057 Anti-HBc-IFU-V3.06-en-others
057 Anti-HBc-IFU-V3.06-en-others
and reporting.
Fax:+86-755-28292740
SUMMARY AND EXPLANATION OF THE TEST
HBV is transmitted through infected body fluids, including blood,
semen, and vaginal fluids (including menstrual blood). It also can
be transmitted from a pregnant woman to her child at or near the
FOR PROFESSIONAL USE ONLY
Store at 2-8°C time of delivery.
Hepatitis B surface antigen (HBsAg) is one of the most frequently
performed tests for HBV. This HBV antigen is the earliest indicator
ef
CAUTION: COMPLETELY READ THE of an active hepatitis B infection. This antigen may be present
INSTRUCTIONS BEFORE PROCEEDING
before symptoms of an HBV infection are present. If this antigen
level remains high for more than 6 months, then people will
probably become a carrier of HBV, meaning people can transmit it
to others throughout people’s life.
er
USE BY present during an active HBV infection. This test determines how
contagious people are. Testing for this antigen can also be used to
monitor the effectiveness of treatment for HBV. Hepatitis B
TEMPERATURE LIMITATION
( STORE AT 2-8 °C) e-antibody (HBeAb) shows that the active stage of the HBV
infection is almost over and people’s risk of being contagious is
greatly reduced. HBeAb is usually present during chronic HBV
e
KIT COMPONENTS
Material Supplies
057 Anti-HBc-IFU-V3.06-en--others 1/4
This document could not be used for the purpose of legal registration.
Please only refer to the current product lot insert enclosed with the kits package for execution and reporting.
2) 2-Point Recalibration
Component 100 tests 50 tests
Via the measurement of calibrators, the predefined master curve is
Magnetic Microbeads: TRIS buffer,
adjusted (recalibrated) to a new, instrument-specific measurement
0.09%NaN3, coated with sheep 2.5 ml 2.0 ml level with each calibration.
anti-FITC polyclonal antibody.
Calibrator Low: phosphate buffer 3) Frequency of Recalibration
containing bovine serum and 2.5 ml 2.0 ml After each exchange of lot (Reagent Integral or Starter
(recommended).
antibody labeled with ABEI, 6.5 ml 4.5 ml
containing BSA, 0.09%NaN3. SPECIMEN COLLECTION AND PREPARATION
All reagents are provided ready-to-use. Sample material: serum
Collect 5.0ml venous blood into Blood Collection Tube. Separate
Reagent Vials in kit box serum by centrifugation after standing whole blood at room
Internal Quality Control: phosphate buffer temperature.
ef
containing bovine serum and anti-HBc, Avoid repeated freezing and thawing. The serum sample can be
2.0 ml
0.09% NaN3. (For target value, refer to frozen and thawed for only two times. Stored samples should be
Quality Control Information data sheet) thoroughly mixed prior to use (Vortex mixer).
Internal quality control is only applicable with MAGLUMI system. .Please ask local representative of SNIBE for more details if
For instructions for use and target value, refer to Quality Control people have any doubt.
er
Information data sheet. User needs to judge results with their own
standards and knowledge. Specimen Conditions
Do not use specimens with the following conditions:
Biomedical Engineering Co., Ltd (SNIBE) or our representative. applicator stick prior to analysis. Use a new applicator stick for
ce
THIS WAY UP System. Refer to the SNIBE service for a more detailed
discussion of onboard sample storage constraints.
CALIBRATION AND TRACEABILITY from the serum separator, red blood cells or clot. Specimens
1) Traceability removed from the separator gel, cells or clot may be stored up
To perform an accurate calibration, we have provided the test to 12 hours at 2-8°C.
calibrators standardized against the WHO 1st International Specimens can be stored up to 30 days frozen at -20° C or
Standard 95/522. colder.
This document could not be used for the purpose of legal registration.
Please only refer to the current product lot insert enclosed with the kits package for execution and reporting.
15 min Incubation
Shipping 400 μl Wash cycle
Before shipping specimens, it is recommended that 3s Measurement
specimens be removed from the serum separator, red blood
cells or clot. When shipped, specimens must be packaged
DILUTION
and labeled in compliance with applicable state, federal and
Samples with concentrations above the measuring range can be
international regulations covering the transport of clinical
diluted. After manual dilution, multiply the result by the dilution
specimens and infectious substances. Specimens must be
factor. After dilution by the analyzers, the analyzer software
Fo
shipped frozen (dry ice).
automatically takes the dilution into account when calculating the
WARNING AND PRECAUTIONS FOR USERS
sample concentration.
The automatic sample dilution is available after dilution settings
For use in IN-VITRO diagnostic procedures only. are done in the MAGLUMI analyzer user software. Please follow
Package insert instructions must be carefully followed. MALGUMI analyzer operating instructions.
Reliability of assay results cannot be guaranteed if there are
rR
any deviations from the instructions in this package insert. QUALITY CONTROL
Observe quality control guidelines for medical laboratories.
Safety Precautions Use suitable controls for in-house quality control. Controls
CAUTION: This product requires the handling of human should be run at least once every 24 hours (a run cannot
specimens. exceed 24 hours), once per reagent kit and after every
All samples, biological reagents and materials used in the calibration. The control intervals should be adapted to each
assay must be considered potentially able to transmit laboratory’s individual requirements. Values obtained should
ef
infectious agents. They should therefore be disposed of in fall within the defined ranges. Each laboratory should
accordance with the prevailing regulations and guidelines of establish guidelines for corrective measures to be taken if
the agencies holding jurisdiction over the laboratory, and the values fall outside the range.
regulations of each country. Disposable materials must be
er
incinerated; liquid waste must be decontaminated with sodium LIMITATIONS OF THE PROCEDURE
hypochlorite at a final concentration of 5% for at least half an 1) Limitations
hour. Any materials to be reused must be autoclaved using an A skillful operation and strict adherence to the instructions are
overkill approach. A minimum of one hour at 121°C is usually necessary to obtain reliable results.
considered adequate, though the users must check the Procedural directions must be followed exactly and careful technique
en
effectiveness of their decontamination cycle by initially must be used to obtain valid results. Any modification of the
validating it and routinely using biological indicators. procedure is likely to alter the results.
It is recommended that all human sourced materials be
Bacterial contamination or repeated freeze-thaw cycles may affect
considered potentially infectious and handled in accordance the test results.
with the 29 CFR. 1910.1030 Occupational exposure to
bloodborne pathogens. Biosafety Level 2 or other appropriate 2) Interfering Substances
ce
pay attention the silicon film still exists on the surface of the kit.
(CLIA) analyzer.
For detailed discussion of handling precautions during system
TEST PROCEDURE
interpreted as follows:
To ensure proper test performance, strictly adhere to the operating
Non-reactive: A result less than 100 index/ml (< 100 index/ml)
instructions of MAGLUMI Fully-auto chemiluminescence
is considered to be negative.
immunoassay (CLIA) analyzer. Each test parameter is identified
Reactive: A result greater than or equal to 100 index/ml is
via a RFID tag on the Reagent Integral. For further information
y
This document could not be used for the purpose of legal registration.
Please only refer to the current product lot insert enclosed with the kits package for execution and reporting.
in the same run, and 30 times for each levels to calculate the
coefficient of variation.
Inter-assay precision
Control Mean(index/ml) SD(index/ml) CV%
Level 1 40.02 2.81 7.03
Level 2 145.32 9.50 6.54
Level 3 395.12 24.62 6.23
rR
2) Analytical Sensitivity
<6 index/ml.
The detection limit represents the lowest analyte level that can be
distinguished from zero.
3) Specificity
ef
4) Recovery
Consider Calibrator High of known concentration as a sample,
dilute it by 1:2 ratio with diluents, and measure the diluted
concentration for 10 times. Then calculate the expected
concentration and recovery of measured concentration. The
recovery should be within 90% -110%.
ce
REFERENCES
1. Maddrey WC. Hepatitis B: an important public health issue. J
Med Virol 2000;61:362-366.
2. Bozkaya H, Akatca US, Ayola B, et al. High degree of
Us
with the severity of liver injury and mutations in the core region.
Am J Gastroenterol 2000;95:2894-2904.
7. Knodell RG, Ishak KG, Black WC, et al. Formulation and
application of a numerical scoring system for assessing
y
This document could not be used for the purpose of legal registration.