Agilent 1100/1200 Series
Liquid Chromatograph
Hardware and Software Installation Checklist
Hardware / Software Installation
Purpose: To ensure that instruments and systems are correctly installed and functioning as designed, in the
customer's facility. Correct installation is the first step in ensuring that instruments and systems operate reliably
over an extended lifetime.
Customer Responsibilities
The customer should ensure that the installation site is prepared in accordance with the specifications contained in
the relevant site preparation document and that the necessary operating supplies, consumables and usage dependent
items such as vials, paper, solvents, etc. are available. A customer representative should be available during the
installation.
Note: Typical installation time is 1.5 hours for a system and 30 minutes for a module.
If installation time was not sold for any section, check that section as not applicable.
Service Engineers Responsibilities
Please only complete/printout pages that relate to the system module(s) that have been installed along with this
page, LC System information and Service Review page and fill the total number of pages that will be included at
the bottom of this page
Additional 1200 SL Modules
If additional SL modules have been included in this installation please insert these module pages into this
document (Please note that the revision number will be different to that of the main document).
Total Number of pages for this Document:____________
Issued: Sept 2008 – Rev. 6.4 Copyright 2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph
Hardware and Software Installation Checklist
LC System Information
System Name: ____________________________
System Location: _________________________
Contact Name: ___________________________
Installation Checklist
Section NOT Applicable
Note:
If one section contains multiple choices for installed Hardware please encircle all modules that are part of the
system and fill in the corresponding Serial Number.
Unpacking and inspection of shipped materials
Note: Shipping containers should not be opened until an Agilent Technologies representative is present
Unpack all boxes and place equipment on bench. To prevent injury, get lifting assistance if needed.
Retain shipping containers and material until installation is complete and performance is verified
Notify order fulfillment of any discrepancies
Unpack and verify condition and completeness of shipment
Check serial numbers and fill in installation documentation S/N fields if necessary
Verify that power cords are correct (must mate with power sockets) and are plugged into power receptacle
When applicable, install computer system and connect all power interconnect cables
Check for the Agilent LC Diagnostic CD-ROM and install this software component
Verify that the instrument is the correct module for the regional voltage (100 V/115 V/230V). Do NOT use
instruments which have an incompatible voltage for ELSD.
Remote Advisor
Ask the Customer is remote advisor is installed
If yes in install the Data Source Software
Issued: Sept 2008 – Rev. 6.4 Copyright 2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph
Hardware and Software Installation Checklist
G1322A/79A/B– Degasser
Section NOT Applicable
1. Record model number
1. Record serial number
Set up the degasser(s)
solvent inlet/outlet
Priming each channel with Isopropanol from solvent bottle to pump inlet
G1310A/11A/12A – Pump
1. Record model number 2. Record model number 3. Record model number
1. Record serial number 2. Record serial number 3. Record serial number
1. Record Firmware 2. Record Firmware 3. Record Firmware
Set up the pump(s)
solvent inlet/outlet
waste drain
remote control (if used)
relay contacts (if installed)
seal wash tubing (if installed)
CAN
Agilent LAN (see note*)
Priming the pump
prepare solvents
turn on
purge
purge mobile phase
monitor the pressure
Issued: Sept 2008 – Rev. 6.4 Copyright 2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph
Hardware and Software Installation Checklist
G1322A/79A/B– Degasser
1. Record model number
1. Record serial number
Set up the degasser(s)
solvent inlet/outlet
Priming each channel with Isopropanol from solvent bottle to pump inlet
Issued: Sept 2008 – Rev. 6.4 Copyright 2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph
Hardware and Software Installation Checklist
G1322A/79A/B– Degasser
Section NOT Applicable
1. Record model number
1. Record serial number
Set up the degasser(s)
solvent inlet/outlet
Priming each channel with Isopropanol from solvent bottle to pump inlet
G1382A/G1376A Capillary Pump with G1379B Micro Degasser
&
G2225A/G2226A Nano pump with G1379B Micro Degasser
1. Record model number 2. Record model number
1. Record serial number 2. Record serial number
1. Record Firmware 2. Record Firmware
Set up the pump
solvent inlet/outlet
waste drain
remote control (if used)
relay contacts (if installed)
CAN
Agilent LAN (see note*)
set up the degasser
solvent inlet/outlet
priming each channel with Isopropanol from solvent bottle to pump inlet
Priming the pump
prepare solvents
turn on
purge
purge mobile phase
monitor the pressure
Issued: Sept 2008 – Rev. 6.4 Copyright 2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph
Hardware and Software Installation Checklist
G1361A – Prep. Pump
1. Record model number 2. Record model number
1. Record serial number 2. Record serial number
1. Record Firmware 2. Record Firmware
Set up the pump(s)
solvent inlet/outlet
waste drain
remote control (if used)
relay contacts (if installed)
seal wash tubing
CAN
Agilent LAN (see note*)
Priming a prep pump
Prime each pump individually
Use the syringe + adapter + tubing and connect to the waste outlet of the pump`s EMPV
Turn ON the pump and electronically switch the EMPV to waste
Manually suck the solvent into and trough the pump
Pump at least 0.5 liter of IPA through the pump
Issued: Sept 2008 – Rev. 6.4 Copyright 2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph
Hardware and Software Installation Checklist
G1328B – Manual Injector or 5065-9922 Prep. Manual Injector
Section NOT Applicable
1. Record model number
1. Record serial number
Setup the injector
start/stop cable to pump
base plate & mounting pole
position valve
inlet/outlet connected
sample loop
vent tubes
G1329A/B – Autosampler G1367A/B/C/D – High Performance Autosampler G1377A - Micro WPS
G2258A – Dual Loop ALS /G2260A - Preparative ALS
Section NOT Applicable
1. Record model number 2. Record model number 3. Record model number
1. Record serial number 2. Record serial number 3. Record serial number
1. Record Firmware 2. Record Firmware 3. Record Firmware
Setup the autosampler(s)
solvent inlet/outlet
waste drain (must be in a straight line down, no bends, kinks or curves
large volume tube
(if installed)
BCD output (if installed)
remote control (if used)
relay contacts (if installed)
CAN
Agilent LAN (see note*)
Issued: Sept 2008 – Rev. 6.4 Copyright 2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph
Hardware and Software Installation Checklist
G1330B - ALS Thermostat
The following section applies if a G1330B - ALS Thermostat is installed together with any of the well-plate-,
autosamplers or fraction collectors to generate a: G1329A/B + G1330B - Thermostat. ALS G1364B/C/D +
G1330B Therm. Fraction Collector G1367B + G1330B - Thermostat. High Performance Autosampler G1377A
+ G1330B - Thermostat. Micro WPS G2260A + G1330B - Thermostat. Preparative ALS
1. Record model number
1. Record serial number
Setup the ALS Thermostat(s) underneath the module(s) that has (have) to be thermostatted.
WARNING: Do not connect the ALS thermostat cable while the line power is connected to the any
of the modules. This would damage the sampler or fraction collector and the thermostat electronics.
air channel adapter and plate
ALS thermostat cable
ALS thermostat waste drain
Issued: Sept 2008 – Rev. 6.4 Copyright 2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph
Hardware and Software Installation Checklist
G1316A/B/C - Column Compartment
1. Record model number
1. Record serial number
1. Record Firmware
Setup the TCC
solvent inlet/outlet
column
waste drain
remote control (if used)
CAN
Issued: Sept 2008 – Rev. 6.4 Copyright 2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph
Hardware and Software Installation Checklist
G1157A/58A/59A/60A/62A/63A Valves
Section NOT Applicable
1. Record model number 2. Record model number 3. Record model number
1. Record serial number 2. Record serial number 3. Record serial number
1. Record Firmware 2. Record Firmware 3. Record Firmware
Setup the Valves
Setup the 1200 Series Valve
solvent inlet/outlet
CAN
CAN-DC-In from other 1200 Module (e.g. G1367B/C/D Autosampler, G1364B/C/D Fraction Collector or
G1361A Prep. Pump)
G1383A Column Organizer
Section NOT Applicable
1. Record model number
1. Record serial number
Setup the column organizer
base plate and organizer plate
rods
traverses
additional holders (if used)
Issued: Sept 2008 – Rev. 6.4 Copyright 2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph
Hardware and Software Installation Checklist
G1314A/B/C/D/E – VWD
1. Record model number
1. Record serial number
1.Record Firmware
Set up the detector
flow cell
solvent inlet/outlet
waste drain
remote control (if used)
analog output (if used)
relay contacts (if installed)
CAN
Agilent LAN (see note*)
Verify the operation of the detector
Perform the Lamp Intensity Test
Perform the Wavelength Calibration Test
Lamp
Intensity Pass/Fail
Zero-order
WL Deviation
calibration 656nm
Deviation
*
Please note that ‘D/E’ detector requires ChemStation B.04.01 and LMD
Issued: Sept 2008 – Rev. 6.4 Copyright 2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph
Hardware and Software Installation Checklist
G1315C/D-DAD / G1365C/D-MWD
Section NOT Applicable
1. Record model number 2. Record model number
1. Record serial number 2. Record serial number
1. Record Firmware 2. Record Firmware
Set up the detector(s)
deuterium lamp (if not installed)
solvent inlet/outlet
waste drain
remote control (if used)
analog output (if used)
relay contacts (if installed)
CAN
Agilent LAN (see note*)
Verify the operation of the detector
Perform the Lamp Intensity Test
Perform the Wavelength Calibration Test
Lamp 1.
Intensity Pass/Fail
1. Alpha
WL Deviation
calibration Beta
Deviation
Lamp 2.
Intensity Pass/Fail
2. Alpha
WL Deviation
calibration Beta
Deviation
Issued: Sept 2008 – Rev. 6.4 Copyright 2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph
Hardware and Software Installation Checklist
G1321A – FLD
1. Record model number
1. Record serial number
1.Record Firmware
Set up the detector
solvent inlet/outlet
waste drain
remote control (if used)
analog output (if used)
CAN
Agilent LAN (see note*)
Verify the operation of the detector
Perform the Lamp Intensity Test & compare profile
Perform the Wavelength
Verification Test using the Raman band.
Lamp
Intensity Pass/Fail
Excitation
WL Deviation
verification Emission
Deviation
If test fails perform Wavelength Calibration.
Issued: Sept 2008 – Rev. 6.4 Copyright 2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph
Hardware and Software Installation Checklist
G1362A – RID
1. Record model number
1. Record serial number
1.Record Firmware
Set up the detector
WARNING: Make sure that all plugs are removed from the ports and the appropriate capillaries
(inlet/outlet/recycling) are connected (maximum backpressure of 5 bar for flow cell).
solvent inlet/outlet
solvent recycle
waste drain
remote control (if used)
analog output (if used)
CAN
Agilent LAN (see note*)
Verify the operation of the detector
Diode
Balance Value
Issued: Sept 2008 – Rev. 6.4 Copyright 2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph
Hardware and Software Installation Checklist
G4218A – ELSD
1. Record model number
1. Record serial number
1.Record Firmware
Set up the detector
Verify that the instrument is the correct module for the regional voltage (100 V/115 V/230V). Do NOT use
instruments which have an incompatible voltage
connect the exhaust line to a fume hood or lab exhaust line
set the correct pressure at the regulator for the pressurized nitrogen line and connect it to the detector
install the nebulizer
install the nebulization chamber
install the capillary connection from detector to HPLC
install the AGP remote connection from detector to HPLC
install the RS-232 connection from detector to Chemstation control PC (if applicable)
connect the siphon overflow to the waste container
optionally connect the analog signal cable
optionally configure the detector in Chemstation
G4218A Checkout
Switch on the instrument
Prepare the instrument:
o If a new column is used, flush the system for 15 minutes AND
o set the detector temperature to 35 °C and equilibrate for 15 minutes
Set up a method for the checkout sample (see user manual for test conditions)
Run a test measurement
Start a run
Attach results to installation documentation
Issued: Sept 2008 – Rev. 6.4 Copyright 2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph
Hardware and Software Installation Checklist
G1364B/C/D – Fraction Collector
1. Record model number 2. Record model number 3. Record model number
1. Record serial number 2. Record serial number 3. Record serial number
1. Record Firmware 2. Record Firmware 3. Record Firmware
Setup the fraction collector(s)
solvent inlet/outlet
waste drain
remote control (if used)
relay contacts (if installed)
CAN
Agilent LAN (see note*)
Issued: Sept 2008 – Rev. 6.4 Copyright 2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph
Hardware and Software Installation Checklist
G1390A – Universal Interface Box (UIB)
Section NOT Applicable
1. Record model number
1. Record serial number
1.Record Firmware
Setup the universal interface box
CAN
Analog In (if used)
GPIO / Analog In (if used)
Relay Contacts (if used)
Note: Agilent LAN connection is made between one (1) of the Agilent 1200 Series System modules and the
Agilent ChemStation PC. If a detector is installed connect the LAN to this module
G4208A Instant Pilot
Section NOT Applicable
1. Record model number
1. Record serial number
1.Record Firmware
Set up the control module
Connect via CAN to available instrument module
G1323B Control Module
Section NOT Applicable
1. Record model number 1. Record model number 1. Firmware
Set up the Control
Module
Connect via CAN to available instrument module
Issued: Sept 2008 – Rev. 6.4 Copyright 2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph
Hardware and Software Installation Checklist
Software Installation (Bundled)
Section NOT Applicable
Software Product:_________________________
Software Revision:________________________
Start Windows
Start ChemStations -> Add Licenses and add the license numbers. You find these numbers in the 'Software
Certificate and Registration Packet' envelopes.
Configure TCP/IP if the instrument will be connected using LAN.
If necessary install and Configure IO Libraries (LC/MS)
If necessary install and configure Agilent Bootp Service for LAN connection.
Use the Configuration Editor and configure your instruments.
Install and Configure Software Add-Ons (if applicable)
If present, install the PC Image software and create a backup using the Backup Solution User Guide as
reference.
If present, install the Agilent LMD software.
Software Installation (Un-Bundled)
Section NOT Applicable
Software Product:_________________________
Software Revision:_____________________
Start Windows
Install ChemStation using the ChemStation Installation guide for reference.
Check if a printer driver is installed on the system
Verify that the regional settings are set to English-US
Install and Configure TCP/IP if the instrument will be connected using LAN.
If necessary install and Configure IO Libraries (LC/MS).
If necessary install and configure AGILENT Bootp Service if the instrument will be connected using LAN Use
the Configuration Editor
Install the Agilent LMD Diagnostic software
Remote Advisor
Ask the Customer is remote advisor is installed
If yes in install the Data Source Software
Issued: Sept 2008 – Rev. 6.4 Copyright 2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph
Hardware and Software Installation Checklist
1200 LC Checkout
Prepare:
1. Isocratic sample 6. detector(s)
2. pump(s) 7. fraction collector(s) (if present)
3. online degasser (s) (if present) 8. Valve(s) (if present)
4. injection device(s) 9. recording device
5. column compartment
Start a run
Attach results to installation documentation
Fill out instrument logbook
Verify connection to the Agilent LC Diagnostic Tool./LMD
Issued: Sept 2008 – Rev. 6.4 Copyright 2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph
Hardware and Software Installation Checklist
Service Review
Affix any reports generated to this Checklist
Record in the instrument logbook date and time install was completed
Explain Agilent warranty for Hardware, Software and return policy for the Instrument service Center
Explain how to log an instrument service call
Explain support services
Explain how to use manuals
Advise customer of additional training options
Advise the customer of the following useful Agilent web sites:
e.g. 1200 Series Firmware (Downloads/patches)
Exclusive offers: www.agilent.com/chem/exclusiveoffers
Library - http://www.agilent.com/chem/library
Education - www.agilent.com/chem/education
eSeminars - www.agilent.com/chem/eseminars
Software Status Bulletins and Patches - www.agilent.com/chem/techsupport
Supplies - www.agilent.com/chem/supplies
Tech Support - www.agilent.com/chem/techsupp
Note: Password is the Registration number for the software
Sales Order Number
_________________________
Service Order (SO) Number
__________________________
Date completed
__________________________
Customer Signature
__________________________
Support Provider Signature
__________________________
Issued: Sept 2008 – Rev. 6.4 Copyright 2008 Agilent Technologies WAD Product Support Select Pages for Required Modules
Agilent 1100/1200 Series
Liquid Chromatograph
Hardware and Software Installation Checklist
*01200-90606* Part Number: 01200-90606 © Agilent Technologies, Inc. 2008
*01200-90606*
01200-90606 Edition 02 / 2008 Agilent Technologies
Printed in Germany Hewlett-Packard-Strasse 8
Issued: Sept 2008 – Rev. 6.4 Copyright 2008 Agilent Technologies WAD Product Support Select Pages for Required Modules