOI | I'S~OTE | XELr~SL XSle~ 0s ez xso ody ued
ri-OL | XoesI~OoE |oz~stI | XOSI~OOT| ¢z~O9T | XSr6~O'ST xoeo~oor | os joowa xT wid
ri~OL | xowst~O0E |oz~stI | XOSI~oOT] sz-O9T | XSr6-o'ST | sr~oz Xoe9-oOr | 8 xt wa
tt~orl | xsr6-ost for~ose | xoe9-oor | rs~oze |xeur~se |oL-orr |xsie~os | ost xso ony
{ww} | uoneoyubew | [ww] | uoneoyuBew| [wu] | uoneoyuBew | [wu] | voneowuBew
reow | rei rou | elon | proyreow | eI IO.
Xp" uoneowuBeyy xe"} uoneoyuBeW
Ziosjowelp soney | 6Lhsejowerpejoney | eLdso}eweIpejoney | szbsoleweIp ejoney | [wu] sonyoelgo
2 soquinu piel $21 Joquinu plol4 91 soquinu piel zz soquinu piel —_jeoveisip|
X0e xX 0z xst XOL Buppon|
seoaidea
008ZWSTable 2: Observable Sample Heights | |
Table 2: Observable Sample Heights [mm]
SMZ1000 + C-PS160 + C-FMC
Mount Mount up When using ae
Objectives normal side down CEP
position position | extension pillar a
Acro 0.5X O~ 24 =— O~ 124 oe
Plan 1X = 0~42 0~ 197 O~ 86
Plan 1X ERGO O~114 0~59 O~214 0~ 103
Plan Apo 0.5X ge | =— O~ 125 O~ 14
Plan Apo 1X. O~ 102 0~47 0~ 202 oes.
ED Plan 1.5X O~111 0~56 0-211 0~ 100
ED Plan 2X O~ 108 0~53 0 ~ 208 O~ 97
HR Plan Apo 0.5X_ | — 0~ 133 a
HR Plan Apo 1X O~ 103 0~48 0~ 203 O~ 92
HR Plan Apo 1.6X, O~112 0~57 O~ 212 O~ 101
SMZ800 + C-PS
When using
onietwes | Nochon | “shan |autay adapter, SAESAL ane
Acro 0.5X = — O~ 58 -
Plan 1X 0~75 0~20 26 ~ 131 0~20
Plan 1X ERGO 0~92 0~37 3 ~ 148 0~37
Plan Apo 0.5X. ose - oe -
Plan Apo 1X O~81 0~ 26 32 ~ 137 0~25
ED Plan |.5X 0~89 0~ 34 40 ~ 145 0~ 34
ED Plan 2X 0~87 0~32 38 ~ 143 O~31
HR Plan Apo 0.5X_ O-11 _ O~ 67 _—
HR Plan Apo 1X 0~ 82 0~27 33 ~ 138 0~27
HR Plan Apo 1.6X, 0~90 0~35 41 ~ 146 0~35NIKON CORPORATION “wat att =
23 Monneuchi3chome, Ctiods Tye 100-831, Japan|
FIO SORTER 2 EEN)
NIKON INSTECH CO.,LTD. /#tatt [> 4YAT ID
Parts id 8, Hg ch Kaas Kanab Kanagawa 710005 pan
et 20-005 nr NAEENIRET 8 BE NL— EHC SR
AUPE WINTER Ra ESE)
SEREaR GU ZERBIBADAIO- GOMRRBLED
exiaze
ARRHI—7 (XR OD. BEL RD
2 S000 AMURNRETRIRG-I- A= 1S
Raat 0681 RERERRMONE/ OHO
RRRIAMI LR GSH). OLN, RII A, RHE, ES, ER]
a AIDS ARENSON 1
NIKON INSTRUMENTS INC.
200 Wot tan Ro Medal NY 117473054, USA
Tab 1-91.547 8900 Fax +1691 470306
NIKON INSTRUMENTS EUROPE B.V.
‘Scene $2, PO. Box222,11TOAE Bachooodop. The Netarnts
Te aia 200-2161
iE ou) 725-2160
aE tt) east
HE 5D 95-0165
‘RHF ce) cont sen)
EE (5) 78-1170
EE HD enti
Fm vei 29-2182
Fax Ot) 225-2100,
Fax nt) 16-008
Fx 52) 954-068,