Faultfinding Procedures CDAS & PFEI

This document provides faultfinding procedures for various error results from the CDAS self-test, including "Stopped", "Errors", and "Error" results for different boards. It describes the purpose of the manual and CDAS self-test. Faultfinding procedures are meant to be followed carefully and feedback provided to improve them. The document also provides context about the CDAS and PFEI boards and their tests, as well as troubleshooting philosophy.

Uploaded by

Michael Kiggins
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
100% found this document useful (1 vote)
221 views35 pages

Faultfinding Procedures CDAS & PFEI

This document provides faultfinding procedures for various error results from the CDAS self-test, including "Stopped", "Errors", and "Error" results for different boards. It describes the purpose of the manual and CDAS self-test. Faultfinding procedures are meant to be followed carefully and feedback provided to improve them. The document also provides context about the CDAS and PFEI boards and their tests, as well as troubleshooting philosophy.

Uploaded by

Michael Kiggins
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 35

Faultfinding procedures CDAS & PFEI

Achieva
Faultfinding procedures CDAS & PFEI

DMR 173803 Rev. 03 CSIP Level 1 (15.1) 1-1


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
(c) 2015 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Faultfinding procedures CDAS & PFEI

Page intentionally left blank

1-2 CSIP Level 1 (15.1) DMR 173803 Rev. 03


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
(c) 2015 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Faultfinding procedures CDAS & PFEI

Contents
1 PREFACE ................................................................................................................................................ 4
1.1 Purpose of this manual ............................................................................................................................ 4
1.2 Faultfinding procedures ........................................................................................................................... 4
1.3 CDAS self-test ......................................................................................................................................... 4
1.4 Faultfinding philosophy ............................................................................................................................ 5
1.5 Compatibility ............................................................................................................................................ 5
1.6 Abbreviations and definitions ................................................................................................................... 6
1.7 Manual history.......................................................................................................................................... 6

2 REQUIRED LEVEL OF CDAS AND PFEI BOARDS ............................................................................. 7

3 KNOWN CDAS AND PFEI ISSUE(S) ..................................................................................................... 8


3.1 CDAS does not boot, all orange LEDs stay on ........................................................................................ 8

4 CDAS TXR BOARD TEST, RESULT: STOPPED .................................................................................. 9

5 PFEI TESTS, RESULT: ERRORS ........................................................................................................16

6 CDAS TXR BOARD TEST, RESULT: ERROR ....................................................................................20

7 CDAS RX BOARD TEST, RESULT: ERROR ......................................................................................22

8 GCI2 BOARD TEST, RESULT: ERROR ..............................................................................................25

9 SCREENSHOTS....................................................................................................................................28

10 OVERVIEW OF LEDS ...........................................................................................................................30

11 MINIMUM START-UP CONFIGURATION OF CDAS AND PFEI .........................................................34

12 VOLTAGE SPECIFICATIONS OF PFEI ...............................................................................................34

13 CDAS START-UP STEPS AND MNEMONICS IN LOG FILE ..............................................................35

14 REPLACEMENT PROCEDURES CDAS AND PFEI BOARDS ...........................................................35

DMR 173803 Rev. 03 CSIP Level 1 (15.1) 1-3


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
(c) 2015 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Faultfinding procedures CDAS & PFEI

1 PREFACE
1.1 PURPOSE OF THIS MANUAL
This document deals with what to do with the outcome of the CDAS self-test.
The following outcomes are dealt with:
 CDAS TXR board test, result: 'Stopped'
 PFEI tests, result: 'Errors'
 CDAS TXR board test, result: 'Error'
 CDAS RX board test, result: 'Error'
 GCI2 board test, result: 'Error'

In the corresponding chapters, you can find what to do when such a message appears. The information is
given in the way of faultfinding procedures.

1.2 FAULTFINDING PROCEDURES


Please follow the faultfinding procedure in the relevant chapter carefully and give your feedback if you see
possible improvements. Send your feedback to [email protected]
Always check on InCenter if you use the latest version of this document.

The answer to a question in a cell of a faultfinding procedure can be "Yes" or "No".


Proceed according to your answer.

Tip: during faultfinding, mark the "Yes" or "No" after each answered question.
In this way, you record the steps in your faultfinding process. This can be important if you consult other
engineers or have to escalate a problem to a next support level.

If you read 'problem solved' in one of the faultfinding procedures this means: the problem of that faultfinding
procedure has been solved. So, the problem with which the faultfinding procedure starts has been solved.

In the faultfinding procedures in this document, the assumption is that only one problem occurs at the same
time. If you solved one problem and another problem remains, use another relevant faultfinding procedure to
solve this problem.

1.3 CDAS SELF-TEST


The CDAS self-test performs most CDAS and all PFEI board tests.
In case the CDAS has been powered off and on, wait 2 minutes for systems < R5 and 4 minutes for systems
≥ R5 before continuing.

The following boards are tested:


 CDAS-TXR (if applicable incl. multi-nuclei TXR) (the multi-nuclei TXR board is identical to the other TXR)
 CDAS-RX (all RX boards)
 CDAS-CLK
 PFEI-CFINT
 PFEI-MRX (both if applicable)
 PFEI-QDD/QBC control board
 PFEI-TRSW
 CDAS-GRADM (if applicable)
 GCI2
 CDAS-PAT (if applicable)
 The HOST-RECON simulation test is performed.
 CBH-2 (if applicable)

1-4 CSIP Level 1 (15.1) DMR 173803 Rev. 03


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
(c) 2015 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Faultfinding procedures CDAS & PFEI

1.4 FAULTFINDING PHILOSOPHY


The faultfinding process is of major importance for corrective maintenance and could be approached as
follows. Any user should like to have system availability up to 100%, without any discontinuity or even an
interruption. With technical systems this is not possible and even a try to achieve this goal would be a major
operation and would cost too much. For this reason planned and corrective maintenance and (after sales)
service are necessities.

Faultfinding and consequently parts replacements and/or adjustment & verification, are essential elements in
this process. For optimal continuity and cost effectiveness in the operation of the system, the 80-20 rule is be
used.

In approximately 80% of the cases, the problems should be solved by field service. This should be done by
the field service engineer (FSE) himself with the help of the available service documentation.
In approximately 20% of the cases, problems are not solvable and/or cannot be allocated within a limited
period. These faults can only be allocated together with dedicated support and/or assistance of specialists.

The goal of this document is to provide faultfinding procedures for the problems that occur in 80% of the
cases. The faultfinding procedures end with the replacement of an FRU, an adjustment or the advice to
escalate the problem to a next support level.

1.5 COMPATIBILITY
This document applies to all systems with a CDAS and a PFEI,
however, not for CDAS PFEI boards with the multi-nuclei option.

Therefore, the faultfinding procedures in this document do not apply for:


 the multi nuclei receive boards in the slots 5 and 15
 the multi nuclei splitter in slot 12
 the multi nuclei coil-int board in the MMD slot of the PFEI

Screenshots of testresults are based on release R3, content for R5 is identical, only lay out is different.
For that reason, no new screenshots are added in this document.

NOTE

CBH-2 board is introduced with R5 systems as successor of next


three CDAS boards: BULK-IF, GRAD-M and PAT (3 in one).

________________

DMR 173803 Rev. 03 CSIP Level 1 (15.1) 1-5


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
(c) 2015 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Faultfinding procedures CDAS & PFEI

1.6 ABBREVIATIONS AND DEFINITIONS


Abbreviation Explanation
ADC Analog to Digital Converter (A/D)
CDAS Control and Data Acquisition System
CBH-2 CDAS Bulk HUB board (version 2)
CFINT board Control Fiber INTerface board
CLK board Clock board
DHCP server Dynamic Host Configuration Protocol
FRU Field Replaceable Unit
FSE Field Service Engineer
FTP File Transfer Protocol
GCI I/O Gradient Control Interface Input/Output board
GCI2 Gradient Control Interface (version 2)
GRADM GRADient and Magnet (control board)
HOST-RECON Host computer to Reconstructor computer
MRX Multi Receiver board
PAT Patient (board / interface)
Patsup board Patient Support board
PFEI PZN Front End Interface
PH Philips Healthcare
PMS Philips Medical Systems
PWR Power
QBC Quad Body Coil
QDD Quad Detune Driver
RX board Receive board
SBC board Single Board Computer
TFTP Trivial File Transfer Protocol
TRSW Transmit/Receive SWitch
TXR board Transmit board

1.7 MANUAL HISTORY

Date Version Name Reason of changes

2008-04-07 8.0 F. Speur, J. Moors Initial version

 Changes of the procedure in chapter 2.


2008-07-15 8.1 F. Speur, J. Moors  Adding of the chapters 9, 10, 11 and 12.
 Several minor changes.

2009-12-01 9.0 F. Speur Update of several procedures


2011-07-29 11.0 F. Speur Update of several procedures, are marked yellow
Update for R5, marked yellow
2013-11-12 13.0 F. Speur This document, DMR 173803, is successor of document
4522 981 37374
Update of replacement procedures and added known
2015-06-17 15.0 F. Speur
CDAS issue, all are marked yellow.
New version due to a publication error, no content
2015-07-29 15.1 F. Speur
changes

1-6 CSIP Level 1 (15.1) DMR 173803 Rev. 03


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
(c) 2015 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Faultfinding procedures CDAS & PFEI

2 REQUIRED LEVEL OF CDAS AND PFEI BOARDS

For certain software versions/option types, the CDAS and PFEI boards have to be of certain 12 NCs,
see table below.
For CDAS and PFEI boards not in the table below, there are no requirements.

Table 1 - Required 12 NCs for CDAS and PFEI boards


Software version / option type Board Required 12 NC of board

Software ≥ Rel.1.8 and <2.0 GRADM-V2 4522 117 9587.x


Software ≥ Rel.2.0 and <2.6.3 SBC 4522 150 3425.3 or
4522 150 4347.x
Software ≥ Rel.2.6.3 SBC 4522 150 3425.3 or
4522 150 4347.x
GRADM-V2 or 4522 117 9587.x or
CBH-2 4522 117 9878.x
Multi Nuclei option MRX-3 or 4522 137 0038.x or
MRX-4 4522 137 0039.x
CFINT-3 or 4522 117 9660.x or
CFINT-4 4522 117 9812.x
SBC 4522 150 3425.3 or
4522 150 4347.x
GRADM-V2 or 4522 117 9587.x or
CBH-2 4522 117 9878.x
32 Channel option SBC 4522 150 3425.3 or
4522 150 4347.x
GRADM-V2 or 4522 117 9587.x or
CBH-2 4522 117 9878.x

NOTE

There is a chance that the “old type” GRADM (4522-117-9410.x) will


function properly when software ≥ R2.6.3 is installed.

________________

NOTE

In case of structural communication link errors between CDAS and PFEI unit the TXR and
CFINT1/3 board can be exchanged for TXR-V2 (4522 117 9725.x) and CFINT-4 (4522 117 9812.x).
Keep in mind that also a new fiber (4522 133 0496.x) should be installed!

TXR-V2 and CFINT-4 board (with new fiber connection)


will only function when software ≥ R2.6.3 is installed
________________

DMR 173803 Rev. 03 CSIP Level 1 (15.1) 1-7


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
(c) 2015 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Faultfinding procedures CDAS & PFEI

3 KNOWN CDAS AND PFEI ISSUE(S)

3.1 CDAS DOES NOT BOOT, ALL ORANGE LEDS STAY ON


Valid for All CDAS systems with CDAS-CLOCK board 4522-117-95898.
Symptom CDAS does not boot, ALL orange LEDs of the CDAS are continuously on after the CDAS is
switched on. In a correct situation, these orange LEDs are only ON for a few seconds and
then switch OFF.
Cause Timing issue between reset- and clock pulse on the CDAS-CLOCK board 4522-117-95898.
Work around Temporary solution is to switch off the CDAS and immediately switch it on.The CDAS should
start up correctly now. If necessary repeat this a few times.

1-8 CSIP Level 1 (15.1) DMR 173803 Rev. 03


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
(c) 2015 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Faultfinding procedures CDAS & PFEI

4 CDAS TXR BOARD TEST, RESULT: STOPPED

This chapter describes the faultfinding procedure to be performed when the CDAS-TXR board test
reports 'Stopped' (Figure 1). If the test CDAS-TXR board test ONLY reports 'Error', see chapter 6.

Figure 1 – CDAS-TXR board test 'Stopped' as result of CDAS selftest

NOTE

The CDAS self-test has to be performed approximately 2 minutes for systems < R5 and
4 minutes for systems ≥ R5 after the CDAS has been powered OFF and ON.

________________

NOTE

The system has to be cabled according to the TPD and has to be powered ON completely
before performing the CDAS self-test.

________________

DMR 173803 Rev. 03 CSIP Level 1 (15.1) 1-9


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
(c) 2015 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Faultfinding procedures CDAS & PFEI

 START faultfinding procedure after the message: 'Stopped' as a result of the TXR board test. (After the CDAS selftest).

1A 1B

Command: ‘ping spectrometer - t’


from Host PC to Spectrometer
(or IP address) results in
message 'Destination host No Go to 4.2A
unreachable or Request timed
out'?

Yes

2A 2B 2C

1. Execute the following 1. Power OFF and ON the Host


commands: PC. Problem of this faultfinding
 Permproc stop recon 2. Power OFF and ON the procedure solved.
 Permproc stop scanner CDAS.
 Permproc start recon Yes
3. Wait 2 / 4 minutes. No For other problems, see other
 Perproc start scanner 4. Run the CDAS self-test. faultfinding procedures.
2. Wait 2 / 4 minutes
3. Search in the log file for the Appears the Message:
string: 'NOMEASPAR’ CDAS-TXR board test:
(See Figure 6 on p. 28). 'Stopped'?

String found? Yes: Go to 8A.

No

3A 3B

1. Power OFF and ON the Host


PC. Problem of this faultfinding
2. Wait 2 / 4 minutes. procedure solved.
3. Run the CDAS self-test. No
For other problems, see other
Appears the message: faultfinding procedures.
CDAS-TXR board test:
'Stopped'?

Yes

1 - 10 CSIP Level 1 (15.1) DMR 173803 Rev. 03


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
(c) 2015 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Faultfinding procedures CDAS & PFEI

4.1A 4B 4C

On the Host PC: 1. Swap the Ethernet cable 1. Replace the defective
the 'ACT/LNK' led flashes between Host PC and SBC, Ethernet cable.
and with the cable between the 2. Power OFF and ON the
the '10/100/1000' LED is on? reconstructor and Host PC. CDAS.
No
2. Power OFF and ON the 3. Power OFF and ON the
CDAS. Host PC.
3. Power OFF and ON the 4. Wait 2 / 4 minutes.
Host PC. 5. Run the CDAS self-test.
Yes 4. Wait 2 / 4 minutes.
5. Run the CDAS self-test. No Appears the message:
4.2A CDAS-TXR board test:
Appears the message: 'Stopped'?
On the SBC board (Figure 11 on CDAS-TXR board test:
p. 32): 'Stopped'? No: Problem solved.
the 'ACT' led flashes green
and Yes: Escalate to
the 'LINK' led is green No next support level.
and
the 'PWR' led is green?

Yes Yes

5A 5B

1. Power OFF and ON the CDAS. Replace the SBC board.


2. Are the orange LEDs on all the CDAS boards always ON?
Yes

No

5C 5D

1. Are the orange LEDs on all the CDAS boards shortly ON and then OFF Replace the CDAS Power
(± 2 sec.), supply.
and is on the Power supply the 'INPUT OK' LED green (Figure 11),
and is on the Power supply the 'FAULT' LED OFF (Figure 11)? No

Yes

6A 6B

1. Power OFF and ON the Host


PC.
2. Power OFF and ON the
CDAS. Problem of this faultfinding
3. Wait 2 / 4 minutes. procedure solved.
4. Run the CDAS self-test. No
For other problems, see other
Appears the message: faultfinding procedures.
CDAS-TXR board test:
'Stopped'?

Yes

DMR 173803 Rev. 03 CSIP Level 1 (15.1) 1 - 11


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
(c) 2015 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Faultfinding procedures CDAS & PFEI

7.1A 7B 7C

Is the software release ≤ 2? 1. Install an ‘Intel’ Ethernet card


in the Host PC.
No: Go to 8A. 2. Swap the SBC with a 12 NC
4522 150 34253 / 43471 or
higher.
3. Power OFF and ON the Host Problem of this faultfinding
Yes PC. procedure solved.
4. Power OFF and ON the
7.2A CDAS. No For other problems, see other
5. Wait 2 / 4 minutes. faultfinding procedures.
Is the Ethernet card in the Host 6. Run the CDAS self-test.
PC an ‘Intel’ card
and is the12 NC of the SBC Appears the message:
No CDAS-TXR board test:
4522 150 34253 / 43471 or
higher? 'Stopped'?

Yes Yes

8A 8B

1. Remove all CDAS boards


except:
 SBC board
 Clock board
 Pat board for systems
≤ R2.6.3 or TXR board for
systems ≥ R3.x and < R5.x
 CDAS Power supply unit Go to: 4.2A.
2. Keep the Ethernet cable to the Yes If the problem repeats: escalate it
SBC board connected. to the next support level.
3. Power ON the CDAS.
4. Wait 2 / 4 minutes.
5. Run the CDAS self-test.

Appears the message:


CDAS-TXR board test: 'Stopped'?

No

9A 9B 9C

1. Install the following boards, 1. Connect the cables to the 1. Install the RX boards.
without connecting the following boards: 2. Install the TXR board(s)
cables:  Bulk and GradM board 3. Restart PFEI.
 Bulk and GradM board or CBH-2 board 4. Restart the GCI I/O unit.
or CBH-2 board 2. Install one RX board , is slot 5. Power ON the CDAS.
2. Power ON the CDAS. independant 6. Wait 2 / 4 minutes.
3. Wait 2 / 4 minutes. 3. Restart GCI I/O unit. 7. Run the CDAS self-test.
4. Run the CDAS self-test. No 4. Power OFF and ON the No
CDAS. Appears the message:
Appears the message: 5. Wait 2 / 4 minutes. CDAS-TXR-board test:
CDAS-TXR board test: 6. Run the CDAS self-test. 'Stopped'?
'Stopped'?
Appears the message: No: Problem solved.
CDAS-TXR board test: Yes: Go to 11C.
'Stopped'?

Yes Yes

1 - 12 CSIP Level 1 (15.1) DMR 173803 Rev. 03


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
(c) 2015 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Faultfinding procedures CDAS & PFEI

Yes

10B 10C

1. Power OFF the GCI I/O unit 1. Install the RX boards.


(switch DG-X1, Figure 9) 2. Install the TXR board.
(in case of Copley 78X 3. Restart PFEI.
switch off the amplifier) 4. Power ON the CDAS.
2. Power OFF and ON the 5. Wait 2 / 4 minutes.
CDAS. 6. Run the CDAS self-test.
3. Wait 2 / 4 minutes.
4. Run the CDAS self-test. No Appears the message:
CDAS-TXR-board test:
Appears the message: 'Stopped'?
CDAS-TXR board test:
'Stopped'? No: Problem solved.
Go to GCI2 test.

Yes: Escalate to next


support level.

Yes

11A 11B

1. Find out which of the just 1. Find out which of the


added boards (Bulk, GradM connected cables of Bulk,
or CBH-2) is defect by Pat or CBH-2 board is
removing them again and defect by removing them
inserting them one by one again and inserting them
and every time run the CDAS one by one and every time
self-test. run the CDAS self-test.
2. If no defect board has been 2. Swap the defective cable. 11C
found, find out if the CDAS 3. Wait 2 / 4 minutes of
rack is defect. powering on the CDAS. 1. Find out which of the just
3. Swap the defective board, or 4. Run the CDAS self-test. added boards (RX or TXR)
the CDAS rack. is defect by removing them
4. Wait 2 / 4 minutes of Appears the message: again and inserting them
powering on the CDAS. CDAS-TXR board test: one by one and every time
5. Run the CDAS self-test. 'Stopped'? run the CDAS self-test.
2. If no defect board has been
Appears the message: No: Problem solved. found, find out if the CDAS
CDAS-TXR board test: Power ON the GCI I/O rack is defect.
'Stopped'? unit. 3. Swap the defective board or
the CDAS rack.
No: Go to 9C. 4. Wait 2 / 4 minutes after
Yes: Escalate to powering on the CDAS
next support level. 5. Run the CDAS self-test.

Appears the message:


CDAS-TXR board test:
Yes 'Stopped'?

No: Problem solved.

Yes: Escalate to
next support level.

DMR 173803 Rev. 03 CSIP Level 1 (15.1) 1 - 13


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
(c) 2015 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Faultfinding procedures CDAS & PFEI

Yes

12A 12B 12C

1. Swap the Ethernet card of 1. Power OFF the CDAS. Problem of this faultfinding
Host PC, with that of the 2. Replace the defective procedure solved.
Reconstructor. Ethernet card of the Host PC.
2. Power ON the Host PC, 3. Power ON the CDAS. 1. Install all CDAS boards.
Reconstructor and CDAS. 4. Wait 2 / 4 minutes. 2. Run the CDAS self-test.
3. Wait 2 / 4 minutes. 5. Power ON the Host PC.
4. Run the CDAS self-test. 6. Run the CDAS self-test. If necessary, see the relevant
faultfinding procedures.
Appears the message: No Appears the message: No
CDAS-TXR board test: CDAS-TXR board test:
'Stopped'? 'Stopped'?

Yes: Escalate to next support


level.

Yes

13A 13B

1. Replace SBC board of Problem of this faultfinding


CDAS. procedure solved.
2. Power ON the CDAS.
3. Power OFF and ON the Host 1. Install all CDAS boards.
PC (required getting 2. Reboot the CDAS.
connection with new SBC 3. Wait 2 / 4 minutes.
board). 4. Run the CDAS self-test.
4. Wait 2 / 4 minutes. No
5. Run the CDAS self-test. If necessary, see the relevant
faultfinding procedures.
Appears the message:
CDAS-TXR board test:
'Stopped'?

Yes

14A 14B

1. Replace the CDAS Clock Problem of this faultfinding


board. procedure solved.
2. Power ON the CDAS.
3. Wait 2 / 4 minutes. 1. Install all CDAS boards.
4. Run the CDAS self-test. 2. Power ON the CDAS.
No 3. Wait 2 / 4 minutes.
Appears the message: 4. Run the CDAS self-test.
CDAS-TXR board test:
'Stopped'? If necessary, see the relevant
faultfinding procedures.

Yes

1 - 14 CSIP Level 1 (15.1) DMR 173803 Rev. 03


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
(c) 2015 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Faultfinding procedures CDAS & PFEI

Yes

15A 15B

1. Replace the CDAS Power Problem of this faultfinding


supply unit. procedure solved.
2. Power ON the CDAS.
3. Wait 2 / 4 minutes. 1. Install all CDAS boards.
4. Run the CDAS self-test. 2. Power ON the CDAS.
No 3. Wait 2 / 4 minutes.
Appears the message: 4. Run the CDAS self-test.
CDAS-TXR board test:
'Stopped'? If necessary, see the relevant
faultfinding procedures.

Yes

16A 16B

1. Replace CDAS rack. Problem of this faultfinding


2. Power ON the CDAS. procedure solved.
3. Wait 2 / 4 minutes.
4. Run the CDAS self-test. 1. Install all CDAS boards.
2. Power ON the CDAS.
Appears the message: No 3. Wait 2 / 4 minutes.
CDAS-TXR board test: 4. Run the CDAS self-test.
'Stopped'?
If necessary, see the relevant
Yes: Escalate to next support level faultfinding procedures.

DMR 173803 Rev. 03 CSIP Level 1 (15.1) 1 - 15


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
(c) 2015 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Faultfinding procedures CDAS & PFEI

5 PFEI TESTS, RESULT: ERRORS

After the run of the CDAS self-test, the result of all PFEI tests (4 or 5x) is: 'Error' (Figure 2).

Figure 2 - CDAS Self-test menu

NOTE

The message of the CDAS TXR board test is not relevant for this faultfinding procedure.
It can be: Passed or Error.

________________

NOTE

The CDAS self-test has to be performed approximately 2 minutes for systems < R5 and
4 minutes for systems ≥ R5 after the CDAS has been powered OFF and ON.

________________

NOTE

The system has to be cabled according to the TPD and has to be powered ON completely
before performing the CDAS self-test.

________________

1 - 16 CSIP Level 1 (15.1) DMR 173803 Rev. 03


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
(c) 2015 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Faultfinding procedures CDAS & PFEI

 START faultfinding procedure after the message 'Error' as result of PFEI board tests (4 or 5x). (After CDAS self-test).

1A 1B

1. Power OFF and ON the


CDAS.
2. Wait 2 / 4 minutes.
3. Run the CDAS self-test.
Problem of this faultfinding
Appears the messages: No procedure solved.
'Error'? as result of the PFEI
boards tests (4 or 5x). (Figure 2)

Yes

2A 2B

1. Mount the wrap around fiber


cable between B1 and B2 or
B12 on the CDAS TXR
board. Go to: 4A
2. Power OFF and ON the
No
CDAS.
3. Run the CDAS-TXR fiber
connector test.
(See Figure 7 on p. 29).

Appears the message:


Fiber test: 'failed'?

Yes

3A 3B

1. Replace the TXR board.


2. Power ON the CDAS.
3. Wait 2 / 4 minutes. No Problem of this faultfinding
4. Run the CDAS self-test. procedure solved.

Appears the message:


'Error'? (4 or 5x) PFEI.

Yes

DMR 173803 Rev. 03 CSIP Level 1 (15.1) 1 - 17


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
(c) 2015 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Faultfinding procedures CDAS & PFEI

Yes

4A 4B

Test the long fiber cable:


1. Disconnect the fibers B1 and
B2 or B10 from the CF_INT
board.
2. Connect the wrap around Go to: 6A.
fiber connector to the
disconnected fibers. No
Note 4A
3. Power OFF and ON CDAS.
4. Wait 2 / 4 minutes.
The fiber test is unreliable for this
5. Run the CDAS-TXR fiber
type of fibers:
connector test.

Appears the message:


Fiber test: 'failed'? See note 4A

Yes

5A 5B If test failed check fiber with a


flashlight , if passed, fiber is ok
1. Replace the long fiber.
2. Power OFF and ON CDAS.
3. Wait 2 / 4 minutes. Problem of this faultfinding
4. Run the CDAS self-test. No procedure solved.

Appears the message: 'Error'


4 times?

Yes

6A 6B

1. Check all voltages of PFEI 1. Powersupply in system filter


unit, for specifications see box is suspected. Check
chapter 12 2. Remove PFEI board one by
No one.

Are all PFEI voltages within


specifications?

Yes

7A 7B

1. Remove all PFEI boards


except the CFINT board.
2. Power OFF and ON CDAS. One of the removed PFEI boards
3. Wait 2 / 4 minutes. is suspected (defect).
No
4. Run the CDAS self-test.

Appears the message: 'Error'


4 times?

Yes

1 - 18 CSIP Level 1 (15.1) DMR 173803 Rev. 03


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
(c) 2015 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Faultfinding procedures CDAS & PFEI

Yes

8A 8B

1. Re-install all boards.


2. Replace the CFINT board.
3. Power OFF and ON CDAS.
4. Wait 2 / 4 minutes.
5. Run the CDAS self-test. No Problem of this faultfinding
procedure solved.
Appears the message: 'Error'
4 times?

Yes: Replace the PFEI rack.

DMR 173803 Rev. 03 CSIP Level 1 (15.1) 1 - 19


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
(c) 2015 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Faultfinding procedures CDAS & PFEI

6 CDAS TXR BOARD TEST, RESULT: ERROR

After the run of the CDAS self-test, the result of the CDAS-TXR board test is: 'Error' (Figure 3).

Figure 3 – CDAS-TXR board test Error as result of CDAS selftest

NOTE

The CDAS self-test has to be performed approximately 2 minutes for systems < R5 and
4 minutes for systems ≥ R5 after the CDAS has been powered OFF and ON.

________________

NOTE

The system has to be cabled according to the TPD and has to be powered ON completely
before performing the CDAS self-test.

________________

1 - 20 CSIP Level 1 (15.1) DMR 173803 Rev. 03


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
(c) 2015 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Faultfinding procedures CDAS & PFEI

 START faultfinding procedure after the message 'Error' as a result of the CDAS TXR board test. (After CDAS self-test)

1A 1B

1. Remove all cables from the


regarding board but leave
fibers connected.
2. Power OFF and ON the
CDAS. Check removed cables
3. Wait 2 / 4 minutes.
4. Run CDAS self-test. No

Appears the message:


CDAS-TXR board test: 'Error'?

Yes

2A 2B 2C

In Test Result Details of CDAS: There is a board access


TXR board access: 'Passed'? problem. Replace the TXR board.
No Other boards have the same No
problem?

Yes Yes

3A 3B

Replace the TXR board. Replace the backplane.

DMR 173803 Rev. 03 CSIP Level 1 (15.1) 1 - 21


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
(c) 2015 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Faultfinding procedures CDAS & PFEI

7 CDAS RX BOARD TEST, RESULT: ERROR

After the run of the CDAS self-test, the result one of the CDAS-RX board tests is: 'Error' (Figure 4).

Figure 4 – CDAS-RX board 1 test Error as result of CDAS selftest

NOTE

The CDAS self-test has to be performed approximately 2 minutes for systems < R5 and
4 minutes for systems ≥ R5 after the CDAS has been powered OFF and ON.

________________

NOTE

The system has to be cabled according to the TPD and has to be powered ON completely
before performing the CDAS self-test.

________________

1 - 22 CSIP Level 1 (15.1) DMR 173803 Rev. 03


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
(c) 2015 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Faultfinding procedures CDAS & PFEI

 START of faultfinding after the message 'Error' (for CDAS RX board test) appears as a result of the CDAS self-test.

1A 1B 1C

Appears in the Test Result Are the RX boards installed in 1. Install the RX boards in the
Details of CDAS the message: the correct slots? correct slots.
RX board access: 'Passed'? 2. Power ON the CDAS.
(See Figure 8). 3. Wait 2 / 4 minutes.
4. Run the CDAS self-test.
No No
If necessary, see relevant
faultfinding procedures.

Yes Yes

2B 2C

1. Swap the suspected board 1. Replace the defective RX


with correct RX board. board.
2. Power ON the CDAS. 2. Power ON the CDAS.
3. Wait 2 / 4 minutes. 3. Wait 2 / 4 minutes.
4. Run the CDAS self-test. 4. Run the CDAS self-test.

Appears the message: If necessary, see relevant


CDAS-RX board tests: 'Error'? No faultfinding procedures.
(For the suspected RX slot).

Yes: Go to 7A.

3A 3B 3C

Are the ‘High pass band’ and Is the 12 NC of the RX-board 1. Install a correct RX board
'Low pass band' within the according to the specifications and/or change the hardware
specifications? No of the MRI system and is its No configuration setting.
hardware configuration correct? 2. Test the system.

Yes Yes

4A 4B 4C

Are the ADC ‘stop band’ and Are all other RX boards within 1. Replace the Clock board.
‘ADC analog wrap’ within specification? 2. Power ON the CDAS.
specifications? 3. Wait 2 / 4 minutes.
4. Run the CDAS self-test.
No No
Appears the message:
CDAS-RX board tests: 'Error'?
(For all RX boards).

No: Problem solved.


Yes: Go to 6A.
Yes Yes

DMR 173803 Rev. 03 CSIP Level 1 (15.1) 1 - 23


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
(c) 2015 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Faultfinding procedures CDAS & PFEI

5A 5B 5C

Result for the demodulator test 1. Swap the RX board from 1. Replace the backplane.
is: 'Passed'? another position into the 2. Power ON the CDAS.
suspected position. 3. Wait 2 / 4 minutes.
2. Power ON the CDAS. 4. Run the CDAS self-test.
3. Wait 2 / 4 minutes.
4. Run the CDAS self-test. If necessary, see relevant
No No faultfinding procedures.
Appears the message:
CDAS-RX board tests:
'Passed'?
(For the suspected position).

Yes Yes

6A 6B

1. Swap the RX board from 1. Replace the defective RX


another position into the board.
suspected slot. 2. Power ON the CDAS.
2. Power ON the CDAS. 3. Wait 2 / 4 minutes.
3. Wait 2 / 4 minutes. 4. Run the CDAS self-test.
No
4. Run the CDAS self-test.
If necessary, see relevant
Appears the message: faultfinding procedures.
CDAS-RX board tests: 'Passed'?
(For the regarding board).

Yes

7A

1. Inspect the suspected slot


(if possible repair it) or
replace the backplane.
2. Power ON the CDAS.
3. Wait 2 / 4 minutes.
4. Run the CDAS self-test.

If necessary, see relevant


faultfinding procedures.

1 - 24 CSIP Level 1 (15.1) DMR 173803 Rev. 03


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
(c) 2015 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Faultfinding procedures CDAS & PFEI

8 GCI2 BOARD TEST, RESULT: ERROR

After the run of the CDAS self-test, the result of the GCI2 board test is: 'Error' (Figure 5).

Figure 5 – GCI2 board test Error as result of CDAS selftest

NOTE

The CDAS self-test has to be performed approximately 2 minutes for systems < R5 and
4 minutes for systems ≥ R5 after the CDAS has been powered OFF and ON.

________________

NOTE

The system has to be cabled according to the TPD and has to be powered ON completely
before performing the CDAS self-test.

________________

DMR 173803 Rev. 03 CSIP Level 1 (15.1) 1 - 25


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
(c) 2015 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Faultfinding procedures CDAS & PFEI

 START of faultfinding after the message 'Error' (for GCI2 board test) appears as a result of the CDAS self-test

1A 1B

1. Power OFF the GCI I/O unit


(switch DG-X1, Figure 9).
2. Power OFF the CDAS.
3. Power ON the GCI I/O unit. Problem of this faultfinding
4. Power ON the CDAS. procedure solved.
5. Wait 2 / 4 minutes. No
6. Run the CDAS self-test.

Is the result of the GCI2 board


test: 'Error'?

Yes

2A 2B

1. Power OFF the GCI I/O unit.


2. Power OFF the CDAS.
3. Replace the optical
transceiver of the GCI I/O unit
(with the spare of the
reconstructor) Problem of this faultfinding
4. Power ON the GCI I/O unit. procedure solved.
5. Power ON the CDAS. No
6. Wait 2 / 4 minutes.
. 7. Run the CDAS self-test.

Is the result of the GCI2 board


test: 'Error'?

Yes

3A 3B

1. Power OFF the GCI I/O unit.


2. Power OFF the CDAS.
3. Replace the transceiver of the
CDAS GradM board or the
transceiver of the B2 Problem of this faultfinding
connector of the CBH-2 procedure solved.
board. No
4. Power ON the GCI I/O unit.
5. Power ON the CDAS.
6. Wait 2 / 4 minutes.
7. Run the CDAS self-test.

Is the result of the GCI2 board


test: 'Error'?

Yes

1 - 26 CSIP Level 1 (15.1) DMR 173803 Rev. 03


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
(c) 2015 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Faultfinding procedures CDAS & PFEI

4A 4B

1. Power OFF the GCI I/O unit.


2. Power OFF the CDAS.
3. Replace the fiber.
4. Power ON the GCI I/O unit. Problem of this faultfinding
5. Power ON the CDAS. procedure solved.
6. Wait 2 / 4 minutes. No
7. Run the CDAS self-test.

Is the result of the GCI2 board


test: 'Error'?

Yes

5A 5B

1. Power OFF the GCI I/O unit.


2. Power OFF the CDAS.
3. Replace the GCI2 board with
a new board.
4. Power ON the GCI I/O unit. Problem of this faultfinding
5. Power ON the CDAS. procedure solved.
6. Wait 2 / 4 minutes. No
7. Run the CDAS self-test.

Is the result of the GCI2 board


test: 'Error'?

Yes

6A 6B

1. Power OFF the GCI I/O unit.


2. Power OFF the CDAS.
3. Replace the CDAS GradM or
CBH-2 board.
4. Power ON the GCI I/O unit. Problem of this faultfinding
5. Power ON the CDAS. procedure solved.
6. Wait 2 / 4 minutes. No
7. Run the CDAS self-test.

Is the result of the GCI2 board


test: 'Error'?

Yes

7A

Escalate to next support level.

DMR 173803 Rev. 03 CSIP Level 1 (15.1) 1 - 27


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
(c) 2015 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Faultfinding procedures CDAS & PFEI

9 SCREENSHOTS

Figure 6 - Message ‘NOMEASPAR’

1 - 28 CSIP Level 1 (15.1) DMR 173803 Rev. 03


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
(c) 2015 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Faultfinding procedures CDAS & PFEI

Figure 7 – CDAS-TXR fiber con. test (I)

Figure 8 - Test result details (of CDAS self-test)

DMR 173803 Rev. 03 CSIP Level 1 (15.1) 1 - 29


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
(c) 2015 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Faultfinding procedures CDAS & PFEI

10 OVERVIEW OF LEDS

The figures below apply during stand-by of the scanner. During scanning other LEDs can lit as well.

Figure 9 – Front view gradient chain interface

+5 V GCI OK ACT
Green Green Orange: low level

+5 V PWR OK SYNC
Green Green Green

or

+5 V SYNC
Green Green

1 - 30 CSIP Level 1 (15.1) DMR 173803 Rev. 03


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
(c) 2015 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Faultfinding procedures CDAS & PFEI

300 V
Figure 10 – Front view PFEI unit Green +5 V
Green
+5 V SUPPLIES OK POWER ON
Green Green Green

Power Power
ON ON
Green Green

DMR 173803 Rev. 03 CSIP Level 1 (15.1) 1 - 31


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
(c) 2015 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Faultfinding procedures CDAS & PFEI

Figure 11 – Front view CDAS rack Achieva < R5

ACT.
Yellow flashing INPUT OK
ACT.
Yellow flashing Green
PWR
Green
FAULT
Out

LINK LINK
Green Green
LINK
Green ACT.
Green flashing

1 - 32 CSIP Level 1 (15.1) DMR 173803 Rev. 03


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
(c) 2015 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Faultfinding procedures CDAS & PFEI

Figure 12 – Front view CDAS rack Achieva ≥ R5

ACT.
Green flashing

LINK INPUT OK
Green Green
PWR
Green
LINK FAULT
Green Out

LINK LINK
Green Green

ACT.
Green flashing

DMR 173803 Rev. 03 CSIP Level 1 (15.1) 1 - 33


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
(c) 2015 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Faultfinding procedures CDAS & PFEI

11 MINIMUM START-UP CONFIGURATION OF CDAS AND PFEI


Table 2 - Boards minimally required to start-up the CDAS or PFEI
Component(s) to start Boards required to start-up

CDAS with SW ≤ R2.6.3 SBC, Clockboard, PAT board and CDAS power supply.

CDAS with SW ≥ R3.x and < R4.x SBC, Clockboard, TXR board and CDAS power supply.

CDAS with SW ≥ R4.x SBC, Clockboard and CDAS power supply.

PFEI CFINT

12 VOLTAGE SPECIFICATIONS OF PFEI


NOTE

The voltage on the PFEI must be within the allowed voltage range in standby mode
of the system, it can be measured on the service connector of the PFEI.
________________

Table 3 – Voltages of PFEI


Nominal Voltage Allowed Voltage range

-5 VA - 5.1….....-4.6 V
-5 VB - 5.1....… -4.6 V
+5 V 4.8.….… 5.4 V
-12 V - 12.5...…-11.3 V
+12 V 11.3 ..… 12.5 V
+30 V 28.4 ..… 30.5 V
+300 V 287…….310 V

1 - 34 CSIP Level 1 (15.1) DMR 173803 Rev. 03


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
(c) 2015 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Faultfinding procedures CDAS & PFEI

13 CDAS START-UP STEPS AND MNEMONICS IN LOG FILE


Table 4 – CDAS steps, actions and mnemonics in log file
Step Action of CDAS Text / mnemonics in log file
no.
1. Reading out the SBC BIOS version. AEADMN(adco_init): SBC BIOS Version: 1.00.03

2. Reading of the configuration files from CFG(Conditional): Reading configuration files


the host, using the FTP connection with
the host.
3. Starting of check of the boards, if CDAS_board(fpga_code): Board with ID: 1179533.7 in slot
necessary downloading of the correct 1 contains FPGA v1 b11
version of the FPGA code.
4. Starting all drivers, beginning with a CDAS(Device): ResetLogic.SignalResetOnJ3 performed
complete reset of all CDAS boards.
5. Starting the time critical devices. AXSTRECHERIF(stretcher): Start the stretched hardware

6. Finishing of driver installation. AXMX(log): 11:06:28.916::AXMX:axmx_init_devs() done

7. Starting of self-tests. CDAS(Device): Receiver1H.Performing self test.

8. CDAS ready for scanning. ADCO(STATE): =>S_NOMEASPAR

14 REPLACEMENT PROCEDURES CDAS AND PFEI BOARDS


After a replacement of CDAS and/or PFEI board always follow the Field Service Framework (FSF)
replacement procedure for specific board.

You can find the procedure in FSF:

1. Select: System Adjustment


2. Select: Test and Tuning procedures
3. Select: Replacement procedure
4. Select: CDAS and PFEI
5. Select: CDAS or PFEI
6. Perform procedure(s) for replaced board(s)
7. Close the doors and covers of all system cabinets

DMR 173803 Rev. 03 CSIP Level 1 (15.1) 1 - 35


This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.