DMTA 10022 01EN - Rev - B 45MG User
DMTA 10022 01EN - Rev - B 45MG User
DMTA-10022-01EN — Rev. B
October 2013
This instruction manual contains essential information on how to use this Olympus product safely and effectively.
Before using this product, thoroughly review this instruction manual. Use the product as instructed.
Keep this instruction manual in a safe, accessible location.
Olympus NDT, 48 Woerd Avenue, Waltham, MA 02453, USA
This document was prepared with particular attention to usage to ensure the
accuracy of the information contained therein, and corresponds to the version of
the product manufactured prior to the date appearing on the title page. There
could, however, be some differences between the manual and the product if the
product was modified thereafter.
All brands are trademarks or registered trademarks of their respective owners and
third party entities.
DMTA-10022-01EN, Rev. B, October 2013
Table of Contents
iv Table of Contents
DMTA-10022-01EN, Rev. B, October 2013
Table of Contents v
DMTA-10022-01EN, Rev. B, October 2013
10. About Custom Setups for Single Element Transducers ................... 165
10.1 Creating a Custom Setup for a Single Element Transducer .............................. 166
10.2 Quickly Adjusting Waveform Parameters For Single Element Transducers . 168
10.3 About the Detection Modes ................................................................................... 170
10.4 About the First Peak ............................................................................................... 172
10.5 About the Pulser Power ......................................................................................... 173
10.6 About the Time-Dependent Gain Curve ............................................................. 174
10.6.1 About the Maximum Gain .......................................................................... 175
10.6.2 About the Initial Gain .................................................................................. 175
10.6.3 About the TDG Slope ................................................................................... 176
10.7 About the Main Bang Blank .................................................................................. 176
vi Table of Contents
DMTA-10022-01EN, Rev. B, October 2013
List of Abbreviations
2-D two-dimensional
AEtoE automatic Echo-to-Echo
AGC automatic gain control
CSV comma separated variables
DB database
DIAG diagnostic
DIFF differential
EFUP environment-friendly use period
ESS electronic stress screening
EXT extended
FRP fiber-reinforced polymer
GB gigabytes
GRN green
HDPE high-density polyethylene
HI high
ID identification
LDPE low-density polyethylene
LOS loss-of-signal
MAX maximum
MB main bang
MEtoE manual Echo-to-Echo
MIL military
MIN minimum
NiMH nickel-metal hydride
PDF portable document format
PRF pulse repetition frequency
PVC polyvinyl chloride
SE single element
STD standard
SW software
TDG time-dependent gain
TFT thin-film transistor (liquid crystal display technology)
TOF time-of-flight
USB universal serial bus
YEL yellow
List of Abbreviations ix
DMTA-10022-01EN, Rev. B, October 2013
x List of Abbreviations
DMTA-10022-01EN, Rev. B, October 2013
Safety-related labels and symbols are attached to the instrument at the locations
shown in Figure i-1 on page 1, and Figure i-2 on page 2. If any or all of the labels or
symbols are missing or illegible, please contact Olympus.
To avoid the risk of electric shock, do not touch the inner conductor of the T/R 1 and
T/R 2 connectors. Up to 200 volts can be present on the inner conductor. The danger
symbol between the Transmit/Receive (T/R) connector markings is shown in
Figure i-1 on page 1.
Serial number
label yynnnnnmm
Where:
yy: last two digits of the production year
nnnnn: five-digit non duplicated incrementing number
representing the nth production unit of this product
mm: production month
Instruction label:
Contains:
The C-Tick label indicates that the product complies with the
applicable standard, and establishes a traceable link between
the equipment and the manufacturer, importer, or agent
responsible for compliance, and for placing it on the
Australian market.
Intended Use
Do not use the 45MG for any purpose other than its intended use. It must never be
used to inspect or examine human or animal body parts.
Instruction Manual
This instruction manual contains complete and essential information on how to use
this Olympus product safely and effectively. Before use, thoroughly review this
instruction manual, and use the product as instructed. Keep this instruction manual
in a safe, accessible location.
Apart from the batteries, the 45MG instrument does not contain any user-serviceable
parts.
Safety
This instrument has been tested according to IEC Publication 61010: Safety
Requirements for Electronic Measuring Apparatus. This instruction manual contains
warnings and safety rules that must be observed by the user to ensure safe operation
of the instrument, and maintain it in safe condition. Please read through these
operating instructions before using the instrument.
Safety Symbols
The following safety symbols might appear on the instrument and in the instruction
manual:
The following safety symbols might appear in the documentation of the instrument:
The following safety symbols could appear in the documentation of the instrument:
The IMPORTANT signal word calls attention to a note that provides important
information, or information essential to the completion of a task.
The NOTE signal word calls attention to an operating procedure, practice, or the like,
which requires special attention. A note also denotes related parenthetical
information that is useful, but not imperative.
The TIP signal word calls attention to a type of note that helps you apply the
techniques and procedures described in the manual to your specific needs, or
provides hints on how to effectively use the capabilities of the product.
Warnings
General Warnings
• Carefully read the instructions contained in this instruction manual prior to
turning on the instrument.
Battery Precautions
• Before disposing of a battery, check your local laws, rules, and regulations, and
follow them accordingly.
• When lithium metal batteries are used, transportation of the batteries are
regulated by the United Nations under the United Nations Recommendations on the
Transport of Dangerous Goods. It is expected that governments, intergovernmental
organizations, and other international organizations shall conform to the
principles laid down in these Regulations, thus contributing to worldwide
harmonization in this field. These international organizations include the
International Civil Aviation organization (ICAO), the International Air Transport
Association (IATA), the International Maritime Organization (IMO), the US
Department of Transportation (USDOT), and others.
Please contact the transporter and confirm current regulations before
transportation of lithium metal battery.
Equipment Disposal
Before disposing of this product, make sure that you follow the regulations and rules
of your local government.
WEEE Directive
China RoHS
This equipment generates and uses radio-frequency energy and, if not installed and
used properly (that is, in strict accordance with the manufacturer’s instructions), may
cause interference. The 45MG has been tested and found to comply with the limits for
an industrial device in accordance with the specifications of the EMC directive.
This equipment has been tested and found to comply with the limits for a Class A
digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to
provide reasonable protection against harmful interference when the equipment is
operated in a commercial environment. This equipment generates, uses, and can
radiate radio frequency energy, and if not installed and used in accordance with the
instruction manual, might cause harmful interference to radio communications.
Operation of this equipment in a residential area is likely to cause harmful
interference, in which case you will be required to correct the interference at your own
expense.
Regulatory Information
The 45MG may display a regulatory screen that lists the specific regulation with
which it complies.
3. Use the up and down arrow keys to scroll through the different REGULATORY
screens.
4. Press [MEAS] to return to the measurement screen.
Warranty Information
Olympus guarantees your Olympus product to be free from defects in materials and
workmanship for a specific period, and in accordance with the conditions specified in
the Olympus NDT Terms and Conditions available at http://www.olympus-
ims.com/en/terms/.
The Olympus warranty only covers equipment that has been used in a proper
manner, as described in this instruction manual, and that has not been subjected to
excessive abuse, attempted unauthorized repair, or modification.
This instruction manual explains the proper operation of your Olympus product. The
information contained herein is intended solely as a teaching aid, and shall not be
used in any particular application without independent testing and/or verification by
Olympus reserves the right to modify any product without incurring the
responsibility for modifying previously manufactured products.
Technical Support
Olympus is firmly committed to providing the highest level of customer service and
product support. If you experience any difficulties when using our product, or if it
fails to operate as described in the documentation, first consult the user’s manual, and
then, if you are still in need of assistance, contact our After-Sales Service. To locate the
nearest service center, visit the Service Centers page at:
http://www.olympus-ims.com.
1. Instrument Description
This chapter describes the main features and hardware components of the 45MG
instrument.
The PDF file for the 45MG Ultrasonic Thickness Gage — User’s Manual (P/N:
DMTA-10022-01EN) is included on the documentation CD (P/N: 45MG-MAN-CD
[U8147024]) that is shipped with the 45MG.
The 45MG by Olympus is a handheld ultrasonic thickness gage designed for a wide
variety of thickness-measurement applications. With the 45MG, you only need access
to one side of a part in order to obtain nondestructive measurements of the thickness
of corroded, pitted, scaled, granular, and other difficult materials.
The 45MG displays a thickness readout and offers an optional A-scan view for
waveform verification. The microprocessor of the 45MG continuously adjusts the
receiver setup so that every measurement is optimized for reliability, range,
sensitivity, and accuracy. An optional advanced internal datalogger can store up to
475000 thickness measurements and 20000 waveforms.
Instrument Description 15
DMTA-10022-01EN, Rev. B, October 2013
With the Single Element option, the 45MG operates with a full line of single element
transducers. Operation with dual element transducers is a standard feature.
Depending on the software options, the 45MG is capable of measuring material
thicknesses between 0.003 in. and 25.0 in. (0.08 mm and 635.0 mm). The temperature
range of measured materials may vary between −4 °F and 932 °F (−20 °C and 500 °C),
depending on the material characteristics, the transducer, and the measurement
mode.
Basic features
• Measurement-related status flags and alarms
• Quarter VGA color transflective LED back-lite display
• Automatic probe recognition for the standard D79X and MTD705 series
transducers
• Warning against calibration doubling (for dual element transducers)
• Calibration for unknown material sound velocity and/or transducer zero
• Fast scan mode with 20 readings per second
• Hold or blank thickness display during loss-of-signal (LOS) conditions
• Hold minimum and maximum functions
• Differential thickness display relative to the set point in absolute values or
percentage ratios
• Selection of password-protected lockout functions
• Selectable resolution: low of 0.01 in. (0.1 mm), standard of 0.001 in. (0.01 mm), or
high (optional) of 0.0001 in. (0.001 mm); [option not available for all transducers]
Optional features
• Single element transducers
• Echo-to-Echo and THRU-COAT measurements
• High Penetration software for low-frequency single element transducers
• A-scan or waveform display
— Real-time A-scan waveform display for verification of critical measurements
— Manual freeze mode with post processing
— Manual zoom and range control of waveform display
— Auto hold on loss of signal (LOS) and auto zoom (measured echo centering)
— Extended blank
— Blank after first received echo in Echo-to-Echo mode
16 Chapter 1
DMTA-10022-01EN, Rev. B, October 2013
The 45MG is a rugged and durable instrument that can be used in harsh
environments. The 45MG was designed to meet the requirement of the IP67 rating
(Ingress Protection rating).
Instrument Description 17
DMTA-10022-01EN, Rev. B, October 2013
Olympus cannot guarantee any level of ingress protection rating once the instrument
seals have been manipulated. You must use sound judgment, and take proper
precautions before exposing the instrument to harsh environments.
To maintain the original level of ingress protection, you are responsible for the proper
care of all routinely exposed membrane seals. Additionally, you are responsible for
returning the instrument to an authorized Olympus service center on an annual basis
to ensure that the instrument seals are properly maintained.
The 45MG front panel features a color display and a keypad. The instrument comes
with a wrist strap. An optional protective rubber boot includes a dust flap seal for the
USB communication connector, strap rings at the four corners, and a stand at the back
of the instrument (see Figure 1-1 on page 18).
Keypad
Stand
Power key
Figure 1-1 The 45MG hardware components – Front, top, and side views
18 Chapter 1
DMTA-10022-01EN, Rev. B, October 2013
1.4 Connectors
Figure 1-2 on page 19 illustrates the possible connections between the 45MG and
external devices.
Transducer
USB connector
MicroSD memory
card
The USB and Transmit/Receive transducer connectors are located on the top of the
45MG (see Figure 1-3 on page 19).
Transmit/Receive transducer
connector 2
Instrument Description 19
DMTA-10022-01EN, Rev. B, October 2013
The external MicroSD memory card slot is located behind the battery door (see
Figure 1-4 on page 20).
Battery compartment
The 45MG comes either with the English or the international keypad (see Figure 1-5
on page 21). The functions are the same for both keypads. On the international
keypad, the text labels on many keys are replaced by pictograms. In this document,
keypad keys are referred to using the English label in bold and within brackets (ex.:
[MEAS]).
20 Chapter 1
DMTA-10022-01EN, Rev. B, October 2013
Each key is labelled according to its primary function. The area immediately above
certain keys contains a secondary key function that can be activated by first pressing
[2nd F]. Throughout this document, references to a secondary function are written as
follows: [2nd F], [Primary] (Secondary). For example, the instruction to activate the
REF Value function is written as follows: “Press [2nd F], [SAVE] (REF Value)”.
The [], [], [], and [] keys, along with the [ENTER] key, are used to select menu
items or screen parameters, and to change parameter values. Use the [MEAS] key at
any time to return to the measurement screen. The yellow keys are related to
calibration.
Table 2 on page 21 lists the key functions available on the 45MG keypad. Many
functions are optional, and may not be available depending on which software
options have been purchased.
Instrument Description 21
DMTA-10022-01EN, Rev. B, October 2013
Up arrow
• In a screen or a list, moves to the previous
element.
• For some parameters, a numerical entry
increases the value.
Down arrow
• In a screen or a list, moves to the next
element.
• For some parameters, a numerical entry
decreases the value.
22 Chapter 1
DMTA-10022-01EN, Rev. B, October 2013
Instrument Description 23
DMTA-10022-01EN, Rev. B, October 2013
24 Chapter 1
DMTA-10022-01EN, Rev. B, October 2013
This chapter describes how to power the 45MG using different power options.
The power indicator is always present on the left side of the screen, and it shows the
type of power source used. The 45MG can be powered by three AA-size batteries, by a
computer through its USB connector, or by a commercially available 5-volt USB
power supply.
When using batteries, the vertical green bar in the power indicator indicates the
remaining battery level (see Figure 2-1 on page 25). Each graduation mark represents
20 % of the level.
When using a computer or a 5-volt USB power supply, the power indicator is
represented by the USB logo, or by the AC logo, respectively (see Figure 2-2 on
page 26).
Figure 2-2 The power indicator when using a computer or an AC power supply
The 45MG can also be operated using three AA-size nickel-metal hydride (NiMH)
rechargeable batteries. The 45MG does not recharge NiMH batteries. To recharge the
batteries, you must use a commercially available external battery charger (not
included).
26 Chapter 2
DMTA-10022-01EN, Rev. B, October 2013
When the batteries are full (100 % level), the battery power indicator will show full
bars.
Batteries
Battery compartment
door
Battery compartment
gasket
Battery compartment
door lock
Do not replace the batteries while the instrument is on. Dispose of used batteries
promptly. Keep batteries out of reach of children. The batteries used in this device
may present a risk of fire or chemical burn if mistreated. Do not disassemble, heat
above 50 °C, or incinerate the batteries.
28 Chapter 2
DMTA-10022-01EN, Rev. B, October 2013
When replacing batteries, make sure that they are full in order to ensure the accuracy
of the estimated remaining battery life shown by the power indicator.
30 Chapter 2
DMTA-10022-01EN, Rev. B, October 2013
The following sections describe the main elements of the 45MG software screens and
menus.
The first screen (see Figure 3-1 on page 31) displays when the Waveform option is
not activated, or when the Waveform option feature is turned off.
OR
The second screen (see Figure 3-2 on page 32) displays when the Waveform
option is purchased and turned on.
Measurement units
Delay Range
Thickness value
Measurement units
Mode and update rate
Gain level
The measurement screen is the main screen of the 45MG software. From anywhere in
the 45MG software, simply press [MEAS] to return to the measurement screen. The
power indicator is always present on the left side of the 45MG screen (see section
“About the Power Indicator” on page 25 for details).
The optional waveform trace, called the A-scan, allows a skilled operator to verify that
the signal used to make a thickness measurement is the correct back-wall echo, and
not noise, material anomaly, or the second multiple echo. The A-scan also enables you
to observe indications that may be too small to be measured by the instrument.
The ID bar (available with the Datalogger option), located at the top of the
measurement screen, contains the ID for the actual thickness-measurement location
and the previously stored value (see Figure 3-3 on page 32). The download indicator
( ) appears when the previously stored thickness measurement comes from a file
rather than a newly acquired value.
32 Chapter 3
DMTA-10022-01EN, Rev. B, October 2013
The indicator ( ) of the optional external MicroSD memory card appears at the top
left side of the screen when a MicroSD memory card is inserted in its slot, which is
located in the battery compartment (see Figure 1-4 on page 20). The 45MG will
recognize an external MicroSD memory card, with a maximum capacity of 2 GB,
upon instrument start-up.
Depending on the context, and on the available functions and options, various
indicators and numeric values appear around the waveform display and around the
main measurement value (see Figure 3-4 on page 33 and Figure 3-5 on page 34). A
help text bar will appear upon certain key presses in order to indicate the keys that
can be used to navigate and make selections in the menu structure.
Rectification indicator
The loss of signal (LOS) appears and the thickness value is cleared when the 45MG no
longer detects ultrasonic echoes (see Figure 3-6 on page 34).
Loss-of-signal (LOS)
indicator (or SAT in the
case of saturation)
In the rare case that the returning signal from the transducer is beyond the upper
receiver-voltage limit, the LOS indicator will be replaced by the SAT flag for
indication of saturation. This situation can normally be corrected by lowering the
pulser voltage.
34 Chapter 3
DMTA-10022-01EN, Rev. B, October 2013
The 45MG displays menus and submenus when you press certain front panel keys.
The menu appears at the top-left corner of the screen (see Figure 3-7 on page 35). If
applicable, a submenu conveniently showing the parameters available for the
highlighted menu command also appears.
The 45MG parameters are logically grouped in parameter screens that can be accessed
using front panel keys or menu commands. Figure 3-8 on page 36 shows the MEAS
parameter screen as an example.
Parameters
The title bar, located at the top of the parameter screen, indicates the parameter
subject. When you access a parameter screen from a menu, a menu button appears on
the left side of the title bar. This menu button can be used to easily return to the
original menu. One or two help text bars appear at the bottom of the screen to indicate
the keys required to select a parameter and edit its value.
36 Chapter 3
DMTA-10022-01EN, Rev. B, October 2013
In the remainder of this document, the above procedure is summarized by the simple
instruction to select a specific parameter or list, and its value. For example:
“In the MEAS screen, set MEAS MODE to THICKNESS.”
The 45MG offers two methods of editing the value of alphanumeric parameters.
Either the virtual keyboard or traditional method can be used. The virtual keyboard
appears on the screen to display all the available characters that can be used (see
section “Editing Text Parameters Using the Virtual Keyboard” on page 37 for details).
With the traditional method, you must select each character from a hidden list of
standard sorted letters, numbers, and special characters (see section “Editing Text
Parameters Using the Traditional Method” on page 39 for details).
Title bar
Parameter value text
box
Virtual keyboard
38 Chapter 3
DMTA-10022-01EN, Rev. B, October 2013
When editing a multiple line parameter value, highlight DONE and press [ENTER] to
move the cursor to the next line. You can also press [2nd F], [] to accept the text and
move the cursor to the next line.
[]
Cycling begins at “A” when the
original character is a letter.
Figure 3-10 The character cycle of the traditional text edit method
2. Use the [] and [] keys to select the character you want to enter. Hold down the
key to quickly cycle through the letters, numbers, and special characters.
3. Use the [] keys to move to the next character.
4. Repeat step 2 and step 3 to enter other characters.
5. If you need to move the position of the cursor in the value text box, use the [] or
[] key.
6. To insert a character at the cursor position, press [CAL ZERO].
The character at the cursor, and all those to the right, move one position to the
right, making space for one new character.
7. To delete the character at the cursor position, press [CAL VEL].
The character at the cursor is deleted, and any characters to the right move to the
left by one position.
8. Press [ENTER] to accept the character string and move to the next parameter.
40 Chapter 3
DMTA-10022-01EN, Rev. B, October 2013
4. Initial Setup
4.1 Setting the User Interface Language and Other System Options
The 45MG can be configured to display the user interface in the following languages:
English, German, French, Spanish, Japanese, Chinese, Russian, Swedish, Italian,
Norwegian, Portuguese, or Czech. You can also set the character delimiting the radix
of a number.
The 45MG includes a beep tone generator that confirms when a key is pressed and
notifies you of an alarm condition. The beeper can be set to On or Off.
To save battery power while the instrument is not in use, enable the inactive time
function so that the instrument automatically turns off when no key has been pressed
and no measurement has been made within a period of about six minutes.
Initial Setup 41
DMTA-10022-01EN, Rev. B, October 2013
The 45MG has a built-in date and time clock. You can set the date and the time and
select their format. The 45MG saves all measurement values with their acquisition
date.
42 Chapter 4
DMTA-10022-01EN, Rev. B, October 2013
3. In the CLOCK screen (see Figure 4-3 on page 43), set the parameters to the
current date and time.
4. Set the desired DATE MODE and HOUR MODE, and then select SET.
Initial Setup 43
DMTA-10022-01EN, Rev. B, October 2013
Many of the display parameters are only visible when the Waveform or Datalogger
options have been activated.
2. In the DISPLAY screen (see Figure 4-4 on page 45), select the desired parameter
and value for the following parameters:
• WAVEFORM ENABLE allows you to activate or deactivate the waveform
display (see section “Waveform Software Option” on page 78 for details).
• ZOOM OPTION allows you to activate or deactivate the zoom function (see
section “Activating the Zoom Function (Available Only with the Waveform
Option)” on page 84 for details).
• RECTIFICATION allows you to select one of the rectification modes (see
section “About Waveform Rectification” on page 80 for details).
• WAVEFORM TRACE is used to select one of the trace types (see section
“About the Waveform Trace” on page 82 for details).
• COLOR SCHEME is used to select either the INDOOR or OUTDOOR
optimized visibility (see section “About Color Schemes” on page 45 for
details).
• DISPLAY BRIGHTNESS is used to select one of the predefined brightness
levels (see section “About Display Brightness” on page 46 for details).
• ID REVIEW LINE is used to activate or deactivate the feature.
44 Chapter 4
DMTA-10022-01EN, Rev. B, October 2013
Initial Setup 45
DMTA-10022-01EN, Rev. B, October 2013
Figure 4-5 Example of the indoor and the outdoor color schemes
The indoor scheme gives the best visibility when the instrument is used indoors or in
low-light conditions. The indoor scheme displays green characters and waveform
trace on a black background.
The outdoor scheme provides the best visibility when the instrument is used in direct
sunlight. The outdoor mode displays black characters and waveform trace on a white
background. For best readability in this document, most screen captures are shown in
the outdoor color scheme.
46 Chapter 4
DMTA-10022-01EN, Rev. B, October 2013
Reducing the display brightness percentage increases the battery life. Battery life
specifications are based on the backlight brightness set to 20 %.
Initial Setup 47
DMTA-10022-01EN, Rev. B, October 2013
Measurement update
rate indicator
The Fast update rate can be as high as 20 Hz, and is dependent on the measurement
type. This option is useful when making high-temperature thickness measurements in
order to help limit transducer contact time, or for applications where the transducer is
scanned over an area in order to determine the minimum thickness.
The 45MG automatically uses the fastest update rate when put into the Minimum or
Maximum mode (see section “Using the Minimum, Maximum, or Min/Max
Thickness Mode” on page 129).
48 Chapter 4
DMTA-10022-01EN, Rev. B, October 2013
The resolution can be reduced in certain applications where the extra precision of the
last digit is not required, or where extremely rough outside or inside surfaces make
the last thickness display digit unreliable.
The Single Element including High Resolution software option (P/N: 45MG-SE
[U8147022]) makes it possible to increase the resolution to either 0.0001 in. or
0.001 mm. High resolution is available for measured thickness smaller than 4 in.
(102 mm). High resolution is not active with low-frequency transducers, or when the
High Penetration software option is activated.
Initial Setup 49
DMTA-10022-01EN, Rev. B, October 2013
50 Chapter 4
DMTA-10022-01EN, Rev. B, October 2013
5. Basic Operation
The following sections describe the basic operation for the 45MG ultrasonic thickness
gage.
The 45MG operates with a full line of single element (optional software) and dual
element transducers. The 45MG automatically recognizes standard D79X dual
element transducers, and automatically loads the appropriate predefined setup. The
predefined setup contains ultrasonic velocity for the stainless steel step block
supplied with the instrument. With dual element transducers, you need to perform a
transducer zero compensation.
For Single Element or High Penetration software options and single element
transducers, you need to manually recall an appropriate setup. The 45MG is factory
shipped with default conditions for the purchased transducers set up using an
approximate sound velocity for the stainless steel test block provided with the
instrument. The default conditions are selected to facilitate instrument usage for your
applications.
Basic Operation 51
DMTA-10022-01EN, Rev. B, October 2013
Figure 5-2 Initial screen with a standard D79X dual element transducer
52 Chapter 5
DMTA-10022-01EN, Rev. B, October 2013
4. For the Single Element software option and a single element transducer, load an
appropriate setup:
a) Press [2nd F], [FREEZE] (XDCR RECALL).
b) In the menu, select the default choice for the probe type that you use (ex.:
DEFAULT SINGLE ELEMENT).
c) In the DEFAULT screen for the type of probe you are using (see the example
shown in Figure 5-3 on page 53), highlight the setup in the list of available
default setups for the transducer that you are using.
Basic Operation 53
DMTA-10022-01EN, Rev. B, October 2013
The setups listed as USER-1 through USER-35 can be renamed for special
applications. Refer to chapter “About Custom Setups for Single Element Transducers”
on page 165 for more details on setups.
d) Press [MEAS] to automatically recall the setup parameters for the chosen
setup and return to the measurement screen.
54 Chapter 5
DMTA-10022-01EN, Rev. B, October 2013
absolute accuracy (better than ±0.004 in. or ±0.10 mm). The procedure only needs
to be performed once for each new transducer and material combination. It is not
necessary to repeat the zero calibration when the transducer temperature
changes; transducer zero compensation is responsible for that task (see section
“Calibrating the Instrument” on page 55 and section “About the Material Sound
Velocity and the Zero Calibrations” on page 59).
The following procedure is illustrated using a dual element probe and a five-step test
block. Refer to section “About the Calibration” on page 54 for more details on the
calibration process.
Basic Operation 55
DMTA-10022-01EN, Rev. B, October 2013
0.500 in.
Figure 5-4 Sound velocity calibration on a five-step test block – Thick part
Figure 5-5 Performing the sound velocity calibration on a five-step test block
56 Chapter 5
DMTA-10022-01EN, Rev. B, October 2013
0.100 in.
Figure 5-6 Sound velocity calibration on a five-step test block – Thin part
3. Press [MEAS] to complete the calibration and return to the measurement screen.
If you turn off the instrument before pressing [MEAS], the velocity will not be
updated to the new value. Instead, the instrument retains the previous value.
Basic Operation 57
DMTA-10022-01EN, Rev. B, October 2013
When the 45MG detects an error in the calibration procedure, it successively displays
the following messages in the help text bar before returning to the measurement
screen:
“Potential wrong echo detected!”
“Invalid calibration results!”
In such case, the velocity is not changed. The probable cause is that an incorrect
thickness value was entered, or the incorrect back-wall echo was detected.
Precision step test blocks are also often used when more than two known thicknesses
are needed (see Figure 5-8 on page 58).
When performing the material sound velocity and zero calibrations, you must use a
test block with the following characteristics:
58 Chapter 5
DMTA-10022-01EN, Rev. B, October 2013
When measurements are made on surfaces that are significantly above room
temperature, the zero point should be recalibrated on a regular basis. This is less
important for transducers with part numbers D790-SM, D791-RM, D797-SM, and
D798 than for other transducers with various types of resin delay lines.
5.2.4 About the Material Sound Velocity and the Zero Calibrations
The 45MG performs a calibration doubling verification to help prevent mis-
calibration on thin samples. Doubling occurs when the instrument measures the time
of flight to the second back-wall echo rather than measuring the time of flight to the
first back-wall echo. The 45MG compares the measured time of flight to the expected
time of flight based on the current sound velocity. The 45MG displays a warning
message if doubling is suspected. Doubling can occur when measuring a thickness
that is below the minimum range of the transducer, or when using a transducer that is
worn out, or getting low in sensitivity.
Basic Operation 59
DMTA-10022-01EN, Rev. B, October 2013
You can also achieve a material sound velocity and zero calibration procedure by
performing the operations in the reverse order, starting with the zero calibration,
followed by the material sound velocity calibration.
60 Chapter 5
DMTA-10022-01EN, Rev. B, October 2013
Message
Calibration
The accuracy of any ultrasonic measurement is only as good as the accuracy and
care with which you calibrate the instrument. The 45MG ships from the factory
with standard setups for a number of transducers and applications. In certain
cases, it may be desirable to optimize these setups for specific measurement
situations. In all cases, it is essential to perform the velocity and zero calibrations
whenever the test material or transducer is changed. Periodic checks with
samples of known thicknesses are recommended to verify that the gage is
operating properly.
Surface roughness of the test piece
The best measurement accuracy is obtained when both the front and back
surfaces of the test piece are smooth. When the contact surface is rough, the
minimum thickness that can be measured will be increased as a result of sound
reverberating in the increased thickness of the couplant layer. Additionally, when
the two test-piece surfaces are rough, the slightly different multiple sound paths
seen by the transducer may cause distortion in the returning echo, resulting in
measurement inaccuracies.
Basic Operation 61
DMTA-10022-01EN, Rev. B, October 2013
Coupling technique
In mode 1 (contact transducer) measurements, the couplant layer thickness is part
of the measurement, and is compensated by a portion of the zero offset. If
maximum accuracy is to be achieved, the coupling technique must be consistent.
In order to accomplish consistent measurements, use a couplant of reasonably
low viscosity, employ only enough couplant to achieve a reasonable reading, and
apply the transducer with uniform pressure. Practice will determine the degree of
moderate to firm pressure needed to produce repeatable readings. In general,
smaller-diameter transducers require less coupling force to squeeze out excess
couplant than larger-diameter transducers. In all modes, tilting the transducer
distorts echoes, and causes inaccurate readings, as noted below.
Curvature of the test piece
A related issue involves the alignment of the transducer with respect to the test
piece. When measuring on curved surfaces, it is important that the transducer be
placed approximately on the centerline of the part, and held steadily as close to
the surface as possible. In some cases, a spring-loaded V-block holder may be
helpful in maintaining this alignment. In general, as the radius of curvature
decreases, the size of the transducer should be reduced, and transducer alignment
becomes more critical. For very small radii, an immersion approach is necessary.
In some cases, it may be useful to observe the waveform display as an aid to
maintaining optimum alignment. Observe the best method for holding a
transducer with the aid of a waveform display. On curved surfaces, it is important
to use only enough couplant to obtain a reading. Excess couplant forms a fillet
between the transducer and the test surface where sound reverberates, and
possibly creates spurious signals that may trigger false readings.
Taper or eccentricity
If the contact surface or back surface of the test piece is tapered or eccentric with
respect to the other, the return echo is distorted due to the variation of the sound
path across the width of the beam. The accuracy of the measurement is therefore
reduced. In severe cases, no measurement is possible.
Acoustic properties of the test material
There are several conditions in certain engineering materials that can potentially
limit the accuracy and range of ultrasonic thickness measurements:
• Sound scattering:
In materials such as cast stainless steel, cast iron, fiberglass, and composites,
sound energy scatters from individual crystallites in the casting, or from
boundaries of dissimilar materials within the fiberglass or composite.
Porosity in any material can have the same effect. Adjust the instrument
sensitivity to prevent detection of these spurious scatter echoes. This
62 Chapter 5
DMTA-10022-01EN, Rev. B, October 2013
compensation can in turn limit the ability to discriminate a valid return echo
from the back wall of the material, thereby restricting the measurement range.
• Sound attenuation or absorption:
In many organic materials, such as low-density plastics and rubbers, sound
energy is attenuated very rapidly at the frequencies used for ultrasonic
gaging. This attenuation typically increases with temperature. The maximum
thickness that can be measured in these materials is often limited by
attenuation.
• Velocity variations:
An ultrasonic thickness measurement is accurate only to the degree that
material sound velocity is consistent with the instrument calibration. Some
materials exhibit significant variations in sound velocity from point to point.
This happens in certain cast metals due to the changes in grain structure that
result from varied cooling rates, and the anisotropy of sound velocity with
respect to grain structure. Fiberglass can show localized velocity variations
due to changes in the resin/fiber ratio. Many plastics and rubbers show a
rapid change in sound velocity with temperature, and as such, velocity
calibration must be performed at the temperature of the location where the
measurements are made.
Phase reversal or phase distortion
The phase or polarity of a returning echo is determined by the relative acoustic
impedances (density × velocity) of the boundary materials. The 45MG performs
computation based on the customary situation, where the test piece is backed by
air or a liquid, both of which have a lower acoustic impedance than metals,
ceramics, or plastics. However, in some specialized cases, such as measurement of
glass or plastic liners over metal, or copper cladding over steel, this impedance
relationship is reversed, and the echo appears phase-reversed. In such cases, it is
necessary to change the appropriate echo detection polarity in order to maintain
accuracy (see section “About the Detection of Echo 1 and Echo 2” on page 179). A
more complex situation can occur in anisotropic or in homogeneous materials,
such as coarse-grain metal castings or certain composites, where material
conditions result in the existence of multiple sound paths within the beam area. In
such cases, phase distortion can create an echo that is neither cleanly positive nor
negative. Careful experimentation with reference standards is necessary in such
cases in order to determine the effects on measurement accuracy.
Basic Operation 63
DMTA-10022-01EN, Rev. B, October 2013
To measure thicknesses
1. Apply couplant to the measurement spot on the test block or tested part.
In general, use a thinner couplant (such as propylene glycol, glycerin, or water) for
smooth material surfaces. Rough surfaces require a more viscous couplant, such as
gel or grease. Special couplants are required for high-temperature applications.
2. Using moderate to firm pressure, couple the tip of the transducer to the surface of
the test material, and keep the transducer as flat as possible on the material
surface (see Figure 5-11 on page 64).
0.500 in.
3. Read the measured thickness value for the tested part (see Figure 5-12 on
page 65).
64 Chapter 5
DMTA-10022-01EN, Rev. B, October 2013
The 45MG optional datalogger is a file-based system, in which one file is opened at a
time. The active file stores a measurement at a thickness-measurement location ID.
Each time you press [SAVE], the displayed value is saved to the active file at the
current ID. The ID is automatically incremented for the next measurement. When you
press [FILE], the name of the active file appears in the ID bar above the menu (see
Figure 5-13 on page 65).
Basic Operation 65
DMTA-10022-01EN, Rev. B, October 2013
The NONAME00 increment-type file, starting with the 001 ID, is the active file by
default when you first use the 45MG, or after resetting the instrument memory.
Various types of files can be created, and IDs can be defined to represent various 1-D,
2-D, or 3-D thickness-measurement locations. When the instrument is restarted, it
automatically opens the last file used.
• When the thickness value is blank, “———” is saved rather than a value.
• When a measurement is already stored at the current ID, the new value
overwrites the old thickness reading unless the overwrite protection is enabled
(see section “Setting the ID Overwrite Protection” on page 114).
• When the ID increment reaches the end of a sequence and cannot be updated,
Last ID appears on the help text bar, a long beep sound is emitted (when the
beeper is active), and the ID on the display remains unchanged.
Refer to section “Datalogger Option” on page 88 for more information on the
datalogger.
66 Chapter 5
DMTA-10022-01EN, Rev. B, October 2013
6. Software Options
The available software options can be used to increase the capability of the already
versatile 45MG (see Table 3 on page 67).
Software Options 67
DMTA-10022-01EN, Rev. B, October 2013
When you order a software option with your original 45MG purchase, the option
comes preactivated. Software options can also be purchased at a later date. The
software options can be activated by entering an activation code in the instrument,
without the need to return the instrument to the factory. (see section “Activating
Software Options” on page 68).
Contact your local Olympus representative for more information about ordering
software options. Refer to Table 3 on page 67 for the software-option part numbers.
Each 45MG has a unique serial number code. A provided option key specific to a
particular 45MG activates the purchased software options only on that specific 45MG
unit. A single-option key is capable of activating one, several, or all software options.
68 Chapter 6
DMTA-10022-01EN, Rev. B, October 2013
Alphanumeric serial
number
Software Options 69
DMTA-10022-01EN, Rev. B, October 2013
With dual element transducers, the 45MG offers three echo detection modes that
allow you to measure thicknesses in various material conditions. A description of
each of the three echo detection modes (STANDARD, optional AUTO E-TO-E, and
MANUAL E-TO-E) follows:
STANDARD
The standard echo detection mode measures the thickness based on the time of
flight between the main bang and the first back-wall echo. Use this mode for
uncoated materials.
The DE-STD indicator appears to the left of the thickness reading, and a
triangular echo detection marker appears at the back-wall echo, below the
waveform display (see Figure 6-2 on page 70).
Echo detection marker Dual element (DE) standard (STD) echo detection mode
70 Chapter 6
DMTA-10022-01EN, Rev. B, October 2013
by an Echo-to-Echo detection bar that indicates the exact pair of back-wall echoes
used to determine the thickness (see Figure 6-3 on page 71). The echo height is
automatically adjusted to a preset level.
MANUAL E-TO-E (optional and only available when Echo-to-Echo and the
Waveform options are activated)
The manual Echo-to-Echo detection mode also measures the thickness using the
time of flight between two consecutive back-wall echoes. However, the gain and
blanking parameters can also be manually adjusted in this mode. This mode is
useful when material conditions produce noisy signals that could render the
automatic mode less effective.
The DE-MEtoE indicator appears to the left of the thickness reading. The Echo-to-
Echo detection bar is similar to the automatic Echo-to-Echo mode, but includes
the adjustable E1 blank bar that indicates the area to exclude for echo detection
(see Figure 6-4 on page 72). Following the E1 blank, the instrument detects the
next echo with an amplitude of at least 20 % of the waveform display height. In
this mode, press [GAIN/WAVE ADJ], and then use the arrow keys to adjust the
EXT BLANK, the E1 BLANK, and the GAIN parameters.
Software Options 71
DMTA-10022-01EN, Rev. B, October 2013
Echo-to-Echo detection
bar
Dual element (DE)
manual Echo-to-Echo
(MEtoE) detection mode
In severe corrosion situations where valid multiple echoes are not present, you must
use either the standard or optional THRU-COAT mode in order to be able to measure
thicknesses.
Dual element transducers can be used in all three modes, and all measurement,
display, and datalogger functions can be used with the echo-to-echo modes. The
internal datalogger recognizes all necessary echo-to-echo information, and saves it in
order to upload and download thickness, waveform, and setup data.
It is not necessary to switch between echo detection modes when measuring both
coated and uncoated areas, because it is possible to measure uncoated wall
thicknesses using an echo-to-echo mode.
72 Chapter 6
DMTA-10022-01EN, Rev. B, October 2013
2. In the MEAS screen, set the MEAS MODE to the desired echo detection mode:
STANDARD, AUTO E-TO-E (optional), or MANUAL E-TO-E (only available
with the Waveform option).
EXT BLANK
The extended blank creates a blanked zone that begins at the left edge of the
waveform display, and in which no signals are detected. In situations where the
second or third pair of back-wall echoes are stronger or cleaner than the first pair,
use the extended blank to control which pair of echoes are used for measurement.
E1 BLANK
The echo 1 (E1) blank runs for a selected interval following the first detected echo.
Use the E1 blank to exclude any unwanted peaks occurring between the first and
second back-wall echoes. Unwanted peaks may be trailing edges of a large first
echo, or shear-wave reflections on thick test pieces. The E1 blank parameter is
only available in the manual Echo-to-Echo detection mode.
Software Options 73
DMTA-10022-01EN, Rev. B, October 2013
3. Use the [] and [] keys to select either the EXT BLANK or the E1 BLANK
parameter.
4. Use the [] and [] keys to adjust the value to exclude unwanted peaks and to
detect the desired echoes.
74 Chapter 6
DMTA-10022-01EN, Rev. B, October 2013
In some cases, an error can occur if you are using a D790 transducer to measure
thicknesses above 0.7 in. (18 mm). Typically, this error is due to a mode-converted
shear-wave echo that may appear before the second back-wall echo. If this unwanted
echo is larger than the second back-wall echo, the gage measures the distance to it,
which produces a thinner reading.
You can usually distinguish the unwanted shear-wave echo from the correct back-wall
echo by examining the waveform display. The distance between the first and the
second back-wall echoes is the same as the distance between the zero thickness point
and the first back-wall echo. If there is an echo between the first two back-wall echoes,
it is probably a mode-converted shear-wave echo. Use the manual echo-to-echo
detection mode techniques, and manually adjust the E1 blank to eliminate this error
(see section “Blanking Adjustments in Manual Echo-to-Echo Detection Mode” on
page 73). Using the D797 transducer beyond 0.7 in. (18 mm) helps eliminate the
possibility of this error.
In some cases, the second or third back-wall echo is smaller in amplitude than
subsequent echoes, which causes the instrument to give a double or triple reading. If
you are using a D790 transducer, this effect may occur around 0.2 in. (5 mm) on flat,
smooth steel samples. If this occurs, it is clearly visible on the waveform display, and
you can work around it using the manual Echo-to-Echo detection mode, or by moving
the extended blank beyond the previously detected first echo.
When the 45MG cannot make an echo-to-echo reading, the LOS flag appears on the
screen. In this case, the waveform display shows that either no echoes are large
enough to be detected, or that only one echo is detectable. In the latter case, the Echo-
to-Echo detection bar begins at the detected echo, but extends indefinitely to the right.
Software Options 75
DMTA-10022-01EN, Rev. B, October 2013
Increase the gain value to make a good echo-to-echo reading. If this does not help, you
can still obtain an approximate measurement by returning to the standard echo
detection mode.
THRU-COAT is an optional function that measures the true metal thickness of coated
or painted parts. This function only requires a single back-wall echo, and is
recommended for heavy corrosion applications where the outside of the material is
coated or painted. If necessary, the measurement of the coating/paint layer can also be
calibrated to precisely measure the thickness of the coating/paint.
In order for the THRU-COAT function to work properly, the thickness of the coating
or paint must be at least 0.005 in. (0.125 mm) thick. The maximum coating/paint
thickness that THRU-COAT can measure is dependent on the type of coating, but
typically greater then 0.080 in. (2 mm).
If the THRU-COAT function does not display a coating thickness, or if the coating
thickness does not seem realistic, THRU-COAT may not be able to properly resolve
the coating thickness. In such cases, the user should try a different measurement
mode, like Echo-to-Echo or standard.
76 Chapter 6
DMTA-10022-01EN, Rev. B, October 2013
Software Options 77
DMTA-10022-01EN, Rev. B, October 2013
9. Using the arrow keys, edit the thickness value to match the known thickness of
the sample.
10. If the coating thickness-measurement accuracy is important to your application,
perform the following actions (omitting this step does not affect the accuracy of
the metal thickness measurement):
a) Press [CAL VEL] again.
b) Couple the transducer to the coated sample.
c) Once the reading is stable, press [ENTER].
d) Using the arrow keys, edit the thickness value to match the known coating
thickness of the coated sample.
11. Press [MEAS] to complete the calibration.
Pressing [2nd F], [CAL VEL] (VEL) opens the VELOCITY screen, in which you can
see and edit the calibrated metal sound velocity. Pressing [2nd F], [CAL VEL] (VEL)
again opens the VELOCITY screen for the calibrated sound velocity through the
coating.
The live Waveform option for the 45MG allows the user to view the live ultrasonic
waveform in order to aid alignment of the transducer during difficult applications.
When this option is activated, the user can switch between the standard thickness
display (see Figure 6-7 on page 79) and the optional waveform thickness display (see
Figure 6-8 on page 80).
• [GAIN/WAVE ADJ] allows the user to make setup adjustments that are not
available in standard mode.
• With dual element transducers (see chapter “Using Advanced Gaging
Features” on page 149):
• Manually adjust the GAIN in 1 dB increments.
• Set an extended blank.
• Set an Echo1 blank in Manual Echo-to-Echo mode.
78 Chapter 6
DMTA-10022-01EN, Rev. B, October 2013
Software Options 79
DMTA-10022-01EN, Rev. B, October 2013
80 Chapter 6
DMTA-10022-01EN, Rev. B, October 2013
Full Half–
Half+ RF
FULL
This mode shows the negative portion of the echo folded around the baseline, so
that both positive and negative waveform lobes are displayed. This mode
provides the best overall representation of position and magnitude for most
thickness-measurement applications. FULL is the default mode for dual element
transducers.
HALF– (NEG indicator)
This mode shows negative waveform lobes as positive, and does not show the
positive lobes.
HALF+ (POS indicator)
This mode shows positive waveform lobes, and does not show the negative
waveform lobes.
RF
This mode shows negative and positive lobes on either side of the baseline. RF is
the default mode for single element transducers.
Software Options 81
DMTA-10022-01EN, Rev. B, October 2013
A filled-in waveform trace is only possible when the waveform rectification is set to
FULL, HALF+, or HALF–.
82 Chapter 6
DMTA-10022-01EN, Rev. B, October 2013
Units
Software Options 83
DMTA-10022-01EN, Rev. B, October 2013
When the [GAIN/WAVE ADJ] parameters are active, the range and delay become
items in the parameter list. Use the [] and [] keys to highlight the range parameter,
and the [] and [] keys to adjust the range. Press [MEAS] to stop adjusting the
range.
When the [GAIN/WAVE ADJ] parameters are active, the range and delay become
items in the parameter list. Use the [] and [] keys to highlight the delay parameter,
and the [] and [] keys to adjust the delay. Press [MEAS] to stop adjusting the delay.
6.4.3.3 Activating the Zoom Function (Available Only with the Waveform Option)
The zoom function automatically and dynamically sets the delay and range values to
optimally track and show the detected echo in the waveform display.
84 Chapter 6
DMTA-10022-01EN, Rev. B, October 2013
The resulting zoomed waveform depends on the current measurement mode. The
zoom for D79X dual element transducers and mode 1 single element transducers
centers the first back-wall echo on the screen (see Figure 6-12 on page 85).
The zoom with single element transducers in mode 2 adjusts the waveform range and
delay so that the interface echo and the first back-wall echo appear on the waveform
display (see Figure 6-13 on page 85).
Software Options 85
DMTA-10022-01EN, Rev. B, October 2013
The zoom with single element transducers in mode 3 adjusts the waveform range and
delay so that the interface echo and the second measured back-wall echo appear on
the waveform display (see Figure 6-14 on page 86).
The optional Single Element High Resolution software allows the 45MG to use single
element direct contact, delay line, and immersion transducers. This enables the 45MG
to be used for precision thickness gaging applications. When this option is activated,
the user can select from a set of default single element transducer setups, or create and
store/recall customers’ single element setups.
The 45MG is not able to automatically recognize which single element transducer is
connected. As such, the appropriate default or customer setup must be recalled for
the single element transducer being used.
86 Chapter 6
DMTA-10022-01EN, Rev. B, October 2013
For single element transducers, the high-resolution software option provides the
added ability to display thickness readings with a high resolution of either 0.0001 in.
or 0.001 mm. High resolution is not available for all transducers and measurement
modes, and is also limited to the maximum thickness. Although the 45MG is able to
display thickness readings in high resolution, the measurement accuracy is highly
dependent on the material, geometry, surface condition, and temperature, and also
needs to be determined on a case-by-case sample evaluation basis.
• Single element transducer in the frequency range of 2.25 MHz to 30.0 MHz
• Thickness measurement below 4.00 in. or 100 mm
High resolution is not available for the following transducers or measurement
conditions:
Software Options 87
DMTA-10022-01EN, Rev. B, October 2013
With a M2008 transducer only, press [2nd F], [CAL ZERO] (Do ZERO) at any time to
auto-adjust the zero offset and compensate for temperature changes in the delay line.
This section describes how to use the 45MG internal datalogger to organize your data.
88 Chapter 6
DMTA-10022-01EN, Rev. B, October 2013
The file also contains header parameters that can be defined to better describe the
content of the file. In the file, you can organize the ID range, select the data format,
and select the saved data. Table 5 on page 90 summarizes the contents of a file, and
indicates where you can find more information.
Software Options 89
DMTA-10022-01EN, Rev. B, October 2013
You can identify the datalogger parameters in the ID bar at the top of the
measurement screen (see Figure 6-16 on page 90).
90 Chapter 6
DMTA-10022-01EN, Rev. B, October 2013
With each measurement, the 45MG also stores a complete description of the
measurement conditions. Table 6 on page 91 describes the additional data stored with
each thickness measurement and with each waveform.
With the datalogger, you can easily create a data file (see section “Creating a Data
File” on page 92), perform a number of file operations (see section “Performing File
Operations” on page 105), and perform data operations (see section “Setting the ID
Overwrite Protection” on page 114).
Software Options 91
DMTA-10022-01EN, Rev. B, October 2013
It is also possible to create a 45MG data file from a computer using the GageView
interface program. Refer to the GageView Interface Program — User’s Manual (P/N:
910-259-EN [U8778347]) for details.
92 Chapter 6
DMTA-10022-01EN, Rev. B, October 2013
At any time, you can press [2nd F], [] or [2nd F], [] to scroll between parameters on
the screen.
• Incremental
• Sequential
• Sequential with Custom Point
• 2-D Matrix Grid
• Boiler
Software Options 93
DMTA-10022-01EN, Rev. B, October 2013
• Increments only digits and letters, not punctuation marks or other special
characters.
• Begins incrementation with the right-most character.
• Extends leftward until reaching the first punctuation mark or special character, or
the left-most character, whichever comes first.
• Increments digits from 0, 1, 2,..., 9. Makes the 9 to 0 transition only after
incrementing the character to the left.
• Increments letters from A, B, C,..., Z. Makes the Z to A transition only after
incrementing the character to the left.
• When an ID cannot be incremented after a reading is saved, the Cannot
increment ID! message momentarily appears in the help text bar. Subsequent
saves overwrite the reading of the last possible ID until the ID value range is
changed.
To configure the gage to increment through a range of numbers several digits wide
while beginning with a single-digit ID number, the maximum number of digit
positions must be entered initially using leading zeroes (see examples in Table 7 on
page 94).
94 Chapter 6
DMTA-10022-01EN, Rev. B, October 2013
Figure 6-18 The CREATE screen for the incremental data file type
Software Options 95
DMTA-10022-01EN, Rev. B, October 2013
96 Chapter 6
DMTA-10022-01EN, Rev. B, October 2013
Figure 6-19 Selecting the ID range for the sequential file type
6.7.2.4 About the Sequential with Custom Point Data File Type
The sequential with custom point (SEQ+CUSTOM PT) data file type is defined by a
starting and an ending ID number, plus a series of custom points. The resulting file is
inclusive of the start and end points, and all points in between. In addition, multiple
thickness readings per ID number location are assigned using the assigned custom
points.
Use the sequential with custom point data file type when you want, for example, to
measure along a pipe or tube on which, at each ID number location, you can take
measurements at the top, bottom, left, and right of the pipe (see an example in Table 9
on page 97).
Software Options 97
DMTA-10022-01EN, Rev. B, October 2013
The allowable number of characters for each custom point depends on the number of
ID characters defined in the start and end ID values. The total number of characters of
the ID value plus the custom points cannot exceed 20. For example, when the start
and end ID values are seven characters long, as shown in the example in Table 9 on
page 97, the maximum allowable length for each custom point is thirteen (20 – 7 = 13).
Figure 6-20 Configuring ID range for a sequential with custom points data file type
98 Chapter 6
DMTA-10022-01EN, Rev. B, October 2013
A 2-D (two-dimensional) sequence begins with the ID number that refers to the first
column and the first row (see Figure 6-21 on page 99). The column (or row) then
increments one value at a time until the sequence reaches the last column (or row)
value, while the other dimension value stays constant. At this point, the other
dimension increments from its first to last value. This continues until the ID number
that refers to the last column and the last row is reached. You have the option to select
whether the columns or the rows will be incremented first.
A grid structure may associate one dimension of the grid with the physical parts
whose wall thickness is to be measured. The particular measurement points on each
part are then associated with the other dimension of the grid (see the example in
Figure 6-22 on page 100).
Software Options 99
DMTA-10022-01EN, Rev. B, October 2013
Measurement location
ID
Row
Column
Alternatively, the rows and columns of a grid may refer to a two-dimensional map of
measurement points on the surface of one part. In this case, a different grid is created
for each part (see the examples in Figure 6-23 on page 101).
100 Chapter 6
DMTA-10022-01EN, Rev. B, October 2013
Name: Elbow
Rows: 01 through 10
Columns: A through E
IDs: Elbow/A0 through
Elbow/E10
Name: Tee
Rows: 1 through 4
Columns: 1 through 3
IDs: Tee/11 through Tee/34
Figure 6-24 Configuring the ID range for a 2-D grid data file type
The 45MG has the capability to add a row or a column, and to change the
incrementation direction after a grid file is created (see section “Editing a File” on
page 108 for details).
Elevation
The first dimension refers to the physical distance from the bottom of the boiler to
the top.
102 Chapter 6
DMTA-10022-01EN, Rev. B, October 2013
Tube number
The second dimension refers to the number of the specific boiler tube to be
inspected.
Custom points
The third dimension refers to the actual thickness reading location at the specified
elevation on the specified tube.
The three dimensions are combined in a single ID number to precisely identify the
exact location of each thickness reading. Table 10 on page 103 shows an example in
which incrementation for the custom points was chosen first, the tube number second,
and the elevation third.
Figure 6-25 Configuring the ID range for a boiler data file type
104 Chapter 6
DMTA-10022-01EN, Rev. B, October 2013
The default file data mode can be set to the most frequently used.
To open a file
1. Press [FILE], and then select OPEN.
2. In the OPEN screen (see Figure 6-27 on page 107):
a) In SORT BY, select the order in which the files appearing on-screen are sorted
(by NAME or by DATE CREATED).
b) In the list of files, select the file you want to open.
The descriptive header for the highlighted file name appears on the lower
section of the display.
c) Select OPEN to return to the measurement screen with the selected file set as
the active file, and the ID number set to the first ID number in the file.
106 Chapter 6
DMTA-10022-01EN, Rev. B, October 2013
The file copy function can only be used to copy an existing file in the internal memory
to the internal memory. Use the file import and export functions to copy data to and
from the internal memory and the external MicroSD card.
To copy a file
1. In the measurement screen, press [FILE], and then select COPY.
2. In the COPY screen (see Figure 6-28 on page 108):
a) In the list, select the source file.
b) In COPY NAME, enter the file name for the destination file.
c) Set COPY THICKNESS DATA? to YES when you also want to copy the
thickness readings from the original file into the new file.
d) Select COPY.
3. Open the newly created file in order to make it the active file (see section
“Opening a File” on page 106.)
• File name
• File description
• Inspector ID
108 Chapter 6
DMTA-10022-01EN, Rev. B, October 2013
• Location note
• Delete protection (on/off)
• End row, column, or point of a grid file
• Incrementing order of a grid file
The edit function does not allow you to edit the file type, and cannot be used to edit
individual measurement identifiers (ID), or actual thickness readings.
When scrolling through the file names, a descriptive header for the highlighted file
name appears on the lower section of the display. This information can assist in
selecting the proper file if you are uncertain of the exact file name.
3. For a grid file, select CONTINUE, and then perform the following actions in the
second page of the EDIT screen (see Figure 6-30 on page 110):
a) Increase the END COLUMN and END ROW values as needed. These values
cannot be decreased.
b) If needed, change the INC 1ST BY value.
c) Select UPDATE.
110 Chapter 6
DMTA-10022-01EN, Rev. B, October 2013
If you select multiple files for deletion, and some of those files are delete-protected,
the 45MG will only delete the files that are not delete-protected.
112 Chapter 6
DMTA-10022-01EN, Rev. B, October 2013
Using the internal memory reset or master reset erases all files, and the data contained
in those files. The deleted files and the data they contain cannot be recovered. The
datalogger will be completely empty after this procedure.
When the ID overwrite protection is enabled, a message appears (see Figure 6-35 on
page 115) on the help text bar when you perform a save to inform you that the existing
thickness readings/waveforms will be overwritten. Select YES to replace the previous
reading with the new one, or NO to leave the original value.
114 Chapter 6
DMTA-10022-01EN, Rev. B, October 2013
Figure 6-36 on page 116 shows an example of the ID review screen and describes its
contents. The area under the waveform is reserved for status flags describing the
displayed stored thickness values. The flags are the same single letter abbreviations
for status words transmitted by the gage using the send commands (see section
“Managing Communications and Data Transfer” on page 187).
Active ID
m: Minimum value
m: MIN / MAX M: Maximum value
116 Chapter 6
DMTA-10022-01EN, Rev. B, October 2013
• Change the active ID to quickly jump to an existing ID. This is useful when you
are using a large file, and it would take too long to locate the desired ID using the
arrow keys.
• Change the active ID to a new ID that does not yet exist in the file. This mode is
useful when you want to include additional measurement points in the active file.
Additional IDs can be added anywhere in the database (beginning, middle, and
end).
4. Press [2nd F], [FILE] (ID#) again, and then edit the ID value (see Figure 6-37 on
page 118).
5. Press [MEAS] to return to the measurement screen with the new active ID.
6. When the edited ID is not in the database, the help text bar message shown in
Figure 6-38 on page 118 appears. Select INSERT to insert the new ID in front of
the active ID.
OR
Select APPEND to add the new ID to the end of the file.
Figure 6-38 The message when the edited ID is not in the database
118 Chapter 6
DMTA-10022-01EN, Rev. B, October 2013
The easiest way to replace a thickness reading is to save a new measurement at the
desired ID in the measurement screen. If you do not want to save a measurement at a
specific ID, press [SAVE] in the measurement screen when you are not taking a
measurement. This stores an LOS condition and —-.—- at the specific ID number.
File summary
Shows basic statistics for the file (minimum thickness and location, maximum
thickness and location, and high and low alarm conditions along with the mean,
median, and standard deviation).
Minimum/Maximum summary
Shows a list of ID number locations that have the minimum and maximum
thickness in a file.
Alarm summary
Shows a list of all the ID number locations where a low alarm and high alarm
occurred.
File comparison
Allows you to select two files and compare them. The first file contains the
previous inspection data, and the second file contains the current inspection data.
The report indicates the maximum wall loss and any area of greater wall
thickness (growth), and their ID number locations.
Minimum Review
Allows you to select a file, and then review all the minimum thickness locations in
the file. In addition to verifying the thickness at all minimum locations, you can
replace these minimum thickness locations if necessary.
To generate a report
1. In the measurement screen, press [FILE], and then select REPORT.
2. In the submenu, select the desired type of report. If you select:
• FILE SUMMARY, go to step 3
• MIN/MAX SUMMARY, go to step 4
• FILE COMPARISON, go to step 5
• ALARM SUMMARY, go to step 6
• MIN REVIEW, go to step 7
3. In the FILE SUMMARY screen (see Figure 6-39 on page 121):
a) Select the file for which you want to create the report.
120 Chapter 6
DMTA-10022-01EN, Rev. B, October 2013
b) Select REPORT.
The FILE SUMMARY report result screen opens (see Figure 6-40 on
page 121).
c) Press [2nd F], [] or [2nd F], [] to move between the #MINS and #MAXS
lists.
d) Select CANCEL to return to the measurement screen, or NEW REPORT to
generate another report.
5. In the FILE COMPARISON screen (see Figure 6-42 on page 123):
a) In the upper list, select the reference file you want to use in the comparison.
b) In the lower list, select the comparison file (containing newer data for the
same measurement points).
c) Select REPORT.
The FILE COMPARISON report result screen opens with the first maximum
wall-loss ID highlighted (see Figure 6-43 on page 123).
122 Chapter 6
DMTA-10022-01EN, Rev. B, October 2013
d) Review the maximum wall-loss location list, and the maximum wall growth
location list.
e) Select CANCEL to return to the measurement screen, or NEW REPORT to
generate another report.
6. In the ALARM SUMMARY screen:
a) Select the file for which you want to generate the report.
b) Select REPORT.
The report page of the ALARM SUMMARY screen opens with the first low
alarm location ID highlighted (see Figure 6-44 on page 124).
124 Chapter 6
DMTA-10022-01EN, Rev. B, October 2013
d) Couple the probe back to the minimum ID location to verify the thickness,
and then press [SAVE] to store the new measurement.
e) Use the [] and [] keys to move to other minimum review list IDs.
Press [MEAS] to exit the minimum review.
126 Chapter 6
DMTA-10022-01EN, Rev. B, October 2013
This chapter describes how to use special 45MG functions and modes. The 45MG has
many thickness-measurement features. Although the features outlined in this section
are not required for basic thickness operation, they can make the gage a more versatile
instrument.
• Activating and Configuring a Differential Mode (see section 7.1 on page 127).
• Using the Minimum, Maximum, or Min/Max Thickness Mode (see section 7.2 on
page 129).
• Preventing False Minimum/Maximum Thickness Readings (see section 7.3 on
page 131).
• Using Alarms (see section 7.4 on page 132).
• Locking the Instrument (see section 7.5 on page 136).
• Freezing the Measurement or Optional Waveform (see section 7.6 on page 138).
The 45MG includes differential modes that can be used to easily compare the actual
measurement with an entered reference value. The actual thickness measurement
appears on the thickness display, and the differential value appears in the differential
display area (see Figure 7-1 on page 128).
Actual thickness
Differential value
Figure 7-1 Normal differential mode (shown with Waveform option activated)
The units and resolution of the differential thickness are the same as those selected for
the thickness measurement.
When you press [SAVE] (with the optional datalogger) while in the NORMAL or
% RATIO differential modes, the 45MG saves the actual thickness value along with
the “D” flag indicating that the Differential mode is active.
• % RATIO: Shows the actual thickness along with the percent difference
from the REF VALUE entered.
128 Chapter 7
DMTA-10022-01EN, Rev. B, October 2013
minimum or maximum thickness value replaces the main thickness display when the
transducer is uncoupled, or when a loss of signal (LOS) occurs. The replacement value
appears in an outlined font.
Figure 7-3 Displaying the minimum and/or maximum thickness (shown with the
Waveform option activated)
130 Chapter 7
DMTA-10022-01EN, Rev. B, October 2013
The fastest display update rate is automatically activated when entering the minimum
or maximum thickness mode. When leaving the mode, the display update rate is
restored to its previous state.
The minimum and maximum thickness modes display the smallest and largest
thickness value measured from the time you activate or reset the minimum mode. The
modes are useful when determining the thinnest/thickest reading obtained while
making a series of readings on a test piece.
A false minimum or maximum reading can occur when you lift the transducer from
the test piece. This is due to excess couplant (particularly on smooth surfaces), which
causes the instrument to read the thickness of a couplant drop when you lift the
transducer.
Any one of the 45MG alarm modes can be activated to help you identify when the
actual thickness measurement is above or below editable reference values.
• The HIGH or LOW alarm indicator flashes with a red background at the bottom
right corner of the measurement screen (see Figure 7-4 on page 132).
• The thickness value appears in red.
• When the beeper is active (see section “Setting the User Interface Language and
Other System Options” on page 41), the 45MG emits a long beep.
The thickness value and alarm indicator only appear in color when the indoor color
scheme is active (see section “About Color Schemes” on page 45 to change the color
scheme).
Previous thickness
Thickness reading
(turns red during an
alarm condition)
High alarm indicator
Figure 7-4 Example of a HIGH alarm indicator (shown with the Waveform option
activated)
132 Chapter 7
DMTA-10022-01EN, Rev. B, October 2013
When the Datalogger option has been purchased, the datalogger records an alarm
condition in the second status box for all stored measurements. An A indicates the
alarm mode, an L indicates a low alarm condition, and an H indicates a high alarm
condition.
There are three different alarm modes (STANDARD, B-SCAN, and REDUCTION
RT):
STANDARD
The standard alarm warns you when the actual measured thickness is below a
low reference value, or above a high reference value. The reference values are
thickness set points using the current instrument units and resolution.
The B-SCAN and REDUCTION RT alarms are only available when the B-scan or
reduction rate (DIFF mode) are activated. These functions must be turned on first
before the B-SCAN and REDUCTION RT alarms can be activated.
B-SCAN
The B-scan alarm mode is similar to the standard alarm mode, except that in the
B-scan alarm mode, lines are shown in the B-scan grid when the reference values
fall within the B-scan thickness range (see Figure 7-5 on page 133). In addition, the
alarms are operational while you review B-scan thicknesses in the B-scan freeze
review mode.
The thickness value and the alarm indicator only appear in color when the indoor
color scheme is active (see section “About Color Schemes” on page 45 to change the
color scheme).
REDUCTION RT
The REDUCTION RT option only appears when the active file is configured with
FILE DATA MODE set to REDUCTION RT. You set the reduction percentage for
the low (YELLOW ALARM) and high (RED ALARM) limits, following which the
instrument displays (see Figure 7-6 on page 134):
• The RED indicator for reduction rates that are greater than or equal to the
RED ALARM value.
• The YEL indicator for reduction rates that are between the YELLOW ALARM
and the RED ALARM values.
• The GRN indicator for reduction rates that are below the YELLOW ALARM
value.
Figure 7-6 YEL (yellow) and RED alarm indicators (shown with the Waveform
option activated)
134 Chapter 7
DMTA-10022-01EN, Rev. B, October 2013
The B-SCAN option only appears when the B-scan mode is active (see section “Using
the B-Scan” on page 156).
The REDUCTION RT option only appears when the active file is configured with the
FILE DATA MODE parameter set to REDUCTION RT (see section “About File Data
Modes” on page 104).
3. When ALARM MODE is set to STANDARD or B-SCAN, set the LOW ALARM
and the HIGH ALARM values.
OR
When ALARM MODE is set to REDUCTION RT, set the YELLOW ALARM and
the RED ALARM values.
4. Press [MEAS] to return to the measurement screen.
Alarm reference values that were entered in one unit system are displayed as the
equivalent value when the alternate units are selected.
The 45MG is equipped with an instrument lock that can be used by a supervisor to
restrict access to selected functions. The supervisor can also enter a password to
prevent other users from unlocking the functions. Once a password has been set, you
must reenter the password before you can lock or unlock any function.
When locking [CAL VEL] and [CAL ZERO], the function ([2nd F], [CAL ZERO]
(Do ZERO)) is still available.
Anytime a user attempts to use a locked function, a message appears in the help bar to
indicate that the function is locked (see Figure 7-8 on page 137).
136 Chapter 7
DMTA-10022-01EN, Rev. B, October 2013
If you forget your password, you can unlock the instrument and deactivate the
password by entering the master password: “OLY45MG”.
To change the password, you must first use the master password to deactivate the
password, and then set a new password.
3. Select DONE to activate the instrument lock and return to the measurement
screen.
Pressing [FREEZE] stops updating the display and optional waveform, and keeps the
waveform and thickness on the display even if you move or uncouple the transducer.
The freeze indicator ( ) appears on the lower right side of the display when the
freeze function is active.
The freeze function is convenient for easily setting the gain parameters, reviewing a
B-scan, or performing a high-temperature measurement without having to keep the
transducer coupled to the test piece.
In order to prevent recording wrong minimum or maximum values, you may also
want to use the freeze function to pause measurements before uncoupling the
transducer from the test piece.
138 Chapter 7
DMTA-10022-01EN, Rev. B, October 2013
Pressing [MEAS] or [SAVE] (when the Datalogger option has been activated) also
unfreezes the display.
140 Chapter 7
DMTA-10022-01EN, Rev. B, October 2013
The MEAS setup is the most commonly used setup menu screen, and is used to access
global parameters concerning the instrument measurement features.
In the MEAS screen, some parameters differ depending on whether a single or dual
element transducer is connected (see Figure 8-1 on page 142).
2. In the MEAS screen for single element transducers, in the MEAS MODE text box,
select the instrument measures and displays from the following options:
• THICKNESS: The thickness of the inspected part.
• VELOCIMETER: The sound velocity in the material of the inspected part.
• TIME OF FLIGHT: The round trip time of flight (TOF) of the sound in the
inspected part
OR
In the MEAS screen for dual element transducers, in the MEAS OPTION text
box, select the measurement method to be used by the instrument from the
following options:
• STANDARD: Measure to the first back-wall echo.
• AUTO ECHO-TO-ECHO (optional): Measure between the successive back-
wall echoes.
• MANUAL ECHO-TO-ECHO (optional with the Waveform): Measure
between the multiple back-wall echoes with the manual control.
3. In the UNIT TYPE text box, select between the INCH (English) or MILLIMETER
(metric) units.
Time-of-flight measurements are always expressed in microseconds.
4. In the RESOLUTION text box, select between STANDARD, LOW, or HIGH (see
section “Changing the Thickness Resolution” on page 49 for details).
5. In the MIN/MAX text box, select either MIN (minimum), MAX (maximum), or
BOTH to activate a mode (see section “Using the Minimum, Maximum, or
142 Chapter 8
DMTA-10022-01EN, Rev. B, October 2013
The MIN/MAX and HOLD BLANK functions are mutually exclusive. You must set
MIN/MAX to OFF in order to change the HOLD BLANK function. Similarly, you
must set HOLD BLANK to BLANK in order to change the MIN/MAX function.
7. In MEASURE RATE, adjust the measurement update rate (see section “Adjusting
the Measurement Update Rate” on page 47 for details).
8. For single element transducers only, set AVERAGING to OFF in order to turn
thickness averaging off, set AVERAGING to ON to perform a running average of
the last five thickness readings, or set AVERAGING to On-QBar to display a
quality measurement Q-bar below the measurement screen, indicating the
stability of the averaged reading.
9. Set ID OVERWRITE PROT to ON (with the Datalogger option only) if you want
to see a confirmation message in the help bar when attempting to save a
measurement reading in an ID that already contains a value (see section “Setting
the ID Overwrite Protection” on page 114 for details).
10. For single element transducers only, set QUICK SETUP RECALL to ON in order
to activate the quick recall of the first four custom setups by using a combination
of the [2nd F] and arrow keys (see section “Quickly Recalling a Custom Setup for
Single Element Transducers” on page 184 for details).
11. For single element transducers only, set AGC to ON to configure the automatic
gain control (AGC) function to automatically bring all measured back-wall echoes
to the same amplitude.
The AGC function works well for most standard thickness gage applications, and is
turned on by default. In some thickness applications, the receiver gain is set at or near
its maximum value. In such cases, turn off the AGC function to prevent reading
instability.
The SYSTEM screen, allows you to configure many 45MG system parameters.
2. Set BEEPER to ON or OFF (see section “Setting the User Interface Language and
Other System Options” on page 41 for details).
3. Set INACTIVE TIME to ON or OFF (see section “Setting the User Interface
Language and Other System Options” on page 41 for details).
4. In the LANGUAGE text box, select the desired user interface language (see
section “Setting the User Interface Language and Other System Options” on
page 41 for details).
5. Set RADIX TYPE to the desired character (period or comma) for separation of the
integer and decimal numbers.
6. Set SAVE KEY DATA to save only the thickness measurement (THICKNESS), or
both the thickness and the waveform (THICKNESS + WF).
144 Chapter 8
DMTA-10022-01EN, Rev. B, October 2013
The calibration and setup parameters are also always saved/sent along with the
thickness measurement.
7. Set DEFAULT FILE MODE to the desired default file data mode when you create
a file (see section “About File Data Modes” on page 104 for details).
8. Set TEXT EDIT MODE to the appropriate option for entering alphanumeric
values. The two available options are the virtual keyboard (VIRTUAL) and the
legacy character wheel selection (TRADITIONAL) [see section “Selecting the
Text Edit Mode” on page 37 for details].
9. Set SCR TO SD CARD to ON to enable the 45MG to create a BMP image file on
the external MicroSD card for the actual screenshot when you press [2nd F],
[SETUP] (see section “Sending a Screen Capture to the External MicroSD Card”
on page 195 for details).
10. Press [MEAS] to return to the measurement screen.
The 45MG features a USB port that can be used to connect the instrument to a
computer. When connected to a computer, the 45MG can send and receive data, or be
remotely controlled by the computer (when the Datalogger option has been
activated).The remote command document and the FTP (file transfer protocol)
document are available upon request.
146 Chapter 8
DMTA-10022-01EN, Rev. B, October 2013
148 Chapter 8
DMTA-10022-01EN, Rev. B, October 2013
• Adjusting the Gain with Dual Element Transducers (see section 9.1 on page 149)
• Adjusting the Extended Blank with Dual Element Transducers (see section 9.2 on
page 151)
• About the B-Scan (see section 9.3 on page 153)
• About the DB Grid (see section 9.4 on page 159)
With D79X series dual element transducers, you can manually adjust the gain by
pressing [GAIN/WAVE ADJ]. The 45MG offers two types of gain adjustments:
• The standard feature allows the user to set the gain to HIGH (+10 dB),
standard (default), and LOW (−6 dB).
• When the waveform (option) is activated, the user can adjust the gain in 1 dB
increments.
The gain value is expressed in decibels (dB), and appears near the lower-right corner
of the display (see Figure 9-1 on page 150).
When activated, the manual gain adjustment also modifies the way in which echoes
are shown on the waveform display (optional). With the default automatic gain, the
measured echo peak is visible on the display, enabling you to readily observe the
measured echo position, independent of its strength or the gain. When you manually
adjust the gain, the displayed echo height changes proportionally to the actual echo
amplitude at the receiver output, enabling you to readily observe gain changes.
1. Press [GAIN/WAVE ADJ] to change the gain from default gain to HIGH (+10 dB).
2. Press [GAIN/WAVE ADJ] again to change the gain to LOW (−6 dB).
3. Press [GAIN/WAVE ADJ] a third time to change the gain back to default gain (no
gain showing).
When the optional waveform is activated:
150 Chapter 9
DMTA-10022-01EN, Rev. B, October 2013
The gain value appears near the lower-right corner of the display. The waveform
display changes to the amplitude proportional to the gain mode.
2. Press [] or [] to adjust the gain in ±1 dB steps.
The gain value and the height of the echoes change accordingly.
• Increase the gain value when the peak amplitude of the echo that should be
detected is below 20 % of the waveform display height.
This prevents the gage from reading too high a value when skipping one
back-wall echo, and from measuring twice the correct thickness (doubling).
• Reduce the gain value when noise peaks are above 20 % of the waveform
display height.
This prevents the gage from detecting a noise peak rather than the correct
back-wall echo.
3. Press [2nd F], [GAIN/WAVE ADJ] again to restore the default gain value.
4. Press [GAIN/WAVE ADJ] again to return to the automatically adjusted echo
height mode.
The default automatic gain mode is indicated by a blank Gain field.
Olympus recommends using the extended blank parameter only if you are an
experienced operator who thoroughly understands the acoustic properties of the
material being measured. Incorrect use of the extended blank can cause the gage to
misread areas of thin material.
Normally, the 45MG searches for echoes down to nearly zero thickness. However,
some special circumstances, such as a high degree of near-surface corrosion,
aluminum material, enclosed flaws, or laminations, can generate echoes that the
instrument may falsely detect as the low thickness. When these echoes are larger than
the sought after back-wall echo, the manual gain adjustment (see section “Adjusting
the Gain with Dual Element Transducers” on page 149) cannot prevent this false
detection. However, the extended blank parameter allows you to define an early
period, during which the 45MG will not perform echo detection, thereby preventing
erroneous measurements.
Waveform adjustment
parameter and value
4. To turn off the extended blank, use the [] key to bring the extended blank value
back down to zero.
152 Chapter 9
DMTA-10022-01EN, Rev. B, October 2013
If the measurement point changes when the extended blank is moved, echoes can
change in height. This is because in the normal waveform display mode, the 45MG
attempts to adjust the height.
The instrument also attempts to make the most accurate measurement by identifying
the beginning of an echo. When the extended blank is positioned within an echo
instead of to its left, the gage is unable to make an accurate detection.
A B-scan is a cross-sectional image of thickness readings. The 45MG can acquire and
display B-scan data (see Figure 9-3 on page 153). When you activate the B-scan, the
thickness reading profile builds up and scrolls on the screen. Once a B-scan is
acquired, you can freeze the screen and review the recorded thickness values.
B-scan
While the B-scan is running, you can save individual thickness readings, the current
B-scan screen (with all thickness values), or the entire scan (up to 10000 readings) [see
section “Saving B-Scans or Thickness Readings (Optional Datalogger)” on page 157].
The B-scan can be actively configured from the B-SCAN screen (see Figure 9-4 on
page 154), which is accessible by pressing [SETUP], and then selecting B-SCAN in the
menu.
DIRECTION:
Choose the B-scan direction to match the direction of transducer movement. A
scan direction arrow appears below the left corner of the B-scan display to
indicate the transducer scan direction (see Figure 9-5 on page 155). The data starts
appearing on-screen in the opposite direction.
LEFT TO RIGHT
The transducer scans the part from left to right, and the data appears on the
right of the screen, scrolling right to left.
RIGHT TO LEFT
The transducer scans the part from right to left, and the data appears on the
left of the screen, scrolling left to right.
154 Chapter 9
DMTA-10022-01EN, Rev. B, October 2013
Loss-of-signal (LOS)
indicator
Scan direction
(left to right) Min/Max marker B-scan range
LOS MODE
Determines how the B-scan behaves when a loss of signal (LOS) occurs.
STOP ON LOS
The B-scan stops scrolling when an LOS occurs. When the instrument regains
the measurements, a thin blank vertical line is inserted into the B-scan to
indicate that an LOS has occurred (see Figure 9-5 on page 155).
CONTINUE ON LOS
The B-scan continues scrolling when an LOS occurs.
FREEZE MODE
Determines which thickness readings are displayed when [FREEZE] is pressed
while a B-scan is active.
SHOW MINIMUM
Displays the thickness reading for the minimum reading collected during the
scan.
SHOW MAXIMUM
Displays the thickness reading for the maximum reading collected during the
scan.
SHOW CURRENT
Displays the last thickness reading made before pressing [FREEZE].
FREEZE REVIEW
When this feature is enabled and a B-scan is active, pressing [FREEZE] freezes the
B-scan image in review mode. In this mode, a vertical line (review marker)
appears, indicating the location of the displayed thickness (see Figure 9-6 on
page 156). The displayed thickness is either the minimum, maximum, or current
thickness, depending on which B-SCAN FREEZE MODE option is selected. Use
the [] and [] keys to move the review marker and read the thickness at the
review marker location.
If the minimum or maximum value moves off the B-scan screen, press [FREEZE] to
center the B-scan and the review marker on the minimum or maximum thickness.
Review marker
Minimum location
marker
156 Chapter 9
DMTA-10022-01EN, Rev. B, October 2013
The B-scan begins drawing the cross section of the material as soon as the first
non-LOS reading is collected.
4. To stop the collection of B-scan data:
Press [FREEZE].
OR
With B-SCAN LOS MODE set to STOP ON LOS, simply uncouple the
transducer from the test material.
5. While the B-scan is frozen, if B-SCAN FREEZE REVIEW is set to ON, use the []
and [] keys to move the review marker, and then read the thickness value
corresponding to the marker position.
6. Press [MEAS] again to reset the B-scan and begin a new B-scan.
7. Refer to section “Saving B-Scans or Thickness Readings (Optional Datalogger)”
on page 157 for details on how to save B-scan data.
158 Chapter 9
DMTA-10022-01EN, Rev. B, October 2013
When a B-scan screen is saved to the datalogger, the gage saves the thickness values
for the data points appearing on the display. All thickness values on a saved B-scan
can be reviewed during ID Review. Recall the saved B-scan, and use the [] and []
keys to review each thickness reading.
You can save a maximum of 10000 thickness readings to a B-scan. When the
maximum number of thickness points reaches 10000, the gage prompts you to save
the B-scan history, or to reset the B-scan without saving.
The database grid (DB grid) is a table representation of 2-D data. This representation
enables you to move freely in any direction on a grid, rather than follow a preset list of
IDs. Instead of automatically incrementing to the next ID location, you can use the
arrow keys to move to a location that is more convenient. You can simultaneously
display the A-scan, the DB grid, and the thickness reading (see Figure 9-7 on
page 160). You can configure the grid cells to show a data cell flag and a background
color corresponding to the range in which their thickness reading falls.
Half-size DB grid
160 Chapter 9
DMTA-10022-01EN, Rev. B, October 2013
The ID number increments in the order assigned when the file was initially set up
independent of the TRANSPOSE GRID value.
5. Set LINEARIZE GRID to ON to display the grid IDs in the linear form (see
Figure 9-10 on page 161).
6. Set DATA CELL FLAG to one of the available options in order to display a single
data flag with each data cell in the DB grid. The data cell flag is a letter that
appears in a small box to the right of the thickness value in the data cell (see
Figure 9-7 on page 160). The available options are:
NONE
No data cell flag appears.
MIN/MAX
“m” indicates a minimum thickness.
“M” indicates a maximum thickness.
ALARM
“L” indicates any kind of low alarm condition, including a standard low
alarm condition.
“H” indicates any high alarm condition.
7. Set GRID COLOR OPTION to ON to activate usage of low, mid, and high-range
colors for the grid cell background.
8. Set LO RANGE COLOR to the desired cell background color (RED, YELLOW, or
GREEN) when the cell thickness value is smaller than the LO RANGE VALUE.
9. Set MID RANGE COLOR to the desired cell background color (RED, YELLOW,
or GREEN) when the cell thickness value is between the LO RANGE VALUE and
the HI RANGE VALUE.
10. Set HI RANGE COLOR to the desired cell background color (RED, YELLOW, or
GREEN) when the cell thickness value is higher than the HI RANGE VALUE.
162 Chapter 9
DMTA-10022-01EN, Rev. B, October 2013
Highlighted cell
4. Press [MEAS] to return to the measurement screen with the current ID number
changed to the ID location selected in the ID review screen.
Inserted cell
164 Chapter 9
DMTA-10022-01EN, Rev. B, October 2013
Once the Single Element High Resolution option has been activated, the 45MG
includes predefined setups for standard single element transducers. In some cases, a
45MG ships from the factory preprogrammed with one or more custom setups to
meet special customer requirements. You can create your own custom setups to meet
the need of a particular single element transducer, or for a particular application.
Predefined and custom setups allow you to quickly change between transducer and
application setups by selecting a previously saved setup.
• Creating a Custom Setup for a Single Element Transducer (see section 10.1 on
page 166).
• Quickly Adjusting Waveform Parameters For Single Element Transducers (see
section 10.2 on page 168).
• About the Detection Modes (see section 10.3 on page 170).
• About the First Peak (see section 10.4 on page 172).
• About the Pulser Power (see section 10.5 on page 173).
• About the Time-Dependent Gain Curve (see section 10.6 on page 174).
• About the Main Bang Blank (see section 10.7 on page 176).
• About the Echo Window (see section 10.8 on page 178).
• Saving Setup Parameters (see section 10.9 on page 183).
• Quickly Recalling a Custom Setup for Single Element Transducers (see
section 10.10 on page 184).
Many of the adjustments that can be made in a setup are interactive. All adjustments
have an effect on the measurement range and/or measurement accuracy of the 45MG.
In most cases, do not attempt adjustments without monitoring the waveform.
Additionally, when establishing a custom setup for a specific application, it is
essential to verify the performance on reference standards representing the materials
and thickness range to be measured.
To reduce the number of changes needing to be made to parameter values, you can
also select an existing single element setup with parameter values that are close to the
desired setup.
166 Chapter 10
DMTA-10022-01EN, Rev. B, October 2013
a) Set DET MODE to the desired detection mode (see section “About the
Detection Modes” on page 170 for details).
Figure 10-1 The ACTIVE screen for the setup of a single element transducer
b) Enter a SETUP NAME that describes the transducer and/or the application to
be used to create the setup.
c) Set MEAS TYPE to the desired measurement type. The available choices are:
• STANDARD: For normal mode 1, 2, or 3 positive or negative peak
measurement.
• FIRST PEAK: For detection of the first of several peaks of similar
amplitudes (see section “About the First Peak” on page 172 for details).
d) Set PROBE TYPE to indicate the transducer type to be used with this setup.
The selected probe type must match the frequency of the transducer being
used for proper pulser/receiver performance.
e) Set VELOCITY to the value of the sound velocity in the material that is to be
tested with this setup (see section “About the Material Sound Velocity and the
Zero Calibrations” on page 59).
f) Set ZERO VALUE to the calibrated zero-offset value (time of flight of the echo
that is not traveling through the material) for this transducer (see section
“Calibrating the Instrument” on page 55).
g) If needed, increase the PULSER POWER to increase the ultrasound wave
penetration in the material. Reduce the value for better near-surface
resolution (see section “About the Pulser Power” on page 173 for details).
h) Set MAX GAIN to the desired maximum gain value (see section “About the
Maximum Gain” on page 175 for details).
i) Set INIT GAIN to the desired initial gain value (see section “About the Initial
Gain” on page 175 for details).
j) Set TDG SLOPE to the desired time-dependent gain slope value (see section
“About the TDG Slope” on page 176 for details).
k) Set MB BLANK to the desired main bang blank time interval (see section
“About the Main Bang Blank” on page 176 for details).
l) Set ECHO WINDOW to the desired time interval (see section “About the
Echo Window” on page 178 for details).
m) Set ECHO 1 DETECT to –SLOPE to detect the negative peak of the first echo,
or to +SLOPE to detect the positive peak of the first echo (see section “About
the Detection of Echo 1 and Echo 2” on page 179 for details).
6. Press [SAVE].
7. In the SAVE SETUP screen:
a) If needed, in the SAVE AS dialog box, edit the setup name.
b) In the SAVE TO list, select the desired custom setup location in which you
want to save the setup.
c) Select SAVE.
8. Press [MEAS] to return to the measurement screen.
The saved setup becomes the active setup.
With single element transducers, you can quickly adjust individual waveform
parameters using the [GAIN/WAVE ADJ] key.
168 Chapter 10
DMTA-10022-01EN, Rev. B, October 2013
Waveform parameter
Waveform parameter value
3. Use the [] or [] key to select the parameter that you want to adjust. The choices
are as follows:
• DET MODE (see section “About the Detection Modes” on page 170 for
details)
• M3 BLANK in mode 3 only (see section “About the Mode 3 Echo Blank” on
page 182 for details)
• IF BLANK in modes 2 and 3 only (see section “About the Interface Blank” on
page 180 for details)
• ECHO 2 DETECT in modes 2 and 3 only (see section “About the Detection of
Echo 1 and Echo 2” on page 179 for details)
• ECHO 1 DETECT (see section “About the Detection of Echo 1 and Echo 2” on
page 179 for details)
• ECHO WINDOW (see section “About the Detection of Echo 1 and Echo 2” on
page 179 for details)
• MB BLANK (see section “About the Main Bang Blank” on page 176 for
details)
• TDG SLOPE (see section “About the TDG Slope” on page 176 for details)
• INIT GAIN (see section “About the Initial Gain” on page 175 for details)
• MAX GAIN (see section “About the Maximum Gain” on page 175 for details)
• PULSER POWER (see section “About the Pulser Power” on page 173 for
details)
• PROBE TYPE
• MEAS TYPE (see step 5.c in section “Creating a Custom Setup for a Single
Element Transducer” on page 166 for details)
4. Use the [] or [] keys to select the value for the selected parameter.
5. Repeat steps 3 and 4 to adjust other parameters.
6. Press [GAIN/WAVE ADJ] again to hide the waveform adjustment parameter.
There are three detection modes (Mode 1, Mode 2, and Mode 3):
Mode 1
Uses a direct contact transducer to measure the time of flight between the main
bang and the first back-wall echo (see Figure 10-3 on page 170).
Main bang
Detection marker
Mode indicator
Mode 2
Measures the time of flight between the interface (or delay line) echo and first
back-wall echo using a delay line or an immersion transducer (see Figure 10-4 on
page 171).
170 Chapter 10
DMTA-10022-01EN, Rev. B, October 2013
Interface echo
Main bang
Detection bar
Mode indicator
Mode 3
Uses a delay line or an immersion transducer to measure the time of flight
between one back-wall echo and the next back-wall echo (see Figure 10-5 on
page 171).
Second back-wall
echo
Interface echo
Detection bar
Mode indicator
Refer to section “About the Echo Window” on page 178 for information on the modes
in relation with the echo window.
With single element transducers, the 45MG normally detects peaks on either the
highest positive or highest negative peak of the RF waveform. This feature works well
for most precision thickness applications.
This normal peak detection can be unstable for applications in which the back-wall
echo is irregular and contains several negative or positive peaks that are close in
amplitude. In such cases, the thickness reading can fluctuate as the gage switches
detection from one peak to another. This can happen, for example, in applications
measuring the length of bolts, or the thickness of the gel coat over fiberglass (see
Figure 10-6 on page 172). In such cases, help stabilize the echo detection and thickness
measurements by selecting the first peak algorithm to detect the first of several peaks
of similar amplitudes (see step 5.c in section “Creating a Custom Setup for a Single
Element Transducer” on page 166).
172 Chapter 10
DMTA-10022-01EN, Rev. B, October 2013
The excitation pulse (main bang) voltage can be set to one of the following values:
60 V, 110 V, 150 V, and 200 V.
Higher voltages may provide greater penetration at the expense of a lower near-
surface resolution, especially in mode 1. Inversely, lower voltages may provide better
near-surface resolution at the expense of a lower penetration.
For most applications, the 110 V value provides the best signal-to-noise ratio for
returning echoes. The pulser power indicates the voltage used to excite the
transducer, and therefore affects the size of the initial pulse (see Figure 10-7 on
page 173), and the amount of energy transmitted into the material.
When the 45MG displays the SAT flag below the thickness reading, the input voltage
from the transducer is above the maximum range, and proper measurements cannot
be made. This can normally be corrected by lowering the PULSER POWER until the
SAT flag no longer appears.
With single element transducers, the 45MG either uses the automatic gain control
(AGC) [see section “Configuring Measurement Parameters” on page 141 for details on
how to activate the AGC] or the time-dependent gain (TDG) functions to
automatically adjust receiver gain to an optimum level when an echo is detected.
The 45MG offers three parameters for drawing a time-dependent gain curve: INIT
GAIN, TDG SLOPE, and MAX GAIN (see Figure 10-8 on page 174). From the initial
gain level, receiver gain slopes up to the maximum gain level at the rate determined
by the TDG SLOPE setting. When you adjust any of the received gain parameters, a
black time-dependent gain curve is displayed that clearly indicates the zones of initial
gain, slope, and maximum gain.
MB BLANK
Time-dependent gain
curve
The TDG curve can be used to optimize near-surface resolution while providing a
higher maximum gain for thicker samples. You can also use the TDG curve in
measurements of highly scattering materials, such as cast metals and fiberglass, in
order to minimize detection of scattered echoes occurring ahead of the back-wall
echo.
174 Chapter 10
DMTA-10022-01EN, Rev. B, October 2013
The maximum available receiver gain can be adjusted from 0.0 dB to 99.0 dB. When
no echo is detected (LOS prompt), the gain rises to the maximum level set by the
initial gain, slope, and maximum gain. When the maximum gain is set too high, the
gage may hang up on transducer noise or other spurious signals; when it is set too
low, returning echoes may not be high enough for detection.
The max gain can never be lower than the initial gain, and has a maximum value of
99.0 dB.
The INIT GAIN setting is most critical in applications where minimum thickness
measurements must be optimized. This setting should always be set with the aid of a
reference standard representing that minimum. In cases where the minimum
thickness capability is less important than penetration, and scatter echoes are not a
problem, the initial gain can be set to be equal to the maximum gain.
The main bang blank is effectively a blank zone that protects the receiver from false
readings generated by the main bang. This blank or dead zone (up to 18 microseconds
from the excitation pulse) prevents the trailing edge echoes of the excitation pulse
from being detected as back-wall or interface echoes. The end of the main bang blank
indicates the point in time at which the gage begins to search for echoes.
In general, set the main bang blank just beyond the point at which the gage hangs up,
and then proceed to test with the transducer both coupled to and uncoupled from the
test material in order to ensure accurate measurements.
In mode 1, however, the length of the main bang blank determines the minimum
thickness that can be measured, and must therefore be positioned with care after
selecting the initial gain level (see Figure 10-9 on page 177). When the main bang
blank is too short, the instrument hangs up on the excitation pulse and readings are
not possible. When the main bang blank is too long, the minimum measurable
thickness is unnecessarily restricted. When using immersion transducers, ensure that
the main bang blank is always set before the interface echo from the shortest water
path.
176 Chapter 10
DMTA-10022-01EN, Rev. B, October 2013
Initial pulse
In mode 2 and mode 3, the setting of the main bang blank is not critical if set to some
point between the end of the excitation pulse and the interface echo (see Figure 10-10
on page 177).
Interface echo
Figure 10-10 Main bang blank position for mode 2 and mode 3
The echo window is the time interval after each main bang, during which the
instrument is enabled to detect echoes. The echo window interval begins at the end of
the main bang blank. The echo window ends at x µs following the main bang in
mode 1, or at x µs following the interface blank in mode 2 and mode 3.
In mode 1, you can usually set the echo window to any value greater than the round-
trip pulse transit time in the thickest or slowest piece of material to be measured (see
Figure 10-11 on page 178). The exact setting is not critical as long as it is long enough
to include the farthest echo of interest.
Echo window
In mode 2 and mode 3, the echo window is limited to the time interval between
successive interface echoes (see Figure 10-12 on page 179). The end of the echo
window must be set ahead of the second interface echo in order to prevent an
incorrect detection, which in turn determines a maximum measurable thickness. In
applications involving mode 2 and mode 3 immersion measurements, the echo
window placement must apply throughout the range of the water paths to be used.
178 Chapter 10
DMTA-10022-01EN, Rev. B, October 2013
Echo window
Interface echo
Interface blank
Refer to Table 12 on page 180 for assistance in selecting the appropriate detection
polarity for a given application.
180 Chapter 10
DMTA-10022-01EN, Rev. B, October 2013
In mode 2, the interface blank prevents the detection of trailing lobes or cycles of the
interface echo which might otherwise be detected as back-wall echoes, which results
in a hang-up condition (see Figure 10-14 on page 181). The interface blank should be
set as short as possible to avoid unnecessarily restricting minimum measurable
thicknesses. The initial gain parameter often helps reduce the interface echo
amplitude, and permits the use of a shorter interface blank. Check the interface blank
settings with the transducer both coupled to and uncoupled from the test material.
In mode 3, the interface blank selects which pair of back-wall echoes are measured
(see Figure 10-15 on page 182). In most conditions, the interface blank would be set
just short of the first back-wall echo. However, as a practical matter, the first back-wall
echo from thin materials is often distorted or lost in the interface echo. With some
challenging geometries (such as tight radii), later pairs of back-wall echoes may be
cleaner than early ones. In such cases, set the interface blank to a length that ensures
detection of a clean and well-defined pair of back-wall echoes, even if the echoes are
not the first two.
Gage reading between back- Interface blank adjusted so the gage measures
wall echoes 1 and 2 between back-wall echoes 2 and 3
182 Chapter 10
DMTA-10022-01EN, Rev. B, October 2013
Because the mode 3 echo blank limits minimum measurable thicknesses, set the
controls to as short a period as possible; typically no longer than a few hundred
nanoseconds. Special cases may be an exception where mode conversion effects on
curved samples cause significant spurious signals between legitimate echo peaks. In
such cases, set the mode 3 echo blank to as long a period as necessary in order to
prevent detection of the spurious signals.
After adjusting selected waveform parameters, it is possible to store the settings for
quick and easy recall. The 45MG can store up to 35 custom setups in its internal
memory.
Setups named SE-USER-1 through SE-USER-35 are empty locations. Use caution
when selecting a location that already contains a custom setup, because the new setup
will replace the old information.
Normally, you can change the setup for a custom application by pressing
[RECALL XDCR] to select the appropriate setup in the list of available setups), and
then pressing [MEAS]. This process is adequate for typical setups that are changed
infrequently. However, when it is necessary to change between two or more custom
setups on a more frequent basis, use the quick setup recall function instead.
When the quick setup recall function is activated, a simple keypad shortcut can be
used to quickly switch to any of the first four custom setups for single element
transducers.
184 Chapter 10
DMTA-10022-01EN, Rev. B, October 2013
This feature only works when a single element transducer is plugged into the 45MG,
and the single element transducer option has also been purchased.
186 Chapter 10
DMTA-10022-01EN, Rev. B, October 2013
This section describes the process through which the 45MG communicates with a
computer to send, receive, import, and export files. The 45MG comes standard with a
USB cable for communication using the USB 2.0 protocol.
GageView is compatible with Windows XP, Windows Vista, and Windows 7. For
more details, refer to the GageView Interface Program — User’s Manual (P/N: 910-259-
EN [U8778347]).
For more information on installing GageView, refer to the GageView Interface Program
— User’s Manual (P/N: 910-259-EN [U8778347]).
2. If you use a communication program other than GageView, refer to the program
documentation for proper configuration of the program for USB communication.
3. Turn the 45MG on.
4. Connect one end of a USB cable to the USB client connector on the top of the
45MG, and connect the other end to a USB port of the computer (see Figure 11-1
on page 188).
USB cable
The first time the 45MG is connected to this computer, the computer alerts you
that a new hardware device has been detected, and asks you if you want to install
the driver. Refer to the GageView Interface Program — User’s Manual (P/N: 910-259-
EN [U8778347]) for more information.
The driver loads, after which you can start using a program like GageView to
communicate with the 45MG.
188 Chapter 11
DMTA-10022-01EN, Rev. B, October 2013
Data transmitted from the 45MG remains in the 45MG internal memory.
190 Chapter 11
DMTA-10022-01EN, Rev. B, October 2013
• At the next survey date, months or years later, retrieve the previously saved
survey thickness data stored in a computer file with ID numbers. This retrieved
data can be used for the following purposes:
— To guide you through the measurement sequence using the ID numbers.
— To compare current values at the measurement site with previously measured
thickness values.
— To manually or automatically verify that the current measurement setup is
identical to the previously used setup.
• Create ID number sequences in a computer, and then download them to the
instrument. This externally created sequence can guide you through the
prescribed measurement location path. The ID sequence created in the computer
must have setup information. The setup can be the gage default setups, or any
other desired sequence of setups.
The data downloaded to the 45MG must be in the exact same format as the data
transmitted. Olympus recommends using the GageView interface program to
handle all functions of interfacing, storing, and creating 45MG data. Please
contact Olympus for information on additional software data management
programs.
192 Chapter 11
DMTA-10022-01EN, Rev. B, October 2013
It is possible to save a screenshot of the full 45MG screen contents to an image file.
This function is useful when you need an exact replica of the display for reporting or
documentation purposes, and can be performed using one of the following two
methods:
Refer to the GageView Interface Program — User’s Manual (P/N: 910-259-EN [U8778347])
for information on how to install and use GageView.
194 Chapter 11
DMTA-10022-01EN, Rev. B, October 2013
Figure 11-5 The Device Tools dialog box with a screen capture
196 Chapter 11
DMTA-10022-01EN, Rev. B, October 2013
198 Chapter 11
DMTA-10022-01EN, Rev. B, October 2013
This section describes how to maintain your 45MG by carrying out routine care and
maintenance.
The 45MG case is sealed to prevent ingress of environmental liquids and dust when
the battery door is closed. However, the instrument should never be immersed in any
fluid.
The 45MG case is designed to withstand normal field use. However, as with any
electronic instrument, damage is possible if the instrument is not handled with
adequate care. Observe the following instructions in particular:
• When connecting cables to the instrument, first align the connector with the
mating receptacle on the instrument (dual element D79X series transducer center
pin protruding downward), and then carefully insert the connector straight into
the receptacle.
• When disconnecting cables from the instrument, first grasp the connector (not the
cable), and then gently pull it out.
• Do not throw or drop the instrument.
• Do not use strong solvents or abrasives to clean the rubber boot, case, keypad, or
display screen.
First, clean the instrument with a dry damp cloth. If necessary, use a wet damp cloth
with a mild detergent to wash the instrument. Dry the instrument before using it.
The ultrasonic transducers used with the 45MG are rugged devices that need little
care. However, they are not indestructible, and some attention to the following items
result in a longer lifespan.
200 Chapter 12
DMTA-10022-01EN, Rev. B, October 2013
The 45MG includes reset functions that can be used to quickly restore the gage to its
default parameters. Resets are useful shortcuts to known configurations. The reset
functions are as follows:
MEASUREMENT RESET
Changes the measurement parameters to the default factory values listed in
Table 14 on page 201.
The internal memory reset permanently deletes all thickness readings/waveforms that
are stored in the memory card. However, this reset does not delete any stored
transducer setups.
CLOCK RESET
Resets the date to 01/01/2010 in the MM/DD/YYYY format, and the time to 12:00
A.M. in the 12-hour format.
COMMUNICATION RESET
Changes the communication settings to the default factory values listed in
Table 15 on page 202.
MASTER RESET
Performs the measurement reset and the internal memory reset in a single step.
The master reset permanently deletes all stored thickness readings/waveforms that
are stored in the internal memory card of the 45MG.
202 Chapter 12
DMTA-10022-01EN, Rev. B, October 2013
The 45MG includes a function that can be used to perform several diagnostic self
tests. The tests can help to localize a suspected hardware problem, or to verify correct
hardware operation. Some tests are designed for Olympus internal testing procedures
during the manufacturing process.
• Keypad test
• Video test
• Internal MicroSD card test (Pass or Fail)
• External MicroSD card test (Pass or Fail)
• Dual element transducer test
• ESS test (Electronic Stress Screening) [for Olympus internal use only]
• B-scan test (for Olympus internal use only)
• Battery test (for Olympus internal use only)
• One-wire test (for Olympus internal use only)
204 Chapter 12
DMTA-10022-01EN, Rev. B, October 2013
• FAIL: Indicates that there is a problem with the memory card. When the
external card fails, reinstall or replace the card, and then restart the
instrument. When the internal card fails, contact Olympus for service.
b) Press [ENTER] to terminate the SD CARD test.
7. When DUAL XDCR TEST is selected, in the DUAL XDCR TEST screen (see
Figure 12-3 on page 205):
a) Press [MEAS] to start the test that measures the time of flight through each
side of a dual element transducer.
The measured parameter values appear.
b) Interpret the Tx and the Rx values as follows:
• Similar values indicate a normal dual element transducer.
• A difference between the values indicates that the delay line wear is not
the same for each element.
• A missing value indicates that a cable is broken, or that the element is not
functioning.
c) Take note of the calculated ZERO VALUE.
d) Press [ENTER] to terminate the dual element transducer test.
The software diagnostic (SW DIAG) function generates an error log documenting all
errors that occurred during instrument operation. Olympus uses this information to
troubleshoot the 45MG operating software.
The STATUS screen lists important information about the instrument. The status
screen displays the following information:
206 Chapter 12
DMTA-10022-01EN, Rev. B, October 2013
When operating the gage, certain error messages may appear. These messages usually
indicate a problem with the operating procedure, but certain messages might indicate
a physical problem with the gage itself. If you do not understand an error message,
contact Olympus for assistance.
The power indicator (in the upper-left corner of the display) contains bars that
indicate the percentage of remaining battery charge level. When the charge level
becomes low, the gage automatically turns off. If the gage turns off immediately after
you turn it on, or if you are unable to turn it on, the batteries are probably depleted.
208 Chapter 12
DMTA-10022-01EN, Rev. B, October 2013
210 Appendix A
DMTA-10022-01EN, Rev. B, October 2013
212 Appendix A
DMTA-10022-01EN, Rev. B, October 2013
a. The maximum thickness measuring capability depends on the transducer type, material
conditions, and temperature.
214 Appendix A
DMTA-10022-01EN, Rev. B, October 2013
216 Appendix A
DMTA-10022-01EN, Rev. B, October 2013
The following table lists the sound velocities in a variety of common materials. Use
this table only as a guide. The actual velocity in these materials may vary significantly
due to a variety of causes, such as composition, preferred crystallographic orientation,
porosity, and temperature. Therefore, for maximum accuracy when establishing the
sound velocity in a given material, first test a sample of the material.
References
Fredericks, J. R. Ultrasonic Engineering, John Wiley & Sons, Inc., New York, 1965.
“Handbook of Chemistry and Physics” Chemical Rubber Co., Cleveland, Ohio, 1963.
Mason, W. P. “Physical Acoustics and the Properties of Solids”, D.Van Nostrand Co., New
York, 1958.
218 Appendix B
DMTA-10022-01EN, Rev. B, October 2013
220 Appendix B
DMTA-10022-01EN, Rev. B, October 2013
Table 32 Couplants
Description Part Number
Glycerin, 2 oz. (0.06 l) Couplant B-2 [U8770023]
Gel type, 12 oz. (0.35 l) Couplant D-12 [U8770026]
Ultratherm-high temperature up to 1000 °F Couplant E-2 [U8770274]
(538 °C), 2 oz. (0.06 l)
Medium temperature up to 500 °F (260 °C), 2 oz. Couplant F-2 [U8770141]
(0.06 l)
222 Appendix C
DMTA-10022-01EN, Rev. B, October 2013
224 Appendix C
DMTA-10022-01EN, Rev. B, October 2013
List of Figures
Figure 6-32 Deleting the data of an ID range in the active file ...................................... 112
Figure 6-33 Warning message when resetting measurements ....................................... 113
Figure 6-34 The MEMORY status screen .......................................................................... 114
Figure 6-35 The ID overwrite protection message ........................................................... 115
Figure 6-36 Identifying the ID review screen ................................................................... 116
Figure 6-37 Editing the ID # edit mode ............................................................................. 118
Figure 6-38 The message when the edited ID is not in the database ............................ 118
Figure 6-39 The FILE SUMMARY report screen .............................................................. 121
Figure 6-40 The FILE SUMMARY report result screen ................................................... 121
Figure 6-41 The MIN/MAX SUMMARY report screen ................................................... 122
Figure 6-42 The FILE COMPARISON report screen ....................................................... 123
Figure 6-43 The FILE COMPARISON report result screen ............................................ 123
Figure 6-44 The ALARM SUMMARY report result screen ............................................ 124
Figure 6-45 The MIN/MAX SUMMARY report result screen ........................................ 124
Figure 6-46 Returning to the measurement screen .......................................................... 125
Figure 7-1 Normal differential mode (shown with Waveform option activated) ..... 128
Figure 7-2 The DIFF screen ............................................................................................... 129
Figure 7-3 Displaying the minimum and/or maximum thickness (shown with the
Waveform option activated) ........................................................................... 130
Figure 7-4 Example of a HIGH alarm indicator (shown with the Waveform option
activated) ........................................................................................................... 132
Figure 7-5 Example of a B-scan alarm mode .................................................................. 133
Figure 7-6 YEL (yellow) and RED alarm indicators (shown with the Waveform option
activated) ........................................................................................................... 134
Figure 7-7 Setting up the STANDARD alarm ................................................................ 135
Figure 7-8 Example of a locked function message in the help bar .............................. 137
Figure 7-9 The PASSWORD screen .................................................................................. 137
Figure 7-10 The LOCKS screen ........................................................................................... 138
Figure 8-1 The MEAS screen ............................................................................................. 142
Figure 8-2 The SYSTEM screen ......................................................................................... 144
Figure 8-3 The COMM screen ........................................................................................... 147
Figure 9-1 Manually adjusting the gain .......................................................................... 150
Figure 9-2 Adjusting the extended blank length ........................................................... 152
Figure 9-3 B-scan example ................................................................................................. 153
Figure 9-4 Changing B-scan parameters ......................................................................... 154
Figure 9-5 B-scan elements ................................................................................................ 155
Figure 9-6 B-scan freeze review elements ....................................................................... 156
Figure 9-7 Example of the half-size DB grid ................................................................... 160
Figure 9-8 Changing DB grid parameters ....................................................................... 160
Figure 9-9 Grid transposition example ............................................................................ 161
Figure 9-10 Example of a linearized DB grid ................................................................... 161
Figure 9-11 The highlighted DB grid cell in the ID review screen ................................ 163
List of Tables
Index
Index 231
DMTA-10022-01EN, Rev. B, October 2013
C compliance
calibration 54, 61 C-Tick (Australia) 4
instrument 55 EMC directive 12
locked 61 FCC (USA) 12
material sound velocity 59 ICES-001 (Canada) 12
test block 58 regulatory information 12
THRU-COAT 77 configuring
transducer zero compensation 59 beeper 41
Canada, ICES-001 compliance 12 communications 145
capturing screen with GageView 193 DB grid 160
caution notes differential mode 128
battery replacement 28 inactive time 41
cannot recover data after measurement reset instrument 141
113 measurement parameters 141
cannot recover deleted file content 110 radix type 41
cannot recover erased data 119 system parameters 144
experienced operator only 151, 166 connections 19
ingress protection guarantee 18 copying
internal memory reset deletes all data 202 file 108
master reset deletes all data 202 corrosion applications 76
replacing setup deletes original data 183 couplant types 64
CE marking 4 coupling technique 62
changing creating
active ID 117 custom setups for SE transducers 166
display settings 44 data file 92
echo detection mode 72 C-Tick mark 4
file data mode 105 curvature of test piece 62
highlighted cell in DB grid 162 custom setups 166
language 41 D
thickness-measurement resolution 49
danger notes
character
electric shock 1
cycle in traditional method 39
data
deleting 40
erasing in active file 119
inserting 40
exchange 189
China RoHS 4, 11
saving 66
cleaning, instrument 200
data file
clock
2-D grid 98
reset 202
boiler 102
setting 42
creating 92
coated material 76
incremental 94
color scheme 45
sequential 96
changing 44
sequential with custom points 97
communication
datalogger
configuring 145
about the 89
reset 196, 202
file based system 89
232 Index
DMTA-10022-01EN, Rev. B, October 2013
Index 233
DMTA-10022-01EN, Rev. B, October 2013
234 Index
DMTA-10022-01EN, Rev. B, October 2013
Index 235
DMTA-10022-01EN, Rev. B, October 2013
236 Index
DMTA-10022-01EN, Rev. B, October 2013
Index 237
DMTA-10022-01EN, Rev. B, October 2013
238 Index