Datex-Ohmeda Cardiocap 5 - Service Manual
Datex-Ohmeda Cardiocap 5 - Service Manual
DN
The Datex-Ohmeda Cardiocap/5 and accessories are indicated for indoor monitoring of hemodynamic (ECG,
impedance respiration, NIBP, temperature, SpO2, and invasive pressure), respiratory (CO2, O2, N20, respiration
rate, anesthetic agent, and agent identification), ventilatory (airway pressure, volume, and flow), and relaxation
status (NMT) of all hospital patients.
With the N-XOSAT option, monitoring of arterial oxygen saturation includes monitoring hospital patients during
conditions of clinical patient motion,
Impedance respiration measurement is indicated for patients ages 3 years and older.
CAUTION: US Federal law restricts this device to sale by or on the order of a licensed medical
practitioner. Outside the USA, check local laws for any restriction that may apply.
Classifications
IEC 60601-1:
= Type of protection against electric shock: Class | equipment.
© Degree of protection against electric shock (indicated by a symbol on the panel beside each connector):
Type BF applied part or Type CF applied part.
* The equipment is not suitable for use in the presence of a flammable anesthetic mixture with air or with
oxygen or nitrous oxide.
» Mode of operation: Continuous.
Datex-Ohmeda assumes no responsibility for the use or reliability of its software on equipment that is not
fumished by Datex-Ohmeda.
Trademarks
Datex®, Ohmeda®, and other trademarks (Cardiocap/5, AS/3, CS/3, S/5, 5/5 Light, D-lite, Pedi-lite, D-fend,
D-fend+, MemCard, ComWheel, EarSat, HexSat, OxyTip, Patient 09, and Patient Spirometry) are the property of
Instrumentanum Corp. or its subsidiaries.
АН other product and company names are the property of their respective owners.
Planned Maintenance 3
Planned Maintenance Instructions
Planned Maintenance Form
Troubleshooting 4
Messages
Troubleshooting Charts
Measurement Parameters 6
Parameter Unit (NESTPR)
Invasive Pressures and Second Temperature Option (N-XP)
Nelicor Pulse Oximetry Option (N-XNSAT)
Datex-Ohmeda Enhanced Pulse Oximetry Option (N-XOSAT)
Airway Gas Options (N-XC, N-XCO, N-XCAiO)
Patient Spirometry Option (N-XV)
NeuroMuscular Transmission Option (N-XNMT)
Service Procedures 7
Repair and Replacement
Checks, Adjustments, and Calibration
Service Menus
Spare Parts 9
Cardiocap/5 Technical Reference Manual
Autoclaving,sterilizing,andeleanin8.............................
cno norris
SEIVICE...ui. ieri
Batteries... inner tres
Special components and modifications
Storage and transport.
DiSPpOS8| .es ee
Table of Figures
Figure 1-1. Cardiocap/5 monitor structure eee nenene eee eeeeee nenene nenene 1-3
Overview
1. OVERVIEW
Read the entire manual and make sure you understand the procedures described before installing,
repairing, or adjusting the monitor. To avoid risks concerning safety or health, strictly observe all safety
precautions.
Related documentation
For information about using the monitor, refer to the following:
Cardiocap/5 User's Guide, Anesthesia
Cardiocap/5 User’s Guide, Critical Care
Cardiocap/5 User’s Reference Manual, Anesthesia
Cardiocap/5 User's Reference Manual, Critical Care
For PCA drawings, circuit diagrams, and component lists, order the PCA Drawings Service Kit. See the
Spare Parts chapter.
1-1
Cardiocap/5 Technical Reference Manual
e All measurement parameter options (and the Recorder option, N-XREC) are factory-configured
and cannot be added after purchase.
= Data collection and management options (N-XNET and N-XDNET) can be added later.
1.2.2 Options for hemodynamic model with airway gas measurement (F-MXG)
The F-MXG measures NIBP, ECG (3-lead and 5-lead), pulse oximetry (Sp05), temperature (T1),
impedance respiration, and airway gases. Gas measurement depends on which airway gas option is
installed (N-XC, N-XCO, or N-XCAiO):
N-XC Carbon Dioxide (CO)
N-XCO CO», N°0, and Patient Oxygen (05)
N-XCAIO CO», anesthetic agents, agent identification, No0, and 0,
The F-MXG can also be equipped with each of these built-in options:
N-XP Two invasive pressure channels and second temperature (T2)
N-XV Patient Spirometry (N-XCO or N-XCAiO option required)
N-XREC Recorder
The F-MXG can also be configured with one of the following options:
N-XNSAT Nellcor® compatible pulse oximetry (SpO.)
N-XOSAT Datex-Ohmeda enhanced pulse oximetry (Sp05)
N-XNMT NeuroMuscular Transmission (NMT) for relaxation measurement (N-XCAIO option required)
1.2.3 Data collection and management options (for F-MX and F-MXG)
For both models, these options can be factory-built or added later as upgrades:
N-XNET Network
N-XDNET Data card and Network
1-2
Overview
FT|
Patient
connectors 7 NESTPR | cw) [Keyboard] [Recorder DC/ pc ac/0c je
ECG > ECG board
e 7
RIA. PCMCIA
=
n + Net code
NIBP トー NIBP board ト [Battery] | LCD
display 1/0 | Serial out
Patient =
Spirometry pb 10.4” board
4 Analog out
Gases P _
CAO
LT _ ___., ¡Mother [Inverter board)
N-XOSAT >| }—{OSAT board I/F board "| Keyboard|
NANSAT pe NSAN board 00 | Baekight board |
Nur bol NM board|
Figure 1-1. Cardiocap/5 monitor structure
The main software and measurement technologies are based on AS/3 hardware and software. Some
parameter-measuring unit boards are interchangeable with AS/3 module boards, however, the units
can not be replaced with the corresponding modules as the hardware of the assemblies is different.
The NESTPR, CAiO, and NMT units are connected to the CPU through the Mother board and
communicate with the CPU over a standard AS/3 module bus. The NSAT or OSAT pulse oximetry unit is
also connected to the CPU through the Mother board and communicates with the CPU over a standard
AS/3 module bus, which is located on the SpO, interface board.
Each parameter measurement board contains a CPU that processes measurement data for the
parameter(s) associated with that board before sending the data to the main CPU.
NESTPR unit
The NESTPR unit contains three boards for measuring hemodynamic parameters:
* The ECG board measures ECG (3-lead and 5-lead) and impedance respiration.
= The STP board measures oxygen saturation, temperature, and invasive blood pressure.
1-3
Cardiocap/5 Technica! Reference Manual
1.3.2 Communication
The CPU communicates with the hemodynamic parameters measuring unit (NESTPR) and airway gas
measuring unit (CAiO) over a standard AS/3 module bus. It is based on the widely-used industry
standard RS485, which uses a differentia! serial method to transfer data and is quite robust.
RS485 serial communication supports multidrop or party line connections. This means all units
connected to the module bus use the same two physical wires for communication purposes. The
module bus uses a 500 kbps data transfer rate.
Communication with the 1/0 board and the DC/DC board takes place over an internal synchronous
serial bus. The same bus also controls display brightness and audio signals by means of D/A
converters located on the CPU board.
Communication with the Net takes place overa separate synchronous serial channel.
Communication with the Recorder takes place over an asynchronous serial channel.
* Controls the brightness of the LCD screen by means of the Backlight board.
* Controls the Inverter board that provides the high voltage for the display backlights.
Two PCMCIA-compatible data card slots on the board are for loading software and transferring data.
1-4
Overview
Distributed processing
The parameter and airway measuring units contain their own microprocessor systems for performing
low level functions, such as waveform filtering and pneumatics control. At the same time, the main
CPU performs higher level tasks (trending and alarm control, for example).
1.3.4 Display
The main CPU directly controls the monitor display, a 10.4 inch color LCD. Supply voltages for the
display are connected via the CPU board and the Backlight board. The inverter board provides high
voltage for the display backlights. The CPU controls the display brightness by adjusting the backlight
voltage.
1-5
Cardiocap/5 Technical Reference Manual
= When displayed beside the 0, value, indicates that the FiO, low-alarm limit is set
below 21%.
* When displayed next to the HR value, indicates that there is a risk that the monitor
counts pacemaker spikes (pacer is set ON R) or the monitor counts T-waves (a
wide QRS is selected).
Type BF applied part (IEC 60601-1). Defibrillator-proof protection against electric shock.
回
Type CF applied part (IEC 60601-1). Defibrillator-proof protection against electric shock.
ーー
=
Main Menu. Located beside the ComWheel to indicate you can open the Main Menu
by pressing the ComWheel when no other menu is displayed.
Power On/Standby.
This battery contains lead. Separate from other waste for disposal according to local
regulations.
Dangerous voltage.
Ethernet connectors.
Alternating current.
Fuse.
1-6
Overview
Symbols on screens
When displayed on the upper left comer of the screen, indicates alarms are silenced.
When in the menu or digit fields, indicates that the alarm source has been turned off.
Sub menu. Selecting an alternative with this symbol in a menu opens a new menu.
y The contents of the package are fragile and have to be handled with care.
И
x
Indicates the correct upright position of the transport package.
1-7
Cardiocap/5 Technical Reference Manual
1.5.1 Warnings
Refer to the User's Reference Manual for additional wamings to be observed while monitoring a patient.
A WARNING indicates a situation in which the user or the patient may be in danger of injury or
death.
Power connection
Before connecting the power cord to the mains outlet, check that the local voltage and frequency
rating corresponds with the rating stated on the device plate on the rear panel of the monitor.
Connect the monitor to a three-wire, grounded, hospital-grade receptacle. Do not remove the
grounding prong from the power plug.
Use an intact power cord. Replace the cord if it is cracked, frayed, broken, or damaged.
Do not apply tension to the power cord. The cord may break.
Do not use extension cords or adapters.
External connection
Connect the monitor only to other monitors or external devices specified by Datex-Ohmeda.
Always perform an electrical safety check and leakage current test of the monitor after service.
Fuse replacement
Replace the fuse with a fuse of the same type and with the same rating.
Explosion hazard
Do not use the monitor in the presence of flammable anesthetics.
Patient safety
Do not test or perform maintenance on the monitor while it is being used on a patient.
Use only cables and accessories approved by Datex-Ohmeda. Do not modify them. Other cables and
accessories may damage the monitor or interfere with measurement.
PACEMAKER PATIENTS: The impedance respiration measurement may cause rate changes in Minute
Ventilation Rate Responsive Pacemakers. Set the pacemaker rate responsive mode off or turn the
impedance respiration measurement off on the monitor.
Temperature probes
To prevent injury, use Datex-Ohmeda temperature probes only.
1-8
Overview
Turn off the power and unplug the power cord before cleaning or service. Make sure the monitor is
completely dry before reconnecting it to the mains outlet.
1.5.2 Cautions
Refer to the User's Reference Manual for additional cautions to be observed while monitoring a patient.
General
US Federal law restricts this device to sale by or on the order of a licensed medical practitioner.
Do not apply presurized airto any outlet or tubing connected to the monitor. Pressure may destroy
sensitive elements.
Use only cables and accessories approved by Datex-Ohmeda. Other cables and accessories may
damage the system or interfere with measurement.
Vibrations during transport may disturb SpO ECG, impedance respiration, and NIBP measurements.
Installation
Leave space behind the monitor to allow proper ventilation.
Before use
Allow two minutes for warm-up and note any error messages or deviations from normal operation.
Do not immerse any part of the monitor in liquid or allow liquid to enter the interior.
Clean the fan dust filter on the rear panel once a month or whenever necessary.
Service
Electrostatic discharge through the PC boards may damage the components. Before handling printed
circuit boards, wear a Static control wrist strap. Handle all boards by their nonconductive edges and
use antistatic containers when transporting them.
Do not break or bypass the patient isolation barrier when testing PC boards.
1-9
Cardiocap/5 Technical Reference Manual
Batteries
There is a lithium battery on the CPU board. Discard broken IC containing the battery according
to
local regulations.
The battery package of the power supply unit in this device contains lead, which is hazardous to the
environment. Dispose of the battery according to local regulations.
To replace the batteries safely, please refer to the instructions in this manual.
* Do not short-circuit the battery terminals. Short-circuiting the battery may produce a very high
current, which damages the battery and may cause injury to personnel.
» Do not dispose of the battery into open flame, nor put the battery near fire, as it may explode.
e Do not disassemble the battery. It contains electrolyte, which may damage clothing or cause
injury to skin or eyes. If exposed to electrolyte, wash the injured area with plenty of water and
contact a doctor.
The manufacturer accepts no responsibility for modifications made to the monitor outside the factory.
Disposal
Dispose of the device and its parts according to local environmental and waste disposal regulations.
1-10
Overview
1.6 Specifications
All product specifications are subject to change without prior notice.
Power supply
Rated voltages and frequencies: 100 to 240 VAC 60/50 Hz
Ailowed voltage fluctuations: + 10%
Maximum power consumption: 80 VA
Fuses (2): T2AH/250V
Environmental conditions
Operating temperature: +10 to +40 °C (50 to 104 °F}
Storage and transport temp: -10 to +50 °C (14 to 122 °F)
Relative humidity: 10 to 85 % noncondensing, in airway 0 to 100 % condensing
Atmospheric pressure: 660 to 1060 hPa (500 to 800 mmHg)
Mechanics
Dimension: 330 mm x 220 mm x 300 mm (width x depth x height)
Weight: <11.2 kg/ <24.8 lbs (F-MXG with all options); <10.2 kg/ <22.6 lbs (F-MX with all options)
LCD display
Display size: 10.4 inch
Display type: Áctive matrix color LCD display
Resolution: 640 x 480
Battery
Type: 12V 2.6AH, lead acid
Back-up battery time: at least 15 minutes when fully charged
Charging time: 5 hours (typical)
The green battery charge status LED is On when the battery is fully charged, on the holding voltage. The
LED flashes when the battery is being charged.
1.6.2 NIBP
Measurement range:
adult 25 to 260 mmHg
child 25 to 195 mmHg
infant 15 to 145 mmHg
Pulse rate range accepted: 30 to 250 bpm
Typical measuring time: adults 23 seconds, infants 20 seconds
1.6.3 Temperature
Measurement range: 10 to 45°C (50 to 113°F)
Measurement accuracy: 25 to 45.0 °C 4.0.1 °C (77 to 113 °F 40.2 OR
10 to 24.9 °C + 0.2 °C (50 to 76.8 °F+ 0.4 °F)
Probe type: Compatible with Datex-Ohmeda temperature probes only
1-11
Cardiocap/5 Technical Reference Manual
1.6.4 ECG
Waveform display (with 50 Hz power supply frequency):
Monitoring filter: 0.5 to 30 Hz
ST filter: 0.05 to 30Hz
Diagnostic filter: 0.05 to 100 Hz
Waveform display (with 60 Hz power supply frequency):
Monitoring filter: 0.5 to 40 H
ST filter: 0.05 to 40 Hz
Diagnostic filter: 0.05 to 100 Hz
Heart rate
Measurement range: 30 to 250 bpm
Measurement accuracy: + 5% or + 5
Pacemaker pulse detection level: 2 to 500 mV
Pacemaker pulse duration: 0.5 to 2 ms
5002
Calibration range: 50 to 100%
Calibrated against functional saturation
Measurement range: 40 to 100%
Measurement accuracy (% SpO, +1 SD):
80 to 100% + 2 digits:
50 to 80% + 3 digits;
Below 50% unspecified
NOTE: SpO, measurement accuracy is based on deep hypoxia studies using Datex-Ohmeda FingerSat
sensors on volunteered subjects. Arterial blood samples were analyzed by a Radiometer OSM
CO-oximeter. Refer to the sensor instructions for specific SpO, accuracy data.
Pulse rate
Measurement range: 30 to 250 bpm
Measurement accuracy: + 5% or + 5 bpm
1-12
Overview
Sp02
Calibration range: 70 to 100%
Calibrated against functional saturation
Measurement range: 1 to 100%
Measurement accuracy (% SpO, +1 SD):
70 to 100% + 2 digits
70 to 100% + 3 digits during conditions of clinical patient motion
Below 70% unspecified
NOTE: SpO, measurement accuracy is statistically derived and correlated to simultaneous arterial
blood gases measured on a Radiometer OSM3 CO-oximeter. Refer to the sensor instructions for
specific accuracy data.
Pulse rate
Measurement range: 30 to 250 bpm
Measurement accuracy: + 2% or + 2 bpm (whichever is greater)
5002
Calibrated against functional saturation
Measurement range: 1 to 100%
Measurement accuracy (% SpO, +1 SD):
{Q to 100% (+ 2 digits to + 3.5 digits, depending on the sensor)
Below 70% unspecified
See the User's Reference Manual (Pulse Oximetry chapter) for a list of approved sensors and
accuracy details.
NOTE: SpO, measurement accuracy is based on testing healthy adult volunteers in induced hypoxia
studies.
1-13
Cardiocap/5 Technical Reference Manual
Pulse rate
Measurement range: 30 to 250 bpm
Measurement accuracy: + 3 digits
Pulse rate
Measurement range: 30 to 250 bpm
Accuracy: +5% or+ 5 bpm
General
Airway humidity: 0 to 100%, condensing
Sampling rate: 200 + 20 ml/min. (sampling line 2-3 m, normal conditions)
Sampling delay: 2.5 seconds typical with a 3 m sampling line
Total system response time: 2.9 seconds typical with a 3 m sampling line, including sampling delay
and rise time
Display update rate: breath-by-breath
Automatic compensation for pressure, CO,-N,0 and CO,-0, collision broadening effect
Warm-up time:
2 minutes for operation with CO,, 0,, and N,0
5 minutes for operation of anesthetic agents
30 minutes for full specifications
Autozeroing interval:
immediately after “Calibrating gas sensor” message and
2,5, 10, 15, 30, 45, 60 minutes after start-up, then every 60 minutes
1-14
Overview
NOTE:
* If CO, concentration is below 0.1%, 0.0% is displayed.
Agent identification
Identification threshold: 0.15 vol% typically
Identification time: < 20 seconds (for pure agents)
Mixture identification threshold for 2nd agent: 0.2 vol% +10% of total conc.
MAC
Range: 0 to 9.9 MAC
Equation:
“WETAA) %ETNXA
MAC(AA) = SABIAM), RETRO
x(AA) 100
where x(AA): Hal = 0.75%, Enf = 1.7%, Iso = 1.15%, Sev = 2.05%, Des = 6.0%.
Cardiocap/5 Technical Reference Manual
Normal conditions
After 30-minute warm-up period:
Ambient temperature: 18 to 28 °C, within + 5 °C of calibration
Ambient pressure: 500 to 800 mmHg, + 50 mmHg of calibration
Ambient humidity: 20 to 80% RH, + 20% RH of calibration
Non-disturbing gases:
Ethanol C,H,OH (< 0.3%)
Acetone (< 0.1%)
Methane CH, (< 0.2%}
Nitrogen N,
Carbon monoxide CO
Nitric oxide NO (< 200 ppm)
Water vapor
Maximum effect on readings:
CO, < 0.2 vol%
0,,N,0 < 2 vol%
Anesthetic agents < 0.15 vol%
Effect of helium: decreases CO, readings < 0.6 vol% typically
© During warm-up, 2 to 10 minutes (anesthetic agents 5-10 minutes) under normal conditions
0, + (2 vol% + 2% of reading)
Agents (Des, Enf, Hal, Iso, Sev) + (0.2 vol% + 10% of reading)
0, + (3 vol% + 3% of reading)
1-16
Overview
Agents (Des, Enf, Hal, Iso, Sev) | + (0.2 vol% + 10% of reading)
Tidal volume
Measurement range 150 to 2000 ml 15 to 300 mi
Resolution 1 πι 1 ml
Accuracy (typical value) + 6% or 30 ml + 6% or4 ml
Minute volume
Measurement range 2 to 20 1/minute 0.5 to 5 I/minute
Resolution 0.1 /minute 0.1 /minute
Accuracy (typical value} + 6% + 6%
Airway pressures
Measurement range -20 to +100 cmH,0 -20 to +100 cmH,O
Resolution 0.5 cmH,0 0.5 cmH,0
Accuracy (typical value) + 1 cmH,0 not applicable
Airway flow
Measurement range 1.5 to 100 I/minute 0.25 to 25 I/minute
Compliance
Measurement range 4 to 100 ml/cmH,0 4 to 100 mi/cmH,0
Resolution 1 mi/cmH,0 0.1 mi/cmH,0
Airway resistance
Measurement range O to 40 cm H,0/1/second O to 40 cm H,0/l/second
Resolution 1 cmH,0/ l/s 1 cmH,0/1/s
| Sensor specifications
Dead space 9.5 mi 2.5 ml
Resistance at 30 1/minute 0.5 cmH,0 not applicable
Resistance at 10 I/minute not applicable 1.0 cmH,0
1-17
Cardiocap/5 Technical Reference Manual
Stimulator
Stimulus pulse: Square wave, constant current
Pulse width: 100, 200 or 300 us
Stimulus current range (supramax and manual): 10 to 70 mA with 5 mA steps
Stimulus current accuracy: 10% or +3 mA (whichever is greater)
Maximum load: 3 kQ
Maximum voltage: 300 V
1-18
Contents
Table of Figures
Figure 2-1. Cardiocap7/5 monitor (F-MXG) iii rene 2-1
Figure 2-2. Patientconnections—Hemodynamic(F-MX)..............................1. 2-2
Figure 2-3. Patient connections—-Hemodynamic with gas (F-MXG) rece 2-3
Figure 2-4.Rearpahel(FMXOG》 eee eee eee 2-4
Figure 2-5. Scavenging through ventilator reservoir eee einen 2-5
Figure 2-6. Gas return to patient circuit in AS/3 ADU ee teen ooo nee 2-6
Figure 2-7. Connection directly to a scavenging system... es 2-6
Installation and Functional Check
2.1 Introduction
This chapter includes the information needed to install and check the monitor. Information for
connecting other equipment, such as a printer or computer, is also included, If you need assistance
concerning the installation, please contact your authorized distributor.
2.2 Installation
(4) _ Insertion slots for memory cards (Data card and Menu card)
A cover for the slots is available. See the Spare Parts chapter later in this manual.
(7) ComWheel
2-1
Cardiocap/5 Technical Reference Manual
( 9) Patient connectors
OJE 5)
NIBP {7} ECG №) $002 -{w}--p1- Щи
b 一
SEE) 4 一
D © © © 000
Figure 2-2. Patient connections—Hemodynamic (F-MX)
(1) MBP
(2) ECG
(3) Spo,
NOTE: Connector type depends or
on which SPO; option is installed:
spo2 14)
Connector for Datex-Ohmeda
ο ο9 enhanced pulse oximetry
Spo, _ (0
| : | Connector
for Nellcor®
| | compatible pulse oximetry
(7) Temperature, T1
2-2
installation and Functional Check
rari
> =D OR
P
Spiromety TİP NMI] it} 二 一
©
Saa oo so
= © ©
Figure 2-3. Patient connections—Hemodynamic with gas (F-MXG)
( 1) Spirometry
(N-XV option)
( 2) NMT(N-XNMT option)
—
( 5) Temperature, T1
)
NOTE:= Connectortype depends on which 1500 option is installed:
sor о σος.
25 Connector for Datex-Ohmeda
sas enhanced pulse oximetry
me BD ον
| | Connector
for Nellcor®
| compatible pulse oximetry
(8) ECG
(9) NIBP
(10) D-fend housing
2-3
Cardiocap/5 Technical Reference Manual
WARNING: Electrical shock hazard. Connect the power cord to a three-wire, grounded, hospital-
grade receptacle only.
CAUTION: Turn off the power before making any rear panel connections.
2-4
Installation and Functional Check
On the rear panel, connect one of the RJ-45 connectors to connector X4 and connect the
> identification Plug to connector
X3.
3. Connect the other RJ-45 connectorto the corresponding Datex-Ohmeda Central Network
connector on the wall box.
4. Poweron the monitor.
5. Confirm that the network symbol and the “Connected to Network” message are displayed on the
upper part of the screen.
παπα,
NOTE: The network symbol does not display if the battery symbol is displayed.
CAUTION: Strong scavenging suction may change the operating pressure of the monitor and cause
inaccurate readings or internal damage.
|
Sample Gas Out
* Connect an exhaust line to the sample gas outlet on the rear panel of the monitor.
e Attach the other end of the line to the ventilator reservoir. Make sure that the reservoir tube
diameter is at least 2-3 times larger than the exhaust line.
2-5
Cardiocap/5 Technical Reference Manual
if you use the Datex-Ohmeda AS/3 Anesthesia Delivery Unit (ADU), connect an optional adapter
(881644, 5/pkg) to the patient breathing tubes.
Connect the sample gas outlet only to an open scavenging system where gas is removed at room
pressure,
CAUTION: Strong scavenging suction may change the operating pressure of the monitor and cause
inaccurate readings or internal damage.
… AE laET
и Sample Gas Out
/
* Connect the exhaust line (733195, 5/pkg) to the sample gas outlet on the monitor.
* Attach a T-fitting (882377, 5/pkg) between the exhaust line and the scavenging tube to prevent
vacuum in the tube.
2-6
Installation and Functional Check
2.3 Interfacing
WARNING: Electrical shock hazard. When you connect equipment to the Cardiocap/5 input and
output connectors, you are configuring a medical system and are responsible for ensuring that
the system complies with IEC/EN 60601-1-1 and with local requirements.
WARNING: Connecting the power supply cord of the printer to the wall socket may cause the
printer leakage current to exceed the limit specified for medical equipment. Always connect the
printer to an appropriate isolation transformer.
WARNING: Connecting the power supply cord of the computer to the wall socket may cause the
computer leakage current to exceed the limit specified for medical equipment. Always connect
the computer to an appropriate separating transformer.
2.3.3 interfacing other devices using the analog/ digital output connector
The analog/ digital output connector (X1) can be used to interface other devices to the Cardiocap/5.
Digital outputs
Defibrillation sync
The ECG generates the defibrillation sync digital output signal. The signal is set to a high state
(logic 1: 2.8 ta 5.0 VDC ) when activated. After 10 ms, it is set back to a low state
(logic O: 0 to 0.8 VDC). The signal is regenerated only after returning to the low state. The delay from
the R-wave peak to the start of the signal is 35 ms maximum.
Nurse call
The nurse call digital output signal is generated by red, yellow, and white alarms. When activated, the
signal is set to a high state and remains at a high state until the alarm situation is over or the Silence
Alarms key is pressed. The high state (logic 1) ranges from 2.8 to 5.0 VDC while the low state (logic 0)
ranges from 0 to 0.8 VDC. The nurse call signal also activates a relay that connects X1 (pins 11 and 12).
2-7
Cardiocap/5 Technical Reference Manual
Analog outputs
Each analog output signal is scaled linearly between -5 and +5 volts. The resolution is 4096 steps
over 10 volts, or approximately 0.00244 volts per step. All signal levels are updated every 10 ms.
Volume The original scale of -2.5 liters to +2.5 liters is scaled between -5 and +5 volts. (Airway gas
sensor failure is applied. See Special indications for analog outputs below.)
Resp The original scale of - 5000 mohms is scaled between -5 and +5 volts.
RR The respiration rate display ofO to 150 breaths per minute is scaled between 0 and +1.5 volts.
T1,72 The original temperature scale of 0 °C to 50 °C is scaled between O and +5 volts.
TEST SIGNALS -5V, OV, +5V
Steady signals with one of the listed values.
TEST 1. Test signal of a triangle shape with a base width of 1 second (0 V minimum; +5 V maximum).
TEST2 Test signal of a triangle shape with a base width of 4 seconds (-5 V minimum; +5 V maximum).
2-8
Installation and Functional Check
Airway gas calibration: During calibration of any gases, a square wave is generated. The minimum
value is 0 volts, the maximum value is +2 volts, the minimum phase length equals the maximum
phase length, and the frequency is 0.25 Hz.
Airway gas zeroing: During zeroing of any gases, a square wave is generated. The minumum value is
0 volts, the maximum value is +5 volts, the minimum phase length equals the maximum phase length,
and the frequency is 0.25 Hz.
Airway gas occlusion: During occlusion of any gases, a triangle-shaped signal is generated with a
base width of4 seconds, a minimum value of 0 volts, and a maximum value of +5 volts.
Airway gas airleak: During an airleak of any gases, a triangle-shaped signal is generated with a base
width of 3 seconds, a minimum value of O volts, and a maximum value of +5 volts.
Airway gas sensor failure: During a sensor failure, a triangle-shaped signal is generated with a base
width of 2 seconds, a minimum value of 0 volts, and a maximum value of +5 volts.
Select Installation.
Select Analog Outputs and set the channels.
fF
2-9
Cardiocap/5 Technical Reference Manual
2-10
Installation and Functional Check
2-12
Installation and Functional Check
The first part of the instructions contains procedures that are performed through the service menu.
These checks are especially recommended if the monitor has been stored and not used for a long
time.
When installing a new monitor, performance checks with a simulator and accessories are typically
enough to ensure the correct function of the monitor. We recommend that you always perform the
complete functional check to confirm that no hardware failures occurred during transport.
The instructions are for the maximum functional configuration of the Cardiocap/5. Perform the
procedures in ascending order and skip items that do not correspond to your monitor configuration.
Recommended tools
Tool | ο... i J Order Number“
Sp0, finger sensor for the pulse oximetry option.
NOTE: With the N-XNSAT option, use only Nellcor sensors (DS- 100A, for example).
With the N-XOSAT option, use only Datex-Ohmeda sensors (OXY-F 1-H, for example).
For standard pulse oximetry, use only Datex-Ohmeda sensors ( OXY-F4-N, for example).
Simulator capable of simulating ECG, RESP, and invBP Obtain locally
Adult NIBP cuff 572435
Adult NIBP hose 877235
Temperature test plug set 884515
MemCard - Menu, English (only for monitors with Data card option) 893860
MemCard - Data, English (only for monitors with Data card option) 887045
Sampling line 3.0 m (only for monitors with gas option) 73319
Н the NMT option (N-XNMT) is installed, you will need these additional items:
2-13
Cardiocap/5 Technical Reference Manual
Check that the green external power indicator/battery charge status LED turns on or starts
flashing.
BS
Switch the monitor on. Check that the monitor starts up as follows:
s If the monitor contains a recorder, start-up information is printed. Verify that the time and
date on the printout are correct.
BS
Check the loudspeaker by adjusting the alarm volume in the Alarms Setup menu:
Press the ComWheel to enter the Main Menu and select:
Alarms Setup
Alarm Volume
Test the whole volume scale from 1 to 10 by turning the ComWheel and check that the alarm
volume changes correspondingly. The alarm sound should be clear and audible at all settings.
Select Main Menu.
BS
Check that the time and date are correct and adjust, if necessary:
Press the ComWheel to enter the Main Menu and select:
Monitor Setup
Time and Date
NOTE: To prevent the loss of trend data, you cannot change the time and date after starting a new
case or admitting a new patient.
NOTE: If the clock shows a time of 0:00 continuously after start-up, the SRAM/Timekeeper battery
on the CPU board needs to be replaced.
BS
2-14
Installation and Functional Check
5. Check that the supply voltages are within the given limit values:
Press the ComWheel to enter the Main Menu and select:
Monitor Setup
Install/ Service (Password 16-4-34)
Service View (Password 26-23-8)
Monitor
Voltages
VDD/TEMP OK 0
CHG ON 1 0
BS
6. Check the back-up battery:
Disconnect the power cord (without switching the monitorto standby).
Check that the monitor continues to run normally on battery power. The battery indicator should
appear in the upper right corner of the screen:
Reconnect the power cord and check that during charging, the charging symbol is displayed and
the battery charge status LED starts flashing:
Charge
E
2-15
Cardiocap/5 Technical Reference Manual
BS
To record waveforms, press the ComWheel to enter the Main Menu and select:
Record/Print
Record Waveforms
Record Wave
Press the Record Waveform/ Stop key and verify that the selected waveforms print.
Press the key again to stop recording.
Press the Record Trend/Stop key and verify that the selected trends print.
Press the key again to stop recording.
BS
Memory card (PCMCIA) test (if N-XDNET option is included)
9. Press the Com Wheel to enter the Main Menu and select:
Monitor Setup
Instali/ Service (Password 16-4-34)
Service View (Password 26-23-8)
Modules
Memory Module
Check that the module is recognized (YES) and that the memory on the memory board and the
PCMCIA controller passed the tests (OK):
BS
10. Select Communication and check that Interface status states Active continuously and the error
counter values on the bottom part of the menu are stable.
BS
11. Select Module Status and insert a memory card labeled “Menu” in the memory card slot.
Check that within one minute:
* The white Menu card symbol appears on the upper right corner of the screen if the battery is
not charging (the battery charging symbol is not present).
Wait until the information regarding the memory card slot is fully updated in the service menu,
then check that the screen shows the Card type is MENU and the File system is ATA. Check that
the rest of the information for the memory card slot is reliable and no errors have been detected.
2-16
Installation and Functional Check
Repeat the test using a Data card in the other memory card slot. Verify that the green Data card
symbol is displayed if the battery charging symbol is not present, the “Data card inserted”
message appears, and the Card type is DATA.).
ES
Network test (if N-XNET or N-XDNET option is included)
12. Connect the Monitor-Network cable and the Identification plug to the monitor. To check the
connection to the network, check the Network connection LEDs between the connectors:
BS
13. Check that the Datex-Ohmeda Network symbol EO. is displayed on the upper right
comer of the screen).
NOTE: If the battery is being charged, the battery charging symbol is displayed instead of the
network symbol.
A message regarding the connection to the Datex-Ohmeda Information Center should appear in
the message field on the screen.
BS
14. To enter the Communication service menu and check Network information, press the ComWheel
to enter the Main Menu and select:
Monitor Setup
Install/ Service (Password 16-4-34)
Service View (Password 26-23-8)
Monitor
Communication
Network
Check that:
* The Location ID number matches the ID plug connected to X3.
e The packets and bytes IN is increasing slowly.
* The packets and bytes OUT is increasing quickly.
» Connections shows the names of the connected networks.
» The data error counters (CRC, Frame, and Transm.) are stable.
NOTE: The counters may show values greater than 0. However, any values increasing continuously
indicate a problem.
ES
15. Check that the hardware error counters (Intern., Missed, FIFO, and Overrun) ail show O (zero). A
value greater than zero for any counter indicates a problem on the CPU board.
BS
2-17
Cardiocap/5 Technical Reference Manual
To return to the Service View menu, select Previous Menu as needed until you can select
Modules and then select ESTP : ECG.
or
* The Timeouts, Bad checksums, and Bad c-s by mod values are not increasing faster than 50
per second.
= The ECG/RESP board memories passed the internal memory test (RAM, ROM, and EEPROM
all show OK).
BS
17. Check that the Power Freq value (the module mains power frequency) has been set according to
the supply frequency. If necessary, change the setting by selecting Power Freq.
BS
18. Check that Resp Available and Resp measurement both show ON.
NOTE: The Resp measurement shows OFF if Resp is not selected to the screen setup.
To set up the screen, press the ComWheel and select:
Monitor Setup
Screen Setup
Waveform Fields or Digit Fields
BS
STP board test
19. To enter the ESTP: STP service menu, press the ComWheel and select:
Monitor Setup
Install/ Service (Password 16-4-34) -
Service View (Password 26-23-8)
Modules
ESTP: STP
Check that the Timeouts, Bad checksums, and Bad c-s by mod values are not increasing faster
than 50 per second.
Check that the STP board memories passed the internal memory test (RAM, ROM, and EEPROM
all show OK).
BS
2-18
Installation and Functional Check
BS
NIBP board test
21. Press Previous Menu as needed to return to the Modules menu, then select NIBP to enter the
NIBP service menu.
Check that the Timeouts, Bad checksums, and Bad c-s by mod values are not increasing faster
than 50 per second.
Check that the NIBP board memories passed the internal memory test (RAM, ROM, and EEPROM
all show OK).
BS
Pulse oximetry test (if N-XOSAT or N-XNSAT option is included)
22. Configure the monitor screen to display the Pleth waveform and other puise oximetry parameters
by pressing the ComWheel and selecting:
Monitor Setup
Screen Setup
Waveform Fields
Select a field (Field1, for example) and choose PLETH as the parameter.
Connect the appropriate SpO, finger sensor to the monitor (use a Nellcor sensor for the N-XNSAT
option; a Datex-Ohmeda sensor for the N-XOSAT option). Attach the sensor to your finger.
Check that an acceptable SpO, reading and a proper SpO, waveform appear on the screen.
AS
23. Enter the SpO, options service menu by pressing the ComWheel and selecting:
Monitor Setup
Install/Service (Password 16-4-34) -
Service View (Password 26-23-8)
Modules
M-NSAT
Check that the Timeouts, Bad checksums, and Bad c-s by mod values are equal to 0 (zero).
Check that ROM shows OK, indicating it passed the internal memory test.
ES
24. Check that the three error indicators (MP-203 Error, QUART Error, and 1/0 Error) show NO.
NOTE: MP-203 Error represents the NSAT or OSAT pulse oximetry board.
BS
2-19
Cardiocap/5 Technical Reference Manual
NOTE: If you just turned on the monitor, wait until the "Calibrating gas sensor" message
disappears from the screen. Then, enter the Gas Unit - General service menu by pressing the
ComWheel and selecting:
Monitor Setup
Instali/Service (Password 16-4-34)
Service View (Password 26-23-8
Modules
Gas Unit
General
Check that the configuration shown corresponds with the configuration of your monitor.
BS
26. Check that the Timeouts, Bad checksums, and Bad c-s by mod values are not increasing faster
than 50 per second. A value increasing faster than this indicates a failure in Module Bus
communication. Refer to the Troubleshooting chapter.
BS
27. Select Previous Menu to return to the Gas Unit menu, then select Gases. Check that the
displayed Ambient value corresponds with the current ambient pressure (+ 20 mmHg).
ES
28. Check that the Amb-Work value in the service menu is within the range of 40-75 mmHg.
ES
2-20
Installation and Functional Check
Check that the NMT header with related information is displayed in the chosen digit field.
BS
30. To enter the NMT service menu, press the ComWheel and select Monitor Setup. Then select:
install/ Service (password 16-4-34)
Service (password 26-23-8)
Modules
NMT
Check that the Timeouts, Bad checksums, and Bad c-s by mod values are not increasing faster
than by 50 per second. Check that the NMT memory passed the internal memory test (RAM, ROM
and EEPROM ail state OK.
JES
31. Check that:
* Inthe digit fields, the NMT header, NMT-related information, and “Cable off” are shown.
Connect an ElectroSensor to the NMT sensor cable and plug the cable into the NMT connector on
the front panel. Check that the digit field message changes to “Measurement OFF” and Cable in
the service menu states EMG and ELECTR. OFF.
#5
32. Perform the stimulus current test:
Connect a 3 kQ resistor between the ElectroSensor stimulus electrode leads (brown and white).
In the NMT service menu, select Start Curr. test. Check that the test was successful with all three
test currents, that is, Current test (mA): states 30 OK, 50 OK, and 70 OK.
Connect the NMT ElectroSensor leads to the NMT simulator. Set the switch on the simulatorto
“Fade off” and turn the knob to “max.” In the Service menu, check that Cable states EMG.
ES
2-21
Cardiocap/5 Technical Reference Manual
33. Start NMT measurement (TOF) by selecting NMT Setup and selecting Start-up. Press Previous
Menu to return to the NMT service menu.
In the digit field, when the “Supramax search” message changes to “Setting reference,” check
that the detected supramaximal current is less than 70 mA (the Current set value on the service
screen is less than 700).
Check that four successive stimulus pulses are generated in 10-second intervals. During each
stimulus pulse, a tone sounds and a small asterisk (**) appears in the digit field. In the digit field,
check that TOF% is within 95-105, Count is 4, and T1% is within 95-105,
In the NMT service menu, check that T1%, T2%, T3%, T4% and Ratio% are all within 950-1059.
Check that the Noise value stays under 50.
ES
34. In the NMT service menu, select NMT Setup and change the stimulus Pulse Width to 100 us.
Then press Previous Menu to return to the NMT service menu.
In the digit field, check that TOF% is still within 95-105, Count is 4, and T1% is within 95-105.
Change the stimulus pulse width to 300 ps and check the same parameters.
BS
. Turn the knob on the NMT simulator
to O (zero). In the NMT service menu, check that the values for
T1%, 12%, T3%, T4% turn to 0 and the Ratio% shows - - -. in the digit field, TOF%= should also
state - - -; Count and T1% should show 0.
BS
36. Turn the NMT simulator knob back to “max.” Change the stimulus mode to Double Burst
Stimulation (select NMT Setup - Stimulus Mode - DBS).
Check that only two stimulus pulses are given in a 10-second interval. In the NMT service menu,
check that T1%, T2%, and Ratio% are still within 950-1059. tn the digit field, check that DBS% is
within 95-105, Count is 2, and T1% is within 95-105.
BS
37. Change the stimulus mode to Single Twitch stimulation (ST). Check that the monitor starts giving
only one stimulus pulse with a 1-second interval. Note the time when the ST stimulation started.
In the service menu, check that the T1% value is within 950-1059. In the digit field, the Count
value should be 1 and T1% within 95-105.
BS
38. Continue single twitch stimulation. Five minutes after the start of ST stimulation, check that the
NMT measurement stops and the “Measurement OFF" message appears in the NMT digit field.
BS
39. Replace the NMT ElectroSensor with the NMT MechanoSensor. In the NMT service menu, check
that Cable states PIEZO.
BS
2-22
Installation and Functional Check
Ας
Connect the patient simulator to the ECG and invBP connectors, connect an SpO, finger sensor to
the Sp0, connector, and connect the temperature test plug to a Temp connector. Attach the 5006
sensor to your finger. Check that:
BS
2-23
Cardiocap/5 Technical Reference Manual
3. Setthe invBP simulatorto +100 mmHg static pressure. Push the ZERO button and check that:
* The invBP waveforms set on the baseline.
BS
4. Check that the “No probe” message is displayed when an Sp0, sensor İs not connected.
Connect an Sp0, finger sensor to the monitor (but not to a finger). Check that “Probe off” (for
standard pulse oximetry), “Check probe” (for N-XOSAT), or "Pulse search" then “Check probe”
{for N-XNSAT) is displayed.
BS
5. Attach the SpO, sensor to your finger and check that the pleth waveform is displayed and the
SpO, value is in the expected range (95-99%).
Check that the HR value is calculated from SpO, when the ECG and InvBP (P1/P2) cables are not
connected.
BS
6. Remove the SpO; sensor from your finger and check that “Probe off” (for standard pulse oximetry)
or “Check probe” (for N-XOSAT or N-XNSAT) ts displayed.
Disconnect the sensor from the monitor and check that the “No probe” message is displayed.
BS
7. Attach an adult NIBP cuff onto your arm and perform one NIBP measurement. Check that the cuff
is identified (Adult appears in the NIBP digit field for a short time).
BS
8. Block the tip of the sampling line with your finger and check that the “Sample line blocked”
message is displayed on the monitor screen within 30 seconds.
Detach the D-fend and check that the "Check D-fend" message is displayed on the monitor
screen within 30 seconds.
ES
9. Reattach the D-fend. Breathe shortly into the sampling line. Check that the CO, waveform moves
up on the screen.
BS
After completing the check, be sure to fill in all necessary documents.
2-24
Functional Check Form
Datex-Ohmeda Cardiocap/5
Customer
Service
Monitor Configuration
Monitor model: Hemodynamic model Hemodynamic model with gas measurement (|
m
F-MX F-MXG
Functional Inspection
OK = Test OK N.A. = Test not applicable Fail = Test Failed
General
LU
Recorder test
©
x
z
>
008
7. Recorder message
ロロ
oo
8. Recording |
000
000
000
10. Communication
i
i
j
}
|1
Network test
0009
0000
0000
13. Communication
000
000
17. Power frequency
0
U
DC
26. Communication LI U
27. Ambient pressure | | Current ambient pressure + 20mmHg |
Page 2
Functional Check Form
29. Recognition | |
| r% dl || 950-1059
||| count || + |
[1% || 1[ 95-105 |
Page 3
Functional Check Form
Performance Checks
General
X
о
=
>
© 00:
1. Screen setup
DO
000
2. Waveforms
oo 0 ooo
AnvBP measurement —
3. Testwith patient simulator
O
Sp02 measurement
4. SpO sensor detection
000
00 0 000
5. Test measurement
€
NIBP measurement |
7. Test measurement and cuff identification
Gas measurement
00
8. D-fend check
9. CQ, waveform
Notes
Signature Date
Page 4
Contents
Functional inspection........
ECG measurement test
Temperature measurement test
Non-invasive blood pressure (NIBP) measurement test
NS
Sp0。 measurementtest ee
Invasive blood pressure measurement test...
OO
Table of Figures
Figure 3-1. Nafion tube and filters to replace in gas unit... 3-3
Planned Maintenance
3. PLANNED MAINTENANCE
3.1 Introduction
These instructions include procedures for planned maintenance (PM) of the Datex-Ohmeda
Cardiocap/5 monitor. Performance of planned maintenance procedures is recommended once each
year after installation of the monitor.
These instructions are for the maximum functional configuration of the monitor. Perform the
procedures in order and skip items that do not correspond with the configuration of your monitor.
Complete instructions on how to perform complex procedures are included in the Service Procedures
chapter later in this manual.
NOTE: The contents of the Planned Maintenance Kits are listed in the Spare Parts chapter.
3-1
Cardiocap/5 Technical Reference Manual
3-2
Planned Maintenance
1. Visual inspection
1. Switch the monitor
to standby and disconnect all cables from the rear of the monitor.
2. Check the external parts of the monitor. For example, check that the power cord receptacle is
intact and the D-fend latch moves properly.
3. Check that the markings of the mains fuses correspond to the fuse plate.
BS
2. Parts replacement
1. Dissamble the frame (see section 7.3.1).
2. Replace these items in the gas unit: occlusion filter, zero absorber, Nafion tubing, and
OM reference filter. Also replace the D-fend, D-fend O-ring, and sampling line.
NOTE: Use only Datex-Ohmeda sampling lines to ensure proper function.
Occlusion filter
OM reference filter
Zero absorber
Nafion tube
Ft
4. Clean the recorder unit and replace the recorder paper, if necessary.
5 Reconnect all cables, then perform a gas sampling system leak test (see section 7.4.7).
BS
3-3
Cardiocap/5 Technical Reference Manual
3. Functional inspection
1. Connect the power cord and check that the Battery charge status LED turns on or flashes.
= The norma! monitoring screen appears and no error messages are displayed.
NOTE: “Check network connectors” is displayed if the N-XNET or N-XDNET option is installed.
e The time and date are displayed; adjust if they are incorrect.
* The battery charging symbol is in the upper right corner of the screen:
一 十
« If the monitor contains a recorder, start-up information prints. Verify the time and date.
Enter the Keyboard service menu (illustrated in the Service Menus chapter) by pressing the
Comwheel and selecting:
Monitor Setup
Install/ Service (password 16-4-34)
Service (password 26-23-8)
Keyboard
To test the function of specific direct access keys, select Dummy Press.
Press the Silence Alarms key and check that the keypress generates a sound from the speaker
and the corresponding text in the service data screen changes color. Repeat for the following
keys: Trends, ECG, NIBP, and Normal Screen.
Go to the Monitor - Voltages service menu (return to the Service View menu and select
Monitor - Voltages). Check that the supply voltages are within the limit values:
3-4
Planned Maintenance
NOTE: You can continue the check while the monitor is powered by batteries.
Configure the monitor screen according to the monitor configuration so that all parameter
information is displayed.
NOTE: The Resp parameter has to be selected in one of the Waveform Fields or Digit Fields
before the respiration measurement can be turned on.
ES
4. ECG measurement test
1. Return to the Service View menu and select Modules - ESTP : ECG to enter the ESTP : ECG service
menu. Check that:
s The Timeouts, Bad checksums, and Bad c-s by mod values are not increasing faster than
50 per second.
* The ECG/RESP board memories passed the internal memory test (that is, RAM, ROM, and
EEPROM all state OK).
Connect a 5-lead ECG cable to the module. Check that the Cable type shows 5 lead. If it shows
3 lead, make sure the 5-lead ECG cable being used contains the necessary wiring for cable
recognition (pins 0, 8, and 9 are connected together).
Check that each Electrode shows OFF and the “Leads Off” message is displayed.
Connect the patient simulator. Check that parameter information is displayed as configured.
Check that the waveforms correspond to the simulator settings. Switch off the simulator and
check that the “Asystole” and “Apnea” messages are displayed.
BS
5. Temperature measurement test
1. Return to the Modules service menu and select ESTP : STP to enter the ESTP : STP service menu.
Check that the STP board memories passed the internal memory test (that is, RAM, ROM, and
EEPROM all state OK).
Check the temperature calibrations with the temperature test plugs. Calibrate if necessary.
BS
3-5
Cardiocap/5 Technical Reference Manual
Check the NIBP calibration with 200 mmHg pressure and calibrate if necessary.
Return to the NIBP service menu and select Pneumatics to enter the Pneumatics service menu.
Check the Watchdog timer activation pressure: the audible signal must activate at 3 to 8 mmHg.
If necessary, adjust the limit with the trimmer on the NIBP board and recalibrate NIBP.
Return to the NIBP service menu and select Watchdog to enter the Watchdog service menu.
Check the NIBP watchdog timer.
Return to the NIBP service menu and select Safety Valve to enter the Safety Valve service menu.
Check the safety valve functions.
NOTE: Make sure the pressure manometer can measure pressures over 300 mmHg. !f such a
pressure manometer is not available, perform the check with an adult cuff that is connected
around a round object, such as a calibration gas bottle.
The Max press and 2 s after stop pressure values for both transducers should be within
290 to 330 mmHg for Adult and 154 to 165 mmHg for Infant.
Connect an infant cuff to the monitor. Start the measurement and check that the infant cuff is
identified correctly. Cancel the measurement.
Attach an adult NIBP cuff to your arm and perform one NIBP measurement. Check that the cuff is
identified correctly (for example, Adult briefly appears in the NIBP digit field).
Check that the module gives a reasonable measurement result.
BS
7. Sp0, measurement test
1. Check that “No probe” is displayed when an SpO, sensor is not connected to the monitor.
Connect an SpO, finger sensor to the monitor (but not to a finger). Check that “Probe off” (for
standard pulse oximetry), “Check probe’ (for N-XOSAT), or "Pulse search" then “Check probe”
(for N-XNSAT) is displayed.
Attach the SpO, sensor to a finger and check that the pleth waveform is displayed and the SpO,
value is in the expected range. Check that the HR value is calculated from SpO, when ECG and
InvBP (P1/P2) cables are not connected.
Remove the SpO, sensor from your finger and check that “Probe off” (for standard pulse oximetry)
or “Check probe” (for N-XOSAT or N-XNSAT) is displayed.
Disconnect the sensor from the monitor and check that the “No probe” message is displayed.
BS
3-6
Planned Maintenance
BS
9. Gas measurement test
NOTE: See the Service Menus chapter for illustrations of the service menus used for the gas
measurement test.
1. Check that the fan in the gas measurement unit is running.
NOTE: If you just turned the monitor ON, you have to wait until the "Calibrating gas sensor"
message disappears from the screen before entering the Gas Unit service menu.
2. Return to the Modules service menu and select Gas Unit - General. Check that the displayed
module configuration corresponds with the configuration of the monitor.
3. Return to the Gas Unit service menu and select Gases. Check that the displayed Ambient value
corresponds with the current ambient pressure (+ 20 mmHg}
BS
10. Anesthetic agent identification test
NOTE: Use only the recommended Datex-Ohmeda calibration gas (see Recommended tools and
accessories earlier in this chapter).
While displaying the Gases service menu, feed the calibration gas continuously for at least
30 seconds and check that the screen shows:
If the value is higher, recalibrate the agent identification and check the value again.
BS
3-7
Cardiocap/5 Technical Reference Manual
25
12. Gas sampling line and D-fend check
1. Block the tip of the sampling line with your finger and check that the “Sample line blocked”
message is displayed on the monitor screen within 30 seconds.
2. Detach the D-fend and check that the"Check D-fend" message is displayed on the monitor screen
within 30 seconds.
3. Breathe once into the sampling line and check that the response time of the CO, curve is normal.
BS
13. Trend test
Check that the monitor is capable of storing the trend information and temporary settings for over
2 minutes while in standby.
BS
14. Watchdog test
In the Watchdog Tests service menu (Monitor - Watchdog Tests), check that the monitor resets and
restarts properly.
BS
15. Recorder test
1. Open the paper compartment cover. Check that the “Recorder: Cover open” message appears on
the screen, then close the cover.
2. Press the Record Waveform/ Stop key on the recorder and check that the recorder starts
recording the selected waveforms. Press the key again to stop recording.
3. Press the Record Trend/ Stop key on the recorder and check that the recorder starts recording the
selected trends. Press the key again to stop recording.
BS
3-8
Planned Maintenance
Check the connection to the network by checking the states of the Network connection LEDS
between the connectors:
Check that the monitor connects to the Datex-Ohmeda Network (that is, the network symbol
is displayed on the upper right corner of the screen).
NOTE: If the battery is being charged, the battery charging symbol is displayed instead of the
network symbol.
A message regarding the connection to the Datex-Ohmeda Information Center should appear in
the message field on the screen.
The counters for data errors (CRC, Frame, Transm.) are stable
NOTE: The counters may show values greater than O (zero), however, a continuous Increase in any
value indicates a problem.
4. Check that the counters for hardware errors (Intern., Missed, FIFO, Overrun) all show O.
25
3-9
Cardiocap/5 Technical Reference Manual
Insert a memory card labeled "Data" into the second slot. Check that the "Data card inserted"
message appears in the message field.
Check that some trend information is available in trend memory for monitored parameters. Erase
the trends. Check that the trends are properly erased.
Reload the trends from the Data card. To do so, select Patient Data - Patient from card. Then,
select the last saved file and select Load.
After the monitor has loaded the data, check that the trends are available again.
Display the Module Status service screen (Modules - More Modules - Memory Module) and
check that:
BS
18. Service log check
1. Enter the Service Log menu and check for possible problems. If the monitor contains a recorder,
record the service log data by selecting Record Log.
2. Clear the content of the service log by selecting Reset Log.
ES
19. Electrical safety check
1, Perform an electrical safety check and leakage current test.
2 Check that the monitor functions normally after performing the electrical safety check.
3. Switch the monitor to standby, disconnect the power cord, and perform final cleaning.
4 Fill in all necessary documents.
BS
3-10
Planned Maintenance Form
Datex-Ohmeda Cardiocap/5
Customer
Service
Monitor Configuration
IDD
second temperature (T2) oximetry
N-XC σου N-XNSAT Nellcor compatible pulse oximetry
다 디니
기니
N-XCO CO,, N,0, Patient Oxygen Data collection and data management options
N-XCAiO CO,, anesthetic agents, agent N-XREC Recorder
identification, N,0, Patient Oxygen
НО
N-XV Patient Spirometry N-XNET Network
OK NA. Fail
1. Visual inspection m | | (|
10000
D-fend Other
Page 1
Planned Maintenance Form
>
©
と
ロロ ロロ ロロ ロロ ロロ ロロ ロロ ロロ ロミ
. Functional inspection
ロロ ロロ ロロ ロロ ロロ ロロ ロロ ロロ ロ
ロロ ロロ ロロロロ
. ECG measurement test
i
|
|
||
i
|
. Invasive blood pressure measurement test
i
|
|
i
i
10. Anesthetic agent identification test
|
|
11 . Spirometry test
ロロ ロロ ロロ ロロ ロ
12. Gas sampling line and D-fend check |
Notes:
Signature:
Page 2
Contents
Chapter 4. Troubleshooting
4.1 Messages............ 00000000 εν οκ οκ κeineиниения 4-1
4.2 Troubleshootingcharts....................................0...
00020 4-7
4.2.1 Star-up troubleshootnE ohen tone on eee ten retence 4-7
4.2.2 General
423 ECG. . . . . . . . OD ON A
4.2.4 Impedancerespiration......................... einen 4-9
4.2.5 Pulseoximetıy(Sp0)),standard.............................. ie 4-9
4.2.6 NIBP ii 4-9
O ON 4-9
4. TROUBLESHOOTING
See the User's Reference Manual for more messages and troubleshooting procedures.
4.1 Messages
MESSAGE POSSIBLE CAUSE ACTION
Acknowl. alarms Acknowledged alarms are silenced (Silence No action required.
silenced Alarms key pressed dunng silencing period).
Air leakage Hose and/or cuff leaking. Damaged cuff, cuff Replace cuff, cuff connector, and/or
connector, or hose double connector. hose double connector.
Hose and cuff OK, but tube disconnected or Connect or replace tube.
damaged.
O-ring damaged or missing. Replace O-ring.
Leakage inside the module. Replace the whole tubing.
Tube disconnected or damaged; valve(s) Fix connections. Replace tubes/valve(s).
damaged.
Apnea deactivated Apnea alarm is silenced until reactivation after No action required.
five breaths.
Artifact(s) Unsuccessful NIBP, Sp02, or ECG measurement Stabilize or warm the patient and start a
because of patient movements or shivering. new measurement.
Back-up batt. failure Discharged or faulty back-up battery. Use mains power for four hours, then
switch to battery power. If the message
reappears, replace the battery.
Cable off NMT or regional block cable is not connected. Check/correct cable connections.
Calibrate Agent ID Agent identification error. Perform gas calibration.
Calibration failed Unsuccessful calibration of P1/P2 (number Turn the transducer to
(invasive blood pressure) field); possibly due to pulsating waveform. sphygmomanometer and try again
(zeroing takes place first).
Gain is beyond the limits (+ 20 % of the default Replace the transducer.
gain).
Calibration not NIBP calibration protection is set to OFF. Open the NIBP Calibration menu and set
protected Protection ON.
| 一
Calibration switch ON EEPROM protection switch (gear wheel) Enables the Protection OFF setting in the
accessed through hole in bottom of module is NIBP Calibration menu. Turn the gear
set for calibration. wheel to select Protection ON if you are
not going to calibrate.
--
4-1
Cardiocap/5 Technical Reference Manual
4-2
Troubleshooting
EEPROM Error EEPROM memory malfunction Perform factory reset. If the problem
(EEPROM is on CPU board). persists, replace the CPU board. Perform
factory reset after replacing the board.
EMG electrodes off (NMT) EMG recording electrodes are off. Check application of electrodes and
Electrodes or stimulus clip is loose. stimulus clip.
ESTP measurement Signal cable is loose or not attached. Check signal cable connections from
removed Mother board to CPU board.
Gas measurement Signal cable is loose or not attached. Check signal cable connections (Mother
removed board to CPU board or Spirometry board
{for CO.) to gas unit).
Leads off One or more cables are disconnected or off the Check ECG cable, all leads, and the
patient. This message may appear during neutral electrode (RL/N).
defibriHation.
一一 一
Long measurement time NIBP measurement ts prolonged over the Reapply the cuff, calm the patient, and
maximum measurement time (two minutes with Start a new measurement.
adult or child inflation limits; one minute with
infant inflation limits).
Measurement off NMT cable is connected, but NMT measurement No action required. Message is no longer
has not started. displayed when measurement starts.
Monitor is overheating Temperature inside monitor is above maximum Check/change the dust filter at the back
limit. of the monitor.
一 一
MVexp << MVinsp Leak in patient circuit between patient and D- Check D-tite connection and D-lite
lite, or in the patient lungs, or leak in tubes from tubing.
D-lite to monitor.
A
MVexp < 0.5 I/min D-lite/Pedi-lite disconnected from ventilator Check D-lite/Pedi-lite connections and
(Mvexp <0.2 1/min with circuit (not shown during Apnea). Gas sampling tubing.
Pedi-lite) is working correctly.
Network connection Too many monitors are active on the network. Check/decrease number of monitors on
down network. |
AAA
4-3
Cardiocap/5 Technical Reference Manual
Poor signal Modulation (red or infrared) < 0.25 %. Patient Make sure patient is not cold.
may be cold.
Pressure measurement Signal cable is loose or not attached. Check signal cable connections from
removed Mother board to CPU board.
Printer error Printer is out of paper or paper is jammed. Printer Check paper, printer connections,
is not online oris not tumed on. Printer cable is power, and cables.
loose or broken.
Printer failure Printer is not responding (see Printer error). Select/use a different printer.
Pulse search Monitor is searching for pulse oximetry signal. No action required.
RAM Error RAM memory malfunction Replace the CPU board. Perform factory
(RAM is on CPU board). reset after replacing the board.
Recorder: cover open Recorder cover open. Close the recorder cover correctly.
Recorder: input voltage +12 Vrec is too high. Check flex-strip cable and connector
high board.
Recorder: input voltage +12 Vrec is too low. Check flex-strip cable and connector
low board.
Recorder: out of paper Out of paper or paper jam. Insert a roll of paper into the recorder or
release paper jam.
Recorder: system error System error. If the problem persists, replace the
(1, 2, or 3) recorder unit.
Recorder: thermal array Recorder overheated. Stop using and allow recorder to cool.
overheat
Reference not stable (NMT) The deviation between the four reference Check for movement.
stimulation responses is too big, causing
reference setting
to fail. This can be caused by
movement artifact and may occur when the
patient ts relaxed.
Replace D-fend Residue build-up on water trap membrane Replace the D-fend.
(decreases airflow).
L
4-4
Troubleshooting
SRAM Error SRAM memory malfunction Restart the monitor. If the problem
(SRAM is on CPU board). persists, replace the battery for SRAM
timekeeper. Perform factory reset after
replacing the battery.
| if message persists, replace CPU board. :
Supramax not found (NMT) Supramaximal stimulus current (70 mA) Stop measurement, reposition the
was not found. stimulating or recording electrodes, and
restart measurement.
Supramax search (NMT) Supramaximal stimulus current search in No action required.
progress.
Temperature error Faulty calibration. Perform calibration. If it does not help,
check that front panel connectors are
properly connected to STP board.
. | . - っ -
Tetanic (NMT) Tetanic stimulation is on. No action required.
| 一一 一
Unable to measure Dia Accurate diastolic pressure not achieved Automatic retrial with increased
because of artifacts, weak pulsation, ete. pressure.
Unable to measure Sys Systolic blood pressure probably higher than the Automatic retrial with increased
maximum inflation pressure. pressure.
Unstable Unsuccessful calibration. Message may appear Repeat calibration.
Lo dunng calibration.
4-5
Cardiocap/5 Technical Reference Manual
4-6
Troubleshooting
y
Check that the green LED on front panel upper right This indicates that +15V VDD is coming from the
comeris lit or flashing when the power cord ts AC/DC unit to the DC/DC board and the battery is
connected to mains. charged.
OK but still no display
!
Check that the alarm LEDs on front panel light after If the LED is lit it indicates that +3.3 V and +5 Vis
pressing the ON/STBY button. coming from the DC/DC board to the CPU.
Alarm LEDs blank after about 20 seconds. Ifthe LEDs only blank for a short moment and then
remain lit, the CPU software didn’t start.
y
Try reloading new software from a PC-card
y
The CPU and parameter units seem to be OK, the
picture is missing for some other reason. Some
voltages may be missing or the display is broken.
4-7
Cardiocap/5 Technical Reference Manual
4.2.2 General
SITUATION POSSIBLE CAUSE ACTION
Keyboard and ComWheel Keyboard and ComWheel cables Detach monitor rear cover and check the
are not working not connected to CPU board. cable connections to the CPU.
Faulty CPU. Replace the CPU if the cable connections
are OK
Keyboard is not working Keyboard cable not properly Detach monitor rear cover and check the
but the ComWheel works connected to the CPU board. keyboard cable connection to CPU.
Faulty key or defective CPU board. Enter service mode and do Dummy Press
keyboard test. Replace the keyboard if
only a single key does not respond.
Replace the CPU if more than one or two
keys do not respond.
No curves in the Wrong configuration set. Ifthe menus can be opened, enter service
waveform fields and no mode and select the configuration set
digits in the digit fields according to the monitor configuration.
(gray fields only)
SRAM back-up battery failure. if the set configuration is lost after a short
switch off period, replace the back-up
battery on the CPU board.
NOTE: The clock goes to 0:0 and the
configuration is lost if the back-up battery
fails or if the SRAM lost data for some
other reason.
Some trend details do Settings for parameter and time Adjust the parameter and time scales to
not appear. scales. see the desired trend detail.
4.2.3 ECG
SITUATION POSSIBLE CAUSE ACTION
HR numerical display No heart rate available. If no ECG waveform, check LEADS OFF
shows '---* message and connect the leads.
If ECG waveform exists, check heart rate
source (For example, in the ECG Setup
menu behind ECG key).
Unacceptable ECG Poor electrode or poor electrode Electrodes from different manufacturers
waveform
|.
skin contact. are used/Too much/little gel is used.
Poor electrode condition. Electrodes are dried out.
Improper site of electrodes. Check that electrodes are not placed over
bones, active muscles, or layers of fat.
improper skin preparation. Remove body hair. Clean attachment site
carefully with alcohol.
Improper bandwidth filter. Check filter.
No ECG trace Waveform not selected on screen. Adjust (Monitor Setup - Screen Setup -
Waveform Fields).
一 —— ———İ
Module not plugged in correctly. Plugin.
4-8
Troubleshooting
4.2.6 NIBP
SITUATION POSSIBLE CAUSE ACTION ο.
No NIBP value displayed NIBP not selected on screen. Check monitor setup.
4.2.7 Temperature
SITUATION POSSIBLE CAUSE ACTION
No temperature Wrong type of probe. Use correct probe.
displayed Temperature out of measurable The range is between 10 and 45 °C.
range.
Temperature calibration not Set the protection ON in the service menu.
protected.
4-9
Cardiocap/5 Technical Reference Manual
4-11
Cardiocap/5 Technica! Reference Manual
No response to breathing Sampling line or water trap loose, blocked, Check/correct sampling line and water trap
or improperly attached. Water trap container blockage and attachment. Empty water
full. trap.
No response to any gas Sampling line, water trap, or intemal tubing Check/correct sampling line, water trap,
blocked orloose, or improperly attached. and intemal tubing blockage and
attachment.
Pump failure. Occlusion or zero valve Enter the appropriate service menu(s) and
malfunction. Supply voltage missing. Serial check/ verify the correct operation of the
communication error. pump, valves, etc.
Sudden increase in gas Water trap malfunction. Check all internal tubing and the interior of
display the water trap for occlusions or leaks.
Replace water trap. Check flow rates.
Significant drift in all Leak in sampling line or internal tubing Check/correct leak in sampling line or
gases (especially in conjunction with readings that internal tubing.
are too low).
Abnormally high (or low) Pressure transducer failure. Repair unit.
response to all gases or
sudden occlusion warning
ETCO, value too low Leak in sampling system. Calibration error. Check ali connections.
High ventilator by-pass flow. Check calibration.
Possible clinical causes: sudden circulation
decrease, pulmonary embolism, very large
dead-space, hyperventilation, or large
shunting.
ETCO, too high D-fend contaminated. Calibration error. Change D-fend.
Possible clinical causes: hypoventitation or Check calibration.
increased metabolism.
Waveform clipped Incorrect scaling, Change scale
No response to breathing Sampling line or water trap loose or blocked. Check all connections,
Sample gas outlet blocked. Air teak. Check that outlet is open.
Possible clinical cause: apnea.
—
ETCO, over scale (>15%) |} Contaminated CO, sensor. Faulty D-fend. Change D-fend.
(not measurable) Possible clinical cause: abnormally high
ETCO, (permissive hypercapnia). |
4-12
Troubleshooting
Strongly vibrating loop Water or secretions tn hoses or D-lite. Suction the patient. Change dry D-lite
Possible clinical cause: mucus in ET tube. and/or empty the water from hoses.
Volumes too large or too Wrong mode vs. sensor selection. Check mode and sensor (D-lite for adult;
small Pedi-lite for pediatric).
Fluctuating Raw Mucus in airway or tubing. Check/clear airway and tubing.
Ventilator exp. valve causes fluctuations
during exp. Flow.
Possible clinical cause: patient breathing
effort against the ventilator; patient-
triggered breaths.
Raw tao high Mucus or kink in tubing. Check/ clear tubing.
Possible clinical cause: asthmatic patient
or bronchospasm.
Raw value invalid: Possible clinical cause: spontaneous
breaths; breathing against ventilator;
patient-tnggered breaths.
=
Ppeak too high Possible clinical cause: bronchospasm or
patient is coughing; breathing against
ventilator; obstructed airways or HME.
Compl value invalid: -- Possible Clinical cause: spontaneous
breaths.
| 一
Static PEEPI not CO, measurement not connected. Static Connect CO, meas. to D-lite. Go to
measured PEEPI measurement not selected, Exp. spirometry Setup.
pause lasted less than 4 seconds.
ud ον
ον
4-13
Cardiocap/5 Technical Reference Manual
Check timeouts,
bad checksums, and
bad c-s by mode
yes
Check RAM, ROM,
and EEPROM
Replace original
input board/cable
assembly.
Connect ElectroSensor to
NMT board stimulator
and start measurement
Replace original
Replace original
Replace original
input board/cable
ММТ sensor. NMT sensor cable.
assembly.
4-14
Frames and Software
5.1 Introduction
Two different frames, or models, are available for Cardiocap/5:
ST analysis
po 一~
NIBP
Impedance respiration
Optional Parameters
N-XP Two invasive pressure channels and second temperature (12)
N-XC CO,
N-XCO CO,, N,0, Patient Oxygen
5-1
Cardiocap/5 Technical Reference Manual
5.2.1 Introduction
The AC/DC unit SR92A720 is a switched-mode power supply made by Efore Ltd.
The AC/DC switched-mode power supply unit SR92A720 is an integrated part of a patient monitoring
device. The SR92A720 power supply is mains connected and can be supplied with a universal single
phase line voltage in the range of 85 to 275 V. The unit has one output line with a maximum
continuous output power of about 63 W.
The aluminum chassis ts classed as IP20. The chassis is connected to protected earth. This makes
the assembly and handling of the power supply unit safer. In addition, the aluminum chassis is used in
thermal management and EMC design.
5-2
Frames and Software
Jinrush J ; To
pe VI VS X122
m | ーー / 7 External
shutdown
signal
и X12-6
ff ーー — m
/ External
feedback
signal
This AC/DC converter contains two converter stages: a preconverter and a postconverter. The
preconverter is an AC/DC converter which together with input low-pass filtering draws almost a
sinusoidal current from the mains. As a result, input line current harmonics are reduced and the power
factor is improved, therefore, this converter stage is typically also called a power factor correction (PFC)
converter. This PFC converter is based on boost topology and its output voltage of about 370 V is fed
to the subsequent DC/DC converter stage. The DC/DC converter uses a forward-type topology to
convert the intermediate voltage of 370 V to an output voltage of 15.7 V (typical value).
PEAK
MOSFET V1
5-3
Cardiocap/5 Technical Reference Manual
The PFC converter controls the intermediate voltage. The intermediate voltage has been set to about
370 V. In addition, there is a small (<10 V,,) ac ripple component across capacitor C9. The frequency
of this AC component is double the line frequency.
|
Isolation | Connector |
, : |
t 1 X12-4 A FISIV
| p + . 全 $ 수 ーー bi .
| | | 1825 '
1 + 1
1 . 1 Ac
| "= I 1
| | 4 |
| | | i X12-1 i
| no ㆍ ~
| в ーー や ツー ーー テキ
|
| i
и! i |
|
|
||
i
1 г
| . : 1 X12-3 : 、
rå à ーー 内 Shutdown
| | comparator
M | и |
5-4
Frames and Software
Output voltage is controlled by a two-loop system. This means that there is a fast inner-voltage-
control loop and a slow outer-voltage-control loop. The slow-voltage control is an external voltage
signal, which, by setting the offset to the inner-control loop, determines the level of the output voltage.
if the external feedback signal is not connected to pin 6 of connector X12, the output voltage
measured from connector X12 settles to approximately 10 V. The external feedback signal cannot
draw the output voltage lower than this 10 V level. The maximum output voltage Jevel that the external
voltage control signal can cause is set by trimmer resistor R52. Typically, the external control signal
sets the output voltage to 15.7 V.
5-5
Cardiocap/5 Technical Reference Manual
5.3.1 Introduction
The DC/DC board converts the output voltage of the AC/DC power supply or the battery voltage to
various supply voltages for the electronics of Cardiocap/5 monitor. Another main task of the board is
battery charging.
ο
ÈmarὋOT
tu.
Lear
q
qLod
나 07 A
5-6
Frames and Software
„ Tim VDD
AC/DC T
Connector VDD |
Recorder
Control Measuring Power
Electronics Electronics Connector
CPU
Signal | — sd
Connector
Fan
ADCHO ... ADCH9 A/D Connector
ADCH10 Converter
11CH 12 BIT
Serial Interface
Power-on sequencing (rising order of the voltages) is controlled by the main processor. The +3.3V (the
supply voltage for the CPU) is switched on first by the control electronics. +5V rises after +3.3V without
CPU control. After that come other voltages. VCC_INT, +5V_INT and +2.5VREF are internal voltages of
the DC/DC board.
5-7
Cardiocap/5 Technical Reference Manual
[VDD
ENA_CHG
= IN high-side G- /nİn. fet
driver ]
+5V-INT | +2.5\ [+2 5VREF
Battery
Charger
一 REFERENCE —
BAT VIN-VCPU
INPE [o]
ENA 3V3
ー N в n-fet
high-side 1 36
driver x +3.3V
Switcher
J
VCC_INT
VDD /
AC/DC VOUT “ “i rov. ANT +5V_INT
ENA 15VB R
> ом умы
high-side と
POW_ON LED
driver мм 158| +15VB
ー Boost |--
Switcher VDD/BAT
+15VB
R ENA VINSV (
Ni h-si G n-fet τον
BATTEST nfet e \-side
ENA 15V
= м high-side G n-fet +15V +15V
driver V Flyback
VS 店 cher -15V
J
ЕМА_ММ12УГ G ー
high-side
driver | vin_12v] +12V +12V
= Switcher
ENA 12V]
+15VD| VFAN m ーー i
REG. ENA ISVD | fm
ENA G n-fet
FAN high-side R
OO driver +15VD
FAN
5-8
Frames and Software
All regulated output voltages of the DC/DC board are short-circuit protected. Protection is based on
current sensing and limiting by a switcher circuit except + 15VD, which is protected by a separate
circuit breaker.
The output voltages are + 14.4 to +15.6V and -15.6 to -14.4V. Both outputs can be loaded at 0.2A
max.
5-9
Cardiocap/5 Technical Reference Manual
On the DC/DC board, the regulated voltage is actually +15VB, which is named +15VD after circuit
breaker electronics. During mains usage, +15VB is taken from VDD via a diode. When the mains
voltage drops, the monitor continues operation on battery. In battery use +15VB is converted from the
battery voltage by a boost type switcher. The boost converter and VDD have common control
electronics. The operation is described further in the next sections.
The boost converter output voltage +15VB ¡is +14.5 to +15.4V. Maximum output current is 3.54.
A boost converter cannot limit its short-circuit output current. To prevent the components from
excessive heating in an overcurrent situation, a signal OC_15VB shuts +15VD consumption. It is also
connected to the CPU to inform the CPU of exceeding the input current limit of the boost switcher.
OC_15VB trip point is 6.8 to 7.9A. The output voltage will drop at a lower input-current level than that.
This is meant to protect the output FET in fault condition.
VDD control
The same control circuit that adjusts + 15VB boost converter also adjusts VDD when the monitor is in
mains use. VDD is adjusted so that + 15VB is at a little higher level than the boost regulator would
regulate it at. When mains and VDD drop, + 15VB fails into the lower level where the boost converter
continues its regulation. To achieve adequate accuracy and to prevent the two adjustments from
overlapping, the control circuits use a common reference voltage as well as a common resistor divider.
A signal TRIM_VDD is connected to AC/DC power supply to adjust VDD. VDD is increasing when
TRIM_VDD is increasing. The difference between the two adjustments of + 15VB is 0.2 to 0.45V, which
means that + 15VB is adjusted to 14.7 to 15.85V when VDD is present.
The current limit is of fold-back type, which means that the trip level of the current is lower when the
output voltage is near to ground. in other words, the load current must be reduced to a lower level than
it was when switched off to enable reconnection. The current limit trip level is 5.1 to 7.2A. Signal
OC_15VD/ becomes active when the current limit is tripped.
The FET and current measuring resistor cause a voltage drop between +15VB and +15VD. This voltage
drop is allowed to be maximum 0.25V at 4A load current. +15VD voltage range is 14.25 to 15.85V at
load currents 0 to 4A.
5-10
Frames and Software
5-11
Cardiocap/5 Technical Reference Manual
if the battery voltage is below the threshold voltage (Vt) the charger enters a trickle-charge mode,
where the charge current is reduced to value It. An overdischarged battery is charged with the trickle
current (It) until the battery voltage reaches the level Vt above which the bulk-charge begins. The
charger output short-circuit current (Isc) is limited by the same resistors as the trickle current.
Maximum charge current for the 12V 2,6Ah battery is 0.78A.
5-12
Frames and Software
A CPU control signal (ICHGLOW) reduces the maximum charge current. This can be used momentarily
to cut the power consumption peaks of the monitor.
Power-on logic.
Pushing the on/stby switch to on position causes +3.3V and +5V supplies to rise. (+3.3V and
+5V on and off sequence: +3.3V rises first and goes off last.) The CPU controls the other power
supplies. Pushing the on/stby switch to standby position does not directly disable the supply
voltages. The switch position information is wired also to the CPU, which controls the power
supplies’ switching off, also +5V and +3.3V supplies.
Over-voltage detection.
+5V and +3.3V are rapidly pulled down by a crowbar circuitry in case of over-voltage. All the other
power supplies are disabled. Over-voltage in + 15VB causes VDD to be switched off by signal
VDD_SHUTDOWN/. Over-voltage information goes also to the CPU by signal OV_15VB/.
Switching a power-resistor load to the battery to enable the CPU to measure approximate
capacity of the battery.
The high-charge states of the battery charger are indicated to the CPU by signal CHG_ON/.
5-13
Cardiocap/5 Technical Reference Manual
5-14
Frames and Software
5.4.9 AD converte r
The analog-to-digital converter is controlled by the CPU via a slow serial data bus. The signals are
SSCLK (clock), SSDOUT (CPU data out), SSDIN (CPU data in), and ADC_CS/ (chip select). The A/D
converter is an 11-channel, 12-bit circuit. It uses external reference voltage. On the DC/DC board, the
reference voltage is RC-filtered from +5V INT supply. The input voltage range is 0 to +5V.
A/D channels
5-15
Cardiocap/5 Technical Reference Manual
* 32kByte NV-SRAM
Hi 用. ren
> i Seated
Fe Уi eh
ο τμ]
O E
ei de ted
ts trat tia).
The CPU board manages power by controlling and monitoring the voltage levels and power
consumption of the monitor. It shuts off supply voltages if over-current is detected.
The CPU board controls the Digital and Analog 1/0, Serial Data interface, and Ethernet interface. In
addition to the external interfaces, the CPU controls the internal Module Bus and Recorder interface. A
Watchdog on the CPU board monitors the operation of the software.
The ComWheel and Matrix keyboard are directly connected to the CPU board. The Real-Time Clock,
Audio generator, and Display controller are located on the CPU board. The CPU board is also equipped
with two PCMCIA-compatible data-card slots for software loading and data transfer purposes.
5-16
Frames and Software
tesi
connector
à ISA A VL-Bus|
- Digital /O
< 러 IDA
Ethernet
controller | p 10BASE-T
RICLK + SRAM ~ 32
S > 个 | |
NV-SRAM Le À | PLD 4x8 bit —» Brightness
serial DAC — Contrast
EPROM — Audio
— > _ >PC-interface
CTRL | 4x8 bit __»Serial I/O 1
5V DATA BUS M DAC H———> Serial I/O 2
一 一 一 vr Recorder
5-17
Cardiocap/5 Technical Reference Manual
The synchronous serial channels are implemented with PLD to minimize the load of the main
processor. Communication with the chips connected to the bus is handled by the PLD registers.
Ethernet ID-block
The Ethernet ID-block can be connected to the 9 pin D-connector on the monitor's rear panel. It
contains an EPROM chip where the bedside specific ID number has been stored. The processor will
read the ID number with PLD by sending the address of the memory location to the chip.
» 8 bitx4 DAC
» 12 bitx4 DAC
* 12 bitx11 ADC
Only the 8 bit x4 DAC is located on the CPU board. It generates audio signals and controls the
brightness and contrast of the display.
The 12 bit x4 DAC on the 1/0 board drives the Analog Output signals.
The 12 bitx11 ADC on the DC/DC board monitors the internal parameters of the monitor, such as
voltages, charging currents, and temperatures.
5-18
Frames and Software
The Module Bus is buffered to RS-485 level on the CPU board. The baud rate of the Module Bus is
500 k baud.
Recorder interface
The Recorder serial interface channel uses both CTS# and RTS# handshake signals. In addition there
is a reset signal for the recorder.
The Recorder serial channel has been implemented on channel D of Exar’s UART ST16C654. The baud
rate is 76.8 k baud.
Computer Interface
The Computer Interface serial channel is buffered to RS-232 level on the 1/0 board. lt can be used for
interfacing some external device to the monitor. The computer interface is using channel A of Exar's
UART ST16C654.
Serial 1/0
The channels B and C of Exar's UART ST16C654 have been reserved for future purposes. They have
been wired to the 1/0 board connector on the CPU board.
Matrix keyboard
The 4x4 matrix keyboard has been implemented with a PLD. One row at a time is set active “0" and
after that the state of each column is read. The state of the column ts “O” if the button corresponding
to that row and column is pressed.
ComWheel
Two pulse generators on the ComWheel change their states in turn when the wheet is rotated. The
direction of rotation is defined by companng the current state to the previous one. The PLD reads the
state of the ComWheel and generates the I/O_INT interrupt if the state of the ComWheel has changed.
The ComWheel push switch is connected to the matrix keyboard.
5-19
Cardiocap/5 Technical Reference Manual
* Alarm signals.
Alarm signals
In addition to audible alarms, the CPU board drives the Alarm LEDs on the monitor front panel and a
NURSE CALL signal on 44-pin 1/0 connector pin 5. The processor's I/O pins directly drive the Alarm
LEDs. The PLD generates the NURSE_CALL signal, which is in high state when the alarm is active. In
addition to the digital nurse-call signal, floating relay contacts are available on pins 11 and 12. The
relay located on the 1/0 board connects the pins when the nurse-call signal is active.
The interrupts generated by the DC/DC board come directly to the processor's 1/0 pins:
5-20
Frames and Software
In addition to the control signals, the supply voltages (VEE_S & V_DISP) for the display are available
on the display connector X7. The supply voltages for the display must be connected ON and OFF ina
certain order compared to the control signals. The display-controller chip controls voltage sequencing.
Display brightness is adjusted with channel D of the 8-bit DAC. The DAC output drives the brightness-
control circuit located on the Backlight board.
The voltage level for the display can be selected with jumper X14.
5.5.6 Memory
DRAM
The CPU board contains two parallel 16 x 4M DRAM chips. The supply voltage for the chips is 3.3V.
When starting the monitor, the program code is loaded into the DRAM from the FLASH memory as the
DRAM memory is much faster than FLASH memory, which is used as a code storage.
When switching off the monitor, the contents ofthe memory area reserved for the variables is copied
to the FLASH memory (8Mbyte). This way the data needed forthe Warm Start is easily stored.
To define if there should be Warm or Cold Start, the elapsed time since the last switch OFF is checked
when the monitor is switched ON. Depending on the elapsed time, data is loaded either from the Store
Flash or from the Code Flash.
An ELAN SC 410 processor generates the signals needed for controlling the DRAM chips.
FLASH
FLASH memory is used for three different purposes:
= Store Flash D9
This is “Warm Start” memory used to store the contents of the DRAM memory area reserved for
the variables when switching off the monitor.
= Boot Flash D9
The Boot Flash means that the uppermost block of D9 memory area is reserved for the Boot code.
The contents of the Boot Flash block cannot be changed without connecting the board to the
factory testing device.
5-21
Cardiocap/5 Technical Reference Manual
PC card
The PC card interface, a Cirrus PCMCIA controller CL-PD6722, drives two PC cards. It is connected to
the ISA bus of the processor.
The PC-Card interface accepts standard +3.3 V or +5 V cards. The controller selects the correct voltage
automatically. A separate VCC and VPP switching matrix connects the voltage to the cards.
Ethernet
The Ethernet controller, a National DP83907, is connected to the ISA bus of the processor. The
1OBASE-T (paired cable) Ethernet interface meets the IEEE802.3 standard.
5.5.8 Audio
Values corresponding to audio waveforms are programmed in PLD memory for generating the audio
signals. The values are fed at the desired speed to a Texas 8-bit DAC converter (TLC5620).
DAC channel A generates the audio waveform and channel B adjusts the volume. The frequency range
of the audio generator is 70Hz - 2500Hz.
Watchdog function
The processor's 1/0 pin |O_CS14 is connected to the Watchdog input of MAX 705. If the state of the
pin does not change at one-second intervals, the control circuit resets the board.
The purpose of the watchdog is to restart the monitor if there is a serious malfunction. This feature is
necessary in two cases: 1) when the main CPU's software is not able to control the monitor and
2) when the software controls the monitor but detects a serious malfunction.
5-22
Frames and Software
5.6.1 Display
NOTE: The LCD display backlight circuit runs on high voitage. Do not touch the inverter board
when powered.
The display consists of 10.4 inch TFT LCD display unit, backlight board, and inverter board.
The CPU board drives the display. The picture data, synchronization, and +5 V supply voltage are
connected from the CPU board to the display unit via the display adapter board.
The +12 V supply voltage connects to the backlight board and inverter board via the CPU board. The
CPU controls the display brightness with three data bits, which adjust the backlight lamp current. The
brightness data is converted to adjusting voltage on the backlight board, which controls the inverter
board output and the brightness of the backlight lamps. The backlight lamps are cold cathode lamps.
+12V
5V a CPU | > に
Board Brightness
33V Data 3 bit ba
Data+sync+5V
Backlights
The power indicator LED on the board is lit constantly when VDD is available. it flashes when battery
charging is active. It is connected directly to the DC/DC board.
The 2.6Ah 12V back-up battery connects to the DC/DC board through the Mother board. The board
contains a 10A fuse in series with the battery cable and a protection diode.
Two D25 connectors connect the airway gas and hemodynamic parameter units to the module bus.
The 26-pin flat cable connector is used for expansion options. A 10-pin flat cable connector connects
the module bus to the CPU board. The Mother board also provides supply voltages to the 1/0 board
through the 16-pin Molex connector.
5-23
Cardiocap/5 Technical Reference Manual
Serial buffers convert the serial TTL data coming from the CPU to RS-232 and vice versa.
A relay connected to the analog/ digital output connector (pins 11 and 12) provides a floating nurse-
call signal. The digital nurse call signal coming from the CPU is on pin 5.
5-24
Frames and Software
* Record the graphical or numerical trend and print up to 24 hours of trend data.
NOTE: The two-button recorder (shown) is for Cardiocap/5 monitors using software version 3.0 or
higher. A one-button recorder was available previously.
(1) Record Waveform/ Stop key. Press to start or stop recording selected real-time waveforms. This
key functions like the Record Start/Stop key on a one-button recorder.
(2) Record Trend/Stop key. Press to start or stop recording the selected graphical or numerical trend
data. Since this key is not available on a one-button recorder, use the options in the Record
Trends menu (Main Menu - Record/Print - Record Trends) to print trend data.
(3) Key to release and open the recorder paper compartment
(4) Recorder paper
When the Record Trend/Stop key orthe Record Waveform/Stop key on the recorder is pressed, the
command to start or stop recording is transferred to the CPU through the serial bus.
The recorder board links supply voltages from the DC/DC board and data from the CPU board to the
recorder. A FET circuit delays the rise. A choke and a 10,000 uF capacitor filter the 12 V voltage. The
delay limits the current peaks from the power supply.
Communication with the main CPU board takes place through +5 V CMOS-level RS232 serial
interface. The recorder supply voltage of +12 VREC is switched on after +5 V is present.
5-25
Cardiocap/5 Technical Reference Manual
5-26
Contents
6.3 Measurement principles for airway gases, spirometry, and NMT.. 6-8
6.3.1. CO02, N20, and agent MEASUTEMENL eren 6-8
6.3.2 02 measurement... ee 에
6.3.3 Patient spirometry measurement
Airway pressure M.M ine
Aiway HOW eee erener sararan
Compliance and airway resistance us 6-11
D-lite™ Flow SENSOT esasan 6-11
6.3.4 NeuroMuscular Transmission (NMT) measurement eeeeeeeee nenene nene 6-12
Nerve stimulation
Tetanic/Post-Tetanic COURL (PTC) een 6-12
RESpoNSe...... ete nné 6-12
Regional block eee 6-13
Watchdog timer
Valves... алина
Powersuppiysection........................ eee eee 6-16
Cardiocap/5 Technical Reference Manual
Table of Figures
Figure 6-1. Hemodynamic parametersmeasurementunit(NESTPR)............................. 6-1
Figure 6-2. Airway gases measurement unit een nete 6-2
Figure 6-3. Absorption of infrared light in the finger... nn 6-4
Figure 6-4. Comparative light absorption... 6-6
Figure 6-5. Extinction versus wavelength graph........... ns 6-6
Figure 6-6. TPX sensor principles 6-8
Figure 6-7. Absorbance of N,0 and CO»... nn 6-8
Figure 6-8. Infrared absorbance of AAs
Figure 6-9.O,measurementprinciple...............................
eee eee
Figure 6-10.Pnnciple of response measurement лилии 6-12
Figure 6-11. Serial communication and opto isolation............ iii iene 6-14
Figure 6-12. NIBP board functional block diagram iene 6-15
Figure 6-13.ECGboardblockdiagram.................................. eee 6-17
Figure 6-14. STP board block diagram een 6-20
Figure 6-15. Pulseoximetrymeasurementblockdiagram............................. 6-21
Figure 6-16. Temperature measurement principles 6-22
Figure 6-17. Pressure measurement principle...
enem 6-23
Figure 6-18. SpO2 interface boargd bjock diagrarn iii 6-26
figure 6-19. Gas Sampling System (N-XC 0ptig0
站 un 6-28
Figure 6-20. Gas sampling system (N-XCO and N-XCAIO Options) ee 6-29
Figure 6-21. Gas tubing 18YOUŤ ui 6-30
Figure 6-22. TPX MeaSUTING UNE inner R Pet 6-31
Figure 6-23. OM measuring unit... eee 6-31
Figure 6-24. PVX measuring Unit... inner 6-32
Figure 6-25.Signalprocessinf................................. emme 6-33
Figure 6-26.Controllogic.............................
eee rre 6-33
Figure 6-2 7. Calibration data stored in EEPROM.............. iene 6-33
Figure 6-28. NMT serial communication and opto isolation...
ie 6-34
Figure 6-29. NMT stimulator block diagram ss 6-34
Measurement Parameters
6. MEASUREMENT PARAMETERS
6.1 Introduction
This chapter provides information about the Cardiocap/5 standard measurement features and all
measurement options.
The pulse oximetry options listed below require hardware and processing that is not part of the
NESTPR unit:
6-1
Cardiocap/5 Technical Reference Manual
N-XC ο
N-XCO e ο 9
N-XCAiO ο e e e e
N-XV * e
N-XNMT ** e
6-2
Measurement Parameters
6.2.1 NIBP
NIBP (Non-Invasive Blood Pressure) is an indirect method for measuring blood pressure.
The NIBP measurement is performed according to the oscillometric measuring principle. The cuff is
inflated with a pressure slightly higher than the presumed systolic pressure and deflated at a speed
based on the patient's pulse. Data are collected from the oscillations caused by the pulsating artery.
Based on these oscillations, values for systolic, mean, and diastolic pressures are calculated.
6.2.2 ECG
Electrocardiography analyzes the electrical activity of the heart by measuring the electrical potential
produced with electrodes placed on the surface of the body.
ECG reflects
6.2.4 Temperature
Temperature is measured by a probe whose resistance varies when the temperature changes, called
NTC (Negative Temperature Coefficient) resistor.
* Applying a constant voltage across the resistor and measuring the current that flows through it.
= Applying a constant current through the resistor and measuring the voltage that is generated
across it.
The two methods are combined in the form ofa voltage divider. The NTC resistor is connected in series
with a normal resistor and a constant voltage is applied across them. The temperature dependent
voltage can be detected at the junction of the resistors, thus producing the temperature signal from
the patient. The signal is amplified by analog amplifiers and further processed by digital electronics.
6-3
Cardiocap/5 Technical Reference Manual
The transducer is a piezo-resistive device that converts the pressure signal to a voltage. The monitor
interprets the voltage signal so that pressure data and pressure waveforms can be displayed.
Intensity of 4 |
transmitted I |
MAMA 1
1
1
i
|
“min
AC-component
i
t
ーーーー . 1т.
l Variable absorption
T tted ; 1
light e 4 1 due to pulse added
I
’ |
volume of arterial
| blood
Arterial blood
Venous blood
ui
De
X
N
TOTS
、
alata
X
Tissue
HEN
>.
M
トー
11111 1 1 |
- 一 一 Time
M
6-4
Measurement Parameters
A general limitation of the above pulse oximetry principle is that due to the use of only two
wavelengths, only two hemoglobin species can be discriminated by the measurement.
Modern pulse oximeters are empirically calibrated against fractional saturation (SaO,frac) or against
functional saturation (SaQ,func) as shown below:
Sensor
SpO> sensors contain the light source LEDs, which are located opposite the photodiode detector.
Different kinds of sensors are available from Datex-Ohmeda, including clip-on and wrap styles.
Pulse rate
The pulse rate calculation is done by peak detection of the plethysmographic pulse wave. The signals
are filtered to reduce noise and checked to separate artifacts.
s The light source consists of red and infrared light-emitting diodes (LEDs).
The two light wavelengths generated by the sensor light source (the red and infrared LEDs) pass
through the tissue at the sensor site. The light is partially absorbed and modulated as it passes
through the tissue.
Arterial blood pulsation at the sensor site modulates transmission of the sensor's light. Since other
fluids and tissues present generally don’t pulsate, they don't modulate the light passing through that
location. The pulsatile portion of the incoming signal is used to detect and isolate the attenuation of
light energy due to arterial blood flow.
6-5
Cardiocap/5 Technical Reference Manual
4 a OX 5 Variable absorption
+ SSIS
> Tee3 LoIDgar ÉU À — (Due to arterialial pulse)
pul
aa 2506)
TIN
ota?
a
2022
一 Arteriali blood absorption
1
5 pa ened eae НН it = i
Venous blood absorption
= レン ΕΣΑΣ ΝΑΟΙ
5 K
【
«a
<
e—— Other tissue absorption
Time
The sensor's photodetector collects and converts the light into an electronic signal. Since 0,Hb and
HHb allow different amounts of light to reach the photodetector at the selected wavelengths, the
electronic signal varies accordingto which light source is “on” (red or infrared) and the oxygenation of
the arterial hemoglobin. The oximeter uses this information to calculate the relative percentage of
OoHb and HHb.
1
(Red) (Infrared)
660 nm 940 nm
\
o
Extinction (10%)
1
| \ 1
/
x
jo
2
|
\
\
HHb
2 1 | ι | 1 1 1 ]
600 700 800 900 1000
Wavelength (nm)
The photodetector sends the electronic signal, which contains the light intensity information, to the
oximeter. The oximeter’s electronic circuitry processes the electronic signal, calculates the SpO2 and
pulse rate values, and displays them on the screen.
Calibration
Datex-Ohmeda enhanced pulse oximetry uses two wavelength ranges, 650 nm to 665 nm and 930
nm to 950 nm, both with an average power of less than 1 mW. These wavelengths are used to
calculate the presence of oxyhemoglobin (O2Hb) and reduced hemoglobin (HHb).
6-6
Measurement Parameters
A CO-oximeter typically uses four or more wavelengths of light and calculates reduced hemoglobin
(HHb), oxyhemogiobin (O2Hb), carboxyhemoglobin (COHb), and methemoglobin (MetHb). Therefore,
pulse oximetry readings and CO-oximetry readings will differ in situations where a patient’s COHb or
MetHb are increased. Increased patient COHb leads to falsely increased SpO in all pulse oximeters.
Datex-Ohmeda enhanced pulse oximetry uses functional calibration, which is represented
mathematically as the percentage of hemoglobin capable of carrying oxygen that is carrying oxygen.
02Hb 02Hb
Functional Sp0 = | —————————— ) x100= | ———— | x 100
Hbrora - COHb - MetHb Ο2Η0 + HHb
The calculation of SpO2 assumes 1.6% carboxyhemoglobin (COHb), 0.4% methemoglobin (MetHb),
and no other pigments. These values are based on the Datex-Ohmeda Pulse Oximeter Empirical
Calibration Study. Appreciable variation from these values will influence SpQ5 accuracy.
6-7
Cardiocap/5 Technical Reference Manual
Temp Sensor
Sample
Chamber
Thermopile
Detectors
Sample Sample
Gas Out Gas In
TPX is a nondispersive infrared analyzer that measures absorption of the gas sample at seven infrared
wavelengths that are selected using optical narrow band filters.
The infrared radiation detectors are thermopiles.
Concentrations of CO, and N,O are calculated from absorption measured at 3 to 5 um.
(arbitrary units)
Absorbance
4.0 4.5
Wavelength (um) 圖
6-8
Measurement Parameters
БезНигапе — >
—,
Nm
ox
(+ = |
E =
se
os
BE
«<P. Enflurane ------- >;
= ーー デー
: !
V Halothane ——__>] |:
ALÍK er ce eta a ee 그 ec
Wavelength (um)
6.3.2 02 measurement
The differential oxygen measuring unit uses the paramagnetic principle in a pneumatic bridge
configuration. The signal picked up with a differential pressure transduceris generated in a measuring
cell with a strong magnetic field that is switched on and off at a frequency of 165 Hz. The output signal
is a DC voltage proportional to the difference in 0, concentration between the two gases to be
measured.
Flectromagnet
Mixture Out
y
Sample in Y
い
Reference
In P
6-9
Cardiocap/5 Technical Reference Manual
Airway pressure
= Peak pressure (Pra)
Airway flow
= Real time flow waveform (V'}
= Compliance (Compl)
The kinetic gas pressure is measured using the Pitot effect. A pressure transducer measures the Pitot
pressure. The obtained pressure signal is linearized and corrected according to the density of the gas.
Flow speed is calculated from these pressure values, then the tidal volume (TV) value is integrated. The
minute volume (MV) value is calculated and averaged using the TV and respiratory rate values.
6-10
Measurement Parameters
n Li
TV. Sp -
Compl
У =
2xdP
Ν р
where: у = уеюсйу (т/зесопа), dP = pressure difference (cmH,0), and p = density (kg/m?)
6-11
Cardiocap/5 Technical Reference Manual
Nerve stimulation |
There are three NMT stimulus modes: Train of Four (TOF), Double Burst 3,3 (DBS) and Single Twitch (ST).
* Train of Four stimulus generates four stimulation pulses at 0.5 second intervals. The response is
measured after each stimulus and the ratio of the fourth and first response of the TOF sequence is
calculated (TOF%). If the first response does not exceed a certain signal tevel, TOF% is not
calculated due to poor accuracy.
* Double burst (3,3) stimulation includes two bursts with a 750 ms interval. Both bursts consist of
three pulses separated by 20 ms intervals. The responses of both bursts are measured, and the
ratio of the second and first response is calculated (DBS%). EMG responses are measured
immediately after the first stimulus pulse of both bursts.
e Single Twitch stimulation generates one stimulation pulse. The response is measured after the
stimulus. To prevent decurarization of the stimulated area, the measurement is automatically
stopped after 5 minutes stimulation in 1-second cycle time.
If the TOF, DBS, or ST measurement cycle was on when tetanic stimulation started, the cycle will
continue after the PTC. After completing the PTC measurement during 1 minute TOF, DBS, or another
PTC measurement is not possible. This is to avoid erroneous readings due to post-tetanic potentiation.
Response
ーー 本 A/D
Timing control | and
_ EMGleadsoff | _| Microprocessor
© Mechano-
Sensor
6-12
Measurement Parameters
Before each stimulation, the sequence offset, noise, and threshold for response detection is
measured. Offset is a baseline of the noise measurement. Noise is calculated by the same algorithm
as the response signal itself. The response detection threshold is calculated based on the noise. If the
response is not greater than the threshold, it is interpreted as no response.
The EMG response is measured as integrated muscle activity. The EMG measurement starts 3 ms after
the stimulation and lasts 15 ms. The 3-ms delay helps prevent the effect of stimulation artifact.
When using a MechanoSensor, response is measured as movement of the thumb, which is the area of
positive signal.
Regional block
A regional block cable can be used as a nerve locator in local anesthesia. A maximum current of
5.0 mA is given every second, every other second, or every third second. The response measurement
is ocular.
6-13
Cardiocap/5 Technical Reference Manual
NIBP Board 」
Receive Data Data
« Receive Data estes 5
5 Send Data river NData 、| る =
a Send Data | E 8
©
8 Send/Receive Send/Receive è
个 2 E
a © o
« Reset | Reset in 品 à
β-
Reset RS485 NResetin | の
Driver
ECG Board P
. : i . 5
< Receive Data | ; ; Receive Data 8
STP Board
Receive Data Data
Receive Data Г RS485
Efe Dri Iver NData+„| 8©
52 я ; Еo |
Data
Send A —— E
Send Data , i
| の Resetin | ©
mes ©
o 2 Reset RS485
: Patient i n Driver bq NResetin | 2
i Isolation :
|
Serial communication between the NESTPR unit and the CPU is done by a RS485-type bus whose
buffers get their supply voltage (+5 VDC) from the DC/DC board. In the isolation section, their supply
voltage (+5 V) is obtained from the isolated power supply. The data transmission rate is 500 kbps.
The serial communication buffers are also controlled by the Reset signal so that when the Reset is
active, the buffer does not transfer data. Reset is also an RS485. Additionally, an auxiliary logic power
reset keeps the reset active for about 500 ms despite the state of reset in the module bus. Time
constant determines the power-up reset time. Components prevent the unit from sending data during
reset.
ECG board communication with the module bus is made through RXD and TXD pins.
6-14
Measurement Parameters
Pump
and 2
Valve Driver 一 一一
. 는 L AD |… + 15 VD
Lu. τν | a Li =, {Converter 一
TN |
L- - a Address Bus |
Powerupl — CPU ペデ |
Reset 80C51FA Address ーー | | ーーー
Decoder | | | |
|
5 A if RS485 las | | k JU |
3p “ {Interface
\ ) Ji Front
3 Address | RAM EPROM Panel | [Software
き| 1 Internal Latch Jr Keys on
2 EEPROM |- Watchdog = -一 -一 |
© (Calibration fy, , | | | | | | | | |
E Data) | sell poe |
の | Databus
Write Protect
Switch
Pressure transducers
The NIBP board contains two piezo-resistive pressure transducers (B1 and B2). Transducer B1
measures the pressure of the blood pressure cuff and the pressure fluctuations caused by arterial wall
movement. Transducer B2 detects the cuff hose type as well as “Cuff loose" and "Cuff occlusion"
conditions. The transducers are temperature-compensated internally. They are supplied by a constant
voltage and their output voltage changes up to 40 mV maximum (50 kPa, 375 mmHg).
Signal processing
The NIBP board is controlled with 80C51FA microprocessor at 16 MHz oscillator frequency. The
microprocessor system ts equipped with its own power-up reset in addition to the external RS485
reset line.
Two signals from the pressure transducers are amplified and sent to the A/D converter. After the
converter, digitized signals are sent to the microprocessor for data processing. Before the converter,
one signal adjusts the offset to the pressure safety level.
6-15
Cardiocap/5 Technical Reference Manual
Software control
Software controls the valves and pump. In addition to ON/OFF signals for each component, a common
power switch for the valves and the pump is used at pump/valve failures.
Memory
NIBP program memory (EPROM) size is 128k x 8. The RAM (32k x 8 bit) stores variable NIBP
measurement values. The EEPROM (64 x 16 bit) stores the calibration values for the pressure
transducers, the pulse valve constants gained during measurements, the PC board identification, and
the module serial number.
Watchdog timer
The NIBP board is equipped with a software-independent safety circuit to disconnect supply voltages
from the pump and the valves if the cuff has been pressurized longer than preset time. The pressure
limit is specified to 5 mmHg. As soon as the cuff pressure rises over 5 mmHg, a timer starts counting.
The timer is adjusted to stop the pump and open the valves in 2 minutes 10 seconds in adult/child
mode and in 1 minute 5 seconds in infant mode.
Valves
Exhaust valves 1 and 2 empty the cuff and the joining chamber after the measurement. Exhaust valve
1 is used as a safety valve in infant mode; it opens at 165 mmHg. Exhaust valve 2 is used as a safety
valve in adult mode; it opens at 320 mmHg.
6-16
Measurement Parameters
| Overload Protection |
ーー =“
R L F C N
Defibrillation
Detector
Pod
=
HP Fiter | ¡| Hp Fiter |
pp | ib
Anti-Alias | Ani Alas
AID
Microprocessor
(uP)
Supply Voltages
Isolation
“EE Cece ве __
σσ Ορίο- quer | ea ECG
spa Board 。
Figure 6-13. ECG board block diagram
6-17
Cardiocap/5 Technical Reference Manual
Analog switches connect leads to amplifiers. The state of the switches depends on the cable type.
Lead-off, noise, and pacemaker are detected by a slew rate detector.
Lower frequency is determined by a high-pass (HP) filter 0.5 Hz (monitor bandwidth) or 0.05 Hz
(diagnostic or ST bandwidth).
ECG filtering
The Cardiocap/5 monitor has three ECG filtering modes:
MONITORING 0.5 to 30 Hz (with 50 Hz reject filter)
0.5 to 40 Hz (with 60 Hz reject filter)
DIAGNOSTIC 0.05 to 100 Hz
ST FILTER 0.05 to 30 Hz (with 50 Hz reject filter)
0.05 to 40 Hz (with 60 Hz reject filter)
Filtering reduces high-frequency noise and low-frequency (for example, respiratory) movement
artifacts.
The monitoring filter is used in normal monitoring. The diagnostic filter is used if more accurate
diagnostic information is needed. The ST filter gives more accurate information of ST segment, but
reduces high-frequency noise.
The high-pass filters 0.5 Hz and 0.05 Hz are done with hardware. The monitor sends a command to
the NESTPR unit determining which corner frequency (0.5 Hz or 0.05 Hz) is to be used.
The 50 Hz and 60 Hz reject filters are low-pass filters with zero at 50 Hz or 60 Hz correspondingly and
they are done with software. They are for the mains supply filtering. When these filters are used, 3 dB
value for low-pass filter is 30 Hz or 40 Hz.
Software filters are not used in diagnostic mode. Then, the upper frequency is limited by hardware and
the -3 dB frequency is 100 Hz.
Respiration section
Analog switches control the current supply source of the impedance respiration measurement. The
lead selection for the 3-lead cable can be seen from the following table:
6-18
Measurement Parameters
When the 5-lead cable is used, the current source is between L-F and the signal is measured from the
N, independently on the lead selection.
The respiration amplifier consists of the operational amplifiers and the components around them. An
analog switch controls the gain of the first stage of the preamplifier.
The synchronous rectifier consists of the analog switches, which detect the respiration signal from 31
kHz amplitude modulated raw signal.
The amplifier stage consists of the differential amplifier and the last amplifier. The differential amplifier
consists of the operational amplifiers and the components around them. This stage is AC-coupled on
both sides for minimizing the offset voltages.
The last amplifier amplifies the signal derived from the differential amplifier stage.
The respiration signal is zeroed at the beginning of the measurement. Zeroing is also used for fast
recovery of the measurement after the motion artifact. This is done in the amplifier section.
NOTE: The respiration measurement is switched OFF for 20 seconds when defibrillation is detected at
the defibrillation detector.
Microprocessor section
The microprocessor contains RAM and EPROM. The processor uses external EEPROM memory. The
microprocessor's internal 8-channel A/D converter converts the ECG signals to digital form.
Isolated section
See STP board later in this chapter.
The patient isolation of ECG is 5 kV.
6-19
Cardiocap/5 Technical Reference Manual
Patient Connectors
Gain Control
Temp. AD Pressure AD
Nonvolatile Memory |
Microprocessor Unit Intensities
2K RAM (internal)
16K RAM (external) Module Bus Data
48K EPROM
Power anti
RS Communication Reset
isolation TT o
Section
Power
| 1
Non-isolation RS-485 Driver RS-485 Driver
Power Reset for for
Section
Module Reset . Data
Power for Module Power for Communication Module Reset Module Data
To NIBP Board
Microprocessor unit
The STP board uses an Intel 80C196KC-16 processor with three A/D converters, external memories,
an 8-bit data bus, a 16 MHz oscillator, an open collector reset, and a watchdog timer. The processor's
internal UART communicates with the CPU board. High speed 1/0 obtains the pulse control sequence
for pulse oximetry measurement. The oscillator provides its timing clock.
6-20
Measurement Parameters
Isolation section
The outputs of the two opto isolators are analog signals in the isolated section, however, the signal is
processed on logical high-low level. Reset line is an open collector type with a pull-up resistor,
allowing the processor to use its internal watchdog function.
1-5-350 MA intensi
LED driving circuit — | ad
IR DC level >
G = 1/4096-1 G=275
Sf ¡Red AC signal,
T F Red AC signal
=
_ RedD(DC level
6-21
Cardiocap/5 Technical Reference Manual
I Offset
6-22
Measurement Parameters
Pressure amplification is realized in the instrumentation amplifier. Gain of the amplifier is set so that
the level of the signal transferred to the A/D converter stays within the measurement range even when
there are circumstantial offsets or offsets caused by the transducer. A filter before the amplifier
attenuates high frequency disturbances.
Pressure Transducer
Vin
| . | 一 | | Instrumentation Amplifier
6-23
Cardiocap/5 Technical Reference Manual
‘lemon
1 +5IN
5 | COMMON
(2 İm
2 COMMON
NO CONNECT
UMLINC |
=—_
ne
Pin 1
7 UNUSED 8 TX- TRANSMIT -
ーー . Connector viewed from solder
9 RX- RECEIVE 10 CTS side of board.
13 NO CONNECT 14 COMMON
6-24
Measurement Parameters
= The connector board/cable assembly consists of the patient connector, the connector board, and
the cable that connects to the pulse oximetry board. The connector board contains the sensor
interface connector, a differential preamplifier, and EMI protection. It provides analog signals to
the pulse oximetry board from the oximeter sensor through a preamplifier inferface cable.
The measured Sp0,, pulse rate values, and status information are transferred from the pulse oximetry
board to the SpO, interface board. Communication between the pulse oximetry board and the SpO,
interface board is established through an RS-232C serial interface. The SpO, interface board transmits
the measurement information to the module bus of the monitor through an RS485 serial interface.
6-25
Cardiocap/5 Technical Reference Manual
Microprocessor .. Power;
1024 byte internal RAM, 32k byte internal ROM; 16MHz isolated section
|
+15V and - 15V
Nonvolatile supply voltages
UART memory to oximetry board
Bidirectional asynchronous
serial communication
to oximetry board
NOTE: Two switches located on the board are not used and have no function with the Cardiocap/5.
RS485 drivers
The Sp02 interface board contains drivers for data and for module reset functions. They drive the
RS485-type serial communication bus between the pulse oximetry unit and the Central Unit. Data
transmission speed of the bus is 500 kbps.
In addition to RS485 bus RESET, a Power-up reset keeps the RESET pin of the CPU active during power
up for about 500 ms despite the state at the RS485 bus RESET. This is used to prevent the sending of
RS485 data during the RESET of the module.
6-26
Measurement Parameters
Opto isolation
The serial communication bus signals between the pulse oximetry board and the Central Unit are
transferred through the patient isolation section by high speed opto couplers.
The UART provides bidirectional asynchronous serial communication between the pulse oximetry
board and interface board CPU.
Nonvolatile memory stores identification information, such as serial number, control number, date, etc.
13(0000000000000) i
25 N000000000000/ 14
Test connector
Pin No Voltage Name Note
1 +5V +5VTEST Supply voltage to the Sp02 interface board
M 2 +5V +5Vn Supply voltage to the pulse oximetry board
3 + 15V +1 5Vn Supply voltage to the pulse oximetry board
4 GND | FGND
5 - -15V - 15Vn
6 - N/C |
6-27
Cardiocap/5 Technical Reference Manual
PVX measuring unit and PVX board for Patient Spirometry (N-XV option)
CPU board
Pneumatic Unit
D-Fend Water Trap
OD
OUT | IN
> JO
INC O Gi
Damping Chamber
OUT IO
Sampling Pump | <
6-28
Measurement Parameters
OM Unit
Pneumatic Unit
O D-Fend Water Trap
(0%
Pressure
< 一 一 一 NafionTube
Transducers OM Ref. Filter
OO)
81 Gas Outlet Connector
T
TPX Unit
OUT
IN
Damping Chamber
Sampling Pump
D-fend™
The sample is drawn through a sampling line to the water trap, where it is Separated into a main flow
and a side flow. The main flow is drawn through into the analyzers. A hydrophobic filter separates this
flow from the patient side. The side flow creates a slight subatmospheric pressure within the D-fend
water trap, causing fluid removed by the hydrophobic filter to collect in the bottle.
Zero valve
The main flow passes through a magnetic valve before proceeding to the analyzers. This valve is
activated to establish the zero points for the TPX and OM units. When the valve is activated, room air is
drawn into the internal system and the gas sensors.
Nafion"" tube ”
A Nafion tube between the water trap and the zero valve balances the sample gas humidity with that of
ambient air. The tube prevents errors caused by the effect of water vapor on gas partial pressure when
humid gases are measured after calibration with dry gases.
Gas analyzers
After passing through the zero valve and Nafion tube, the gas passes through the TPX and the OM
units. The oxygen sensor has two inputs. One input accepts the main flow and the other draws in room
air for reference. Both gas flows exit from a single port.
6-29
Cardiocap/5 Technical Reference Manual
Pneumatic unit
The pneumatic unit contains a zeroing valve, an occlusion valve, and tubing connections. A series of
restrictors and chambers form a pneumatic filter to prevent pressure oscillations from the pump to
reach the measuring units. Zeroing and occlusion valve connections to room air include a dust filter.
Occlusion valve
The valve is activated when the sampling line becomes occluded. The main flow is diverted to the side
flow of the D-fend water trap to remove the occlusion quickly.
Pump
6-30
Measurement Parameters
OM board
The Oxygen board contains the specific electronics for the oxygen sensor. Sample flow measurement
and sampling system pressure sensors are on this board. It also contains EEPROMs that store
calibration data of both the IPX and OM sensors.
6-31
Cardiocap/5 Technical Reference Manual
6.6.4 PVX measuring unit and board (Patient Spirometry, N-XV option)
NOTE: Never apply overpressure or negative pressure of more than 300 cmH,0 to the flow and volume
tubing. Differential pressure maximum is 25 cmH,0 on one port at a time (when connecting tubes, for
example).
Patient Spirometry provides patient respiration monitoring capabilities using D-lite and Pedi-lite flow
sensors. When Patient Spirometry is used, a special sensor (D-lite) replaces the normal airway
adapter in the patient circuit. Double lumen tubing is attached to the two connectors on the adapter
and on the monitor front panel.
The measurement unit consists of airway connections, two pressure transducers, valves, and
preamplifiers. The preamplifiers connect to the A/D converter on the main monitor CPU.
Measurement is based on measuring the kinetic gas pressure and is performed using the Pitot effect.
A pressure transducer measures the Pitot pressure. The signal is linearized and corrected according to
the density of the gas. Flow speed is calculated from the pressure and TV is integrated from it.
The breathing flow of a patient passing through the D-lite adapter creates a pressure difference. This
pressure difference is measured by pressure transducer B1. Overpressure and negative pressure in
airways are measured by pressure transducer B2.
PVX board
The Spirometry board connects to the Oxygen board. It contains pressure sensors for airway pressure
and flow measurement differential pressure, and the preamplifiers for them. An EEPROM stores
spirometry calibration data.
6-32
Measurement Parameters
ーー | 때 Lamp Current I
E Sensing |
Fan
一] 5
MUX A/D a
and - 5
ВиНег v À 4 3
TT, =
Gas Press -— po A/D
iff | MUX R |
Diff Press — Temp Sensor | m CPU eset +
OM Board < 80C196NT Data
I
|
PVXBoard | PVA Valves 1
PR 77707 m
Factory calibration EEPROM | 2
data for OM and TPX PRON | 77 CPU __ | 85485 2το
80C196NT Driver | |
PVX Board AE |
Factory calibration | |
data for PVX [ eeprom | — |
= Ly
6-33
Cardiocap/5 Technical Reference Manual
» Microprocessor
for stimulation and measurement control, and for counting measurement results.
= Serial communication.
The serial bus speed is 500 kbps and the bus itself is half duplex (data can be transferred in both
directions but only one way at a time).
Serial communication
NMT Board
Receive data |
5 KV | RS-485 o
9 q Receive data Send data Driver 8
떻 Send data » | | 2
3. Send/Receive Opto Isolation Send/Recei 5
三
et | Resetin 5
RS-485 3
Driver a.
Patient
Isolation
Stimulator
The constant current stimulator generates pulses whose amplitude is independent of the load. The
main components of the stimulator are a transformer, a capacitor and a transistor. The transformer
produces a high voltage which charges the capacitor and the transistor adjusts the pulse width and
amplitude of the current.
Power
— y supply
Current
80C196KD_
Constant
4 Adjustment |. PWM
current HSO
Current
generator
|
control o τε κν οσον
lo
Continuous
stimulation a
congestion
6-34
Measurement Parameters
OMAN =
EMG signal -
Not used
Stimulus +
Stimulus -
©
©
Ground
Not connected
LU
Sensor identification
OG
+5V
Mechanical signal
a
©
3(0000000000000) 1
25 N000000000000/ 14
6-35
Cardiocap/5 Technical Reference Manual
6-36
Contents
Table of Figures
Figure 7-1. CPU battery replacement eee erener teta nene 7-6
Figure 7-2. SpO2 interface DO . . . . {1 qJg0.6.66 0.000 0000000008 1-8
Figure 7-3.NIBPboard..................................
ーー 7-10
Figure 7-4. Calibration SWitCh resen RES 7-12
Figure 7-5. Attaching the regulator ee 7-17
Figure 7-6. Connecting a sampling line and feeding gas... 7-17
Service Procedures
7. SERVICE PROCEDURES
For service advice from Datex-Ohmeda, please provide the unit serial number, full type designation,
and a detailed fault description.
CAUTION: Only trained personnel with the appropriate tools and equipment shall perform the tests
and repairs outlined in this chapter. Unauthorized service may void warranty of the unit.
CAUTION: Electrostatic discharge through the PC boards may damage the components. Before
replacing PC boards, wear a static-contral wrist strap. Handle all PC boards by their non-conductive
edges and use anti-static containers when transporting them.
CAUTION: When reassembling the monitor, make sure all cables and tubing are reconnected properly
and are not pinched between other components.
1. Insert the service software PC card into the data card slot closest to the display.
2. Poweron the monitor. Both LEDs illuminate and the screen remains blank for approximately two
minutes while loading.
3. When the normal monitoring screen is displayed, remove the card.
7-1
Cardiocap/5 Technical Reference Manual
Tools
= Screwdriver (pozidrive)
s Hexagon driver (3.0 mm)
e Pliers
= Cable tie
CAUTION: When reassembling the monitor, make sure all cables and tubing are connected properly.
Disconnect the Battery- Mother board (red & black color) cable from the Mother board.
Disconnect the 1/0
- Mother board flat cable from the Mother board.
wo
Disconnect the |/O-CPU board flat cable from the CPU board.
ὃν
Disconnect the expansion unit-Mother board cable (if N-XNSAT, N-XOSAT, or NMT is installed).
の
Remove the 4 screws that hold the Mother board and the metal bracket to the inner-parts unit.
の
Twist straight the metal clamp that holds the Mother board on the inner-parts unit.
7. Move the Mother board with the metal bracket out of the way.
7-2
Service Procedures
Remove 4 screws from the corners of the STP board and detach the STP board.
Push the backplate unit forward and bend the retainers in both sides to loosen the unit.
Carefully pull out the backplate unit.
Disconnect the CPU-NET board cable from the 1/0 board.
Disconnect the fan power cable from the DC/DC board.
OI
To remove the sample gas outlet tube, unfasten the lock washer from the gas outlet connector
(X6) and pry the connector out ofthe cover.
7-3
Cardiocap/5 Technical Reference Manual
Unfasten the 3 screws and washers that hold the locking chock and remove the chock.
if installed, disconnect the hook-and-loop strap.
Disconnect the 2 tubes from the rear of the airway gas unit.
oF
Fan replacement
NOTE: You need to detach the pump unit to access the fan power cable.
1. Detach the fan by carefully bending the fastening springs.
2. Disconnect the fan power cable from the CPU board.
7-4
Service Procedures
Disconnectthe DC/DC-Motherboard (red & black color) cable from the Mother board.
Disconnectthe DC/DC-Battery cable from the DC/DC board.
Disconnect the Fan power cable from the DC/DC board.
お
Remove the 3 screws from the heat sink from the backside of the inner-part unit.
の
Remove the 6 screws that hold the board to the holder plate from the corners of the DC/DC board.
の
7-5
Cardiocap/5 Technical Reference Manual
Before replacing the battery, determine the monitor software version. Look on the small sticker on the
rear of the monitor. The software version follows the dash after the part number (for example, in
898716-1.0, the monitor software is version 1.0; in 8987 16-2.4, it is version 2.4). If you cannot
determine the software version from the sticker, look at the Service View screen: Press the ComWheel
and select Monitor Setup - Install/ Service (16-4-34) - Service (26-23-8).
For version 1.0, you must upgrade to the latest monitor software (see the Spare Parts chapter).
For version 2.4 or earlier, after replacing the battery you must reconfigure the monitor and, if the
N-XNET or N-XDNET option is installed, load the N-XNET or N-XDNET software (see Spare Parts).
For version 2.5 orlater, replace the battery as described below.
Turn off the monitor and disconnect the power cord. Remove the cover. Use a screwdriver to
carefully detach the battery from the D12 circuit on the CPU board. Align and install the new
battery as shown below.
Perform a factory reset (from the Service View menu, select Monitor - Factory Reset).
Set the time and user configurations. See the User's Reference Manual.
Turn off the monitor for 5 minutes. Turn it on and check that the time and user configurations are OK.
7-6
Service Procedures
7-7
Cardiocap/5 Technical Reference Manual
— 路
OSATO OSAT ©
NSATO
Header for
33 on OSAT board
ΠΠΠΠΠΠΗΗ CO
NSAT Connector OSAT Connector
NSATO NSAT ©
Jumper installed for OSAT
Jumper removed for NSAT
OSATO OSAT ©
Not used 一 可 |]
7-8
Service Procedures
7-9
Cardiocap/5 Technical Reference Manual
To enter the Service View menu, from which you access all other service menus:
1. Press the ComWheel and select Monitor Setup.
2. Select Install/ Service and enter the password (16-4-34).
3. Select Service and enter the password (26-23-8)
The Service View menu is displayed.
NOTE: Refer to the Service Menus chapter for illustrations and detailed explanations of the service
menus and screens.
Tube to transducer B2
Timmer
Tools
* Troubleshooting extension cable (884298)
1. Connect the NIBP board to the NESTPR connector on the mother board using the troubleshooting
extension cable.
2. Switch onthe monitor.
To enter the NIBP Pneumatics menu from the Service View menu, select:
Modules
NIBP
Pneumatics
4. Apply 5 mmHg pressure to pressure transducer B2 with a precision manometer and adjust the
trimmer to change the ADS value on the screen from negative to positive.
7-10
Service Procedures
if the error for pressure channel B1 is larger than specified above, recalibrate.
The error for pressure channel B2 may be twice as large because it does not effect blood pressure
measurement accuracy. When the B2 error is larger than specified above, however, recalibration is
recommended to ensure the best possible operation.
7-11
Cardiocap/5 Technical Reference Manual
CAUTION: Turning the gear wheel more than 1/4 turn can damage the calibration switch.
Calibrating switch
levers must be placed
in the slot of the slide.
Detach the NIBP hoses from the front panel connector to enable proper zeroing.
To enter the NIBP Calibration service menu from the Service View menu, select
Modules - NIBP - Calibrations.
lf Protection is OFF and Calibration is highlighted, go to step 7. Otherwise, continue with step 4.
Place the monitor on its left side. Insert a medium-sized, flat-head screwdriver into the hole that
accesses the gear wheel for the calibration protection switches (see Figure 7-4).
Turn the gear wheel approximately 1/4 turn counterclockwise until Protection is highlighted on
the screen. Select Protection OFF and push the ComWheel.
Turn the gear wheel approximately 1/4 turn clockwise until Calibration is highlighted. The
"Calibration not protected” message appears in the NIBP digit field.
With Calibration highlighted, push the ComWheel to start the calibration. The "Zeroing” message
will appear. After successful zeroing, the pressure adjustment bar will appear.
Connect an external mercury manometer with pump to the NIBP connector through both tubes
of the hose (transducers B1 and B2 must be calibrated simultaneously). Pump up to about
200 mmHg pressure, according to the manometer, to the NIBP tubing. Calibration is possible
in the range of 150 to 300 mmHg.
Verify that both pressure values on the screen match the manometer reading. If not, adjust by
turning the ComWheel. When the values are equal, push the ComWheel to confirm the calibration.
When the calibration data has been saved, the "Calibrated" message will appear.
10. Place the monitor on its left side. Insert a medium-sized, flat-head screwdriver into the hole that
accesses the gear wheel for the calibration protection switches (see Figure 7-4).
7-12
Service Procedures
11. Turn the gear wheel approximately 1/4 turn counterclockwise until Protection is highlighted.
Choose Protection ON and push the ComWheel. Turn the gear wheel approximately 1/4 turn
clockwise until Calibration and Protection become faint.
12. Perform the NIBP calibration check (earlier in this chapter) to verify the new calibration.
NOTE: 25 °C and 45 °C test plugs are needed for the temperature calibration.
Tools
e Temperature calibration test plugs, 25 °C and 45 °C (884515)
1. To enter the STP Calibrations service menu from the Service View menu, select
Modules - ESTP: STP - Calibrations
2. Place the monitor on its left side. Insert a medium-sized, flat-head screwdriver into the hole that
accesses the gear wheel for the calibration protection switches (see Figure 7-4).
CAUTION: Turning the gear wheel more than 1/4 turn can damage the calibration switch.
3. Turn the gear wheel approximately 1/4 turn counterclockwise until Protection is highlighted on
the screen. Select Protection OFF and push the ComWheel.
4. Turn the gear wheel approximately 1/4 turn clockwise to highlight Calibrate T1 and choose
Calibrate T1 (or highlight Calibrate T2 and choose Calibrate T2).
5. Insert a calibration plug (25 °C) into the appropriate connector (T1 or T2). Push the ComWheel.
Remove the plug and insert a calibration plug (45 °C) into the same connector. Push the ComWheel.
7. Toreset the protection, turn the gear wheel approximately 1/4 turn counterclockwise until
Protection is highlighted. Select Protection ON and push the ComWheel. Turn the gear wheel
approximately 1/4 turn clockwise until Calibration and Protection become faint.
Connect a pressure manometer with a pump to the transducer dome. Pump a pressure of
100 mmHg to 300 mmHg to the transducer. The recommended pressure is 200 mmHg.
7. Turnthe ComWheel to adjust the pressure on the display to match the pressure reading on the
manometer, then push the ComWheel. A tolerance of + 1 mmHg is allowed.
The “Calibrated” message will be displayed.
7-13
Cardiocap/5 Technical Reference Manual
The following checks should help in localizing the fault. Whenever you suspect the sampling system
and always after you work on the sampling system, check and (if necessary) adjust the flow rate.
CAUTION: The special internal sample tube is mechanically fragile. Sharp bends will cause teaks.
CAUTION: Because of the complex and sensitive construction of the oxygen measuring unit, no repairs
should be attempted inside the unit. If a fault is found in the measuring unit, the entire airway gas
measurement unit should be replaced. Send the faulty unit to Datex-Ohmeda for repair.
In cases of no response to O, or strong drift, check the tubing for loose connections, blockages, and
leaks.
CAUTION: To avoid permanent damage to the pressure transducer, never apply overpressure to the
oxygen measuring unit.
CAUTION: The TPX measuring unit can only be calibrated and repaired at the factory. In case of failure,
the entire airway gas unit should be replaced and sent to Datex-Ohmeda for repair.
NOTE: Never apply overpressure or negative pressure of more than 300 cmH,0 to the flow and volume
tubing. Also never apply differential pressure of more than 25 mmHg on one PVX connection at a time.
7-14
Service Procedures
2. Prevent the monitor from performing the normal occlusion functions, such as controlling the
valves, by turning the pump OFF, then ON again from the menu.
3. Open the rear cover and block the reference gas connector.
Connect a flow cassette with high-flow resistance (50/1.1) to the end of the sampling line.
5. Check the Amb-Work value that is displayed. When the value exceeds 170 mmHg, connect the
other port of the flow cassette to the sample-gas-out connector and switch the pump off.
6. Wait until pressure inside the sampling system stabilizes, then check the Amb-Work value. The
value, or pressure inside the sampling system, should not drop more than 6 mmHg in one minute.
7、 fthe pressure drops more, first ensure the connections you have made and then repeat the test.
Wait untilthe Sample Flow value shown in the Gases service menu is near 200 ml/minute.
Connect a flowmeter to the 3-meter sampling line (use a 2-meter line for the N-XCAiO and N-XV
options). Check that the sample flow reading on the flowmeter is within 180-220 ml/minute.
4. Connect the flowmeter to the reference gas connector on the front panel to check the reference
flow of the oxygen measuring unit. Check that the reference flow reading on the flowmeter is within
31 to 45 ml/minute.
7-15
Cardiocap/5 Technical Reference Manual
Spirometry calibration
The PVX measuring unit is calibrated at the factory. Spirometry calibration is not regularly needed in
routine clinical use, but is recommended each year as part of the Planned Maintenance procedures.
Perform the calibration with adult values using the D-lite and with pediatric values using the Pedi-lite.
Use a calibration pump or a spirometry tester (884202). Readings are observed and adjusted from
the Spirometry Service Data screen. Calibration values are saved to the EEPROM on the PVX board.
To improve accuracy, the endotracheal tube and all accessories normally in use should be attached
during the calibration.
8. Connect a sampling line to the sampling line port on the flow sensor (D-lite or Pedi-lite).
9. Attach a 1-liter calibrated syringe (spirometry tester) to the flow sensor.
10. in the Spirometry service menu, select Zero PVX to zero the flow.
11. Select the Sensor Type (adult for D-lite or pediatric for Pedi-lite).
12. Pump air in, simulating a normal breathing rate, until the readings stabilize.
13. Observe the tidal volumes (TVol Exp and TVol Insp values). If both values are 1000 ml +6% (D-lite)
or 300 ml +6% (Pedi-lite), calibration is not needed.
14. If calibration is needed, adjust the Exp Flow Gain and Insp Flow Gain values in proportion to the
difference between the measured values and the spirometry tester reading:
* Select Exp Flow Gain, continue to pump air in, and turn the ComWheel to increase or
decrease the gain until the TVol Exp reading is within the specified range. Press the
ComWheel to store the adjusted value.
* Select insp FLow Gain, continue to pump air in, and turn the ComWheel to increase or
decrease the gain until the TVol Insp reading is within the specified range. Press the
ComWheel to store the adjusted value.
7-16
Service Procedures
Gas calibration is not available during the first five minutes of monitoring and during a gas sampling
warning.
NOTE:
* During gas calibration, % units are always used for CO» regardless of selected measuring units.
e Anesthetic agent is always calibrated with Desflurane.
3. Attach a new sampling line to the water trap. Connect the loose end of the sampling line to the
regulator on the gas container.
7-17
Cardiocap/5 Technical Reference Manual
4. Toenter the Gas Calibration menu, press the ComWheel and select Parameters - Airway Gas.
The Gas Calibration menu item remains gray (inactive) until the ‘Calibrating gas sensor message
is no longer displayed.
5. Select Gas Calibration.
Wait until the ‘Zero ok’ and then ‘Feed gas’ messages appear on the screen after each gas.
If the ‘Zero error’ message is displayed, press the Normal Screen key and repeat the calibration
procedure. If the problem persists, the unit may require service.
7. Open the regulator and feed the calibration gas until ‘Adjust’ appears, then close the valve.
NOTE: When the monitor is in automatic agent ID mode and it detects a calibration gas and no
anesthetic agent (when using the Datex-Ohmeda Quick Cal calibration gas, for example), the ‘Cal
gas found’ message is displayed. If the identification sensor fails, the ‘Agent id inop’ message is
displayed.
8. Check that the displayed gas values match the values on the calibration gas container.
Calibration adjustments
Calibration adjustments may be required if the gas values displayed during calibration do not match
the values on the calibration gas container.
co2 [500 | 3. Turn the ComWheel until the displayed value matches
02 52.1A| 10.00 the desired value. Press the ComWheel to confirm the
N20 314A adjustment.
Last calibration:
o 0
Е to match
calibration gas.
Autozeroing intervals
Autozeroing intervals after startup are as follows: every 5 minutes for the first 15 minutes, then every
15 minutes for the next 45 minutes, and every hour thereafter. Autozeroing may also occur after agent
selection.
7-18
Contents
8.11 M-NSAT menu and data (N-XNSAT or N-XOSAT options) ......... 8-31
Table of Figures
Figure 8-1. Service menu structure -ἷἷιιιιι---.. 8-3
Service Menus
8. SERVICE MENUS
8.1 Introduction
This chapter contains illustrations and descriptions of the service menus that are accessed from the
Service View menu. You use the service menus to check monitor functions and troubleshoot during
possible failures.
The Cardiocap/5 service menu structure is similarto the service menu structure used in other Datex-
Ohmeda monitors. As a result, some items in the menus are not used in the Cardiocap/5. The service
data values displayed in the menu illustrations may differ depending on the monitor configuration.
NOTE: There is no service menu for the recorder (N-XREC option).
Monitor Setup
Main Menu
(O ©
Give password. Give password.
8-1
Cardiocap/5 Technical Reference Manual
Modules
Carias = outlaader Software
i EEE
Rag Vers CPU serial nbr:
0DU00015 :
Record Vers CPU test date: code: level: -------
, 1999-12-07 0895196 O01
Frevious Heru PLD level:
Configuration:
Mode] #14
Keyboard 1
?
Keyboard 2
?
E È
-More-
Scroll Vers Select this option, then turn the ComWheel to scroll the field on the right side.
-More- indicates there are more lines to be viewed.
Record Vers Prints the software versions and other information on the recorder paper.
8-2
Service Menus
| Service View |
НЕ мя トー wwws|
一 Configuration |
| Country Settings
es |.čs トト cases |
—| spiromety|
し | gpee トー 66500|
— ESPSP |__ Calbrations |
TP 나 |
ae |
Jl Sea
He | |
| Butons/teds
A Preumatios |
一 Watehdog |
e トー Nursewp |
e
— More Modules | - © Ballery * |
Memory Module. |
}
レー Service Log |
8-3
Cardiocap/5 Technical Reference Manual
| -Mare- |
Factory Reset —_ Restores factory default settings and clears data memories. Select this option if
monitor software ts replaced or if the SRAM/Timekeeper IC or its battery is replaced.
PEL 55
UNO
tetera ποπ
Le
Dr 92
+15vĎ „24
UN
SE „05
-15Y ㆍ 05
+2, SYREF 32
00 .
TEMP
BATYOLT ES
a
VO/TEHP OK
CHG ON
ala ale
μα ge. 15va gy-
Voltages Voltages measured by the CPU board. These values are used during the planned
maintenance and service checks to check each supply voltage. The numbers shown
here are directive only. lo pet accurate voltage values these must be measured
according to the instructions for the Planned Maintenance Check ta chapter 3.
Temp ( C) The internal temperature measured hy the temperature sensor an the DC/DC board.
8-4
Service Menus
2. Select Communication.
Previous; Nenu
Connectians
Network (with N-XNET or N-XDNET option) Shows information about communication with the
Datex-Ohmeda Network.
* Statistics In shows the number of received data packets and data bytes
» Statistics Out shows the number of transmitted data packets and data bytes.
Computer If. Shows information about communication through the 1/0 board RS-232 serial
connector (X2).
Interface status shows whether the interface through the connector is opened,
active, or closed.
* Opened indicates that the hardware and software for the interface is running
but there is no connection, or that there have been errors in using the interface.
* Active indicates that the interface Is operating normally.
* Closed indicates that the necessary hardware is not present.
8-5
Cardiocap/5 Technical Reference Manual
Querrun
|
Access Mode When set to Normal (the default), the PC can only read service menu data from the
monitor.
When set to Extended (future option), the PC can also activate the monitor's service
menu functions.
NOTE: Patient monitoring is not possible when Extended is selected. The Access
Mode returns to Normal if the monitor is restarted or if the service menu is closed.
8-6
Service Menus
Network: Net
Previous Menu
= Neither (NONE)
8-7
Cardiocap/5 Technical Reference Manual
asd
WD tros UEL
Rio
Previous Menu
Watchdog This test ensures that the watchdog functions properly. This test shows running
seconds with an accuracy of 0.1 seconds.
When the safety element is working properly, the monitor restarts 1.5 seconds after
the start of the test.
In case of a malfunction, >20 s is displayed and the test will interrupt. In this case,
the fault is in the safety element on the CPU board.
WD by Overload = This test ensures the functionality of a feature in which the software controls the
monitor and detects an overload situation in the main CPU.
When the feature is working properly, the monitor will restart 15 seconds after the
start of the test.
WD by UPI Not used in Cardiocap/5.
8-8
Service Menus
|
Language Sets the language used in this monitor.
National Reqs Sets specific software features to fulfill national requirements in Germany or
France.
Power Frequency Sets the power frequency (50 Hz or 60 Hz). This setting is used to filter out possible
power frequency interference from the parameter measurements.
Time format Sets the time format of the real-time clock (24 1/12 Π).
8-9
Cardiocap/5 Technical Reference Manual
Control wheel
Er |
O Counterclockwise
e Control wheel. Press counts when the ComWheel is pushed. Clockwise and Counterclockwise
count when the ComWheel is turned.
NOTE: Counters in this page are automatically zeroed when you return to the previous menu.
8-10
Service Menus
ESTR : STE
PPT
PP B
MM
NIBP 8
HHT
H-NSAT
More modules
Previous Menu
-Nore-
The modules menu provides access to the measurement parameter menus used for the Cardiocap/5:
Gas Unit
ESTP : ECG
ESTP: SIP
NIBP
NMT
More Modules
Each of these menus and screens are illustrated and described in the sections that fallow.
NOTE: The P/PT, PP, and COP selections in the Modules menu are not used in Cardiocap/5.
Cardiocap/5 Technical Reference Manual
not available
available
Timgouts
Bad checksum 또
Bad c-s by m
Timeouts. Bad checksums. and Bad c-s by mod are cumulative numbers that indicate the following:
* Bad checksums: how many times communication from the measurement unit to the CPU
broke down.
8-12
Service Menus
Counting starts at power on and is reset to zero at power off. Nonzero values do not indicate a
fallure, but continuous counting (more than 50 per second) indicates a serial communication
failure. Also, other measurement units can cause these numbers to rise.
epi as bc
GCC
Fall time Maas
L MÁ
Pump ctrl
Zero valve ctr! sample Flow 201.4 Z 1.2 ménin
Gain
Decl valve ctr]
Ampient FS Ht Soe a A ll
Record Data --- ba -— M3
Previous Menu
Pump ON я 9625
Lanp ON ® 1050
Fan Ch
Zero valve MEAS 059 valve
Temp ТРА 22.6 CPU 30.9 OM 27,
Time after power on CR TES
Sample gain adj Adjusts the sampling pump gain (that is, adjusts the sample flow
measurement.
Fall time Meas Activates the fall time measurement.
Record Data Prints the shown service data and board information (id. serial number and
software id.) onto the optional recorder, N-XREC.
ID unrel. The shown value tells how unreliable the identification is. With pure agent the
value ts normally 50.
8-13
Cardiocap/5 Technical Reference Manual
Sample Flow Calculated from differential pressure and adjusted by the measurement unit.
Zero 一Value aS measured during initialization when the pump is off.
Gain—sample flow measurement can be calibrated by adjusting the gain.
Ambient Ambient pressure is measured every 30 min.
Amb-Work Ambient pressure—sampling system internal pressure.
Fall time CO, and O, in ms. For N,0 and AA, same as CO...
C0,-0, Delay in ms. No delay between CO,, N,0, and agents.
Pump Can be toggled ON/OFF.
PWM output 0-100% is shown. Pump voltage ts also shown.
Lamp The state, PWM control, and current of the lamp are shown.
Fan The state of the fan is shown.
Zero and Occl valve Can be toggled between measurement state (MEAS) and zeroing/occiusion
states (ZERO/OCCL).
Temp Temperatures measured by the measurement unit fram TPX, CPU, and OM.
Time after power on Elapsed time in minutes after power on.
8-14
Service Menus
N2 injection Selects On (1) or Off (0); setting is zero (0) during normal operation.
Record Data Prints the shown service data and board information (id. serial number and
software id.) to the recorder (N-XREC option).
Spirometry calibration
Spirometry calibration is not needed in routine clinical use, but is recommended as part of the yearly
Planned Maintenance procedures. Use a calibration pump or spirometry tester to check the
calibration. Check the TVol Exp and TVol Insp readings on the Spirometry Service Data screen and, if
calibration is needed, adjust the flow gains accordingly. Gain scaling is 1.000 when factory settings
are in effect. Calibration values are sent to the measurement unit during adjustment.
When you select Exp Flow Gain, Insp Flow Gain, VCO2 Gain, VO2 Gain, Y deadspace, or N2
injection, a box for adjusting the value appears. Tuming the Comwheel adjusts the value. Pressing the
ComWheel permanenily stores the adjusted value in the measurement unit's EEPROM.
Aw Press Gain is the value used in the software. It can be adjusted, but not permanently stared.
Exp Gain and Insp Gain apply to the selected sensor. Exp and Insp Flow Gains are scaled as
(user gain)/(factory gain). They can be adjusted in the user service menu.
Common Offset is the compensation factor for the pressure-difference reading of the difference
sensor when applying an equal pressure on both sides of the sensor.
Ref. Condition shows the reference conditions in which results are located. With calibration
pump or spirometry tester, the results are always in ATP. If breathing is detected (EtCO, > 1.0%),
results are according to Flow & Vol Setup selection.
TVol Exp, TVol Insp, MVol Exp, and MVol Insp--the values shown on this screen are the same as
the values shown on the main display.
Y-deadspace is the geometric volume in ml between the Y-piece and the D lite. Default is 5 ml for
a standard Y-piece (as delivered with Siemens 9000 ventilators). Used for VCO, and VO, delay
time corrections.
N2 injection--set it to 1 during laboratory tests with a spirometry tester and infection of N, gas (for
scientific validations use only). At power on and during normal use, the value is always zero.
8-15
Cardiocap/5 Technical Reference Manual
Rp
Available OFF
Measurement ON
Ánp ¿ero CFF
Value ーーー
Timeouts 0 RAM 매
Bad checksums - O- ROM OK
Bad c-s by mo O. EEFFOM ОК
Quick zero is ON when the signal in any of the three internal amplifiers goes beyond scale, and
therefore, a capacitor connected to the related channel discharges overvoltage. At least one Quick zero
value is OFF when using a 3-lead cable; all three are OFF when using a 5-lead cable. Quick zero also
takes place when the lead is changed in 3-lead measurement.
Resp Available indicates that FCG hardware ts capable of measunng impedance respiration.
Amp Zero shows ON when zeroing of the respiration amplifier takes place.
fordefinitions of Timeouts, Bad checksums, Bad c-s by mod, RAM. ROM, and EEPROM. see
NIBP service menu, data, and checks later in this chapter.
8-16
Service Menus
TT. 3. select
ECG Setup.
OFF OFF OFF OFF
] RAM UK
"s Oh
1 EEPROM OK
H ter Filters the ECG signal high-frequency noise and slow-respiratory artifacts.
Monit (monitor) filter is used in routine monitoring. It effectively filters the artifacts
caused by the electrosurgery unit and respiration.
Diagn (diagnostic) filter is used if more accurate waveform information is needed (of
P-wave or AV biock, for example). The diagnostic filter is more susceptible to high
frequencies and baseline wander than the monitor filter.
STfilt (ST filter) permits more accurate information of ST segment. It filters the high-
frequency artifacts caused by electrosurgery unit but catches the slow changes in ST
segment. The ST filter is more susceptible to baseline wander than the monitor filter.
Pacemaker Selects how to display the pacing pulse of a cardiac pacemaker. The selections are
Show, Hide, ON R and Sensit.
ON R does not filter pacing pulses away from LCG data. This improves ECG monitoring
with A-V pacemaker patients, as QRS complexes are counted even if the pacing pulse
hits the QRS complex. During asystole pacing pulses may be counted as heartbeats.
Sensit uses more sensitive pacemaker detection. Pacemaker spike is displayed on
ECG.
QRS Type QRS detection and HR calculation may be affected by the shape of the QRS
complex,
— Ired int.
Fed int.
Hr ーーー EC gain
Cale OM IDC
Probe OH ROC
Probe ctt ME 9311)
Pre gain
Temp error UN
Temp test UFF
Protect key BEF
Protect mode (N
Configuration STP
Timeouts ü RAM GE
Bad checksums O ROM DE
Bad c-s by mac DEEPROM OK
Record Data Records, or prints, the service data shown and board information (ID, serial number
and software 1D) on the recorder paper.
temp Test Activates the automatic temperature test for temperature channels T1 and T2. The result
appears in the service data screen. NOTE: Select Temp Test twice to start the test.
8-18
Service Menus
red int. (infrared intensity) range is 40 to 255 and Red int. (red intensity) range ts 40 to 255.
NOTE: Balance between LEDs Is adjusted by changing the infrared/red intensity.
Protect key is normally OFF. It turns to ON when the button at the bottom of the module is pressed.
Protect mode is normally ON. It turns to OFF when Protection is switched to OFF for temperature
calibration in the STP Calibrations service menu.
Configuration shows the chosen module configuration: IP, ST, or STP.
For definitions of Timeouts, Bad checksums, Bad c-s by mod, RAM, ROM, and EEPROM, refer to
NIBP service menu, data, and checks later in this chapter.
Protection Protects the configuration and temperature calibrations. Can be set ON (protected) and
OFF only when protect button at the bottom of the module ts pressed.
Set Config Sets the configuration to TP, ST, or STP. The setting is possible only when Protection is
OFF. This setting should be checked if the STP board is replaced.
Calibrate T1 / Calibrate T2
Calibrates temperature channels 11 or 12. Calibration is possible only when Protection is
OFF. Accurate test plugs of 25 °C and 45 °C values are required. See Temperature
calibration in the Service Procedures chapter.
Calibrate P1 / Calibrate P2
Calibrates the invasive blood pressure channels PT and P2. A pressure transducer (with
an appropriate cable) and a pressure manometer ate required. See Temperature
calibration in the Service Procedures chapter.
8-19
Cardiocap/5 Technical Reference Manual
de
B
le
Previous Menu
Timsouts Q BAN
Bad checksums - à ROM
Bad c-s by mod O EEPROM
The submenus shown tn this menu (NIBP Demo, Calibrations, etc.) are described later in this section.
Previous Menu Closes this menu and displays the previous menu.
Zero Pressure at auto zeroing multiplied by 10 and changes between +20 and -20
mmHg. Absolute pressure is the sum of Pressure and Zero.
Protect handle Hardware protection for EEPROM memory (always ON during normal operation). If it
is OFF, data cannot be read from or written to EEPROM; only the calibration
protection can be set or reset by software. You can change the protection setting in
the Calibration menu by turning the gear wheel (located inside a small hole in the
bottom of the monitor) to enable Protection ON/OFF.
+15 V power The condition of the +15 Vdirty supply voltage for the pump and valves. It exists
(ON) or not (OFF), depending on service menu function. The supply voltage can be
turned ON by selecting the previous Menu and then the desired menu again.
ADO ... AD7 The values of each of the eight A/D converter channels.
Timeouts, Bad checksums, and Bad c-s by mod are cumulative numbers that indicate the following:
Timeouts--how many times the module did not respond to an inquiry from the
monitor.
Bad checksums--how many umes communication from the module to the monitor
failed.
8-20
Service Menus
Counting starts at power up and resets to zero at power off. Nonzero values do not
indicate a failure, but continuous counting (more than 50 per second) indicates a
serial communication failure or that the module is not installed. Other modules can
cause communication errors that cause these numbers to rise.
Wave Recording Selects the recording option. If ON is selected, the pressure signals are recorded in
real time onto the recorder paper.
Remove menu Widens the displayed waveform area.
Previous Menu Closes this menu and displays the previous menu. İf Remove menu was selected,
you can also just press the ComWheel.
8-21
Cardiocap/5 Technical Reference Manual
EL se]
e a エー イニ モー イー
III
Calibration Check SFF
LE
Protection CEF
O
O
ONE
Calibration
pe ağaCn
Frevicus Meru
ED
Calibration Calibrates NIBP. This selection ts available only when Protection is OFF.
For details, see NIBP calibration in the Service Procedures chapter.
2. Select Active Leak Test ON. Pressure of 260 mmHg is pumped into the cuff automatically.
3. Wait for several seconds until the pressure stabilizes. If the pressure reading drops more than
5 mmHg per minute, detect and fix leaking point(s).
8-22
Service Menus
Bi
Q
則
AS U RAM OK
(SET checksuns O RON UA
Bad c-s by mod O EEPROM DE
Start Test Starts the Safety Valve Test: changes to Stop Test after you start the test.
Max. press and 2 s after stop show the measured values at Safety Valve test.
Max. press The pressure at which the safety valve opens. Itis normally 310 + 15 mmHg for
adult and 150 mmHg+ 15 mmHg
for infant.
2 s after stop The pressure at 2 seconds after the pump has stopped. It is normally >280 mmHg
for adult and >120 mmHg for mřant. If the value is less, check leakage by the active
leak test.
8-23
Cardiocap/5 Technical Reference Manual
Protect handle OM
Cali t OFF
Timeouts Q BAH 매
Eac checksums ΠΜ 내
Bad č-s by mod O EEPRON CK
Start Test Starts the Pulse Valve test; changes to Stop Test after you start the test.
Set Valve Lets you adjust the opening of the pulse valve.
The pressure rises beyond 240 mmHg and stops. The pulse valve opens. The time it takes for the
pressure to go from 240 mmHg to 50 mmHg is counted and displayed on the screen.
3. Tostop the test, select Stop Test.
The valve can be adjusted between O (fully closed) and 255 (fully open). To adjust, choose Set Valve
and push the ComWheel. Adjust the pulse-valve value by turning the ComWheel, then push it to set
the value.
The Interval 240 mmHg -> 50 mmHg time should be less than 60 seconds when the valve is 150,
and less than 10 when fully opened (255). When fully closed (0), the system should be airtight and the
pressure does not drop. Depending on an individual, the pulse valve may remain closed up to 100,
approximately.
If the measured time deviates much from those above, then the pulse valve or its tubes are faulty.
8-24
Service Menus
OM
OFF
ng STAT.
nt a à
OFF
Timeouts 2 RAM pi
Bari checksums ROM ον
Bal C-s by mod EEPROM ОК
The Auto ON/OFF, Manual ON/OFF, STAT ON/OFF, and Measur. ON/OFF menu selections have no
effect.
8-25
Cardiocap/5 Technical Reference Manual
CIN
OFF
ON
0 РАН 1
O ROM i
0 EEPROM OK
Start Pump Starts the pump; changes to Stop Pump when the pump is on.
Open Exh1 Opens exhaust valve 1; changes to Close Exh1 when the valve is open.
Open Exh2 Opens exhaust valve 2; changes to Close Exh2 when the valve is open.
Set Valve Adjusts the pulse-valve opening between 0 (fully closed) and 255 (fully open). To
adjust, push the ComWheel, turn it to adjust the value shown on the screen, and
push it to set the value.
8-26
Service Menus
Frevious Menu
0H
(FF
où
Watchdog Interval Hs
11060 나 을 0. RAM Ok -
Bad Chécksums C RDM Us
EL SS
Test ADULT Tests the watchdog timer in adult mode (120 to 140 seconds).
Test INFANT Tests the watchdog timer in infant mode (about 60 to 70 seconds).
Stop Test Stops the test.
8-27
Cardiocap/5 Technical Reference Manual
RAM OK
Oo ROW. Ok
DETTE TT 0. . EEPROM OK
Start Curr. test A test that checks the difference between a given current and the measured
current.
Connect a 3 kQ resistance between the stimulus electrodes before starting the
test. All currents that are checked will be displayed on the service data screen.
After the current value, the test status is shown.
if the test is not passed, factory calibration is needed. Contact Datex-Ohmeda.
T1%/ref NMT MechanoSensor setting. Always use a setting of 3 for normal monitoring.
Settings 1 and 2 are for research purposes only.
Record Data Prints the displayed service data and circuit board information (id., serial number,
and software id.) to the optional Recorder.
8-28
Service Menus
T1%, etc. The measured response. A value of 1000 corresponds to 100%. Each is a ratio of
the Absolute T1, etc. to the Reference (Ref). As drugs are administered, the
typical response for T1 lessens, then drops more rapidly for each successive
measured response (12%, 13%, then T4%).
Ratio% The common label for TOF/DBS%.
Absolute T1, etc. The voltage measured from the A/D converter. The area-under-the-curve of the
response to stimulation as the muscle-relaxant anesthetic is being administered.
Ref (Reference) The area-under-the-curve calculation from the response of the NMT
stimulation before the administration of inhibiting anesthetics.
Noise The interference just before the measurement. A typical value is < 10. Represents
the measurement of the area-under-the-curve of the response when no stimulus
iS given, but the circuitry noise is picking up some response. The noise factor
varies from sensor to sensor.
Offset An average of the noise measurement; the voltage baseline. An offset of 2520 is
2520 mV, or 2.5V. This is the measureable baseline voltage of the sensor and
varies from sensor to sensor.
Curr set The selected current, a value of 700 corresponds to 70 mA.
Curr meas The measured current, a value of 700 corresponds ta 70 mA.
Pulses Pulses the module has produced.
Bad checksums--how many times communication from the NMF unit to the
monitor failed.
Bad c-s by mod -how many communication errors the NMT unit detected.
Counting starts at power on and Is reset to zero at power off. Nonzero values do
not indicate a failure, but continuous counting (more than 50 per second)
indicates a Serial communication failure or that the NMT unit is not installed. Also,
communicaton errors from other modules can cause these numbers to rise.
8-29
Cardiocap/5 Technical Reference Manual
Stimulus Mode Selects the mode of stimulation: TOF (train of four), OBS (double burst
stimulation), or ST (single twitch).
Cycle Time Sets the ume intervals for the selected Stimulus Mode. For TOF and DBS, the
selections are Manual, 10 , 12, 15, or 20 seconds. and 1, 5, or 15 minutes. The
ST cycle times are 1, 10, and 20 seconds.
When the cycle time is less than one minute, the Count must be below the Count
limit in two Successive measurements before the Recovery Note Is enabled. To
give the note, the number of Counts must exceed or match the limit twice in one
minute. The note disappears when two measurements are below the limit again.
lf the cycle time is One minute or more, or ifthe measurement is done manually,
at least one Count must be below the chosen limit before the Recovery Note is
enabled. To give the note, one Count must exceed or match the Count limit. The
note disappears when one measurement is below the limit again.
Current Sets the stimulus current, which ts either supramaximal (automatic start-up
search) or manually selected from 10 mA through 70 mA at 5 mA intervals.
Pulse Width Increases or decreases the effect of the stimulation. Selections are 100 jis.
200 ys, and 300 ps.
8-30
Service Menus
NoProbe
4 fs seSearch
NoPu1se
CheckProbe
HP-203 Error
QUART Error
1/0 Error
? Timeouts — 0 Erro
“ ¢Bad checksums 2: ROM OK
| 839.0-8 00 0060 Е
PR Pulse rate value [bpm] calculated from pleth.
Timeouts -how many times the unit did not respond to an inquiry from the monitor.
Bad checksums--how many times communication from the SpO» unit to the
monitor failed.
Bad c-s by mod -number of communication errors detected by the SpO» unit.
Counting starts at power on and is reset to zero at power off. Nonzero values do not
indicate a failure, but continuous counting (> 50 per second) indicates a serial
communication failure or that the SpO unit is not installed. Also, communicaton
errors from other modules can cause these numbers to rise.
ROM Indicates whether the checksum in the EPROM is in accordance with the checksum
calculated by the software. The state is 0K, Fail or ? (unit not installed or
communication error)
8-31
Cardiocap/5 Technical Reference Manual
Memory Module
icus Meru
NOTE: The Interface and Battery selections in the More Modules menu are not used in Cardiocap/5.
8-32
Service Menus
Car q tura
Fil E Systen
Čard size
Card used
Card full
Card empty
Read errůr
Urita error №
Module present Indicates whether the module is firmly attached to the monitor (YES or NO).
Module active indicates whether module Services are available (YES or NO).
Read error Indicates whether reading from the card failed (YES ar NO).
Write error Indicates whether wating to the card failed (YES or NO).
8-33
Cardiocap/5 Technical Reference Manual
Timeouts
Length err
fuplicated
Communication screen
Interface status
The status of the data link between the monitor and the memory module. If memory module is
properly attached, the status should always be ACTIVE. if status blinks between ACTIVE and
CLOSED, a communications error has occurred: remove module briefly, and insert it back into the
monitor frame to check if the error disappears.
Message types The type of data packets sent (Ix) and received (Rx) since the last monitor start:
Module status—number of sent/received data packets that relate to the memory module status.
Packets total--total number of data packets sent/received since last monitor start.
Bytes total—total number of data bytes sent/recetved since last monitor start.
Timeouts
-- The number of time-outs that occurred in memory module data transmission.
Chksum err—!he number of checksum errors in data packets from memory module.
Length err The number of data packets with erroneous length from the memory module.
Duplicated - [he number of duplicate data packets from the memory module.
8-34
Service Menus
Reset Log Clears the service log content (recommended after a performed maintenance).
Reset RunTime Resets the monitor's run-time counter. The counter is located in the main CPU
board's SRAM and basically counts the amount of time the monitor has been on
since leaving the factory.
Error History
. | To enter this service menu from
Service Log Error History the Service View menu, select
Last errors: Service Log.
Fast cold start EAD De s
Event History
GSP watch-dog ti ον
Alarm History
Fast cold start fad
Scroll Last Er
GSP watch-dag oa t
SCro11 Countrs El
i Fast cold start (off 1gs < 20
1999-0ct-20 1
Record Log ime
„ast errors Shows the last monitor errors and the time of their occurrence.
Errorcounters Contains counters for each different (detected) monitor error. The time of
occurrence of the last error is shown beside each counter.
Fast cold start Gives the number of erroneous cold starts when power was off < 20 seconds. The
cause can be a failing lead-acid battery or a software problem that was solved by
the hardware watchdog circultry.
Last log reset Shows the date and the time of last service log reset.
8-35
Cardiocap/5 Technical Reference Manual
Event History
+ To enter this service menu from the
Service Log Event History Service View menu:
Running hours: 1068 。
η κ Μο ους i. Select Service Log.
Last events: 2. Select Event History.
La
à Mi e ali Vee
a
Last events Shows the fast events and the time of their occurrence.
Event counters Contains counters for each different (detected) event. The time the last event
occurred is shown beside each counter.
Lastlogreset Shows the date and the time of the last service log reset.
Alarm History
Shows the monitor alarm log. The alarm log is reset if the monitor is in standby for more than 15 minutes.
A To enter this service menu from the Service View menu:
Alarm History
| | 1. Select Service Log.
MR RAL a 2. Select Event History.
8-36
Contents
9. SPARE PARTS
Use the Order Number when ordering parts. Item numbers are used only to identify parts shown in the
illustrations.
NOTE: Remanufactured or exchange parts may be available for some measurement units or boards.
Request information when you order the part.
Push buttons
Button (Silence Alarms) Order Number Button (Normal Screen) Order Number
Danish, Silence Alarms 896958 Danish, Normal Screen 896877
Dutch, Silence Alarms 896839 Dutch, Normal Screen 896870
English, Silence Alarms 896836 English, Normal Screen 896867
Finnish, Silence Alarms 896843 Finnish, Normal Screen 896874
9-2
Spare Parts
Screw, pozidrive
の
M3x12 (1)
ET
vE
ox
Sa
요크
=>
A
u
る
- =
の
Se
az
So
5
wo
=
a
Ss
Nm
一
Screw, pozidrive M3x8 ( 5)
we
==
の の
зо
2
=>
D 十
aga
M
=
©
o
vo
oo
=
ac
Z
©
>
x
©
9-3
Cardiocap/5 Technical Reference Manual
9-4
Spare Parts
Screw, pozidrive M3x6 (6)
©
©
— ©
© E
o の
< >
© г
= B
2 ©
Screw, pozidrive M3x8 (2)
š ©
= =
- Y
=
©D
o
[ορ]
A
Screw, pozidrive M4x6 (1)
|
DS
DL
Connector lock screw M3 (4)
SD, |
Screw, pozidrive M3x6 (2)
+
©
CN
>
+
=
qa
- 숨
©
+o $È e
2
323 358
о © © > DO
İT
(15
—
DOm © e x ©
Г,
o = = Ge
|
o ~ 一 一-
Co
ag
o
W
2
© от
=
=
è
—_ © o
o B à の
ーー od y < 8 8
nc — 8 70 x
Doo a 2E
- / =
る 2
の a
9-5
Cardiocap/5 Technical Reference Manual
31 LCD display unit, complete with display shield (28) (1D 894745) 897500
32 Backlight for LCD display (includes 2 bulbs) 572791
33 Service Kit, Display adapter board and mounting foam 6050-0006-020
34 Inverter board 6050-0005-752
35 Backlight board (ID 896222) 897502
36 Lithium battery for SRAM on CPU board (ID 197230) 897506
37 CPU board (ID 895196) 897499
NOTE: The appropriate software service kit(s) are required when
reptacing the CPU board (and may be required when replacing the
battery, depending on the monitor software level):
1. Service Kit, Cardiocap/5 Monitor Software: XANE: 6050-0005-821
Dutch or
English
French XCCA: 6050-0005-819
German
2. Service Kit, Cardiocap/5 Monitor Software: XANE: 6050-0005-992
English or
Italian
Portuguese XCCA: 6050-0005-993
Spanish
3. Service Kit, Cardiocap/5 Monitor Software: XANE: 6050-0005-994
Danish
| or
English
Finnish XCCA: 6050-0005-995
Norwegian
Swedish
4. Service Kit, Cardiocap/5 Monitor Software: XANE: 6050-0006-860
English or
Polish
XCCA: 6050-0006-858
5. Service Kit, Cardiocap/5 N-XNET Software 6050-0005-820
(order if N-XNET is installed)
9-6
Spare Parts
MES 7
CR
X
+
TICA
7
/
N
N
9-7
Cardiocap/5 Technica! Reference Manual
기
Screw, pozidrive plastite
M2.5x10 (8) % É serew, pozidrive plastite
M2.5x10 (12)
9-8
Spare Parts
|| Backlight board
[lrInverter board |
N-XNSAT | NSAT io VE | | board (60 - kota rd
9-9
Cardiocap/5 Technical Reference Manual
9-10
Spare Parts
一
Intet ring (2)-T
9-12
Spare Parts
101
9-13
Cardiocap/5 Technical Reference Manual
© © (© (©
9-14
Spare Parts
Screw, pozidrive
- M3x6 (4)
Header
{for J3 on OSAT board)
9-15
Cardiocap/5 Technical Reference Manual
9-16