Ultrasonic Testing Procedure
Ultrasonic Testing Procedure
ULTRASONIC TESTING
PROCEDURE
CONTENTS
COVER PAGE 1
CONTENTS 2
1 AIM / SCOPE 3
2 DEFINITIONS 3
3 REFERENCES 3
4 SAFETY 3
5 PROCEDURE REQUIREMENTS 3
6 PERSONNEL QUALIFICATION 4
7 EQUIPMENT AND ACCESSORIES 4
8 EQUIPMENT CALIBRATION 6
9 PRE - TEST REQUIREMENTS 10
10 TEST REQUIREMENTS 10
11 METHOD FOR SIZING INDICATIONS 12
12 METHOD FOR DISCRIMINATING GEOMETRIC
FROM FLAW INDICATIONS 12
13 EVALUATION (ACCEPTANCE CRITERIA) 12
14 RECORDS 13
APPENDIX
B ACCEPTANCE CRITERIA 18
ATTACHMENT
UT REPORT FORM
This procedure describes the requirements for Ultrasonic Testing on weldments of plates and
pipes (Except TKY joints and category D Nozzle welds), to ASME Boiler and Pressure Vessel
Code, Section V, Non-destructive Examination, and similar codes.
The typical weldments that are covered with this procedure is limited to a minimum thickness of 10
mm and a minimum diameter of 10” NPS.
2.0 DEFINITIONS
UT Ultrasonic Testing
FSH Full Screen Height
DAC Distance Amplitude Correction
Client with whom Stork-Cooperheat has a direct contractual agreement
Company End User / Owner
3.0 REFERENCES
4.0 SAFETY
4.1 Stork-Cooperheat safety requirements and project specific safety instructions shall be met at
all times.
4.2 The area shall be well ventilated while carrying out the test specifically in confined areas.
Procedure Qualification – When procedure qualification is specified by the referencing code section
a change of a requirement in Table 1 identified as essential variable from the specified value, or
range of values, shall require re-qualification of the written procedure by demonstration.
Table – 1
6.1 Personnel performing ultrasonic testing shall be qualified in accordance with Stork-
Cooperheat procedure as referred in item 3.0 herein.
6.2 Personnel having other qualification such as PCN may use this procedure with specific
written approval of the Owner representative.
6.3 Personnel involved in ultrasonic testing shall be qualified to minimum NDT UT Level II.
7.1 Equipment
a) The ultrasonic flaw detector shall be of the pulse-echo type having A-scan presentation,
capable of generating frequencies over the range of at least 1 to 5 MHz with a valid
calibration certificate.
The ultrasonic instrument shall provide linear vertical presentation within ± 5% of the
full screen height for 20% to 80% of the calibrated screen height (base line to
maximum calibrated screen points).
Method
Position an angle beam on a calibration block so that indications from the ½ and ¾ T
holes gives a 2:1 ratio of amplitudes between the two indications. Adjust the gain so
that the larger indication is set at 80% of full screen height (FSH). Without moving the
search unit adjusts gain to successively set the larger indication from 100% to 20% of
full screen height in 2 dB increments and read the smaller indication at each setting.
The reading shall be 50% of the larger amplitude within 5% of FSH.
Screen height linearity check shall be performed at intervals not to exceed 3 months
for analog type instruments and one year for digital type instruments or prior to first
use thereafter.
The ultrasonic instrument shall utilise an amplitude control accurate over its useful
range to ± 20% of the nominal amplitude ratio, to allow measurement of indications
beyond the linear range of the vertical display on the screen.
The amplitude control linearity shall be performed at intervals not to exceed 3 months,
for analog type instruments and one year for digital type instruments or prior to first
use thereafter.
Method
Position an angle beam search unit on a basic calibration block so that the indication
from the ½ T side drilled hole is peaked on the screen. When adjusting the gain, the
indications shall fall within the specified limits below:
e) The proper functioning of the equipment shall be checked and the equipment shall be
calibrated by the use of the calibration standard as a minimum, at the following
stages:
If during any check it is determined that the testing equipment is not functioning
properly, the entire product that has been tested since the last valid equipment shall
be re-examined.
b) Probe unit with contoured contact wedges may be used to aid ultrasonic coupling. Such
cases, calibration shall be done with the contact wedges used during the examination.
c) The crystal sizes for angle beam probes may be 10 mm diameter or 8 x 9 mm, and 10
mm to 24 mm diameter for straight beam probes.
e) The sizes and frequencies other than stated in c) and d) above may be used to assure
adequate penetration or better resolution.
The basic calibration reflectors shall be used to establish a primary reference response of
the equipment.
The material from which the block is fabricated shall be of the same product form,
and material specification or equivalent P-number grouping as the material(s) being
examined. Calibration blocks for dissimilar metal welds, the material selection shall
be based on the material on the side of the weld from which the examination will be
conducted
When two or more base material thickness are involved, the calibration block
thickness shall be determined by the average thickness of the weld
The finish on the surfaces of the block shall be representative of the surface finishes
on the components to be examined.
The basic calibration reflectors may be located either in the component material or in
a basic calibration block.
Note: For each increase in weld thickness of 50mm or fraction thereof over
100mm, the hole diameter shall increase 1.5mm.
Calibration shall include the complete ultrasonic examination system. The original calibration
must be performed on the basic calibration block. Checks shall be made to verify the sweep
range calibration and distance amplitude correction. Checks shall include the entire
examination system.
a) Position the angle beam probe towards the 100 mm quadrant and
maximise the echo response by moving the probe front and back
parallel to the sides of the calibration block, without any angulations.
(Rotation of the beam into the corner formed by the hole and the side of
the block may produce higher amplitude at a longer beam path; this
beam path shall not be used for calibration).
b) To start with, adjust the initial echo at zero division on the screen.
f) Using only delay control, bring the first back reflection to zero scale
position (the initial response need not coincide with zero position of the
scale).
g) Now the tenth scale division of the screen represents 100 mm metal
path in steel.
b) To start with, adjust the initial echo at zero division on the screen.
c) Set the first and second back reflections to 2.5 and tenth scale divisions
for the probe position towards 25 mm quadrant by adjusting both range
and delay controls. Then for verifying the calibrations keep the probe
towards 50 mm. Now the first back reflection shall be at fifth scale
division.
d) Now the tenth division of the screen represents the 100 mm steel metal
path distance.
a) Position the probe for maximum response from ¼ T hole and adjust the
gain setting to provide an indication height of 80% (± 5% of full screen
height) and mark the peak of the indication on the screen
b) Position the probe for maximum response from ½ T hole and mark the
peak of the indication on the screen
c) Position the probe for maximum response from ¾ T hole and mark the
peak of the indication on the screen.
d) Connect the screen marks for the side-drilled holes to provide the
distance amplitude curve.
a) To start with, adjust the initial echo at zero division on the screen
b) Position the straight beam probe on the IIW-V1 block 25 mm thick side
and maximise the first reflection.
b) Adjust the first, second, third and fourth back reflections to 2.5, 5.0, 7.5
and tenth scale divisions by using range and delay controls.
d) Now the tenth division of the screen represents the 100 mm metal path
distance in steel.
Calibration confirmation shall be performed prior to use of the system in the thickness range
under examination. A calibration check shall verify the sweep range calibration and distance-
amplitude correction.
If any point on the DAC curve has moved on the sweep line more than 10% of the sweep
division reading, correct the sweep range calibration and note the correction in the
examination record. If recordable reflectors are noted on the data sheets, those data
sheets shall be voided, a new calibration shall be recorded, and the voided examinations
shall be repeated
If any point on the distance-amplitude correction (DAC) curve has decreased 20% or 2
dB of its amplitude, all data sheets since the last calibration or calibration check shall be
marked void. A new calibration shall be made and recorded and the voided examination
areas shall be re-examined. If any point on the distance-amplitude correction (DAC)
curve has increased by more than 20% or 2 dB of its amplitude, all recorded indications
since the last valid calibration or calibration check shall be evaluated with the corrected
calibration, and their values shall be changed on the data sheets.
A calibration check on at least three of the basic reflectors in the basic calibration block shall
be made at the beginning , every 4 hours and at the end of each examination or whenever
examination personnel are changed and at any time that malfunctioning is suspected.
8.5 Beam Spread Angle Determination for Angle Beam Probe by 20dB Method
a) Set the maximum amplitude from the ¼ T hole at 100% of full screen.
b) Move probe toward the hole until indication equals 10% of full screen. Mark beam
centre line ‘toward’ position on the block.
c) Set back the maximum amplitude from the ¼ T hole at 100% of full screen. Move
probe away from the hole until indication equals 10% of full screen. Mark beam centre line
‘away’ position on the block.
d) Position the search unit for 100% of full screen amplitude from the ¼ T hole. Move
probe right without pivoting the beam toward the reflector until the indication equals 10% of
full screen. Mark the beam centre line ‘right’ position on the block (when manually
positioning the search unit, a straight edge may be used to guide the search unit while
moving to the right and left to ensure that axial positioning and beam alignment are
maintained).
e) Move probe left without pivoting the beam toward the reflector until the indication
equals 10% of full screen. Mark the beam centreline ‘left’ position on the block.
f) Repeat setting and measurements of ‘toward’, ‘away’, ‘right’, and ‘left’ positions
marked on the block for the ½ T and ¾T side drilled holes.
g) Record the dimensions from ‘toward’, ‘away’, ‘right’, and ‘left’ positions marked on
the block for the ¼ T and ¾T side drilled holes.
9.1 Client shall supply sufficient information to the technician enabling the item for testing to be
identified.
9.3 The test surface shall be free of irregularities; loose foreign matter or coatings, which may
interfere with ultrasonic wave transmission.
10.1 Procedure
Variation from this test procedure shall only be undertaken with the prior written approval of
the Client. Each variation shall be noted on the ultrasonic report and a relevant test
technique or procedure produced, if required.
10.2 Couplant
10.2.1 Couplant such as Chemetall UCA-2M, Ultragel, cellulose paste or equivalent may be
used.
Couplant used on nickel base alloys shall not contain more than 250ppm couplant
used on austenitic stainless steel or titanium shall not contain more than 250ppm of
halides (chlorides plus fluorides).
10.2.2 Specific consideration shall be given to the service and material of the component
tested. Such considerations shall include, but not limited to oil or ‘raw’ water where
such could be injurious or hazardous in certain applications and materials.
The rate of search unit movement for examination shall not exceed 150 mm / second unless
calibration shall be verified at intended scanning speed.
10.4.1 The volume shall be examined by moving the search unit over the examination
surface so as to scan the entire examination volume. Each pass of the probe shall
overlap a minimum of 10% of the probe (piezo-electric element dimension
perpendicular to the direction of the scan).
10.4.2 For weld testing, scanning shall be done from both the sides of same face of material.
In case both the sides of same face are inaccessible the information shall be recorded
in the report.
10.4.3 The entire volume of the weld needs to be covered, if necessary two or more angle
beams need to be used to ensure full volume coverage.
For plate lamination testing, using straight beam probe, the area to be tested shall be
scanned in grid pattern. The grid dimensions shall be 225 mm width by 225 mm breadth
unless otherwise previously specified and agreed with the Client.
Scanning shall be performed at a gain setting of at least two times the primary reference
level.
When the referencing code requires detection and evaluation of all indications exceeding
20% DAC, the gain should be increased an additional amount so that no calibration reflector
indication is less than 40%FSH. As an alternate, the scanning sensitivity level may be set at
14dB higher than the reference level gain setting.
Scanning surface shall be free of weld spatter, dirt and corrosion products. Forged or rolled
surfaces having reasonably smooth surfaces are acceptable.
10.9 Scanning of weld
The scanning of the adjacent base metal shall be performed to detect reflectors that
might affect interpretation of angle beam results, and is not to be used as an
acceptance-rejection examination. Locations and areas of such reflectors shall be
recorded.
10.9.2 Scanning for reflectors oriented parallel to the weld using angle beam probe
The angle beam shall be directed approximately right angles to the weld axis from two
directions where possible. The probe shall be manipulated so that the ultrasonic
energy passes through the required volumes of weld and adjacent base metal.
10.9.3 Scanning for reflectors oriented transverse to the weld using angle beam probe
The angle beam shall be directed essentially parallel to the weld axis. The search unit
shall be manipulated so that the angle beam passes through the required volumes of
weld and adjacent base metal specified by the referencing code section.
Used to size defect dimensions which are larger than the beam, such as the length of lack of
sidewall fusion in a weld. The probe is moved off the end of the defect until the signal amplitude is
reduced by 50 % (6dB). The position of the centre of the probe is then marked onto the material
surface. The probe is now in a position where the beam is half on and half off the defect. If this is
repeated at the other end of the defect then the distance between the marks represents its length.
12.1 All ultrasonic reflectors do not indicate flaws as certain metallurgical discontinuities and
geometric conditions may produce indications that are not relevant. Included in this category
are plate segregates in the heat affected zone that become reflective after fabrication. Under
straight beam examination, these may appear as spot or line indications. Under angle beam
examination, indications that are determined to originate from surface conditions such as
weld root geometry may be classified as geometric indications.
12.2 The identity, maximum amplitude, location, and extent of reflector causing a geometric
indication shall be recorded.
12.3 Interpret the area containing the reflector in accordance with the applicable examination
procedure. Plot and verify the reflector coordinates. Prepare a cross-sectional sketch
showing reflector position and surface discontinuities such as root.
12.4 Review fabrication or weld preparation drawings.
13.1 It shall be responsibility of NDT UT Level II to evaluate and report within the limits of criteria
as per appendix B.
Any reflector, which causes an indication in excess of 20% of DAC, shall be investigated to
the extent that it can be evaluated in terms of the acceptance criteria of the reference code
section.
For examination using other than DAC, evaluation shall be according to the requirements of
the reference code section.
14.0 RECORDS
14.2 Reports shall contain the following information as a minimum. This would include (where
available):
a) Client
b) PO / request no.
c) Stork-Cooperheat job no.
d) Line/Drawing/Component and / or joint identification
e) Acceptance criteria
f) Procedure and revision
g) Dimensions and / or area of test
h) Material/weld type and condition
i) Material Surface temperature
j) Calibration block and serial number
k) Test equipment type and serial number
l) Details of probe, frequency and size
m) Remarks/reference Sketches
n) Couplant used and brand name
o) Search unit cable used, type and length
p) Instrument reference gain
q) Calibration data
r) Interpretation and test results
s) Technician name, qualification level, signature & stamp
t) Date of test
14.3 Probe details shall include the type, crystal dimensions, nominal angles and frequencies.
The evaluation sensitivity / reference level shall also be included. The scanning sensitivity /
test level shall be in accordance with item 8.0 herein.
14.4 The test results shall include details (when available) of the item under test, welded joint or
plate tested, welding operator identification and the outcome of the test as detailed in item
8.0 herein.
14.5 When testing plate / parent material for thickness measurement / corrosion monitoring, any
relevant discontinuity noted shall be recorded and the size and depth from the scanned
surface noted. Any internal pitting or corrosion found during the testing shall also be
recorded.
14.6 Recorded / rejected reflectors shall have the following information noted:
a) Test number
b) Discontinuity type
c) Length of reflector
d) Reference height of reflector
e) Location to datum of reflector
f) Assessment per acceptance criteria
g) Sketches of reflector cross-sectional positions and / or location drawings
14.7 All reports shall be handed over against signature in the pink copy to the Client
representative and maintained at site as evidence for the handover of report. The Original,
white and blue copies shall gat signed by all parties. The blue copy will be on job file, the
white copy for the Client and the original is meant for invoicing. At the stage of handing over
of the report only the pink copy will be at site, which will only have two signatures at that
stage, Stork-Cooperheat and the Client.
14.8 Stork-Cooperheat responsibility shall be limited to ensuring the integrity of the weld or part
within the confines of the ultrasonic test requirements of the governing specification.
System Calibration
(Refer item 8.2)
Fig. 1
25 mm
0 1 2 3 4 5 6 7 8 9 10
Fig. 2a
0 1 2 3 4 5 6 7 8 9 10
100 mm R
Fig. 2b
0 1 2 3 4 5 6 7 8 9 10
25 mm R 50 mm R
Fig. 3a FIRST SECOND
ECHO REPEAT
0 1 2 3 4 5 6 7 8 9 10
Calibration
check
Fig. 3b
0 1 2 3 4 5 6 7 8 9 10
+
¼T + ½ T
5/4 T
¾ T +
+
100
90
80
70
60
50
40
30
20
10
0
0 1 2 3 4 5 6 7 8 9 10