Weighing Indicator
User manual
User Manual
safety instruction
For safety operation pls. follow the safety instruction.
WARNING
WARNING
Pls. make sure the weighing display well earthing
WARNING
Page 2 of 27
User Manual
1.2 Technical parameter
1.3 Outline and installation picture
1.4 Battery
2. Installation and Calibration
2.1 connection indicator with load cell
2.2 communication interface
3. Basic operation
3.1 Key and display
3.2 Power on
3.3 Zero setting
3.4 TARE
3.5 HOLD
3.6 TOTAL
3.7 10 times high resolutions
3.8 Up and Low limit alarm
3.9 Print function
4 Calibration & parameter setting
4.1 Enter calibration
4.2. Step of calibration operation:
4.3 Application parameter setting
4.4 Communication setting
4.5 Application setting
4.6 Exit setting
5.Output data format
5.1 Computer continuous sending format
5.2 Big display continuous sending format
5.3 Serial interface reception command:
5.4 Print output format
5.5 Print the accumulated output format
6. Maintenance
6.1 Regular Error and maintain method
6.2 Daily maintenance
Page 3 of 27
User Manual
6.3 Restore default parameters
1. Summary
》General weighing: zero tare
》Animal weighing: Peakhold. Datahold, Autohold
》Accumulation
》Optional by RS232
》Low battery remind
》Power off automatically
》Stimulating voltage:+3.3 VDC
》A/D converting speed: 1 0 SPS
》Load signal range: 0~12.8mV
》load capacity: it can connect 4 pcs 350 load cell at most
Ω
》weight unit: kg.
》 : 3000e
Resolution
》Interval: 1/2/5/10/20/50
》Display: 6digits LED/LCD,word height: 20.3mm
》key: ON/OFF TOTAL HOLD TARE ZERO SET
Page 4 of 27
User Manual
》Interface:RS232C Baud rate optional 1200/2400/4800/9600
》Operation temperature: 10~40℃
》 Storage temperature : 20~+60℃
》optional power: 4
V/4Ah rechargeable battery
110/220VAC
1.3 Outline and installation picture
1.4 Battery
Page 5 of 27
User Manual
1. when you use the internal battery first time,you should charge the
battery 1012 hours, to prevent low voltage resulted from self leakage of
battery.
2. when the red battery light is on and flashes,it means low battery You
should charge battery in time.
3. Charge time: 1012 hours And it works 45 hours
4. When the battery light turns green, it means fully charged
5. If you don’t use the battery long time, take out the battery to protect the
indicator from battery leakage
6. In order to keep the battery in best using condition, it is suggest that you
fully discharge the battery every month, the method is that using the
indicator till it is automatically power off.
2. Installation and Calibration
it can connect four pcs 350Ωload cell at most, both four and six line load
cell are ok. To make it simple, we use quick connector
Or standard plug. As belows
Page 6 of 27
User Manual
Quick connector connection
PIN2TXD
5 4 3 2 1 PIN3RXD
PIN5 GND
9 8 7 6
3. Basic operation
Page 7 of 27
User Manual
LP7516 Weighing indicator display instruction
LED display instruction
Weighing data display
Kg Weight unit kg
HOLD Hold the data
Tare Display tare status
Page 8 of 27
User Manual
Net Display net weight
Gross Display gross weight
Display data keep still
zero, indicating zero weight
Battery Using battery
Hi Over Limit
OK Within Limit
Lo Below Limit
Total Accumulation
Count Counting function
Key’s function
ON/OFF, EXIT AND SAVE SETTING
LEFT
RIGHT
DOWN
UP
CONFIRM, GO TO NEXT STEP
Page 9 of 27
User Manual
Work together with
“on/off” enter and exit
SET calibration
ZERO Clear weight within zero range
1.At Gross mode, tare the loaded weight
TARE 2.At Net mode, display gross weight after
deduct tare
HOLD Enter and exit “hold” mold
TOTAL accumulating operation
Press it for 2 seconds to power on or power
ON/OFF
off
Page 10
of 27
User Manual
3.2 Power on
Power on and indicator perform selfchecking and go to weighing mode.
Within zero range, press “zero”, indicator weighing is cleared. When
Indicator is not stable, zero is unworkable.
3.4 TARE
At the gross weight mode, if the weight is stable, pls. press ”Tare” key , the
indicator will take the loaded weight as tare, and show net weight,
At this time the gross mode will change to net mode. The “ net” and “tare”
light is on, and the net weight is zero
3.5 HOLD
C11=0 “hold” function unworkable
C11=1 PEAK HOLD
Press” HOLD” key, the Hold light is on, and show the Maximum data on the
Page 11
of 27
User Manual
weighing indicator. Press “ HOLD” key again to exit the hold function
C11=2 Datahold
Press” HOLD” key, the Hold light is on, and show the data on the weighing
indicator. Press “ HOLD” key again to exit the hold function
C11=3 Autohold
If the weight on the scales above 20d and keep stable, the indicator will show
the data for 6 seconds and the “ hold” light Is on , after 6 seconds the
indicator back to general weighing, and the “ hold” light is off
C11=4 Special Animal weighing function
Press” Hold” key, the indicator will show” LOC” for 3 seconds, the “ hold”
light is on, During the 3 seconds, the indicator will catch the average weight
and show it.
Press” HOLD” key again to exit it
3.6 TOTAL
Accumulation operation
At Zero mode, load weight till stable, Press “TOTAL” key go to
accumulating
Mode,” total” light on, display” n001”, and then display loaded weight;
unload weight , back to zero, load weight again till stable. Press ”TOTAL”,
display”n002”
Then show the loaded weight. Repeat it maximum 999 times.
Check the total weight operation:
Press “SET” hold it then press “TOTAL” At the same time, display ”n**”,
(accumulating times) then display total weight.
Page 12
of 27
User Manual
There are 8 data totally. It shows the first 4 digital. then the last 4 digital
For example, the first 4 digital is”0012”, the last 4 digital is”34,56”
It means the actual weight is “1234.56”
At TOTAL (accumulate) mode, Press “TOTAL” key the indicator show“
clrn”, it means don’t clear the total Weight, Press “ key confirm it
and exit ; if clear total weight, Press “ ” “ ”
Press “ SET” and “TARE” key at the same time, you will get 10 times high
resolutions. And it back to normal weighing after 3 seconds.
Pls. set C13= Up limit, C14=Low Limit, when the weight is over up limit,
the “HI” light will on, and indicator will make a sound to alarm.; when the
weight is below than the low limit, the “LO” light will on.. when the weight
is within the limit, the “OK” light is ok.
When the data is stable, connection with printer, it will be printed after
Page 13
of 27
User Manual
press“ ”1 second.
Note:print the gross weight when at tare mode , if the net weight is zero.
Can not print.
setting
There have two methods to enter the setting menu:
1. when the switch” CAL” is off, Press” ” then press” at the
same time, hold it, you will enter C08C39 setting.
2. Take out the sealing screw on the back of indicator, then press
down the “span” Press still and then press at the same time ,
you will enter C01C39 setting.
Page 14
of 27
User Manual
According to the second method which can enter setting menu, C01C39
C01 UNIT
[C01 ] Press
[C1 1] unit is kg
Press , go to next step
C02 Set decimal digits
[C02 ] Press
option:0/1/2/3/4
[C2 0] no decimal point
[C2 1] one decimal point
[C2 2] two decimal point
[C2 3] three decimal point
Press , go to next step
C03 Division setting
[C03 ] Press
[C3 1] d=1
[C3 2] d=2
[C3 5] d=5
[C3 10] d=10
[C3 20] d=20
[C3 50] d=50
Press , go to next step
C04 Maximum capacity
Page 15
of 27
User Manual
For example:max weighing 100kg:
set [0100.00]
Press , go to next step
C05 Zero calibration
option:0=noncalibration zero 1=need calibration zero
calibration zero please choose 1 and ensure scale is empty and “stable” light
is on. countdown[CAL 10]~[CAL 0], then the indicator will
show[0.00](example for two decimal point).
C06 Loading calibration
, show [SPAn ],
Basic on max capacity setting, add suitable weight on scale. close to the max
capacity, heavier than 10% max at least.
For example: the weight is 80kg
As bellows:
[0080.00]
[CAL 9]
……
[0080.00]
[CALEnd]
count down over, indicator shows loaded weight , loading calibration finish.
Press
C07 Default parameters setting
[C07 0] nonrestore default parameters
[C07 1] restore default parameters
Note: after the above parameters setting finish, please do not set default
parameters often, avoid the original setting parameters lost.
Page 16
of 27
User Manual
C08 warning tone
[C8 1] open warning tone
[C8 0] close warning tone
C09 Power off automatically
[C9 0] Nonpower off
[C9 10] keep still within 10 min. power off automatically
[C9 30] keep still within 30 min. power off automatically
[C9 60] keep still within 60 min. power off automatically
C10 Power saving setting
[C10 0] close backlight
[C10 1] close backlight after 1 minute
[C10 2] always backlight
C11 Hold
[C11 0] No Hold function
[C11 1] Peck hold
[C11 2] Data hold
[C11 3] Autohold
[C11 4] Animal weighing
C12 Hold time ( if you choose C11=4, you can set the time )
[C12 3] 3 seconds
[C12 5] 5 seconds
C13 Upper limit alarm value
C14 Low limit alarm value
C15 Check inner code
Page 17
of 27
User Manual
C18 Serial interface setting
[C18 0] No sending
[C18 1] Big display
[C18 2] Print format output
[C18 3] Command mode(Z =zero T=tare R=Reply weight
[C18 4] continuous sending
C19 BAUD RATE
[C19 0] 1200bit/s
[C19 1] 2400bit/s
[C19 2] 4800bit/s
[C19 3] 9600bit/s
C20 Manually Zero
[C20 00] no Manually Zero
[C20 01] Manually Zero range ±1% Max. capacity
[C20 02] Manually Zero range ±2% Max. capacity
[C20 04] Manually Zero range ±4% Max. capacity
[C20 10] Manually Zero range ±10% Max. capacity
[C20 20] Manually Zero range ±20% Max. capacity
[C20100] Manually Zero range ±100% Max. capacity
C21 Initially zero
[C21 0] No initially zero
[C21 1] Initially zero range±1% Max. capacity
[C21 2] Initially zero range±2% Max. capacity
Page 18
of 27
User Manual
[C21 5] Initially zero range±5% Max. capacity
[C21 10] Initially zero range±10% Max. capacity
[C21 20] Initially zero range±20% Max. capacity
C22 Zero tracking range
[C22 0.0] No zero tracking
[C22 0.5] ±0.5d
[C22 1.0] ±1.0d
[C22 2.0] ±2.0d
[C22 3.0] ±3.0d
[C22 4.0] ±4.0d
[C22 5.0] ±5.0d
C23 Zero tracking time
[C23 0] No zero tracking
[C23 1] 1 second
[C23 2] 2 seconds
[C23 3] 3 seconds
C24 Overload range
[C24 09] Over 9d than max. capacity
C25 Negative display
[C25 00] Less than 9d
[C25 10] Less 10% Max. capacity
[C25 20] Less 20% Max. capacity
[C25 50] Less 50% Max. capacity
[C25100] Less 100% Max. capacity
C26 Standstill time
[C26 0] Quick
[C26 1] Medium
[C26 2] Slow
C27 Standstill range
[C27 1] ±1d
[C27 2] ±2d
[C27 5] ±5d
Page 19
of 27
User Manual
[C27 10] ±10d
C28 Dynamic filter
[C28 0] close dynamic filter
[C28 1] Low dynamic filter
[C28 3] Medium dynamic filter
[C28 5] High dynamic filter
C29 Noisy filter
[C29 0] Close noisy filter
[C29 1] Low noisy filter
[C29 2] Medium filter
[C29 3] High filter
For example [C10 1], Press , confirm it then press to exit and
save it
Page 20
of 27
User Manual
5.Output data format
, , CR L
F
S 1 S 2 S 3 Data S 4
S1: weight status, ST= standstill, US= not standstill, OL= overload
S2: weight mode, GS=gross mode, NT=net mode
S3: weight of positive and negative, “+” or ” –“
S4: measurement unit, “kg” or “lb”
Data: weight value, including decimal point
CR: carriage return
LF: line feed
Output continuous format
S S S S C
C
T W W W X X X X X X X X X X X X K
R
X A B C S
1 2 3 4 5 6
Page 21
of 27
User Manual
State A
Bits0,1,2
0 1 2 Decimal point position
1 0 0 XXXXXX0
0 1 0 XXXXXXX
1 1 0 XXXXX.X
0 0 1 XXXX.XX
1 0 1 XXX.XXX
Bits3,4 Division
0 1 X1
1 0 X2
State B
BitsS function
Bits0 gross=0, net=1
Bits1 symbol:positive =0,negative =1
Bits2 overload(or lower zero)=1
Bits3 dynamic=1
Bits4 unit:lb=0, kg=1
Bits5 Constant 1
Bits6 Constant 0
State C
Bit2 Bit1 Bit0 unit
0 0 0 Kg or lb
0 0 1 g
0 1 0 t
Bit 3 printing=1
Extend
Bit 4
display=1
Bit 5 Constant 1
Bit 6 Constant 0
Page 22
of 27
User Manual
RS232COM serial interface can receive simple ASCII command.
Command word and role as follows:
Command name role
T Tare off command Save and clear tare
Z Zero command Zero the gross weight
P Print command Print the weight
R Read gross/ net Read gross/net weight
weight
5.4Print output format
NO. 004 (NO.)
G.W: 8.88kg (gross,example for two decimal point)
T.W: 2.88kg (tare)
N.W: 6.00kg (net)
NO. 004 (NO.)
Total: 003 (accumulate times,example for 3 times is 003)
Total.W: 2.88kg (accumulate weight)
Page 23
of 27
User Manual
6. Maintenance
Page 24
of 27
User Manual
during calibration, the Added weight at least
added weights not 10%of Max. capacity,
enough Recommend the weights is
ERR2
6080% the Max. capacity
6.2
Daily maintenance
1. In order to ensure indicator display clearly and prolong use life, the
indicator should not be placed directly on sunlight.
2. Load cell and indicator should be well connected , the system should
have a good ground, away from strong electric field, magnetic field.
3. Do not use indicator outside in rainy, better keep it power off.
4. Power off firstly while plug and unplug
Page 25
of 27
User Manual
Enter setting menu, set C07= 1,press then press exit saving
setting, all parameters will be back to default setting.
Note:Pls. do not restore default parameter easily if you are not professional
and have not scale calibration.
Default parameter form
parameter instruction Default value
C01 Calibration unit 1
C02 decimal digits 0
C03 Division value 1
C04 Max capacity 10000
C05 Empty scales calibration 0
C06 Capacity calibration 0
C07 restore the default parameters 0
C08 Warning tone 1
C09 Automatic power off 0
C10 Power saving mode 0
C11 Hold function 0
C12 Animal weighing mode 0
C13 Upper limit warning 000000
C14 Lower limit warning 000000
C15 Inner code display
Page 26
of 27
User Manual
C16 Date
C17 Time
C18 Serial interface data output methord 0
C19 Serial interface Baud rate 3=9600
C20 Manual zero setting 2
C21 Initical zero setting 10
C22 Automatic zero tracking range 0.5
C23 Automatic zero tracking time 1
C24 Verload range 9
C25 Negative display range 10
C26 Standstill time 1
C27 Standstill range 2
C28 Dynamic filter 0
C29 Noisy filter 2
C30~C40 Reseverd menu