0% found this document useful (0 votes)
267 views30 pages

Arcovis 3000 S - R9 en Rev 9

This document provides technical specifications for an Arcovis 3000 S/R 9'' C-arm mobile unit with a 9'' image intensifier and stationary/rotating anode tube. It includes details on electrical data, environmental conditions, mechanical data, user interface languages, radiological data, and exposure modes.

Uploaded by

Hristo
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
267 views30 pages

Arcovis 3000 S - R9 en Rev 9

This document provides technical specifications for an Arcovis 3000 S/R 9'' C-arm mobile unit with a 9'' image intensifier and stationary/rotating anode tube. It includes details on electrical data, environmental conditions, mechanical data, user interface languages, radiological data, and exposure modes.

Uploaded by

Hristo
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 30

Arcovis 3000 S/R 9’’ – C-arm mobile unit with 9’’ I.I.

and stationary / rotating anode tube

Product Data
Electrical data
Features Data
Power Supply 230 Vac ±10% standard monophase
105/115/125/220/240 Vac ±10% monophase on request
Frequency 50/60 Hz ±5 Hz
Absorbed current see following schedule
Line compensation Automatic
Line resistance <0,4 Ω @230 V <0,2 Ω @115 V
Standard mains plug 16A @ 230 Vac
Supply cable 8m
Insulation class Class I with applied parts type B
Use conditions Continuous working with intermitting load
Classification according to the IPx0
liquids seepage
Safety in presence of The equipment is not type AP or APG
inflammable anesthetic gases

Absorbed current
Fixed anode version Rotating anode version
in fluoroscopy 4,5 A @ 230 Vac - 7,5 A @ 115 Vac 5 A @ 230 Vac – 10 A @ 115 Vac
in radiography 20 A @ 230 Vac – 28 A @ 115 Vac 20 A @ 230 Vac – 30 A @ 115 Vac
in stand-by 1,0 A @ 230 Vac - 1,7 A @ 115 Vac 1,0 A @ 230 Vac - 2,0 A @ 115 Vac

Environmental conditions
Environmental Factor In normal use Warehouse and transport
Temperature from 10 °C to 40 °C from -25°C to 70°C
Relative Humidity from 30 % to 75 % non- from 10% to 90% non-
condensing condensing
Pressure from 700 hPa to 1060 hPa from 500 hPa to 1060 hPa

Product data Arcovis 3000 S/R 9” – Rev. 9 Page 1 of 30


Total equipment filtration
Description Data
Monobloc 1,4 mmAl
Permanent additional filter 1 mmAl
Collimator 0 mmAl
Monobloc cover 0,1 mmAl
Total filtration of monobloc 2,5 mm Al
group
Permanent additional DAP 0,3 mmAl
meter filtration
Total filtration 2,8 mmAl

Mechanical data C-arm stand 9"


Dimensions in transport position Data
Width (1) 860 mm (33,9 in)
Depth (2) 1950 mm (76,8 in)
Height (3) 1810 mm (7,3 in) (camera 1k)
1780 mm (70,1 in) (camera 0,5k)
Control console height (4) 830 mm (32.7 in)

Description Data
Vertical run (A) Total run 500 mm (19,69 in), motorized in 60 s

Horizontal run (B) Manual, 210 mm (8,27 in)

Arm rotation around the (C) Manual rotation, ± 270°, motorized rotation ± 30°
horizontal axis (optional)

Orbital rotation (D) Manual, 135° (+93° ÷ - 42°)

Arm group overview (E) Manual, ± 12,5°

Useful space (F) 770 mm (30,31 in)

Arm depth (G) 690 mm (27,17 in)

Product data Arcovis 3000 S/R 9” – Rev. 9 Page 2 of 30


Description Fixed Anode Data Rotating Anode Data
S.I.D. (H) 988 mm (38,90 in) 980 mm (38,6 in)
Focus-skin distance (I) 218 mm (8,58 in) 200 mm (7,9 in)
Min. distance from floor (L) 172 mm (6,77 in) 150 mm (5,9 in)
Floor-focus distance (M) 339 mm (13,35 in) 360 mm (14,2 in)
Floor-skin distance (N) 557 mm (21,93 in) 560 mm (22,0 in)
Min. II center-floor distance in (O) 954 mm (37,56 in) 950 mm (37,4 in)
oblique projection
Width in oblique projection (P) 1603 mm (63,11 in) 1660 mm (65,4 in)

Description Data
Equipment movement Rear drive wheels with manual control from the operator,
front castor. Manual parking brake.
Rear wheels diameter 125 x 40 mm (4,92 x 1,57 in)
Front wheels diameter 80 x 30 mm (3,15 x 1,18 in)
Protection against the cables Cable pusher on all wheels of the C-arm stand.
squeezing

Description Fixed Anode Version Rotating Anode Version


Weight 340 Kg (749,57 lb) 360 Kg (793,66 lb)

Overall dimensions 9"

Product data Arcovis 3000 S/R 9” – Rev. 9 Page 3 of 30


“Basic” display station
Version for 1+330, 1+2700 memories

Load limit for the upper shelf: 2 Kg


Load limit for the lower shelf: 10 Kg

Product data Arcovis 3000 S/R 9” – Rev. 9 Page 4 of 30


Display station "Basic"
Version for RTP 800 memory

Load limit for the upper shelf: 2 Kg


Load limit for the lower shelf: 10 Kg

Product data Arcovis 3000 S/R 9” – Rev. 9 Page 5 of 30


"Premium" display station
Version for RTP 1000 and HRP 3000 memories

Load limit for the upper shelf: 2 Kg


Load limit for the lower shelf: 10 Kg

Product data Arcovis 3000 S/R 9” – Rev. 9 Page 6 of 30


User's interface and languages
Description Data
User's interface Membrane keyboard with alphanumeric touch-screen 5.7” LCD display
for all the operative parameters and error messages. Microprocessor
management. Keyboard can be rotated of ±60°
Selectable languages Italian, English, French. German, Spanish, Russian (selection by set-up)

Radiological data

Fixed anode monobloc


Description Fixed anode version
Generator, power in DC current 2,5 kW @ 100 kV (100 kV, 25 mA, 100 ms)
Generator, max power in DC current 3,5 kW (100 kV, 35 mA) mAs < 2,2 @ 230 Vac
2,5kW (100 kV, 25 mA) mAs < 2,2 @ 115 Vac)
Max high voltage (fluoroscopy and radiography) 110 kVp
Inverter frequency 40 kHz
Max. current in continuous fluoroscopy 8,0 mA (standard curve)
Max. current in “SNAPSHOT” fluoroscopy 10 mA
Max. current in radiography 25 mA @115 Vac – 35 mA @230 Vac
Max. mAs in radiography 90 mAs @115 Vac – 125 mAs @230 Vac
Max. fluoroscopy time at 75 W (75 kV-1 mA) continuous
Max. fluoroscopy time at 280 W (80 kV-3,5 A) 44 min
Max. fluoroscopy time at 400 W (100 kV-4 mA) 29 min
Max. fluoroscopy time at 525 W (70 kV-7,5 mA) 24 min
Max. fluoroscopy time at 550 W (110 kV-5 mA) 21 min
Max. fluoroscopy time H.U Safety after 21 min of fluoroscopy @110 kV,
5 mA (550 W).
Cooling at 30° 300 min

Rotating anode monobloc


Description Rotating anode version
Generator, power in DC current 2,5 kW @100 kV (100 kV, 25 mA, 100 ms)
Generator, max power in DC current 5k W (100 kV, 50 mA) mAs < 2,2 @230 Vac;
3,5 kW (100 kV, 35 mA) mAs < 2,2 @115 Vac
Generator, max power in DC current (with HRP) max 7,2 kW (120 kV 60 mA, pulse mode 25 fps)
Max high voltage (fluoroscopy and radiography) 120 kVp
Inverter Frequency 40 kHz
Max. current in continuous fluoroscopy 8,0 mA (standard curve) – 15 mA (RTP 800
and RTP 1000 memories)
Max current in HCF with HRP 30 mA
Max. current in “SNAPSHOT” fluoroscopy 12 mA (0,5K camera) – 30 mA (1K camera)
Max current in pulsed fluorography with HRP 60 mA @ 230 Vac – 45 mA @115 Vac
“Snapshot”
Product data Arcovis 3000 S/R 9” – Rev. 9 Page 7 of 30
Max current in digital graphy-mode with HRP 60 mA @ 230 Vac – 45 mA @115 Vac
Max. current in radiography 35 mA @ 115 Vac – 50 mA @ 230 Vac
Max. mAs in radiography 90 mAs @115 Vac – 125 mAs @230V ac
Max. fluoroscopy time at 75 W (75 kV-1 mA) continuous
Max. fluoroscopy time at 280 W (80 kV-3,5 A) 54 min
Max. fluoroscopy time at 400 W (100 kV-4 mA) 38 min
Max. fluoroscopy time at 525 W (70 kV-7,5 mA) 31 min
Max. fluoroscopy time at 600 W (120 kV-5 mA) 28 min
Max. fluoroscopy time H.U Safety after 28 min of fluoroscopy @120
kV, 5 mA (600 W).
Cooling at 30° 420 min

Exposure mode: fluoroscopy

Continuous fluoroscopy
Description Fixed Anode Version
Small focus (IEC 336) 0,5 mm
kV variation range 40 - 110 kV
mA variation range 0,25 - 8 mA
kV-mA relationship Standard: 40 kV / 0,5 mA, 80 kV / 7,6 mA , 110 kV / 8 mA
Storage Last Image Hold
Safety timer Audible alarm after 5’ of uninterrupted or cumulative load
application¹. X-ray stop after uninterrupted 10’ of load
application¹ (not cumulative).
¹ defined as electricity provisions to the x-ray tube anode (IEC 60601-2-54).

Description Rotating Anode Version


Small focus (IEC 336) 0,3 mm
kV variation range 40 - 120 kV
mA variation range 0,25 - 8 mA (15 mA with RTP 800 and RTP 1000 memories)
kV-mA relationship Standard: 40 kV / 0,5 mA, 80 kV / 7,6 mA, 120 kV / 8 mA
Storage Last Image Hold
Safety timer Audible alarm after 5’ of uninterrupted or cumulative load
application¹. X-ray stop after uninterrupted 10’ of load
application¹ (not cumulative).
¹ defined as electricity provisions to the x-ray tube anode (IEC 60601-2-54).

Product data Arcovis 3000 S/R 9” – Rev. 9 Page 8 of 30


Pulsed fluoroscopy
0,5K² Camera
Description Fixed Anode Version
Small focus (IEC 336) 0,5 mm
Exposure frequency (selectable 2 imm/s, 1 imm/s, 1 imm/2s, 1 imm/3s, 1 imm/5s (with 1+330,
from console) 1+2700 memories)
Xr flash time Min. time for the best image quality
Other features as per Continuous Fluoroscopy

Description Rotating Anode Version


Small focus (IEC 336) 0,3 mm
Exposure frequency 2 imm/s, 1 imm/s, 1 imm/2s, 1 imm/3s, 1 imm/5s (1+330, 1+2700
(selectable from console) memories)
Xr flash time Min. time for the best image quality
Other features as per Continuous Fluoroscopy

1K² Camera
Description Fixed Anode Version
Small focus (IEC 336) 0,5 mm
Exposure frequency 2/s, 1/s, 1/3s (for RTP 1000 memory)
(selectable from console) 12/s, 6/s, 3/s, 1/s (fo RTP 800 memory)
Xr flash time Min. time for the best image quality
Other features as per Continuous Fluoroscopy

Description Rotating Anode Version


Small focus (IEC 336) 0,3 mm
Exposure frequency 2/s, 1/s, 1/3s (for RTP 1000 memory)
(selectable from console) 12/s, 6/s, 3/s, 1/s (fo RTP 800 memory)
Xr flash time Min. time for the best image quality
Other features as per Continuous Fluoroscopy

HCF Fluoroscopy (HRP memory)


Description Data
Exposure frequency From 1 img/s to 25 img/s (according to the selected ’Average’
(selectable from console) weight).
Xr flash time Min. time for the best image quality.
Focus 0,6 mm Large Focus
mA variation range 15/30 mA
Images acquisition Performed by integrating the selected images number in the
memory set-up.
Images storage Manual
Other features as per Continuous Fluoroscopy

Product data Arcovis 3000 S/R 9” – Rev. 9 Page 9 of 30


Pulsed fluorography (HRP memory)
Description Data
Exposure frequency (selectable from From 1 img/s to 25 img/s (according to the selected
memory) ‘Average’ weight).
mA variation range 15/30/45/60 mA @ 230 Vac, 15/30/45 mA @ 115
Vac
Focus 0,6 mm Large Focus
Xr flash time Min. time for the best image quality.
Images acquisition Integration of 16 continuous readings.
Images storage Automatic storage into the RAM memory or in
nonvolatile memory.
Other features as per Continuous Fluoroscopy

Snapshot Fluoroscopy
0,5K² Camera
Description Fixed Anode Version
mA variation range 1-10 mA
Xr flash time <1s
Acquisition obtained from the integration of 16 continuative readings.
Automatic storage in non-volatile RAM memory according to the memory.
Other features as per Continuous Fluoroscopy

Description Rotating Anode Version


mA variation range 1-12 mA
Xr flash time <1s
Acquisition obtained from the integration of 16 continuative readings.
Automatic storage in non-volatile RAM memory according to the memory.
Other features as per Continuous Fluoroscopy

1K² Camera
Description Fixed Anode Version
mA variation range 1-10 mA
Xr flash time <1s
Acquisition obtained from the integration of 16 continuative readings.
Automatic storage in non-volatile RAM memory according to the memory.
Other features as per Continuous Fluoroscopy

Description Rotating Anode Version


mA variation range 2,5-30 mA
ma variation range (with HRP) 60 mA @230 Vac – 45 mA @ 115 Vac
Xr flash time <1s
Acquisition obtained from the integration of 16 continuative readings.
Automatic storage in non-volatile RAM memory according to the memory.
Other features as per Continuous Fluoroscopy
Product data Arcovis 3000 S/R 9” – Rev. 9 Page 10 of 30
(1/2) mA Fluoroscopy
Description Fixed Anode Version
mA variation range 0,25-4 mA
kV-mA relationship 40 kV / 0,25 mA, 80 kV / 3,8 mA, 110 kV / 4 mA
Other features as per Continuous Fluoroscopy

Description Rotating Anode Version


mA variation range 0,25-4 mA
kV-mA relationship 40 kV / 0,25mA, 80 kV / 3,8 mA, 120 kV / 4 mA
Other features as per Continuous Fluoroscopy

APR Fluoroscopy
0,5K² Camera
Description Fixed Anode Version
APR standard 40 kV / 0,5 mA, 80 kV / 7,6 mA, 110 kV / 8 mA
APR 1 (Fine anatomic parts or 40 kV / 0,7 mA, 80 kV / 6,4 mA, 110 kV / 6,6 mA
for pediatric use)
APR 2 (lungs) 40 kV / 0,5 mA, 80 kV / 7 mA, 110 kV / 6 mA
The curves cannot be modified by the operator.

Description Rotating Anode Version


APR standard 40 kV / 0,5 mA, 80 kV / 7,6 mA, 120 kV / 8 mA
APR 1 (Fine anatomic parts or 40 kV / 0,7 mA, 80 kV / 6,4 mA, 120 kV / 6,6 mA
for pediatric use)
APR 2 (lungs) 40 kV / 0,5 mA, 80 kV / 7 mA, 120 kV / 5,5 mA
The curves cannot be modified by the operator.

1K² Camera
Description Fixed Anode Version
APR standard 40 kV / 0,5 mA, 80 kV / 7,6 mA, 110 kV / 8 mA
APR 1 (Fine anatomic parts or 40 kV / 0,7 mA, 80 kV / 6,4 mA, 110 kV / 6,6 mA
for pediatric use)
APR 2 (lungs) 40 kV / 0,5 mA, 80 kV / 7 mA, 110 kV / 6 mA
The curves cannot be modified by the operator.

Description Rotating Anode Version


APR standard 40 kV / 0,5 mA, 80 kV / 7,6 mA, 120 kV / 8 mA
APR 1 (Fine anatomic parts or 40 kV / 0,7 mA, 80 kV / 6,4 mA, 120 kV / 6,6 mA
for pediatric use)
APR 2 (lungs) 40 kV / 0,5 mA, 80 kV / 7 mA, 120 kV / 5,5 mA
APR 3 (Sturdy patient) 40 kV / 1 mA, 70 kV / 15 mA, 120 kV / 9 mA
The curves cannot be modified by the operator.

Product data Arcovis 3000 S/R 9” – Rev. 9 Page 11 of 30


Exposure mode: radiography

Two points technique, kV and mAs selection

Equipment with mains voltage at 230Vac


Description Fixed Anode Version
Large Focus (IEC 336) 1,5 mm
kVp range 40-110 kV
mA range 25 mA fixed from 40 kV to 100 kV; 22 mA at 110 kV
mAs range 1 – 125 mAs from 40 to 100 kV, 1 – 100 mAs from 101 to 110 kV
in 42 steps, curve R’20
HiRad mA range (mAs < 2,2) 35 mA fixed from 40 kV to 100 kV; 31 mA at 110 kV
HiRad Exposure times range 28-90 ms
Exposure times range 0,04-5 s
Duty cycle Calculated according to the anode dissipation

Description Rotating Anode Version


Large Focus (IEC 336) 0,6 mm
kVp range 40-120 kV
mA range 25 mA fixed from 40 kV to 100 kV; 20 mA at 120 kV
mAs range 1 – 125 mAs from 40 to 100 kV, 1 – 100 mAs from 101 to 120 kV
at 42 steps, curve R’20
HiRad mA range (mAs < 2,2) 50 mA fixed from 40 kV to 100 kV; 30 mA at 120 kV
HiRad Exposure times range 20-64 ms
Exposure times range 0,04-5 s
Duty cycle Calculated according to the anode dissipation

Equipment with mains voltage at 115 Vac


Description Fixed Anode Version
Large Focus (IEC 336) 1,5 mm
kVp range 40-110 kV
mA range 18 mA fixed from 40 kV to 100 kV; 16 mA at 110 kV
mAs range 1 – 90 mAs from 40 to 100 kV, 1 – 80 mAs from 101 to 110 kV
HiRad mA range (mAs < 2,2) 25 mA fixed from 40 kV to 100 kV; 22 mA at 110 kV
HiRad exposure times range 40-128 ms
Exposure times range 0,04-5 s
Duty cycle Calculated according to the anode dissipation

Product data Arcovis 3000 S/R 9” – Rev. 9 Page 12 of 30


Description Rotating Anode Version
Large Focus (IEC 336) 0,6 mm
kVp range 40-120 kV
mA range 18 mA from 40 kV to 100 kV; 15 mA at 120 kV
mAs range 1 – 90 mAs from 40 to 100 kV, 1 – 71 mAs from 101 to 120 kV
HiRad mA range (mAs < 2,2) 35 mA fixed from 40 kV to 100 kV; 20 mA at 12 0kV
HiRad exposure times range 28-90 ms
Exposure times range 0,04-5 s
Duty cycle Calculated according to the anode dissipation

APR Radiography
Description Norm Dose¹ Dose¹ Dose¹
Set mGy mGy mGy
APR 1 (Head) 77 kV-56 mAs 6,769 74 kV – 45 5,026 80 kV – 71 mAs 9,261
mAs
APR 2 110 kV – 11 2,462 107 kV – 9 1,911 110 kV – 14 mAs 3,154
(Lungs) mAs mAs
APR 3 85 kV – 22 3,153 82 kV – 28 3,777 88 kV – 18 mAs 2,741
(Pelvis) mAs mAs
¹ The dose measurement (in mGy) has been performed in compliance with IEC 60601-1-3 § 5.2.4.2. and
60601-2-54 §203.5.2.4.5.101 Skin dose level.

Product data Arcovis 3000 S/R 9” – Rev. 9 Page 13 of 30


Fixed anode X-ray group

I-40S 3,5 RF Monobloc


Description Data
Monobloc model I-40S 3,5 RF
Max. power (100 kV – 35 mA) (IEC 60601-1) 3,5 kW
Max. tube voltage 110 KV
Ripple at the max. power <2%
kV Rise time at max. power <1 ms
Mechanical housing features
Half-value layer @75 kV 2,0 mmAl
Min. inherent filtration @75 kV 1,4 mmAl
Weight 15,5 kg (34,17 lb)
Thermal housing features
Available thermal capacity (RX) 590 kJ (790 kHU)
Total thermal capacity 900 kJ (1215 kHU)
Thermal safety 60°C ±5°C
Compensation lung 410 cm³ (25.015 cubic inch)
Continuous thermal dissipation 80 W, 109 HU/s, 6400 HU/min
X-ray tube filament power supply - Max 500 mA
current (rms)
Leakage radiation (IEC 60601-1-3) <1 mGy/h

Fixed anode X-ray insert IAE F112


Description Data
Nominal high voltage 120 kVP
Max. filament current 4,3 A
Nominal focus dimension: small focus 0,5 mm
Nominal focus dimension: large focus 1,5 mm
Nominal anodic power: small focus 780 W
Nominal anodic power: large focus 4200 W
Anode material Tungsten
Anode inclination angle 12°
Thermal anode capacity 40 kJ (54 kHU)
Anode max. thermal dissipation 500 W (670 HU/s)
Min. inherent filtration 0,5 mm Al
Tube material Glass

Product data Arcovis 3000 S/R 9” – Rev. 9 Page 14 of 30


Rotating anode X-ray group

I-40R 5 RF Monobloc
Description Data
Monobloc model I-40R 5 RF
Max. power (100 kV – 50 mA) (IEC 60601-1) 5 kW
Max. tube voltage 120 KV
Ripple at the max. power <2%
kV Rise time at max. power <1 ms
Mechanical housing features
Half-value layer @75 kV 2,0 mmAl
Min. inherent filtration @75 kV 1,4 mmAl
Weight 19 kg (41,89 lb)
Thermal housing features
Available thermal capacity (RX) 650 kJ (870 kHU)
Total thermal capacity 940 kJ (1259 kHU)
Thermal safety 60°C ±5°C
Compensation lung 410 cm³ (25.015 cubic inch)
Continuous thermal dissipation 85 W, 115 HU/s, 6900 HU/min
X-ray tube filament power supply - Max. 500 mA
current (rms)
Leakage radiation (IEC 60601-1-3) <1 mGy/h

X-Ray insert model IAE X20P 0,3-0,6


Description Data
Max. peak voltage 120 kVP
Max. filament current 5,4 A
Nominal focus dimension: small focus 0,3 mm
Nominal focus dimension: large focus 0,6 mm
Nominal anodic power: small focus 5 kW
Nominal anodic power: large focus 17 kW
Anode material Rhenium / Tungsten / Molybdenum
Anodic diameter 64 mm
Anode inclination angle 10°
Thermal anode capacity 150 kJ (200 kHU)
Anode continuous thermal dissipation 300 W (24 kHU/min, 18 kJ/min)
Anode max. thermal dissipation 500 W (40 kHU/min)
Min. inherent filtration 0,7 mm Al
Tube material glass
Speed of rotation of the anode 3000 rpm

Product data Arcovis 3000 S/R 9” – Rev. 9 Page 15 of 30


Collimator
Description Data
Model with iris R605/027F/DASM
Model with iris and parallel shutters R605/027E/DASM
Fields dimension (DF 100 cm- 39”): round field 5±23 cm
delimitation
Fields dimension (DF 100 cm- 39”): elliptical 0±23 cm
field delimitation
Leakage radiation (EN60601-1-3) < 1 mGy/h
Inherent filtration (EN60601-1-3) 0 mm
Classification EN60601-1 par.6:
Protection against electrical hazards Class I
Protection against direct and indirect contacts equipment with applied part Type B
Protection against water seepage common protection (IPXO)

Image Intensifier 9" (1)


Description Triple 9/6/4"
Brand Thales
Model TH 9428 HP2H542 VR13
Fields Number 3
Nominal Input Diameter 230 mm (9,06 in)
Output Image Diameter 20 mm (0,79 in)
Output Window Thickness 3,6 mm (0,14 in)
Useful Input Field Size 215/160/120 mm (8,46/6,30/4,72 in)
Typical Resolution (Central) 48/56/64 lp/cm
Conversion Factor (Cd/m-2/mR/s-1) 240/120/60
Contrast Ratio 23:1/25:1/30:1
Integral Image Distortion 4% / 2% / 1%
Differential Distortion at 90% radius 15% / 6% / 3%
DQE at 59.5 keV 65%
"All metal" Technology Yes
Input Screen "Hi-Res." Yes
MTF at 10 Lp/cm 60/65/70%
MTF at 20 Lp/cm 25/30/40%
Low frequency drop LDF 7/6/5%
(1) According to IEC standard IEC from 1262-1 to 1262-6

Fixed antiscattering grid


Line rate 60 L/cm
Ratio 10:1
Focus 100 cm
Material Fiber interspacer and carbon cover

Product data Arcovis 3000 S/R 9” – Rev. 9 Page 16 of 30


TV chain

CCD 0,5K x 0,5K camera


Camera for L.I.H., 1+330, 1+2700 memories
Description Data
Camera technology CCD at low persistence of ½” (470.000 pixels)
Video standard CCIR 625/50 Hz interlaced with matrix 752x582 pixels
Aspect ratio 4:3
Band width 20 MHz ± 3 dB
Signal-noise ratio 65 dB
Resolution 20 lines-pairs (on 6” image intensifier)
Gamma correction 0,4 or 1
Automatic video level compensation Yes
Dynamic contrast Shading Yes
Video output A/D converter 10 bit

Camera CCD 1K x 1K mvBlueCOUGAR


Camera per memoria RTP 800
Description Data
Model mvBlueCOUGAR-X122G
Resolution 1280 x 960
Mpixels 1.2
Max frame rate 31 Hz
Binning 8/2/8/2
Shutter type Global
Sensor size 1/3"
Pixel size 3.75 x 3.75 µm
Readout type Progressive
Exposure time 10 µs - 20 s
ADC resolution / output 14 bit → 14, 12, 8 bit
EMVA1288 Yes
Trigger (HW /SW) Yes
Pipelined global shutter in trigger mode Yes

Product data Arcovis 3000 S/R 9” – Rev. 9 Page 17 of 30


Thales CCD 1K x 1K camera
Camera for RTP 1000 and HRP 3000 memories
Description Data
Model Thales L103, L108, L115
Camera technology CCD interline progressive scanning
Iris Motorized and manual
Image resolution 1000 x 1000 pixels
Signal-noise ratio 62 dB
Digital video out 12 bits
Max frame rate 30 fps
Min exposure time 33 ms @ 1000 x 1000
Sensibility Motorized lens: > 64 LSB/Cd/m2
Manual lens: 18 LSB/Cd/m2
Iris opening range Motorized lens: from F/1.5 to F/11
Manual lens: from F/2.8 to F/11
CCD errors in operative normal Correct
conditions

Product data Arcovis 3000 S/R 9” – Rev. 9 Page 18 of 30


Image processor

Memories
Features L.I.H. 1+330 1+2700
Images number L.I.H. Only Ram image L.I.H. +330 L.I.H. + 2700
(nonvolatile images) (nonvolatile images)
Image format 768x576x12 bits 50 Hz 768x576x12 bit 50 Hz 768x576x12 bit 50 Hz
256 gray levels 256 grey levels 256 grey levels
A/D Converter 8 bits 10 bit 10 bit
D/A Converter 8 bits 8 bit 8 bit
Sampling frequency 15 MHZ 15 MHZ 15 MHZ
Monitors number 1 2 2
Recursive Filter, YES k= 0, 2, 4, 8, 16 YES k= 0, 2, 4, 8, 16 YES k= 0, 2, 4, 8, 16
noises reduction
Digital rotation in YES YES YES
real time
Vertical image YES, in combination YES, in combination YES, in combination
Inversion with the rotation with the rotation with the rotation
Horizontal image YES YES YES
inversion
Grey scale NO YES YES
Inversion
Frame rate fps Single store image Single store image Single store image
acquisition
Image edge (EDGE) YES YES YES
Patient data editing NO YES YES
Video input Composite Video signal Composite Video signal Composite Video
standard CCIR 1 Vpp standard CCIR 1 Vpp signal standard CCIR 1
end 75 Ohm end 75 Ohm Vpp end 75 Ohm
Video output BNC, Standard CCIR 1 BNC, Standard CCIR 1 BNC, Composite Video
Vpp composite video Vpp composite video signal standard CCIR 1
signal 75 Ohm signal 75 Ohm Vpp end 75 Ohm
termination termination
Power supply 20 V to 36V dc 10 Vdc, 36V dc 10 Vdc, 36V dc
Dimensions BOX (26x10,2x6,5 cm) BOX (26x10,2x6,5) BOX (26x10,2x6,5)
[10.24x4.02x2.56 in] [10.24x4.02x2.56 in] [10.24x4.02x2.56 in]

Product data Arcovis 3000 S/R 9” – Rev. 9 Page 19 of 30


RTP 800 memory
Features Data
Image acquirement Digital CCD camera 960x960 up to 12 bit (GigE).
Continuous and pulsed Fluoroscopy. Digital
Snapshot.
Image storage 1024 x 1024 x 12 bit
Image storage on H.D. About 55.000
Video output 1 x HDMI 1960x1200
Pulsed fluoroscopy Yes
Electronic rotation at 1° steps Yes
Horizontal and vertical inversion Yes
Grey scale inversion Yes
Brightness and contrast Yes
Recursive filter Yes (1, 2, 4, 8, 16)
Smart Filter with “Motion Detection” Yes
Edge enhancement Smooth, Normal, Sharp in Yes
post processing
Virtual collimator Yes
Electronic collimator (rectangular and circular) Yes
Electronic zoom factor from 1, 2 to 3 Yes
Electronic lens factor from 1, 2 to 3 Yes
Overview Yes 4, 9, 16 images
Text editing Yes
Dose report Yes
Patients archive Yes
Length measures Yes
Length calibration on reference object Yes
Angles measures Yes
Stenosis measure Yes
Interface for network Ethernet TCP/IP Yes
Adapter of Wireless network 802.11 a/g/n Optional
Data transmission towards PACS Yes
DICOM options DICOM VERIFY (SCU/SCP), DICOM STORAGE
(SCU), DICOM WORK LIST (SCU), DICOM PRINT
(SCU), DICOM CDR/DVD, DICOM
QUERY/RETRIEVE (SCU), DICOM MPPS (CPU),
DICOM STORAGE COMMITMENT (SCU)

Product data Arcovis 3000 S/R 9” – Rev. 9 Page 20 of 30


RTP 1000 memory
Features Data
Image acquirement CCD Digital camera 1024x1024 up to 12 bit
(GigE). Pulsed and contiunuos fluoroscopy.
“Cine” sequence up to 25 fps. Digital Snapshot
Image storage 1024 x 1024 x 12 bit
Image storage on H.D. About 110.000
Video output 2 x DVI 1280x1024
Programmed acquirement sequences Yes, 1, 3, 6, 12, 25 fps
Pulsed fluoroscopy Yes
Electronic rotation at 1° steps Yes
Horizontal and vertical inversion Yes
Grey scale inversion Yes
Brightness and contrast Yes
Recursive filter Yes (1, 2, 4, 8, 16)
Smart Filter with “Motion Detection” Yes
Edge enhancement Smooth, Normal, Sharp in Yes
post processing
Cineloop review Yes
Virtual shutter Yes
Electronic collimator (rectangular and circular) Yes
Max. opacity fluoroscopy acquirement Yes
Subtraction in real time (DSA) with Yes
manual/automatic mask
Shifting pixels Yes
Land marking Yes
Electronic zoom factor from 1, 2 to 3 Yes
Electronic lens factor from 1, 2 to 3 Yes
Overview Yes 4, 9, 16 images
Text editing Yes
Dose report Yes
Patients archive Yes
Length measures Yes
Length calibration on reference object Yes
Angles measures Yes
Stenosis measure Yes
Interface for network Ethernet TCP/IP Yes
Adapter of Wireless network 802.11 a/g/n Optional
Infrared remote control Yes
Data transmission towards PACS Yes
DICOM options DICOM VERIFY (SCU/SCP), DICOM STORAGE
(SCU), DICOM WORK LIST (SCU), DICOM PRINT
(SCU), DICOM CDR/DVD, DICOM
Product data Arcovis 3000 S/R 9” – Rev. 9 Page 21 of 30
QUERY/RETRIEVE (SCU), DICOM MPPS (CPU),
DICOM STORAGE COMMITMENT (SCU)

HRP 3000 memory


Features Data
Image acquirement Digital CCD camera 1024x1024 up to 12 bit
(GigE). Continuous and pulsed fluoroscopy.
Pulsed at high dose up to 25 fps (HCF). “Cine”
sequence at high dose up to 25 fps. Digital
Snapshot.
Image storage 1024 x 1024 x 12 bit
Image storage on H.D. About 110.000
Video output 2 x DVI 1280x1024
Programmed acquirement sequences Yes, 1, 3, 6, 12, 25 fps
Pulsed fluoroscopy Yes
Electronic rotation at 1° steps Yes
Horizontal and vertical inversion Yes
Grey scale inversion Yes
Brightness and contrast Yes
Recursive filter Yes (1, 2, 4, 8, 16)
Smart Filter with “Motion Detection” Yes
Edge enhancement Smooth, Normal, Sharp in Yes
post processing
Cineloop review Yes
Virtual shutter Yes
Electronic collimator (rectangular and circular) Yes
Max. opacity fluoroscopy acquirement Yes
Subtraction in real time (DSA) with Yes
manual/automatic mask
Shifting pixels Yes
Land marking Yes
Electronic zoom factor from 1, 2 to 3 Yes
Electronic lens factor from 1, 2 to 3 Yes
Overview Yes 4, 9, 16 images
Text editing Yes
Dose report Yes
Patients archive Yes
Length measures Yes
Length calibration on reference object Yes
Angles measures Yes
Stenosis measure Yes
Interface for network Ethernet TCP/IP Yes
Adapter of Wireless network 802.11 a/g/n Optional
Product data Arcovis 3000 S/R 9” – Rev. 9 Page 22 of 30
Infrared remote control Yes
Data transmission towards PACS Yes
DICOM options DICOM VERIFY (SCU/SCP), DICOM STORAGE
(SCU), DICOM WORK LIST (SCU), DICOM
PRINT (SCU), DICOM CDR/DVD, DICOM
QUERY/RETRIEVE (SCU), DICOM MPPS
(CPU), DICOM STORAGE COMMITMENT (SCU)

DICOM Terminology
Terminology Description
DICOM VERIFY (SCU/SCP) It allows the connection check, in both ways, with DICOM
units present on the network.
DICOM STORAGE (SCU) It allows to send the images to a DICOM server for filing.
DICOM WORKLIST (SCU) It queries and receives from a DICOM server the patients list
to be examined on the acquirement system.
DICOM PRINT (SCU) It sends to DICOM printer the images to be printed in
panoramic way and through film composer.
DICOM CDR/DVD (media It burn on CD or DVD the patients images with the
Interchange) possibility to add a display program.
DICOM MPPS-Modality It informs the server that the examination is in operation and
Performed Procedure Step (SCU) then that it has been completed by sending the reference
indication to the images and, if available, the total
acquirement dose.
DICOM STORAGE It asks and wait for the confirmation to the storage server
COMMITMENT (SCU) that the images sent are saved in a safe way by allowing the
cancellation of the acquirement unit from the file.
DICOM QUERY/RETRIEVE It queries and receives from a server the images of a patient
(SCU) data for consultation.
SCU (SERVICE CLASS UNIT) Unit that asks a DICOM service to a unit SCP (SERVICE
CLASS PROVIDER) that is able to supply such service.

Product data Arcovis 3000 S/R 9” – Rev. 9 Page 23 of 30


Monitors

Basic color monitor 19"


Standard monitor with memories L.I.H., 1+330, 1+2700
Description Data
Model EM19TFTI/MCII-V
Type 19” LCD TFT anti-glare panel
Display angle Hor: 170° Vert: 170°
Contrast Ratio 800:1
Resolution 1280 x 1024
Pixel dimensions 0.294 mm x 0.294 mm
Display color Base color
Brightness Max. luminance 270 cd/m2 @ 6500 °K
Aspect Ratio 4:3
Response time -
Video Standard VGA / DVI-D
Connectors D-Sub 15p. HD ; DVI-D, BNC for composite video (B/N)
Brightness / Contrast OSD menu, frontal pushbutton
Power supply 110-240 V~ 50-60 Hz
Absorption 75 VAmax
Weight 9.8 Kg
Mounting interface VESA 100x100

Monochrome monitor 19"


Optional monitor with memories L.I.H., 1+330, 1+2700.
Description Data
Model EM19TFTI/MII-V
Type 19” LCD TFT B/N anti-glare panel
Display angle Hor: 170° Vert: 170°
Contrast Ratio 900:1
Resolution 1280 x 1024
Pixel dimensions 0.294 mm x 0.294 mm
Display color Monochrome
Brightness Max. luminance 800 nit (1000 nit typ.)
Aspect Ratio 4:3
Response time -
Video Standard VGA / DVI
Connectors D-Sub 15p. HD ; DVI-D, BNC video composite (B/N)
Brightness / Contrast OSD menu, frontal pushbutton
Power supply 110-240 V~ 50-60 Hz
Absorption max 85 W
Weight 8,6 kg
Mounting interface VESA 100x100
Product data Arcovis 3000 S/R 9” – Rev. 9 Page 24 of 30
Monochrome monitor 19"
Standard monitor with memories RTP 1000 and HRP 3000
Description Data
Model MM190D
Type 19” TFT active matrix LCD, anti-glare and hard coated
Display angle Hor: 170° Vert: 170°
Contrast Ratio 1000 (typical)
Resolution 1280 x 1024
Pixel dimensions 0.294 mm x 0.2 mm
Display color Monochrome
Brightness Max. luminance 1300 cd/m2 (typical)
Aspect Ratio 4:3
Response time 20 ms (Tr+Tf)
Video Standard VGA / DVI
Connectors D-Sub 15p. HD ; DVI-D, BNC
Brightness / Contrast OSD menu
Power supply 100~250 Vac, 1.2 A, 60/50 Hz, universal
Absorption < 80 W,
Weight 5,4 kg
Mounting interface VESA 100x100

Basic color monitor 24”


Standard monitor only for RTP 800 memory
Description Data
Brand EIZO
Model EV2456
Type 24” LCD TFT
Display angle Hor: 170° Vert: 170°
Contrast Ratio 1000:1
Resolution 1920 x 1200 pixel
Pixel dimensions 0.27 x 0.27 mm
Display color 16 milion colors
Brightness Max. luminance 350 cd/m2
Aspect Ratio 16:10
Response time 5 ms (grey to grey)
Input signals Display Port, HDMI, DVI-D, DSub
Brightness / Contrast OSD menu, frontal pushbutton
Power supply 110-240 V~ 50-60 Hz
Absorption 44 W max
Weight 5,7 Kg
Mounting interface VESA 100x100

Product data Arcovis 3000 S/R 9” – Rev. 9 Page 25 of 30


Color monitor 24” touch screen high brightness
Optional monitor only for RTP 800 memory
Description Data
Brand Olorin
Model MLG24004HPCAP
Type 24” LCD PCAP touch screen
Display angle Hor: 178° Vert: 178°
Contrast Ratio 1000:1
Resolution 1920 x 1200 pixel
Backlight LED
Display color 16 milion colors
Brightness Max. luminance 900 cd/m2
Aspect Ratio 16:10
Response time 14 ms (grey to grey)
Input signals VGA, DVI, HDMI, Display Port, USB Hub, USB
Brightness / Contrast Via OSD
Power supply External adapter 24 V (110-240 V~ 50-60 Hz for the adapter)
Absorption <60 W
Weight 6,8 Kg
Mounting interface VESA 100x100

Laser targeting device for dose reduction (only on I.I. tube) (option)
Description Data
Wave Length 645 ±10 nm
Optical power of the laser diode < 1 mW
Safety class 60825-1:2007 Class 2

Dose meter with ionizing chamber (DAP Meter) (option)


Description Data
Model, Type DIAMENTOR CM-T
Measurement category Dose area product
Measuring unit μGy m ²
Digital resolution 0,01 μGy m²
Measurement category Dose area product rate
Measuring unit μGy m²/s
Digital resolution 0,01 μGy m²/s
Dose rate range 0,10 a 400 μGy/s
Dose area product range rate 0,01 a 4500 μGy m²/s
Tube voltage 40 to 120 kV
Chamber: Type TA34037
Dimension of chamber field Ø 7,2 cm
Max energy dependence ± 8 % according to IEC60580, Table 6
Min. inherent filtration < 0,3 mmAl / 70 kV
Product data Arcovis 3000 S/R 9” – Rev. 9 Page 26 of 30
Thermal dose meter printer (option)
Description Data
Type Movable printer
Model Custom Print's
Printing method Thermal printing line
Resolution 203 dpi
Printing speed 50 mm/s*
Paper width (mm) 58 mm
Roll dimensions (mm) 57.5 ±1
Print area 48 mm
Interface RS-232
Power Supply 9/50 Vdc / 0,6 A
Operating temperature 0±50 °C
Humidity storage 10 ± 85 %, there must be no condensation
Dimensions (WxDxH) 146 x 88 x 65 mm
Weight 340 gr
Safety EN60950
* it depends on the printing typology and the environment temperature

Thermal printer (option)


Description Data
Model Sony UP971AD / UP991AD
Print method Direct thermal printer
Resolution 325 dpi
Grey levels 8 bits (256 levels) processing
Picture elements Digital: 7680 x 2560 dots, Video NTSC: 720 x 504 dots,
Video PAL: 720 x 604 dots
Paper size Paper width of 210 mm
Picture area Digital: 600 x 200 mm (max)
Video: STD NTSC: 182 x 144 mm PAL: 188 x 140 mm SIDE
NTSC: 244 x 184 mm PAL: 244 x 183 mm
Picture memory Digital: 2816 x 7680 x 8 (bit)
Video: 6 frame (720 x 608 x 8 bit for one frame)
Printing time Approx. 8 seconds/image (in standard mode)
Interface Hi-Speed USB (USB 2.0) (x1)
VIDEO INPUT: BNC type (x1)
NTSC or PAL composite video signals 1.0Vp-p, 75ohm
(NTSC/PAL automatically discriminated)
VIDEO OUTPUT: BNC type (x1) Loop-through
REMOTE: Stereo mini jack (x1)
Power requirements AC 100 V to 240 V, 50/60 Hz
Dimensions (W x H x D) 316 x 132,5 x 265 mm
Weight 7 kg
Product data Arcovis 3000 S/R 9” – Rev. 9 Page 27 of 30
DVD Recorder (option)
Description Data
Model Sony, DVD Recorder HVO-550MD
Recording video format MPEG-4 AVC/H.264
Recording audio format AC-3/AAC LC
Recording file format AVCHD/MP4
Recording media Internal HDD (500GB), DVD-R, External USB Storage, Network
(CIFS)
Recording resolution 1280 × 720/59.94p, 1280 × 720/50p, 720 × 480/59.94i, 720 ×
576/50i
Recording bit rate (HD) 14 Mbps (Best), 8 Mbps (High), 4 Mbps (Standard)
Recording bit rate (SD) 5 Mbps (Best), 3 Mbps (High), 2 Mbps (Standard)
Input connectors HDMI (Type A) (1), DVI-D (DVI 19-pin) (1), S VIDEO (Mini DIN 4-pin
type) (1), VIDEO (BNC type) (1) AUDIO (Stereo mini jack) (1) DC IN
(DIN 3-pin)
Output connectors HDMI (Type A) (1), DVI-D (DVI 19-pin) (1), S VIDEO (Mini DIN 4-pin
type) (1), VIDEO (BNC type) (1) AUDIO (Stereo mini jack) (1)
Other interfaces USB (Type A) (3), USB (Type B) (1), Network (RJ- 45, 1000 Base-
T/100 Base-TX) (1)), REMOTE RS-232C (D-sub 9-pin) (1), REMOTE
contact switch (stereo mini jack) (2) REMOTE MONITOR (RJ-45) (1),
Equipotential
Power requirements DC from +12 V to +24 V (supplied from adapter AC AC-81MD)
Input current 3.5 A to 1.8 A
Weight 3.2 kg (7 lb. 0.88 oz.)
Dimensions 212.0 × 287.7 × 105.5 mm (8 3/8 × 11 3/8 × 4 1/4 in.)

Product data Arcovis 3000 S/R 9” – Rev. 9 Page 28 of 30


Medical Image Capture Device (option)
Available only with SBFM memory version
Description Data
Model MediCap USB200. Digital medical Image capture device. Saves digital
images and video sources to a USB flash drive or USB hard drive.
Equipped with USB Data: 2GB, 350 Kb per image, about 5700 image
Case Rugged metal case with high-impact plastic faceplate
Buttons Sealed membrane, fluid resistant
Video PAL / NTSC (switch selectable), S-video or composite BNC connections
for both input and output
Foot Switch Hands-free capture. Standard 3.5 mini jack
Mechanical Size: 240 x 200 x 63 mm / 9.5 x 9.7 x 2.5” Weight: 1.7 Kg / 3 lb
Environmental Storage: -40° to +85° C - Operating: -20°C to +40° C
Power 100/240 Vac 50-60 Hz 20 W 3-prong AC jack
Media USB flash drives (MediCapture brand recommended) or external USB
hard drive
Image Formats DICOM, JPEG, TIFF, PNG
Video Formats MPEG2 PS, DVD-quality.
Quality Levels:
Low (DVD-) 352 x 240 pixels, 1.5 Mbps Variable Bit Rate (VBR)
Normal (DVD) NTSC: 720 x 480 pixels, 3.5 Mbps VBR, PAL: 720 x 576 pixels, 3.5 Mbps
VBR
High (DVD+) NTSC: 720 x 480 pixels, 8 Mbps VBR, PAL: 720 x 576 pixels, 8 Mbps
VBR
Audio format MPEG1 L2, 2 channels, 16 bits/channel, 48 kHz sampling 256 kbps bitrate
Image Size 1024 x 768 pixels, 800 x 600 pixels, 640 x 480 pixels
Certifications IEC, UL, CSA, FCC

Product data Arcovis 3000 S/R 9” – Rev. 9 Page 29 of 30


Configuration and accessories
Description Mounting 9"
X-ray handswitch with extensible cable till 4 mt Standard
Fluoroscopy control by triple footswitch Standard
Cassette-holder for radiography on cassette 24x30 cm Optional
Cassette-holder for radiography on cassette 18x24 cm Optional
Sterilizable drapes set Optional
Thermal dry film printer Sony UP 971 / UP 991 or equivalent Optional
Medical Image Capture Device MediCap USB200 Optional
USB Data: 4 GB, 350 Kb per image, about 11400 images Optional
DVD Recorder Sony HVO-550MD Optional
Laser Targeting device (only on I.I. tube) Optional
Dose meter with ionization chamber (DAP) Diamentor CM-T Optional
Printer for dose meter (on equipment) Optional
Collimator with iris and parallel shutter Optional
Movement handles on “C” arm Optional
Movement handle on I.I. tube Optional

Standards and regulations


CE symbol grants the product compliance to the European Directive for Medical
0051 Devices 93/42/EEC and its revised versions as a class IIB device

Note: Products are continuously under review in the light of technical advancement. The actual specification
may therefore be subject to improvement or modification without notice.

Company with Quality System certified


by
VILLA SISTEMI MEDICALI s.p.a.
20090 BUCCINASCO (MI) – ITALY,
Via delle Azalee, 3
Tel. +39-02-488591, Fax +39-02-4881844
ISO 9001:2015 ISO 13485:2016

Product data Arcovis 3000 S/R 9” – Rev. 9 Page 30 of 30