INSTRUCTIONS MANUAL
CODE 80392 REV 0 DEC 2019 (It can be modified without notice)
SOFTWARE
M.WAVE PROFESSIONAL
V-1100 / VR-2000
USER MANUAL
(FOR SPECTROPHOTOMETERS
V-1100 & VR-2000)
4312001
J.P. SELECTA s.a.u. Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España Tel 34 937 700 877 Fax 34 937 702 362
e-mail: [email protected] - website: http://www.jpselecta.es
Contents
Functions ............................................................................................... 1
Main Functions ............................................................................... 1
Spectrum Processing Function .............................................. 2
System Check and Calibration Function ......................... 3
Starting.................................................................................................. 3
PC System Requirements............................................................. 3
Install UV-Vis Analyst............................................................. 4
Uninstall UV-Vis Analyst ........................................................ 6
Run UV-Vis Analyst ...................................................................... 6
Setup Communication Port ........................................................ 7
Introduction ........................................................................................ 7
Main Interface ............................................................................... 7
Menus Bar and Tools Bar .......................................................... 8
Operation ............................................................................................. 13
Single Wavelength Photometric Measurement .............. 13
Fixed Point Measurement ........................................................ 14
Multi-wavelength Photometric Measurement ............ 14
Concentration Measurement................................................ 17
Wavelength Scanning .................................................................. 22
Time Scanning (Kinetic Analysis) ................................... 38
DNA/Protein Measurement ........................................................ 41
Appendix................................................................................................ 45
Methods of Quantitative Analysis ................................... 45
UV-Vis Analyst
Functions
This section introduces the functions of the UV-Vis
Analyst.
Main Functions
Single wavelength photometric measurement
Go to a desired wavelength quickly and conveniently.
Photometric value display mode can be changed (%T or
Abs).
Fixed Points Measurement
Multi-wavelength Photometric Measurement
Up to 20 wavelength points can be set up.
Results will be grouped into a table format
automatically.
Concentration Measurement
2 methods to set up the regression curve.
Up to 20 standards to set up the regression curve.
The UV-Vis Analyst will calculate the working curve
using a linear equation that fits the data. Enter
factor values to generate regression curve.
3 methods for curve fit.
Linear fit, Quadratic fit and Cubic fit.
Wavelength Scanning
Allow user to set scan step (0.1, 0.2, 0.5, 1.0 and
5.0nm).
Spectrum display mode can be changed (Wavelength-%T
or Wavelength-Abs).
Peaks and valleys will be automatically detected
after scanning (User can define the peak threshold).
Powerful spectrum processing functions are provided.
Time Scanning
1
UV-Vis Analyst
Allow user to set scan Interval (0.5, 1.0, 2.0, 5.0,
10, 30 and 60s).
Spectrum display mode can be changed (Time-%T or
Time-Abs).
Peaks and valleys will be automatically detected
after scanning (User can define the peak threshold).
Powerful spectrum processing functions are provided.
DNA/Protein Measurement
7 methods to measure DNA/Protein.
Wavelength points and ratios can be set up.
Results will be grouped into a table format
automatically.
Spectrum Processing Function
Trace a Spectrum
The cursor can be moved to a desired point in the
spectrum displayed on the screen and the photometric
data at this point is displayed.
Automatic Peak Detection
After a scanning is complete, peaks and valleys can be
automatically detected and listed in a table format.
They will also be labeled on the spectrum.
Scale Expansion
Simultaneous expansion of the X and Y axes are provided
with the “Zoom” function. Display range can also be
changed though the “Display Setup” function.
Differentiation
You can calculate and display the first through to the
fourth derivative spectrum for a given spectrum.
Derivative spectrum is useful for enhancing spectrum
data that are not readily apparent in an absorbance
spectrum.
Calculate Spectrum
You can calculate addition, subtraction, multiplication
and division between two spectrum with the resulting
data displayed on the screen.
2
UV-Vis Analyst
System Check and Calibration
Function
Instrument Validity Check
Up to 10 wavelength points can be set up in the
instrument validity mode. Two methods can be selected
(Photometric Validity measurement and Wavelength
Validity measurement) and tolerance can be entered.
Results will be grouped into a table format
automatically.
Dark Current Check
You can resample the dark current of the instrument.
Spectrum Bandwidth Check
A special scan for checking spectrum Bandwidth and it
will calculate the spectrum Bandwidth value
automatically.
Energy of Light Sources Check
It allows scan the energy of light sources with a fixed
gain (0-7).
Reset Wavelength
It affords to relocate the 656.1nm.
Starting
This section introduces how to start to use the UV-Vis
Analyst.
PC System Requirements
Pentium or above PC;
CD-ROM;
USB Ports.
512MB Memory(1GB or Above is strongly recommended);
50MB or above hard disc Space;
Microsoft Windows XP/Vista/7/8.
3
UV-Vis Analyst
Install UV-Vis Analyst
Note: Please disconnect the USB Cable and
unplug the USB Key before install the
UV-Vis Analyst software.
1. Put UV-Vis Analyst disc in the CD-ROM;
2. Double Click to open the CD-ROM, and then, double
click Setup.exe which is under the root directory of
CD to start installation , click Next;
3. Choose install path, then click Next to copy files
to PC;
4
UV-Vis Analyst
4. After finished copy all files of UV-Vis Analyst, it
will start install the USB Drive. Click Next;
< Back Next > Cancel
5. Select “I accept this agreement”. Click Next to copy
files to PC;
< Back Next > Cancel
6. Click Finish to finish install USB Drive;
5
UV-Vis Analyst
< Back Finish Cancel
7. Click Finish to finish all the setup.
Uninstall UV-Vis Analyst
Start→All Programs →UV-Vis Analyst→Uninstall UV-Vis
Analyst to remove.
Run UV-Vis Analyst
Note: Before you run UV-Vis Analyst, please
check these things as follow:
Computer has connected to spectrophotometer by
using the USB cable;
Spectrophotometer is on main interface.
WL: 500.0nm 10:00:00
D2
W
Main Menu
Basic Mode
Quantitative Mode
Wavelength Scan
Kinetics Mode
DNA/Protein Mode
Multi-Wavelength Mode
System Utility
Cancel Move Dn Select
There are two ways to start the UV-Vis Analyst:
Double-click shortcut icon on the desktop.
6
UV-Vis Analyst
Start→All Programs→UV-Vis Analyst→UV-Vis Analyst.
Setup Communication Port
On the UV-Photometer menu, click Comm. Hub Setup
appears the following box, select the RS232 Port and
Baud Rate (38400), click OK.
Introduction
We will introduce the UV-Vis Analyst in this chapter.
Main Interface
This is the main interface after start.
7
UV-Vis Analyst
Menu Bar Toolbar
Data Zone
Status Bar
Menus Bar and Tools Bar
Menu bar and Toolbar are both provided in the software
offering you two ways to select a desired function.
On the menu bar, use your keypad or mouse to select
the desired function.
Almost all the functions listed in the menu bar can
be reached by clicking a corresponding button in the
toolbar.
Main Menu Sub Menu Tool Function
New a Fixed
File New Points
New a
Measurement
Wavelength
New a Time Scan
Scan
Measurement
Measurement
8
UV-Vis Analyst
New a
DNA/Protein
New a
Measurement
Instrument
Open… Open a
Validity
spectrum/data
Close Close current
file
measurement
Save Save current
Save As… measurement
Save current
Open file from measurement
Open a file as
UV-Photometer a new file name
saved
Export Exportin
data or
instrument
method
Print… Print test
Print Setup… report
Setup printer
Exit Exit UV-Vis
Status Bar analyst
Display/Hide
Status of status
Displaybar
status
View
Spectrophotomet
Status font of
Setup font of
er spectrophotomet
status
Customize Define bar
the
er
information of
9 display and
print
UV-Vis Analyst
Peaks Mark peak value
Valleys Mark valley
Magnify value
Magnify the
Restore area selected
Restore the
Search default
Search
parameters
peak/valley one
Link Connect to the
fdisplay
by one
Spectrophotomet
Reset Instrument
Reset
er
Spectrophotomet parameters of
Escape Stop current
er instrument
measurement
View dark Retest the dark
UV-
Current
Set Amplifier current
Reset amplifier
Photomete
Locate 656.1nm Relocate
r
Calibrate 656.1nm
Scan system
System Baseline
Automatic Blank baseline
Do blank
Calibration
Slit Bandwidth Set slit
*
Set Unit bandwidth
Set unit (0.5,
1.0, 2.0, 4.0)
10
UV-Vis Analyst
Turn on/off W Turn on/off W
lamp
Turn on/off D2 lamp
Turn on/off D2
lamp
D2/W Switch lamp
Set switch
Point
Comm. Port point
Setup of D2/W
comm.
Setup
Change Password port
Set/Change
Locate Cell ** login
Locatepassword
cell (1-
Auto-
Setup Multicell 8) to light
Setup Multicell
sample
** path
Autorun ** Measure multi
Start samples
Start a
automatically
measurement
Scan Stop Stop a
Service measurement
Measure
Display Range spectrum and
Setup scan
Settings
scan energy
display
Peak Height Define
parameters
peak/valley
Add Add two
Compute
threshold
spectrum
Sub Subtract one
spectrum from
11 another
UV-Vis Analyst
Multiply Multiply two
Divide spectrum
Divide one
Moving Window spectrum
Smooth a from
Averaging another
spectrum
Savitzky-Golay Smooth a with
Smoothing the method
spectrum with
Derivate Derivative of a
Filter Moving
the Window
method
Resample spectrum
Resample a
Averaging
Savitzky-Golay
New Window spectrum
New a
Smoothing
Cascade measurement
Multi windows
Filter
window
display as
in a
Window Tile Multi windows
current
cascade
Arrange Icons display
Arrange in
alla
tile
icons
Split Split minimized
display
Help About UV-Vis area
Display the
Analyst information
Setup
about the UV-
measurement
Modify a
Vis Analyst
parameters
measurement
Delete results
result
selected
12
UV-Vis Analyst
Set and Goto
one wavelength
Display
Instrument CPU
Delete current
information
Spectrum
Display result
as mode %T
Display result
as mode
Undo Abs
Scale
Operation
This chapter introduces how to use UV-Vis Analyst.
Single Wavelength Photometric
Measurement
The UV-Vis Analyst provides a convenient method to
measure photometric value at a fixed wavelength.
1. Click on the toolbar, appears Goto specified
wavelength.
2. Key in the desired wavelength position, click Goto.
The minimum wavelength step is 0.1nm in a range from
13
UV-Vis Analyst
190-1100nm.
3. Do blank.
Single Beam: Place a reference in the
compartment, click Zero.
Double Beam: Skip to step 4.
4. Measure Sample.
Single Beam: Place a sample in the sample
compartment.
Double Beam: Place a reference in the reference
compartment and place a sample in the sample
compartment.
The photometric value will display in the Readout
box.
Fixed Point Measurement
This UV-Vis Analyst performs fixed wavelength
measurement at 1-20 points and how to analyze unknown
compounds against calibration standards.
Multi-wavelength Photometric Measurement
1. Click on the toolbar, appears follow form.
14
UV-Vis Analyst
2. Click the Method tab.
3. Type the number of wavelength points in the Number
of WL Points box, or click the up/down arrows next
to the box set the wavelength points. Leave the two
boxes Calculate Concentration and Use Standard
Samples.
4. Key in the wavelength in the Wavelength box.
5. Click the Sample tab. It will display the following.
The control menu contains six buttons: Start,
Delete, Modify, Recalculate, Data Font and Print.
15
UV-Vis Analyst
6. Do blank.
Single Beam: Place a reference in the
compartment, click .
Double Beam: Skip to step 7.
7. Measure Sample.
Single Beam: Place a sample in the sample
compartment.
Double Beam: Place a reference in the reference
compartment and place a sample in the sample
compartment.
Click to run a new measurement. The display
will change to the following. Key in the sample name
in the Name box.
16
UV-Vis Analyst
8. Click OK. The photometric data for sample will be
listed in the Sample table.
9. Repeat steps 7-8 to measure all samples.
Concentration Measurement
Set Up Linear Regression Curve
There are two methods available to set up the linear
regression curve. You can use standards to set up the
regression curve or just key in the parameters
manually. Use the following steps to select the method
you wish to use.
1. Click on the toolbar.
17
UV-Vis Analyst
2. Click the Method tab.
3. Enter the number of wavelength points in the Number
of Points box.
4. Key in the wavelengths in the Wavelength boxes.
5. Tick the Calculate Concentration check box to
activate concentration calculation.
6. Set up the linear regression curve.
Method 1: Set up the linear regression curve with
prepared standards.
(1) Tick the Use Standard Samples check box.
(2) Click the Standard tab.
(3) Do blank.
Single Beam: Place a reference in the
compartment, click .
Double Beam: Skip to step (4).
(4) Measure Standard Sample.
Single Beam: Place a sample in the sample
compartment.
Double Beam: Place a reference in the reference
compartment and place a sample(Standard 1) in
the sample compartment.
Click Start to run a measurement.
(5) Key in the concentration value of Standard 1 in
the Conc. box.
(6) Key in the sample name for the standard in the
Name box.
(7) Click OK. The photometric data, △A and
concentration will be shown in the standard table.
(8) Repeat steps (4)-(7) to measure all the prepared
standards.
18
UV-Vis Analyst
(9) Click down arrow in Curve Fit box to select curve
fit method.
Method 2: Input the factor of the linear regression
curve.
(1) Leave the Use Standard Samples check box.
(2) Click down arrow in Curve Fit box to select curve
fit method.
(3) Input the factor of the linear regression curve.
7. Click Fitting tab to view the linear regression
curve. Click Display Setting tab to set the display
parameters and unit of concentration.
19
UV-Vis Analyst
Measure Concentration by Using The Linear Regression
Curve
The following procedure shows how to measure
concentration of samples.
20
UV-Vis Analyst
1. Set up linear regression curve or click to open
a file of linear regression curve (*.QUA).
2. Click the Sample tab.
3. Do blank.
Single Beam: Place a reference in the
compartment, click .
Double Beam: Skip to step 4.
4. Measure Sample.
Single Beam: Place a sample in the sample
compartment.
Double Beam: Place a reference in the reference
compartment and place a sample in the sample
compartment.
Click to run a new measurement. Type the sample
name in the Name box. The default is Sample-1.
5. Click OK. The photometric result for Sample-1 will
be listed in the sample data. Delta Abs. and
concentration value of Sample-1 will also be
displayed in columns 3 and 4.
6. Repeat steps 4-5 to measure remaining samples.
21
UV-Vis Analyst
Wavelength Scanning
This chapter describes how to collect a spectrum while
using Wavelength Scan function.
Scan Sample
1. Click on the toolbar to new a sample scan
measurement, appears the following form.
22
UV-Vis Analyst
2. Click on the toolbar, appears the following
form. Input start wavelength in From box (range:
190-1100nm), end wavelength in To box (range: 190-
1100nm), select scan interval (0.1, 0.2, 0.5, 1.0,
2.0 or 5.0nm) and Filter times (5, 10, 30 or 50),
click OK.
3. Click on the toolbar to select %Transmittance
23
UV-Vis Analyst
mode or click to select Absorbance mode.
4. Click on the toolbar to set display parameters.
5. Scan baseline.
Single Beam: Place a reference in the
compartment, click .
Double Beam: Skip to step 6.
6. Measure Sample.
Single Beam: Place a sample in the sample
compartment.
Double Beam: Place a reference in the reference
compartment and place a sample in the sample
compartment.
Click to scan sample, the real time spectrum
will be displayed. Click to cancel while
scanning.
24
UV-Vis Analyst
Auto List Peaks and Valleys
Click on the toolbar to set the peak/Valley
threshold (range: 0 to 1.000, step: 0.001), Input the
threshold value, click OK. Click to list peaks and
click to list valleys.
25
UV-Vis Analyst
Rescale
Click on the toolbar to set the new parameters for
display.
Original Scales
Click on the toolbar to restore the default display
settings.
Zoom Selected Area
Click on the toolbar to activate zoom function.
Position the cursor in the upper-left corner of the
area you want to select. Hold the left mouse button to
drag the cursor to outline the spectrum area you want
to enlarge. Release the mouse button. The part of the
spectrum which is displayed within the outlined area
26
UV-Vis Analyst
will be enlarged. Click to undo scale. To cancel
zoom to click again.
27
UV-Vis Analyst
Trace a Spectrum
Click on the toolbar, a crosshair cursor appears,
move the cursor on the spectrum. Move the crosshair
cursor left or right on the spectrum. The data in the
cursor window indicate the X-axis and Y-axis values for
the current cursor location. Click key “ESC” to release
the crosshair cursor.
28
UV-Vis Analyst
Select a Spectrum as Current
As UV-Vis Application Software can display several
spectrum overlaid on the screen, you should specify the
spectrum you wish to process. Click the down arrow on
the toolbar. All spectrums will be listed in the pull-
down menu. Click the spectrum you want to select. Its
name will be listed in the Name Box and it will be
referred to as Current Spectrum.
Derivative
Click on the toolbar. The following dialogue box
appears. Key in the class of derivative (1-10,
depending on whether 1st, 2nd, … 10th derivative is
required) and type a name for the result spectrum, then
click OK. The result spectrum will be displayed
29
UV-Vis Analyst
overlaid with the original one.
Moving Window Averaging
Click on the toolbar. Appears following form. Click
up/down arrow of the Range box to select range value,
key a file name in the Name box, click OK. The result
spectrum will be displayed overlaid with the original
one.
30
UV-Vis Analyst
Savitzky-Golay Smoothing Filter
On the Computer menu, click Savitzky-Golay Smoothing
Filter. Appears following form. Click up/down arrow to
select the parameters, key a file name in the Name of
Result box, click OK. The result spectrum will be
displayed overlaid with the original one.
31
UV-Vis Analyst
Resample
Click on the toolbar. The following dialogue box
will be displayed. Click Up/Down arrow to select Sample
times. Click OK. The new spectrum displays.
32
UV-Vis Analyst
Spectrum Addition
Spectrum addition can assist in the development of
artificial spectrum in multi-component mixtures.
Click on the toolbar. The following dialogue box
will be displayed. Click the down arrow next to File 1
to select a spectrum and define it as source 1. Select
a spectrum for File 2 in the same way. It will not
allow you to select the same spectrum twice. Key in a
name for the Result spectrum and click OK. The result
spectrum will be displayed on the screen.
Note:UV-Vis Analyst will only add, subtract,
multiply and divide two spectrums that are
already displayed on the screen. Before
arithmetic processing, load or collect two
spectrums from memory.
33
UV-Vis Analyst
Spectrum Subtraction
Subtracting one spectrum from another has been a
classical technique to offset spectrum interference
from the spectrum of interest.
Click on the toolbar. The following dialogue box
will be displayed. Click the down arrow next to File 1
to select a spectrum and define it as source 1. Select
a spectrum for File 2 in the same way. It will not
allow you to select the same spectrum twice. Key in a
name for the Result spectrum and click OK. The result
spectrum will be displayed on the screen.
34
UV-Vis Analyst
Spectrum Multiplication
Multiplying spectrum can assist in the development of
artificial structure of spectrum in multi-component
mixtures.
Click on the toolbar. The following dialogue box
will be displayed. Click the down arrow next to File 1
to select a spectrum and define it as source 1. Select
a spectrum for File 2 in the same way. It will not
allow you to select the same spectrum twice. Key in a
name for the Result spectrum and click OK. The result
spectrum will be displayed on the screen.
35
UV-Vis Analyst
Spectrum Division
Dividing one spectrum from another has been a classical
technique to offset spectrum interference from the
spectrum of interest.
Click on the toolbar. The following dialogue box
will be displayed. Click the down arrow next to File 1
to select a spectrum and define it as source 1. Select
a spectrum for File 2 in the same way. It will not
allow you to select the same spectrum twice. Key in a
name for the Result spectrum and click OK. The result
spectrum will be displayed on the screen.
36
UV-Vis Analyst
Unload a Spectrum
Select the spectrum you want to unload as the Current
Spectrum, Click on the toolbar to remove the
spectrum from the display.
Define Display Information
Click on the toolbar, appears the Settings to
display and print the spectra form, click the Legend
tab, type the information for display.
37
UV-Vis Analyst
Edit Print Information
Click on the toolbar, appears the Settings to
display and print the spectra form, click the Print
tab, type the information for print out.
Time Scanning (Kinetic Analysis)
This chapter tells you how to obtain the absorbance or
transmittance value for a sample as a function of time
at a given wavelength.
Scan Sample
1. Click on the toolbar, the following dialog box
will appear.
38
UV-Vis Analyst
2. Click on the toolbar to select the %transmittance
mode or click to select the absorbance mode.
3. Click on the toolbar. A dialogue box will be
displayed. Key in the wavelength, total time (in
seconds) and scan step in the above dialog box. The
wavelength range should be within 190 to 1100 nm. The
upper limit for total time is 100000 seconds. Seven
scan intervals can be selected from 0.5S, 1S, 2S, 5S,
10S, 30S and 60S. Click OK.
4. Do blank.
Single Beam: Place a reference in the
39
UV-Vis Analyst
compartment, click .
Double Beam: Skip to step 5.
5. Measure Sample.
Single Beam: Place a sample in the sample
compartment.
Double Beam: Place a reference in the reference
compartment and place a sample in the sample
compartment.
Click on the toolbar. The instrument will start
scanning automatically. The graph will be displayed on
the screen during time scanning. You can stop scanning
by clicking .
Calculate Rate
Click on the toolbar, appears the Settings to
display and print the spectra form, click the Dynamic
Analysis tab, type the begin time in Time Begin box,
40
UV-Vis Analyst
type the end time in Time End box, and type the K
factor in K Factor box, click Calculate, the result
will be displayed.
Define Display Information
Click on the toolbar, appears the Settings to
display and print the spectra form, click the Legend
tab, type the information for display.
Edit Print Information
Click on the toolbar, appears the Settings to
display and print the spectra form, click the Legend
tab, type the information for display.
DNA/Protein Measurement
This chapter describes how to perform DNA/Protein
41
UV-Vis Analyst
measurement.
DNA/Protein Measurement
1. Click on the toolbar, the following dialog box
will appear.
2. Click the down arrows of the method to select the test
method. Key in the wavelength position in the
Wavelength box. Key in the value of DNA/Protein Conc.
3. Click the Sample tab. It will display the following.
The control menu contains six buttons: Start, Delete,
Modify, Recalculate, Font and Print.
42
UV-Vis Analyst
4. Do blank.
Single Beam: Place a reference in the
compartment, click .
Double Beam: Skip to step 5.
5. Measure Sample.
Single Beam: Place a sample in the sample
compartment.
Double Beam: Place a reference in the reference
compartment and place a sample in the sample
compartment.
Click Start or to run a new measurement. The
display will change to the following.
43
UV-Vis Analyst
6. The UV-Vis Analyst will read the photometric value of
sample 1 at the fixed wavelength automatically. Key
in the sample name in the Name box. Click OK after
the measurement is complete. The photometric data for
sample 1 will be listed in the sample table.
7. Repeat steps 5-6 to test all samples.
44
UV-Vis Analyst
Appendix
Methods of Quantitative Analysis
Single Wavelength Method :Abs.=A1
Double Wavelengths Method :Abs.=m*A1-n*A2
Three Wavelengths Method :Abs.=A1-(WL1-WL2)*(A2-
A3)/(WL2-WL3)-A3
45
UV-VIS ANALYST FOR SPECTROPHOTOMETER